Product information about Medtronic.Kappa 700/600 Series
software and pacemakers is presented in two separate guides.
The Pacemaker Reference Guide (PRG) is a supplementary guide
that provides detailed information on Medtronic.Kappa 700/600
Series pacemakers.
The Pacemaker Programming Guide (PPG) accompanies
Medtronic.Kappa 700/600 Series software and contains
instructions on how to use the programmer and the programming
software.
About this Guide
This supplementary guide describes in detail, how the pacemaker
operates and specifies the capabilities of each model.
How to Use This Guide
■
Describes the pacing modes, rate response options, special
therapy features, telemetry types, and data collection options.
In some cases, guidelines are given on how to configure the
pacemaker operation.
■
Contains troubleshooting information for electrical and
hemodynamic problems.
■
Specifies parameter and data collection capabilities, longevity
projections, and mechanical and electrical specifications.
■
Provides general warnings and cautions, potential
interference sources, and general indications for pacing.
■
Contains a glossary of terms.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
iii
How to Use This Guide
About the Pacemaker Programming Guide
This guide presents the following information to use the 9790
programmer.
■
How to setup and configure the programmer and access
on-line help.
■
How to start a patient session, use the various follow-up
features during the session, and properly end the session.
■
How to use checklist to streamline a follow-up session.
■
How to view and print the patient’s ECG and EGM waveform
traces.
■
How to configure the pacemaker to collect diagnostic data
and how to retrieve and view this information.
■
How to measure stimulation thresholds and sensing levels.
■
How to program parameter values and verify rate response
parameters settings.
iv
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Special Notice F-2
Warnings F-4
Precautions F-6
Potential Complications F-11
G. Environmental Interference
Hospital or Medical Environment Interference G-2
Home and Job Environment Interference G-6
H. Glossary
I. Index
viii
9
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Understanding
Pacemaker Operation
Chapters 1 - 9 provide detailed
information about the operation of
the Kappa 700/600 Series
pacemakers.
Pacing Modes
Rate Response
Pacemaker Timing
Lead/Cardiac Tissue Interface
Special Therapy Options
Telemetry Data
Miscellaneous Operations
Diagnostics
Troubleshooting The
Pacing System
Pacing Modes
This chapter provides information about the modes
available with the pacemaker.
1
1
Introduction
Mode Selection
Decision Tree:
Mode Pertinency Tables
Indications and Usage
Contraindications
DDDR Mode
DDD Mode
DDIR Mode
DDI Mode
DVIR Mode
DVI Mode
1-2
1-4
1-8
1-9
1-10
1-11
1-12
1-13
1-14
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-7
1-5
VDD Mode
AAIR / ADIR Modes
AAI / ADI Modes
VVIR / VDIR Modes
VVI / VDI Modes
AAT / VVT Modes
DOOR / AOOR / VOOR
Modes
DOO / AOO / VOO Modes
ODO / OAO / OVO Modes
1-15
1-16
1-17
1-18
1-19
1-20
1-21
1-22
1-23
1-1
10
Pacing Modes
Introduction
Introduction
Pacing Mode Selection
This chapter provides an introduction to pacemaker modes as an
aid to pacing mode selection. The chapter is organized as follows:
Mode Selection Decision Tree – This decision tree, based on the
1991 ACC/AHA guidelines for pacemaker implantation,
1
provides a simple means of identifying pacing modes appropriate
for given indications.
Mode Pertinency Tables – These tables show which features and
parameters apply to each commonly used pacing mode.
Mode Descriptions – These descriptions provide indications and
contraindications for modes available with the pacemaker and
brief descriptions of how these modes operate.
NBG Pacing Codes
The pacemaker modes are defined in NBG Code.2 Each five-letter
NBG code describes a specific type of operation for implantable
pacemakers. For simplicity, this manual uses only the first three or
four letters, such as DDD, DDIR, DVIR, and so forth. Figure 1-1
describes the first four letters of the NBG code.
11
1-2
1
Dreifus LS, Fisch C, Griffin JC, et al. Guidelines for implantation of cardiac
pacemakers and antiarrhythmia devices. A report of the American College of
Cardiology/American Heart Association Task Force on Assessment of
Diagnostic and Therapeutic Cardiovascular Procedures (Committee on
Pacemaker Implantation).
18: 1-13.
2
Bernstein A., et al., “The NASPE/BPEG Pacemaker Code,” PACE, 10(4), Jul-Aug
1987. (“NBG” stands for The North American Society of Pacing and
Electrophysiology [NASPE] and the British Pacing and Electrophysiology
Group [BPEG] Generic. NBG’s five-letter code supersedes the ICHD Code.
Journal of the American College of Cardiology.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1991;
Pacing Modes
Introduction
CHAMBER PACED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
CHAMBER SENSED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
Figure 1-1.
DDDR
NBG Pacing Codes
MODE OF RESPONSE
T = Triggered
I = Inhibited
D = Double (Both)
O = None
PROGRAMMABLE/
RATE RESPONSE
P = Programmable
M = Multiprogrammable
C = Communicating
R = Rate Responsive
O = None
Further Information
The mode descriptions in this chapter provide only a basic
overview of each mode. For further details on the rate response,
timing, and therapy capabilities refer to “Rate Response” on
page 2-1, “Pacemaker Timing” on page 3-1, and “Special Therapy
Options” on page 5-1, respectively.
12
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-3
Pacing Modes
Mode Selection Decision Tree:
Mode Selection Decision Tree:
Figure 1-2 shows a basic decision tree used to select the pacing mode. In the shaded boxes the
preferred mode(s) is listed and the alternate mode(s) appears below the dashed line.
Symptomatic
Bradycardia
(e.g., persistent
atrial fibrillation,
inexcitable atrium)
Is SA node conduction
presently adequate?
Ye sNoNo
AAI
DDD
AAIR
DDDR
VVIR
VVI
No
Ye s
Can the atrium be sensed
and/or paced reliably?
Is AV conduction
presently adequate?
AAIR
DDDR
Figure 1-2.
Ye s
DDD
DDDR
VDD
Mode Selection Tree
No
(e.g., complete or
transient AV block)
Is SA node conduction
presently adequate?
Ye s
DDDR
DDIR
13
1-4
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacing Modes
Mode Pertinency Tables
black
✓
es.
e
✓✓
✓✓✓ ✓✓
✓✓✓✓ ✓
✓✓✓✓
. Automatic PVARP is available in the DDDR, DDD, and VDD mod
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
b
See Chapter 5 for operational descriptions of special therapy options. Search AV is described in Chapter 3.
Sinus Preference ✓
Sleep Function✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓
Rate Response
ADL Rate✓–✓–✓––✓✓–––✓✓–––
Rate Profile Optimization✓–✓–✓––✓✓–––✓✓–––
Activity Threshold✓–✓–✓––✓✓–––✓✓–––
Activity Acceleration✓–✓–✓––✓✓–––✓✓–––
See Chapter 2 for operational descriptions of rate response features.
Activity Deceleration✓–✓–✓––✓✓ –––✓✓–––
a
b
Indications and Usage
Medtronic.Kappa 700/600 Series pacemakers are indicated for the
following:
■
Rate adaptive pacing in patients who may benefit from
increased pacing rates concurrent with increases in activity.
■
Accepted patient conditions warranting chronic cardiac
pacing which include:
–Symptomatic paroxysmal or permanent second or third-
degree AV block.
–Symptomatic bilateral bundle branch block.
–Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders.
–Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
Pacing Modes
Indications and Usage
16
–Vasovagal syndromes or hypersensitive carotid sinus
syndromes.
Medtronic.Kappa 700/600 Series pacemakers are also indicated
for dual chamber and atrial tracking modes in patients who may
benefit from maintenance of AV synchrony. Dual chamber modes
are specifically indicated for treatment of conduction disorders
that require restoration of both rate and AV synchrony, which
include:
■
Various degrees of AV block to maintain the atrial
contribution to cardiac output.
■
VVI intolerance (e.g., pacemaker syndrome) in the presence of
persistent sinus rhythm.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-7
Pacing Modes
Contraindications
Contraindications
Medtronic.Kappa 700/600 Series pacemakers are contraindicated
for the following applications:
■
Dual chamber atrial pacing in patients with chronic refractory
atrial tachyarrhythmias.
■
Asynchronous pacing in the presence (or likelihood) of
competitive paced and intrinsic rhythms.
■
Unipolar pacing for patients with an implanted cardioverterdefibrillator (ICD) because it may cause unwanted delivery or
inhibition of ICD therapy.
17
1-8
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
DDDR Mode
Pacing Modes
DDDR Mode
In the DDDR mode, the pacemaker tracks the faster of the intrinsic
atrial rate or the sensor-indicated rate. If the intrinsic rate is faster,
the DDDR mode provides atrial synchronous pacing; otherwise,
AV sequential pacing occurs at the sensor-indicated rate.
■
Rate limits for atrial tracking (Upper Tracking Rate)1 and
sensor tracking (Upper Sensor Rate) are separately
programmable.
■
The AV intervals that follow sensed atrial events (SAV) and
paced atrial events (PAV) are separately programmable, and
they can be programmed to shorten with increasing rates
(Rate Adaptive AV) or to change with intrinsic conduction
times (Search AV).
■
A nonrefractory sensed event in either chamber inhibits
pacing in that chamber. A ventricular nonrefractory sensed
event in the VA interval that is not preceded by an atrial sense
(AS or AR) is a pacemaker-defined PVC, and starts a new VA
interval.
18
Sensor-indicated
Interval
A
P
V
P
Parameters:
Lower Rate = 60 ppm (1000 ms)PAV Interval = 200 msPVARP = 280 ms
Sensor-indicated Rate = 90 ppm (667 ms) SAV Interval = 170 ms
A
P
V
S
Figure 1-3.
1
Example of DDDR Mode Operation
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
Sensor-indicated
Interval
A
S
V
S
A
P
V
P
value.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
S
V
P
200 ms
1-9
Pacing Modes
DDD Mode
DDD Mode
The DDD mode provides atrial synchronous pacing in the
presence of intrinsic atrial activity; otherwise, AV sequential
pacing occurs at the Lower Rate.
■
Each atrial paced or nonrefractory atrial sensed event starts an
AV interval and a lower rate interval. The AV intervals that
follow sensed atrial events (SAV) and paced atrial events
(PAV) are separately programmable, and the SAV may be
optionally programmed to shorten with increasing rate (Rate
Adaptive AV) or to change with intrinsic conduction times
(Search AV).
■
A ventricular paced events may track atrial sensed events up
to the programmed Upper Tracking Rate.
■
A ventricular nonrefractory sensed event in the VA interval
1
that is not preceded by an atrial sense (AS or AR) is a
pacemaker-defined PVC, and starts a new VA interval.
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
A
S
V
S
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Lower Rate Interval
200 ms
A
P
DDIR Mode
Pacing Modes
DDIR Mode
The DDIR mode provides dual chamber, sensor-driven,
atrioventricular (AV) sequential pacing for heart rate variation
without atrial tracking.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval.
That is, ventricular paced events after such sensed atrial
events occur at the sensor-indicated rate. The following
ventriculoatrial (VA) interval may be extended slightly to
avoid an increasing atrial paced rate.
■
A ventricular nonrefractory sensed event in the VA interval
starts a new VA interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Sensor-indicated
VA Interval
V
P
Sensor-indicated
A
A
P
P
V
P
Interval
200 ms
A
P
1-11
20
Pacing Modes
DDI Mode
DDI Mode
Lower Rate Interval
The DDI mode provides dual chamber atrioventricular (AV)
sequential pacing with atrial sensing but without atrial tracking.
■
Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval.
Ventricular paced events after such sensed atrial events occur
at the Lower Rate.
■
A ventricular nonrefractory sensed event in the
ventriculoatrial (VA) interval starts a new VA interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
S
V
P
V
P
Example of DDI Mode Operation
200 ms
A
P
DVIR Mode
Pacing Modes
DVIR Mode
The DVIR mode provides AV sequential pacing at the sensorindicated rate unless inhibited by ventricular sensed events.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
The DVIR mode ignores intrinsic atrial events. Sensing occurs
only in the ventricle. A ventricular nonrefractory sensed event
during the ventriculoatrial (VA) interval starts a new VA
interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-13
Pacing Modes
DVI Mode
DVI Mode
The DVI mode provides dual chamber AV sequential pacing
without atrial sensing/tracking.
■
Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited.
■
Sensing occurs only in the ventricle, and intrinsic atrial events
are ignored. A ventricular nonrefractory sensed event during
the VA interval starts a new ventriculoatrial (VA) interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
VDD Mode
Pacing Modes
VDD Mode
The VDD mode provides atrial synchronous pacing (or VVI
pacing at the Lower Rate). The ventricle is paced synchronously
up to the programmed Upper Tracking Rate.
1
Sensing occurs in
both the atrium and ventricle, but pacing occurs only in the
ventricle.
■
To promote atrial synchronous pacing at slow rates, a sensed
atrial event occurring near the end of the Lower Rate interval
will be followed by the programmed maximum SAV interval.
The result is an extension of the ventricular lower rate.
■
A ventricular nonrefractory sensed event in the V-V interval
that is not preceded by an atrial sense (AS or AR) is a
pacemaker-defined PVC, and it starts a new V-V interval.
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
Example of VDD Operation
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
S
V
P
A
S
200 ms
1-15
Pacing Modes
AAIR / ADIR Modes
AAIR / ADIR Modes
The AAIR mode provides atrial-based rate responsive pacing in
patients with intact AV conduction. Sensing and pacing occur
only in the atrium. In the absence of sensed events, the chamber is
paced at the sensor-indicated rate.
The ADIR mode operates the same as the AAIR mode except that
events sensed in the ventricle are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the conducted
ventricular rhythm without affecting atrial pacing.
In the AAIR and ADIR modes, atrial refractory sensed
Note:
events do not restart the Upper Sensor Rate interval.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
A
P
200 ms
AAI / ADI Modes
The AAI mode provides single chamber inhibited atrial pacing.
Sensing and pacing occur only in the atrium. Pacing occurs at the
programmed Pacing Rate unless inhibited by sensed events.
The ADI mode operates the same as the AAI mode except that
events sensed in the ventricle are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the conducted
ventricular rhythm without affecting atrial pacing.
Pacing Modes
AAI / ADI Modes
Pacing Rate Interval
A
P
Parameters:
Pacing Rate = 75ppm (800 ms)
A
R
Figure 1-11.
Pacing Rate Interval
A
P
A
S
Example of AAI Mode Operation
8006004002001000
A
P
200 ms
26
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-17
Pacing Modes
VVIR / VDIR Modes
VVIR / VDIR Modes
The VVIR mode provides ventricular rate responsive pacing in
patients for whom atrial-based pacing is deemed unnecessary or
inappropriate. In the absence of sensed events, the ventricle is
paced at the sensor-indicated rate.
The VDIR mode operates the same as the VVIR mode except that
events sensed in the atrium are recorded by the diagnostics. When
used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the
underlying atrial rhythm without affecting ventricular pacing.
Note:
events restart the Upper Sensor Rate interval.
In the VVIR and VDIR modes, ventricular refractory sensed
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
VVI / VDI Modes
The VVI mode provides single chamber inhibited pacing at the
programmed Pacing Rate unless inhibited by sensed events.
Sensing occurs only in the ventricle.
The VDI mode operates the same as the VVI mode except that
events sensed in the atrium are recorded by the diagnostics. When
used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the
underlying atrial rhythm without affecting ventricular pacing.
Pacing Modes
VVI / VDI Modes
Pacing Rate Interval
V
P
Parameters:
Pacing Rate = 60 ppm (1000 ms)
Ventricular Refractory Period = 300 ms
Figure 1-13.
Pacing Rate Interval
V
P
V
S
Example of VVI Mode Operation
200 ms
V
P
28
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-19
Pacing Modes
AAT / VVT Modes
AAT / VVT Modes
Pacing occurs at the programmed Pacing Rate, but a nonrefractory
sensed event triggers an immediate pacing output (rather than
inhibiting such output). Except that pacing outputs occur when
events are sensed, the triggered modes operate identically to the
corresponding inhibited modes.
Note:
ms (200 ppm) from the previous paced event. Temporary
programmed triggered pacing is not limited to 300 ms (200 ppm).
Programmed triggered pacing will not occur faster than 300
Pacing Rate Interval
V
P
Parameters:
Pacing Rate = 60 ppm (1000 ms)
Ventricular Refractory Period = 300 ms
V
R
Figure 1-14.
Pacing Rate Interval
V
P
T
P
Example of VVT Mode Operation
V
P
200 ms
29
1-20
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
DOOR / AOOR / VOOR Modes
The DOOR, AOOR, and VOOR modes operate as follows:
■
The DOOR mode provides asynchronous AV sequential
pacing at the sensor-indicated rate. Intrinsic events are
ignored.
■
The AOOR and VOOR modes provide single chamber pacing
at the sensor-indicated rate. Intrinsic events are ignored.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-21
Pacing Modes
DOO / AOO / VOO Modes
DOO / AOO / VOO Modes
The DOO, AOO, and VOO modes operate as follows:
■
The DOO mode provides AV sequential pacing at the
programmed Lower Rate with no inhibition by intrinsic
events.
■
The AOO and VOO modes provide pacing at the
programmed Pacing Rate with no inhibition by intrinsic
events in the applicable chamber.
Besides being directly programmable, the DOO mode is the
Magnet mode of the corresponding dual chamber modes, except
for the VDD mode, which is the VOO mode. AOO and VOO
modes are the Magnet modes of the corresponding atrial and
ventricular single chamber modes, respectively.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
ODO / OAO / OVO Modes
Pacing Modes
ODO / OAO / OVO Modes
w
Warning:
patients. For such patients, use the programmer’s inhibit function
for brief interruption of outputs.
In the ODO, OAO, and OVO modes, sensing occurs in the
designated chamber(s). When used in conjunction with Marker
Channel telemetry and concurrent ECG, these modes may be used
to observe underlying rhythms.
■
■
Never program these modes for pacemaker-dependent
Blanking periods in these modes are automatically minimized
to maximize the sensing window(s). Thus, Marker Channel
telemetry may display sense markers for cardiac events (for
example, far-field R waves) that otherwise would not appear
due to longer blanking.
No timing intervals or refractory periods are used.
32
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
1-23
Pacing Modes
ODO / OAO / OVO Modes
33
1-24
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Rate Response
This chapter describes how the pacemaker’s
automatic rate response features operate and how to
individualize rate response. The sections are as
follows:
2
2
Introduction to Rate
Responsive Pacing
Preset Rate Response
at Implant
Rate Profile Optimization
Operation
Individualizing Rate Profile
Optimization
2-4
2-6
2-11
2-2
34
Activity Sensor Operation
Manual Control of Rate Profile
Optimization
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-19
2-13
2-1
Rate Response
Introduction to Rate Responsive Pacing
Introduction to Rate Responsive Pacing
Rate Response
The pacemaker may provide appropriate rate response for
patients who require cardiac pacing support at both submaximal
and maximal rates. To achieve appropriate rate response, the
pacemaker provides activity sensor-driven pacing with rate
response control at the submaximal and maximal rate ranges.
Submaximal rates are moderate pacing rates obtained during
typical daily activities, such as walking or daily chores. Maximal
rates are rates at or near the upper rate obtained during vigorous
activities.
The pacemaker provides appropriate rate response by employing
the following operations:
■
Three programmable rates control the submaximal and
maximal rate ranges: Lower Rate, ADL Rate (Activities of
Daily Living Rate), and Upper Sensor Rate. The ADL Rate is
equivalent to the average target rate that the patient achieves
for moderate activities.
35
2-2
■
Independent control of rate response is provided in both the
submaximal and maximal rate ranges.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Rate Response
Introduction to Rate Responsive Pacing
Automatic Features
For models implanted in a rate responsive mode, the pacemaker
automatically enables rate response after implant and
automatically adjusts rate response, if necessary, once each day.
■
During the first 30 minutes after implant, pacing occurs at the
implanted mode but without rate response. 30 minutes after
implant, nominal rate response operation is enabled.
■
Once each day, rate response is assessed and adjusted, if
necessary, in the submaximal or maximal rate ranges. The
assessment is based on comparing the pacemaker’s historical
sensor-indicated rate profiles against a clinician prescribed
target rate profile of the patient. If the rate profiles differ, rate
response is adjusted slightly in the appropriate rate range,
and the assessment is repeated again the next day.
For Further Information
Refer to “Rate Profile Optimization Operation” on page 2-6 for
information on how the pacemaker optimizes rate response.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-3
36
Rate Response
Preset Rate Response at Implant
Preset Rate Response at Implant
Overview
The pacemaker is shipped in the DDDR mode for dual chamber
rate responsive models and the VVIR mode for single chamber
rate responsive models. It effectively operates in a nonrate
responsive mode—DDD for dual chamber models and VVI for
single chamber models—until implant detection is completed,
which is typically 30 minutes after implant. Thereafter, the
pacemaker automatically enables rate responsive pacing.
Consequently, no programming is required for nominal rate
response operation.
Rate response is enabled for other operations in nonrate
Note:
responsive modes (e.g., DDD mode switching to DDIR).
Three Pacing Rate Controls
37
2-4
If customization of rate response is desired, three pacing rates are
provided to control the submaximal and maximal rate ranges:
■
Lower Rate defines the slowest rate at which pacing occurs in
the absence of a sinus rate or physical activity.
■
ADL Rate (Activities of Daily Living Rate) is the approximate
rate that the patient’s heart is expected to reach during
moderate exercise.
■
Upper Sensor Rate provides the upper limit for the sensordriven rate during vigorous exercise.
Refer to “Rates” on page 3-2 for additional considerations when
selecting pacemaker rates.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Rate Response
Preset Rate Response at Implant
Independent Control of Submaximal and
Maximal Rates
The pacemaker automatically assesses and independently adjusts,
if necessary, rate response levels for both submaximal and
maximal rate ranges on a daily basis using rate profile data. This
operation allows the pacemaker to respond to changes in
submaximal or maximal exercise levels without affecting rate
response in the other rate range. This independent control is
accomplished by comparing rate profiles of the patient’s sensorindicated rates against a nominal or clinician prescribed target
rate profile.
■
If the rate profile data indicates that sensor-indicated rates are
higher than rates targeted for in the submaximal and/or
maximal ranges, the pacemaker will decrease rate response
slightly in the respective range.
■
If rates are lower than targeted for, rate response is increased
slightly.
38
Starting Rate Response Immediately
In situations where the clinician wishes to start rate responsive
pacing before the 30 minute implant detection period is
completed, perform the following steps:
Program Implant Detection to “Off/Complete”.
1.
Configure pace and sense lead polarities and Lead Monitor.
2.
Verify that Rate Profile Optimization is On.
3.
Verify the appropriate ADL Response, Exertion Response,
4.
Activity Threshold, Activity Acceleration, and Activity
Deceleration settings.
For Further Information
Refer to “Rate Profile Optimization Operation” on page 2-6 and
“Individualizing Rate Profile Optimization” on page 2-11.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-5
Rate Response
Rate Profile Optimization Operation
Rate Profile Optimization Operation
Overview
When Rate Profile Optimization is programmed On, the
pacemaker can adapt submaximal and maximal rate response
levels once each day by comparing the patient’s current sensor
rate profiles against a nominal or clinician prescribed target rate
profile. This feature is intended to provide automatic and
independent monitoring of rate response at both submaximal
rates for daily patient activities, such as walking and daily chores,
and at maximal rates for vigorous patient activities.
Optimization can be individualized to the patient’s activity levels.
Refer to “Individualizing Rate Profile Optimization” later in this
chapter.
Optimization can also operate in the background when a nonrate
responsive mode is programmed. This can provide appropriate
rate response to patient activity if rate response is needed later or
for certain therapy features, such as mode switching to a nonatrial
tracking rate responsive mode.
39
2-6
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Rate Response
Rate Profile Optimization Operation
Submaximal and Maximal Rate Control
The pacemaker maintains a linear relationship between the sensor
input and the sensor-indicated rate in both the submaximal and
maximal rate ranges. Optimization controls how rapidly and to
what level the sensor-indicated rate increases and decreases in
these two rate ranges. The three programmable rate controls
[Lower Rate, ADL Rate (Activities of Daily Living Rate), and
Upper Sensor Rate] define the rate ranges, see Figure 2-1.
■
Submaximal rates are moderate pacing rates achieved during
typical daily patient activities. These rates are at or near the
ADL Rate.
■
Maximal rates are rates at or near the Upper Sensor Rate
achieved during vigorous activities.
Optimization also controls rate responsiveness for rates below the
submaximal rate range (i.e., at or near the Lower Rate).
40
30%
20%
10%
Indicated % of Time
Lower
Rate
Figure 2-1.
Submaximal
Rate
Range
ADL
Rate
Submaximal and Maximal Rate Ranges
Maximal
Rate
Range
Sensor Rate
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Upper
2-7
Rate Response
Rate Profile Optimization Operation
Optimization Using Rate Profiles
Optimization of rate response occurs independently in both the
submaximal and maximal rate ranges. Rate response to activity
sensor changes are assessed daily based on the following rate
profile data:
Sensor-indicated rate profile – An actual rate versus time
distribution of the patient’s averaged sensor-indicated rates. Once
each day, the pacemaker collects a daily sensor rate profile and
cumulates the data into a monthly average. Both the daily and
monthly rate profiles are assessed each day to determine if
adjustments to rate response are required. The monthly sensor
rate profile is automatically stored in the Sensor Indicated Rate
Profile diagnostic.
Target rate profile – A programmable rate versus time
distribution of the patient’s desired rates. The ADL Response and
Exertion Response parameters define the percent of time the
sensor-indicated rate is in the submaximal and maximal ranges,
respectively.
2-8
41
30%
Daily Sensor
20%
10%
Indicated % of Time
30%
20%
10%
Indicated % of Time
Rate Profile
Rate
Monthly Sensor
Rate Profile
Rate
Figure 2-2.
C
o
m
p
a
r
e
e
r
Co
a
p
m
Indicated % of Time
Comparing Rate Profiles
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
30%
Target Rate
Profile
20%
10%
Rate
Rate Response
Rate Profile Optimization Operation
Daily Optimization of Rate Response
Once each day, the pacemaker evaluates the percent of time the
sensor rate is in the submaximal and maximal rate ranges by
comparing the daily and monthly sensor-indicated rate profiles
against the target rate profile, see Figure 2-2. From this
comparison, the pacemaker automatically adjusts rate response in
the submaximal and maximal ranges, if necessary, based on the
following criteria:
■
If the sensor rate profiles show a higher percent of time spent
pacing than the target rate profile, rate response for the
pertinent rate range is set to be less responsive. Conversely, if
a lower percent of time spent pacing is profiled than targeted
for, rate response is set to be more responsive, see Figure 2-3.
■
If the sensor rate profiles match the target rate profile or the
daily and monthly sensor rate profiles contradict each other,
no rate response adjustments occur.
The goal of this operation is to keep the patient’s sensor rate
profiles equivalent to the target rate profile.
42
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-9
Rate Response
Rate Profile Optimization Operation
30%
25%
20%
15%
10%
Indicated % of Time
5%
60-70
Lower
Rate
=
Target Rate Profile
=
Sensor Rate Profile
Less paced rates
than targeted for,
rate response in the
maximal range will be
adjusted to be
more aggressive
Maximal
Rate Range
120-130
130-140
Upper
Sensor Rate
70-80
Figure 2-3.
More paced rates than
targeted for, rate response
in the submaximal range
will be adjusted to be
less aggressive
Submaximal
Rate Range
80-9090-100 100-110 110-120
ADL
Rate
Optimizing Rate Response
Adaptations in Optimization Operation
43
2-10
The pacemaker adapts rate response more rapidly for the first ten
days after Optimization is first activated post-implant or after
certain rate response parameters are manually reprogrammed
(e.g., Lower Rate, ADL Rate, Upper Sensor Rate, ADL Response,
or Exertion Response). The intent is to quickly match rate
response to the target rate profile under these circumstances.
Optimization also adapts rate response more rapidly when the
difference between the sensor rate profile and target rate profile is
significant or when a need for less rate response is indicated.
Optimization is skipped on any day that a device interrogation or
parameter programming occurs.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Individualizing Rate Profile Optimization
Individualizing Rate Profile Optimization
Overview
For Kappa 700 Series pacemakers, the clinician can prescribe a
target rate profile using the ADL Response and Exertion Response
parameters to match the patient’s life-style or activity levels. The
programmable ADL Response parameter alters the targeted rate
distribution in the submaximal rate range, while the Exertion
Response parameter alters the rate distribution in the maximal
rate range. Optimization takes approximately one month to
incrementally adjust rate response to match the individualized
target rate profile.
Submaximal Rate Profiles
The nominal setting for the ADL Response parameter is
“3-Moderately Active”. Programming a more active setting
redefines the target rate profile to spend more time pacing at or
above the ADL Rate, thereby increasing rate responsiveness in the
submaximal rate range. Programming a less active setting
redefines the rate profile to spend less time pacing at or above the
ADL Rate, thereby decreasing rate responsiveness.
Rate Response
44
Maximal Rate Profiles
The nominal setting for the Exertion Response parameter is
“3-Moderately Frequent”. Programming a more frequent setting
redefines the target rate profile to spend more time pacing near
the Upper Sensor Rate, thereby increasing rate responsiveness in
the maximal rate range. Programming a less frequent setting
redefines the rate profile to spend less time pacing near the Upper
Sensor Rate, thereby decreasing rate responsiveness.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-11
Rate Response
Individualizing Rate Profile Optimization
Programming Guidelines
If it is necessary to adjust rate response from the nominal setting,
first verify that the three rate controls are appropriate for the
patient. The Lower Rate should provide sufficient cardiac support
at rest; the ADL Rate should be set for submaximal rate support
during typical daily activities; and the Upper Sensor Rate should
limit the rate during maximal exertion. If these rate control
settings are appropriate, the ADL Response and/or Exertion
Response settings can then be adjusted based on the guidelines
below in Table 2-1:
Tabl e 2- 1.
Rate Range Select Settings
Submaximal or
Moderate Rates
Maximal or High RatesRate Response
a
If a higher Exertion Response setting has not produced the desired rate
response, increase the ADL Response setting.
ADL Response and Exertion Response Guidelines
ADL Response
Rate Response
Too Aggressive
Rate Response
Too Low
Too Aggressive
Rate Response
Too Low
Lower Number
(Less Active)
Higher Number
(More Active)
Exertion Response
Lower Number
(Less Frequent
Exertion)
Higher Number
(More Frequent
Exertion)
a
For more detailed programming guidelines, refer to Table E-12
and Table E-13 which list the targeted time spent pacing for the
five ADL Response and Exertion Response settings.
45
2-12
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Activity Sensor Operation
Overview
Activity sensor based pacing is controlled by the following
programmable parameters:
■
Activity Threshold determines the minimum intensity of
detected physical activity to which the pacemaker responds.
■
Activity Acceleration and Activity Deceleration times control
how rapidly the pacing rate changes in response to increased
or decreased activity. One programmable Activity
Deceleration setting, “Exercise” provides an extended
deceleration period following prolonged exercise.
Note that Activity Threshold, Activity Acceleration, and Activity
Deceleration are automatically set to nominal settings 30 minutes
after implant or can be manually programmed.
Rate Response
Activity Sensor Operation
46
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-13
Rate Response
Activity Sensor Operation
How Activity Threshold Influences Rate
A piezoelectric crystal, bonded to the pacemaker circuitry, is
deflected by physical motion. The activity sensor converts
detected motion into electrical signals. The programmed Activity
Threshold screens out activity signals below the selected setting.
Detected sensor signals vary from patient to patient due to body
structure, placement of pacemaker, and so forth. Only sensor
signals whose amplitude exceeds the programmed Activity
Threshold (as shown in Figure 2-4) are used in computing the
sensor-indicated rate. The lower the Activity Threshold, the
smaller the signal required to influence the sensor-indicated rate.
Settings
High
Med/High
Med/Low
Low
2-14
Activity
Sensor
Output
Time
Activity Threshold = Medium/Low
Figure 2-4.
Activity Sensor Signal (Threshold Set to Medium/Low)
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Low
Med/Low
Med/High
High
47
Rate Response
Activity Sensor Operation
Evaluating the Activity Threshold Setting
Walking increases the pacing rate; sitting results in pacing at or
near the programmed Lower Rate. Use Table 2-2 below as a guide
for selecting an appropriate setting.
Table 2-2.
Programmable
Settings
LowResponds to most body activity, including
Medium/Low Limited response to minimal exertion;
Medium/High Limited response to moderate body
High Responds to only vigorous body movements
Activity Threshold Guidelines
Typical Rate Performance
minimal exertion.
responds to moderate or greater exertion.
movements and exertion.
and exertion.
How Activity Acceleration and Deceleration
Influence Rate
Activity Acceleration and Activity Deceleration times control how
rapidly the pacing rate changes in response to increased or
decreased physical activity. One programmable Activity
Deceleration setting, “Exercise” provides an extended
deceleration period following prolonged exercise.
48
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-15
Rate Response
Activity Sensor Operation
■
Activity Acceleration time is the time required to achieve
approximately 90% of the difference between the current rate
and a higher steady-state rate consistent with the current level
of activity. Figure 2-5 shows a graphic representation of the
acceleration curves at the onset of strenuous exercise.
Upper
Sensor
Rate
Activity Acceleration
Programmable Settings
15 Seconds
30 Seconds
60 Seconds
Rate Range
Lower
Rate
49
2-16
Time (Minutes)
Figure 2-5.
Activity Acceleration Curves
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
210453
Rate Response
Activity Sensor Operation
■
Activity Deceleration time is the time required to achieve
approximately 90% of the difference between the current rate
and a lower steady-state rate consistent with the current level
of activity. Figure 2-6 shows a graphic representation of the
deceleration curves at an abrupt cessation of strenuous
exercise.
Upper
Sensor
Rate
Rate Range
Lower
Rate
Activity Deceleration
Programmable Settings
2.5 Minutes
5 Minutes
10 Minutes
50
654321078910
Time (Minutes)
Figure 2-6.
Activity Deceleration Curves
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-17
Rate Response
Activity Sensor Operation
Exercise Deceleration Operation
Activity Deceleration programmed to “Exercise” extends the rate
slowing period following an exercise episode, providing up to
20 minutes of rate deceleration. When it is programmed on, the
pacemaker uses activity sensor data to detect periods of vigorous,
prolonged exercise. At the end of such an exercise period, the
pacemaker uses a longer deceleration curve for the central portion
of the programmed rate range. The actual deceleration rate is
determined dynamically based on the intensity and duration of
exercise and the new level of activity. Figure 2-7 shows the
composite deceleration curve that applies after the abrupt
cessation of sustained exercise.
Upper
Sensor
Rate
Rate Range
Lower
Rate
5 Minute Deceleration Curve
Begins Exercise Deceleration
Ends Exercise
Deceleration
Time (Minutes)
Figure 2-7.
Exercise Deceleration
5 Minute
Deceleration
Curve
1210864201416182022
51
2-18
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Rate Response
Manual Control of Rate Profile Optimization
Manual Control of Rate Profile Optimization
Overview
As an alternative to automatic Rate Profile Optimization, a
programmer assisted Exercise test can be used to manually set
rate response for the submaximal and maximal rates. The Exercise
test is used to immediately set rate response to certain levels. Rate
response parameters remain set to their programmed values if
Optimization is Off. When Optimization is On, it can adjust these
parameters once each day.
Evaluate and Program Rate Response
The Exercise test is used to evaluate the patient’s rate response
and allow manual programming of two rate response control
parameters:
■
ADL Setpoint (Activities of Daily Living Setpoint) determines
the minimum sensor response to pace at the ADL Rate, which
falls within the submaximal rate range.
52
■
UR Setpoint (Upper Rate Setpoint) determines the minimum
sensor response to pace at the Upper Sensor Rate, which is at
the upper limit of the maximal rate range.
The programmed ADL Setpoint setting must less than the
Note:
UR Setpoint setting.
Refer to Medtronic.Kappa 700/600 Series Pacemaker Programming
Guide for programming instructions.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
2-19
Rate Response
Manual Control of Rate Profile Optimization
53
2-20
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
This chapter describes the programmable and
nonprogrammable timing intervals that govern the
operation of the pacemaker.
3
3
Rates
AV Intervals
Rate Adaptive AV
Search AV and Diagnostic
Blanking Periods
Refractory Periods
High Rate Atrial Tracking
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-2
3-12
3-16
3-26
3-29
3-20
3-40
3-1
54
Pacemaker Timing
Rates
Rates
Overview
The following programmable rates control timing in the
pacemaker:
■
Normal operating rates:
–Lower rate
–ADL rate
–Upper tracking rate
–Upper sensor rate
■
Other operating rates:
–Sleep rate (for Sleep function)
–Hysteresis rate (for single chamber demand modes)
–Sinus preference zone (for Sinus Preference)
55
3-2
–Intervention rate (for Rate Drop Response)
Additionally, rates calculated by the pacemaker are used for some
operations. These are:
■
Sensor-indicated rate
■
Mean atrial rate
The other operating rates are described in “Special Therapy
Options” on page 5-1 along with the functions that use them. The
normal rates are described in this chapter.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Rates
A–A and V–V Timing
A–A Timing – In all modes that pace the atrium, the pacemaker
times from atrial event to atrial event (A–A timing). This timing
method mimics a natural sinus rhythm, producing A-A intervals
that are nearly equal, except when timing is interrupted by:
■
PACs in the DDIR and DDI modes,
■
PVCs in the DDDR, DDD, DDIR, and DDI modes (PVC
Response operation),
■
A ventricular sensed event during the VA interval in the DVIR
and DVI modes, or
■
An atrial refractory sensed event that triggers an NCAP
extension.
VA intervals vary due to adjustments by A-A timing operations in
order to achieve sensor-indicated or lower rate operation in the
presence of varying AV conduction.
56
V–V Timing – In modes that do not pace the atrium (e.g., VDD or
VDIR) or single chamber ventricular modes, the pacemaker times
from ventricular event to ventricular event (V–V timing).
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-3
Pacemaker Timing
Rates
Lower Rate IntervalLower Rate Interval
Lower Rate
The programmed Lower Rate defines the slowest rate at which
pacing occurs during a mode’s basic operation. In rate responsive
modes, in the absence of sensor-detected activity, the sensorindicated rate is equal to the programmed Lower Rate.
SAV Interval = 180 msVentricular Refractory Period = 240 ms
S
S
Example of Lower Rate Operation
P
P
200 ms
57
3-4
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Rates
Operating Lower Rate
Under certain circumstances, the programmed Lower Rate may
be overridden by an operating lower rate that is higher or lower
than the programmed value. The following rates may become the
operating lower rate:
■
Switching from and back to atrial tracking mode (for Mode
Switch)
■
Sinus preference zone (for Sinus Preference)
■
Sleep rate (for Sleep function)
■
Intervention rate (for Rate Drop Response)
■
Hysteresis rate (for single chamber modes)
■
Threshold margin test rate of 100 ppm
■
Magnet mode rate of 85 ppm
■
Elective replacement indicator rate of 65 ppm
58
Selecting a Lower Rate
Program the Lower Rate to maintain adequate heart rates during
periods of inactivity or during pauses in atrial rhythms in the
DDDR, DDD, VDD, AAIR, ADIR, AAI, and ADI modes.
In the VDD mode, atrial tracking near the Lower Rate may
Note:
result in V–V intervals that exceed the Lower Rate interval. This is
normal operation.
Lower Rates from 120 to 130 ppm are intended for pediatric
patients. Lower Rates below 50 ppm and above 100 ppm are
primarily intended for diagnostic purposes.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-5
Pacemaker Timing
Rates
Sensor-Indicated Rate
The sensor-indicated rate is the basic pacing rate in all rate
responsive modes (DDDR, DDIR, DVIR, DOOR, VVIR, VDIR,
VOOR, AAIR, ADIR, and AOOR). It is determined by the
pacemaker based on the sensor-detected level of patient activity
and the programmed rate response parameters. The sensorindicated rate will never be greater than the Upper Sensor Rate or
less than the Lower Rate.
Sensor-Indicated
Interval
SensorSensorSensor
P
DDDR
A
V
Parameters:
Sensor-Indicated Rate = 90 ppm (667 ms)
PAV Interval = 200 msPVARP = 300 ms
SAV Interval = 190 msVentricular Refractory Period = 220 ms
P
Figure 3-2.
P
S
Example of Sensor-Indicated Rate Operation
S
Sensor-Indicated
Interval
S
P
P
200 ms
59
3-6
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
In rate responsive modes, the sensor-indicated rate tracks the
activity sensor, which is detected by the piezoelectric crystal
sensor’s frequency and amplitude.
■
In dual chamber rate responsive modes, the sensor-indicated
interval is the AS–AP or AP–AP interval.
■
In single chamber rate responsive modes, the sensorindicated interval is the A–A or V–V interval. In these modes,
sensor-indicated rate intervals start with a sensed or paced
event in the chamber being paced.
Sensor Indicated Rate Effect on Other
Intervals
The sensor-indicated rate is used to determine the values of
certain other timing intervals. These intervals are:
■
Rate adaptive paced AV (PAV) interval
■
Sensor-varied PVARP (even in nonrate responsive DDD and
VDD modes)
Rates
60
■
PVARP extension (sensor-corroboration before PMT
intervention)
ADL Rate
The ADL Rate (Activities of Daily Living Rate) is the target rate
which the patient’s heart rate is expected to reach during
moderate exercise.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-7
Pacemaker Timing
Rates
Upper Tracking Rate
The programmable Upper Tracking Rate is the maximum rate at
which the ventricle may be paced in response to sensed atrial
events in the DDDR, DDD, and VDD modes. Sensed atrial events
below the Upper Tracking Rate will be tracked at a 1:1 ratio, but
sensed events above the Upper Tracking Rate will result in
pacemaker Wenckebach (for example, 6:5, 4:3, 3:2, or 2:1 block).
The Upper Tracking Rate usually should be programmed to a
value less than the 2:1 block rate. Refer to “High Rate Atrial
Tracking” on page 3-40 for details.
Upper Tracking Rate
Parameters:
Sensor-indicated Rate =
75 ppm (800 ms)
Upper Tracking Rate =
100 ppm (600 ms)
SAV Interval = 200 ms
3-8
S
DDD
A
V
Figure 3-3.
P
Example of Upper Tracking Rate (Wenckebach) Operation
S
P
Upper Sensor Rate
In rate responsive modes, the programmable Upper Sensor Rate
provides the upper limit for the sensor-indicated rate during
physical activity, particularly during vigorous exercise. In the
DDDR mode, the Upper Sensor Rate may be higher than, lower
than, or the same as the Upper Tracking Rate.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
200 ms
61
Pacemaker Timing
Rates
Programming Considerations and
Restrictions
ADL Rate – It is recommended that the ADL Rate be 10 ppm less
than the Upper Sensor Rate or 20 ppm greater than the Lower
Rate. Although, programming the ADL Rate above or below these
limits is permitted.
Upper Rates – Programming a combination of high Upper Sensor
Rate and Upper Tracking Rate and a long refractory period may
result in a shorter “sensing window.” Loss of sensing in such cases
could result in competitive pacing (unless Non-Competitive
Atrial Pacing is programmed On). See “Non-Competitive Atrial
Pacing” on page 5-9 for more information.
Programming the Upper Tracking Rate to a value greater than the
Upper Sensor Rate permits the atrial rhythm to be tracked to a rate
higher than the sensor-driven rate.
The Upper Sensor Rate and/or Upper Tracking Rate must be
greater than the Lower Rate. The Upper Sensor Rate must be
greater than or equal to the ADL Rate.
62
Rate Limit
An internal circuit, independent of the pacing timers, limits single
chamber atrial or ventricular pacing rates to 200 ppm for most
single component failures. For dual chamber modes, atrial and
ventricular rates are limited independently to 200 ppm. The rate
limit is automatically disabled during temporary pacing in the
AAI, ADI, AAT, AOO, VVI, VDI, VVT, and VOO modes to allow
high rate pacing for diagnostic or therapeutic purposes.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-9
Pacemaker Timing
Rates
Possible Atrial Competition at High Rates
At high sensor-driven rates in the DDDR and DDIR modes,
sensor-driven pacing may approximate the intrinsic atrial rate,
with some intrinsic atrial events falling into the PVARP. This could
result in asynchronous pacing with the potential for competitive
atrial pacing. Consider the potential for asynchronous pacing at
high rates before selecting an Upper Sensor Rate, especially for
patients known to be susceptible to induction of atrial
tachyarrhythmias. Weigh the benefits of high rate sensor-driven
pacing against the potential for competitive pacing.
Use of the Rate Adaptive AV feature and sensor-varied or
Note:
automatic PVARP can reduce the likelihood of the type of
asynchronous pacing described above. In the DDDR mode, Sinus
Preference and NCAP can also be considered.
Mean Atrial Rate
The mean atrial rate (MAR) is a running average of the atrial rate
for use by the Rate Adaptive AV and automatic PVARP features.
The average uses all A–A intervals (except AS–AP or AR–AP
intervals). In order to respond quickly to rapidly increasing atrial
rates, the average gives preference to shorter A–A intervals over
longer intervals when calculating the MAR. Figure 3-4 shows how
the MAR tracks an increasing intrinsic atrial rate.
63
3-10
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
200
180
160
140
120
Rate (bpm)
100
80
60
05101520253035
Pacemaker Timing
Rates
Atrial Rate Increasing by 2 bpm/beat
MAR
Intrinsic Rate
Time (Seconds)
Figure 3-4.
Increasing Mean Atrial Rate
64
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-11
Pacemaker Timing
AV Intervals
AV Intervals
Overview
In dual chamber modes, the AV intervals determine the time
between the occurrence of an atrial event and the scheduled
delivery of a ventricular stimulus. Separate AV intervals for paced
and sensed atrial events are available. The lengths of these
intervals may be programmed to fixed values or (optionally) rate
adaptive or therapeutically determined.
Paced AV Interval (PAV) – PAV follows an atrial pace in the
DDDR, DDD, DDIR, DDI, DVIR, DVI, DOOR, and DOO modes.
The PAV interval duration may differ from the programmed value
due to one of the following operations:
■
Rate Adaptive AV
■
Search AV
■
Ventricular Safety Pacing
65
3-12
■
Non-Competitive Atrial Pacing
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
AV Intervals
PAV
Interval
P
P
DDD
Interval
P
A
V
Figure 3-5.
PAV
PAVPAV
P
Example of PAV Interval Operation
Sensed AV Interval (SAV) – SAV follows an atrial sensed event in
atrial synchronous pacing modes (DDDR, DDD, and VDD). The
SAV interval duration may differ from the programmed value due
to one of the following operations:
■
Rate Adaptive AV
66
■
Automatic PVARP
■
Search AV
■
Wenckebach
For Wenckebach operation, the SAV is extended to avoid violation
of the Upper Tracking Rate or the total atrial refractory period
while tracking a fast intrinsic atrial rate.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-13
Pacemaker Timing
AV Intervals
SAV
Interval
S
DDD
Interval
S
A
V
Figure 3-6.
SAV
SAVSAV
PP
Example of SAV Interval Operation
Selecting PAV and SAV
When programming AV intervals, the general hemodynamic goal
is to assure that, to the extent possible, left-atrial systole is
completed before left-ventricular systole begins. To achieve this,
the AV interval durations may be adjusted independently of each
other.
67
3-14
■
To accommodate the difference in interatrial conduction
times, the SAV usually should be programmed to a shorter
duration than the PAV, typically 30 to 50 ms shorter. If an SAV
greater than the PAV is selected, the programmer notes that
this is not usual, but the selected values may be programmed
if clinically warranted.
■
When the SAV is longer than the PAV, a V pace following an
atrial sense will always occur after the full SAV, even when the
sensor-indicated rate or Lower Rate interval expires first.
■
In certain patients, short AV intervals may be used as a
prophylaxis for AV nodal or accessory pathway reentrant
tachycardias in dual chamber modes.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
AV Intervals
■
AV intervals should be programmed to allow for normal AV
conduction and ventricular depolarization in sick sinus
patients.
■
Long PAV intervals (greater than or equal to 250 ms) should
be used with caution. If intrinsic ventricular events occur and
are not sensed, a long PAV may result in pacing into the
ventricle’s relative refractory period, including the Twave, or
loss of AV synchrony, which may precipitate retrograde
activation of the atria with corresponding hemodynamic
consequences and symptoms.
68
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-15
Pacemaker Timing
Rate Adaptive AV
Rate Adaptive AV
Overview
In the normal heart, AV conduction times tend to shorten as the
heart rate increases and lengthen as the heart rate decreases. The
Rate Adaptive AV (RAAV) feature, available in the DDDR, DDD,
DDIR, DVIR, DOOR, and VDD modes, mimics this physiologic
response. When RAAV is programmed On, the pacemaker
shortens AV intervals for atrial rates within a programmed rate
range. This feature provides increased opportunity for atrial
sensing, as follows:
■
■
Shortened SAV intervals increase the tracking range at fast
atrial rates by shortening the total atrial refractory period
(TARP) and increasing the 2:1 block rate. Refer to “Total Atrial
Refractory Period (TARP)” on page 3-34 and “High Rate
Atrial Tracking” on page 3-40 for more information.
Shortened PAV intervals lengthen the atrial sensing window
of the VA interval at higher sensor-driven rates.
69
3-16
RAAV will not shorten AV intervals to less than 30 ms.
Note:
Programming for Rate Adaptive AV
For RAAV operation, the SAV and PAV are programmed (as
applicable) to the values desired for low rates. Three additional
programmable parameters control how AV intervals are adjusted
at higher rates:
Start Rate – RAAV operation of shortening SAV and PAV
intervals begins at this rate.
Stop Rate – The shortest SAV and PAV occur at this rate and at all
higher rates, up to the upper rate limits.
Maximum Offset – The maximum amount of time (in ms) by
which the SAV and PAV intervals can be shortened.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
240
220
200
Pacemaker Timing
Rate Adaptive AV
The PAV minus the Maximum Offset gives the shortest PAV
interval at the Stop Rate (e.g., 200 ms - 100 ms = 100 ms).
Subtracting the Maximum Offset from the SAV gives the shortest
SAV interval (e.g., 170 ms - 100 ms = 70 ms).
Figure 3-7 shows how the SAV and PAV intervals are linearly
shortened as the rate increases from below the Start Rate to above
the Stop Rate.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Rate Adaptive AV Operation (DDDR Mode)
3-17
Pacemaker Timing
Rate Adaptive AV
RAAV Operations
Shortening of the AV interval(s) occurs when the appropriate rate
exceeds the programmed Start Rate, as follows:
SAV – The mean atrial rate determines SAV adjustments. Because
of how the mean atrial rate is calculated:
■
SAV adjustments will lag during rapid increases or decreases
in intrinsic atrial rates.
■
The SAV is not adjusted for isolated events (PACs).
■
AS–AP or AR–AP intervals may affect the SAV value since
these intervals are not used in the mean atrial rate calculation.
PAV – The sensor-indicated rate determines PAV adjustments.
The approximate difference between programmed SAV and PAV
is maintained as the SAV and PAV intervals are adjusted.
71
3-18
Programming Considerations and
Restrictions
Search AV – RAAV will adjust AV intervals in conjunction with
Search AV. Search AV will first determine the appropriate AV
intervals and RAAV will further modify the AV intervals. Note
that with RAAV On, the adaptive Search AV operation is limited
to the RAAV minimum of 30 ms.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Rate Adaptive AV
RAAV and Sick Sinus Syndrome
If RAAV is activated for a sick sinus syndrome patient whose PAV
and SAV have been programmed to promote AV conduction,
consider the following:
■
The rate at which AV conduction is lost should not be too low
(i.e., below 90 bpm).
■
Review of the AV Conduction Histogram diagnostic data may
aid in appropriate programming of Start Rate and Stop Rate
to maintain AV conduction as long as possible.
72
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-19
Pacemaker Timing
Search AV and Diagnostic
Search AV and Diagnostic
Overview
The Search AV feature is intended for patients with intact or
intermittent AV conduction and is available in the DDDR, DDD,
DDIR, DDI, DVIR, DVI, and VDD modes for dual chamber Kappa
700 Series pacemakers. The pacemaker searches for the patient’s
intrinsic AV conduction time and adjusts the SAV and PAV
intervals either longer or shorter to promote intrinsic activation of
the ventricles and to track fast atrial rates. When Rate Adaptive
AV is active, the pacemaker will also adjust the SAV and PAV
intervals relative to the rate adaptive values.
Two methods for adapting AV intervals are available:
■
Adjusting the AV intervals adaptively as intrinsic AV
conduction times vary.
■
Adjusting AV intervals at a fixed hysteresis value to sensed
ventricular events.
73
3-20
Programming to Adaptive AV
Programming Search AV to “Adaptive” operation requires the
setting of the Maximum Offset parameter. This parameter defines
the maximum amount of time (in ms) that the operating SAV and
PAV intervals can be lengthened.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Search AV and Diagnostic
Adaptive AV Operation
The pacemaker attempts to keep intrinsic conducted events in an
“AV delay window” that precedes scheduled paced events. The
AV delay window is set to promote intrinsic conduction of the
ventricles, but end early enough to avoid fusion or pseudo-fusion
beats if pacing is necessary, see Figure 3-8 below.
Previous 16 AV events, 8 or more VS events are within 15 ms of the
scheduled VP, thus PAV and SAV are extended by 31 ms to
promote intrinsic conduction.
PAV is now
181 ms
74
Previous 16 AV events, 8 or more VS events occur 55 ms
before the scheduled VP, thus PAV and SAV are shortened by
8 ms to limit long AV delays if ventricular pacing is needed.
Parameters:
DDDRSAV = 120 ms
Lower Rate = 60 ppmPAV = 150 ms
Sensor-Indicated Rate = 90 ppm
Figure 3-8.
Adaptive Search AV Operation
Maximum Offset = 110 ms
To determine when intrinsic conducted events occur, the
pacemaker assesses the 16 most recent AV conduction sequences
that start with a nonrefractory atrial sense (DDDR, DDD, and
VDD modes) or an atrial pace (DDDR, DDD, DDIR, DDI, DVIR,
and DVI modes) and end with a ventricular pace or a
nonrefractory ventricular sense.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
PAV is now
173 ms
200 ms
3-21
Pacemaker Timing
Search AV and Diagnostic
Search criteria of AV conduction times – The measured AV
conduction times are classified as either “too short” or “too long”.
■
Where too short means 8 or more of the last 16 ventricular
sensed events occurred within 15 ms of the scheduled
ventricular pace.
■
Where too long means 8 or more of the last 16 ventricular
sensed events occurred more than 55 ms before the scheduled
ventricular pace.
Resultant adjustment of SAV and PAV intervals – If AV
conduction times are classified as too short, the pacemaker will
lengthen the operating SAV and PAV intervals by 31 ms for the
next 16 pacing cycles to promote intrinsic conduction. The
maximum that the SAV and PAV can be lengthened is limited by
the Search AV Maximum Offset parameter.
If the previously 16 AV intervals were lengthened and are now
classified as too long, the operating SAV and PAV intervals will be
shortened by 8 ms for the next 16 pacing cycles. Shortening of the
SAV and PAV is limited by the programmed SAV and PAV values
or the RAAV Maximum Offset parameter, if RAAV is On.
75
3-22
Suspension of Adaptive AV Operation
If maximum extension of the SAV and PAV intervals is reached
and the search criteria indicates that the AV intervals are still too
short, the pacemaker will suspend all AV interval adjustments for
1 hour and reset the SAV and PAV intervals to the programmed or
current RAAV values. Subsequent AV interval adjustments will be
successively suspended at 2, 4, 8, and a maximum of 16 hours,
provided the maximum AV interval length is indicated as too
short.
When the search criteria indicates that the AV intervals are too
long, the current suspension period is terminated and
adjustments of AV intervals resumes.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Search AV and Diagnostic
Programming to Fixed AV Hysteresis
Programming Search AV to a specific AV hysteresis delay (in ms)
defines the amount of time the operating SAV and PAV intervals
can be lengthened.
Fixed AV Hysteresis Operation
The pacemaker modulates the AV intervals according to whether
or not the previous AV sequence started with an atrial event and
ended in a ventricular pace or sense.
■
If the AV sequence ended with a ventricular sense, the
programmed SAV and PAV intervals for the next pacing cycle
will be lengthened by the programmed fixed AV hysteresis
delay value to promote intrinsic conduction.
■
If the AV sequence ended with a ventricular pace, the
programmed SAV and PAV intervals are restored to the
programmed or RAAV values for the next pacing cycle.
76
After any A to VS event, SAV and PAV are
extended by 60 ms to promote intrinsic
conduction.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
After any A to VP event, SAV and PAV return
to programmed values to limit long AV
delays if ventricular pacing is needed.
PAV is now
210 ms
SAV = 120 ms
PAV = 150 ms
AV Hysteresis set to 60 ms
Hysteresis Search AV Operation
200 ms
3-23
Pacemaker Timing
Search AV and Diagnostic
Programming Considerations and
Restrictions
Automatic PVARP – When automatic PVARP is active and Search
AV is set to Adaptive, the pacemaker will ignore conduction times
that are the result of automatic PVARP shortening of the SAV
interval.
Rate Adaptive AV – Rate Adaptive AV (RAAV) will adjust AV
intervals in conjunction with Search AV. Search AV will first
determine the appropriate AV intervals and RAAV will further
modify the AV intervals. Note that with RAAV On, the adaptive
Search AV operation is limited to the RAAV minimum of 30 ms.
When RAAV is active between the RAAV Start Rate and Stop Rate,
Search AV will use the operating rate adaptive SAV and PAV
intervals when assessing and adjusting the AV intervals. At rates
below the RAAV Start Rate or when RAAV is Off, Search AV will
use the programmed SAV and PAV values.
77
3-24
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Search AV and Diagnostic
Recording AV Interval Adaptations
AV interval diagnostics record data about AV operations for the
Search AV feature. Also, you have the option to record detailed
data about AV operations.
Automatic Search AV Histogram
Programing AV Search parameters automatically initiates
recording of data by the Search AV Histogram diagnostic. This
histogram shows the number of A–VS, VS from Search, and A–VP
intervals versus rate. A histogram can be displayed or printed
from the Data icon.
Detail AV Interval Histogram
You have the option to record detailed histogram data on AV
intervals for the Search AV feature. The clinician-selected AV Interval
Histogram shows the number of AS–VS and AP–VS intervals
versus rate. (Refractory sensed events are not included.)
78
AV Interval Histograms record data about AV operations for the
Search AV feature.
Clearing AV interval Data
AV interval episode data is normally cleared by the pacemaker
one hour after a programming session.
Also, you can select the option to clear data immediately. Be sure
to save the session data or print the episode report before ending
the patient session.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-25
Pacemaker Timing
Blanking Periods
Blanking Periods
Blanking periods disable sensing for a programmable or
nonprogrammable interval. Signals that are blanked may
originate in either chamber or from outside sources such as noise
from muscle movement.
DDD
34
12
P
A
V
Note: Black bars indicate blanking
periods.
1. Nonprogrammable Atrial Blanking
P
2. Programmable Post-Ventricular
Atrial Blanking
3. Programmable Ventricular
Blanking
P
4. Nonprogrammable Ventricular
Blanking
79
3-26
Figure 3-10.
Example of Dual Chamber Blanking Operation
Nonprogrammable Blanking Periods
Immediately following a sensed or paced event in either chamber,
sensing for that chamber is blanked for a nonprogrammable
period that may vary from 50 to 100 ms. The actual duration of the
blanking period is determined dynamically by the pacemaker,
based on the strength and duration of the signal. Dynamic
blanking prevents sensing the same signal twice, while
minimizing total blanking time.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Blanking Periods
Post-Ventricular Atrial Blanking
The programmable Post-Ventricular Atrial Blanking (PVAB)
period, used in the DDDR, DDD, DDIR, DDI, VDD, VDIR, and
VDI modes, prevents sensing of ventricular paced events or farfield R waves on the atrial lead. Any ventricular event (paced or
sensed) starts the PVAB, which is also the first portion of the PostVentricular Atrial Refractory period (PVARP). PVAB is limited to
values equal to or less than the programmed PVARP, except in the
VDIR and VDI modes since PVARP does not apply to these
modes.
PVAB is programmed to a value less than or equal to
Note:
PVARP.
Ventricular Blanking
The programmable Ventricular Blanking period, which follows an
atrial pacing stimulus in the DDDR, DDD, DDIR, DDI, DVIR, and
DVI modes, prevents ventricular inhibition or ventricular safety
pacing due to sensing of the atrial stimulus on the ventricular lead
(crosstalk). The Ventricular Blanking period also applies to the
ADIR and ADI modes to prevent sensing of the atrial stimulation.
80
■
Long blanking periods (36 ms or greater) increase the
possibility of unsensed ventricular events.
■
Long blanking periods used in conjunction with long PAV
intervals (250 ms or greater) may result in pacing into the
T wave when intrinsic ventricular events are blanked and not
sensed. PAV values (200 ms or less) should reduce the
possibility of T wave pacing.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-27
Pacemaker Timing
Blanking Periods
Single Chamber Atrial Blanking
The programmable single chamber atrial blanking period, used in
the AAIR, ADIR, AAI, ADI, and AAT mode, prevents sensing of
far-field R waves. It is started by a paced, sensed, or refractory
sensed atrial event.
Atrial Blanking must be programmed at least 50 ms less
Note:
than the Atrial Refractory Period.
81
3-28
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Refractory Periods
Overview
A refractory period is an interval during which an intrinsic event
sensed on a particular lead channel cannot start certain timing
intervals. Each refractory period begins with a blanking period,
during which no sensing occurs. During the unblanked portion of
a refractory period, sensing occurs, but sensed events may not
directly affect timing operations. Refractory periods are intended
to prevent certain timing intervals from being started by
inappropriate signals such as retrograde P waves, far-field
R waves, or electrical noise.
Though they may not start timing intervals, refractory sensed
events are monitored by the pacemaker, and they affect the
operation of PVC Response, Mode Switch, Rate Adaptive AV
operation, automatic PVARP, Non-Competitive Atrial Pacing, and
other features for which the periodicity or number of sensed
events are pertinent. Refractory sensed events are included on
Marker Channel recordings.
Pacemaker Timing
Refractory Periods
82
Post-Ventricular Atrial Refractory Period
The Post-Ventricular Atrial Refractory Period (PVARP) follows a
paced, sensed, or refractory sensed ventricular event in the
DDDR, DDD, DDIR, DDI, and VDD modes. It is intended
primarily to prevent the sensing of retrograde P waves that might
promote Pacemaker-Mediated Tachycardias (PMTs) in atrial
tracking modes. In the DDIR and DDI modes, PVARP prevents
atrial inhibition from retrograde P waves.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-29
Pacemaker Timing
Refractory Periods
The first portion of the PVARP is the programmable PostVentricular Atrial Blanking period (PVAB). During the remainder
of the PVARP, intrinsic atrial events may be sensed as refractory
sensed events (AR) and identified on Marker Channel recordings,
but they do not affect stimulus timing.
■
In the DDDR, DDD, and VDD modes, an SAV is not started.
■
In the DDDR, DDD, DDIR, and DDI modes, the scheduled
atrial pace is not inhibited.
PVARP
P
DDD
A
P
3-30
V
P
Figure 3-11.
Example of PVARP Operation
The duration of the PVARP may be selected as follows:
■
The PVARP should be programmed to a value greater than
the patient’s ventriculoatrial (VA) retrograde time when
retrograde conduction is present.
■
Excessively long PVARPs may induce 2:1 block at high
intrinsic rates in atrial tracking modes (DDDR, DDD, and
VDD).
■
To reduce the 2:1 block point, PVARP can be set to vary based
on the sensor-indicated rate (sensor-varied PVARP) or the
mean atrial rate (automatic PVARP).
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
83
Pacemaker Timing
Refractory Periods
Sensor-Varied PVARP
When sensor-varied PVARP is programmed, the pacemaker
determines a value for the PVARP based on the sensor-indicated
rate. The intended purpose of the sensor-varied PVARP depends
upon the mode:
■
In the DDDR, DDD, and VDD modes, sensor-varied PVARP
is intended to do the following:
–Enhance protection against PMT at lower rates by
providing longer PVARPs at low sensor-indicated rates.
–Allow tracking of higher atrial rates (that is, provide a
higher 2:1 block rate) by shortening the PVARP at high
sensor-indicated rates.
■
In the DDIR mode, the sensor-varied PVARP is intended to
promote AV synchrony by preventing inhibition of atrial
pacing by an atrial sense early in the VA interval. It also
reduces the likelihood of competitive atrial pacing at high
sensor-indicated rates.
84
Upper
Sensor
Rate
Lower
Rate
Figure 3-12.
400 ms300 ms400 ms
PAVPVARPPAVPVARP
200 ms
Sensor-Varied PVARP Operation (DDDR Mode)
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-31
Pacemaker Timing
Refractory Periods
Determining Sensor-Varied PVARP
The pacemaker determines the duration of the sensor-varied
PVARP as follows:
■
In the DDDR, DDD, and VDD modes, the sensor-varied
PVARP is limited to 400 ms at low rates and the programmed
PVAB at high rates (as shown in Figure 3-12).
■
In the DDIR mode, the sensor-varied PVARP is approximately
400 ms at low rates and the programmed PVAB at high rates.
■
In the DDDR, DDD, DDIR, and VDD modes, the sensorvaried PVARP is automatically adjusted to maintain a 300 ms
sensing window (as shown in Figure 3-12).
Automatic PVARP
When automatic PVARP is programmed, the pacemaker
determines a value for the PVARP based on the mean atrial rate
(which is an average of all A-A intervals except those starting with
an atrial sense or atrial refractory sense and ending with an atrial
pace). In the DDDR, DDD, and VDD modes, automatic PVARP is
intended to provide a higher 2:1 block rate by shortening the
PVARP and SAV (if necessary) at higher tracking rates and protect
against PMTs at lower rates by providing a longer PVARP.
85
3-32
Determining Automatic PVARP
The pacemaker determines the duration of the automatic PVARP
as follows:
■
After every four pacing cycles, a 2:1 block rate is calculated
that is 30 bpm above the current mean atrial rate.
■
PVARP is then adjusted so the total atrial refractory period
equals the calculated 2:1 block rate. The programmable
Minimum PVARP parameter controls the minimum value
that PVARP can be shortened to.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
Refractory Periods
■
If the minimum PVARP value is reached and the 2:1 block rate
is still too low, the SAV interval can be shortened to increase
the 2:1 block rate. The minimum SAV that can be set is the rate
adaptive SAV value (i.e., the programmed SAV value minus
the RAAV Maximum Offset value).
The minimum adjustable 2:1 block rate is 100 ppm. The maximum
adjustable 2:1 block rate is the Upper Tracking Rate plus 35 ppm.
If Mode Switch is On, the maximum 2:1 block rate can be Detect
Rate if this rate is less than the Upper Tracking Rate calculation.
Programming Restrictions for Automatic
PVARP
Automatic PVARP – For Kappa 600 Series pacemakers, automatic
PVARP is not available.
Rate Drop Response – When programmed On in the DDD mode,
automatic PVARP is not available.
86
Spontaneous PVARP Extension
The programmed PVARP duration, the sensor-varied PVARP, and
the automatic PVARP may be overridden by the PVC Response
and PMT Intervention features, as follows:
■
When the PVC Response feature is programmed On and a
pacemaker-defined PVC occurs, the PVARP is forced to
400 ms for one cycle if a lesser value is in effect.
■
When PMT Intervention is programmed On and a
pacemaker-defined PMT is detected, the PVARP is forced to
400 ms for one cycle after the ninth paced ventricular event of
the PMT.
Refer to “PMT Intervention” on page 5-12 and “PVC Response”
on page 5-16 for further details on the PMT Intervention and PVC
Response features and their interactions with PVARP.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-33
Pacemaker Timing
Refractory Periods
Total Atrial Refractory Period (TARP)
In dual chamber modes that sense in the atrium, the Total Atrial
Refractory Period (TARP) is the sum of two intervals, as follows:
AV Interval – The AV interval begins with an atrial event and
ends with a ventricular event. The first portion is a
nonprogrammable blanking period. Its complete duration is
determined as follows:
■
In the DDDR, DDD, and VDD modes, the PAV or SAV interval
is the AV interval.
■
In the DDIR and DDI modes, the AV interval starts with the
first atrial sensed event in the VA interval or with an atrial
pacing stimulus; it ends when the PAV expires, even when
ventricular pacing is inhibited.
Post-Ventricular Atrial Refractory Period (PVARP) – The
PVARP is described earlier in this chapter.
3-34
TA RPTAR P
SAV + PVARP
SAV
PVARP
Figure 3-13.
Total Atrial Refractory Period
During atrial tracking, TARP = SAV + PVARP, and its duration
determines the rate at which 2:1 block occurs. Refer to “High Rate
Atrial Tracking” on page 3-40 for more information.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
SAV
PVARP
87
Pacemaker Timing
Refractory Periods
Ventricular Refractory Period
The programmable Ventricular Refractory Period (VRP) follows
paced, sensed, and refractory sensed ventricular events (including
PVCs) in all dual chamber and ventricular modes that sense in the
ventricle. The VRP is intended to prevent sensing of the Twave or
a PVC. The first portion of the VRP is a nonprogrammable
blanking period. A ventricular refractory sensed event affects
pacemaker timing as follows:
■
Ventricular blanking and refractory periods restart in all
modes.
■
In the DDDR, DDD, and VDD modes, the upper tracking rate
interval, PVARP, and PVAB also restart.
■
In the VVIR and VDIR modes, the upper sensor rate interval
restarts.
In dual chamber modes, the VRP should be programmed
Note:
shorter than the PVARP.
88
Figure 3-14.
P
DDD
A
V
P
Example of Ventricular Refractory Period Operation
P
VRP
In dual chamber modes, a ventricular refractory sensed event
does not affect a scheduled sensor-driven or lower rate atrial
output. Thus, a sensor-driven atrial output pulse will initiate a
PAV with a ventricular output pulse following, unless inhibited.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-35
Pacemaker Timing
Refractory Periods
Atrial Refractory Period (Single Chamber)
The programmable Atrial Refractory Period (ARP) follows paced,
sensed, and refractory sensed atrial events. The ARP is used in the
AAIR, ADIR, AAI, ADI, and AAT modes. It is intended to prevent
inhibition due to far-field R wave sensing. The first portion of the
ARP is a programmable blanking period. The ARP should be
programmed to a value long enough (150 ms or greater) to
prevent far-field R wave sensing but short enough to ensure atrial
sensing up to the programmed Upper Sensor Rate.
Noise Reversion
When sensing occurs during the Atrial Refractory Period (ARP) or
Ventricular Refractory Period (VRP), the refractory period (and its
blanking period) are restarted. The operation associated with
continuous refractory sensing in the ARP or VRP is called noise
reversion. Multiple restarts of the ARP or VRP (continuous noise
reversion) do not inhibit scheduled pacing. Pacemaker behavior
during continuous noise reversion is as follows:
89
3-36
■
Pacing occurs at the sensor-indicated rate for all rate
responsive modes (except VVIR and VDIR).
■
Pacing occurs at the programmed Lower Rate for all nonrateresponsive modes (including VVIR and VDIR).
On the ECG, noise reversion may be difficult to distinguish from
loss of sensing, but Marker Channel recordings will show
refractory sense markers when noise reversion occurs.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Example of Noise Reversion in DDDR at Sensor-Indicated Rate.
Atrial sensing during the PVARP, or refractory period
Note:
following an atrial paced or sensed event in the DDDR, DDD,
DDIR, DDI, or VDD modes does not restart the refractory period.
An atrial refractory sensed event, however, will start a short
blanking period of 50 to 100 ms depending on the signal strength
and duration of the atrial event.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-37
90
Pacemaker Timing
Refractory Periods
VVIR
V
SensorSensor
Parameters:
Lower Rate = 60 ppm (1000 ms)Upper Sensor Rate = 120 ppm (500 ms)
Ventricular Refractory Period = 240 ms
Lower Rate
RRRRP
P
P
200 ms
3-38
Figure 3-16.
Example of Noise Reversion in VVIR at Lower Rate.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
91
Pacemaker Timing
Refractory Periods
Preventing Noise Sensing
Noise reversion may be caused by electromagnetic interference
(EMI], myopotentials, excessively high output settings, or low
sensitivity settings. When it has been identified, noise reversion
usually can be reduced or eliminated by one of the following
actions:
■
Reprogram sensitivity to a less sensitive setting (higher
numerical value) or program Sensing Assurance on to
monitor and if necessary, adjust the sensitivity value. Refer to
“Sensing Assurance and Diagnostic” on page 4-31.
■
Reprogram sensing polarity to bipolar polarity (if available).
■
Reduce the amplitude and/or pulse width in the same or
opposite chamber.
■
Program Capture Management to Adaptive to monitor
ventricular capture thresholds and, if necessary, adjust
amplitude and pulse width values. Refer to “Capture
Management and Diagnostic” on page 4-15.
92
■
Remove patient from EMI environment.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-39
Pacemaker Timing
High Rate Atrial Tracking
High Rate Atrial Tracking
Overview
In the DDDR, DDD, and VDD modes, the fastest atrial rate the
pacemaker can track is determined by the total atrial refractory
period (TARP), which is the sum of the SAV and the PVARP.
Pacemaker behavior at high atrial rates in these modes is
determined by the relationship between the TARP and the
interval corresponding to the Upper Tracking Rate. In the DDDR
mode, the interval corresponding to the Upper Sensor Rate also
must be considered.
2:1 Block
When the intrinsic atrial interval is shorter than the TARP, some
atrial events will fall in the PVARP and not be tracked. At the rate
where this first occurs, ventricular tracking occurs only on
alternate beats, and 2:1 block ensues. In the DDD and VDD
modes, the ventricular pacing rate drops precipitously.
93
3-40
■
When sensor-varied PVARP or automatic PVARP is selected,
the 2:1 block rate may occur at a higher rate during activity
due to shortening of the PVARP and the SAV (automatic
PVARP only), thus increasing atrial tracking.
■
When Rate Adaptive AV operation is selected, the SAV
shortens at high atrial rates, shortening the TARP and raising
the 2:1 block rate.
■
When the 2:1 block rate is less than the Upper Tracking Rate,
the Upper Tracking Rate cannot be achieved.
■
In the DDDR mode, pacing at the sensor-indicated rate may
prevent a precipitous rate drop at the 2:1 block point when
activity is present.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
High Rate Atrial Tracking
■
For patients with a documented propensity for prolonged or
sustained atrial fibrillation or flutter, the clinician can select
Upper Tracking Rate, SAV, and PVARP values that induce
2:1 block at a desired rate (2:1 block rate = 60,000/TARP).
Alternatives for controlling rates in these patients include use
of the Mode Switch feature and DDIR mode pacing.
■
In the DDDR mode, atrial competition may occur if Upper
Sensor Rate exceeds the 2:1 block rate.
Pacemaker Wenckebach
When the 2:1 block rate exceeds the programmed Upper Tracking
Rate, pacemaker Wenckebach may occur. When the intrinsic rate
exceeds the Upper Tracking Rate, a pacing stimulus at the
expiration of the SAV would violate the upper tracking rate. The
pacemaker therefore extends the SAV until the upper tracking rate
interval expires. Subsequent SAVs require greater extension, until
an atrial event falls in the PVARP and is not tracked.
■
In the DDDR, DDD, and VDD modes, the result normally is a
fixed ratio between atrial and ventricular rates (3:2, 4:3, and so
forth).
94
■
In the DDDR mode, the pacemaker Wenckebach rate may be
smoothed by sensor-driven ventricular pacing, thereby
overriding the fixed ratio.
The following example shows how pacemaker Wenckebach
operation occurs in the DDDR, DDD, or VDD modes.
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-41
Pacemaker Timing
High Rate Atrial Tracking
Upper Tracking
Rate Interval
S
DDDR
A
V
P
Parameters:
Sensor-Indicated Rate = 90 ppm (667 ms) PVARP = 300 ms
PAV Interval = 230 msUpper Tracking Rate = 100 ppm(600 ms)
SAV Interval = 200 ms
S
Figure 3-17.
Upper Tracking
Rate Interval
P
S
Upper Tracking
Rate Interval
P
SAV
Interval
S
Example of Pacemaker Wenckebach Operation
Sensor
P
R
P
P
200 ms
95
3-42
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Pacemaker Timing
High Rate Atrial Tracking
High Rate Operation in the DDDR Mode
Table 3-1 summarizes how the total atrial refractory period
(TARP), the Upper Tracking Rate (UTR) interval, and the Upper
Sensor Rate (USR) interval may interact at high atrial rates in the
DDDR mode.
Table 3-1.
Relationship
Between TARP and
Upper Rate
Intervals
TARP > both USR
and UTR intervals
USR interval > TARP
> UTR interval
USR interval > UTR
interval > TARP
UTR interval > both
USR interval and
TA R P
a
Unless the Non-Competitive Atrial Pacing is On, see “Non-Competitive Atrial
Pacing” on page 5-9.
Upper Rates Interaction With TARP
Wenckebach
Before 2:1
Block
nonoyes
nonono
yesyesno
yesyesyes
Achieve
Upper
Tracking
Rate
Potential
Atrial
Competition
a
a
96
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
3-43
Pacemaker Timing
High Rate Atrial Tracking
97
3-44
Medtronic.Kappa 700/600 Series Pacemaker Reference Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.