Medtronic KDR651 Technical Manual

KAPPA® DR650 SERIES
198798002A_view.pdf
Pacemaker Models KDR651/653/656
Product Information Manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
Kappa®650 Series Pacemaker Product Information Manual
Models KDR651, KDR653, and KDR656
3
This Product Information Manual is primarily intended as an implantation manual. For programming information see the Pacemaker Reference Guide that accompanies the programmer software. It is primarily intended as a follow-up manual, and contains further information on therapeutic and diagnostic features, troubleshooting information, follow-up precautionary information, and complete reference information.
The following are trademarks of Medtronic:
Capture Management, FAST, Implant Detection, Kappa, Marker Channel, Medtronic, Medtronic Kappa, Medtronic Vision, Rate Profile Optimization, Sensing Assurance, and Vision.
4

Table of Contents

Chapter 1 - Prescribing the Pacemaker 7
Device Description 8 Indications and Usage 9 Contraindications 10 Warnings and Precautions 10 Co-implantation with an Implantable Defibrillator 20 Adverse Events 22 Clinical Studies 25
Chapter 2 - Implanting the Pacemaker 29
Implantation Procedures 30 Implant Documentation 39 Parameter Programming at Implant 40 Medical Therapy Interactions 45 Assistance 48
Chapter 3 - Description 49
Pacing Mode Operations 50 Rate Responsive Pacing 52 Timing Operations 54 Pacing and Sensing Operations 58 Special Therapy Options 65
Chapter 4 - Pacemaker Follow-up 71
Pacemaker Telemetr y 72 Other Operations 75
5
Diagnostics 79 General Recommendations 80
Pacemaker Specifications 81
Lead Requirements, Compatibility 82 Radiopaque Identification 83 Emergency Parameter Settings 84 Shipping and Nominal Parameter Settings 85 Electrical Reset Parameter Settings 89 Elective Replacement Indicator 93 Magnet Mode Conditions 93 Longevity Projections 94 Programmable Parameters 96 Nonprogrammable Parameters 103 Temporary Parameters 104 Tel e m e tr y M a r k e rs 10 5 Electrograms 105 Automatic Diagnostics 105 Clinician-Selected Diagnostics 106 Patient Information 106 Battery and Lead Telemetered Information 107 Battery Parameters 108 Mechanical Dimensions 108
Appendix 109
NBG Codes 110 Special Notice 111
Index 113
6
Prescribing the Pacemaker
Chapter 1 - Prescribing the Pacemaker
Device Description 8
Indications and Usage 9
Contraindications 10
Warnings and Precautions 10
Co-implantation with an Implantable Defibrillator 20
Adverse Events 22
Clinical Studies 25
7
Prescribing the Pacemaker

Device Description

Medtronic Kappa 650 Series pacemakers (KDR650 Series) are dual chamber, multiprogrammable, rate responsive, implantable pacemakers, intended for a variety of bradycardia pacing applications. Rate response is controlled through an activity-based sensor. The following models are available:
Model Polarity Primary Leads
K
DR651 Bipolar/Unipolar IS-1
KDR653 Bipolar/Unipolar Low-profile 3.2 mm bipolar
K
DR656 Unipolar Unipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992).
DR650 Series pacemakers are programmed using the Medtronic
K Vision software Model 9953 and a Medtronic Model 9790 programmer. For programming instructions, refer to the Pacemaker Programming Guide that accompanies Medtronic Kappa 700/650 Series software.
a
or IS-1
BI
a
BI
8
Prescribing the Pacemaker

Indications and Usage

KDR650 Series pacemakers are indicated for the following:
Rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity.
Accepted patient conditions warranting chronic cardiac pacing which include:
– Symptomatic paroxysmal or permanent second or
third-degree AV block. – Symptomatic bilateral bundle branch block. – Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders. – Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
KDR650 Series pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include:
Various degrees of AV block to maintain the atrial contribution to cardiac output.
VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm.
9
Prescribing the Pacemaker

Contraindications

KDR650 Series pacemakers are contraindicated for the following applications:
Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias.
Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms.
Unipolar pacing for patients with an implanted cardioverter­defibrillator (ICD) because it may cause unwanted delivery or inhibition of ICD therapy. See “Co-implantation with an Implantable Defibrillator” on page 20.

Warnings and Precautions

Rate responsive modes. Do not use rate responsive modes in those patients who cannot tolerate pacing rates above the programmed Lower Rate.
Single chamber atrial modes. Do not use single chamber atrial modes in patients with impaired AV nodal conduction because ventricular capture cannot be assured.

Pacemaker Dependent Patients

Diagnostic modes. Never program diagnostic modes (ODO, OVO, and OAO) for pacemaker-dependent patients. For such patients, use the programmer’s inhibit function for brief interruption of outputs.
Electrogram (EGM) of the patient’s intrinsic activity should be obtained with care since the patient is without pacing support when using the programmer’s inhibit function.
Polarity override. Overriding the bipolar verification prompt with bipolar polarity when a unipolar lead is connected results
10
Prescribing the Pacemaker
in no pacing output. See “Manually Programming Polarity” on page 61 for further information.
A false bipolar pathway on a unipolar lead, a possible occurrence with bipolar 3.2 mm connector pacemakers, may result in a loss of output. See the warning in “Automatic Polarity Configuration” on page 58 for further information.
Loss of capture during threshold margin test (TMT) at a 20% reduction in amplitude indicates that the stimulation safety margin is inadequate. Consider increasing the pacing amplitude and/or pulse width. See “Magnet Operation” on page 75 for further information on the Threshold Margin Test.
Ventricular safety pacing should always be used for pacemaker-dependent patients. See “Ventricular Safety Pacing” on page 68 for further information.
Capture Management will not program ventricular outputs above 5.0 V or 1.0 ms. If the patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually. See “Capture Management” on page 64 for further information

Medical Therapy

Therapeutic Diathermy can cause fibrillation, burning of the myocardium, and irreversible damage to the pulse generator due to induced currents.
Magnetic resonance imaging of pacemaker patients has resulted in significant adverse effects. See “Magnetic Resonance Imaging (MRI)” on page 16 for further information.

Storage and Resterilization

Medtronic pacemakers are intended for single use only. Do not resterilize and re-implant explanted pacemakers.
The chart below gives recommendations on handling and storing the package. Medtronic has sterilized the pacemaker with ethylene oxide
11
Prescribing the Pacemaker
prior to shipment. Resterilizing the pacemaker is necessary if the seal on the sterile package is broken. Resterilization does not affect the “Use Before” date.
Handling and Storage: Acceptable Unacceptable
Store and transport within Environmental Temperature limits:
°F (- 18°C) to 131°F (55°C).
0 Note: A full or partial electrical reset
condition may occur at temperatures below 0°F (- 18°C). See “Electrical Reset Parameter Settings” on page 89.
Resterilization: Acceptable Unacceptable
Resterilize if the sterile package seal is broken. Place the device in an ethylene oxide permeable package and resterilize with ethylene oxide. Allow the device to aerate ethylene oxide residues. See sterilizer instructions for details. Use an acceptable method for determining sterility, such as biological indicators.
Do not implant the device if it has been dropped on a hard surface from a height of 12 inches (30 cm) or more.
Do not resterilize the device or the torque wrench using:
an autoclave,
gamma radiation,
organic cleaning agents, e.g., alcohol, acetone, etc.,
ultra-sonic cleaners.
Do not exceed 140 17 psi (103 kPa) when sterilizing.
Do not resterilize the device more than twice.

Lead Evaluation and Lead Connection

Connector compatibility. Do not use any lead with this pacemaker without first verifying connector compatibility. Using incompatible leads can damage the connector or result in a leaking or intermittent connection.
°F (60°C) or
12
Prescribing the Pacemaker
Pacing and sensing safety margins. Consider lead maturation when choosing pacing amplitudes, pacing pulse widths, and sensing levels. See “Manual Programming” on page 42.
Hex wrench. Do not use a hex wrench with a blue handle or a right- angled hex wrench. These wrenches have torque capabilities greater than is designed for the lead connector. See “Connection Procedure” on page 33 for lead connection instructions.

Programming and Pacemaker Operation

Epicardial leads. Ventricular epicardial leads have not been determined appropriate for use with the Capture Management feature. Therefore, Capture Management should be programmed Off if epicardial leads are implanted with
DR650 Series pacemakers.
K
Shipping values. Do not use shipping values for pacing amplitude and sensitivity without verifying that they provide adequate safety margins for the patient.
Constant current devices. To test the performance of the lead, Medtronic recommends using a constant voltage device such as the Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA). Medtronic does not recommend using a constant current device such as the Medtronic Model 5880A or 5375 External Pacemaker because the K pacemakers have constant voltage output circuits.
Crosstalk occurs in dual chamber systems when atrial pacing output pulses are sensed by the ventricular lead. Crosstalk results in self-inhibition and is more likely to occur at high sensor-driven pacing rates, high atrial amplitudes, and wide atrial pulse widths. To prevent self-inhibition caused by crosstalk, program Ventricular Safety Pacing (VSP) On or lengthen the Ventricular Blanking period.
DR650 Series
13
Prescribing the Pacemaker
Elective Replacement Indicator (ERI). When ERI is set, the pacemaker must be replaced within three months. See “Elective Replacement Indicator” on page 77 for more information.
Full electrical reset is indicated by VVI pacing at a rate of 65 ppm without the elective replacement indicator set. See “Electrical Reset” on page 77 for more information.
Slow retrograde conduction, especially with conduction time greater than 400 ms, may induce pacemaker-mediated tachycardia (PMT).
PMT intervention. Even with the feature turned On, PMTs may still require clinical intervention such as pacemaker reprogramming, magnet application, drug therapy, or lead evaluation. See “PMT Intervention” on page 67 for further information.
Lead Monitor. If the Lead Monitor detects out-of-range lead impedance, investigate lead integrity more thoroughly.
Rate Increases
Twiddler’s syndrome, i.e., patient manipulation of the device after implant, may cause the pacing rate to increase temporarily if the pacemaker is programmed to a rate responsive mode.
Muscle stimulation, e.g., due to unipolar pacing, may result in pacing at rates up to the Upper Sensor Rate in rate responsive modes.
Unipolar Sensing
Continuous myopotentials cause reversion to asynchronous operation when sensed in the refractory period. Sensing of myopotentials is more likely when atrial sensitivity settings of
0.5 through 1.0 mV and ventricular sensitivity settings of
1.0 and 1.4 mV are programmed.
14
Prescribing the Pacemaker

Environmental and Medical Therapy Hazards

Patients should be directed to exercise reasonable caution in avoidance of devices which generate a strong electric or magnetic field. If the pacemaker inhibits or reverts to asynchronous operation at the programmed pacing rate or at the magnet rate while in the presence of electromagnetic interference (EMI), moving away from the source or turning it off will allow the pacemaker to return to its normal mode of operation.
Hospital and Medical Environments
Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or may cause asynchronous or inhibited pacemaker operation. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pacemaker and leads as possible. See “Electrosurgical Cautery” on page 47 for more information.
External defibrillation may damage the pacemaker or may result in temporary and/or permanent myocardial damage at the electrode-tissue interface as well as temporary or permanent elevated pacing thresholds. Attempt to minimize current flowing through the pacemaker and lead system by following these precautions when using external defibrillation on a pacemaker patient:
– Position defibrillation paddles as far from the pacemaker as
possible (minimum of 5 inches [13 cm]). Attempt to minimize current flowing through the pacemaker and leads by positioning the defibrillation paddles perpendicular to the implanted pacemaker/lead system.
– Use the lowest clinically appropriate energy output (watt
seconds).
– Confirm pacemaker function following any defibrillation.
High energy radiation sources such as cobalt 60 or gamma radiation should not be directed at the pacemaker. If a patient requires radiation therapy in the vicinity of the pacemaker,
15
Prescribing the Pacemaker
place lead shielding over the device to prevent radiation damage and confirm its function after treatment.
Lithotripsy may permanently damage the pacemaker if the device is at the focal point of the lithotripsy beam. If lithotripsy must be used, program the pacemaker to a single chamber nonrate responsive mode (VVI/AAI or VOO/AOO) prior to treatment; and keep the pacemaker at least 1 to 2 inches (2.5 to 5 cm) away from the focal point of the lithotripsy beam.
Magnetic resonance imaging (MRI). Pacemaker patients subjected to MRI should be closely monitored and programmed parameters should be verified upon cessation of MRI. MRI of pacemaker patients should be carefully weighed against the potential adverse affects. Clinicians should carefully weigh the decision to use MRI with pacemaker patients. Limited studies of the effects of MRI on pacemakers have shown that:
– Magnetic and radio frequency (RF) fields produced by MRI
may adversely affect the operation of the pacemaker and may inhibit pacing output.
– Magnetic fields may activate magnet mode operation and
cause asynchronous pacing.
– Reported
1
effects of MRI on pacing include increased
ventricular pacing beyond the rate limit.
1
Holmes, Hayes, Gray, et al. The effects of magnetic resonance imaging on implantable pulse generators. PA C E. 1986; 9 (3): 360-370.
16
Prescribing the Pacemaker
Radiofrequency ablation procedure in a patient with a
DR650 Series pacemaker may cause any of the following:
K – Asynchronous pacing above or below the
programmed rate. – Reversion to an asynchronous operation. – Pacemaker electrical reset. – Premature triggering of the elective replacement indicator.
RF ablation risks may be minimized by:
1. Programming a non-rate responsive, asynchronous pacing mode prior to the RF ablation procedure.
2. Avoiding direct contact between the ablation catheter and the implanted lead or pacemaker.
3. Positioning the ground plate so that the current pathway does not pass through or near the pacemaker system, i.e., place the ground plate under the patient’s buttocks or legs.
4. Having a Medtronic programmer available for temporary pacing.
5. Having defibrillation equipment available.
Home and Occupational Environments
High voltage power transmission lines may generate enough EMI to interfere with pacemaker operation if approached too closely.
Communication equipment such as microwave transmitters, linear power amplifiers, or high-power amateur transmitters may generate enough EMI to interfere with pacemaker operation if approached too closely.
Commercial electrical equipment such as arc welders, induction furnaces, or resistance welders may generate enough EMI to interfere with pacemaker operation if approached too closely.
17
Prescribing the Pacemaker
Home appliances which are in good working order and properly grounded do not usually produce enough EMI to interfere with pacemaker operation. There are reports of pacemaker disturbances caused by electric hand tools or electric razors used directly over the pacemaker implant site.
Electronic article surveillance (EAS) equipment such as retail theft prevention systems may interact with pacemakers. Patients should be advised to walk directly through and not to remain near an EAS system longer than is necessary.
Cellular Phones
Note: Testing was not performed on the K
to similarity in design and function to the K data collected for the K
DR700 Series models therefore supports the
DR650 Series models due
DR700 Series models. The
safety and efficacy claims for the KDR650 Series models and is included here for reference purposes.
DR700 Series pacemakers have been tested to the frequency
K ranges used by the cellular phones included in Table 1. Based on this testing, these pacemakers should not be affected by the normal operation of such cellular phones.
These pacemakers contain a filter that allows usage, without interaction, of all cellular phones having one of the transmission technologies listed in Table 1. These transmission technologies represent most of the cellular phones in use worldwide. Patients can contact their local cellular phone service provider to confirm that the provider uses one of these technologies.
18
Prescribing the Pacemaker
Table 1. Cellular Phone Transmission Technologies
Transmission Technology Frequency Range
Analog
FM (Frequency Modulation) 824 - 849 MHz
Digital TDMA
North American Standards
TDMA - 11 Hz 806 - 821 MHz
NADC
PCS
International Standards
GSM
DCS
a
b
(TDMA - 50 Hz)
c
1900
d
e
1800
824 - 849 MHz
1850 - 1910 MHz
880 - 915 MHz
1710 - 1785 MHz
Digital CDMA
CDMA - DS
a
Time Division Multiple Access
b
North American Digital Cellular
c
Personal Communication System
d
Global System for Mobile Communications
e
Digital Cellular System
f
Code Division Multiple Access - Direct Sequence
f
824 - 849 MHz
19
Prescribing the Pacemaker

Co-implantation with an Implantable Defibrillator

An implantable defibrillator may be implanted concomitantly with a bipolar pacemaker.
– The use of unipolar-only Model KDR656 and the
DR650 Series bipolar models implanted with unipolar
K leads is contraindicated for patients having an implantable defibrillator.
– Follow implant protocol and precautions for pacemaker
and defibrillator lead placement. Ensure the pacemaker is configured to be compatible with the defibrillator.

Programming Considerations

Note the following programming considerations for patients who have an implantable defibrillator.
Only bipolar pacing should be used with these patients. In some cases, pacing in the unipolar configuration may cause the defibrillator either to deliver inappropriate therapy or to withhold appropriate therapy.
Polarity is automatically configured during Implant Detection (see “Automatic Polarity Configuration” on page 58 of the Product Information Manual). If lead integrity is suspect, confirmation of the automatically programmed polarities should be made after completion of Implant Detection in order to assure that bipolar polarities have been programmed appropriately.
The implantable cardiac defibrillator (ICD) should be turned off during pacemaker implantation procedures until lead polarities have been configured and confirmed. This is to prevent possible back-up unipolar paces from triggering the ICD.
Lead Monitor should not be programmed to Adaptive. When a prevalence of out-of-range lead impedance paces is detected,
20
Prescribing the Pacemaker
the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Transtelephonic Monitor should be programmed to Off. If it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Although these pacemakers are designed to be compatible with implantable defibrillators, the potential does exist for a defibrillation pulse to reset them.
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection. If lead integrity is suspect, confirmation of bipolar polarity should be made after completion of Implant Detection.
21
Prescribing the Pacemaker

Adverse Events

The Medtronic Kappa 700 Series devices were evaluated in a multicenter prospective study (43 investigational centers, 15 centers in the US) of the adaptive features and rate response of the device. Clinical study of the Medtronic Kappa 700 Series of pacemakers included 288 devices implanted in 285 patients worldwide.
Note: Clinical studies were not performed on the K series models due to similarity in design and function to the K Series models. The clinical data collected for the K models therefore supports the safety and efficacy claims for the
DR650 Series models and is included here for reference purposes.
K
There were a total of six deaths in the KDR700 Series study; all were reviewed and judged to be non-device related by a clinical events committee comprising clinical investigators and Medtronic clinical evaluation managers. Two were attributed to ventricular arrhythmia, one to respiratory failure, the fourth to respiratory insufficiency due to chronic obstructive pulmonary disease, the fifth to a mesotelioma, and the sixth to multi-system organ failure.
Eight devices were explanted: three due to pocket infection, one due to infection of the electrode, one from lead/connector mismatch, one patient had a psychosomatic disorder, one patient required the implant of a dual chamber ICD, and one patient continued with vasovagal symptoms and the therapy did not provide sufficient benefit.
DR650 Series
DR700
DR700 Series

Observed Adverse Events

A total of 355 adverse events were reported. The device-related events (182 events) are listed in descending order of frequency in Ta bl e 2 .
22
Prescribing the Pacemaker
Tab le 2. Adverse Events Reported in Four or More Patients- Complicationsa
(Comps) and Observations
All patients implanted (n=288 devices in 285 patients, 133 device years)
c
% of
Patients
with Comps (n=285)
Event Total
Number
of
Events
(Patients)
Any adverse
355 (168) 17.2% 0.45 52% 2.22
b
(Obs)
Comps
per
Device-
Year
(n=133)
% of
Patients
with Obs
(n=285)
Obs per
Device-
Year
(n=133)
events
Any device-related
182 (118) 10.9% 0.31 34% 1.06
events
Pain at pocket site 32 (31) 10.9% 0.24
Other 23 (21) 1.1% 0.02 6.3% 0.15
Inappropriate
11 (11) 3.9% 0.08
programming
Atrial lead
11 (10) 3.6% 0.08
dislodgment
Programmer/ software anomaly
11 (8) 2.8% 0.08
d
Pocket infection 7 (6) 0.7% 0.02 1.4% 0.03
Intermittent
6 (6) 0.7% 0.02 1.4% 0.03
undersensing
Palpitations 6 (6) 2.1% 0.05
Pocket hematoma 6 (6) 0.4% 0.01 1.8% 0.04
Ventricular lead
6 (6) 2.1% 0.05
dislodgment
Elevated pacing
4 (4) 0.7% 0.02 0.7% 0.02
thresholds
Syncope 4 (4) 1.4% 0.03
23
Prescribing the Pacemaker
a
Complications included those adverse events which required invasive measures to correct (e.g., surgical intervention), and were related to the presence of the pacing system or procedure.
b
Observations included those adverse events which did not require invasive measures to resolve, and were related to the presence of the pacing system or procedure.
c
Where present, a number in parentheses indicates the number of patients with the event.
d
Programmer software anomalies observed: screen lock-ups while saving data to diskette (8); problems printing reports outside of a patient session (2); and an incorrect parameter setting on a printout (1), which occurred in an earlier version of the software—software changes were made to eliminate reoccurrence.
The following other adverse events were reported, but occurred in three or fewer patients: angina pectoris; atrial flutter (paroxysmal)/ atrial fibrillation; bipolar short circuit pathway; chest pain; chest pain (non-specific); dizziness; dyspnea/shortness of breath; exit block; failure to capture/loss of capture; false negative capture detection; far field R-wave sensing; fatigue/tiredness; hypotension; inadequate lead/pacemaker connection; infection of electrode; lead/connector mismatch; lead insertion route problem; lead insulation failure; migration of pulse generator; myopotential interference; other oversensing; pacemaker mediated tachycardia; pacemaker syndrome; pectoral muscle stimulation; penetration of myocardium by lead; phrenic nerve/diaphragm muscle stimulation; pneumothorax; swelling pocket site; tachycardia (atrial); thrombus formation at lead; ventricular ectopy; ventricular tachycardia (non-sustained); ventricular tachycardia (sustained).
The following adverse events were deemed not device related (173 events were reported): angina pectoris; atrial fibrillation; atrial flutter (paroxysmal); atrial flutter (persistent); atrial tachycardia; chest pain; chest pain (non-specific); congestive heart failure; dizziness; dyspnea/shortness of breath; fatigue/tiredness; hypertension; hypotension; insufficient cardiac output; myocardial infarction (acute); palpitations; syncope; ventricular ectopy; ventricular fibrillation; ventricular tachycardia (non-sustained); ventricular tachycardia (sustained).
24
Prescribing the Pacemaker

Potential Adverse Events

Adverse events (in alphabetical order), including those reported in Table 2, associated with pacing systems include:
Cardiac perforation
Cardiac tamponade
Death
Erosion through the skin
Hematoma/seroma
Infection
Myopotential sensing
Nerve and muscle stimulation
Rejection phenomena (local tissue reaction, fibrotic tissue formation, pulse generator migration)
Threshold elevation
Transvenous lead-related thrombosis

Clinical Studies

The Medtronic Kappa 700 Series devices were evaluated in a multicenter prospective study (43 investigational centers, 15 centers in the US) of the adaptive features and rate response of the device.
Note: Clinical studies were not performed on the K models due to similarity in design and function to the K models. The clinical data collected for the K therefore supports the safety and efficacy claims for the K series models and is included here for reference purposes.
DR650 series
DR700 series
DR700 series models
DR650
25
Prescribing the Pacemaker

Methods

This study compared the following features of the Medtronic Kappa 700 Series pacemakers to historical controls:
Rate Response
Capture Management
Automatic Polarity Configuration
Sensing Assurance
Patient data were collected at implant, pre-discharge, two weeks, one month, two and/or three months, and six months post implant. Patients were evaluated utilizing a modified version of the Minnesota Pacemaker Response Exercise Protocol (MPREP
1
) at their one month visit. Evaluation of rate response performance for the Medtronic Kappa 700 Series pacemaker was conducted using the Metabolic Chronotropic model described by Wilkoff as applied by
2
. Automatic polarity configuration data were collected at implant.
Kay Sensing Assurance and Capture Management data were collected at each follow-up.

Description of Patients

Patients enrolled in the study represented a general dual chamber pacing population.

Results of the Study

Table 3 summarizes the results of the clinical study. The incidence of complications was found to be similar to that experienced by similar devices. The performance of the automatic polarity configuration, Capture Management, Sensing Assurance, and rate response features were found to meet study objectives.
1
Benditt, David G. M, Editor, Rate Adaptive Pacing, Blackwell Scientific Publications, Boston. 1993: 63-65.
2
Kay, Neal G., “Quantitation of Chronotropic Response: Comparison of Methods for Rate-Modulating Permanent Pacemakers”, JACC 20(7):1533-41, Dec 1992.
26
Prescribing the Pacemaker
The slope of the exercise rate response (1.0 target slope) was less than 0.65 for 26 of 87 (30%) of patients.
All patients implanted (n=288 devices in 285 patients, 133 device years)
Primary Objectives
Automatic Polarity Configuration (n with loss of output / N leads)
Total Leads 0% (0/546) [0%, 0.55%] ≤5%
Unipolar 0% (0/107) [0%, 2.8%] ≤5%
Bipolar 0% (0/439) [0%, 0.7%] ≤5%
Sensing Assurance
(n with loss of sensing or oversensing / N device years)
Atrial 13.5% (18/133) [8.8%, 20.5%] 35.7%
Ventricular 0.8% (1/133) [0.2%, 4.1%] 9.2%
Capture Management (n with all causes loss of capture / N device years)
Loss of capture 5.3% (7/133) [2.6%, 10.5%] ≤10.7%
Slope of MPREP rate response at 1 month (n=87 patients)
Mean 0.81 [0.76, 0.86] [0.65, 1.35]
Table 3. Effectiveness Analysis
Percent of
events
% (n/N)
95%
Confidence
interval
Criteria:
Upper
95% CI
27
Prescribing the Pacemaker
The Medtronic Kappa 700 Series pacemaker’s Rate Profile Optimization (RPO) governs sensor indicated rate (SIR) output. This function is identical in the K
DR650 Series pacemakers. Figure 1
shows the SIR vs. the Wilkoff predicted heart rate achieved using the RPO feature during the MPREP tests at 1 month.
All patients reaching Anaerobic Threshold, N=87
Expected (Wilkoff) rate, mean and 95% CI
100
90
80
70
CI
60
50
40
30
SIR (normalized)
20
10
0
0 204060
%
5
9
r
pe
p
U
n
a
e
M
Lower 95% CI
E
N
(
R
I
S
W
(
d
e
t
c
e
p
x
)
7
8
=
80
MPREP Workload (normalized)
Figure 1. Sensor Indicated Rate (SIR) vs. Expected Rate at One Month
R
I
S
)
f
f
o
k
l
i
100
28
Implanting the Pacemaker
Chapter 2 - Implanting the Pacemaker
Implantation Procedures 30
Implant Documentation 39
Parameter Programming at Implant 40
Medical Therapy Interactions 45
Assistance 48
29
Implanting the Pacemaker

Implantation Procedures

Testing Leads and Pacemaker

Equipment to Test the Pacemaker and Lead
To test the performance of the lead, Medtronic recommends using a constant voltage device such as the Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA).
Caution: Do not use constant current devices (such as the Medtronic external pacemaker Models 5880A, 5375, 5348, or 5346) to test lead performance. They may damage the pacemaker’s constant voltage output circuits.
For further procedures on determining thresholds and analyzing pacemaker operation, consult the PSA technical manual.
Note: Wait at least 15 minutes after implanting screw-in leads (nonsteroid) before measuring final stimulation thresholds and intracardiac sensing potentials.
Determining Stimulation Threshold
While testing the lead system prior to implant, Medtronic recommends taking and verifying threshold measurements in both the unipolar and bipolar polarities for all of the pacemaker models. Stimulation thresholds less than 1.0 V at a 0.5 ms pulse width are recommended for acute ventricular leads. Atrial leads may have slightly higher stimulation thresholds.
Test for the following configurations:
Lead tip to case (for unipolar pacing and sensing).
Lead tip to lead ring (for bipolar pacing and sensing).
If the stimulation threshold of a mature chronic lead exceeds 2.5 V at a pulse width of 0.5 ms, consider replacing the lead.
Note: Low-profile 3.2 mm bipolar and IS-1 BI leads can be connected to the pacemaker in only one way, i.e., distal (tip)
30
Loading...
+ 90 hidden pages