MEDTRONIC DBS
CHARGING
CONTENTS
EXPERIENCE
The 37651 charging system
Quick guide
Charging system components
Checking the battery status
Charger information screen
MEDTRONIC DBS
Recharge frequency in days
75th percentile 90th percentile
CHARGING
EXPERIENCE
The 37651 charging system
The Medtronic 37651 DBS charging system guarantees patient comfort and
safety during the charging experience thanks to:
■ Temperature control during recharging
The system constantly monitors the temperature and the antenna connection.
If the temperature rises above a certain level:
- The system displays a warning message
- The charging power is reduced until
the antenna cools down
■ Temperature control function
- Monitors the antenna temperature every minute
- Automatically shuts o when the temperature rises above 38.3°C
- Charging restarts automatically when the temperature returns to a safe level.
■ Adhesive disc to improve the attachment of the antenna during charging (Ref.: 040618)
1 3 42
These double-sided adhesive discs provide better attachment of the antenna during charging,
keeping the head of the antenna from moving unintentionally and thus improving the charging
time. It is recommended not to place the adhesive directly on the skin.
■ Long-term battery performance
The proven
*
long-term performance of the battery used in Activa™ RC neurostimulators shows
that patients will experience a similar charging time and frequency over the entire service life of
the device.
The recharge frequency corresponds
30
25
20
15
10
5
0
1050
Time from implant in years
50th percentile25th percentile
to maximum time allowed between
recharge events. In practice more
regular charge behaviour might be seen.
The percentile corresponds to patient
groups that need dierent energy levels
1
Medtronic data on le, document
number: NDHF1172-159780
1
The Activa™ RC can maintain consistent recharge intervals over 15 years to minimally impact the
patient experience
*
.
For more information about the charging system, see the relevant product and therapy manuals.
*
Medtronic data on le, document number: NDHF1172-159780
1
.