The following are trademarks or registered trademarks of Medtronic, Inc. in the United States and other countries: Stealth-Midas™, Midas Rex™, StealthStation™, O-arm™, IPC™, ClearView™, and Synergy™. All other
trademarks, service marks, registered trademarks or registered service marks are the property of their respective owners in the United States and other countries.
Stealth-Midas motors
1
1 IPC Connection
2 Cable
3 Tracker
Electric motor: EM200N Pneumatic motor: PM700N
4 Secondary lock
5 Collet
2
1 Hose
2 Tracker
3 Pneumatic motor
Navigable dissecting tools
All tools can be navigated with straight, angled, variable, or double lock attachments.
3
10 cm attachment 14 cm attachment 15 cm attachment
4
4 Collet
5 Foot switch connection
5
6
10BA40D
SP12MH30T
10MH30
ClearView tools Only navigate tools that have Stealth-Midas on the label.
14BA40 14BA50 14MH30
7
15TA23L 15MH22
SP14MH30T

Straight aachment and dissecng tool assembly
1. Slide attachment over the motor collet with arrows aligned (see Step 1).
2. Insert the tool into the attachment (see Step 2). An audible click conrms the tool is seated.
3. Rotate the attachment until the arrow is aligned with the lock symbol (see Step 3). You will hear two clicks as the attachment is rotated. Gently pull the tool to
ensure that it is locked.
4. Rotate the secondary lock by twisting in the direction of the lock arrow (see Step 4).
Step 1 Step 2
Step 3 Step 4
Angled attachment and dissecting tool assembly
1. Insert a dissecting tool into the unlocked attachment (see Step 1).
2. Rotate the tool lock until the arrow is aligned with the lock symbol (see Step 2).
3. Slide the attachment onto the motor collet and rotate to the locked position (see Step 3). Verify that both tool and attachment are now in locked position. See
“Angled attachment orientation” section for orientation relative to tracker.
4. Rotate the secondary lock by twisting in the direction of the lock arrow (see Step 4).
Step 1 Step 2
Step 3 Step 4
1

Angled attachment orientation
Note: The angled attachments can be connected in two dierent orientations.
• To assemble the attachment so that the tracker faces left, insert the attachment facing up relative to the tracker (see Step 1). Rotate the attachment to the
locked position and the tracker will now face to the left (see Step 2).
Step 1 Step 2
Angled Left - Unlocked Angled Left - Locked
• To assemble the attachment so that the tracker faces right, insert the attachment facing down relative to the tracker (see Step 1). Rotate the attachment to the
locked position and the tracker will now face to the right (see Step 2).
Step 1 Step 2
Angled Right - Unlocked Angled Right - Locked
2