Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
ENTRUST® D153ATG, D153DRG,
D153VRC
Reference Manual
A guide to the operation and programming of the EnTrust Implantable Cardioverter Defibrillator
systems
The following list includes trademarks or registered trademarks of Medtronic in the
United States and possibly in other countries. All other trademarks are the property
of their respective owners.
Index ................................................................... 489
10Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Introduction
About this manual
This manual describes the operation and intended use of the EnTrust Model D153ATG,
D153DRG, and D153VRC systems.
EnTrust Model D153ATG devices, referred to as “AT” devices in this manual, provide atrial
and ventricular tachyarrhythmia detection and therapy and a full range of dual chamber
bradycardia pacing modes and associated features. Unless otherwise noted, all information
in this manual applies to AT devices.
EnTrust Model D153DRG devices, referred to as “DR” devices in this manual, provide atrial
tachyarrhythmia monitoring, dual chamber tachyarrhythmia detection, ventricular
tachyarrhythmia therapy, and a full range of dual chamber bradycardia pacing modes and
associated features. Information in this manual that describes atrial tachyarrhythmia
detection features (other than atrial monitoring) and atrial tachyarrhythmia therapy features
does not apply to DR devices.
EnTrust Model D153VRC devices, referred to as “VR” devices in this manual, provide
ventricular tachyarrhythmia detection, ventricular tachyarrhythmia therapy, and single
chamber bradycardia pacing. Information in this manual that describes atrial
tachyarrhythmia detection, atrial tachyarrhythmia therapy, dual chamber detection, dual
chamber pacing features, and atrial pacing modes does not apply to VR devices.
Programmer hardware and screen images
The screen image examples in this manual show the Medtronic CareLink Model 2090
programmer screen. Wherever possible, these screen images show the application for an
AT device (EnTrust Model D153ATG).
The information provided in this manual about using the programmer assumes the
Medtronic CareLink Model 2090 Programmer is used. For information about using the
Model 9790C Programmer, see the 9790/9790C Programmer Instruction Manual.
Manual conventions
Throughout this document, the word “device” refers to the implanted EnTrust device.
The symbol in parameter tables indicates the Medtronic nominal value for that parameter.
On-screen buttons are shown with the name of the button surrounded by brackets: [Button
Name].
Reference Manual11
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Additional literature
Before implanting the device, it is strongly recommended that you take the following actions:
●
Refer to the product literature packaged with the device for information about
prescribing the device.
●
Thoroughly read the technical manuals for the leads used with the device.
●
Discuss the procedure and the device with the patient and any other interested parties,
and provide them with any patient information packaged with the device.
Technical support
Medtronic employs highly trained representatives and engineers located throughout the
world to serve you and, upon request, to provide training to qualified hospital personnel in
the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical
consultation to product users.
For more information, contact your local Medtronic representative, or call or write Medtronic
at the appropriate telephone number or address listed on the back cover.
Customer education
Medtronic invites physicians to attend an educational seminar on the device. The course
describes indications for use, system functions, implant procedures, and patient
management.
References
The primary reference for background information is Zacouto FI, Guize LJ. Fundamentals
of Orthorhythmic Pacing. In: Luderitz B, ed. Cardiac Pacing Diagnostic and TherapeuticTools. New York: Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
●
Estes M, Manolis AS, Wang P, Eds. Implantable Cardioverter-Defibrillators. New York,
NY: Marcel Dekker, Inc. 1994.
Singer I, Ed. Implantable Cardioverter-Defibrillator. Armonk, NY: Futura Publishing Co.
1994.
●
Singer I, Barold SS, Camm AJ, Eds. Nonpharmacological Therapy of Arrhythmias forthe 21st Century: The State of the Art. Armonk, NY: Futura Publishing Co. 1998.
●
Stadler RW, Gunderson BD, Gillberg JM. An Adaptive Interval-Based Algorithm for
Withholding ICD Therapy During Sinus Tachycardia. Pace. 2003; 26:1189–1201.
12Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Notice
The Patient Information screen of the programmer software application is provided as an
informational tool for the end user. The user is responsible for accurate input of patient
information into the software. Medtronic makes no representation as to the accuracy or
completeness of the patient information that end users enter into the Patient Information
screen. Medtronic SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULT FROM THE
USE OF THE PATIENT INFORMATION SUPPLIED BY END USERS TO THE SOFTWARE.
Reference Manual13
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Part I
Quick overview
1 Quick reference
1.1 Physical characteristics, AT and DR devices
Table 1. Device physical characteristics
Volume
Mass63 g
H x W x D
Surface area of device can58 cm
Radiopaque IDPNR
Materials in contact with human tissuecTitanium, polyurethane, silicone rubber
BatteryLithium silver vanadium oxide hybrid
a
b
c
a
b
Volume with connector holes unplugged.
Grommets may protrude slightly beyond the can surface.
These materials have been successfully tested for the ability to avoid biological incompatibility. The device does
not produce an injurious temperature in the surrounding tissue during normal operation.
3
33 cm
62 mm x 51 mm x 13 mm
2
14Reference Manual
6
1
2
3
4
5
Medtronic ENTRUST
®
Figure 1. Connector and suture holes
D153ATG, D153DRG, D153VRC
1 IS-1 connector port, V
2 IS-1 connector port, A
3 DF-1 connector port, SVC (HVX)
4 DF-1 connector port, RV (HVB)
5 Device Active Can electrode, Can (HVA)
6 Suture holes
1.2 Physical characteristics, VR devices
Table 2. Device physical characteristics
Volume
Mass63 g
H x W x D
Surface area of device can58 cm
a
b
Radiopaque IDPNT
Materials in contact with human tissuecTitanium, polyurethane, silicone rubber
BatteryLithium silver vanadium oxide hybrid
a
Volume with connector holes unplugged.
b
Grommets may protrude slightly beyond the can surface.
c
These materials have been successfully tested for the ability to avoid biological incompatibility. The device does
not produce an injurious temperature in the surrounding tissue during normal operation.
4 Device Active Can electrode, Can (HVA)
5 Suture holes
1.3 Replacement indicators
Battery voltage and messages about replacement status appear on the programmer display
and on printed reports. Table 3 lists the Elective Replacement Indicator (ERI) and the End
of Life (EOL) conditions.
Table 3. Replacement indicators
Elective Replacement Indicator (ERI)≤ 2.61 V on three consecutive daily automatic
measurements
End of Life (EOL)3 months after ERI
ERI date – The Quick Look and Battery and Lead Measurements screens display the date
when the battery reached ERI.
EOL indication – If the programmer indicates that the device is at EOL, replace the device
immediately.
Post-ERI conditions – EOL device status is defined as three months following an ERI
indication assuming the following post-ERI conditions: 100% DDD pacing (VVI in VR
16Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
devices) at 60 bpm, 2.5 V, 0.4 ms; 500 Ω pacing load; and six full-energy charges. EOL
may be indicated before the end of three months if the device exceeds these conditions.
1.4 Longevity projections
Device longevity is affected by how certain features are programmed, such as
Pre-arrhythmia EGM. See Section 4.12, “Optimizing device longevity”, page 55.
1.4.1 AT and DR device longevity projections
The following longevity estimates are based on accelerated battery discharge data and
device modeling, as specified.
These models assume the default automatic capacitor formation setting. As a guideline,
each full energy charge decreases device longevity by approximately 28 days.
Table 4. Projected longevity in years with 0.4 ms pulse width and 60 bpm pacing rate
Maximum energy
charging fre-
Pacing
DDD, 0%Semi-annualOff7.37.37.37.3
DDD, 15%Semi-annualOff7.06.97.17.0
DDD, 50%Semi-annualOff6.46.16.86.6
AAI<=>DDD,
(MVP mode)
50% atrial,
5% ventricular
quency
QuarterlyOff6.36.36.36.3
QuarterlyOff6.16.06.26.1
QuarterlyOff5.75.46.05.8
Semi-annualOff6.96.77.17.0
QuarterlyOff6.05.96.26.1
a
Pre-arrhythmia
EGM storage
On7.07.07.07.0
On6.16.16.16.1
On6.86.76.96.8
On5.95.86.06.0
On6.25.96.66.4
On5.55.35.85.6
On6.76.56.96.8
On5.95.76.05.9
500 Ω pacing
impedance
b
2.5 V3.0 V2.5 V3.0 V
900 Ω pacing
impedance
Reference Manual17
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 4. Projected longevity in years with 0.4 ms pulse width and 60 bpm pacing rate
(continued)
500 Ω pacing
impedance
b
2.5 V3.0 V2.5 V3.0 V
Pacing
Maximum energy
charging frequency
a
Pre-arrhythmia
EGM storage
DDD, 100%Semi-annualOff5.75.36.46.0
On5.65.16.25.9
QuarterlyOff5.14.85.65.4
On5.04.65.55.2
a
Maximum energy charging frequency may include full-energy therapy shocks or capacitor formations.
b
The data provided for programming Pre-arrhythmia EGM on is based on a 6 month period (two 3 month follow-up
intervals) over the life of the device. Additional use of Pre-arrhythmia EGM reduces longevity by approximately
28% or 3.4 months per year.
900 Ω pacing
impedance
1.4.2 VR device longevity projections
The following longevity estimates are based on accelerated battery discharge data and
device modeling, as specified.
These models assume the default automatic capacitor formation setting. As a guideline,
each full energy charge decreases device longevity by approximately 40 days.
Table 5. Projected longevity in years with VVI pacing mode, 0.4 ms pulse width, and 60 bpm
pacing rate
500 Ω pacing
impedance
b
2.5 V3.0 V2.5 V3.0 V
% Pacing
Maximum energy
charging frequency
a
Pre-arrhythmia
EGM storage
0%Semi-annualOff9.89.89.89.8
On9.59.59.59.5
QuarterlyOff8.28.28.28.2
On7.97.97.97.9
15%Semi-annualOff9.69.59.79.6
On9.39.29.49.3
QuarterlyOff8.07.98.18.0
On7.87.77.87.8
50%Semi-annualOff9.08.79.49.2
On8.78.59.18.9
QuarterlyOff7.67.47.97.7
On7.47.27.67.5
900 Ω pacing
impedance
18Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 5. Projected longevity in years with VVI pacing mode, 0.4 ms pulse width, and 60 bpm
pacing rate (continued)
500 Ω pacing
impedance
b
2.5 V3.0 V2.5 V3.0 V
% Pacing
Maximum energy
charging frequency
a
Pre-arrhythmia
EGM storage
100%Semi-annualOff8.37.99.08.6
On8.17.68.78.4
QuarterlyOff7.16.87.67.3
On6.96.67.47.1
a
Maximum energy charging frequency may include full-energy therapy shocks or capacitor formations.
b
The data provided for programming Pre-arrhythmia EGM on is based on a 6 month period (two 3 month follow-up
intervals) over the life of the device. Additional use of Pre-arrhythmia EGM reduces longevity by approximately
34% or 4.1 months per year.
900 Ω pacing
impedance
1.5 Magnet application
When a magnet is placed near the device, the device responds as shown in Table 6. When
the magnet is removed, the device returns to its programmed operations.
Note: Before implant and for the first 6 hours after implant, the device will not sound audible
tones when a magnet is placed over the device.
Table 6. Effects of magnet application on the device
Pacing modeAs programmed
Pacing rate and intervalAs programmed
Tachyarrhythmia detectionSuspended
Patient Alert audible tones (20 s or
less)
With programmable alerts enabled:
●
Continuous tone (Test)
●
On/off intermittent tone (seek follow-up)
●
High/low dual tone (urgent follow-up)
a
b
c
With programmable alerts disabled:
●
No tone
●
High/low dual tone (urgent follow-up)
a
Rate Response adjustments are suspended while a Patient Alert tone sounds.
b
Detection resumes if telemetry between the device and the programmer is established and the application
software is running.
c
The Test tone does not sound if “VF Detection OFF, 3+ VF or 3+ FVT Rx Off” is the only alert enabled.
1.6 Typical charge times
The most recent capacitor charge time appears on the programmer display and on printed
reports. You can evaluate charge time using the Charge/Dump test (see Table 7).
Reference Manual19
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 7. Typical full energy charge times with fully formed capacitors
At Beginning of Life (BOL)6.7 s
At Elective Replacement (ERI)9.0 s (AT and DR devices)
9.7 s (VR devices)
1.7 High-voltage therapy energy
The stored energy of the device is derived from the peak capacitor voltage and is always
greater than the energy delivered by the device. Table 8 compares the programmed energy
levels delivered by the device to the energy levels stored in the capacitors before delivery.
Table 8. Delivered (programmed) and stored energy levels
Delivered/Programmed
30 J34 J6.7 s
28 J31 J6.3 s
26 J29 J5.8 s
24 J27 J5.4 s
22 J25 J4.9 s
20 J23 J4.5 s
18 J20 J4.0 s
16 J18 J3.6 s
15 J17 J3.4 s
14 J16 J3.1 s
13 J15 J2.9 s
12 J14 J2.7 s
11 J12 J2.5 s
10 J11 J2.2 s
9 J10 J2.0 s
8 J9.1 J1.8 s
7 J7.9 J1.6 s
6 J6.8 J1.3 s
5 J5.8 J1.1 s
4 J4.6 J0.9 s
3 J3.4 J0.7 s
2 J2.3 J0.4 s
1.8 J2.1 J0.4 s
1.6 J1.9 J0.4 s
1.4 J1.7 J0.3 s
1.2 J1.4 J0.3 s
a
Stored
a
a
Charge Time
b
20Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 8. Delivered (programmed) and stored energy levelsa (continued)
Delivered/Programmed
a
1.0 J1.2 J0.2 s
0.8 J1.0 J0.2 s
0.6 J0.7 J0.1 s
0.4 J0.5 J0.1 s
a
Delivered energy values are based on measurements at the connector block during high-voltage therapies into
a 75 Ω load. Stored energy values indicate the energy on the capacitor at the end of charging.
b
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth of a
second.
Flashback memory interval data
before the latest AT/AF episode
9.25 min12.5 min13.75 min
45 s45 s51 s
30 s30 s30 s
3.75 min——
1.0 min1.5 min—
1.25 min1.25 min1.4 min
2000 events
(includes both Aand V-events)
2000 events
(includes both Aand V-events)
2000 events
(includes both Aand V-events)
2000 events
(includes both Aand V-events)
2000 events
(V-events only)
—
Reference Manual21
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 10. VT/VF episode counters
Counter data typeAT and DR devicesVR devices
Counts of each VT/VF episode type
Counts of each SVT episode type
(VT/VF therapy withheld)
●
VF
●
FVT
●
VT
●
VT Monitor
●
VT-NS
●
Runs of PVCs
●
Single PVCs
●
Runs of VRS paces
●
Single VRS paces
●
AF/Afl
●
Sinus Tach
●
Other 1:1 SVTs
●
V. Stability
●
Onset
●
VF
●
FVT
●
VT
●
VT Monitor
●
VT-NS
●
Runs of PVCs
●
Single PVCs
●
Runs of VRS paces
●
Single VRS paces
●
V. Stability
●
Onset
●
Wavelet
Table 11. VT/VF therapy counters
Counter data typeAT, DR and VR devices
VT/VF therapy summary counters
VT/VF therapy efficacy counters
●
ATP-terminated episodes
●
Shock-terminated episodes
●
Total VT/VF shocks
●
Aborted charges
For VF Rx1–Rx6 and ATP during/before charging:
●
Delivered therapy counts
●
Successful therapy counts
For FVT Rx1–Rx6:
●
Delivered therapy counts
●
Successful therapy counts
●
Counts of episodes accelerated to VF
For VT Rx1–Rx6:
●
Delivered therapy counts
●
Successful therapy counts
●
Counts of episodes accelerated by 60 ms or to FVT or VF
Table 12. AT/AF episode counters
Counter data typeAT devicesDR devices
AT/AF summary data
●
Percent of time in AT/AF
●
Average time in AT/AF per
day
●
Percentage of AT/AF epi-
●
Percent of time in AT/AF
●
Average time in AT/AF
per day
sodes terminated by ATP
Average number per day of each
AT/AF episode type
●
Monitored AT/AF
●
Treated AT/AF
●
Non-sustained AT
●
Monitored AT/AF
●
Non-sustained AT
22Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 12. AT/AF episode counters (continued)
Counter data typeAT devicesDR devices
Percent of time in each kind of pacing
Number of AT/AF episodes, presented in different groupings
a
This counter includes any instance when the device identifies AT/AF Onset. Therefore, the total number of
●
Atrial pacing
●
Atrial intervention pacing
●
Grouped by duration
●
Grouped by start time
a
episodes in this counter may exceed the number of detected AT/AF episodes recorded by the device.
●
Atrial pacing
●
a
Grouped by duration
●
Grouped by start time
Table 13. AT/AF therapy counters
Counter data typeAT devices
Number of AT/AF episodes treated
and the percentage terminated,
presented in different groupings
Counts of different AT/AF therapy
types
●
Grouped by detection zone and therapy
●
Grouped by atrial cycle length
●
Delivered ATP sequences
●
Aborted ATP sequences
●
Delivered automatic atrial CV shocks
●
Automatic atrial CV shocks that failed to terminate the
episode
●
Delivered patient-activated shocks
●
Patient-activated shocks that failed to terminate the episode
a
a
Table 14. Battery and lead measurement data
AT devicesDR devicesVR devices
●
Battery voltage
●
Last capacitor formation
●
Last charge
●
Lead impedance:
– A. pacing
– RV pacing
– RV defib
– SVC defib (if used)
●
R-wave amplitude
●
P-wave amplitude
●
Last high-voltage therapy
●
Sensing integrity counter
●
Atrial Lead Position Check
●
Battery voltage
●
Last capacitor formation
●
Last charge
●
Lead impedance:
– A. pacing
– RV pacing
– RV defib
– SVC defib (if used)
●
R-wave amplitude
●
P-wave amplitude
●
Last high-voltage therapy
●
Sensing integrity counter
●
Battery voltage
●
Last capacitor formation
●
Last charge
●
Lead impedance:
– RV pacing
– RV defib
– SVC defib (if used)
●
R-wave amplitude
●
Last high-voltage therapy
●
Sensing integrity counter
results
Reference Manual23
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 15. Lead performance trend data
14 days of daily measurements, 80 weeks of weekly minimum and maximum measurements, highest value, lowest value, value at implant, and latest value.
AT and DR devicesVR devices
Lead impedance measurements:
●
A. Pacing
●
RV pacing
●
RV defibrillation
●
SVC defibrillation (if used)
Sensing amplitude measurements:
●
P-wave
●
R-wave
Lead impedance measurements:
●
RV pacing
●
RV defibrillation
●
SVC defibrillation (if used)
Sensing amplitude measurements:
●
R-wave
Table 16. Cardiac Compass trend data
Printed report showing up to 14 months of measurement trends and summary data.
AT and DR devicesVR devices
●
Annotations of interrogations, programming,
and remote sessions
●
VT and VF episodes per day
●
High-voltage therapies delivered per day
●
Ventricular rate during VT or VF
●
Episodes of non-sustained tachycardia per
day
●
Heart rate variability
●
Total daily time in AF or AT
●
Ventricular rate during AF or AT
●
Percent pacing per day
●
Patient activity
●
Average day and night ventricular heart rate
●
Annotations of interrogations, programming,
and remote sessions
●
VT and VF episodes per day
●
High-voltage therapies delivered per day
●
Ventricular rate during VT or VF
●
Episodes of non-sustained tachycardia per
day
●
Heart rate variability
●
Percent pacing per day
●
Patient activity
●
Average day and night ventricular heart rate
24Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 17. Rate Histograms report
Data typeAT devicesDR devicesVR devices
Graphs displaying the
percent of time in each
rate range for the listed
conditions:
●
Atrial pacing and
atrial sensing
●
Ventricular pacing
and ventricular
sensing
●
Ventricular pacing
a
●
Atrial pacing and
atrial sensing
●
Ventricular pacing
and ventricular
sensing
●
a
Ventricular pacing
and ventricular
sensing
and ventricular
sensing during
AT/AF
Percent of time for
each event type:
a
If more than 2% of atrial sensed events are identified as far-field R-waves, the general percentage range (either
“2% to 5%” or “> 5%”) is reported above the atrial rate histogram.
b
In AT and DR devices, if the programmed pacing mode during the reporting period was a dual chamber mode,
b
●
AS-VS events
●
AS-VP events
●
AP-VS events
●
AP-VP events
the report displays the AS-VS, AS-VP, AP-VS, and AP-VP event sequence data. If a single chamber mode was
programmed, the report displays the percent of time spent pacing and sensing. MVP modes (AAIR<=>DDDR
and AAI<=>DDD) are considered dual chamber modes for this purpose.
●
AS-VS events
●
AS-VP events
●
AP-VS events
●
AP-VP events
●
VP events
●
VS events
Table 18. Patient Alert event data
Log of events that triggered Patient Alert notifications. Each log entry includes the following information:
●
Date when the event first occurred (since the last interrogation)
●
Description of event that triggered the Patient Alert
●
Programmed threshold for the Patient Alert, if applicable
Reference Manual25
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
2 The EnTrust system
2.1 System overview
EnTrust Implantable Cardioverter Defibrillator (ICD) systems are implantable medical
devices that automatically detect and treat episodes of ventricular fibrillation (VF),
ventricular tachycardia (VT), fast ventricular tachycardia (FVT), and bradyarrhythmia. The
EnTrust AT device also detects and treats atrial tachyarrhythmia episodes.
Each EnTrust ICD system includes three major components: the implanted device, the
leads connecting the device to the patient’s heart, and the Medtronic programmer with
EnTrust application software installed.
2.1.1 Implanted device
The device senses the electrical activity of the patient’s heart using the sensing electrodes
of the implanted leads. It analyzes the heart rhythm based on selectable sensing and
detection parameters. If the device detects a ventricular tachyarrhythmia, it can deliver
defibrillation, cardioversion, or antitachycardia pacing therapy to the patient’s heart. If the
device identifies a bradyarrhythmia, it delivers bradycardia pacing therapy.
2.1.2 Leads
The device can be used with transvenous or epicardial defibrillation leads. The lead system
should consist of bipolar or paired unipolar1 leads for pacing and sensing and one or two
high-voltage cardioversion/defibrillation electrodes. The pacing and sensing electrodes
sense cardiac activity and deliver pacing stimuli. AT and DR devices require both atrial and
ventricular leads to be implanted. VR devices require only ventricular leads. The device
case may optionally be used as an Active Can high voltage electrode. You can enable or
disable this option using the programmable Active Can parameter.
2.1.3 Programmer and software
The Medtronic programmer and EnTrust application software allow you to perform the
following tasks:
●
configure the detection, therapy, and bradycardia features for your patient
●
perform electrophysiological studies and system tests
●
monitor, display, or print patient cardiac activity information
●
view patient and device diagnostic data
1
With an appropriate unipolar to bipolar adapter kit.
26Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
The EnTrust device and application software are compatible with the Medtronic CareLink
Model 2090 programmer with a Model 2067 or 2067L programming head.
2.1.4 Patient assistant
Patients with implanted AT devices can use the Model 2696 InCheck Patient Assistant to
perform the following functions:
●
Verify whether the implanted device has detected a suspected atrial tachyarrhythmia.
●
Initiate recording of cardiac event data in the device memory.
●
Request delivery of atrial cardioversion therapy (if the device is programmed to allow
patient-activated cardioversion).
Note: Patient-activated cardioversion is only delivered if the implanted device is
currently detecting an AT/AF episode.
2.1.5 Detecting ventricular tachyarrhythmias
The device monitors the cardiac rhythm for short ventricular intervals that may indicate the
presence of VF, VT, or FVT.
You can program the device to distinguish between true ventricular arrhythmias and rapidly
conducted supraventricular tachycardia (SVT) and to withhold therapy for SVT.
AT and DR devices also have the ability to detect double tachycardias (unrelated ventricular
arrhythmias occurring simultaneously with SVTs) so that therapy is not withheld for a
ventricular arrhythmia in the presence of an SVT.
2.1.6 Treating ventricular tachyarrhythmias
The device treats detected VF episodes by delivering a biphasic defibrillation shock. If the
VF episode persists, up to 5 more individually programmed defibrillation shocks can be
delivered.
You also have the option of delivering 1 sequence of ATP therapy before or during charging
for a VF therapy. This option can prevent delivery of painful shocks for episodes that are
detected as VF but can be terminated by pacing therapy.
The device treats detected VT episodes by delivering either a Ramp, Ramp+, or Burst
antitachycardia pacing therapy or a biphasic cardioversion shock synchronized to a
ventricular depolarization. If the VT episode persists, up to 5 more individually programmed
VT therapies can be delivered.
Reference Manual27
Medtronic ENTRUST
The device treats detected FVT episodes by delivering either a Ramp, Ramp+, or Burst
antitachycardia pacing therapy or a biphasic cardioversion shock synchronized to a
ventricular depolarization. If the FVT episode persists, up to 5 more individually
programmed FVT therapies can be delivered.
®
D153ATG, D153DRG, D153VRC
2.1.7 Detecting atrial tachyarrhythmias
If there is no ventricular episode in progress, the device applies the AT/AF detection
algorithm, which detects AT/AF episodes by examining the atrial rate and the relationship
between atrial and ventricular events. Both AT and DR devices can detect AT/AF episodes,
but only AT devices can respond to detected AT/AF episodes with programmed atrial
tachyarrhythmia therapies. AT devices also provide an additional detection zone for Fast
AT/AF episodes. This second zone allows the device to treat a second, faster atrial
tachyarrhythmia with a separately programmable set of therapies.
2.1.8 Treating atrial tachyarrhythmias
The device treats detected AT/AF episodes by delivering Burst+, Ramp or 50 Hz Burst
antitachycardia pacing therapy or by delivering an atrial cardioversion. Each sustained
AT/AF episode can be treated with up to 5 automatic therapies per detection zone: 3
antitachycardia pacing therapies and 2 atrial cardioversion therapies. Patient-activated
cardioversion is also available to treat AT/AF episodes.
2.1.9 Treating bradycardia
The device provides rate responsive pacing to treat bradycardia. An internal accelerometer
senses the patient’s physical activity, allowing the device to increase and decrease the
pacing rate in response to changes in the level of activity.
VR devices provide single chamber ventricular pacing modes. AT and DR devices provide
dual chamber pacing, single chamber pacing, and MVP (Managed Ventricular Pacing)
modes. The MVP modes switch between single chamber atrial pacing and dual chamber
pacing to promote intrinsic conduction and manage unnecessary right ventricular pacing.
28Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
2.1.10 Monitoring for real-time and stored data
The device and programmer provide real-time information on detection and therapy
parameters and status during a patient session. The device also provides accumulated data
on device operation, including stored electrograms, detected and treated tachyarrhythmia
episodes, bradycardia interventions, and the efficacy of therapy. The Cardiac Compass
report provides up to 14 months of clinically significant data, including arrhythmia episodes,
shocks delivered, physical activity, heart rate, and bradycardia pacing activities. The Rate
Histograms report shows the percent of time that cardiac events occurred at different heart
rates. In AT and DR devices, this report also shows the distribution of ventricular heart rates
during AT/AF episodes.
All of this information can be printed and retained in the patient’s file or saved in electronic
format on a floppy diskette.
2.1.11 Conducting electrophysiologic tests
You can use the system to conduct non-invasive electrophysiologic studies, including
manual delivery of therapies, to manage an induced or spontaneous tachyarrhythmia.
2.1.12 Alerting the patient to system events
You can use the programmable Patient Alert monitoring feature to notify the patient with
audible tones if certain conditions occur that are related to the leads, battery, charge time,
or therapies. The patient can then respond based on your prescribed instructions.
2.2 Indications and usage, AT devices
The device is indicated for use in ICD patients with atrial tachyarrhythmias, or who are at
significant risk of developing atrial tachyarrhythmias. The device is intended to provide
ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of
life-threatening ventricular arrhythmias.
In addition, the device is intended to provide pacing, cardioversion, and defibrillation for
treatment of patients with
The use of the device has not been demonstrated to decrease the morbidity related to
atrial tachyarrhythmias.
●
The effectiveness of high-frequency burst pacing (atrial 50 Hz Burst therapy) in
terminating device classified atrial tachycardia (AT) was found to be 17%, and in
terminating device classified atrial fibrillation (AF) was found to be 16.8%, in the VT/AT
patient population studied.
●
The effectiveness of high-frequency burst pacing (atrial 50 Hz Burst therapy) in
terminating device classified atrial tachycardia (AT) was found to be 11.7%, and in
terminating device classified atrial fibrillation (AF) was found to be 18.2%, in the AF-only
patient population studied.
2.3 Indications and usage, DR and VR devices
The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia
pacing and ventricular defibrillation for automated treatment of life-threatening ventricular
arrhythmias.
2.4 Contraindications, AT devices
The device is contraindicated for patients experiencing any of the following conditions:
●
tachyarrhythmias with transient or reversible causes including, but not limited to, the
following: acute myocardial infarction, digitalis intoxication, drowning, electric shock,
electrolyte imbalance, hypoxia, or sepsis
●
incessant ventricular tachycardia or ventricular fibrillation
●
primary disorder of chronic atrial tachyarrhythmia with no concomitant VT or VF
●
present implant of a unipolar implantable pulse generator
●
primary disorder of bradyarrhythmia
2.5 Contraindications, DR and VR devices
The device is contraindicated for patients experiencing any of the following conditions:
●
tachyarrhythmias with transient or reversible causes including, but not limited to, the
following: acute myocardial infarction, digitalis intoxication, drowning, electric shock,
electrolyte imbalance, hypoxia, or sepsis
●
incessant ventricular tachycardia or ventricular fibrillation
●
present implant of a unipolar implantable pulse generator
●
primary disorder of bradyarrhythmia or atrial arrhythmia
30Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
2.6 Patient screening
Other optional screening procedures may include exercise stress testing to determine the
patient’s maximum sinus rate and cardiac catheterization to determine if there is a need for
concomitant surgery or medical therapy.
Reference Manual31
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
3 Emergency therapy
3.1 Overview of emergency therapies
The device provides emergency defibrillation, cardioversion, fixed burst pacing, and VVI
pacing therapy. You can gain access to emergency therapies by pressing the on-screen
[Emergency] button, or by pressing the red mechanical button on the programmer display
panel. To return to other programming functions from an Emergency screen, select
[Exit Emergency].
Effect on system operation – The device suspends the automatic detection features when
emergency defibrillation, cardioversion, or fixed burst pacing therapies are delivered.
Detection is not suspended during emergency VVI pacing. Remove the programming head
or press [Resume] to enable detection again.
Aborting an emergency therapy – You can immediately terminate an emergency
defibrillation or emergency cardioversion therapy by selecting [ABORT]. To stop an
emergency fixed burst therapy, remove the programmer stylus from the [BURST Press and
Hold] button. To terminate emergency VVI pacing, reprogram the bradycardia pacing
parameters from the Parameters screen.
Mechanical Emergency buttons on the Model 9790C programmer – If you press the
red mechanical Emergency button on the programmer display panel, the programmer
displays the Emergency screen. The mechanical yellow-on-blue deliver button activates
the emergency therapy displayed on the programmer screen. This button functions only
when the Emergency screen is displayed.
Mechanical Emergency VVI button on the Medtronic CareLink Model 2090
programmer – If you press the red Emergency VVI button on the programmer display panel,
the device initiates Emergency VVI pacing and the programmer displays the Emergency
screen.
Temporary parameter values – Emergency tachyarrhythmia therapies use temporary
parameter values that do not change the programmed parameters of the device. After the
tachyarrhythmia therapy is complete, the device reverts to its programmed values.
Note: When you enable emergency VVI pacing, the programmed bradycardia pacing
parameters are changed to the emergency settings.
32Reference Manual
3
4
2
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
3.2 Delivering an emergency defibrillation therapy
The default emergency therapy is a full-energy defibrillation. When you select
[Emergency] and [DELIVER], the device charges and delivers a biphasic full-energy shock.
The programmer sets the emergency defibrillation energy to its maximum value each time
you select [Emergency] or select the [Defibrillation] option from an Emergency screen.
3.2.1 Parameters
Energy – Amount of energy delivered to the heart
by the therapy.
Pathwaya – Direction the electrical current flows
through the heart.
a
If Active Can is Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway.
0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20 … 30 J
B>AX (fixed)
3.2.2 How to deliver emergency defibrillation
1. Position the programming head over the device.
2. Select [Emergency].
Reference Manual33
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
3. Accept the defibrillation energy shown on the screen, or select a new Energy value.
4. Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly positioned,
and select [Retry] or [Cancel].
3.3 Delivering an emergency cardioversion therapy
When you initiate an emergency cardioversion therapy, the device charges its capacitors
to the selected energy and attempts to deliver therapy synchronized with a sensed
tachyarrhythmia event. If the cardioversion therapy cannot be synchronized, it is aborted.
See Section 7.7.5.5, “Synchronizing cardioversion after charging”, page 188.
3.3.1 Parameters
Energy – Amount of energy delivered to the heart by the
therapy.
Pathwaya – Direction the electrical current flows through
the heart.
a
If Active Can is Off, the HVA (Can) electrode is not used as part of the high-voltage delivery pathway.
0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20 …
30 J
B>AX (fixed)
34Reference Manual
234
5
Medtronic ENTRUST
®
3.3.2 How to deliver emergency cardioversion
D153ATG, D153DRG, D153VRC
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the cardioversion energy shown on the screen, or select a new Energy value.
5. Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly positioned,
and select [Retry] or [Cancel].
3.4 Delivering emergency fixed burst pacing
Emergency fixed burst pacing delivers maximum output pacing pulses to the ventricle at a
selectable interval. The therapy continues for as long as you keep the programmer stylus
on the [BURST Press and Hold] button.
Reference Manual35
2
534
Medtronic ENTRUST
3.4.1 Parameters
®
D153ATG, D153DRG, D153VRC
Interval – Time interval between pacing pulses delivered during the fixed burst therapy.
RV Amplitude – Voltage of the ventricular pacing
pulses delivered during the fixed burst therapy.
RV Pulse Width – Duration of the ventricular pacing
pulses delivered during the fixed burst therapy.
100; 110 … 350 … 600 ms
8 V (fixed)
1.5 ms (fixed)
3.4.2 How to deliver emergency fixed burst pacing
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval shown on the screen, or a new Interval value.
5. Select [BURST Press and Hold].
If delivery is not confirmed, the programmer displays an error window. Verify that the
programming head is properly positioned. Select [OK] from the window, and reselect
[BURST Press and Hold].
36Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
3.5 Enabling emergency VVI pacing
Emergency VVI pacing programs the device to deliver high-output ventricular pacing. You
can initiate emergency VVI pacing from the Emergency screen or by pressing the red
mechanical button on the Medtronic CareLink Model 2090 Programmer display panel. To
disable emergency VVI pacing, reprogram the bradycardia pacing parameters from the
Parameters screen.
3.5.1 Parameters
Pacing Mode – NBG Codea for the pacing mode provided during
during periods of inactivity.
RV Amplitude – Voltage of the ventricular pacing pulses delivered
during emergency VVI pacing.
RV Pulse Width – Duration of the ventricular pacing pulses deliv-
ered during emergency VVI pacing.
V. Blank Post VP – Time interval during which sensing is disabled
after a pacing pulse.
Rate Hysteresis – Enables tracking of intrinsic heart rate below
programmed Lower Rate to prevent pacing during extended periods of inactivity, such as when a patient is sleeping.
V. Rate Stabilization – Modifies the pacing rate to eliminate the
long pause that typically follows a premature ventricular contraction.
a
N–North American Society of Pacing and Electrophysiology (NASPE), B–British Pacing and Electrophysiology
Group (BPEG), G–Generic Pacemaker Code
VVI
70 bpm
6 V
1.5 ms
240 ms
Off
Off
Reference Manual37
3
4
2
Medtronic ENTRUST
®
3.5.2 How to enable emergency VVI pacing
D153ATG, D153DRG, D153VRC
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM] to change the pacing parameters to the emergency VVI settings.
If programming is not confirmed, verify that the programming head is properly
positioned, and select [Retry] or [Cancel].
38Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Part II
Device implant and patient follow-up procedures
4 Implanting the device
4.1 Overview
Proper surgical procedures and sterile techniques are the responsibility of the physician.
The following procedures are provided for information only. Each physician must apply the
information in these procedures according to professional medical training and experience.
The implant procedure includes the following steps:
●
Section 4.3, “Preparing for an implant”, page 40
●
Section 4.4, “Verify lead and connector compatibility”, page 41
●
Section 4.5, “Position the leads”, page 42
●
Section 4.6, “Test the lead system”, page 43
●
Section 4.7, “Connect the leads to the device”, page 44
●
Section 4.8, “Testing ventricular defibrillation operation and effectiveness”, page 47
●
Section 4.9, “Position and secure the device”, page 51
●
Section 4.10, “Completing the implant procedure”, page 53
For information about replacing a previously implanted device, see Section 4.11, “Replacing
a device”, page 54.
4.2 Considerations
Review the following information before implanting the device:
Electrical isolation during implant – Do not allow the patient to have contact with
grounded equipment that might produce electrical current leakage during implant. Electrical
current leakage may induce arrhythmias that may result in the patient’s death.
Reference Manual39
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
External defibrillation equipment – Keep external defibrillation equipment nearby for
immediate use whenever arrhythmias are possible or intentionally induced during device
testing, implant procedures, or post-implant testing.
Lead coils and Active Can electrodes – Lead coils and Active Can electrodes in contact
during a high-voltage therapy may cause electrical current to bypass the heart, possibly
damaging the device and leads. While the device is connected to the leads, verify that
therapeutic electrodes, stylets, or guide wires are not touching or connected by an
accessory low impedance conductive pathway. Move objects made from conductive
materials (for example, an implanted guide wire) well away from all electrodes before
delivering a high-voltage shock.
Use by date – Do not implant the device after the “Use by” date on the package label.
Battery longevity may be reduced.
4.3 Preparing for an implant
4.3.1 Equipment and sterile supplies for an implant
The following is the necessary equipment for an implant:
●
Medtronic CareLink Model 2090 programmer with a Model 2067 or 2067L programming
head
●
EnTrust Model 9987 software application
●
Model 2290 or 8090 Analyzer or equivalent pacing system analyzer
●
external defibrillator
The following are the necessary sterile supplies for an implant:
●
implantable device and lead system components
●
programming head sleeve
Note: If a sterilized programming head is used during an implant, a sterile programming
head sleeve is not necessary.
●
analyzer cables
●
lead introducers appropriate for the lead system
●
extra stylets of appropriate length and shape
4.3.2 How to set up the programmer and start the application
1. Set up the programmer as described in the instructions provided with the programmer.
2. Install the EnTrust Model 9987 software on the programmer if it is not already installed.
3. Place the programming head over the device, and start the application. Select the
device model, or select [Find Patient…].
Note: The programmer automatically interrogates the device when the application starts.
40Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4.3.3 How to prepare the device for implant
Before opening the sterile package, prepare the device for implant:
1. Interrogate the device. Print an initial interrogation report.
Note: If the programmer reports that an electrical reset occurred, do not implant the
device. Contact a Medtronic representative.
2. Check the initial interrogation report or the Quick Look screen to confirm that the battery
voltage is at least 3.0 V at room temperature.
If the device has recently delivered a high-voltage charge or if the device has been
exposed to low temperatures, the battery voltage will be temporarily lower and
capacitor charge time may increase. Allow the device to warm to room temperature
and check the battery voltage again. If an acceptable battery voltage cannot be
obtained, contact a Medtronic representative.
3. Set the internal clock of the device (see Section 10.2.3).
4. Perform a manual capacitor formation (see Section 11.7.2 for additional information).
a. Dump any charge on the capacitors.
b. Perform a test charge to full energy.
c. Retrieve the charge data.
d. Do not dump the stored charge. Allow the stored charge to dissipate for at least
10 min; the dissipation forms the capacitors.
e. If the reported charge time is unacceptable, contact a Medtronic representative.
5. Program the therapy and pacing parameters to values appropriate for the patient (see
Chapter 8 and Chapter 7). Ensure that tachyarrhythmia detection is not programmed
to On (see Section 6.1).
Note: Do not enable the Atrial Preference Pacing feature or a rate responsive pacing
mode before implanting the device. Doing so may result in a pacing rate that is faster
than expected.
4.4 Verify lead and connector compatibility
Select a compatible lead. Refer to the following table.
Table 19. Lead and connector compatibility
PortPrimary leadLead adaptor
RV, SVCDF-1
A, VIS-1a bipolar5866-24M for 5 mm paired unipolar
a
DF-1 refers to the international standard ISO 11318:2002. IS-1 refers to ISO 5841-3:2000.
Reference Manual41
a
6707 for 6.5 mm cardioversion/defibrillation lead
5866-24M for 5 mm bifurcated
5866-38M for IS-1 unipolar
5866-40M for Medtronic 3.2 mm low-profile
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4.5 Position the leads
Implant transvenous leads according to the supplied instructions unless suitable chronic
leads are already in place. Do not use any lead with this device without first verifying
connector compatibility. Transvenous or epicardial leads may be used. For AT and DR
devices, a bipolar atrial lead with closely spaced pacing and sensing electrodes is
recommended.
4.5.1 Using transvenous leads
Use standard transvenous implant techniques to position the ventricular lead tip in the right
ventricular apex. For DR and AT devices, use standard transvenous implant techniques to
position the atrial pacing lead tip high on the right atrial appendage.
Follow the general guidelines below for initial positioning of other transvenous leads (the
final positions are determined by defibrillation efficacy tests):
●
SVC (HVX) lead: Place the lead tip high in the innominate vein, approximately 5 cm
proximal to the right atrium (RA) and SVC junction.
●
SQ patch: Place the patch along the left mid-axillary, centered over the fourth-to-fifth
intercostal space.
Note: An SQ patch is not implanted transvenously, but it is used with transvenous lead
systems.
●
CS lead: Advance the lead tip to a position just under the left atrial appendage, if
possible.
If using a subclavian approach, position the lead using a more lateral approach to avoid
pinching the lead body between the clavicle and the first rib.
Warning: Pinching the lead can damage the lead conductor or insulation, which may cause
unwanted high-voltage therapies or result in the loss of sensing or pacing therapy.
4.5.2 Using epicardial leads
A variety of surgical approaches can be used to implant epicardial leads, including a limited
left thoracotomy or median sternotomy. A typical placement may use an anterior right
ventricular patch as the RV (HVB) and a posterolateral left ventricular patch as the SVC
(HVX).
Follow these general guidelines for positioning epicardial leads:
●
If unipolar epicardial pacing leads are used, position the electrodes about 1 to 2 cm
apart to reduce electromagnetic interference, and route the leads together with several
loose twists.
42Reference Manual
Medtronic ENTRUST
●
Place the patches so that they encompass the maximum amount of cardiac mass and
®
D153ATG, D153DRG, D153VRC
they have approximately equal amounts of mass between them.
●
Ensure that the patches do not overlap and that the electrode portions do not touch.
●
Avoid placing extra-pericardial patches over the phrenic nerve.
●
Suture the smooth face of each patch lead against the epicardium or pericardium in
locations that produce optimal defibrillation.
4.5.3 Surgical incisions
A single-incision submuscular or subcutaneous approach is recommended when the device
is implanted in the pectoral region. Make the implant pocket about 1.5 times the size of the
device.
Submuscular implant – An incision extending over the deltoid-pectoral groove typically
provides access to the cephalic and subclavian veins as well as to the implant pocket. Place
the device sufficiently medial to the humeral head to avoid interference with shoulder
motion.
Subcutaneous implant – A transverse incision typically permits isolation of the cephalic
vein. Place the device in a far medial position to keep the leads away from the axilla. Make
sure that the upper edge of the device remains inferior to the incision.
4.6 Test the lead system
Sensing and pacing tests include the following measurements:
●
P-wave amplitudes (AT and DR devices only)
●
R-wave amplitudes
●
slew rate
●
pacing threshold
●
pacing lead impedance
Medtronic recommends that you use a Model 2290 or 8090 Analyzer to perform sensing
and pacing measurements. If you use a Pacing System Analyzer (PSA), such as the Model
5311 or 5311B, perform both atrial and ventricular measurements using the ventricular
channel of the PSA.
Refer to the technical manual supplied with the PSA for details about performing sensing
and pacing measurements.
Reference Manual43
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4.6.1 Parameters
See Table 20 for information about acceptable measured sensing and pacing values at
implant.
≤ 1.5 V (atrial)≤ 3.0 V (atrial)
≤ 1.0 V (ventricular)≤ 3.0 V (ventricular)
a
The measured pacing lead impedance is a reflection of measuring equipment and lead technology. Refer to the
lead technical manual for acceptable impedance values.
b
Chronic leads are leads implanted for 30 days or more.
a
b
4.6.2 Considerations
Note: Do not measure the intracardiac EGM telemetered from the device to assess sensing.
When measuring sensing and pacing values, measure between the tip (cathode) and ring
or coil (anode) of each bipolar pacing/sensing lead.
For paired unipolar epicardial pacing leads, either electrode can be the cathode; use the
configuration that yields the lower pacing threshold.
4.7 Connect the leads to the device
Warning: Verify that the lead connections are secure. Loose lead connections may result
in inappropriate sensing and failure to deliver arrhythmia therapy.
Caution: Use only the wrench supplied with the device. The wrench is designed to prevent
damage to the device from overtightening a setscrew.
See the following figures for information about lead connections. Figure 3 shows the AT and
DR device ports for lead connections. Figure 4 shows the VR device ports for lead
connections.
44Reference Manual
1
2
3
4
5
2
3
4
1
Medtronic ENTRUST
Figure 3. AT and DR device ports for lead connections
®
D153ATG, D153DRG, D153VRC
1 IS-1 connector port, V
2 IS-1 connector port, A
3 DF-1 connector port, SVC (HVX)
Note: For easier lead insertion, insert the ventricular IS-1 connector leg first.
4 DF-1 connector port, RV (HVB)
5 Device can electrode, Can (HVA)
3 DF-1 connector port, RV (HVB)
4 Device can electrode, Can (HVA)
Reference Manual45
a
b
1
2
Medtronic ENTRUST
4.7.1 How to connect the lead to the device
®
D153ATG, D153DRG, D153VRC
1. Insert the wrench into a grommet on the connector port.
a. Check that the setscrew is retracted from the connector port. If the connector port
is obstructed, retract the setscrew to clear it. Do not disengage the setscrew from
the connector block.
b. Leave the wrench in the setscrew until the lead is secure. This allows a pathway
for venting trapped air when the lead is inserted.
2. Push the lead or plug into the connector port until the lead pin is clearly visible in the
pin viewing area. No sealant is required, but sterile water may be used as a lubricant.
3. Tighten the setscrew by turning the wrench to the right until the wrench clicks.
46Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4. Repeat these steps for each connector port.
5. Gently pull on the lead to confirm the connection.
4.8 Testing ventricular defibrillation operation and
effectiveness
Warning: Ensure that an external defibrillator is charged for a rescue shock before
delivering a test shock.
Demonstrate reliable defibrillation effectiveness with the implanted lead system by using
your preferred method to establish that a 10 J (minimum) safety margin exists.
4.8.1 High-voltage implant values
See Table 21 for information about measured high-voltage therapy values recommended
at implant.
Table 21. High-voltage therapy values recommended at implant
4.8.2 How to prepare for defibrillation threshold testing
1. Place the programming head over the device and start a patient session. Interrogate
the device if this has not already been completed.
2. Observe the Marker Channel annotations to verify that the device is sensing properly.
3. Conduct a manual Lead Impedance Test to verify defibrillation lead connections.
Perform this test with the device in the surgical pocket. Keep the surgical pocket very
moist. If the lead impedance is out of range, perform one or more of the following tasks:
●
Recheck the lead connections and lead electrode placement.
●
Repeat the lead impedance measurements.
●
Inspect the EGM for abnormalities.
●
Measure the defibrillation impedance with a manual test shock.
4. Program the device to detect VF with an adequate safety margin. An adequate safety
margin is typically 1.2 mV sensitivity.
Reference Manual47
87561341149321211
10
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4.8.3 How to perform defibrillation threshold testing using T-Shock
1. Select Tests > EP Study.
2. Select T-Shock from the Inductions/Therapies box.
3. Select [Resume at DELIVER] to resume arrhythmia detection after induction delivery.
4. Select [Adjust Permanent…].
48Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
5. Set the VF detection, RV Sensitivity, and VF therapy parameters. VF Enable should
be On, and the Energy parameter for VF Therapies 2–6 should be set to the maximum
value.
6. Select [Program].
7. Select [Close].
8. Select Enable.
9. Select [DELIVER T-Shock]. If necessary, you can abort an induction or therapy in
progress by selecting [ABORT].
10. Observe the live rhythm monitor for proper post-shock sensing.
11. Use the [Adjust Permanent…] button to program a new energy level, if desired.
12. Wait until the on-screen timer reaches 5 minutes, then repeat Step 8 through
Step 12 as needed.
13. Select the Params icon and disable VF, FVT, and VT detection before closing the
pocket.
14. Interrogate the device, and print an episode summary report.
Reference Manual49
86131142457912311
10
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4.8.4 How to perform defibrillation threshold testing using 50 Hz Burst
1. Select Tests > EP Study.
2. Select 50 Hz Burst from the Inductions/Therapies box.
3. Select [RV] as the chamber for induction/therapy.
4. Select [Resume at BURST] to resume arrhythmia detection after induction delivery.
5. Select [Adjust Permanent…].
50Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6. Set the VF detection, RV Sensitivity, and VF therapy parameters. VF Enable should
be On, and the Energy parameter for VF Therapies 2–6 should be set to the maximum
value.
7. Select [Program].
8. Select [Close].
9. Press and hold [50 Hz BURST Press and Hold]. If necessary, you can abort an
induction or therapy in progress by selecting [ABORT].
10. Observe the live rhythm monitor for proper post-shock sensing.
11. Use the [Adjust Permanent…] button to program a new energy level, if desired.
12. Wait until the on-screen timer reaches 5 minutes, then repeat Step 9 through
Step 12 as needed
13. Select the Params icon and disable VF, FVT, and VT detection before closing the
pocket.
14. Interrogate the device, and print an episode summary report.
4.9 Position and secure the device
Cautions:
●
If no SVC electrode is implanted, the pin plug provided with the device must be secured
in the SVC port to avoid damage to the device.
●
Program tachyarrhythmia detection to Off or Monitor before closing the pocket to avoid
inappropriate detection and shock while closing the pocket.
Note: Implant the device within 5 cm (2 in) of the surface of the skin to optimize post-implant
ambulatory monitoring. The side of the device engraved with the Medtronic logo should face
toward the skin to optimize the Patient Alert feature.
Reference Manual51
4
2
Medtronic ENTRUST
®
4.9.1 How to position and secure the device
D153ATG, D153DRG, D153VRC
1. Verify that each lead connector pin or plug is fully inserted into the connector port and
that all setscrews are tight.
2. To prevent twisting of the lead body, rotate the device to loosely wrap the excess lead
length. Do not kink the lead body.
3. Place the device and leads into the surgical pocket.
4. Suture the device securely within the pocket. Use non-absorbable sutures. Secure the
device to minimize post-implant rotation and migration. Use a surgical needle to
penetrate the suture holes on the device (indicated by arrows in the drawing).
5. Suture the pocket incision closed.
52Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4.10 Completing the implant procedure
Warning: For AT and DR devices, do not enable the Other 1:1 SVTs PR Logic detection
criterion until the atrial lead has matured, approximately one month after implant. This
criterion may inappropriately withhold therapy if atrial sensing is compromised by an
unstable or dislodged atrial lead.
After implanting the device, obtain an x-ray of the patient to verify device and lead
placement. To complete programming the device, select parameters that are appropriate
for the patient.
Note: The high-voltage capacitors should be formed (or conditioned) periodically to
maintain quick charging for high-voltage therapy. You can use the Automatic Capacitor
Formation feature to ensure that they are formed regularly. See Section 5.3, “Optimizing
charge time”, page 58 for details.
4.10.1 How to complete programming the device
1. Enable tachyarrhythmia detection and the desired tachyarrhythmia therapies.
2. Perform a final VF induction, and allow the implanted system to detect and treat the
arrhythmia.
3. Program defibrillation and sensitivity parameters to values appropriate for the patient
after defibrillation threshold testing is complete.
4. Monitor the patient after the implant, and take x-rays as soon as possible to document
and assess the location of the leads.
5. Program patient information (see Section 10.12.1).
6. Configure the Patient Alert feature (see Section 10.4.3).
7. Set up data collection parameters (see Section 10.2.3).
Before the patient is discharged from the hospital, assess the performance of the implanted
device and leads.
4.10.2 How to assess the performance of the device and leads
1. If any tachyarrhythmia therapies are enabled while the patient is in the hospital,
interrogate the device after any spontaneous episodes to evaluate the detection and
therapy parameter settings.
2. If the patient has not experienced spontaneous episodes, you may induce the clinical
tachyarrhythmias using the non-invasive EP Study features to further assess the
performance of the system. See Chapter 12, “Conducting electrophysiologic studies”,
page 404.
Reference Manual53
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
3. Recheck pacing and sensing values, and adjust if necessary.
4. Interrogate the device, and print a Final Report to document the postoperative
programmed device status.
4.11 Replacing a device
If you are replacing a previously implanted device, disable detection before explanting.
When implanting the device with a chronic lead system, perform the following evaluations
to ensure appropriate detection and therapy:
●
Check the integrity of the chronic high-voltage leads with a test shock, chest x-ray, and
inspection.
●
Inspect the chronic lead connector pins for signs of pitting or corrosion.
●
Perform chronic pacing and sensing measurements.
●
Measure lead impedances.
●
Test defibrillation efficacy.
●
Confirm adequate sensing during VF.
●
Ensure proper fit of the lead connectors in the device connector block.
Notes:
●
To meet the implant requirements, it may be necessary to reposition or replace the
chronic leads or to add a third high-voltage electrode.
●
Any unused leads that remain implanted must be capped.
4.11.1 How to explant and replace a device
1. Program tachyarrhythmia detection to Off or Monitor to avoid potential inappropriate
shocks to the patient or implanter while handling the device (see Chapter 6).
2. Program the device to a non-rate responsive mode to avoid potential rate increases
while handling the device (see Section 8.1.4).
3. For AT devices, disable Atrial Preference Pacing (see Section 8.14.4).
4. Dissect the lead and the device free from the surgical pocket. Do not nick or breach
the lead insulation.
5. Use a wrench to loosen the setscrews in the connector port.
6. Gently pull the lead out of the connector port.
7. Evaluate the condition of the lead. Replace the lead if the electrical integrity is not
acceptable or if the lead connector pin is pitted or corroded. Return the explanted lead
to Medtronic for analysis and disposal.
8. Connect the lead to the replacement device.
Note: A lead adaptor may be needed to connect the lead to the replacement device.
Contact a Medtronic representative for questions about lead adaptor compatibility.
54Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
9. Evaluate sensing, pacing, and defibrillation efficacy. Use the replacement device.
10. After confirming acceptable electrical measurements, suture the pocket incision
closed.
11. Return the explanted device to Medtronic for analysis and disposal.
4.12 Optimizing device longevity
The following factors may affect the longevity of the device:
●
a greater number of high-voltage charges, either for therapy or for capacitor formation
●
higher programmed cardioversion or defibrillation therapy energies
●
a greater number of bradycardia paced events, for example, an increase in the pacing
rate
●
use of Atrial Preference Pacing (AT devices only), which increases the number of
bradycardia paced events
●
higher programmed pacing amplitude or pulse width
●
pacing outputs or high-voltage therapy energy increased to compensate for a decrease
in pacing and high-voltage lead impedance
●
use of the Pre-arrhythmia EGM storage or Holter telemetry features
For device longevity estimates, see Section 1.4.
4.12.1 Considerations
Shock energy – If the defibrillation threshold of the patient allows for an appropriate safety
margin (at least 10 J, after the acute implant period), consider programming the first
ventricular shock energy below the maximum value. Always program all subsequent shocks
to the full available energy.
Note: If VF Detection is enabled, consider programming the Energy parameter for the first
VF therapy to at least 20 J if you enable ATP During Charging. The charge times for therapies
with energies less than 20 J may be too brief for the device to detect when ATP During
Charging has terminated an arrhythmia.
Pacing outputs – If the patient’s pacing threshold allows for an appropriate safety margin
(at least a factor of 2 after the acute implant period), consider decreasing the pacing outputs.
Always consider the patient’s access to regular follow-up care in selecting a safety margin
for chronic pacing.
Pacing mode – If the patient’s intrinsic rhythm allows for appropriate rate support, you can
decrease the pacing burden by programming the Mode, Rate Response, and AV Interval
parameters to promote intrinsic activation or conduction.
Reference Manual55
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Pre-arrhythmia EGM storage – When Pre-arrhythmia EGM storage is enabled, the device
collects up to 10 s of EGM information before the onset or detection of the tachyarrhythmia.
However, using Pre-arrhythmia EGM storage reduces longevity by approximately 28% or
by 3.4 months per year in AT and DR devices2, and by approximately 34% or 4.1 months
per year in VR devices3. In a patient who uniformly repeats the same onset mechanisms,
the greatest clinical benefit of Pre-arrhythmia EGM storage is achieved after a few episodes
are captured.
To maximize the effectiveness of the Pre-arrhythmia EGM feature and optimize device
longevity, consider these programming options:
●
Enable Pre-arrhythmia EGM to capture possible changes in the onset mechanism
following significant clinical adjustments, for example, device implant, medication
changes, and surgical procedures.
●
Disable Pre-arrhythmia EGM once you have successfully captured the information of
interest.
Note: When Pre-arrhythmia EGM is disabled, the device starts storing EGM information
after the third tachyarrhythmia event occurs. However, the device still records up to 20 s of
information before the onset or detection of the tachyarrhythmia, including interval
measurements and Marker Channel annotations. In addition, the most recent
tachyarrhythmia episodes also provide Flashback interval data.
2
Based on device modeling with 50% atrial pacing, 5% ventricular pacing.
3
Based on device modeling with 15% ventricular pacing.
56Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
5 Conducting a patient follow-up session
5.1 Patient follow-up guidelines
During the first few months after receiving a new device, the patient may require close
monitoring. Schedule office visits at least every 3 months to monitor the condition of the
device and leads and to verify that the device is configured appropriately for the patient.
The Quick Look screen, which is displayed when you start the application, provides a good
beginning for the follow-up. The Quick Look screen allows you to perform these functions:
●
Verify that the device is functioning correctly.
●
Review the clinical performance and device trends.
●
Print appropriate reports to compare the results to the patient’s history and to retain for
future reference.
Note: To print an Initial Interrogation Report, you may need to change Initial Report
preferences and restart the session. See Section 9.12.4, “How to set Initial Reports
preferences”, page 317.
Note: The Checklist feature provides a standard list of tasks to perform at a complete
follow-up visit. You can also customize your own checklists. See Section 9.3, “Streamlining
follow-up and implant sessions with Checklist”, page 282.
5.2 Verifying the status of the implanted system
To verify that the device and leads are functioning correctly, review the following information
from the Quick Look screen, and perform follow-up tests as indicated:
●
Review the displayed battery voltage for comparison to the Elective Replacement
Indicator (ERI) value (see Section 1.3). If the displayed battery voltage is at or below
the displayed ERI value or if the ERI indicator is displayed, schedule an appointment to
replace the device.
Note: Battery voltage may be low if high-voltage charging has occurred within 24 hours.
●
If the programmer displays the EOL indicator for low battery voltage or excessive charge
time, the device should be replaced immediately.
●
Review the last full energy charge. For information about adjusting the capacitor
formation interval, see Section 5.3, “Optimizing charge time”, page 58.
●
Review the lead impedance values for inappropriate values or significant changes since
the last follow-up. See Section 11.5, “Measuring lead impedance”, page 396.
Reference Manual57
Medtronic ENTRUST
●
Perform P-wave and R-wave amplitude tests for comparison to previous P-wave and
®
D153ATG, D153DRG, D153VRC
R-wave amplitude measurements. See Section 11.6, “Performing a Sensing Test”,
page 398.
●
To review trends in sensing and impedance measurements, select the [>>] button from
the lead status area of the Quick Look screen. The programmer displays a detailed
history of automatic sensing and impedance measurements. See Section 10.11,
“Viewing Lead Performance Trends data”, page 377.
5.3 Optimizing charge time
The high-voltage capacitors should be formed (or conditioned) periodically to maintain quick
charging for high-voltage therapy. You can use the Automatic Capacitor Formation feature
to ensure that they are formed regularly.
The device can be programmed to automatically adjust the capacitor formation interval for
interactive management of charge time and device longevity.
Note: The device capacitors have not been fully formed since manufacture. Capacitors
should be manually formed before programming the Automatic Capacitor Formation Interval
to reduce the device charge time. See Section 11.7.2, “How to perform a Charge/Dump
test”, page 402.
See Section 5.3.5, “Details about managing charge time”, page 61.
5.3.1 Parameters
Minimum Auto Cap Formation IntervalAuto ; 1; 2 … 6 months
5.3.2 Considerations
Formation interval and longevity – A shorter formation interval provides faster charge
times by optimizing the efficiency of the capacitors. However, each capacitor formation
includes a full energy charge, reducing the longevity of the device.
Assess the patient’s requirements for faster therapy delivery relative to the effect on device
longevity. Each full energy charge decreases the longevity of AT and DR devices by
approximately 28 days and decreases the longevity of VR devices by approximately 40
days.
Programming a new formation interval – When you program a new automatic formation
interval, always confirm that the charge time is adequate at present. Either perform a manual
capacitor formation or evaluate a recent full energy charge time recorded in the Battery and
Lead Measurements display.
58Reference Manual
134
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Capacitor formation and Patient Alert – You can use the Patient Alert monitoring feature
to receive prompt notice if a long charge time has occurred.
5.3.3 How to evaluate charging performance
1. Perform a Charge/Dump test, and review the charge time (see Section 11.7.2, “How
to perform a Charge/Dump test”, page 402).
2. Allow the charge to dissipate for 10 min.
3. Select [DUMP Capacitors].
4. Perform another Charge Time test, and review the second charge time.
●
If the second charge time is clinically acceptable, consider reducing the Automatic
Capacitor Formation Interval (see Section 5.3.4, “How to program the Automatic
Capacitor Formation Interval”, page 60).
●
If the second charge time is not clinically acceptable, contact your Medtronic
representative.
Reference Manual59
12534
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
5.3.4 How to program the Automatic Capacitor Formation Interval
1. Select the Params icon.
2. Select the Therapies… field for VF, FVT, or VT.
3. Select Auto Cap Formation….
4. Select the Minimum Auto Cap Formation Interval.
5. Return to the Parameters screen and select [Program].
60Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
5.3.5 Details about managing charge time
The device provides the fastest charging and most prompt therapy delivery just following a
capacitor formation. During the time between formations, the capacitors gradually lose their
efficiency. This results in longer charge times until the next formation.
The charge time also gradually increases over the life of the device as the battery is
depleted. This occurs independently of capacitor formation.
Capacitor formation consists of 2 steps: First, the capacitors are charged to their full energy.
Then the charge is allowed to dissipate for at least 10 min.
The device records the time and date when a full-energy charge dissipates for 10 min without
interruption, and reports it on the programmer as a capacitor formation.
5.3.5.1 Smart Auto Cap
Therapeutic charging can also contribute to the efficiency of the capacitors. This occurs
because the capacitors are partially conditioned each time they are charged (for example,
during a full energy defibrillation). In these situations, there is less need for the next
scheduled capacitor formation. Automatic capacitor formation is not enabled until the device
is implanted.
More frequent formations decrease the device longevity, with little benefit to the charge
time. To optimize the effectiveness of the automatic capacitor formations, the device
postpones the schedule when full energy charging has occurred:
●
Whenever the capacitor formation interval is reprogrammed, the device resets the
automatic capacitor formation interval clock.
●
After a manual capacitor formation, the device resets the automatic capacitor formation
interval clock.
●
After an incidental capacitor formation (a full energy charge that dissipates for 10 min),
the device resets the automatic capacitor formation interval clock.
●
After a full energy charge is delivered or dumped, the device extends the automatic
capacitor formation interval clock by up to 2 months. The total of these extensions will
not exceed the programmed automatic capacitor formation interval.
Parameter set to “Auto” – When the capacitor formation interval parameter is
programmed to Auto, the device maintains a 6-month schedule for capacitor formation until
the battery is close to end of life.
If a charging period for therapy or capacitor formation exceeds 16 s, the device changes to
a 1-month schedule for capacitor formation. If a second charging period exceeds 16 s, the
device sets the EOL status message, indicating that the device should be replaced
immediately.
Reference Manual61
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
The rules for postponing the scheduled automatic formation after a full energy charge
remain in effect if the device changes to a 1-month schedule.
5.4 Verifying accurate detection and appropriate therapy
To verify that the device is providing effective tachyarrhythmia detection and therapy, review
the following information from the Quick Look screen, and investigate as indicated:
●
Review Quick Look Observations that relate to patient history and device operation. To
display more detailed information about an observation (when available), select the
observation and then select the [>>] button.
●
Review any Patient Alerts listed in the Observations of the Quick Look screen. For the
most detailed information about Patient Alerts, select Patient Alert Events from the Data
icon.
●
Check stored episode records for appropriate sensing and detection of arrhythmias
(see Section 10.5, “Viewing Arrhythmia Episode data”, page 342).
●
Check stored SVT episode records for appropriate identification of SVTs.
5.4.1 Considerations
Review the following information before verifying detection and therapy.
Caution: Do not program the device to decrease oversensing without ensuring that
appropriate sensing is maintained (see Section 6.2, “Setting up sensing”, page 70).
Flashback memory – In addition to the episode records and stored electrograms, use
Flashback memory and interval plots to investigate the accuracy and specificity of atrial and
ventricular detection.
Episode misidentification – If the episode records indicate that false detections have
occurred, the Sensing Integrity Counter may help in determining the prevalence of
oversensing. For more information, see Section 10.10.2.2, “Sensing integrity counter”,
page 377.
If the device is oversensing, consider these options:
●
Assess the lead integrity and lead connections.
●
Assess the sensitivity threshold.
62Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
If the episode records reveal that a stable monomorphic VT has been identified and treated
as VF, consider these options to improve the detection accuracy, which may also improve
the therapy accuracy:
●
Review the Interval Plot for the episode, and adjust the VF Interval if necessary. Use
caution when reprogramming the VF Interval because changes to this value can
adversely affect VF detection.
●
Program the ATP parameter for VF therapies to ATP During Charging (see
Section 7.5.4).
●
Enable FVT via VF detection (see Section 6.6.4).
If the SVT episode records include episodes of true VT, review the SVT episode records to
identify the SVT detection criterion that withheld detection. Adjust the SVT detection criteria
parameters as necessary.
5.5 Verifying effective bradycardia pacing
To verify that the device is sensing and pacing appropriately, review the following
information from the Quick Look screen, and investigate as indicated:
●
Confirm that the patient is receiving adequate cardiac support for daily living activities.
●
Review the pacing conduction history for comparison to the patient history. A sharp
increase in the paced beats percentage may indicate a need for investigation and
analysis.
●
Review the Cardiac Compass report for comparison to patient history (see Section 10.7,
“Using Cardiac Compass to view long-term clinical trends”, page 356).
●
Conduct pacing threshold tests (see Section 11.3, “Measuring pacing thresholds”,
page 388) to verify that the programmed pacing outputs provide a sufficient safety
margin.
5.5.1 Considerations
Review the following information for AT and DR devices before verifying bradycardia
pacing.
Ventricular Pacing – If the ventricle is predominantly paced and the patient exhibits
adequate ventricular response, consider these options:
●
Program the device to an MVP mode (AAIR<=>DDDR or AAI<=>DDD). These modes
are designed to manage unnecessary right ventricular pacing.
●
Decrease the Lower Rate.
●
Increase the Sensed AV interval.
●
Increase the Paced AV interval.
Reference Manual63
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Atrial Pacing – If the conduction history shows a predominance of atrial pacing despite a
healthy sinus response, consider these options to decrease the atrial pacing burden:
●
Decrease the Lower Rate.
●
Decrease the Rate Response or increase the Activity Threshold.
●
Decrease Paced AV and enable Rate Adaptive AV.
Conduction history – The reported percentages in the conduction history may not add up
to 100 because of rounding.
Note: Rate histograms may also be used to assess pacing and sensing history. The pacing
and sensing history on the Quick Look screen may not match the data in the histograms
because of premature atrial contractions (PACs) and A:V dissociation.
64Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Part III
Configuring the device for the patient
6 Detecting tachyarrhythmias
6.1 Detection overview
EnTrust devices provide several different atrial and ventricular tachyarrhythmia detection
capabilities.
●
AT devices provide AT/AF detection, AT/AF monitoring, VF/FVT/VT detection, and VT
monitoring.
●
DR devices provide AT/AF monitoring, VF/FVT/VT detection, and VT monitoring.
●
VR devices provide VF/FVT/VT detection and VT monitoring.
The device detects atrial tachyarrhythmias (AT/AF) by examining the atrial rate and the
relationship between atrial and ventricular events. If the atrial rate is fast enough and if
enough evidence of an atrial tachyarrhythmia is accumulated, the device detects an AT/AF
episode.
AT/AF detection can be programmed to Monitor or On. If AT/AF detection is programmed
to Monitor, the device collects data about the incidence and duration of atrial arrhythmias,
but it does not treat episodes with atrial therapies. If AT/AF detection is programmed to On,
detected AT/AF episodes can be treated with atrial therapies (see Chapter 7, “Treating
tachyarrhythmia episodes”, page 133). The AT/AF Detection Interval, whether
programmed to Monitor or On, also triggers mode-switching in the device.
The device detects ventricular tachyarrhythmias (VF, VT, and FVT) by comparing the time
intervals between sensed ventricular events to a set of programmed detection intervals. If
enough intervals are shorter than the programmed intervals, the device detects a
tachyarrhythmia.
Reference Manual65
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
In addition to VF, VT, and FVT detection zones, the device provides a VT monitoring zone,
which functions as an independent VT zone. When the device detects a VT episode in the
VT monitoring zone, it records the episode but does not deliver VT therapy.
To discriminate between rapidly conducted SVTs (for example, sinus tachycardia or atrial
fibrillation) and ventricular tachyarrhythmias, the device provides several detection
enhancements such as the following detection criteria:
●
PR Logic with enhanced Sinus Tach criterion in AT and DR devices
●
Wavelet Dynamic Discrimination in VR devices
●
Stability
●
Onset
Figure 5 shows how the ventricular tachyarrhythmia detection features interact in AT and
DR devices during initial detection. Figure 6 shows how the ventricular tachyarrhythmia
detection features interact in VR devices during initial detection. During redetection of
ventricular tachyarrhythmias, devices do not apply Onset, PR Logic, or Wavelet detection
criteria.
Note: Detection functions can be disabled by programming the VF, FVT, VT, and VT Monitor
detection parameters to Off. For an example, see Section 6.4.4. AT/AF detection can only
be programmed to On or Monitor.
66Reference Manual
Ventricular Event
No
No
No
No
Yes
Yes
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Update counts and pattern information
Is the interval in the
VF, FVT, or VT detection zone?
Has High Rate Timeout suspended detection
enhancements?
Does Stability reset the VT event count?
(VT and FVT via VT detection only)
Is the Onset criterion allowing VT event counting?
(VT and FVT via VT detection only)
Has a tachyarrhythmia event count reached
an Initial Beats to Detect criterion?
Are one or more PR Logic criteria enabled?
Is the median ventricular interval less than
the SVT V. Limit?
Is there a double tachycardia in progress?
Are one or more PR Logic criteria withholding
detection?
No/ Suspended
by High Rate
Timeout
Yes
Yes
Tachy
Episode
detected
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Figure 5. How ventricular tachyarrhythmia detection features in AT and DR devices interact
during initial detection
Reference Manual67
Ventricular Event
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
Yes
Update counts information
Is the interval in the VF, FVT, or VT detection zone?
Has High Rate Timeout suspended detection enhancements?
Does Stability reset the VT event count? (VT and FVT via VT
detection only)
No
Yes
Is the Onset criterion allowing VT event counting? (VT and FVT via
VT detection only)
Has a tachyarrhythmia event count reached an Initial Beats to Detect
criterion?
Is Wavelet enabled?
No / Suspended by High
Rate Timeout
Is the median ventricular interval less than the SVT V. Limit?
Is Wavelet withholding detection?
Tachy
Episode
detected
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Figure 6. How ventricular tachyarrhythmia detection features in VR devices interact during
initial detection
68Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.1.1 Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops classifying and counting
tachyarrhythmia intervals. Sensing and bradycardia pacing remain active, and the
programmed detection settings are not modified.
Note: Suspending detection during a VT Monitor episode terminates the data storage. If
the ventricular episode continues following suspension, a new episode is detected.
Note: If detection is suspended during an AT/AF episode, a new episode is not recorded
when detection resumes. Redetection of the episode begins when detection resumes.
Detection is suspended during the following occurrences:
●
The device senses the presence of a strong magnet. The programmer head contains
a magnet which suspends detection. Once telemetry between the device and
programmer is established, detection resumes.
●
The device is performing any of the manual system tests. Detection automatically
resumes once the test is complete.
●
The device is performing an EP Studies induction. You can choose to have the device
automatically resume detection after delivering the induction.
●
The device is delivering a manual therapy or emergency therapy. You can resume
detection by selecting the [Resume] button or by removing the programming head from
the device.
●
You have selected the on-screen [Suspend] button. You can resume detection by
selecting the [Resume] button or by removing the programming head from the device.
●
The device is performing automatic daily lead impedance measurements. Detection
resumes when the measurements are complete.
Note: The automatic daily lead impedance measurements are not performed during
detected VT/VF episodes. These measurements are, however, performed during VT
Monitor episodes.
●
The device is delivering an automatic tachyarrhythmia therapy (including capacitor
charging for defibrillation and cardioversion). However, the device does continue to
confirm detected ventricular episodes during charging. Detection resumes when the
therapy is complete.
Note: The device suspends VT detection (and Combined Count detection, see
Section 6.8, “Detecting tachyarrhythmia episodes with combined count”, page 101) for
17 events following a defibrillation therapy delivered in response to a detected VF.
4
4
If the defibrillation therapy is delivered as a result of either a High Rate Timeout Therapy operation or
Progressive Episode Therapies operation, VT detection is not suspended (see Section 6.15.4, “Details about
High Rate Timeout”, page 132).
Reference Manual69
Medtronic ENTRUST
●
The device is charging for Automatic Capacitor Formation. Detection resumes when
®
D153ATG, D153DRG, D153VRC
charging is complete.
6.2 Setting up sensing
The device provides bipolar sensing in both the atrium and ventricle using the sensing
electrodes of the implanted atrial and ventricular leads. You can adjust the sensitivity to
intracardiac signals using independent atrial and ventricular sensitivity settings. These
settings define the minimum electrical amplitude recognized by the device as an atrial or
ventricular sensed event.
Dual chamber atrial and ventricular sensing is available only in AT and DR devices. Only
ventricular sensing is available in VR devices.
Proper sensing is essential for the safe and effective use of the device. To provide
appropriate sensing, the device uses the following features:
●
short (30 ms) cross-chamber blanking after paced events
●
auto-adjusting atrial and ventricular sensing thresholds
●
programmable options for PVAB
See Section 6.2.4, “Details about sensing”, page 72.
6.2.1 Parameters
RV Sensitivity – Minimum amplitude of electrical signal that regis-
ters as a sensed ventricular event.
A. Sensitivity – Minimum amplitude of electrical signal that regis-
ters as a sensed atrial event.
0.15; 0.3 ; 0.45; 0.6; 0.9;
1.2 mV
0.15; 0.3 ; 0.45; 0.6; 0.9;
1.2; 1.5; 2.1 mV
6.2.2 Considerations
Review the following information before programming sensing parameters.
Sensitivity thresholds – The programmed sensitivity thresholds apply to all features
related to sensing, including detection, bradycardia pacing, and the Sensing Test.
Bradycardia pacing and sensing – A combination of high pacing pulse width or high
amplitude with a low sensitivity threshold may cause inappropriate sensing across
chambers or in the same chamber. Programming a lower pulse width, lower amplitude,
longer pace blanking, or a higher sensitivity threshold may eliminate this inappropriate
sensing.
Dual chamber sensing and bradycardia pacing modes (AT and DR devices) – The
device senses in both the atrium and the ventricle at all times, except when the programmed
70Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
bradycardia pacing mode is DOO, VOO, or AOO. When the pacing mode is programmed
to DOO or VOO, there is no sensing in the ventricle. When the pacing mode is programmed
to DOO or AOO, there is no sensing in the atrium.
Recommended ventricular sensitivity threshold – A ventricular sensitivity threshold of
0.3 mV is recommended to maximize the probability of detecting VF and to limit the
possibility of oversensing and cross-chamber sensing.
High ventricular sensitivity threshold – Setting RV Sensitivity to a value greater than
0.6 mV is not recommended except for testing purposes. Doing this may cause
undersensing, which may cause any of the following situations:
●
delayed or aborted cardioversion therapy
●
delayed defibrillation therapy (when VF confirmation is active)
●
asynchronous pacing
●
underdetection of tachyarrhythmias
Testing sensitivity after reprogramming – If you change the ventricular sensitivity
threshold, test for proper sensing. If appropriate, test for proper detection by inducing VF
and allowing the device to automatically detect and treat the arrhythmia.
Sensing during VF – Always verify that the device senses properly during VF. If the device
is not sensing or detecting properly, disable detection and therapies, and evaluate the
system (making sure to monitor the patient for life-threatening arrhythmias until you enable
detection and therapies again). You may need to reposition or replace the ventricular
sensing lead to achieve proper sensing.
Low sensitivity threshold – If you set a sensitivity parameter to its most sensitive value
of 0.15 mV, the device will be more susceptible to EMI, cross-chamber sensing, and
oversensing.
Recommended atrial sensitivity threshold – An atrial sensitivity threshold of 0.3 mV is
recommended to optimize the effectiveness of atrial detection and pacing operations while
limiting the possibility of oversensing and cross-chamber sensing.
High atrial sensitivity threshold – If you set the A. Sensitivity value too high, the device
may not provide reliable sensing of P-waves during AT/AF episodes and sinus rhythm.
Atrial pacing and ventricular sensing – If you program the device to an atrial pacing
mode, make sure that it does not sense atrial pacing pulses as ventricular events.
Atrial lead selection – Atrial leads with minimal (10 mm) tip-to-ring spacing may reduce
far-field R-wave sensing.
Repositioning the atrial lead – If reprogramming the atrial sensitivity threshold does not
provide reliable atrial sensing during AT/AF episodes and sinus rhythm, you may need to
reposition or replace the atrial sensing lead.
Reference Manual71
1
2
3
Medtronic ENTRUST
®
6.2.3 How to program sensitivity
D153ATG, D153DRG, D153VRC
1. Select the Params icon.
2. Select the desired A. Sensitivity and RV Sensitivity parameters.
3. Select [PROGRAM].
6.2.4 Details about sensing
6.2.4.1 Blanking periods
Blanking periods inhibit sensing after paced and sensed events. These periods help prevent
the sensing of device pacing, cardioversion and defibrillation pulses, post-pacing
depolarization, T-waves, and multiple sensing of the same event. The blanking periods after
paced events are longer than or equal to those after sensed events to avoid sensing the
atrial and ventricular depolarizations.
Table 22 shows the duration of the fixed blanking periods. See Section 8.1, “Providing basic
pacing therapy”, page 198, for more information about programmable pace blanking
periods.
72Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Table 22. Duration of fixed blanking periods
Cross-chamber blanking after atrial or ventricular pacing pulse30 ms
Atrial and ventricular blanking after delivered cardioversion or defibrillation
therapy
520 ms
The device does not sense electrical signals during blanking periods, except during the
post-ventricular atrial blanking (PVAB) period. See Section 6.2.4.3, “Post-ventricular atrial
blanking and far-field R-waves”, page 75, for details about PVAB.
6.2.4.2 Auto-adjusting sensitivity thresholds
The device automatically adjusts sensitivity thresholds after certain paced and sensed
events to help reduce oversensing from T-waves, cross-chamber events, and pacing.
In Figure 7, sensitivity thresholds are adjusted after different types of events when the
nominal PVAB method is programmed.
Note: The way the device adjusts the atrial sensitivity threshold depends upon the
programmed method of PVAB. See Section 6.2.4.4 for details about the different PVAB
methods.
Reference Manual73
A
P
V
S
A
S
V
S
A
S
V
P
Sensitivity Threshold
Rectified and
Filtered A. EGM
Rectified and
Filtered RV EGM
Marker Channel
12
3
4
5
Medtronic ENTRUST
®
Figure 7. Auto-adjusting sensitivity thresholds
D153ATG, D153DRG, D153VRC
1 After an atrial sensed event, the atrial sensitivity threshold increases to 75% of the EGM peak
2 After a ventricular sensed event, the ventricular sensitivity threshold increases to 75% of the
3 After an atrial paced event, the device does not adjust the atrial sensitivity threshold. The
4 After a ventricular paced event, the atrial sensitivity threshold increases to 4x the programmed
5 After the ventricular pace blanking period is finished, the ventricular threshold increases to 4.5x
5
If the programmed ventricular sensitivity threshold exceeds 0.3 mV, the threshold is adjusted to 1.25x the
programmed value.
6
If the programmed atrial sensitivity threshold exceeds 1.2 mV, the threshold is adjusted to 1.25x the
programmed value.
74Reference Manual
(maximum: 8x the programmed value).
EGM peak (maximum: 10x the programmed value).
ventricular sensitivity threshold increases to 0.45 mV.
value6 (maximum: 2.0 mV).
the programmed value (maximum: 1.8 mV).
5
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.2.4.3 Post-ventricular atrial blanking and far-field R-waves
Nominally, atrial sensing is active during the post-ventricular atrial blanking (PVAB) interval.
When the device senses an atrial event in the PVAB interval, the event is ignored by
bradycardia pacing features. However, this event is detected by arrhythmia detection
features. See Section 6.3, “Detecting atrial tachyarrhythmias”, page 78, for details about
far-field R-wave sensing.
Note: Atrial events in the PVAB interval are not used by NCAP, PVC Response, Atrial Rate
Stabilization, Rate Adaptive AV, and PMT Intervention features.7 See Chapter 8 for details.
The PVAB interval can be programmed to 3 different methods of operation. The 3
programmable PVAB methods are Partial (the nominal method), Partial+, and Absolute.
The programmed method of PVAB affects how the atrial auto-adjusting sensitivity threshold
operates.
6.2.4.4 Post-ventricular atrial blanking methods
The auto-adjusting sensitivity threshold operates in the same way if either the Partial method
(the nominal method) or the Absolute method is programmed (see Figure 7). However, the
auto-adjusting sensitivity threshold operates in a different way if the Partial+ method is
programmed (see Figure 8). When the Partial+ method is programmed, the atrial sensitivity
threshold is increased (less sensitive) following ventricular events to provide
amplitude-based discrimination between far-field R-waves and intrinsic atrial events.
The following are details about the 3 PVAB methods:
Partial PVAB (nominal) – When the Partial PVAB method is programmed, atrial sensed
events in the PVAB interval are ignored by bradycardia pacing features but are used by
arrhythmia detection features.
See Section 8.2.6.4, “Post Ventricular Atrial Blanking interval (dual chamber)”, page 210,
for details about the Partial PVAB method.
Absolute PVAB – When the Absolute PVAB method is programmed, atrial sensed events
in the PVAB interval are not used by either arrhythmia detection features or bradycardia
pacing features. This method is recommended only for addressing complications not
addressed by the other PVAB methods.
See Section 8.2.6.4, “Post Ventricular Atrial Blanking interval (dual chamber)”, page 210,
for details about programming the Absolute PVAB method.
Partial+ PVAB – When the Partial+ PVAB method is programmed, atrial sensed events in
the PVAB interval are ignored by bradycardia pacing features but are used by arrhythmia
detection features. The atrial sensitivity threshold is increased during the PVAB interval
7
PVAB does not affect Mode Switch unless Absolute is the programmed PVAB method.
Reference Manual75
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
following a ventricular paced or sensed event for a period of time. This period of time, called
the desensitization period, reduces the chances of sensing far-field R-waves. Extending
the PVAB interval may affect intrinsic and far-field R-wave sensing.
Following a ventricular sensed or paced event, the length of the desensitization period is
similar to the length of the programmed PVAB interval. The length of the desensitization
period is 40, 60, 80, or 100 ms. The length of the desensitization period is determined by
the longest of these 4 lengths that does not exceed the length of the PVAB interval. The
desensitization period following a ventricular paced or sensed event reverts to the atrial
sensing threshold.
See Section 8.2.6.4, “Post Ventricular Atrial Blanking interval (dual chamber)”, page 210,
for details about the Partial+ PVAB method.
In Figure 8, sensitivity thresholds are adjusted after different types of events when the
Partial+ PVAB method is programmed.
76Reference Manual
A
S
A
P
V
S
V
S
12
3
Sensitivity
Threshold
Rectified and
Filtered A. EGM
Marker Channel
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Figure 8. Auto-adjusting sensitivity thresholds with Partial+ PVAB
1 After an atrial sensed event, the atrial sensitivity threshold increases to 75% of the EGM peak
2 After a ventricular event, the atrial sensitivity threshold is adjusted to the level to which it was
3 After an atrial paced event, the device adjusts the atrial sensitivity threshold to the minimum
Table 23. Minimum adjusted atrial threshold
If the programmed atrial sensitivity isThen the minimum adjusted threshold is
0.15 mV0.45 mV
0.3 mV0.6 mV
0.45 mV0.9 mV
Reference Manual77
(maximum: 8x the programmed value; minimum: see Table 23).
adjusted for the previous atrial event. If no atrial event occurred during the most recent V-V
interval, the atrial sensitivity is adjusted according to the minimum adjusted atrial threshold (see
Table 23).
If the programmed atrial sensitivity isThen the minimum adjusted threshold is
0.6 mV1.2 mV
0.9–1.5 mV1.8 mV
2.1 mV2.1 mV
See “Partial+ PVAB”, page 211, for details about programming the Partial+ PVAB method.
6.2.4.5 Refractory periods
During a refractory period, the device senses normally but classifies sensed events as
refractory and limits its response to these events. Pacing refractory periods prevent
inappropriately sensed signals, such as far-field R-waves (ventricular events sensed in the
atrium) or electrical noise from triggering certain pacing timing intervals.
Synchronization refractory periods help prevent the device from delivering cardioversion
and defibrillation therapies at inappropriate times. See Section 7.5.5.7, “Synchronizing
subsequent defibrillation therapies”, page 172 and Section 7.7.5.5, “Synchronizing
cardioversion after charging”, page 188.
Note: Refractory periods do not affect tachyarrhythmia detection.
6.3 Detecting atrial tachyarrhythmias
The device detects atrial tachyarrhythmias (AT/AF) by examining the atrial rate and the
relationship between atrial and ventricular events. If the atrial rate is fast enough and if
enough evidence of an atrial tachyarrhythmia is accumulated, the device detects an AT/AF
episode.
AT/AF detection is available in AT and DR devices. In DR devices, AT/AF detection can
only be programmed to Monitor. If AT/AF detection is programmed to Monitor, the device
collects data about the incidence and duration of atrial arrhythmias, but it does not treat
episodes with atrial therapies.
If AT/AF Detection is programmed to On, detected AT/AF episodes can be treated with atrial
therapies (see Chapter 7, “Treating tachyarrhythmia episodes”, page 133). The AT/AF
Detection Interval, whether programmed to Monitor or On, also triggers mode-switching in
the device.
78Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.3.1 Parameters
AT/AF DetectionOn; Monitor (AT devices)
Monitor (DR devices)
Number of detection zones1 ; 2 (AT devices)
AT/AF Interval (Rate)150; 160 … 350 … 450 ms
Fast AT/AF A. Interval (Rate)150; 160 … 200 … 250 ms (AT devices)
6.3.2 Considerations
Review the following information before programming AT/AF Detection.
Two-zone AT/AF Detection – If the patient exhibits two distinct atrial tachyarrhythmias,
you can program the device to detect each separately. This allows you to deliver a separate
therapy set for each tachyarrhythmia. Set the number of detection zones to 2, and program
the AT/AF A. Interval (Rate) and Fast AT/AF A. Interval (Rate) parameters to values
appropriate for each arrhythmia.
Atrial therapies and AT/AF Detection – If all atrial therapies are programmed Off and you
change the AT/AF Detection parameter value from Monitor to On, the programmer
automatically sets the first 2 AT/AF therapies to nominal or previously programmed settings.
6.3.3 Restrictions
Review the following information before programming AT/AF detection.
VF detection backup during AT/AF – To ensure VF detection backup during AT/AF
episodes, AT/AF Detection cannot be On unless VF Detection is also On.
Asynchronous pacing mode – Atrial detection must be programmed to Monitor when the
programmed pacing mode is DOO, VOO, or AOO.
Reference Manual79
142
3
a
b
5
Medtronic ENTRUST
®
6.3.4 How to program AT/AF Detection
D153ATG, D153DRG, D153VRC
1. Select the Params icon.
2. Select a value for AT/AF Detection.
3. Select a value for the AT/AF Interval (Rate).
80Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
4. To program Fast AT/AF Detection (only if AT/AF Detection is programmed to On),
select the Therapies… field for AT/AF to open the AT/AF Detection and Therapies
window.
a. Set Zones to 2.
b. Select a value for Fast AT/AF A. Interval (Rate).
5. Return to the Parameters screen and select [PROGRAM].
6.3.5 Details about AT/AF detection
Atrial tachyarrhythmia (AT/AF) detection has 4 phases:
●
AT/AF onset
●
initial AT/AF detection
●
sustained AT/AF detection
●
AT/AF episode termination
The device begins storing AT/AF data when the AT/AF onset criteria are met. After the
AT/AF onset criteria are met, the rhythm is monitored for initial AT/AF detection criteria and
then sustained AT/AF detection criteria. Once the rhythm is sustained, it is monitored until
it meets AT/AF termination criteria if AT/AF Detection is programmed to Monitor. If AT/AF
Detection is programmed to On and the device delivers an atrial therapy, the rhythm is
monitored for redetection and termination criteria.
The device uses several detection criteria throughout the phases of AT/AF detection. For
more information about AT/AF detection criteria, see Section 6.3.5.2, “AT/AF detection
criteria”, page 83.
Note: AT/AF detection can occur only if the device has not detected a ventricular
tachyarrhythmia. If an AT/AF episode is in progress and VF, VT, or FVT detection occurs,
AT/AF detection is suspended, and AT/AF therapies are disabled until the ventricular
episode terminates.
6.3.5.1 AT/AF detection phases
The 4 phases of AT/AF detection are discussed in detail as follows.
AT/AF onset – AT/AF onset occurs when the median atrial interval is less than the AT/AF
detection interval and the AT/AF evidence counter has counted at least 3 ventricular events
in which the A:V pattern shows evidence of an atrial tachyarrhythmia. The device begins
storing episode data after AT/AF onset occurs.
Reference Manual81
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Notes:
●
AT/AF onset data is used to calculate the percentage of time that the patient was
experiencing AT/AF.
●
If Mode Switch is enabled, the device starts mode switch operations when AT/AF onset
is detected. See Section 8.8, “Mode Switch”, page 240.
Initial AT/AF detection – Initial AT/AF detection occurs when the median atrial interval is
less than the AT/AF detection interval and the AT/AF evidence counter has counted at least
32 ventricular events in which the A:V pattern shows evidence of an atrial tachyarrhythmia.
If the device is programmed for two-zone AT/AF detection, the device identifies the episode
in the episode log as a Fast AT/AF episode if the median atrial interval is less than the
programmed Fast AT/AF detection interval.
Once initial AT/AF detection occurs, the AT/AF episode continues until episode termination.
Sustained AT/AF detection – Sustained AT/AF detection begins after initial detection.
During sustained detection, the device monitors the cardiac rhythm for changes or episode
termination. Sustained detection continues until the AT/AF episode termination criteria are
met.
When AT/AF Detection is programmed to On, the device can respond to detected episodes
with tachyarrhythmia therapies. For more information about the conditions required to
deliver atrial therapies, see Section 7.1, “Controlling atrial therapy sequencing”,
page 133. If AT/AF Detection is programmed to Monitor, the device delivers no atrial
therapy, but monitors the episode until the AT/AF episode ends.
If the device delivers an atrial therapy, the AT/AF evidence counter is reset to zero and
redetection begins. The AT/AF episode is redetected if the median atrial interval is less than
the AT/AF detection interval and the AT/AF evidence counter reaches 32.
If the device detects a ventricular tachyarrhythmia episode during an AT/AF episode, it
suspends AT/AF detection until the ventricular episode terminates. When AT/AF detection
resumes, the AT/AF evidence counter is reset to zero and must reach 32 before the atrial
episode is redetected.
AT/AF episode termination – Once the device detects an atrial tachyarrhythmia, the
episode is considered ongoing until episode termination. The device classifies an AT/AF
episode as terminated when one of the following conditions occurs:
●
The device identifies sinus rhythm using the sinus rhythm criterion.
●
The rhythm remains unclassified for 3 min because the median atrial interval is greater
than the AT/AF detection interval or the AT/AF evidence counter is less than 27.
If AT/AF episode termination is detected, the AT/AF evidence counter is reset to zero.
82Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.3.5.2 AT/AF detection criteria
The device applies several criteria during AT/AF detection to help identify the presence of
an atrial tachyarrhythmia. These criteria include the median atrial interval, the AT/AF
evidence counter, the far-field R-wave criterion, and the sinus rhythm criterion.
Median atrial interval – The device continually updates the median atrial interval. This
interval is calculated by finding the median of the 12 most recent atrial intervals. The last
12 atrial intervals are sorted in numerical order, and the median atrial interval is the larger
of the middle 2 values in the set. The median atrial interval must be less than the
programmed AT/AF detection interval for AT/AF detection to occur.
AT/AF evidence counter – The AT/AF evidence counter accumulates evidence of an atrial
arrhythmia based on the number and location of atrial events during ventricular intervals.
On each ventricular event, the device updates the counter as follows:
●
If A:V pattern information indicates the presence of an atrial arrhythmia (that is, greater
than 1:1 conduction in the absence of far-field R-wave sensing), the device increments
the counter by one.
●
If the counter was incremented on the previous ventricular interval because of A:V
pattern information, the device increments the counter by one.
●
If the AT/AF episode terminates or the device delivers an atrial therapy, the counter
resets to zero.
●
If none of the previous conditions is true, the device subtracts one from the counter.
Far-field R-wave criterion – If there are 2 atrial events in a ventricular interval, the device
analyzes A:V pattern information to determine if one of the atrial events is actually a far-field
R-wave. The device identifies a sensed far-field R-wave if it detects both of the following:
●
a short-long pattern of A-A intervals
●
a short AV interval (< 60 ms) or a short VA interval (< 160 ms)
Sinus rhythm criterion – The Sinus rhythm criterion identifies normal sinus rhythm (or a
paced rhythm) if 5 consecutive beats exhibit the A:V pattern of sinus rhythm. If this criterion
is satisfied after initial AT/AF detection, the device identifies episode termination, and the
AT/AF evidence counter resets to zero.
Note: The device also applies the far-field R-wave criterion to identify normal sinus rhythm
in the presence of far-field R-wave sensing.
Reference Manual83
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.4 Detecting VF episodes
The device detects VF episodes by examining the cardiac rhythm for short ventricular
intervals. If a predetermined number of intervals occur that are short enough to be
considered VF events, the device detects VF and delivers the first programmed VF therapy.
After therapy delivery, the device continues to evaluate the ventricular rhythm to determine
if the episode is ongoing.
6.4.1 Parameters
VF Detection – Turns VF Detection on or off.On ; OFF
VF Initial Beats to Detect – Number of beats to detect: number
of VF events the device must count to detect a VF episode.
VF Beats to Redetect – Number of beats to redetect: number of
VF events the device must count to redetect a continuing VF after
a therapy.
VF V. Interval (Rate) – V-V intervals shorter than this value are
counted as VF events.
Review the following information before programming VF detection parameters.
VF Interval minimum setting – To ensure proper VF Detection, do not program the VF
Interval less than 300 ms.
VF Interval maximum setting – Programming the VF Interval to a value greater than
350 ms may cause inappropriate detection of rapidly conducted atrial fibrillation as VF or
FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter,
which is more sensitive than the consecutive VT event counter.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia
interval, you should program the VF, FVT, and VT Intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Beats
to Redetect lower than the VF and VT Initial Beats to Detect.
Enabling VF Detection – When VF Detection enabled, the device initiates these functions:
●
enables Automatic Capacitor Formation
●
starts recording Cardiac Compass data
●
starts recording lead performance trends (starting at 2:15 AM, by the device clock)
●
clears all bradycardia pacing counters
84Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
VF detection and PR Logic criteria – You can program the device to distinguish between
SVT and VF Detection by enabling the PR Logic detection criteria. Note that the SVT Limit
must be programmed shorter than the VF Interval in order for the PR Logic criteria to affect
VF detection. See Section 6.10, “Enhancing detection with PR Logic criteria”, page 107.
Double tachycardia detection – When any PR Logic detection criterion is enabled, the
device also enables double tachycardia detection (VF, VT, or FVT in the presence of an
SVT). See Section 6.14, “Detecting double tachycardias”, page 129.
6.4.3 Restrictions
Review the following information before programming VF detection parameters.
AT/AF detection – If AT/AF Detection is On, the VF Initial Beats to Detect must be
programmed to a value less than or equal to the AT/AF evidence counter threshold, which
is 32. The VF Initial Beats to Detect must be programmed to a value less than or equal to
24/32 when AT/AF Detection is On. See Section 6.3, “Detecting atrial tachyarrhythmias”,
page 78.
Tachyarrhythmia detection and bradycardia pacing – To ensure reliable ventricular
tachyarrhythmia detection, the programmer regulates the values available for bradycardia
pacing and tachyarrhythmia detection.
Paced AV and temporary Lower Rate interval automatic adjustment – When VF or
VT detection is enabled, the device automatically adjusts the temporary Lower Rate interval
and PAV interval when needed. This adjustment maintains a ventricular event to atrial pace
interval that is at least 30 ms greater than the ventricular detection interval.
Asynchronous pacing mode – All ventricular detection must be programmed Off and
AT/AF Detection must be programmed to Monitor when the programmed pacing mode is
DOO, VOO, or AOO.
VF detection backup – To ensure VF detection backup during VT and FVT episodes, VT
and FVT detection cannot be enabled unless VF detection is also enabled.
Reference Manual85
3
241
Medtronic ENTRUST
®
6.4.4 How to program VF Detection
D153ATG, D153DRG, D153VRC
1. Select the Params icon.
2. Select Detection (V.)… from the Parameters screen.
3. Select the desired values for enabling or disabling VF Detection, VF Initial Beats to
Detect, VF Beats to Redect, and VF V. Interval (Rate).
4. Return to the Parameters screen and select [PROGRAM].
86Reference Manual
A
S
A
S
V SV SV
S
V
S
V SV
S
V
S
A
R
A
R
A
R
A
R
A
R
1 2 3 4 5 66 7 8 9 10 11 12 13 1415 1617 18
F SF SF SF SF SF SF SF SF SF
S
F SF SF
S
F SF SF
S
F
S
F
D
200 ms
VF event count
ECG
Marker Channel
VF interval
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.4.5 Details about VF detection
The device detects VF by counting the number of VF events, which are V-V intervals shorter
than the programmed VF V. Interval (Rate). On each event, the device counts the number
of recent VF events. The number of recent events examined is called the VF Detection
window. The size of the VF Detection window is the second number in the programmed VF
Initial Beats to Detect (for example, 24 events if the VF Initial Beats to Detect is 18/24).
The threshold for detecting VF is the first number in the programmed VF Initial Beats to
Detect (for example, 18 events if the VF Initial Beats to Detect is 18/24). This threshold is
always 75% of the VF Detection window. That is, if 75% of the events in the VF Detection
window are VF events, the device detects a VF episode (see Figure 9).
After the device detects VF, it delivers the first programmed VF therapy. Following the
therapy, if the number of VF events reaches the programmed VF Beats to Redetect, the
device redetects VF and delivers the next programmed VF therapy.
Note: The device can also detect VF episodes using the combined count detection criterion
(see Section 6.8, “Detecting tachyarrhythmia episodes with combined count”, page 101).
Figure 9. Device detects VF episode
1 VF starts, and the device begins counting VF events (intervals shorter than the programmed VF
2 A ventricular interval occurs outside the VF Detection zone. The VF event count is not
3 The VF event count reaches the programmed VF Initial Beats to Detect value of 18 events out
Reference Manual87
Interval).
incremented.
of 24, and the device detects VF.
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.5 Detecting VT episodes
The device detects VT episodes by examining the cardiac rhythm for short ventricular
intervals. If enough intervals occur that are short enough to be considered VT events (but
are not VF or FVT events), the device detects VT and delivers the first programmed VT
therapy. After therapy delivery, the device continues to evaluate the ventricular rhythm to
determine if the episode is ongoing.
You can program the device to detect and record VT episodes without treating them by
using the VT Monitor zone. This zone is independent from other detection zones. It allows
you to collect data without delivering therapy or affecting VF detection. See Section 6.7,
“Monitoring VT episodes”, page 96 for details about the VT Monitor zone.
See Section 6.5.5, “Details about VT detection”, page 91.
6.5.1 Parameters
VT Detection – Turns VT Detection on or off.On; OFF
VT Initial Beats to Detect – Number of beats to detect:
number of VT events the device must count to detect a
VT episode.
VT Beats to Redetect – Number of beats to redetect:
number of VT events the device must count to redetect a
continuing VT after a therapy.
VT V. Interval (Rate) – V-V intervals shorter than this
value are counted as VT events.
12; 16 … 52; 76; 100
4; 8; 12 … 52
280; 290 … 400 … 650 ms
6.5.2 Considerations
Review the following information before programming VT detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia
interval, program the VF, FVT, and VT V. Intervals (Rate)at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Beats
to Redetect lower than the VF and VT Initial Beats to Detect.
VT detection, AF/Afl, and Sinus Tach in AT and DR devices – When you program
VT Detection to On in AT and DR devices, the AF/Afl and Sinus Tach parameters are also
automatically set to On. See Section 6.10, “Enhancing detection with PR Logic criteria”,
page 107 for details about AF/Afl and Sinus Tach parameters.
VT detection and Wavelet in VR devices – When you program VT Detection to On in VR
devices, the Wavelet parameters are also automatically programmed to Monitor. See
88Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Section 6.11, “Enhancing detection with Wavelet”, page 114 for details about Wavelet
parameters.
VT detection and combined count detection – When VT Detection is On, the device
applies the combined count detection criterion to help speed detection of rhythms that
fluctuate between detection zones. combined count detection is disabled if VT Detection is
Off. See Section 6.8, “Detecting tachyarrhythmia episodes with combined count”,
page 101.
VT detection and rapidly conducted SVTs – You can program the device to distinguish
between SVT and VT detection by enabling the PR Logic (AT and DR devices only), Wavelet
(VR devices only), Onset, or Stability detection criteria. See Section 6.10, “Enhancing
detection with PR Logic criteria”, page 107, Section 6.11, “Enhancing detection with
Wavelet”, page 114; Section 6.12, “Enhancing VT detection with the Onset criterion”,
page 122; and Section 6.13, “Enhancing VT detection with the Stability criterion”,
page 126 for details.
Double tachycardia detection – When any PR Logic detection criterion is enabled, the
device also enables double tachycardia detection (VF, VT, or FVT in the presence of an
SVT; see page Section 6.14, “Detecting double tachycardias”, page 129).
6.5.3 Restrictions
Review the following information before programming VT detection parameters.
AT/AF detection – If AT/AF Detection is On, the VT Initial Beats to Detect must be
programmed to a value less than the AT/AF evidence counter threshold, which is 32. The
VT Initial Beats to Detect must be programmed to a value less than or equal to 28 when
AT/AF Detection is On. See Section 6.3, “Detecting atrial tachyarrhythmias”, page 78.
Tachyarrhythmia detection and bradycardia pacing – To ensure reliable ventricular
tachyarrhythmia detection, the programmer regulates the values available for bradycardia
pacing and tachyarrhythmia detection.
Paced AV and temporary Lower Rate interval automatic adjustment – If VF or VT
detection is enabled, the device automatically adjusts the temporary Lower Rate interval
and PAV interval when needed. This adjustment maintains a ventricular event to atrial pace
interval that is at least 30 ms greater than the ventricular detection interval.
Asynchronous pacing mode – All ventricular detection must be programmed Off and
AT/AF Detection must be programmed to Monitor when the programmed pacing mode is
DOO, VOO, or AOO.
VF detection backup – To ensure VF detection backup during VT and FVT episodes, VT
and FVT Detection cannot be enabled unless VF Detection is also enabled.
Reference Manual89
3
142
Medtronic ENTRUST
®
6.5.4 How to program VT Detection
D153ATG, D153DRG, D153VRC
1. Select the Params icon.
2. Select Detection (V.)… from the Parameters screen.
3. Select the desired values for enabling or disabling VT Detection, VT Initial Beats to
Detect, VT Beats to Redect, and VT V. Interval (Rate).
4. Return to the Parameters screen and select [PROGRAM].
90Reference Manual
A
S
A
S
V SV
S
V
S
V
S
A RA RA RA RA RA RA
R
10 12 34 5
T
S
T ST ST ST ST
S
V
S
A RA
R
A RA RA RA RA RA RA RA RA RA RA
R
76 8 910 11 12 13 14 15 16
200 ms
T ST
S
T ST ST ST ST ST ST ST
S
T
D
VT event count
ECG
VT interval
VT event count
ECG
VT interval
Marker Channel
Marker Channel
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.5.5 Details about VT detection
The device detects VT by counting the number of consecutive VT events. A VT event is a
V-V interval shorter than the programmed VT Interval but greater than or equal to the VF
Interval. If the number of consecutive VT events reaches the programmed VT Initial Beats
to Detect, the device detects VT (see Figure 10).
The VT event count resets to zero whenever an interval occurs that is greater than or equal
to the programmed VT Interval. The count remains at the current value if an interval is shorter
than the programmed VF Interval.
After the device detects VT, it delivers the first programmed VT therapy. Following the
therapy, if the VT event counter reaches the VT Beats to Redetect, the device redetects VT
and delivers the next programmed therapy.
Note: The device can also detect VT episodes using the combined count detection criterion
(see Section 6.8, “Detecting tachyarrhythmia episodes with combined count”, page 101).
Figure 10. Device detects VT episode
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval but greater than or equal to the VF Interval).
Reference Manual91
Medtronic ENTRUST
2 A ventricular interval occurs outside the VT detection zone. The VT event count resets to zero.
3 The VT event count reaches the programmed VT Initial Beats to Detect of 16 events, and the
device detects VT.
®
D153ATG, D153DRG, D153VRC
6.6 Detecting FVT episodes
The device detects episodes of Fast Ventricular Tachycardia (FVT) by examining the
cardiac rhythm for short ventricular intervals. If enough intervals occur in the programmed
FVT Detection zone, the device detects FVT and delivers the first programmed FVT therapy.
After therapy delivery, the device continues to evaluate the ventricular rhythm to determine
if the episode is ongoing. To make sure it delivers sufficiently aggressive therapies, the
device can merge the programmed detection zones during redetection to increase
sensitivity.
6.6.1 Parameters
FVT Detection – Enables FVT detection via the VF or the
VT detection algorithm.
FVT V. Interval (Rate) – V-V intervals between this value
and the programmed VF V. Interval (Rate) are marked as
FVT events.
OFF ; via VF; via VT
200; 210 … 240 … 600 ms
6.6.2 Considerations
Review the following information before programming FVT detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia
interval, program the VF, FVT, and VT V. Intervals (Rate) at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Beats
to Redetect lower than the VF and VT Initial Beats to Detect.
FVT Detection enabled – Your choice for an appropriate setting for FVT Detection should
depend on the patient’s VF and VT cycle lengths. After determining a reliably sensitive VF
Interval, consider the following suggestions:
●
If the patient presents with a clinical VT Interval that is in the VF zone, select via VF to
ensure reliable detection of VF. (VT Detection need not be enabled at all.)
●
If the patient presents with 2 clinical VT Intervals that are both outside the VF zone,
select via VT to allow for correct classification of the faster VT and offer a separate
therapy regimen for each VT.
●
If the patient presents with only 1 clinical VT interval that is outside the VF zone, enable
VF and VT Detection only, and set FVT Detection to Off.
92Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
FVT Detection and PR Logic criteria – You can program the device to distinguish between
SVT and FVT detection by enabling the PR Logic criteria. Note that the SVT Limit must be
programmed shorter than the VF Interval for the PR Logic criteria to affect FVT via VF
Detection.
Double tachycardia detection – When any PR Logic criterion is enabled, the device also
enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT, see
Section 6.14, “Detecting double tachycardias”, page 129).
6.6.3 Restrictions
Review the following information before programming FVT detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To ensure reliable ventricular
tachyarrhythmia detection, the programmer regulates the values available for bradycardia
pacing and tachyarrhythmia detection.
VF detection backup – To ensure VF detection backup during VT and FVT episodes, VT
and FVT detection cannot be enabled unless VF detection is also enabled.
Asynchronous pacing mode – All ventricular detection must be programmed Off and
AT/AF Detection must be programmed to Monitor when the programmed pacing mode is
DOO, VOO, or AOO.
FVT detection parameters – To ensure reliable ventricular tachyarrhythmia detection, the
programmer regulates the values available for the FVT parameter as follows:
●
If FVT Detection is set to via VT, VT Detection must be set to On.
●
If FVT Detection is set to via VF, the FVT Interval must be programmed to a value shorter
than the VF V. Interval (Rate).
●
If FVT Detection is set to via VT, the FVT V. Interval (Rate) must be programmed to a
value greater than the VF V. Interval (Rate) and less than the VT V. Interval (Rate).
Reference Manual93
3
142
Medtronic ENTRUST
®
6.6.4 How to program FVT Detection
D153ATG, D153DRG, D153VRC
1. Select the Params icon.
2. Select Detection (V.)… from the Parameters screen.
3. Select the desired values for enabling or disabling FVT Detection and FVT V. Interval
(Rate).
4. Return to the Parameters screen and select [PROGRAM].
94Reference Manual
200 ms
13 14 15 16 17 18
A RA RA
R
A RA RA RA RA RA
R
T
F
T F
T FT FT FT FT FV
S
1123 4
A
S
A
S
A RA RA
R
A
R
V
S
V
S
V
S
T ST
F
T FT FT FT
F
1
2
3
11 12
A RA RA
R
T FT FT
F
VF Event
Count
VF and FVT
Intervals
ECG
Marker Channel
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.6.5 Details about FVT detection
You can program the device to detect FVT episodes using the VF or VT Detection zone and
VF or VT Initial Beats to Detect.
When FVT Detection is set to via VF, a V-V interval within the FVT Detection zone is marked
as an “FVT via VF” event. When the count is reached for the programmable VF Initial Beats
to Detect, the device reviews the last 8 intervals:
●
If any one of the last 8 intervals is in the VF Detection zone, the device detects the
episode as VF.
●
If all of the last 8 intervals are outside the VF Detection zone, the device detects the
episode as FVT (see Figure 11).
When FVT Detection is set to via VT, a V-V interval within the FVT Detection zone is marked
as an “FVT via VT” event. When the count is reached for the programmable VT Initial Beats
to Detect, the device reviews the last 8 intervals:
●
If any one of the last 8 intervals is in the VF or FVT Detection zones, the device detects
the episode as FVT.
●
If all of the last 8 intervals are outside the FVT and VF Detection zones, the device
detects the episode as VT.
Note: The device can also detect FVT episodes using the combined count detection
criterion (see Section 6.8, “Detecting tachyarrhythmia episodes with combined count”,
page 101).
Figure 11. Device detects FVT via VF episodes
Reference Manual95
1 A fast ventricular tachycardia starts, and the first event falls into the FVT Detection zone.
2 The second event of the FVT episode has an interval that falls into the VT Detection zone. The
VF event count is not incremented.
3 The device detects FVT after the VF event count reaches the VF Initial Beats to Detect.
Before detection
After VF detection
After FVT detection
FVT set to “via VF”
VF and FVT zones merge, leaving a
larger VF zone.
All zones remain unchanged.
FVT set to “via VT”
VT and FVT zones merge, leaving a
larger FVT zone.
VT and FVT zones merge, leaving a
larger FVT zone.
Detection Intervals: VF Interval: 320 ms, FVT Interval: 280 ms / 360 ms, VT
Interval: 400 ms
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.6.5.1 Zone merging after detection
To ensure the device delivers sufficiently aggressive therapies during an extended or highly
variable tachyarrhythmia episode, the device merges detection zones during redetection in
some instances, as shown in Figure 12. The merged zone configuration uses the event
counting and therapies for the faster arrhythmia and remains in effect until episode
termination.
Figure 12. FVT zone merging
6.7 Monitoring VT episodes
The VT Monitor zone provides a programmable, diagnostic zone used for monitoring
tachyarrhythmias. This zone detects and records VT as VT Monitor episodes and does not
deliver VT therapy.
The VT Monitor zone can provide additional data about a patient’s condition by detecting
arrhythmias that are not programmed for detection by the VF, FVT, and VT Detection zones.
See Section 6.7.5, “Details about VT Monitor detection”, page 99.
96Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.7.1 Parameters
VT Monitor – Enables the VT Monitor.Monitor ; Off
VT Monitor Initial Beats to Detect – Number of
beats to detect: number of VT Monitor events that
the device must count to detect and classify a VT
Monitor event.
VT Monitor V. Interval (Rate) – V-V intervals
shorter than this value are counted as VT Monitor
events.
16; 20 … 56; 80; 110; 130
280; 290 … 450 … 650 ms
6.7.2 Considerations
Review the following information before programming VT Monitor.
VF Detection – VF Detection must be enabled to have VT Monitor enabled.
Combined count detection – Combined count detection does not include VT Monitor
events.
Sustained rhythm type classification – When a ventricular tachyarrhythmia occurs in the
VT Monitor zone, monitoring classification stops until the ventricular arrhythmia termination
criteria are met. See Section 6.7.5, “Details about VT Monitor detection”, page 99 for
details about monitoring classification.
SVT discrimination features – The enabled SVT discrimination features are applied in
the VT Monitor zone.
6.7.3 Restrictions
Review the following information before programming VT Monitor parameters.
AT/AF detection – If AT/AF Detection is On, the VT Monitor Initial Beats to Detect must be
less than or equal to 32, which is the AT/AF evidence counter threshold. See Section 6.3,
“Detecting atrial tachyarrhythmias”, page 78.
VF detection backup – To ensure VF detection backup during VT, FVT, and VT Monitor
episodes, if VF Detection is disabled, VT Detection, FVT Detection, and VT Monitor must
also be disabled.
VT Monitor Initial Beats to Detect – The VT Monitor Initial Beats to Detect must be greater
than the VF and VT Initial Beats to Detect.
Pacing rate – The pacing (lower rate, upper tracking rate, and upper sensor rate) interval
must be programmed 10 ms greater than the VT Monitor Interval (Rate).
Reference Manual97
3
142
Medtronic ENTRUST
®
6.7.4 How to program VT Monitor
D153ATG, D153DRG, D153VRC
1. Select the Params icon.
2. Select Detection (V.)… from the Parameters screen.
3. Select the desired values for enabling or disabling Monitor Detection, Monitor Initial
Beats to Detect, and Monitor V. Interval (Rate).
4. Return to the Parameters screen and select [PROGRAM].
98Reference Manual
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
6.7.5 Details about VT Monitor detection
The VT Monitor zone functions as an independent VT zone, operating with a lower rate
cutoff than any of the enabled detection zones. The enabled SVT discrimination features
are applied to the VT Monitor zone.
VT Monitor detection features a programmable number of intervals to detect sustained
arrhythmias. When the VT Monitor event count reaches the programmed number of
intervals, the rhythm is considered sustained. For AT and DR devices, VT Monitor detection
classifies sustained rhythms as either ventricular tachycardia (VTM) or double tachycardia
(VTM+SVT). For VR devices, VT Monitor Detection classifies sustained rhythms as
ventricular tachycardia (VTM).
Note: A double tachycardia (VTM+SVT) is a detected arrhythmia that comprises a
ventricular tachyarrhythmia (VT) with a simultaneous SVT. Double tachycardia detection
ensures that the PR Logic criteria do not compromise ventricular arrhythmia monitoring.
Once a ventricular tachycardia or double tachycardia is classified, further classification
stops until that sustained rhythm ends or terminates (see Figure 13). Until a sustained
rhythm is detected, a rhythm is classified as a rejected SVT rhythm, and VT Monitor
continues to look for ventricular tachycardia and double tachycardia. The rejection is based
on enabled features.
The device compares ventricular intervals to the VT Monitor Interval to identify when a VT
Monitor episode has terminated. Any one of the following criteria terminates an episode:
●
8 consecutive intervals greater than or equal to the VT Monitor Interval.
●
20 s elapse without the median ventricular interval shorter than the programmed VT
Monitor Interval.
●
Detection of a VT, FVT, or VF rhythm. Detection of a VT, FVT, or VF rhythm (or a SVT
rhythm in the VT, FVT, or VF Detection zones) terminates VT Monitor operation until
the rhythm terminates.
Reference Manual99
200 ms
1234
A
S
A
R
A RARA
R
V
S
V SV SV SV SV
S
13 181920000
A
R
A
R
A RA RA RA RA
R
TSV SV SV TV SV SV
S
12
3
Monitor Interval
Monitor Event
Count
Marker Channel
ECG
Medtronic ENTRUST
®
D153ATG, D153DRG, D153VRC
Figure 13. Device detects and monitors VT
1 A VT starts in the VT Monitor zone, and the device begins counting VT events.
2 The VT Monitor event count reaches the programmed VT Monitor Initial Beats to Detect of 20
events, and the device detects a VT Monitor episode.
3 After detecting the VT Monitor episode, the device monitors the episode until termination or
detection of VT, FVT, or VF rhythms.
8
6.7.5.1 Effects of enabling VT Monitor
Enabling VT Monitor causes the following detection operations to work differently:
VF, VT, and FVT detection – The VT Monitor zone is an independent zone, and it has no
effect on the VF, VT, or FVT zones.
PR Logic, Stability, and Onset criteria – Before the device detects a VTM or a VTM+SVT
episode, the PR Logic and Stability criteria, if enabled, are applied. If a VT Monitor episode
accelerates into the VF, VT, or FVT detection zone, the device continues to apply PR Logic
criteria as initial VF, VT, or FVT detection begins. Onset and Stability are independently
tracked between the VT Monitor and VT Detection zones.
Wavelet criterion – If Wavelet is programmed to Monitor, and VT Monitor is enabled, SVTs
are detected as VT Monitor events.
8
Detection of a VT, FVT, or VF rhythm (or SVT rhythm in the VT, FVT, or VF Detection zones) terminates
monitoring operation until the rhythm terminates.
100Reference Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.