Sonicision™ Curved Jaw Cordless Ultrasonic
Dissection System
Indications for Use
The Sonicision™ curved jaw cordless ultrasonic
dissection device is indicated for soft tissue
incisions when bleeding control and minimal
thermal injury are desired. The device can be
used as an adjunct to or substitute for
electrosurgery, lasers, and steel scalpels in
general, plastic, gynecologic, urologic, and
other open and endoscopic procedures. The
Sonicision™ curved jaw cordless ultrasonic
dissection device can be used to coagulate
isolated vessels up to and including 5 mm in
diameter.
The Sonicision™ 13 cm device is also indicated
for use in otorhinolaryngologic (ENT)
procedures.
Contraindications
• The device is not indicated for incising
bone.
• The device is not to be used for
contraceptive tubal occlusion.
en
Clinical Benefits
The Sonicision™ curved jaw cordless ultrasonic
dissection device and compatible accessories
perform simultaneous cutting and
coagulation for soft tissue, and isolated vessels
up to and including 5 mm in diameter.
Ultrasonic dissectors have been shown to
reduce operative times and blood loss
compared to conventional techniques (cold
steel and manual ligation).
13
en
Section I – The Curved Jaw Cordless
Ultrasonic Dissection System
Overview
The Sonicision™ curved jaw cordless ultrasonic
dissection system enables cordless
transection, dissection, and coagulation of
tissue during general surgery. The device can
be used in open procedures, or endoscopically
when used with a compatible 5 mm trocar.
When assembled, electrical power supplied by
the battery pack is converted to ultrasonic
energy in the generator. This energy is
transferred to the active blade, which is used
to accomplish the intended clinical use.
Intensity of the ultrasonic energy is controlled
by the clinician using the dual-mode energy
button. Minimum power mode (energy button
depressed to the first position) produces
slower dissection when increased hemostasis
is desired. Maximum power mode (energy
button fully depressed) produces faster
dissection.
The Curved Jaw Cordless Ultrasonic
Dissection System Components
The Sonicision™ curved jaw cordless ultrasonic
dissection system has both single-use and
14
reusable components. Sonicision™ reusable
generators and reusable battery insertion
guides must be sterilized before use. Validated
sterilization methods and parameters are
provided in this document. Sonicision™
curved jaw cordless ultrasonic dissectors are
supplied sterile. They are intended for singlepatient use and cannot be adequately cleaned
and sterilized for reuse.
The assembled dissector, reusable generator,
and reusable battery pack are referred to as
the Sonicision™ curved jaw cordless ultrasonic
dissection device in this document. The
Sonicision™ curved jaw cordless ultrasonic
dissection system refers to all components,
including the assembled device, reusable
battery insertion guide, and battery charger.
Each component is further described in this
section.
The Sonicision™ curved jaw cordless ultrasonic
dissection system has no components made
with natural rubber latex.
Curved Jaw Cordless Ultrasonic Dissector
en
1
2
8
Clamping jaw– Holds tissue against
active blade
ཱActive curved blade– Delivers energy
effect
ིShaft
ཱིShaft-rotation wheel (black)– Rotates
shaft 360°
ུSpeaker– Emits system status tones
ཱུDual-mode energy button (blue)–
Selects energy activation in minimum
or maximum mode
ྲྀJaw lever (black)– Closes and opens
clamping jaw
3
4
5
6
7
10
9
ཷShaft measurement marks– Measures
up to 10 cm from the distal end (not
available on all models)
ླྀBattery door (black)– Closes and seals
battery compartment
ཹBattery door release button (black)–
Opens the battery door
The Sonicision™ curved jaw cordless ultrasonic
dissector is a sterile, single-use component to
which the generator and battery pack attach.
This component provides control for device
functions such as selecting power levels, blade
placement and position, grasping,
coagulating, and dissecting tissue.
15
en
Reusable Generator
up to 150 times, after which the generator is
no longer usable and must be replaced.
1
Reusable Battery Pack
2
LED indicator– Color of LED indicates
system status (see Device Status and
Troubleshooting on page 24 for color
indications and descriptions)
ཱTorque knob– Used to connect the
generator to the dissector
Generators are provided non-sterile and must
be sterilized before each use. Use a steam
autoclave to safely sterilize the generator. See
Cleaning and Sterilizing the Battery Insertion
Guide and Generator on page 34.
The Sonicision™ reusable generator converts
electrical power from the battery pack to
ultrasonic energy to drive the active blade. The
generator is designed to be sterilized and used
16
1
Battery-pack handle – Aids in holding
and positioning the battery pack for
insertion and extraction from the
dissector or battery charger
Battery packs are provided non-sterile and
must be cleaned, disinfected, and charged
prior to each use. See Cleaning and Disinfecting the Battery Pack on page 41.
The Sonicision™ reusable battery pack is a
rechargeable, lithium-ion battery that
provides power to the Sonicision™ curved jaw
cordless ultrasonic dissection device. The
battery pack is designed to be used up to 200
times. It is compatible only with the
Sonicision™ curved jaw cordless ultrasonic
dissection device and battery charger.
Battery Charger
1
en
Reusable Battery Insertion Guide
2
3
Charging bays (4)
ཱBattery-charger power-status light
ིBattery-pack status display
The battery charger connects to a standard
wall outlet and charges the reusable battery
packs. Status displays on the battery charger
indicate the status of each battery pack. See
Charging Indications and Troubleshooting on
page 44.
The Sonicision™ reusable battery insertion
guide is intended to facilitate the aseptic
transfer of the non-sterile Sonicision™ battery
pack into the Sonicision™ curved jaw cordless
ultrasonic dissection device.
The battery insertion guides are provided nonsterile and must be sterilized before each use.
See Cleaning and Sterilizing the Battery Insertion Guide and Generator on page 34.
17
en
Section II – Before a Procedure
Warning
Safe use of the Sonicision™ curved jaw
cordless ultrasonic dissection system
requires the user to read, understand, and
follow the accompanying instructions.
The dissector cannot be adequately cleaned
or sterilized for safe reuse and is, therefore,
intended for single use. Attempts to clean
and sterilize the dissector for reuse may
result in infection or product-failure risks to
the patient and operator.
Visually inspect all system components for
breaks, cracks, nicks, or other damage. Do
not use damaged components. Use of
damaged components may result in injury
to the patient or operator.
Use caution when handling the assembled
device. The assembled device is capable of
creating dissecting energy when the dualmode energy button is depressed. Avoid fire
and unintended burns by always keeping
the distal end of the device in sight and away
from direct contact with adjacent tissue,
textiles, and flammable materials during and
immediately after use.
18
Warning
Infection hazard Non-sterile components
may contaminate the sterile field and place
the patient at risk for infection.
• Do not use the dissector if the sterile
package is opened, damaged, or has
exceeded the expiration date.
• Generators and battery insertion guides
are provided non-sterile and must be
cleaned and sterilized prior to first use
and for each reuse. See Cleaning and
Sterilizing the Battery Insertion Guide and
Generator on page 34.
• Battery packs are non-sterile and cannot
be sterilized. The battery pack must be
aseptically inserted into the dissector
according to instructions. See Inserting the Battery Pack on page 22.
Fire and Burn Hazard The clamping jaw,
active blade, and distal 6 cm of the shaft are
hot during, and for a period of time after
activating the device. Avoid the potential for
fire and unintended burns by always
keeping the distal end of the device in sight
and away from direct contact with adjacent
tissue, textiles, and other flammable
materials during and for a period of time
after activation.
en
Warning
Avoid accidental activation by installing the
generator before installing the battery pack.
Accidental activation of the Sonicision™
curved jaw cordless ultrasonic dissection
device can cause serious injury to the patient
or surgical team.
No modification of this equipment is
allowed.
Caution
Use only Sonicision™ components listed in
the compatibility table on the front cover.
Components from other manufacturers are
not compatible with this system and may
cause injury to the patient and operator.
Do not bend or sharpen the blade. Altering
the blade will compromise the performance
of the device, which may result in patient or
user injury.
Caution
Successful hemostasis may require adjunct
measures when ultrasonic instruments are
used to incise or coagulate solid organs (for
example, liver parenchyma). Due to the
difficulty of visualizing internal structures,
avoid transecting large masses of tissue in a
single activation.
Ensure that the battery pack is sufficiently
charged before use. See Charging a Battery Pack on page 43.
Use of a battery pack that is not fully charged
can result in a procedural delay.
If generators and battery insertion guides
are to be sterilized outside of the medical
facility, ensure packaging protects items
from potential damage and contamination
during transport.
Note: Ensure the availability of backup
equipment to protect against unexpected
system failure and provide additional battery
capacity.
Note: This device is intended to be used with a
5 mm trocar when used endoscopically. Verify
19
en
proper trocar size and compatibility prior to
using the device in a procedure.
Note: Ensure the battery pack connector is
dry. Inserting the battery pack into the device
or the battery charger when the connector is
wet may damage the components.
Note: The Sonicision™ curved jaw cordless
ultrasonic dissection system is intended to be
used in a hospital environment.
Note: Battery packs are shipped with a low
state of charge and must be charged upon
receipt.
Note: To avoid electrostatic-damage to a
battery pack, do not touch the battery pack
connector.
Before Assembly
Battery insertion guides and generators must
be prepared before each use in accordance
with the cleaning and sterilizing instructions.
See Cleaning and Sterilizing the Battery Insertion Guide and Generator on page 34.
20
Assemble the Sonicision™ Curved Jaw
Cordless Ultrasonic Dissection Device
Complete assembly of a working Sonicision™
curved jaw cordless ultrasonic dissection
device requires:
• Single-use dissector
• Reusable generator
• Reusable battery pack
• Reusable battery insertion guide
Assembly
en
1.21.1
1.3
1.51.41.6
21
en
Attaching the Generator
1. Circulating Person: Transfer the dissector,
battery insertion guide(s), and generator
aseptically into the sterile field.
2. Scrubbed Person: Slide the generator into
the opening until it makes contact with the
dissector (step 1.1). Hold the dissector
shaft-rotation wheel. Turn the torque knob
clockwise until it clicks twice (step 1.2).
Inserting the Battery Pack
Note: Attach the generator to the dissector
prior to inserting the battery pack.
1. Scrubbed Person: Hold the handle with
the open battery compartment up. Open
the battery compartment door.
2. Scrubbed Person: Seat the sterile battery
insertion guide onto the open dissector
handle and hold firmly (step 1.3).
3. Circulating Person: Orient the battery
pack as shown with the battery insertion
guide. Insert the battery pack through the
battery insertion guide into the dissector
handle (step 1.4).
4. Circulating Person: After the battery pack
is seated in the dissector handle, remove
22
the contaminated battery insertion guide
from the sterile field (step 1.5).
5. Scrubbed Person: Do not touch the
battery pack. Close the battery
compartment door from the outside until
there is tactile feedback and an audible
click (step 1.6).
A series of audible tones and an
illuminated green LED on the generator
indicate correct assembly.
6. In order to ensure the device meets
required essential performance, it should
be tested after assembly and each battery
exchange as follows:
With the clamping jaws open, test the
minimum-power mode (energy button
depressed to first position) and maximumpower mode (energy button fully
depressed). If the device generates
different pulsating tones for each power
mode and a green LED is present, assembly
is complete and the device is ready for use.
If inaudible or undifferentiated tones or a
non-green LED are present, refer to Device Status and Troubleshooting on page 24.
Selecting a Power Level
The dual-mode energy button activates the
ultrasonic power to the active blade. Two
power settings are available:
• Minimum-power mode
– Energy button depressed to first position
(2)
– A low-pitch, slow pulsating tone
indicates minimum-power activation
– Produces slower dissection when
increased hemostasis is needed
• Maximum-power mode
– Energy button fully depressed (3)
– A high-pitch, fast pulsating tone
indicates maximum-power activation
– Produces faster dissection
Dual Mode Energy Button Positions
1
2
3
Off
ཱMinimum-power mode
ིMaximum-power mode
en
23
en
Device Status and Troubleshooting
The generator LED indicator changes color to
signify the status of the Sonicision™ curved
jaw cordless ultrasonic dissection device. The
LED color in combination with various audible
tones indicate the operational status of the
system. The table below provides an
explanation of system conditions indicated by
a combination of LED colors and emitted
tones. Where applicable, troubleshooting tips
are provided.
LED indicatorToneStatusPossible remedy
Six pulses - start up toneSystem is ready.N/A
Pulsing - low
Green
Pulsing - high
No audio (during assembly
or activation)
Purple and green
alternating
(flashing)
24
Six pulses - start up tone
Note: When performing remedy actions,
follow instructions provided in this user's
guide for maintaining the sterile field.
Note: Test device after each remedy step for
functionality.
Activation of minimum power
(energy button depressed to
first position).
Activation of maximum
power (energy button fully
depressed).
Device is not functional.
Generator is on its last surgical
use.
Replace the dissector.
Press activation button to
N/A
N/A
begin use.
LED indicatorToneStatusPossible remedy
Purple (flashing)Three pulses
Red
(during assembly)
LED not
illuminated
(during assembly)
Yellow (flashing)
Three pulses
No audio
Six pulses - start up tone
Nine pulsesBattery charge is low (< 15%).
Four descending tonesLow-battery reminder.
Twelve pulses followed by a
constant tone
Generator has reached its
maximum use life and is no
longer functional.
System error during assembly.
The device is not functional.
Device is not functional.
Battery pack is depleted. The
device is not functional.
Replace the generator.
1) Fully detach and reattach
the generator.
Ensure two clicks occur
when tightening the
generator.
2) Replace the battery pack.
3) Replace the generator.
4) Replace the dissector.
1) Replace the battery pack.
2) Replace the generator.
1) Replace the battery pack.
2) Replace the generator.
Replace the battery pack.
en
25
en
LED indicatorToneStatusPossible remedy
1) Clean jaw and blade. Test
with open jaw.
Red
(during activation)
Green
(after activation)
Red (during use)No audio
LED not
illuminated
26
Three pulses during
activation
Audio may or may not
occur
System error during use.
Device is not functional.
System error during use.
Device is not functional.
Device is not functional.
2) Fully detach and reattach
the generator.
Ensure two clicks occur
when tightening the
generator.
3) Replace the dissector.
4) Replace the generator.
Replace all components:
dissector, generator, and
battery pack.
1) Replace the battery pack.
2) Replace the dissector.
3) Replace the generator.
Section III – During a Procedure
Warning
The safe and effective use of ultrasonic
energy depends on many factors solely
under the control of the properly trained
clinician.
Contact between the active blade and other
metal objects (for example, hemostats, clips,
staples, and retractors) may result in product
damage, such as a broken blade. Pieces of a
broken blade may fall into the surgical cavity
causing unintended tissue damage.
Applying excessive torque to the jaws while
activating energy on tissue may result in
device damage, such as a broken blade.
en
Warning
Use the device with caution when operating
in close proximity to external or implanted
electronic medical devices. Electromagnetic
interference produced by the proximity of
the ultrasonic generator can potentially
cause a device to enter an unsafe mode.
There is also a possibility that a nearby
electronic device could alter the
performance of the Sonicision™ curved jaw
cordless ultrasonic dissection device.
Consult the manufacturer of the external or
implanted electronic device or responsible
hospital department for further information
when use is planned in close proximity to
such devices.
Fire and Burn Hazard – The clamping jaw,
active blade, and distal 6 cm of the shaft are
hot during and immediately after activating
the device. Avoid the potential for fire and
unintended burns by always keeping the
distal end of the device in sight and away
from direct contact with adjacent tissue,
textiles, and other flammable materials
during and immediately after use.
27
en
Warning
When using the 13 cm device for ENT
procedures, take particular care to avoid
unintended contact with thermally sensitive
tissues such as nerves and the parathyroid
glands. Adjunctive use of a nervemonitoring device is recommended during
nerve-sparing procedures.
Do not activate the instrument with the
clamping jaw closed unless tissue is present.
This could damage the device jaws and
cause injury to the patient.
Infection hazard – Battery insertion guides
and generators must be cleaned and
sterilized before each use in accordance with
the cleaning and sterilizing instructions. See
Cleaning and Sterilizing the Battery Insertion
Guide and Generator on page 34. Non-sterile
battery insertion guides and generators will
contaminate the sterile field and place the
patient at risk for infection.
Examine the vessel for complete hemostasis.
If hemostasis is not achieved, use
appropriate techniques to maintain
hemostasis. Failure to inspect the vessel may
cause serious injury to the patient.
28
Warning
Smoke plume and aerosols generated
during procedures can be potentially
harmful to patients and the perioperative
staff. Use of a smoke evacuation system is
recommended to remove harmful fumes
and particulates produced by heated tissue.
Caution
Close the instrument jaws before inserting or
extracting from a trocar to prevent damage
to the jaws and trocar.
Portable and mobile RF communications
equipment can affect the device and battery
charger. Refer to the Guidance and
Declaration tables for Electromagnetic Compatibility (EMC) on page 53.
Use caution during surgical cases in which
patients exhibit certain types of vascular
pathology (atherosclerosis, aneurysmal
vessels, and so forth). For best results, apply
the energy to unaffected tissue.
Place vessels and tissue in the center of the
jaws to assure optimal hemostatic effect. Do
not place tissue in the jaw hinge.
Caution
Do not apply excessive force to the jaw lever
when grasping tissue. Excessive force
between the clamping jaw and active blade
may result in unintended damage to tissue
and failure of the device to operate as
intended.
Do not release the dual-mode energy button
until tissue is completely coagulated and/or
divided.
The safety and efficacy of ultrasonic devices
has not been established for use on lung
parenchyma.
Keep the jaws clean during use. Buildup of
eschar and tissue may reduce the
effectiveness of the dissection and
coagulation functions, and cause
abnormally high temperatures at the distal
end of the device. During the procedure,
carefully wipe the clamping jaw and active
blade with a wet gauze, or submerse the
distal end in a sterile saline bath and activate
with the clamping jaw open as needed.
Note: This guide and the equipment it
describes are for use only by qualified medical
professionals trained in the particular
en
technique and surgical procedure to be
performed.
29
en
Changing a Battery Pack During a Procedure
30
Warning
Do not use a contaminated battery insertion
guide from the previous assembly. A new
sterile battery insertion guide must be used
for each procedural battery exchange. Refer
to Cleaning and Sterilizing the Battery Insertion Guide and Generator on page 34 for
cleaning and sterilization instructions.
1. Scrubbed Person: Hold the device with
the battery compartment up. Press the
battery door release button to open the
compartment (step 2.1).
2. Scrubbed Person: Place the new sterile
battery insertion guide over the exposed
battery pack and seat it onto the open
dissector handle and hold firmly (step 2.2).
3. Circulating Person: Grasp the battery-
pack handle and pull the battery pack
through the insertion guide out of the
dissector and into a clean, non-sterile
environment (step 2.3).
4. Circulating Person: Orient the new
battery pack as shown with the battery
insertion guide. Insert the battery pack
through the battery insertion guide into
the dissector handle (step 2.4).
en
5. Circulating Person: After the battery pack
is seated in the dissector handle, remove
the contaminated battery insertion guide
from the sterile field (step 2.5).
6. Scrubbed Person: Do not touch the
battery pack. Close the battery
compartment door from the outside until
there is tactile feedback and an audible
click (step 2.6).
7. With the clamping jaws open, test the
minimum-power mode (energy button
depressed to first position) and maximumpower mode (energy button fully
depressed) as described in Assemble the
Sonicision™ Curved Jaw Cordless Ultrasonic
Dissection Device on page 20.
31
en
Section IV – After a Procedure
Disassembly
3.23.1
3.3
32
3.4
en
Warning
Disassembly of the device with these
instructions should only occur after the
procedure is complete and the sterile field is
broken.
A thorough understanding of the principles
and practices of electrical equipment is
essential to avoid shock and hazards to the
operators and/or the device and accessories.
Ensure that electrical isolation and outlet
grounding are not compromised.
The battery charger does not have a main
power switch. During any observed unsafe
operating conditions, immediately
disconnect the battery charger’s AC power
plug from the wall outlet.
If required by local codes, connect the
battery charger’s equipotential post to the
hospital connector with an equipotential
cable. The battery charger’s equipotential
post is located on the back panel.
Follow appropriate safety precautions when
connecting the AC power plug. Ensure
access to the power plug is unobstructed at
all times.
Warning
The battery charger and power cord are nonsterile and cannot be sterilized.
Caution
Take particular care to keep the battery pack
clean.
Remove the Battery Pack
1. Hold the dissector handle with the battery
compartment up. Press the battery door
release button (step 3.1).
2. Use the battery-pack handle to pull the
battery pack out into a clean environment
(step 3.2).
Refer to Cleaning and Disinfecting the Battery Pack on page 41 for cleaning and
charging instructions.
Remove the Generator
1. To loosen the torque knob, hold the shaft-
rotation wheel firmly with one hand while
turning the torque knob counterclockwise
with the other hand (step 3.3).
33
en
2. Continue to turn the torque knob
counterclockwise until the generator can
be removed from the dissector (step 3.4).
After Use
Reprocess the generator and battery insertion
guides as soon as possible following use. If
reprocessing cannot be performed
immediately, cover the components with a
moist towel.
Reusable System Components
Generators and battery insertion guides are
designed to be sterilized up to 150 times. The
battery pack is designed to be used up to 200
times. See Limited Warranty on page 76 for
more information.
34
Cleaning and Sterilizing the Battery
Insertion Guide and Generator
Warning
These instructions have been validated by
Covidien as capable of preparing the
Sonicision™ reusable battery insertion
guides and reusable generators for use. It is
the responsibility of the processor to ensure
that sterilization is performed using
equipment, materials, and trained personnel
that will achieve the desired results. Any
deviation by the processor from the
instructions provided should be properly
evaluated for effectiveness and potential
adverse consequences.
Infection hazard – Non-sterile components
will contaminate the sterile field and place
the patient at risk for infection. Generators
and battery insertion guides are provided
non-sterile and must be cleaned and
sterilized before first use and each reuse.
en
Warning
Use only cleaning detergents specified in
these instructions. Use of cleaning
detergents not specified in these
instructions will compromise sterility and
damage components.
Thoroughly clean the battery insertion guide
and generator. Attempting to sterilize
contaminated components will compromise
sterility.
Use only the sterilization method that is
recommended and qualified as described in
these instructions. Do not use immediate-
use (flash) steam sterilization, hydrogen
peroxide gas plasma technology (such as
STERRAD™* systems), ethylene oxide
(EO), or gamma sterilization to sterilize
any components of the Sonicision™
system. Exposure to these alternate forms
of sterilization will damage the
equipment and compromise sterility.
Clean and sterilize the reusable generator
and the reusable battery insertion guide
before each use. Reprocess as soon as
possible following each use. If processing
cannot be completed immediately, cover the
components with a moist towel.
Warning
Battery insertion guides and generators that
are to be stored before sterilization must be
cleaned before storage. Storing
contaminated battery insertion guides and
generators may compromise their ability to
be sterilized.
Caution
Do not use abrasive cleaning detergents as
they may scratch the surface of the
generator and battery insertion guide.
The cleaning and sterilization processes
described in this section have been validated
by Covidien as capable of preparing
Sonicision™ reusable generators and
Sonicision™ reusable battery insertion guides.
Sterility of the final product relies on the use of
properly qualified personnel, equipment, and
tools.
Cleaning
Use a cleaning solution of pH-neutral, pHneutral enzymatic, or mild alkaline detergent
to clean the generator and battery insertion
guides.
35
en
Preparation for Cleaning - Gross
Decontamination
1. Visually inspect the battery insertion
guides and generator. Do not use the
components if they are broken, cracked,
nicked, or have any other damage.
Replace damaged components.
2. Under warm (32–40°C, 90–104°F) running
tap water scrub all components with a
small, soft, nylon-bristle, general
instrument- cleaning brush to remove
surgical debris.
For hard to reach crevices, use a 4.0 mm
soft nylon-bristle brush for the battery
insertion guide, and use a 12.0 mm soft
nylon-bristle brush for the generator.
3. Rinse under warm (32–40°C, 90–104°F)
running tap water for a minimum of 1
minute.
4. Prepare a cleaning solution according to
the manufacturer’s instructions.
Note: The battery insertion guides and the
generator are tested for material compatibility
with cleaners that have a pH range from
neutral to 10.9.
36
Choose Manual Cleaning or Automatic
Cleaning.
Manual Cleaning
1. Immerse the battery insertion guides and
generator in a warm (32–40°C, 90–104°F)
detergent bath for 10 minutes.
2. Use a small, soft, nylon-bristle, general
instrument-cleaning brush to scrub each
component in the detergent bath for a
minimum of 1 minute.
For hard to reach crevices, use a 4.0 mm
soft nylon-bristle brush for the battery
insertion guide, and use a 12.0 mm soft
nylon-bristle brush for the generator.
3. Rinse each component under warm
(32–40°C, 90–104°F) running tap water for
a minimum of 1 minute.
4. Perform a final rinse under warm purified
water for a minimum of 1 minute for each
component.
5. Use a clean, soft, lint-free cloth to dry the
components.
Automatic Cleaning
Run the automatic cleaning cycle. Refer to the
following Automatic Cleaning Recommendations table on page 38.
Caution
To avoid mechanical damage, arrange the
contents of the washer so that the generator
and battery insertion guides will not contact
other items during the wash cycle.
Do not wash the generator in an ultrasonic
washer.
en
37
en
Automatic Cleaning Recommendations
TreatmentTime (MM:SS)TemperatureChemical
Pre-wash01:00Cold tap waterN/A
Wash03:00Hot tap waterpH-neutral, pH-neutral enzymatic, or
Rinse00:15Hot tap waterN/A
Wash02:00Hot tap waterpH-neutral, pH-neutral enzymatic, or
Rinse00:15Hot tap waterN/A
Thermal rinse01:00Heated purified water
Dry06:00High setting 95°C
Note: The components have been qualified to withstand the following parameters: Pre-wash of 4 minutes,
total wash of 10 minutes, hot tap water rinse of 5 minutes, thermal rinse of 5 minutes at 95°C (203°F), and dry
time of 30 minutes.
82°C (180°F)
(203°F)
mild alkaline detergent diluted per the
manufacturer’s specifications
mild alkaline detergent diluted per the
manufacturer’s specifications
N/A
N/A
Inspection
Visually inspect the battery insertion guides
and generator for cleanliness. Repeat the
cleaning cycle if necessary.
38
Pre-Sterilization Wipe Down
(Optional)
In facilities that require additional preparation
prior to sterilization, wipe the surfaces of the
battery insertion guides and generator with an
isopropyl alcohol disinfectant (70%).
Sterilizing the Reusable Generator and
the Battery Insertion Guides
Caution
Ensure that the selected sterile barrier and
process indicators have been qualified for
the sterilization exposure temperature and
time that will be used. The minimum
sterilization temperatures and times in each
row of the recommended cycle parameters
have been validated for achieving product
sterility, and the Sonicision™ products have
been qualified to withstand repeated
exposure to the maximum temperatures and
times listed. It is the responsibility of the
user to select the appropriate sterile barrier
and process indicators for each sterilization
cycle.
en
Warning
The following list of sterilization barriers
have been validated for effective sterilization
of the generator and battery insertion
guides. Do not use alternative sterilization
barriers or additional absorbent or
protective materials as this may
compromise sterility or functionality of
the components.
1. After cleaning, use the manufacturer’s
instructions for the sterile barrier selected:
• Single Polypropylene Wrap - Wrap
each component individually in a
compatible polypropylene sterilization
wrap.
• Double Polypropylene Wrap - Wrap
each component individually in two
compatible polypropylene sterilization
wraps.
• Single Breathable Polyethylene Pouch - Place each component
individually in its own single compatible
breathable polyethylene pouch and
seal. Place the pouch in the autoclave
with the breathable side of the pouch
facing up.
39
en
• Double Breathable Polyethylene
Pouch - Place each component
individually within two compatible
breathable polyethylene pouches, and
seal. The breathable side of the inside
pouch must be aligned with the
breathable side of the outside pouch.
Place the pouches in the autoclave with
the breathable sides of the pouches
facing up.
• Rigid Steam Sterilization Container*
Place the components in a compatible
rigid steam sterilization container.
Ensure that the individual components
are not touching, and the container is
loaded according to manufacturer’s
recommendations.
*Validated with an Aesculap™* rigid
container.
2. Place the generators and battery insertion
guides into the autoclave. Do not stack the
generators and battery insertion guides.
Note: When sterilizing multiple
instruments in one autoclave cycle, do not
exceed the maximum load specified by the
manufacturer.
3. Select one of the sterilization cycles in the
following Steam Sterilization Cycles table.
40
Warning
Generators and battery insertion guides can
not be sterilized with rigid containers using a
steam sterilization cycle of Gravity 132°C
(270°F).
en
Steam Sterilization Cycles
Cycle typeExposure
Pre-vacuum132°C (270°F)420–40Wrap(s), pouch(es), and rigid container
Pre-vacuum135°C (275°F)316–40Wrap(s), pouch(es), and rigid container
Gravity132°C (270°F)1515–40Wrap(s) and pouch(es)
Gravity135°C (275°F)1030 Wrap(s), pouch(es), and rigid container
Note: The generator and battery insertion guide have been tested to allow exposure to temperatures up to
135°C (275°F) for 18 minutes, in order to allow for variations in local requirements and regulations for
sterilization protocols.
temperature
Exposure
time
(minutes)
Dry time
(minutes)
Compatible sterile barriers
Cleaning and Disinfecting the Battery
Pack
Warning
Do not immerse the battery pack.
Do not disassemble or modify the battery
pack.
Inspect the battery pack for damage before
use. Do not use a damaged battery pack.
Warning
The battery pack is non-sterile, and cannot
be sterilized. The battery pack will be
damaged if sterilization is attempted.
Do not rinse the battery pack under running
water or submerge the battery pack in water.
Do not use instrument lubricant on the
battery pack.
41
en
Caution
The battery pack should be cleaned and
disinfected after each use and before
charging. Clean and disinfect the battery
pack as instructed.
Note: Do not expose the battery pack
connector to excessive moisture, conductive
materials, or contaminants.
Cleaning the Battery Pack
1. Moisten a lint-free cloth with a disinfectant
solution or use a disinfectant wipe.
Use one of the following disinfectants:
• Ethyl alcohol or isopropyl alcohol (70%)
• Hydrogen peroxide wipe (for example,
TM*
Oxivir
)
2. Using a clean disinfectant wipe, wipe all
external surfaces of the battery pack until
visually clean. Ensure the surface remains
thoroughly wet by continuously wiping for
the contact time recommended per the
manufacturer's instructions.
• If using 70% ethyl alcohol or isopropyl
alcohol, ensure a contact time of 1
minute.
42
• Pay special attention to the battery pack
handle.
• Avoid contact with the battery pack
connector.
• Use additional wipes if needed.
3. Use a lint-free cloth to dry all exposed
surfaces of the battery pack, taking care to
avoid damage to the electrical contacts.
4. Visually inspect the battery pack. If it is not
clean, repeat cleaning steps above.
Disinfecting the Battery Pack
1. Moisten a lint-free cloth with a disinfectant
solution or use a disinfectant wipe.
Use one of the following disinfectants:
• Ethyl alcohol or isopropyl alcohol (70%)
• Hydrogen peroxide wipe (for example,
TM*
Oxivir
)
2. Using a clean disinfectant wipe, wipe all
external surfaces of the battery pack.
Ensure the surface remains thoroughly wet
by continuously wiping for the contact
time recommended per the
manufacturer's instructions.
• If using 70% ethyl alcohol or isopropyl
alcohol, ensure a contact time of 1
minute.
• Pay special attention to the battery pack
handle.
• Avoid contact with the battery pack
connector.
• Use additional wipes if needed.
3. Use a lint-free cloth to dry all exposed
surfaces of the battery pack, taking care to
avoid damage to the electrical contacts.
Caution
Store the battery charger in a clean dry place
to avoid contamination from blood or other
body fluids.
Battery packs must be cleaned and
disinfected before being placed on the
battery charger to avoid potential
contamination of the charging bay.
Charging a Battery Pack
1. Connect the power cord to the battery
charger.
en
2. Connect the power cord to a grounded
electrical outlet. The green power-status
light will illuminate when the battery
charger has power.
3. Place the battery pack into a charging bay
and ensure the battery pack is fully seated.
The charging cycle will start automatically
when the battery pack is inserted correctly
into the battery charger. Refer to Charging Indications and Troubleshooting on page 44
for more information.
Charging cycle time depends on the
amount of charge remaining in the battery
pack. An empty battery pack typically takes
3 hours to reach a full charge.
43
en
4. To remove a battery pack from a battery
charger bay, pull the battery pack straight
up.
Charging Indications and
Troubleshooting
Flashing green bars—Active
charging. Five solid bars indicate a
full charge.
Flashing green bars with a
number—The white number shows
the surgical uses remaining. This
number is present when there are 20
or fewer surgical uses remaining for
the battery pack.
White battery with “0”—The
battery pack has exceeded the
number of surgical uses and has
reached its end-of-life. The battery
pack must be replaced. Dispose of
the old battery pack as required by
local, state, and government
regulations.
44
Red battery—The battery pack has a
fault and is non-functional.
Troubleshooting— Reinsert the
battery pack into a different charger
bay. If the error indication remains,
then the battery pack is nonfunctional and must be replaced.
Contact Covidien Customer Service.
Yellow caution—The charging bay
has a fault and is non-functional.
Troubleshooting— Unplug and replug the battery-charger power cord.
If the error indication remains, that
battery charger bay is nonfunctional.
Contact Covidien Customer Service.
Yellow caution on all displays and
red power-status light—the
battery charger has a system fault
and is non-functional.
Y
Troubleshooting— Unplug and replug the battery-charger power cord.
If the error indication remains, then
the battery charger is nonfunctional.
Contact Covidien Customer Service.
Battery Pack Storage
The service life of the battery pack is
monitored by the battery charger. Store the
battery pack in the battery charger between
uses. If the battery pack is stored off of the
charger, the battery pack should be returned
to the charger before use to verify
functionality and state of charge, and recharge
as necessary. See Environmental Parameters on
page 46.
Optional Maintenance of the Battery
Charger
Warning
Electric shock hazard – Always unplug the
battery charger before cleaning.
Do not immerse the battery charger.
Caution
Do not rinse the battery charger under
running water or submerge the battery
charger in water.
Do not use instrument lubricant on the
battery charger.
Caution
Do not sterilize the battery packs or the
battery charger. The sterilization process will
damage these components.
Do not spray or pour cleaning solutions or
other liquids directly onto the battery
charger or battery charger bays. Product
damage may result.
1. Remove the battery charger power cord
from the wall outlet.
2. Wipe down all external surfaces of the
battery charger with a water-dampened
lint-free cloth to completely remove any
gross debris or dust.
3. The battery charger may also be wiped
with a hydrogen peroxide based wipe such
TM*
as Oxivir
or ethyl or isopropyl alcohol
(70%).
Avoid contact with the battery charger
connectors.
4. Use a lint free cloth to dry all exposed
surfaces of the battery charger, taking care
to avoid damage to the connectors.
en
45
en
Caution
Ensure that the battery charger is completely
dry before connecting to power or inserting
a battery pack to charge.
Environmental Parameters
Operation and Storage
Ambient
temperature
range
Relative
humidity
Atmospheric
pressure
46
System
(except
Battery
Charger)
Battery
Charger
15% to 85% noncondensing
700 millibars to 1060
millibars
10°C to 30°C
(50°F to 86°F)
15°C to 30°C
(59°F to 86°F)
Transport
Ambient
temperature
range
Relative
humidity
-30°C to 60°C (-22°F to
140°F)
15% to 90% noncondensing
Note: Store battery packs at 18°C to 25°C (64°F
to 77°F) in a dry, ventilated area to reduce
possibility of lowered battery-service life from
elevated temperatures.
Note: If the battery pack is stored off of the
charger, the battery pack should be returned
to the charger before use to verify
functionality, state of charge, and recharge as
necessary.
Recycling and Disposal
Curved Jaw Cordless Ultrasonic Dissector
Discard biologically contaminated single-use
electronic devices in accordance with your
institution's hazardous medical waste
procedures and local regulatory requirements.
Reusable Battery Pack
Covidien is committed to a healthy world and
recommends that Sonicision™ reusable
battery packs be recycled following
decontamination.
Within the U.S. and Canada
Covidien has provided funding to support the
recycling of Sonicision™ reusable battery
packs. Visit www.call2recycle.org for more
information and to receive a prepaid
collection kit.
Outside the U.S. and Canada
Covidien recommends that Sonicision™
reusable battery packs be recycled in
accordance with your local laws, rules and
national regulations.
en
Caution
Failure to properly recycle or discard the
single-use or reusable system components
may lead to environmental contamination.
Reusable Generator, Reusable Battery
Insertion Guide, and Battery Charger
Recycle or discard decontaminated reusable
system components in accordance with your
institution's waste procedures and local
regulatory requirements.
Residual Risk Summary
While every attempt has been made to reduce
patient and user risks, all surgeries using this
system carry some residual risk, even when
used by trained physicians. It is also to be
noted that some of the listed risks are inherent
in any surgical procedure and are not unique
to the system. The potential adverse events
associated with the use of the ultrasonic
dissection system include, but are not limited
to, the following risks:
• Arrhythmia or cardiac arrest (interaction
with implanted electronic device)
• Bleeding
• Burn (thermal, external, bowel or to critical
ENT structures)
• Carcinogen exposure (surgical plume)
• Chemical exposure (damaged battery)
• Electric shock
• Foreign body in patient
• Foreign body reaction
• Infection
• Nerve damage
• Perforation (vessel or organ)
• Thromboembolism
• Tissue damage
Note: Any serious incident that has occurred
in relation to the device should be reported to
the manufacturer and necessary regulatory
authority (if European Union, the competent
authority of the Member State) in which the
user and/or patient is established.
48
Section V – Pre-Clinical Studies
Note: There is no animal data qualified to
predict the effectiveness of this device in
sealing vessels containing atherosclerotic
plaque. Therefore, this device should not be
used on vessels containing atherosclerotic
plaque.
Clearance of the Sonicision™ curved jaw
cordless ultrasonic dissection device in the
United States was not based on human clinical
testing; instead, the device was investigated in
a chronic in vivo porcine animal model. The
results showed that no animals studied
experienced any hemostatic complications
related to the device during the 21-day
survival period. A variety of tissue types and
vessels were evaluated (shown below).
Vessel typeVessel location Vessel size range
A/V bundleOmentum < 1.0 mm
Gastrosplenic3.0 mm – 3.5 mm; up to 10.0 mm bundle
Short gastric bundleup to 6.5 mm bundle
L/R ovarian bundlebundles up to 4.0 mm
L/R broad ligamentbundles up to 3.0 mm
ArterySplenic artery4.5 mm – 5.0 mm
Left renal artery3.0 mm – 5.0 mm
Note: Complete study data on file at Covidien.
en
49
en
Section VI – Technical Description
All specifications are nominal and subject to change due to variability. A specification referred to
as “typical” is within ± 20% of a stated value at room temperature (25°C/77°F) and a nominal
input voltage.
Product Codes
The assembled Sonicision™ curved jaw cordless ultrasonic dissector, reusable generator, and
reusable battery pack are referred to as the Sonicision™ curved jaw cordless ultrasonic dissection
device in this document. The Sonicision™ curved jaw cordless ultrasonic dissection system refers
to all components, including the assembled device, battery insertion guide, and battery charger.
SCDA13 Sonicision™ Curved Jaw Cordless Ultrasonic Dissector 5 mm – 13 cm
SCDA26 Sonicision™ Curved Jaw Cordless Ultrasonic Dissector 5 mm – 26 cm
SCDA39 Sonicision™ Curved Jaw Cordless Ultrasonic Dissector 5 mm – 39 cm
SCDA48 Sonicision™ Curved Jaw Cordless Ultrasonic Dissector 5 mm – 48 cm
SCDA1316.7 cm x 3.9 cm x 25.8 cm (13 cm shaft)0.15 kg
SCDA2616.7 cm x 3.9 cm x 38.9 cm (26 cm shaft)0.17 kg
SCDA3916.7 cm x 3.9 cm x 51.1 cm (39 cm shaft)0.18 kg
SCDA4816.7 cm x 3.9 cm x 60.7 cm (48 cm shaft)0.19 kg
•Active blade length14.5 mm
Reusable generator5.7 cm x 3.9 cm x 11.7 cm0.16 kg
Reusable battery pack9.5 cm x 2.5 cm x 4.1 cm0.14 kg
Battery charger29.0 cm x 13.0 cm x 10.2 cm1.13 kg
Battery insertion guide3.6 cm x 18.9 cm x 5.5 cm0.06 kg
en
51
en
ComponentLengthWeight
Battery charger cord3.0 m 0.34 kg
Standards
The Sonicision™ curved jaw cordless ultrasonic dissection system conforms to all applicable
clauses of AAMI ES60601-1:2012, IEC 60601-1:2012 (Ed. 3.1), IEC 60601-1-6:2013, IEC 60601-12:2014, and CAN/CSA C22.2#60601-1*BEI Issued:2014/02/01 Ed:3 Medical Electrical Equipment Part 1.
In addition, the Sonicision™ reusable battery pack conforms to IEC 62133:2012, IEC 62133-2:2017,
IEC 62281:2004, and UN DOT 38.3 (Ed. 5).
Cleaning and sterilization validations of Sonicision™ reusable components conform to ISO
17664:2004, ISO 17665:2013, AAMI TIR12:2010, AAMI TIR30:2011, AAMI ST77:2013, and AAMI
ST81:2010.
52
en
Electromagnetic Compatibility (EMC)
Electromagnetic Compatibility (IEC 60601-1-2)
The Sonicision™ curved jaw cordless ultrasonic dissection device and battery charger comply
with IEC 60601-1-2 specifications regarding electromagnetic compatibility.
The use of accessories other than those specified in the user’s guide and instructions for use may
result in increased emissions or decreased immunity of the Sonicision™ curved jaw cordless
ultrasonic dissection device and battery charger.
Medical equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect medical electrical equipment.
The Sonicision™ curved jaw cordless ultrasonic dissection device and battery charger should not
be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary,
the Sonicision™ curved jaw cordless ultrasonic dissection device and battery charger should be
observed to verify normal operation in the configuration in which they will be used.
Refer to the Guidance and Declaration tables for compliance levels of IEC 60601-1-2 3rd Ed.
(2007).
53
en
Guidance and manufacturer’s declaration - electromagnetic emissions
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment specified below. The customer or the user of the Sonicision™ curved
jaw cordless ultrasonic dissection device should assure that it is used in such an environment.
Group 1The Sonicision™ curved jaw cordless ultrasonic
Class AThe Sonicision™ curved jaw cordless ultrasonic
Not applicable
Not applicable
dissection device uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
dissection device is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment specified below. The customer or the user of the Sonicision™ curved
jaw cordless ultrasonic dissection device should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance level Electromagnetic
Electrostatic discharge
(ESD)
IEC 61000-4-2± 8 kV air± 8 kV air
Electrical fast transient/
burst
IEC 61000-4-4
± 6 kV contact± 6 kV contactFloors should be wood,
± 2 kV for power supply lines
± 1 kV for input/output lines
(Continued)
Does not apply as
the Sonicision™
curved jaw cordless
ultrasonic
dissection device is
battery powered
and has no signal
leads.
environment guidance
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Not applicable
55
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment specified below. The customer or the user of the Sonicision™ curved
jaw cordless ultrasonic dissection device should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance level Electromagnetic
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
<5% U
T
Does not apply as
the Sonicision™
curved jaw cordless
ultrasonic
dissection device is
battery powered.
(>95% dip in UT) for 0.5 cycle
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
IEC 61000-4-11
40% U
T
(60% dip in UT) for 5 cycles
70% U
T
(30% dip in UT) for 25 cycles
<5% U
T
Does not apply as
the Sonicision™
curved jaw cordless
ultrasonic
dissection device is
battery powered
and has no signal
leads.
(>95% dip in UT) for 5 sec
(Continued)
56
environment guidance
Not applicable
Not applicable
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment specified below. The customer or the user of the Sonicision™ curved
jaw cordless ultrasonic dissection device should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance level Electromagnetic
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
Note: U
is the mains voltage prior to application of the test level.
T
3 A/m3 A/mPower frequency
environment guidance
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
57
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment specified below. The customer or the user of the Sonicision™ curved
jaw cordless ultrasonic dissection device should assure that it is used in such an environment.
Immunity test IEC 60601 test
Conducted RF
IEC 61000-4-6
level
3 Vrms
150 kHz to 80
MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
58
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
Not applicable
should be used no closer to any part of the
Sonicision™ curved jaw cordless ultrasonic dissection
device, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
Not Applicable
3 V/md=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
b
.
(Continued)
a
,
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment specified below. The customer or the user of the Sonicision™ curved
jaw cordless ultrasonic dissection device should assure that it is used in such an environment.
Interference may occur in the vicinity of equipment
marked with the following symbol.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Sonicision™ curved jaw cordless ultrasonic dissection device is used exceeds the applicable RF compliance
level above, the Sonicision™ curved jaw cordless ultrasonic dissection device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Sonicision™ curved jaw cordless ultrasonic dissection device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 v/m.
59
en
Recommended separation distances between portable and mobile RF communication equipment
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the Sonicision™ curved jaw cordless ultrasonic dissection device can help prevent
electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Sonicision™ curved jaw cordless ultrasonic
dissection device as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter (W)
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
60
and the Sonicision™ curved jaw cordless ultrasonic dissection device
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz80 MHz to 800 MHz
d=1.2√P
0.010.12 m0.12 m0.23 m
0.10.37 m0.37 m0.74 m
11.2 m1.2 m2.3 m
103.7 m3.7 m7.4 m
10012 m12 m23 m
(Continued)
800 MHz to 2.5 GHz
d=2.3√P
en
Recommended separation distances between portable and mobile RF communication equipment
The Sonicision™ curved jaw cordless ultrasonic dissection device is intended for use in the
electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the Sonicision™ curved jaw cordless ultrasonic dissection device can help prevent
electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Sonicision™ curved jaw cordless ultrasonic
dissection device as recommended below, according to the maximum output power of the
communications equipment.
Note 1: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
and the Sonicision™ curved jaw cordless ultrasonic dissection device
Guidance and manufacturer’s declaration - electromagnetic emissions
Group 1The battery charger uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
(Continued)
61
en
Guidance and manufacturer’s declaration - electromagnetic emissions
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
Guidance and manufacturer’s declaration - electromagnetic immunity
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
Immunity testIEC 60601 test
Electrostatic discharge
(ESD)
IEC 61000-4-2± 8 kV air± 8 kV air
62
Class AThe battery charger is suitable for use in all
establishments other than domestic and those
Class A
Complies
level
± 6 kV contact± 6 kV contactFloors should be wood, concrete, or
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Compliance
level
(Continued)
Electromagnetic environment -
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
guidance
Guidance and manufacturer’s declaration - electromagnetic immunity
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
Immunity testIEC 60601 test
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV differential
mode
± 2 kV common
mode
level
Compliance
level
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV
differential
mode
± 2 kV common
mode
(Continued)
Electromagnetic environment -
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
guidance
en
63
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
Immunity testIEC 60601 test
<5% U
level
T
(>95% dip in UT)
for 0.5 cycle
40% U
T
Voltage dips, short
interruptions, and voltage
(60% dip in UT) for
5 cycles
variations on power supply
input lines
IEC 61000-4-11
70% U
T
(30% dip in UT) for
25 cycles
<5% U
T
(>95% dip in UT)
for 5 sec
Power frequency (50/
3 A/m3 A/mPower frequency magnetic fields should
60 Hz) magnetic field
IEC 61000-4-8
Note: U
is the mains voltage prior to application of the test level.
T
64
Compliance
level
<5% U
T
(>95% dip in UT)
for 0.5 cycle
40% U
T
(60% dip in UT)
for 5 cycles
70% U
T
(30% dip in UT)
for 25 cycles
<5% U
T
(>95% dip in UT)
for 5 sec
Electromagnetic environment -
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the battery
charger requires continued operation
during power mains interruptions, it is
recommended that the battery charger be
powered from an uninterruptible power
supply or a battery pack.
guidance
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Guidance and manufacturer’s declaration - electromagnetic immunity
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Vrms
3 V/m
should be used no closer to any part of the battery
charger, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d=1.2√P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
the compliance level in each frequency range
a
, should be less than
b
.
(Continued)
en
65
en
Guidance and manufacturer’s declaration - electromagnetic immunity
The battery charger is intended for use in the electromagnetic environment specified below. The
customer or the user of the battery charger should assure that it is used in such an environment.
Interference may occur in the vicinity of equipment
marked with the following symbol.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
battery charger is used exceeds the applicable RF compliance level above, the battery charger should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the battery charger.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
66
en
Recommended separation distances between portable and mobile RF communication equipment
The battery charger is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the battery charger can help prevent
electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the battery charger as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter (W)
0.010.12 m0.12 m0.23 m
0.10.37 m0.37 m0.74 m
11.2 m1.2 m2.3 m
103.7 m3.7 m7.4 m
10012 m12 m23 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
150 kHz to 80 MHz
and the battery charger
Separation distance according to frequency of transmitter (m)
d=1.2√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to 2.5 GHz
d=2.3√P
67
en
Recommended separation distances between portable and mobile RF communication equipment
Frequency ranges and text condition on transmitter bands
a. Pulse modulation is defined as a square wave input with a 50% duty cycle at the listed frequency.
Note 1: The minimum dwell time during test is 8 seconds.
Note 2: A minimum separation distance of 0.3 meters should be maintained between any device
transmitting in this band and the battery charger. This includes devices such as mobile phones, PDAs,
Wireless LANs, RFID, and Bluetooth™*.
The Sonicision™ curved jaw cordless ultrasonic dissection device and battery charger provide a high
degree of protection against electric shock, particularly regarding allowable leakage currents.
Accessible conductive parts of the battery charger cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth conductor. The
Sonicision™ curved jaw cordless ultrasonic dissection device is type BF isolated (floating) output but it is
not intended for direct cardiac application.
Internally Powered Equipment; Defib-Proof Type BF Applied Part
69
en
Enclosure Ingress Protection (IP Classification)
ComponentClassification
Assembled
device
Battery
charger
Input Power
Battery charger
100–120 VAC, 50 Hz/60 Hz ± 10%, Mains current 2 A maximum
220–240 VAC, 50 Hz/60 Hz ± 10%, Mains current 1 A maximum
IP2XThe battery charger is constructed to prevent finger contact with
dissection device is constructed so that vertically falling drops shall
have no harmful effects when the device enclosure is tilted at any
angle up to 15 degrees on either side of the vertical.
hazardous parts within the charger, and penetration of an
enclosure by a solid foreign object with a diameter equal to or
greater than 12.5 mm.
Reusable Battery Pack
Battery-pack SKU SCBA
Battery-pack typeRechargeable lithium-ion
Mode of insertion Manual
Number of cells2
Cell voltage3.6 VDC nominal
Battery-pack voltage7.2 VDC
Discharge rateMaximum 10 A
Battery-pack capacitySee Sonicision™ reusable battery pack label for battery rating
Watt-hour ratingSee Sonicision™ reusable battery pack label for battery rating
Battery-charge time (typical)180 minutes or less
Battery-pack lifeDesigned to be used up to 200 times.
Battery storageStore battery packs at 18–25°C (64–77°F) in a dry, ventilated area.
information
information
Elevated temperatures can result in reduced battery-service life. If
the battery pack is stored off of the charger, the battery pack should
be returned to the charger before use to verify functionality, state of
charge, and recharge as necessary.
en
71
en
Duty Cycle (Mode of Operation)
Caution
Use of a duty cycle no greater than 33% (3 seconds on, 5 seconds off and a rest period of 1
minute after 60 activations). Duty cycles above 33% increase the risk of overheating, increasing
the risk of injury to the operator and failure of the device to operate as intended.
Use of the above mentioned duty cycle for an expected use of 300 activations in a procedure
may result in temperatures up to and including 46.2°C (115°F) along the proximal portion of the
shaft up to 6 cm from the distal end. Further distal of this point is considered an applied part
intended to supply heat and will generate higher temperatures to achieve clinical tissue effect.
A temperature of 46.2°C (115°F) is within the safe maximum touch temperature range as
specified by IEC 60601-1 (2012/R, 2014) Edition 3.1 Table 24.
The battery charger is equipped with a power cord (unshielded). Should the AC power cord need
to be replaced, the replacement plug/cable/receptacle must meet or exceed the following
specifications:
100–120 VAC
Electrical ratings125 VAC, 10 A
CableSJT18/3, maximum length 3.05 m (10 ft)
Unit receptacleIEC female
220–240 VAC
Electrical ratings250 VAC, 6 A
CableH05VVF3G1.0 VDE, maximum length 4.57 m (15 ft)
Unit receptacleIEC female
73
en
Equipotential Ground Connection
An equipotential ground connection is provided to allow optional connection of the battery
charger to ground.
USB Connection
The USB port on the back of the battery charger is for factory use only.
Alert Tones
The audio levels stated below represent activation and system-alert tones at a distance of 38-51
cm. (Refer to Device Status and Troubleshooting on page 24 for additional information on alert
tones.)
Volume (non-adjustable)50–80 dBA
Frequency250–1150 Hz
Duration50–750 ms
74
Battery Charger Fuses
Type5 mm x 20 mm
Voltage250 VAC
Current2.5 A
Operating speedFast blow
Interrupt rating35 A at 250 VAC
en
75
en
Limited Warranty
Covidien warrants each covered product listed below to be free from defects in material and
workmanship for normal use as specifically described herein. Covidien’s obligation under this
limited warranty is limited to the repair or replacement, at its sole option, of any product, or part
thereof, which has been returned to it (or its authorized distributor) within the applicable time
period shown below after delivery of the product to the original purchaser, and which
examination discloses, to Covidien’s satisfaction, that the product does not conform to this
warranty. This limited warranty does not apply to any product, or part thereof, which has been
repaired, or altered in a way so as, in Covidien’s judgment, to affect its stability or reliability. In
addition, this limited warranty does not cover misuse, neglect, or accident, such as but not
limited to: improper cleaning or sterilizing (for example, ultrasonic cleaner, immediate-use (flash)
steam sterilization, hydrogen peroxide, ethylene oxide [EO]), improper handling (for example,
dropping), non-intended use (for example, incising bone, contraceptive tubal occlusion),
refurbishing, or attempted reuse of single use components.
The warranty periods for the Sonicision™ curved jaw cordless ultrasonic dissection system is as
follows:
One year from date of shipment, 150 sterilization
cycles, or 150 surgical uses (whichever comes first)
One year from date of shipment, or 200 surgical uses
(whichever comes first)
One year from date of shipment
Sterility only as stated on packaging
One year from date of shipment or 150 sterilization
cycles (whichever comes first)
en
Notwithstanding any other provision herein or in any other document or communication,
Covidien’s liability with respect to this limited warranty and the products sold hereunder shall be
limited to the aggregate purchase price for the products sold to the customer. This limited
warranty is non-transferable and runs only to the original purchaser of the covered product(s).
There are no warranties which extend beyond the terms hereof. Covidien disclaims any liability
hereunder or elsewhere in connection with the sale of products and for any form of indirect, tort,
or consequential damages.
For purposes of this limited warranty, Covidien shall consider a sterilization cycle to be following
the steps contained in the Sterilizing Section of this user's guide.
This limited warranty and the rights and obligations hereunder shall be construed under and
governed by the laws of the State of Colorado, USA. The sole forum for resolving disputes arising
under or relating in any way to this limited warranty is the District Court of the County of Boulder,
State of Colorado, USA.
Covidien reserves the right to make changes in covered products built or sold by it at any time
without incurring any obligation to make the same or similar changes to equipment previously
built or sold by it.
THE OBLIGATION TO REPAIR OR REPLACE A DEFECTIVE OR NONPERFORMING PRODUCT IS THE
SOLE REMEDY OF THE CUSTOMER UNDER THIS LIMITED WARRANTY. EXCEPT AS EXPRESSLY
PROVIDED HEREIN, COVIDIEN DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS
OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO PRODUCTS, INCLUDING WITHOUT
LIMITATION ALL IMPLIED WARRANTIES, WARRANTIES OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE.
77
en
Technical Support
Within the U.S.+1 800 255 8522
Outside the U.S.To contact your local Covidien support, go to
78
http://www.medtronic.com/covidien/support/service-centers
or call +1 303 476 7996
California Proposition 65 Statement
For information regarding California Proposition 65, please refer to www.covidien.com/
caprop65.
China RoHS Compliance
Hazardous Substance Table
en
(Part Name)
(Pb)(Hg)(Cd)(CR(VI))
SCGAA
XOO OOO
SCDAXX
OOO OOO
SCBA
XOO OXX
CBCA
XOO OXX
(Toxic and Hazardous Substances or Elements)
(PBB)
(PBDE)
O:
GB/T 26572.
79
en
O: indicates that the content of the toxic and hazardous substance in all the Homogenous
Materials of the part is below the concentration limit requirement as described in GB/T 26572.
X:
GB/T 26572 ( SJ/T 11364
X: indicates that the content of the toxic and hazardous substance in at least one Homogenous
Material of the part exceeds the concentration limit requirement as described in GB/T 26572.
(Information provided in this table is solely for compliance with SJ/T 11364 and the Management
Measures for the Control of Pollution by Electronic Information Products.)
80
)
Not made with
natural rubber latex
Consult
Instructions
for Use
Caution
ÑPT00101127_Ó
Part No. PT00101127
COVIDIEN, COVIDIEN with logo, and Covidien logo
and Positive Results for Life are U.S. and
internationally registered trademarks of Covidien
AG. ™* brands are trademarks of their respective
owners. Other brands are trademarks of a
Covidien company.
May be covered by U.S. Patents:
www.covidien.com/patents