Medtronic DDMC3D1 Reference Guide

EVERA MRI™ SURESCAN™ DR
Family of MR Conditional digital dual-chamber implantable cardioverter defibrillators with SureScan™ Technology
Reference Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
EVERA MRI™ SURESCAN™ DR
Reference Manual
A reference manual for the Medtronic Evera MRI SureScan DR family of digital dual-chamber implantable cardioverter defibrillators.
Active Can, ATP During Charging, Capture Management, Cardiac Compass, CareAlert, CareLink, ChargeSaver, Conexus, Evera, Evera MRI, Export, Flashback, Integrity, Intrinsic, Marker Channel, Marquis, Medtronic, Medtronic CareAlert, Medtronic CareLink, MVP, OptiVol, Paceart, PR Logic, Quick Look, Reactive ATP, SessionSync, SureScan, Switchback, T-Shock, TherapyGuide
Medtronic
EVERA MRI™ SURESCAN™ DR
Contents
1 Introduction .......................................................... 8
1.1 Introduction ......................................................... 8
1.2 Evera MRI SureScan DR ICD feature model matrix ...................... 10
2 Conducting a patient session with the programmer .................... 11
2.1 Establishing telemetry between the device and the programmer ........... 11
2.2 Starting and ending a patient session ................................. 16
2.3 Display screen features .............................................. 20
2.4 Delivering an emergency tachyarrhythmia therapy ....................... 25
2.5 Enabling emergency VVI pacing ...................................... 27
2.6 Suspending and resuming tachyarrhythmia detection .................... 28
2.7 Monitoring cardiac activity with the Live Rhythm Monitor .................. 30
2.8 Navigating a patient session with Checklist ............................. 39
2.9 Programming device parameters ...................................... 41
2.10 Saving and retrieving a set of parameter values ......................... 42
2.11 Using TherapyGuide to select parameter values ......................... 44
2.12 Storing patient information ........................................... 47
2.13 Printing reports ..................................................... 51
2.14 Transferring data to Paceart with SessionSync .......................... 53
2.15 Saving and retrieving device data ..................................... 60
2.16 Patient follow-up guidelines ........................................... 63
2.17 Optimizing device longevity ........................................... 67
3 Diagnostic data features ............................................. 71
3.1 Quick Look II summary data .......................................... 71
3.2 Medtronic CareAlert events and notifications ............................ 75
3.3 RV Lead Integrity Alert ............................................... 84
3.4 Device and lead performance data .................................... 91
3.5 OptiVol 2.0 fluid status monitoring .................................... 100
3.6 Cardiac Compass Trends ........................................... 105
3.7 Heart Failure Management Report .................................... 111
3.8 Arrhythmia Episodes data ........................................... 118
Reference Manual 5
Medtronic
3.9 Episode and therapy counters ....................................... 128
3.10 Flashback Memory data ............................................ 133
3.11 Rate Drop Response episodes ....................................... 134
3.12 Rate Histograms ................................................... 137
3.13 Automatic device status monitoring ................................... 139
4 Pacing features ..................................................... 143
4.1 Sensing .......................................................... 143
4.2 Basic pacing ...................................................... 153
4.3 Managed Ventricular Pacing (MVP) ................................... 163
4.4 Rate Response .................................................... 170
4.5 Capture Management .............................................. 177
4.6 Rate Adaptive AV .................................................. 188
4.7 Auto PVARP ...................................................... 190
4.8 Rate Drop Response ............................................... 192
4.9 Rate Hysteresis .................................................... 198
4.10 Sleep feature ...................................................... 200
4.11 Non-Competitive Atrial Pacing ....................................... 201
4.12 PMT Intervention .................................................. 203
4.13 PVC Response .................................................... 204
4.14 Ventricular Safety Pacing ............................................ 206
4.15 Mode Switch ...................................................... 208
4.16 Conducted AF Response ........................................... 211
4.17 Atrial Rate Stabilization ............................................. 213
4.18 Atrial Preference Pacing ............................................ 216
4.19 Post-Mode Switch Overdrive Pacing .................................. 220
4.20 Post Shock Pacing ................................................. 223
4.21 Post VT/VF Shock Overdrive Pacing .................................. 224
4.22 Ventricular Rate Stabilization ........................................ 225
EVERA MRI™ SURESCAN™ DR
5 Tachyarrhythmia detection features .................................. 228
5.1 AT/AF detection ................................................... 228
5.2 VT/VF detection ................................................... 234
5.3 PR Logic .......................................................... 249
5.4 Wavelet .......................................................... 253
5.5 Onset ............................................................ 260
6 Reference Manual
Medtronic
5.6 Stability ........................................................... 266
5.7 High Rate Timeout ................................................. 268
5.8 TWave Discrimination .............................................. 270
5.9 RV Lead Noise Discrimination ....................................... 274
6 Tachyarrhythmia therapy features ................................... 279
6.1 Atrial therapy scheduling ............................................ 279
6.2 Atrial ATP therapies ................................................ 286
6.3 Atrial cardioversion ................................................. 296
6.4 Patient-activated atrial cardioversion .................................. 302
6.5 VF therapies ...................................................... 307
6.6 Ventricular ATP therapies ........................................... 317
6.7 Ventricular cardioversion ............................................ 327
6.8 Progressive Episode Therapies ...................................... 336
7 System test and EP Study features .................................. 338
7.1 Underlying Rhythm Test ............................................. 338
7.2 Pacing Threshold Test .............................................. 339
7.3 Wavelet Test ...................................................... 341
7.4 Lead Impedance Test ............................................... 344
7.5 Sensing Test ...................................................... 345
7.6 Charge/Dump Test ................................................. 347
7.7 Arrhythmia inductions with EP Studies ................................ 348
7.8 Manual therapies with EP Studies .................................... 356
EVERA MRI™ SURESCAN™ DR
Glossary ................................................................ 359
Index ................................................................... 370
Reference Manual 7
Medtronic
EVERA MRI™ SURESCAN™ DR

1 Introduction

1.1 Introduction

This manual describes the operation and intended use of features offered by Medtronic Evera MRI SureScan DR ICD devices.
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan system to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection and all user-defined diagnostics. Before performing an MRI scan, refer to the MRI technical manual.
Throughout this manual, the word “device” refers to the implanted DR ICD device.
The programmer screen image examples in this manual are provided for reference only and may not match the final software.
The names of on-screen buttons are shown within brackets: [Button Name].
Tables in the feature programming sections summarize how to navigate to screens with programmable parameters for the feature. As shown in the example in Table 1, each table row lists a parameter or group of parameters with the path to a specific screen on the programmer. If the navigation path is the same for related parameters, the path is not repeated in the table. Additional rows are included for parameters that appear on different screens. Groups of parameters, such as “ATP parameters”, include the word “parameters”. Individual parameters, such as “Energy” and “Pathway”, do not.

Table 1. How to navigate to VF therapies parameters

Parameters Path
VF therapies (Rx1 through Rx6) VF Therapy Status (On, Off) Energy Pathway
ATP parameters (Rx1) Params > VF Therapies… > ATP…
ChargeSaver parameters (ATP in Rx1) Params > VF Therapies… > ATP… > During
Shared Settings (V. ATP and V. Therapies) Params > VF Therapies… > Shared Settings…
The feature programming sections also include programming considerations. For detailed information about parameter settings, see the device manual for the specific device.
8 Reference Manual
Params > VF Therapies…
Charging > ChargeSaver…
Medtronic
EVERA MRI™ SURESCAN™ DR

1.1.1 Product literature

Before implanting the device, it is recommended that you take the following actions:
Read the product literature for information about prescribing, implanting, and using the device and conducting a patient follow-up session.
Thoroughly read the technical manuals for the leads used with the device. Also read the technical manuals for other system components.
Discuss the device and implant procedure with the patient and any other interested parties, and give them any patient information materials packaged with the device.
The following manuals and documents also contain information about the device:
MRI technical manual – This manual provides MRI-specific procedures and warnings and precautions.
Device manual – Each device model has a separate device manual. The manual contains the specific features for the model, indications and contraindications, warnings and precautions, instructions for implanting the device, quick reference specifications, and parameter tables.
Explanation of symbols – This document defines the symbols that may appear on the device package. Refer to the package label to see which symbols apply specifically to this device.
Radio regulatory compliance insert – This document provides Federal Communications Commission (FCC) regulations and compliance information for the transmitter in this active implantable medical device.
Medical Procedure and EMI Warnings and Precautions Manual for Health Care Professionals – This manual provides warnings, precautions, and guidance for health care
professionals who perform medical therapies and diagnostic procedures on cardiac device patients. The manual also provides patient education information related to sources of electromagnetic interference (EMI) at home, at work, and in other environments.

1.1.2 Technical support

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users.
For more information, contact your local Medtronic representative or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
Reference Manual 9
Medtronic
EVERA MRI™ SURESCAN™ DR

1.1.3 Notice

The Patient Information screen of the programmer software application is provided as an informational tool for the end user. The user is responsible for accurate input of patient information into the software. Medtronic makes no representation as to the accuracy or completeness of the patient information that end users enter into the Patient Information screen. Medtronic SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY THAT RESULT FROM THE USE OF THE PATIENT INFORMATION SUPPLIED BY END USERS TO THE SOFTWARE.
For more information, see Section 2.12, “Storing patient information”, page 47.

1.2 Evera MRI SureScan DR ICD feature model matrix

Feature availability for each device model is marked with an “X” in the corresponding column.

Table 2. Product feature relationship

Evera MRI XT DR
SureScan
Features
Optivol 2.0 fluid status monitoring X
Heart Failure Management Report X
DDMB1D4 DDMB1D1
Evera MRI S DR SureScan
DDMC3D4 DDMC3D1
Note: All other features described in this manual apply to all Evera MRI SureScan DR ICD
devices.
10 Reference Manual
Medtronic EVERA MRI™ SURESCAN™ DR

2 Conducting a patient session with the programmer

2.1 Establishing telemetry between the device and the programmer

You can conduct a patient session using wireless or nonwireless telemetry.
You cannot switch between wireless and nonwireless telemetry during a patient session. If you are conducting a patient session and you wish to change the telemetry mode, you must end that session and start a new session using the alternative mode.
Refer to the programmer reference guide for information about setting up the programmer for a patient session.

2.1.1 Using Conexus wireless telemetry

To establish wireless telemetry, use the Medtronic programmer with Conexus telemetry or the Medtronic Conexus Activator. The Conexus Activator is a hand-held, battery-powered communication device. It enables wireless telemetry in a Conexus-compatible heart device independently of a programmer. Once wireless telemetry is established, a practitioner can use the programmer to conduct a session without using the programming head.
Conexus wireless telemetry uses the Medical Implant Communications Service (MICS) radiofrequency band, which is designated worldwide for medical devices. Using this band protects devices against interference from home electronics, such as microwaves, cell phones, and baby monitors.
Conexus wireless telemetry is designed for use during implant and follow-up sessions.
At implant, Conexus wireless telemetry enables you to:
interrogate the patient’s wireless device without using a programming head (no programming head in the sterile field)
maintain connectivity during induction and delivery of therapies
program the device anytime during the procedure while maintaining continuous patient monitoring
During follow-up sessions, Conexus wireless telemetry maintains continuous communication between the device and the programmer.
1
1
Medical Implant Communications Service (MICS) is also referred to as the core 402-405 MHz band of the Medical Device Radiocommunication Service (MedRadio).
Reference Manual 11
Medtronic
During remote follow-up sessions with the CareLink Network, Conexus wireless telemetry automatically transmits comprehensive arrhythmia and diagnostic device data. It transmits wirelessly without patient involvement.
EVERA MRI™ SURESCAN™ DR
2.1.1.1 How to activate wireless telemetry
1. Turn on the programmer.
Make sure the “Allow wireless communication” check box in the Find Patient window is selected.
2. Use the Conexus Activator, or briefly place the programming head over the device to activate wireless telemetry in the device.
When wireless telemetry is first established during a session, the telemetry status indicator in the upper left corner of the task bar changes from the programming head icon to the wireless telemetry icon, as shown in Figure 1 .

Figure 1. Wireless telemetry icon on the task bar

1 Wireless telemetry icon
The indicator bar on the icon displays the strength of the wireless signal. At least 3 of the green lights must be illuminated to ensure reliable telemetry has been established.
If you are using the Conexus Activator, press the blue button to activate wireless telemetry. A green light illuminates when you have successfully communicated with the device.
2.1.1.2 Conexus wireless timer operation
Once you initiate wireless telemetry, the device sends a signal to the programmer and remains active for a period of time. A response from the programmer establishes communication, and the device appears in the Find Patient window when that window is open. If the programmer touch pen is not used within 5 min, the Find Patient window closes and the Select Model screen appears.
When you end a session, you have a short period of time where you can re-interrogate the device before the session ends completely.
12 Reference Manual
Medtronic
EVERA MRI™ SURESCAN™ DR
The inactivity timer is 2 hours for the first week after implant, at which time it changes to 5 min automatically. If the Extend Wireless Telemetry Session feature is available, you have the option to extend the inactivity timer from 5 min to 2 hours in any patient session. When the session is ended, the timer reverts to 5 min. This feature reduces device battery depletion.
If electrical interference disrupts a session, the programmer attempts to re-establish communication with the device. If you are unable to re-establish communication between the device and the programmer, use the Conexus Activator or programming head to reactivate wireless telemetry in the device to resume or start a session.
2.1.1.3 How to maintain reliable telemetry
You can expect reliable wireless telemetry between the implanted device and the programmer in a typical examination room or operating room. If other electrical equipment is in the area, the system is designed to maintain effective communication between the device and the programmer at distances up to 2 m (6.5 feet). The system should not interfere with other electronic equipment in the area.
If you are having trouble maintaining consistent, reliable telemetry, take one or more of the following actions:
Adjust the angle of the programmer screen. The telemetry antenna is part of the programmer display screen structure; slight movements of the screen may improve the telemetry link.
Change the position of the programmer so that the space between the programmer screen and the patient is relatively free of obstruction. Make sure that nothing is between the programmer and the patient.
Shorten the distance between the programmer and the patient.
Remove any sources of electromagnetic interference (EMI) that may be affecting the telemetry signal.
2.1.1.4 Session inactivity safeguards
If you or the patient moves away from the programmer, the system guards against unintentional programming in the following ways:
After 2 min of programmer inactivity, the system displays the patient’s name or ID number, if available, and device information. You are then required to confirm that the correct patient is in the follow-up session before you can process a programming command.
After 2 hours of programmer inactivity in an implant session or a follow-up session, the device transitions to Standby mode.
Reference Manual 13
Medtronic
EVERA MRI™ SURESCAN™ DR
Standby mode – Use Standby mode when you need a period of inactivity in a patient session. In this mode, live waveforms are turned off, and the programmer telemetry status indicator shows no telemetry link. Programmer functions are limited.
You can activate and deactivate Standby mode in the device manually:
To activate Standby mode, select the wireless telemetry icon on the task bar.
To deactivate Standby mode and reactivate wireless telemetry with the programmer, select the wireless telemetry icon on the task bar or place either the Conexus Activator or the programming head over the device.
After a brief period of time in Standby mode, wireless telemetry is deactivated within the device. To recover the wireless telemetry session, use either the Conexus Activator or the programming head.
Standby mode is also deactivated when you attempt to program parameters, interrogate the device, or conduct testing or emergency operations. The programmer screen displays the Verify Patient warning. To deactivate Standby mode and resume a patient session, verify that the session is with the intended patient, select the “Allow communication with” check box, and select [Continue].
Note: To use Holter telemetry to transmit EGM and Marker Channel data during a Conexus telemetry session, you first must activate Standby mode.
2.1.1.5 Device standby mode
When the device is in standby mode, live waveforms are turned off, the programmer telemetry status indicator shows no telemetry link, and programmer functions are limited. You can use standby mode when you need a period of inactivity in a patient session.
To activate and deactivate Standby mode in the device manually:
To activate Standby mode, select the wireless telemetry icon on the task bar.
To deactivate Standby mode and reactivate wireless telemetry with the programmer, select the wireless telemetry icon on the task bar or place either the Conexus Activator or the programming head over the device.
After a brief period of time in Standby mode, wireless telemetry is deactivated within the device. To recover the wireless telemetry session, use either the Conexus Activator or the programming head.
Standby mode is also deactivated when you attempt to program parameters, interrogate the device, or conduct testing or emergency operations. The programmer screen displays the Verify Patient warning. To deactivate Standby mode and resume a patient session, verify that the session is with the intended patient, select the “Allow communication with” check box, and select [Continue].
14 Reference Manual
Medtronic
EVERA MRI™ SURESCAN™ DR
Note: To use Holter telemetry to transmit EGM and Marker Channel data during a Conexus telemetry session, you first must activate Standby mode.
2.1.1.6 Managing wireless telemetry session duration
Reducing wireless telemetry session duration may help reduce the energy requirements placed on the battery.
At implant and in the first 7 days after implant, the wireless telemetry session duration is 2 hours to provide time for implant, programming, and patient follow-up tasks. The device transitions to Standby mode after 2 hours of programmer inactivity.
7 days after implant, the wireless telemetry session duration is reduced automatically to 5 min to help preserve device longevity. The device transitions to Standby mode after 5 min of programmer inactivity. If patient follow-up tasks require it and if the Extend Wireless Telemetry Session feature is available, you can optionally extend the wireless telemetry for a patient session to 2 hours by selecting Session > Extend Wireless Telemetry Session and then selecting [Continue]. At the end of the session, the wireless telemetry session duration for the patient reverts automatically to 5 min.
2.1.1.7 How to maintain patient safety and privacy
During a wireless telemetry session, all other programmers are blocked from communicating or initiating a session with the patient’s implanted device. Implanted devices in other patients are locked out from any communication or programming occurring during the patient’s session.
When you are using wireless telemetry, the patient’s name is displayed on the Command bar of the programmer screen. If you have not entered the patient’s name, the patient’s ID number appears. If the patient’s name or ID has not been entered, then “Patient name is not entered” appears on the Command bar. Enter the patient’s name and ID number as early as possible to assist with patient identification when using wireless telemetry.

2.1.2 Using nonwireless telemetry

Some Medtronic programmers feature only nonwireless telemetry. If your Medtronic programmer features both Conexus wireless telemetry and nonwireless telemetry, you must choose to use nonwireless telemetry.
Note: You must have a Medtronic programming head to use nonwireless telemetry.
Reference Manual 15
Medtronic
EVERA MRI™ SURESCAN™ DR
2.1.2.1 How to establish nonwireless telemetry
1. Turn on the programmer.
If you are using a programmer with Conexus wireless telemetry, make sure the “Allow wireless communication” check box in the Find Patient window is not selected. The check box does not appear if you are using a programmer without Conexus wireless telemetry.
2. Place the programming head over the device to activate nonwireless telemetry in the device.
When nonwireless telemetry is established during a session, the telemetry status indicator on the task bar displays the programming head icon, as shown in Figure 2.

Figure 2. Programming head icon on the task bar

1 Programming head icon
Note: The magnet in the programming head can suspend tachyarrhythmia detection. However, when telemetry between the device and the programmer is established, detection is not suspended.
When telemetry is established, the amber light on the programming head turns off, and 1 or more of the green indicator lights on the programming head illuminate. To ensure that the proper telemetry is established, position the programming head over the device so at least 2 of the green lights illuminate. If the programming head slides off the patient, the session does not terminate. Place the programming head back over the device to resume programming or interrogating the device.
Note: More information about the general use of the programming head is available in the programmer reference guide.

2.2 Starting and ending a patient session

The programmer interrogates the patient’s device at the start of a patient session. Because the programmer collects and stores data on a session-by-session basis, start a new session for each patient. End the previous session before starting a session with another patient.
16 Reference Manual
Medtronic
EVERA MRI™ SURESCAN™ DR
Caution: A programmer failure (for example, a faulty touch pen) could result in inappropriate programming or the inability to terminate an action or an activity in process. In the event of a programmer failure, immediately turn the programmer power off to deactivate telemetry and terminate any programmer-controlled activity in process.
Caution: During a wireless telemetry session, verify that you have selected the appropriate patient before proceeding with the session, and maintain visual contact with the patient for the duration of the session. If you select the wrong patient and continue with the session, you may inadvertently program the wrong patient’s device.
Caution: Do not leave the programmer unattended while a wireless telemetry session is in progress. Maintain control of the programmer during the session to prevent inadvertent communication with the patient’s device.
Note: During an initial interrogation, only Emergency programmer functions are available.

2.2.1 How to start a patient session using wireless telemetry

1. Select [Find Patient…] from the Select Model window.
2. Select the “Allow wireless communication” check box on the Find Patient window.
3. Use the Conexus Activator, or briefly place the programming head over the device to activate wireless telemetry in the device.
Notes:
When the Conexus Activator is used to activate telemetry in the device, the programmer launches the patient session without suspending tachyarrhythmia detection. Placing a magnet near the device, however, suspends tachyarrhythmia detection.
When the programming head is used to activate telemetry in the device, the programmer automatically launches the patient session with tachyarrhythmia detection suspended. Detection remains suspended as long as the programming head is over the device. If tachyarrhythmia detection is programmed on, a warning reminds you that tachyarrhythmia detection is suspended.
4. Select the appropriate patient from the Patient Name list on the Find Patient window.
Note: The programmer lists all patients with wireless-activated implantable devices within telemetry range.
5. Select [Start].
Reference Manual 17
Medtronic
EVERA MRI™ SURESCAN™ DR

2.2.2 How to start a patient session using nonwireless telemetry

1. Select [Find Patient…] from the Select Model window.
2. If you are using a Medtronic programmer with Conexus wireless telemetry, make sure that the “Allow wireless communication” check box on the Find Patient window is not selected. If you start a session with the programming head over the patient’s device and the “Allow wireless communication” check box is selected, the system initiates a wireless telemetry session and automatically interrogates the device. If you are using a Medtronic programmer without Conexus wireless telemetry, the “Allow wireless communication” check box does not appear on the Find Patient window.
3. Place the programming head over the device and the nonwireless session automatically begins.

2.2.3 Device and telemetry effects during a patient session

Tachyarrhythmia detection during a wireless telemetry session – If you place a
programming head over the device, the magnet in the programming head always suspends tachyarrhythmia detection.
Tachyarrhythmia detection during a nonwireless telemetry session – If you place a programming head over the device and telemetry is established, the magnet in the programming head does not suspend tachyarrhythmia detection.
Episodes in progress during a wireless telemetry session – If you attempt to initiate a patient session when a detected arrhythmia episode is in progress, the device treats the arrhythmia normally. If telemetry has not been established, the magnet inside the programming head causes the device to suspend detection when the programming head is placed over the device.
Episodes in progress during a nonwireless telemetry session – After telemetry has been established and you position the programming head over the device when a detected arrhythmia episode is in progress, the device treats the arrhythmia normally. If telemetry has not been established and you position the programming head over the device, the magnet inside the programming head causes the device to suspend detection.
Capacitor charging during a wireless telemetry session – Interference caused by capacitor charging may affect telemetry between the device and the programmer. This interference could result in a temporary loss of telemetry indicator lights as shown on the programmer task bar and a temporary loss in Marker transmissions. It could also temporarily affect the ability to send programming commands. Ensure that the greatest number of telemetry strength indicator lights are illuminated on the programmer task bar to help improve telemetry reliability before any manual or automatic capacitor charging.
18 Reference Manual
Medtronic
Capacitor charging during a nonwireless telemetry session – Interference caused by capacitor charging may affect telemetry between the device and the programmer. The programming head indicator lights may turn off during charging periods. It is normal for the lights to turn off on the programming head.
Note: The programming head “P” button is disabled during all EP study and manual system tests. During tachyarrhythmia inductions, the programming head “I” button is also disabled.
Marker transmissions during a wireless telemetry session – The device continuously transmits Marker Channel and supplementary marker data while telemetry is established. The device stops these transmissions when telemetry is interrupted. If Holter telemetry is programmed to On, the device transmits telemetry at all times except during a Conexus wireless telemetry session. To use Holter telemetry during a Conexus wireless telemetry session, you must first activate Standby mode.
Marker transmissions during a nonwireless telemetry session – The device continuously transmits Marker Channel and supplementary marker data while telemetry is established and the programming head is positioned over the device. The device stops these transmissions when you lift the programming head, unless the Holter Telemetry feature is programmed to On. If Holter Telemetry is programmed to On, the device transmits Marker Channel and supplementary marker data regardless of the position of the programming head.
Device longevity and wireless telemetry – In typical patient session and device operation scenarios, wireless telemetry has no significant effect on device longevity.
EVERA MRI™ SURESCAN™ DR

2.2.4 How to interrogate the device during the session

At the start of the patient session, the programmer interrogates the device. You can manually interrogate the device at any time during the patient session by performing the following steps:
1. Select [Interrogate…] from the Command bar. In a nonwireless session, you can also interrogate the device by pressing the “I” button on the programming head.
2. To gather information collected since the last patient session, select the Since Last Session option from the interrogation window. To gather all of the information from the device, select the All option.
3. Select [Start].
Note: You cannot manually interrogate the device during an emergency programmer operation. You must select [Exit Emergency] before you can manually interrogate the device.
Reference Manual 19
Medtronic
EVERA MRI™ SURESCAN™ DR

2.2.5 How to end a patient session

1. To review a list of programming changes made during this session, select Session > Changes This Session.
2. To print a record of the changes, select [Print…].
3. Select [End Session…].
4. To save the session data to a USB flash drive or a disk, if available, select [Save To Media…].
5. To end the session and return to the Select Model screen, select [End Session…] and then [End Now].

2.3 Display screen features

The programmer display screen is an interface that displays text and graphics. It is also a control panel that displays buttons and menu options that you can select by using the touch pen.
The main elements of a typical display screen during a patient session are shown in Figure 3.
20 Reference Manual
Medtronic

Figure 3. Main elements of a display screen

EVERA MRI™ SURESCAN™ DR
1 Task bar 2 Status bar 3 Live Rhythm Monitor window
4 Task area 5 Command bar 6 Tool palette

2.3.1 Task bar

The display screen features a task bar at the top of the screen. You can use the task bar to note the status of programmer-specific features such as the analyzer.
The task bar also includes a graphical representation of the telemetry strength indicator. In a wireless telemetry session, selecting the wireless telemetry icon breaks the telemetry link. Selecting it again restores the telemetry link. If you are conducting a nonwireless telemetry session, the task bar includes a graphical representation of the telemetry strength light array on the programming head.

Figure 4. Task bar display

Reference Manual 21
1 2 3 4 6
5
Medtronic
EVERA MRI™ SURESCAN™ DR
1 Telemetry icon and telemetry strength
indicator (wireless telemetry shown) 2 SessionSync icon 3 Disk icon (for some Medtronic programmer
models)
4 USB icon 5 Analyzer icon 6 Device icon
If the SessionSync feature is installed on the programmer, the programmer task bar displays an icon that indicates the status of the SessionSync feature. For complete information on viewing the status of the SessionSync feature from the programmer task bar, see Section 2.14, “Transferring data to Paceart with SessionSync”, page 53.

2.3.2 Status bar

When the device has been interrogated, you can use the status bar at the top of the display screen (located immediately below the task bar) to perform some basic functions and to note the current status of the device.

Figure 5. Status bar display

1 Currently active pacing mode 2 Programmed detection and therapy configuration 3 Buttons used to resume or suspend detection 4 Automatic detection status 5 Indicator that a tachyarrhythmia episode is in progress 6 Either the current episode, therapy, or manual operation status, or the device name and model
number

2.3.3 Live Rhythm Monitor window

The Live Rhythm Monitor window displays ECG, Leadless ECG, Marker Channel, and telemetered EGM waveform traces. In addition to waveform traces, the Live Rhythm Monitor window shows the following information:
The heart rate and the rate interval appear if telemetry has been established with the device.
The annotations above the waveform trace show the point at which parameters are programmed.
22 Reference Manual
Medtronic
The Live Rhythm Monitor window appears in the partial view by default. You can expand this window to its full size by selecting the small square button in the upper right corner of the window or by selecting [Adjust…]. For more information, see Section 2.7, “Monitoring cardiac activity with the Live Rhythm Monitor”, page 30.
EVERA MRI™ SURESCAN™ DR

2.3.4 Task area

The portion of the screen between the Live Rhythm Monitor window near the top of the screen and the command bar at the bottom of the screen changes according to the task or function you select.
One example of a task area is the Parameters screen, which is used to view and program device parameters as described in Section 2.9, “Programming device parameters”, page 41.

Figure 6. Task area of the screen

2.3.5 Tool palette

The buttons and icons along the right edge of the screen are referred to as the “tool palette”. You can use these tools to display a task or function screen. After starting a patient session, the tool palette is displayed on all but the Emergency or Live Rhythm Monitor Adjust… screens, making it quick and easy to move to the desired task or function.
Each of the icons acts like a button. To select an icon, touch the icon with the touch pen. Each option in the tool palette is described in Table 3.
Reference Manual 23
Medtronic

Table 3. Tool palette options

The [Freeze] button captures a segment of the Live Rhythm Monitor dis­play.
The [Strips…] button accesses the waveform strips saved since the start of the session.
The [Adjust…] button opens a window of options for adjusting the Live Rhythm Monitor display.
The Checklist icon opens the Checklist screen for simplified navigation through a set of follow-up tasks. The Checklist [>>] button navigates to the next task in the Checklist.
The Data icon displays options for viewing device information and diag­nostic data.
The Params icon displays the Parameters screen for viewing and pro­gramming device parameters.
The Tests icon displays options for performing system tests and EP stud­ies.
EVERA MRI™ SURESCAN™ DR
The Reports icon displays options for printing reports.
The Patient icon displays options for accessing the TherapyGuide screen or the Patient Information screen.
The Session icon displays options for adjusting preferences, viewing parameter changes made during the session, saving data, and ending the session.

2.3.6 Buttons

Buttons, such as those shown in Figure 7, respond when you “select” them by touching them with the tip of the touch pen.

Figure 7. Display screen buttons

Buttons with a less distinctly shaded label are inactive and do not respond if you select them.
24 Reference Manual
Medtronic
EVERA MRI™ SURESCAN™ DR
Selecting a button with the touch pen causes one of the following responses:
Buttons such as the [PROGRAM] button execute a command directly.
Buttons such as the [Save…] and [Get…] buttons open a window that prompts another action. The labels on these buttons end with an ellipsis.
A procedure may instruct you to “press and hold” a button. In such cases, touch the tip of the touch pen to the button and continue to maintain pressure against the button. The button continues to respond to the touch pen until you remove the touch pen from the button.

2.3.7 Command bar

The bar at the bottom of the screen always shows the buttons for programming Emergency parameters, interrogating the device, and ending the patient session.
If the programmer is using wireless telemetry, the patient may be identified on the command bar of the programmer screen. Depending on the programmed patient information, one of the following text fields appears:
the patient name
the patient ID, if the patient name was not entered
the message “(Patient name not entered)”, if neither the name nor the ID was entered
Note: The [Interrogate…] and [End Session…] buttons do not appear on the Emergency screen.

Figure 8. Command bar

2.4 Delivering an emergency tachyarrhythmia therapy

You can use emergency defibrillation, cardioversion, and fixed burst pacing therapies to quickly treat ventricular tachyarrhythmia episodes during a patient session. Emergency defibrillation therapy delivers a high-voltage biphasic shock at the selected energy level. Emergency cardioversion therapy also delivers a high-voltage biphasic shock, but it must be synchronized to a ventricular event. Emergency fixed burst pacing therapy delivers maximum output pacing pulses to the ventricle at the selected interval.
Reference Manual 25
Medtronic
EVERA MRI™ SURESCAN™ DR

2.4.1 Considerations for emergency tachyarrhythmia therapies

Tachyarrhythmia detection during emergency tachyarrhythmia therapies – The
device suspends the tachyarrhythmia detection features when emergency defibrillation, cardioversion, or fixed burst pacing therapies are delivered. Select [Resume] to re-enable tachyarrhythmia detection.
Temporary parameter values – Emergency tachyarrhythmia therapies use temporary parameter values that do not change the programmed parameters of the device. After the tachyarrhythmia therapy is complete, the device reverts to its programmed parameter values.
Aborting an emergency tachyarrhythmia therapy – You can immediately terminate an emergency defibrillation or emergency cardioversion therapy by selecting [ABORT]. To stop an emergency fixed burst therapy, remove the touch pen from [BURST Press and Hold].
Emergency tachyarrhythmia therapies and MRI SureScan – If you deliver any emergency tachyarrhythmia therapy when MRI SureScan is programmed to On, MRI SureScan is automatically programmed to Off.

2.4.2 How to deliver an emergency tachyarrhythmia therapy

1. Establish telemetry with the device.
2. Select [Emergency].
3. Select the type of emergency therapy to deliver: Defibrillation, Cardioversion, or Fixed Burst.
4. Accept the therapy parameters shown on the screen, or select new values.
26 Reference Manual
Medtronic
EVERA MRI™ SURESCAN™ DR
5. For defibrillation and cardioversion therapy, select [DELIVER]. For fixed burst therapy, select [BURST Press and Hold] and hold the touch pen over the button for as long as you want to deliver the therapy.

2.5 Enabling emergency VVI pacing

You can use emergency VVI pacing to quickly enable 70 bpm, high-output ventricular pacing to restore ventricular support in an emergency situation.
Note: If you enable emergency VVI pacing when MRI SureScan is programmed to On, MRI SureScan is automatically programmed to Off.

2.5.1 Considerations for emergency VVI pacing

Parameter values – Emergency VVI pacing reprograms pacing parameters to emergency
settings. For a list of the emergency VVI parameter settings, see the device manual for the device. To terminate emergency VVI pacing, you must reprogram pacing parameters from the Parameters screen.

2.5.2 How to enable emergency VVI pacing

1. During a patient session, establish telemetry with the device.
2. Press the red VVI button on the programmer to enable emergency VVI pacing.
Depending on your model of Medtronic programmer, the emergency VVI button is:
a mechanical red button to the left of the programmer screen, on the programmer bezel.
a red button on the programmer button panel, above the programmer screen.
Note: On all programmers, an [Emergency] button is implemented in the software and is available on the display screen (see the following procedure for instructions regarding this button).
How to enable emergency VVI pacing with the on-screen [Emergency] button –
Perform the following steps to enable emergency VVI pacing with the on-screen [Emergency] button:
Reference Manual 27
Medtronic
1. Establish telemetry with the device.
2. Select [Emergency].
3. Select VVI Pacing.
4. Select [PROGRAM].
EVERA MRI™ SURESCAN™ DR

2.6 Suspending and resuming tachyarrhythmia detection

It may be necessary to turn off tachyarrhythmia detection in some situations. For example, during emergency therapies and some EP study tests, therapies are delivered manually, and detection and episode storage are not needed. Also, certain types of surgery, including electrocautery surgery, RF ablation, and lithotripsy, can cause the device to detect tachyarrhythmias inappropriately and possibly deliver inappropriate therapy.
When detection is suspended, the device temporarily stops the process of classifying intervals for tachyarrhythmia detection. Sensing and bradycardia pacing remain active, and the programmed detection settings are not modified. When the device resumes detection, it does so at the previously programmed detection settings. The Suspend/Resume function applies to both atrial and ventricular tachyarrhythmia detection.
Note: If MRI SureScan is programmed to On, tachyarrhythmia detection and Medtronic CareAlert events (including audible alerts) are suspended.

2.6.1 Considerations for suspending detection

If you suspend detection during a tachyarrhythmia detection process but before detection has occurred, the initial detection never occurs. When you resume, detection starts over.
28 Reference Manual
Medtronic
If you suspend detection after a tachyarrhythmia detection has occurred and resume detection before the tachyarrhythmia episode terminates, redetection works differently for each type of episode, as follows:
AT/AF episodes – If you suspend detection during a detected AT/AF episode and then resume detection before the episode terminates, detection starts over for the same episode.
Note: Suspending tachyarrhythmia detection does not affect Mode Switch. A Mode Switch may occur whether or not tachyarrhythmia detection has been suspended.
VT/FVT/VF episodes – If you suspend detection while a therapy is being delivered, the device finishes delivering the therapy that is in progress but does not redetect until you resume detection. If you resume detection before the episode terminates, the device begins redetection, and the episode is redetected if the programmed Beats to Redetect value is reached.
VT Monitor episodes – If you suspend detection during a detected VT Monitor episode, and then resume detection before the episode terminates, there will be episode data storage for 2 episodes with the first episode terminated while the rate is still fast.
EVERA MRI™ SURESCAN™ DR

2.6.2 How to suspend or resume detection with the programmer

Figure 9. [Suspend] and [Resume] buttons

The [Suspend] and [Resume] buttons can be used whenever there is telemetry with the device and the device software is running.
1. To suspend detection, select [Suspend]. The programmer displays a SUSPENDED annotation on the status bar.
2. To resume detection, select [Resume].

2.6.3 How to suspend or resume detection with a magnet

1. To suspend detection, place the magnet (such as the Model 9466 Tachy Patient Magnet) over the device.
2. To resume detection, remove the magnet from over the device.
Note: The programming head contains a magnet. When the programmer is using wireless telemetry, you can suspend detection by placing the programming head over the device. For more information, see Section 2.1, “Establishing telemetry between the device and the programmer”, page 11.
Reference Manual 29
Medtronic
EVERA MRI™ SURESCAN™ DR

2.7 Monitoring cardiac activity with the Live Rhythm Monitor

The Live Rhythm Monitor window displays ECG, Leadless ECG (LECG), Marker Channel with marker annotations, and telemetered EGM waveform traces on the programmer screen. The Live Rhythm Monitor window also displays the patient heart rate and interval in the upper-left corner of the window. You can view live waveform traces, freeze waveform traces, record live waveform traces to the programmer’s strip chart recorder or Electronic Strip Chart (eStrip) recorder, whichever is available, and recall any saved waveform strips prior to ending a patient session.
By default, the Live Rhythm Monitor window appears in partial view. You can expand this window to its full size by selecting the small square button in the upper-right corner of the window or by selecting the [Adjust…] button. Waveform traces display depending on which waveform source is selected and how waveform traces have been arranged in the full-screen view.

2.7.1 Types of live waveform traces

2.7.1.1 Leadless ECG (LECG) waveform
Leadless ECG simplifies and expedites patient follow-up sessions by providing an alternative to obtaining an ECG signal without the need to connect surface leads to the patient. Leadless ECG is available in the clinic and at remote locations where the CareLink Network is available.
Leadless ECG provides a far-field view of cardiac activity without connecting leads to the patient. The Leadless ECG (LECG) waveform displays an approximation of a surface ECG signal through the Can to SVC source. The Can to SVC source is available only when an SVC coil is present. You can also choose to display the waveform from the RVcoil to Aring source or the Can to Aring source on the LECG channel. This signal is telemetered from the device and is selected from the programmable LECG source when you set up data collection.
2.7.1.2 ECG waveforms
The ECG Lead I, ECG Lead II, and ECG Lead III waveforms display ECG signals that are detected using skin electrodes attached to the patient. The ECG cable attached to these electrodes must be connected to the programmer.
30 Reference Manual
Loading...
+ 356 hidden pages