Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and
possibly in other countries. All other trademarks are the property of their respective owners.
Medtronic
1 Medical procedures and EMI precautions
1.1 Introduction
This manual is intended for physicians and other health care professionals who treat patients who have implanted
Medtronic Cardiac Implantable Electronic Devices (CIED)1. To view this manual online or to download it, refer to
the Medtronic Manual Library at www.medtronic.com/manuals.
Section 1.2 of this document is useful to health care professionals who perform medical procedures on patients
with Medtronic implanted cardiac device systems and who consult with the patients’ cardiologists. This section
provides warnings, precautions, and guidance related to medical therapies and diagnostic procedures that may
cause serious injury to a patient, interfere with a Medtronic implanted cardiac device system, or permanently
damage the system. Some common medical procedures that pose no risk are also listed.
Section 1.3 provides precautions and other information related to electromagnetic interference (EMI) that is helpful
to patients in their daily living. Health care professionals can review the information with their patients and use it as
a reference for post-implant consultations.
For additional guidance on medical procedures or potential EMI scenarios you are concerned about that are not
addressed in this manual, customers can contact the following resources:
●
Customers in the United States can contact Medtronic Technical Services at +1 800 723 4636 for ICDs,
or +1 800 505 4636 for pacemakers. You may also submit questions to tshelp@Medtronic.com or your
Medtronic representative.
●
Customers outside of the United States can contact a Medtronic representative.
1.2 Warnings, precautions, and guidance for clinicians performing medical procedures
on cardiac device patients
Ablation (RF ablation or microwave ablation) – Ablation is a surgical technique in which radio frequency (RF)
or microwave energy is used to destroy cells by creating heat. Ablation used in cardiac device patients may result
in, but is not limited to, induced ventricular tachyarrhythmias, oversensing, unintended tissue damage, device
damage, or device malfunction. Pulse-modulated ablation systems may pose higher risk for induced ventricular
tachyarrhythmias. Medtronic cardiac devices are designed to withstand exposure to ablation energy. To mitigate
risks, observe the following precautions:
●
Ensure that temporary pacing and defibrillation equipment is available.
●
Avoid direct contact between the ablation catheter and the implanted system.
●
Position the return electrode patch so that the electrical current pathway does not pass through or near the
device and lead systems.
●
Always monitor the patient during ablation with at least two separate methods, such as arterial pressure
display, ECG, manual monitoring of the patient’s rhythm (taking pulse) or monitor by some other means such
as ear or finger pulse oximetry, or Doppler pulse detection.
To avoid or mitigate the effects of oversensing, consider the following precautions:
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. If a programmer is used and ablation
causes a device reset, the cardiac device resumes detection. After the ablation procedure, remove the magnet
or restore device parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for example, DOO). After
the ablation procedure, restore device parameters.
Capsule endoscopy, pH capsule procedures – Capsule endoscopy is a procedure in which a capsule
containing a tiny camera is swallowed by the patient to take pictures of the patient’s digestive tract. Capsule
endoscopy and pH capsule procedures should pose no risk of electromagnetic interference.
Dental procedures – Dental equipment, such as ultrasonic scalers, drills, and pulp testers, poses no risk of
electromagnetic interference. Keep a cardiac device at least 15 cm (6 in) away from magnets, such as magnets
found in dental office pillow headrests.
Diagnostic radiology (CT scans, fluoroscopy, mammograms, x-rays) – Diagnostic radiology refers to the
following medical procedures:
●
Computerized axial tomography (CT or CAT scan)
●
Fluoroscopy (an x-ray procedure that makes it possible to see internal organs in motion by producing a video
image)
●
Mammograms
●
X-rays (radiography, such as chest x-rays)
Normally, the accumulated dose from diagnostic radiology is not sufficient to damage the device. If the device is
not directly exposed to the radiation beam, no risk of interference with device operation occurs. However, if the
device is directly in a CT scan beam, see the following precautions in “CT scan”. Similar interference may be
observed for some forms of high-intensity fluoroscopy.
CT scan – A CT scan is a computerized process in which two-dimensional x-ray images are used to create a
three-dimensional x-ray image. If the device is not directly in the CT scan beam, the device is not affected. If the
device is directly in the CT scan beam, oversensing may occur for the duration of time the device is in the beam.
If the device will be in the beam for longer than 4 s, to avoid or mitigate the effects of oversensing, consider the
following precautions:
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. After completing the CT scan,
remove the magnet or restore device parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for example, DOO). After
completing the CT scan, restore device parameters.
Diagnostic ultrasound – Diagnostic ultrasound is an imaging technique that is used to visualize muscles and
internal organs, their size, structures, and motion as well as any pathological lesions. It also is used for fetal
monitoring and to detect and measure blood flow. Diagnostic ultrasound, such as echocardiogram, poses no risk
of electromagnetic interference. For precautions about therapeutic ultrasound, see “Diathermy treatment
(including therapeutic ultrasound)”.
Diathermy treatment (including therapeutic ultrasound) – Diathermy is a treatment that involves the
therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and
therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device
patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system.
Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused
ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the
body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm
(6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing
away from the device and lead system.
Electrolysis – Electrolysis is the permanent removal of hair by using an electrified needle (AC or DC) that is
inserted into the hair follicle. Electrolysis introduces electrical current into the body, which may cause oversensing.
Evaluate any possible risks associated with oversensing with the patient’s medical condition. To avoid or mitigate
the effects of oversensing, consider the following precautions:
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. After completing electrolysis, remove
the magnet or restore device parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for example, DOO). After
completing electrolysis, restore device parameters.
4
Electrosurgery – Electrosurgery (including electrocautery, electrosurgical cautery, Medtronic Advanced Energy
surgical incision technology, and hyfrecator) is a process in which an electric probe is used to control bleeding, to
cut tissue, or to remove unwanted tissue. Electrosurgery used on cardiac device patients may result in, but is not
limited to, oversensing, unintended tissue damage, tachyarrhythmias, device damage, or device malfunction. If
electrosurgery cannot be avoided, consider the following precautions:
●
Ensure that temporary pacing and defibrillation equipment is available.
●
Use a bipolar electrosurgery system or Medtronic Advanced Energy surgical incision technology, if possible.
If a unipolar electrosurgery system is used, position the return electrode patch so that the electrical current
pathway does not pass through or within 15 cm (6 in) of the device and lead system.
●
Do not apply unipolar electrosurgery within 15 cm (6 in) of the device and lead system.
●
Use short, intermittent, and irregular bursts at the lowest clinically appropriate energy levels.
●
Always monitor the patient during electrosurgery. If the ECG tracing is not clear due to interference, manually
monitor the patient’s rhythm (take pulse); alternatively, monitor by some other means such as ear or finger
pulse oximetry, Doppler pulse detection, or arterial pressure display.
To avoid or mitigate the effects of oversensing, consider the following precautions:
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. If a programmer is used and
electrosurgery causes a device reset, the cardiac device resumes detection. After completing electrosurgery,
remove the magnet or restore device parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for example, DOO). After
completing electrosurgery, restore device parameters.
External defibrillation and cardioversion – External defibrillation and cardioversion are therapies that deliver
an electrical shock to the heart to convert an abnormal heart rhythm to a normal rhythm.
Medtronic cardiac devices are designed to withstand exposure to external defibrillation and cardioversion. While
damage to an implanted system from an external shock is rare, the probability increases with increased energy
levels. These procedures may also temporarily or permanently elevate pacing thresholds or temporarily or
permanently damage the myocardium. If external defibrillation or cardioversion are required, consider the
following precautions:
●
Use the lowest clinically appropriate energy.
●
Position the patches or paddles a minimum of 15 cm (6 in) away from the device.
●
Position the patches or paddles perpendicular to the device and lead system.
●
If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device, use a Medtronic
programmer to evaluate the device and lead system.
Hyperbaric therapy (including hyperbaric oxygen therapy, or HBOT) – Hyperbaric therapy is the medical use
of air or 100% oxygen at a higher pressure than atmospheric pressure. Hyperbaric therapies with pressures
exceeding 4.0 ATA, approximately 30 m (100 ft) of seawater, may affect device function or cause device damage.
To avoid or mitigate risks, do not expose implanted devices to pressures exceeding 4.0 ATA.
Lithotripsy – Lithotripsy is a medical procedure that uses mechanical shock waves to break up kidney or
gallbladder stones. If the device is at the focal point of the lithotripter beam, lithotripsy may permanently damage
the device. If lithotripsy is required, keep the focal point of the lithotripter beam a minimum distance of 2.5 cm (1 in)
away from the device. To avoid or mitigate the effects of oversensing, consider the following precautions:
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. After completing lithotripsy
treatment, remove the magnet or restore device parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for example, DOO). After
completing lithotripsy treatment, restore device parameters.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create
an internal view of the body. Do not conduct MRI scans on patients who have this device or lead implanted. MRI
scans may result in serious injury, induction of tachyarrhythmias, or implanted system malfunction or damage.
5
Radiotherapy – Radiotherapy is a cancer treatment that uses radiation to control cell growth. When performing
radiotherapy, take precautions to avoid oversensing, device damage, and device operational errors, as described
in the following sections:
●
Oversensing – If the patient undergoes radiotherapy treatment and the average dose rate at the device
exceeds 1 cGy/min, the device may inappropriately sense direct or scattered radiation as cardiac activity for
the duration of the procedure. To avoid or mitigate the effects of oversensing, consider these precautions:
– Suspend tachyarrhythmia detection by using a magnet or a programmer. After completing radiotherapy
treatment, remove the magnet or restore device parameters.
– If appropriate for the patient, program the device to an asynchronous pacing mode (for example, DOO).
After completing radiotherapy treatment, restore device parameters.
●
Device damage – Exposing the device to high doses of direct or scattered radiation from any source that
results in an accumulated dose greater than 500 cGy may damage the device. Damage may not be
immediately apparent. If a patient requires radiation therapy from any source, do not expose the device to
radiation that exceeds an accumulated dose of 500 cGy. To limit device exposure, use appropriate shielding
or other measures. For patients who are undergoing multiple radiation treatments, consider the accumulated
dose to the device from previous exposures.
Note: Normally, the accumulated dose from diagnostic radiology is not sufficient to damage the device. See
“Diagnostic radiology” for precautions.
●
Device operational errors – Exposing the device to scattered neutrons may cause electrical reset of the device,
errors in device functionality, errors in diagnostic data, or loss of diagnostic data. To help reduce the chance
of electrical reset due to neutron exposure, deliver radiotherapy treatment by using photon beam energies less
than or equal to 10 MV. The use of conventional x-ray shielding during radiotherapy does not protect the device
from the effects of neutrons. If photon beam energies exceed 10 MV, Medtronic recommends interrogating the
device immediately after radiotherapy treatment. An electrical reset requires reprogramming of device
parameters. Electron beam treatments that do not produce neutrons do not cause electrical reset of the
device.
Stereotaxis – Stereotaxis is a catheter navigation platform that allows clinicians to steer catheter-based
diagnostic and therapeutic devices throughout the body by using magnetic navigation. During a stereotaxis
procedure, the magnetic field may activate the magnet detection sensor in the implanted device, which suspends
tachyarrhythmia detection. The device resumes normal programmed operation after the procedure.
Transcutaneous electrical nerve stimulation (TENS) – TENS (including neuromuscular electrical stimulation
or NMES) is a pain control technique that uses electrical impulses passed through the skin to stimulate nerves. A
TENS device is not recommended for in-home use by cardiac device patients due to a potential for oversensing,
inappropriate therapy, inhibition of pacing, or asynchronous pacing. If a TENS device is determined to be medically
necessary, contact a Medtronic representative for more information.
Transurethral needle ablation (TUNA) and Transurethral Microwave Therapy (TUMT) – TUNA and TUMT
are surgical procedures used for benign prostatic hyperplasia (BPH) in which precisely focused energy is used to
ablate prostate tissue. Patients with implanted cardiac devices may conditionally undergo procedures that use a
TUNA or TUMT system. To avoid affecting cardiac device function when performing a TUNA or TUMT procedure,
position the return electrode on the lower back or lower extremity at least 15 cm (6 in) away from the implanted
device and lead system.
1.3 Warnings, precautions, and guidance related to electromagnetic interference (EMI)
for cardiac device patients
Many cardiac device patients resume their normal daily activities after full recovery from surgery. However, there
may be certain situations that patients need to avoid. Because a cardiac device is designed to sense the electrical
activity of the heart, the device may sense a strong electromagnetic energy field outside of the body and deliver
a therapy that is not needed or withhold a therapy that is needed. The following sections provide important
information to share with patients about electrical equipment or environments that may cause interference with
their implanted cardiac device.
6
For additional guidance about EMI, customers can contact the following resources:
●
Customers in the United States can contact either of the following telephone numbers: for pacemakers,
contact Medtronic Technical Services at +1 800 505 4636; for ICDs, contact Medtronic Technical Services
at +1 800 723 4636. You may also submit questions to tshelp@Medtronic.com or your Medtronic
representative.
●
Customers outside of the United States can contact a Medtronic representative.
General EMI guidelines for patients – Patients should observe the following general guidelines regarding EMI:
●
Area restrictions – Before entering an area where signs are posted prohibiting persons with an implanted
cardiac device, such as a pacemaker or ICD, consult with your doctor.
●
Symptoms of EMI – If you become dizzy or feel rapid or irregular heartbeats while using an electrical item,
release whatever you are touching or move away from the item. The cardiac device should immediately return
to normal operation. If symptoms do not improve when you move away from the item, consult with your doctor.
If you have an ICD and you receive a therapy shock while using an electrical item, release the item or move
away from it, then consult with your doctor.
●
Proper grounding of electrical items – To avoid interference from electrical current that may leak from
improperly grounded electrical items and pass through the body, observe the following precautions:
– Make sure that all electrical items are properly wired and grounded.
– Make sure that electrical supply lines for swimming pools and hot tubs are properly installed and grounded
according to local and national electrical code requirements.
Wireless communication devices – Wireless communication devices include transmitters that can affect
cardiac devices. When using wireless communication devices, keep them at least 15 cm (6 in) away from your
cardiac device. The following items are examples of such devices:
●
Handheld cellular, mobile, or cordless telephones (wireless telephones); two-way pagers; personal digital
assistants (PDAs); smartphones; and mobile email devices
●
Wireless-enabled devices such as laptop, notebook, or tablet computers; network routers; MP3 players;
e-readers; gaming consoles; televisions; DVD players; and headsets
●
Remote keyless entry and remote car starter devices
Using wireless telephones – Cardiac devices have been tested with many types of wireless telephone
technologies to ensure that they operate correctly while a wireless phone is in use. Keep a cardiac device at least
15 cm (6 in) away from the antenna of a handheld wireless telephone (for example, by holding the telephone to the
ear farthest away from the cardiac device). Do not carry the telephone in a pocket over the device or in a shoulder
bag near the device.
Household and hobby items with motors or magnets and other items that cause EMI – Household and
hobby items that have motors or magnets or that generate electromagnetic energy fields could interfere with a
cardiac device. Keep a cardiac device at least 15 cm (6 in) away from the following items:
●
Handheld kitchen appliances, such as electric mixers
●
Sewing machines and sergers
●
Personal care items, such as corded handheld hair dryers, corded electric shavers, electric or ultrasonic
toothbrushes (base charger), or back massagers
●
Items that contain magnets, such as bingo wands, mechanic’s extractor wands, magnetic bracelets, magnetic
clasps, magnetic chair pads, or stereo speakers
●
Remote controller of radio-controlled toys
●
Two-way walkie-talkies (less than 3 W)
The following household and hobby items require special precautions:
●
Boat motors – Keep a cardiac device at least 30 cm (12 in) away from electric trolling motors or
gasoline-powered boat motors.
●
Electronic body fat scale – Using this type of scale is not recommended for cardiac device patients because
it passes electricity through the body and can interfere with the device.
●
Electronic pet fences or invisible fences – Keep a cardiac device at least 30 cm (12 in) away from the buried
wire and the indoor antenna of electronic pet fences or invisible fences.
●
Handheld metal detectors – Keep a cardiac device at least 60 cm (24 in) away from the detector end.
7
●
Home-use electric kilns – Keep a cardiac device at least 60 cm (24 in) away from home-use electric kilns.
●
Induction cook tops – An induction cook top uses an alternating magnetic field to generate heat. Keep a
cardiac device at least 60 cm (24 in) away from the heating zone when the induction cook top is turned on.
●
Magnetic mattress pads or pillows – Items containing magnets can interfere with the normal operation of a
cardiac device if they are within 15 cm (6 in) of the device. Avoid using magnetic mattress pads or pillows
because they cannot easily be kept away from the device.
●
Portable electric generators up to 20 kW – Keep a cardiac device at least 30 cm (12 in) away from portable
electric generators.
●
UPS (uninterruptible power source) up to 200 A – Keep a cardiac device at least 30 cm (12 in) away from a UPS.
If the UPS is operating by battery source, keep a cardiac device at least 45 cm (18 in) away.
Home power tools – Most home power tools should not affect cardiac devices. Consider the following
common-sense guidelines:
●
Keep all equipment in good working order to avoid electrical shock.
●
Be certain that plug-in tools are properly grounded (or double insulated). Using a ground fault interrupter outlet
is a good safety measure (this inexpensive device prevents a sustained electrical shock).
Some home power tools could affect cardiac device operation. Consider the following guidelines to reduce the
possibility of interference:
●
Electric yard and handheld power tools (plug-in and cordless) – Keep a cardiac device at least 15 cm (6 in)
away from such tools.
●
Soldering guns and demagnetizers – Keep a cardiac device at least 30 cm (12 in) away from these tools.
●
Gasoline-powered tools and gasoline-powered yard equipment – Keep a cardiac device at least 30 cm (12 in)
away from components of the ignition system. Turn off the motor before making adjustments.
●
Car engine repair – Turn off car engines before making any adjustments. When the engine is running, keep a
cardiac device at least 30 cm (12 in) away from components of the ignition system.
Industrial equipment – After recovering from implant surgery, you likely will be able to return to work, school, or
daily routine. However, if you will be using or working near high-voltage equipment, sources of high electrical
current, magnetic fields, or other EMI sources that may affect device operation, consult with your doctor. You may
need to avoid using, or working near, the following types of industrial equipment:
●
Electric furnaces used in the manufacturing of steel
●
Induction heating equipment and induction furnaces, such as kilns
●
Industrial magnets or large magnets, such as those used in surface grinding and electromagnetic cranes
●
Dielectric heaters used in industry to heat plastic and dry glue in furniture manufacturing
●
Electric arc and resistance welding equipment
●
Broadcasting antennas of AM, FM, shortwave radio, and TV stations
●
Microwave transmitters. Note that microwave ovens are unlikely to affect cardiac devices.
●
Power plants, large generators, and transmission lines. Note that lower voltage distribution lines for homes and
businesses are unlikely to affect cardiac devices.
Radio transmitters – Determining a safe distance between the antenna of a radio transmitter and a cardiac
device depends on many factors such as transmitter power, frequency, and the antenna type. If the transmitter
power is high or if the antenna cannot be directed away from a cardiac device, you may need to stay farther away
from the antenna. Refer to the following guidelines for different types of radio transmitters:
●
Two-way radio transmitter (less than 3 W) – Keep a cardiac device at least 15 cm (6 in) away from the antenna.
●
Portable transmitter (3 to 15 W) – Keep a cardiac device at least 30 cm (12 in) away from the antenna.
●
Commercial and government vehicle-mounted transmitters (15 to 30 W) – Keep a cardiac device at least
60 cm (24 in) away from the antenna.
●
Other transmitters (125 to 250 W) – Keep a cardiac device at least 2.75 m (9 ft) away from the antenna.
For transmission power levels higher than 250 W, contact a Medtronic representative for more information.
8
Security systems – When passing through security systems, follow these precautions:
●
Electronic antitheft systems, such as in a store or library, and point-of-entry control systems, such as gates or
readers that include radio frequency identification equipment – These systems should not affect a cardiac
device, but as a precaution, do not linger near or lean against such systems. Simply walk through these
systems at a normal pace. If you are near an electronic antitheft or entry control system and experience
symptoms, promptly move away from the equipment. After you move away from the equipment, the cardiac
device resumes its previous state of operation.
●
Airport, courthouse, and jail security systems – Given the short duration of security screening, it is unlikely that
metal detectors (walk-through archways and handheld wands) and full body imaging scanners (also called
millimeter wave scanners and three-dimensional imaging scanners) in airports, courthouses, and jails will
affect a cardiac device. When encountering these security systems, follow these guidelines:
– Always carry your cardiac device ID card. If a cardiac device sets off a metal detector or security system,
show your ID card to the security operator.
– Minimize the risk of temporary interference with your cardiac device while going through the security
screening process by not touching metal surfaces around any screening equipment.
– Do not stop or linger in a walk-through archway; simply walk through the archway at a normal pace.
– If a handheld wand is used, ask the security operator not to hold it over or wave it back and forth over your
cardiac device.
– If you have concerns about security screening methods, show your cardiac device ID card to the security
operator, request alternative screening, and then follow the security operator’s instructions.
Toll-free in the USA (24-hour technical
consultation for physicians and medical
professionals)
Bradycardia: +1 800 505 4636
Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl
Route du Molliau 31
Case Postale 84
CH-1131 Tolochenaz
Switzerland
+41 21 802 7000