Medtronic Guardian User Manual

User Guide
Guardian Sensor (3)
- 1 -
English
Introduction
The Guardian™ Sensor (3) glucose sensor is part of your Continuous Glucose Monitoring (CGM) system. The sensor continuously converts tiny amounts of glucose from the interstitial fluid under your skin into an electronic signal. Your system then uses these signals to provide sensor glucose values.
C
D
A
B
A. pedestal B. needle
housing C. sensor D. clear liner
Potential risks related to sensor use
General risks with sensor use include:
• Skin irritation or other reactions
• Bruising
• Discomfort
• Redness
• Bleeding
• Pain
• Rash
• Infection
• Raised bump
• Appearance of a small “freckle-like” dot where needle was inserted
• Allergic reaction
• Fainting secondary to anxiety or fear of needle insertion
• Soreness or tenderness
• Swelling at insertion site
• Sensor fracture, breakage or damage
• Minimal blood splatter associated with sensor needle removal
• Residual redness associated with adhesive or tapes or both
• Scarring
Indications for use
The Guardian Sensor (3) is intended for use with the Medtronic MiniMed® 670G system to continuously monitor glucose levels in persons with diabetes. It is intended to be used for detecting trends and tracking patterns in persons aged fourteen years and older, and to be used by the MiniMed 670G system to automatically adjust basal insulin levels. It is indicated for use as an adjunctive device to complement, not replace, information
- 2 -
obtained from standard blood glucose monitoring devices. The sensor is intended for single use and requires a prescription. The Guardian Sensor (3) is indicated for 7 days of continuous use.
Contraindications
None known.
Assistance
Department Telephone Number
24 Hour HelpLine (calls within the United States)
800 646 4633
24 Hour HelpLine (calls outside the United States)
+1 818 576 5555
Website www.medtronicdiabetes.com
General warnings
Read this entire user guide before attempting to insert the sensor. The one-press serter (MMT-7512) does not work the same as other Medtronic insertion devices. Failure to follow directions or using a different serter may result in improper insertion, pain, or injury.
The Guardian Sensor (3) was developed for use with and the performance evaluated with the approved system only. The sensor should not be used as part of unapproved systems, as it may provide inaccurate sensor glucose readings.
The sensor is designed to work with approved transmitters only. It is not interchangeable with transmitters and recorders that are not compatible with the sensor. Connecting your sensor to a transmitter or recorder that is not approved for use with the sensor may cause damage to the components or inaccurate sensor glucose values.
Do not make therapy decisions based on sensor glucose values because sensor glucose and blood glucose (BG) values may differ. If your sensor glucose reading is low or high, or if you feel symptoms of low or high glucose, do the following prior to making therapy decisions. Confirm your blood glucose with your meter using a fingerstick blood sample.
Taking medications with acetaminophen, such as Tylenol®, fever reducers, or cold medicine, while wearing the sensor may falsely raise your sensor glucose readings. The level of inaccuracy depends on the amount of acetaminophen active in your body and may be different for each person. Always check the label of
- 3 -
any medications to confirm whether acetaminophen is an active ingredient.
Do not expose your sensor to MRI equipment, diathermy devices, or other devices that generate strong magnetic fields as the performance of the sensor has not been evaluated under those conditions and may be unsafe. If your sensor is inadvertently exposed to a strong magnetic field, discontinue use and contact the 24 Hour HelpLine for further assistance.
A retractable needle is attached to the sensor and minimal blood splatter may occur. If you are a healthcare professional or caregiver, wrap sterile gauze around the sensor to minimize contact with blood. Keep as much distance as possible between you and the patient when removing the needle.
Keep the needle housing within sight at all times to avoid an accidental needlestick or puncture.
Always inspect the packaging for damage prior to use. Sensors are sterile and non-pyrogenic, unless the package has been opened or damaged. Do not use the sensor if the sterile package has been opened or damaged. Use of an unsterile sensor can cause site infection.
This product contains small parts and may pose a choking hazard for children.
Watch for bleeding at the insertion site (under, around, or on top of the sensor).
If bleeding occurs, do the following:
1. Apply steady pressure, using sterile gauze or a clean cloth placed on top of the sensor, for up to three minutes. The use of unsterile gauze can cause site infection.
2. If bleeding stops, connect the transmitter (or recorder) to the sensor.
If bleeding does not stop, do not connect the transmitter to the sensor because blood can get into the transmitter connector, and could damage the device.
If bleeding continues, causes excessive pain or discomfort, or is significantly visible in the plastic base of the sensor, do the following:
plastic base
1. Remove the sensor and continue to apply steady pressure until the bleeding stops. Discard the sensor in a sharps container.
- 4 -
2. Check the site for redness, bleeding, irritation, pain, tenderness, or inflammation. Treat based on instructions from your healthcare professional.
3. Insert a new sensor in a different location.
General precautions
Wash your hands with soap and water before inserting the sensor to help prevent site infection.
Wear gloves when inserting the sensor into someone other than yourself to avoid contact with patient blood.
Do not insert the sensor through tape. Inserting the sensor through tape may cause improper sensor insertion and function.
Only use alcohol to prepare the insertion site, to ensure that residue is not left on the skin.
Rotate the sensor insertion site so that sites do not become overused.
Discard used sensors and needle housings in a sharps container after each use to avoid accidental needlestick or puncture.
Do not clean, resterilize, or try to extract the needle from the needle housing. An accidental needlestick or puncture may occur.
Do not reuse sensors. Reuse of a sensor may cause damage to the sensor surface and lead to inaccurate glucose values, site irritation, or infection.
Where to insert the sensor
CAUTION: Avoid the 2 inch (5.0 cm) area around the navel to help ensure a comfortable insertion site and to help with sensor adhesion.
Choose an insertion site that has an adequate amount of subcutaneous fat. Shown here are the best body areas (shaded) for sensor insertion.
Loading...
+ 14 hidden pages