Medtronic DS1TD32 Instructions for Use

Instructions for Use 175042EN E
Midas Rex™ Legend™
Depth Stop Attachment and Tools
EN To obtain a copy of this manual in your local language contact your local Medtronic representative or go to
manuals.medtronic.com.
CS Máte-li zájem okopii této příručky ve svém jazyce, kontaktujte místního zástupce společnosti Medtronic nebo
navštivte webové stránky manuals.medtronic.com.
manuals medtronic.com.
DE Eine Kopie dieses Handbuchs in der jeweiligen Landessprache kann vom örtlichen Medtronic-Vertreter
angefordert oder unter manuals.medtronic.com bezogen werden.
EL Για να λάβετε αντίγραφο του παρόντος εγχειριδίου στην τοπική σας γλώσσα, επικοινωνήστε με τον τοπικό
αντιπρόσωπο της Medtronic ή επισκεφθείτε την ιστοσελίδα manuals.medtronic.com.
ES Para obtener una copia de este manual en su idioma local, póngase en contacto con su representante local de
Medtronic o visite manuals.medtronic.com.
FI Jos haluat tämän käyttöoppaan omalla kielelläsi, ota yhteyttä paikalliseen Medtronicin edustajaan tai siirry
osoitteeseen manuals.medtronic.com.
FR Pour obtenir un exemplaire de ce manuel dans votre langue, veuillez contacter votre représentant Medtronic
local ou visiter la page manuals.medtronic.com.
HR Da biste dobili kopiju ovog priručnika na svom jeziku, obratite se lokalnom zastupniku tvrtke Medtronic ili
posjetite web-mjesto manuals.medtronic.com.
HU Ha szeretne egy saját nyelvű példányt a jelen kézikönyvből, vegye fel a kapcsolatot a Medtronic helyi
képviseletével, vagy látogasson el a manuals.medtronic.com weboldalra.
IT Per ottenere una copia del presente manuale nella lingua locale, contattare il rappresentante Medtronic di
zona o visitare la pagina manuals.medtronic.com.
KK
Осы нұсқаулықтың өзіңіздің жергілікті тіліңіздегі көшірмесін алу үшін, жергілікті Medtronic өкіліне хабарласыңыз немесе мына веб-сайтқа өтіңіз: manuals.medtronic.com.
MK За да добиете копија од овој прирачник на вашиот локален јазик, контактирајте со вашиот локален
претставник на Medtronic или појдете на manuals.medtronic.com.
NL Neem contact op met uw vertegenwoordiger van Medtronic of ga naar manuals.medtronic.com voor een
exemplaar van deze handleiding in uw taal.
NO Du kan få et eksemplar av denne håndboken på ditt språk ved å kontakte din lokale Medtronic-representant
eller gå til manuals.medtronic.com.
PL Aby uzyskać egzemplarz niniejszego podręcznika w wybranym języku, należy się skontaktować z lokalnym
przedstawicielem rmy Medtronic lub odwiedzić stronę manuals.medtronic.com.
PT-BR Para obter uma cópia deste manual em seu idioma local, entre em contato com o representante local da
Medtronic ou acesse manuals.medtronic.com.
PT-PT Para obter uma cópia deste manual no seu idioma local, contacte o seu representante local Medtronic ou visite
manuals.medtronic.com.
RO Pentru a obține o copie a acestui manual în limba locală, contactați reprezentantul local Medtronic sau
accesați manuals.medtronic.com.
RU Для получения экземпляра этого руководства на вашем языке обратитесь к местному представителю
Medtronic или посетите сайт manuals.medtronic.com.
SK Ak chcete získať kópiu tejto príručky vo svojom jazyku, kontaktujte miestneho zástupcu spoločnosti
Medtronic alebo navštívte stránku manuals.medtronic.com.
SL Če želite izvod tega priročnika v svojem jeziku, se obrnite na lokalnega predstavnika za Medtronic ali pa ga
poiščite na spletnem mestu manuals.medtronic.com.
SR Da biste dobili kopiju ovog priručnika za korisnike na svom jeziku, kontaktirajte lokalno predstavništvo
kompanije Medtronic ili posetite lokaciju manuals.medtronic.com.
SV För att erhålla en kopia av denna handbok på ditt språk kontaktar du din lokala representant för Medtronic eller går
till manuals.medtronic.com.
TR Bu kılavuzun kendi dilinizde bir kopyasını almak için, bölgenizdeki Medtronic temsilcinize başvurun veya
manuals.medtronic.com adresine gidin.
UK Щоб отримати копію цього посібника вашою місцевою мовою, зв’яжіться зі своїм місцевим
представником Medtronic або відвідайте веб-сайт manuals.medtronic.com.
Customer service
For further information regarding the use of this product or to report any problems, please contact Medtronic Customer Service, or contact your local distributor:
Medtronic Powered Surgical Solutions
4620 North Beach Street Fort Worth, Texas 76137 USA US Help Line (800-335-9557)
RS.DFWrepairs@medtronic.com
International Service: International Customers should contact their local Medtronic representative. The information contained in this document is accurate at time of publication. Medtronic reserves the right to make
changes to the product described in this manual. Refer to manuals.medtronic.com for the current version.
The following are trademarks or registered trademarksof Medtronic, Inc. in the United States and other countr ies: IPC™, Legend™, Midas Rex™, MR7™, Visualase™, and Stealth Autoguide™. All other trademarks, service marks, registered trademarks or registered service marks are the property of their respective owners in the United States and other countries.
EN
EN NOTE: Attach these instructions to your IPC/Legend/MR7/Stealth Autoguide/Visualase Instruction
Manual or User’s Guide. Refer to your instruction manual or user’s guide for additional indications, contraindications, safety precautions, and more detailed instructions for use. This attachment is designed for use with the Legend depth stop dissecting tools, the Stealth Autoguide, and the Legend electric and pneumatic motors. Refer to the dissecting tool’s labeling for information on specic operating requirements and limitations, and to the Stealth Autoguide manual for interface details.
Indications for use
The Midas Rex™ Legend™ depth stop attachment and tools are indicated for the incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.
Device description
The Midas Rex™ Legend™ depth stop attachment and tools are intended to create a cranial access hole of a known diameter and depth.
Contraindications
None.
Warnings
W1 The attachment shall only be used after the proper cleaning and sterilization protocol has been
followed.
W2 Failure to fully lock the attachment and dissecting tools may lead to product detachment and/
or patient injury.
W3 Drilling with trajectories above the angle limits provided in Table 2 may lead to damage of sub-
cranial tissue and patient injury.
W4 Verify the drill depth setting on the attachment prior to drilling. Contacting the brain or other
tissue with a rotating dissecting tool may lead to patient or user injury. W5 Applying high lateral forces while drilling may result in loss of accuracy and/or injury to the patient. W6 Dissecting tools are single use only. Reprocessing and re-using a dissecting tool may cause
injury to the patient. W7 Perform proper ushing and irrigation during drilling to prevent entrance of debris into the
brain and potentially cause injury. W8 Inspect the attachment before and after each use for damage, wear, or illegible markings. W9 Using an attachment and/or dissecting tool that appears unstable may cause loss of accuracy
and/or patient injury. W10 Using interface products others than those approved for use in this IFU may lead to serious
patient injury. W11 Selec ting the wrong dissec ting tool may cause instrumentation incompatibility and/or loss of accuracy. W12 Contact between a rotating dissecting tool and the drill guide or other equipment may damage
the dissecting tool and may cause dissecting tool instability or loss of accuracy. W13 Excessive drilling forces may cause dissecting tool instability or loss of accuracy. W14 The dissecting tool performance may be compromised when operated outside of the recommended
speed or pressure setting and may cause patient injury.
Installation
Note: If any damage, wear, rust, or corrosion is observed with the product or if action is required beyond
cleaning, do not use the product and return to Medtronic for service.
1. Slide attachment over the motor collet, aligning the triangular arrows on the attachment and the motor case. An audible click, heard and perceptible by touch, conrms that the attachment is fully seated (Figure 1).
1
1
EN
2. Insert the tool into the attachment with a slight
2
rotational motion (Figure 2). An audible click, heard and perceptible by touch, conrms that the tool is fully seated.
3. Rotate the attachment in the direction indicated
3
by arrow on the attachment until the attachment alignment mark is directly in line with the locked symbol (Figure 3). You will hear two clicks as the attachment is rotated.
4. Gently pull on the tool to ensure that it is locked into the assembly. Free run the drill assembly prior to use to check for proper function and dissecting tool stability.
Drilling
The Legend depth stop attachment and dissecting tools are intended to be used with the Stealth Autoguide drill guides to create cranial access holes. Refer to Table 1 for procedure, dissecting tool, and drill guide compatibility.
Table 1: Drill guide compatibility
Procedure Attachment Dissecting tool Hole diameter
Biopsy
Biopsy DS1TD75 7.5mm 28322
ASDS01
sEEG bolt placement DS1TD24 2.4mm 28326
Visualase bone anchor placement DS1TD32L 3.2mm 28324
DS1TD32 3.2mm 28326
Stealth Autoguide drill
guide number
After following installation steps 1 through 4, prepare for drilling with the following steps:
1. To adjust the drill depth setting, rotate collar on
4
the attachment so that the arrow aligns with the desired drill depth (Figure 4).
Note: The numbers correspond to the drill depth
Collar
in millimeters.
2. Insert the drill assembly into corresponding drill
5a
5b
guide per Table 1 for the given procedure (Figure 5a).
3. Power the drill assembly per the following: (1) For electric motors: set the Integrated Power
Console (IPC) speed between 60,000 RPM to 75,000 RPM
(2) For pneumatic motors: set the pneumatic foot
(PCU) control pressure between 80 psi to 120 psi
4. Provide a slight forward drilling force until the shoulder of the attachment comes in contact with the top of the drill guide (Figure 5b). Use irrigation
Drill guide
Shoulder
while drilling.
5. Stop Legend motor before removing drill assembly from drill guide.
6. Verify that the hole has been fully created. If not, readjust the depth setting and repeat steps 2-5.
2
Drill angle limitations
Warnings: Refer to warning W3.
Dissecting tool Drill angle limit
DS1TD24
EN
Table 2: Drill angle limits
DS1TD32
DS1TD32L
DS1TD75
30°
10°
Removal
1. Hold the motor in palm of hand. Rotate the attachment to the unlocked position. In this position, the arrows on the attachment and motor will line up.
2. Remove the dissecting tool from the attachment and discard the tool per local regulations.
3. Use thumb and index nger to lift the attachment o of the motor.
Cleaning and sterilization
Devices should be cleaned and then sterilized per the instructions below: Medtronic recommends incineration of devices that have directly contacted patients suspected or conrmed
with a Transmissible Spongiform Encephalopathy (TSE)/ Creutzfeldt-Jakob disease (CJD) diagnosis.
Reprocessing instructions
Warnings Do not soak/submerge devices.
Do not use ultrasound to clean devices. Do not use chlorine based or corrosive cleaning agents such as bleach, lye,
acetone, sodium hypochlorite/bleach, sodium hydroxide, formic acid, or solutions containing glutaraldehyde.
The use of a washer-disinfector for cleaning may cause a degradation in performance. Allow an adequate cooling period after steam sterilization. Use only nylon cleaning brushes. Non-nylon cleaning brushes leave residue that
may prevent the dissecting tool from being secured properly in the handpiece.
Limitations on reprocessing
Point of use Reprocessing begins at the point of use. Reprocess the device within 30
Containment and transportation
The product is provided non-sterile and must be cleaned and sterilized before rst use and any reuse.
minutes of use. Do not allow blood, debris, or bodily uids to dry on the device. Remove excess soil using running, cold tap water 10 - 22°C (50 - 72°F). Tap water is dened as potable water with a hardness value of <150 mg/L.
It is recommended that devices are reprocessed as soon as is practical following use.
3
EN
Reprocessing instructions
Preparation for decontamination
Preclean instructions prior to use in an automated washer
Detach the attachment from the motor. Remove and discard the tool. Do not soak devices.
Review the washer-disinfector warning above, before using this cleaning method. Manually rinse attachment/tubes under tap water, until no visible soil is noticed. Actuate moving parts through their full range of motion while under running
tap water. While rinsing under running tap water, an appropriately sized nylon brush may
be used internally or externally to aid in cleaning attachments. Remove devices from instrument trays before placing into washer baskets. Attachment must be placed in the washer with the tube in the fully extended
position. Orient devices following recommendations of the washer/disinfector manufacturers. Clean per one of the recommended washer cycles below. After cleaning per the recommended washer cycle below, visually examine the
attachment for cleanliness.
Cleaning: Automated Caution: Do not use ultrasonic cleaner. Review the washer-disinfector warning above, before using
this cleaning method.
Automated wash cycle using a neutral enzymatic detergent (pH 6.0-8.0)
Phase Recirculation time Water temperature Detergent type
Pre-wash 2 minutes (02:00) Cold tap water Not applicable
Wash 5 minutes (05:00) 66°C (151°F) (set point)
Rinse 1 minute (01:00) Hot tap water Not applicable
Neutral enzymatic detergent, pH 6.0-8.0
Automated wash cycle using an alkaline detergent (pH 8.0 -10.5)
Phase Recirculation time Water temperature Detergent type
Pre-wash 2 minutes (02:00) Cold tap water Not applicable
Wash 5 minutes (05:00)
Rinse 1 minute (01:00) Hot tap water Not applicable
Thermal rinse 1 minute (01:00)
Puried water rinse
4
1 minute 30 seconds (01:30)
43°C (109°F) (set point)
90°C (194°F) (set point)
66°C (151°F) (set point)
Alkaline detergent with pH 8.0 - 10.5
Not applicable
Not applicable
Reprocessing instructions
Cleaning: Manual 1. Thoroughly rinse the attachment under cold running tap water
(10–22°C/50–72°F) to remove any visible soil. Actuate all moveable parts through their full range of motion while rinsing. Use a nylon brush to aid in cleaning. Give particular attention to crevices and other areas that present a challenge to cleaning.
2. Using tap water, prepare a neutral enzymatic cleaner like Steris Prolystica 2x concentrate enzymatic cleaner following the manufacturer’s recommendation of 1.0mL/L(1/8oz/gallon) at a temperature of 23°C(73°F) or use an equivalent neutral pH cleaner following that manufacturer’s recommendations.
3. Wipe the attachment with a cloth dampened with the prepared cleaner.
4. Use a nylon brush dampened with the prepared cleaner to clean the external surfaces of the attachment. Actuate all moveable parts through their full range of motion to allow the cleaner to thoroughly clean the attachment.
5. Use a 3.2mm diameter nylon brush dampened with the prepared cleaner to clean the inner diameter of the attachment. Push the brush through the attachment from rear to front to loosen and remove debris trapped inside.
6. Thoroughly rinse the attachment under warm running tap water (23–43°C/73–109°F). Actuate all moveable parts through the full range of motion while rinsing. Flush both ends to remove cleaner.
7. Following the warm tap water rinse, thoroughly rinse the attachment with room temperature (25°C/77°F) puried water (deionized, reverse osmoses, or equivalent) for a minimum of 30 seconds.
8. Verify that the attachment is visually clean after manual cleaning. If visible soil remains, repeat cleaning.
Drying and lubrication
Note: Medtronic no longer recommends using the Legend attachment cleaning
nozzle (PA120), as this may cause some attachments to overheat. Thoroughly dry attachments. An air gun may be used to blow moisture out
from rear to front of attachment. Using an aerosol spray lubricant (such as Pana Spray), perform the following
steps to lubricate the attachment.
1. Holding the can approximately 10-15cm (3-6 in.) away from the attachment, spray all components that move, rotate, or slide with three quick squirts.
2. Articulate movable components to ensure proper lubrication.
3. Remove excess lubricant with a clean cloth.
Packaging Place the attachment in the Stealth Autoguide instrument tray. Follow the
sterilization instructions for Stealth Autoguide found in the Stealth AutoguideIFU.
Sterilization Place the attachment in the Stealth Autoguide instrument tray.
Follow the sterilization instructions for Stealth Autoguide system found in the Stealth Autoguide IFU.
Maintenance, inspection, and testing
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
Verify functionality prior to re-use.
Storage Store with other sterile devices.
EN
5
EN
Return policy for devices exposed to Transmissible Spongiform Encephalopathies (TSE)
Reusable devices that have been used on patients with suspected Creutzfeldt-Jakob Disease (CJD) or other Transmissible Spongiform Encephalopathies (TSE) should be quarantined and not reused until a diagnosis is conrmed or excluded. Medtronic Powered Surgical Solutions (MPSS) will not authorize or accept the return of MPSS products that directly contacted a patient or are contaminated with a patient’s body uids that is suspected or conrmed with TSE or CJD diagnosis. Furthermore, MPSS recommends that all MPSS product used on a patient conrmed with or suspected of a TSE/CJD diagnosis be incinerated. If TSE/CJD is excluded as a diagnosis, the quarantined reusuable equipment may be returned for use after appropriate cleaning, decontamination and sterilization. Hospital personnel should contact their infection control personnel for current procedures and policy for reusable equipment processing. Contact your Sales Representative for temporary equipment while original equipment is quarantined or for replacement of product incinerated under this policy. Contact MPSS Regulatory Aairs Department for additional information regarding TSE/CJD contamination.
Limited warranty* (U.S. customers only)
A. This Limited Warranty provides the following assurance for the customer who purchases a Legend
Depth Stop Attachment. This Limited Warranty is extended only to the buyer purchasing the Legend Depth Stop Attachment directly from Medtronic or from its aliate or its authorized distributor or representative.
1. Should the Attachment fail to function to Medtronic’s published specications due to defect in materials or workmanship during the term of this Limited Warranty (ninety [90] days from the date of shipment of the Attachment to customer), Medtronic will either repair or replace the Attachment or any portion there of at no charge.
B. To qualify for this Limited Warranty, the following conditions must be met:
1. The product must be used on or before its “Use By” or “Use Before” date, if applicable.
2. The Attachment must be used in accordance with its labeling and may not be altered or subjected to misuse, abuse, accident or improper handling.
3. Medtronic must be notied in writing within thirty (30) days following discovery of a defect.
4. The Attachment must be returned to Medtronic within thirty (30) days of Medtronic receiving notice as provided for in (3) above.
5. Upon examination of the Attachment by Medtronic, Medtronic shall have determined that: (a) the Attachment was not repaired or altered by anyone other than Medtronic or its authorized representative, (b) the Attachment was not operated under conditions other than normal use, and (c) the prescribed periodic maintenance and services, if applicable, have been performed on the Attachment.
C. This Limited Warranty is limited to its express terms. THIS LIMITED WARRANTY IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESSED OR IMPLIED WHETHER STATUTORY OR OTHERWISE, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL MEDTRONIC BE LIABLE FOR ANY CONSEQUENTIAL, INCIDENTAL, PROSPECTIVE OR OTHER SIMILAR DAMAGE RESULTING FROM A DEFECT, FAILURE, OR MALFUNCTION OF THE IPC SYSTEM, WHETHER A CLAIM FOR SUCH DAMAGE IS BASED UPON THE WARRANTY, CONTRACT, NEGLIGENCE OR OTHERWISE.
D. The exclusions and limitations set out above are not intended to, and should not be construed
so as to, contravene mandatory provisions of applicable law. Users may benet from statutory warranty rights under legislation governing the sale of consumer goods. If any part or term of this Limited Warranty is held by any court of competent jurisdiction to be illegal, unenforceable, or in conict with applicable law, the validity of the remaining portion of the Limited Warranty shall not be aected, and all rights and obligations shall be construed and enforced as if this Limited Warranty did not contain the particular part or term held to be invalid.
*This Limited Warranty is provided by Medtronic Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, Texas 76137-4116. It applies only in the United States. Areas outside the United States should contact their local Medtronic representative for terms of the warranty.
6
Symbols
EC REP
Compliant with European Council Directive MDD 93/42/EEC
Authorized Representative in the European Community
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Consult Instructions for Use
Manufacturer
Use By Date
Catalog Number
Use With
Attachment
Lot Number
Diameter
Dissecting Tool
Single Use
Locked
Do not use if package is damaged
Package Quantity
Non-sterile
Sterilized using Irradiation
Serial Number
175042EN E
EC REP
2019-11
May be covered by U.S. Patents:
Medtronic.com/patents
©2019 Medtronic, Inc.
Made in USA. Printed in USA.
Medtronic Powered Surgical Solutions
4620 North Beach Street
Fort Worth, Texas 76137 USA
+1 800 643 2773
Medtronic B.V.
Earl Bakkenstraat 10
6422 PJ Heerlen The Netherlands +31 45 566 8000
Medtronic Australian Sponsor:
Medtronic Australasia Pty Ltd
2 Alma Road
Macquarie Park, NSW 2113 Australia
medtronic.com
manuals.medtronic.com
Loading...