Medtronic D334DRG Errata

Software Errata Sheet for the Model SW009 Software for the Medtronic 2090 CareLink Programmer
The following table lists certain conditions that may occur during use of the Model SW009 Software. The table also provides the resolution for these conditions.
Condition Observed Behavior Suggested Corrective Action
Termination of in-clinic session: During communication
with the implantable device, including interrogations, tests, Emergency selection, or while programming permanent or test parameters, the programmer may display a system error and terminate the session. This error has a low frequency of occurrence.
The following are trademarks of Medtronic: CareLink, Medtronic, Medtronic CareLink.
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The Programmer displays a black screen with a message indicating that an error has occurred. This terminates the session.
Note: The red Emergency VVI button is available.
Turn the Programmer Off and On to restart the session.