Medtronic DDPC3D1 Reference Guide

Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Family of MR Conditional digital single-chamber and dual-chamber implantable cardioverter defibrillators with SureScan™ technology
Reference Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Reference Manual
A reference manual for the Medtronic Cobalt™ XT DR, Cobalt™ DR, Crome™ DR, Cobalt™ XT VR, Cobalt™ VR and Crome™ VR families of digital implantable cardioverter defibrillators with SureScan technology
Medtronic, Medtronic with rising man logo, and Medtronic logo are trademarks of Medtronic. Third-party trademarks (“TM*”) belong to their respective owners. The following list includes trademarks or registered trademarks of a Medtronic entity in the United States and/or in other countries.
Active Can™, ATP During Charging™, Capture Management™, Cardiac Compass™, CareAlert™, CareLink™, ChargeSaver™, Cobalt™, Crome™, Flashback™, Intrinsic ATP™, Marquis™, Medtronic CareAlert™, Medtronic CareLink™, MVP™, OptiVol™, PR Logic™, Quick Look™, Reactive ATP™, SmartShock™, SureScan™, Switchback™, T-Shock™, TruAF™
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Contents
1 Introduction ............................................................ 8
1.1 About the product literature .......................................... 8
1.2 Device features .................................................... 9
2 Patient follow-up guidelines ............................................ 11
2.1 In-clinic follow-up appointments and remote monitoring ................ 11
2.2 Optimizing device longevity ........................................ 16
3 Diagnostic data features ............................................... 21
3.1 Quick Look summary data ......................................... 21
3.2 Medtronic CareAlert Monitoring ..................................... 23
3.3 OptiVol 2.0 Fluid Status Monitoring .................................. 30
3.4 RV Lead Integrity Alert ............................................. 35
3.5 Cardiac Compass Trends .......................................... 39
3.6 Heart Failure Management Report .................................. 47
3.7 Arrhythmia Episodes data .......................................... 49
3.8 Episode and therapy counters ...................................... 54
3.9 Flashback data ................................................... 57
3.10 Rate Drop Response episodes ..................................... 58
3.11 MVP mode switches data .......................................... 59
3.12 Rate Histograms ................................................. 59
3.13 Device and lead performance data .................................. 62
3.14 Automatic device status monitoring .................................. 66
4 Pacing features ....................................................... 69
4.1 Sensing in DR devices ............................................ 69
4.2 Sensing in VR devices ............................................. 79
4.3 Basic pacing for DR devices ........................................ 84
4.4 Basic pacing for VR devices ........................................ 95
4.5 Managed Ventricular Pacing (MVP) .................................. 98
4.6 Rate Response .................................................. 104
4.7 Capture Management ............................................ 109
4.8 Rate Adaptive AV ................................................ 120
4.9 Auto PVARP .................................................... 121
Reference Manual 5
Medtronic
4.10 Rate Drop Response ............................................. 124
4.11 Rate Hysteresis ................................................. 128
4.12 Sleep feature ................................................... 130
4.13 Non-Competitive Atrial Pacing ..................................... 131
4.14 PMT Intervention ................................................ 133
4.15 PVC Response .................................................. 135
4.16 Ventricular Safety Pacing ......................................... 136
4.17 Mode Switch .................................................... 138
4.18 Conducted AF Response in DR devices ............................ 141
4.19 Conducted AF Response in VR devices ............................ 143
4.20 Atrial Rate Stabilization ........................................... 145
4.21 Atrial Preference Pacing .......................................... 148
4.22 Post-Mode Switch Overdrive Pacing ................................ 152
4.23 Post Shock Pacing in DR devices .................................. 155
4.24 Post Shock Pacing in VR devices .................................. 156
4.25 Post VT/VF Shock Overdrive Pacing ................................ 156
4.26 Ventricular Rate Stabilization in DR devices ......................... 157
4.27 Ventricular Rate Stabilization in VR devices .......................... 159
5 Tachyarrhythmia detection features .................................... 163
5.1 AT/AF detection in DR devices ..................................... 163
5.2 AF detection in VR devices: the TruAF Detection algorithm ............ 169
5.3 VT/VF detection in DR devices .................................... 174
5.4 VT/VF detection in VR devices ..................................... 187
5.5 PR Logic ....................................................... 199
5.6 Wavelet ........................................................ 203
5.7 Onset .......................................................... 209
5.8 Stability ........................................................ 216
5.9 High Rate Timeout ............................................... 219
5.10 TWave Discrimination ............................................ 221
5.11 RV Lead Noise Discrimination ..................................... 224
6 Tachyarrhythmia therapy features ..................................... 228
6.1 Atrial therapy scheduling .......................................... 228
6.2 Atrial ATP therapies .............................................. 234
6.3 Atrial cardioversion ............................................... 245
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
6 Reference Manual
Medtronic
6.4 VF therapies in DR devices ........................................ 251
6.5 VF therapies in VR devices ........................................ 261
6.6 Ventricular ATP therapies ......................................... 272
6.7 Ventricular cardioversion in DR devices ............................. 285
6.8 Ventricular cardioversion in VR devices ............................. 294
6.9 Progressive Episode Therapies .................................... 302
Glossary ................................................................ 304
Index ................................................................... 313
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Reference Manual 7
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

1 Introduction

1.1 About the product literature

This manual describes the operation and intended use of features offered by Medtronic Cobalt XT DR and VR, Cobalt DR and VR, and Crome DR and VR MRI SureScan devices.
Throughout this manual, the word “device” refers to the implanted cardioverter defibrillator.
Unless otherwise noted, all device features described in this manual apply to the Cobalt XT DR device or the Cobalt XT VR device, respectively. To determine which features are available for another model in the Cobalt XT DR/VR, Cobalt DR/VR, and Crome DR/VR families, refer to Table 1.

1.1.1 Product literature

Before implanting the device, it is recommended that you take the following actions:
• Read the product literature for information about prescribing, implanting, and using the device and conducting a patient follow-up session.
• Thoroughly read the technical manuals for the leads used with the device. Also read the technical manuals for other system components.
• Discuss the device and implant procedure with the patient and any other interested parties, and give them any patient information materials packaged with the device.
Additional information about the device is provided in the following documents:
MRI technical manual – This manual provides MRI-specific procedures and warnings and precautions.
Implantable device app help – The help explains how to use the implantable device app to program the device settings and view the stored device data.
Device manual – This manual contains model-specific feature information, indications and contraindications, warnings and precautions, instructions for implanting the device, quick reference specifications, and parameter tables.
Explanation of symbols – This document defines the symbols that may appear on the device package. Refer to the package label to see which symbols apply specifically to this device.
Medical Procedure and EMI Warnings, Precautions, and Guidance Manual for Health Care Professionals – This manual provides warnings, precautions, and guidance for health
care professionals who perform medical therapies and diagnostic procedures on cardiac device patients. This manual also includes information about hazards from sources of
8 Reference Manual
Medtronic Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
electromagnetic interference (EMI) in the patient’s home, recreational environments, and occupational environments.
Radio regulatory compliance insert – This document provides compliance information related to the radio components of the device.

1.1.2 Technical support

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users.
For more information, contact your local Medtronic representative or call or write Medtronic at the appropriate telephone number or address listed on the back cover.

1.2 Device features

The following table lists the features in the Cobalt XT DR/VR, Cobalt DR/VR, and Crome DR/VR device families that are not available in all devices. Features that are omitted from this list apply to all DR devices, to all VR devices, or to every device in the portfolio.

Table 1. Product feature relationship

Cobalt XT MRI SureScan, Cobalt MRI SureScan, and Crome MRI
SureScan
DDPA2D4
Features
AF Alerts X X
TruAF Detection algorithm
AT/AF Detection X X
Atrial ATP X X
Atrial Cardiover­sion
Atrial 50 Hz Burst In-Office
Heart Failure Man­agement Report
OptiVol 2.0 fluid status monitoring
DDPA2D1
X X
X X
X X
X X
X X
DDPB3D4 DDPB3D1
DDPC3D4 DDPC3D1
DVPA2D4 DVPA2D1
DVPB3D4 DVPB3D1
DVPC3D4 DVPC3D1
Reference Manual 9
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Table 1. Product feature relationship (continued)
Cobalt XT MRI SureScan, Cobalt MRI SureScan, and Crome MRI
SureScan
DDPA2D4
Features
Reactive ATP algo­rithm
SmartShock 2.0 technology
SmartShock 2.0+ technology with Intrinsic ATP algo­rithm
a
Includes the following features: Confirmation+, PR Logic (DR devices only), RV Lead
a
b
DDPA2D1
X X
X X X X
X X
DDPB3D4 DDPB3D1
DDPC3D4 DDPC3D1
DVPA2D4 DVPA2D1
DVPB3D4 DVPB3D1
DVPC3D4 DVPC3D1
Integrity Alert, RV Lead Noise Discrimination, SVT Discriminators in VF Zone, TWave Discrimination, and Wavelet
b
Includes the following features: Confirmation+, Intrinsic ATP algorithm, PR Logic (DR devices only), RV Lead Integrity Alert, RV Lead Noise Discrimination, SVT Discriminators in VF Zone, TWave Discrimination, and Wavelet
10 Reference Manual
Medtronic Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

2 Patient follow-up guidelines

2.1 In-clinic follow-up appointments and remote monitoring

Schedule regular in-clinic follow-up appointments with the patient throughout the service life of the device. For patients enrolled in the Medtronic CareLink Network, remote monitoring can replace the need for some in-clinic follow-up appointments. With remote monitoring, data from a patient’s implanted device is sent to the Medtronic CareLink Network, and you can review the transmitted data on the Medtronic CareLink Network website. Schedule in-clinic follow-up appointments and CareLink transmissions as follows:
• Schedule an in-clinic follow-up appointment within 72 hours of implant so that the patient can be checked for lead dislodgment, wound healing, and postoperative complications.
• Schedule an in-clinic follow-up appointment within 2 to 12 weeks after implant to evaluate the condition of the patient, the device, and the leads, and to verify that the device is configured appropriately for the patient.
• Schedule routine CareLink transmissions or in-clinic follow-up appointments every 3 to 6 months with in-clinic follow-up appointments occurring at least annually.
• When the device battery approaches Recommended Replacement Time (RRT), schedule CareLink transmissions or in-clinic follow-up appointments every 1 to 3 months.
• Schedule in-clinic follow-up appointments as needed (for example, if data from a CareLink transmission indicates that the patient’s device requires adjustment).

2.1.1 Remote monitoring options

The device provides automatic wireless remote monitoring, if available, through a patient-owned smart phone or tablet running a mobile app (referred to as the patient app) or through a home communicator. This device features Bluetooth® wireless technology1. Routine monitoring transmissions occur according to a schedule that you create on the Medtronic CareLink Network website. In addition, the CareAlert Monitoring feature provides automatic, unscheduled transmissions for specific clinical or device status events (see Section 3.2, Medtronic CareAlert Monitoring, page 23). Patients can also send unscheduled transmissions.
1
The Bluetooth® word mark is a registered trademark of Bluetooth SIG, Inc. and any use of this mark by Medtronic is under license.
Reference Manual 11
Medtronic
Note: When viewing a CareLink transmission, the data collected since the last session is presented differently than it is for a patient session. For a CareLink transmission, the last session is defined as either the last patient session or the last CareLink transmission. During an in-clinic appointment, the last session is defined as the last patient session.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

2.1.2 Follow-up process

The process for conducting a follow-up evaluation, either during an in-clinic appointment or with a CareLink transmission, includes the following steps:
1. Review the patient’s presenting rhythm.
2. Verify the status of the implanted system.
3. Verify the clinical effectiveness of the implanted system.
4. During an in-clinic follow-up appointment, adjust device parameters as necessary.
5. If evaluating data remotely, schedule an in-clinic follow-up appointment as necessary.

2.1.3 Reviewing the presenting rhythm

The presenting rhythm may indicate the presence of undersensing, far-field oversensing, or loss of capture. The presence of these basic pacing issues can affect the delivery of therapy. These issues can often be resolved by making basic programming changes.
Review the presenting rhythm as follows:
• During an in-clinic follow-up appointment, view the Live Rhythm Monitor and record the EGM and marker traces.
• For remote monitoring, review the EGM data that was recorded at the time of the CareLink transmission.
Viewing this information can help you identify any issues with the patient’s presenting rhythm. It may be necessary to adjust the pacing parameters.

2.1.4 Verifying the status of the implanted system

To verify the status of the implanted system, perform the following tasks:
• Assess the battery status.
• Check lead measurements and trend data.
• Review any Quick Look observations about the device and lead status.
12 Reference Manual
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
2.1.4.1 Assessing the battery status
To assess the status of the device battery, review the REMAINING LONGEVITY estimate on the Quick Look screen. If the device battery has reached a replacement threshold, the associated indicator is displayed.
To see more detail about the battery status, including battery voltage, tap REMAINING LONGEVITY on the Quick Look screen to see the device data provided by the BATTERY AND LEAD MEASUREMENTS window.
Warning: Replace the device immediately if the End of Service (EOS) indicator is displayed. The device may lose the ability to pace, sense, and deliver therapy adequately after the battery reaches End of Service.
If the Recommended Replacement Time (RRT) indicator or the Elective Replacement Indicator (ERI) is displayed, or if the battery voltage is at or below the displayed RRT voltage, contact your Medtronic representative and schedule a replacement procedure with your patient. For more information about the replacement indicators, see Section 3.13, Device and lead performance data, page 62.
2.1.4.2 Checking lead measurements and trend data
In-clinic follow-up appointment – During an in-clinic follow-up appointment, you can
check the status of the implanted leads. On the Quick Look screen you can review up to 12 months of the following lead trend data:
• Lead impedance
– Pacing (Ω)
– Defib (Ω)
• THRESHOLD (V @ ms)—shows P-wave and R-wave threshold trends
• AMPLITUDE (mV)—shows capture amplitude trends
For a more detailed history of each trend, tap LEAD TRENDS. For more information about the automatic collection of these trends, see Section 3.13,
Device and lead performance data, page 62. To gather real-time information about the performance of the device and leads, perform the
following tests:
• LEAD IMPEDANCE
• PACING THRESHOLD
• SENSING
For more information about these tests, refer to the implantable device app help.
Reference Manual 13
Medtronic
Evaluating a CareLink transmission – When evaluating a CareLink transmission, you can check the status of implanted leads by reviewing the most recent lead impedance, capture threshold, and sensing amplitude measurements on the Quick Look report. Compare these values to the patient history and to the trend data provided on the Lead Trends reports.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
2.1.4.3 Reviewing Quick Look observations about the status of the device, the leads, and the alert events
The Quick Look data includes observations that are based on an analysis of the programmed parameters and the collected data. Observations may include information about the status of the device and the battery and the integrity of the implanted leads or potential issues with the programmed parameter settings. If Medtronic CareAlert Monitoring is enabled, any alert events detected by the device are presented in the OBSERVATIONS section of the Quick Look screen. Review these observations and check related reports for evidence of a problem with the device or the leads.

2.1.5 Verifying the clinical effectiveness of the implanted system

You can use the information available on the Quick Look screen and the saved reports to assess whether the device is providing adequate clinical support for the patient.
2.1.5.1 Reviewing Quick Look observations about clinical status
The Quick Look data includes observations about noteworthy or abnormal patient conditions such as low patient activity, unexpectedly high rates, or high arrhythmia burden. If you enable Medtronic CareAlert Monitoring, the OBSERVATIONS section of the Quick Look screen lists alert events detected by the device. To evaluate the clinical effectiveness of the implanted system, review this event data and check the related data.
2.1.5.2 Assessing the effectiveness of pacing therapy
1. Review the pacing percentages in the Quick Look data. To assess the patient’s pacing and sensing history in more detail, review the rate histograms data. For more information, see Section 3.12, Rate Histograms, page 59.
2. Review Cardiac Compass Trends data and compare it with the patient history. Cardiac Compass Trends data can help you to determine whether changes in the patient’s activity, pacing therapies, and arrhythmias have occurred during the past 14 months. For more information about the data collected by the Cardiac Compass feature, see Section 3.5, Cardiac Compass Trends, page 39.
14 Reference Manual
Medtronic
3. Evaluate the patient’s pacing thresholds by reviewing capture threshold trend data. During an in-clinic follow-up appointment, you can also perform a Pacing Threshold Test. Refer to the implantable device app help for more information about conducting these tests.
4. Check the programmed pacing parameters to confirm that they provide an appropriate safety margin.
5. During in-clinic follow-up appointments, interview the patient to confirm that the patient is receiving adequate cardiac support for activities of daily living.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
2.1.5.3 Assessing tachyarrhythmia detection
1. Review the Quick Look data for the counts of each kind of tachyarrhythmia episode.
2. Review the Cardiac Compass Trends data and the Rate Histograms data to assess the frequency of arrhythmias and device therapies in consideration of patient condition. For more information, see Section 3.5, Cardiac Compass Trends, page 39 and Section 3.12, Rate Histograms, page 59.
Episode misidentification – If the episode records indicate that the device has misidentified the patient’s rhythm, carefully review the tachyarrhythmia episode and sensing integrity data, the Cardiac Compass Trends data, and the data stored for other episodes. Consider adjusting the detection parameters and the SVT detection criteria as needed. For more information about how to view sensing integrity data, see Section 4.1, Sensing in DR devices, page 69 and Section 4.2, Sensing in VR devices, page 79.
Caution: Use caution when reprogramming the detection or sensing parameters to ensure that changes do not adversely affect VF detection. Ensure that appropriate sensing is maintained. For more information, see Section 4.1, Sensing in DR devices, page 69 and Section 4.2, Sensing in VR devices, page 79.
2.1.5.4 Assessing tachyarrhythmia therapy
1. Review any Medtronic CareAlert notifications in the Quick Look OBSERVATIONS section that relate to therapy delivery. To see detailed information about Medtronic CareAlert notifications, tap DATA > CareAlert EVENTS from the Menu button.
2. Check tachyarrhythmia episode records to determine the effectiveness of therapies that have been delivered.
3. Adjust the therapy parameters as needed.
Reference Manual 15
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

2.1.6 Adjusting device parameters

Adjust the pacing, tachyarrhythmia detection, tachyarrhythmia therapy, and diagnostic data parameters as needed to address any issues identified during the follow-up appointment.
Caution: Use caution when reprogramming the detection or sensing parameters to ensure that appropriate sensing is maintained. For more information, see Section 4.1, Sensing in DR devices, page 69 and Section 4.2, Sensing in VR devices, page 79.

2.1.7 Scheduling an in-clinic follow-up appointment

Data transmitted to the Medtronic CareLink Network may indicate the need to schedule an in-clinic follow-up appointment with your patient in addition to the patient’s regularly scheduled appointments. You may need to perform manual tests, adjust device parameters, or assess lead status more directly. The following table shows an example of how data from a CareLink transmission may be used to make scheduling decisions.

Table 2. Example: Responses to different kinds of CareLink transmissions

Device and lead status
Normal Normal According to the regular schedule
Normal Abnormal, but no urgent or emer-
Normal Abnormal, urgent condition Within 1 week
Normal Abnormal, emergency condition Immediately
Abnormal Any Immediately
Clinical status
gency conditions
When to schedule an in-clinic fol­low-up appointment
According to the regular schedule

2.2 Optimizing device longevity

Optimizing device longevity is a desirable goal because it can reduce the frequency of device replacement for patients. Optimizing device longevity requires balancing the benefit of device therapy and diagnostic features with the energy requirements placed on the battery as a result of these features.
To view the REMAINING LONGEVITY estimate for the device, refer to the Quick Look screen.
Note: The remaining longevity estimate is updated when parameters are reprogrammed and when the device is interrogated.
The following sections describe strategies that can help reduce the energy requirements placed on the battery.
16 Reference Manual
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

2.2.1 Promoting intrinsic AV conduction

MVP (Managed Ventricular Pacing) – The MVP feature promotes AV conduction by
reducing unnecessary right ventricular pacing. The primary benefit of the MVP feature is therapeutic, but it may also preserve device longevity as a result of a decrease in the percentage of pacing. For more information, see Section 4.5, Managed Ventricular Pacing (MVP), page 98.
Promoting AV conduction with longer AV intervals – Another method of promoting AV conduction is to increase the paced AV interval and the sensed AV interval. This action allows intrinsic conduction to occur before a ventricular pace. Fewer pacing pulses may help to preserve device longevity. For more information, see Section 4.3, Basic pacing for DR devices, page 84.

2.2.2 Managing pacing outputs

Capture Management – The Capture Management feature provides the device with
automatic monitoring and follow-up capabilities for managing pacing thresholds. This feature is designed to monitor the pacing threshold and, optionally, to adjust the pacing outputs to maintain capture. Programming the Capture Management feature allows the device to set the pacing amplitude just high enough to maintain capture while preserving battery energy. For more information, see Section 4.7, Capture Management, page 109 .
Manual optimization of amplitude and pulse width – If you choose to program the Capture Management feature to Off, you can optimize the patient’s pacing output parameters manually. Perform a Pacing Threshold Test to determine the patient’s pacing thresholds. Select amplitude and pulse width settings that provide an adequate safety margin above the patient’s pacing threshold. These actions decrease the pacing outputs and preserve battery energy. Refer to the implantable device app help for more information about performing a Pacing Threshold Test.
Pacing rate – The more paced events that are delivered, the more device longevity is reduced. Make sure that you have not programmed an unnecessarily high pacing rate for the patient. Carefully consider using features that increase the bradycardia pacing rate. Use features such as Atrial Preference Pacing (APP), Conducted AF Response, and Rate Response only for patients who can receive therapeutic benefit from the feature.

2.2.3 Optimizing tachyarrhythmia therapy settings

Defibrillation – To treat ventricular fibrillation episodes, the device may deliver defibrillation
therapy to terminate the episode and restore the patient’s normal sinus rhythm. The device can be programmed to deliver a sequence of up to 6 defibrillation therapies. Although defibrillation therapy expends a high level of energy, VF therapies should be programmed to the maximum energy level.
Reference Manual 17
Medtronic
Atrial cardioversion – You can program a DR device to deliver automatic atrial cardioversion (CV) therapies to treat atrial tachyarrhythmia episodes. If you choose to treat the patient with atrial CV therapies, you can extend device longevity by carefully considering how you program the following parameters: the number of shocks delivered during a 24-hour cycle and the Episode Duration before CV. It is recommended that CV therapy be set to full energy to terminate the arrhythmia. For more information, see Section 6.3, Atrial cardioversion, page 245.
Ventricular cardioversion – If you are providing ventricular cardioversion therapies for the patient, consider programming the therapy energy to a value lower than the maximum energy but high enough to terminate the VT. However, at least one VT therapy and one FVT therapy in a sequence should be programmed to the maximum energy level. For more information, see Section 6.7, Ventricular cardioversion in DR devices, page 285 and Section 6.8, Ventricular cardioversion in VR devices, page 294.
FVT via VF detection – An FVT detection zone can be used to detect and treat a VT episode that is in the rate zone for VF. This approach can help maintain reliable detection of VF while allowing ATP to be delivered for fast VT episodes. For more information, see Section 5.3, VT/
VF detection in DR devices, page 174 and Section 5.4, VT/VF detection in VR devices, page 187.
Antitachycardia pacing (ATP) – ATP therapies interrupt the tachycardia episode and restore the patient’s normal sinus rhythm. ATP therapies deliver pacing pulses instead of the high-voltage shocks that are delivered in cardioversion therapy and defibrillation.
ATP therapy requires less battery energy than cardioversion or defibrillation. For some patients, you may be able to program the device to deliver ATP therapies before delivering high-voltage therapies.
For more information about ATP and atrial episodes, see Section 6.2, Atrial ATP therapies,
page 234. For more information about ATP and ventricular episodes, see Section 6.6, Ventricular ATP therapies, page 272.
Delivering ATP before the first defibrillation – You can program the device to deliver ATP as the first defibrillation therapy before the device is charged. This action can prevent delivery of high-voltage therapy for rhythms that can be terminated by ATP (rapid, monomorphic VT, for example).
If you program the ChargeSaver feature to On, the device can also automatically switch to the ATP Before Charging operation. This switch allows the device to attempt a sequence of ATP therapy before charging the capacitors to treat a detected VF episode. For more information, see Section 6.4, VF therapies in DR devices, page 251 and Section 6.5, VF therapies in VR devices, page 261.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
18 Reference Manual
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

2.2.4 Disabling atrial sensitivity

Atrial sensitivity – For DR devices, when an atrial lead is not present or when atrial
monitoring is not needed due to chronic AT/AF, program Atrial Sensitivity to Off before changing the mode to VVI or VVIR to preserve battery energy.
Note: When Atrial Sensitivity is programmed to Off, AT/AF monitoring and PR Logic are disabled.

2.2.5 Considering how diagnostic features with data storage impact longevity

Pre-arrhythmia EGM storage – When the device is set to nominal parameters, it provides
continuous Pre-arrhythmia EGM storage with minimal longevity impact. If the parameters for EGM storage or sense polarity are changed, the device may be required to keep additional EGM circuitry enabled at all times, which reduces device longevity. You can preserve longevity by ensuring that the stored EGM channels use EGM SOURCE values that match the Sense Polarity values.
If it is necessary to store EGM from a source that does not match the sense polarity, consider the following information about the Pre-arrhythmia EGM feature:
• The Pre-arrhythmia EGM feature allows the device to collect up to 20 s of EGM data before the onset of VT/VF, VT Monitor, SVT, VT-NS, and Fast A&V episodes.
• The Pre-arrhythmia EGM feature does not apply to AT/AF episodes. The device stores approximately 4 s of EGM before AT/AF detection, regardless of the Pre-arrhythmia EGM storage setting.
• When Pre-arrhythmia EGM storage is set to Off, the device begins to store EGM information for VT/VF, VT Monitor, and SVT episodes after the third tachyarrhythmia event occurs. Though EGM is not recorded before the start of the arrhythmia, the device still records up to 20 s of data before the onset or the detection of the episode. This data includes interval measurements and markers. In addition, Flashback data is stored for the most recent tachyarrhythmia episodes.
• For a patient with uniform tachyarrhythmia onset mechanisms, the greatest benefit of Pre-arrhythmia EGM storage is obtained after capturing a few episodes. Program Pre-arrhythmia EGM to Off after you have obtained the data of interest.
Holter telemetry – Extended use of the Holter telemetry feature decreases device longevity. The Holter telemetry feature continues to transmit EGM and marker data for the programmed time duration, regardless of whether the patient connector is positioned over the device.
Medtronic CareLink remote transmissions – When scheduling Medtronic CareLink remote transmissions, be aware that increasing the frequency of remote transmissions
Reference Manual 19
Medtronic
reduces the service life of the implanted device. Refer to the device manual for more information about the estimated effect on a specific device model. To conserve battery energy, schedule the lowest frequency of remote transmissions that still allows for the desired monitoring of your patient’s device.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

2.2.6 Extended wireless telemetry session

If the implantable device app detects no activity for 45 min, a pop-up window prompts you to extend the session. If you do not tap EXTEND to extend the session, the patient session ends in 15 min.
20 Reference Manual
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

3 Diagnostic data features

3.1 Quick Look summary data

At the start of a patient session, it is useful to quickly view summary information about device operation and the patient’s condition. This overview can help you to determine whether you need to look more closely at diagnostic data or reprogram the device to optimize therapy for the patient.
The Quick Look data summarizes the most important indicators of system operation and the patient’s condition. These indicators include device and lead status data, pacing therapy information, arrhythmia episode data, and system-defined observations.
You can view Quick Look data on the Quick Look screen, which is displayed at the beginning of a patient session. To return to the Quick Look screen from another screen, tap Quick Look from the Menu button. For more information about using the Quick Look screen, refer to the implantable device app help.
Note: The Quick Look screen shows information collected since the last patient session and stored in device memory. Programming changes made during the current session can affect the Quick Look observations.

3.1.1 Quick Look device and lead status information

The Quick Look data includes the following information about device and lead status:
• Estimate of remaining battery longevity
• Trends of the weekly average impedance, capture threshold, and wave amplitude measurements
• Most recent measured values for impedance, capture threshold, and wave amplitude
Reference Manual 21
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

3.1.2 Quick Look pacing therapy information

The Quick Look data includes the following information about pacing therapy, listed under the % OF TIME label:
• For DR devices, the percentage of time in atrial pacing (AP) since the last patient session; and for DR and VR devices, the percentage of ventricular pacing (VP) since the last patient session
• For DR devices, the percentage of time pacing in AS-VP event sequences
• For DR devices, an indicator that the pacing mode is programmed to an MVP mode (MVP On) or to another pacing mode (MVP Off)
3.1.2.1 Quick Look arrhythmia episode information
The Quick Look data includes the following information about arrhythmia episodes since the last patient session:
• Percentage of time spent in AT/AF
• Number of VT/VF episodes treated with tachyarrhythmia therapy
• Number of AT/AF episodes treated with tachyarrhythmia therapy
• Number of delivered shocks
• Number of monitored VT episodes
• Number of monitored SVT episodes
• Number of monitored AT/AF episodes
3.1.2.2 Quick Look observations
Observations are based on an analysis of programmed parameters and data collected since the last session. The following types of observations may occur:
• Device status observations inform you of conditions that affect device operation and require attention. Examples of such conditions include Recommended Replacement Time (RRT) or the occurrence of a device reset.
• Lead status observations report any potential issues with the sensing integrity of the leads, possible lead dislodgments, and abnormal Capture Management results. You may also be warned about possible inconsistencies in the programming of lead polarity.
• Parameter observations warn of any inconsistencies in the programming of detection and therapy parameters. An example is certain parameter settings resulting in a therapy being disabled.
22 Reference Manual
Medtronic
• Diagnostic data observations report noteworthy arrhythmia episodes. Examples include arrhythmias of different types occurring together and episodes for which therapies were unsuccessful. Conditions that prevent diagnostic data from being collected effectively are also reported.
• Medtronic CareAlert observations can report system or device performance conditions and certain heart rhythm conditions. For more information, see Section 3.2, Medtronic CareAlert Monitoring, page 23.
• Clinical status observations alert you to abnormal patient conditions, such as low activity rates, unexpectedly high heart rates, high arrhythmia burden, or fluid accumulation.
If you tap an observation that includes additional details, a > button appears at the end of the OBSERVATIONS field. To see the observation details, tap the > button.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

3.2 Medtronic CareAlert Monitoring

Important clinical management and system performance events can occur between scheduled patient sessions. These events can relate to clinical management data, to inappropriately programmed settings, or to system performance issues that must be investigated. The early detection and notification of these events, if they occur, enable you to intervene promptly with appropriate care for your patient.
The device continuously monitors for a specified set of clinical management and system performance events that can occur between scheduled follow-up sessions. If the device detects that such an event has occurred and if alerting parameters are programmed to On, Medtronic CareAlert Monitoring responds in the following ways:
• Patient alert
• Wireless signal and network transmission of event information
Medtronic CareAlert Monitoring continuously monitors for alert events. If an event occurs, CareAlert Monitoring sends alert data about specified clinical management or system performance events to the Medtronic CareLink Network (if available). Alert data is sent through the patient’s personal wireless device or through a Medtronic home communicator.
• Notification of alert event
Depending on the severity of the alert condition, you can set up Medtronic CareAlert notifications through the CareLink Network (if available). You can hold the alert for routine review on the CareLink website, or you can be notified by the following notification methods, according to your preferences:
– Email
– Text message
Reference Manual 23
Medtronic Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
– Pager
– Automated voice message
– Telephone call
Note: CareAlert notification methods, when available, vary from geography to geography.

3.2.1 Operation of Medtronic CareAlert Monitoring

If a clinical event or a system performance event occurs, and the device is programmed to notify you over the CareLink Network, the CareAlert Monitoring feature transmits alert data from the device to the patient app or the home communicator. The patient app or the home communicator then sends the alert data to the CareLink Network. The CareLink Network records the alert, then notifies you according to your preferences.
If an alert data transmission from the device is unsuccessful, the CareAlert Monitoring feature tries to establish communication with the patient app or the home communicator every 3 hours until the data transmission is successful. If the alert data remains untransmitted for 72 hours, the device emits a tone at the Alert Time that you select for your patient or at intervals unique to some alerts as described in Section 3.2.3.1 and Section 3.2.3.2.
Note: After a wireless alert signal is successfully transmitted, the device does not retransmit data for that alert until the alert is cleared. There is one exception to this rule: a device reset alert is transmitted each time a device reset occurs. There are 2 exceptions to this rule: a device reset alert is transmitted each time a device reset occurs, and an OptiVol Event alert is transmitted each time the patient’s fluid level exceeds the OptiVol Threshold following a reset of the OptiVol 2.0 Fluid Index (see Section 3.3, OptiVol 2.0 Fluid Status Monitoring, page 30).
The CareAlert notification methods (any one or a combination of voice message, text message, pager, email, live call, or website-only) can be programmed to On or Off, on a per-patient and per-clinic basis, and according to time of day.

3.2.2 Operation of Medtronic CareAlert events

Medtronic CareAlert events are either programmable or always on:
• Clinician-defined alerts can be programmed on or off.
• System-defined alerts are always on.
Alerts are displayed in the OBSERVATIONS area of the Quick Look screen.
2
CareAlert notification methods, when available, vary from geography to geography.
24 Reference Manual
Medtronic
When an alert is initiated, the device sounds a tone either at a selected time of day or at a fixed time interval. The tone then sounds each day at the selected time or interval until it is cleared through interrogation by the implantable device manager. Active tones also sound when the patient magnet is placed over the device. You can view alert details during a patient session.
Notes:
• A CareLink transmission does not clear an alert tone. The tone will continue to sound until the alert is cleared in the office (or if the DEVICE TONE is set to Off).
• Once an alert has been transmitted over the CareLink Network (if available), further transmissions for that alert condition will not occur until the alert is cleared.
• OptiVol alert tones stop sounding if the OptiVol 2.0 Fluid Index reaches 0. If the Fluid Index threshold is crossed again, another CareLink transmission is initiated and OptiVol alert tones sound at the programmed times.
• All alerts are cleared automatically when the device is interrogated with a device manager.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
3.2.2.1 Patient alert process
The device can be programmed to sound a tone to alert the patient when a clinical management event or a system performance event occurs. However, the patient will not be able to determine the nature of the alert. Instruct your patients to call the clinic if they hear a device tone.
The alert event determines when the device tone will sound: when the event occurs, at a programmed time of day, or at fixed intervals. Device tones sound for up to 30 s, and they are slightly louder than the ambient noise in a typical living room.
For more information about patient alerts, refer to the implantable device app help.
3.2.2.2 Instructing the patient
It is important that patients understand that they may hear device tones emitted from their implanted device. They must know what to do when a device tone sounds.
Warning: Make sure that patients understand that they must not carry, store, or leave the patient magnet positioned over the device. Device operation is temporarily impaired when the magnet is placed over the device and it must be moved away from the device to restore normal operation.
• Instruct patients to contact you immediately if they hear ANY tones from the device.
Reference Manual 25
Medtronic
• Tell patients the time of day that you have programmed an alert tone to sound. If a tone sounds, patients should expect it to sound every day at that time until the alert is cleared (or until DEVICE TONE is programmed to Off).
• Advise patients that if an alert is not successfully transmitted to their personal mobile devices or to a Medtronic home communicator within 72 hours, a tone sounds each day at the programmed alert time.
• Advise patients that the alert time does not adjust for time zone changes.
• Advise patients that they may hear a steady test tone or any active alert tones if they are in the presence of a strong electromagnetic field, such as the field within a store theft detector. Advise patients that the device operation is temporarily impaired in these situations and that they should move away from the source of the interference to restore normal device operation.
Patients should also understand the purpose of the patient magnet and how and when to use it. Make sure that they know that current patient alerts sound when the patient magnet is placed over the device. Demonstrate how to place the patient magnet over the device to replay the alert tones, and review the patient magnet manual with them.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs

3.2.3 Clinical management and system performance event alerts

3.2.3.1 Clinical management event alerts
Possible OptiVol fluid accumulation – This alert indicates that the OptiVol 2.0 Fluid Index
has met or exceeded the programmed OptiVol Threshold.
AT/AF Daily Burden – This alert indicates that the cumulative time in AT/AF exceeds the programmed threshold.
Avg. V. Rate During AT/AF – This alert indicates that the average ventricular rate during a selectable duration of AT/AF exceeds the programmed threshold.
Monitored VT Episode Detected – This alert indicates that 1 or more monitored VT episodes have been detected.
Daily VT/VF Episodes – This alert indicates that the total number of daily VT/VF episodes is greater than or equal to 3 episodes.
Weekly ATP Delivered Episodes – This alert indicates that the number of episodes treated with antitachycardia pacing is greater than or equal to the programmed threshold, per week.
Number of Shocks Delivered in an Episode – This alert indicates that the number of shocks delivered in a VT/VF episode is greater than or equal to the programmed Number of Shocks Threshold value.
26 Reference Manual
Medtronic
Cumulative Right Ventricular Pacing > 40% – This alert indicates that the cumulative percentage of ventricular pacing over the most recent 7 days was greater than 40%.
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
3.2.3.2 System performance event alerts
RV Lead Integrity – This alert indicates that an RV lead problem is suspected, which could
indicate lead fracture. The device immediately sounds an alert tone that lasts for 30 s. This tone repeats every 4 hours, beginning at the next scheduled 4-hour time interval, and at the programmed Alert Time….
RV Lead Noise – This alert indicates that noise was detected on the RV lead, which could indicate lead fracture, breached lead insulation, lead dislodgement, or improper lead connection. The device sounds an alert tone 3 min after a lead noise episode is detected. This tone repeats every 4 hours, beginning at the next scheduled 4-hour time interval, and at the programmed Alert Time….
Lead Impedance Out of Range… – These alerts indicate that the daily lead impedance measurement for one of the implanted leads is out of range. These alerts could indicate that a lead has dislodged or is improperly connected. For all leads, the device sounds an alert tone at the programmed Alert Time…. For the RV Pacing, RV Defibrillation and SVC Defibrillation leads, the device sounds an alert tone and repeats this tone every 4 hours, beginning at the next scheduled 4-hour time interval.
Capture Management High Threshold… – These alerts indicate high capture threshold measurements. During Right Ventricular Capture Management and Atrial Capture Management operation, an alert occurs when 3 consecutive daily measurements are high.
Low Battery Voltage RRT – This alert indicates that the daily automatic battery voltage measurement has been at or below the Recommended Replacement Time (RRT) voltage level for 3 consecutive days. When RRT has been met, the device has less than 3 months to the Elective Replacement Indicator (ERI).
Excessive Charge Time EOS… – This alert indicates that the charging period equals or exceeds the charge time threshold.
VF Detection OFF, 3+VF or 3+FVT Rx Off – These alerts indicate that one or more of the following conditions has occurred for at least 6 hours since the last programming: VF detection has been turned off; 3 or more VF therapies have been turned off; or FVT detection is programmed to FVT via VF and 3 or more FVT therapies have been turned off. Note that this alert sounds immediately and then every 6 hours until cleared.
For details about programmable settings for a particular parameter, see the device manual for the specific device.
Reference Manual 27
Medtronic Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
3.2.3.3 Non-programmable system performance event alerts
Pacing Mode DOO, VOO, or AOO – This alert indicates that the device is programmed to
a DOO, a VOO, or an AOO pacing mode and, as such, does not deliver tachyarrhythmia therapy. The device sounds a tone daily at the programmed time.
Active Can Off without SVC – This alert indicates that the Active Can feature is disabled without an SVC lead in place, which does not provide a viable defibrillation pathway. The device sounds a tone daily at the programmed time.
Rapid Current Drain EOS indicated – This alert indicates that the battery voltage is declining faster than expected at the time that low battery Recommended Replacement Time (RRT) has been detected. This situation reduces the 3-month Prolonged Service Period to device End of Service (EOS). If this alert condition is met, the device sounds a continuous patient alert tone for 30 s. This patient alert tone repeats every 4 hours, beginning at the next scheduled 4-hour time interval, and at the programmed Alert Time. Immediately contact your Medtronic representative if a Rapid Current Drain EOS alert occurs.
Unsuccessful Wireless Transmission – This alert indicates that the device attempted a wireless transmission but that the transmission was still unsuccessful after a 72-hour period.

3.2.4 Programming Alerts

The Medtronic CareAlert SETUP screen shows either a Lead/Device Integrity Alerts view or a Clinical Management Alerts view. To toggle between views, tap either Clinical Management Alerts… or Lead/Device Integrity Alerts….
Table 3. How to navigate to parameters for Lead/Device Integrity Alerts from the Menu
a
button
Parameters Path
Wireless Telemetry with Monitor Alert Time…
RV Lead alert parameters:
RV Lead Integrity RV Lead Noise
PARAMETERS > Alert…
PARAMETERS > Alert… > RV Lead… WIRE­LESS ALERT
28 Reference Manual
Medtronic Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Table 3. How to navigate to parameters for Lead/Device Integrity Alerts from the Menu buttona (continued)
Parameters Path
Lead Impedance Out of Range alert param­eters:
A. Pacing ENABLE A. Pacing LESS THAN A. Pacing GREATER THAN RV Pacing ENABLE RV Pacing LESS THAN RV Pacing GREATER THAN RV Defibrillation ENABLE RV Defibrillation LESS THAN RV Defibrillation GREATER THAN SVC Defibrillation ENABLE SVC Defibrillation LESS THAN SVC Defibrillation GREATER THAN
Capture Management High Threshold parameters:
A. Capture ENABLE RV Capture ENABLE
Low Battery Voltage RRT Excessive Charge Time EOS… VF Detection OFF, 3+ VF or 3+ FVT Rx Off.
a
The Lead/Device Integrity Alerts… label shares a toggle field with the Clinical
PARAMETERS > Alert… > Lead/Device Integrity Alerts… > Lead Impedance Out of Range… WIRELESS ALERT
PARAMETERS > Alert… > Lead/Device Integrity Alerts… > Capture Management High Threshold… WIRELESS ALERT
PARAMETERS > Alert… > Lead/Device Integrity Alerts… WIRELESS ALERT
Management Alerts… label.
Table 4. How to navigate to parameters for Clinical Management Alerts from the Menu
a
button
Parameters Path
Wireless Telemetry with Monitor Alert Time (OptiVol)… Alert Time…
OptiVol 2.0 Fluid Settings parameters
OptiVol Alert Enable OptiVol Threshold
AT/AF Burden and Rate Settings parameters
AT/AF Daily Burden ENABLE AT/AF Daily Burden BURDEN Avg. V. Rate During AT/AF ENABLE Avg. V. Rate During AT/AF BURDEN Avg. V. Rate During AT/AF V. RATE
PARAMETERS > Alert…
PARAMETERS > Alert… > OptiVol 2.0 Fluid Set­tings… WIRELESS ALERT
PARAMETERS > Alert… > AT/AF Burden and Rate Settings… WIRELESS ALERT
Reference Manual 29
Medtronic
Cobalt™ XT / Cobalt™ / Crome™ DR/VR MRI SureScan™ ICDs
Table 4. How to navigate to parameters for Clinical Management Alerts from the Menu buttona (continued)
Parameters Path
VT/VF Episodes and Therapies parameters
Monitored VT Episode Detected ENABLE Daily VT/VF Episodes ENABLE Weekly ATP Delivered Episodes ENABLE Weekly ATP Delivered Episodes THRESHOLDS Number of Shocks Delivered in an Episode ENA­BLE Number of Shocks Delivered in an Episode THRESHOLDS
Cumulative Right Ventricular Pacing > 40% PARAMETERS > Alert…
a
The Clinical Management Alerts… label shares a toggle field with the Lead/Device
PARAMETERS > Alert… > VT/VF Episodes and Therapies… WIRELESS ALERT
Integrity Alerts… label.
Transmitting alerts – The ability to transmit Medtronic alerts to the CareLink Network is programmable only when the Wireless Telemetry with Monitor parameter on the Medtronic CareAlert SETUP screen is programmed to On.
Repetitive alerts – If a programmable alert is triggered so often that it loses its clinical value, you can adjust the alert threshold, program the device to improve therapy effectiveness, or turn off the alert.

3.2.5 Evaluation of alert events

The Medtronic CareAlert EVENTS log categorizes alert events as CareAlert EVENTS and OptiVol EVENTS. For each alert event, a log entry includes the date and time of the alert, a description of the event, and the measurement or information that caused the event. Up to 15 CareAlert events and the last 7 OptiVol events are stored.
Caution: Verify lead integrity when evaluating OptiVol events. Loss of RVcoil integrity due to lead fracture or insulation defect may adversely affect OptiVol events.
From the Menu button, tap DATA > CareAlert EVENTS to see the alert events stored in the Medtronic CareAlert EVENTS log.

3.3 OptiVol 2.0 Fluid Status Monitoring

Clinical studies have shown that lung congestion is a primary complication associated with heart failure and is a frequent cause of repeated hospital admissions.
30 Reference Manual
Loading...
+ 296 hidden pages