Medtronic H1701A08 Instructions for Use

HOW SUPPLIED

The reusable T rocar handle is provided as a clean but NOT STERILE assembly consisting of the handle, blade holder, end cap and rod; the disposable blade and tip sets are provided separately. Each disposable blade and tip set consists of one (1) clean but NOT STERILE blade and matching tip in sizes 8 mm, 10 mm, 12 mm, 14 mm, 16 mm, 18 mm, 20 mm, 22 mm, and 26 mm.

HANDLING/STERILIZATION

Adequate aeration is required following ETO Sterilization. After each use, discard the blade and tip. The trocar handle should be
disassembled and scrubbed with a brush and mild detergent. Particular attention should be given to the threaded and knurled areas of the instrument.

WARRANTY

Medtronic, Inc. warrants that at the time of manufacture, reasonable care was used in the manufacture of the trocar. No other warranty, including any warranty of merchantability or fitness for purpose is either expressed or implied. This warranty is exclusive and in lieu of all others. No representative of Medtronic, Inc. may change any of the foregoing and the purchaser (and/or user) accepts the product subject to all terms herein stated.

CAUTION

Federal law (USA) restricts this device to sale by or on the order of a physician.
3
Medtronic, Inc. 710 Medtronic Parkway NE Minneapolis, MN 55432-56 04 USA
© Medtronic, Inc. 2001 All Rights Reserved UC199403975aML 1200845-002A March 2001
HANCOCK TROCAR HANDLING INSTRUCTIONS

DESCRIPTION OF MATERIALS

The Hancock Trocar employs disposable stainless steel cylindrical blades which correspond to the size and geometry of the Hancock Apical Left Ventricle Connectors. The trocar creates a clean, uniform circular ventriculotomy with minimum trauma to surrounding tissues and retains the myocardial plug to simplify removal.

INDICATIONS

The Hancock Trocar is recommended for use with Hancock Apical Left Ventricle Connectors.

WARNINGS

Care should be taken not to damage the chordae tendineae or the ventricular septum when excising the myocardial plug.
Blades and tips are intended for single use only. Examine the trocar before use to verify correct assembly and secure
attachment of all components.

PRECAUTIONS

For each procedure, sterilize a single trocar handle, end cap and blade holder in addition to a matching blade and tip for each size connector which may be selected for implantation.
Aseptically assemble the appropriate blade and tip on the trocar handle after
the specific size of connector has been selected.
The opening created by the trocar must correspo nd to the size of connector to be used in order to aid hemostasis and minimize distortion of the connector.

DIRECTIONS FOR USE

The trocar must be disassembled before cleaning and sterilization. After sterilization the trocar handle and the appropriate size blade and tip should be assembled aseptically as shown in illustrations.
Specific recommendations pertaining to the use of the trocar are included in the surgical techniques detailed in the Hancock Apical Left Ventricle Connector handling instructions.
STEP I
HANDLE
END CAP
STEP IV
HANDLE
ROD
BLADE HOLDER
TIP
BLADE
Attach the end cap to the handle by turning the end cap clockwise.
STEP II
Attach the blade to the blade holder by turning the blade clockwise. CAUTION: Do not hold the blade near the cutting edge.
BLADE
BLADE HOLDER
STEP III
HANDLE
BLADE HOLDER
Slide the blade holder onto the shaft of the handle until the blade holder is fully seated inside the handle.
BLADE
Attach the tip to the threaded shaft of the handle by turning the tip clockwise until tight.
STEP V
BLADE
ROD
TIP
HANDLE
END CAP
BLADE HOLDER
Completed trocar assembly.

CLINICAL EXPERIENCE

Devices employing similar principles as the Hancock Trocar have been used extensively to create uniform openings in the heart and vessels.

ACCESSORIES

Replacement blade and tip sets are available in sizes 8 mm to 26 mm.