Medtronic AT501 Addendum

Page 1
AT 5 0 0
DDDRP Pacing System Software Supplement Software Model 9968 5.0
System Reference Guide Supplement
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician or properly licensed practitioner.
Page 2
Page 3
Software Model 9968 5.0 0
AT500 DDDRP Pacing System 0 Software Supplement 0
A guide to using supplemental software for programming the AT500 DDDRP Pacing System
Page 4
The following are trademarks of Medtronic.
Medtronic, Marker Channel, Decision Channel, PR Logic, T-Shock, Quick Look, Quick Links, Active Can, and Flashback.
Page 5
Table of contents
About this software supplement 7
New software capabilities 8
1 Preferences update 9
Preferences 10
2 Waveform trace enhancements 13
Adjusting and Configuring Waveforms 14
AT500 DDDRP Pacing System Software Supplement
Page 6
Page 7

About this software supplement

This System Reference Guide supplement covers version 5.0 updates for software Model 9968. This guide supplements information provided in the AT500 DDDRP Pacing System Reference Guide.
Introduction
7
AT500 DDDRP Pacing System Software Supplement
Page 8
8
Introduction

New software capabilities

This software includes updates to Preference features and Waveform trace enhancements for the AT500 DDDRP Pacing System.
AT500 DDDRP Pacing System Software Supplement
Page 9
Preferences 10

Preferences update1

1
AT500 DDDRP Pacing System Software Supplement
Page 10
10
Chapter 1

Preferences

Preferences
Select printing preferences
Printing Preferences
A new menu item in the session icon allows you to access options for printing reports. Selections are retained between sessions.
Selecting the Preferences... in the session icon displays the printing options.
from this list.
Printing Preferences let you select the number of copies, default printer, and whether to print now or later. Printing Preferences are retained between sessions. The options are:
Pop up these options when any Print button is selected displays print options when any Print button is selected (default)
Number of copies (default is 1)
Printer
Full Size
Programmer (default)
Print report
Now (default)
Later from Print Queue
AT500 DDDRP Pacing System Software Supplement
Page 11
Preferences update
Preferences
For one step printing (excluding custom reports, AT/AF episodes, and VT episodes) from any screen, press [Print...] once, using the following procedure:
Procedure for one-step printing
1. Select the Session icon.
2. Select the Preferences... option.
3. Verify that Printing is selected on the list under Index.
4. Verify that Pop up these options when any Print Button is selected is not selected.
5. Select the Printer.
6. Select Now under Print Report.
7. Select [Ok].
11
AT500 DDDRP Pacing System Software Supplement
Page 12
Page 13

Waveform trace enhancements2

Adjusting and Configuring Waveforms 14
2
AT500 DDDRP Pacing System Software Supplement
Page 14
14
Chapter 2

Adjusting and Configuring Waveforms

Adjusting and Configuring Waveforms
The programmer Adjust window has been modified, and a waveform adjustment button bar has been added to each waveform for greater flexibility in adjusting waveform options.
The Adjust Window
Options available for adjusting the waveform traces using the Adjust window are as follows:
Clipping Select this checkbox to truncate the tops and
bottoms of waveform traces on the display at a 22 mm boundary. A checkmark () indicates that the filter is turned on. Default setting is ().
ECG Filter Select this checkbox to turn the ECG filter on or off.
Show Artifacts
Sweep speed Select the Sweep Speed field to display the sweep
Normalize Select this button to equalize the trace spacing and
Calibrate Select this button to send a reference signal to
OK Select this button to close the Adjust window.
A checkmark () indicates that the filter is turned on. The filter may improve clarity of the ECG in the presence of interference. Default setting is ().
Select this checkbox to enable or disable the pacing artifact enhancement function. A checkmark () indicates that the function is enabled.
speed options (12.5, 25, 50, and 100 mm/sec). To change the sweep speed, select the desired option.
adjust the size of each trace to the default setting.
Marker Channel and EGM traces on a recording.
AT500 DDDRP Pacing System Software Supplement
Page 15
Waveform trace enhancements
Adjusting and Configuring Waveforms
To Open the Adjust Window
1. Select the [Adjust...] button near the top of the tool palette. This expands the live rhythm display window (if it is not already expanded) and displays the Adjust window.
15
AT500 DDDRP Pacing System Software Supplement
Page 16
16
Chapter 2
Adjusting and Configuring Waveforms
To Adjust Clipping
1. Select the Clipping checkbox in the Adjust window.
[Clipping] button
Selecting the Clipping checkbox causes EGM and ECG waveforms to be truncated at the 22 mm boundary, with a flat line drawn at the boundary wherever a waveform is clipped.
To Select the ECG Filter
1. Select the ECG Filter checkbox to turn the ECG filter on or off. A
checkmark () indicates that the filter is turned on.
ECG Filter checkbox
Show Artifact checkbox
The ECG filter on may improve clarity of both the displayed and printed ECG in the presence of interference. The filter affects the ECG detection bandwidth as follows:
Filter Off (no checkmark) – Bandwidth is 0.05 to 100 Hz
Filter On (checkmark) – Bandwidth is 0.5 to 40 Hz
AT500 DDDRP Pacing System Software Supplement
Page 17
Waveform trace enhancements
Adjusting and Configuring Waveforms
Enabling Artifact Enhancement
Select the Show Artifacts checkbox to enable or disable the pacing artifact enhancement function. A checkmark () indicates that the function is enabled.
The two trace illustrations below show how the ECG trace appears with and without this feature enabled.
Artifact Enhancement Enabled (Checkmark)
Artifact Enhancement Disabled (No Checkmark)
Selecting a Different Sweep Speed
17
From the Adjust window, you can set the trace sweep speed to one of four settings: 12.5, 25, 50, and 100 mm/sec.
To Change the Trace Display Sweep Speed
Select the Sweep Speed field in the Adjust window and select the desired speed.
Sweep Speed Field
Menu of Choices
AT500 DDDRP Pacing System Software Supplement
Page 18
18
Chapter 2
Adjusting and Configuring Waveforms
To Normalize the Rhythm Monitor
Select the [Normalize] button in the Adjust window. This refreshes the Rhythm Monitor screen and restores each trace to its original default size, with equal vertical spacing between the traces.
The Waveform Adjustment Button Bar
A waveform adjustment button bar has been added to each waveform for greater flexibility in adjusting waveform options. To select a button, touch it with the stylus.
Normalize button
Waveform Adjustment Buttons
Increase size – press and release to increase the size of the
waveform.
Normalize – refreshes the waveform trace and restores it to
its original default size.
Decrease size – press and release to decrease the size of
the waveform.
Waveform source – use to select waveform source.
Waveform print selection – Select to enable waveform for
printing. Up to two waveforms can be enabled.
AT500 DDDRP Pacing System Software Supplement
Page 19
[Normalize] button
[Size] buttons
Waveform trace enhancements
Adjusting and Configuring Waveforms
To Adjust the Size or Amplitude of a Waveform
1. Select or in the waveform adjustment button bar for the trace you wish to adjust.
2. Press the appropriate button to increase or decrease waveform size. Each button may be pressed up to six times for EGM waveforms, and up to 4 times for ECG waveforms.
To Normalize a Waveform
To return the waveform to its default size, press the
[Normalize] button.
Selecting a Trace Source
You can change the order in which traces appear in the monitor window. One method is by changing the “source” of one or more of the traces as described below. Also, you can use the stylus to “drag” a trace to the desired location as described on page 22.
19
Note: The EGM 1 and EGM 2 sources reflect the programmed electrogram sources, which are selectable from the
Data - Collection Setup screen.
To Change the Source of a Trace
1. Select the [Source] button from the waveform adjustment button bar.
[Source] button
Traces may be superimposed on one another. In this example above, both the ECG Lead II and Marker Annotations make up the first trace. Note that the name of only one source appears at the left edge of the screen.
To display the name of the other source, “tap” the trace name at the left edge of the screen. It will change to show the source of the other trace.
AT500 DDDRP Pacing System Software Supplement
Page 20
20
Chapter 2
Adjusting and Configuring Waveforms
2. Select the source field for the trace you want to change.
Source field
3. From the drop-down menu of Source options, select the
desired source.
Selecting a source option causes the source of the trace presently displayed to change positions with the trace source you selected.
Color Coding the Traces
To Change the Color of a Trace
1. Open the Adjust... window. Select the color field in the waveform
trace area.
2. Select the color field for the trace you want to change.
Color Field
3. Select the desired color from the drop-down menu of color options.
AT500 DDDRP Pacing System Software Supplement
Page 21
Printing disabled for
this waveform
Waveform trace enhancements
Adjusting and Configuring Waveforms
Selecting waveforms for Chart Recorder Printing
At any time during a patient session, you can initiate a continuous, live recording of the patient’s ECG, EGM, and Marker Channel traces on the programmer’s chart recorder.
By default, the topmost ECG and the topmost EGM waveforms on the programmer display are enabled for printing (below).
To Select Waveforms for Chart Recorder Printing
1. Select the print button next to a print-enabled waveform that you do not want to print.
21
Printing enabled for
this waveform
This disables printing for that waveform. All other waveforms are then available for selection.
2. Select the blank button next to the waveform you want to print.
The blank button will then change to a print button, indicating that the waveform is enabled for printing.
AT500 DDDRP Pacing System Software Supplement
Page 22
22
Chapter 2
Adjusting and Configuring Waveforms
Arranging the Traces
In addition to ordering the traces by selecting the source as described on page 19, you can use the stylus to “drag” a trace to the desired position. The procedure below shows how to move the Marker Annotation trace from its current position to a position over the ECG Lead II trace.
To Move a Trace Using the Stylus
1. Press and hold the stylus against the name of the trace you want
to move.
2. Without lifting the stylus, drag the box now appearing around the
trace name and waveform adjustment button bar to the desired location.
Note: To change the position of a superimposed trace by dragging it, the source that you want to move must show at the left edge of the screen.
3. When you have the box positioned where you want the trace to
appear, lift the stylus. If you are positioning the trace over another, the trace will “snap” into position.
To equalize the spacing between the traces in the new arrangement, select the [Normalize] button in the Adjust window.
3
2
1
AT500 DDDRP Pacing System Software Supplement
Page 23

Index

A
Adjust WIndow 14 Adjust Window
opening 15 Adjustment Buttons 18 Artifact viewing enhancement 17 Artifacts
enabling 17
pacing 17
C
Clipping
adjusting 16 [Clipping] button (Live Rhythm Monitor) 14 Color coding waveform traces 20
E
ECG
adjusting sweep speed 17
filter 16 [ECG Filter] button (Live Rhythm Monitor) 14
F
Filter, ECG 16
N
Normalize
waveform 19 [Normalize] button (Live Rhythm Monitor) 14, 18
O
one-step printing 11
[Show Artifacts] button (Live Rhythm Monitor) 14 Size
waveform
adjusting 19 [Size] button (Live Rhythm Monitor) 18 Sweep speed
adjusting 17
[Sweep Speed] button (Live Rhythm Monitor) 14
T
Trace
arranging on the display 22 color coding 20 selecting the source 19
V
Viewing pacing artifacts 17
W
Waveform
arranging traces 22
normalize 19 waveform 18 Waveform Adjustment Button Bar
Adjustment Button Bar 18 Waveform Adjustment Buttons 18 [Waveform print selection] button (Live Rhythm
Monitor) 18
Waveform size
adjusting 19 [Waveform source] button (Live Rhythm Monitor) 18 Waveforms
selecting for Chart Recorder printing 21
P
preferences
printing 10
Printing
one-step 11 preferences 10 waveforms 21
S
Selecting
ECG filter 16 waveform trace color 20
AT500 DDDRP Pacing System Software Supplement
Page 24
Page 25
Page 26
Page 27
Page 28
World Headquarters
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-514-4879
Asia
Medtronic International Ltd. Suite 1602 16/F, Manulife Plaza The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: 1-800-505-4636 Tachycardia: 1-800-723-4636
Europe/Africa/Middle East Headquarters
Medtronic Europe Sàrl Route du Molliau 31 Case Postale CH-1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Medtronic E.C. Authorized Representative/Distributed by
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Australia
Medtronic Australasia Pty. Ltd. Unit4/446VictoriaRoad Gladesville NSW 2111 Australia Tel. 61-2-9879-5999 Fax 61-2-9879-5100
Canada
Medtronic of Canada Ltd. 6733 Kitimat Road Mississauga, Ontario L5N 1W3 Tel. 905-826-6020 Fax 905-826-6620 Toll-free in Canada: 1-800-268-5346
Japan
Medtronic Japan Solid Square West Tower 6F, 580 Horikawa-cho, Saiwai-ku, Kawasaki, Kanagawa 210-0913 Japan Tel. 81-44-540-6112 Fax 81-44-540-6200
*221447001*
© Medtronic, Inc. 2003 All Rights Reserved
UCX221447001 221447001 March 2003