IMPORTANT INFORMATION ON THE ATLANTIS® ANTERIOR CERVICAL PLATE
SYSTEM
PURPOSE
The ATLANTIS® Anterior Cervical Plate System components are temporary implants that are intended for anterior interbody
screw fixation of the cervical spine during the development of a cervical spinal fusion.
DESCRIPTION
The ATLANTIS® Anterior Cervical Plate System consists of a variety of shapes and sizes of bone plates (set screws and
washers are pre-assembled to the plates), screws, and associated instruments. Fixation is provided by bone screws inserted
into the vertebral body of the cervical spine using an anterior approach.
The ATLANTIS® Anterior Cervical Plate System implant components are made from titanium alloy, with certain plates having
subcomponents manufactured from shape memory alloys (Nitinol-NiTi). Stainless steel and titanium implant components must
not be used together in a construct. Implied warranties of merchantability and fitness for a particular purpose or use are
specifically excluded. Do not use any of the ATLANTIS® Anterior Cervical Plate System components with the components from
any other system or manufacturer.
INDICATIONS
Properly used, this system is intended for anterior interbody screw fixation from C2 to T1. The indications and contraindications
of spinal instrumentation systems should be well understood by the surgeon. The system is indicated for use in the temporary
stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease
(as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic
studies), 2) trauma (including fractures), 3) tumors, 4) deformity (defined as kyphosis, lordosis, or scoliosis), 5) pseudarthrosis,
and/or 6) failed previous fusions.
NOTA BENE: This device system is intended for anterior cervical intervertebral body fusions only.
WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical,
thoracic, or lumbar spine.
CONTRAINDICATIONS
Contraindications include, but are not limited to:
▪ Infection local to the operative site.
▪ Signs of local inflammation.
▪ Fever or leukocytosis.
▪ Morbid obesity.
▪ Pregnancy.
▪ Mental illness.
▪ Any medical or surgical condition which would preclude the potential benefit of spinal implant surgery, such as the elevation
of sedimentation rate unexplained by other diseases, elevation of white blood count (WBC), or a marked left shift in the
WBC differential count.
▪ Suspected or documented metal allergy or intolerance.
▪ Any case not needing a bone graft and fusion or where fracture healing is not required.
▪ Any case requiring the mixing of metals from different components.
▪ Any patient having inadequate tissue coverage over the operative site or where there is inadequate bone stock, bone
quality, or anatomical definition.
▪ Any case not described in the Indications.
▪ Any patient unwilling to cooperate with the post-operative instructions.
▪ Any time implant utilization would interfere with anatomical structures or expected physiological performance.
NOTA BENE: Although not absolute contraindications, conditions to be considered as potential factors for not using
this device include:
▪ Severe bone resorption
▪ Osteomalacia
▪ Severe osteoporosis
POTENTIAL ADVERSE EVENTS
All of the possible adverse events associated with spinal fusion surgery without instrumentation are possible. With
instrumentation, a listing of potential adverse events includes, but is not limited to:
▪ Early or late loosening of any or all of the components.
▪ Disassembly, bending, and/or breakage of any or all of the components.
▪ Foreign body (allergic) reaction to implants, debris, corrosion products, graft material, including metallosis, staining, tumor
formation, and/or auto-immune disease.
▪ Pressure on the skin from component parts in patients with inadequate tissue coverage over the implant possibly causing
skin penetration, irritation, and/or pain.
▪ Bursitis and tissue damage caused by improper positioning and placement of implants or instruments.
▪ Post-operative change in spinal curvature, loss of correction, height, and/or reduction.
▪ Infection.
▪ Dural tears.
▪ Loss of neurological function, including paralysis (complete or incomplete), dysesthesias, hyperesthesia, anesthesia,
paraesthesia, appearance of radiculopathy, and/or the development or continuation of pain, numbness, neuroma, or tingling
sensation.
▪ Neuropathy, neurological deficits (transient or permanent), bilateral paraplegia, reflex deficits, and/or arachnoiditis.
▪ Loss of bowel and/or bladder control or other types of urological system compromise.
▪ Scar formation possibly causing neurological compromise around nerves and/or pain.
▪ Fracture, microfracture, resorption, damage, or penetration of any spinal bone and/or bone graft or bone graft harvest site
at, above, and/or below the level of surgery.
▪ Interference with roentgenographic, CT, and/or MR imaging because of the presence of the implants.
▪ Non-union (or pseudarthrosis), delayed union, and mal-union.
▪ Cessation of any potential growth of the operated portion of the spine.
▪ Loss of spinal mobility or function.
▪ Inability to perform the activities of daily living.
▪ Bone loss or decrease in bone density, possibly caused by stress shielding.
▪ Graft donor site complications including pain, fracture, or wound healing problems.
▪ Atelectasis, ileus, gastritis, herniated nucleus pulposus, and/or retropulsed graft.
▪ Hemorrhage, hematoma, seroma, embolism, edema, stroke, excessive bleeding, phlebitis, wound necrosis, wound
dehiscence, or damage to blood vessels.
▪ Gastrointestinal and/or reproductive system compromise, including sterility and loss of consortium.
▪ Development of respiratory problems (e.g., pulmonary embolism, bronchitis, pneumonia, etc.).
▪ Change in mental status.
▪ Death.
Note: Additional surgery may be necessary to correct some of these anticipated adverse events.
WARNINGS AND PRECAUTIONS
A successful result is not always achieved in every surgical case. This fact is especially true in spinal surgery where many
extenuating circumstances may compromise the results. The ATLANTIS® Anterior Cervical Plate System is only a temporary
implant used for the correction and stabilization of the spine. This system is also intended to be used to augment the
development of a spinal fusion by providing temporary stabilization. This device system is not intended to be the sole means of
spinal support. Bone grafting must be part of the spinal fusion procedure in which the ATLANTIS® Anterior Cervical Plate
System is utilized. Use of this product without a bone graft or in cases that develop into a non-union will not be successful. This
spinal implant cannot withstand body loads without the support of bone. In this event, bending, loosening, disassembly and/or
breakage of the device(s) will eventually occur. Preoperative planning and operating procedures including knowledge of surgical
techniques, proper reduction, and proper selection and placement of the implant are important considerations in the successful
utilization of the ATLANTIS® Anterior Cervical Plate by the surgeon. Further, the proper selection and compliance of the patient
will greatly affect the results. Patients who smoke have been shown to have an increased incidence of non-unions. These
patients should be advised of this fact and warned of this consequence. Obese, malnourished, and/or alcohol and/or other drug
abuse patients are also not good candidates for spine fusion. Patients with poor muscle and bone quality and/or nerve paralysis
are also not good candidates for spine fusion. The implants are not prostheses.
A device that has been implanted should never be reused, reprocessed or resterilized under any circumstances. Sterile
packaged devices should also never be resterilized. Reuse, reprocessing, or resterilization may compromise the structural
integrity of these implants and create a risk of contamination of the implants which could result in patient injury, illness, or death.
PHYSICIAN NOTE: Although the physician is the learned intermediary between the company and
the patient, the important medical information given in this document should be conveyed to the
patient.
FOR US AUDIENCES ONLY
CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER
OF A PHYSICIAN.
IMPLANT SELECTION
The selection of the proper size, shape, and design of the implant for each patient is crucial to the success of the procedure.
Metallic surgical implants are subject to repeated stresses in use, and their strength is limited by the need to adapt the design to
the size and shape of human bones. Unless great care is taken in patient selection, proper placement of the implant, and
postoperative management to minimize stresses on the implant, such stresses may cause metal fatigue and consequent
breakage, bending or loosening of the device before the healing process is complete, which may result in further injury or the
need to remove the device prematurely.
PREOPERATIVE
▪ Only patients that meet the criteria described in the indications should be selected.
▪ Patient conditions and/or predispositions such as those addressed in the aforementioned contraindications should be
avoided.
▪ Care should be used in the handling and storage of the implant components. The implants should not be scratched or
otherwise damaged. Implants and instruments should be protected during storage especially from corrosive environments.
▪ The type of construct to be assembled for the case should be determined prior to beginning the surgery. An adequate
inventory of implants should be available at the time of surgery; normally a quantity in excess of what is expected to be
used.
▪ Since mechanical parts are involved, the surgeon should be familiar with the various components before using the
equipment and should personally assemble the devices to verify all parts and necessary instruments are present before the
surgery begins. The ATLANTIS® Anterior Cervical Plate System components are not to be combined with the components
from another manufacturer. Different metal types should not be used together.
▪ All components and instruments should be cleaned and sterilized before use. Additional sterile components should be
available in case of an unexpected need.
INTRAOPERATIVE
▪ Any instruction manuals should be carefully followed.
▪ At all times, extreme caution should be used around the spinal cord and nerve roots. Damage to nerves will cause loss of
neurological functions.
▪ When the configuration of the bone cannot be fitted with an available temporary internal fixation device, and contouring is
absolutely necessary, it is recommended that such contouring be gradual and great care be used to avoid notching or
scratching the surface of the device(s). The components should not be repeatedly or excessively bent any more than
absolutely necessary. The components should not be reverse bent at the same location.
▪ The implant surfaces should not be scratched or notched, since such actions may reduce the functional strength of the
construct.
▪ Bone grafts must be placed in the area to be fused and the graft must be extended from the upper to the lower vertebrae to
be fused.
▪ Bone cement should not be used since this material will make removal of the components difficult or impossible. The heat
generated from the curing process may also cause neurologic damage and bone necrosis.
▪ Before closing the soft tissues, all of the screws should be seated onto the plate. Recheck the tightness of all screws after
finishing to make sure that none has loosened during the tightening of the other screws. Also, secure the locking
mechanism into place to cover the screw heads.
POSTOPERATIVE
The physician’s postoperative directions and warnings to the patient, and the corresponding patient compliance, are extremely
important.
▪ Detailed instructions on the use and limitations of the device should be given to the patient. If partial weight-bearing is
recommended or required prior to firm bony union, the patient must be warned that bending, loosening, or breakage of the
components are complications which can occur as a result of excessive or early weight-bearing or excessive muscular
activity. The risk of bending, loosening, or breakage of a temporary internal fixation device during postoperative
rehabilitation may be increased if the patient is active, or if the patient is debilitated, demented, or otherwise unable to use
crutches or other such weight supporting devices. The patient should be warned to avoid falls or sudden jolts in spinal
position.
▪ To allow the maximum chances for a successful surgical result, the patient or device should not be exposed to mechanical
vibrations that may loosen the device construct. The patient should be warned of this possibility and instructed to limit and
restrict physical activities, especially lifting and twisting motions and any type of sport participation. The patient should be
advised not to smoke or consume alcohol during the bone graft healing process.
▪ The patient should be advised of their inability to bend at the point of spinal fusion and taught to compensate for this
permanent physical restriction in body motion.
▪ If a non-union develops or if the components loosen, bend, and/or break, the device(s) should be revised and/or removed
immediately before serious injury occurs. Failure to immobilize a delayed or non-union of bone will result in excessive and
repeated stresses on the implant. By the mechanism of fatigue, these stresses can cause eventual bending, loosening, or
breakage of the device(s). It is important that immobilization of the spinal surgical site be maintained until firm bony union is
established and confirmed by roentgenographic examination. The patient must be adequately warned of these hazards and
closely supervised to ensure cooperation until bony union is confirmed.
▪ The ATLANTIS
are designed to stabilize the operative site during the normal healing process. After the spine is fused, these devices serve
no functional purpose and should be removed. In most patients, removal is indicated because the implants are not intended
to transfer or support forces developed during normal activities. If the device is not removed following completion of its
intended use, one or more of the following complications may occur: (1) corrosion with localized tissue reaction or pain, (2)
migration of implant position possibly resulting in injury, (3) risk of additional injury from post-operative trauma, (4) bending,
loosening, and/or breakage, which could make removal impractical or difficult, (5) pain, discomfort, or abnormal sensations
due to the presence of the device, (6) possible increased risk of infection, and (7) bone loss due to stress shielding.
®
Anterior Cervical Plate System implants are temporary internal fixation devices. Internal fixation devices
▪ While the surgeon must make the final decision on implant removal, it is the position of the Orthopedic Surgical
Manufacturers Association that whenever possible and practical for the individual patient, bone fixation devices should be
removed once their service as an aid to healing is accomplished, particularly in younger and more active patients. Any
decision to remove the device should take into consideration the potential risk to the patient of a second surgical procedure
and the difficulty of removal. Implant removal should be followed by adequate postoperative management to avoid fracture.
▪ Any retrieved devices should be treated in such a manner that reuse in another surgical procedure is not possible. As with
all orthopedic implants, none of the ATLANTIS® Anterior Cervical Plate System components should ever be reused under
any circumstances.
PACKAGING
Devices may be supplied in a sterile or non-sterile form. Packages for each of the components should be intact upon receipt.
Once the seal on the sterile package has been broken, the product should not be re-sterilized. If a loaner or consignment
system is used, all sets should be carefully checked for completeness and all components should be carefully checked to
ensure that there is no damage prior to use. Damaged packages or products should not be used, and should be returned to
Medtronic.
CLEANING AND DECONTAMINATION
Remove all packaging material prior to disassembly (if applicable), cleaning, and sterilization. All reusable instruments must be
disassembled (if applicable) and thoroughly cleaned prior to sterilization or if applicable, cleaned and decontaminated before
being returned to Medtronic. Always immediately clean and re-sterilize all instruments used in surgery. Cleaning instructions
and associated disassembly instructions (if applicable) can be found at http://manuals.medtronic.com.
Note: certain cleaning solutions such as those containing formalin, glutaraldehyde, bleach, and/or other alkaline cleaners may
damage some devices, particularly instruments. These solutions should
disassembly before cleaning.
All products should be treated with care. Improper use or handling may lead to damage and/or possible improper functioning of
the device.
not be used. Also, many instruments require
STERILIZATION
Unless marked sterile and clearly labeled as such in an unopened sterile package provided by the company, all implants used in
surgery must be sterilized by the hospital prior to use. Remove all packaging materials prior to sterilization. Only sterile products
should be placed in the operative field. Unless specified elsewhere, these products are recommended to be steam sterilized by
the hospital using one of the sets of the following process parameters:
Table 1: Sterilization Cycle Parameters for the United States and its Territories:
*For Medical Facilities Located Outside the United States and its Territories. Some non-U.S. health care authorities
recommend sterilization according to these parameters so as to minimize the potential risk of transmission of Creutzfeldt-Jakob
disease, especially of surgical instruments that could come into contact with the central nervous system.
MINIMUM DRY TIME
1
Table 2: Sterilization Cycle Parameters for Medical Facilities Outside the United States and Its Territories:
The minimum dry times were validated using sterilizers having vacuum drying capabilities. Drying cycles using ambient
atmospheric pressure may require longer dry times. Refer to the sterilizer manufacturer’s recommendations.
NOTE: Because of the many variables involved in sterilization, each medical facility should calibrate and verify the sterilization
process (e.g. temperatures, times) used for their equipment.
PRODUCT COMPLAINTS
Any health care professional (e.g., customer or user of this system of products) who has any complaints or who has
experienced any dissatisfaction in the product quality, identity, durability, reliability, safety, effectiveness, and/or performance,
should notify the distributor or Medtronic. Further, if any of the implanted spinal system component(s) ever “malfunctions” (i.e.,
does not meet any of its performance specifications or otherwise does not perform as intended), or is suspected of doing so, the
distributor should be notified immediately. If any Medtronic product ever malfunctions and may have caused or contributed to
the death or serious injury of a patient, the distributor should be notified immediately by telephone, fax, or written
correspondence. When filing a complaint, provide the component(s) name and number, lot number(s), your name and address,
the nature of the complaint, and notification of whether or not a written report from the distributor is requested.
MRI INFORMATION
The ATLANTIS® Anterior Cervical Plate System has not been evaluated for safety, heating, migration, or compatibility in the
magnetic resonance environment.
FURTHER INFORMATION
Recommended directions for use of this system (surgical operative techniques) are available at no charge upon request. If
further information is needed or required, contact Medtronic.