Medtronic KDR603 Errata

Software Errata Sheet for the Kappa 700, Model 9953 Software Version 7.0 for the Medtronic 2090 CareLink Programmer:
The following table lists certain conditions that may occur during use of the Model 9953 Software Version 7.0. The table also provides the Resolution for these conditions.
Condition Observed Behavior Suggested Corrective Action
When programmed to a dual chamber mode with Auto PVARP On, select a single chamber atrial mode (Note: ARP value is set pending to a default value of 400 ms).
Select a new parameter value for ARP.
When Rate Drop Response is programmed to On (only valid for DDD and DDI mode), perform Emergency programming.
The following are trademarks of Medtronic: Medtronic, Medtronic CareLink, Kappa, Vitatron
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The pending ARP value remains 400 ms, even if another ARP value is selected.
After Emergency programming has gone into effect, interlock symbols will appear on the screen with a non-pertinent parameter.
Program Mode, which also programs the ARP to the default value.
Select a new ARP value and perform another programming.
Select interrogate button to reinterrogate the device and refresh the display.
© Medtronic, Inc. 2007 M931845A001A 2007-09-20