Caution: Federal Law (USA) restricts this device
to sale by or on the order of a physician (or
properly licensed practitioner).
Medtronic Kappa® 700
Series Pacemaker Product
Information Manual
Models KDR701/703/706, KDR721, and
DR731/733
K
This Product Information Manual is primarily intended as an
implantation manual. For programming information, see the
Pacemaker Reference Guide that accompanies the programmer
software. It is primarily intended as a follow-up manual, and contains
further information on therapeutic and diagnostic features,
troubleshooting information, follow-up precautionary information,
and complete reference information.
Device Description 8
Indications and Usage 9
Contraindications 10
Warnings and Precautions 11
Co-implantation with an Implantable Defibrillator 21
Adverse Events 23
Clinical Studies 27
Chapter 2 - Implanting the Pacemaker 31
Implantation Procedures 32
Implant Documentation 42
Parameter Programming at Implant 43
Medical Therapy Interactions 49
Assistance 52
Chapter 3 - Description 53
Pacing Mode Operations 54
Rate Responsive Pacing 57
Timing Operations 58
Pacing and Sensing Operations 64
Special Therapy Options 71
Chapter 4 - Pacemaker Follow-up 79
Pacemaker Telemetry 80
Other Operations 83
5
Diagnostics 87
General Recommendations 89
Appendix 91
NBG Codes 92
Special Notice 93
Pacemaker Specifications 95
Lead Requirements, Compatibility 96
Radiopaque Identification 97
Emergency Parameter Settings 98
Shipping and Nominal Parameter Settings 99
Electrical Reset Parameter Settings 103
Elective Replacement Indicator 108
Magnet Mode Conditions 108
Longevity Projections 109
Programmable Parameters 116
Nonprogrammable Parameters 125
Temporary Parameters 126
Tel e m e t r y M a r k e r s12 7
Electrograms 127
Automatic Diagnostics 127
Clinician-Selected Diagnostics 128
Patient Information 128
Battery and Lead Telemetered Information 129
Battery Parameters 130
Mechanical Dimensions 131
Index 133
6
Chapter 1 - Prescribing the Pacemaker
Device Description8
Indications and Usage 9
Contraindications 10
Warnings and Precautions 11
Co-implantation with an Implantable Defibrillator21
Adverse Events23
Clinical Studies27
7
Device Description
Device Description
Medtronic Kappa 700 Series pacemakers (KDR700 Series) are dual
chamber multiprogrammable, rate responsive, implantable
pacemakers, intended for a variety of bradycardia pacing
applications. Rate response is controlled through an activity-based
sensor. The following models are available:
ModelPolarityPrimary Leads
K
DR701, KDR721,
DR731
and K
KDR703 and
DR733
K
K
DR706UnipolarUnipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992).
DR700 Series pacemakers are programmed using the Medtronic
K
Bipolar/UnipolarIS-1
Bipolar/UnipolarLow-profile 3.2 mm bipolar
Vision software Model 9953 and a Medtronic Model 9790
programmer. For programming instructions, refer to the Pacemaker Programming Guide, which accompanies Medtronic Kappa 700
Series software.
a
BI
or IS-1a BI
8
Indications and Usage
Indications and Usage
KDR700 Series pacemakers are indicated for the following:
■
Rate adaptive pacing in patients who may benefit from
increased pacing rates concurrent with increases in activity.
■
Accepted patient conditions warranting chronic cardiac pacing
which include:
– Symptomatic paroxysmal or permanent second or
third-degree AV block.
– Symptomatic bilateral bundle branch block.
– Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders.
– Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
– Vasovagal syndromes or hypersensitive carotid sinus
syndromes.
DR700 Series pacemakers are also indicated for dual chamber and
K
atrial tracking modes in patients who may benefit from maintenance
of AV synchrony. Dual chamber modes are specifically indicated for
treatment of conduction disorders that require restoration of both rate
and AV synchrony, which include:
■
Various degrees of AV block to maintain the atrial contribution
to cardiac output.
■
VVI intolerance (e.g., pacemaker syndrome) in the presence of
persistent sinus rhythm.
9
Contraindications
Contraindications
KDR700 Series pacemakers are contraindicated for the following
applications:
■
Dual chamber atrial pacing in patients with chronic refractory
atrial tachyarrhythmias.
■
Asynchronous pacing in the presence (or likelihood) of
competitive paced and intrinsic rhythms.
■
Unipolar pacing for patients with an implanted
cardioverter-defibrillator (ICD) because it may cause
unwanted delivery or inhibition of ICD therapy. See “Coimplantation with an Implantable Defibrillator” on page 21.
10
Warnings and Precautions
Warnings and Precautions
■
Rate responsive modes. Do not use rate responsive modes
in those patients who cannot tolerate pacing rates above the
programmed Lower Rate.
■
Single chamber atrial modes. Do not use single chamber
atrial modes in patients with impaired AV nodal conduction
because ventricular capture cannot be assured.
Pacemaker Dependent Patients
■
Diagnostic modes. Never program diagnostic modes (ODO,
OVO, and OAO) for pacemaker-dependent patients. For such
patients, use the programmer’s inhibit function for brief
interruption of outputs.
■
Electrogram (EGM) of the patient’s intrinsic activity should be
obtained with care since the patient is without pacing support
when using the programmer’s inhibit function.
■
Polarity override. Overriding the bipolar verification prompt
with bipolar polarity when a unipolar lead is connected results
in no pacing output. See “Manually Programming Polarity” on
page 67 for further information.
■
A false bipolar pathway on a unipolar lead, a possible
occurrence with bipolar 3.2 mm connector pacemakers, may
result in a loss of output. See the warning in “Automatic
Polarity Configuration” on page 64 for further information.
11
Warnings and Precautions
■
Loss of capture during threshold margin test (TMT) at a
20% reduction in amplitude indicates that the stimulation
safety margin is inadequate. Consider increasing the pacing
amplitude and/or pulse width. See “Magnet Operation” on
page 83 for further information on the Threshold Margin Test.
■
Ventricular safety pacing should always be used for
pacemaker-dependent patients. See “Ventricular Safety
Pacing” on page 74 for further information.
■
Capture Management will not program ventricular outputs
above 5.0 V or 1.0 ms. If the patient needs a pacing output
higher than 5.0 V or 1.0 ms, program Amplitude and Pulse
Width manually. See “Capture Management” on page 70 for
further information.
Medical Therapy
■
Therapeutic Diathermy can cause fibrillation, burning of
the myocardium, and irreversible damage to the pulse
generator due to induced currents.
■
Magnetic resonance imaging of pacemaker patients has
resulted in significant adverse effects. See “Magnetic
Resonance Imaging (MRI)” on page 17 for further information.
Storage and Resterilization
Medtronic pacemakers are intended for single use only. Do not
resterilize and re-implant explanted pacemakers.
The chart below gives recommendations on handling and storing the
package. Medtronic has sterilized the pacemaker with ethylene oxide
prior to shipment. Resterilizing the pacemaker is necessary if the
seal on the sterile package is broken. Resterilization does not affect
the “Use Before” date.
12
Warnings and Precautions
Handling and Storage: AcceptableUnacceptable
Store and transport within
Environmental Temperature limits:
°F (- 18°C) to + 131°F (55°C).
0
Note: A full or partial electrical reset
condition may occur at temperatures
below 0°F (- 18°C). See “Electrical
Reset Parameter Settings” on
page 103.
Resterilization: AcceptableUnacceptable
Do not implant the device if it
has been dropped on a hard
surface from a height of
12 inches (30 cm) or more.
Resterilize if the sterile package
seal is broken. Place the device in
an ethylene oxide permeable
package and resterilize with
ethylene oxide. Allow the device to
aerate ethylene oxide residues.
See sterilizer instructions for
details. Use an acceptable method
for determining sterility, such as
biological indicators.
Do not resterilize the device or the
torque wrench using:
Do not exceed 140
17 psi (103 kPa) when sterilizing.
Do not resterilize the device more
than twice.
°F (60°C) or
13
Warnings and Precautions
Lead Evaluation and Lead Connection
■
Connector compatibility. Do not use any lead with this
pacemaker without first verifying connector compatibility.
Using incompatible leads can damage the connector or result
in a leaking or intermittent connection.
■
Pacing and sensing safety margins. Consider lead
maturation when choosing pacing amplitudes, pacing pulse
widths, and sensing levels. See “Manual Programming” on
page 45.
■
Hex wrench. Do not use a hex wrench with a blue handle or a
right- angled hex wrench. These wrenches have torque
capabilities greater than is designed for the lead connector.
See “Connection Procedure” on page 36 for lead connection
instructions.
Programming and Pacemaker Operation
■
Epicardial leads. Ventricular epicardial leads have not been
determined appropriate for use with the Capture Management
feature. Therefore, Capture Management should be
programmed Off if epicardial leads are implanted with K
Series pacemakers.
■
Shipping values. Do not use shipping values for pacing
amplitude and sensitivity without verifying that they provide
adequate safety margins for the patient.
■
Constant current devices. To test the performance of the
lead, Medtronic recommends using a constant voltage device
such as the Medtronic Model 5311B (or equivalent) Pacing
System Analyser (PSA). Medtronic does not recommend using
a constant current device such as the Medtronic Model 5880A
or 5375 External Pacemaker because the Kappa DR
pacemaker has constant voltage output circuits.
■
Crosstalk occurs in dual chamber systems when atrial pacing
output pulses are sensed by the ventricular lead. Crosstalk
DR700
14
Warnings and Precautions
results in self-inhibition and is more likely to occur at high
sensor-driven pacing rates, high atrial amplitudes, and wide
atrial pulse widths. To prevent self-inhibition caused by
crosstalk, program Ventricular Safety Pacing (VSP) On or
lengthen the Ventricular Blanking period.
■
Elective Replacement Indicator (ERI). When ERI is set, the
pacemaker must be replaced within three months. See
“Elective Replacement Indicator” on page 86 for more
information.
■
Full electrical reset is indicated by VVI pacing at a rate of
65 ppm without the elective replacement indicator set. See
“Electrical Reset” on page 85 for more information.
■
Slow retrograde conduction, especially with conduction time
greater than 400 ms, may induce pacemaker-mediated
tachycardia (PMT).
■
PMT intervention. Even with the feature turned On, PMTs
may still require clinical intervention such as pacemaker
reprogramming, magnet application, drug therapy, or lead
evaluation. See “PMT Intervention” on page 73 for further
information.
■
Lead Monitor. If the Lead Monitor detects out-of-range lead
impedance, investigate lead integrity more thoroughly.
Rate Increases
■
Twiddler’s syndrome, i.e., patient manipulation of the device
after implant, may cause the pacing rate to increase
temporarily if the pacemaker is programmed to a rate
responsive mode.
■
Muscle stimulation, e.g., due to unipolar pacing, may result
in pacing at rates up to the Upper Sensor Rate in rate
responsive modes.
15
Warnings and Precautions
Unipolar Sensing
■
Continuous myopotentials cause reversion to asynchronous
operation when sensed in the refractory period. Sensing of
myopotentials is more likely when atrial sensitivity settings of
0.5 through 1.0 mV and ventricular sensitivity settings of
1.0 and 1.4 mV are programmed.
Environmental and Medical Therapy Hazards
Patients should be directed to exercise reasonable caution in
avoidance of devices which generate a strong electric or magnetic
field. If the pacemaker inhibits or reverts to asynchronous operation
at the programmed pacing rate or at the magnet rate while in the
presence of electromagnetic interference (EMI), moving away from
the source or turning it off will allow the pacemaker to return to its
normal mode of operation.
Hospital and Medical Environments
■
Electrosurgical cautery could induce ventricular arrhythmias
and/or fibrillation, or may cause asynchronous or inhibited
pacemaker operation. If use of electrocautery is necessary, the
current path and ground plate should be kept as far away from
the pacemaker and leads as possible. See “Electrosurgical
Cautery” on page 50 for more information.
■
External defibrillation may damage the pacemaker or may
result in temporary and/or permanent myocardial damage at
the electrode-tissue interface as well as temporary or
permanent elevated pacing thresholds. Attempt to minimize
current flowing through the pacemaker and lead system by
following these precautions when using external defibrillation
on a pacemaker patient:
– Position defibrillation paddles as far from the pacemaker as
possible (minimum of 5 inches [13 cm]). Attempt to
minimize current flowing through the pacemaker and leads
16
Warnings and Precautions
by positioning the defibrillation paddles perpendicular to
the implanted pacemaker/lead system.
– Use the lowest clinically appropriate energy output (watt
seconds).
– Confirm pacemaker function following any defibrillation.
■
High energy radiation sources such as cobalt 60 or gamma
radiation should not be directed at the pacemaker. If a patient
requires radiation therapy in the vicinity of the pacemaker,
place lead shielding over the device to prevent radiation
damage and confirm its function after treatment.
■
Lithotripsy may permanently damage the pacemaker if the
device is at the focal point of the lithotripsy beam. If lithotripsy
must be used, program the pacemaker to a single chamber
nonrate responsive mode (VVI/AAI or VOO/AOO) prior to
treatment; and keep the pacemaker at least 1 to 2 inches
(2.5 to 5 cm) away from the focal point of the lithotripsy beam.
■
Magnetic resonance imaging (MRI). Pacemaker patients
subjected to MRI should be closely monitored and
programmed parameters should be verified upon cessation of
MRI. MRI of pacemaker patients should be carefully weighed
against the potential adverse affects. Clinicians should
carefully weigh the decision to use MRI with pacemaker
patients. Limited studies of the effects of MRI on pacemakers
have shown that:
– Magnetic and radio frequency (RF) fields produced by MRI
may adversely affect the operation of the pacemaker and
may inhibit pacing output.
– Magnetic fields may activate magnet mode operation and
cause asynchronous pacing.
17
Warnings and Precautions
– Reported1 effects of MRI on pacing include increased
ventricular pacing beyond the rate limit.
■
Radiofrequency ablation procedure in a patient with a
DR700 Series pacemaker may cause any of the following:
K
– Asynchronous pacing above or below the programmed
rate.
– Reversion to an asynchronous operation.
– Pacemaker electrical reset.
– Premature triggering of the elective replacement indicator.
RF ablation risks may be minimized by:
1. Programming a non-rate responsive, asynchronous pacing
mode prior to the RF ablation procedure.
2. Avoiding direct contact between the ablation catheter and the
implanted lead or pacemaker.
3. Positioning the ground plate so that the current pathway does
not pass through or near the pacemaker system, i.e., place the
ground plate under the patient's buttocks or legs.
4. Having a Medtronic programmer available for temporary
pacing.
5. Having defibrillation equipment available.
Home and Occupational Environments
■
High voltage power transmission lines may generate
enough EMI to interfere with pacemaker operation if
approached too closely.
■
Communication equipment such as microwave transmitters,
linear power amplifiers, or high-power amateur transmitters
may generate enough EMI to interfere with pacemaker
operation if approached too closely.
1
Holmes, Hayes, Gray, et al. The effects of magnetic resonance imaging on implantable
pulse generators. PA C E. 1986; 9 (3): 360-370.
18
Warnings and Precautions
■
Commercial electrical equipment such as arc welders,
induction furnaces, or resistance welders may generate
enough EMI to interfere with pacemaker operation if
approached too closely.
■
Home appliances which are in good working order and
properly grounded do not usually produce enough EMI to
interfere with pacemaker operation. There are reports of
pacemaker disturbances caused by electric hand tools or
electric razors used directly over the pacemaker implant site.
■
Electronic article surveillance (EAS) equipment such as
retail theft prevention systems may interact with pacemakers.
Patients should be advised to walk directly through and not to
remain near an EAS system longer than is necessary.
Cellular Phones
DR700 Series pacemakers have been tested to the frequency
K
ranges used by the cellular phones included in Table 1. Based on this
testing, these pacemakers should not be affected by the normal
operation of such cellular phones.
These pacemakers contain a filter that allows usage, without
interaction, of all cellular phones having one of the transmission
technologies listed in Table 1. These transmission technologies
represent most of the cellular phones in use worldwide. Patients can
contact their local cellular phone service provider to confirm that the
provider uses one of these technologies.
19
Warnings and Precautions
Table 1. Cellular Phone Transmission Technologies
Transmission TechnologyFrequency Range
Analog
FM (Frequency Modulation)824 - 849 MHz
Digital TDMA
North American Standards
TDMA - 11 Hz806 - 821 MHz
NADC
PCS
International Standards
GSM
DCS
a
b
(TDMA - 50 Hz)
c
1900
d
e
1800
824 - 849 MHz
1850 - 1910 MHz
880 - 915 MHz
1710 - 1785 MHz
Digital CDMA
CDMA - DS
a
Time Division Multiple Access
b
North American Digital Cellular
c
Personal Communication System
d
Global System for Mobile Communications
e
Digital Cellular System
f
Code Division Multiple Access - Direct Sequence
f
20
824 - 849 MHz
Co-implantation with an Implantable Defibrillator
Co-implantation with an Implantable
Defibrillator
■
An implantable defibrillator may be implanted concomitantly
with a bipolar pacemaker.
– The use of unipolar-only Model KDR706 and the KDR700
Series bipolar models implanted with unipolar leads is
contraindicated for patients having an implantable
defibrillator.
– Follow implant protocol and precautions for pacemaker
and defibrillator lead placement. Ensure that the
pacemaker is configured to be compatible with the
defibrillator.
Programming Considerations
Note the following programming considerations for patients who have
an implantable defibrillator.
■
Only bipolar pacing should be used with these patients. In
some cases, pacing in the unipolar configuration may cause
the defibrillator either to deliver inappropriate therapy or to
withhold appropriate therapy.
■
Polarity is automatically configured during Implant Detection
(see “Automatic Polarity Configuration” on page 64 of the
Product Information Manual). If lead integrity is suspect,
confirmation of the automatically programmed polarities
should be made after completion of Implant Detection in order
to assure that bipolar polarities have been programmed
appropriately.
■
The implantable cardiac defibrillator (ICD) should be turned off
during pacemaker implantation procedures until lead polarities
have been configured and confirmed. This is to prevent
possible back-up unipolar paces from triggering the ICD.
21
Co-implantation with an Implantable Defibrillator
■
Lead Monitor should not be programmed to Adaptive. When a
prevalence of out-of-range lead impedance paces is detected,
the monitor automatically reprograms the selected lead(s) to
unipolar polarity. Pacing in the unipolar configuration may
cause the defibrillator either to provoke inappropriate therapy
or to withhold appropriate therapy.
■
Transtelephonic Monitor should be programmed to Off. If it is
programmed On, the pacing polarity is temporarily set to
unipolar when the magnet is applied. Pacing in the unipolar
configuration may cause the defibrillator either to provoke
inappropriate therapy or to withhold appropriate therapy.
■
Although these pacemakers are designed to be compatible
with implantable defibrillators, the potential does exist for a
defibrillation pulse to reset them.
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular
bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection.
If lead integrity is suspect, confirmation of bipolar polarity
should be made after completion of Implant Detection.
22
Adverse Events
Adverse Events
The Medtronic Kappa 700 Series devices were evaluated in a
multicenter prospective study (43 investigational centers, 15 centers
in the US) of the adaptive features and rate response of the device.
Clinical study of the Medtronic Kappa 700 Series of pacemakers
included 288 devices implanted in 285 patients worldwide.
There were a total of six deaths in the study; all were reviewed and
judged to be non-device related by a clinical events committee
comprising clinical investigators and Medtronic clinical evaluation
managers. Two were attributed to ventricular arrhythmia, one to
respiratory failure, the fourth to respiratory insufficiency due to
chronic obstructive pulmonary disease, the fifth to a mesotelioma,
and the sixth to multi-system organ failure.
Eight devices were explanted: three due to pocket infection, one
due to infection of the electrode, one from lead/connector mismatch,
one patient had a psychosomatic disorder, one patient required
the implant of a dual chamber ICD, and one patient continued
with vasovagal symptoms and the therapy did not provide
sufficient benefit.
Observed Adverse Events
A total of 355 adverse events were reported. The device-related
events (182 events) are listed in descending order of frequency in
Ta bl e 2 .
23
Adverse Events
Table 2. Adverse Events Reported in Four or More Patients-
Complications
a
(Comps) and Observationsb (Obs)
All patients implanted (n=288 devices in 285 patients, 133 device years)
c
% of
Patients
with
Comps
(n=285)
Comps
per
Device-
Year
(n=133)
% of
Patients
with Obs
(n=285)
Obs per
Device-
Year
(n=133)
EventTotal
Number
of
Events
(Patients)
Any adverse
355 (168)17.2%0.4552%2.22
events
Any device-related
182 (118)10.9%0.3134%1.06
events
Pain at pocket site32 (31)——10.9%0.24
Other23 (21)1.1%0.026.3%0.15
Inappropriate
11 (11)——3.9%0.08
programming
Atrial lead
11 (10)3.6%0.08——
dislodgment
Programmer/
software anomaly
11 (8)——2.8%0.08
d
Pocket infection7 (6)0.7%0.021.4%0.03
Intermittent
6 (6)0.7%0.021.4%0.03
undersensing
Palpitations6 (6)——2.1%0.05
Pocket hematoma6 (6)0.4%0.011.8%0.04
Ventricular lead
6 (6)2.1%0.05——
dislodgment
Elevated pacing
4 (4)0.7%0.020.7%0.02
thresholds
Syncope4 (4)——1.4%0.03
24
Adverse Events
a
Complications included those adverse events which required invasive measures to
correct (e.g., surgical intervention), and were related to the presence of the pacing
system or procedure.
b
Observations included those adverse events which did not require invasive measures
to resolve, and were related to the presence of the pacing system or procedure.
c
Where present, a number in parentheses indicates the number of patients with
the event.
d
Programmer software anomalies observed: screen lock-ups while saving data to
diskette (8); problems printing reports outside of a patient session (2); and an incorrect
parameter setting on a printout (1), which occurred in an earlier version of the
software—software changes were made to eliminate reoccurrence.
The following other adverse events were reported, but occurred in
three or fewer patients: angina pectoris; atrial flutter (paroxysmal)/
atrial fibrillation; bipolar short circuit pathway; chest pain; chest pain
(non-specific); dizziness; dyspnea/shortness of breath; exit block;
failure to capture/loss of capture; false negative capture detection; far
field R-wave sensing; fatigue/tiredness; hypotension; inadequate
lead/pacemaker connection; infection of electrode; lead/connector
mismatch; lead insertion route problem; lead insulation failure;
migration of pulse generator; myopotential interference; other
oversensing; pacemaker mediated tachycardia; pacemaker
syndrome; pectoral muscle stimulation; penetration of myocardium
by lead; phrenic nerve/diaphragm muscle stimulation;
pneumothorax; swelling pocket site; tachycardia (atrial); thrombus
formation at lead; ventricular ectopy; ventricular tachycardia
(non-sustained); ventricular tachycardia (sustained).
25
Adverse Events
The following adverse events were deemed not device related
(173 events were reported): angina pectoris; atrial fibrillation; atrial
flutter (paroxysmal); atrial flutter (persistent); atrial tachycardia; chest
pain; chest pain (non-specific); congestive heart failure; dizziness;
dyspnea/shortness of breath; fatigue/tiredness; hypertension;
hypotension; insufficient cardiac output; myocardial infarction
(acute); palpitations; syncope; ventricular ectopy; ventricular
fibrillation; ventricular tachycardia (non-sustained); ventricular
tachycardia (sustained).
Potential Adverse Events
Adverse events (in alphabetical order), including those reported in
Table 2, associated with pacing systems include:
The Medtronic Kappa 700 Series devices were evaluated in a
multicenter prospective study (43 investigational centers, 15 centers
in the US) of the adaptive features and rate response of the device.
Methods
This study compared the following features of the Medtronic Kappa
700 Series pacemakers to historical controls:
■
Rate Response
■
Capture Management
■
Automatic Polarity Configuration
■
Sensing Assurance
Patient data were collected at implant, pre-discharge, two weeks,
one month, two and/or three months, and six months post implant.
Patients were evaluated utilizing a modified version of the Minnesota
Pacemaker Response Exercise Protocol (MPREP
month visit. Evaluation of rate response performance for the
Medtronic Kappa 700 Series pacemaker was conducted using the
Metabolic Chronotropic model described by Wilkoff as applied by
2
. Automatic polarity configuration data were collected at implant.
Kay
Sensing Assurance and Capture Management data were collected at
each follow-up.
Description of Patients
Patients enrolled in the study represented a general dual chamber
pacing population.
1
) at their one
1
Benditt, David G. M, Editor, Rate Adaptive Pacing, Blackwell Scientific Publications,
Boston. 1993: 63-65.
2
Kay, Neal G., “Quantitation of Chronotropic Response: Comparison of Methods for
Rate-Modulating Permanent Pacemakers”, JACC 20(7):1533-41, Dec 1992.
27
Clinical Studies
Results of the Study
Table 3 summarizes the results of the clinical study. The incidence of
complications was found to be similar to that experienced by similar
devices. The performance of the automatic polarity configuration,
Capture Management, Sensing Assurance, and rate response
features were found to meet study objectives.
The slope of the exercise rate response (1.0 target slope) was less
than 0.65 for 26 of 87 (30%) of patients.
28
Clinical Studies
All patients implanted (n=288 devices in 285 patients, 133 device years)
Primary Objectives
Automatic Polarity Configuration (n with loss of output / N leads)
Total Leads0% (0/546)[0%, 0.55%]≤5%
Unipolar0% (0/107)[0%, 2.8%]≤5%
Bipolar0% (0/439)[0%, 0.7%]≤5%
Sensing Assurance
(n with loss of sensing or oversensing / N device years)
Atrial13.5% (18/133)[8.8%, 20.5%]≤35.7%
Ventricular0.8% (1/133)[0.2%, 4.1%]≤9.2%
Capture Management (n with all causes loss of capture / N device years)
Loss of capture5.3% (7/133)[2.6%, 10.5%]≤10.7%
Slope of MPREP rate response at 1 month (n=87 patients)
Mean0.81[0.76, 0.86][0.65, 1.35]
Table 3. Effectiveness Analysis
Percent of
events
% (n/N)
95%
Confidence
interval
Criteria:
Upper
95% CI
The Medtronic Kappa 700 Series pacemaker’s Rate Profile
Optimization (RPO) governs sensor indicated rate (SIR) output.
Figure 1 shows the SIR vs. the Wilkoff predicted heart rate achieved
using the RPO feature during the MPREP tests at 1 month.
29
Clinical Studies
All patients reaching Anaerobic Threshold, N=87
Expected (Wilkoff) rate, mean and 95% CI
100
90
80
70
60
50
40
30
SIR (normalized)
20
10
0
0 204060
p
p
U
I
C
5%
9
r
e
n
a
e
M
Lower 95% CI
S
e
t
c
e
p
x
E
)
7
8
=
N
(
R
I
MPREP Workload (normalized)
Figure 1. Sensor Indicated Rate (SIR) vs. Expected Rate at One Month
)
f
f
o
k
l
i
W
(
d
80
R
I
S
100
30
Loading...
+ 110 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.