Medtronic KDR706 Technical Manual

Page 1
KAPPA® DR700 SERIES
197881003A__view.pdf
PacemakerModels KDR700/720/730
Product Information Manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
Page 2
Page 3
Medtronic Kappa® 700 Series Pacemaker Product Information Manual
Models KDR701/703/706, KDR721, and
DR731/733
K
Page 4
This Product Information Manual is primarily intended as an implantation manual. For programming information, see the Pacemaker Reference Guide that accompanies the programmer software. It is primarily intended as a follow-up manual, and contains further information on therapeutic and diagnostic features, troubleshooting information, follow-up precautionary information, and complete reference information.
The following are trademarks of Medtronic:
Auto PVARP, Capture Management, FAST, Implant Detection, Kappa, Key Parameter History, Marker Channel, Medtronic, Medtronic Kappa, Medtronic Vision, Rate Profile Optimization, Remote Assistant, Search AV, Sensing Assurance, Sinus Preference, and Vision.
Page 5

Table of contents

Chapter 1 - Prescribing the Pacemaker 7
Device Description 8 Indications and Usage 9 Contraindications 10 Warnings and Precautions 11 Co-implantation with an Implantable Defibrillator 21 Adverse Events 23 Clinical Studies 27
Chapter 2 - Implanting the Pacemaker 31
Implantation Procedures 32 Implant Documentation 42 Parameter Programming at Implant 43 Medical Therapy Interactions 49 Assistance 52
Chapter 3 - Description 53
Pacing Mode Operations 54 Rate Responsive Pacing 57 Timing Operations 58 Pacing and Sensing Operations 64 Special Therapy Options 71
Chapter 4 - Pacemaker Follow-up 79
Pacemaker Telemetry 80 Other Operations 83
5
Page 6
Diagnostics 87 General Recommendations 89
Appendix 91
NBG Codes 92 Special Notice 93
Pacemaker Specifications 95
Lead Requirements, Compatibility 96 Radiopaque Identification 97 Emergency Parameter Settings 98 Shipping and Nominal Parameter Settings 99 Electrical Reset Parameter Settings 103 Elective Replacement Indicator 108 Magnet Mode Conditions 108 Longevity Projections 109 Programmable Parameters 116 Nonprogrammable Parameters 125 Temporary Parameters 126 Tel e m e t r y M a r k e r s 12 7 Electrograms 127 Automatic Diagnostics 127 Clinician-Selected Diagnostics 128 Patient Information 128 Battery and Lead Telemetered Information 129 Battery Parameters 130 Mechanical Dimensions 131
Index 133
6
Page 7
Chapter 1 - Prescribing the Pacemaker
Device Description 8
Indications and Usage 9
Contraindications 10
Warnings and Precautions 11
Co-implantation with an Implantable Defibrillator 21
Adverse Events 23
Clinical Studies 27
7
Page 8

Device Description

Device Description
Medtronic Kappa 700 Series pacemakers (KDR700 Series) are dual chamber multiprogrammable, rate responsive, implantable pacemakers, intended for a variety of bradycardia pacing applications. Rate response is controlled through an activity-based sensor. The following models are available:
Model Polarity Primary Leads
K
DR701, KDR721,
DR731
and K
KDR703 and
DR733
K
K
DR706 Unipolar Unipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992).
DR700 Series pacemakers are programmed using the Medtronic
K
Bipolar/Unipolar IS-1
Bipolar/Unipolar Low-profile 3.2 mm bipolar
Vision software Model 9953 and a Medtronic Model 9790 programmer. For programming instructions, refer to the Pacemaker Programming Guide, which accompanies Medtronic Kappa 700 Series software.
a
BI
or IS-1a BI
8
Page 9

Indications and Usage

Indications and Usage
KDR700 Series pacemakers are indicated for the following:
Rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity.
Accepted patient conditions warranting chronic cardiac pacing which include:
– Symptomatic paroxysmal or permanent second or
third-degree AV block. – Symptomatic bilateral bundle branch block. – Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders. – Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias. – Vasovagal syndromes or hypersensitive carotid sinus
syndromes.
DR700 Series pacemakers are also indicated for dual chamber and
K atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include:
Various degrees of AV block to maintain the atrial contribution to cardiac output.
VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm.
9
Page 10

Contraindications

Contraindications
KDR700 Series pacemakers are contraindicated for the following applications:
Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias.
Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms.
Unipolar pacing for patients with an implanted cardioverter-defibrillator (ICD) because it may cause unwanted delivery or inhibition of ICD therapy. See “Co­implantation with an Implantable Defibrillator” on page 21.
10
Page 11

Warnings and Precautions

Warnings and Precautions
Rate responsive modes. Do not use rate responsive modes in those patients who cannot tolerate pacing rates above the programmed Lower Rate.
Single chamber atrial modes. Do not use single chamber atrial modes in patients with impaired AV nodal conduction because ventricular capture cannot be assured.

Pacemaker Dependent Patients

Diagnostic modes. Never program diagnostic modes (ODO, OVO, and OAO) for pacemaker-dependent patients. For such patients, use the programmer’s inhibit function for brief interruption of outputs.
Electrogram (EGM) of the patient’s intrinsic activity should be obtained with care since the patient is without pacing support when using the programmer’s inhibit function.
Polarity override. Overriding the bipolar verification prompt with bipolar polarity when a unipolar lead is connected results in no pacing output. See “Manually Programming Polarity” on page 67 for further information.
A false bipolar pathway on a unipolar lead, a possible occurrence with bipolar 3.2 mm connector pacemakers, may result in a loss of output. See the warning in “Automatic Polarity Configuration” on page 64 for further information.
11
Page 12
Warnings and Precautions
Loss of capture during threshold margin test (TMT) at a 20% reduction in amplitude indicates that the stimulation safety margin is inadequate. Consider increasing the pacing amplitude and/or pulse width. See “Magnet Operation” on page 83 for further information on the Threshold Margin Test.
Ventricular safety pacing should always be used for pacemaker-dependent patients. See “Ventricular Safety Pacing” on page 74 for further information.
Capture Management will not program ventricular outputs above 5.0 V or 1.0 ms. If the patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually. See “Capture Management” on page 70 for further information.

Medical Therapy

Therapeutic Diathermy can cause fibrillation, burning of the myocardium, and irreversible damage to the pulse generator due to induced currents.
Magnetic resonance imaging of pacemaker patients has resulted in significant adverse effects. See “Magnetic Resonance Imaging (MRI)” on page 17 for further information.

Storage and Resterilization

Medtronic pacemakers are intended for single use only. Do not resterilize and re-implant explanted pacemakers.
The chart below gives recommendations on handling and storing the package. Medtronic has sterilized the pacemaker with ethylene oxide prior to shipment. Resterilizing the pacemaker is necessary if the seal on the sterile package is broken. Resterilization does not affect the “Use Before” date.
12
Page 13
Warnings and Precautions
Handling and Storage: Acceptable Unacceptable
Store and transport within Environmental Temperature limits:
°F (- 18°C) to + 131°F (55°C).
0 Note: A full or partial electrical reset
condition may occur at temperatures below 0°F (- 18°C). See “Electrical Reset Parameter Settings” on page 103.
Resterilization: Acceptable Unacceptable
Do not implant the device if it
has been dropped on a hard surface from a height of 12 inches (30 cm) or more.
Resterilize if the sterile package seal is broken. Place the device in an ethylene oxide permeable package and resterilize with ethylene oxide. Allow the device to aerate ethylene oxide residues. See sterilizer instructions for details. Use an acceptable method for determining sterility, such as biological indicators.
Do not resterilize the device or the torque wrench using:
an autoclave,
gamma radiation,
organic cleaning agents, e.g., alcohol, acetone, etc.,
ultra-sonic cleaners.
Do not exceed 140 17 psi (103 kPa) when sterilizing.
Do not resterilize the device more than twice.
°F (60°C) or
13
Page 14
Warnings and Precautions

Lead Evaluation and Lead Connection

Connector compatibility. Do not use any lead with this pacemaker without first verifying connector compatibility. Using incompatible leads can damage the connector or result in a leaking or intermittent connection.
Pacing and sensing safety margins. Consider lead maturation when choosing pacing amplitudes, pacing pulse widths, and sensing levels. See “Manual Programming” on page 45.
Hex wrench. Do not use a hex wrench with a blue handle or a right- angled hex wrench. These wrenches have torque capabilities greater than is designed for the lead connector. See “Connection Procedure” on page 36 for lead connection instructions.

Programming and Pacemaker Operation

Epicardial leads. Ventricular epicardial leads have not been determined appropriate for use with the Capture Management feature. Therefore, Capture Management should be programmed Off if epicardial leads are implanted with K Series pacemakers.
Shipping values. Do not use shipping values for pacing amplitude and sensitivity without verifying that they provide adequate safety margins for the patient.
Constant current devices. To test the performance of the lead, Medtronic recommends using a constant voltage device such as the Medtronic Model 5311B (or equivalent) Pacing System Analyser (PSA). Medtronic does not recommend using a constant current device such as the Medtronic Model 5880A or 5375 External Pacemaker because the Kappa DR pacemaker has constant voltage output circuits.
Crosstalk occurs in dual chamber systems when atrial pacing output pulses are sensed by the ventricular lead. Crosstalk
DR700
14
Page 15
Warnings and Precautions
results in self-inhibition and is more likely to occur at high sensor-driven pacing rates, high atrial amplitudes, and wide atrial pulse widths. To prevent self-inhibition caused by crosstalk, program Ventricular Safety Pacing (VSP) On or lengthen the Ventricular Blanking period.
Elective Replacement Indicator (ERI). When ERI is set, the pacemaker must be replaced within three months. See “Elective Replacement Indicator” on page 86 for more information.
Full electrical reset is indicated by VVI pacing at a rate of 65 ppm without the elective replacement indicator set. See “Electrical Reset” on page 85 for more information.
Slow retrograde conduction, especially with conduction time greater than 400 ms, may induce pacemaker-mediated tachycardia (PMT).
PMT intervention. Even with the feature turned On, PMTs may still require clinical intervention such as pacemaker reprogramming, magnet application, drug therapy, or lead evaluation. See “PMT Intervention” on page 73 for further information.
Lead Monitor. If the Lead Monitor detects out-of-range lead impedance, investigate lead integrity more thoroughly.
Rate Increases
Twiddler’s syndrome, i.e., patient manipulation of the device after implant, may cause the pacing rate to increase temporarily if the pacemaker is programmed to a rate responsive mode.
Muscle stimulation, e.g., due to unipolar pacing, may result in pacing at rates up to the Upper Sensor Rate in rate responsive modes.
15
Page 16
Warnings and Precautions
Unipolar Sensing
Continuous myopotentials cause reversion to asynchronous operation when sensed in the refractory period. Sensing of myopotentials is more likely when atrial sensitivity settings of
0.5 through 1.0 mV and ventricular sensitivity settings of
1.0 and 1.4 mV are programmed.

Environmental and Medical Therapy Hazards

Patients should be directed to exercise reasonable caution in avoidance of devices which generate a strong electric or magnetic field. If the pacemaker inhibits or reverts to asynchronous operation at the programmed pacing rate or at the magnet rate while in the presence of electromagnetic interference (EMI), moving away from the source or turning it off will allow the pacemaker to return to its normal mode of operation.
Hospital and Medical Environments
Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or may cause asynchronous or inhibited pacemaker operation. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pacemaker and leads as possible. See “Electrosurgical Cautery” on page 50 for more information.
External defibrillation may damage the pacemaker or may result in temporary and/or permanent myocardial damage at the electrode-tissue interface as well as temporary or permanent elevated pacing thresholds. Attempt to minimize current flowing through the pacemaker and lead system by following these precautions when using external defibrillation on a pacemaker patient:
– Position defibrillation paddles as far from the pacemaker as
possible (minimum of 5 inches [13 cm]). Attempt to minimize current flowing through the pacemaker and leads
16
Page 17
Warnings and Precautions
by positioning the defibrillation paddles perpendicular to the implanted pacemaker/lead system.
– Use the lowest clinically appropriate energy output (watt
seconds).
– Confirm pacemaker function following any defibrillation.
High energy radiation sources such as cobalt 60 or gamma radiation should not be directed at the pacemaker. If a patient requires radiation therapy in the vicinity of the pacemaker, place lead shielding over the device to prevent radiation damage and confirm its function after treatment.
Lithotripsy may permanently damage the pacemaker if the device is at the focal point of the lithotripsy beam. If lithotripsy must be used, program the pacemaker to a single chamber nonrate responsive mode (VVI/AAI or VOO/AOO) prior to treatment; and keep the pacemaker at least 1 to 2 inches (2.5 to 5 cm) away from the focal point of the lithotripsy beam.
Magnetic resonance imaging (MRI). Pacemaker patients subjected to MRI should be closely monitored and programmed parameters should be verified upon cessation of MRI. MRI of pacemaker patients should be carefully weighed against the potential adverse affects. Clinicians should carefully weigh the decision to use MRI with pacemaker patients. Limited studies of the effects of MRI on pacemakers have shown that:
– Magnetic and radio frequency (RF) fields produced by MRI
may adversely affect the operation of the pacemaker and may inhibit pacing output.
– Magnetic fields may activate magnet mode operation and
cause asynchronous pacing.
17
Page 18
Warnings and Precautions
– Reported1 effects of MRI on pacing include increased
ventricular pacing beyond the rate limit.
Radiofrequency ablation procedure in a patient with a
DR700 Series pacemaker may cause any of the following:
K – Asynchronous pacing above or below the programmed
rate. – Reversion to an asynchronous operation. – Pacemaker electrical reset. – Premature triggering of the elective replacement indicator.
RF ablation risks may be minimized by:
1. Programming a non-rate responsive, asynchronous pacing mode prior to the RF ablation procedure.
2. Avoiding direct contact between the ablation catheter and the implanted lead or pacemaker.
3. Positioning the ground plate so that the current pathway does not pass through or near the pacemaker system, i.e., place the ground plate under the patient's buttocks or legs.
4. Having a Medtronic programmer available for temporary pacing.
5. Having defibrillation equipment available.
Home and Occupational Environments
High voltage power transmission lines may generate enough EMI to interfere with pacemaker operation if approached too closely.
Communication equipment such as microwave transmitters, linear power amplifiers, or high-power amateur transmitters may generate enough EMI to interfere with pacemaker operation if approached too closely.
1
Holmes, Hayes, Gray, et al. The effects of magnetic resonance imaging on implantable pulse generators. PA C E. 1986; 9 (3): 360-370.
18
Page 19
Warnings and Precautions
Commercial electrical equipment such as arc welders, induction furnaces, or resistance welders may generate enough EMI to interfere with pacemaker operation if approached too closely.
Home appliances which are in good working order and properly grounded do not usually produce enough EMI to interfere with pacemaker operation. There are reports of pacemaker disturbances caused by electric hand tools or electric razors used directly over the pacemaker implant site.
Electronic article surveillance (EAS) equipment such as retail theft prevention systems may interact with pacemakers. Patients should be advised to walk directly through and not to remain near an EAS system longer than is necessary.
Cellular Phones
DR700 Series pacemakers have been tested to the frequency
K ranges used by the cellular phones included in Table 1. Based on this testing, these pacemakers should not be affected by the normal operation of such cellular phones.
These pacemakers contain a filter that allows usage, without interaction, of all cellular phones having one of the transmission technologies listed in Table 1. These transmission technologies represent most of the cellular phones in use worldwide. Patients can contact their local cellular phone service provider to confirm that the provider uses one of these technologies.
19
Page 20
Warnings and Precautions
Table 1. Cellular Phone Transmission Technologies
Transmission Technology Frequency Range
Analog
FM (Frequency Modulation) 824 - 849 MHz
Digital TDMA
North American Standards
TDMA - 11 Hz 806 - 821 MHz
NADC
PCS
International Standards
GSM
DCS
a
b
(TDMA - 50 Hz)
c
1900
d
e
1800
824 - 849 MHz
1850 - 1910 MHz
880 - 915 MHz
1710 - 1785 MHz
Digital CDMA
CDMA - DS
a
Time Division Multiple Access
b
North American Digital Cellular
c
Personal Communication System
d
Global System for Mobile Communications
e
Digital Cellular System
f
Code Division Multiple Access - Direct Sequence
f
20
824 - 849 MHz
Page 21

Co-implantation with an Implantable Defibrillator

Co-implantation with an Implantable Defibrillator
An implantable defibrillator may be implanted concomitantly with a bipolar pacemaker.
– The use of unipolar-only Model KDR706 and the KDR700
Series bipolar models implanted with unipolar leads is contraindicated for patients having an implantable defibrillator.
– Follow implant protocol and precautions for pacemaker
and defibrillator lead placement. Ensure that the pacemaker is configured to be compatible with the defibrillator.

Programming Considerations

Note the following programming considerations for patients who have an implantable defibrillator.
Only bipolar pacing should be used with these patients. In some cases, pacing in the unipolar configuration may cause the defibrillator either to deliver inappropriate therapy or to withhold appropriate therapy.
Polarity is automatically configured during Implant Detection (see “Automatic Polarity Configuration” on page 64 of the Product Information Manual). If lead integrity is suspect, confirmation of the automatically programmed polarities should be made after completion of Implant Detection in order to assure that bipolar polarities have been programmed appropriately.
The implantable cardiac defibrillator (ICD) should be turned off during pacemaker implantation procedures until lead polarities have been configured and confirmed. This is to prevent possible back-up unipolar paces from triggering the ICD.
21
Page 22
Co-implantation with an Implantable Defibrillator
Lead Monitor should not be programmed to Adaptive. When a prevalence of out-of-range lead impedance paces is detected, the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Transtelephonic Monitor should be programmed to Off. If it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Although these pacemakers are designed to be compatible with implantable defibrillators, the potential does exist for a defibrillation pulse to reset them.
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection. If lead integrity is suspect, confirmation of bipolar polarity should be made after completion of Implant Detection.
22
Page 23

Adverse Events

Adverse Events
The Medtronic Kappa 700 Series devices were evaluated in a multicenter prospective study (43 investigational centers, 15 centers in the US) of the adaptive features and rate response of the device. Clinical study of the Medtronic Kappa 700 Series of pacemakers included 288 devices implanted in 285 patients worldwide.
There were a total of six deaths in the study; all were reviewed and judged to be non-device related by a clinical events committee comprising clinical investigators and Medtronic clinical evaluation managers. Two were attributed to ventricular arrhythmia, one to respiratory failure, the fourth to respiratory insufficiency due to chronic obstructive pulmonary disease, the fifth to a mesotelioma, and the sixth to multi-system organ failure.
Eight devices were explanted: three due to pocket infection, one due to infection of the electrode, one from lead/connector mismatch, one patient had a psychosomatic disorder, one patient required the implant of a dual chamber ICD, and one patient continued with vasovagal symptoms and the therapy did not provide sufficient benefit.

Observed Adverse Events

A total of 355 adverse events were reported. The device-related events (182 events) are listed in descending order of frequency in Ta bl e 2 .
23
Page 24
Adverse Events
Table 2. Adverse Events Reported in Four or More Patients-
Complications
a
(Comps) and Observationsb (Obs)
All patients implanted (n=288 devices in 285 patients, 133 device years)
c
% of
Patients
with Comps (n=285)
Comps
per
Device-
Year
(n=133)
% of
Patients
with Obs
(n=285)
Obs per
Device-
Year
(n=133)
Event Total
Number
of
Events
(Patients)
Any adverse
355 (168) 17.2% 0.45 52% 2.22
events
Any device-related
182 (118) 10.9% 0.31 34% 1.06
events
Pain at pocket site 32 (31) 10.9% 0.24
Other 23 (21) 1.1% 0.02 6.3% 0.15
Inappropriate
11 (11) 3.9% 0.08
programming
Atrial lead
11 (10) 3.6% 0.08
dislodgment
Programmer/ software anomaly
11 (8) 2.8% 0.08
d
Pocket infection 7 (6) 0.7% 0.02 1.4% 0.03
Intermittent
6 (6) 0.7% 0.02 1.4% 0.03
undersensing
Palpitations 6 (6) 2.1% 0.05
Pocket hematoma 6 (6) 0.4% 0.01 1.8% 0.04
Ventricular lead
6 (6) 2.1% 0.05
dislodgment
Elevated pacing
4 (4) 0.7% 0.02 0.7% 0.02
thresholds
Syncope 4 (4) 1.4% 0.03
24
Page 25
Adverse Events
a
Complications included those adverse events which required invasive measures to correct (e.g., surgical intervention), and were related to the presence of the pacing system or procedure.
b
Observations included those adverse events which did not require invasive measures to resolve, and were related to the presence of the pacing system or procedure.
c
Where present, a number in parentheses indicates the number of patients with the event.
d
Programmer software anomalies observed: screen lock-ups while saving data to diskette (8); problems printing reports outside of a patient session (2); and an incorrect parameter setting on a printout (1), which occurred in an earlier version of the software—software changes were made to eliminate reoccurrence.
The following other adverse events were reported, but occurred in three or fewer patients: angina pectoris; atrial flutter (paroxysmal)/ atrial fibrillation; bipolar short circuit pathway; chest pain; chest pain (non-specific); dizziness; dyspnea/shortness of breath; exit block; failure to capture/loss of capture; false negative capture detection; far field R-wave sensing; fatigue/tiredness; hypotension; inadequate lead/pacemaker connection; infection of electrode; lead/connector mismatch; lead insertion route problem; lead insulation failure; migration of pulse generator; myopotential interference; other oversensing; pacemaker mediated tachycardia; pacemaker syndrome; pectoral muscle stimulation; penetration of myocardium by lead; phrenic nerve/diaphragm muscle stimulation; pneumothorax; swelling pocket site; tachycardia (atrial); thrombus formation at lead; ventricular ectopy; ventricular tachycardia (non-sustained); ventricular tachycardia (sustained).
25
Page 26
Adverse Events
The following adverse events were deemed not device related (173 events were reported): angina pectoris; atrial fibrillation; atrial flutter (paroxysmal); atrial flutter (persistent); atrial tachycardia; chest pain; chest pain (non-specific); congestive heart failure; dizziness; dyspnea/shortness of breath; fatigue/tiredness; hypertension; hypotension; insufficient cardiac output; myocardial infarction (acute); palpitations; syncope; ventricular ectopy; ventricular fibrillation; ventricular tachycardia (non-sustained); ventricular tachycardia (sustained).

Potential Adverse Events

Adverse events (in alphabetical order), including those reported in Table 2, associated with pacing systems include:
Cardiac perforation
Cardiac tamponade
Death
Erosion through the skin
Hematoma/seroma
Infection
Myopotential sensing
Nerve and muscle stimulation
Rejection phenomena (local tissue reaction, fibrotic tissue formation, pulse generator migration)
Threshold elevation
Transvenous lead-related thrombosis
26
Page 27

Clinical Studies

Clinical Studies
The Medtronic Kappa 700 Series devices were evaluated in a multicenter prospective study (43 investigational centers, 15 centers in the US) of the adaptive features and rate response of the device.

Methods

This study compared the following features of the Medtronic Kappa 700 Series pacemakers to historical controls:
Rate Response
Capture Management
Automatic Polarity Configuration
Sensing Assurance
Patient data were collected at implant, pre-discharge, two weeks, one month, two and/or three months, and six months post implant. Patients were evaluated utilizing a modified version of the Minnesota Pacemaker Response Exercise Protocol (MPREP month visit. Evaluation of rate response performance for the Medtronic Kappa 700 Series pacemaker was conducted using the Metabolic Chronotropic model described by Wilkoff as applied by
2
. Automatic polarity configuration data were collected at implant.
Kay Sensing Assurance and Capture Management data were collected at each follow-up.

Description of Patients

Patients enrolled in the study represented a general dual chamber pacing population.
1
) at their one
1
Benditt, David G. M, Editor, Rate Adaptive Pacing, Blackwell Scientific Publications, Boston. 1993: 63-65.
2
Kay, Neal G., “Quantitation of Chronotropic Response: Comparison of Methods for Rate-Modulating Permanent Pacemakers”, JACC 20(7):1533-41, Dec 1992.
27
Page 28
Clinical Studies

Results of the Study

Table 3 summarizes the results of the clinical study. The incidence of complications was found to be similar to that experienced by similar devices. The performance of the automatic polarity configuration, Capture Management, Sensing Assurance, and rate response features were found to meet study objectives.
The slope of the exercise rate response (1.0 target slope) was less than 0.65 for 26 of 87 (30%) of patients.
28
Page 29
Clinical Studies
All patients implanted (n=288 devices in 285 patients, 133 device years)
Primary Objectives
Automatic Polarity Configuration (n with loss of output / N leads)
Total Leads 0% (0/546) [0%, 0.55%] ≤5%
Unipolar 0% (0/107) [0%, 2.8%] ≤5%
Bipolar 0% (0/439) [0%, 0.7%] ≤5%
Sensing Assurance
(n with loss of sensing or oversensing / N device years)
Atrial 13.5% (18/133) [8.8%, 20.5%] ≤35.7%
Ventricular 0.8% (1/133) [0.2%, 4.1%] ≤9.2%
Capture Management (n with all causes loss of capture / N device years)
Loss of capture 5.3% (7/133) [2.6%, 10.5%] ≤10.7%
Slope of MPREP rate response at 1 month (n=87 patients)
Mean 0.81 [0.76, 0.86] [0.65, 1.35]
Table 3. Effectiveness Analysis
Percent of
events
% (n/N)
95%
Confidence
interval
Criteria:
Upper
95% CI
The Medtronic Kappa 700 Series pacemaker’s Rate Profile Optimization (RPO) governs sensor indicated rate (SIR) output. Figure 1 shows the SIR vs. the Wilkoff predicted heart rate achieved using the RPO feature during the MPREP tests at 1 month.
29
Page 30
Clinical Studies
All patients reaching Anaerobic Threshold, N=87
Expected (Wilkoff) rate, mean and 95% CI
100
90
80
70
60
50
40
30
SIR (normalized)
20
10
0
0 204060
p
p
U
I
C
5%
9
r
e
n
a
e
M
Lower 95% CI
S
e
t
c
e
p
x
E
)
7
8
=
N
(
R
I
MPREP Workload (normalized)
Figure 1. Sensor Indicated Rate (SIR) vs. Expected Rate at One Month
)
f
f
o
k
l
i
W
(
d
80
R
I
S
100
30
Page 31
Chapter 2 - Implanting the Pacemaker
Implantation Procedures 32
Implant Documentation 42
Parameter Programming at Implant 43
Medical Therapy Interactions 49
Assistance 52
31
Page 32

Implantation Procedures

Implantation Procedures

Testing Leads and Pacemaker

Equipment to Test the Pacemaker and Lead
To test the performance of the lead, Medtronic recommends using a constant voltage device such as the Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA).
Caution: Do not use constant current devices (such as the Medtronic external pacemaker Models 5880A, 5375, 5348, or 5346) to test lead performance. They may damage the pacemaker’s constant voltage output circuits.
For further procedures on determining thresholds and analyzing pacemaker operation, consult the PSA technical manual.
Note: Wait at least 15 minutes after implanting screw-in leads (nonsteroid) before measuring final stimulation thresholds and intracardiac sensing potentials.
Determining Stimulation Threshold
While testing the lead system prior to implant, Medtronic recommends taking and verifying threshold measurements in both the unipolar and bipolar polarities for all of the pacemaker models. Stimulation thresholds less than 1.0 V at a 0.5 ms pulse width are recommended for acute ventricular leads. Atrial leads may have slightly higher stimulation thresholds.
Test for the following configurations:
Lead tip to case (for unipolar pacing and sensing).
Lead tip to lead ring (for bipolar pacing and sensing).
If the stimulation threshold of a mature chronic lead exceeds 2.5 V at a pulse width of 0.5 ms, consider replacing the lead.
Note: Low-profile 3.2 mm bipolar and IS-1 BI leads can be connected to the pacemaker in only one way, i.e., distal (tip)
32
Page 33
Implantation Procedures
electrode to the negative terminal. Because of this, the negative clip of the PSA connecting cable must be connected to the lead connector pin for correct threshold measurement. (See PSA technical manual for more information.)
Determining Lead Impedance
Acute lead impedance or pacing impedance for standard leads should typically fall between 250 and 1000 ohms, although it may fall outside this range for short periods of time.
Acute lead impedance for leads designed with impedance at or greater than 1000 ohms should typically fall between 500 and 2500 ohms. High impedances with these leads are acceptable as long as stimulation and sensing thresholds are acceptable.
If the lead or pacing impedance remains outside of the recommended range, reposition or replace the lead.
Determining Intracardiac Sensing Potentials
The intracardiac signal must be of sufficient strength to be sensed by the pacemaker. Intracardiac signals on chronic leads have a lesser magnitude and slew rate.
Note: Acceptable stimulation thresholds are not an indication of adequate sensing potentials.
33
Page 34
Implantation Procedures
Analyzing Pacemaker Operation
Before connecting the leads, check the pacemaker for proper operation with a Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA). Note that specifications for device operation apply at body temperature. As a result, some variance may occur if measurements are taken at room temperature.
During this test, the setscrew(s) in the connector assembly must contact the PSA’s cable probes. Use the appropriate cable probes to match the connector ports for K
DR700 Series models.
34
Page 35
Implantation Procedures

Connecting Leads to the Pacemaker

KDR700 Series pacemakers have two lead connector ports. All models except the unipolar-only Model K connector contact aligned behind one another. (See Figure 2.)
DR706 have a tip and ring
Tip Electrodes
Ring Electrodes
Atrial Port
Ventricular Port
Figure 2. Example of Bipolar Connector Assembly for
Model K
DR701
The Models KDR701, KDR721, KDR731 use a setscrew to contact the lead tip electrode, but do not use a setscrew to contact the ring electrode. The Models K
DR703 and KDR733 contact the lead with
setscrews at the tip and ring electrodes.
For information regarding acceptable lead type for each model, refer to “Lead Requirements, Compatibility” on page 96.
35
Page 36
Implantation Procedures
Connection Procedure
Connecting the leads to the pacemaker consists of three steps:
1. Insert the provided torque wrench into the bisected grommet as shown in Figure 3.
a. Check that the setscrew is retracted from the connector
port by visual inspection of the opening. If a setscrew is obstructing the pathway, rotate the setscrew counterclockwise one or two revolutions until the opening is no longer obstructed.
Note: Counterclockwise rotation beyond this point may
disengage the setscrew from the connector block.
b. Eliminate the “piston effect” (difficulty in inserting the lead
due to compression of trapped air in the connector port) when the lead is inserted by leaving the torque wrench in the distal grommet until the lead is secure. This allows a pathway for venting trapped air when the lead is inserted. The rubber connector seal maintains the torque wrench in position.
36
Page 37
Implantation Procedures
a
Figure 3. Connector Port Setscrew Preparation
b
2. Push the lead into the connector port until the lead pin is
clearly visible in the pin viewing area as indicated by arrows in Figure 4 and described in the table below.
Note: Lubricating the lead end with sterile water may ease lead insertion.
Lead-Pin Tip Placement In Pin Viewing Area
Models K K
Models K
Model K
DR701,
DR721, KDR731
DR703, KDR733 IS-1 BI Just becomes visible
DR706 5 or 6 mm unipolar Visible at end of area
IS-1 BI Visible at end of area
Low-profile
3.2 mm bipolar
Visible at end of area
37
Page 38
Implantation Procedures
KDR701, KDR721, KDR731
K
DR706
K
DR703, KDR733
(IS-1 BI leads)
Figure 4. Lead Insertion Viewing Area
KDR703, KDR733
(with 3.2 mm bipolar leads)
38
Page 39
Implantation Procedures
3. Tighten the setscrew by turning clockwise until the torque wrench clicks as indicated in Figure 5. The setscrew(s) must
be engaged for pacing to begin. Do not pull on the lead until all setscrews have been tightened.
KDR701, KDR721, KDR731
K
DR706
K
DR703, KDR733
Figure 5. Tightening Setscrews
Caution: Do not overtighten the setscrew(s). Do not use blue handled or right angled hex wrenches, since they have torque capabilities greater than is designed for this connector. The torque wrench supplied with the pacemaker is designed to slip when overtorqued to prevent damage to the socket or the setscrews. Bending the wrench may break it.
39
Page 40
Implantation Procedures

Pacemaker Implantation

Proper placement can help facilitate lead wrap and prevent muscle stimulation and device migration.
Pacemaker Orientation
The pacemaker may be used in either right or left pectoral implants. Either side of the shield can face the skin to facilitate excess lead wrap. However, pacemakers with an insulative coating must be implanted with the uncoated side facing the patient’s skin.
Avoiding Muscle Stimulation
If the patient experiences muscle stimulation while being paced in the unipolar configuration, program the device to lower output settings, such as a lesser Amplitude and/or narrower Pulse Width, without compromising the patient’s stimulation safety margin.
Securing the Pacemaker (Suture Hole)
A suture placed through the suture hole located in the connector assembly area helps secure the pacemaker in the subcutaneous pocket, thus minimizing post-implant rotation and migration of the device. Use normal surgical needles to penetrate the suture hole.

Replacing an Implanted Pacemaker

When replacing a pacemaker:
1. Program the pacemaker to a nonrate responsive mode prior to explantation if previously programmed to a rate responsive mode. This avoids any potential rate increase while handling the pacemaker.
40
Page 41
Implantation Procedures
2. Insert the proper wrench through each slit in the rubber grommet and loosen each setscrew by turning counterclockwise.
3. Gently retract the lead from the pacemaker connector.
4. If the lead pin shows signs of pitting or corrosion, the entire
lead should be discarded and replaced with a new lead to assure the integrity of the pacing system.
5. Measure stimulation thresholds and sensing potentials.
6. Pacemaker-dependent patients will be without pacing support when the lead is disconnected. Therefore, have a temporary external pacing instrument available for use.
Caution: Electrosurgical cautery could induce ventricular
arrhythmias and/or fibrillation, or may cause asynchronous or inhibited pacemaker operation. Refer to “Electrosurgical Cautery” on page 50 for information on electrocautery use.
When replacing a Medtronic pacemaker or a pacemaker from another manufacturer, the physician may require a lead adaptor kit to connect the pacemaker to the chronic lead. Use the appropriate hex wrench or screwdriver to disconnect the chronic lead. Medtronic provides such wrenches in adaptor kits for use with pacemakers manufactured by Biotronik, Coratomic, CPI, Intermedics, Pacesetter, Telectronics/Cordis, and Vitatron. See the Medtronic publication Pacemaker and Connector Encyclopedia under “Adaptors and Accessories” for the appropriate adaptor kits.
For questions about lead and pacemaker compatibility, consult your Medtronic representative, or call Medtronic Technical Services.
41
Page 42

Implant Documentation

Disposal of Pacemaker

When replacing the pacemaker (due to battery depletion or explanting the pacemaker at the death of the patient who is to be cremated), return the pacemaker to Medtronic for analysis and disposal. Medtronic recommends cleaning and sterilizing the pacemakers before returning. See the back cover for mailing addresses.
Implant Documentation

Pacemaker Registration

A registration form is included in the shipping package for each Medtronic implantable pacemaker. Upon completion of the registration form by the clinician, this form serves as a permanent record of facts related to the implanted device. A copy of this form should be returned to Medtronic, where vital information will be transferred to a wallet-size, plastic-coated pacemaker Identification Card which will be mailed directly to the patient. A temporary identification card is included in the shipping package for use by the patient until the permanent card arrives. Refer to the back cover for mailing addresses.

Establishing a Patient Record

At the time of implant the clinician should establish a record of the patient’s pacemaker programmed parameter settings for later reference. A patient record can give the clinician an accurate account of the patient’s medical history at subsequent follow-ups when evaluating the pacemaker’s performance. The registration form should not serve this purpose since the form is intended for registering the patient and pacemaker with Medtronic.
42
Page 43

Parameter Programming at Implant

Parameter Programming at Implant

Shipping Parameters

The pacemaker is configured to a set of shipping parameters at the time of manufacture. At implant the device operates at these settings unless reprogrammed prior to implant. Refer to“Shipping and Nominal Parameter Settings” on page 99 for specific shipping parameter settings.
In the event that the pacemaker is not configured to shipping settings at implant, the device may have changed to partial or full electrical reset settings due to extreme temperatures or intense electromagnetic interference.
Caution: If, at implant, a programmer interrogation indicates the pacemaker has been reset to VVI, 65 ppm full electrical reset settings, use the programmer to reset the device and reprogram the previous mode and pacing operations. If, after a reset, a message appears on the programmer screen/printout indicating the Date/Time memory has been altered, contact your Medtronic representative for further information.

Implant Detection

Implant Detection is a 30-minute period, beginning at lead insertion, during which the pacemaker verifies that a lead(s) has been connected by measuring lead impedance. After 30 minutes of continuous lead connection, the pacemaker completes Implant Detection and activates the pacemaker’s automatic features listed in “Automatic Operation” on page 44.
43
Page 44
Parameter Programming at Implant

Automatic Operation

At the completion of Implant Detection, these features are automatically enabled or begin adaptation from their shipping settings.
Operating polarities, which are determined during Implant Detection, are confirmed at its completion.
Mode becomes DDDR (though shipped in the DDDR mode, the device effectively operates in the DDD mode until Implant Detection is completed).
Rate Profile Optimization is enabled.
Capture Management is enabled, and Ventricular Amplitude and Pulse Width become adaptive.
Sensing Assurance is enabled, and Sensitivity becomes adaptive.
Search AV is enabled, and Paced AV and Sensed AV become adaptive.
Diagnostics are enabled.
If the physician manually programs Implant Detection to Off/ Complete during the 30-minute detect time, the automatic operation of the above features will be activated at that time. If a programming session is initiated at any time during Implant Detection, the pacemaker will reset Implant Detection.
44
Page 45
Parameter Programming at Implant

Manual Programming

If pacing rates, pacing outputs, and sensitivities are manually programmed at implant, the physician should be aware of the following programming considerations:
Pacing Rates
Program the Lower Rate to a setting that permits the patient to sleep comfortably or program the Sleep Function On. Program the ADL Rate to provide adequate cardiac output during typical patient activity. Program the Upper Sensor Rate and Upper Tracking Rate to provide cardiac output that meets the patient’s metabolic demand during exercise without provoking symptoms (e.g., angina, palpitations, etc.).
Note: For patients with stable angina, program the Upper Sensor Rate and Upper Tracking Rate to a value below the rate at which angina occurs.
Programming a combination of high Upper Sensor Rate and Upper Tracking Rate and a long refractory period may result in a shorter “sensing window.” Loss of sensing in such cases could result in competitive pacing. Programming PVARP to “varied” or “auto” may minimize competitive pacing.
Programming the Upper Tracking Rate to a value greater than the Upper Sensor Rate has the effect of allowing the atrial rhythm to be tracked to a rate higher than the sensor-driven rate.
45
Page 46
Parameter Programming at Implant
Pacing Outputs
Strive for acute ventricular stimulation thresholds of 1.0 V amplitudes at 0.5 ms pulse widths. Atrial stimulation thresholds may be slightly higher. Do not use chronic atrial or ventricular leads if the stimulation thresholds exceed 2.5 V amplitude at 0.5 ms pulse width.
The clinician should consider programming the pacemaker’s atrial and ventricular outputs to allow 2:1 voltage safety margins above the measured stimulation threshold values. Capture Management can also be programmed to provide ongoing ventricular capture monitoring and adjustment.
Sensitivity
Intracardiac signal amplitudes and slew rates decrease during the lead maturation process. The clinician should consider 2:1 or 3:1 atrial and ventricular sensitivity safety margins when selecting sensitivity settings, especially with acute leads. After establishing the intracardiac sensing thresholds, the clinician may wish to program the atrial and ventricular sensitivities to more sensitive settings (i.e., lower numerical settings than the measured sensing thresholds). Weigh these lower settings against the possibility of oversensing noise and electromagnetic interference, especially with unipolar sensing.
Ventricular sensitivities of 1.0 mV or 1.4 mV with wide atrial pulse widths or high atrial amplitudes may result in ventricular safety pacing with some lead systems at high sensor-driven pacing rates. Reprogramming ventricular sensitivity to a less sensitive setting (higher numerical value) is one option under such circumstances. Other options include programming a longer Ventricular Blanking period.
Note: Adequate intracardiac signals should be present in both the unipolar and bipolar polarities when using bipolar leads. Measure sensing potentials in the unipolar polarity (lead tip to case) and bipolar polarity (lead tip to lead ring) prior to connecting the leads.
46
Page 47
Parameter Programming at Implant
Self-Inhibition
Self-inhibition in single chamber modes results from the sensing of the pacing output pulse. This is more likely to occur with high amplitudes, wide pulse widths, and low sensitivity settings. Reprogramming the Sensitivity to a less sensitive setting (higher numeric value) will usually resolve the situation.
In addition, inhibition can occur when far-field R-waves are sensed in AAIR, ADIR, AAI, and ADI modes. This can be prevented by proper programming of the single chamber Atrial Refractory Period and Atrial Blanking.
Muscle Stimulation with Unipolar Pacing
Under certain circumstances (e.g., high output settings), pacemaker­induced muscle stimulation may occur at the pocket site with the pacemaker programmed to ventricular unipolar pacing. Pacemaker­induced muscle stimulation may be effectively controlled and/or eliminated by programming Pulse Width to a narrower setting or programming a lower Amplitude.
47
Page 48
Parameter Programming at Implant
Adapting Other Parameters
In addition to programming those parameters discussed above, the physician should consider adjusting other nominal parameters based on the patient’s history, medical condition, lifestyle, and hemodynamic potential and need.
Rate response parameters: Rate Profile Optimization (ADL Response and Exertion Response), Activity Threshold, Activity Acceleration, and Activity Deceleration.
Timing intervals: Sensed AV (SAV) interval and Paced AV (PAV) interval, Rate Adaptive AV (RAAV), and Search AV.
Refractory and blanking periods: Ventricular Refractory Period, Ventricular Blanking, Post-Ventricular Atrial Refractory Period and Post-Ventricular Atrial Blanking.
In addition, the physician should consider programming several other pacing features appropriate to the patient’s special pacing needs. These features might include:
Mode Switch for managing atrial arrhythmias,
Rate Drop Response for backup pacing during drops in heart rate,
Sinus Preference for a greater range of sinus activation of the ventricles, and
Sleep Function and Single Chamber Hysteresis for other pacing operations during inactivity.
Note: Non-Competitive Atrial Pacing, PVC Response, and Ventricular Safety Pacing are On/Off features that are programmed On at the time of manufacture.
48
Page 49

Medical Therapy Interactions

Medical Therapy Interactions
This section provides details on possible interactions between pacemaker therapy and other medical therapies.

Implantable Defibrillators

Some pacemaker patients may also require therapy from an implantable defibrillator. Only bipolar pacing should be used with these patients. In some cases, pacing in the unipolar configuration may cause the defibrillator either to deliver inappropriate therapy or to withhold appropriate therapy.
Regarding use of K implantable defibrillator, note the following:
The use of unipolar-only Model KDR706 and the KDR700 Series bipolar models implanted with unipolar leads is contraindicated for patients having an implantable defibrillator.
KDR700 Series bipolar models implanted with bipolar leads may be used in patients having an implantable defibrillator. With these pacemakers, however, polarity is automatically configured during Implant Detection. (See “Automatic Polarity Configuration” on page 64.) If lead integrity is suspect, confirmation of the automatically programmed polarities should be made after completion of Implant Detection in order to assure that bipolar polarities have been programmed appropriately.
The implantable cardiac defibrillator (ICD) should be turned off during pacemaker implantation procedures until lead polarities have been configured and confirmed. This is to prevent possible back-up unipolar paces from triggering the ICD.
Although these pacemakers are designed to be compatible with implantable defibrillators, the potential does exist for a defibrillation pulse to reset them.
DR700 Series models in patients having an
49
Page 50
Medical Therapy Interactions
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection. If lead integrity is suspect, confirmation of bipolar polarity should be made after completion of Implant Detection.
Follow implant protocol and precautions for pacemaker and defibrillator lead placement. Ensure the pacemaker is configured to be compatible with the defibrillator.
Warni ng: Lead Monitor should not be programmed to Adaptive. When a prevalence of out-of-range lead impedance paces is detected, the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Warni ng: Programming Transtelephonic Monitor On is contraindicated for patients with implantable defibrillators. When it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.

Electrosurgical Cautery

Electrocautery may cause permanent loss of output prior to explanting the pacemaker and can induce ventricular fibrillation. The electrocautery tip should never be used within 6 inches (15 cm) of an implanted pacemaker/lead system.
The use of electrocautery may cause any of the following: triggered or inhibited output, a temporary pause in output, permanent loss of output, a rate in excess of the 200 ppm rate limit, reversion to asynchronous operation, an electrical reset, or premature triggering of the elective replacement indicator.
50
Page 51
Medical Therapy Interactions
Because of these potential complications, alternatives to electrocautery should be used when available.
Effects to the pacemaker vary considerably based on the type of electrocautery unit, coagulation and cutting current settings, current pathway from the cautery tip to the indifferent plate, and the pacemaker/lead system. Electrocautery units could also interfere with ECG monitoring equipment. Monitor cardiac activity via blood pressure or pressure pulses when using electrocautery.
If electrocautery cannot be avoided, follow these steps to minimize complications:
1. Program the device to the VOO/AOO (non-rate responsive) mode beforehand and avoid direct contact with the pacemaker or leads to avoid induction of tachyarrhythmias and/or fibrillation.
2. Position the ground plate so that the current pathway does not pass through or near the pacemaker system, i.e., place ground plate under the patient’s buttocks or legs during abdominal surgery.
3. Use short, intermittent, and irregular bursts at the lowest feasible energy levels.
4. Use a bipolar electrocautery system, where possible.
5. Have temporary pacing and defibrillation equipment available.

X-ray and Fluoroscopy

X-ray and fluoroscopy testing of pacemakers similar to the KDR700 Series pacemakers by exposure to diagnostic X-ray or fluoroscopic radiation has not affected those pacemakers.
51
Page 52

Assistance

Assistance

Assistance Information

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise. For more information, contact your local Medtronic representative or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
52
Page 53
Chapter 3 - Description
Pacing Mode Operations 54
Rate Responsive Pacing 57
Timing Operations 58
Pacing and Sensing Operations 64
Special Therapy Options 71
53
Page 54

Pacing Mode Operations

Pacing Mode Operations
A brief description of programmable mode operations is given below.
Warni ngs:
Do not use rate responsive modes in those patients who cannot tolerate pacing rates above the programmed Lower Rate.
Do not use single chamber atrial modes in patients with impaired AV nodal conduction because ventricular capture cannot be assured.

DDDR / DDD Modes

In the presence of sensed atrial events, the ventricle will track the atrium in both the DDDR and DDD modes. Both chambers are paced at the sensor-indicated rate (DDDR) or the programmed Lower Rate (DDD) in the absence of sensed intrinsic events. An atrial sensed event inhibits atrial pacing and initiates a ventricular pace after the Sensed AV (SAV) interval. An atrial paced event initiates a ventricular pace after the Paced AV (PAV) interval. A ventricular sensed event during either AV interval inhibits the ventricular pace.

DDIR / DDI Modes

Sensed atrial events are not tracked by the ventricle; such events however, inhibit the scheduled atrial pace. Both chambers are paced at the sensor-indicated rate (DDIR) or the programmed Lower Rate (DDI). An atrial paced event initiates a ventricular pace after the Paced AV (PAV) interval, unless inhibited by a sensed ventricular event.

DVIR / DVI Modes

Both chambers are paced at the sensor-indicated rate (DVIR) or the programmed Lower Rate (DVI) in the absence of sensed ventricular events. An atrial paced event initiates a ventricular pace after the Paced AV (PAV) interval, unless inhibited by a sensed ventricular event.
54
Page 55
Pacing Mode Operations

VDD Mode

In the presence of sensed atrial events, the ventricle will track the atrium in the VDD mode. The ventricle is paced at the programmed Lower Rate in the absence of sensed intrinsic events. A sensed atrial event initiates a ventricular pace after the Sensed AV (SAV) interval unless inhibited by a sensed ventricular event. The SAV interval may delay the ventricular pace at a rate slower than the programmed Lower Rate. No pacing occurs in the atrium.

VVIR / VDIR, AAIR / ADIR Modes

The ventricle is paced at the sensor-indicated rate (VVIR and VDIR) in the absence of intrinsic ventricular events. Events in the atrium are also sensed and recorded for diagnostic purposes (VDIR). The atrium is paced at the sensor-indicated rate (AAIR and ADIR) in the absence of intrinsic atrial events. Events in the ventricle are also sensed and recorded for diagnostic purposes (ADIR).

VVI / VDI, AAI / ADI, VVT / AAT Modes

The ventricle is paced at the programmed Lower Rate (VVI and VDI) in the absence of intrinsic ventricular events. Events in the atrium are also sensed and recorded for diagnostic purposes (VDI). The atrium is paced at the programmed Lower Rate (AAI and ADI) in the absence of intrinsic atrial events. Events in the ventricle are also sensed and recorded for diagnostic purposes (ADI).
Sensed ventricular events in the VVT mode trigger pacing pulses. Sensed atrial events in the AAT mode trigger pacing pulses.
Note: Programmed triggered pacing will not occur faster than 300 ms (200 ppm) from the previous paced event. Temporary programmed triggered pacing is not limited to 300 ms (200 ppm).
55
Page 56
Pacing Mode Operations

DOOR / DOO, VOOR / VOO, AOOR / AOO Asynchronous Modes

The DOOR mode paces the atrium and ventricle sequentially at the sensor-indicated rate, whereas the DOO mode paces both chambers at the programmed Lower Rate. The VOOR mode paces the ventricle at the sensor-indicated rate, whereas the VOO mode paces at the programmed Lower Rate. The AOOR mode paces the atrium at the sensor-indicated rate, whereas the AOO mode paces at the programmed Lower Rate.

ODO / OVO / OAO Diagnostic Modes

The ODO mode senses both intrinsic atrial and ventricular events. The OVO mode senses only intrinsic ventricular events, whereas the OAO mode senses only intrinsic atrial events.
Warni ng: These modes should never be programmed for pacemaker-dependent patients. For such patients, the programmer’s “Inhibit” function may be used for brief interruption of outputs.
56
Page 57

Rate Responsive Pacing

Rate Responsive Pacing
The KDR700 Series pacemakers use activity-based pacing, which varies the pacing rate in response to the patient’s detected physical motion. In rate responsive modes, the pacing rate varies between the Lower Rate and Upper Sensor Rate limits and is dependent on the following programmable parameters:
Activity Threshold establishes the minimum signal amplitude above which activity signals are detected by the activity circuit.
Activity Acceleration and Deceleration times determine how quickly the rate will rise and fall as a result of changes in patient physical activity. Deceleration can also extend the rate fall by providing up to 20 minutes of rate deceleration based on the intensity and duration of exercise.
Rate response is initialized and DDDR pacing becomes effective upon completion of Implant Detection.

Rate Profile Optimization

Rate Profile Optimization controls how rapidly and to what level the sensor-indicated rate increases and decreases within submaximal and maximal rate ranges:
Submaximal rates are moderate pacing rates achieved during typical daily patient activities. These rates are at or near the ADL (Activities of Daily Living) Rate.
Maximal rates are rates at or near the Upper Sensor Rate achieved during vigorous activities.
Rate response for rates below the submaximal rate range (i.e., at or near the Lower Rate) are also controlled by Optimization.
Rate response function is optimized in the submaximal and maximal rate ranges by a daily comparison of data in two rate profiles:
Sensor rate profile is an actual rate distribution of the patient’s averaged sensor-indicated rates.
57
Page 58

Timing Operations

Target rate profile is a programmable rate distribution of the patient’s desired rates. The ADL Response and Exertion Response parameters define the time spent pacing in the submaximal and maximal ranges, respectively.
Once each day, the pacemaker evaluates the percent of time spent pacing in both the submaximal and maximal rate ranges by comparing the sensor rate profile against the target rate profile. From this comparison, the pacemaker automatically adjusts rate response in both rate ranges.
Timing Operations

A-A Timing and Mean Atrial Rate

KDR700 Series pacemakers use A-A timing, that is, in all dual chamber modes that pace the atrium, the pacemaker mimics sinus rhythm by timing from atrial event to atrial event. The pacemaker keeps a running average of all A-A intervals (except those starting with an atrial sense or atrial refractory sense and ending with an atrial pace) which is called the Mean Atrial Rate (MAR). The Mean Atrial Rate is used by the pacemaker in Rate Adaptive AV and automatic Post-Ventricular Atrial Refractory Period (PVARP) operation.

Rates

The pacemaker has four rates that determine the patient's pacing rate.
The programmable Lower Rate is the minimum pacing rate for a given mode.
The programmable Upper Tracking Rate is the maximum ventricular rate at which the atrium is tracked in the DDDR, DDD, and VDD mode.
ADL Rate (Activities of Daily Living Rate) is the average target rate that the patient’s heart rate is expected to reach during moderate exercise.
58
Page 59
Timing Operations
Upper Sensor Rate provides the upper limit for the sensor-driven rate during physical activity, particularly during vigorous exercise.

AV Intervals

The pacemaker has separately programmable paced and sensed AV intervals that determine the duration between an atrial event and the delivery of a ventricular stimulus in dual chamber modes.
The programmable Paced AV (PAV) interval defines the duration from paced atrial events to paced ventricular events. The ventricle is paced after the PAV expires unless inhibited by a sensed ventricular event.
Note: PAV duration may differ from the programmed setting due to Mode Switch, Rate Adaptive AV, Search AV, Ventricular Safety Pacing, Non-Competitive Atrial Pacing or Capture Management operation.
The programmable Sensed AV (SAV) interval defines the duration from sensed atrial events to paced ventricular events. The ventricle is paced after the SAV expires unless inhibited by a sensed ventricular event. SAV maintains 1:1 AV tracking up to the Upper Tracking Rate (UTR) or the total atrial refractory period (which is the sum of the SAV and the PVARP).
Note: SAV duration may differ from the programmed setting due to Mode Switch, Rate Adaptive AV, Search AV, automatic PVARP, or Capture Management operation or may be extended as the result of Wenckebach operation if the intrinsic atrial rate exceeds the UTR.
59
Page 60
Timing Operations

Rate Adaptive AV

The programmable On or Off Rate Adaptive AV (RAAV) feature provides rate-variable PAV and SAV intervals. The variable PAV shortens or lengthens to the sensor-indicated rate, whereas the variable SAV shortens or lengthens to the Mean Atrial Rate. The variable SAV promotes 1:1 AV tracking to sensed atrial events at higher rates. Several programmable parameters define RAAV operation.
Start Rate defines the rate at which RAAV operation begins.
Stop Rate defines the rate above which the PAV and SAV intervals stop shortening and remain at the minimum AV interval.
Max Offset defines the maximum amount of time by which the PAV and SAV intervals can be shortened.
As the sensor-indicated rate increases above the Start Rate, the PAV gradually shortens until the programmed Max Offset is reached at the Stop Rate. As the Mean Atrial Rate increases above the Start Rate, the SAV gradually shortens until the programmed Max Offset is reached at the Stop Rate.
The approximate interval difference between the programmed PAV and SAV is maintained throughout the rate range when RAAV is active. The pacemaker operates at the programmed PAV and SAV at rates below the programmed Start Rate.
Notes:
The 2:1 block rate increases to a higher rate as the SAV component of the TARP shortens.
The sensing window after the PVARP lengthens as the PAV shortens at higher rates.
60
Page 61
Timing Operations

Search AV

The Search AV feature provides the physician with two methods for measuring and responding to AV conduction in patients with intact or intermittent AV conduction. By adjusting the programmed AV intervals to the patient’s intrinsic AV conduction, the pacemaker can mimic normal physiology and promote tracking of fast atrial rates.
When Search AV is programmed to Adaptive, the pacemaker evaluates conduction every 16 AV sequences. Based on the measured conduction times for the 16 AV sequences, the pacemaker classifies the AV complexes as too short, too long, or adequate. If the complexes are adequate, no adjustment is made, but if they are too short or long, the pacemaker adds or subtracts to the SAV and PAV for the next 16 pacing cycles to promote intrinsic activation. The programmable Maximum Offset defines the maximum amount by which the PAV and/or SAV intervals can be lengthened.
If Rate Adaptive AV is On, the pacemaker adjusts the AV intervals above the Rate Adaptive AV curve to promote intrinsic conduction. If Rate Adaptive AV is Off, it adjusts the intervals above the programmed PAV and SAV values.
Search AV, programmed to adaptive, becomes operational automatically when Implant Detection is completed.
When Search AV is programmed to a fixed AV hysteresis delay value, the pacemaker modulates the AV intervals according to whether or not the previous complex ended in a ventricular pace or sense. If a ventricular sense, the SAV and PAV for the next cycle will be lengthened by the programmed fixed AV hysteresis delay value.
Note: The pacemaker cannot operate adaptively and at a fixed hysteresis setting simultaneously.
61
Page 62
Timing Operations

Upper Rate Behaviors (2:1 Block Rate, Wenckebach)

The 2:1 block rate occurs when every other atrial event falls within the PVARP. Such refractory sensed events do not initiate an SAV and, thus, are not synchronized to a ventricular paced event, thereby producing a 2:1 block rate. If the 2:1 block rate exceeds the programmed Upper Tracking Rate, the SAV is extended to limit the ventricular rate at the Upper Tracking Rate, thereby producing pacemaker Wenckebach operation.

Refractory and Blanking Periods

The pacemaker has several refractory periods to avoid inappropriate restarting of certain timing intervals.
The programmable Post-Ventricular Atrial Refractory Period (PVARP) prevents tracking of retrograde P waves. In the DDIR and DDI modes, PVARP prevents atrial inhibition from retrograde P waves. The PVARP occurs after paced, sensed, and refractory sensed ventricular events and is set to at least 400 ms after a pacemaker-defined PVC.
The first portion of the PVARP is a programmable Post-Ventricular Atrial Blanking period which disables atrial sensing after paced, sensed, and refractory sensed ventricular events.
– The programmable sensor-varied PVARP protects against
PMTs at lower sensor-indicated rates by lengthening the PVARP and provides a higher 2:1 block rate at higher sensor-indicated rates by shortening the PVARP. Sensor-varied PVARP is based on the sensor-indicated pacing rate and varies from approximately 400 ms at the lower rate to no less than the Post-Ventricular Atrial Blanking period at higher sensor rates.
The programmable automatic PVARP protects against
PMTs and against transitory 2:1 block in tracking modes. Automatic PVARP maintains an appropriately high 2:1 block rate by calculating a new target 2:1 block point,
62
Page 63
Timing Operations
based on the Mean Atrial Rate plus 30 bpm, every four pacing cycles. The pacemaker then adjusts the PVARP to produce a Total Atrial Refractory Period equal to the target 2:1 block point. The minimum 2:1 block point is 100 bpm, and the maximum is the Upper Tracking Rate plus 35 bpm. The programmable Minimum PVARP parameter defines the limit to which the PVARP can be adjusted.
The programmable Ventricular Refractory Period (VRP) prevents sensing of T-waves or PVCs. The VRP occurs after paced, sensed, and refractory sensed ventricular events in dual chamber and ventricular single chamber modes.
The first portion of the VRP is a nonprogrammable Ventricular Blanking period in which ventricular sensing is disabled after paced, sensed, and refractory sensed ventricular events. The Ventricular Blanking period varies dynamically from 50 to 100 ms based on the strength and duration of the ventricular event.
The programmable Ventricular Blanking period disables ventricular sensing in dual chamber modes to prevent potential crosstalk from paced atrial events.
The programmable Atrial Refractory Period (ARP) prevents atrial inhibition due to sensed far-field R waves or noise. The ARP occurs after paced, sensed, and refractory sensed atrial events in single chamber atrial modes.
The first portion of the ARP is a programmable Atrial Blanking period that disables atrial sensing after paced, sensed, and refractory sensed atrial events.
The first portion of the PAV and SAV intervals is a nonprogrammable Atrial Blanking period in which atrial sensing cannot take place after paced or sensed atrial events. The Atrial Blanking period varies dynamically from 50 to 100 ms based on the strength and duration of the atrial event.
63
Page 64

Pacing and Sensing Operations

Pacing and Sensing Operations

Polarity

Unipolar polarity uses the lead tip as the active electrode and the pacemaker case as the common electrode. Bipolar polarity uses the lead tip as the active electrode and the lead ring as the common electrode.
With the K polarities are automatically or manually programmable and are independently programmable for the atrial and ventricular channels. The Model K
Automatic Polarity Configuration
With bipolar pacemakers, polarity settings are selected automatically during the pacemaker’s 30-minute Implant Detection period that begins at lead insertion. During Implant Detection, with either bipolar or unipolar leads implanted, and with the Configure or Adaptive parameter under Lead Monitor active, the pacemaker continuously measures impedance as follows:
DR700 Series bipolar pacemakers, pacing and sensing
DR706 provides unipolar pacing and sensing only.
The pacemaker issues a bipolar pace and immediately checks the pace for high impedance.
– If the pace is within range, it is considered an acceptable
bipolar pace.
– If high impedance is found, the pacemaker immediately
follows the bipolar pace with a backup unipolar pace.
If a unipolar backup pace is issued, the pacemaker checks it for high impedance.
– If the pace is within range, it is considered an acceptable
unipolar pace.
– If high impedance is found, the pacemaker assumes a lead
is not attached and restarts Implant Detection.
Note: During polarity configuration, the pacemaker also checks for low impedance paces but does not follow them with
64
Page 65
Pacing and Sensing Operations
unipolar backup paces. To avoid possible loss of capture due to continuous low impedance pacing, the pacemaker sets unipolar polarity when 3 of 16 paces are detected as low impedance during the first phase of configuration (described below).
Warning: If the Implant Detection feature identifies a bipolar, high-impedance path, backup unipolar pacing will be delivered. This may be inappropriate for patients with an implantable defibrillator (ICD). See “Notes” on page 66.
Polarity configuration for bipolar pacemakers takes place in two phases:
Initial Configuration Phase – Five minutes after lead connection, if a high impedance unipolar pace has not reset Implant Detection, the pacemaker issues three asynchronous paces at magnet rate and determines if they are bipolar or unipolar.
– Two of three paces must be bipolar for initial pacing and
sensing polarities to be determined bipolar for the Confirmation Phase.
– Two of three paces must be unipolar for initial pacing and
sensing polarities to be set to unipolar and remain unipolar during and after Implant Detection.
Confirmation Phase – If bipolar polarity is determined at five minutes, the remaining 25 minutes of Implant Detection is used to confirm bipolar polarity.
The pacemaker continues to monitor each pace as
described above, providing backup unipolar paces for high impedance bipolar paces.
The pacemaker checks for a programmed number of
out-of-range (high or low impedance) paces that indicate bipolar pacing is in jeopardy.
The pacemaker confirms bipolar pacing and sensing
polarity at the end of 25 minutes by examining three
65
Page 66
Pacing and Sensing Operations
asynchronous paces as described above for the Initial Configuration Phase.
Note: If a prevalence of out-of-range bipolar paces is detected during the Confirmation Phase, the pacemaker reconfigures pacing and sensing polarities to unipolar and the asynchronous paces are not delivered.
Warni ng: When implanting a Model K
DR703 or KDR733 pacemaker,
ensure that both the tip and ring setscrews are properly engaged and all electrical contacts are sealed in order to prevent possible electrical leakage between the tip and ring contacts. Such leakage may cause the pacemaker to falsely identify a unipolar lead as bipolar, resulting in a loss of output. For the same reason, ensure that electrical contacts are sealed when using lead extenders or adaptors with all bipolar models.
Notes:
During Implant Detection, the unipolar-only Model KDR706 pacemaker examines unipolar pacing impedance to determine lead connection.
When implanting a bipolar pacemaker in a patient with an implantable defibrillator, the physician has the option to manually program polarities to avoid the possible occurrence of unipolar backup paces during Implant Detection.
66
Page 67
Pacing and Sensing Operations
Manually Programming Polarity
Pacing and Sensing Polarities for the bipolar models can also be manually programmed. However, the Configure selection under Lead Monitor, which controls automatic configuration, must be “turned off” by programming the Monitor Only selection instead.
When pacing and/or sensing polarity is manually programmed from unipolar to bipolar, the pacemaker verifies whether a functioning bipolar lead is connected by measuring lead impedance. If lead impedance is outside the accepted programmed impedance range for bipolar polarity (between 200 and 4000 ohms), the lead is assumed to be unipolar and pacing and/or sensing polarity remains set to unipolar.
Warni ng: If the clinician overrides the bipolar lead verification and programs bipolar polarity when a unipolar lead is connected to the pacemaker, no pacing output results.

Lead Monitor

The Lead Monitor feature enables the pacemaker to monitor lead integrity and adapt bipolar pacing and sensing to unipolar when bipolar lead integrity is in doubt. It also controls automatic configuration of lead polarities at implant. Lead Monitor is available in all pacing modes.
The Lead Monitor feature monitors lead integrity by looking for high and low impedance on each pace. The Atrial and Ventricular Lead Monitor can be programmed to operate as follows:
Configure – provides automatic polarity configuration of bipolar pacemakers (bipolar pacemakers are shipped in Configure).
67
Page 68
Pacing and Sensing Operations
Adaptive – monitors bipolar paces and provides backup unipolar
paces when impedance is high
– switches pacing and sensing polarities from bipolar to
unipolar when a prevalence of out-of-range (high and low impedance) paces is detected
– provides automatic polarity configuration if the pacemaker
is programmed to Adaptive prior to implant
Monitor Only – monitors unipolar and bipolar paces to determine if they are out of range (high and low impedance). Unipolar-only Model K
DR706 is shipped in and only
programmable to Monitor Only.
When Lead Monitor is set to Adaptive or Monitor Only, and the pacemaker determines a lead is out of range, the pacemaker issues a status warning that appears on the programmer screen at the next interrogation. However, a status warning is not issued at any time during the configuration process.
After automatic polarity configuration is complete, Lead Monitor is automatically set to Monitor Only for bipolar and unipolar configurations. If, however, the Adaptive setting was programmed prior to implant, bipolar configurations are set to Adaptive, but unipolar configurations are set to Monitor Only.
Warni ng: Lead Monitor should not be programmed to Adaptive for patients with an implantable defibrillator. When a prevalence of out-of-range lead impedance paces is detected, the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Caution: If the Lead Monitor detects out-of-range lead impedance, investigate lead integrity more thoroughly.
68
Page 69
Pacing and Sensing Operations

Sensitivity

Sensitivity determines the minimum intracardiac signal that the pacemaker can detect when intrinsic atrial and ventricular events occur. With K or manually programmable.
Sensing Assurance, the pacemaker's automatic sensing feature, adjusts sensitivities within defined limits. At the completion of Implant Detection, the pacemaker enables Sensing Assurance and begins continual monitoring of the peak amplitude of sensed signals. In response to monitoring, it automatically increases or decreases Sensitivity to maintain an adequate sensing margin with respect to the patient's sensed P and R waves.
During provocative myopotential testing of the Medtronic Kappa 700 Series pacemakers having Sensing Assurance On, 56% (28/50) of patients experienced atrial oversensing while 7.3% (4/55) experienced ventricular oversensing. The atrial oversensing rate is comparable to that experienced by Medtronic’s Thera devices which had an atrial oversensing rate of 52.4% (75/143) during provocative testing. The ventricular oversensing rate of Medtronic Kappa 700 Series pacemakers compared favorably to the ventricular oversensing rate experienced by Medtronic Thera pacemakers –
44.1% (63/143).
For guidelines regarding manually programming Sensitivity settings, refer to “Parameter Programming at Implant” on page 43.
Atrial and Ventricular sensitivities are separately programmable. They can also be programmed on a temporary basis.
Note: Atrial Sensitivity with unipolar sensing polarity is restricted to
0.5 mV to limit oversensing of muscle noise or electromagnetic interference.
DR700 Series pacemakers, sensitivity is automatically

Pacing Output

Amplitude and Pulse Width determine the pacing pulse strength necessary to capture the atrium and ventricle. Amplitude and Pulse
69
Page 70
Pacing and Sensing Operations
Width are both manually programmable for the atrial and ventricular channels or automatically programmable for the ventricular channel through the Capture Management feature. Amplitude and Pulse Width can also be programmed on a temporary basis. Refer to “Parameter Programming at Implant” on page 43 for pacing output guidelines.

Capture Management

The Capture Management feature provides automatic monitoring of ventricular pacing thresholds and automatic adjustment of Amplitude and Pulse Width to maintain capture.
When Capture Management is programmed to Monitor Only, the pacemaker periodically causes paces to be delivered (affecting pacemaker timing temporarily if necessary). The pacemaker then monitors the paces by changing first amplitude and then pulse width to find two points that lie on the strength duration curve that define the boundary between settings that capture and those that do not. How often the pacemaker performs this pacing threshold search is determined by the programmable Capture Test Frequency parameter. This parameter determines how often the pacing threshold search will be initiated and provides for retry if the test is delayed.
When Capture Management is programmed to Adaptive, the pacemaker responds to monitoring by adapting ventricular amplitude and pulse width using the following programmable parameters:
Amplitude Margin and Pulse Width Margin – the pacemaker­determined threshold multiplied by the selected safety factor.
Minimum Adapted Amplitude and Minimum Adapted Pulse Width – the lower limit to which Amplitude and Pulse Width can be set by the pacemaker during adaptation.
Acute Phase Days Remaining – time in days during which the pacemaker will not decrease output settings below the initially programmed settings. This parameter is used during the lead maturation period.
70
Page 71

Special Therapy Options

Capture Management becomes operational automatically when Implant Detection is completed.
Warni ng: Capture Management will not program ventricular outputs above 5.0 V or 1.0 ms. If the patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually.
Note: In the event of partial or complete lead dislodgment, Capture Management may not prevent loss of capture.
Special Therapy Options
The pacemaker has several therapy features for managing atrial and ventricular arrhythmias and other pacing operations. A brief description of these features is given below.

Mode Switch

Mode Switch is a programmable On or Off feature designed to prevent tracking of paroxysmal atrial tachycardias in the DDDR, DDD, and VDD modes. Whenever a rapid atrial rhythm is detected, the pacemaker switches from an atrial tracking mode to a non-atrial tracking mode until the atrial tachycardia ceases. Mode switching occurs as follows:
Atrial Tracking Mode Non-Atrial Tracking Mode
DDDR DDIR
DDD DDIR
VDD VDIR
→ ←
→ ←
→ ←
In determining when to mode switch, the pacemaker continuously monitors A-A intervals for a rapid atrial rate. When it determines that the atrial rate equals or exceeds the programmed Detect Rate (the rate at which pacemaker-defined atrial tachycardia starts) and is sustained for the programmed Detect Duration (time period
71
Page 72
Special Therapy Options
confirming continuation of the tachycardia), the pacemaker switches mode and gradually changes the ventricular rate to the sensor­indicated rate. When seven A-A intervals occur below the Upper Tracking Rate, or five consecutive atrial paces occur, the pacemaker switches back to the programmed atrial tracking mode.
In monitoring the atrial rate for tachycardias, the pacemaker searches for:
any four of the last seven A-A intervals that are shorter than the programmed Detect Rate interval, and
eight consecutive A-A intervals that measure less than two times the total atrial blanking period (AV interval + PVAB) and less than two times the Detect Rate interval. If they are found, the pacemaker extends the PVARP for one beat to expose the blanked atrial flutter that may be occurring. This search occurs if the Blanked Flutter Search parameter is On.
Note: Mode Switch is not recommended for patients with chronic refractory atrial tachyarrhythmias (i.e., atrial tachycardia, atrial fibrillation, atrial flutter).

Non-Competitive Atrial Pacing

Non-Competitive Atrial Pacing (NCAP) is a programmable On or Off feature in the DDDR and DDD modes. NCAP prevents an atrial pace from falling within the atrium’s relative refractory period, thereby preventing a potential atrial tachycardia. A refractory sensed atrial event within the PVARP starts a 300-ms NCAP period during which no atrial pacing may occur. The scheduled atrial pace is delayed until the NCAP period expires, thus preventing pacing within the atrium’s relative refractory period. The Paced AV following NCAP operation may be shortened to a minimum of 30 ms to maintain a stable ventricular rate.
Note: Even when NCAP is programmed Off, NCAP operation is invoked automatically in the DDDR mode for one cycle when PVC Response or PMT Intervention operations occur.
72
Page 73
Special Therapy Options

PMT Intervention

Pacemaker-Mediated Tachycardia (PMT) Intervention is a programmable On or Off feature that provides automatic detection and interruption of pacemaker-defined PMTs. The pacemaker detects a PMT episode as eight consecutive ventricular pace-to-atrial sense events with VA intervals less than 400 ms. In addition, the pacemaker corroborates the episode as a PMT if, on the eighth consecutive pace, the Sensor Rate is at or below the ADL rate. When an episode is detected and corroborated, the pacemaker intervenes by forcing a 400-ms PVARP for one pacing cycle. Extending the PVARP ensures that the next sensed atrial event within 400 ms will be refractory, thus interrupting atrial tracking for one cycle, which may stop the PMT.
Note: A sinus tachycardia could cause a feature intervention, resulting in a single P wave falling in the PVARP and therefore not being tracked by the pacemaker.
After intervention occurs, PMT detection is automatically suspended for 90 seconds to prevent unnecessary intervention for fast intrinsic atrial rates.
Warni ng: Even with the feature turned On, PMTs may require clinical intervention such as pacemaker reprogramming, magnet application, drug therapy, or lead evaluation.
73
Page 74
Special Therapy Options

PVC Response

Premature Ventricular Contraction (PVC) Response is a programmable On or Off feature that prevents tracking of retrograde P waves generated by PVCs in the DDDR, DDD, and VDD modes. PVC Response prevents inhibition of atrial pacing resulting from retrograde P waves generated by PVCs in the DDIR and DDI modes. Any ventricular event-to-ventricular refractory or nonrefractory sense event with no intervening atrial event starts a minimum PVARP of 400 ms. A minimum PVARP of 400 ms causes retrograde P waves to be sensed as refractory events, thereby preventing retrograde P waves from starting an SAV interval which could result in a PMT. Note that an SAV interval is not started in the DDIR and DDI modes.

Ventricular Safety Pacing

Ventricular Safety Pacing (VSP) is a programmable On or Off feature that prevents ventricular asystole due to inappropriate inhibition of ventricular pacing by crosstalk or ventricular oversensing. A sensed ventricular event within 110 ms after an atrial pacing pulse results in a ventricular pacing pulse at 110 ms or the programmed or rate adaptive PAV interval, whichever expires first. If the sensed event was, in fact, a ventricular depolarization, the ventricular pacing pulse at 110 ms falls harmlessly into the absolute refractory period of the ventricle.
Warni ng: VSP should always be programmed On for pacemaker­dependent patients.
74
Page 75
Special Therapy Options

Rate Drop Response

Rate Drop Response (RDR) is a programmable On or Off feature that provides backup pacing for patients who experience symptomatic episodes of heart rate drop. RDR intervenes by pacing both chambers at an elevated rate (programmed Intervention Rate) for a specific duration (programmed Intervention Duration). At the conclusion of the Intervention Duration, the pacing rate gradually returns to the programmed Lower Rate. Two programmable detection methods may be selected, either separately or in tandem:
Low Rate Detect, in which the pacemaker monitors for a programmable number of paces at the Lower Rate (Detection Beats) required to trigger an intervention.
Drop Detect, in which the pacemaker monitors a rate drop which must meet two programmable requirements to trigger an intervention. The programmable Drop Size is the number of beats the rate must fall within the Detection Window (programmable amount of time monitored for a rate drop) in order for an RDR episode to be detected. The programmable Drop Rate is the threshold that the rate must fall below during the Detection Window in order to trigger an intervention.
RDR is programmable in the DDD or DDI mode only.
75
Page 76
Special Therapy Options

Sinus Preference

Sinus Preference, programmable On or Off in the DDDR mode only, searches for or tracks an intrinsic sinus rate that is lower than the sensor-indicated rate. This operation is determined by two programmable parameters:
The Search Interval is the amount of time spent pacing between sinus search operations.
The Sinus Preference Zone is the maximum allowable reduction in rate relative to the sensor-indicated rate.
Two modalities determine how a sinus rate may emerge. During sinus search operation, the pacemaker searches for a lower intrinsic rate within the Sinus Preference Zone. If the sinus rate is present, it is tracked by the ventricle. If it is not, pacing occurs for 8 beats at the maximum allowable reduced rate as defined by the Sinus Preference Zone, then the rate gradually returns to the sensor-indicated rate.
In sinus breakthrough operation, a sinus rate may be tracked below the sensor-indicated rate as long as it remains within the Sinus Preference Zone. If the intrinsic rate falls below the Sinus Preference Zone, the pacemaker paces at the maximum allowable reduced rate for eight paces and gradually returns to the sensor-indicated rate.
76
Page 77
Special Therapy Options

Sleep Function

Sleep Function is a programmable On or Off feature that reduces the paced rhythm during sleep. It replaces the programmed Lower Rate with a slower Sleep Rate during a programmable sleep period. At the programmed Bed Time, the pacemaker gradually lowers the operating lower rate from the programmed Lower Rate to the programmed Sleep Rate over a 30 minute period. The Sleep Rate becomes the operating lower rate until the programmed Wake Time occurs, when the pacemaker gradually raises the operating lower rate from the Sleep Rate to the Lower Rate over a 30 minute period.
Note: For rate responsive modes, the Sleep Rate becomes the minimum rate during the sleep period. However, in the presence of sensor-detected activity, the sensor-indicated rate overrides the Sleep Rate.

Single Chamber Hysteresis

Single Chamber Hysteresis allows tracking of an appropriate intrinsic rhythm below the Lower Rate during extended periods of inactivity such as sleep. When a nonrefractory sensed event occurs, sensed events that follow are allowed to occur below the programmed Lower Rate to the programmed Hysteresis Rate. As long as the intrinsic rate is above the Hysteresis Rate, pacing is inhibited. When the intrinsic rate drops to the Hysteresis Rate, escape pacing results and the programmed Lower Rate returns as the operating lower rate until the next nonrefractory sensed event occurs.
77
Page 78
Page 79
Chapter 4 - Pacemaker Follow-up
Pacemaker Telemetry 80
Other Operations 83
Diagnostics 87
General Recommendations 89
79
Page 80

Pacemaker Telemetry

Pacemaker Telemetry
Pacemaker telemetry is established by placing the programming head over the implant site and moving the head along the axis shown in Figure 6 until the head lights indicate telemetry is possible. Pacemaker telemetry can be used for routine monitoring. A brief description of the types of telemetry is given below.
Axis
Figure 6. Positioning the Programming Head

Parameter Summary

Upon interrogation by the programmer, the pacemaker telemeters the programmed parameter settings along with battery status and device identification.
Note: The parameters may change due to a full or partial electrical reset, elective replacement indication, Rate Profile Optimization, automatic Sensitivity, Capture Management, or Lead Monitor operation. See the appropriate topic for more information.

Battery and Lead Information

The pacemaker provides Battery and Lead Information for both measured and calculated values of the pacing output pulse(s), the lead(s), and the battery. The pacemaker’s Battery and Lead Information is accessed via the programmer.
80
Page 81
Pacemaker Telemetry

Telemetry Markers (Extended Markers)

Telemetry markers depict cardiac events that are useful in evaluating pacemaker operations.
Standard Markers (Marker Channel Telemetry) annotate pacemaker paced, sensed, and refractory sensed events in any mode.
Therapy Trace Markers provide depictions of therapy interventions that alter timing intervals.

Electrogram (EGM)

The Electrogram (EGM) is a real-time intracardiac waveform detected by the atrial or ventricular lead. Atrial and ventricular waveforms can be displayed simultaneously with Dual EGM (simultaneous display of the independent channels) or Summed EGM (the electrical summation of both channels). The waveform(s) is accompanied by standard Marker Channel telemetry annotation.
Warni ng: An EGM of the patient’s intrinsic activity should be obtained with care since the patient is without pacing support when using the programmer’s inhibit function.
Notes:
The simulated pacing pulses on the EGM should not be interpreted as representing the actual timing or amplitude of paced events. Use an ECG to determine such events.
Intracardiac amplitudes measured on the EGM should not be used to assess the adequacy of the programmed Sensitivity setting. Use the sensing test instead.

Transtelephonic Monitor

The Transtelephonic Monitor is a programmable On or Off feature intended for use with remote pacemaker monitoring services. When the Transtelephonic Monitor is programmed On, the Threshold Margin Test is delayed for five seconds upon magnet application to
81
Page 82
Pacemaker Telemetry
enhance communication with transtelephonic equipment. If the pacing polarity is programmed to bipolar, it is temporarily set to unipolar to provide improved ECG artifact detection for the remote monitoring equipment. (The programmed polarity is restored when the magnet is removed.) When Transtelephonic Monitor is programmed Off, the Threshold Margin Test is not delayed upon magnet application, and conventional transtelephonic monitoring can occur.
Warni ng: Programming Transtelephonic Monitor On is contraindicated for patients with implantable defibrillators. When it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.

Extended Telemetry

Extended Telemetry provides frequent patient monitoring in the programmed mode. When Extended Telemetry is programmed On, the pacemaker automatically transmits the programmed telemetry type (i.e., Electrogram, Marker Channel telemetry) when the programmer head or monitoring receiver (with or without a magnet) is positioned over the pacemaker. Extended Telemetry is automatically programmed Off after 48 hours or can be programmed Off via the programmer.
82
Page 83

Other Operations

Other Operations

Magnet Operation

Magnet operation includes an initial capture safety margin test, an automated self test, and magnet mode operation with telemetry:
The Threshold Margin Test (TMT) provides a check for loss of capture. It is performed at 100 ppm with pulse amplitudes reduced by 20% on the third pulse.
Warning: Loss of capture during TMT may indicate that the stimulation threshold safety margin is inadequate. Perform a pacing threshold test and reprogram outputs to establish a 2:1 voltage safety margin, if necessary.
The Fully Automated Self Test (FAST) is a pattern of pacing pulses that indicates either of the following events has occurred:
– Lead Monitor has detected an out-of-range lead, and, if
programmed to Adaptive, has switched polarities from bipolar to unipolar.
– Capture Management has measured unusually high
ventricular thresholds.
Beginning on the cycle following TMT, the pacemaker delivers two pacing pulses at 85 ppm followed by two pulses at 65 ppm and then repeats the same sequence once again.
Magnet Mode forces the pacemaker’s programmed operation to an asynchronous mode (DOO in dual chamber configurations, VOO in the VDD mode, and VOO/AOO in single chamber configurations). Magnet Mode rate is 85 ppm for all of the above modes. However, the rate is 65 ppm if an electrical reset occurs or the elective replacement indicator is set.
83
Page 84
Other Operations
Magnet Operation Using the Transtelephonic Magnet
When a transtelephonic magnet is placed over the pacemaker, a TMT is automatically initiated. A FAST, if programmed On, immediately follows the TMT if any of the events mentioned above that initiate the test have occurred. When the FAST (or TMT when a FAST is not initiated) is completed, pacing is forced to Magnet Mode operation until the magnet is removed.
Magnet Operation Using the Programming Head
The programming head does not initiate a TMT when placed over the pacemaker. The clinician must perform a Magnet Test in order to obtain a TMT. A FAST, if programmed On, will follow the TMT if initiated by any of the events described above. When the FAST (or TMT when a FAST is not initiated) is completed, pacing is forced to Magnet Mode operation until the Magnet Test is terminated or the programming head is removed.
FAST Reporting Using the Remote Assistant
The FAST is available with the patient-activated hand-held Remote Assistant when the Remote Assistant diagnostic is off, and FAST has been programmed On.
With the Remote Assistant, the FAST results indicating a lead or threshold problem are reported by two beeps in a low tone, and a yellow light appears on the underside of the activator. When no problems exist, the FAST results are reported by three beeps, and a green light appears instead.
FAST with the Remote Assistant also reports if the ERI has been set or an electrical reset has occurred (indicated by two beeps and a yellow light from the activator).

Temporary Parameters

Several pacemaker parameters can be programmed temporarily during a patient session. These include: Mode, Rate, Amplitude, Pulse Width, Sensitivity, and AV intervals.
84
Page 85
Other Operations

Electrical Reset

A partial Electrical Reset can occur when the pacemaker is exposed to cold temperatures, electrocautery, internal or external defibrillation. Under these conditions, the pacemaker resumes pacing at the previously programmed mode and preserves many of the primary pacing parameter settings in operation prior to the partial reset. See “Electrical Reset Parameter Settings” on page 103 for a complete list of the preserved settings as well as full Electrical Reset settings.
If the temperature or electrical disturbance is severe enough, the pacemaker can reset to full Electrical Reset settings, in which case the pacemaker reverts to VVI mode at a pacing rate of 65 ppm and restarts Implant Detection. To restore the pacemaker to its previous mode and pacing operations, it must be reset and manually reprogrammed via the programmer.
Note: If after a reset a message appears on the programmer screen or printout indicating the Date/Time memory has been altered, contact your Medtronic representative for further information.
85
Page 86
Other Operations

Elective Replacement Indicator

The Elective Replacement Indicator (ERI) forewarns when the battery is nearing depletion. When the pacemaker detects that the battery voltage has dropped below 2.59 V, the pacemaker sets the ERI status and reverts to VVI operation at a rate of 65 ppm. Pacing
is maintained, in VVI mode, even in pacemakers previously programmed to the AAIR, ADIR, AAI, ADI, AAT, AOOR, or AOO mode.
Caution: When ERI is set and verified, the pacemaker must be
replaced within three months.
High-rate pacing over several years, or with the battery partially depleted, may trigger the ERI prematurely.
Continuous pacing at high rates (i.e., greater than 100 ppm) and high output settings (i.e., pulse widths greater than 1.0 ms and/or amplitudes greater than 5.0 V) may cause the battery to deplete rapidly and set ERI. While operating at ERI, the battery voltage may rise above 2.59 V, allowing the pacemaker to be reset and reprogrammed.
Before and after the use of high-rate pacing, the clinician should interrogate the pacemaker to determine the battery status and ERI condition.

Emergency Pacing

Emergency pacing provides VVI pacing at high output settings in emergency situations for pacemaker-dependent patients. See “Emergency Parameter Settings” on page 98.
86
Page 87

Diagnostics

Rate Limit

In the event of a single component failure, the absolute rate limit for atrial and ventricular rates are held independently to 200 ppm. Rate limit is automatically overridden in temporary single chamber modes for high rate pacing.
Diagnostics
The pacemaker has automatic data summary diagnostics, some of which have an accompanying clinician-selected diagnostic to collect further detailed data. At the completion of Implant Detection, the pacemaker enables diagnostic collection.

Automatic Diagnostics / Clinician-Selected Diagnostics

Heart Rate Histograms. Automatically collect short and long term heart rate and percent-paced data for the atrium and ventricle.
AV Conduction Histogram. Automatically collects short and long term AV conduction sequences.
Search AV Histogram. Automatically collects conduction sequences resulting from the Search AV function.
Sensor Indicated Rate Profile. Automatically collects sensor rate data.
Atrial and Ventricular High Rate Episodes. Automatically collect basic data on atrial and ventricular high rates and, optionally, clinician-selected detailed data of atrial, ventricular, or summed electrograms with a rate trend.
Lead Diagnostics. Chronic Lead Impedance Trend automatically collects long-term chronic lead impedance data. The automatic Lead Monitor Counters present the impedance status of every beat. Chronic lead impedance data is collected in all pacing modes except for triggered pacing modes (AAT
87
Page 88
Diagnostics
and VVT). This data is not collected in non-pacing modes (OAO, OVO, and ODO).
Key Parameter History. Automatically records the status of key parameters pertaining to the last eight programming sessions.
Sensitivity Trend. Automatically collects the minimum and maximum sensitivities used to operate automatic sensitivity.
Capture Management Trend. Automatically collects information about the range of ventricular thresholds and outputs over the life of the device and, optionally, clinician-selected detailed data reporting the results of pacing threshold tests.
Mode Switch Episodes. Automatically collect basic data on mode switch episodes, or optionally, clinician-selected electrogram with a rate trend.
Sinus Preference Episodes. Automatically collects data from searches for the sinus rate and intrinsic sinus breakthrough episodes.
Rate Drop Response Episodes. Automatically collects basic data on rate drops and Rate Drop Response therapy intervention, and, optionally, clinician-selected rate trend information for Rate Drop Response.
Custom Rate Trend. Clinician-selected collection of heart rate and percent-paced trends over a beat-to-beat or 24-hour period.
AV Interval Histogram. Clinician-selected option allows you to program the pacemaker to collect AV interval data for display in the form of a histogram.
Lead Monitor Impedance Trend. Provides trend data on the integrity of the lead.
Sensitivity Detail. Provides a detailed view of Sensing Assurance operation by recording sensitivity values at 1-hour intervals.
1
88
Page 89

General Recommendations

Remote-Assistant Symptoms. Data for symptom diagnosis is collected by the clinician-selected atrial and ventricular beat­to-beat rate trend with optional EGM.
Data collection is initiated by the patient using the Remote Assistant patient activator. The patient will hear a clicking sound from the Remote Assistant when initiating a telemetry link with the pacemaker. If the link is successful, the Remote Assistant responds with three beeps, and a green light appears on the underside of the activator. If communication is unsuccessful, it responds with two beeps in a lower tone, and a yellow light appears instead.
Remote-Assistant Exercise. Records heart rate data for a trend graph during a particular exercise activity.
General Recommendations
During the first few months after receiving a new pacing system the patient may require close monitoring. In the U.S.A., the physician may wish to use the individual state’s or Medicare’s maximum frequency table for dual chamber pacemakers under Guideline I, i.e., every two months from the second month through three years post-implant, and monthly thereafter (Ref. Carriers Manual, Part 3, Claims Process, Transmittal No. 1051, or a later publication, if available). Note that Guideline I applies to those pacing systems (i.e., the pacemaker and the lead used) for which 5 years of clinical data are not yet available.
1
If you selected Sensitivity Detail as one of the setup options for Capture Management Detail, the recording interval is 2 hours rather than 1 hour.
89
Page 90
Page 91

Appendix

Appendix
NBG Codes
Special Notice 93
92
91
Page 92

NBG Codes

NBG Codes
The KDR700 Series pacemakers may be described as DDDR pacemakers using the NBG
1
code (NBG code stands for “The North American Society of Pacing and Electrophysiology [NASPE] and the British Pacing and Electrophysiology Group [BPEG] Generic”). The five-letter code, which supersedes the ICHD Code, describes the operation of implantable pacemakers. For simplicity, this manual uses only the first three or four letters, e.g., DDIR, DVIR, DDD, etc. The following chart describes the first four letters of the NBG code.
CHAMBER PACED
V
=
Ventricle
A
=
Atrium
D
=
Dual Chamber
S
=
Single Chamber
O
=
None
CHAMBER SENSED
V
=
Ventricle
A
=
Atrium
D
=
Dual Chamber
S
=
Single Chamber
O
=
None
DDDR
MODE OF RESPONSE
T
=
Triggered
I
=
Inhibited
D
=
Double (Both)
O
=
None
PROGRAMMABLE/ RATE RESPONSE
P
=
Programmable
M
=
Multiprogrammable
C
=
Communicating
R
=
Rate Responsive
O
=
None
1
Bernstein A., et al., The NASPE/BPEG Pulse Generator Code, Pace, 10 (4), Jul-Aug 1987.
92
Page 93

Special Notice

Special Notice
Medtronic implantable pacemakers and leads are implanted in the extremely hostile environment of the human body. This environment places severe limitations on the design and function of the pacemaker and lead. These limitations unavoidably reduce the potential performance and longevity of the pacemaker and lead despite the exercise of due care in design, component selection, manufacture and testing prior to sale. Reference is hereby made to published data on predictable failure rates of the pacemaker and lead and their implantation which have either been furnished by Medtronic or are otherwise available to physicians. The pacemaker includes an inseparable power source which will ultimately cease to function due to exhaustion or premature failure thereby necessitating removal of the pacemaker. The lead is necessarily very small in diameter and very flexible, which unavoidably increases the likelihood of breakage or breach of its insulation. Other reasons for failure of the pacemaker or lead include, but are not limited to: body rejection phenomena; change in performance characteristics due to component changes or failures; unusual physiological variations in patients; lead displacement; lead fracture; fibrotic tissue formation; elevated thresholds; medical complications, such as myocardial infarction occurring at the site of the electrode; erosion of the pacemaker or lead through body tissue or interference from transmitters, tools, appliances, instruments, equipment or other devices which use electricity or electromagnetic wave transmission. Consequently, no representation or warranty is made that failure or cessation of function of the pacemaker or lead will not occur, that the body will not react adversely to the implantation of the pacemaker or lead, that medical complications will not follow the implantation of the pacemaker or lead or that the pacemaker and lead will, in all cases, restore adequate cardiac function.
93
Page 94
Page 95

Pacemaker Specifications

Pacemaker Specifications
Lead Requirements, Compatibility 96
Radiopaque Identification 97
Emergency Parameter Settings 98
Shipping and Nominal Parameter Settings 99
Electrical Reset Parameter Settings 103
Elective Replacement Indicator 108
Magnet Mode Conditions 108
Longevity Projections 109
Programmable Parameters 116
Nonprogrammable Parameters 125
Temporary Parameters 126
Telemetry Markers 127
Electrograms 127
Automatic Diagnostics 127
Clinician-Selected Diagnostics 128
Patient Information 128
Battery and Lead Telemetered Information 129
Battery Parameters 130
Mechanical Dimensions 131
95
Page 96

Lead Requirements, Compatibility

Lead Requirements, Compatibility
The chart below lists the available models by polarity type with the required lead for proper operation. Refer to “Implant Documentation” on page 42.
Model Polarity Primary Leads Connector
K
DR701 DR721
K
DR731
K
DR703
K
DR733
K
DR706 Unipolar Unipolar 5 or 6 mm Setscrew for
K
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992) whereby pulse generators and leads so designated are assured of meeting the electrical and mechanical parameters specified in the IS-1 International Standard.
Bipolar/Unipolar IS-1
Bipolar/Unipolar Low-profile 3.2 mm
a
BI
bipolar or IS-1a BI
Setscrew for tip electrode
Setscrews for tip and ring electrodes
tip electrode
96
Page 97

Radiopaque Identification

Radiopaque Identification
Listed below are the pacemaker’s radiopaque identifications, which are used to identify the pacemaker with fluoroscopy. An X-ray showing the location of the radiopaque for Model K Figure 7.
Models Radiopaque Models Radiopaque
DR701 PGU KDR721 PGR
K
K
DR703 PGY KDR731 PHB
K
DR706 PGW KDR733 PHD
PGU
Figure 7. Model KDR701 X-Ray
DR701 is given in
97
Page 98

Emergency Parameter Settings

Emergency Parameter Settings
Listed below in Table 4 are the pacemaker’s emergency parameter settings.
Table 4. Emergency Settings
Mode VVI
Pacing Rate 70 ppm
Ventricular
Amplitude 7.5 V
Pulse Width 1.5 ms
Sensitivity 2.8 mV
Pacing Polarity Unipolar
Sensing Polarity Unipolar
Lead Monitor Monitor Only
Capture Management Off
Ventricular Refractory Period 330 ms
Single Chamber Hysteresis Off
98
Page 99

Shipping and Nominal Parameter Settings

Shipping and Nominal Parameter Settings
Listed below in Table 5 are the pacemaker’s shipping and nominal parameter settings.
Note: “Adaptive” indicates that the parameter value is adapted during operation. “Unchanged” indicates that the parameter setting is unchanged by nominal programming.
Table 5. Shipping and Nominal Settings
Parameter Shipping Medtronic
a
Mode DDD (DDDR
Mode Switch On On
Detect Rate 175 bpm 175 bpm
Detect Duration No Delay No Delay
Blanked Flutter Search On On
Lower Rate 60 ppm 60 ppm
Upper Tracking Rate 120 ppm 120 ppm
Upper Sensor Rate 120 ppm 120 ppm
ADL Rate 95 ppm 95 ppm
Rate Profile Optimization On
ADL Response 3-Mod. Active 3-Mod. Active
Exertion Response 3-Mod. Frequent 3-Mod. Frequent
Activity Threshold Med. Low Unchanged
Activity Acceleration 30 sec Unchanged
Activity Deceleration Exercise Unchanged
a
) DDDR
Nominal
Unchanged
99
Page 100
Shipping and Nominal Parameter Settings
Table 5. Shipping and Nominal Settings (Continued)
Parameter Shipping Medtronic
Atrial Lead
Amplitude 3.5 V 3.5 V
Pulse Width 0.4 ms 0.4 ms
Sensitivity 0.5 mV (Adaptive
Sensing Assurance On
Pace Polarity Configure
Sense Polarity Configure
Lead Monitor Configure
a
a
Unipolar
Unipolar
c
a
c
d
Monitor Only
or
or
or
c
Notify if < 200 ohms Unchanged
Notify if > 4000 ohms 4000 ohms
Monitor Sensitivity 8 8
Nominal
a
) 0.5 mVb (Adaptive)
On
Unchanged
Unchanged
Unchanged
Ventricular Lead
a
Amplitude 3.5 V (Adaptive
Pulse Width 0.4 ms (Adaptive
Sensitivity 2.8 mV (Adaptive
Sensing Assurance On
a
Pace Polarity Configure
Unipolar
Sense Polarity Configure
Unipolar
Lead Monitor Configure
Monitor Only
)3.5V
a
)0.4msb (Adaptive)
a
)2.8mVb (Adaptive)
a
or
c
a
or
c
d
or
c
Notify if < 200 ohms Unchanged
Notify if > 4000 ohms 4000 ohms
Monitor Sensitivity 8 8
100
b
(Adaptive)
On
Unchanged
Unchanged
Unchanged
Loading...