CoreValve™ Evolut™ PRO System
CAUTION: Investigational device. Limited by Federal (or United States) law to investigational use.
See below or contact LifeLine Technical Support (1 877 526 7890) for MRI safety information for the following bioprosthesis models:
TAV-MDT2-23-C
TAV-MDT2-26-C
TAV-MDT2-29-C
MRI safety information
Nonclinical testing and modeling have demonstrated that the bioprosthesis is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:
Static magnetic field of 1.5 T and 3.0 T
Maximum spatial field gradient of 2500 gauss/cm (25 T/m)
Maximum MR system reported, whole body averaged specific absorption rate (SAR)
of 2.0 W/kg (Normal Operating Mode)
Based on nonclinical testing and modeling, under the scan conditions defined above, the bioprosthesis is expected to produce a maximum in vivo temperature rise of less than
4.0°C after 15 minutes of continuous scanning. Based on nonclinical data, the image artifact caused by the device will extend no greater than 7 mm from the bioprosthesis when imaged with a gradient echo pulse sequence and a 3.0 T MRI system.
Scanning under the conditions defined above may be performed immediately after implantation.
The presence of other implants or medical circumstances of the patient may require lower limits on some or all of the above parameters.