Medtronic KSR403 Reference Guide

KAPPA®400 SERIES AND DX2
PACEMAKERS
Model 9952
Volume II, Pacemaker Reference Guide
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
Volume II - Pacemaker Reference Guide
A Guide To Understanding the Medtronic.Kappa
DR
K
401, KDR403,
SR
K
401, KSR403,
Pacemakers
®
This guide also applies to Medtronic DX2 7970 and 7972 pacemakers as described in Section I at the back of this guide.
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
4
Checklist, Fast Path, Kappa, Marker Channel, Medtronic, Medtronic.Kappa, Medtronic.Vision, Rate Profile Optimization, Remote Assistant, Significant Events, Sinus Preference, Quick Look, and Vision are all trademarks of Medtronic, Inc.
5

Table of Contents

How To Use This Guide ix
10. Pacing Modes
Introduction 10-2 Mode Selection Decision Tree: 10-4 Mode Pertinency Tables 10-5 Indications/Intended Use 10-7 Contraindications 10-8 DDDR Mode 10-9 DDD Mode 10-10 DDIR Mode 10-11 DDI Mode 10-12 DVIR Mode 10-13 DVI Mode 10-14 VDD Mode 10-15 AAIR / ADIR Modes 10-16 AAI / ADI Modes 10-17 VVIR / VDIR Modes 10-18 VVI / VDI Modes 10-19 AAT / VVT Modes 10-20 DOOR / AOOR / VOOR Modes 10-21 DOO / AOO / VOO Modes 10-22 ODO / OAO / OVO Modes 10-23
Table of Contents
11. Rate Response
Introduction to Rate Responsive Pacing 11-2 Preset Rate Response at Implant 11-4 Automatic Sensor Initialization 11-6 Rate Profile Optimization Individualizing Rate Profile Optimization 11-13 Integrated Sensor Operation 11-15 Minute Ventilation Sensor Operation 11-18 Activity Sensor Operation 11-20 Manual Control of Optimization and MV Sensor 11-26
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
6
Operation 11-9
iii
Table of Contents
12. Pacemaker Timing
Rates 12-2 AV Intervals 12-12 Blanking Periods 12-19 Refractory Periods 12-21 High Rate Atrial Tracking 12-31
13. Lead / Cardiac Tissue Interface
Implant Detection 13-2 Bipolar Lead Monitor 13-5 Selecting Pacing Parameters 13-9 Selecting Sensing Parameters 13-12 Transtelephonic Capture Verification with TMT 13-15
14. Special Therapy Options
Mode Switch and Diagnostic 14-2 Non-Competitive Atrial Pacing 14-6 PMT Intervention 14-9 PVC Response 14-12 Ventricular Safety Pacing 14-15 Sinus Preference
TM
and Diagnostic 14-17 Sleep Function 14-21 Single Chamber Hysteresis 14-23 Rate Drop Response and Diagnostic 14-25
15. Telemetry Data
Parameter Summary 15-2 Patient Data 15-4 Battery and Lead Information 15-6 Marker Channel
TM
Telemetry 15-8 Extended Markers 15-9 Intracardiac and MV Electrograms 15-10 Extended Telemetry 15-14
iv
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
7
16. Miscellaneous Operations
Magnet Mode Operation 16-2 Temporary Programming 16-5 Electrical Reset 16-7 Elective Replacement Indicator (ERI) 16-9 Emergency Pacing 16-10
17. Diagnostics
Introduction to Diagnostics 17-2 Atrial and Ventricular Rate Histograms 17-8 AV Conduction Histogram 17-10 Sensor Indicated Rate Profile 17-13 Atrial and Ventricular High Rate Diagnostics 17-14 Remote Assistant
TM
Diagnostic 17-19
Custom Rate Trend 17-21
18. Troubleshooting the Pacing System
Troubleshooting Strategy 18-2 Troubleshooting Hemodynamic Problems 18-6 Handling, Storage, and Resterilization 18-8 Pacemaker Longevity 18-9 Replacing the Pacemaker 18-11 Patient Information and Service 18-12
Table of Contents
A. Pacemaker Description
Basic Description A-2 Lead Compatibility A-3 Radiopaque Identification A-4 Mechanical Dimensions A-6 Pacemaker Connectors A-7
B. Preset Parameter Settings
Preset Parameter Values B-2
C. Longevity Projections
Longevity Projections C-2 Elective Replacement Indicator C-5 Battery Specifications C-6
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
v
8
Table of Contents
D. Telemetry and Diagnostic Values
Magnet Mode Operations D-2 Telemetry Functions D-3 Automatic Diagnostics D-6 Clinician-Selectable Diagnostics D-7 Cardiac Event Counters D-9
E. Parameter Values and Restrictions
Programmable Modes and Parameters E-2 Nonprogrammable Parameters E-8 Temporary Modes and Parameters E-9 Rate Response Programming Guidelines E-10 Programming Restrictions and Requirements E-12 Timing Reference E-15
F. Warnings and Precautions
Special Notice F-2 Warnings F-3 Precautions F-5 Potential Complications F-9
G. Environmental Interference
Hospital or Medical Environment Interference G-2 Home and Job Environment Interference G-7
H. Pacing Indications
Indications in Acquired Atrioventricular Block in Adults H-2 Indications After Myocardial Infarction H-4 Indications in Bifascicular and Trifascicular Block H-5 Indications in Sinus Node Dysfunction H-6 Indications in Hypersensitive Carotid Sinus Syndrome and Neurovas-
cular Syndromes H-7 Indications in Children H-8
vi
9
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
I. Supplement: DX2 Models 7970 and 7972
How To Use This Supplement I-2 Pacemaker Differences I-3 Low MV Monitor I-5 Radiopaque Identification I-6 Pacemaker Connectors I-7 Longevity Projections I-8 Battery Specifications I-10 Lead Compatibility I-10 Mechanical Dimensions I-10
J. Glossary
K. Index
Table of Contents
10
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
vii

How To Use This Guide

Guide Is Organized Into Two Volume Set
Product information about Medtronic.Kappa® 400 Series pacemakers and Vision software is organized into a two-volume set:
Volume I, which contains programming information is comprised of Chapters 1 through 9.
Volume II, which contains pacemaker information includes Chapters 10 through 18 and Sections A through J.
Note that implant information is also provided in the product information manual that is packaged with each Kappa pacemaker.
What Is Contained In This Guide - Volume II
How To Use This Guide
12
This guide describes how the pacemaker operates and specifies the capabilities of each model.
Chapters 10 through 17 describe the pacing modes, rate response options, special therapy features, telemetry types, and data collection options. In some cases, guidelines are given of how to configure these pacemaker operations.
Chapter 18 contains troubleshooting information for electrical and hemodynamic problems.
Sections A through E specify parameter and data collection capabilities, longevity projections, and mechanical and electrical specifications.
Sections F through H provide general warnings and cautions, potential interference with the pacemaker, and general indications for pacing. A supplement covering the clinical DX2 pacemakers, a glossary, and an index follow section H.
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
ix
How To Use This Guide
What Is Contained In The Programming Guide - Volume I
Volume I instructs the clinician on the Model 9790/C Programmer and managing Kappa pacemaker patients. Chapters 1 through 9 cover the following topics:
how to setup, configure, and operate the programmer and accessing on-line help
using programmer features to conduct patient follow-up sessions
viewing patient ECGs and electrocardiograms
retrieving pacemaker collected data and configuring new data collection options
evaluating pacemaker parameter settings and programming new settings, if necessary
initializing the minute ventilation sensor and verifying rate response operation
13
x
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide

Understanding Pacemaker Operation

Chapters 10 - 18 provide detailed information about the operation of the Kappa 400 series pacemakers.
Pacing Modes
Rate Response
Pacemaker Timing
16
Lead/Cardiac Tissue Interface
Special Therapy Options
Telemetry Data
Miscellaneous Operations
Diagnostics
Troubleshooting The Pacing System
Pacing Modes
General Desciption
This chapter provides information about the modes available with the pacemaker. The sections are as follows:
10
Introduction
Mode Selection Decision Tree:
Mode Pertinency Tables
Indications/Intended Use
Contraindications
DDDR Mode
DDD Mode
DDIR Mode
DDI Mode
DVIR Mode
DVI Mode
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
10-2
10-4
10-5
10-7
10-8
10-9
10-10
10-11
10-12
10-13
10-14
VDD Mode
AAIR / ADIR Modes
AAI / ADI Modes
VVIR / VDIR Modes
VVI / VDI Modes
AAT / VVT Modes
DOOR / AOOR / VOOR Modes
DOO / AOO / VOO Modes
ODO / OAO / OVO Modes
10-15
10-16
10-17
10-18
10-19
10-20
10-21
10-22
10-23
10-1
18
Pacing Modes
Introduction
Introduction
Pacing Mode Selection
This chapter provides an introduction to pacemaker modes as an aid to pacing mode selection. The chapter is organized as follows:
Mode Selection Decision Tree – This decision tree, based on the
1991 ACC/AHA guidelines for pacemaker implantation, provides a simple means of identifying pacing modes appropriate for given indications. More specific indications for bradycardia-related pacing are described in Section H.
Mode Pertinency Tables – These tables show which features and
parameters apply to each commonly used pacing mode.
Mode Descriptions – These descriptions provide indications
and contraindications for modes available with the pacemaker and brief descriptions of how these modes operate.
1
19
10-2
1. Dreifus LS, Fisch C, Griffin JC, et al. Guidelines for implantation of cardiac pacemakers and antiarrhythmia devices. A report of the American College of Cardiology/American Heart Association Task Force on Assessment of Diagnostic and Therapeutic Cardiovascular Procedures (Committee on Pacemaker Implantation). 18: 1-13.
Journal of the American College of Cardiology.
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
1991;
Pacing Modes
Introduction
NBG Pacing Codes
The pacemaker modes are defined in NBG Code.1 Each five-letter NBG code describes a specific type of operation for implantable pacemakers. For simplicity, this manual uses only the first three or four letters, such as DDD, DDIR, DVIR, and so forth. Figure 10-1 describes the first four letters of the NBG code.
CHAMBER PACED
V = Ventricle A = Atrium D = Dual Chamber S = Single Chamber O = None
CHAMBER SENSED
V = Ventricle A = Atrium D = Dual Chamber S = Single Chamber O = None
Figure 10-1.
DDDR
NBG Pacing Codes
MODE OF RESPONSE
T = Triggered I = Inhibited D = Double (Both) O = None
PROGRAMMABLE/ RATE RESPONSE
P = Programmable M = Multiprogrammable C = Communicating R = Rate Responsive O = None
Further Information
The mode descriptions in this chapter provide only a basic overview of each mode. For further details on the rate response, timing, and therapy capabilities refer to Chapter 11, Chapter 12, and Chapter 14, respectively.
20
1. Bernstein A., et al., “The NASPE/BPEG Pacemaker Code,” PACE, 10(4), Jul-Aug
1987. (“NBG” stands for “The North American Society of Pacing and Electrophysiology [NASPE] and the British Pacing and Electrophysiology Group [BPEG] Generic.” NBG’s five-letter code supersedes the ICHD Code.
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
10-3
Pacing Modes
Mode Selection Decision Tree:
Mode Selection Decision Tree:
Figure 10-2 shows a basic decision tree used to select the pacing mode. In the shaded boxes the preferred mode(s) are listed and the alternate mode(s) appears below the dashed line.
Symptomatic
Bradycardia
(e.g., persistent
atrial fibrillation,
inexcitable atrium)
is SA node conduction
Ye s NoNo Yes
AAI
DDD AAIR
DDDR
No
VVIR
VVI
presently adequate?
Can the atrium be sensed
and/or paced reliably?
presently adequate?
AAIR
DDDR
Figure 10-2.
Ye s
is AV conduction
Mode Selection Tree
is SA node conduction
presently adequate?
DDD
DDDR
VDD
NoYe s
(e.g., complete or transient AV block)
DDDR
DDIR
21
10-4
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
Pacing Modes
Mode Pertinency Tables
*
–––
✓✓
–––
✓✓
––
✓ ✓ ✓ ✓✓
✓✓✓✓ ✓
22
Mode Pertinency Tables
Pacing Parameters Available For Each Mode
Table 10-1.
Table 10-1 and Table 10-2 indicate which pacing parameters and features apply to each commonly used pacing mode
by black check marks. Dashes indicate parameters that are programmable but not pertinent to basic mode operation.
These parameters are programmable when mode switch, RAAV, or sensor-varied PVARP are active or when
background rate response is operative.
Basic Operation Parameters
✓ ✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓
Pacing Parameter DDDR DDD DDIR DDI DVIR DVI VDD VVIR VDIR VVI VDI VVT AAIR ADIR AAI ADI AAT
Lower Rate
✓✓
✓✓
Upper Tracking Rate
✓ ✓✓✓✓✓
Sensed AV Interval
Paced AV Interval
Upper Sensor Rate
✓✓✓ ✓ ✓
✓ ✓✓✓
PVARP
Rate Adaptive AV
✓ ✓✓✓
PVAB
Atrial Blanking
Atrial Refractory Period
✓ ✓✓✓✓✓
✓ ✓✓✓✓✓✓✓✓✓✓✓
Ventricular Blanking (After AP)
Ventricular Refractory Period
* See Chapter 12 for operational descriptions of these timing parameters.
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
Standard pacing and sensing parameters (amplitude, pulse width, sensitivity, and pace/sense polarities) are
Note:
† Sensor-varied PVARP available in the DDDR, DDD, DDIR, and VDD modes.
not shown here but are found in Section E. Also, asynchronous modes are not shown.
10-5
Pacing Modes
Mode Pertinency Tables
*
Features Available For Each Mode
Special Therapy Options
✓✓✓ ✓✓ ✓
✓✓
Rate Response Therapy Options
–––
–––
–––
–––
✓✓
✓✓
✓✓
✓✓
–––
–––
–––
–––
✓✓
✓✓
✓✓
✓✓
––
––
––
––
–––
–––
✓✓
✓✓
–––
–––
✓✓
✓✓
––
––
23
Table 10-2.
10-6
✓✓
✓✓
Fe atu re DDDR DDD DDIR DDI DVIR DVI VDD VVIR VDIR VVI VDI VVT AAIR ADIR AAI ADI AAT
Managing Atrial Rhythms
Mode Switch
Non-Competitive Atrial Pacing
✓✓
Rhythms
PMT Intervention
Managing Ventricular
✓ ✓✓✓✓✓
✓ ✓✓✓
Ventricular Safety Pacing
Special Pacing Operations
Rate Drop Response
PVC Response
Single Chamber Hysteresis
✓ ✓✓✓✓✓✓✓✓✓✓✓✓✓✓
Sinus Preference ✓Sleep Function
Integrated Sensor
Rate Profile Optimization
ADL Rate
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
Activity Threshold
Activity Sensor
Minute Ventilation Sensor
* See Chapter 14 for operational descriptions of special therapy options.
† See Chapter 11 for operational descriptions of rate response features.
Indications/Intended Use
Kappa pacemakers are indicated for the following:
Rate adaptive pacing in patients who may benefit from
increased pacing rates concurrent with increases in activity and/or minute ventilation.
Accepted patient conditions warranting chronic cardiac
pacing which include:
Symptomatic paroxysmal or permanent second or third-
degree AV block.
Symptomatic bilateral bundle branch block.
Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction disorders.
Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic tachyarrhythmias.
Pacing Modes
Indications/Intended Use
24
Vasovagal syndromes or hypersensitive carotid sinus
syndromes.
Kappa pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include:
Various degrees of AV block to maintain the atrial
contribution to cardiac output.
VVI intolerance (e.g., pacemaker syndrome) in the presence of
persistent sinus rhythm.
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
10-7
Pacing Modes
Contraindications
Contraindications
Kappa pacemakers are contraindicated for the following applications:
Dual chamber atrial pacing in patients with chronic refractory
atrial tachyarrhythmias.
Asynchronous pacing in the presence (or likelihood) of
competitive paced and intrinsic rhythms.
Co-implant in a patient with an implanted cardioverter-
defibrillator (ICD) because it may cause unwanted delivery of ICD therapy.
25
10-8
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
DDDR Mode
Pacing Modes
DDDR Mode
In the DDDR mode, the pacemaker tracks the faster of the intrinsic atrial rate or the sensor-indicated rate. If the intrinsic rate is faster, the DDDR mode provides atrial synchronous pacing; otherwise, AV sequential pacing occurs at the sensor-indicated rate.
Rate limits for atrial tracking (Upper Tracking Rate)1 and
sensor tracking (Upper Sensor Rate) are separately programmable.
The AV intervals that follow sensed atrial events (SAV) and
paced atrial events (PAV) are separately programmable, and they can be programmed to shorten with increasing rates (Rate Adaptive AV).
A nonrefractory sensed event in either chamber inhibits
pacing in that chamber. A ventricular nonrefractory sensed event in the VA interval that is not preceded by an atrial sense (AS or AR) may be a pacemaker-defined PVC, and starts a new VA interval.
26
Sensor-indicated
Interval
A P
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) PAV Interval = 200 ms PVARP = 280 ms Sensor-indicated Rate = 90 ppm (667 ms) SAV Interval = 170 ms
A P
V S
Figure 10-3.
Example of DDDR Mode Operation
Sensor-indicated
Interval
A S
V S
A P
V P
1. The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser value.
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
A S
V P
200 ms
10-9
Pacing Modes
DDD Mode
DDD Mode
The DDD mode provides atrial synchronous pacing in the presence of intrinsic atrial activity; otherwise, AV sequential pacing occurs at the Lower Rate.
Each atrial paced or nonrefractory atrial sensed event starts an
AV interval and a lower rate interval. The AV intervals that follow sensed atrial events (SAV) and paced atrial events (PAV) are separately programmable, and the SAV may be optionally programmed to shorten with increasing rate (Rate Adaptive AV).
Ventricular paced events may track atrial sensed events up to
the programmed Upper Tracking Rate.
A ventricular nonrefractory sensed event in the VA interval
1
that is not preceded by an atrial sense (AS or AR) may be a pacemaker-defined PVC, and starts a new VA interval.
10-10
Lower Rate Interval
A P
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) PAV Interval = 200 ms
Figure 10-4.
A P
V S
SAV Interval = 170 ms
Example of DDD Mode Operation
1. The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser value.
Lower Rate Interval
A S
V S
200 ms
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
A P
27
DDIR Mode
Pacing Modes
DDIR Mode
The DDIR mode provides dual chamber, sensor-driven, atrioventricular (AV) sequential pacing for heart rate variation without atrial tracking.
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless inhibited.
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval. That is, ventricular paced events after such sensed atrial events occur at the sensor-indicated rate.
A ventricular nonrefractory sensed event in the
ventriculoatrial (VA) interval starts a new VA interval.
Sensor-indicated
Interval
A P
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) PAV Interval = 200 ms Sensor-indicated Rate = 90 ppm (667 ms)
A P
Figure 10-5.
Sensor-indicated
Interval
A S
V P
Example of DDIR Mode Operation
Sensor-indicated
VA Interval
V P
Sensor-indicated
A
A P
P
V P
Interval
200 ms
A P
28
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
10-11
Pacing Modes
DDI Mode
DDI Mode
Lower Rate Interval
The DDI mode provides dual chamber atrioventricular (AV) sequential pacing with atrial sensing but without atrial tracking.
Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited.
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval. Ventricular paced events after such sensed atrial events occur at the Lower Rate.
A ventricular nonrefractory sensed event in the
ventriculoatrial (VA) interval starts a new VA interval.
Lower Rate
Lower Rate Interval
VA Interval
29
A P
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) PAV Interval = 200 ms
A P
Figure 10-6.
10-12
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
A S
V P
V P
Example of DDI Mode Operation
200 ms
A P
DVIR Mode
Pacing Modes
DVIR Mode
The DVIR mode provides AV sequential pacing at the sensor­indicated rate unless inhibited by ventricular sensed events.
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless inhibited.
The DVIR mode ignores intrinsic atrial events. Sensing occurs
only in the ventricle. A ventricular nonrefractory sensed event during the ventriculoatrial (VA) interval starts a new VA interval.
Sensor-indicated
Interval
A P
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) PAV Interval = 200 ms Sensor-indicated Rate = 90 ppm (667 ms)
A P
V S
Figure 10-7.
Example of DVIR Mode Operation
Sensor-indicated
V S
VA Interval
Sensor-indicated
A P
V P
Interval
A P
200 ms
30
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
10-13
Pacing Modes
DVI Mode
DVI Mode
Lower Rate Interval
The DVI mode provides dual chamber AV sequential pacing without atrial sensing/tracking.
Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited.
Sensing occurs only in the ventricle, and intrinsic atrial events
are ignored. A ventricular nonrefractory sensed event during the VA interval starts a new ventriculoatrial (VA) interval.
Lower Rate
VA Interval
10-14
A P
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) PAV Interval = 200 ms
Figure 10-8.
A P
V S
Example of DVI Mode Operation
V S
A P
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
V P
200 ms
31
VDD Mode
Pacing Modes
VDD Mode
The VDD mode provides atrial synchronous pacing (or VVI pacing at the Lower Rate). The ventricle is paced synchronously up to the programmed Upper Tracking Rate.
1
Sensing occurs in both the atrium and ventricle, but pacing occurs only in the ventricle.
To promote atrial synchronous pacing at slow rates, a sensed
atrial event occurring near the end of the lower rate interval will be followed by the programmed maximum SAV interval. The result is an extension of the ventricular lower rate.
A ventricular nonrefractory sensed event in the V-V interval
that is not preceded by an atrial sense (AS or AR) may be a pacemaker-defined PVC, and it starts a new V-V interval.
Lower Rate Interval
SAV
Interval
A S
V P
Parameters:
Lower Rate = 60 ppm (1000 ms) SAV Interval = 200 ms Upper Tracking Rate = 120 ppm (500 ms) PVARP = 250 ms
A S
V P
Figure 10-9.
1. The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
Example of VDD Operation
A S
V P
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
A S
200 ms
10-15
32
Pacing Modes
AAIR / ADIR Modes
AAIR / ADIR Modes
The AAIR mode provides atrial-based rate responsive pacing in patients with intact AV conduction. Sensing and pacing occur only in the atrium. In the absence of sensed events, the chamber is paced at the sensor-indicated rate.
The ADIR mode operates the same as the AAIR mode except that events sensed in the ventricle are recorded by the diagnostics. When used in conjunction with Marker Channel telemetry and concurrent ECG, this mode may be used to observe the conducted ventricular rhythm without affecting atrial pacing.
In the AAIR and ADIR modes, atrial refractory sensed
Note:
events do not restart the upper sensor rate interval.
Sensor-indicated Interval Sensor-indicated Interval
33
10-16
A P
Operation
Sensor-indicated Rate = 75 ppm (800 ms) Upper Sensor Rate = 100 ppm (600 ms)
A R
:
A P
Figure 10-10.
A S
Example of AAIR Mode Operation
A P
200 ms
Medtronic.Kappa® 400 Series Pacemaker Information and Programming Guide
Loading...
+ 294 hidden pages