Medtronic BM130 Instructions for Use

MIDAS REX™ ELECTRIC BONE MILL BM110 & BM120
User’s Guide

Table of Contents

Symbols .................................................................................................................................................................... 3
When The System Arrives .................................................................................................................................... 3
Warnings and Precautions .................................................................................................................................. 3
System Description ................................................................................................................................................ 5
Indications .................................................................................................................................................... 5
Contraindications ...................................................................................................................................... 5
Setup And Operation ............................................................................................................................................ 5
Operative Procedure ................................................................................................................................. 5
Post-Operative Procedure ....................................................................................................................... 6
Cleaning .................................................................................................................................................................... 6
Sterilization ............................................................................................................................................................... 6
Limited Warranty .................................................................................................................................................... 7
Service ........................................................................................................................................................................ 7
Troubleshooting ..................................................................................................................................................... 7
Technical Specications ....................................................................................................................................... 8
Midas Rex™ Electric Bone Mill Components ............................................................................................... 8
Electromagnetic Compatability ........................................................................................................................ 9
It is important that you read this entire manual before you use the Midas Rex
Electric Bone Mill. It is unsafe to use this device before you have read and are
thoroughly familiar with this document.
2

Symbols

REF
SN
D
the device, which could result in patient injury, illness, or death.
STERILE EO
Sterilized using ethylene oxide.
LOT
Medtronic.Com for instructions on proper disposal of this product.
EC REP
710044
10 mins
Bone Mill
! USA
P/N
+ XX °C
- XX °C

When the System Arrives

Caution
Follow instructions for use
Catalog number
Serial number
Date of manufacture
Manufacturer
Fuse For single patient use only. Do not re-use, re-process, or re-sterilize this product. Re-use, re-processing or re-sterilization may compromise the structural integrity of the device and/or create a risk of contamination of
Lot number
O position of on-o switch
Run
Moving parts can crush and cut. Keep hands clear.
RF transmitter (interference may occur)
Do not dispose of this product in the unsorted municipal waste stream. Dispose of this product according to local regulations. See http://recycling.
Unpacking and Inspection
As the box is unpacked, check o the contents of the box against the items listed on the packing slip. If the contents are incomplete or if there is damage, notify Customer Service. If the shipping container is damaged, or the cushioning material shows signs of stress, notify the carrier as well. Keep the shipping materials for carrier inspection.
Aer unpacking your new Midas Rex™ Electric Bone Mill, save the cartons and packing material. If the instrument is to be shipped from one location to another, the carefully designed shipping package will provide proper protection.

Warnings and Precautions

It is important that the Midas Rex™ Electric Bone Mill operator be familiar with this manual: its precautions, procedures and safety issues. Four labels are used in this manual to identify important concerns, conditions, or procedures:
WARNING
Describes serious adverse reactions and potential safety hazards that can occur during the proper or improper use of a device.
IPX1
IPX2
Max ON 45 s / Min OFF
Authorized representative in the European Community
is device complies with medical device directive 93/42/EEC
Conforms to ANSI/AAMI ES 60601-1, IEC/EN 60601-1 Certied to CSA C22.2 No. 60601-1
Caution: federal law in the U.S.A. Restricts the sale, distribution or use of this device to, by or on the order of a licensed medical practitioner.
Protected against vertically dripping water
Protected against vertically dripping water when the enclosure is inclined 15°
EN60601-1 duty cycle - maximum on time 45 seconds / minimum o time 10 minutes
Bone Mill
Sterilized using irradiation
Quantity
For U.S. Audience only
Part number
PRECAUTION
Any special care to be exercised by a practitioner or patient for the safe and eective use of the device.
IMPORTANT
Identies conditions or practices which require attention for most desirable results.
NOTE
Identies special information or to clarify / emphasize important instructions.
ROHS - Environmental friendly use period - China (SJ/T11364-2006.)
Temperature Limitation
MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL
3
WARNING
1. It is important that the Electric Bone Mill operator be familiar with this manual, its precautions, procedures and safety issues.
2. is medical device complies with EN60601-1-2 safety standard for electromagnetic compatibility, requirements and test. However, if this equipment is operated in the presence of high levels of electromagnetic interference (EMI) or highly sensitive equipment, interference may be encountered and the user should take whatever steps are necessary to eliminate or reduce the source of the interference. Diminished performance may lengthen operating time for anesthetized patient.
3. Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this Guide.
4. Portable and mobile RF communications equipment can aect Medical Electrical Equipment.
5. Use of accessories and cables other than those specied and sold by Medtronic may result in increased emissions and decreased immunity of this unit.
6. A system that is not functioning properly should not be used until all necessary repairs have been made and the unit is tested to ensure that it is functioning in accordance with Medtronic specications.
7. Always inspect the components before and aer use for any damage. If damage is observed, do not use the damaged part until it is repaired or replaced.
8. DO NOT operate the Midas Rex™ Electric Bone Mill base without the bowl and cap secured in place.
9. Verify reusable device was sterilized prior to use. If not sterilized, do not use.
10. Achieve electrical grounding reliability with proper connections. Connect the Bone Mill to hospital grade receptacles only.
11. DO NOT operate the Midas Rex™ Electric Bone Mill in the presence of ammable anesthetics. Avoid potential ignition or explosion of gases.
12. Take care to process bone to the desired partical size. DO NOT over process bone.
13. e Midas Rex™ Electric Bone Mill bowl, cap, and spatula are provided STERILE and are intended for SINGLE PATIENT USE ONLY. DO NOT RESTERILIZE THESE COMPONENTS. MEDTRONIC ASSUMES NO LIABILITY FOR PRODUCTS WHICH HAVE BEEN RESTERILIZED BY HEALTH CARE FACILITIES. In discarding the disposables, follow health-care facility guidelines on proper disposal of contaminated materials.
14. Disconnect power to the Midas Rex™ Electric Bone Mill base before cleaning the unit to avoid electrical shock.
15. Aer each procedure, properly clean all reusable system components.
16. Sterilizing and drying the Midas Rex™ Electric Bone Mill Base before storing decreases the likelihood of cross-contamination.
17. All service must be performed by Medtronic-qualied personnel only.
18. DO NOT operate the Midas Rex™ Electric Bone Mill before following proper cleaning and sterilization protocols.
19. Properly dispose of single-use devices removed from sterile packages. Devices lose sterility upon removal from packaging.
20. DO NOT come in contact with the interior of the disposable bowl or cap during the milling of bone.
21. DO NOT operate the Midas Rex™ Electric Bone Mill in the presence of Magnetic Resonance Imaging Devices.
22. DO NOT operate the Bone Mill without eye wear protection.
23. DO NOT use any other components except for Medtronic Midas Rex™ Electric Bone Mill disposables.
24. Do not use accessory if package is opened or damaged. A broken seal oers no protection against cross-contamination.
25. No modication of this equipment is allowed.
26. To avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth.
PRECAUTION
1. e addition of any liquid to the milling process should be administered in small volume to prevent leakage.
2. Improper sterilization or overloading of the Midas Rex™ Electric Bone Mill can lead to permanent motor damage.
3. e Midas Rex™ Electric Bone Mill must not be overloaded. DO NOT process more that 10 cc of bone at one time. e maximum starting size of the pieces should not exceed 1.5 cc. Additional bone gra may be processed in separate batches or loads.
4. DO NOT use Midas Rex™ Electric Bone Mill Disposables to process more than 6 bone pieces approximately 1.5 cm3 in size at one time.
5. DO NOT use Midas Rex™ Electric Bone Mill Disposables to process more than a total of 40 cc of bone during any one surgical procedure.
6. DO NOT immerse and/or soak the Bone Mill Base or Console.
7. DO NOT use chlorine based or corrosive cleaning agents such as bleach, acetone, solutions containing glutaraldehyde, etc.
8. e equipment should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment should be observed to verify normal operation in the conguration in which it will be used.
9. Take care when placing base unit on a surgical drape, that the base is in a stable position and that the cable is safely routed.
10. DO NOT obstruct access to the power cord connection at the rear of the console.
4
1

System Description

1
2
3
5
4
5
1
3
MIDAS REX®
BONE MILL
2
e Midas Rex™ Electric Bone Mill is composed of a non sterile console and base with sterile single use disposable components. e sterile single use disposable components consist of one bowl, one cap, and one spatula. A Midas Rex Console provides the power to drive the base and disposable components.
Indications
Electric Bone Mill
Disposables
Midas Rex Electric Bone Mill disposable components
2
3
e Midas Rex™ Electric Bone Mill is intended to mill bone, producing bone particles 1 to 5 mm in size.
Contraindications
ere are no known contraindications.

Setup and Operation

Base with bowl and cap.
1. Cap
2. Bowl
3. Base
4. Run switch
5. Disposable bowl release buttons.
1. Bowl
2. Spatula
3. Cap
Operative Procedure
MIDAS REX®
BONE MILL
1. Connect the base cable to console.
Correct
2. Press bowl “rmly” onto Midas Rex Electric Bone Mill Base until an audible click (lock) is heard. “Observe” that both of the “Release Buttons” are fully ou
NOTE
Incorrect
t.
Console
BM200 bowl and cap are not compatible with this unit.
1. Power Switch (lighted indicator = On / O = O).
2. Bone mill base to console control cable.
3. LED (illumination indicates “mill present”).
MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL
3. Lay harvested bone specimens into bowl.
PRECAUTION
Please see Precautions 1, 3, 4, and 5.
5
10 cc Before Milling 10 cc Aer Milling
WARNING
DO NOT operate the Midas Rex™ Electric Bone Mill base without the bowl and cap secured in place.
4. Secure Cap onto Bowl by rotating counterclockwise. DO NOT activate Bone Mill without securing Cap.
5. To process bone, press the Midas Rex Electric Bone Mill run switch. Continue to run until desired bone particle size is achieved.
IMPORTANT
Milling time is dependent on the density of the patient’s bone, the ratio of cancellous and cortical bone, the size of the individual particles and the batch amount of the bone to be processed. Always check bone aer 10-15 seconds to gauge quality of milling.
WARNING
Take care to process bone to the desired partical size. DO NOT over process bone. Over processed bone may lead to thermal osteonecrosis.
NOTE
If the blade becomes inoperable, (jams or runs very slowly) stop operation of the Midas Rex™ Electric Bone Mill and remove bowl cap and use the spatula to free the bone and blade. Remove some of the bone if needed to obtain proper operating speed.
6. To remove bone particles:
• stop operation of the Midas Rex Electric Bone Mill
• press release buttons and remove bowl and cap from base.
• remove cap from bowl
• use spatula to remove bone particles from bowl and cap.
NOTE
Aer removing bone particles from Bowl, the milling process may be repeated with additional bone specimens, as needed, (Up to a total of 40 cc of bone may be processed by the Bone Mill Disposable during any one surgical procedure.)

Post-Operative Procedure

1. Turn power O at console and disconnect power cable.
2. Press release buttons and remove bowl from base.
3. Discard bowl, cap and spatula.
4. Clean and sterilize Midas Rex Electric Bone Mill Base according to directions in cleaning section of manual.
5. Clean console

Cleaning & Sterilization

Warnings and Precautions
Limitations
INSTRUCTIONS
Manual cleaning of Midas Rex™ Electric Bone Mill Base (BM110) and Midas Rex™ Bone Mill Console (BM120)
Sterilization of Midas Rex™ Electric Bone Mill Base (BM110)
(Temperatureand times are minimum requirement.)
Please see Precautions 2, 5 and 6.
Aer cleaning and sterilization, verify functionality prior to re-use. Midas Rex™ Electric Bone Mill Bowl, Cap and Spatula are single use devices. Opened but unused Bone Mill Disposables may not be re-sterilized under any sterilization method.
Wipe all surfaces of the Midas Rex™ Electric Bone Mill with a cloth dampened with a neutral enzymatic detergent (pH 6.0 - 8.0). Dry thoroughly. Sterilizing and drying the Midas Rex™ Electric Bone Mill Base before storing decreases the likelihood of cross-contamination.
Steam Sterilization:
Cycle: Temperature: Time: Drying:
*Reference: for items contaminated with TSE agents steam sterilization under a temperature of 273-278°F/134-137°C may be repeated for either a total of six 3 minute cycles OR a single cycle for 18 minutes as referenced in NHS Estates HTM 2010 parts 4 & 6: Appendix 2, items contaminated with TSE agents.
100% EtO Sterilization Parameters: Not Validated
Gravity 132°C 10 min > 25 min
Pre-vac 132°C 4 min > 20 min
Pre-vac* 134-137°C 3 min > 20 min
PRECAUTION
Please see Precautions 1, 3, 4, and 5.
6
ese sterilization guidelines are only recommended guidelines
receptacles.
788-6441
and may vary among hospitals depending upon the type of sterilization equipment and possible pathogens in their particular areas. ese guidelines represent exposure time and not total cycle time. e recommended sterilization parameters may be modied based on standard hospital sterilization protocol. e bone mill base should be sterilized in accordance with the guidelines in the Midas Rex™ Electric Bone Mill user’s guide. It must be allowed to cool prior to use. e metal portion of the bone mill base must be less than 40ºC prior to use. Cooling time required will vary based on sterilization containers used, wrapping method, etc. Use of the bone mill base while still hot from sterilization (ash sterilization) may cause the motor to overheat and fail.

Limited Warranty*

(U.S. Customers Only)
A. is Limited Warranty provides the following assurance to the
purchaser of a Medtronic Midas Rex™ Electric Bone Mill. e Midas Rex™ Electric Bone Mill includes the Console, and Base (hereaer referred to as “Components”), and the Disposable Bowl, Cap and Spatula (hereaer referred to as “Accessory”) and jointly referred to as Midas Rex™ Electric Bone Mill, unless specically noted.
(1) Should a Component fail to function within normal tolerances
due to a defect in materials or workmanship within a period of one (1) year from the date of shipment of the new Component, or 180 days from the date of shipment of the serviced or refurbished Component to the purchaser (but not less than the one year period from the date of shipment of the new Component), or
(2) Should an Accessory fail to function within normal tolerances
due to a defect in materials or workmanship within a period of 90 days from the date of shipment of a new Accessory or serviced or refurbished Accessory to the purchaser.
(3) Medtronic will at its option: (a) issue a credit to the purchaser of
the replacement Component or Accessory equal to the Purchase Price, as dened in Subsection A(4), against the purchase of any same Component or Accessory requested as its replacement, or, (b) provide a functionally comparable replacement Component or Accessory at no charge.
(4) As used herein, Purchase Price shall mean the lesser of the
net invoiced price of the original or current functionally comparable, or replacement Component or Accessory.
B. To qualify for this Limited Warranty, these conditions must be met:
(1) e Midas Rex™ Electric Bone Mill must be used in accordance
with the labeling and Instruction Manual provided with the products.
(2) Only Midas Rex™ Electric Bone Mill Accessories that are
manufactured, authorized, and quality controlled by Medtronic may be used.
(3) No reprocessed Accessory sets may be used. (4) e Midas Rex™ Electric Bone Mill may not be damaged by
accident, abuse, misuse or tampering.
(5) e Midas Rex™ Electric Bone Mill or Midas Rex™ Electric
Bone Mill Accessories may not be modied without the writte permission of Medtronic.
C. is Limited Warranty is limited to its express terms. In particular:
(1) Except as expressly provided by this Limited Warranty,
MEDTRONIC IS NOT RESPONSIBLE FOR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES BASED ON ANY DEFECT, FAILURE OR MALFUNCTION OF THE Midas Rex™ Electric Bone Mill TO FUNCTION WITHIN
NORMAL TOLERANCES WHETHER THE CLAIM IS BASED ON WARRANTY, CONTRACT, NEGLIGENCE, STRICT LIABILITY, OR OTHER TORT OR OTHERWISE.
MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL
n
(2) is Limited Warranty is made only to the purchaser. AS TO
ALL OTHERS, MEDTRONIC MAKES NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, WHETHER ARISING FROM STATUTE, COMMON LAW, CUSTOM OR OTHERWISE.
(3) e exclusions and limitations set out above are not intended
to, and should not be construed so as to contravene mandatory provisions of applicable law. If any part or term of this Limited Warranty is held to be illegal, unenforceable or in conict with applicable law by a court of competent jurisdiction, the validity of the remaining portions of the Limited Warranty shall not be aected, and all rights and obligations shall be construed and enforced as if this Limited Warranty did not contain the particular part or term held to be invalid. is Limited Warranty gives the purchaser specic legal rights. e purchaser may also have other rights that vary from state to state.
(4) No such express or implied warranty to the purchaser shall
extend beyond the period specied in A(1) above.
(5) is Limited Warranty shall be the exclusive remedy available to
any person.
(6) No person has any authority to bind Medtronic to any
representation, condition, or warranty, except this Limited Warranty.
*is Limited Warranty is provided by Medtronic Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, Texas 76137-4116. It applies only in the United States. Areas outside the United States should contact their local Medtronic representative for terms of the warranty.

Service

e Midas Rex™ Electric Bone Mill is not user-serviceable, and no preventive maintenance or calibration procedures are required except for the directions listed in this manual.
PRECAUTION
Please see Precaution 8

Troubleshooting

SYMPTOM ISSUE ACTION e blades do not
turn with the run switch depressed:
If there is still a problem with the Midas Rex™ Electric Bone Mill Contact Medtronic Customer Service at 1-800-335-9557 or FAX: 817-
Loose connection
Bone stuck under blade.
Ensure that the power cable and the control switch cable are fully engaged in the
Ensure that power is O and using spatula rotate blade in a clockwise direction.
7
Technical Specications
Size Base 4.95”W x 4.35”H x 5.92”D
(126mm x 111mm x 151mm)
Console 8.00”W x 2.76”H x 9.09”D
(204mm x 70mm x 231mm)
Weight Base 2.73lbs (1.24kg)
Console 3.60 lbs (1.63kg) EN 60601-1 Duty Cycle Max ON 45s/Min OFF 10 Mins Recommended Installation: Table Top External Power Source: 100-240VAC~, 50/60 Hz, 250 VA Max Internal Fuse T4A, 250VAC Break 10A, 250VAC Front Panel Control: Power (On-O) Switch Front Panel Indicators: Power, Mill Present Authorized Service Organization: Medtronic Xomed, Inc. Environmental Requirements: Operating:
Temperature: +10 °C to +40 °C
Humidity: 30% to 75% RH
Barometric Pressure: 700 to 1060 hPa Storage and Transportation:
Temperature: -40 °C to +70 °C
Humidity: 10% to 95% RH
Barometric Pressure: 500 to 1060 hPa Disposable Components: Bowl, Cap, and Spatula Conforms to ANSI/AAMI ES 60601-1, IEC/EN 60601-1. Certied to CSA C22.2 No. 60601-1, and is not for continuous operation. Protection Against Water Ingress
Essential Performance:
Console, IPX1 Base, IPX2 ere is no essential performance per IEC
60601-1 based on the risk analysis of this product.

Midas Rex™ Electric Bone Mill Components

BM110 BM110 Midas Rex™ Electric Bone Mill Base BM120 BM120 Midas Rex Electric Bone Mill Console BM130 Bone Mill Instrument Tray EA600 North America 110 VAC Power Supply Cord EA603 Japanese Power Supply Cord EA611 Italian Power Supply Cord EA604 Chinese Power Supply Cord EA607 India / South Africa Power Supply Cord EA608 Argentina Power Supply Cord EA609 Israel Power Supply Cord EA610 Danish Power Supply Cord EA606 UK 240 VAC Power Supply Cord EA602 Continental Europe 230 VAC Power Supply Cord EA601 International 115/230 VAC Power Supply EA605 Australia 240 VAC Power Supply Cord 175025EN Midas Rex™ Electric Bone Mill User’s Guide, English 175025MLIU Midas Rex™ Electric Bone Mill User’s Guide, 15 Language 11270037 Fuse
DISPOSABLES BM210 Bowl, Cap and Spatula
8

Electromagnetic Compatability

Environment of Intended Use: Professional healthcare facility environment
Guidance and manufacturer’s declaration – electromagnetic emissions
e Midas Rex™ Electric Bone Mill is intended for use in the electromagnetic environment specied below. e customer or the user of the Midas Rex™ Electric Bone Mill should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage uctuations/
icker emissions
IEC 6100-3-3
e Midas Rex™ Electric Bone Mill is intended for use in the electromagnetic environment specied below. e customer or the user of the Midas Rex™ Electric Bone Mill should assure that it is used in such an environment.
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic eld
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Group 1 e Midas Rex™ Electric Bone Mill uses RF energy only for its internal function. erefore, its RF emissions are very low and are not
Class A e Midas Rex™ Electric Bone Mill is suitable for use in all establishments, other than domestic establishment and those directly
Class A
Complies
likely to cause any interference in nearby electronic equipment.
connected to the public low voltage power supply network that supplies buildings for domestic purpose.
Guidance and manufacturer’s declaration – electromagnetic immunities - part I
± 8 kV contact ± 15 kV air
± 2 kV for power supply lines ± 1 kV for input/output lines
± 1 kV line to line ± 2 kV line to earth
0% UT (100% dip in UT) for 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT (100% dip in UT) for 1 cycle at 0°
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 0.5 sec
0% UT (100% dip in UT) for 5 sec
30 A/m 30 A/m Power frequency magnetic elds should be at levels characteristic of a typical location in a
± 8 kV contact ± 15 kV air
± 2 kV for power supply lines ± 1 kV for input/output lines
± 1 kV line to line ± 2 kV line to earth
0% UT (100% dip in UT) for 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT (100% dip in UT) for 1 cycle at 0°
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 0.5 sec
0% UT (100% dip in UT) for 5 sec
e relative humidity should be at least 5%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Midas Rex™ Electric Bone Mill requires continued operation during power mains interruptions, it is recommended that the Midas Rex™ Electric Bone Mill be powered from an uninterruptible power supply or a battery.
typical commercial or hospital environment.
MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL
9
Guidance and manufacturer’s declaration – electromagnetic immunities - part II
e Midas Rex™ Electric Bone Mill is intended for use in the electromagnetic environment specied below. e customer or the user of the Midas Rex™ Electric Bone Mill should assure that it is used in such an environment.
Immunity test IEC 60601-1-2 test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
NOTE: ese guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects and people.
level
3 Vr ms 150 kHz to 80 MHz
6 Vr ms 150 kHz to 80 MHz in ISM bands
3 V/m 80 MHz to 2.7 GHz
9–28 V/m Spot frequencies 385 MHz–5.785 GHz Pulse modulation
Compliance level Electromagnetic environment - guidance
3 Vr ms 150 kHz to 80 MHz
6 Vr ms 150 kHz to 80 MHz in ISM bands
3 V/m 80 MHz to 2.7 GHz
9–28 V/m Spot frequencies 385 MHz–5.785 GHz Pulse modulation
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30cm (12 inches) to any part of the Midas Rex™ Electric Bone Mill, including cables specied by the manufacturer. Othewise, degradation of the performance of this equipment could result.
Portable and mobile RF communications equipment should be used no closer to any part of the Midas Rex™ Electric Bone Mill including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer, E is the immunity test levels in volt per meter (V/m), and d is the recommended separation distance in meters (m).
Interference may occur in the vicinity of equipment marked with the following symbol:
d = (6/E)√P
Recommended separation distances between
portable and mobile RF communications equipment and the Midas Rex™ Electric Bone Mill
e Midas Rex™ Electric Bone Mill is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. e customer or the user of the Midas Rex™ Electric Bone Mill can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Midas Rex™ Electric Bone Mill as recommended below, according to the maximum output power of the communications equipment.
Rated
maximum
output
power of
transmitter
P (W)
0.01 0.03 0.03 0.07 0.03 0.03 0.03 0.07
0.1 0.07 0.07 0.21 0.07 0.07 0.07 0.21
1 0.22 0.22 0.67 0.22 0.22 0.22 0.67
10 0.7 0.7 2.12 0.7 0.7 0.7 2.12
100 2.2 2.2 6.7 2.2 2.2 2.2 6.7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: ese guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects and people.
380 MHz–390MHz
d = 0.22√P
430 MHz–470MHz
d = 0.22√P
Separation distance according to frequency of transmitter m
704 MHz–787MHz
d = 0.67√P
800 MHz–960MHz
d = 0.22√P
1.7 GHz–1.99GHz d = 0.22√P
2.4 GHz–2.57GHz d = 0.22√P
5.1 GHz–5.8GHz d = 0.67√P
10
U.S. Customers
FAX: 817-788-6489
oce or distributor.
Medtronic Powered Surgical Solutions 4620 North Beach Street Fort Worth, TX 76137 USA medtronic.com manuals.medtronic.com
Orders:
Customer Service 1-800-433-7080
Repair Department
1-800-335-9557 817-788-6440 FAX: 817-788-6441
International Customers
International Customers should contact their local Medtronic
Should your console or related equipment malfunction, (please adhere to the following guidelines)
1. A return goods authorization number is required.
2. Please package the console as safely as possible and insure the shipment.
3. A copy of your purchase order is required. Make certain the purchase order includes the following:
• Shipping and billing information
• Purchase order number
• Contact person
• Phone number
• Description of malfunction
• Your Medtronic account number.
4. Please indicate preferred method of return shipment. Otherwise the unit will be shipped back via UPS ground.
5. When the malfunctioning unit is not covered by warranty, Medtronic will contact your facility promptly with a service cost estimate if requested. e customer will be responsible for freight charges on non-warranted units.
The information contained in this document is accurate at time of publication. Medtronic reserves the right to make changes to the product described in this manual. Refer to manuals.medtronic.com for the current version.
MIDAS REX™ ELECTRIC BONE MILLMIDAS REX™ ELECTRIC BONE MILL
11
M988773A001 A
EC REP
2018 - 09
© 2018 Medtronic, Inc.
Manufactured by:
Medtronic Powered Surgical Solutions 4620 North Beach Street Fort Worth, TX 76137 USA Telephone: 817-788-6400 Toll-Free: 1-800-643-2773 Fax: 817-788-6401 medtronic.com
Orders:
Telephone: 817-788-6400 Toll-Free: 1-800-433-7080 Fax: 817-788-6489
Repairs:
Telephone: 817-788-6440 Toll-Free: 1-800-335-9557 Fax: 817-788-6441
EC Authorized Representative:
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands
Europe/Middle East/Africa:
Medtronic International Trading Sàrl
Route du Molliau 31
Case Postale 84
CH- 1131 Tolochenaz
Switzerland
Australia:
Medtronic Australasia Pty Ltd
5 Alma Road
Macquarie Park, NSW 2113
Australia
Canada:
Medtronic of Canada Ltd
99 Hereford Street
Brampton, Ontario L6Y 0R3
Canada
+1 905 460 3800
Technical manuals:
medtronic.com/manuals
Loading...