Medtronic ATDRS1 Programming Guide

Page 1
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™ SW043
Programmer software
Pacemaker Programming Guide Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Page 2
Page 3
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™ SW043
Pacemaker Programming Guide
A guide to programming the Attesta MRI SureScan and Sphera MRI Surescan pacemakers using the Medtronic SW043 application software on a Medtronic programmer.
Page 4
Adapta, Attesta, Capture Management, Marker Channel, Medtronic, MVP, Quick Look, Relia, Sensia, Sphera, SureScan, TherapyGuide, Versa
Page 5
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Contents

1 Introduction .......................................................... 7
1.1 Introduction ......................................................... 7
2 Using the programmer ............................................... 10
2.1 The display screen .................................................. 10
2.2 Setting up your programmer .......................................... 13
2.3 Streamlining implant and follow-up sessions with Checklist ............... 14
2.4 Connecting the programmer to skin electrodes .......................... 16
2.5 Positioning and using the programming head ........................... 17
2.6 Enabling emergency VVI pacing ...................................... 20
2.7 Recording an ECG strip .............................................. 21
3 Conducting a patient session ........................................ 23
3.1 Therapy and programmer features ..................................... 23
3.2 Automatically adapting parameters .................................... 24
3.3 Starting a patient session ............................................ 25
3.4 Proceeding with session tasks ........................................ 29
3.5 Interrogating the pacemaker .......................................... 32
3.6 Taking a quick look at pacemaker operation ............................. 33
3.7 Viewing Battery and Lead Measurements .............................. 35
3.8 Checking the present parameter settings ............................... 37
3.9 Viewing and programming patient information stored in the pacemaker ..... 37
3.10 Recording an ECG Strip of magnet operation ........................... 38
3.11 Checking the patient’s underlying rhythm ............................... 39
3.12 Printing reports ..................................................... 41
3.13 Saving session data ................................................. 43
3.14 Ending a patient session ............................................. 45
4 Viewing the patient’s ECG and EGM traces ............................ 49
4.1 Viewing the ECG and other rhythm waveforms .......................... 49
4.2 Adjusting and arranging the waveform traces ............................ 51
4.3 Waveform strips .................................................... 56
Pacemaker Programming Guide 5
Page 6
Medtronic
5 Collecting diagnostic data ........................................... 59
5.1 About diagnostic data collection ....................................... 59
5.2 Viewing the collected data ........................................... 62
5.3 Automatically collected data displays .................................. 63
5.4 Clinician-selected detailed data displays ............................... 72
5.5 Programming data collection options .................................. 78
5.6 Clearing data from the pacemaker ..................................... 80
6 Evaluating parameter settings ........................................ 81
6.1 Measuring stimulation thresholds ...................................... 81
6.2 Determining a sensitivity setting ....................................... 92
6.3 Using temporary programming to evaluate parameter settings ............. 96
7 Programming pacemaker parameters ................................. 98
7.1 Programming parameters ............................................ 98
7.2 Using TherapyGuide to select parameter values ........................ 106
7.3 Saving/retrieving a set of parameter values ............................ 108
7.4 Retrieving Key Parameter History information .......................... 109
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
8 Rate Response setup ............................................... 111
8.1 Programming Rate Response parameters ............................. 111
8.2 Using Exercise test to verify Rate Response ........................... 112
9 Using the EP Studies function ....................................... 116
9.1 About the EP Studies function ....................................... 116
9.2 Starting the EP Studies function ...................................... 122
9.3 Setting up and executing a PES protocol .............................. 125
9.4 Setting up and executing a Burst protocol ............................. 128
9.5 Setting up and executing a VOO Burst ................................ 130
9.6 PES parameter definitions and values ................................. 131
9.7 Burst parameter definitions and values ................................ 134
Index ................................................................... 136
6 Pacemaker Programming Guide
Page 7
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

1 Introduction

1.1 Introduction

The information in the Attesta MRI SureScan / Sphera MRI SureScan Pacemaker Programming Guide (PPG) applies to the following devices:
Attesta DR MRI SureScan model ATDR01
Attesta L DR MRI SureScan model ATDRL1
Attesta S DR MRI SureScan model ATDRS1
Attesta SR MRI SureScan model ATSR01
Sphera DR MRI SureScan model SPDR01
Sphera L DR MRI SureScan model SPDRL1
Sphera SR MRI SureScan model SPSR01

1.1.1 How to use this guide

Product information about Attesta MRI SureScan and Sphera MRI Surescan pacemakers and the associated Medtronic Model SW043 software for the programmer is presented in 2 documents.
The reference manual provides detailed information on the pacemakers.
This pacemaker programming guide contains instructions on how to use the programmer and the programming software.
1.1.1.1 About the reference manual
The reference manual describes in detail how the pacemakers operate and specifies the capabilities of the pacemakers. This manual includes the following information:
Describes the pacing modes, rate response options, special therapy features, telemetry types, and data collection options. In some cases, guidelines are given on how to configure the pacemaker operation.
Contains troubleshooting information for electrical and hemodynamic problems.
Contains a glossary of terms.
Pacemaker Programming Guide 7
Page 8
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
1.1.1.2 About the pacemaker programming guide
The pacemaker programming guide describes how to program Attesta MRI SureScan and Sphera MRI Surescan pacemakers using a programmer. The manual presents the following information:
How to set up and configure the programmer.
How to start a patient session, use the various follow-up features during the session, and properly end the session.
How to view and print the patient’s ECG and EGM waveform traces.
How to configure the pacemaker to collect diagnostic data and how to retrieve and view this information.
How to measure stimulation thresholds and sensing levels.
How to use TherapyGuide to obtain suggested parameter values.
How to program parameter values and verify rate response parameters settings.
How to run EP Studies.
1.1.1.3 Manual conventions
This pacemaker programming guide contains instructions on how to use the programmer and the programming software. Throughout this manual, the work ’device’ refers to the implanted pacemaker. The names of screen titles, screen text, and interactive elements are shown in bold type. Navigation paths to software screens or programmable parameters are shown with a “>” character betrween steps in the path (for example, Data > Graphs and
Tables > Heart Rate Histograms > Open Data).
1.1.1.4 Additional manuals
Additional manuals and documents with information about the programmer and implanted device:
MRI technical manual – This manual provides MRI-specific procedures and warnings and precautions.
Device manual – This manual contains model-specific feature information, indications and contraindications, warnings and precautions, instructions for implanting the device, quick reference specifications, and parameter tables.
Explanation of symbols – This document defines the symbols that may appear on the device package. Refer to the package label to see which symbols apply specifically to this device.
8 Pacemaker Programming Guide
Page 9
Medtronic
Medical Procedure and EMI Warnings and Precautions Manual for Health Care Professionals – This manual provides warnings, precautions, and guidance for health care
professionals who perform medical therapies and diagnostic procedures on cardiac device patients. The manual also provides patient education information related to sources of electromagnetic interference (EMI) at home, at work, and in other environments.
Radio regulatory compliance information – This document provides compliance information related to the radio components of the device.
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Pacemaker Programming Guide 9
Page 10
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

2 Using the programmer

2.1 The display screen

The programmer display screen is an interactive device that not only displays information in the form of both text and graphics but also functions as a control panel by displaying buttons and menu options that you can select using the touch pen.

2.1.1 Features and conventions of the display screen

This section describes the features and conventions of the display screen. The Therapy Parameters screen below shows the main elements of the typical screen.

Figure 1. Main elements of a display screen

1 Status bar 2 Live Rhythm Monitor window 3 Task area 4 Buttons
10 Pacemaker Programming Guide
5 Command bar 6 Active fields 7 TherapyGuide 8 Tool palette
Page 11
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
2.1.1.1 The status bar
The status bar at the top of the screen shows this information:
The current pacing mode
When any one of a number of test conditions is occurring
The pacemaker model
2.1.1.2 The Live Rhythm Monitor window
This window displays a partial view of the ECG, Marker Channel, and telemetered EGM waveform traces. To expand this window to its full size, tap the small square button in the upper-right corner of the window below the status bar, or tap Adjust….
Heart rate and rate interval show if the programming head is positioned over the pacemaker.
Annotations above the waveform trace show the point of programmed parameter changes.
The waveform trace or traces shown depend on the selected task screen and how traces have been arranged in the full-screen view. Refer to Section 4.1, “Viewing the ECG and other rhythm waveforms”, page 49 for information about the Live Rhythm Monitor.
2.1.1.3 The task area
The portion of screen between the Live Rhythm Monitor window at the top and the command bar at the bottom changes according to the task or function you select. The example in Figure 1 shows the Therapy Parameters screen for programming pacemaker parameter settings. This task area would appear much different if you selected, for example, the
Threshold Test Setup screen.
2.1.1.4 Buttons
Tap buttons with the touch pen to operate the programmer.
Buttons may directly execute a command, such as the PROGRAM button, or they may open a window that prompts another action. Usually such buttons have a label ending with an ellipsis, such as the Save… or Get… buttons.
A procedure may instruct you to “tap and hold” a button. In such cases you must hold the tip of the touch pen on the button for a duration determined by the procedure.
Pacemaker Programming Guide 11
Page 12
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
2.1.1.5 The command bar
The bar at the bottom of the screen includes 3 buttons:
tap Emergency to program VVI pacing.
tap Interrogate… to interrogate the device.
tap End Session… to end a patient session
For more information on these functions, see Section 2.6, “Enabling emergency VVI pacing”, page 20, Section 3.5, “Interrogating the pacemaker”, page 32, and Section 3.14, “Ending a patient session”, page 45.
2.1.1.6 Active fields
An active field is displayed in bolded text, values, or symbols. It is active by default, or it becomes active in response to the task or function you select.
Tap an active field to access its range of programmable values, or to select it to be programmed.
2.1.1.7 TherapyGuide
TherapyGuide offers a simple, clinically focused method to obtain suggested parameter values. At implant or at an early follow-up appointment, information can be entered about the patient’s clinical conditions. Based on those inputs, the programmer suggests parameter values. The suggestions are based on clinical studies, literature, current practice, and physician feedback.
For more information about TherapyGuide, see Section 7.2.
2.1.1.8 The tool palette
The collection of buttons and icons along the edge of the screen is referred to as the tool palette. These are the controls you use to choose the task or function screen you want displayed. Once you have started a patient session, the tool palette is always displayed, making it quick and easy to move to the desired task or function.
Each of the icons acts like a button. To select an icon, touch the icon with the touch pen. The < symbol adjacent to some of the icons indicates that selecting one of these icons opens a menu of related options. The icons without the < symbol directly open a task screen.
12 Pacemaker Programming Guide
Page 13
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Table 1. Tool palette for the desktop

Tool Selecting the tool (button or icon)
Freeze Freezes a segment of the live rhythm display.
A frozen strip can be viewed and printed (but not saved) between patient ses­sions. Markers and EGM traces are not present between patient sessions.
Strips… Accesses the rhythm strips saved during the patient session.
No saved strips are available between patient sessions.
Adjust… Opens a window of options for adjusting the Live Rhythm Monitor.
Additional adjustment options are present during a patient session.
Select Model Displays the Select Model screen. Use this screen to select a pacemaker
model and start a patient session.
Print Queue Displays the Print Queue screen. Use this screen to select a printer and to print
or delete print jobs.
< Programmer Displays the programmer setup options, including Preferences, Time and Date,
Artifact Detection, Software, Programmer Profile, Other Software, and Tools.
Analyzer Starts an Analyzer session.
Note: When some functions are active on the display, selecting a tool button or icon will have
no effect. Closing the active window restores operation of the tool palette.
The tool palette for the desktop differs from the tool palette during a patient session. For a description of the icons and buttons available during a patient session, see Section 3.3, “Starting a patient session”, page 25.

2.2 Setting up your programmer

2.2.1 Adjusting the programmer time and date

Use this procedure if the programmer displays or prints the incorrect time or date:
1. Tap Programmer > Time and Date to navigate to the Programmer Time and Date screen.
2. Tap the arrows to change the values for Hours, Minutes, Month, Day, and Year. Hold the touchpen on an arrow to quickly index through the range of available values.
Note: A 24-hour clock is used for the Hours field.
3. Tap Apply to program the corrected time and date.
Pacemaker Programming Guide 13
Page 14
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

2.2.2 Using audible tones

Certain events in the operation of the programmer result in an audible signal. The following tones alert the user to the success or failure of an action.
A two-tone beep (low-to-high) indicates confirmation of an Interrogate or a Program command.
A single low-tone beep indicates that an Interrogate, Program, or Emergency command was not confirmed. It may also indicate that the selected command cannot be executed.
A single, short beep coincides with pressing the Interrogate or the Program button. It also occurs upon automatic identification of the pacemaker.
2.2.2.1 How to turn the audible tones on or off
1. Tap Programmer > Preferences.
2. Tap Audio ON or Audio OFF as desired.

2.3 Streamlining implant and follow-up sessions with Checklist

Checklist allows you to catalog and list tasks that are performed during implant and follow-up sessions. You start with the first task and continue through each task in sequential order. When you select a task in the checklist, the programmer displays the screen associated with the task. When you complete the task, you can either proceed directly to the screen associated with the next task in the checklist or return to the checklist. Two standard checklists are provided: the Medtronic Standard Implant checklist and the Medtronic Standard Followup checklist. In addition to these standard checklists, you can create customized checklists.

2.3.1 How to select a checklist

1. Tap Checklist.
Note: When starting a new session, the checklist used during the previous
programming session is active.
2. Review the tasks in the Select Task list for the active checklist.
3. Tap the Checklist field to choose a different checklist.
14 Pacemaker Programming Guide
Page 15
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

2.3.2 How to use a checklist

1. Tap >> Go To Task to perform the first task on the Select Task: list.
Note: If you do not wish to perform the first task on the Select Task: list, tap on your
preferred task and then tap >> Go To Task.
2. Tap >> to the right of the Checklist icon to perform the next task in the Select Task: list.
3. Repeat Step 2 to work through the checklist tasks. To disregard a task, tap >> until your reach your preferred task on the checklist.
Note: When you reach the final task in a checklist, >> becomes inactive. Tap the Checklist icon to view the checklist at any time.
Note: The Checklist function displays a check mark next to tasks that you have performed
during the current patient session. If you wish to perform one of these tasks again, tap on that task then tap >> Go To Task.

2.3.3 How to create a custom checklist

1. Tap Checklist.
2. Tap Create/Edit….
3. Select an existing list from the Copy From: dropdown list.
4. Tap on tasks in the box on the left to copy them to the box on the right. Note that a task can be added more than once to a custom checklist.
If you do not wish to place a new task at the end of the list in the box on the right, perform the following procedure:
a. Tap on the task in the box on the right which you want the new task to follow.
b. Tap on the new task in the box on the left. It will be copied below the selected task
in the box on the right.
5. To delete a task, tap the task in box on the right, then tap Delete Item.
6. Tap Save As… to assign a name to the new checklist.
7. Tap Save.
Pacemaker Programming Guide 15
Page 16
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

2.3.4 How to edit a custom checklist

1. Tap the Checklist icon.
2. Select the checklist that you wish to edit from the Checklist: dropdown list.
Note: You cannot edit the Master Checklist.
3. Tap Create/Edit….
4. Tap on tasks in the box on the left in the order that you want them to be copied to the box on the right.
5. Perform these steps to add tasks to the box on the right that are not in the box on the left:
a. Select Master Checklist from the Copy From: dropdown list.
b. Tap on tasks in the box on the left in the order that you wish them to be copied to the
box on the right.
6. Perform these steps to add a task to a position before the end of your edited checklist:
a. Tap the task in the box on the right that the new task will follow.
b. Tap a task in the box on the left. It will appear under the highlighted task in the box
on the right.
7. To delete a task, tap the task then tap Delete Item.
8. Perform these steps to save the edited checklist:
a. Tap Save As….
b. Tap the checklist name in the Checklist - Save As window.
c. Tap Save.

2.4 Connecting the programmer to skin electrodes

At the start of a patient session, the programmer must be connected to skin electrodes on the patient. The ECG display and measurement functions will not operate without detection of the surface ECG signal.
Use an electrode lead to connect each skin electrode to the appropriate port on the ECG cable. Typical electrode placement is shown below. Use standard procedures for attaching disposable skin electrodes to the patient.
16 Pacemaker Programming Guide
Page 17
( ) = IEC coding
Labeled C on some cables
1
RA (R) LA (L)
V1 (C)
RL (N)
LL (F)
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™

Figure 2. Connecting the programmer to skin electrodes

Refer to your programmer manual for more details on this procedure.
SW043

2.5 Positioning and using the programming head

In many of the procedures described in the following chapters, you will be directed to position the programming head over the patient’s implanted device.

2.5.1 When to position the programming head

During a patient session, properly position the programming head over the implanted device before any of the following actions:
Selection of any command that initiates a programming transmission. The programming head must be held in position until completion of the transmission, which is usually indicated by a confirmation message.
Selection of any command that initiates data transmission from the implantable device. The programming head should be held steady until data reception is complete, which is usually indicated by a confirmation message.
Selection of a measurement function that requires the implantable device to be operating asynchronously as a result of the programming head magnet.
Pacemaker Programming Guide 17
Page 18
Medtronic
For any temporarily programmed state or function or for reception of continuous data such as Marker Channel telemetry or EGM waveforms, the programming head must be held in place over the implanted device for the duration of the function or until termination is desired. Lifting the programming head cancels a temporary program and terminates continuous telemetry. The implantable device reverts to permanently programmed values.
Caution: Do not position the programming head over an implanted device during electrocautery or defibrillation procedures.
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
2.5.1.1 How to position the programming head
The programming head must be properly positioned as described below to establish telemetry with the device. An incorrectly positioned programming head can result in the failure of a transmitted command and the loss of telemetry.

Figure 3. Positioning the programming head

1. Hold the programming head directly against the patient’s skin with the face of the programming head parallel to the implanted device.
2. Position the programming head so that the amber light in the light array goes out and one or more of the green lights come on. Move the head a little in each direction to find the position that lights the greatest number of green lights, indicating that telemetry has been established with the device. This is the optimum position.
18 Pacemaker Programming Guide
Page 19
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™

Figure 4. Light array indicator

SW043
1 Light array 2 Green
3 Amber/green

2.5.2 Effect of the programming head on pacemaker operation

If model selection and application (software) loading are complete, positioning the programming head over a pacemaker does not put the pacemaker in the magnet mode of operation. The programmer automatically sends a Cancel Magnet command to the pacemaker, which causes it to continue to operate as programmed.
An exception to this operation can occur if the programming head does not establish a telemetry link with the pacemaker (because of strong electrical interference or because the programming head is improperly positioned). In such cases, positioning the programming head would result in magnet mode operation until a telemetry link is established. You can easily verify the present pacing mode by observing the Status Line at the top of the screen.
Pacemaker operation returns to its programmed state about two seconds after you remove the programming head from its position over the pacemaker.

2.5.3 Observing magnet mode operation

To observe magnet mode operation during a patient session, you must conduct the Magnet test described in Section 2.7, “Recording an ECG strip”, page 21. Initiating the Magnet test results in a Threshold Margin Test (TMT) and causes the pacemaker to operate in the magnet mode.
Note: Before you select the pacemaker model, positioning the programming head over the pacemaker results in magnet mode operation.
Pacemaker Programming Guide 19
Page 20
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

2.5.4 Program and Interrogate buttons

To initiate the Program and Interrogate commands, you have the option to use the on-screen PROGRAM and Interrogate buttons or the Program P and Interrogate I buttons on the programming head.

Figure 5. Alternative Program and Interrogate programming head buttons

1 I or Interrogate button 2 Light array
Note: The I or P button on the programming head is active only when its counterpart is displayed as an active button on the display screen.
3 P or Program button

2.5.5 Automatic interrogation at the start of a session

At the start of the session, the programmer automatically interrogates the patient’s pacemaker for most of the data stored in the pacemaker. For this process, you must position the programming head over the implanted device and hold it steady in place until the interrogation is complete.
At the bottom center of the screen you will see an indicator showing the progress of the interrogation. Because this interrogation retrieves most of the data stored in the pacemaker, the process may take up to a minute.

2.6 Enabling emergency VVI pacing

You can use emergency VVI pacing to quickly enable 70 bpm, high-output ventricular pacing to restore ventricular support in an emergency situation.

2.6.1 Considerations for emergency VVI pacing

Parameter values – Emergency VVI pacing reprograms pacing parameters to emergency
settings. For a list of the emergency VVI parameter settings, see the implant manual for the device. To terminate emergency VVI pacing, you must reprogram pacing parameters from the Parameters screen.
20 Pacemaker Programming Guide
Page 21
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
MRI SureScan – MRI SureScan feature operation is disabled when emergency VVI pacing is programmed.

2.6.2 How to enable emergency VVI pacing

1. During a patient session, establish telemetry with the device.
2. Press the red VVI button on the programmer to enable emergency VVI pacing.
Depending on your model of Medtronic programmer, the emergency VVI button is:
a mechanical red button to the left of the programmer screen, on the programmer bezel.
a red button on the programmer button panel, above the programmer screen.
Note: The following procedure applies to programmers that have an on-screen Emergency button.
How to enable emergency VVI pacing with the on-screen Emergency button –
Perform the following steps to enable emergency VVI pacing with the on-screen Emergency button:
1. Establish telemetry with the device.
2. Tap Emergency.
3. Tap OK when the Emergency Program – Successful pop-up window appears. The parameters change automatically to the VVI mode.

2.7 Recording an ECG strip

At any time during a patient session, you can initiate a continuous, real-time ECG recording by pressing the desired paper speed button.

Figure 6. Printer buttons on the programmer

Pacemaker Programming Guide 21
Page 22
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
1 50 mm/s 2 25 mm/s
3 12.5 mm/s 4 Paper advance
50, 25, or 12.5 mm/s – Three buttons allow for selection of a desired paper speed for ECG, Marker Channel telemetry, and EGM recording. Pressing a button selects the printer speed and turns on the adjacent indicator light. Pressing a lit button stops the printout. The selected paper speed is printed once along the top edge of the chart recording. Paper speed for text printing is not affected by these buttons.
Paper advance – Pressing the paper advance button advances the printer paper to its next perforation for tearing off.
To stop the recording, press the same paper speed button again. Before you tear off the ECG strip, press the paper advance button to advance the strip to a perforation.

2.7.1 About the ECG recording

Because the printed recording provides a higher resolution, it may show artifacts and events that do not appear on the display.
Annotation of executed commands – Information on the ECG printout includes an indication of when certain commands to the pacemaker occurred. When confirmation of the command is received, the command name is printed at the appropriate point in the margin above the waveform grid. A recording made during the use of the following test functions shows the programmed test values as they are programmed: Magnet, Underlying Rhythm, Threshold, Sensing, EP Studies, and Temporary.
Programming arrows – The point at which the command occurred is marked above the ECG trace by a down or up arrow. A down arrow indicates a transmitted command from the programmer (programming the device); an up arrow indicates a telemetry response from the pacemaker (confirming the programming).
Marker Channel and EGM telemetry – If the programming head is positioned over the patient’s pacemaker, the recording includes a trace or traces of the telemetry being received from the pacemaker. In this example, the patient’s ECG is accompanied by Marker and EGM telemetry.
ECG and EGM trace adjustment – The ECG and EGM are recorded according to control settings accessible from the Live Rhythm Monitor window on the programmer screen (see Section 4.2, “Adjusting and arranging the waveform traces”, page 51). The ECG source (Lead I, II, or III) and the EGM source are each selected using the button bars appearing with each waveform trace.
Missing markers – A programming command or interrogation momentarily interrupts the transmission of Marker Channel telemetry. This interruption may result in missing markers.
22 Pacemaker Programming Guide
Page 23
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3 Conducting a patient session

3.1 Therapy and programmer features

3.1.1 Therapy features

Note: Certain features described in the beginning of this chapter are not offered by all
models in the Adapta/Versa/Sensia/Relia series. To find applicable features for a specific model, refer to the Implant Manual for that pacemaker.
Atrial Preference Pacing – Atrial rhythm management feature that adapts the pacing rate to slightly higher than the intrinsic sinus rate.
MVP (Managed Ventricular Pacing) – Promotes intrinsic conduction by reducing unnecessary right ventricular pacing. MVP operates when the programmed mode is either AAIR<=>DDDR or AAI<=>DDD.
Rate Response User Interface – Provides a graphical display to aid in programming rate response. When Rate Profile Optimization parameters are programmed, rate response undergoes an immediate change.
TherapyGuide – Provides a set of suggested parameter values based on patient conditions that were identified by the physician. TherapyGuide does not replace a physician’s expert judgment. The physician is free to accept, reject, or modify any of the suggested parameter values.

3.1.2 Programmer features

Auto cancel magnet – Cancels the magnet automatically when you place the programmer
head over the pacemaker. With Magnet Test, you can view and record magnet operation.
Auto-interrogation – Occurs automatically at the start of a session. The programmer automatically prints a report of the interrogated information unless this feature is turned off.
Checklist feature – You can advance quickly to the next task to be done in a patient session. Clinicians can create custom checklists that streamline task selection in follow-up and implant sessions.
Full page-size reports – You can choose to connect an external printer for printing full page-size reports.
Live rhythm waveform display – You can quickly tailor a multiple-trace display of the patient’s ECG, atrial and ventricular EGMs, and Marker Channel signals.
Pacemaker Programming Guide 23
Page 24
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Saved session data – You can save data from a session on a disk, which allows you to import session data into a database or review it using Read From Media.

3.2 Automatically adapting parameters

This section lists some of the automatically adapting parameters:
Rate Profile Optimization
Automatic Implant Detection
Capture Management
Sensing Assurance
Lead Monitor
Search AV+
Automatic PVARP

3.2.1 Rate Profile Optimization

When Rate Profile Optimization is programmed On, the pacemaker can adapt ADL and exertion rate response levels once each day by comparing the patient’s current sensor rate profiles against a prescribed target rate profile. This feature is intended to provide automatic and independent ADL rate response for moderate activities such as walking and daily chores, and exertion rate response for vigorous exercise.

3.2.2 Automatic Implant Detection

The Automatic Implant Detection feature does the following:
It determines that the pacemaker has been implanted and that its leads are stable, and it sets lead polarities.
At implant it initializes Sensing Assurance, Capture Management, Rate Response, MVP, and diagnostic data collection.

3.2.3 Capture Management

The Atrial Capture Management and Ventricular Capture Management features check the patient’s pacing thresholds at regular intervals. Using these threshold measurements, the pacemaker can determine whether pacing pulses in that chamber are capturing the heart. Optionally, it makes adjustments to the amplitude and pulse width parameters based on these measurements.
24 Pacemaker Programming Guide
Page 25
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.2.4 Sensing Assurance

The Sensing Assurance feature, when active for a specific chamber, allows the pacemaker to change the sensitivity threshold for that chamber to track changes in the sensed amplitude.

3.2.5 Lead Monitor

This feature monitors lead integrity by measuring and recording lead impedance. Optionally, it can switch either lead (or both leads) from bipolar to unipolar polarity if lead impedance is out of range.

3.2.6 Search AV+

The Search AV+ feature is intended to promote intrinsic ventricular activation in patients with intact or intermittent AV conduction and prevent inappropriate therapy in patients without conduction. Search AV+ is available for dual chamber pacemakers when the operating mode is DDDR, DDD, DDIR, DDI, DVIR, DVI, or VDD. The pacemaker searches for the patient’s intrinsic AV conduction time and adjusts the SAV and PAV intervals either longer or shorter to promote intrinsic activation of the ventricles. When Rate Adaptive AV is active, the pacemaker also adjusts the SAV and PAV intervals relative to the rate adaptive values. If the pacemaker does not observe intrinsic ventricular activation during its periodic searches over the course of a week, it turns off the Search AV+ feature.

3.2.7 Automatic PVARP

When automatic PVARP is programmed, the pacemaker determines a value for the PVARP based on the mean atrial rate (which is an average of all A-A intervals except those starting with an atrial sense or atrial refractory sense and ending with an atrial pace). When the operating mode is DDDR, DDD, or VDD, automatic PVARP is intended to provide a higher 2:1 block rate by shortening the PVARP and SAV (if necessary) at higher tracking rates and protect against PMTs at lower rates by providing a longer PVARP.

3.3 Starting a patient session

Because the programmer collects and stores data on a session-by-session basis, it is important to correctly start and end each session. This section describes how to start a patient session. To end a session, refer to Section 3.14, “Ending a patient session”, page 45.
Pacemaker Programming Guide 25
Page 26
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™

3.3.1 The tool palette during a patient session

Table 2. Tool palette during a patient session

Tool Selecting the tool (button or icon)
Freezes a segment of the live rhythm display.
Accesses the rhythm strips saved since the start of the session.
Opens a window of options for adjusting the live rhythm display.
Opens the Checklist screen for simplified navigation through a set of follow-up tasks. >> navigates to the next task in a Checklist.
Displays the following options for retrieving information about the patient’s pace­maker and its operation and for setting up or clearing the data collection func­tions: Quick Look II - Initial Interrogation Graphs and Tables Battery and Lead Measurements Data Collection Setup/Clear
Displays the Therapy Parameters screen.
SW043
Displays the following pacing system test options: Magnet Underlying Rhythm Threshold Exercise Sensing Temporary EP Studies
Displays the following options for printing and saving generated reports: Available Reports Print Queue
Displays the following options for entering patient information: TherapyGuide Patient Information Key Parameter History
Displays the following options for setting session preferences and saving ses­sion data to a diskette or to a USB flash drive: Preferences Save To Media…
26 Pacemaker Programming Guide
Page 27
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3.3.2 Starting a session with Find Patient…
Perform this procedure to start a Medtronic SW043 programmer software session Find Patient….
Note: Contact your Medtronic representative if your programmer does not have the
Medtronic SW043 programmer software installed on it.
1. With the Select Model screen displayed, position the programming head over the patient’s pacemaker and hold it in place. If telemetry is established with the device, the session starts, initial interrogation begins, and you can proceed to Step 3.
Notes:
If telemetry is not established with the device, a short self-test sequence takes place and the Find Patient screen displays. If this occurs, tap Cancel to display the Select Model screen.
If the programmer is operating, but the Select Model screen is not displayed, tap Select Model.
If Select Model does not display, the previous patient session was not ended properly. Refer to Section 3.14, “Ending a patient session”, page 45.
2. Tap Find Patient… at the bottom of the screen.
After 40 to 50 s of internal software loading, the session starts and initial interrogation begins.
3. Hold the programming head in place until initial interrogation is 100% complete.
Refer to Section 3.4, “Proceeding with session tasks”, page 29 for information on selecting the functions or tasks you have planned for the session.

3.3.3 About automatic interrogation

At the start of a patient session, the programmer automatically attempts to interrogate the patient’s pacemaker to retrieve data that might be needed during the session.
Note: You must position the programming head over the pacemaker and hold it in place until the automatic interrogation is complete.
Clinician-selected diagnostic data is not included in the initial interrogation. The programmer interrogates that data when you choose to view the clinician-selected data on-screen or when you tap Interrogate… and select the Collected Data on the Interrogate How Much? screen.
Pacemaker Programming Guide 27
Page 28
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
The interrogation time may take about a minute because of the amount of information stored in the pacemaker. A status gauge at the bottom of the screen shows progress of the interrogation. Tap Stop at any time to cancel the interrogation.
Note: Except for use of Emergency or Stop, you cannot proceed with session activities until the initial interrogation is 100% complete. If you tap Stop or lift the programming head before the interrogation is complete, the programmer will display a message to interrogate the pacemaker if it requires pacemaker data that was not collected during the interrogation.
You also can manually interrogate the pacemaker at any time during the patient session (see Section 3.5, “Interrogating the pacemaker”, page 32).
Warning messages – A warning message may display in response to an interrogation. Examples of warning messages include:
RRT (Recommended Replacement Time)
ERI (Elective Replacement Indicator)
POR (Power On Reset) or Full Electrical Reset
Interrogation interrupted or unsuccessful
Lead warnings
You must acknowledge the message before proceeding with other activities. For an RRT, ERI, POR, or lead message, you can attempt to clear them by following the instructions in the message. Otherwise, you can close the message box and clear the condition later. For more information about clearing these conditions, see Section 7.1, “Programming parameters”, page 98.

3.3.4 Initial Interrogation Report

Following a successful interrogation, a report based on this interrogation is printed automatically. Session preferences allow you to turn this function on or off. For more information on session preferences, see Section 3.4, “Proceeding with session tasks”, page 29). If the function is on, the same preferences allow you to choose whether or not to include the Arrhythmia Summary and Permanent Parameters sections in the report. Depending on the programmed status of the implanted pacemaker, this report can include the following:
History (clinical conditions for TherapyGuide)
Patient/Pacemaker Information
Pacemaker Status
Last Measured Threshold
28 Pacemaker Programming Guide
Page 29
Medtronic
P and R-Wave Measurements
Threshold Trend
Impedance Trend
Parameter Summary
Clinical Status
Long Term Heart Rate Histograms
Pacing % counters
Event counters
High Rate Episodes Summaries
Arrhythmia Summary, including:
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
– High Rate Episodes Summaries
– V. Rate During Atrial Arrhythmias
– Atrial Arrhythmia Durations
– Atrial Arrhythmia Trend
All Permanent Parameters
SW043

3.4 Proceeding with session tasks

After you begin a patient session, a short period of internal software loading occurs before the first task screen appears.

3.4.1 The first task screen

The first task screen to appear is the Quick Look II screen. See Section 3.6, “Taking a quick look at pacemaker operation”, page 33 for more information.

3.4.2 Selecting another task or function

To proceed with the session, tap the desired task or function from the button and icon options grouped along the edge of the screen. This group of buttons and icons is referred to as the “tool palette.” It enables you to quickly and easily display a desired task or function screen.
Note: When some functions are active, the tool palette is inactivated. The tool palette is reactivated when these function windows are closed.
Pacemaker Programming Guide 29
Page 30
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.4.3 Consider using Checklist

During a follow-up session, you can use the tool palette to select tasks or functions in any order as you proceed through the session. However, if you use a particular follow-up routine or protocol, you can configure and use the Checklist feature to streamline the session. For more information, see Figure 1, page 10.
With Checklist, you can advance quickly to each successive task in a follow-up session. You also can create custom checklists that can be used for task selection during an implant procedure.

3.4.4 Verifying or changing session preferences

At the start of a patient session, or anytime during the session, tap the Session icon to access preference settings that apply to printing reports and viewing waveform traces during test procedures. Preference settings are saved and will not be canceled when the session ends.
3.4.4.1 Printing preferences
Printing preferences let you set up how reports are printed when you tap Print…. You can choose to display the Print - Options window, print the report immediately, or add the report to the print queue. You can also enable or disable the printer Trace Mode. With the Trace
Mode enabled, the programmer prints programmed variables and their values.
3.4.4.2 How to set up printing preferences
Perform this procedure to set up your printing preferences:
1. Tap Session > Preferences.
2. Select the check box for Pop up these options when any Print button is selected as follows:
If a check mark appears, the print setup window will appear each time you tap Print….
If there is no check mark, Print… immediately prints the report or transfers it to the print queue.
3. Tap the Number of copies: field to select 1, 2, 3, or 4 copies to be printed.
4. Select one of the following options for Printer:
Full Size: This option requires that a compatible, external printer be connected to the programmer.
30 Pacemaker Programming Guide
Page 31
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Note: Tap the Printer field and choose a printer from the list. The printer you select must be compatible with the input requirements of your connected printer.
Programmer Strip: This option prints reports on the programmer’s strip chart printer.
5. Select one of the following options for printing the report:
Now: The report prints immediately when you tap Print….
Later from Print Queue: The report is sent to the print queue when you tap Print…. For information about the print queue, see Section 3.12, “Printing reports”,
page 41.
6. Select the check box for Print continuously (Trace Mode) as follows:
If a check mark appears, the programmer prints programmed variables and their values. The data in this case is not printed in a report format.
If there is no check mark, the programmer does not print automatically.
3.4.4.3 Initial Report preference
The Initial Report preference lets you choose whether or not an initial interrogation report is automatically printed at the start of each patient session following a successful interrogation.
3.4.4.4 How to select your Initial Report preference
Perform this procedure to set up your Initial Report preferences:
1. Tap the Session > Preferences > Initial Report.
2. Select the check box for Always print at session start as follows:
If a check mark appears, the programmer automatically prints the initial interrogation report.
If there is no check mark, the programmer does not automatically print the initial interrogation report.
3. Select the Include: check boxes if you wish to include an Arrhythmia Summary, the Permanent Parameters (this includes programmed parameter values), or both.
Pacemaker Programming Guide 31
Page 32
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3.4.4.5 Tests preference
The Tests preference lets you choose how waveform traces are displayed during a selected follow-up test. You can choose whether the live rhythm display automatically displays the EGM for the heart chamber being tested (atrium or ventricle) or does not change the arrangement of the rhythm traces.
3.4.4.6 How to select your preference for the trace display
Perform this procedure to set up your trace display preferences:
1. Tap Session > Preferences > Tests .
2. Select either Switch EGM to match selected test chamber or Do not change
waveform arrangement.

3.5 Interrogating the pacemaker

At the start of the patient session the programmer performs an automatic interrogation to retrieve most of the information stored in the pacemaker.
You can also manually interrogate the pacemaker at any time during the patient session. In some cases, a pop-up window may request that you do so.

3.5.1 How to interrogate the pacemaker

1. Tap Interrogate... or press the I button on the programming head.
2. From the window of options, select the type of information you want to retrieve. You can select more than one option.
Collected data can be interrogated only once during a patient session. (AGAIN) indicates that this data was retrieved in a previous interrogation for this session.
3. Position the programming head and tap Start or press the I button on the programming head.
Hold the programming head in place until the interrogation is 100% complete as shown by the status gauge at the bottom of the screen.
Tap Stop to cancel the interrogation.
32 Pacemaker Programming Guide
Page 33
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.6 Taking a quick look at pacemaker operation

Quick Look II provides a summary or overview of pacemaker operation by displaying the essential information on one screen. This data was retrieved during the initial interrogation of the session.

3.6.1 Quick Look II screen

The Quick Look II screen appears at the start of a patient session or when you tap Data > QuickLook II.
Note: Each >> on the Quick Look II screen is a link to diagnostic data related to the
information with which it is associated.

3.6.2 Viewing data on the Quick Look II screen

The Quick Look II screen displays the following information and information access options.
Remaining Longevity – This is an estimate of the time in years or months remaining until pacemaker replacement is required. This estimate is based on the programmed parameter settings and events recorded by the pacemaker since the last patient session. For more information, tap the information icon to the left of the Remaining Longevity label to read the
Explanation of Remaining Longevity Estimates.
Threshold trend – The graph shows the average weekly atrial and ventricular pacing
thresholds over the last year. If Capture Management is enabled, the last measured threshold is shown in the box to the right of the graph. Tap >> to view threshold trend data in the Collected Data - Capture Management Trend window. For more information about Capture Management Trends, see Section 5.3, “Automatically collected data displays”, page 63.
Impedance trend – The graph shows the average weekly atrial and ventricular measured impedances over the last year. The lead impedances measured during initial interrogation are shown in the box to the right of the graph. Tap >> to view lead impedance trend data in the Collected Data - Lead Impedance window. For more information about Impedance Trends, see Section 5.3, “Automatically collected data displays”, page 63.
P Wave – If Atrial Sensing Assurance is on, this shows the current P Wave amplitude range measured by the pacemaker within the last week. Tap >> to view P Wave amplitude trend data in the Collected Data - Sensitivity Trends window. For more information about P Wave amplitude trend, see Section 5.3, “Automatically collected data displays”, page 63.
R Wave – This shows the current R Wave amplitude range if Ventricular Sensing Assurance is on. Tap >> to view R Wave amplitude trend data in the Collected Data - Sensitivity
Pacemaker Programming Guide 33
Page 34
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Trends window. For more information about R Wave amplitude trend, see Section 5.3, “Automatically collected data displays”, page 63.
Mode – This shows the currently programmed mode.
Lower Rate – This is the slowest rate at which pacing occurs during a mode’s basic
operation. In rate-responsive modes, in the absence of sensor-detected activity, the sensor-indicated rate is equal to the programmed lower rate.
Upper Track Rate – This is the maximum rate at which the ventricle may be paced in response to sensed atrial events in the DDDR, DDD, and VDD modes.
AT/AF – This shows the average number of hours each day in which atrial arrhythmia episodes are occurring and the total percentage of patient time that is spent in atrial arrhythmias. Tap >> to view atrial arrhythmia trend data in the Collected Data - Atrial Arrhythmia Trend screen. For more information about the atrial arrhythmia trend, see Section 5.3, “Automatically collected data displays”, page 63.
Pacing – The pacing and sensing event sequences applicable to the programmed pacing mode (see Table 3) are listed under this label. These are shown as the percentage of events recorded since the last patient session.
An indicator following the event sequences shows whether the pacemaker was operating in one of the following conditions between the previous session and the current session:
MVP On
MVP Off
Search AV+ On
Note: These messages are not changed by programming during the current session.
Tap >> to view the Collected Data - Heart Rate Histograms window. For more information about the histogram displays, see Section 5.3, “Automatically collected data displays”, page 63.
Note: The information under the Pacing label and the Observations label shows data collected since the last patient session. This data is cleared from pacemaker memory after the session has ended. After the data is cleared, it cannot be recalled.

Table 3. Event sequence categories

Dual chamber pacing modes (and ADI, ADIR, VDI, VDIR, VDD)
AS-VS Atrial Sense - Ventricular Sense AS-VP Atrial Sense - Ventricular Pace AP-VS Atrial Pace - Ventricular Sense AP-VP Atrial Pace - Ventricular Pace
34 Pacemaker Programming Guide
Page 35
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Table 3. Event sequence categories (continued)
Single chamber pacing modes
Paced Atrial or Ventricular Pace Sensed Atrial or Ventricular Sense
Observations – Observations summarize diagnostic data collected since the last patient
session. Each observation reports the number and category of significant events recorded by diagnostic monitoring functions (see Table 4).
To view a graph or table showing the details of an observation, tap the observation then tap >>. Refer to Section 5.2, “Viewing the collected data”, page 62 for information about viewing the data recorded by the monitoring functions.

Table 4. Observation monitoring functions

Functions and criteria used for observation reporting
Atrial Lead Monitor Lead impedance outside Min/Max settings Ventricular Lead Monitor Lead impedance outside Min/Max settings Atrial High Rate Episodes One or more episodes detected Ventricular High Rate Episodes One or more episodes detected Rate Drop Response Episodes One or more episodes detected RRT/ERI Monitor Occurrence of RRT/ERI conditions Electrical Reset Monitor Occurrence of reset conditions Capture Management High ventricular or atrial threshold measurement
Abort threshold searches prior to RRT/ERI Capture Management unable to run

3.7 Viewing Battery and Lead Measurements

Tap Data > Battery and Lead Measurements to view the Battery and Lead Measurements screen. This screen captures information about the pacemaker battery and
the lead system based on real-time measurements and calculations made at the time of pacemaker interrogation.
Note: The values measured for the pacemaker battery and the lead system can change between interrogations.
Pacemaker Programming Guide 35
Page 36
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.7.1 Pacemaker Battery and Lead Measurements

The Battery and Lead Measurements screen includes the following information.
Battery Status – Displays OK or Replace Pacer based on battery voltage and internal impedance measurements. The Date of Implant is shown below the battery status.
Remaining Longevity – This is a calculated estimate of the time remaining until pacemaker
replacement will be required. This calculated estimate is based on the programmed parameter settings and event data accumulated by the pacemaker since the previous patient session.
Tap the information icon to the right of the Remaining Longevity label to read the
Explanation of Remaining Longevity Estimates.
Caution: Pacemaker replacement should not be based on the estimated remaining
longevity. For this decision, use only the RRT (Recommended Replacement Time), ERI (Elective Replacement Indicator), or the REPLACE PACER battery status message.
Battery Voltage, Current, Impedance – These values show the measured pacemaker battery voltage, the present current drain on the pacemaker battery averaged over a pacing cycle, and the battery’s impedance.
Atrial Lead and Ventricular Lead – Lead information includes the following:
Amplitude The measured amplitude of a pacing pulse Pulse Width Present programmed pulse width setting Output Energy Output energy contained in a single pacing pulse Measured Current The measured current in the pacing lead during delivery of a pacing pulse Measured Impe-
dance Pace Polarity The present lead electrode configuration (unipolar or bipolar) used for pacing
The measured electrical impedance presented by the pacing lead and elec­trode/tissue interface

3.7.2 Updating the displayed data

It takes a few seconds to update the battery and lead measurements data. Perform this procedure:
1. Position the programming head and hold it in place.
2. Tap Measure Again.
3. Hold the programming head in place for the duration of measurement as indicated in the Measure - In Progress pop-up window.
36 Pacemaker Programming Guide
Page 37
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.8 Checking the present parameter settings

Tap Params to view the Parameters - Therapy screen. This screen displays the pacemaker’s programmed parameter settings.
If the label for a parameter or option includes an ellipsis (for example, Rate Response…), there are subordinate parameters that are not displayed on this screen.
For more information and a table of subordinate therapy parameters, see Section 7.1.5, “How to program subordinate parameters”, page 101.

3.9 Viewing and programming patient information stored in the pacemaker

Pacemakers can store patient-related information that you can view and print during a patient session. This information typically is programmed into the pacemaker at the time of implant, but it can be revised at any time.
When you program the patient’s clinical conditions (Age and History) into pacemaker memory, they are available to the TherapyGuide feature. Refer to Section 7.2, “Using TherapyGuide to select parameter values”, page 106.
Note: Tap Patient > TherapyGuide to display TherapyGuide window.

3.9.1 How to program patient information

1. Tap Patient > Patient Information
2. On the Patient Information screen, tap the data field in which you want to enter or change the data.
This action displays selectable options or the on-screen keyboard for entering data.
3. Tap the desired information option, or type the desired information, and then tap Enter on the on-screen keyboard.
4. Repeat steps 1 and 2 for any field to which you want to add, delete, or change the data.
5. To define the patient’s clinical conditions, which are made available to TherapyGuide, take the following actions:
Tap the data field for Age, and tap the patient’s age range.
Tap the History… field to open the History window. Tap the appropriate clinical conditions, then tap OK.
Pacemaker Programming Guide 37
Page 38
Medtronic
6. When you have finished, position the programming head and tap PROGRAM.
This action programs into the pacemaker all the data that displayed on the Patient Information screen.
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.10 Recording an ECG Strip of magnet operation

Positioning the programming head does not cause the pacemaker to operate in its magnet mode. To record or view magnet operation, you must use the Magnet test.

3.10.1 Magnet test

The Magnet test allows you to start and stop magnet operation in the pacemaker. Options let you collect an ECG strip of magnet operation and, if desired, non-magnet operation.
Note: At any time during a Magnet test, lift the programming head from over the patient’s pacemaker for at least 2 s to restore pacemaker operation to its permanent status. Take this action in the event of programmer malfunction, loss of power, or the absence of an appropriate command confirmation.

3.10.2 Procedure for conducting a Magnet test

1. Tap Tests > Magnet to display the Magnet - Test Setup screen.
2. To start magnet operation, position the programming head and tap START Magnet.
Observe that the live rhythm display shows the point at which magnet operation begins (“DOO, 85 bpm” for example). At this point, the pacemaker performs a Threshold Margin Test.
If you have chosen to collect an ECG strip (see Section 3.10.3), a pop-up window shows progress during the collection process. Tap Stop Collection if you want to stop the ECG collection before it completes. Tapping this button does not stop magnet operation.
3. To stop magnet operation, tap STOP Magnet.
A pop-up window gives you the option to collect an ECG strip of non-magnet operation. Tap Yes to collect a non-magnet strip or tap No to close the window.
38 Pacemaker Programming Guide
Page 39
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.10.3 Collecting an ECG strip

The programmer automatically collects an ECG strip during the Magnet test unless you cancel this option.
A check mark appearing in the Magnet Strip check box indicates that a strip will be collected. This is the default status when you open the screen.
To change the length of the strip, tap the time field and choose the collection time you desire.
If you do not want automatic strip collection, tap the Magnet Strip check box to clear the check mark.

3.10.4 About the collected ECG strips

If you have chosen to collect a Magnet strip or both a Magnet and Non-Magnet strip, the strips are stored by the programmer for viewing and printing. A Magnet Strip and Non-Magnet Strip icon appears at the bottom of the screen to indicate when strips are available for viewing and printing.
To view a collected ECG strip, tap either icon. Refer to Section 4.3, “Waveform strips”, page 56for information about using the strip viewing feature and about printing waveform strips and reports.
To print a Magnet test report, tap Print… and then choose the desired print options.

3.11 Checking the patient’s underlying rhythm

To evaluate a patient’s underlying rhythm or determine the patient’s intrinsic heart rate, use the Underlying Rhythm test. This test provides two instruments to evaluate the patient’s underlying rhythm:
the Inhibit test
the Manual Rate Decrease test
Caution: The use of this test function is intended for diagnostic and test purposes. It should be used only under conditions of careful patient monitoring and control.
Note: At any point during the Underlying Rhythm test, lifting the programming head away from the site of the pacemaker for at least 2 s will restore normal pacemaker operation. Lift the programming head immediately in the event of a programmer malfunction or loss of power.
Pacemaker Programming Guide 39
Page 40
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.11.1 Procedure for checking the patient’s underlying rhythm

Note: The Inhibit mode shuts off pacemaker output for the test duration.
3.11.1.1 How to check underlying rhythm using Inhibit
1. Tap Tests > Underlying Rhythm to display the Underlying Rhythm Test (Manual Rate Decrease or Inhibit) screen.
2. To stop the output of pacing stimuli:
a. Position the programming head.
b. Tap and hold INHIBIT Press and Hold.
Pacemaker blanking periods are shortened during this test to increase the period during which cardiac events can be sensed.
3. To restore permanent pacing, release INHIBIT Press and Hold.
3.11.1.2 How to check underlying rhythm using manual rate decrease
1. Tap Tests > Underlying Rhythm to display the Underlying Rhythm Test (Manual Rate Decrease or Inhibit) screen.
2. Choose the desired test pacing mode and the starting value for the lower rate:
a. Tap the Mode field under the Test Value label to display the test mode options.
b. Tap on the desired test mode from the Temp. Mode list.
c. Tap the Lower Rate arrows to adjust the test value to a rate setting close to the
patient’s present pacing rate.
3. Tap START Test to engage the test mode and lower rate.
4. Tap the down arrow to gradually decrease the lower rate until the patient’s underlying rhythm emerges. (Tap the up arrow to increase the lower rate if desired.)
5. To end the test, tap STOP and Restore.

3.11.2 Collected ECG strips

During either type of Underlying Rhythm Test, the programmer automatically collects and saves a 10 s ECG strip. A Test Strip icon appears near the bottom of the screen (after the test has ended) to indicate its availability.
40 Pacemaker Programming Guide
Page 41
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
To view a collected ECG strip for either the Manual Rate Decrease test or the Inhibit test, tap the Test Strip icon for the respective test. Refer to Section 4.3, “Waveform strips”, page 56 for information about using the strip viewing feature and about printing waveform strips and reports.
To print an Underlying Rhythm Test report for either the Manual Rate Decrease test or the
Inhibit test, tap Print… for that test, then choose the desired print options.

3.12 Printing reports

You can print data generated during a patient session by tapping Print… or by selecting a report from the Available Reports window. You can also print reports that are in the print queue.
3.12.1 Using Print…
On most task screens, Print… becomes active when a task is executed or a function generates report data. Tap Print to print a formatted data report associated with the current task. Print… appears near the bottom of all task screens except the Parameters - Therapy screen.
Note: On the Parameters - Therapy screen, an icon of a printer can be found to the right of Undo.
Tap Print… or the printer icon to perform one of the following tasks, as determined by your session preferences settings (see Section 3.4.4):
print the associated report.
send the report to the print queue to be printed later.
open the Print - Options window (or the Output Pacemaker Memory window for the Device Memory Report).
For information about session preferences and print options, see Section 3.4, “Proceeding with session tasks”, page 29.

3.12.2 Printing from the Available Reports window

During a patient session, the programmer automatically generates and saves the following reports:
The Initial Interrogation report
The Final report
Pacemaker Programming Guide 41
Page 42
Medtronic
Reports generated by the use of the Tests functions.
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Note: If a test function is used more than once, a report for only the most recent use is available.
Device Memory report (This is a special report intended for use only by Medtronic technical support personnel.)
At any time during the session, you can view the list of reports that are available for printing, and select the ones you want to print immediately or save in the print queue for printing later in the session or after the session has ended.
3.12.2.1 How to print from the list of Available Reports
1. Tap Reports > Available Reports.
2. From the list of available reports, tap the report you wish to print.
3. Tap Print… > Print Now or Print… > Print Later. This will place the report in the print queue.

3.12.3 Printing reports held in the Print Queue

When you tap Print… > Print Now or Print… > Print Later to print a report, control of the print job is passed on to the print queue shown in the Session Print Queue window.
The Session Print Queue window shows the status of any print job. You can either print or delete print jobs in the queue. All print jobs held for printing are displayed, and up to 25 print jobs already printed are reported.
Note: When you end a patient session, the print queue is still available. It lists any reports held from that session and previous sessions.
3.12.3.1 How to print or delete reports from the Print Queue
1. Tap Reports > Print Queue .
2. From the list of reports, tap the report you want to print or delete.
Caution: A deleted report cannot be recovered.
3. Tap Print to print the report, or tap Delete to delete the report from the Session Print Queue report list.
42 Pacemaker Programming Guide
Page 43
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.13 Saving session data

Any time during a patient session, you can save data from the session on a diskette or on a USB flash drive.
The saved data is listed in Table 5. Saved session data can be reloaded on the programmer for analysis.

Table 5. Saved session data

Feature name Information exported
Therapy Parameters Initial interrogated values
Last programmed values Patient Information Last programmed values Battery and Lead Measurements Last measured values Threshold Tests
Sensing Tests
Exercise Test Results for the last test conducted Automatic Diagnostics Event Counters
Device Memory Retrieved from interrogation performed only for sav-
a
Manual test results are saved only if you have saved the results.
a
a
Last results for each test type conducted (for each
chamber tested)
Last results for each test type conducted (for each
chamber tested)
Atrial High Rate Episodes
Ventricular High Rate Episodes
Mode Switch Episodes
Rate Drop Response Episodes
ing session data

3.13.1 How to save session data on a diskette

The diskette drive is functional only when no USB flash drive is inserted in the programmer. When no flash drives are inserted, the diskette indicator is green to indicate that the diskette drive is available for use. Inserting a diskette allows all session data and PDF file saving operations to take place on the diskette.
1. Tap Session > Save to Media… to display the Save To Media - Insert Media window.
Note: A similar procedure allows you to save session data when ending a patient session, but it requires that you end the session immediately after doing that. For more information, refer to Section 3.14, “Ending a patient session”, page 45.
2. Insert a formatted, IBM-compatible, 90 mm (3.5 inch) diskette into the diskette drive.
Pacemaker Programming Guide 43
Page 44
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3. Position the programming head and tap Start.
This initiates an interrogation and then starts the transfer of data to the media. The programming head must be held in position until the interrogation is complete. Depending on the amount of session data being saved, this process may take several minutes. A gauge on the screen indicates the progress of the data transfer to the media.
Note: If the diskette is full, the programmer will notify you to insert another diskette.
4. When the transfer of data is complete, remove the diskette from the programmer.
The programmer automatically assigns a unique file name so that it cannot overwrite any existing session file. To determine which file was saved for a particular session, use an IBM-compatible computer to view the media file directory. The directory lists the date and time each file was created.
You can continue with the patient session after saving the data.

3.13.2 How to save session data on a USB flash drive

When a USB flash drive is inserted in the programmer, the diskette drive becomes unavailable. Diskettes may be used when no USB flash drive is inserted.
Any operation that normally uses a diskette to read or write data (such as saving session data, reloading session data, or saving reports to a PDF file) uses a USB flash drive if it is inserted.
Insert only one writable flash drive at a time. Inserting additional flash drives results in an error during data-saving operations. This condition is indicated by the USB indicator being grayed out.
Notes:
While a Save to Media action is in progress, the progress indicator and the message Saving Session… display. The progress indicator displays the completion percentage. Before removing the flash drive, wait a few seconds after the progress indicator shows 100%.
To ensure the integrity and security of patient information, use a flash drive that is reserved to store programmer data.
A USB flash drive should be inserted only if the programmer is powered on. Insert a writable USB flash drive in the programmer using any available USB port. A slight delay may occur while the flash drive is authorized. The USB indicator on the task bar will turn green to indicate that the USB flash drive is available for use, and the diskette icon will become grayed out to indicate that the diskette drive is not available.
44 Pacemaker Programming Guide
Page 45
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Flash drives should not be inserted or removed during the following operations:
programming a device
saving data to or reading data from a diskette
saving a report as a PDF file

3.14 Ending a patient session

Note: Before you end a patient session, print a copy of the session data in case the session
data is accidentally cleared without first being saved to media. For more information, refer to Section 3.12.
Note: The pacemaker will be unresponsive to the application of a magnet for one hour following the use of a programmer unless the session is ended with the command option to immediately clear data collected in the pacemaker. The default command for ending a session allows the pacemaker to retain collected data for one hour, during which use of a magnet has no effect. See Section 3.14.3.
Because the programmer collects and stores data on a session-by-session basis, it is important to correctly end a session when you finish. You should not begin using the programmer with a different patient until you have completed the procedure described in this section.
Note: If the programmer detects a serial number that is different from the one acquired during the initial interrogation, it will automatically end the current session.
Once you end a session, you can still access the Print Queue window to print reports from the session. Prior to ending the session, you have the following options:
Save the session data on a diskette.
Save the session data on a USB flash drive.
Change the way the pacemaker clears collected data.

3.14.1 How to end a patient session

1. Tap End Session… in the command bar at the bottom of the screen. The End Session window is displayed.
2. Note the Warnings! field at the top of the End Session window. It informs you of any action you should take or any information you should know before you end the session. To return to the session, tap Cancel.
Pacemaker Programming Guide 45
Page 46
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3. Before ending the session, consider the options available from the End Session window.
To save the session data on a diskette, see Section 3.14.2.1.
To save the session data on a USB flash drive, see Section 3.14.2.2.
To change the settings for how collected data is cleared from the pacemaker when you end the session, see Section 3.14.3.
Note: To accomplish any of these tasks, proceed to its listed procedure. Do not tap End Now in Step 4 at this point if you want to save the session data or change data
collections settings.
4. To end the session now and return to the Select Model screen, tap End Now.

3.14.2 Saving session data

Before you end a patient session, you can save data from the session on a diskette or USB flash drive. For a list of the data saved, see Section 3.13, “Saving session data”, page 43. Saved session data can be reloaded on the programmer for analysis. For information about reloading session data, refer to Section 3.13, “Saving session data”, page 43.
3.14.2.1 How to save session data on a diskette
Note: When using this procedure, you must end the patient session after saving session
data on a diskette. If you choose to initiate the save by selecting the Session icon, you can continue the session (see Section 3.13, “Saving session data”, page 43).
1. Insert a 90 mm (3.5 inch) computer diskette into the disk drive on the right side of the programmer. Use a formatted diskette that is IBM-compatible.
2. From the End Session window (see Section 3.14.1), tap Save To Media….
Note: If the diskette is full, the programmer will notify you that you need to insert a different diskette.
46 Pacemaker Programming Guide
Page 47
Medtronic
3. Position the programming head and tap Start from the Save To Media window.
This initiates an interrogation and then starts the transfer of data to the diskette. The programming head must be held in position until the interrogation is complete. Depending on the amount of session data being saved, this process may take several minutes. A gauge on the screen indicates the progress of the data transfer to the diskette.
Note: After session data has been saved, you can change the options for clearing data (see Section 3.14.3). However, you cannot cancel the End Session sequence and return to the session.
4. When the transfer of data is complete, tap End Now to end the patient session. Remove the diskette from the programmer.
Note: The programmer automatically assigns a unique file name so that it cannot overwrite any existing session file. To determine which file was saved for a particular session, use an IBM-compatible computer to view the diskette file directory. The directory lists the date and time each file was created.
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3.14.2.2 How to save session data on a USB flash drive
Note: When using this procedure, you must end the patient session after saving session
data on a flash drive. If you choose to initiate the save by tapping the Session icon, you can continue the session (see Section 3.13, “Saving session data”, page 43).
1. Insert a USB flash drive in the programmer.
2. From the End Session window (see Section 3.14.1), tap Save To Media….
3. Position the programming head and tap Start from the Save To Media window.
This initiates an interrogation and then starts the transfer of data to the USB flash drive. The programming head must be held in position until the interrogation is complete.
Note: After session data has been saved, you can change the options for clearing data (see Section 3.14.3). However, you cannot cancel the End Session sequence and return to the session.
4. When the transfer of data is complete, tap End Now to end the patient session.
Note: The programmer automatically assigns a unique file name so that it cannot overwrite any existing session file. To determine which file was saved for a particular session, use an IBM-compatible computer to view the contents of the flash drive. The directory lists the date and time each file was created.
Pacemaker Programming Guide 47
Page 48
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

3.14.3 Clearing pacemaker data

By default, the diagnostic data collected in the implanted pacemaker is cleared one hour after you end a patient session. Collection of new data then starts. You can restart a session within one hour of ending it and still retrieve the collected data. Alternatively, you can select Now in the Clear Data window to clear the collected data and start collection of new data immediately when you end the session and lift the programming head. For more information, see Section 5.6, “Clearing data from the pacemaker”, page 80.
48 Pacemaker Programming Guide
Page 49
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

4 Viewing the patient’s ECG and EGM traces

4.1 Viewing the ECG and other rhythm waveforms

During a patient session, the programmer continuously displays the patient’s ECG (or other selected waveform trace) in the Live Rhythm Monitor window at the top of the screen.
On the programmer, the size of the window automatically changes between a dual-trace size and a triple-trace size.
The instructions on the following pages describe how you can:
Expand the monitor window to its full-screen size.
Adjust the size, position, and color of the waveform traces.
Choose a different sweep speed, display pacing artifacts or turn them off, and engage an interference filter.
Program pacemaker telemetry to a different EGM mode.

4.1.1 Expanding the Live Rhythm Monitor window

To view all of the available waveform traces, you can expand the Live Rhythm Monitor window to its full size with the tap of a button. The expanded screen covers the task area of the current screen.
To expand the Live Rhythm Monitor window to its maximum size, tap the maximize button in the upper-right corner of the partial-view window, displayed to the left of Freeze.
To collapse the Live Rhythm Monitor window to its minimum size, tap the minimize button in the upper-right corner of the full-view window, displayed to the left of Freeze.

4.1.2 About the waveform traces

The programmer can display the traces described in Table 6. The ECG Leads (I, II, and III) are always available if the ECG leads are connected. Marker Annotation and Marker Intervals are displayed if the programming head is positioned over the pacemaker.
The telemetry modes for the traces are programmable. For information about changing telemetry modes, see Section 4.2, “Adjusting and arranging the waveform traces”, page 51.
Pacemaker Programming Guide 49
Page 50
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Note: A programming command or interrogation momentarily interrupts the transmission of marker annotations. This interruption can result in missing markers on the trace display.

Table 6. Waveform trace information

Trace Description
ECG Lead I ECG Lead II ECG Lead III
Marker Annotation
Marker Intervals
a
(ms)
Programmable Trace (Telemetry Mode)
a
Since these traces depend on telemetry received from the pacemaker, they are not displayed unless the programming head is positioned over the pacemaker.
ECG signals are detected via skin electrodes attached to the patient. The programmer must be connected to these electrodes with the ECG cable as described in Section 2.4, “Connecting the programmer to skin elec­trodes”, page 16.
a
Marker annotations depict pacemaker operation by showing events as they occur within the pacemaker. These annotations, which are coded as shown below, are intended to facilitate ECG interpretation. The typical position of the Marker Annotation trace is superimposed on an ECG trace.
Dual Chamber Models
AP - Atrial pace AS - Atrial sense AR - Atrial sense during refractory
Single Chamber Models
P - Pace S - Sense
SR - Sense during refractory VP - Ventricular pace VS - Ventricular sense VR - Ventricular sense during refractory
Other Markers
TP - Triggered pace (AAT or VVT pacing mode) MS - Mode Switch episode (marks the beginning and end) ER - Error marker (indicates a marker signal that could not be decoded)
The programmer automatically measures the interval between pace and sense markers and displays these intervals (in milliseconds) as one of the traces. For dual chamber and ADI, ADIR, VDI, and VDIR modes, the trace displays the A–V interval and the V–V interval. For single chamber modes, the trace displays the A–A or V–V interval depending on the chamber being paced.
The Telemetry Mode feature lets you program the pacemaker to transmit one of the following types of EGM signals to be displayed in the Live Rhythm Monitor window:
- Atrial EGM
- Ventricular EGM
- Dual EGM
- Summed EGM Note: The Telemetry Mode feature is not available when the pacemaker is at RRT/ERI (Recommended Replacement Time). Also, the Telemetry Mode feature can be programmed Off.
50 Pacemaker Programming Guide
Page 51
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

4.2 Adjusting and arranging the waveform traces

Tap Adjust… to display a button bar associated with each waveform trace (Figure 7) and the options for viewing waveform traces. This button bar provides the controls for tailoring the Live Rhythm Monitor window.

4.2.1 Changing trace size, source, and print status

The button bar appearing below the source title of each waveform trace provides controls for adjusting the waveform size, selecting the trace source, and choosing whether or not the trace will appear on the printed chart recording.

Figure 7. Waveform trace adjustment buttons

1 Increases waveform size 2 Returns waveform to its default size 3 Decreases waveform size
4 Displays options for selecting the trace
source
5 Indicates the 2 waveform traces selected for
printing (chart recorder)
4.2.1.1 Adjusting trace size (amplitude)
1. Tap the up or down arrow on the associated button bar to increase or decrease the size (or amplitude) of the trace waveforms. ECG and EGM traces can be adjusted by a maximum of 4 steps.
2. Tap the return button to return the waveform trace to its default size.
4.2.1.2 Changing the trace source
You can change the order in which traces appear in the monitor window. One method is by changing the “source” of one or more of the traces as described below. You also can use the touch pen to “drag” a trace to the desired location as described on Section 4.2.9.
Pacemaker Programming Guide 51
Page 52
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
4.2.1.3 How to change the source of a displayed trace
1. Tap the trace source button from the associated button bar.
Note: Traces can be superimposed on each other. For example, both ECG Lead II and Marker Annotation sources can display as a composite trace, but the name of only
one source appears on the screen.
Tap the trace name to see the trace name behind it. Tap the trace name again to see that it changes each time you tap.
2. Select the desired source from the drop-down menu of Source options.
The trace for the newly selected source appears in place of the previously displayed trace. If the monitor window is fully expanded, note that the original trace and the selected trace have switched positions.
Note: You can also rearrange the trace positions by dragging them with the touch pen. Refer to Section 4.2.9.
4.2.1.4 Selecting traces to be printed by the chart recorder
Initiating a continuous trace recording on the programmer chart recorder results in a printout of Marker Channel telemetry and 2 waveform traces based on selections made as described below. Only ECG and EGM traces can be selected for printing.
4.2.1.5 How to select an ECG or EGM trace to print

Figure 8. Buttons to enable and disable trace printing

1. Tap the printing enabled button next to the print-enabled trace that you do not want to print. This disables printing for that trace.
2. Tap the printing disabled button next to the trace you want to print. This enables printing for that trace.
52 Pacemaker Programming Guide
Page 53
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

4.2.2 Displaying additional adjustment options

The Adjust… window contains additional controls that let you tailor the Live Rhythm Monitor to your specific needs. Once certain display attributes have been adjusted, the programmer will maintain these settings from one patient session to another.
Tap Adjust… near the top of the tool palette. This action:
Expands the Live Rhythm Monitor window (if it is not already expanded).
Displays options for changing the trace color coding.
Opens the Adjust… window.
Instructions in the following sections describe using the trace color options and each of the options in the Adjust… window.
Table 7. The Adjust… window options
Clipping Select this check box to truncate the tops and bottoms of waveform traces
at a 22 mm boundary (see Section 4.2.4). A check mark indicates that this option is selected.
ECG Filter Select this check box to turn the ECG filter on or off. A check mark indicates
that the filter is on (see Section 4.2.5).
Show Artifacts Select this check box to enable or disable the pacing artifact enhancement
function. A check mark indicates that the function is enabled (see Sec­tion 4.2.6).
Sweep speed: Tap the Sweep speed: field to display options for changing the sweep
speed (see Section 4.2.7). Sweep speed options are 12.5, 25, 50, and 100 mm/s.
Telemetry Mode: Tap the Telemetry Mode: field to display options for programming the type
of waveform telemetry transmitted by the device (see Section 4.2.8).
Normalize Tap Normalize to equalize the trace spacing and to adjust the size of each
trace to the default setting.
Calibrate symbol Tap the calibrate symbol to calibrate the external device connected to the
analog output.
OK Tap OK to close the Adjust… window.

4.2.3 Color coding the waveform traces

The Live Rhythm Monitor with the Adjust… window open provides options for color coding the displayed traces as described below.
1. Tap Adjust… to open the Adjust… window.
2. Tap the color field for the trace you want to change.
3. Select the desired color from the drop-down menu of color options.
Pacemaker Programming Guide 53
Page 54
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

4.2.4 Enabling or disabling waveform clipping

The Clipping option in the Adjust… window vertically truncates the waveform trace, allowing you to limit the vertical space that a trace can occupy on the screen and on the chart recording. With the Clipping option selected, the vertical space allotted to each trace is 22 mm on the screen and 25 mm on the printout. Waveforms that exceed this limit will be “clipped.”
To enable or disable waveform clipping, select the Clipping check box in the Adjust… window. A check mark indicates that the clipping option is enabled.

4.2.5 Using the ECG Filter option

In the presence of interference, the ECG filter may improve clarity of both the displayed and printed ECG. The filter affects the ECG detection frequency bandwidth as follows:
Filter Off (no check mark) – Bandwidth = 0.05 to 100 Hz
Filter On (check mark) – Bandwidth = 0.5 to 40 Hz
To turn the ECG filter on or off, select the ECG Filter check box in the Adjust… window. A check mark indicates that the filter is turned on.

4.2.6 Enabling or disabling artifact enhancement

The Show Artifacts option allows you to enable or disable the artifact enhancement function, which shows the position of each pacing stimulus on the ECG traces. To enable or disable the Show Artifacts option, select the Show Artifacts check box in the Adjust… window. A check mark indicates that this function is enabled.

4.2.7 Selecting a different sweep speed

From the Adjust… window, you can set the trace sweep speed to one of four settings: 12.5, 25, 50, and 100 mm/s. These sweep speeds apply only to the Live Rhythm Monitor display. The default setting is 25 mm/s.
1. Tap the Sweep Speed: field in the Adjust… window.
2. From the drop-down menu of options, select the desired sweep speed.

4.2.8 Programming a different telemetry mode

In contrast to the traces that automatically appear in the Live Rhythm Monitor window (the ECG, Marker Annotation, and Marker Intervals traces), the EGM trace can be programmed to any one of the Telemetry Mode: options (see Table 8).
54 Pacemaker Programming Guide
Page 55
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
1. Position the programming head over the patient’s implanted device. Observe that marker annotations are appearing on the screen.
2. Tap the Telemetry Mode: field in the Adjust… window.
3. From the drop-down menu of options, select the desired EGM telemetry (see Table 8).

Table 8. Programmable telemetry modes

Atrial EGM The intracardiac signal detected in the atria by the atrial pacing lead Ventricular EGM The intracardiac signal detected in the ventricle by the ventricular pacing
lead
Dual EGM Display atrial and ventricular EGM traces. Since the device transmits both
EGMs over a single telemetry channel in a “time-sharing” fashion, resolution of the EGM signals in this case is reduced.
Summed EGM The atrial and ventricular EGM signals combined (added together) into a
single trace.
Off No EGM trace is displayed.
Note: For single-chamber models, the mode choices are EGM or Off.

4.2.9 Arranging waveform traces using the touch pen

In addition to arranging waveform traces by selecting the source as described in Section 4.2.1.3, you can use the touch pen to “drag” a trace to the desired position. The procedure in the following example shows how to move the Marker Annotation trace from its position over the ECG Lead II trace to a position over the Ventricular EGM trace.
1. First decide which trace you want to move. If the trace name is not displayed (for example, Marker Annotation), tap the name of the superimposed trace to display the hidden name.
2. Press and hold the touch pen against the name of the trace you want to move.
3. Without lifting the touch pen, drag the box now appearing around the trace name to the desired location.
4. When you have the box positioned where you want the trace to appear, lift the touch pen. If you are positioning the trace over another, the trace will “snap” into position.
5. To equalize the spacing between the traces in the new arrangement, tap Normalize in the Adjust… window.
Pacemaker Programming Guide 55
Page 56
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

4.3 Waveform strips

4.3.1 Freezing and analyzing a waveform strip

The Freeze feature enables you to “freeze” the last 15 s of all waveform traces that show in the expanded monitor window. Tapping Freeze at the top of the tool palette captures the previous 15 s of the trace signals and opens the frozen trace viewing window shown below.
Controls below the viewing window include the caliper controls described in the next section and the following buttons:
Adjust… Tap to open a window of waveform adjustment controls (see Section 4.3.1.3). Strips… Tap to view a list of other frozen waveform strips (see Section 4.3.2). Save Tap to save the waveform strip presently displayed (see Section 4.3.1.4). Delete Tap to delete a saved waveform strip. This button is active only if you are viewing
a saved waveform strip.
Print… Tap to print the waveform strip presently displayed (see Section 4.3.1.5). Close Tap to close the frozen waveform viewing window. If you have not saved the trace,
a pop-up window will remind you to save or delete the trace.
4.3.1.1 Using the on-screen calipers
The caliper positioning buttons control the frozen trace viewing window by letting you move each of the 2 vertical cursors appearing in the window to any desired position. The cursors thus act as calipers allowing you to measure the time interval between events. The caliper measurement is displayed in milliseconds in the upper-left corner of the window.
To use the caliper positioning buttons, alternately press and release the appropriate button to make small movements, or press and hold the button to make larger movements.
4.3.1.2 How to view other portions of the frozen waveform display
1. Use the vertical scroll bar to scroll the display up or down to view other waveform traces.
Tap the scroll arrows to scroll the traces up or down in small increments. The scroll box shows the relative position of the window with respect to the total height of the strip. Touch and drag the box to scroll the strip up or down.
2. Use the horizontal scroll bar, which operates like the vertical scroll bar, to move the display to the right or left to see other portions of the 15 s strip.
56 Pacemaker Programming Guide
Page 57
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
4.3.1.3 Adjusting the frozen traces
The button bar next to the trace source name provides options for adjusting the trace size and selecting a different trace source. Selecting Adjust… opens the Adjust… window, which provides additional control options. In both cases, these are some of the same controls available for the Live Rhythm Monitor window. Refer to Section 4.2, “Adjusting and arranging the waveform traces”, page 51 for information about using the following Adjust controls:
Using the button bar to adjust trace size or select a trace source
Using the waveform Clipping option
Selecting a Sweep Speed:
Using the Normalize button
1
4.3.1.4 Saving a frozen waveform strip
You can save the frozen waveform strip you are viewing by tapping Save. You may then recall the saved strip for viewing or printing later (see Section 4.3.2).
4.3.1.5 Printing the frozen strip
To print a frozen strip, tap Print… near the bottom of the Frozen Strip window. Depending on the Session preference settings, tapping Print… does one of the following:
Immediately initiates printing of the selected strip.
Sends the strip to the print queue to be printed later.
Opens a window of print options.
For more information about session preferences, see Section 3.4, “Proceeding with session tasks”, page 29.

4.3.2 Recalling and viewing waveform strips

Prior to ending the patient session, you can recall and view any waveform strip collected and saved during the session. Such strips may be strips saved during a test such as the Magnet test, Threshold test, or Underlying Rhythm test, or a strip saved during use of the Freeze feature.
1
Sweep Speed: options for frozen traces are 12.5, 25, 50, 100, and 200 mm/s.
Pacemaker Programming Guide 57
Page 58
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
4.3.2.1 How to view a previously collected strip
1. Tap Strips… near the top of the tool palette or in the lower-left corner of the Frozen Strip viewing window.
2. From the Other Strips… window now displayed, tap Collected by programmer.
3. From the list of strips in the selection field, tap the strip you want to view.
It may be necessary to use the scroll bar at the right side of the field if there are more than five strips available.
4. Tap Open.
For instructions on using the Frozen Strip viewing window, refer to Section 4.3.1.
58 Pacemaker Programming Guide
Page 59
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5 Collecting diagnostic data

5.1 About diagnostic data collection

To aid in the evaluation of pacemaker operation and delivery of pacing therapy, the pacemaker records several types of information pertaining to the patient’s heart rate and certain device operations. This recording process occurs automatically and continuously between patient sessions.
You can retrieve the recorded information and view it in a graphic format. You can also print the information in its graphic format, with or without a tabular summary of the numerical data.
This chapter does not list the diagnostics available for each pacemaker model. To find applicable diagnostics for a specific pacemaker model, refer to the Implant Manual for that pacemaker.

5.1.1 Automatic data collection

Once implanted, the pacemaker records data and presents it in a number of ways, depending on the programmed pacing mode and the pacemaker model.
Patient’s Heart Rate (histogram and data format)
Patient’s AV Conduction Status (histogram and data format)
Search AV+ Histogram (histogram and date format)
Sensor Indicated Rate Profile
High Rate Episodes
Rate Drop Response Episodes
Atrial Arrhythmia Trend
Atrial Arrhythmia Durations (printed only, Initial Interrogation Report)
Ventricular Rate During Atrial Arrhythmias (histogram and data format)
Capture Management Trends
Sensitivity Trends
Lead Impedance (Chronic Trends and Impedance Data)
Key Parameter History
Pacemaker Programming Guide 59
Page 60
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.1.2 Clinician-selected data collection

In addition to the automatically collected data, the clinician can choose to have the pacemaker collect additional data by programming one of the following options:
Custom Rate Trend
A. Capture Management Detail
V. Capture Management Detail
High Rate Detail
Rate Drop Response Detail

5.1.3 Important points about using the data function

5.1.3.1 Data collection is automatic
Data collection starts at device implant when the device detects the presence of an acceptable lead impedance. Thereafter, data collection is automatic throughout the life of the device. If Implant Detection is restarted, data collection is suspended until implant detection is complete or programmed off.
5.1.3.2 Programmable setup options
The programmer gives you options to adjust how the device collects its various types of data. For more information, refer to Section 5.5, “Programming data collection options”, page 78.
5.1.3.3 Time and date notations
The time and date information that appears on the display or printout of the diagnostic data is based on current programmer time settings. You can verify that the programmer time and date are set correctly by tapping Freeze. The current settings appear at the top of the strip viewing window. For instructions on setting the clock/calendar function, see Section 2.2, “Setting up your programmer”, page 13.
5.1.3.4 Effects of the programming head or a magnet
Data collection is suspended while the programming head or a magnet is positioned over the implanted device. If data collection is suspended during an episode, the word “Suspended” appears in the data listed for that episode. If a patient session has been initiated, data collection is suspended until one hour after the patient session has ended. At this time histogram and episode data are automatically cleared, and new data collection begins.
60 Pacemaker Programming Guide
Page 61
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
For trend data, a dashed vertical line indicates that data collection was suspended at that time.
5.1.3.5 Automatically cleared diagnostics
The following diagnostics are cleared following each patient session: histograms, episode data, sensitivity trend, and clinician-selected diagnostics. By default, the data are cleared one hour after the end of the patient session, at which time collection of new data begins. You can choose to start new data collection immediately upon ending the session (see Section 5.6, “Clearing data from the pacemaker”, page 80).
Note: The minimum action that constitutes a patient session is interrogation of the device.
Note: Sensor Indicated Rate Profile data are cleared only when programming certain rate
response parameters or another rate responsive mode.
5.1.3.6 Manually cleared trend data
Certain types of trend data are not cleared automatically. Unless you choose to clear these specific trends, collection occurs continuously as follows:
Chronic lead trend data are collected for the most recent 14 months and typically should not be cleared unless the lead is replaced. Clearing the lead trend data for a particular chamber (atrial or ventricular), clears both the Lead Impedance and Capture Management Trend data for that chamber.
The Atrial Arrhythmia Trend can show up to the most recent 6 months of collected data.
See Section 5.6, “Clearing data from the pacemaker”, page 80 for information about clearing trend data.
5.1.3.7 Effects of reaching RRT
There is no effect on data collection when the device reaches RRT (Recommended Replacement Time). Data continues to accumulate, and it can be retrieved for viewing and printing. There is no effect on diagnostic data if an RRT condition is reset.
5.1.3.8 Effects of reaching ERI
Data collection stops when the device reaches ERI (Elective Replacement Indicator). Accumulated data can still be retrieved for viewing and printing. If the ERI condition is reset, diagnostic data is cleared and data collection resumes one hour after the end of the programming session (or immediately, if programmed).
Pacemaker Programming Guide 61
Page 62
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
5.1.3.9 Effect of an electrical reset
All recorded diagnostic data is lost in an electrical reset.
5.1.3.10 Sense events not included in data collection
Data collected for the Rate Histogram, AV Conduction Histogram, High Rate Episodes, Search AV+ Histogram, and Ventricular Rate During Atrial Arrhythmias include ventricular paces triggered by the Ventricular Safety Pacing feature. However, the ventricular sense events that trigger these outputs are not included in the collected data. In the view of an episode marker strip, for example, the triggering sense (VS) event is not shown, but the resulting triggered pace is shown.
5.1.3.11 Using the Exercise Test and EP Studies functions
The use of the Exercise Test and EP Studies functions disables and clears data collected for the clinician-selected diagnostic. Interrogate the diagnostic data before using these functions. Once interrogated, the collected data are available on the programmer throughout the patient session, but are lost when you end the session. View and print the collected data before ending the session (see Section 5.2, “Viewing the collected data”, page 62).
The procedure for using the Exercise Test includes a step that automatically restores the data collection settings. If this step is not completed or is not successful, you must reprogram the clinician-selected diagnostic on the Data Collection Setup screen (see Section 5.5, “Programming data collection options”, page 78).
Conducting an Exercise test clears clinician-selected diagnostic data immediately from pacemaker memory.
Using the EP Studies function clears clinician-selected diagnostic EGM data immediately from pacemaker memory. The EGM type is set to Off after using the EP Studies function.

5.2 Viewing the collected data

To see the types of collected data that you can view, tap Data > Graphs and Tables to open the Collected Data - Graphs and Tables screen.
The list of available collected data varies according to pacemaker model, the programmed pacing mode, and the programmed option for Clinician Selected data collection. See Section 5.3, “Automatically collected data displays”, page 63 for more information.
62 Pacemaker Programming Guide
Page 63
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.2.1 Viewing procedure

From the Collected Data - Graphs and Tables screen, you can view data as follows:
1. Tap on an item in the Data Type list.
2. Tap Open Data.

5.2.2 Viewing tools

Some data windows allow you to “zoom in” on a selected segment of the plotted data to show more detailed information.
Use the right and left scroll buttons to position the zoom region over the data that you want to see in detail.
To display details of the zoom region, tap the Zoom In (+) icon. On some screens, tapping the Zoom In (+) icon has the same effect as tapping the Strip icon.
If the selected data detail includes more information than can be displayed in the window, use the scroll buttons to navigate back and forth across the full data detail.
To return to the full data display screen, tap the Zoom Out (–) icon.
For some data, the details of the zoom region are presented as a recorded strip of event markers.
If the Strip icon is displayed, tap it to view the collected EGM or strip of event markers.

5.3 Automatically collected data displays

5.3.1 Automatic data collection

Patient’s Heart Rate (histogram and data format)
Patient’s AV Conduction Status (histogram and data format)
Search AV+ Histogram (histogram and data format)
Sensor Indicated Rate Profile
High Rate Episodes
Rate Drop Response Episodes
Atrial Arrhythmia Trend
Atrial Arrhythmia Durations (printed only, Initial Interrogation Report)
Ventricular Rate During Atrial Arrhythmias (histogram and data format)
Capture Management Trends
Pacemaker Programming Guide 63
Page 64
Medtronic
Sensitivity Trends
Lead Impedance (Chronic Trends and Impedance Data)
Key Parameter History
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.3.2 Heart Rate Histograms

5.3.2.1 Description
Heart Rate Histograms shows at a glance the range of the patient’s heart rate (to a resolution of 10 bpm) recorded since the last patient session and how the number of heart beats counted during this period are distributed over that rate range. The histograms are intended as an aid for evaluating rate response parameter settings and chronotropic incompetence.
To view the Heart Rate Histograms, tap Data > Graphs and Tables > Heart Rate
Histograms > Open Data.
5.3.2.2 Data collection and presentation
The Heart Rate Histograms can be displayed for two intervals of data collection:
Short-term heart rate histograms represent data collected over the last 3 to 6 days.
Long-term heart rate histograms represent data collected since the last follow-up.
Heart Rate Histograms indicate the percentage of rate events counted during the monitoring period that fall into 10 bpm segment in the range from 40 bpm to 180 bpm (the “< 40” bpm group also includes events occurring below 40 bpm; the “180 >” bpm group also includes events occurring above 180 bpm). Each 10 bpm event group is coded to show the percentage of paced events and the percentage of sensed events.
Both atrial and ventricular histograms display the pertinent programmed parameter settings. Ventricular histograms also display the recorded number of single PVCs and PVC runs. Atrial histograms display the number of PAC runs.
For single-chamber pacemakers, Heart Rate Histograms data is available for the paced chamber.
5.3.2.3 Atrial and ventricular histograms
If data were collected during device operation in a dual chamber mode (including ADIR, ADI, VDIR, VDI, and VDD modes), atrial and ventricular histograms can be viewed by tapping the Histogram: field in the Collected Data - Heart Rate Histograms window.
64 Pacemaker Programming Guide
Page 65
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
5.3.2.4 Setup options
Data collection can be set to include or exclude refractory senses (events sensed during the refractory period). See Section 5.5, “Programming data collection options”, page 78.

5.3.3 AV Conduction Histograms

5.3.3.1 Description
Like the Heart Rate Histograms, AV Conduction Histograms indicate the percentage of the total number of heart beats counted since the last patient session occurred in each of the displayed 10 bpm rate groups. AV Conduction Histograms are available only if a dual chamber mode (including ADIR, ADI, VDIR, VDI, and VDD modes) has been programmed.
To view the A-V Conduction Histograms, tap Data > Graphs and Tables > A-V Conduction
Histograms > Open Data.
5.3.3.2 AV conduction sequence categories
Each 10 bpm rate group is coded to show the portion of the events in that group that fall into the different AV conduction sequence categories. The AV conduction sequence categories are as follows:
AS–VS Atrial Sense - Ventricular Sense
AS–VP Atrial Sense - Ventricular Pace
AP–VS Atrial Pace - Ventricular Sense
AP–VP Atrial Pace - Ventricular Pace
a
Sequences of AS-VP do not apply to the ADIR and ADI modes.
b
Sequences of AP-VS do not apply to the VDD, VDI, and VDIR modes.
a
b
a,b
A rate group on the histogram that represents less than 5% of the counted events may be coded to the AV sequence with the highest percentage of events.
Short-term AV conduction histograms represent data collected over the last 3 to 6 days.
Long-term AV conduction histograms represent data collected since the last follow-up.
Refer to the reference manual for additional information about collecting AV Conduction Histogram data.
Pacemaker Programming Guide 65
Page 66
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.3.4 Search AV+ Histogram

The Search AV+ Histogram data is automatically collected when Search AV+ is programmed On. A Search AV+ Histogram shows the percentage of the following intervals versus rate in every rate range:
A–VS: AV sequences ending in a ventricular sense within the programmed paced AV or sensed AV interval.
VS from Search: AV sequences ending in a ventricular sense due to the extension of the AV interval by Search AV+.
A–VP: AV sequences ending in a ventricular pace.
To view the Search AV+ Histogram, tap Data > Graphs and Tables > Search AV+ Histogram > Open Data.
Refer to the reference manual for additional information about collecting Search AV+ Histogram data.

5.3.5 Sensor Indicated Rate Profile

The Sensor Indicated Rate Profile records the rate that is derived from activity sensor operation. Because this diagnostic shows the measured sensor rate profile, it aids in evaluating the Rate Profile Optimization feature. It indicates the percentage of time spend in individual rate ranges.
Rates are recorded in 10 bpm ranges from 40 bpm to 180 bpm. Rates outside this range are recorded in 2 additional ranges. Rates below 40 bpm are counted in the lowest range. Rates above 180 bpm are counted in the highest range.
To view the Collected Data - Sensor Indicated Rate Profile window, tap Data > Graphs
and Tables > Sensor Indicated Rate Profile > Open Data.
Note: Sensor Indicated Rate Profile data is cleared only when programming certain rate
response parameters or another rate responsive mode.
Refer to the reference manual for additional information about collecting Sensor Indicated Rate Profile data.

5.3.6 High Rate Episodes

High Rate Episodes data include the date and time, the maximum rate in both chambers, the average ventricular rate, and the duration of episodes. On printed reports, the High Rate Episodes data will also include the sensor rate at the time the episode was detected.
66 Pacemaker Programming Guide
Page 67
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
The programmed Collection Method determines whether episode data are updated or data stops when the limit is reached. The options are listed below:
Rolling: Data for the first, fastest, longest, and the most recent 13 episodes are collected until the pacemaker is interrogated.
Frozen: Data are collected for the first 16 episodes that occur after the device is programmed. Data collection stops after 16 episodes; however, the Episodes Detected count continues.
Episodes are defined by the programmed collected data setup options. For more information about these options, see Section 5.5, “Programming data collection options”, page 78.
The maximum number of high rate episodes the pacemaker will count is 16,777,215 for atrial episodes and the same count for ventricular episodes. Once a limit is reached, the count is frozen.
To view the Collected Data - High Rate Episodes window, tap Data > Graphs and Tables > High Rate Episodes > Open Data. This window has check boxes that allow you to view atrial data, ventricular data, or both types. For single-chamber pacemakers, High Rate Episodes data is available for the paced chamber.
If Mode Switch is programmed On, the atrial data collected will be based on Mode Switch episodes. If Mode Switch is programmed Off, the atrial data collected will be based on independent Atrial High Rate detection criteria.
When the clinician-selected data is High Rate Detail, the screen indicates that the applicable episodes have EGM and/or Trend data. To view the additional data, tap the episode and then tap Open Episode.
Note: When Mode Switch is programmed On, the diagnostic parameter Collection Delay sets up a minimum length that mode switching episodes must satisfy to be included in High Rate data collection. (Collection Delay does not affect Mode Switch therapy, but it allows the user to filter out short episodes from the high rate episode collection.)
If you tap >>, the Episode Collection Criteria window opens listing the criteria for collection of atrial and ventricular episodes.
Refer to the reference manual for additional information about collecting High Rate Episodes data.
Pacemaker Programming Guide 67
Page 68
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
5.3.6.1 Clinician-selected data collection option
High Rate Detail is one of the optional clinician-selected diagnostics discussed in Section 5.4, “Clinician-selected detailed data displays”, page 72.

5.3.7 Rate Drop Response Episodes

Rate Drop Response Episodes records the operation of the Rate Drop Response function (no data will be collected if Rate Drop Response function is Off). Depending on the programmed settings for Rate Drop Response, it will record the number of Low Rate Episodes, Rate Drop Episodes, or both. Low Rate and Rate Drop Episodes are defined as follows:
Low Rate Episode: When a programmed number of lower rate paced events occurs with no intervening non-refractory sensed events.
Rate Drop Episode: Defined by programmed values for Drop Rate and Drop Size.
– Drop Rate: Intrinsic ventricular rate must decrease to, or falls below, this rate.
– Drop Size: Number of beats per minute the ventricular intrinsic rate must decrease
within the detection window time.
– Detection Window: Maximum allowable time for the rate drop criteria to be met.
If more than 255 Rate Drop Response episodes occur, the Rate Drop Response count stops recording.
The date/time, type, and intrinsic detection rate of the initial and the 9 most recent episodes can be accessed from the Graphs and Tables screen.
To view the Additional Data, tap the episode, and then tap Open Episode. Trend data is collected only when the clinician-selected data is Rate Drop Response Detail.
5.3.7.1 Clinician-selected data collection option
Rate Drop Response Detail is one of the optional clinician-selected diagnostics discussed in Section 5.4, “Clinician-selected detailed data displays”, page 72.

5.3.8 Atrial Arrhythmia Trend

5.3.8.1 Description
The Atrial Arrhythmia Trend shows a log of how much time each day that the patient spends in atrial arrhythmias. Collected over a period of weeks or months, this information indicates when there is an upward or downward trend in this type of cardiac activity.
68 Pacemaker Programming Guide
Page 69
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
The Episode Trigger is Mode Switch if Mode Switch is enabled. Otherwise, it is the programmed atrial high rate (detection rate) value.
5.3.8.2 Data collection and presentation
This display is a graph of daily measurements of the length of time that the implanted device detects atrial arrhythmias. Included with the trend display is the time period covered by the accumulated data, the pertinent programmed parameter settings, and the date of the last patient follow-up session.
The Atrial Arrhythmia Trend can show up to 6 months of collected data, and it is a rolling trend that updates continuously. An option allows you to clear the accumulated data if desired (see Section 5.6, “Clearing data from the pacemaker”, page 80).
To view the Collected Data - Atrial Arrhythmia Trend window, tap Data > Graphs and Tables > Atrial Arrhythmia Trend
Tap >> to the right of High Rate Episodes to open the Collected Data - High Rate Episodes diagnostic window. See Section 5.3.6 for more details.
Tap >> to the right of V. Rate During A. Arrhythmias to open the Collected Data -
Ventricular Rate During Atrial Arrhythmias diagnostic window.

5.3.9 Atrial Arrhythmia Durations

This diagnostic data appears only in the Initial Interrogation Report under Arrhythmia Summary. Each atrial arrhythmia episode is counted under one of these ranges of episode duration:
72 hr
24 hr to <72 hr
12 hr to <24 hr
4 hr to <12 hr
1 hr to <4 hr
10 min to <1 hr
1 min to <10 min
<1 min
To print the Initial Interrogation Report, tap Reports > Available Reports > Initial Interrogation Report > Print…
Pacemaker Programming Guide 69
Page 70
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.3.10 Ventricular Rate During Atrial Arrhythmias

5.3.10.1 Description
The Ventricular Rate During Atrial Arrhythmias histogram shows a profile of ventricular rate recorded during the atrial arrhythmias that have been recorded since the last patient session. This information can show how the ventricular rate is controlled during atrial arrhythmias.
To view the Collected Data - Ventricular Rate During Atrial Arrhythmias window, tap
Data > Graphs and Tables > V. Rate During A. Arrhythmias > Open Data.
5.3.10.2 Data collection and presentation
During each atrial arrhythmia, the implanted device collects ventricular rate data in addition to atrial data. This histogram shows the percentage of the total number of ventricular beats counted during these arrhythmias that falls into each of eight 20 bpm rate groups in the measured rate range. The beats within each rate group are coded to show the proportion of paced and sensed beats.
Information included with the histogram lists the pertinent programmed parameter settings and summary information about the recorded atrial arrhythmias during which the ventricular rate information was recorded.
5.3.10.3 Setup options
Data collection can be set to include or exclude refractory senses (events sensed during the refractory period). See Section 5.5, “Programming data collection options”, page 78.

5.3.11 Capture Management Trend

Capture Management Trend data are collected automatically when Capture Management is programmed On. This trend records pacing threshold history for the most recent 14 months. When VCM and ACM are enabled, the pacemaker collects the following data every 7 days:
Maximum Threshold
Average Threshold
Max Adapted Amplitude
Capture Management Setting (Adaptive or Monitor Only)
To open the Collected Data - Capture Management Trend window, tap Data > Graphs and Tables > Capture Management Trend > Open Data.
70 Pacemaker Programming Guide
Page 71
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
To view Atrial or Ventricular Capture Management Trend data, select Atrial or Ventricular from the Trend: field.
To open the Ventricular Capture Management - Last Measured Threshold window, tap >> to the right of the Measured Threshold label. This graph presents the results of the most recent ambulatory ventricular pacing threshold search. This includes the measured strength duration curve and the curve corresponding to a 2X Amplitude Margin.
Note: >> is not available when viewing the Atrial Collected Data - Capture Management Trend window.
Refer to the reference manual for additional information about collecting Capture Management Trend data.
5.3.11.1 Clinician-selected data collection option
Atrial and Ventricular Capture Management Detail are two of the optional clinician-selected diagnostics discussed in Section 5.4, “Clinician-selected detailed data displays”, page 72.

5.3.12 Sensitivity Trend

Sensitivity Trend data is collected automatically when Sensing Assurance is programmed On. The pacemaker records the following information for both chambers every 7 days:
Minimum Sensitivity
Maximum Sensitivity
Whether sensitivity adjustments reached the lower limit (most sensitive)
To view the Collected Data - Sensitivity Trends window, tap Data > Graphs and Tables > Sensitivity Trend > Open Data.
Tap the Trend: field to view trends for P-Wave Amplitude, R-Wave Amplitude, Sensitivity - Atrial, or Sensitivity - Ventricular.
Refer to the reference manual for additional information about collecting Sensitivity Trend data.

5.3.13 Lead Impedance

The Collected Data - Lead Impedance window displays lead impedance trend and polarity switch data collected for the previous 14 months. Data includes high impedance paces and low impedance paces from the Lead Monitor function.
Pacemaker Programming Guide 71
Page 72
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Measurements are taken every 3 hours in each chamber that is being paced. The maximum, minimum, and average lead impedances are recorded every 7 days. Depending on which chambers are paced, the following data can be selected from the Trends: field:
Atrial Impedance - Chronic
Ventricular Impedance - Chronic
Atrial Impedance Data
Ventricular Impedance Data
Lead impedance trends collection can be programmed On or Off via the Data Collection - Setup screen.
To view the Collected Data - Lead Impedance window, tap Data > Graphs and Tables > Lead Impedance > Open Data.
Refer to the reference manual for additional information about collecting lead impedance data.

5.4 Clinician-selected detailed data displays

Only one type of clinician-selected data can be programmed On at one time. The following options for clinician-selected data are accessed from the Data Collection - Setup screen:
Custom Rate Trend
A. Capture Management Detail
V. Capture Management Detail
High Rate Detail
Rate Drop Response Detail
To choose the type of clinican-selected data collected by the device, tap Data > Data Collection Setup/Clear and then tap the field under Clinician Selected…. Tap an option
from the drop-down list, and configure the settings for that option. Tap PROGRAM to program any updated settings to the device.
72 Pacemaker Programming Guide
Page 73
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.4.1 Custom Rate Trend (rate versus time)

The pacemaker can be set to record the patient’s heart rate for a beat-to-beat, 1-hour, or 24-hour monitoring period, depending on the programmed “duration.”
Beat-to-Beat duration: Records beat-to-beat data for several minutes depending on the mode, rate, and number of refractory events. Shows the maximum, mean, and minimum rates.
1-hour duration: The pacemaker records the patient’s heart rate every 2 s and accumulates one hour of data.
24-hour duration: The pacemaker measures heart rate every 60 s.
5.4.1.1 About displaying Custom Rate Trends
Custom Rate Trends can be displayed by tapping Data > Graphs and Tables > Custom Rate Trends > Open Data. If data was collected during dual chamber pacing (including
ADIR, ADI, VDIR, VDI, VDD, and ODO modes), atrial and ventricular graphs are recorded.
5.4.1.2 About Zoom-In
Zoom-In lets you view a segment of the Rate vs. Time graph in more detail. If the duration of recorded data is less than the zoom region, the zoom-in feature automatically is in effect when the data is displayed.
On the detailed display that shows a segment of the Rate vs. Time graph, the data samples are coded to indicate the percentage of the beats in each sample that are paced. .
Refer to the reference manual for additional information about collecting Custom Rate Trend data.
Pacemaker Programming Guide 73
Page 74
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
5.4.1.3 About Custom Rate Trend parameters
Choosing Rate Trend clinician-selected data involves programming the parameters shown in Table 9.

Table 9. Clinician-selected Rate Trend diagnostic parameters

Parameter Definition Settings
Duration This parameter programs the duration of data
collection to beat-to-beat, 1 hour, or 24 hours.
Collection Method “Rolling” programs the pacemaker to record
continuously, so that new data overwrites the old data once memory capacity is reached. “Frozen” programs the pacemaker to record data until its memory capacity is reached. Recording then stops, and the collected data is frozen.
Include Refractory Senses?
(Refractory Sensed Events) Programs the pacemaker to include or exclude events sensed within the refractory period when it counts heart rate events.
Beat-to-Beat, 1 hr, 24 hr
Rolling, Frozen
Exclude, Include

5.4.2 Atrial Capture Management Detail

Atrial Capture Management Detail data provides a record of the pacing threshold amplitude at a pulse width of 0.40 ms. The pacemaker records the following information for the most recent 359 atrial pacing threshold searches:
Date and Time
Amplitude Threshold
Amplitude Margin
Adapted or Monitor Only Amplitude
Atrial Capture Management Detail data can be displayed by tapping Data > Graphs and Tables > Clinician Selected - A. Capture Management > Open Data.
A beat-to-beat trend for the most recent pacing threshold search is available. If you select the Strip icon, the trend shows the most recent 1000 events in the search.
The results of a pacing threshold search can be displayed by selecting the Zoom In icon. The pacing threshold search result can be selected from the list on the screen.
Clinicians have the option to record an EGM (atrial, ventricular, or summed EGM). No EGM is available if the last pacing threshold search was aborted.
Refer to the reference manual for additional information about collecting Atrial Capture Management Detail data.
74 Pacemaker Programming Guide
Page 75
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.4.3 Ventricular Capture Management Detail

Ventricular Capture Management Detail data provide a record of the pacing threshold pulse width and amplitude. The pacemaker records the following information for the most recent 668 ventricular pacing threshold searches:
Date and Time
Amplitude Threshold
Pulse Width Threshold
Amplitude Margin
Adapted or Monitor Only Amplitude
Ventricular Capture Management Detail data can be displayed by tapping Data > Graphs and Tables > Clinician Selected - V. Capture Management > Open Data.
A beat-to-beat trend for the most recent pacing threshold search is available. If you select the Strip icon, the trend shows the most recent 1000 events in the search.
The results of a pacing threshold search can be displayed by selecting the Zoom In icon. The pacing threshold search result can be selected from the list on the screen.
Clinicians have the option to record an EGM (atrial, ventricular, or summed EGM). No EGM is available if the last pacing threshold search aborted before finding the amplitude threshold.
Refer to the reference manual for additional information about collecting Ventricular Capture Management Detail data.

5.4.4 High Rate Detail

The pacemaker collects a beat-to-beat trend at the onset and at the termination of selected episodes. For some episodes, it also collects EGM.
Episode selection depends on the diagnostic parameters for automatic high rate episodes, which are automatic diagnostics discussed in Section 5.3, “Automatically collected data displays”, page 63. One of these automatic parameters, Collection Method, controls whether High Rate data collection is Rolling or Frozen. Rolling data collection continues to be updated. Frozen data collection stops when a limit is reached.
The specific data collected for High Rate Detail is controlled by these diagnostic parameters:
High Rate Type: AHR, VHR, or both
EGM Type: AEGM, VEGM, Summed, or Off
Pacemaker Programming Guide 75
Page 76
Medtronic
EGM Allocation: how many episodes with EGM collection and how the EGM is aligned
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
relative to the episode onset
Pre-detection Timeout: the time limit in which pre-onset EGM can be collected
For single-chamber pacemakers, High Rate Detail data is available for the paced chamber.
Trend data is collected whether or not EGM is collected. The number of episodes of trend data collected depends on Collection Method and High Rate Type:
If High Rate Type is AHR or VHR, for Rolling collection, trend data is collected for the first episode and is updated for the most recent 13 episodes until the pacemaker is interrogated. For Frozen collection, trend data is collected for the first 16 episodes.
If High Rate Type is AHR and VHR, Rolling trend data is collected for the most recent 8 AHR and 8 VHR episodes. Frozen trend data is collected for the first 8 AHR and 8 VHR episodes.
The choices for EGM Allocation depend on how the Collection Method and High Rate Type parameters have been set. You should set those parameters before selecting the EGM Allocation.
When Rolling data collection is used for pre-onset EGM, the total length of EGM collected is less than it is for the other options:
Rolling data collection for post-onset EGM
Frozen data collection for pre-onset or post-onset EGM
In Figure 9 you can see an example of the EGM Allocation window with the Collection Method set to Rolling and High Rate Type set to AHR&VHR. Each value in the window represents a number of episodes and the number of seconds of pre-onset and post-onset EGM to be collected for each episode. If High Rate Type set to AHR&VHR, the number of episodes is split evenly between atrial episodes and ventricular episodes.
76 Pacemaker Programming Guide
Page 77
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Figure 9. EGM Allocation window: atrial and ventricular collection

1 High rate type is AHR and VHR. 2 4 episodes with 8 s (pre-onset) and 0 s (post-onset). 3 4 episodes with 4 s (pre-onset) and 4 s (post-onset).
Refer to the implant manual for EGM Allocation values related to the settings of High Rate Type and Collection Method.
Note: If Rolling collection is programmed and several episodes of one type occur before episodes of the other type occur, EGM collection for the first type will use all allocated EGM storage until the other episode type occurs. For example, if an allocation of 4 EGM episodes is programmed and both AHR and VHR High Rate Types are selected, an occurrence of 4 consecutive VHR episodes results in EGM collection for all 4 VHR episodes. If any AHR episodes subsequently occur, EGM data for the first 2 AHR episodes replaces the EGM data of the oldest 2 VHR episodes. Thereafter, EGM data are allocated equally between the 2 triggers.
EGM can be collected with a programmable Pre-detection Timeout of 1 to 12 weeks, in 1-week increments, or of 14 to 24 weeks, in 2-week increments. If episodes occur before the programmable timeout, pre-onset and post-onset EGM can be collected. If episodes occur after the programmable timeout, only post-onset EGM can be collected.
Note: If pre-onset EGM and Rolling collection are programmed, and the EGM Pre-detection Timeout expires, EGM collection will occur for an extra episode. For example, if EGM Allocation is programmed to “4 for 4/4 secs”, the user would go from 4 episodes with EGM (prior to the timeout) to 5 episodes with EGM (after the timeout).
Pacemaker Programming Guide 77
Page 78
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
5.4.4.1 How to display a High Rate Detail screen
1. Tap Data > Graphs and Tables > High Rate Episodes > Open Data.
2. Select an episode with EGM and Trend data or an episode with Trend data only.
Note: Episodes with EGM data also have Trend data.
3. Tap Open Episode.
4. Tap the Zoom In icon or the Strip icon from the Episode Onset screen to display episode onset data.

5.4.5 Rate Drop Response Detail

Data is collected only if the Rate Drop Response feature has been programmed On. When Rate Drop Response Detail is the clinician-selected data, the pacemaker monitors the designated chamber for activity related to the operation of the Rate Drop Response. A beat-to-beat trend is collected for the first and the 4 most recent episodes. Each trend can record a maximum of 680 events and is frozen when the detection criteria are met. Rate Drop Response episodes are defined by the programmed settings for the Rate Drop Response feature.
You may include or exclude refractory sensed events from the data.
5.4.5.1 How to display a Rate Drop Response Detail screen
1. Tap Data > Graphs and Tables > Rate Drop Response Episodes > Open Data.
2. Select an episode with Trend data.
3. Tap Open Episode.

5.5 Programming data collection options

The Data Collection - Setup screen displays settings to collect device data. To view this screen, tap Data > Data Collection Setup/Clear.
Tap the applicable field to display subordinate parameters or information about data collection.

5.5.1 Setup options for automatic data collection

Data is collected based on the programmed mode, parameter values, and setup options. Table 10 gives an overview of setup options necessary to collect certain types of data.
78 Pacemaker Programming Guide
Page 79
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™

Table 10. Data collection setup options

Data type Setup options Parameters
Rate Histo­grams
Atrial High Rate Episodes (Mode Switch Off)
Atrial High Rate Episodes (Mode Switch On)
Ventricular High Rate Epi­sodes
Clinician Selected
a
# indicates a programmable parameter.
b
The Termination Rate has the same value as the programmable Detection Rate.
c
The Ventricular Rate for ending collection has the same value as the programmable Ventricular Rate (for collection).
Include or Exclude refractory sensed events from heart rate histograms.
Include Refractory Exclude Refractory
Collect if:
Atrial Rate >= #a bpm Detection Rate For more than #a seconds Detection Duration
End Collection if:
Atrial Rate < #b bpm Termination Rate (not programmable)
for # a beats Termination Beats Frozen or Rolling Collection Collection Method
Based on programmed Mode Switch therapy parameters (Mode Switch, Detect Rate, Detect Duration, and Blanked Flutter Search)
Filter out short Mode Switch episodes Collection Delay after Mode Switch Frozen or Rolling Collection Collection Method
Collect if:
Ventricular Rate >= #a bpm Ventricular Rate (for collection)
for #a beats for (X beats) End Collection if:
Ventricular Rate < #c bpm Ventricular Rate (not programmable)
for #a beats for (Y beats) Filter out conducted SVT episodes SVT Filter Frozen or Rolling Collection Collection Method
Select one of the following: Off Custom Rate Trend Duration, Collection Method,
Refractory Senses (Include or
Exclude) Atrial Capture Management Detail EGM Collection Ventricular Capture Management Detail EGM Collection High Rate Detail High Rate Type, EGM Type, Allocation,
Predetection Timeout Rate Drop Response Detail Refractory Senses (Include or
Exclude)
SW043
Pacemaker Programming Guide 79
Page 80
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

5.5.2 Programming a clinician-selected data collection option

5.5.2.1 How to program options for data collection
1. Tap Data > Data Collection Setup/Clear.
2. Verify the parameter settings.
3. (optional) Tap the Clinician Selected… field and choose another option. Verify all of the parameter settings.
4. Tap PROGRAM.

5.6 Clearing data from the pacemaker

Unless another setting is chosen, data is automatically cleared one hour after the end of the programming session. The End Session window provides access to options that let you make the following changes to the default settings:
The following data is not automatically cleared. However, you can choose to have it cleared.
– Atrial Lead Trend (Lead Impedance and Pacing Threshold Trends)
– Ventricular Lead Trend (Lead Impedance and Pacing Threshold Trends)
– Atrial Arrhythmia Trend
Note: Chronic lead trend data typically should not be cleared unless the pacing lead has been replaced.
Selecting the Now option causes the pacemaker to clear the collected data and start collection of new data immediately when you end the session. New data collection starts when you lift the programming head.
Open the Clear Data window by tapping Data > Data Collection Setup/Clear. Clearing data is discussed in more detail in Section 3.14, “Ending a patient session”, page 45.
Chronic Ventricular Lead Trend data should be cleared only if the lead is replaced. Clearing Ventricular Chronic Lead data also clears automatic Ventricular Capture Management Trend data.
Note: Sensor Indicated Rate Profile data is cleared only when you program certain rate response parameters or another rate-responsive mode.
80 Pacemaker Programming Guide
Page 81
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6 Evaluating parameter settings

6.1 Measuring stimulation thresholds

The Threshold test function allows you to measure the patient’s stimulation thresholds to determine appropriate settings for amplitude and pulse width. To begin a threshold test, you must select the desired test type, heart chamber, and test parameter values from the Threshold Test Setup screen.

6.1.1 The Threshold tests setup screen

To display the Threshold tests setup screen, tap Tests > Threshold.
Note: During the execution of a threshold measurement, lifting the programming head from
over the pacemaker for at least two seconds will return the pacemaker to its programmed state. This action should be taken in the event of a programmer malfunction, loss of power, or the absence of an appropriate command confirmation.

6.1.2 Setting up a threshold measurement test

Before you execute a threshold test, choose the test type and parameter options that will provide appropriate threshold information.
6.1.2.1 How to set up a threshold test
1. Tap Tests > Threshold to display the Threshold tests setup screen.
2. Choose a setting for each of the 3 setup options: Test Type, Chamber, and Decrement after. Note that not all setup options apply to every test and thus, are not displayed.
3. Choose a test setting for each displayed parameter (see Table 11).
4. Execute the test you selected in step 2, above:
Capture Management: see Section 6.1.4.
Strength Duration: see Section 6.1.3.
Amplitude - Auto Decrement: see Section 6.1.5.
Pulse Width - Auto Decrement: see Section 6.1.5.
Manual - Amplitude & Pulse Width: see Section 6.1.6.
Pacemaker Programming Guide 81
Page 82
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Table 11. Threshold test parameters

Test parameteraValue selection
Mode Select the pacing mode you want to be in effect during the threshold test. The
available mode options depend on the programmed mode.
Lower Rate Select a lower rate value high enough to provide consistent pacing during the
test. Loss of capture cannot be determined if the pacemaker is inhibited.
AV delay If the selected test mode is DDD, DDI, or DOO, select an AV delay value short
enough to provide a consistently paced ventricular rhythm. If the mode is DDD, the selected value applies to both Paced AV and Sensed AV intervals.
Amplitude If the selected test type is Amplitude - Auto Decrement, Strength-Duration, or
Manual, select the value at which you want the test to start. The value must be high enough to provide consistent capture. If the selected test type is Pulse Width - Auto Decrement, select the value you want to be in effect during measurement of the pulse width threshold.
Pulse width If the selected test type is Pulse Width - Auto Decrement or Manual, select the
value at which you want the test to start. The value must be high enough to provide consistent capture. If the selected test type is Amplitude - Auto Decrement or Strength-Duration, select the value you want to be in effect during measurement of the amplitude threshold. If the selected test type is Strength-Duration, pulse width cannot be less than
0.4 ms.
a
If the Ventricular Capture Management test type is chosen, the test parameters are fixed to pacemaker configured values.

6.1.3 Executing a Strength Duration threshold test

If you completed the setup steps in Section 6.1.2 and you selected Strength Duration as the threshold test type, use this procedure to execute the test. The results of this test are displayed as a strength duration curve, described in Figure 10.
82 Pacemaker Programming Guide
Page 83
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Figure 10. Strength duration curve

1 The “X” marks the present programmed settings. 2 The square marks the pending amplitude and pulse width settings. 3 The computed strength duration curve based on the threshold test results. Only amplitude and
pulse width values that intersect above the curve are projected to result in capture; any point of intersection falling in the shaded area below the curve will not result in capture.
6.1.3.1 Test restrictions
Any one of the following conditions will prevent or abort execution of the Strength Duration threshold test.
Implant detection is still operative.
Recommended Replacement Time (RRT) is set.
Tap TEST Press and Hold to open a message window that informs you of the restricting condition.
Pacemaker Programming Guide 83
Page 84
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
6.1.3.2 Test procedure
There are 3 parts to the Strength Duration test. Complete all 3 parts in the order listed.
Part I: Measuring the amplitude threshold
Part II: Measuring the pulse width threshold
Part III: Displaying the strength-duration curve
6.1.3.3 How to measure the amplitude threshold (Part I)
1. Position the programming head, and hold it steady for the duration of this procedure.
2. Press and hold TEST Press and Hold.
The programmer automatically reduces pulse amplitude one programmable value every 2 to 15 pacing pulses (as set during setup). The programmed value and the point at which it takes effect shows at the top of the Live Rhythm Monitor.
In cases where pacing is inhibited, the test reduces pulse amplitude one value every eight seconds.
3. Carefully observe the patient’s ECG for loss of capture.
4. At the moment you see loss of capture, release TEST Press and Hold.
The Strength Duration test screen shows the results of the amplitude threshold measurement. The point at which capture is recovered is considered the amplitude threshold.
Tap the Test Strip icon if you want to view or print a 10 s ECG strip starting 8 s before the test ended. If you want to restart the test, tap Return to Setup.
Note: If the displayed threshold value is not the lowest programmed test value that maintained capture, tap the Threshold Amplitude field to change the threshold value.
6.1.3.4 How to measure the pulse width threshold (Part II)
1. With the programming head in position, press and hold TEST Press and Hold.
The programmer automatically reduces pulse width one programmable value at a time. The amplitude setting in effect is twice the threshold value measured in Part I. The starting pulse width is approximately one-half the value used in Part I or 0.40 ms, whichever is greater.
2. Carefully observe the patient’s ECG for loss of capture.
84 Pacemaker Programming Guide
Page 85
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3. At the moment you see loss of capture, release TEST Press and Hold.
The screen now shows the results of both amplitude and pulse width threshold measurements. The point at which capture is recovered is considered the pulse width threshold.
Select the Test Strip icon if you want to view or print a 10 ms ECG strip starting 8 ms before the end of the test.
Note: If the displayed threshold value is not the lowest programmed test value that maintained capture, adjust the Pulse Width value in the test setup screen.
Tap Return to Setup to restart the test.
6.1.3.5 How to display the strength duration curve (Part III)
1. Tap Draw Curve.
2. This action causes the programmer to calculate and draw the strength-duration curve based on the amplitude and pulse width threshold test results.
Note: For information about the strength duration curve, see Figure 10.
Options on this screen allow you to:
Program amplitude and pulse width values (see Section 6.1.3.6). After a Strength Duration Test, you can program settings for ambulatory capture management operation also.
View the 10 ms ECG strip collected during either threshold measurement
Print a Strength Duration Threshold Test Report, which includes the ECG strips
Return to the Test Setup screen, which clears this test data from the screen
Note: If certain pacing parameter values have changed during the session, the programmer may display a pop-up Warning window advising you that Battery and Lead parameters should be remeasured. This information is needed for the calculation of the estimated battery life appearing on the Test Results screen. Tap Yes to allow the programmer to make these real-time measurements.
6.1.3.6 Selecting and programming amplitude and pulse width values
The programmer displays the pending threshold point on the strength curve based on the pending amplitude safety margin value or output settings.
From the Strength Duration Test: Curve screen, you can program the suggested amplitude and pulse width values, or select and program different values.
Pacemaker Programming Guide 85
Page 86
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
If you do not wish to program the suggested values, you can change them:
Tap the Amplitude Safety Margin of field. Select an adequate safety margin value that is above the amplitude and pulse width at the pacing threshold. The programmer then selects the nearest amplitude value that maintains the lowest possible drain on the pacemaker to correspond to your selected safety margin. It displays this value under the Pending column. Note that rounding to the nearest upper value may set a slightly higher safety margin.
Tap the Amplitude or Pulse Width field to display the programmable values. Select the desired value for pulse width or amplitude. Note that the programmer may recalculate the amplitude safety margin value to the nearest decimal value to approximate the selected amplitude value (for example, 3.20 X).
Tap the Capture… field (if available) if you want to change Capture Management parameters.
To program the capture, amplitude and pulse width values, position the programming head, and tap PROGRAM.
Note: The Estimated battery life of value is an estimate of the average time remaining until the pacemaker requires replacement. This estimate is based on the amplitude and pulse width values selected for programming, as well as data accumulated by the pacemaker since the previous patient session.
Note: The ambulatory capture management feature will discontinue ambulatory threshold measurements if amplitudes greater than 5.0 V or pulse widths greater than 1.0 ms are programmed.

6.1.4 Executing a Ventricular Capture Management threshold test

If you selected Ventricular Capture Management as the threshold test type, use this procedure to execute the test.
6.1.4.1 Test restrictions
Any of the following conditions prevents execution of the Ventricular Capture Management test. When you tap Tests > Threshold, then select Capture Management and Ventricle, a message informs you of restricting conditions, if any:
Implant detection is still operative.
Amplitude is set to a value greater than 5.0 V.
Pulse width is set to a value greater than 1.0 ms.
86 Pacemaker Programming Guide
Page 87
Medtronic
Pacing mode that does not support ventricular pacing and sensing (these include
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
asynchronous modes, non-pacing modes, triggered modes, and atrial modes).
Recommended Replacement Time (RRT) is set.
6.1.4.2 Test procedure
The programmer will determine test values for mode and lower rate that are in effect before determining amplitude and pulse width thresholds.
Note: The patient should be at a resting rate before you execute this test. To verify a resting rate, the test performs an inactivity check to determine if pacing or sensing is at a low rate. If a high sensed or paced rate detected, the programmer will stop the test during the first stage of the test.
6.1.4.3 How to execute a Ventricular Capture Management test
1. Position the programming head and hold it steady for the duration of the test.
2. Tap START Test to start the test sequence, which proceeds automatically (see Section 6.1.4.4).
Note: To interrupt the test at any time during the automatic execution sequence, tap Abort & Restore. Previous parameter settings will be restored.
3. Upon successful completion, the test graphically displays the threshold measurement results as a strength duration curve.
Refer to Figure 10 for information about the strength-duration curve.
6.1.4.4 About the automatic test sequence
The following paragraphs describe each phase of the automatic test sequence, which starts when you tap START Test in step 2 of the preceding procedure.
The test applies a series of ventricular support cycles and test paces at each amplitude and pulse width value to determine if capture is present. Each series has 1 to 3 sets of support cycles, with each set followed by a test pace and a backup pace:
The support cycles are at the currently programmed amplitude and pulse width (including ventricular safety paces and nonrefractory ventricular sensed events).
The test pace is at a reduced amplitude or pulse width value, followed 110 ms later by a backup pace at the programmed amplitude and at a pulse width of 1.0 ms
Pacemaker Programming Guide 87
Page 88
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
During each phase of threshold measurements, the test annotates on the live rhythm window the test paces and whether each test pace maintained capture (CAP), lost capture (LOC), or capture could not be determined (Ignore).
Note: The programmer temporarily shortens AV intervals during the test so the pacemaker predominately paces the ventricle from conducted events.
How amplitude threshold is measured – The test first measures the stimulation threshold by reducing amplitude at a 0.4 ms pulse width. The test starts at the last ambulatory amplitude threshold measurement or at 0.75 V if the threshold has not been measured before.
Amplitude values are automatically reduced by each programmable value until loss of capture occurs. Amplitude is then increased to regain capture. The point at which capture is recovered is considered the amplitude threshold. The programmer then plots this point on the strength-duration graph.
How pulse width threshold is measured – The test next measures the stimulation threshold by reducing pulse width. The previously determined amplitude threshold value is doubled and the test starts at the last ambulatory pulse width threshold measurement. Alternately, the test will start at 0.21 ms if the threshold has not been measured before.
Pulse width values are automatically reduced by each programmable value until loss of capture occurs. Pulse width is then increased to regain capture. The point at which capture is recovered is considered the pulse width threshold. This point is plotted on the strength duration graph.
How test results are displayed – The programmer now displays a strength duration curve based on the measured amplitude and pulse width thresholds. The programmer also displays on the strength duration curve the points where the capture threshold exists for current and pending output settings (amplitude and pulse width).
Options on this screen allow you to:
Program settings for ambulatory capture management operation, particularly the amplitude margin value.
View or print the 10 s ECG strip collected during either threshold measurement.
Print a Ventricular Capture Management Threshold test report, which includes the ECG strips.
Return to the Test Setup screen, which clears this test data from the screen.
Note: If certain pacing parameter values have changed during the session, the programmer may display a pop-up Warning window advising you that Battery and Lead parameters should be remeasured. This information is needed to calculate the estimated battery life. Tap Yes to allow the programmer to make these real-time measurements.
88 Pacemaker Programming Guide
Page 89
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
Displaying ambulatory test results – Optionally, the programmer can also display a second strength-duration curve based on the last successful threshold search measurement taken by the ambulatory capture management feature. This option can be used to compare the ambulatory threshold results against the in-office test results.
6.1.4.5 Selecting and programming the Amplitude Safety Margin
From the test results screen, you can program the amplitude safety margin value by tapping Amplitude Margin. You may also program the operation of the ambulatory Capture Management feature by tapping Capture…. This opens the Capture Management window, which allows you to select several ambulatory Capture Management parameters.
6.1.4.6 How to change the Amplitude Safety Margin
1. Tap Amplitude Margin to display the safety margin options.
2. Select the desired value.
Note that the programmer displays the nearest amplitude value that corresponds to the selected safety margin and displays this value along with the pending value for pulse width under the Pending column.
The programmer also displays the pending threshold point on the strength-duration curve based on this pending safety margin.
Note: The ambulatory Capture Management feature will discontinue threshold measurements if amplitudes greater than 5.0 V or pulse widths greater than 1.0 ms are programmed.
The Ventricular Capture Management Threshold test results provide the following programmable options:
Capture…: select this field to change the operation of the ambulatory capture management feature.
Amplitude Margin: select this field to change the safety margin. Selecting an amplitude safety margin automatically displays the amplitude value that attains this safety margin.
Note: The Estimated battery life of value is an estimate of the average time remaining until pacemaker replacement is required. This estimate is based on the amplitude and pulse width values selected for programming as well as data accumulated by the pacemaker since the previous patient session.
Pacemaker Programming Guide 89
Page 90
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6.1.5 Executing an Auto Decrement (Amplitude or Pulse Width) threshold test

If you completed the setup steps on Section 6.1.2 and you selected Amplitude - Auto Decrement or Pulse Width - Auto Decrement as the threshold Test Type, use this
procedure to execute the test. The test results screen provides the option to program permanent values for amplitude and pulse width.
6.1.5.1 How to execute an Auto Decrement threshold test
1. Position the programming head and hold it steady in position for the duration of this procedure.
2. Press and hold TEST Press and Hold.
The programmer automatically reduces the test parameter (amplitude or pulse width) one programmable step every 2 to 15 pacing pulses (as set during setup). The programmed value and the point at which it takes effect shows at the top of the live rhythm display.
3. Carefully observe the patient’s ECG for loss of capture.
4. At the moment you see loss of capture, release TEST Press and Hold.
The Test Results screen shows the results of the threshold measurement.
Tap the Test Strip icon if you want to view or print a 10 s ECG strip starting 8 s before the test ended.
Note: If the displayed threshold value is not the lowest programmed test value that maintained capture, tap Threshold to change the threshold value.
Tap Print… to print a Threshold Test Report. The Threshold Test Report includes the ECG strip.
If you want to restart the test, tap Return to Setup.
6.1.5.2 How to program Amplitude and Pulse Width values
1. Tap Amplitude in the Permanent column and choose the desired setting from the values displayed.
2. Tap Pulse Width in the Permanent column and choose the desired setting from the values displayed.
3. Position the programming head and tap PROGRAM.
90 Pacemaker Programming Guide
Page 91
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6.1.6 Conducting a Manual Threshold test

If you completed the setup steps on Section 6.1.2 and you selected Manual - Amplitude & Pulse Width as the threshold test type, use this procedure to execute the test.
6.1.6.1 How to execute a Manual Threshold test
1. Position the programming head, and hold it steady for the duration of this procedure.
2. Tap START Test to program the selected test values.
3. Tap the arrows to change the amplitude or pulse width setting.
The newly displayed value is automatically programmed. The value and the point at which it was programmed show at the top of the live rhythm display.
4. Observe the patient’s ECG for loss of capture as you adjust the amplitude and pulse width settings.
By raising and lowering the parameter values, you can determine the lowest value or set of values that maintain capture.
5. To stop the test and return to permanent parameter settings, tap STOP and Restore.
6. Tap Save/Print… to display the Save/Print Test Values screen.
From the Save/Print Test Values screen, you can:
View the sequence of all test values programmed during execution of the manual threshold test.
View or print a 10 s ECG strip that starts 8 s prior to the test termination. (Tap the STOP row from the field of tested values and then tap the Test Strip icon.)
Save one or more of the programmed test values you want to include on the Manual Threshold Test report.
6.1.6.2 How to save test values for the printed report
1. Select each row to be saved.
Or, select the Select All check box to choose all test values.
If you do not want a selected value to be saved, select that row a second time.
2. Tap Save.
The test values saved are marked in the right-most column.
3. Print the Manual Threshold Test Report. Tap Print….
Pacemaker Programming Guide 91
Page 92
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6.2 Determining a sensitivity setting

The Sensing (P/R wave amplitude) test lets you determine an appropriate setting for pacemaker sensitivity. You can select an automatic or a manual test method to determine the present level of P wave or R wave signals. A Sensing test begins at the Sensing Test Setup screen.

6.2.1 The Sensing Test setup screen

To display the Sensing Test Setup screen, select Tests > Sensing.
Caution: The Sensing test requires that the pacemaker be programmed temporarily to a
pacing rate that will result in predominantly intrinsic heart activity. This is necessary to allow sensing during execution of the test. For this reason, use of the Sensing test should be limited to patients having an intrinsic rhythm that will provide adequate support during execution of the test.
Note: During the execution of a measurement, lifting the programming head from over the pacemaker for at least 2 s will return the pacemaker to its programmed state. This action should be taken in the event of a programmer malfunction, loss of power, or the absence of an appropriate command confirmation.

6.2.2 Checking the patient’s intrinsic rate

To properly conduct the Sensing test, it is helpful to first evaluate the patient’s intrinsic rhythm. Refer to Section 3.11, “Checking the patient’s underlying rhythm”, page 39.
When you observe the patient’s intrinsic rhythm, note the rate and whether or not each intrinsic depolarization is accompanied by a Sense marker (AS or VS). If Sense markers do not appear, the present setting for pacemaker sensitivity may be too high. A test sensitivity value that provides consistent sensing is required to successfully run the Sensing test.

6.2.3 Executing an automatic Sensing test

This test automatically measures and reports the present level of R wave or P wave sensing and includes the option to program a sensitivity value based on the test results.
6.2.3.1 How to execute the sensing measurement
1. Tap Tests > Sensing to display the Sensing Test Setup screen.
2. Choose P Wave Amplitude or R Wave Amplitude from the Test Type field, or choose the desired Chamber (see Table 12).
92 Pacemaker Programming Guide
Page 93
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043
3. Choose a desired Test Value for each of the test parameters listed in Table 13.
4. Position the programming head and tap Start Measurement.
Continue to hold the programming head steady for the duration of the automatic testing process.
Caution: If the patient’s condition indicates that pacing support is needed, tap Abort Test or lift the programming head for at least 2 s to terminate the test.
5. Read the messages displayed in the Test Results window. For information about test result messages, refer to Section 6.2.4.
During execution of the Sensing test, the pacemaker determines whether or not sensing is lost as it steps sensitivity through a range of programmable settings above the selected test value. If sensing is lost when the sensitivity is stepped from 2.0 mV to 2.8 mV for example, the measured amplitude is stated to be between 2.0 and 2.8 mV
Note: Because a special test circuit in the pacemaker performs this test, pacemaker operating parameters do not vary from the selected test values, unless the intrinsic signal is significantly larger than the sensitivity test value. In such cases, the test value for sensitivity is automatically adjusted upward to 1.00 mV for atrial tests and 2.80 mV for ventricular tests. The ECG and Marker Channel traces will not show any changes in pacemaker operation during the test.

Table 12. Sensing Test setup options

Setup option Available settings
Test Type Select Test Type to display the following options:
P Wave Amplitude:
Select this test to determine the level of P wave sensing in the atrium. R Wave Amplitude: Select this test to determine the level of R wave sensing in the ventricle. Manual: Select this test to manually adjust sensitivity to determine the setting at which sensing is lost.
Chamber Atrium, Ventricle:
You can select these options as an alternative to selecting P Wave Amplitude or R Wave Amplitude as the test type. For these automatic tests, the test type and chamber options are linked. If you selected Manual as the test type, select the chamber in which you want to conduct the test.
Pacemaker Programming Guide 93
Page 94
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Table 13. Sensing Test parameters

Test parameter Value selection
Mode Select the pacing mode you want to be in effect during the Sensing test. The
available options depend on the programmed pacing mode.
Lower Rate Adjust rate to a value low enough to obtain consistent sensing. The Sensing test
will not run unless the pacemaker is predominately inhibited.
AV Delay If the test mode is DDD or VDD, adjust this interval to a value long enough to
allow normal conduction. If the mode is DDD, the selected value applies to the paced AV interval and sensed AV interval during execution of the test.
Sensitivity Select a sensitivity value that provides consistent sensing. This setting would
typically be left to the same value as the present programmed value.

6.2.4 Automatic Sensing test results

Test Successful Measured P Wave Amplitude between:
1.00 mV — 1.40 mV
Retest Required Measured P Wave Amplitude > 5.60 mV
For more precise measurement, select Return to Setup button and rerun at a higher sensi­tivity value.
Measured P Wave Amplitude between:
1.00 mV — 1.40 mV For more precise measurement, select Return to Setup button and rerun at a lower sensitiv­ity value.
Test Aborted No sense markers were received during the test. Please lower the test rate or increase the AV delay to allow patient intrinsic beats and rerun the test.
The results of a successfully executed test dis­play the P wave or R wave amplitude as being between 2 sensitivity settings. You can use this information to determine an appropriate sensi­tivity setting.
This message indicates that the Test Value for Sensitivity is too low.
Loss of sensing did not occur within the range of values tested above the selected test value. In this case, you should select a higher test value for sensitivity and conduct the test again.
This message indicates that the Test Value for Sensitivity is too high.
If the test determines that sensing was lost at the next setting above the test value, the pro­grammer displays this message. In this case, select a lower value for sensitivity and conduct the test again.
This message indicates that Sense mark­ers were not present.
Telemetered Sense markers (AS or VS) were not present to start the test. Verify on the live rhythm display that sensing of the patient’s intrinsic rhythm is occurring. Adjustment of the lower rate or the Sensitivity test value may be required.
94 Pacemaker Programming Guide
Page 95
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6.2.5 Executing a Manual Sensing test

Using this test, you can determine the level of P wave or R wave sensing by manually adjusting temporarily programmed values of atrial or ventricular sensitivity.
6.2.5.1 How to execute a Manual Sensing measurement
1. Tap Tests > Sensing.
2. Choose Manual as the Test Type, then choose the desired Chamber.
3. Choose a desired Test Value for each of the test parameters.
4. Tap START Test to program the selected test values.
5. Tap the arrows to change the sensitivity setting.
The newly displayed value is automatically programmed. The value and the point at which it was programmed shows at the top of the live rhythm display.
6. Observe the patient’s ECG for loss of sensing as you adjust sensitivity. Loss of sensing is indicated by the loss of Sense markers in the chamber being tested.
By raising and lowering the sensitivity setting, you can determine the approximate sensed P wave or R wave amplitude, which is the range between the highest sensitivity setting that provided sensing and the one above it.
7. To stop the test and return to permanent parameter settings, tap Stop and Restore.
8. Tap Save/Print… to display the Save/Print Test Values screen.
From the Save/Print Test Values screen, you can:
View the sequence of all values programmed during execution of the manual sensing test.
View or print a 10 s ECG strip that starts 8 s prior to the test termination. (Tap the “Stop” row from the field of tested values and then tap the Test Strip icon.)
Save one or more of the programmed test values you want to include on the Manual Sensing Test Report.
6.2.5.2 How to save test values for the printed report
1. Select each row to be saved.
Or, select the Select All check box to choose all test values.
If you do not want a selected value to be saved, select that row a second time.
Pacemaker Programming Guide 95
Page 96
Medtronic
2. Tap Save.
The test values saved are marked in the right-most column.
Print the Manual Sensing test report. Tap Print….
Note: Only those rows that are selected or saved are printed.
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6.3 Using temporary programming to evaluate parameter settings

The Temporary test lets you evaluate different parameter settings or use high-rate pacing while having the option to quickly and easily return to the original settings. The test settings remain in effect only while you press TEST Press and Hold.

6.3.1 The Temporary Test setup screen

To display the Temporary Test setup screen, tap Tests > Temporary.
Warning: High-rate stimulation of the ventricles could result in ventricular tachycardia or
fibrillation. Application of temporary high-rate pacing should be performed only under careful patient monitoring and control.

6.3.2 Procedure for conducting a Temporary Test

1. Tap Tests > Temporary to display the Temporary - Test Setup screen.
2. Select the desired test chamber and the parameter settings you want to test. Availability of temporary pacing mode options depend on the programmed mode setting.
3. Position the programming head and continue to hold it in place for the duration of this test.
4. To initiate the test values, select and hold TEST Press and Hold.
5. To end the test, release TEST Press and Hold.
6. Tap the Test Strip icon to view or print a 10 s ECG strip starting 8 s before the temporary test ended.
7. Tap Print… to print a Temporary Test Report, which includes the 10 s ECG strip.
96 Pacemaker Programming Guide
Page 97
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

6.3.3 Programming a high temporary pacing rate

Warning: High-rate stimulation of ventricles could result in ventricular tachycardia or
fibrillation. Application of temporary high-rate pacing should be performed only under careful patient monitoring and control.
1. Tap Tests > Temporary to display the Temporary - Test Setup screen and select the desired test chamber and a single chamber test mode.
High pacing rates above 180 bpm are available only if you select a single chamber pacing mode (AAI/VVI, AOO/VOO, AAT/VVT, ADI/VDI).
2. Tap the Enable Rates > 180 bpm check box so that a check mark shows.
This action displays the High Rate Pacing Warning! window.
3. Tap OK to proceed or Cancel to discontinue this procedure. Tapping OK allows you access to pacing rates up to 400 bpm.
4. Select the desired rate and any other parameter setting you want to temporarily initiate.
5. Position the programming head and continue to hold it in place for the duration of this test.
6. To initiate the test values, press and hold TEST Press and Hold.
7. To end the test, release TEST Press and Hold.
8. Tap the Test Strip icon to view or print a 10 s ECG strip starting 8 s before the temporary test ended.
9. Tap Print… to print a Temporary Test Report, which includes the 10 s ECG strip.
Pacemaker Programming Guide 97
Page 98
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

7 Programming pacemaker parameters

7.1 Programming parameters

7.1.1 Selecting the Therapy Parameters screen

To view the present parameter settings or to program changes, tap Params. This displays the Parameters - Therapy screen. This screen allows you to select and program new parameter values applicable to the interrogated pacemaker model.

7.1.2 Programming, printing, and clearing parameter values

Programming parameters – To program pending values for one or more parameters, tap PROGRAM. This button is active only when one or more parameter values have been made
pending and no parameter values are restricted (see Section 7.1.3).
Similarly, the Program key on the programming head will respond only when PROGRAM is active on the screen.
Printing parameters – To print the currently displayed parameter values, tap the printer icon.
Clearing parameters – To clear all pending parameter values and restore the previously interrogated values on the screen, tap Undo.
Note: On primary and subordinate parameter value windows, tap Undo Pending to clear only the pending parameter values associated with that window. See Section 7.1.4.1.

7.1.3 Parameter restrictions

Certain combinations of parameter values are restricted because they are invalid or result in undesirable interactions. The programmer recognizes these combinations and may not allow programming until all parameter conflicts are resolved and all parameter selection requirements are met. A symbol that provides the status of a parameter value appears next to the value in the selection window. The symbols in Table 14 can appear next to a parameter value.
98 Pacemaker Programming Guide
Page 99
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

Table 14. Symbols that appear with parameter values

Symbol Explanation
Interlock – Indicates that the parameter value conflicts with the setting of another present
or pending value. Select another value or resolve the conflicting parameter value before programming the parameter.
Warning – Indicates that a warning message is available regarding that value. The mes­sage can be viewed either by selecting the message button or by reselecting that param­eter. In the latter case, the warning is displayed as a warning note in the selection window. These parameter values can be programmed.
Adaptive – Indicates that the programmed value can be changed automatically by the device. The symbol does not necessarily indicate that the parameter value has been adapted from a previously programmed value, only that it is able to be adapted.
Nominal – Indicates that the value is the Medtronic nominal value.
Programmed – Indicates that the value is the programmed value.
7.1.3.1 Adaptive parameter values
The adaptive symbol appearing next to a parameter value on the Parameters - Therapy screen indicates that the programmed value can be changed automatically by the pacemaker. The symbol does not necessarily indicate that the parameter value has been adapted from a previously programmed value—only that it can be adapted.
For example, the adaptive symbol appears next to the Ventricular Amplitude value, which is adaptive if Ventricular Capture Management is in the Adaptive mode. This is true whether the Adaptive mode is programmed or pending for Ventricular Capture Management.
If Ventricular Capture Management is in the Monitor Only mode (which is not adaptive), no adaptive symbol appears next to the Ventricular Amplitude value.
Pacemaker Programming Guide 99
Page 100
Medtronic
ATTESTA MRI™ SURESCAN™ / SPHERA MRI™ SURESCAN™
SW043

7.1.4 How to program parameters

All pacing therapy parameters are changed by selecting the parameter value field (the white box containing the value).
Notes:
If you intend to program a different mode, select the pacing mode first. Selecting a mode displays the parameters pertinent to that mode.
Some parameter value fields open windows that display subordinate parameters. Refer to Section 7.1.5.
For example, MVP operates when the programmed mode is either AAI<=>DDD or AAIR<=>DDDR.
7.1.4.1 How to program parameter values
1. Tap the parameter value field next to Mode.
2. From the parameter value window, select the desired value. For MVP operation, select either the AAI<=>DDD mode or the AAIR<=>DDDR mode.
Selecting a value closes the window. The selected value appears in place of the present value and is denoted as “pending” by a dashed rectangle.
Note: To clear a pending value from the window, select the parameter value field and then tap Undo Pending. To close the value window without selecting a value, tap Close or OK.
3. Repeat steps 1 and 2 to include other parameters you want to program.
4. Position the programming head and tap PROGRAM, or press the program button on the programming head.
To program all pending parameter values, tap PROGRAM. Hold the programming head over the pacemaker until the confirmation message appears in the displayed ECG strip.
100 Pacemaker Programming Guide
Loading...