Medtronic ATLGIQE Operator's Manual

Page 1
Quick Setup Guide
autoLog IQ
Autotransfusion System
Page 2
Read the entire Medtronic autoLog IQ™ Autotransfusion System User Manual before using the autoLog IQ system. This Quick Setup
Guide is not a substitute for the complete User Manual. It is only intended to assist the operator in the setup of the autoLog IQ autotransfusion system.
Page 3
1. Attach the step-down connector to the bottom of the collection reservoir. Ensure that the connector is fully secured.
2. Close the clamp on the step-down connector before the kit setup is complete. Ensure that the clamp is fully closed.
3.1 . Place the collection reservoir into the reservoir holder. Attach a vacuum line to the yellow-capped por t on the collection reservoir.
Blood collection system setup
3.2. Attach the other end of the vacuum line to the vacuum
overow trap of the device.
1English
Page 4
Blood collection system setup
2English
4. Attach the blue-capped suction/anticoagulant line
connector to either the blue-capped or white- capped inlet port on top of the collection reservoir.
6. Turn on the device power by pressing
.
5. Close the roller clamp on the suction/anticoagulant line.
7. Turn on the vacuum by pressing
.
Page 5
8. If necessar y, adjust the vacuum setting using the Vacuum
drop- down menu on the Main screen.
9. Spike the anticoagulant c ontainer. If the container is nonvented, open the vent cover on the drip chamber.
10 .1. Before blood collection, open the roller clamp and prime
the collection reservoir with a minimum of 100 mL of anticoagulant solution.
Blood collection system setup
10.2. Reduce the anticoagulant ow to a ratio of approximately
15 mL of solution to 100 mL of blood.
3English
Page 6
Wash kit setup
4English
1. The message appears: Kit is not installed. Lock the
bowl cover.
3. Place the wash kit in the cradle arm and remove the wash ki t c over.
2. Attach the saline wash solution containers to a lower IV pole hook.
4. Remove the waste bag from the wash kit. Conrm that the drain valve is closed on the waste bag.
Page 7
5. Install the waste bag onto the waste bag pins.
6. Remove the holding bag from the wash kit and hang it on an upper IV pole hook. Close the clamps on the 2 red­capped outlet lines that each include a membrane port.
7. Place the bowl in the c entrifuge chamber with the yellow­capped waste line outlet tubing facing the waste bag. Refer to the markings on the top panel.
Wash kit setup
8. Align the centrifuge markings with the vertical assemblies on the bowl. To lock the bowl, simultaneously press down on the bowl and turn it clock wise until a click is heard.
5English
Page 8
Wash kit setup
6English
9. Insert the yellow-capped connector into the yellow-capped
inlet por t on the waste bag. Conrm that there are no kinks
in the waste line.
11.1. Place the manifold into the recessed cavity. Place the 3 manifold tubing lines into the tubing guides, and ensure that each tube is fully seated in each guide.
10. The message appears: Insert the kit in the valve and
the pump. Press to resume.
11. 2 . Plac e the single tubing line of the manifold into the air detector. Close the manifold cover. Rotate the manifold cover latch until the lock indicators are aligned.
Page 9
12. Pull the pump lever and hold it open. Stretch the pump tubing around the pump head and over the pump outlet tubing guide. Release the pump lever.
13. Press
. The message appears: Connect the saline line
and reservoir. Press to resume.
14. Spike the middle manifold tubing into the saline bag or bags. Note: If using one saline bag, conrm that the clamp on the unused saline line is closed.
Wash kit setup
15. Connect the left-hand tube of the manifold to the blue- capped step- down connector at the bot tom of the reservoir, and then open the clamp on the step-down connector.
7English
Page 10
Blood processing setup
8English
1. Press . The message appears: Machine stopped.
Press to enable self- star t.
2.2. The device automatic ally starts when it detects
approximately 80 0 mL of blood in the reservoir for at
least 5 sec onds. Note: To begin the ll before 800 mL is collected, press .
2.1 . Press
to activate the self-star t mode. The message
appears: Self-start enabled. Press to manually
start.
Page 11
Page 12
Medtronic, Inc.
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000 LifeLine Technical Support, 24hour consultation service: 1 877 526 789 0
Medtronic Perfusion Systems
7611 Northland Drive Minneapolis, MN 55428 USA +1 763 391 900 0 Customer service and product orders: 1 800 854 3570 www.perfusionsystems.com
Plexus Manufacturing Sdn Bhd
Bayan Lepas Free Industrial Zone Phase II, 11900 Bayan Lepas Penang Malaysia +60 4 6321000
Medtronic B.V.
Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands +31 45 566 80 00
Canada
Medtronic of Canada Ltd
99 Hereford Street Brampton, Ontario L6Y 0R3 Canada 1 800 26 8 5346
© 2019 Medtronic M999097A001 Rev A UC201708091a EN
*M999097A001*
Loading...