The following are trademarks or registered trademarks of Medtronic, Inc. in the United States and other countries: Midas Rex®, Stealth-Midas™, StealthStation™,
IPC™, ClearView™, Legend™, Synergy™, and O-arm™. All other trademarks, service marks, registered trademarks or registered service marks are the property of their
respective owners in the United States and other countries.
Indications for use .................................................................................................................................................................. 3
No Latex Policy ....................................................................................................................................................................... 3
Electric Stealth-Midas motor .............................................................................................................................................. 3
Dissecting tools and attachments .................................................................................................................................... 5
Cleaning and sterilization .................................................................................................................................................. 16
Connection to IPC ............................................................................................................................................................. 4
Instrument management: Adding Stealth-Midas to a procedure .................................................................10
Verifying the Stealth-Midas .........................................................................................................................................11
Selecting tip ......................................................................................................................................................................13
Dissecting toolDisposable cutting tool that connects to Stealth-Midas motor
IPCIntegrated Power Console, used to power the Electric Stealth-Midas motor
Tool listList of navigable items in StealthStation software
Indications for use
The Stealth-Midas System is indicated for the drilling, burring and removal of hard tissue and bone in spinal surgical procedures. Computer-assisted surgery and
its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for
any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, or
vertebra, can be identied relative to a CT- or MR-based model, uoroscopic images, or digitized landmarks of the anatomy.
Device description
The Stealth-Midas is a powered drill that will remove soft tissue, hard tissue, bone, and biomaterials during surgical procedures. The permanently attached tracker
will allow for intraoperative navigation of spinal procedures using the StealthStation.
This device is intended for use by physicians trained in the procedures described.
Warnings
W1 It is important that the Stealth-Midas system operator be familiar with the system User’s Guide, its precautions, procedures and safety issues.
W2 The Stealth-Midas system and its associate applications should be used only by qualied medical professionals who are thoroughly trained and experienced
in performing surgery with Medtronic computer-assisted surgery systems.
W3 The Stealth-Midas system and its associated applications should be used only as an adjunct for surgical guidance. It is not a replacement for the surgeon’s
knowledge, expertise, or judgment.
W4 Do not place Electric Stealth-Midas handpiece in the proximity of magnetic eld, such as magnetic drape and MRI equipment, to avoid inadvertent
handpiece activation.
W5 Ensure that the Reference Frame is properly secured to anatomy before navigating. Neglecting to verify that the Reference Frame is secured could result in
navigational inaccuracy if the hardware moves in relation to the anatomy after registration is complete.
W6 Operate the Stealth-Midas only after appropriate anatomical landmarks and the intended surgical site have been conrmed.
W7 During navigation and after decompression and/or distraction, frequently conrm navigational accuracy by touching the tip of the Stealth-Midas on known
anatomical points and comparing the position of the Stealth-Midas tip in the image with its physical location.
W8 Always inspect the components before and after use for any damage. If damage is observed, do not use damaged part until it is repaired or replaced.
W9 Heavy side loads and/or long operating periods may cause the device to overheat.
W10 Always secure the secondary lock to eliminate motion between the dissecting tool and the tracker, which could impact navigation accuracy.
W11 The Stealth-Midas Motor should only be operated when the attachment is in the locked position.
W12 DO NOT verify the handpiece for navigation with the handpiece running to prevent device damage or deposit metal shavings on the surgical site.
W13 The Stealth-Midas is a precision surgical instrument. With precision instruments, excessive force, beyond the design strength of the instruments, can be
caused by even simple manual overloading.
W14 Do not use excessive force to pry or push bone with the attachment or dissecting tool during surgery.
W15 Carefully inspect dissecting tool both prior to and following each use for signs of excessive wear, fragmentation, eccentricities or other defects. Replace any
suspicious dissecting tools with a new one prior to use.
W16 Use only dissecting tools specically designed for use with this drill system. Match the nomenclature and color code on the tool packaging to the same
nomenclature and color code on the attachment. Conrm that the dissecting tool packaging label has the Stealth-Midas logo () if
navigation is required.
W17 Do not use the Stealth-Midas system before proper cleaning and sterilization.
W18 The tracker is permanently axed to the Stealth-Midas motor and should not be removed at any time.
W19 If system navigation seems inaccurate and recommended steps to restore accuracy are not successful, abort use of the system.
W20 Always re-verify the Stealth-Midas in the software when changing dissecting tools, attachments, or attachment orientation.
W21 Clinical Navigation Software Application prevents the simultaneous localization with multiple trackers with incompatible geometries. Use only one
Navigated Stealth-Midas Motor per procedure.
W22 Do not apply force when verifying dissecting tools.
W23 Refer to the IPC Instructions for Use, Clearview Insert, and Synergy Spine & Trauma Pocket Guide for additional instructions, warnings, and cautions.
No Latex Policy
Medtronic Powered Surgical Solutions guarantees that all Stealth-Midas, Legend equipment and products identied in this manual manufactured by and/or for
Medtronic Powered Surgical Solutions do not contain latex.
Medtronic Powered Surgical Solutions’ packaging materials, labels, pouches, package inserts or similar items do not contain latex.
3
Page 6
Electric Stealth-Midas motor
Components
The Electric Stealth-Midas motor (Figure 1) is a high speed, high torque, navigable electric motor used to dissect bone and biomaterial at selectable speeds from
200 to 75,000 rpm.
The following instructions for the Electric Stealth-Midas are in addition to the general assembly instructions found in the IPC instructions for use. Complete IPC
setup, then continue to the instruction below.
1
Electric Stealth-Midas
1. IPC connection
2. Cable
3. Tracker
2
1. Electric Stealth-Midas motor connection ports
2. Multifunction foot pedal connection port
IPC Connector Panel
4. Secondary lock
5. Collet
Connection to IPC
Locate the Electric Stealth-Midas connection port on the IPC connector
panel (Figure 2) and insert the connector.
Note: Refer to IPC Instructions for Use for more detailed instructions on
operation of IPC and touch-screen.
Technical specications
Electric Stealth-Midas Motor EM200N
Size17.4 cm Length x 1.65 cm Diameter
Weight130g
Speed75,000 rpm Forward/Reverse
Duty Cycle for Applied PartFor use in operating room temperatures up to 40°C (104°F), the Electric Stealth-Midas motor is rated for 3 minutes at 60,000 rpm
followed by 25 minutes of rest.
For normal operating room temperatures (typically 20°C/68°F), the Electric Stealth-Midas motor is rated for cutting indenitely at
60,000 rpm.
4
Page 7
Dissecting tools and attachments
Straight attachment assembly
Caution: Match the nomenclature and color code on the tool packaging to the same nomenclature and color code on the attachment. Conrm that the dissecting
tool packaging label has the Stealth-Midas logo () if navigation is required.
Notes:
• An attachment will not seat on the motor if the arrows on the collet ats are not in alignment.
• The Electric Stealth-Motor will not run properly unless the attachment is in the locked position.
• Conrm that the dissecting tool packaging label has the Stealth-Midas logo (Stealth-Midas) if navigation is required.
• Smoke and/or excessive heat may be generated if the attachment is not in the fully locked position. This may result in thermal injury to the surgeon or sta.
1. Slide a straight attachment over the motor collet aligning triangular arrows on the attachment and the motor case (Figure 3). An audible click, heard and
perceptible by touch, conrms that the attachment is fully seated.
2. Insert the tool into the attachment with a slight rotational motion (Figure 4). An audible click, heard and perceptible to the touch, conrms that the tool is fully
seated.
3
3. Rotate the attachment in the direction indicated by arrow until the attachment alignment mark is directly in line with the locked symbol (Figure 5). You will
hear two clicks as the attachment is rotated.
4. Rotate the secondary lock to the locked position (Figure 6).
5. Gently pull on the tool to ensure that it is locked into the handpiece.
Note: Tool should rotate freely. If not, unlock the assembly, re-seat the tool and re-lock the attachment.
1. Hold the motor in palm of hand. Rotate the secondary lock to the unlocked position.
2. Rotate the attachment to the unlocked position.
3. Remove the dissecting tool from the attachment and discard the tool. For double lock attachments, push the sleeve on the attachment distally before rotating.
4. Use thumb and index nger to lift the attachment o of the motor.
5
Page 8
Angled attachment assembly
Notes:
• A dissecting tool may be installed and locked in the attachment before the angled attachment is installed onto the motor.
• Angled and straight attachments with the same length, marking and color band share the same dissecting tools.
• The Electric Stealth-Midas motor will not run properly unless the attachment is in the locked position.
• Always secure the secondary lock when using angled attachments to eliminate motion between the dissecting tool and the tracker on the motor, which could
impact navigation accuracy.
• Conrm that the dissecting tool packaging label has the Stealth-Midas logo () if navigation is required.
• Smoke and/or excessive heat may be generated if attachment is not in the fully locked position. This may result in thermal injury to the surgeon or sta.
1. With the tool lock in the unlocked position, insert a dissecting tool into the angled attachment with a slight rotational motion (Figure 7). An audible click,
heard and perceptible by touch, conrms that the tool is fully seated.
2. Rotate the tool lock in the direction indicated by arrow until the tool lock alignment mark is directly in line with the locked symbol (Figure 8).
7
3. Gently pull on the tool to ensure that it is locked into the handpiece.
4. Slide the angled attachment over the motor collet aligning triangular arrows on the attachment and the motor case. An audible click, heard and perceptible by
touch, conrms that the attachment is fully seated.
5. Rotate the attachment in the direction indicated by the arrow until attachment alignment mark is directly in line with the locked symbol. You will hear two
clicks as the attachment is rotated.
6. Verify that both the attachment to motor alignment mark and the tool lock alignment mark are directly in line with the locked symbol (Figure 9).
7. Rotate the secondary lock to the locked position (Figure 10). This secondary lock is required to eliminate motion between the dissecting tool and the tracker
on the motor, which could impact navigation accuracy. Toggle the attachment to conrm that the secondary lock is holding the tool securely.
Angled attachment assembly: Tool/attachment lock to motor Secondary lock tightens attachment to motor
8
10
6
Page 9
Angled attachment alignment
Note: Angled attachments can connect in two dierent orientations.
1. To assemble the attachment so that the tracker faces to the left, insert the attachment facing up relative to the tracker. Rotate the attachment in the direction
indicated by the arrow until attachment alignment mark is directly in line with the locked symbol. You will hear two clicks as the attachment is rotated. The
tracker will now face to the left (Figure 11).
2. To assemble the attachment so that the tracker faces to the right, insert the attachment facing down relative to the tracker. Rotate the attachment in the
direction indicated by the arrow until attachment alignment mark is directly in line with the locked symbol. You will hear two clicks as the attachment is
rotated. The tracker will now face to the right (Figure 12).
11
Angled attachment in left orientation Angled attachment in right orientation
UnlockedLockedUnlockedLocked
12
3. Note that the direction of the locked angled attachment will match the label on the tracker (Figure 13). This information will be used when selecting the
appropriate navigation tool card.
13
Angled leftAngled right
Angled attachment direction matches tracker label
7
Page 10
Angled attachment disassembly
1. Rotate the Tool Lock to the unlocked position to remove the tool from the attachment.
2. Rotate the secondary lock to the unlocked position.
3. Rotate the attachment to the unlocked position and lift attachment o of the motor. For double lock attachments, push the sleeve on the attachment distally
before rotating.
ClearView tool assembly
Caution: Conrm that the dissecting tool packaging label has the Stealth-Midas logo () if navigation is required.
1. Slide the ClearView tool over the motor collet aligning triangular arrows on the tool hub and the motor case.
Note: The ClearView tools can connect in two dierent orientations, but will only navigate with the tool facing away from the tracker.
2. To assemble the ClearView tool, insert the tool facing to the left (Figure 14).
3. Rotate the attachment in the direction indicated by the arrow until attachment alignment mark is directly in line with the locked symbol. You will hear two
clicks as the attachment is rotated. The tracker will now face away from the tool (Figure 15).
4. Rotate the secondary lock to the locked position (Figure 16). This secondary lock is required to eliminate motion between the dissecting tool and the tracker
on the motor, which could impact navigation accuracy. Toggle the tool to conrm that the secondary lock is holding the tool securely.
5. Attach irrigation tubing (Figure 17).
14
1617
Unlocked position of ClearView tool Locked position of ClearView tool
Secondary lock attachment to motor ClearView tool with irrigation tubing attached
ClearView attachment disassembly
1. Remove irrigation tubing.
2. Rotate the secondary lock to the unlocked position.
3. Rotate the attachment to the unlocked position on the motor case.
4. Remove the attachment from the motor.
15
18
Navigable dissecting tools and attachments
Dissecting Tools
Use only dissecting tools specically designed for use with this drill
system. Conrm that the dissecting tool packaging label has the
Stealth-Midas logo () if navigation is required
(Figure 18).
Midas tool label with Stealth-Midas logo
Attachments
The Stealth-Midas is compatible with all Midas attachment types that correspond to a navigable tool: straight, angled, variable, or double-locking.
Note:
• Variable attachments do not change the tip location of the dissecting tool.
• Adjusting variable tube exposure will not be represented on navigation image.
Sterile sphere assembly
Place single-use sterile spheres on each of the tracker’s four stems. Push each sphere onto the stem until it “clicks” into place. Make sure the sphere is rmly seated
on the stem before navigating.
8
Page 11
Navigation
The following instructions for the Stealth-Midas are in addition to the general navigation instructions found in the Synergy Spine & Trauma Pocket Guide and
StealthStation System Manual. User should consult the Synergy Spine & Trauma Pocket Guide and StealthStation System Manual for more detailed instruction on
the StealthStation System.
Synergy spine software workow
Complete the following steps before navigating with the Stealth-Midas:
Synergy spine software workow
Select Synergy Spine & Trauma.Select Surgeon.
Select Procedure: Select the procedure type by selecting the imaging
modality: Stealth-Midas uses Optical Tracking.
Verify Instruments: Check that desired Stealth-Midas tool card is selected and
matches physical hardware.
Set up Equipment: Ensure that all lines to essential equipment are green. If
lines are dashed orange, check connections/cables.
Acquire Scan: The system will remain on this screen until the O-arm system
image acquisition step has been performed.
9
Page 12
Instrument management: Adding Stealth-Midas to a procedure
The Verify Instruments screen displays tool cards (Tool List) for the instruments congured for the procedure. The Tool List may be scrolled using the up or down
arrows at the bottom of the list. Select the tool card that matches the attachment and dissecting tool that is connected to the Stealth-Midas motor.
Note: Only one Stealth-Midas dissecting tool and attachment combination can be added to a procedure. Adding a new dissecting tool/attachment combination
will automatically remove any other Stealth-Midas tool that had been selected.
Caution: If a tool or attachment orientation is changed during a procedure, you must return to the Add or Remove Instruments screen to change the tool in the
software.
To add or remove instruments to the procedure, click [Add or Remove Instruments] (Figure 19) to display the Add or Remove Instruments dialog (Figure 20).
19
Verify Instrument Screen: Select Add or Remove instruments if a dierent Stealth-Midas tool is desired
The Instruments Used in Procedure list displays prominently on the left. The Instruments Not Used in Procedure list displays on the right. Each list can be
scrolled using the up or down arrows located at the bottom of each list. Use the search box located at the top of the Instruments Not Used in Procedure list to
search for instruments without scrolling. Click [Search] and enter text for the instrument desired using the virtual keyboard onscreen or the system keyboard. The
list is updated as text is entered or removed from the search box. Click [Done] to return to the Verify Instruments task. To add instruments to the Instruments Not Used in Procedure list, click [Add/Remove at this Site] to open the Admin panel to the Instruments tab (Figure 20).
10
Page 13
Add or Remove Instrument Dialog: If desired instrument is not listed, it can be added by selecting Add/Remove at Site
20
Add or remove instruments used at the site from the Instruments tab. Then click [Done] to return to the Add or Remove Instruments dialog (Figure 21).
21
Add or Remove Instrument from this site: Press done to return to Verify Instruments screen
Verifying the Stealth-Midas
The Verify Instruments screen (Figure 19) displays the tool card for the instruments selected for the procedure. Each tool card indicates whether those
instruments are within the camera’s tracking view (main part of card is green) and whether it is veried (bottom part of the card).
Warnings:
• Physically inspect instruments for any defects. Never attempt to use a damaged instrument.
• Always verify an instrument before use.
• Always re-verify an instrument if you have changed tool or changed attachment orientations.
• Do not apply force when verifying dissecting tool.
11
Page 14
Conrm that the tool card selected matches the physical set-up of the Stealth-Midas (Figure 22).
22
You can verify instruments either now or later at the Navigate task. To verify an instrument:
1. Conrm that the correct reference frame and Stealth-Midas tool cards are pictured in the Verify Instruments screen.
2. Click [Camera] and conrm that both the Stealth-Midas and the reference frame are visible in the Tracking View.
3. Insert the tip of the Stealth-Midas tool into the bottom of the divot in the reference frame (Figure 23). Keep the instrument as perpendicular to the divot as
possible.
4. Hold the Stealth-Midas steady for 2 seconds.
5. When the Stealth-Midas is veried, the system makes a positive sound (chime), the bottom of the instrument’s tool card turns green, and changes to Veried
status. When an instrument verication fails, the system makes a negative sound (bonk).
6. Click the right arrow to advance to the next task.
Conrm that the tool card selected matches the Stealth-Midas hardware
23
Verify Instrument Screen: Stealth-Midas is being inserted into reference frame divot
12
Page 15
Starting navigation
Warning: Frequently conrm navigation accuracy and system responsiveness during live navigation. Use the Stealth-Midas while not rotating to touch several
bony anatomy landmarks and conrm that the locations identied on the images match the locations touched on the patient. If accuracy degrades, re-register the
patient. For O-arm and 3D procedures, re-acquire the exam and re-register the patient.
Caution: Ensure the Reference Frame is properly secured to anatomy. Neglecting to verify that the Reference Frame is secured could result in navigational
inaccuracy if the hardware moves in relation to the anatomy after registration is complete.
The Stealth-Midas with tool and attachment selected for the procedure is veried and automatically tracked when placed in the camera’s eld of view. If you have
not already veried the Stealth-Midas, do so now by holding its tip in the reference frame divot.
During navigation, the software updates the 2D and 3D views in real time to reect changes in the position of the Stealth-Midas’ tip on the anatomy. When
navigation is active, the crosshairs display green. When navigation is frozen, the crosshairs display red. Tracked instrument status indicators display in green at the
bottom right corner of the image area. The names of instruments that are not tracked are displayed in red.
To display the image modication buttons, click the Modify Images tab. Click the tab again to hide the buttons. Refer to the Synergy Spine & Trauma Pocket
Guide for more details.
In the Standard Prole, the default layout for the navigation screen is 4-on-1 with the coronal, axial, sagittal, and 3D model images displayed (Figure 24). To change
the layout, click the Control Panel tab, click [View Settings] and then select a layout. You can customize up to three view settings per procedure for quick cycling.
Use the Trajectory and Probe’s Eye views to visualize progression along the instrument axis (Figure 25).
• Trajectory View 1 is paraxial. The view is dened by a plane rotated 90 from the plane that denes Trajectory View 1.
• Trajectory View 2 is parasagittal or para-coronal. The view is dened by the plane containing the probe and running parallel to the long axis of the patient’s
body.
• The Probe’s Eye View shows the plane that is located at the probe tip, perpendicular to the probe. This perspective mimics what the surgeon would see if
looking down the shaft of the Stealth-Midas.
Click [Select Projection] to add a projection onto the tip of the Stealth-Midas’ virtual display. The tip projection will extend outward along the instrument’s linear
trajectory and is adjustable in length and width using on-screen controls (Figure 26).
Selecting tip
Click [Select Tip] to choose to display the CAD model for the selected Stealth-Midas dissecting tool and attachment (Figure 27).
26
Stealth-Midas with a cylinder projection added Stealth-Midas displayed as a CAD model
27
13
Page 16
Changing views
Click [Cycle Views] to cycle through the image presets selected in View Settings (Figure 28).
28
Click cycle views to display alternate views
Planning
Click [Select Plan] to choose which plan to display on the images. The plans are color coded, and multiple plans can be displayed (Figure 29).
29
Stealth-Midas with plan selected
14
Page 17
Troubleshooting
IssuePossible CauseAction
Motor is too hot to touch/holdInadequate cool down period following
sterilization
Attachment transferring heat to the motor.Switch attachments to determine whether the heat is being
Heavy side loading during dissection.Discontinue use and rest the motor by using it intermittently or
Inadequate irrigationEnsure adequate irrigation to surgical site during bone dissection.
Tool is dicult to remove from attachment Aging of attachment
Use of reprocessed tools
Use of an unauthorized refurbisher
Improper cleaningClean the attachment thoroughly according to the instructions in
Attachment will not seat properly on the
motor
Motor does not runCables not properly connectedEnsure motor and foot control cables are properly connected.
Motor with attachment rotates, but an
abnormal noise is heard.
Motor collet ats are not alignedUse the Stealth-Midas motor wrench to rotate the at closest to
Speed setting is too lowEnsure that a speed greater than 3000 rpm is selected.
Attachment not properly installed and locked
onto the motor
Internal failure of motor and/or console.Change motor or console to isolate the problem.
Foot control not properly functioningCheck for obstruction under the foot pedal.
Cables damagedCheck cable for cracks, splits, or bent connector pins.
Bearings are wornChange the attachment to isolate the location of the problem.
Poor electrical connectionCheck all connections from the electrical source to console.
Internal failure of motor, console, or cableChange motor or console to isolate the failing component.
Attachment not properly installedRemove and reinstall the attachment and dissecting tool.
Motor must be allowed to cool down following steam
sterilization.
generated by the motor or the attachment.
wrap the motor with a moist sterile towel.
Contact Customer Service.
this manual.
Change tool.
the motor case until its marker is aligned with the marker on the
at farthest away from the motor case.
Remove and reinstall the attachment and dissecting tool to
ensure proper installation.
Ensure motor and foot control cables are properly connected.
Clearview tool or irrigation does not run
properly.
System does not navigateSpheres not properly seated on postsReseat spheres
Camera does not see spheresReposition navigation camera. Ensure spheres on Stealth-Midas
Instrument not veriedVerify instrument by placing the Stealth-Midas tip in the
Non-navigated dissecting tool selectedEnsure tool has Stealth-Midas logo on packaging.
Incorrect attachment or tool selected in
software
Incorrect angled attachment orientation
selected
Angled attachment not in fully locked position Conrm that attachment to motor lock and secondary lock are
Incorrect ClearView orientation.Conrm ClearView tool faces away from tracker.
Refer to Midas Rex Clearview User’s Guide.
face the camera and that the sphere surface is clean.
reference frame divot.
Conrm that the dissecting tool in the Stealth-Midas matches
what was selected in the software
Conrm that angled attachment orientation matches what was
selected in the software.
both secure.
15
Page 18
Cleaning and sterilization
Electric Stealth-Midas
Cleaning and Sterilization instructions are subject to change. Up to date instructions are available online at manuals.medtronic.com.
Warnings and Precautions• Do not remove tracker from motor.
• Remove Navigation spheres prior to cleaning and sterilization.
• Do not soak/submerge devices.
• Do not use ultrasonic cleaners to clean devices.
• Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach,
sodium hydroxide, formic acid, or solutions containing glutaraldehyde.
• The use of a washer-disinfector for cleaning may cause a degradation in performance.
• Allow an adequate cooling period after steam sterilization.
LimitationsVerify functionality prior to re-use.
Point of UseFollow hospital procedures.
Containment and
Transportation
Preparation for
Decontamination
Cleaning: Automated (Do NOT
use ultrasonic washer)
Cleaning: Manual• Wipe all external surfaces of the motor and cable with a cloth dampened with a neutral enzymatic detergent, pH 6.0-
Drying and Lubrication1. Thoroughly dry motor. An air gun may be used to blow moisture out.
DisinfectionFollow hospital procedures.
PackagingFor sterilization, place devices in instrument tray. Instrument tray should be wrapped with up to two layers of 1-ply
Sterilization (Temperatures are
minimum required, times are
minimum required)
It is recommended that devices are reprocessed as soon as is practical following use.
Follow hospital procedures.
• Review the washer-disinfector warning above, before using this cleaning method.
• Remove devices from instrument trays before placing into washer baskets.
• Orient devices following recommendations of the washer/disinfector manufacturers.
Recommended
Washer Cycle
8.0.
• Brush motor case and collet with a nylon brush dampened with a neutral enzymatic detergent.
• Rinse motor thoroughly under running water, collet end pointed down. Dry collet and motor with towel.
• Verify that devices are visually clean after manual cleaning.
Note: Medtronic no longer recommends using the cleaning nozzle (PA120), as this may cause some devices to overheat.
2. Using an aerosol spray lubricant (such as Pana Spray), perform the following steps to lubricate motor:
• Holding the can approximately 10-15 cm (3-6 inches) away from the motor, spray all components that move, rotate, or
slide with three quick squirts.
• Remove excess lubricant with a clean cloth.
polypropylene wrap. For use in the US, a FDA-cleared wrap should be used.
Steam Sterilization
CyclePre -VacPre-VacPre-Vac
Temperature132 °C134 °C134 °C
Time4 minutes18 minutes3 minutes
Drying20 minutes20 minutes15 minutes
STERRADDo not use low temperature hydrogen peroxide gas plasma sterilization due to lumen
LIQUID CHEMICAL
STERILIZATION
100% EtO Sterilization Parameters
Preconditioning55-59 °C, 70 ±5 % relative humidity, 30 minutes
Do not use liquid chemical sterilants due to immersion procedure.
Relative Humidity70 +/- 5 %
Ethylene oxide concentration725 +/- 25 mg/L
Gas exposure time (full-cycle)4 hours
Aeration55-59 °C, 18 hours
Detergent Type and Concentration
16
Page 19
Maintenance, Inspection and
Testing
• Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is
repaired.
• Verify functionality prior to re-use.
StorageStore with other sterile devices.
Additional InformationNone
Note: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparing the product for re-use. It remains the responsibility of the processor
to ensure that the reprocessing is actually performed using equipment, materials and personnel in the reprocessing facility to achieve the desired result. This normally requires
validation and routine monitoring of the process.
Note: All validations performed per current AAMI TIR12, Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device
manufacturers.
Medtronic recommends incineration of devices that have directly contacted patients suspected or conrmed with prions or a Transmissible Spongiform Encephalopathy (TSE) such as
Creutzfeldt-Jakob disease (CJD).
Legend Attachments
Cleaning and Sterilization instructions are subject to change. Up to date instructions are available online at manuals.medtronic.com.
Warnings and Precautions• Do not soak/submerge devices.
• Do not use ultrasonic cleaners to clean devices.
• Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach,
sodium hydroxide, formic acid, or solutions containing glutaraldehyde.
• The use of a washer-disinfector for cleaning may cause a premature degradation in performance.
• Allow an adequate cooling period after steam sterilization.
• Use ONLY nylon cleaning brushes. Non-nylon cleaning brushes leave residue that may prevent the tool from being
secured properly in the handpiece.
LimitationsVerify functionality prior to re-use.
Point of UseFollow hospital procedures.
Containment and
It is recommended that devices are reprocessed as soon as is practical following use.
Transportation
Preparation for
Follow hospital procedures.
Decontamination
Cleaning: Automated
(Do NOT use ultrasonic washer)
• Review the washer-disinfector warning above, before using this cleaning method.
• Manually rinse attachments under tap water, until no visible soil is noticed, before placing them into the automatic
washer.
• Remove devices from instrument trays before placing into washer baskets.
• Orient devices following recommendations of the washer/disinfector manufacturers.
Recommended
Washer Cycle
PhaseRecirculation Time Water
Temperature
Detergent Type and Concentration
Pre-Wash2 minutesCold tap waterNot applicable
Wash5 minutes66 °C (set point)Neutral enzymatic detergent, pH 6.0-8.0
Rinse1 minuteHot tap waterNot applicable
Cleaning: Manual1. Wipe all attachments with a cloth, dampened with a surgical instrument cleaning solution.
2. Attachments may be mechanically agitated in cleaning solution, but not soaked or immersed.
3. A nylon brush dampened with a surgical instrument cleaning solution may be used to clean the external surfaces and
internal connecting surfaces of the attachments.
4. Straight attachments have special cleaning brushes sized to the attachment’s internal diameter. Push the brush wet
with surgical instrument cleaning solution through the attachment from rear to front to loosen and remove debris
trapped inside.
5. Move any moveable parts back and forth to allow solution to thoroughly clean attachment.
6. Rinse thoroughly with tap water.
Drying and Lubrication1. Thoroughly dry attachments. An air gun may be used to blow moisture out from rear to front of attachment.
Note: Medtronic no longer recommends using the Stealth-Midas attachment cleaning nozzle (PA120), as this may cause
some attachments to overheat.
2. Using an aerosol spray lubricant (such as Pana Spray), perform the following steps to lubricate attachments:
• Holding the can approximately 10-15 cm (3-6 inches) away from the attachment, spray all components that move,
rotate, or slide with three quick squirts.
• Articulate movable components to ensure proper lubrication.
• Remove excess lubricant with a clean cloth.
DisinfectionFollow hospital procedures.
PackagingFor sterilization, place devices in instrument tray. Instrument tray should be wrapped with up to two layers of 1-ply
polypropylene wrap. For use in the US, a FDA-cleared wrap should be used.
17
Page 20
Sterilization (Temperatures are
minimum required, times are
minimum required)
Maintenance, Inspection and
Testing
StorageStore with other sterile devices.
Additional InformationNone
Note: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparing the product for re-use. It remains the responsibility of the processor
to ensure that the reprocessing is actually performed using equipment, materials and personnel in the reprocessing facility to achieve the desired result. This normally requires validation and routine monitoring of the process.
Note: All validations performed per current AAMI TIR12, Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device
manufacturers.
Steam Sterilization
CyclePre -VacPre-VacPre-Vac
Temperature132 °C134 °C134 °C
Time4 minutes18 minutes3 minutes
Drying15 minutes20 minutes10 minutes
Medtronic recommends incineration of devices that have directly contacted patients suspected or conrmed with prions or a Transmissible
Spongiform Encephalopathy (TSE) such as Creutzfeldt-Jakob disease (CJD).
STERRADDo not use low temperature hydrogen peroxide gas plasma sterilization due to lumen
internal diameter and length restrictions.
LIQUID CHEMICAL
STERILIZATION
100% EtO Sterilization Parameters
Preconditioning55-59 °C, 70 ±5% relative humidity, 30 min
SterilizationTemperature55-59 °C
• Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is
repaired.
• Verify functionality prior to re-use.
Do not use liquid chemical sterilants due to immersion procedure.
Relative Humidity70 ±5 %
Ethylene oxide concentration725 +/- 25 mg/L
Gas exposure time (full-cycle)4 hours
Aeration55-59 °C, 18 hours
Preventative maintenance
For the Electric Stealth-Midas System, the EHS System Preventive Maintenance Manual will assist you in getting the greatest ownership value from your StealthMidas EHS System, while helping to maximize its performance, safety and reliability. The scheduled preventive maintenance/service program is in addition to the
required routine cleaning after each use. Please refer to the preventive maintenance manual for the specic steps necessary to maintain the Electric Stealth-Midas
System.
Limited warranty
A. This Limited Warranty provides the following assurance for the customer who purchases a Medtronic Stealth-Midas System. This Limited Warranty is extended
only to the buyer purchasing the Electric Stealth-Midas directly from Medtronic or from its aliate or its authorized distributor or representative. The Electric
Stealth-Midas includes the motor with integrated tracker and motor cable.
i. Should a System Component fail to function to Medtronic’s published specications during the term of this Limited Warranty [one (1) year from the
date of sale of a new System Component or ninety (90) days from the date of sale of a refurbished or used System Component], Medtronic will either
repair or replace the Motor Component or any portion thereof.
ii. Should a Single Use Component fail to function to Medtronic’s published specications prior to its “use by” date, Medtronic will replace the Single Use
Component.
B. To qualify for this Limited Warranty, the following conditions must be met:
i. The Product must be used on or before its “Use By” or “Use Before” date, if applicable.
ii. The Product must be used in accordance with its labeling and may not be altered or subjected to misuse, abuse, accident or improper handling.
iii. Medtronic must be notied in writing within thirty (30) days following discovery of defect.
iv. The Product must be returned to Medtronic within thirty (30) days of Medtronic receiving notice as provided for in (iii) above.
v. Upon examination of the Product by Medtronic, Medtronic shall have determined that: (i) the Product was not repaired or altered by anyone other
than Medtronic or its authorized representative, (ii) the Product was not operated under conditions other than normal use, and (iii) the prescribed
periodic maintenance and services, if applicable, have been performed on the Product.
C. This Limited Warranty is limited to its express terms. THIS LIMITED WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED WHETHER
STATUTORY OR OTHERWISE, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. In no event shall
Medtronic be liable for any consequential, incidental, prospective or other similar damage resulting from a defect, failure, or malfunction of the IPC System,
whether a claim for such damage is based upon the warranty, contract, negligence or otherwise
D. The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law.
Users may benet from statutory warranty rights under legislation governing the sale of consumer goods. If any part or term of this Limited Warranty is held by
any court of competent jurisdiction to be illegal, unenforceable, or in conict with applicable law, the validity of the remaining portion of the Limited Warranty
shall not be aected, and all rights and obligations shall be construed and enforced as if this Limited Warranty did not contain the particular part or term held
to be invalid.
18
Page 21
Symbols
EUR · USA · JPN · AUS
IPX1
IPX7
DEHP
NIM
EC REP
ACC
STERILE
On O ButtonPackage ContentsFollow Instructions
For Use
EMC Compliance MarkDo Not OilDo Not ImmerseCaution: Federal law
Fuse
Use By Date
Protected Against
Vertical Water Drops
Protected Against The
Eects Of Temporary
Immersion In Water
F
OscillateAuthorized
ForwardPrecaution: Pinch Hazard.
Conforms to ANSI/AAMI
ES 60601-1, IEC/EN
60601-1. Certied to CSA
C22.2 No.601.1
(USA) restricts this device
to sale by or on the order
of a physician.
Representative in the
European Community
Keep Fingers Clear Of
Rollers
>120 VAC
AccessoryType BF Applied Part
AC PowerStart/StopFoot Pedal ConnectorProtective Earth
OutputRF Transmitter
(Interference May
Occur)
Is Approximately Equal
To
Non-SterileStim Bur ConnectorRight Foot Control
Consult Instructions
for Use
R
BUR
Quantity
Not Greater Than
120vac
EHS
NIM Console ConnectorTop Foot Control Unit
Electrical High Speed
Handpiece Connector
Reverse
Fine Irrigant
Adjustment
Left Foot Control Unit
Button / Mode Button
Unit Button / Control
Button
Button / Handpiece
Button
LockedIf the single use symbol
PHT
PHT Contains DEHP (di2-ethyl hexyl phthalate)
Equipotential Ground
Connector
Use With
RoHS - Environmental
friendly use period China (SJ/T11364-2006)
Do Not Dispose Of This
Product In The Unsorted
Municipal Waste
Stream. Dispose Of This
Product According to
Local Regulations. See
Recycling. Medtronic.
Com For Instructions On
Proper Disposal Of This
Product.
is on the device label
then this device is
designed for single
patient use only. Do
not reuse, reprocess, or
resterilize this product.
Reuse, reprocessing,
or resterilization may
compromise the
structural integrity of the
device and/or create a
risk of contamination of
the device, which could
result in patient injury,
illness, or death.
19
Page 22
Symbols
REGULATOR
REF
LOT
STERILE R
SN
!USA
Dissecting Tool
Caution
Fr
UnlockedHandpieceAdapterMotor
World Wide Standard
for Medical Tubing
Diameter
Accessory
Accessory
Regulator
ON
<120s
XX° C
>X° C
L
Applied Part Duty CycleDate of ManufactureAttachment
OFF
>180s
Catalog NumberManufacturer
Lot NumberBrush
Serial NumberErgonomically
engineered
USA Only
Skeeter
1
Pump Head 1
Pump Head 2Instrument Case
Bone Mill
2
Angled attachment
with tracker oriented to
the left
<XX° C
Temperature LimitationCompliant with
XX° C
R
Angled attachment with
tracker oriented to the
right
European Council
Directive MDD 93/42/
EEC.
REFURBISHED
Bone MillControl Unit
Sterilized Using Irradiation
Multi-Use Disposable Attachment
Dissecting Tool
Lubricant/Diuser
Dissecting tool compatible with StealthMidas Navigation
Refurbished
20
Page 23
The information contained in this document is accurate at time of publication. Medtronic reserves the right to make changes to the product described in this
manual. Refer to manuals.medtronic.com for the current version.
Page 24
Medtronic Powered
EC REP
Surgical Solutions
4620 North Beach Street
Fort Worth, TX 76137
USA
medtronic.com
+1 800 643 2773
Authorized Representative in the
European Community