Product information about Medtronic.Kappa 700/600 Series
software and pacemakers is presented in two separate guides.
The Pacemaker Reference Guide (PRG) is a supplementary guide
that provides detailed information on Medtronic.Kappa 700/600
Series pacemakers.
The Pacemaker Programming Guide (PPG) accompanies
Medtronic.Kappa 700/600 Series software and contains
instructions on how to use the programmer and the programming
software.
About this Guide
This supplementary guide describes in detail, how the pacemaker
operates and specifies the capabilities of each model.
How to Use This Guide
■
Describes the pacing modes, rate response options, special
therapy features, telemetry types, and data collection options.
In some cases, guidelines are given on how to configure the
pacemaker operation.
■
Contains troubleshooting information for electrical and
hemodynamic problems.
■
Specifies parameter and data collection capabilities, longevity
projections, and mechanical and electrical specifications.
■
Provides general warnings and cautions, potential
interference sources, and general indications for pacing.
■
Contains a glossary of terms.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
iii
How to Use This Guide
About the Pacemaker Programming Guide
This guide presents the following information to use the 9790
programmer.
■
How to setup and configure the programmer and access
on-line help.
■
How to start a patient session, use the various follow-up
features during the session, and properly end the session.
■
How to use checklist to streamline a follow-up session.
■
How to view and print the patient’s ECG and EGM waveform
traces.
■
How to configure the pacemaker to collect diagnostic data
and how to retrieve and view this information.
■
How to measure stimulation thresholds and sensing levels.
■
How to program parameter values and verify rate response
parameters settings.
iv
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Special Notice F-2
Warnings F-4
Precautions F-6
Potential Complications F-11
G. Environmental Interference
Hospital or Medical Environment Interference G-2
Home and Job Environment Interference G-6
H. Glossary
I. Index
viii
9
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Understanding
Pacemaker Operation
Chapters 1 - 9 provide detailed
information about the operation of
the Kappa 700/600 Series
pacemakers.
Pacing Modes
Rate Response
Pacemaker Timing
Lead/Cardiac Tissue Interface
Special Therapy Options
Telemetry Data
Miscellaneous Operations
Diagnostics
Troubleshooting The
Pacing System
Pacing Modes
This chapter provides information about the modes
available with the pacemaker.
1
1
Introduction
Mode Selection
Decision Tree:
Mode Pertinency Tables
Indications and Usage
Contraindications
DDDR Mode
DDD Mode
DDIR Mode
DDI Mode
DVIR Mode
DVI Mode
1-2
1-4
1-8
1-9
1-10
1-11
1-12
1-13
1-14
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-7
1-5
VDD Mode
AAIR / ADIR Modes
AAI / ADI Modes
VVIR / VDIR Modes
VVI / VDI Modes
AAT / VVT Modes
DOOR / AOOR / VOOR
Modes
DOO / AOO / VOO Modes
ODO / OAO / OVO Modes
1-15
1-16
1-17
1-18
1-19
1-20
1-21
1-22
1-23
1-1
10
Pacing Modes
Introduction
Introduction
Pacing Mode Selection
This chapter provides an introduction to pacemaker modes as an
aid to pacing mode selection. The chapter is organized as follows:
Mode Selection Decision Tree – This decision tree, based on the
1991 ACC/AHA guidelines for pacemaker implantation,
1
provides a simple means of identifying pacing modes appropriate
for given indications.
Mode Pertinency Tables – These tables show which features and
parameters apply to each commonly used pacing mode.
Mode Descriptions – These descriptions provide indications and
contraindications for modes available with the pacemaker and
brief descriptions of how these modes operate.
NBG Pacing Codes
The pacemaker modes are defined in NBG Code.2 Each five-letter
NBG code describes a specific type of operation for implantable
pacemakers. For simplicity, this manual uses only the first three or
four letters, such as DDD, DDIR, DVIR, and so forth. Figure 1-1
describes the first four letters of the NBG code.
11
1-2
1
Dreifus LS, Fisch C, Griffin JC, et al. Guidelines for implantation of cardiac
pacemakers and antiarrhythmia devices. A report of the American College of
Cardiology/American Heart Association Task Force on Assessment of
Diagnostic and Therapeutic Cardiovascular Procedures (Committee on
Pacemaker Implantation).
18: 1-13.
2
Bernstein A., et al., “The NASPE/BPEG Pacemaker Code,” PACE, 10(4), Jul-Aug
1987. (“NBG” stands for The North American Society of Pacing and
Electrophysiology [NASPE] and the British Pacing and Electrophysiology
Group [BPEG] Generic. NBG’s five-letter code supersedes the ICHD Code.
Journal of the American College of Cardiology.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1991;
Pacing Modes
Introduction
CHAMBER PACED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
CHAMBER SENSED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
Figure 1-1.
DDDR
NBG Pacing Codes
MODE OF RESPONSE
T = Triggered
I = Inhibited
D = Double (Both)
O = None
PROGRAMMABLE/
RATE RESPONSE
P = Programmable
M = Multiprogrammable
C = Communicating
R = Rate Responsive
O = None
Further Information
The mode descriptions in this chapter provide only a basic
overview of each mode. For further details on the rate response,
timing, and therapy capabilities refer to “Rate Response” on
page 2-1, “Pacemaker Timing” on page 3-1, and “Special Therapy
Options” on page 5-1, respectively.
12
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-3
Pacing Modes
Mode Selection Decision Tree:
Mode Selection Decision Tree:
Figure 1-2 shows a basic decision tree used to select the pacing mode. In the shaded boxes the
preferred mode(s) is listed and the alternate mode(s) appears below the dashed line.
Symptomatic
Bradycardia
(e.g., persistent
atrial fibrillation,
inexcitable atrium)
Is SA node conduction
presently adequate?
Ye sNoNo
AAI
DDD
AAIR
DDDR
VVIR
VVI
No
Ye s
Can the atrium be sensed
and/or paced reliably?
Is AV conduction
presently adequate?
AAIR
DDDR
Figure 1-2.
Ye s
DDD
DDDR
VDD
Mode Selection Tree
No
(e.g., complete or
transient AV block)
Is SA node conduction
presently adequate?
Ye s
DDDR
DDIR
13
1-4
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacing Modes
Mode Pertinency Tables
black
✓
es.
e
✓✓
✓✓✓ ✓✓
✓✓✓✓ ✓
✓✓✓✓
. Automatic PVARP is available in the DDDR, DDD, and VDD mod
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
b
See Chapter 5 for operational descriptions of special therapy options. Search AV is described in Chapter 3.
Sinus Preference ✓
Sleep Function✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓
Rate Response
ADL Rate✓–✓–✓––✓✓–––✓✓–––
Rate Profile Optimization✓–✓–✓––✓✓–––✓✓–––
Activity Threshold✓–✓–✓––✓✓–––✓✓–––
Activity Acceleration✓–✓–✓––✓✓–––✓✓–––
See Chapter 2 for operational descriptions of rate response features.
Activity Deceleration✓–✓–✓––✓✓ –––✓✓–––
a
b
Indications and Usage
Medtronic.Kappa 700/600 Series pacemakers are indicated for the
following:
■
Rate adaptive pacing in patients who may benefit from
increased pacing rates concurrent with increases in activity.
■
Accepted patient conditions warranting chronic cardiac
pacing which include:
–Symptomatic paroxysmal or permanent second or third-
degree AV block.
–Symptomatic bilateral bundle branch block.
–Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders.
–Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
Pacing Modes
Indications and Usage
16
–Vasovagal syndromes or hypersensitive carotid sinus
syndromes.
Medtronic.Kappa 700/600 Series pacemakers are also indicated
for dual chamber and atrial tracking modes in patients who may
benefit from maintenance of AV synchrony. Dual chamber modes
are specifically indicated for treatment of conduction disorders
that require restoration of both rate and AV synchrony, which
include:
■
Various degrees of AV block to maintain the atrial
contribution to cardiac output.
■
VVI intolerance (e.g., pacemaker syndrome) in the presence of
persistent sinus rhythm.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-7
Pacing Modes
Contraindications
Contraindications
Medtronic.Kappa 700/600 Series pacemakers are contraindicated
for the following applications:
■
Dual chamber atrial pacing in patients with chronic refractory
atrial tachyarrhythmias.
■
Asynchronous pacing in the presence (or likelihood) of
competitive paced and intrinsic rhythms.
■
Unipolar pacing for patients with an implanted cardioverterdefibrillator (ICD) because it may cause unwanted delivery or
inhibition of ICD therapy.
17
1-8
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
DDDR Mode
Pacing Modes
DDDR Mode
In the DDDR mode, the pacemaker tracks the faster of the intrinsic
atrial rate or the sensor-indicated rate. If the intrinsic rate is faster,
the DDDR mode provides atrial synchronous pacing; otherwise,
AV sequential pacing occurs at the sensor-indicated rate.
■
Rate limits for atrial tracking (Upper Tracking Rate)1 and
sensor tracking (Upper Sensor Rate) are separately
programmable.
■
The AV intervals that follow sensed atrial events (SAV) and
paced atrial events (PAV) are separately programmable, and
they can be programmed to shorten with increasing rates
(Rate Adaptive AV) or to change with intrinsic conduction
times (Search AV).
■
A nonrefractory sensed event in either chamber inhibits
pacing in that chamber. A ventricular nonrefractory sensed
event in the VA interval that is not preceded by an atrial sense
(AS or AR) is a pacemaker-defined PVC, and starts a new VA
interval.
18
Sensor-indicated
Interval
A
P
V
P
Parameters:
Lower Rate = 60 ppm (1000 ms)PAV Interval = 200 msPVARP = 280 ms
Sensor-indicated Rate = 90 ppm (667 ms) SAV Interval = 170 ms
A
P
V
S
Figure 1-3.
1
Example of DDDR Mode Operation
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
Sensor-indicated
Interval
A
S
V
S
A
P
V
P
value.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
S
V
P
200 ms
1-9
Pacing Modes
DDD Mode
DDD Mode
The DDD mode provides atrial synchronous pacing in the
presence of intrinsic atrial activity; otherwise, AV sequential
pacing occurs at the Lower Rate.
■
Each atrial paced or nonrefractory atrial sensed event starts an
AV interval and a lower rate interval. The AV intervals that
follow sensed atrial events (SAV) and paced atrial events
(PAV) are separately programmable, and the SAV may be
optionally programmed to shorten with increasing rate (Rate
Adaptive AV) or to change with intrinsic conduction times
(Search AV).
■
A ventricular paced events may track atrial sensed events up
to the programmed Upper Tracking Rate.
■
A ventricular nonrefractory sensed event in the VA interval
1
that is not preceded by an atrial sense (AS or AR) is a
pacemaker-defined PVC, and starts a new VA interval.
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
A
S
V
S
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Lower Rate Interval
200 ms
A
P
DDIR Mode
Pacing Modes
DDIR Mode
The DDIR mode provides dual chamber, sensor-driven,
atrioventricular (AV) sequential pacing for heart rate variation
without atrial tracking.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval.
That is, ventricular paced events after such sensed atrial
events occur at the sensor-indicated rate. The following
ventriculoatrial (VA) interval may be extended slightly to
avoid an increasing atrial paced rate.
■
A ventricular nonrefractory sensed event in the VA interval
starts a new VA interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Sensor-indicated
VA Interval
V
P
Sensor-indicated
A
A
P
P
V
P
Interval
200 ms
A
P
1-11
20
Pacing Modes
DDI Mode
DDI Mode
Lower Rate Interval
The DDI mode provides dual chamber atrioventricular (AV)
sequential pacing with atrial sensing but without atrial tracking.
■
Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval.
Ventricular paced events after such sensed atrial events occur
at the Lower Rate.
■
A ventricular nonrefractory sensed event in the
ventriculoatrial (VA) interval starts a new VA interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
S
V
P
V
P
Example of DDI Mode Operation
200 ms
A
P
DVIR Mode
Pacing Modes
DVIR Mode
The DVIR mode provides AV sequential pacing at the sensorindicated rate unless inhibited by ventricular sensed events.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
The DVIR mode ignores intrinsic atrial events. Sensing occurs
only in the ventricle. A ventricular nonrefractory sensed event
during the ventriculoatrial (VA) interval starts a new VA
interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-13
Pacing Modes
DVI Mode
DVI Mode
The DVI mode provides dual chamber AV sequential pacing
without atrial sensing/tracking.
■
Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited.
■
Sensing occurs only in the ventricle, and intrinsic atrial events
are ignored. A ventricular nonrefractory sensed event during
the VA interval starts a new ventriculoatrial (VA) interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
VDD Mode
Pacing Modes
VDD Mode
The VDD mode provides atrial synchronous pacing (or VVI
pacing at the Lower Rate). The ventricle is paced synchronously
up to the programmed Upper Tracking Rate.
1
Sensing occurs in
both the atrium and ventricle, but pacing occurs only in the
ventricle.
■
To promote atrial synchronous pacing at slow rates, a sensed
atrial event occurring near the end of the Lower Rate interval
will be followed by the programmed maximum SAV interval.
The result is an extension of the ventricular lower rate.
■
A ventricular nonrefractory sensed event in the V-V interval
that is not preceded by an atrial sense (AS or AR) is a
pacemaker-defined PVC, and it starts a new V-V interval.
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
Example of VDD Operation
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
S
V
P
A
S
200 ms
1-15
Pacing Modes
AAIR / ADIR Modes
AAIR / ADIR Modes
The AAIR mode provides atrial-based rate responsive pacing in
patients with intact AV conduction. Sensing and pacing occur
only in the atrium. In the absence of sensed events, the chamber is
paced at the sensor-indicated rate.
The ADIR mode operates the same as the AAIR mode except that
events sensed in the ventricle are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the conducted
ventricular rhythm without affecting atrial pacing.
In the AAIR and ADIR modes, atrial refractory sensed
Note:
events do not restart the Upper Sensor Rate interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
P
200 ms
AAI / ADI Modes
The AAI mode provides single chamber inhibited atrial pacing.
Sensing and pacing occur only in the atrium. Pacing occurs at the
programmed Pacing Rate unless inhibited by sensed events.
The ADI mode operates the same as the AAI mode except that
events sensed in the ventricle are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the conducted
ventricular rhythm without affecting atrial pacing.
Pacing Modes
AAI / ADI Modes
Pacing Rate Interval
A
P
Parameters:
Pacing Rate = 75ppm (800 ms)
A
R
Figure 1-11.
Pacing Rate Interval
A
P
A
S
Example of AAI Mode Operation
8006004002001000
A
P
200 ms
26
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-17
Pacing Modes
VVIR / VDIR Modes
VVIR / VDIR Modes
The VVIR mode provides ventricular rate responsive pacing in
patients for whom atrial-based pacing is deemed unnecessary or
inappropriate. In the absence of sensed events, the ventricle is
paced at the sensor-indicated rate.
The VDIR mode operates the same as the VVIR mode except that
events sensed in the atrium are recorded by the diagnostics. When
used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the
underlying atrial rhythm without affecting ventricular pacing.
Note:
events restart the Upper Sensor Rate interval.
In the VVIR and VDIR modes, ventricular refractory sensed