TPV 22 and TPV 25 Transcatheter Pulmonary Valves and
Delivery Catheter System
Instructions for Use
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Medtronic, Medtronic with rising man logo, and Medtronic logo are trademarks of
Medtronic. Third-party trademarks (“TM*”) belong to their respective owners. The following
list includes trademarks or registered trademarks of a Medtronic entity in the United States
and/or in other countries.
AOA™, Harmony™
Explanation of symbols on package labeling
MR Conditional
For US audiences only
Model
Size
Serial number
Consult instructions for use at this website
Catalog number
Lot number
Manufacturer
Do not resterilize
Do not reuse
Do not use if package is damaged
Sterilized using ethylene oxide
Sterile LC: Device has been sterilized using liquid chemical sterilants according to EN/ISO 14160
Use-by date
Quantity
Temperature limit
Nonpyrogenic
Keep dry
Do not freeze
Fragile, handle with care
Keep away from sunlight
Maximum guidewire diameter
3
Temperature limit maintained
Temperature limit exceeded
Date of manufacture
Manufactured in
4
Device
Model
TPV 22
HARMONY-22
TPV 25
HARMONY-25
Delivery catheter system
HARMONY-DCS
1.0 Device description
The Harmony TPV system consists of a self-expanding transcatheter pulmonary valve and a
delivery catheter system.
Table 1: Model numbers
1.1 Transcatheter pulmonary valve (TPV)
The TPV consists of a porcine pericardial valve that is preserved in buffered 0.2%
glutaraldehyde and sutured within a Nitinol frame that is sewn onto a polyester knit fabric. The
inflow end of the TPV features an attachment suture loop on each crown to thread onto the
delivery catheter system coil during loading. The TPV is treated with an alpha amino oleic acid
antimineralization process (AOA), which has been shown to mitigate leaflet calcification in
animal studies. A final sterilization step is performed using a 0.2% glutaraldehyde sterilant in
which the TPV is preserved and packaged until used.
1.2 Patient anatomical criteria
Caution: The Harmony TPV bioprosthesis size must be appropriate to fit the patient’s anatomy
measured using a perimeter base framework. Proper sizing of the device is the responsibility of
the physician. Failure to implant a device within the sizing matrix could lead to adverse effects
such as those listed in Chapter 5.
Caution: The Harmony device is not intended for patients previously treated with an RV-PA
conduit or previously implanted bioprosthesis.
The Harmony TPV bioprosthesis is available in two sizes (TPV 22, model number HARMONY22 and TPV 25, model number HARMONY-25), and each is appropriate for a range of patient
main pulmonary artery (PA) sizes (measured with ECG-gated CTA at the end of the diastolic
phase, i.e. 90% time point in the cardiac cycle) as shown in Figure 1 and Figure 2 respectively.
Please also note the following points:
• this device is intended to be implanted in a section of the main PA (between the RVOT and
main PA bifurcation) with proper distal, proximal and axial dimensions to ensure sufficient
oversizing of the device with respect to the anatomy
• the perimeter values along the main PA should be compared to those of the device at all
possible implant locations or in all possible implant scenarios within the implant zone (i.e.,
within the main PA and between the RVOT and main PA bifurcation)
5
Harmony TPV22
Anatomical Size (perimeter-derived
Harmony TPV25
Anatomical Size (perimeter-derived
•the patient’s main PA should have cross-sectional perimeters described in Figure 1 or Figure
2 on the distal and proximal parts of the PA within the implant zone of the main PA
• the combined RVOT and overall PA length should be longer than or equal to the device
length
• The patient’s venous anatomy should accommodate an 8.33 mm (25Fr) delivery catheter
system
PA section
diameter)
(1) Outflow
Distal PA/Bifurcation
(2) Valve Housing>22 mm
(3) Inflow
Proximal PA/RVOT
Figure 1: TPV 22 (Nitinol frame profile) with dimensions and HARMONY-22 sizing matrix
PA section
22 mm - 28 mm
23 mm - 39 mm
diameter)
Figure 2: TPV 25 (Nitinol frame profile) with dimensions and HARMONY-25 sizing matrix
(1) Outflow
Distal PA/Bifurcation
(2) Valve Housing>25 mm
(3) Inflow
Proximal PA/RVOT
25 mm - 38 mm
32 mm - 48 mm
6
1.3 Delivery catheter system (DCS)
The delivery catheter system (DCS) has a braided outer shaft with a polytetrafluoroethylene
(PTFE) lined capsule in which the TPV is housed. The DCS has a soft, tapered distal tip. The
TPV is attached to the distal end of the DCS by the DCS coil and is protected by the capsule
during delivery. The deployment of the self-expanding TPV is controlled by pulling back the
outer shaft, allowing the TPV to open. Rotating the proximal handle on the proximal end of the
DCS rotates the DCS coil and releases the TPV for final deployment.
The DCS has a nominal outside diameter of 8.33 mm (25 Fr) and a nominal effective length of
101 cm. The DCS is compatible with an 0.889 mm (0.035 in) intravascular guidewire.
1. Distal tip
2. Guidewire lumen
3. Delivery catheter system coil
4. Capsule
5. Outer shaft
Figure 3: Delivery catheter system
7
6. Hemostasis sleeve
7. Stopcock
8. Hemostasis valve body
9. Hemostasis actuator
10. Proximal handle
11. Proximal handle actuator
12. Guidewire luer
13. Loading funnel halves (packaged with delivery catheter system)
14. Capsule support tube (preloaded on the delivery catheter system)
8
2.0 Indications
The Harmony Transcatheter Pulmonary Valve (TPV) System is indicated for use in the
management of pediatric and adult patients with severe pulmonary regurgitation (i.e., severe
pulmonary regurgitation as determined by echocardiography and/or pulmonary regurgitant
fraction ≥ 30% as determined by cardiac magnetic resonance imaging) who have a native or
surgically-repaired right ventricular outflow tract and are clinically indicated for surgical
pulmonary valve replacement.
9
3.0 Contraindications
The following are contraindications for the use of this device:
• Active bacterial endocarditis or other active infections
• Known intolerance to Nitinol (titanium or nickel) or an anticoagulation/antiplatelet regimen
10
4.0 Warnings and precautions
Carefully read all warnings, precautions, and instructions for use for all components of the
system before use. Failure to read and follow all instructions or failure to observe all stated
warnings could cause serious injury or death to the patient.
4.1 Warnings
4.1.1 General
• Implantation of the Harmony TPV system should be performed only by physicians who have
received Harmony TPV system training.
• The transcatheter pulmonary valve (TPV) is to be used only in conjunction with the Harmony
delivery catheter system (DCS).
• This procedure should only be performed where emergency pulmonary valve surgery can be
performed promptly.
•Do not use any of the Harmony TPV system components if any of the following has
occurred:
• It has been dropped, damaged, or mishandled in any way
• The Use By date has elapsed
4.1.2 Transcatheter pulmonary valve (TPV)
• This device was designed for single use only. Do not reuse, reprocess, or resterilize the TPV.
Reuse, reprocessing, or resterilization may compromise the structural integrity of the device
and/or create a risk of contamination of the device, which could result in patient injury,
illness, or death.
•Do not resterilize the TPV by any method. Exposure of the device and container to
irradiation, steam, ethylene oxide, or other chemical sterilants renders the device unfit for
use.
• The device is packaged with a temperature sensor. Do not freeze the device. Do not expose
the device to extreme temperatures. Do not use the device if the arrow on the sensor points to
the symbol that indicates that the temperature limit has been exceeded.
•Do not use the device if any of the following have occurred:
11
• The tamper-evident seal is broken.
• The serial number tag does not match the container label.
• The arrow on the sensor points to the symbol that indicates that the temperature limit
has been exceeded.
• The device is not completely covered by the storage solution.
• Do not contact any of the Harmony TPV system components with cotton or cotton swabs.
• Do not expose any of the Harmony TPV system components to organic solvents, such as
alcohol.
• Do not introduce air into the catheter.
• Do not expose the device to solutions other than the storage and rinse solutions.
• Do not add or apply antibiotics to the device, the storage solution, or the rinse solution.
• Do not allow the device to dry. Maintain tissue moisture with irrigation or immersion.
• Do not attempt to repair a damaged device.
• Do not handle the valve leaflet tissue or use forceps to manipulate the valve leaflet tissue.
• Do not attempt to recapture the device once deployment has begun.
• Do not attempt to retrieve the TPV if any one of the outflow TPV struts is protruding from
the capsule. If any one of the outflow TPV struts has deployed from the capsule,
the TPV must be released from the catheter before the catheter can be withdrawn.
•Do not attempt post-implant balloon dilatation (PID) of the TPV during the procedure, which
may cause damage to or failure of the TPV leading to injury to the patient resulting in
reintervention.
4.1.3 Delivery catheter system (DCS)
• This device was designed for single use only. Do not reuse, reprocess, or resterilize the DCS.
Reuse, reprocessing, or resterilization may compromise the structural integrity of the device
and/or create a risk of contamination of the device, which could result in patient injury,
illness, or death.
12
• Do not reuse or resterilize the DCS.
• If resistance is met, do not advance the guidewire, DCS, or any other component without
first determining the cause and taking remedial action.
•Do not remove the guidewire from the DCS at any time during the procedure.
4.2 Precautions
4.2.1 General
• Clinical long-term durability has not been established for the Harmony TPV. Evaluate the
TPV performance as needed during patient follow-up.
• The safety and effectiveness of Harmony TPV implantation in patients with pre-existing
prosthetic heart valve or prosthetic ring in any position has not been demonstrated.
• The Harmony TPV system has not been studied in female patients of child-bearing potential
with positive pregnancy.
4.2.2 Before Use
• Exposure to glutaraldehyde may cause irritation of the skin, eyes, nose, and throat. Avoid
prolonged or repeated exposure to the chemical vapor. Use only with adequate ventilation. If
skin contact occurs, immediately flush the affected area with water (for a minimum of 15
minutes) and seek medical attention immediately.
• The TPV and the glutaraldehyde storage solution are sterile. The outside of the TPV
container is nonsterile and must not be placed in the sterile field.
• The TPV and DCS should be used only in a sterile catheterization laboratory (cath lab)
environment. Ensure that sterile technique is used at all times.
• Strictly follow the TPV rinsing procedure.
• For TPV 25: Ensure that all green sutures have been removed from the attachment suture
loops on the TPV before loading onto the DCS.
• Prevent contamination of the TPV, its storage solution, and the DCS with glove powder.
• Verify the orientation of the TPV before loading it onto the DCS. The inflow end of the TPV
with attachment suture loops must be loaded first.
13
• Do not place excessive pressure on the TPV during loading.
• Inspect the sealed DCS packaging before opening. If the seal is broken or the packaging has
been damaged, sterility cannot be assured.
• Proper functioning of the DCS depends on its integrity. Use caution when handling the DCS.
Damage may result from kinking, stretching, or forceful wiping of the DCS.
• This DCS is not recommended to be used for pressure measurement or delivery of fluids.
• Carefully flush the DCS and maintain tight DCS connections to avoid the introduction of air
bubbles.
4.2.3 During Use
• The TPV segment is rigid and may make navigation through vessels difficult.
• Do not advance any portion of the DCS under resistance. Identify the cause of resistance
using fluoroscopy and take appropriate action to remedy the problem before continuing to
advance the DCS.
• Careful management of the guidewire is recommended to avoid dislodgement of the TPV
during DCS removal.
• Once deployment is initiated, retrieval of the TPV from the patient is not recommended.
Retrieval of a partially deployed valve may cause mechanical failure of the delivery catheter
system or may cause injury to the patient. Refer to Section 5.0 for a list of potential adverse
events associated with the Harmony TPV implantation.
• During deployment, the DCS can be advanced or withdrawn prior to the outflow struts
protruding from the capsule. Once the TPV struts contact the anatomy during deployment, it
is not recommended to reposition the device. Advancing the catheter forward once the TPV
struts make contact with the anatomy may lead to an undesired deployment or may cause
damage to or failure of the TPV and injury to the patient. Refer to Section 5.0 for a list of
potential adverse events associated with the Harmony TPV implantation.
• Physicians should use judgment when considering repositioning of the TPV (for example,
using a snare or forceps) once deployment is complete. Repositioning the bioprosthesis is not
recommended, except in cases where imminent serious harm or death is possible (for
example, occlusion of the main, left, or right pulmonary artery). Repositioning of a deployed
valve may cause damage to or failure of the TPV and injury to the patient. Refer to Section
5.0 for a list of potential adverse events associated with the Harmony TPV implantation.
14
• Ensure the capsule is closed before DCS removal. If increased resistance is encountered
when removing the DCS through the introducer sheath, do not force passage. Increased
resistance may indicate a problem and forced passage may result in damage to the device and
harm to the patient. If the cause of resistance cannot be determined or corrected, remove the
DCS and introducer sheath as a single unit over the guidewire, and inspect the DCS and
confirm that it is complete.
• If there is a risk of coronary artery compression, assess the risk and take the necessary
precautions.
• Endocarditis is a potential adverse event associated with all bioprosthetic valves (Chapter 5).
Patients should make their health care providers aware that they have a bioprosthetic valve
before any procedure. Postprocedure, administer appropriate antibiotic prophylaxis as needed
for patients at risk for prosthetic valve infection and endocarditis.
• Prophylactic antibiotic therapy is recommended for patients receiving a TPV before
undergoing dental procedures.
• Postprocedure, administer anticoagulation and/or antiplatelet therapy per physician/clinical
judgment and/or institutional protocol.
• Excessive contrast media may cause renal failure. Preprocedure, measure the patient’s
creatinine level. During the procedure, monitor contrast media usage.
• Conduct the procedure under fluoroscopy. Fluoroscopic procedures are associated with the
risk of radiation damage to the skin, which may be painful, disfiguring, and long-term.
15
5.0 Potential complications/adverse events
Potential risks associated with the implantation of the Harmony TPV may include, but are not limited
to, the following:
• Death
• Valve dysfunction
• Tissue deterioration
• Hematoma
• Heart failure
• Cerebrovascular incident
• Perforation
• Rupture of the RVOT
• Compression of the aortic root
• Compression of the coronary arteries
• Sepsis
• Pseudoaneurysm
• Erosion
• Stent fracture
• Arrhythmias
• Device embolization or migration
• Pulmonary embolism
• Occlusion of a pulmonary artery
• Laceration or rupture of blood vessels
16
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.