Medtronic KDR651 Technical Manual

Page 1
KAPPA® DR650 SERIES
198798002A_view.pdf
Pacemaker Models KDR651/653/656
Product Information Manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
Page 2
Page 3
Kappa®650 Series Pacemaker Product Information Manual
Models KDR651, KDR653, and KDR656
3
Page 4
This Product Information Manual is primarily intended as an implantation manual. For programming information see the Pacemaker Reference Guide that accompanies the programmer software. It is primarily intended as a follow-up manual, and contains further information on therapeutic and diagnostic features, troubleshooting information, follow-up precautionary information, and complete reference information.
The following are trademarks of Medtronic:
Capture Management, FAST, Implant Detection, Kappa, Marker Channel, Medtronic, Medtronic Kappa, Medtronic Vision, Rate Profile Optimization, Sensing Assurance, and Vision.
4
Page 5

Table of Contents

Chapter 1 - Prescribing the Pacemaker 7
Device Description 8 Indications and Usage 9 Contraindications 10 Warnings and Precautions 10 Co-implantation with an Implantable Defibrillator 20 Adverse Events 22 Clinical Studies 25
Chapter 2 - Implanting the Pacemaker 29
Implantation Procedures 30 Implant Documentation 39 Parameter Programming at Implant 40 Medical Therapy Interactions 45 Assistance 48
Chapter 3 - Description 49
Pacing Mode Operations 50 Rate Responsive Pacing 52 Timing Operations 54 Pacing and Sensing Operations 58 Special Therapy Options 65
Chapter 4 - Pacemaker Follow-up 71
Pacemaker Telemetr y 72 Other Operations 75
5
Page 6
Diagnostics 79 General Recommendations 80
Pacemaker Specifications 81
Lead Requirements, Compatibility 82 Radiopaque Identification 83 Emergency Parameter Settings 84 Shipping and Nominal Parameter Settings 85 Electrical Reset Parameter Settings 89 Elective Replacement Indicator 93 Magnet Mode Conditions 93 Longevity Projections 94 Programmable Parameters 96 Nonprogrammable Parameters 103 Temporary Parameters 104 Tel e m e tr y M a r k e rs 10 5 Electrograms 105 Automatic Diagnostics 105 Clinician-Selected Diagnostics 106 Patient Information 106 Battery and Lead Telemetered Information 107 Battery Parameters 108 Mechanical Dimensions 108
Appendix 109
NBG Codes 110 Special Notice 111
Index 113
6
Page 7
Prescribing the Pacemaker
Chapter 1 - Prescribing the Pacemaker
Device Description 8
Indications and Usage 9
Contraindications 10
Warnings and Precautions 10
Co-implantation with an Implantable Defibrillator 20
Adverse Events 22
Clinical Studies 25
7
Page 8
Prescribing the Pacemaker

Device Description

Medtronic Kappa 650 Series pacemakers (KDR650 Series) are dual chamber, multiprogrammable, rate responsive, implantable pacemakers, intended for a variety of bradycardia pacing applications. Rate response is controlled through an activity-based sensor. The following models are available:
Model Polarity Primary Leads
K
DR651 Bipolar/Unipolar IS-1
KDR653 Bipolar/Unipolar Low-profile 3.2 mm bipolar
K
DR656 Unipolar Unipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992).
DR650 Series pacemakers are programmed using the Medtronic
K Vision software Model 9953 and a Medtronic Model 9790 programmer. For programming instructions, refer to the Pacemaker Programming Guide that accompanies Medtronic Kappa 700/650 Series software.
a
or IS-1
BI
a
BI
8
Page 9
Prescribing the Pacemaker

Indications and Usage

KDR650 Series pacemakers are indicated for the following:
Rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity.
Accepted patient conditions warranting chronic cardiac pacing which include:
– Symptomatic paroxysmal or permanent second or
third-degree AV block. – Symptomatic bilateral bundle branch block. – Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders. – Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
KDR650 Series pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include:
Various degrees of AV block to maintain the atrial contribution to cardiac output.
VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm.
9
Page 10
Prescribing the Pacemaker

Contraindications

KDR650 Series pacemakers are contraindicated for the following applications:
Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias.
Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms.
Unipolar pacing for patients with an implanted cardioverter­defibrillator (ICD) because it may cause unwanted delivery or inhibition of ICD therapy. See “Co-implantation with an Implantable Defibrillator” on page 20.

Warnings and Precautions

Rate responsive modes. Do not use rate responsive modes in those patients who cannot tolerate pacing rates above the programmed Lower Rate.
Single chamber atrial modes. Do not use single chamber atrial modes in patients with impaired AV nodal conduction because ventricular capture cannot be assured.

Pacemaker Dependent Patients

Diagnostic modes. Never program diagnostic modes (ODO, OVO, and OAO) for pacemaker-dependent patients. For such patients, use the programmer’s inhibit function for brief interruption of outputs.
Electrogram (EGM) of the patient’s intrinsic activity should be obtained with care since the patient is without pacing support when using the programmer’s inhibit function.
Polarity override. Overriding the bipolar verification prompt with bipolar polarity when a unipolar lead is connected results
10
Page 11
Prescribing the Pacemaker
in no pacing output. See “Manually Programming Polarity” on page 61 for further information.
A false bipolar pathway on a unipolar lead, a possible occurrence with bipolar 3.2 mm connector pacemakers, may result in a loss of output. See the warning in “Automatic Polarity Configuration” on page 58 for further information.
Loss of capture during threshold margin test (TMT) at a 20% reduction in amplitude indicates that the stimulation safety margin is inadequate. Consider increasing the pacing amplitude and/or pulse width. See “Magnet Operation” on page 75 for further information on the Threshold Margin Test.
Ventricular safety pacing should always be used for pacemaker-dependent patients. See “Ventricular Safety Pacing” on page 68 for further information.
Capture Management will not program ventricular outputs above 5.0 V or 1.0 ms. If the patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually. See “Capture Management” on page 64 for further information

Medical Therapy

Therapeutic Diathermy can cause fibrillation, burning of the myocardium, and irreversible damage to the pulse generator due to induced currents.
Magnetic resonance imaging of pacemaker patients has resulted in significant adverse effects. See “Magnetic Resonance Imaging (MRI)” on page 16 for further information.

Storage and Resterilization

Medtronic pacemakers are intended for single use only. Do not resterilize and re-implant explanted pacemakers.
The chart below gives recommendations on handling and storing the package. Medtronic has sterilized the pacemaker with ethylene oxide
11
Page 12
Prescribing the Pacemaker
prior to shipment. Resterilizing the pacemaker is necessary if the seal on the sterile package is broken. Resterilization does not affect the “Use Before” date.
Handling and Storage: Acceptable Unacceptable
Store and transport within Environmental Temperature limits:
°F (- 18°C) to 131°F (55°C).
0 Note: A full or partial electrical reset
condition may occur at temperatures below 0°F (- 18°C). See “Electrical Reset Parameter Settings” on page 89.
Resterilization: Acceptable Unacceptable
Resterilize if the sterile package seal is broken. Place the device in an ethylene oxide permeable package and resterilize with ethylene oxide. Allow the device to aerate ethylene oxide residues. See sterilizer instructions for details. Use an acceptable method for determining sterility, such as biological indicators.
Do not implant the device if it has been dropped on a hard surface from a height of 12 inches (30 cm) or more.
Do not resterilize the device or the torque wrench using:
an autoclave,
gamma radiation,
organic cleaning agents, e.g., alcohol, acetone, etc.,
ultra-sonic cleaners.
Do not exceed 140 17 psi (103 kPa) when sterilizing.
Do not resterilize the device more than twice.

Lead Evaluation and Lead Connection

Connector compatibility. Do not use any lead with this pacemaker without first verifying connector compatibility. Using incompatible leads can damage the connector or result in a leaking or intermittent connection.
°F (60°C) or
12
Page 13
Prescribing the Pacemaker
Pacing and sensing safety margins. Consider lead maturation when choosing pacing amplitudes, pacing pulse widths, and sensing levels. See “Manual Programming” on page 42.
Hex wrench. Do not use a hex wrench with a blue handle or a right- angled hex wrench. These wrenches have torque capabilities greater than is designed for the lead connector. See “Connection Procedure” on page 33 for lead connection instructions.

Programming and Pacemaker Operation

Epicardial leads. Ventricular epicardial leads have not been determined appropriate for use with the Capture Management feature. Therefore, Capture Management should be programmed Off if epicardial leads are implanted with
DR650 Series pacemakers.
K
Shipping values. Do not use shipping values for pacing amplitude and sensitivity without verifying that they provide adequate safety margins for the patient.
Constant current devices. To test the performance of the lead, Medtronic recommends using a constant voltage device such as the Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA). Medtronic does not recommend using a constant current device such as the Medtronic Model 5880A or 5375 External Pacemaker because the K pacemakers have constant voltage output circuits.
Crosstalk occurs in dual chamber systems when atrial pacing output pulses are sensed by the ventricular lead. Crosstalk results in self-inhibition and is more likely to occur at high sensor-driven pacing rates, high atrial amplitudes, and wide atrial pulse widths. To prevent self-inhibition caused by crosstalk, program Ventricular Safety Pacing (VSP) On or lengthen the Ventricular Blanking period.
DR650 Series
13
Page 14
Prescribing the Pacemaker
Elective Replacement Indicator (ERI). When ERI is set, the pacemaker must be replaced within three months. See “Elective Replacement Indicator” on page 77 for more information.
Full electrical reset is indicated by VVI pacing at a rate of 65 ppm without the elective replacement indicator set. See “Electrical Reset” on page 77 for more information.
Slow retrograde conduction, especially with conduction time greater than 400 ms, may induce pacemaker-mediated tachycardia (PMT).
PMT intervention. Even with the feature turned On, PMTs may still require clinical intervention such as pacemaker reprogramming, magnet application, drug therapy, or lead evaluation. See “PMT Intervention” on page 67 for further information.
Lead Monitor. If the Lead Monitor detects out-of-range lead impedance, investigate lead integrity more thoroughly.
Rate Increases
Twiddler’s syndrome, i.e., patient manipulation of the device after implant, may cause the pacing rate to increase temporarily if the pacemaker is programmed to a rate responsive mode.
Muscle stimulation, e.g., due to unipolar pacing, may result in pacing at rates up to the Upper Sensor Rate in rate responsive modes.
Unipolar Sensing
Continuous myopotentials cause reversion to asynchronous operation when sensed in the refractory period. Sensing of myopotentials is more likely when atrial sensitivity settings of
0.5 through 1.0 mV and ventricular sensitivity settings of
1.0 and 1.4 mV are programmed.
14
Page 15
Prescribing the Pacemaker

Environmental and Medical Therapy Hazards

Patients should be directed to exercise reasonable caution in avoidance of devices which generate a strong electric or magnetic field. If the pacemaker inhibits or reverts to asynchronous operation at the programmed pacing rate or at the magnet rate while in the presence of electromagnetic interference (EMI), moving away from the source or turning it off will allow the pacemaker to return to its normal mode of operation.
Hospital and Medical Environments
Electrosurgical cautery could induce ventricular arrhythmias and/or fibrillation, or may cause asynchronous or inhibited pacemaker operation. If use of electrocautery is necessary, the current path and ground plate should be kept as far away from the pacemaker and leads as possible. See “Electrosurgical Cautery” on page 47 for more information.
External defibrillation may damage the pacemaker or may result in temporary and/or permanent myocardial damage at the electrode-tissue interface as well as temporary or permanent elevated pacing thresholds. Attempt to minimize current flowing through the pacemaker and lead system by following these precautions when using external defibrillation on a pacemaker patient:
– Position defibrillation paddles as far from the pacemaker as
possible (minimum of 5 inches [13 cm]). Attempt to minimize current flowing through the pacemaker and leads by positioning the defibrillation paddles perpendicular to the implanted pacemaker/lead system.
– Use the lowest clinically appropriate energy output (watt
seconds).
– Confirm pacemaker function following any defibrillation.
High energy radiation sources such as cobalt 60 or gamma radiation should not be directed at the pacemaker. If a patient requires radiation therapy in the vicinity of the pacemaker,
15
Page 16
Prescribing the Pacemaker
place lead shielding over the device to prevent radiation damage and confirm its function after treatment.
Lithotripsy may permanently damage the pacemaker if the device is at the focal point of the lithotripsy beam. If lithotripsy must be used, program the pacemaker to a single chamber nonrate responsive mode (VVI/AAI or VOO/AOO) prior to treatment; and keep the pacemaker at least 1 to 2 inches (2.5 to 5 cm) away from the focal point of the lithotripsy beam.
Magnetic resonance imaging (MRI). Pacemaker patients subjected to MRI should be closely monitored and programmed parameters should be verified upon cessation of MRI. MRI of pacemaker patients should be carefully weighed against the potential adverse affects. Clinicians should carefully weigh the decision to use MRI with pacemaker patients. Limited studies of the effects of MRI on pacemakers have shown that:
– Magnetic and radio frequency (RF) fields produced by MRI
may adversely affect the operation of the pacemaker and may inhibit pacing output.
– Magnetic fields may activate magnet mode operation and
cause asynchronous pacing.
– Reported
1
effects of MRI on pacing include increased
ventricular pacing beyond the rate limit.
1
Holmes, Hayes, Gray, et al. The effects of magnetic resonance imaging on implantable pulse generators. PA C E. 1986; 9 (3): 360-370.
16
Page 17
Prescribing the Pacemaker
Radiofrequency ablation procedure in a patient with a
DR650 Series pacemaker may cause any of the following:
K – Asynchronous pacing above or below the
programmed rate. – Reversion to an asynchronous operation. – Pacemaker electrical reset. – Premature triggering of the elective replacement indicator.
RF ablation risks may be minimized by:
1. Programming a non-rate responsive, asynchronous pacing mode prior to the RF ablation procedure.
2. Avoiding direct contact between the ablation catheter and the implanted lead or pacemaker.
3. Positioning the ground plate so that the current pathway does not pass through or near the pacemaker system, i.e., place the ground plate under the patient’s buttocks or legs.
4. Having a Medtronic programmer available for temporary pacing.
5. Having defibrillation equipment available.
Home and Occupational Environments
High voltage power transmission lines may generate enough EMI to interfere with pacemaker operation if approached too closely.
Communication equipment such as microwave transmitters, linear power amplifiers, or high-power amateur transmitters may generate enough EMI to interfere with pacemaker operation if approached too closely.
Commercial electrical equipment such as arc welders, induction furnaces, or resistance welders may generate enough EMI to interfere with pacemaker operation if approached too closely.
17
Page 18
Prescribing the Pacemaker
Home appliances which are in good working order and properly grounded do not usually produce enough EMI to interfere with pacemaker operation. There are reports of pacemaker disturbances caused by electric hand tools or electric razors used directly over the pacemaker implant site.
Electronic article surveillance (EAS) equipment such as retail theft prevention systems may interact with pacemakers. Patients should be advised to walk directly through and not to remain near an EAS system longer than is necessary.
Cellular Phones
Note: Testing was not performed on the K
to similarity in design and function to the K data collected for the K
DR700 Series models therefore supports the
DR650 Series models due
DR700 Series models. The
safety and efficacy claims for the KDR650 Series models and is included here for reference purposes.
DR700 Series pacemakers have been tested to the frequency
K ranges used by the cellular phones included in Table 1. Based on this testing, these pacemakers should not be affected by the normal operation of such cellular phones.
These pacemakers contain a filter that allows usage, without interaction, of all cellular phones having one of the transmission technologies listed in Table 1. These transmission technologies represent most of the cellular phones in use worldwide. Patients can contact their local cellular phone service provider to confirm that the provider uses one of these technologies.
18
Page 19
Prescribing the Pacemaker
Table 1. Cellular Phone Transmission Technologies
Transmission Technology Frequency Range
Analog
FM (Frequency Modulation) 824 - 849 MHz
Digital TDMA
North American Standards
TDMA - 11 Hz 806 - 821 MHz
NADC
PCS
International Standards
GSM
DCS
a
b
(TDMA - 50 Hz)
c
1900
d
e
1800
824 - 849 MHz
1850 - 1910 MHz
880 - 915 MHz
1710 - 1785 MHz
Digital CDMA
CDMA - DS
a
Time Division Multiple Access
b
North American Digital Cellular
c
Personal Communication System
d
Global System for Mobile Communications
e
Digital Cellular System
f
Code Division Multiple Access - Direct Sequence
f
824 - 849 MHz
19
Page 20
Prescribing the Pacemaker

Co-implantation with an Implantable Defibrillator

An implantable defibrillator may be implanted concomitantly with a bipolar pacemaker.
– The use of unipolar-only Model KDR656 and the
DR650 Series bipolar models implanted with unipolar
K leads is contraindicated for patients having an implantable defibrillator.
– Follow implant protocol and precautions for pacemaker
and defibrillator lead placement. Ensure the pacemaker is configured to be compatible with the defibrillator.

Programming Considerations

Note the following programming considerations for patients who have an implantable defibrillator.
Only bipolar pacing should be used with these patients. In some cases, pacing in the unipolar configuration may cause the defibrillator either to deliver inappropriate therapy or to withhold appropriate therapy.
Polarity is automatically configured during Implant Detection (see “Automatic Polarity Configuration” on page 58 of the Product Information Manual). If lead integrity is suspect, confirmation of the automatically programmed polarities should be made after completion of Implant Detection in order to assure that bipolar polarities have been programmed appropriately.
The implantable cardiac defibrillator (ICD) should be turned off during pacemaker implantation procedures until lead polarities have been configured and confirmed. This is to prevent possible back-up unipolar paces from triggering the ICD.
Lead Monitor should not be programmed to Adaptive. When a prevalence of out-of-range lead impedance paces is detected,
20
Page 21
Prescribing the Pacemaker
the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Transtelephonic Monitor should be programmed to Off. If it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Although these pacemakers are designed to be compatible with implantable defibrillators, the potential does exist for a defibrillation pulse to reset them.
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection. If lead integrity is suspect, confirmation of bipolar polarity should be made after completion of Implant Detection.
21
Page 22
Prescribing the Pacemaker

Adverse Events

The Medtronic Kappa 700 Series devices were evaluated in a multicenter prospective study (43 investigational centers, 15 centers in the US) of the adaptive features and rate response of the device. Clinical study of the Medtronic Kappa 700 Series of pacemakers included 288 devices implanted in 285 patients worldwide.
Note: Clinical studies were not performed on the K series models due to similarity in design and function to the K Series models. The clinical data collected for the K models therefore supports the safety and efficacy claims for the
DR650 Series models and is included here for reference purposes.
K
There were a total of six deaths in the KDR700 Series study; all were reviewed and judged to be non-device related by a clinical events committee comprising clinical investigators and Medtronic clinical evaluation managers. Two were attributed to ventricular arrhythmia, one to respiratory failure, the fourth to respiratory insufficiency due to chronic obstructive pulmonary disease, the fifth to a mesotelioma, and the sixth to multi-system organ failure.
Eight devices were explanted: three due to pocket infection, one due to infection of the electrode, one from lead/connector mismatch, one patient had a psychosomatic disorder, one patient required the implant of a dual chamber ICD, and one patient continued with vasovagal symptoms and the therapy did not provide sufficient benefit.
DR650 Series
DR700
DR700 Series

Observed Adverse Events

A total of 355 adverse events were reported. The device-related events (182 events) are listed in descending order of frequency in Ta bl e 2 .
22
Page 23
Prescribing the Pacemaker
Tab le 2. Adverse Events Reported in Four or More Patients- Complicationsa
(Comps) and Observations
All patients implanted (n=288 devices in 285 patients, 133 device years)
c
% of
Patients
with Comps (n=285)
Event Total
Number
of
Events
(Patients)
Any adverse
355 (168) 17.2% 0.45 52% 2.22
b
(Obs)
Comps
per
Device-
Year
(n=133)
% of
Patients
with Obs
(n=285)
Obs per
Device-
Year
(n=133)
events
Any device-related
182 (118) 10.9% 0.31 34% 1.06
events
Pain at pocket site 32 (31) 10.9% 0.24
Other 23 (21) 1.1% 0.02 6.3% 0.15
Inappropriate
11 (11) 3.9% 0.08
programming
Atrial lead
11 (10) 3.6% 0.08
dislodgment
Programmer/ software anomaly
11 (8) 2.8% 0.08
d
Pocket infection 7 (6) 0.7% 0.02 1.4% 0.03
Intermittent
6 (6) 0.7% 0.02 1.4% 0.03
undersensing
Palpitations 6 (6) 2.1% 0.05
Pocket hematoma 6 (6) 0.4% 0.01 1.8% 0.04
Ventricular lead
6 (6) 2.1% 0.05
dislodgment
Elevated pacing
4 (4) 0.7% 0.02 0.7% 0.02
thresholds
Syncope 4 (4) 1.4% 0.03
23
Page 24
Prescribing the Pacemaker
a
Complications included those adverse events which required invasive measures to correct (e.g., surgical intervention), and were related to the presence of the pacing system or procedure.
b
Observations included those adverse events which did not require invasive measures to resolve, and were related to the presence of the pacing system or procedure.
c
Where present, a number in parentheses indicates the number of patients with the event.
d
Programmer software anomalies observed: screen lock-ups while saving data to diskette (8); problems printing reports outside of a patient session (2); and an incorrect parameter setting on a printout (1), which occurred in an earlier version of the software—software changes were made to eliminate reoccurrence.
The following other adverse events were reported, but occurred in three or fewer patients: angina pectoris; atrial flutter (paroxysmal)/ atrial fibrillation; bipolar short circuit pathway; chest pain; chest pain (non-specific); dizziness; dyspnea/shortness of breath; exit block; failure to capture/loss of capture; false negative capture detection; far field R-wave sensing; fatigue/tiredness; hypotension; inadequate lead/pacemaker connection; infection of electrode; lead/connector mismatch; lead insertion route problem; lead insulation failure; migration of pulse generator; myopotential interference; other oversensing; pacemaker mediated tachycardia; pacemaker syndrome; pectoral muscle stimulation; penetration of myocardium by lead; phrenic nerve/diaphragm muscle stimulation; pneumothorax; swelling pocket site; tachycardia (atrial); thrombus formation at lead; ventricular ectopy; ventricular tachycardia (non-sustained); ventricular tachycardia (sustained).
The following adverse events were deemed not device related (173 events were reported): angina pectoris; atrial fibrillation; atrial flutter (paroxysmal); atrial flutter (persistent); atrial tachycardia; chest pain; chest pain (non-specific); congestive heart failure; dizziness; dyspnea/shortness of breath; fatigue/tiredness; hypertension; hypotension; insufficient cardiac output; myocardial infarction (acute); palpitations; syncope; ventricular ectopy; ventricular fibrillation; ventricular tachycardia (non-sustained); ventricular tachycardia (sustained).
24
Page 25
Prescribing the Pacemaker

Potential Adverse Events

Adverse events (in alphabetical order), including those reported in Table 2, associated with pacing systems include:
Cardiac perforation
Cardiac tamponade
Death
Erosion through the skin
Hematoma/seroma
Infection
Myopotential sensing
Nerve and muscle stimulation
Rejection phenomena (local tissue reaction, fibrotic tissue formation, pulse generator migration)
Threshold elevation
Transvenous lead-related thrombosis

Clinical Studies

The Medtronic Kappa 700 Series devices were evaluated in a multicenter prospective study (43 investigational centers, 15 centers in the US) of the adaptive features and rate response of the device.
Note: Clinical studies were not performed on the K models due to similarity in design and function to the K models. The clinical data collected for the K therefore supports the safety and efficacy claims for the K series models and is included here for reference purposes.
DR650 series
DR700 series
DR700 series models
DR650
25
Page 26
Prescribing the Pacemaker

Methods

This study compared the following features of the Medtronic Kappa 700 Series pacemakers to historical controls:
Rate Response
Capture Management
Automatic Polarity Configuration
Sensing Assurance
Patient data were collected at implant, pre-discharge, two weeks, one month, two and/or three months, and six months post implant. Patients were evaluated utilizing a modified version of the Minnesota Pacemaker Response Exercise Protocol (MPREP
1
) at their one month visit. Evaluation of rate response performance for the Medtronic Kappa 700 Series pacemaker was conducted using the Metabolic Chronotropic model described by Wilkoff as applied by
2
. Automatic polarity configuration data were collected at implant.
Kay Sensing Assurance and Capture Management data were collected at each follow-up.

Description of Patients

Patients enrolled in the study represented a general dual chamber pacing population.

Results of the Study

Table 3 summarizes the results of the clinical study. The incidence of complications was found to be similar to that experienced by similar devices. The performance of the automatic polarity configuration, Capture Management, Sensing Assurance, and rate response features were found to meet study objectives.
1
Benditt, David G. M, Editor, Rate Adaptive Pacing, Blackwell Scientific Publications, Boston. 1993: 63-65.
2
Kay, Neal G., “Quantitation of Chronotropic Response: Comparison of Methods for Rate-Modulating Permanent Pacemakers”, JACC 20(7):1533-41, Dec 1992.
26
Page 27
Prescribing the Pacemaker
The slope of the exercise rate response (1.0 target slope) was less than 0.65 for 26 of 87 (30%) of patients.
All patients implanted (n=288 devices in 285 patients, 133 device years)
Primary Objectives
Automatic Polarity Configuration (n with loss of output / N leads)
Total Leads 0% (0/546) [0%, 0.55%] ≤5%
Unipolar 0% (0/107) [0%, 2.8%] ≤5%
Bipolar 0% (0/439) [0%, 0.7%] ≤5%
Sensing Assurance
(n with loss of sensing or oversensing / N device years)
Atrial 13.5% (18/133) [8.8%, 20.5%] 35.7%
Ventricular 0.8% (1/133) [0.2%, 4.1%] 9.2%
Capture Management (n with all causes loss of capture / N device years)
Loss of capture 5.3% (7/133) [2.6%, 10.5%] ≤10.7%
Slope of MPREP rate response at 1 month (n=87 patients)
Mean 0.81 [0.76, 0.86] [0.65, 1.35]
Table 3. Effectiveness Analysis
Percent of
events
% (n/N)
95%
Confidence
interval
Criteria:
Upper
95% CI
27
Page 28
Prescribing the Pacemaker
The Medtronic Kappa 700 Series pacemaker’s Rate Profile Optimization (RPO) governs sensor indicated rate (SIR) output. This function is identical in the K
DR650 Series pacemakers. Figure 1
shows the SIR vs. the Wilkoff predicted heart rate achieved using the RPO feature during the MPREP tests at 1 month.
All patients reaching Anaerobic Threshold, N=87
Expected (Wilkoff) rate, mean and 95% CI
100
90
80
70
CI
60
50
40
30
SIR (normalized)
20
10
0
0 204060
%
5
9
r
pe
p
U
n
a
e
M
Lower 95% CI
E
N
(
R
I
S
W
(
d
e
t
c
e
p
x
)
7
8
=
80
MPREP Workload (normalized)
Figure 1. Sensor Indicated Rate (SIR) vs. Expected Rate at One Month
R
I
S
)
f
f
o
k
l
i
100
28
Page 29
Implanting the Pacemaker
Chapter 2 - Implanting the Pacemaker
Implantation Procedures 30
Implant Documentation 39
Parameter Programming at Implant 40
Medical Therapy Interactions 45
Assistance 48
29
Page 30
Implanting the Pacemaker

Implantation Procedures

Testing Leads and Pacemaker

Equipment to Test the Pacemaker and Lead
To test the performance of the lead, Medtronic recommends using a constant voltage device such as the Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA).
Caution: Do not use constant current devices (such as the Medtronic external pacemaker Models 5880A, 5375, 5348, or 5346) to test lead performance. They may damage the pacemaker’s constant voltage output circuits.
For further procedures on determining thresholds and analyzing pacemaker operation, consult the PSA technical manual.
Note: Wait at least 15 minutes after implanting screw-in leads (nonsteroid) before measuring final stimulation thresholds and intracardiac sensing potentials.
Determining Stimulation Threshold
While testing the lead system prior to implant, Medtronic recommends taking and verifying threshold measurements in both the unipolar and bipolar polarities for all of the pacemaker models. Stimulation thresholds less than 1.0 V at a 0.5 ms pulse width are recommended for acute ventricular leads. Atrial leads may have slightly higher stimulation thresholds.
Test for the following configurations:
Lead tip to case (for unipolar pacing and sensing).
Lead tip to lead ring (for bipolar pacing and sensing).
If the stimulation threshold of a mature chronic lead exceeds 2.5 V at a pulse width of 0.5 ms, consider replacing the lead.
Note: Low-profile 3.2 mm bipolar and IS-1 BI leads can be connected to the pacemaker in only one way, i.e., distal (tip)
30
Page 31
Implanting the Pacemaker
electrode to the negative terminal. Because of this, the negative clip of the PSA connecting cable must be connected to the lead connector pin for correct threshold measurement. (See PSA technical manual for more information.)
Determining Lead Impedance
Acute lead impedance or pacing impedance for standard leads should typically fall between 250 and 1000 ohms, although it may fall outside this range for short periods of time.
Acute lead impedance for leads designed with impedance at or greater than 1000 ohms should typically fall between 500 and 2500 ohms. High impedances with these leads are acceptable as long as stimulation and sensing thresholds are acceptable.
If the lead or pacing impedance remains outside of the recommended range, reposition or replace the lead.
Determining Intracardiac Sensing Potentials
The intracardiac signal must be of sufficient strength to be sensed by the pacemaker. Intracardiac signals on chronic leads have a lesser magnitude and slew rate.
Note: Acceptable stimulation thresholds are not an indication of adequate sensing potentials.
31
Page 32
Implanting the Pacemaker
Analyzing Pacemaker Operation
Before connecting the leads, check the pacemaker for proper operation with a Medtronic Model 5311B (or equivalent) Pacing System Analyzer (PSA). Note that specifications for device operation apply at body temperature. As a result, some variance may occur if measurements are taken at room temperature.
During this test, the setscrew(s) in the connector assembly must contact the PSA’s cable probes. Use the appropriate cable probes to match the connector ports for the K
DR650 Series models.

Connecting Leads to the Pacemaker

The Model KDR651 connector assembly accepts IS-1 BI leads. The Model K BI leads. The Model KDR656 accepts 5 or 6 mm unipolar leads or IS-1 leads with the appropriate sizing sleeves.
The K ports. Models K contact aligned behind one another. (See Figure 2.)
DR653 accepts either low-profile 3.2 mm bipolar leads or IS-1
DR650 Series pacemakers each have two lead connector
DR651 and KDR653 have a tip and ring connector
32
Page 33
Implanting the Pacemaker
Tip Electrodes
Ring Electrodes
Atrial Port
Ventricular Por t
Figure 2. Example of Bipolar Connector Assembly for Model KDR651
Model KDR651 uses a setscrew to contact the lead tip electrode, but does not use a setscrew to contact the ring electrode. Model K
DR653
contacts the lead with setscrews at the tip and ring electrodes.
Connection Procedure
Lead connection consists of three steps:
1. Insert the provided torque wrench into the bisected grommet as shown in Figure 3.
a. Check that the setscrew is retracted from the connector
port by visual inspection of the opening. If a setscrew is obstructing the pathway, rotate the setscrew counterclockwise one or two revolutions until the opening is no longer obstructed.
Note: Counterclockwise rotation beyond this point may disengage the setscrew from the connector block.
b. Eliminate the “piston effect” (difficulty in inserting the lead
due to compression of trapped air in the connector port) when the lead is inserted by leaving the torque wrench in the distal grommet until the lead is secure. This allows a pathway for venting trapped air when the lead is inserted.
33
Page 34
Implanting the Pacemaker
The rubber connector seal maintains the torque wrench in position.
a
Figure 3. Connector Port Setscrew Preparation
b
2. Push the lead into the connector port until the lead pin is
clearly visible in the pin viewing area as indicated by arrows in Figure 4 and described in the table below.
Note: Lubricating the lead end with sterile water may ease lead insertion.
Lead-Pin Tip Placement In Pin Viewing Area
Model K
DR651 IS-1 BI Visible at end of area
Model K
DR653 IS-1 BI Just becomes visible
Low-profile
3.2 mm bipolar
DR656 5 or 6 mm unipolar Visible at end of area
Model K
Visible at end of area
34
Page 35
Implanting the Pacemaker
KDR651
K
DR656
K
DR653
(IS-1 BI leads)
Figure 4. Lead Insertion Viewing Area
(with 3.2 mm bipolar leads)
KDR653
3. Tighten the setscrew by turning clockwise until the torque wrench clicks as indicated in Figure 5. The setscrew(s) must
be engaged for pacing to begin. Do not pull on the lead until all setscrews have been tightened.
35
Page 36
Implanting the Pacemaker
KDR651
DR656
K
DR653
K
Figure 5. Tightening Setscrews
Caution: Do not overtighten the setscrew(s). Do not use blue handled or right angled hex wrenches, since they have torque capabilities greater than is designed for this connector. The torque wrench supplied with the pacemaker is designed to slip when over-torqued to prevent damage to the socket or the setscrews. Bending the wrench may break it.
36
Page 37
Implanting the Pacemaker

Pacemaker Implantation

Proper placement can help facilitate lead wrap and prevent muscle stimulation and device migration.
Pacemaker Orientation
The pacemaker may be used in either right or left pectoral implants. Either side of the shield can face the skin to facilitate excess lead wrap. However, pacemakers with an insulative coating must be implanted with the uncoated side facing the patient’s skin.
Avoiding Muscle Stimulation
If the patient experiences muscle stimulation while being paced in the unipolar configuration, program the device to lower output settings, such as a lesser Amplitude and/or narrower Pulse Width, without compromising the patient’s stimulation safety margin.
Securing the Pacemaker (Suture Hole)
A suture placed through the suture hole located in the connector assembly area helps secure the pacemaker in the subcutaneous pocket, thus minimizing post-implant rotation and migration of the device. Use normal surgical needles to penetrate the suture hole.

Replacing an Implanted Pacemaker

When replacing a pacemaker:
1. Program the pacemaker to a nonrate responsive mode prior to explantation if previously programmed to a rate responsive mode. This avoids any potential rate increase while handling the pacemaker.
2. Insert the proper wrench through each slit in the rubber grommet and loosen each setscrew by turning counterclockwise.
3. Gently retract the lead from the pacemaker connector.
37
Page 38
Implanting the Pacemaker
4. If the lead pin shows signs of pitting or corrosion, the entire lead should be discarded and replaced with a new lead to assure the integrity of the pacing system.
5. Measure stimulation thresholds and sensing potentials.
6. Pacemaker-dependent patients will be without pacing support when the lead is disconnected. Therefore, have a temporary external pacing instrument available for use.
Caution: Electrosurgical cautery could induce ventricular
arrhythmias and/or fibrillation, or may cause asynchronous or inhibited pacemaker operation. Refer to “Electrosurgical Cautery” on page 47 for information on electrocautery use.
When replacing a Medtronic pacemaker or a pacemaker from another manufacturer, the physician may require a lead adaptor kit to connect the pacemaker to the chronic lead. Use the appropriate hex wrench or screwdriver to disconnect the chronic lead. Medtronic provides such wrenches in adaptor kits for use with pacemakers manufactured by Biotronik, Coratomic, Guidant, Intermedics, St. Jude/Pacesetter, Telectronics/Cordis, and Vitatron. See the Medtronic publication Pacemaker and Connector Encyclopedia under “Adaptors and Accessories” for the appropriate adaptor kits.
For questions about lead and pacemaker compatibility, consult your Medtronic representative, or call Medtronic Technical Services.
38
Page 39
Implanting the Pacemaker

Disposal of Pacemaker

When replacing the pacemaker (due to battery depletion or explanting the pacemaker at the death of the patient who is to be cremated), return the pacemaker to Medtronic for analysis and disposal. Medtronic recommends cleaning and sterilizing the pacemakers before returning. See the back cover for mailing addresses.

Implant Documentation

Pacemaker Registration

A registration form is included in the shipping package for each Medtronic implantable pacemaker. Upon completion of the registration form by the clinician, this form serves as a permanent record of facts related to the implanted device. A copy of this form should be returned to Medtronic, where vital information will be transferred to a wallet-size, plastic-coated pacemaker Identification Card which will be mailed directly to the patient. A temporary identification card is included in the shipping package for use by the patient until the permanent card arrives. Refer to the back cover for mailing addresses.

Establishing a Patient Record

At the time of implant the clinician should establish a record of the patient’s pacemaker programmed parameter settings for later reference. A patient record can give the clinician an accurate account of the patient’s medical history at subsequent follow-ups when evaluating the pacemaker’s performance. The registration form should not serve this purpose since the form is intended for registering the patient and pacemaker with Medtronic.
39
Page 40
Implanting the Pacemaker

Parameter Programming at Implant

Shipping Parameters

The pacemaker is configured to a set of shipping parameters at the time of manufacture. At implant the device operates at these settings unless reprogrammed prior to implant. Refer to“Shipping and Nominal Parameter Settings” on page 85 for specific shipping parameter settings.
In the event that the pacemaker is not configured to shipping settings at implant, the device may have changed to partial or full electrical reset settings due to extreme temperatures or intense electromagnetic interference.
Caution: If, at implant, a programmer interrogation indicates the pacemaker has been reset to VVI, 65 ppm full electrical reset settings, use the programmer to reset the device and reprogram the previous mode and pacing operations. If, after a reset, a message appears on the programmer screen/printout indicating the Date/Time memory has been altered, contact your Medtronic representative for further information.

Implant Detection

Implant Detection is a 30-minute period, beginning at lead insertion, during which the pacemaker verifies that a lead(s) has been connected by measuring lead impedance. After 30 minutes of continuous lead connection, the pacemaker completes Implant Detection and activates the pacemaker’s automatic features listed in “Automatic Operation” on page 41.
40
Page 41
Implanting the Pacemaker

Automatic Operation

At the completion of Implant Detection, these features are automatically enabled or begin adaptation from their shipping settings.
Operating polarities, which are determined during Implant Detection, are confirmed at its completion.
Mode becomes DDDR (though shipped in the DDDR mode, the device effectively operates in the DDD mode until Implant Detection is completed).
Rate Profile Optimization is enabled.
Capture Management is enabled and begins monitoring ventricular pacing thresholds.
Sensing Assurance is enabled, and Sensitivity becomes adaptive.
Diagnostics are enabled.
If the physician manually programs Implant Detection to Off/ Complete during the 30-minute detect time, the automatic operation of the above features will be activated at that time. If a programming session is initiated at any time during Implant Detection, the pacemaker will reset Implant Detection.
41
Page 42
Implanting the Pacemaker

Manual Programming

If pacing rates, pacing outputs, and sensitivities are manually programmed at implant, the physician should be aware of the following programming considerations:
Pacing Rates
Program the Lower Rate to a setting that permits the patient to sleep comfortably or program the Sleep Function On. Program ADL Rate to provide adequate cardiac output during typical patient activity. Program the Upper Sensor Rate and Upper Tracking Rate to provide cardiac output that meets the patient’s metabolic demand during exercise without provoking symptoms (e.g., angina).
Note: For patients with stable angina, program the Upper Sensor Rate and Upper Tracking Rate to a value below the rate at which angina occurs.
Programming a combination of high Upper Sensor Rate and Upper Tracking Rate and a long refractory period may result in a shorter “sensing window.” Loss of sensing in such cases could result in competitive pacing. Programming PVARP to “varied” may minimize competitive pacing.
Programming the Upper Tracking Rate to a value greater than the Upper Sensor Rate has the effect of allowing the atrial rhythm to be tracked to a rate higher than the sensor-driven rate.
Pacing Outputs
Strive for acute ventricular stimulation thresholds of 1.0 V amplitudes at 0.5 ms pulse widths. Atrial stimulation thresholds may be slightly higher. Do not use chronic atrial or ventricular leads if the stimulation thresholds exceed 2.5 V amplitude at 0.5 ms pulse width.
The clinician should consider programming the pacemaker’s atrial and ventricular outputs to allow 2:1 voltage safety margins above the measured stimulation threshold values. Capture Management can
42
Page 43
Implanting the Pacemaker
also be programmed to provide ongoing ventricular capture monitoring and adjustment.
Capture Management can also be programmed to provide ongoing ventricular capture monitoring and adjustment.
Sensitivity
Intracardiac signal amplitudes and slew rates decrease during the lead maturation process. The clinician should consider 2:1 or 3:1 atrial and ventricular sensitivity safety margins when selecting sensitivity settings, especially with acute leads. After establishing the intracardiac sensing thresholds, the clinician may wish to program the Atrial and Ventricular Sensitivities to more sensitive settings (i.e., lower numerical settings than the measured sensing thresholds). Weigh these lower settings against the possibility of oversensing noise and electromagnetic interference, especially with unipolar sensing.
Ventricular Sensitivities of 1.0 mV or 1.4 mV with wide Atrial Pulse Widths or high Atrial Amplitudes may result in ventricular safety pacing with some lead systems at high sensor-driven pacing rates. Reprogramming ventricular Sensitivity to a less sensitive setting (higher numerical value) is one option under such circumstances. Other options include programming a longer Ventricular Blanking period.
Note: Adequate intracardiac signals should be present in both the unipolar and bipolar polarities when using bipolar leads. Measure sensing potentials in the unipolar polarity (lead tip to case) and bipolar polarity (lead tip to lead ring) prior to connecting the leads.
43
Page 44
Implanting the Pacemaker
Self-Inhibition
Self-inhibition in single chamber modes results from the sensing of the pacing output pulse. This is more likely to occur with high amplitudes, wide pulse widths, and low sensitivity settings. Reprogramming the Sensitivity to a less sensitive setting (higher numeric value) will usually resolve the situation.
In addition, inhibition can occur when far-field R-waves are sensed in AAIR, ADIR, AAI, and ADI modes. This can be prevented by proper programming of the single chamber Atrial Refractory Period and Atrial Blanking.
Muscle Stimulation with Unipolar Pacing
Under certain circumstances (e.g., high output settings), pacemaker­induced muscle stimulation may occur at the pocket site with the pacemaker programmed to ventricular unipolar pacing. Pacemaker­induced muscle stimulation may be effectively controlled and/or eliminated by programming Pulse Width to a narrower setting or programming a lower Amplitude.
Adapting Other Parameters
In addition to programming those parameters discussed above, the physician should consider adjusting other nominal parameters based on the patient’s history, medical condition, lifestyle, and hemodynamic potential and need.
Rate response parameters: Rate Profile Optimization (ADL Response and Exertion Response), Activity Threshold, Activity Acceleration, and Activity Deceleration.
Timing intervals: Sensed AV (SAV) interval and Paced AV (PAV) interval, and Rate Adaptive AV (RAAV).
Refractory and blanking periods: Ventricular Refractory Period, Ventricular Blanking, Post-Ventricular Atrial Refractory Period and Post-Ventricular Atrial Blanking.
44
Page 45
Implanting the Pacemaker
In addition, the physician should consider programming several other pacing features appropriate to the patient’s special pacing needs. These features might include:
Mode Switch for managing atrial arrhythmias, and
Sleep Function and Single Chamber Hysteresis for other pacing operations during inactivity.
Note: Non-Competitive Atrial Pacing, PVC Response, and Ventricular Safety Pacing are On/Off features that are programmed On at the time of manufacture.

Medical Therapy Interactions

This section provides details on possible interactions between pacemaker therapy and other medical therapies.

Implantable Defibrillators

Some pacemaker patients may also require therapy from an implantable defibrillator. Only bipolar pacing should be used with these patients. In some cases, pacing in the unipolar configuration may cause the defibrillator either to deliver inappropriate therapy or to withhold appropriate therapy.
Regarding use of K implantable defibrillator, note the following:
The use of unipolar-only Model KDR656 and the Models
DR651 and KDR653 implanted with unipolar leads is
K contraindicated for patients having an implantable defibrillator.
The Models KDR651 and KDR653 implanted with bipolar leads may be used in patients having an implantable defibrillator. With these pacemakers, however, polarity is automatically configured during Implant Detection. (See “Automatic Polarity Configuration” on page 58.) If lead integrity is suspect, confirmation of the automatically programmed polarities should be made after completion of Implant Detection in order
DR650 Series models in patients having an
45
Page 46
Implanting the Pacemaker
to assure that bipolar polarities have been programmed appropriately.
The implantable cardiac defibrillator (ICD) should be turned off during pacemaker implantation procedures until lead polarities have been configured and confirmed. This is to prevent possible back-up unipolar paces from triggering the ICD.
Although these pacemakers are designed to be compatible with implantable defibrillators, the potential does exist for a defibrillation pulse to reset them.
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection. If lead integrity is suspect, confirmation of bipolar polarity should be made after completion of Implant Detection.
Follow implant protocol and precautions for pacemaker and defibrillator lead placement. Ensure the pacemaker is configured to be compatible with the defibrillator.
Warning: Lead Monitor should not be programmed to Adaptive. When a prevalence of out-of-range lead impedance paces is detected, the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Warning: Programming Transtelephonic Monitor On is contraindicated for patients with implantable defibrillators. When it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
46
Page 47
Implanting the Pacemaker

Electrosurgical Cautery

Electrocautery may cause permanent loss of output prior to explanting the pacemaker and can induce ventricular fibrillation. The electrocautery tip should never be used within 6 inches (15 cm) of an implanted pacemaker/lead system.
The use of electrocautery may cause any of the following: triggered or inhibited output, a temporary pause in output, permanent loss of output, a rate in excess of the 200 ppm rate limit, reversion to asynchronous operation, an electrical reset, or premature triggering of the elective replacement indicator.
Because of these potential complications, alternatives to electrocautery should be used when available.
Effects to the pacemaker vary considerably based on the type of electrocautery unit, coagulation and cutting current settings, current pathway from the cautery tip to the indifferent plate, and the pacemaker/lead system. Electrocautery units could also interfere with ECG monitoring equipment. Monitor cardiac activity via blood pressure or pressure pulses when using electrocautery.
If electrocautery cannot be avoided, follow these steps to minimize complications:
1. Program the device to the VOO/AOO (non-rate responsive) mode beforehand and avoid direct contact with the pacemaker or leads to avoid induction of tachyarrhythmias and/or fibrillation.
2. Position the ground plate so that the current pathway does not pass through or near the pacemaker system, i.e., place ground plate under the patient’s buttocks or legs during abdominal surgery.
3. Use short, intermittent, and irregular bursts at the lowest feasible energy levels.
4. Use a bipolar electrocautery system, where possible.
5. Have temporary pacing and defibrillation equipment available.
47
Page 48
Implanting the Pacemaker

X-ray and Fluoroscopy

X-ray and fluoroscopy testing of pacemakers similar to the
DR650 Series pacemakers by exposure to diagnostic X-ray or
K fluoroscopic radiation has not affected those pacemakers.

Assistance

Assistance Information

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise. For more information, contact your local Medtronic representative or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
48
Page 49
Chapter 3 - Description
Pacing Mode Operations 50
Rate Responsive Pacing 52
Timing Operations 54
Pacing and Sensing Operations 58
Special Therapy Options 65
Description
49
Page 50
Description

Pacing Mode Operations

A brief description of programmable mode operations is given below.
Warnings :
Do not use rate responsive modes in those patients who cannot tolerate pacing rates above the programmed Lower Rate.
Do not use single chamber atrial modes in patients with impaired AV nodal conduction because ventricular capture cannot be assured.

DDDR / DDD Modes

In the presence of sensed atrial events, the ventricle will track the atrium in both the DDDR and DDD modes. Both chambers are paced at the sensor-indicated rate (DDDR) or the programmed Lower Rate (DDD) in the absence of sensed intrinsic events. An atrial sensed event inhibits atrial pacing and initiates a ventricular pace after the Sensed AV (SAV) interval. An atrial paced event initiates a ventricular pace after the Paced AV (PAV) interval. A ventricular sensed event during either AV interval inhibits the ventricular pace.

DDIR / DDI Modes

Sensed atrial events are not tracked by the ventricle; such events however, inhibit the scheduled atrial pace. Both chambers are paced at the sensor-indicated rate (DDIR) or the programmed Lower Rate (DDI). An atrial paced event initiates a ventricular pace after the Paced AV (PAV) interval, unless inhibited by a sensed ventricular event.

DVIR / DVI Modes

Both chambers are paced at the sensor-indicated rate (DVIR) or the programmed Lower Rate (DVI) in the absence of sensed ventricular events. An atrial paced event initiates a ventricular pace after the
50
Page 51
Description
Paced AV (PAV) interval, unless inhibited by a sensed ventricular event.

VDD Mode

In the presence of sensed atrial events, the ventricle will track the atrium in the VDD mode. The ventricle is paced at the programmed Lower Rate in the absence of sensed intrinsic events. A sensed atrial event initiates a ventricular pace after the Sensed AV (SAV) interval unless inhibited by a sensed ventricular event. The SAV interval may delay the ventricular pace at a rate slower than the programmed Lower Rate. No pacing occurs in the atrium.

VVIR / VDIR, AAIR / ADIR Modes

The ventricle is paced at the sensor-indicated rate (VVIR and VDIR) in the absence of intrinsic ventricular events. Events in the atrium are also sensed and recorded for diagnostic purposes (VDIR). The atrium is paced at the sensor-indicated rate (AAIR and ADIR) in the absence of intrinsic atrial events. Events in the ventricle are also sensed and recorded for diagnostic purposes (ADIR).

VVI / VDI, AAI / ADI, VVT / AAT Modes

The ventricle is paced at the programmed Lower Rate (VVI and VDI) in the absence of intrinsic ventricular events. Events in the atrium are also sensed and recorded for diagnostic purposes (VDI). The atrium is paced at the programmed Lower Rate (AAI and ADI) in the absence of intrinsic atrial events. Events in the ventricle are also sensed and recorded for diagnostic purposes (ADI).
Sensed ventricular events in the VVT mode trigger pacing pulses. Sensed atrial events in the AAT mode trigger pacing pulses.
Note: Programmed triggered pacing will not occur faster than 300 ms (200 ppm) from the previous paced event. Temporary programmed triggered pacing is not limited to 300 ms (200 ppm).
51
Page 52
Description

DOOR / DOO, VOOR / VOO, AOOR / AOO Asynchronous Modes

The DOOR mode paces the atrium and ventricle sequentially at the sensor-indicated rate, whereas the DOO mode paces both chambers at the programmed Lower Rate. The VOOR mode paces the ventricle at the sensor-indicated rate, whereas the VOO mode paces at the programmed Lower Rate. The AOOR mode paces the atrium at the sensor-indicated rate, whereas the AOO mode paces at the programmed Lower Rate.

ODO / OVO / OAO Diagnostic Modes

The ODO mode senses both intrinsic atrial and ventricular events. The OVO mode senses only intrinsic ventricular events, whereas the OAO mode senses only intrinsic atrial events.
Warning: These modes should never be programmed for pacemaker-dependent patients. For such patients, the programmer’s “Inhibit” function may be used for brief interruption of outputs.

Rate Responsive Pacing

The KDR650 Series pacemakers use activity-based pacing, which varies the pacing rate in response to the patient’s detected physical motion. In rate responsive modes, the pacing rate varies between the Lower Rate and Upper Sensor Rate limits and is dependent on the following programmable parameters:
Activity Threshold establishes the minimum signal amplitude above which activity signals are detected by the activity circuit.
Activity Acceleration and Deceleration times determine how quickly the rate will rise and fall as a result of changes in patient physical activity. Deceleration can also extend the rate fall by providing up to 20 minutes of rate deceleration based on the intensity and duration of exercise.
52
Page 53
Description
Rate response is initialized and DDDR pacing becomes effective upon completion of Implant Detection.

Rate Profile Optimization

Rate Profile Optimization controls how rapidly and to what level the sensor-indicated rate increases and decreases within submaximal and maximal rate ranges:
Submaximal rates are moderate pacing rates achieved during typical daily patient activities. These rates are at or near the ADL (Activities of Daily Living) Rate.
Maximal rates are rates at or near the Upper Sensor Rate achieved during vigorous activities.
Rate response for rates below the submaximal rate range (i.e., at or near the Lower Rate) are also controlled by Optimization.
Rate response function is optimized in the submaximal and maximal rate ranges by a daily comparison of data in two rate profiles:
Sensor rate profile is an actual rate distribution of the patient’s averaged sensor-indicated rates.
Target Rate Profile is a programmable rate distribution of the patient’s desired rates. The ADL Response and Exertion Response parameters define the time spent pacing in the submaximal and maximal ranges, respectively.
Once each day, the pacemaker evaluates the percent of time spent pacing in both the submaximal and maximal rate ranges by comparing the sensor rate profile against the target rate profile. From this comparison, the pacemaker automatically adjusts rate response in both rate ranges.
53
Page 54
Description

Timing Operations

A-A Timing and Mean Atrial Rate

KDR650 Series pacemakers use A-A timing, that is, in all dual chamber modes that pace the atrium, the pacemaker mimics sinus rhythm by timing from atrial event to atrial event. The pacemaker keeps a running average of all A-A intervals (except those starting with an atrial sense or atrial refractory sense and ending with an atrial pace) which is called the Mean Atrial Rate (MAR). The Mean Atrial Rate is used by the pacemaker in Rate Adaptive AV operation.

Rates

The KDR650 Series pacemakers have four rates that determine the patient’s pacing rate.
The programmable Lower Rate is the minimum pacing rate for a given mode.
The programmable Upper Tracking Rate is the maximum ventricular rate at which the atrium is tracked in the DDDR, DDD, and VDD mode.
ADL Rate (Activities of Daily Living Rate) is the average target rate that the patient’s heart rate is expected to reach during moderate exercise.
Upper Sensor Rate provides the upper limit for the sensor-driven rate during physical activity, particularly during vigorous exercise.

AV Intervals

The pacemaker has separately programmable paced and sensed AV intervals that determine the duration between an atrial event and the delivery of a ventricular stimulus in dual chamber modes.
The programmable Paced AV (PAV) interval defines the duration from paced atrial events to paced ventricular events.
54
Page 55
Description
The ventricle is paced after the PAV expires unless inhibited by a sensed ventricular event.
Note: PAV duration may differ from the programmed setting due to Mode Switch, Rate Adaptive AV, Ventricular Safety Pacing, Non-Competitive Atrial Pacing or Capture Management operation.
The programmable Sensed AV (SAV) interval defines the duration from sensed atrial events to paced ventricular events. The ventricle is paced after the SAV expires unless inhibited by a sensed ventricular event. SAV maintains 1:1 AV tracking up to the Upper Tracking Rate (UTR) or the total atrial refractory period (which is the sum of the SAV and the PVARP).
Note: SAV duration may differ from the programmed setting due to Mode Switch, Rate Adaptive AV, or Capture Management operation or may be extended as the result of Wenckebach operation if the intrinsic atrial rate exceeds the UTR.

Rate Adaptive AV

The programmable On or Off Rate Adaptive AV (RAAV) feature provides rate-variable PAV and SAV intervals. The variable PAV shortens or lengthens to the sensor-indicated rate, whereas the variable SAV shortens or lengthens to the Mean Atrial Rate. The variable SAV promotes 1:1 AV tracking to sensed atrial events at higher rates. Several programmable parameters define RAAV operation.
Start Rate defines the rate at which RAAV operation begins.
Stop Rate defines the rate above which the PAV and SAV intervals stop shortening and remain at the minimum AV interval.
Max Offset defines the maximum amount of time by which the PAV and SAV intervals can be shortened.
55
Page 56
Description
As the sensor-indicated rate increases above the Start Rate, the PAV gradually shortens until the programmed Max Offset is reached at the Stop Rate. As the Mean Atrial Rate increases above the Start Rate, the SAV gradually shortens until the programmed Max Offset is reached at the Stop Rate.
The approximate interval difference between the programmed PAV and SAV is maintained throughout the rate range when RAAV is active. The pacemaker operates at the programmed PAV and SAV at rates below the programmed Start Rate.
Notes:
The 2:1 block rate increases to a higher rate as the SAV component of the TARP shortens.
The sensing window after the PVARP lengthens as the PAV shortens at higher rates.

Upper Rate Behaviors (2:1 Block Rate, Wenckebach)

The 2:1 block rate occurs when every other atrial event falls within the PVARP. Such refractory sensed events do not initiate an SAV and, thus, are not synchronized to a ventricular paced event, thereby producing a 2:1 block rate. If the 2:1 block rate exceeds the programmed Upper Tracking Rate, the SAV is extended to limit the ventricular rate at the Upper Tracking Rate, thereby producing pacemaker Wenckebach operation.

Refractory and Blanking Periods

The pacemaker has several refractory periods to avoid inappropriate restarting of certain timing intervals.
The programmable Post-Ventricular Atrial Refractory Period (PVARP) prevents tracking of retrograde P waves. In the DDIR and DDI modes, PVARP prevents atrial inhibition from retrograde P waves. The PVARP occurs after paced, sensed,
56
Page 57
Description
and refractory sensed ventricular events and is set to at least 400 ms after a pacemaker-defined PVC.
The first portion of the PVARP is a programmable Post­Ventricular Atrial Blanking period which disables atrial sensing after paced, sensed, and refractory sensed ventricular events.
– The programmable sensor-varied PVARP protects against
PMTs at lower sensor-indicated rates by lengthening the PVARP and provides a higher 2:1 block rate at higher sensor-indicated rates by shortening the PVARP. Sensor-varied PVARP is based on the sensor-indicated pacing rate and varies from approximately 400 ms at the lower rate to no less than the Post-Ventricular Atrial Blanking period at higher sensor rates.
The programmable Ventricular Refractory Period (VRP) prevents sensing of T-waves or PVCs. The VRP occurs after paced, sensed, and refractory sensed ventricular events in dual chamber and ventricular single chamber modes.
The first portion of the VRP is a nonprogrammable Ventricular Blanking period in which ventricular sensing is disabled after paced, sensed, and refractory sensed ventricular events. The Ventricular Blanking period varies dynamically from 50 to 100 ms based on the strength and duration of the ventricular event.
The programmable Ventricular Blanking period disables ventricular sensing in dual chamber modes to prevent potential crosstalk from paced atrial events.
The programmable Atrial Refractory Period (ARP) prevents atrial inhibition due to sensed far-field R waves or noise. The ARP occurs after paced, sensed, and refractory sensed atrial events in single chamber atrial modes.
The first portion of the ARP is a programmable Atrial Blanking period that disables atrial sensing after paced, sensed, and refractory sensed atrial events.
57
Page 58
Description
The first portion of the PAV and SAV intervals is a nonprogrammable Atrial Blanking period in which atrial sensing cannot take place after paced or sensed atrial events. The Atrial Blanking period varies dynamically from 50 to 100 ms based on the strength and duration of the atrial event.

Pacing and Sensing Operations

Polarity

Unipolar polarity uses the lead tip as the active electrode and the pacemaker case as the common electrode. Bipolar polarity uses the lead tip as the active electrode and the lead ring as the common electrode.
With the bipolar Model K sensing polarities are automatically or manually programmable, and in dual chamber pacemakers polarity is independently programmable for the atrial and ventricular channels. The Model
DR656 provides unipolar pacing and sensing only.
K
Automatic Polarity Configuration
With bipolar pacemakers, polarity settings are selected automatically during the pacemaker’s 30-minute Implant Detection period that begins at lead insertion. During Implant Detection, with either bipolar or unipolar leads implanted, and with the Configure or Adaptive parameter under Lead Monitor active, the pacemaker continuously measures impedance as follows:
The pacemaker issues a bipolar pace and immediately checks the pace for high impedance.
– If the pace is within range, it is considered an acceptable
bipolar pace.
– If high impedance is found, the pacemaker immediately
follows the bipolar pace with a backup unipolar pace.
DR651 and KDR653 pacemakers, pacing and
58
Page 59
Description
If a unipolar backup pace is issued, the pacemaker checks it for high impedance.
– If the pace is within range, it is considered an acceptable
unipolar pace.
– If high impedance is found, the pacemaker assumes a lead
is not attached and restarts Implant Detection.
Note: During polarity configuration, the pacemaker also checks for low impedance paces but does not follow them with unipolar backup paces. To avoid possible loss of capture due to continuous low impedance pacing, the pacemaker sets unipolar polarity when 3 of 16 paces are detected as low impedance during the first phase of configuration (described below).
Warning: If the Implant Detection feature identifies a bipolar, high-impedance path, backup unipolar pacing will be delivered. This may be inappropriate for patients with an implantable defibrillator (ICD). See “Notes” on page 3-14.
Polarity configuration for bipolar pacemakers takes place in two phases:
Initial Configuration Phase – Five minutes after lead connection, if a high impedance unipolar pace has not reset Implant Detection, the pacemaker issues three asynchronous paces at magnet rate and determines if they are bipolar or unipolar.
– Two of three paces must be bipolar for initial pacing and
sensing polarities to be determined bipolar for the Confirmation Phase.
– Two of three paces must be unipolar for initial pacing and
sensing polarities to be set to unipolar and remain unipolar during and after Implant Detection.
Confirmation Phase – If bipolar polarity is determined at five minutes, the remaining 25 minutes of Implant Detection is used to confirm bipolar polarity.
59
Page 60
Description
– The pacemaker continues to monitor each pace as
described above, providing backup unipolar paces for high impedance bipolar paces.
– The pacemaker checks for a programmed number of out-
of-range (high or low impedance) paces that indicate bipolar pacing is in jeopardy.
– The pacemaker confirms bipolar pacing and sensing
polarity at the end of 25 minutes by examining three asynchronous paces as described above for the Initial Configuration Phase.
Note: If a prevalence of out-of-range bipolar paces is detected during the Confirmation Phase, the pacemaker reconfigures pacing and sensing polarities to unipolar and the asynchronous paces are not delivered.
Warning: When implanting a Model K
DR653 pacemaker, ensure that
both the tip and ring setscrews are properly engaged and all electrical contacts are sealed in order to prevent possible electrical leakage between the tip and ring contacts. Such leakage may cause the pacemaker to falsely identify a unipolar lead as bipolar, resulting in a loss of output. For the same reason, ensure that electrical contacts are sealed when using lead extenders or adaptors with all bipolar models.
Notes:
During Implant Detection, the unipolar-only Model KDR656 pacemaker examines unipolar pacing impedance to determine lead connection.
When implanting a bipolar pacemaker in a patient with an implantable defibrillator, the physician has the option to manually program polarities to avoid the possible occurrence of unipolar backup paces during Implant Detection.
60
Page 61
Description
Manually Programming Polarity
Pacing and Sensing Polarities for the Models KDR651 and KDR653 can also be manually programmed. However, the Configure selection under Lead Monitor, which controls automatic configuration, must be “turned off” by programming the Monitor Only selection instead.
When pacing and/or sensing polarity is manually programmed from unipolar to bipolar, the pacemaker verifies whether a functioning bipolar lead is connected by measuring lead impedance. If lead impedance is outside the accepted programmed impedance range for bipolar polarity (between 200 and 4000 ohms), the lead is assumed to be unipolar and pacing and/or sensing polarity remains set to unipolar.
Warning: If the clinician overrides the bipolar lead verification and programs bipolar polarity when a unipolar lead is connected to the pacemaker, no pacing output results.

Lead Monitor

The Lead Monitor feature enables the pacemaker to monitor lead integrity and adapt bipolar pacing and sensing to unipolar when bipolar lead integrity is in doubt. It also controls automatic configuration of lead polarities at implant. Lead Monitor is available in all pacing modes.
The Lead Monitor feature monitors lead integrity by looking for high and low impedance on each pace. The Atrial and Ventricular Lead Monitor can be programmed to operate as follows:
Configure – provides automatic polarity configuration of bipolar pacemakers (bipolar pacemakers are shipped in Configure).
Adaptive – monitors bipolar paces and provides backup unipolar
paces when impedance is high
– switches pacing and sensing polarities from bipolar to
unipolar when a prevalence of out-of-range (high and low impedance) paces is detected
61
Page 62
Description
– provides automatic polarity configuration if the pacemaker
is programmed to Adaptive prior to implant
Monitor Only – monitors unipolar and bipolar paces to determine if they are out of range (high and low impedance). Unipolar-only Model K
DR656 is shipped in and only
programmable to Monitor Only.
When Lead Monitor is set to Adaptive or Monitor Only, and the pacemaker determines a lead is out of range, the pacemaker issues a status warning that appears on the programmer screen at the next interrogation. However, a status warning is not issued at any time during the configuration process.
After automatic polarity configuration is complete, Lead Monitor is automatically set to Monitor Only for bipolar and unipolar configurations. If, however, the Adaptive setting was programmed prior to implant, bipolar configurations are set to Adaptive, but unipolar configurations are set to Monitor Only.
Warning: Lead Monitor should not be programmed to Adaptive for patients with implantable defibrillators. When there is a prevalence of out-of-range lead impedance paces detected, the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.
Caution: If the Lead Monitor detects out-of-range lead impedance, investigate lead integrity more thoroughly.

Sensitivity

Sensitivity determines the minimum intracardiac signal that the pacemaker can detect when intrinsic atrial and ventricular events occur. With K or manually programmable.
Sensing Assurance, the pacemaker’s automatic sensing feature, adjusts sensitivities within defined limits. At the completion of Implant Detection, the pacemaker enables Sensing Assurance and begins
DR650 Series pacemakers, sensitivity is automatically
62
Page 63
Description
continual monitoring of the peak amplitude of sensed signals. In response to monitoring, it automatically increases or decreases Sensitivity to maintain an adequate sensing margin with respect to the patient’s sensed P and R waves.
During provocative myopotential testing of the Medtronic Kappa 700 Series pacemakers having Sensing Assurance On, 56% (28/50) of patients experienced atrial oversensing while 7.3% (4/55) experienced ventricular oversensing. The atrial oversensing rate is comparable to that experienced by Medtronic’s Thera devices which had an atrial oversensing rate of 52.4% (75/143) during provocative testing. The ventricular oversensing rate of Medtronic Kappa 700 Series pacemakers compared favorably to the ventricular oversensing rate experienced by Medtronic Thera pacemakers –
44.1% (63/143).
For guidelines regarding manually programming Sensitivity settings, refer to “Parameter Programming at Implant” on page 40.
Atrial and Ventricular sensitivities are separately programmable. They can also be programmed on a temporary basis.
Note: Atrial Sensitivity with unipolar sensing polarity is restricted to
0.5 mV to limit oversensing of muscle noise or electromagnetic interference.

Pacing Output

Amplitude and Pulse Width determine the pacing pulse strength necessary to capture the atrium and ventricle. Amplitude and Pulse Width are programmable for the atrial and ventricular channels. Amplitude and Pulse Width can also be programmed on a temporary basis. Refer to “Parameter Programming at Implant” on page 40 for pacing output guidelines.
63
Page 64
Description

Capture Management

The Capture Management feature provides automatic monitoring of ventricular pacing thresholds and automatic adjustment of Amplitude and Pulse Width to maintain capture.
When Capture Management is programmed to Monitor Only, the pacemaker periodically causes paces to be delivered (affecting pacemaker timing temporarily if necessary). The pacemaker then monitors the paces by changing first amplitude and then pulse width to find two points that lie on the strength duration curve that define the boundary between settings that capture and those that do not. How often the pacemaker performs this pacing threshold search is determined by the programmable Capture Test Frequency parameter. This parameter determines how often the pacing threshold search will be initiated and provides for retry if the test is delayed.
When Capture Management is programmed to Adaptive, the pacemaker responds to monitoring by adapting ventricular amplitude and pulse width using the following programmable parameters:
Amplitude Margin and Pulse Width Margin – the pacemaker­determined threshold multiplied by the selected safety factor.
Minimum Adapted Amplitude and Minimum Adapted Pulse Width – the lowest Amplitude and Pulse Width values that the pacemaker can suggest for programming after a pacing threshold search has been performed.
Acute Phase Days Remaining – time in days when the pacemaker will never suggest the use of output settings less than the operating Amplitude and Pulse Width values. This parameter is used during the lead maturation period.
Capture Management becomes operational automatically when Implant Detection is completed.
Warning: Capture Management will not program ventricular outputs above 5.0 V or 1.0 ms. If the patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually.
64
Page 65
Description
Note: In the event of partial or complete lead dislodgment, Capture Management may not prevent loss of capture.

Special Therapy Options

The pacemaker has several therapy features for managing atrial and ventricular arrhythmias and other pacing operations. A brief description of these features is given below.

Mode Switch

Mode Switch is a programmable On or Off feature designed to prevent tracking of paroxysmal atrial tachycardias in the DDDR, DDD, and VDD modes. Whenever a rapid atrial rhythm is detected, the pacemaker switches from an atrial tracking mode to a non-atrial tracking mode until the atrial tachycardia ceases. Mode switching occurs as follows:
Atrial Tracking Mode Non-Atrial Tracking Mode
DDDR DDIR
DDD DDIR
VDD VDIR
In determining when to mode switch, the pacemaker continuously monitors A-A intervals for a rapid atrial rate. When it determines that the atrial rate equals or exceeds the programmed Detect Rate, the rate at which pacemaker-defined atrial tachycardia starts, the pacemaker switches mode and gradually changes the ventricular rate to the sensor-indicated rate. When seven A-A intervals occur below the Upper Tracking Rate, or five consecutive atrial paces occur, the pacemaker switches back to the programmed atrial tracking mode.-
→ ←
→ ←
→ ←
65
Page 66
Description
In monitoring the atrial rate for tachycardias, the pacemaker searches for:
any four of the last seven A-A intervals that are shorter than the programmed Detect Rate interval, and
eight consecutive A-A intervals that measure less than two times the total atrial blanking period (AV interval + PVAB) and less than two times the Detect Rate interval. If they are found, the pacemaker extends the PVARP for one beat to expose the blanked atrial flutter that may be occurring. This search occurs if the Blanked Flutter Search parameter is On.
Note: Mode Switch is not recommended for patients with chronic refractory atrial tachyarrhythmias (i.e., atrial tachycardia, atrial fibrillation, atrial flutter).

Non-Competitive Atrial Pacing

Non-Competitive Atrial Pacing (NCAP) is a programmable On or Off feature in the DDDR and DDD modes. NCAP prevents an atrial pace from falling within the atrium’s relative refractory period, thereby preventing a potential atrial tachycardia. A refractory sensed atrial event within the PVARP starts a 300-ms NCAP period during which no atrial pacing may occur. The scheduled atrial pace is delayed until the NCAP period expires, thus preventing pacing within the atrium’s relative refractory period. The Paced AV following NCAP operation may be shortened to a minimum of 30 ms to maintain a stable ventricular rate.
Note: Even when NCAP is programmed Off, NCAP operation is invoked automatically in the DDDR mode for one cycle when PVC Response or PMT Intervention operations occur.
66
Page 67
Description

PMT Intervention

Pacemaker-Mediated Tachycardia (PMT) Intervention is a programmable On or Off feature that provides automatic detection and interruption of pacemaker-defined PMTs. The pacemaker detects a PMT episode as eight consecutive ventricular pace-to-atrial sense events with VA intervals less than 400 ms. In addition, the pacemaker corroborates the episode as a PMT if, on the eighth consecutive pace, the Sensor Rate is at or below the ADL rate. When an episode is detected and corroborated, the pacemaker intervenes by forcing a 400-ms PVARP for one pacing cycle. Extending the PVARP ensures that the next sensed atrial event within 400 ms will be refractory, thus interrupting atrial tracking for one cycle, which may stop the PMT.
Note: A sinus tachycardia could cause a feature intervention, resulting in a single P wave falling in the PVARP and therefore not being tracked by the pacemaker.
After intervention occurs, PMT detection is automatically suspended for 90 seconds to prevent unnecessary intervention for fast intrinsic atrial rates.
Caution: Even with the feature turned On, PMTs may require clinical intervention such as pacemaker reprogramming, magnet application, drug therapy, or lead evaluation.

PVC Response

Premature Ventricular Contraction (PVC) Response is a programmable On or Off feature that prevents tracking of retrograde P waves generated by PVCs in the DDDR, DDD, and VDD modes. PVC Response prevents inhibition of atrial pacing resulting from retrograde P waves generated by PVCs in the DDIR and DDI modes. Any ventricular event-to-ventricular refractory or nonrefractory sense event with no intervening atrial event starts a minimum PVARP of 400 ms. A minimum PVARP of 400 ms causes retrograde P waves to be sensed as refractory events, thereby preventing retrograde P waves
67
Page 68
Description
from starting an SAV interval which could result in a PMT. Note that an SAV interval is not started in the DDIR and DDI modes.

Ventricular Safety Pacing

Ventricular Safety Pacing (VSP) is a programmable On or Off feature that prevents ventricular asystole due to inappropriate inhibition of ventricular pacing by crosstalk or ventricular oversensing. A sensed ventricular event within 110 ms after an atrial pacing pulse results in a ventricular pacing pulse at 110 ms or the programmed or rate adaptive PAV interval, whichever expires first. If the sensed event was, in fact, a ventricular depolarization, the ventricular pacing pulse at 110 ms falls harmlessly into the absolute refractory period of the ventricle.
Warning: VSP should always be programmed On for pacemaker­dependent patients.

Sleep Function

Sleep Function is a programmable On or Off feature that reduces the paced rhythm during sleep. It replaces the programmed Lower Rate with a slower Sleep Rate during a programmable sleep period. At the programmed Bed Time, the pacemaker gradually lowers the operating lower rate from the programmed Lower Rate to the programmed Sleep Rate over a 30 minute period. The Sleep Rate becomes the operating lower rate until the programmed Wake Time occurs, when the pacemaker gradually raises the operating lower rate from the Sleep Rate to the Lower Rate over a 30 minute period.
Note: For rate responsive modes, the Sleep Rate becomes the minimum rate during the sleep period. However, in the presence of sensor-detected activity, the sensor-indicated rate overrides the Sleep Rate.
68
Page 69
Description

Single Chamber Hysteresis

Single Chamber Hysteresis allows tracking of an appropriate intrinsic rhythm below the Lower Rate during extended periods of inactivity such as sleep. When a nonrefractory sensed event occurs, sensed events that follow are allowed to occur below the programmed Lower Rate to the programmed Hysteresis Rate. As long as the intrinsic rate is above the Hysteresis Rate, pacing is inhibited. When the intrinsic rate drops to the Hysteresis Rate, escape pacing results and the programmed Lower Rate returns as the operating lower rate until the next nonrefractory sensed event occurs.
69
Page 70
Page 71
Chapter 4 - Pacemaker Follow-up
Pacemaker Telemetry 72
Other Operations 75
Diagnostics 79
General Recommendations 80
Pacemaker Follow-up
71
Page 72
Pacemaker Follow-up

Pacemaker Telemetry

Pacemaker telemetry is established by placing the programming head over the implant site and moving the head along the axis shown in Figure 6 until the head lights indicate telemetry is possible. Pacemaker telemetry can be used for routine monitoring. A brief description of the types of telemetry is given below.
Axis
Figure 6. Positioning the Programming Head

Parameter Summary

Upon interrogation by the programmer, the pacemaker telemeters the programmed parameter settings along with battery status and device identification.
Note: The parameters may change due to a full or partial electrical reset, elective replacement indication, Rate Profile Optimization, automatic Sensitivity, Capture Management, or Lead Monitor operation. See the appropriate topic for more information.

Battery and Lead Information

The pacemaker provides Battery and Lead Information for both measured and calculated values of the pacing output pulse(s), the lead(s), and the battery. The pacemaker’s Battery and Lead Information is accessed via the programmer.
72
Page 73
Pacemaker Follow-up

Telemetry Markers (Extended Markers)

Telemetry markers depict cardiac events that are useful in evaluating pacemaker operations.
Standard Markers (Marker Channel Telemetry) annotate pacemaker paced, sensed, and refractory sensed events in any mode.
Therapy Trace Markers provide depictions of therapy interventions that alter timing intervals.

Electrogram (EGM)

The Electrogram (EGM) is a real-time intracardiac waveform detected by the atrial or ventricular lead. Atrial and ventricular waveforms can be displayed simultaneously with Dual EGM (simultaneous display of the independent channels) or Summed EGM (the electrical summation of both channels). The waveform(s) is accompanied by standard Marker Channel telemetry annotation.
Warning: An EGM of the patient’s intrinsic activity should be obtained with care since the patient is without pacing support when using the programmer’s inhibit function.
Notes:
The simulated pacing pulses on the EGM should not be interpreted as representing the actual timing or amplitude of paced events. Use an ECG to determine such events.
Intracardiac amplitudes measured on the EGM should not be used to assess the adequacy of the programmed Sensitivity setting. Use the sensing test instead.
73
Page 74
Pacemaker Follow-up

Transtelephonic Monitor

The Transtelephonic Monitor is a programmable On or Off feature intended for use with remote pacemaker monitoring services. When the Transtelephonic Monitor is programmed On, the Threshold Margin Test is delayed for five seconds upon magnet application to enhance communication with transtelephonic equipment. If the pacing polarity is programmed to bipolar, it is temporarily set to unipolar to provide improved ECG artifact detection for the remote monitoring equipment. (The programmed polarity is restored when the magnet is removed.) When Transtelephonic Monitor is programmed Off, the Threshold Margin Test is not delayed upon magnet application, and conventional transtelephonic monitoring can occur.
Warning: Programming Transtelephonic Monitor On is contraindicated for patients with implantable defibrillators. When it is programmed On, the pacing polarity is temporarily set to unipolar when the magnet is applied. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.

Extended Telemetry

Extended Telemetry provides frequent patient monitoring in the programmed mode. When Extended Telemetry is programmed On, the pacemaker automatically transmits the programmed telemetry type (i.e., Electrogram, Marker Channel telemetry) when the programmer head or monitoring receiver (with or without a magnet) is positioned over the pacemaker. Extended Telemetry is automatically programmed Off after 48 hours or can be programmed Off via the programmer.
74
Page 75
Pacemaker Follow-up

Other Operations

Magnet Operation

Magnet operation includes an initial capture safety margin test, an automated self test, and magnet mode operation with telemetry:
The Threshold Margin Test (TMT) provides a check for loss of capture. It is performed at 100 ppm with pulse amplitudes reduced by 20% on the third pulse.
Warning: Loss of capture during TMT may indicate that the stimulation threshold safety margin is inadequate. Perform a pacing threshold test and reprogram outputs to establish a 2:1 voltage safety margin, if necessary.
The Fully Automated Self Test (FAST) is a pattern of pacing pulses that indicates either of the following events has occurred:
– Lead Monitor has detected an out-of-range lead, and, if
programmed to Adaptive, has switched polarities from bipolar to unipolar.
– Capture Management has measured unusually high
ventricular thresholds.
Beginning on the cycle following TMT, the pacemaker delivers two pacing pulses at 85 ppm followed by two pulses at 65 ppm and then repeats the same sequence once again.
Magnet Mode forces the pacemaker’s programmed operation to an asynchronous mode (DOO in dual chamber configurations, VOO in the VDD mode, and VOO/AOO in single chamber configurations). Magnet Mode rate is 85 ppm for all of the above modes. However, the rate is 65 ppm if an electrical reset occurs or the elective replacement indicator is set.
75
Page 76
Pacemaker Follow-up
Magnet Operation Using the Transtelephonic Magnet
When a transtelephonic magnet is placed over the pacemaker, a TMT is automatically initiated. A FAST, if programmed On, immediately follows the TMT if any of the events mentioned above that initiate the test have occurred. When the FAST (or TMT when a FAST is not initiated) is completed, pacing is forced to Magnet Mode operation until the magnet is removed.
Magnet Operation Using the Programming Head
The programming head does not initiate a TMT when placed over the pacemaker. The clinician must perform a Magnet Test in order to obtain a TMT. A FAST, if programmed On, will follow the TMT if initiated by any of the events described above. When the FAST (or TMT when a FAST is not initiated) is completed, pacing is forced to Magnet Mode operation until the Magnet Test is terminated or the programming head is removed.
FAST Reporting Using the Remote Assistant
With the patient-activated hand-held Remote Assistant, the FAST results indicating a lead or threshold problem are reported by two beeps in a low tone, and a yellow light appears on the underside of the activator. When no problems exist, the FAST results are reported by three beeps, and a green light appears instead.
FAST with the Remote Assistant also reports if the ERI has been set or an electrical reset has occurred (indicated by two beeps and a yellow light from the activator).

Temporary Parameters

Several pacemaker parameters can be programmed temporarily during a patient session. These include: Mode, Rate, Amplitude, Pulse Width, and Sensitivity, and AV intervals.
76
Page 77
Pacemaker Follow-up

Electrical Reset

A partial Electrical Reset can occur when the pacemaker is exposed to cold temperatures, electrocautery, internal or external defibrillation. Under these conditions, the pacemaker resumes pacing at the previously programmed mode and preserves many of the primary pacing parameter settings in operation prior to the partial reset. See “Electrical Reset Parameter Settings” on page 89 for a complete list of the preserved settings as well as full Electrical Reset settings.
If the temperature or electrical disturbance is severe enough, the pacemaker can reset to full Electrical Reset settings, in which case the pacemaker reverts to VVI mode at a pacing rate of 65 ppm and restarts Implant Detection. To restore the pacemaker to its previous mode and pacing operations, it must be reset and manually reprogrammed via the programmer.
Note: If after a reset a message appears on the programmer screen or printout indicating the Date/Time memory has been altered, contact your Medtronic representative for further information.

Elective Replacement Indicator

The Elective Replacement Indicator (ERI) forewarns when the battery is nearing depletion. When the pacemaker detects that the battery voltage has dropped below 2.59 V, the pacemaker sets the ERI status and reverts to VVI operation at a rate of 65 ppm. Pacing
is maintained, in VVI mode, even in pacemakers previously programmed to the AAIR, ADIR, AAI, ADI, AAT, AOOR, or AOO mode.
Caution: When ERI is set and verified, the pacemaker must be
replaced within three months.
High-rate pacing over several years, or with the battery partially depleted, may trigger the ERI prematurely.
Continuous pacing at high rates (i.e., greater than 100 ppm) and high output settings (i.e., pulse widths greater than 1.0 ms and/or
77
Page 78
Pacemaker Follow-up
amplitudes greater than 5.0 V) may cause the battery to deplete rapidly and set ERI. While operating at ERI, the battery voltage may rise above 2.59 V, allowing the pacemaker to be reset and reprogrammed.
Before and after the use of high-rate pacing, the clinician should interrogate the pacemaker to determine the battery status and ERI condition.

Emergency Pacing

Emergency pacing provides VVI pacing at high output settings in emergency situations for pacemaker-dependent patients. See “Electrical Reset Parameter Settings” on page 89.

Rate Limit

In the event of a single component failure, the absolute rate limit for atrial and ventricular rates are held independently to 200 ppm. Rate limit is automatically overridden in temporary single chamber modes for high rate pacing.
78
Page 79
Pacemaker Follow-up

Diagnostics

The pacemaker has automatic data summary diagnostics, some of which have an accompanying clinician-selected diagnostic to collect further detailed data. At the completion of Implant Detection, the pacemaker enables diagnostic collection.

Automatic Diagnostics / Clinician-Selected Diagnostics

Heart Rate Histograms. Automatically collect short and long term heart rate and percent-paced data for the atrium and ventricle.
AV Conduction Histogram. Automatically collects short and long term AV conduction sequences.
Sensor Indicated Rate Profile. Automatically collects sensor rate data.
Atrial and Ventricular High Rate Episodes. Automatically collect basic data on atrial and ventricular high rates and, optionally, clinician-selected detailed data of atrial, ventricular, or summed electrograms with a rate trend.
Lead Diagnostics. Chronic Lead Impedance Trend automatically collects long-term chronic lead impedance data. The automatic Lead Monitor Counters present the impedance status of every beat. Chronic lead impedance data is collected in all pacing modes except for triggered pacing modes (AAT and VVT). This data is not collected in non-pacing modes (OAO, OVO, and ODO).
Sensitivity Trend. Automatically collects the minimum and maximum sensitivities used to operate automatic sensitivity.
Capture Management Trend. Automatically collects information about the range of ventricular thresholds and outputs over the life of the device.
79
Page 80
Pacemaker Follow-up
Mode Switch Episodes. Automatically collect basic data on mode switch episodes, or optionally, clinician-selected electrogram with a rate trend.
Custom Rate Trend. Clinician-selected collection of heart rate and percent-paced trends over a beat-to-beat 1-hour or 24-hour period.

General Recommendations

During the first few months after receiving a new pacing system the patient may require close monitoring. In the U.S.A., the physician may wish to use the individual state’s or Medicare’s maximum frequency table for dual chamber pacemakers under Guideline I, i.e., every two months from the second month through three years post-implant, and monthly thereafter (Ref. Carriers Manual, Part 3, Claims Process, Transmittal No. 1051, or a later publication, if available). Note that Guideline I applies to those pacing systems (i.e., the pacemaker and the lead used) for which 5 years of clinical data are not yet available.
80
Page 81

Pacemaker Specifications

Pacemaker Specifications
Lead Requirements, Compatibility 82
Radiopaque Identification 83
Emergency Parameter Settings 84
Shipping and Nominal Parameter Settings 85
Electrical Reset Parameter Settings 89
Elective Replacement Indicator 93
Magnet Mode Conditions 93
Longevity Projections 94
Programmable Parameters 96
Nonprogrammable Parameters 103
Temporary Parameters 104
Telemetry Markers 105
Electrograms 105
Automatic Diagnostics 105
Clinician-Selected Diagnostics 106
Patient Information 106
Battery and Lead Telemetered Information 107
Battery Parameters 108
Mechanical Dimensions 108
81
Page 82
Pacemaker Specifications

Lead Requirements, Compatibility

The chart below lists the available models by polarity type with the required lead for proper operation. Refer to “Connecting Leads to the Pacemaker” on page 32.
Models Polarity Primary Leads
K
DR651 Bipolar/Unipolar IS-1
KDR653 Bipolar/Unipolar Low-profile 3.2 mm bipolar or IS-1a BI
K
DR656 Unipolar Unipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992) whereby pulse generators and leads so designated are assured of meeting the electrical and mechanical parameters specified in the IS-1 International Standard.
a
BI
82
Page 83
Pacemaker Specifications

Radiopaque Identification

Listed below are the pacemaker’s radiopaque identifications, which are used to identify the pacemaker with fluoroscopy. An X-ray showing the location of the radiopaque for Model K Figure 7.
Models Radiopaque
DR651 PLJ
K
K
DR653 PLK
K
DR656 PLL
PLJ
Figure 7. Model KDR651 X-Ray
DR651 is given in
83
Page 84
Pacemaker Specifications

Emergency Parameter Settings

Listed below in Table 4 are the pacemaker’s emergency parameter settings.
Table 4. Emergency Settings
Mode VVI
Pacing Rate 70 ppm
Ventricular
Amplitude 7.5 V
Pulse Width 1.5 ms
Sensitivity 2.8 mV
Pacing Polarity Unipolar
Sensing Polarity Unipolar
Lead Monitor Monitor Only
Capture Management Off
Ventricular Refractory Period 330 ms
Single Chamber Hysteresis Off
84
Page 85
Pacemaker Specifications

Shipping and Nominal Parameter Settings

Listed below in Table 5 are the pacemaker’s shipping and nominal parameter settings.
Note: “Adaptive” indicates that the parameter value is adapted during operation. “Unchanged” indicates that the parameter setting is unchanged by nominal programming.
Table 5. Shipping and Nominal Settings
Parameter Shipping Medtronic
Mode DDDR
Mode Switch On On
Detect Rate 175 bpm 175 bpm
Detect Duration No Delay No Delay
Blanked Flutter Search On On
Lower Rate 60 ppm 60 ppm
Upper Tracking Rate 120 ppm 120 ppm
Upper Sensor Rate 120 ppm 120 ppm
ADL Rate 95 ppm 95 ppm
Rate Profile Optimization On
ADL Response 3-Mod. Active 3-Mod. Active
Exertion Response 3-Mod. Frequent 3-Mod. Frequent
Activity Threshold Med. Low Unchanged
Activity Acceleration 30 sec Unchanged
Activity Deceleration Exercise Unchanged
a
a
Nominal
DDDR
Unchanged
85
Page 86
Pacemaker Specifications
Tab le 5. Shipping and Nominal Settings (Continued)
Parameter Shipping Medtronic
Atrial Lead
Amplitude 3.5 V 3.5 V
Pulse Width 0.4 ms 0.4 ms
Sensitivity 0.5 mV (Adaptive
Sensing Assurance On
Pace Polarity Configure
Sense Polarity Configure
Lead Monitor Configure
a
a
or
c
Unipolar
Unipolar
a
or
c
d
or
Monitor Only
c
Notify if < 200 ohms Unchanged
Notify if > 4000 ohms 4000 ohms
Monitor Sensitivity 8 8
Nominal
a
) 0.5 mVb (Adaptive)
On
Unchanged
Unchanged
Unchanged
Ventricular Lead
Amplitude 3.5 V 3.5 V
Pulse Width 0.4 ms 0.4 ms
b
b
Sensitivity 2.8 mV (Adaptive) 2.8 mVb (Adaptive)
Sensing Assurance On On
Pace Polarity Configure
Unipolar
Sense Polarity Configure
Unipolar
Lead Monitor Configure
Monitor Only
or
c
a
or
c
d
or
c
Unchanged
Unchanged
Unchanged
a
Notify if < 200 ohms Unchanged
Notify if > 4000 ohms 4000 ohms
Monitor Sensitivity 8 8
86
Page 87
Pacemaker Specifications
Tab le 5. Shipping and Nominal Settings (Continued)
Parameter Shipping Medtronic
Capture Management Adaptive
Amplitude Margin 1.5x (times) 1.5x (times)
Pulse Width Margin 1.5x (times) 1.5x (times)
Minimum Adapted Amplitude
Minimum Adapted Pulse Width
Capture Test Frequency
Acute Phase Days Remaining
Paced AV (PAV) 150 ms 150 ms
Sensed AV (SAV) 120 ms 120 ms
RAAV Off Off
PVARP 310 ms 310 ms
Atrial Blanking (PVAB) 180 ms 180 ms
Ventricular Refractory Period
Ventricular Blanking (after atrial pace)
Sleep Off Off
Non-Competitive Atrial Pacing
Single Chamber Hysteresis
PMT Intervention Off Off
PVC Response On On
Ventricular Safety Pacing On On
2.5 V 2.5 V
0.4 ms 0.4 ms
Day at Rest Day at Rest
112 days Unchanged
230 ms 230 ms
28 ms 28 ms
On On
Off Unchanged
a
Nominal
Adaptive
87
Page 88
Pacemaker Specifications
Tab le 5. Shipping and Nominal Settings (Continued)
Parameter Shipping Medtronic
Implant Detection On/Restart Unchanged
Transtelephonic Monitor Off Unchanged
Extended Telemetry Off Unchanged
Extended Marker Standard Unchanged
FAST Indicators On Unchanged
Serial Number Factory Set Unchanged
Clinician-Selected
Custom Rate Trend Unchanged
Diagnostic
a
At the completion of the 30-minute Implant Detection period, Mode becomes DDDR; Rate Profile Optimization is enabled; Polarities are automatically configured for all bipolar models; Capture Management is enabled, and Ventricular Amplitude and Pulse Width become Adaptive; Sensing Assurance is enabled, and Sensitivity becomes Adaptive.
b
Value from which adaptive adjustment begins when nominals are programmed.
c
Unipolar-only Model KDR656
d
Models KDR651 and KDR653
Nominal
88
Page 89
Pacemaker Specifications

Electrical Reset Parameter Settings

Listed below in Table 6 are the pacemaker’s full and partial electrical reset parameter settings.
Note: “Unchanged” indicates that the parameter setting is unchanged by a reset.
Table 6. Electrical Reset Parameter Settings
Parameter Partial Electrical
Reset
Mode Unchanged VVI
Mode Switch Unchanged Off
Detect Rate 175 bpm
Detect Duration No Delay
Blanked Flutter Search Unchanged
Lower Rate Unchanged 65 ppm
Upper Tracking Rate Unchanged 120 ppm
Upper Sensor Rate Unchanged 120 ppm
ADL Rate Unchanged 95 ppm
Rate Profile Optimization Unchanged Off
ADL Response 3-Mod. Active
Exertion Response 3-Mod. Frequent
Activity Threshold Med. Low Med. Low
Activity Acceleration 30 sec 30 sec
Activity Deceleration Exercise Exercise
Full Electrical Reset
89
Page 90
Pacemaker Specifications
Table 6. Electrical Reset Parameter Settings (Continued)
Parameter Partial Electrical
Reset
Atrial Lead
Amplitude Unchanged 5.0 V
Pulse Width Unchanged 0.4 ms
Sensitivity Unchanged 0.5 mV
Sensing Assurance Unchanged Off
Pace Polarity Unchanged Configure
Sense Polarity Unchanged Configurea or
Lead Monitor Unchanged Configure or
Notify if < 200 ohms 200 ohms
Notify if > 4000 ohms 4000 ohms
Monitor Sensitivity 8 8
Ventricular Lead
Amplitude Unchanged 5.0 V
Pulse Width Unchanged 0.4 ms
Sensitivity Unchanged 2.8 mV
Sensing Assurance Unchanged Off
Pace Polarity Unchanged Configure
Sense Polarity Unchanged Configurea or
Lead Monitor Unchanged Configure or
Notify if < 200 ohms 200 ohms
Notify if > 4000 ohms 4000 ohms
Monitor Sensitivity 8 8
Full Electrical Reset
Unipolar
Unipolar
Monitor Only
Unipolar
Unipolar
Monitor Only
a
or
b
b
b
a
or
b
b
b
90
Page 91
Pacemaker Specifications
Table 6. Electrical Reset Parameter Settings (Continued)
Parameter Partial Electrical
Reset
Capture Management Unchanged Off
Amplitude Margin Unchanged
Pulse Width Amplitude Unchanged
Minimum Adapted
Unchanged
Amplitude
Minimum Adapted
Unchanged
Pulse Width
Capture Test
Day at Rest
Frequency
Acute Phase Days
112 Days
c
Remaining
Paced AV (PAV) 150 ms 150 ms
Sensed AV (SAV) 120 ms 120 ms
RAAV Unchanged Off
Start Rate 80 ppm
Stop Rate 120 ppm
Max Offset -40 ms
PVARP Unchanged 310 ms
Atrial Blanking (PVAB) 180 ms 180 ms
Ventricular Blanking 28 ms 28 ms
Ventricular Refractory
230 ms 230 ms
Period
Sleep Off Off
NCAP Unchanged Off
Single Chamber
Unchanged Off
Hysteresis
PMT Intervention Unchanged Off
PVC Response Unchanged On
Full Electrical Reset
d
91
Page 92
Pacemaker Specifications
Table 6. Electrical Reset Parameter Settings (Continued)
Parameter Partial Electrical
Reset
Ventricular Safety Pacing Unchanged On
Implant Detection Unchanged On/Restart
Transtelephonic Monitor Unchanged Off
Extended Telemetry Off Off
Extended Marker Standard Standard
FAST Indicators On On
Serial Number Unchanged Zeros
Full Electrical Reset
Clinician-Selected
Off Off
Diagnostic
a
Bipolar Models KDR651 and KDR653 revert to Implant Detection dur ing which polarity is automatically configured.
b
Unipolar-only Model KDR656
c
If a reset occurs after the completion of Acute Phase Days Remaining, the parameter will default to 42 days.
d
Sensor varied PVARP is disabled at a full electrical reset.
92
Page 93
Pacemaker Specifications

Elective Replacement Indicator

Listed below in Table 7 are the pacemaker’s Elective Replacement Indicators for nonmagnet and magnet modes.
Table 7. Elective Replacement Indicator Conditions
Nonmagnet Mode VVI mode at 65 ppm rate
Magnet Mode VOO mode at 65 ppm rate
Telemetry Replacement message on programmer
Battery/Lead Information Replacement message and displayed
battery voltage on programmer

Magnet Mode Conditions

Listed below in Table 8 are the pacemaker’s Magnet Mode Conditions in dual and single chamber modes.
Table 8. Magnet Mode Conditions
Dual Chamber Modes DOO mode at 85 ppm
VDD Mode VOO mode at 85 ppm
Single Chamber Modes VOO/AOO mode at 85 ppm
93
Page 94
Pacemaker Specifications

Longevity Projections

Listed below in Table 9 are the longevity projections from implant to the Elective Replacement Indicator (ERI) for the K
DR650 Series
pacemakers. The estimated time varies based on the lead impedance and the percent paced.
Table 9. Projected Longevity from Implant to ERI
Programmed Settings Percent
Mode: DDDR Lead Impedance
Lower Rate: 60 ppm 500
Nominal Outputs
Amplitudes: 3.5 V Pulse Widths: 0.4 ms
Low Outputs
Amplitudes: 2.5 V Pulse Widths: 0.4 ms
Lower Ventricular
Outputs
A. Amplitude: 2.5 V V. Amplitude 1.5 V Pulse Widths: 0.4 ms
Paced
100%
50%
100%
50%
100%
50%
Projected Longevity
(Years)
ohms
6.3
7.7
7.7
8.6
8.1
8.9
600
ohms
6.6
8.0
8.0
8.8
8.3
9.0
1000
ohms
7.7
8.8
8.6
9.2
8.8
9.4
Warning: When the Elective Replacement Indicator (ERI) is set, the pacemaker will continue to operate at ERI conditions for a minimum of three months until the battery depletes.
Note: The values above are based on calculations using deliverable battery capacity. These values are estimates of longevity projections from implant to ERI and are intended to assist the clinician in understanding the effects of various pacing conditions on battery
94
Page 95
Pacemaker Specifications
longevity. These values should not be interpreted as precise projections.
Listed below in Table 10 are the longevity projections from ERI to erratic pacing for the K
DR650 Series pacemakers. At most
programmed settings, virtually all (i.e., 99.9%) of these pacemakers will achieve a minimum of three months longevity after ERI is set.
Note: The average time in the table below refers to the projected mean longevity of pacemakers at the stated conditions.
Table 10. Longevity from ERI to Erratic Pacing
Previous Programmed Settings Before ERI
Lower Rate: 60 ppm Amplitudes: 3.5 V Pulse Widths: 0.4 ms
Mode: DDDR 100% 5.0 11.0
Mode: VVIR 100% 4.0 8.5
a
Programmed settings which lead to increased current demands in VVI, 65-ppm operating condition after ERI may slightly decrease average projections. Increased current demands post-ERI could be caused by an increase in the percent paced and/ or the increase in pacing rate.
b
Current data show that combinations of extreme pacing conditions (high output values [≥ 5.0 V] and high pulse widths, and /or low impedance values [300 ohms] may increase the average projection in the DDDR/DDD modes and decrease the average projection in the AAI/AAIR and VVI/VVIR modes.
Percent
Paced
Projected Longevity
(Months)
Lead Impedance
500 ohms
Minimum Average
a,b
95
Page 96
Pacemaker Specifications

Programmable Parameters

Listed below in Table 11 through Table 19 are the pacemaker’s programmable modes and parameters.
Table 11. Pacing Modes and Rates
Parameter Capability Notes
Pacing Mode DDDR, DDD, VDD, DDIR,
Mode Switch On, Off DDDR, DDD, VDD
Detect Rate 120, 125, 130, … 200 bpm
Detection Duration No Delay Nonprogrammable
Blanked Flutter Search
Lower Rate 30, 35, 40, … 175 ppm
Upper Tracking Rate 80, 90, 95, … 180 ppm DDDR, DDD, VDD
Upper Sensor Rate 80, 90, 95, … 180 ppm
DDI, DVIR, DVI, DOOR, DOO, VVIR, VDIR, VVI, VDI, VVT, VOOR, VOO, AAIR, ADIR, AAI, ADI, AAT, AOOR, AOO, ODO, OVO, OAO
On, Off
(except 65 and 85 ppm)
modes
modes
Rate responsive modes or DDD and VDD modes with mode switching
96
Page 97
Pacemaker Specifications
Table 12. Rate Response Parameters
Parameter Capability Notes
ADL Rate 60, 65, 70, … 180 ppm Rate responsive
modes or DDD and VDD modes with mode switching
Rate Profile Optimization
ADL Response 1-Inactive
Exertion Response 1-Infrequent
ADL Setpoint 5, 6, 7, … 40,
UR Setpoint 15, 16, 17, … 40,
Activity Threshold Low, Medium Low,
Activity Acceleration 15, 30, 60 sec
Activity Deceleration 2.5, 5.0, 10 min, Exercise
On, Off, Rate responsive
2-Less Active 3-Mod. Active 4-More Active 5-Very Active
2-Less Frequent 3-Mod. Frequent 4-More Frequent 5-Very Frequent
42, 44, 46, … 80
42, 44, 46, … 80, 85, 90, 95, … 180
Medium High, High
modes
Programmable from the Exercise test only
Programmable from the Exercise test only
97
Page 98
Pacemaker Specifications
Table 13. Atrial Outputs, Sensitivity, and Polarity
Parameter Capability Notes
Amplitude
a
0.5, 0.75, 1.0 … 4.0 V
4.5, 5.0, 5.5, 6.0 V, 7.5 V
Pulse Width 0.12, 0.15, 0.21, 0.27,
0.34, 0.40, 0.46, 0.52,
0.64, 0.76, 1.00, 1.25,
1.50 ms
Sensitivity 0.18, 0.25, 0.35, 0.5, 0.7,
1.0, 1.4, 2.0, 2.8, 4.0 mV
0.18, 0.25, 0.35 mV atrial bipolar only
Sensing Assurance On, Off
Pacing Polarity Bipolar, Unipolar,
Configure
DR656 is
Model K unipolar only. Configure is displayed but is not selectable.
Sensing Polarity Bipolar, Unipolar,
Configure
DR656 is
Model K unipolar only. Configure is displayed but is not selectable.
Lead Monitor Configure, Monitor Only,
Adaptive, Off
DR656 only
Model K operates in Monitor Only
Notify if < 200 ohms Nonprogrammable
Notify if > 1000, 2000, 3000,
4000 ohms
Monitor Sensitivity 2, 3,4, … 16
a
Tolerance for amplitudes from 0.5 V through 6.0 V is ± 10%, and for 7.5 V is
-20/+0%. Tolerances are based on 37°C and a 500-ohm load.
98
Page 99
Pacemaker Specifications
Table 14. Ventricular Outputs, Sensitivity, and Polarity
Parameter Capability Notes
Amplitude
a
0.5, 0.75, 1.0 … 4.0 V
4.5, 5.0, 5.5, 6.0 V, 7.5 V
0.625, 0.875, 1.125,
1.375, 1.625, and
1.875 V can be set by Capture Management. Values are displayed, but not selectable.
Pulse Width 0.12, 0.15, 0.21, 0.27,
0.34, 0.40, 0.46, 0.52,
0.64, 0.76, 1.00, 1.25,
1.50 ms
Sensitivity 1.0, 1.4, 2.0, 2.8, 4.0, 5.6,
8.0, 11.2 mV
Sensing Assurance On, Off
Pacing Polarity Bipolar, Unipolar,
Configure
DR656 is
Model K unipolar only. Configure is displayed but is not selectable.
Sensing Polarity Bipolar, Unipolar,
Configure
DR656 is
Model K unipolar only. Configure is displayed but is not selectable.
Lead Monitor Configure, Monitor Only,
Adaptive, Off
DR656 only
Model K operates in Monitor Only
Notify if < 200 ohms Nonprogrammable
Notify if > 1000, 2000, 3000,
4000 ohms
Monitor Sensitivity 2, 3,4, … 16
a
Tolerance for amplitudes from 0.5 V through 6.0 V is ± 10%, and for 7.5 V is
-20/+0%. Tolerances are based on 37°C and a 500-ohm load.
99
Page 100
Pacemaker Specifications
Table 15. Capture Management
Parameter Capability Notes
Ventricular Capture
Off, Monitor Only
Management
Amplitude Margin 1x, 1.25x, 1.5x, 2x,
2.5x (times)
Pulse Width Margin 1x, 1.5x, 2x, 2.5x, 3x,
4x, 5x (times)
Minimum Adapted Amplitude
Minimum Adapted Pulse Width
0.5, 0.625, 0.75, 1.0 … 2.0 V
2.25, 2.50 ... 4.0 V, 4.5, 5.0 V
0.12, 0.15, 0.21, 0.27, 0.34,
0.40, 0.46, 0.52, 0.64, 0.76,
1.00 ms
Capture Test Frequency
Test Capture Every 15, 30 Minutes
1, 2, 4, 8, 12 hours, Day at Rest, Day at …, 7, 14, 28, 42 Days at ...
At 12:00 am, 12:15 am,
12:30 am, … 11:45 pm
Retry if Test
Ye s , No
Delayed?
Acute Phase Days Remaining
a
When programmed to Monitor Only, the subparameters shown in the table above are programmable for use by the pacemaker in providing suggested output values.
Off, 7, 14, 21, … 84 days 112, 140, 168, … 252 days
a
, Adaptive
Applicable only for Day(s) at … parameter
100
Loading...