Medtronic AMP015020001 Instructions for Use

INSTRUCTIONS FOR USE
CAUTION: Federal (USA) law restricts this product for sale by or on the order of a physician.
0110081 Rev. 04 Medtronic, Inc., Minneapolis , MN
1 of 8
Manufacturer
INSTRUCTIONS FOR USE
DESCRIPTION
The AMPHIRION™ DEEP percutaneous transluminal angioplasty (PTA) balloon dilatation catheter i s a cathe ter wi th a semi-compliant inflatable balloon mounted at the d istal tip. Dil atation balloon catheters are used to exert radial force to dilate narrow vessel segments. The AMPHIRION™ DEEP PTA catheter is available in different balloon sizes. Nominal balloon diameter and length are printed on the hub.
BALLOON CONSTRUCTION
Each balloon inflates to its nominal diameter and length at 7ATM. The working pressure range for the balloon is between the nominal pressure and the rated burst pressure. All balloons distend to sizes above the nominal size at pressures greater than the nominal pressure. Consult the Compliance Charts included in the catheter packaging for the diameters of the balloons at given pressures.
Note: The rated burst pressure is printed on the package label. In vitro testing has shown that with 95% confidence, 99.9% of the balloons will not burst at or below the rated burst pressure. Balloons should not be inflated in excess of the rated burst pressure. The nominal balloon diameters were measured in vitro and can present a maximum statistical deviation of ±10%.
A single central radiopaque marker or 2 radiopaque markers are available in order to correctly position the balloon under fluoroscopy. The markers are placed on the shaft under the balloon itself, defining its cylindrical area in the double marker version and placed in the middle of the cylindrical area in the single marker version. The catheter tip is tapered to ease entry into peripheral arteries and to facilitate the crossing of tight stenoses.
0110081 Rev. 04 Medtronic, Inc., Minneapolis , MN
CATHETER CONSTRUCTION
The AMPHIRION™ DEEP PTA balloon dilatation catheter is a dual lumen device. The lumen marked "WIRE" is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guide wire with a maximum outer diameter of 0.014 in. The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the dilatation balloon with a mixture of contrast medium and saline solution. The catheter tapers beneath the balloon segment to achieve the lowest possible deflated profile.
One or 2 radiopaque marker bands are placed under the balloon segment of the catheter shaft to provide visual reference points for balloon positioning within the vessel. The distal catheter shaft is hydrophilic coated (balloon included).
INDICATIONS FOR USE
The AMPHIRION DEEP PTA balloon dilatation catheter up to 120 mm balloon length is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra­popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
The AMPHIRION DEEP PTA balloon dilatation catheter in 150 mm and 210 mm balloon length is intended to d ilate stenoses in the femoral, popliteal, and infra-popliteal arteries.
CONTRAINDICATIONS
Inability to cross target lesion with a guide wire.
The AMPHIRION DEEP PTA balloon dilation catheter is contraindicated for use in coronary arteries or the neurovasculature.
WARNINGS
- This device is designed and intended for single use only. DO NOT RESTERILIZE
2 of 8
Manufacturer
INSTRUCTIONS FOR USE
AND/OR REUSE. Reuse or resterilization may create a risk of contamination of the device and/or cause patient infection or crossinfection, including, but not limited to , the transmiss ion of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient. Reuse or resterilization may compromise the structural integrity of the device and/or lead to device failure, which, in turn, may result in patient injury, illness, or death. Medtronic will not be responsible for any direct, incidental, or consequential damages resulting from resterilization or reuse.
- To reduce the potential for vessel damage, the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the stenosis.
- When the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.
- Do not manipulate the balloon in inflated state. The position of the balloon catheter may only be changed with the guide wire in place.
- If resistance occurs during manipulation, the cause must first be ascertained by fluoroscopy, road mapping or DSA before the balloon is moved backwards or forwards.
- The guide wire may under no circumstances be moved during inflation of the balloon.
- The balloon must be completely deflated before pulling it out of the dilated area.
- Balloon pressure should not exceed the rated burst pressure. The rated burst pressure is based on the results of in­vitro tes ting. A t least 99.9% of th e
0110081 Rev. 04 Medtronic, Inc., Minneapolis , MN
balloons (with a 95% confidence) will not burst at or below their rated burst pressure. Use of a pressure-monitoring device is recommended to prevent over pressurization.
- Use only a mixture of contrast medium and saline solution to fill the balloon (1:1). Never use air or any gaseous medium to inflate the balloon.
- Use the catheter prior to the Use Before date specified on the package.
PRECAUTIONS
- Only interventionalists wh o have sufficient experience should carry out PTA with this balloon catheter. A thorough understanding of the technical principles, clinical applications and risks associated with percutaneous transluminal angioplasty is necessary before using this product.
- AMPHIRION DEEP PTA balloon dilatation catheters should be used with caution for procedures involving calcified lesions or synthetic vascular grafts due to the abrasive nature of these lesions.
- Allergic reactions to contrast medium should be identified before PTA.
- The general technical requirements for catheter insertion must be observed at all times. This includes flushing the components with sterile, isotonic saline solution prior to use and the usual prophylactic, systemic heparinization.
- Catheter applications vary and the technique must be selected on the basis of the patient’s condition and the experience of the interventionalist.
CAUTION: Larger models of AMPHIRION DEEP PTA balloon dilatation catheters may exhibit slower deflation times particularly on long catheter shafts.
3 of 8
Manufacturer
Loading...
+ 5 hidden pages