Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Medtronic, Medtronic with rising man logo, and Medtronic logo are trademarks of Medtronic. Third-party trademarks (“TM*”) belong to their respective owners. The
following list includes trademarks or registered trademarks of a Medtronic entity in the United States and/or in other countries.
The Medtronic DiamondTemp FASTR RF generator (model CEDTG200) is part of the DiamondTemp ablation system, which also
includes the DiamondTemp ablation catheter, DiamondTemp irrigation pump, DiamondTemp catheter-to-RF generator (RFG) cable,
DiamondTemp GenConnect cable, DiamondTemp EGM cable, and DiamondTemp irrigation tubing set.
The DiamondTemp ablation system is designed to deliver radiofrequency (RF) energy to the cardiac anatomy via the DiamondTemp
catheter. The DiamondTemp RF generator provides RF energy and temperature monitoring functions, as well as control and
communication to the DiamondTemp irrigation pump and commercially available external devices, such as cardiac stimulators,
electrophysiology (EP) recording systems, and EP navigational and mapping systems like the EnSite Precision™ or EnSite™
Velocity™ Cardiac Mapping System (Abbott, MN).
The generator operates in temperature control mode. The desired catheter tip-to-tissue temperature is selected by the user.
Thermocouples in the catheter tip provide temperature feedback and the generator automatically adjusts the power output to maintain
the desired tip-to-tissue temperature.
The generator (Figure 1) has a touch-screen display, control buttons, and a control knob for modifying and controlling ablation
parameters during the procedure. Ablation parameters such as temperature, power, impedance, duration, and irrigation flow rate are
displayed on the front panel and can be recorded and saved by the generator in a format that can be downloaded to a computer or a
USB flash drive.
Audible indicators, visual indicators, and informational messages (with codes) are sounded out or displayed on the touch-screen
panel of the generator. A foot switch is also included with the generator and may be used as an option to start or stop RF energy
delivery.
The generator can be placed on a cart or a table top.
The generator has a power ramp time programmed to reach temperature set-point in approximately 1 second.
Note: There are many factors that influence the time to reach the temperature set-point: tissue contact, tissue morphology and
thickness, etc. The actual ramp time to reach maximum power may be greater than the programmed ramp time.
Figure 1. DiamondTemp RF Generator
1 Touch-screen display
2 Control knob
2 Indications for use
The DiamondTemp RF generator is designed for use with the DiamondTemp ablation system. Refer to the DiamondTemp catheter
instructions for the indications for use. Carefully review the specific indications, contraindications, warnings, precautions, and adverse
events included with the DiamondTemp catheter before using the generator.
3 Principle of operation
The generator operates by delivering RF energy, via the catheter, to discrete regions of the cardiac anatomy. The application of RF
energy causes localized thermal injury, which results in a conduction block at the targeted location.
4
4 Contraindications
The contraindications listed in the DiamondTemp catheter instructions apply to the use of the DiamondTemp generator. Carefully
review the specific indications, contraindications, warnings, precautions, and adverse events included with the DiamondTemp
catheter before using the generator.
5 Warnings and precautions
• Refer to the DiamondTemp catheter instructions for warnings and precautions related to the use of the DiamondTemp ablation
system.
• The generator materials are not compatible with magnetic resonance imaging (MRI).
• The generator is designed for use only with the DiamondTemp catheter, catheter-to-RFG cable, irrigation pump, tubing set, and
GenConnect cable. Use of other devices or improper cables may cause conditions in the operation of the generator that may result
in hazards to the patient.
• There is a direct electrical connection to the catheter electrodes through the EGM connection cables or the generator. Improper
use may be dangerous for the patient.
• Do not touch the accessible contacts of the connectors while touching the patient.
• To avoid unintentional skin burns to the patient at the return pad site during RF energy delivery, do the following:
– Minimize the distance between the return pad and the operating field.
– Minimize skin-to-skin contact between parts of the patient’s body by covering these areas with dry gauze.
– When using multiple ablation devices, remove those devices not actively in use from patient contact. In all cases, monitoring
systems incorporating high frequency current-limiting devices are recommended.
– Ensure the entire area of the return pad makes reliable contact with the patient’s body during ablation. Using the RF START
button to initiate an ablation results in continuous activation of RF energy output for the programmed duration. Do not remove
the return pad while the system is ablating.
• Radiofrequency temperature and power should not exceed the maximum limits specified in the catheter instructions.
• The Service Communication port and Remote Control port on the rear panel are for service only and no connections should be
made during normal use.
• The USB port on the rear panel is used to export generator data stored in internal memory. This data contains ablation data from
previous cases. Exporting of the data should only be performed when the system is not in use.
• Failure of the generator could result in an unintended increase of output power. Monitor informational messages displayed by the
generator and the irrigation pump.
• It is the responsibility of hospital personnel to use compatible return pads as specified in Chapter 9 and Chapter 10 of this user
manual. Improper return pad use may result in skin burns to the patient. When applying the return pad, position the lead such that
contact with the patient or other leads is avoided.
• The generator should be placed on a hard, level surface and not stacked on other equipment.
• Do not modify the generator or accessory components. Modifications may reduce system effectiveness and impact patient
health.
• The generator should not be used with a Booker box or equivalent devices.
• All devices connected to the generator must be safe for the patient per specifications in IEC 60601-1 and IEC 60601-2-2. Improper
use may be dangerous for the patient.
• When using the generator with the DiamondTemp catheter, the generator can be operated only in temperature control mode.
• The patient should not come into contact with grounded metal parts or metal parts that have considerable ground capacity (for
example, operating table).
• To avoid the risk of explosion, do not use the generator in the presence of flammable anesthetics or gases.
• To avoid the risk of electrical shock or fire, do not expose the generator to excessive moisture, especially when power is
connected. Make sure that the connectors to the catheter and cables are completely dry before connecting.
• The rear panel of the generator should remain readily accessible, so the power cord may be easily detached from the mains power
module, should that become necessary as part of providing mains isolation.
• Ensure that the return pad is positioned and connected properly. RF application will not start if the return pad is missing or
incorrectly connected to the generator.
• Verify effective contact between the patient and the return pad whenever the patient is repositioned. Patient movement may
disrupt return pad contact, resulting in patient injury or extended procedure times.
• Before increasing any ablation settings, such as temperature set-point or ablation duration, or before performing several
consecutive ablations with short off-periods, recheck the adherence of the return pad and its connections.
• Loss of contact between the return pad and the patient results in an error message, as controlled by the impedance cut-off
set-point (11, Figure 7).
5
• Read and follow the manufacturer’s instructions for use with the return pad. Use only return pads that meet appropriate regulatory
requirements.
• All reusable accessories should be visually inspected on a regular basis to avoid system malfunction and maintain patient safety.
No broken insulation of wires is allowed.
• Once the generator output START button is pressed, it will remain energized until either the front panel RF STOP button is pressed
or until the foot switch is released.
• Before use, inspect the generator and the packaging to verify that no damage has occurred. Do not use damaged products.
• Do not immerse the generator in any liquid or expose the generator to steam autoclave or ethylene oxide (EtO) sterilization.
• To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground.
• Electromagnetic interference (EMI) from the environment or produced by the generator may adversely affect the performance of
other equipment, including that of EP recording, mapping, or navigation systems. Excessive EMI may cause the generator to enter
a Safe State, which requires a reboot. The generator is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Note: The emission characteristics of this equipment make it suitable for use in industrial areas and hospitals. It is not intended
for use in a residential environment where this equipment might not offer adequate protection to radiofrequency communication
services. Use in such environments might require mitigation measures, such as relocating or reorienting the equipment.
• Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating
normally.
• Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could
result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of the DiamondTemp system, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
• Position connecting cables to avoid contact with the patient and other electrical leads.
• If for any reason (including EM interference) the generator loses communication with the irrigation pump, or if the irrigation pump
flow becomes 0 mL/min, the generator displays a message indicating that the condition should be corrected before any other
steps are taken.
• During use of the RF generator, pay attention to all messages, error codes, warnings, indicators, and tones, and exercise caution
as needed.
• Do not attempt ablation without the use of the irrigation pump. Before attempting ablation, make sure the pump flow rate is at the
minimum continuous flow and the pump is actively communicating with the generator.
• If the DiamondTemp system is used in conjunction with a compatible mapping system and GenConnect box (or similar connection
box), consult their respective instruction manuals to ensure correct connectivity and use. Construct the 3D anatomic map of the
region of interest only after all mapping catheters, the DiamondTemp catheter, and all respective cables and neutral electrodes
(including the ablation return pad) are completely and properly connected and positioned at the cardiac location of interest. The
subsequent addition of catheters or electrodes may lead to inaccurate anatomic mapping and may require remapping.
• Do not stack more than two generator shipping boxes on top of each other.
• If a serious incident related to the device occurs, immediately report the incident to Medtronic and the applicable competent
authority or regulatory body.
• Carefully inspect the package before opening. If the package has been damaged or opened, do not use and contact your
Medtronic representative.
• If you find information in this manual that is incorrect or illegible, contact your Medtronic representative.
6 Potential adverse events
The potential adverse events that may be associated with ablation procedures can vary greatly in frequency and severity and may
necessitate additional medical intervention, including surgery. Carefully review the specific indications, contraindications, warnings,
precautions, and adverse events included with each DiamondTemp catheter before using the DiamondTemp generator.
7 How supplied
The DiamondTemp generator is supplied with the following accessories and documentation. The generator and accessories are
packaged nonsterile and are not intended for sterilization.
• Foot switch
• Power cord
• User manual
6
The following additional accessories may be used with the DiamondTemp ablation system. Read all applicable instructions before
use.
• DiamondTemp EGM connecting cable, 4 male 2.0 mm shrouded connectors (supplied separately)
• Return pad - single-use, nonsterile, adult patient return pads that utilize high-moisture, conductive adhesive with 2.7 m (9 ft) cord,
for example the Valleylab E7507, E7507DB, or equivalent (not supplied)
Note: The term "return pad" is used throughout this manual and on the RF generator user interface. A return pad is also commonly
called dispersive electrode, dispersive indifferent patch (DIP), grounding pad, patient return electrode, or passive/plate electrode.
Note: Ensure that the connector on the return pad cable is compatible with the DiamondTemp RF generator port.
• Grounding cable, equipotential connector, DIN 42801, or equivalent (not supplied)
• Serial communication cable, shielded, DB-9 connector, EIA RS-232, or equivalent (not supplied)
• Video output cable, shielded, DB15 connector, VGA, or equivalent (not supplied)
8 Unpacking the generator
Carefully remove the generator from the shipping container. Unpack the user manual and all accessories to be used with the
generator. Inspect all items for possible damage during shipment. If any items are damaged, do not use them and contact a Medtronic
representative.
9 Setting up the generator
The generator is intended for use with the DiamondTemp catheter-to-RFG cable only (supplied separately). The catheter-to-RFG
cable connects the generator to the catheter. The catheter-to-RFG cable plugs into the front of the generator. When using the
DiamondTemp ablation system with a compatible mapping system, the GenConnect cable (supplied separately) can be connected
between the catheter-to-RFG cable and the generator to the GenConnect box (or similar connection box) (Section 10.3). Consult the
respective instructions for the catheter-to-RFG and GenConnect cables. Do not use cables with devices not indicated in the
respective instruction manuals.
The generator is designed for use with a one return pad, which connects to the front of the generator. Apply the return pad per its
instructions and according to the guidance in Section 9.2 of this user manual.
The generator is intended for use with the DiamondTemp irrigation pump and DiamondTemp irrigation tubing set only (supplied
separately). Included with the irrigation pump is a standard Ethernet cable to connect the irrigation pump to the generator. (See the
irrigation pump user manual for more information.)
9.1 Technical safety inspections
The following inspections for safety must be performed at first use and after each repair.
• Perform a general inspection of the generator:
– Assure the front and rear panels are not loose.
– The touch-screen display must not show signs of damage.
• Check that the selector knob is securely connected and able to easily rotate.
• Perform a visual and functional inspection of the accessories:
– Check the power cord and connector for any damage.
– Check the connecting cable input for the return pad electrode.
– Check the foot switch and its connector to ensure that it is not damaged and that it functions correctly.
• Check that the catheter (supplied separately) is connected appropriately.
• Check the return pad connector on the front panel of the generator for proper connection.
• Check the power cord connector for damage.
• Before operation, ensure that all connections are established and checked.
• The rear panel of the generator should remain readily accessible so that the power cord may be easily detached from the mains
power module. The means of isolating equipment from mains are provided by a power entry module with detachable power cord.
9.2 Handling the return pad
One (1) return pad is required in order to operate the generator. The generator does not deliver RF energy without a return pad properly
connected. If the return pad is or becomes disconnected, the generator will either not allow entry into RF energy delivery mode or will
terminate the delivery of RF energy with a ‘High Impedance’ informational code. The RF generator does not implement a continuity
monitor or a contact quality monitor for the return pad. For optimal performance, the self-adhesive return pad should be applied close
7
to the operating field. Figure 2 shows the recommended locations for the return pad. Ensure that the entire area of return pad is reliably
attached to a suitably prepared and appropriate area of the patient’s body, as defined by the return pad manufacturer. Use only a
compatible return pad (see Chapter 7).
Figure 2. Return Pad Positioning
10 Directions for use
Note: The screen images shown in this user manual are representative of what is seen on-screen with the software; actual images
may differ slightly.
10.1 Electrical connections
The generator operates using line power of 100 to 240 V~ 50/60 Hz. To avoid the risk of electric shock, connect only to a hospital-grade
grounded power outlet. The means of isolating equipment from mains are provided by a power entry module with detachable power
cord. The generator is not battery powered and cannot be moved during use.
10.2 Turning the generator on and off
Before operation, the generator must be placed on a stable surface with adequate circulation of air around the device to avoid
overheating. Additionally, the generator must be protected from moisture, contamination, and contact with flammable or explosive
substances.
Ensure that the power cord is plugged in to a power line of 100 to 240 V~ 50/60 Hz. Connect the power cord only to a hospital-grade
grounded power outlet and to the rear AC power cord inlet, as depicted in Figure 3. The grounding cable serves the purpose of
equipment potential equalization and should be connected to the equipotential (grounding) stud located on the rear panel of the
generator. The other end of the grounding cable should be connected to the grounding stud of another medical electrical equipment.
The daisy-chaining of grounding cables ensures that relevant medical electrical equipment operates at the same potential, thereby
reducing the chances of electrostatic or AC potential build-up. The power switch is located on the back panel of the generator near
the power cord inlet. Turn on the switch at the rear of the generator. Next, to start the generator, depress the soft power button on the
lower-left front panel of the generator, as shown in Figure 4. This button appears green when the generator is on, and amber when the
generator is in standby mode. An image with the Epix Therapeutics logo will appear on the touch-screen display on the front of the
generator after this button is depressed. Once turned on, the generator will perform a self-test. Once the self-test has passed, an
audible tone is heard, and the touch-screen display will show the Advanced Settings screen in standby mode with the word
“STANDBY” displayed in the top center of the screen.
8
The generator has four screen options that will be used during the procedure:
• Advanced Settings screen
• Contact Impedance screen
• Ablation Treatment screen
• Configuration screen
Unless otherwise instructed by an on-screen message, the device should be powered down by first pressing the soft power on/off
button on the front panel (1, Figure 4), not the rear power switch. Once the device returns to standby mode (front panel button
illuminated amber in color), the device may be completely powered off by switching the rear panel power switch (8, Figure 3) to the
0 position.
Figure 3. RF Generator Rear Panel
1 EP mapping system connection for EGM output and pacing
2 Video Output port
3 Serial Output port
4 USB port
5 Foot switch plug
Figure 4. RF Generator Front Panel
6 Equipotential (grounding) stud
7 AC power cord inlet
8 Power switch
9 Pump control port (using Ethernet cable)
1 Soft power on/standby button
2 Touch-screen display
3 Power indicator – green when power is on
4 Fault indicator – red when a fault condition is encountered
5 Catheter-to-RFG cable port
6 Return pad connection port
7 RF STOP - generator stop button
8 RF START - generator start button
9
10.3 Connecting the cables and accessory components
Use only the cables, accessories, and catheters provided or recommended. Refer to Chapter 7 for a list of supplied components and
compatible devices.
The irrigation pump and the generator must be used in communication mode. To establish this mode, plug the Ethernet cable provided
with the irrigation pump into the port on the rear panel of the pump, and connect the other end to the Pump Control port on the rear
panel of the generator (9, Figure 3). Once communication is established between the two devices, a communication icon (5,
Figure 7) will be displayed on the upper portion of the generator touch screen.
Note: When communication is established between the two devices, modifications to the flow rate or flow rate set-points on the
generator or pump will be enacted. However, the generator should be used to control the irrigation pump during the procedure.
Once the return pad has been adhered to the patient (Section 9.2), the return pad connector may be plugged into the front panel of
the generator (6, Figure 4).
Plug the sterile catheter-to-RFG cable into the catheter interface cable port on the front panel of the generator, marked by the catheter
icon (5, Figure 4). The multi-pin connector of the catheter-to-RFG cable, which has a green band on the connector to match the green
connector on the generator, will lock into place once plugged into the generator. In order to disconnect the catheter-to-RFG cable plug,
its corrugated sleeve must be slid back. When ready to connect the catheter, plug the distal end of the sterile catheter-to-RFG cable
into the back connector of the catheter. Refer to the catheter-to-RFG cable instructions for more detailed information.
Signals from the electrodes located on the tip of the catheter are passed through the generator and can be accessed for intracardiac
electrogram recording, stimulation, and location guidance purposes from the output port labeled with the ECG/Pace symbol on the
rear panel of the generator (1, Figure 3). An EGM cable can be plugged into the output port to route these signals to the appropriate
recording and mapping system. The signals on the output port are filtered; however, electromagnetic interference from the
environment or from the generator may affect recording and mapping systems. Connectivity to such systems should be made with
caution, avoiding unnecessary cable loops. Also, it is recommended to use monitoring systems that incorporate high frequency
current-limiting devices.
The DiamondTemp system can also be used with a compatible cardiac mapping system (such as the Abbott EnSite™ system). When
connecting the DiamondTemp system to the mapping system, use a GenConnect box (or similar connection box). For this purpose,
use the GenConnect cable. Connect one end of the GenConnect cable to the catheter-to-RFG cable. Connect the distal end of the
GenConnect cable to the generator. Connect the grey 9-pin connector to the catheter input of the GenConnect box. Connect the black
14-pin connector to the RF generator output of the GenConnect box. Figure 5 and Figure 6 provide connectivity diagrams for using
the DiamondTemp catheter and DiamondTemp RF generator, without the mapping system and with the mapping system. Confirm the
correct connectivity by using the 3-D navigation function of the mapping system. Use the mapping system and the GenConnect box
according to their respective instructions. Connect the return pad directly to the generator.
Figure 5. Connectivity Diagram between Catheter and RF Generator
1 DiamondTemp RF generator
2 26-pin male connector to RF generator
3 DiamondTemp cath-to-RFG cable
4 19-pin male connector to DiamondTemp ablation catheter
5 DiamondTemp ablation catheter
10
Figure 6. Connectivity Diagram between Catheter and RF Generator When Using a Mapping and Navigation System
6 Ablation return pad
7 14-pin twist connector
8 GenConnect box (oriented up-side down, for purposes of
illustration)
9 Amplifier
10.4 Advanced Settings screen
The generator will first display the Advanced Settings screen (Figure 7) upon powering-up on the touch-screen display panel. This
screen can also be accessed at any time during the procedure from the RF Ablation Treatment screen. The Advanced Settings screen
allows for generator and pump settings to be established, for presets to be saved, and for a new case to be started for each patient.
The default pump settings are 2 mL/min for low flow and 8 mL/min for high flow. A catheter does not need to be connected to the
generator to use the Advanced Settings screen.
7 Settings for Pre-Cool, Ramp Time, Duration and Post-Cool
8 “Treatment” button: touch to proceed to the Ablation
Treatment screen
9 “Set-up” button: touch to proceed to the Set-Up screen
10 Export data button
11 Start GenConnect auto calibration
The following ablation and flow rate parameters can be set up for the start of each case or established as a preset configuration.
Recommended settings are defined in Table 1.
• GenConnect auto-calibration process (11, Figure 7): Press this button to initiate the auto-calibration feature. The system will
detect absence or improper connection of a GenConnect unit or respective cable. When everything is properly connected, the
system will complete the calibration process and activate the Treatment button (8). If the auto-calibration fails, or if it is not
attempted, the Treatment button (8) is grayed out and further steps cannot be taken.
• Low and high irrigation pump flow rates, in mL/min (1, 2 in Figure 7): The value of the low irrigation flow rates can be set on
the side of the Advanced Settings screen by using the up or down arrows to change the value of the flow rate for each box. The
value of the high irrigation flow rate, which is only used during RF ablation, can also be set on the same screen.
Note: These functions will only be active (indicated by blue text) when communication is established between the generator and
irrigation pump.
• Power set-point, in Watts: With DiamondTemp catheters, the generator operates only in temperature control mode. This means
the generator controls the power level automatically, so that the temperature set-point is reached.
Note: The user cannot define or control the actual level of RF power delivered by the generator.
A maximum power set-point is available and may be adjusted up or down using the arrows next to the Power display. During
temperature-controlled RF energy delivery, the generator automatically adjusts the power only up to levels equal to or less than
this maximum power set-point. The recommended setting, which is also the default setting, is 50 W. By design, the generator
cannot deliver more than 50 W.
• Temperature set-point, in Celsius (4, Figure 7): The value of the temperature control set-point can be established by turning
the knob to the desired level.
• Timing for each stage of the ablation, in seconds (7, Figure 7): To adjust the timing for each stage of treatment, touch the
button for the desired stage located below the graph in the center of the screen. Once selected, the active stage will be highlighted
blue and the up and down arrows to the right can be used to increase or decrease the time increments in seconds.
– The pre- and post-cooling stages set the amount of time that the irrigation pump delivers irrigation before and after the ablation
to cool the tissue interface. No RF energy is delivered during the pre-cooling or post-cooling stage. At this stage, the generator
will set the irrigation pump flow rate to “High”.
– The set-point for the duration of the ablation can also be established on this screen. This number can be set at 1 second
increments between 0 and 999 seconds.
12
Table 1. Recommended Generator and Irrigation Pump Settings
RECOMMENDED/DEFAULT GENERATOR SETTING
Operational ModeTemperature Control
Maximum Temperature Set-Point60°C
Maximum Power Setting50 W
Maximum Ablation Duration45 seconds
DEFAULT PUMP SETTING
Irrigation Flow Rate during ablation8 mL/min
Minimum continuous flow rate2 mL/min
Preset memory keys
The generator has four (4) programmable preset memory keys (6, Figure 7), which can save and retrieve any desired combination of
preset parameters. To program a preset memory key, first set up the desired preset values as previously described, then push and hold
the preset key for 3 seconds. After holding the preset key for the required time, there will be an audible tone and the key will turn blue
to indicate successful programming of the preset values. The saved preset parameters can be retrieved from the Advanced Settings
screen or the Ablation Treatment screen by selecting the desired key. All presets are saved in the generator internal memory, even
when the generator is powered down between cases.
Case ID and starting a new case
A case is defined as a set of ablations performed in a patient or in a particular chamber of the heart. A new case is started each time
the generator is powered up, each time the new case button is selected. The generator saves relevant ablation parameters for each
patient into a case file with this unique identifier (see “Exporting Data” for additional details on saving ablation data).
To proceed to the Ablation Treatment screen, touch the “Treatment” button (8, Figure 7) in the lower right corner of the screen. The
Set-Up screen can also be navigated to by pressing the Set-Up icon (tools image, 9, Figure 7) at the bottom right-hand side of the
Advanced Setting screen.
Exporting Data
Data from each ablation procedure is saved to the internal memory of the generator. Each unique case ID created will have a unique
folder that stores the ablation data during the procedure. An external hard drive or memory stick with a USB connector can be
connected to the USB port on the rear panel of the generator (4, Figure 3 ). Once connected, the “Export Data” button (10, Figure 7)
will become active on the touch-screen display panel. Selecting this button will allow all data stored on the generator to be exported
to the connected USB device. Exporting data should only be performed when the system is not in use. The generator has storage
capacity for at least 1000 case files. Exporting the data does not clear the internal memory. If the internal memory reaches >80%
capacity (on the Configuration screen), contact a Medtronic representative.
10.5 Configuration screen
The Configuration screen (Figure 8) can be accessed from either the Advanced Settings screen or the Ablation Treatment screen by
pressing the Configuration icon at the bottom right-hand side of either screen (with the tools icon).
13
Figure 8. Configuration Screen
The Configuration screen allows the generator system and display preferences to be adjusted using the following buttons:
• Date and Time: Touch the part of the time or date to be changed. The active box will have a purple border. Press the up and down
arrows to set the date and time. Time can be selected in 12-hour or 24-hour format.
Note: If the date and time setting reverts to a time in the past (e.g. Jan 01, 2003) when the generator is powered off, contact your
Medtronic representative.
• Volume: Touch the green volume bar to set the desired volume.
• Brightness: Touch the yellow brightness bar to set the desired brightness.
• Language: Press the up and down arrows to choose a language.
Note: Once a new language is selected, the generator will restart and display information in the new language.
• Save & Exit: Press to save and exit back to the prior screen.
• Cancel: Press the Cancel button to exit without saving changes.
The Configuration screen also displays the Generator Serial Number, Software Version, Date of Manufacture (DOM), and the
percentage of memory available.
10.6 Contact Impedance screen
The Contact Impedance screen (Figure 9) can be accessed from the Advanced Settings screen. When a catheter is correctly
connected to the generator, the text at the top center of this screen will change from “STANDBY” to “READY”, indicating that ablation
may be performed. In READY mode, before initiating RF energy delivery, the Contact Impedance screen displays the tissue-contact
impedance vs. time. This impedance may be utilized to monitor the level of contact between tissue and the RF electrode. A higher
impedance value may indicate better tissue contact. However, caution should be exercised as other parameters, such as
electrograms, fluoroscopic or intracardiac echo images, should also be monitored.
Figure 9 shows an example of the Contact Impedance when the electrode is in good contact with tissue. The real-time trace shows
the history of tissue contact impedance over the previous 30 seconds.
Tissue contact impedance is not available during pre-cool, ablation, and post-cool phases. Once the RF START button is pressed,
the front panel screen toggles to the Ablation Treatment screen (Section 10.7).
The Ablation Treatment screen (Figure 10) can only be accessed from the Contact Impedance screen (Section 10.6), after the RF
START button is pressed, or during READY mode by selecting the ‘Graph’ tab.
The Ablation Treatment screen displays the real-time parameters (irrigation flow rate, RF impedance, RF impedance relative change
with respect to RF impedance start value, catheter highest temperature from all thermocouples, and ablation power and ablation
duration) across the top of the screen in four boxes (1, Figure 10). On this screen, purple color indicates duration, green color indicates
RF impedance, yellow color indicates power and orange color indicates temperature. The green Impedance field also includes the
relative change in RF impedance during the course of RF delivery. The relative change is displayed as a signed percent variable (%)
and it is computed relative to the initial value of RF impedance at the start of RF delivery. A change with a negative sign indicates a
relative drop in RF impedance with respect to its initial value.
On the right-hand side of the Ablation Treatment screen, the top display section in the “Settings” segment is outlined in orange and
labeled “Temperature,” if the generator is in temperature control mode (2, Figure 10). The temperature value represents the
4 Graphical display area
5 Message field
15
temperature set-point to be used during temperature control ablation. Rotating the knob adjusts the temperature set-point up or down.
These settings can be adjusted at any point during the procedure.
The Duration set-point (3, Figure 10) displays the ablation duration set-point value in seconds. This value can be increased or
decreased at any time during the procedure by using the up and down arrows located just below it. The maximum duration setting in
the generator is 999 seconds. This time does not include the pre-cooling or post-cooling duration. An ablation will automatically
terminate if this ablation duration set-point is reached.
Once correct communication has been established between the generator and the irrigation pump, irrigation flow rate levels can be
selected by touching the desired level. The irrigation flow set-points for the low and high levels can be modified remotely with the
generator by selecting the desired level and using the up and down arrows to modify the level to the desired value. With the
communication link established, the generator will automatically switch the irrigation pump between the low and high flow rate before
ablation, and then back down to the prior flow rate after the ablation has terminated. The irrigation flow from the pump can also be
stopped by touching the “STOP PUMP” button on the touch-screen panel.
Note: When the communication link is established between the generator and irrigation pump, modifications to the flow rate or flow
rate set-points on either the generator or pump will be enacted. Use the generator to control the irrigation pump during the procedure.
A preset memory value can also be selected by touching the desired preset button (6, Figure 7) to recall the stored preset value to one
of the predetermined ablation settings. Selecting a preset will change the current ablation and flow rate preset values on the generator
to those stored in the preset. If any of the values stored in the preset are changed on the generator after it is selected, the preset will
become inactive and the generator functionality will default to the adjusted parameters.
To start an ablation at the selected values on the generator, press the green “Start” button located below the control knob on the front
panel of the generator (8, Figure 4). Alternatively, the foot switch can be activated, if one is connected. Once the ablation duration
set-point has been reached, the ablation will be stopped. To terminate an ablation sequence before reaching the duration set-point,
press the red “Stop” button located above the control knob on the front panel of the generator (7, Figure 4) or alternatively, release the
foot switch pedal. The ablation duration set-point will still remain at the current value. The ablation process can be restarted by
pressing the “Start” button again. The top of the screen will display “RF ON”, and a constant audible tone will sound at all times while
RF energy is being delivered.
Once an ablation is initiated, a real-time graph of the temperature, power, and impedance (in Celsius, Watt and Ohm, respectively, on
the vertical axis) versus time (in seconds, on the horizontal axis) will be created in the “GRAPH” tab on the display screen (4,
Figure 10). The real-time graph displays temperature in orange, power in yellow, and impedance in green.
The “ABLATION DATA” and “CASE DATA” tabs of the screen contain summary information on the following parameters (Table 2) for
the current case:
Table 2. Ablation and Case Data Information
PARAMETERUNITS
DateNA
Total number of ablations in procedureNA
Total ablation time min
Average ablation duration s
RF power maximum W
RF power average W
Composite temperature range°C
Composite temperature average°C
RF Impedance start/end values Ω
Maximum/Average RF impedance Ω
Relative RF impedance change%
Total infused fluid volume mL
Tissue contact impedance maximum Ω
Tissue contact impedance average Ω
To access this information, touch the “ABLATION DATA” tab (Figure 11) or “CASE DATA” tab (Figure 12) to make it active (indicated
by a blue highlighting of the tab). This will display a summary of the case and ablation information in the area where the ablation graph
was located. Additionally, a summary of parameter values reached during the last ablation is also provided. The Last Ablation
summary provides data about: Mode (such as Temperature control), last ablation duration, power, impedance, temperature and
contact impedance values. To return to the graph, touch the tab labeled “GRAPH”. If any ablation is started while the tab is active, the
display window will automatically switch back to the graph display.
16
Figure 11. Ablation Data Tab
Figure 12. Case Data Tab
From the Ablation Treatment screen, it is possible to access the Advanced Settings screen by selecting the “Advanced Settings”
button at the lower-left corner of the screen. It is also possible to access the Configuration screen from the Ablation Treatment screen
by selecting the Configuration screen icon at the lower-right corner of the screen.
10.8 Indicators and informational messages
During the procedure, informational messages are displayed in a separate pop-up window or in the message field (5, Figure 10) at the
bottom of the touch-screen display on the generator. The displayed message contains an error code, an error message, and (for some
errors) recommended actions. Table 3 shows informational messages that may occur during use with recommended actions. Table 4
shows messages that do not have actions for the user to perform. For these messages, if it cannot be cleared, power down and restart
the generator. Report to Medtronic if the condition occurs repeatedly.
Some messages may indicate that there has been a hardware or software undesired operating condition of the generator. In addition,
the red indicator light on the upper-right front panel of the generator will turn on. If such an informational message occurs during
operation, attempt to restart the generator using the soft power on/standby switch on the front panel of the generator.
17
The informational messages are displayed with an error code (such as, E1003, P3032, or W110). The codes are useful for advanced
engineering troubleshooting. If it is necessary to call a Medtronic representative for support, it can be helpful to provide the error code
to the engineering team. If an error code is seen that is not listed in the tables below, contact a Medtronic representative.
Table 3. Generator Codes, Message Descriptions, and Recommended Actions
Error CodeError MessageRecommended Action
E1002EEPROM COMMUNICATION FAILURE Disconnect and reconnect catheter cable.
If problem persists, replace catheter.
E1003EXPIRED CATHETERReplace catheter.
E1006FOOTSWITCH FAILUREDisconnect and reconnect the Foot
E1014MISSING RETURN PADPlug in or replace the return pad. If problem
persists, cycle power to the Generator.
E1015MISSING CATHETERPlug in or replace catheter. If problem per-
sists, replace the Foot Switch.
E1022DISK SPACE LOWContact a Medtronic representative to free
up disk space.
E1023DISK SPACE FULLContact a Medtronic representative to free
up disk space.
E1024, E1053, E1059LOW IMPEDANCERecheck all cable, return pad, and catheter
connections. If problem persists, reboot
the RFG. If problem persists, replace cath-
eter, cables, or return pad.
E1025UNSUPPORTED ADAPTER CABLEDisconnect and reconnect catheter cable.
If failure recurs, replace catheter cable.
E1030, E1032PUMP NOT DELIVERINGCheck RFG to Pump connection. Discon-
nect and reconnect Ethernet cable. If prob-
lem persists, reboot the Pump.
E1031PUMP FLOW RATE LOWCheck RFG to Pump connection. Discon-
nect and reconnect Ethernet cable. If prob-
lem persists, reboot the Pump.
E1033PUMP GENERATED ERRORRefer to the Pump manual.
E1034EEPROM DATA INVALIDDisconnect and reconnect catheter. If fail-
ure recurs, replace catheter.
E1036GEN CONNECT NO CALIBRATIONDisconnect and reconnect the GenCon-
nect Cable. Perform GenConnect calibra-
tion. If problem persists, replace GenCon-
nect Cable and/or Generator Connection
Box E (or EPT GenConnect Box). Perform
GenConnect calibration again.
E1037RF OVER TEMPERATURECatheter may need replacement.
E1038GEN CONNECT CALIBRATION FAILEDDisconnect and reconnect the GenCon-
nect Cable. Perform GenConnect calibration. If problem persists, replace GenConnect Cable and/or GenConnect Box). Per-
form GenConnect calibration again. If
problem persists, reboot the RFG.
E1044PUMP NOT PRESENTDisconnect and reconnect the Ethernet
cable to the Pump and RFG. Reboot Pump
and RFG. If problem persists, replace the
Ethernet cable.
E1045RF TRANSITION PUMP NOT PRESENTDisconnect and reconnect the Ethernet
cable to the Pump and RFG. Rebook
Pump. If problem persists, replace the
Ethernet cable.
E1046-E1051TEMPERATURE SENSOR FAILUREReplace the catheter.
18
Table 3. Generator Codes, Message Descriptions, and Recommended Actions (continued)
Error CodeError MessageRecommended Action
E1052TEMPERATURE ABOVE SET POINTCheck irrigation flow. Replace catheter if
E1054EXCESSIVE DELIVERY TEMPERATURECatheter may need replacement.
E1055TEMPERATURE LIMIT REACHEDCatheter may need replacement.
E1056IMPEDANCE OUT OF RANGERecheck all cable, return pad, and catheter
connections. If problem persists, reboot
the RFG. If problem persists, replace cath-
E1057LOW IMPEDANCECatheter may need replacement.
E1058, E1060, E1065HIGH IMPEDANCECables may be disconnected, check con-
E1061PUMP FLOW OFFTurn on flow.
E1066NO VALIDATED CATHETERDisconnect and reconnect catheter. If
problem persists, replace the catheter.
E1067High RF current limit exceeded at return
pad
Confirm adequate return pad contact to
patient. Ablation may start once Heating
Factor falls below 30 A2s (message W05 is
cleared). For more information on ablation
duration and power settings, see Sec-
E1068High RF current limit exceeded - wait to
start ablation
Confirm adequate return pad contact to
patient. Ablation may start once Heating
Factor falls below 30 A2s (message W05 is
cleared). For more information on ablation
duration and power settings, see Sec-
F2999Unrecoverable software failure. Shut down
the system via the rear power switcha.
W05RF current high at return pad - RF output
limited
Restart the system after shutting down. If
problem persists, contact Medtronic.
Confirm adequate return pad contact to
patient. For more information on ablation
duration and power settings, see Sec-
W102, W104LOSS OF TISSUE CONTACTReposition catheter for better contact.
W105-W109LOSS OF CONTACTReposition catheter for better contact.
W103IMPEDANCE OUT OF RANGECatheter tip electrode may be inside intro-
ducer sheath. Reposition catheter or
W110-W111INEFFICIENT IRRIGATIONCheck pump flow or catheter contact to tis-
sue. Replace catheter and irrigation tube if
a
This message is displayed only in English.
condition persists.
eter, cables, or return pad.
nections.
tion 16.4.
tion 16.4.
tion 16.4.
check catheter connectivity.
necessary.
Table 4. Generator Codes and Message Descriptions
Error CodeError Message
E1004-E1005CURRENT DISCREPANCY
E1026EXCESSIVE RF OUTPUT CURRENT
E1027EXCESSIVE RF FORWARD POWER
E1028EXCESSIVE RF VOLTAGE
E1035RAMP SET POINT DISCREPANCY
E1039MEASURED POWER
E1042RF POWER DISCREPANCY
E1062-E1063VOLTAGE LIMIT EXCEEDED
F2001ADAPTOR
F2004EXCESSIVE SUPPLY CURRENT
19
Table 4. Generator Codes and Message Descriptions (continued)
P3017, P3021, P3037-P3038INTERNAL LOAD TEST FAILURE
P3032, P3035-P3036SOFTWARE FAILURE
P3039-P3044SELF TEST FAILURE
P3045POST CONTACT INDICATION LOAD
P3046POST CONTACT INDICATION PHASE
P3047STUCK RF ON BUTTON
10.9 Operation sequence for the generator
1. Before operation, the generator must be placed on a stable surface with adequate circulation of air around the device to avoid
overheating. Additionally, the generator must be protected from moisture, contamination, and contact with flammable or
explosive substances.
2. Plug in and turn on the generator.
3. Create and check all connections.
Note: The generator will recognize the catheter when the correct connection is made with the appropriate catheter-to-RFG
cable. Refer to the message bar for information on any missing connections. It is not required to notify the generator or select the
type of catheter connected.
4. Verify that the generator and irrigation pump are communicating by noting that the communication icon on the top of the generator
touch screen is illuminated.
5. Verify that the irrigation pump is fully operational, according to the irrigation pump user manual.
6. On the Advanced Settings screen, configure any procedure parameters, as desired. Use caution and sound medical reasoning
when deciding to change any of these parameters. Maximum continuous ablation duration at a single site should not exceed 60 s.
Note: To start a new procedure immediately after a prior one without turning off the generator, press the “NEW CASE” button on
the generator touch screen.
7. Check or modify any of the preset memory keys, if desired.
8. Enter patient information.
9. Touch the GenConnect auto-calibration button and wait until this process completes successfully.
10. Touch the configuration icon at the bottom right corner of the Advanced Settings screen to advance to the Configuration screen.
11. Ensure the generator configuration preferences are accurate. Modify configuration preferences, if desired.
12. Touch the “SAVE & EXIT” button at the bottom right corner of the generator touch screen to return to the Advanced Settings
screen.
13. When the preceding steps are complete, touch the “TREATMENT” button on the Advanced Settings screen to begin the
procedure.
14. Ensure that the generator displays “READY” in the top center field shaded green on the screen.
15. The generator enters the Contact Impedance screen. Monitor the electrode-tissue contact so that it is suitable for the goals of
the procedure.
16. Use the controls of the generator to establish the desired irrigation flow rate through the catheter. See the DiamondTemp catheter
and irrigation pump instructions for full information on setting up and operating the irrigation pump, as well as appropriate
irrigation flow rates during the ablation procedure.
Note: When communication is established between the two devices, modifications to the flow rate or flow rate set-points on
either the generator or pump will be enacted. Use the generator to control the irrigation pump during the procedure.
Note: It is recommended that the controls of the irrigation pump only be used in three cases: (1) during initial flush of the tubing
set and preparation of the catheter, (2) if air bubbles are detected in the tubing set, and (3) in case of an urgent need to change
the irrigation flow rate or stop the irrigation pump. Refer to the irrigation pump user manual for full information on set-up and
operation of the pump.
17. Set the desired ablation parameters on the generator or select the desired preset memory key. Use caution and sound medical
reasoning when deciding to change any of these parameters. Maximum continuous ablation duration at a single site should not
exceed 60 s.
20
Note: Read the complete DiamondTemp catheter manual for full instructions for use of the catheter, including ablation settings.
18. To start an ablation at the selected values on the generator, press the green “Start” button located below the control knob on the
front panel of the generator. If the ablation duration set-point has been reached, the current ablation will be stopped. To terminate
an ablation sequence before reaching the duration set-point, press the red “Stop” button located above the control knob on the
front panel of the generator. The ablation duration set-point will still remain at the current value.
Note: During an ablation, the programmed set-points for the ablation temperature level and duration can be adjusted, without
interrupting the ablation process, by using the knob or duration up and down keys, respectively.
Note: With the communication link established between the irrigation pump and generator, the high irrigation flow rate used for
ablation will be initiated automatically. After the ablation termination and the post-cooling phase, the irrigation pump will
automatically return to the irrigation flow rate that was on before the start of that ablation.
19. To restart subsequent ablations, first ensure that the appropriate ablation settings are displayed on the generator touch screen.
The ablation process can be initiated again by pressing the “Start” button on the generator front panel.
20. After the desired therapy has been delivered to the patient and the DiamondTemp catheter has been removed, the irrigation flow
may be stopped by touching the “Stop Pump” button on the touch-screen display.
21. Procedural ablation parameter data is automatically saved to the internal memory of the generator. To export this data after the
case, return to the Advanced Settings screen and connect an external hard drive or memory stick with a USB connector to the
USB port on the rear panel of the generator. Once connected, the “Export Data” button will become active on the touch-screen
display. Selecting this button will allow all data stored on the generator to be exported to the connected USB device.
22. Unless otherwise instructed by an on-screen message, the device should be powered down by first pressing the soft power
button on the front panel, not the rear power switch. Once the device returns to standby mode (front panel button illuminated
amber in color), the device may be completely powered off by switching the rear panel power switch to the 0 position.
11 DiamondTemp ablation system architecture and cybersecurity
Figure 13. DiamondTemp Ablation System Architecture Diagram
1 Specially Configured Maintenance USB Thumb Drive
2 USB Port
3 Power Cable
4 RS-232 Output to EP Recording System
5 Static IP Output only Ethernet Port
6 RF Generator
7 Logged Measurements
8 Program & Settings
9 Pump Control port
The DiamondTemp generator and irrigation pump devices are not intended for use on or to be connected to a computer network and
do not accept wireless or unknown physical connections. A system architecture diagram for the DiamondTemp ablation system is
presented in Figure 13. The Pump Control port is dedicated only for communications between the DiamondTemp generator and
DiamondTemp irrigation pump. The USB port is used to export data when the system is not in use.
10 CAT 5 Cable (Pump Flow Rate Commands &
Acknowledgements)
11 Irrigation Pump
12 Motor
13 Saline Bag
14 Irrigation Tubing
15 Patient
16 Catheter
21
Any suspected compromise of the DiamondTemp ablation system’s cybersecurity, from such events as unauthorized access,
computer virus infection, or inadvertent connection to a network, should be reported to Medtronic. A proper course of action,
determined by Medtronic and the end user, should be determined before the system can be further used.
Software upgrades are to be performed only by Medtronic or authorized personnel.
The DiamondTemp ablation system contains the following commercial, open source, or off-the-shelf software:
• QT Third party graphics library
• ALSA Linux Sounds generation library
• SQLite Database Interface library
• Customized Linux Operating System including TCP/IP stack
12 Maintenance and service
12.1 Cleaning
The DiamondTemp generator must be protected from moisture, contamination, and contact with flammable or explosive substances.
The generator and foot switch surface may be cleaned with non-flammable and non-explosive agents only, according to the following
steps. Follow universal precautions for protective apparel when handling and cleaning contaminated instruments. Make sure no fluids
or moisture enter the interior of the generator during cleaning.
1. Before cleaning, turn off the generator. Disconnect the power cord from the electrical power source and from the rear of the
generator. Disconnect all other cables and peripherals.
2. Wipe the generator enclosure with a clean, soft, nonlinting cloth dampened with a pH neutral detergent.
3. Wipe again with distilled or sterilized water.
4. Wipe dry with a clean, soft, nonlinting cloth.
5. To disinfect, use broad spectrum disinfectants like phenolics, aldehydes, or alcohols. Recommended disinfectants are
Glutaraldehyde 1%, 70% IPA, sodium hypochlorite (0.1%) or equivalent. Follow the manufacturer’s instruction for use to use the
disinfectant.
Do not immerse the generator in any liquid or expose the generator to steam autoclave or ethylene oxide (EtO) sterilization.
For instructions on the use, cleaning, or sterilization of any accessory cables to the generator, refer to the cable instructions.
12.2 Maintenance
There are no user-serviceable parts in the generator, except as noted in the instructions below. The generator requires no
adjustments, calibrations, or regularly-scheduled maintenance.
While maintenance is not required, local standards and regulations should be followed with respect to periodic performance
verification. If the generator requires repairs or is defective, the unit should be returned to the manufacturer for service or replacement.
Contact a Medtronic representative for details.
12.3 Replacing fuses
To replace a fuse in the generator:
1. Remove the power cord.
2. Use a small blade screwdriver or similar tool to unlatch the
fuse holder door at the top of the fuse holder.
3. Use a small blade screwdriver or similar tool to remove the
red fuse block from the fuse holder.
4. Remove the defective fuse(s) and replace with the correct
size, rating, and type (replace with Littelfuse 2183.15P or
equivalent). To avoid the risk of fire, use only the specified
fuse.
5. Reinstall the fuse block, close the fuse holder door, and
reinstall the power cord.
6. Switch on the RF generator and confirm it powers up.
13 Storage
Disconnect the generator from a power source for long-term storage. For additional storage information, see Section 16.9
22
14 Disposal
Refer to local requirements regarding the disposal of the generator and accessories.
15 Guidance and manufacturer’s declarations
Table 5. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The generator is intended for use in the electromagnetic environment specified below. The customer or the user of the generator
should assure that it is used in such an environment.
Emissions TestCompli-
ance
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 6. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The generator is intended for use in the electromagnetic environment specified below. The customer or the user of the generator
should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance levelElectromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2±2 kV, ±4 kV, ±8 kV,
Electrical fast transient/burst
IEC 61000-4-4±2 kV @ 100 kHz repeti-
SurgePower inputs:Power inputs:Mains power quality should be that of a typical com-
IEC 61000-4-5±0.5 kV, ±1 kV, ±2 kV
Voltage dips, short interruptions and voltage variations on power supply
input lines
IEC 61000-4-110% UT; 1 cycle and
Note: UT is the a.c. mains voltage before application of the test level.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-850 Hz or 60 Hz50 Hz or 60 Hz
±8 kV contact±8 kV contactFloors should be wood, concrete or ceramic tile. If
±15 kV air
±2 kV @ 100 kHz repeti-
tion frequency for power
supply lines
tion frequency for
input/output lines
±0.5 kV, ±1 kV Line-toLine
Line-to-Ground
Signal input/outputs:Signal input/outputs:
±2 kV Line-to-Ground±2 kV Line-to-Ground
Voltage Dips:Voltage Dips:Mains power quality should be that of a typical com0% UT; 0.5 cycle0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315° phase angles
70% UT; 25/30 cycles
Single phase: at 0°Single phase: at 0°
Voltage interruptions:Voltage interruptions:
0% UT; 250/300 cycle0% UT; 250/300 cycle
30 A/m30 A/mPower frequency magnetic fields should be at levels
Group 1The generator must emit electromagnetic energy in order
Class AThe generator is suitable for use in all establishments
Class A
Complies
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±2 kV @ 100 kHz repetition frequency for power
supply lines
±2 kV @ 100 kHz repetition frequency for
input/output lines
±0.5 kV, ±1 kV Line-toLine
±0.5 kV, ±1 kV, ±2 kV
Line-to-Ground
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315° phase angles
0% UT; 1 cycle and
70% UT; 25/30 cycles
Electromagnetic environment—Guidance
to perform its intended function. Nearby electronic equipment may be affected.
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
mercial or hospital environment.
mercial or hospital environment.
If the user of the generator requires continued oper-
ation during power mains interruptions, it is recommended that the generator be powered from an
uninterruptible power supply or battery.
characteristic of a typical location in a typical commercial or hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of the generator including cables, than the
recommended separation distance calculated from the equation:
d=6/E×√PWhere:
d is the separation in meters
P is the maximum output power rating of the transmitter in watts (W) according to the manufacturer
E is the Compliance Test Level indicated above.
Interference may occur in the vicinity of equipment marked with the following
symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the generator or any of its components are used exceeds the applicable RF compliance level
above, the generator should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating components or the generator.
b
Within the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 7. Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Generator
The generator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the generator can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the generator as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power
of transmitter
(W)
150 kHz to 80 MHz
d = 1.2√P
Separation distance according to frequency of transmitter
(m)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2√P
d = 2.3√P
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
24
Table 7. Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Generator
(continued)
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be
determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 – At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 – These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
16 Technical specifications
16.1 Mains input
• Input voltage/current/frequency: 100 - 240 Vac / 2.5 A - 1.4 A / 50/60 Hz
16.2 Fuses
• Replace fuses with Littelfuse 2183.15P or equivalent
16.3 RF output
• RF frequency: 460 kHz ± 1%, quasi-sinusoidal
• Max rated power: 50 W into 50 – 288 Ω load
• Maximum current:
– Heating Factor of 0-30 A2s: 1 ARMS
– Heating Factor of >30 A2s: 0.8 ARMS
• Maximum voltage: 120 V
• Maximum Heating Factor generated in any 60 second period: 40 A2s
RMS
16.4 Heating Factor
This generator model contains software that tracks RF current (A) applied to the return pad from the catheter tip over the duration of
the ablation. This parameter is called ‘Heating Factor’ with units A2seconds (A2s).
Heating Factor calculations are tied to informational messages W05, E1067, and E1068 in Table 3. Full RF output current (1 ARMS)
is allowed when Heating Factor is less than 30 A2s. Current output is reduced to 0.8 ARMS when Heating Factor goes above 30 A2s,
in alignment with emission of W05 warning message. RF ablation is terminated when Heating Factor reaches 40 A2s (error E1067)
and may not commence again until Heating Factor drops below 30 A2s (message W05 is cleared) to allow for heat dissipation at the
return pad site.
Extended duration of high RF power output may result in excessive heating of the skin at the return pad site. To reach the maximum
power set-point, current is modulated based on RF impedance. For an example of this relationship and its effect on Heating Factor see
Table 8. At lower RF impedance values, higher current is required. At higher RF impedances, lower current is required, and the warning
and error messages will not be seen (N/A in Table 8).
Table 8. Heating Factor and Current Limitations
Single Ablation Parameters
Maximum RF Power (Watts) Average RF Impedance (Ω) W05 (Heating Factor 30 A2s)E1067 (Heating Factor
50603651.6
508048N/A
5010060N/A
50101 and aboveN/AN/A
405541.356.9
408060N/A
4081 and aboveN/AN/A
Duration to warning / error (seconds)
40 A2s)
16.5 Electrical safety
• IEC 60601 Class 1 equipment
• Rated for continuous operation
• Accessory rated voltage: 120 V
minimum
RMS
25
• Applied parts classification
– Catheter - type CF defibrillation-proof
– Neutral (Indifferent/Return) Electrode – type F high frequency isolated
16.6 Settings
• Temperature: 40°C to 80°C
• RF Power: 0 to 50 Watts
• RF On Time: 0 to 999 seconds
• RF Ramp Time: 0 to 30 seconds
16.7 Measurement accuracy
• Temperature: ± 1.5 °C
• Power: ± 7% or ± 2 W of setting
• Impedance: ± 10% or ± 5 Ω whichever is greater from 35 to 300 Ω
• Time: ± 1 sec of setting
16.8 Mechanical specifications
• Size: 38 cm x 45 cm x 22 cm
• Weight: less than 11 kg
• Foot switch cable length: 355 cm ± 15 cm
Moisture protection rating
The generator and foot switch comply with the following international electrical safety ratings with regard to water as required by IEC
60601-1:
• Generator: IPX0, not water resistant
• Foot switch: IPX8, immersion resistant
16.9 Environmental specifications
• Note: Allow the system to equilibrate to room temperature for a minimum of 4 hours if the system has been stored at temperatures
outside of the operational temperature range.
• Operational temperature: 15°C to 30°C (59°F to 86°F)
• Operational humidity: 30% to 75% relative humidity (noncondensing)
• Operational pressure: 70 to 106 kPa
• Storage temperature: 15°C to 30°C (59°F to 86°F)
• Stacking height: Maximum of two shipping boxes on top of each other
16.10 Output power curves
Figure 14. Power vs. Impedance RF Generator
26
1 Power [W]
2 Impedance [Ω]
3 Full Power vs. Impedance
4 Half Power vs. Impedance
17 Limited warranty
The following limited warranty applies to customers within the United States only:
A. This limited warranty provides the following assurance to the customer of the Medtronic DiamondTemp RF generator, irrigation
pump, GenConnect cable, and EGM cable, with reusable parts (foot switch, ethernet cable, and power cord), hereafter collectively
referred to as the DiamondTemp ablation system. Subject to the limitations herein, Medtronic warrants the DiamondTemp ablation
system sold to the customer will be free from defects in materials and workmanship under normal usage for a period of 12 months from
the delivery date at the customer’s facility.
B. Should the DiamondTemp ablation system fail to meet the above warranty, Medtronic will at its option, repair or replace such
DiamondTemp ablation system, or any portion thereof. For the limited warranty to apply, the following conditions must be met:
(1) Medtronic must be notified of and confirm the failure of the alleged defect within 60 days after discovery of the defect.(2) The
DiamondTemp ablation system must not have been repaired or altered outside of authorized personnel at Medtronic.
(3) THE DIAMONDTEMP ABLATION SYSTEM MUST BE USED IN ACCORDANCE WITH LABELING AND NOT ALTERED OR
SUBJECTED TO MISUSE, ABUSE, IMPROPER HANDLING, OR ACCIDENT, INCLUDING, BUT NOT LIMITED TO, FIRE, FLOOD
OR NATURAL DISASTERS.C. At the discretion of Medtronic, parts or assemblies used or installed as part of the DiamondTemp
ablation system may be either new or rebuilt of equal or improved quality. All parts removed or replaced during maintenance become
the property of Medtronic.
D. This limited warranty is limited to its express terms. In particular:
(1) Except as expressly provided by this limited warranty, MEDTRONIC EXPRESSLY DISCLAIMS ANY REPRESENTATION OR
WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WHETHER AS TO MERCHANTABILITY, FITNESS FOR A PARTICULAR
PURPOSE, OR ANY OTHER MATTER. THE REMEDIES SET FORTH IN THIS WARRANTY ARE THE EXCLUSIVE REMEDIES
AVAILABLE TO THE CUSTOMER OR ANY THIRD PARTY FOR BREACH OF WARRANTY. MEDTRONIC SHALL HAVE NO
LIABILITY TO ANY PERSON FOR INCIDENTAL, CONSEQUENTIAL, OR PUNITIVE DAMAGES OF ANY DESCRIPTION,
WHETHER ARISING OUT OF WARRANTY, OTHER CONTRACT, TORT, OR OTHERWISE.
(2) Except as expressly provided by this limited warranty, MEDTRONIC IS NOT RESPONSIBLE FOR ANY DIRECT, INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES BASED ON ANY DEFECT, MALFUNCTION, OR FAILURE OF THE DEVICE TO
FUNCTION WITHIN NORMAL TOLERANCE, WHETHER THE CLAIM IS BASED ON WARRANTY, CONTRACT, TORT, OR
OTHERWISE.
E. The exclusions and limitations set out above are not intended to, and should not be construed so as to contravene mandatory
provisions of applicable law. If any part or term of this limited warranty is held to be illegal, unenforceable, or in conflict with applicable
law by a court of competent jurisdiction, the validity of the remaining portions of the limited warranty shall not be affected, and all rights
and obligations shall be construed and enforced as if this limited warranty did not contain the particular part or term held to be invalid.
This limited warranty gives the customer specific legal rights. The customer may also have other rights that vary from state to state.
F. No person has any authority to bind Medtronic to any representation, condition, or warranty except this limited warranty.
G. This limited warranty is not applicable to accessories or products used with the DiamondTemp ablation system unless specifically
noted.
18 Glossary of symbols
The following table defines symbols that are used on packaging and product labeling. Refer to the labels to determine which symbols
apply to this product and for the product-specific information, such as the date of manufacture.
Symbol
Standard/Standard title or
reference
21 CFR 801.109
ISO 15223-1
EN 50419
IEC 60601-1
a
d
b
c
Symbol title/Reference numberExplanatory text
Prescription onlyUSA Federal law restricts this
device to sale by or on the order
of a licensed healthcare practitioner.
Consult instructions for use
(clause 5.4.3)
Consult instructions for use at
this website:
www.medtronic.com/manuals
Recycle: Electronic Equipment Do NOT throw in trash
Follow instructions for use
(Table D2, Symbol 10)
Refer to instruction manual/booklet (blue symbol)
27
Symbol
Standard/Standard title or
reference
ISO 15223-1
d
Symbol title/Reference numberExplanatory text
Catalog number
(clause 5.1.6)
Indicates the manufacturer’s
catalog number so the device
can be identified
ISO 15223-1
d
Serial number
(clause 5.1.7)
Indicates the manufacturer’s
serial number so that the device
can be identified
ISO 15223-1
N/AManufactured in/ manufactur-
ISO 15223-1
d
Manufacturer
(clause 5.1.1)
Indicates the medical device
manufacturer
Indicates where the device was
d
Date of manufacture
(clause 5.1.3)
ing site
manufactured
Indicates the date when the
medical device was manufactured
ISO 15223-1
d
Keep dry
(clause 5.3.4)
Indicates a medical device that
needs to be protected from
moisture
ISO 15223-1
d
Do not use if package is damaged
(clause 5.2.8)
Indicates a medical device that
should not be used if the package has been damaged or
opened
ISO 15223-1
d
Temperature limit
(clause 5.3.7)
Indicates the temperature limits
to which the medical device can
be safely exposed
N/AStorage temperature limitIndicates the required temper-
ature range for storing the
device
N/ATransit temperature limitIndicates the required temper-
ature range for transporting the
device
ISO 15223-1
d
Humidity limitation
(clause 5.3.8)
Indicates the range of humidity
to which the medical device can
be safely exposed
ISO 15223-1
d
Atmospheric pressure limitation
(clause 5.3.9)
Indicates the range of atmospheric pressure to which the
medical device can be safely
exposed
To indicate elevated, potentially
dangerous, levels of non-ionizing radiation
To indicate that items shall not
be vertically stacked beyond
the specified number
ISO 7000
e
Equipotentiality
(symbol 5021)
To identify the terminals which,
when connected together, bring
the various parts of an equipment or of a system to the same
potential, not necessarily being
the earth (ground) potential,
e.g. for local bonding
ISO 7000
ISO 7000
e
e
Foot switch
(symbol 5114)
Alternating current
To identify a foot switch or a
connection for a foot switch
Alternating current
(symbol 5032)
ISO 7000
e
Defibrillation Proof Type CF
Applied Part
(symbol 5336)
To identify a defibrillation-proof
type CF applied part complying
with IEC 60601-1.
28
Symbol
Standard/Standard title or
reference
ISO 7000
ISO 7000
e
e
Symbol title/Reference numberExplanatory text
Stand by
(symbol 5009)
OFF
Power in Standby (lit amber) or
ON (lit green)
Power OFF
(symbol 5008)
ISO 7000
e
ON
Power ON
(symbol 5007)
N/AN/AFuses
N/AN/AEGM Output – Filtered
ECG / Pace
N/AN/AVideo Output
N/AN/ASerial Output
N/AN/AService Communication port
(not for clinical use)
N/AN/ARemote Control port (for future
generation use)
N/AN/APump Control port
N/AN/AUSB port (when system is not in
use)
N/AN/ACatheter connection
N/AN/ANeutral Electrode connection
N/AN/AFloating Neutral Electrode con-
nection
IEC 60529
IEC 60529
f
f
International Protection (IP)
Code
International Protection (IP)
Code
Indicates the product is not
water resistant (generator)
Indicates the product is able to
be submerged in water (foot
switch)
N/APackage contentsIndicates the components
included in the device package
N/AGeneratorIndicates that the type of device
is RF generator
N/AAccessoriesIndicates that accessories are
included in the device package
ISO 7000
e
Product documentation
(symbol 0419)
Indicates that product documentation is included in the
device package
N/AN/AIndicates the device Stop but-
ton (red symbol)
N/AN/AIndicates the device Start but-
ton (green symbol)
N/AN/APower cord
ISO 15223-1
d
Lot number
(clause 5.1.5)
Indicates the manufacturer’s lot
number so that the lot can be
identified
29
Symbol
Standard/Standard title or
reference
ISO 15223-1
d
Symbol title/Reference numberExplanatory text
Fragile, handle with care
(clause 5.3.1)
Indicates the device is fragile
and should be handled with
care
N/ACompliance markIndicates conformance to appli-
cable standards
a
21 CFR 801.109: United States Code of Federal Regulations, Title 21, Food and Drugs
b
EN 50419: Marking of electrical and electronic equipment in accordance with Article 11(2) of Directive 2002/96/EC (WEEE)
c
IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
d
ISO 15223-1: Medical devices – Symbols to be used with medical device labels, labeling and information to be supplied
e
ISO 7000: Graphical symbols for use on equipment
f
IEC 60529: Degrees of protection provided by enclosures (IP Code)