Software Errata Sheet for the Kappa 400, Model 9952
Software for the Medtronic 2090 CareLink Programmer:
The following table lists certain conditions that may occur during use of the Model 9952
Software. The table also provides the Resolution for these conditions.
Condition Observed Behavior Suggested Corrective Action
When Rate Drop Response is
programmed to On (only valid for
DDD and DDI mode), perform
Emergency programming.
The following are trademarks of Medtronic: Medtronic, Medtronic CareLink, Kappa, Vitatron
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
After Emergency programming
has gone into effect, interlock
symbols will appear on the
screen with a non-pertinent
parameter.
Select interrogate button to
reinterrogate the device and
refresh the display.