Medtronic AF01R-R Addendum

Reprocessing Instructions for Legend Devices with Alkaline Detergents in an Automated Washer
Addendum: Integrated Power Console (IPC™) System User’s Guide 68E4066, 175027EN, M726750B246
Note: These instructions are supplemental to the Reprocessing Instructions provided in the Integrated Power Console (IPC™) System User’s Guide for the Legend EHS™, Legend EHS Stylus™, Stylus Touch™, and Legend™ Attachments and Tubes. All Warnings, Precautions, and Limitations included in the Integrated Power Console (IPC™) System User’s Guide apply.
Devices cleaned via the following procedure should be packaged, sterilized, maintained, inspected and tested and stored per the Reprocessing Instructions provided in the Integrated Power Console (IPC™) System User’s Guide prior to use.
Cleaning Instructions for Legend EHS™, Legend EHS Stylus™, and Stylus Touch™
• Review the washer-disinfector warnings in the IPC System User’s Guide before using this cleaning method.
• A nylon brush may be used to aid in cleaning.
• Remove devices from instrument trays before placing into washer.
• Orient devices following recommendations of the washer/disinfector manufacturer.
• Clean per recommended washer cycle below.
• After cleaning per the recommended washer cycle below, visually examine the devices for cleanliness.
Cleaning Instructions for Legend™ Attachments and Tubes
• Review the washer-disinfector warnings in the IPC System User’s Guide before using this cleaning method.
• Rinse attachments/tubes under running tap water, until no visible soil is noticed.
• Attachments with moving parts should be actuated through their full range of motion under running tap water.
• While rinsing under running tap water, an appropriately sized nylon brush may be used internally or externally to aid in cleaning attachments.
• Remove devices from instrument trays before placing into washer.
• Variable exposure attachments (AVAXX, AVAXXDK, AVSXX) must be placed in the washer with their tube in the fully extended position.
• Orient devices following recommendations of the washer/disinfector manufacturer.
• Clean per recommended washer cycle below.
• After cleaning per the recommended washer cycle below, visually examine the attachments/tubes for cleanliness.
Recommended Washer Cycle
Automated Cleaning - Alkaline Detergent (8.0 -10.5 pH)
Phase Recirculation Time Water Temperature Detergent Type
Pre-wash 2 minutes (02:00) Cold tap water Not applicable
Wash 5 minutes (05:00) 43°C (set point) Alkaline detergent with pH 8.0 - 10.5
Rinse 1 minute (01:00) Hot tap water Not applicable
Thermal Rinse 1 minute (01:00) 90°C Not applicable
Puried Water Rinse 1 minute 30 seconds (01:30) 66°C (set point) Not applicable
Reprocessing Instructions for Legend Devices with Alkaline Detergents in an Automated Washer
Addendum: Integrated Power Console (IPC™) System User’s Guide 68E4066, 175027EN, M726750B246
Note: These instructions are supplemental to the Reprocessing Instructions provided in the Integrated Power Console (IPC™) System User’s Guide for the Legend EHS™, Legend EHS Stylus™, Stylus Touch™, and Legend™ Attachments and Tubes. All Warnings, Precautions, and Limitations included in the Integrated Power Console (IPC™) System User’s Guide apply.
Devices cleaned via the following procedure should be packaged, sterilized, maintained, inspected and tested and stored per the Reprocessing Instructions provided in the Integrated Power Console (IPC™) System User’s Guide prior to use.
Cleaning Instructions for Legend EHS™, Legend EHS Stylus™, and Stylus Touch™
• Review the washer-disinfector warnings in the IPC System User’s Guide before using this cleaning method.
• A nylon brush may be used to aid in cleaning.
• Remove devices from instrument trays before placing into washer.
• Orient devices following recommendations of the washer/disinfector manufacturer.
• Clean per recommended washer cycle below.
• After cleaning per the recommended washer cycle below, visually examine the devices for cleanliness.
Cleaning Instructions for Legend™ Attachments and Tubes
• Review the washer-disinfector warnings in the IPC System User’s Guide before using this cleaning method.
• Rinse attachments/tubes under running tap water, until no visible soil is noticed.
• Attachments with moving parts should be actuated through their full range of motion under running tap water.
• While rinsing under running tap water, an appropriately sized nylon brush may be used internally or externally to aid in cleaning attachments.
• Remove devices from instrument trays before placing into washer.
• Variable exposure attachments (AVAXX, AVAXXDK, AVSXX) must be placed in the washer with their tube in the fully extended position.
• Orient devices following recommendations of the washer/disinfector manufacturer.
• Clean per recommended washer cycle below.
• After cleaning per the recommended washer cycle below, visually examine the attachments/tubes for cleanliness.
Recommended Washer Cycle
Automated Cleaning - Alkaline Detergent (8.0 -10.5 pH)
Phase Recirculation Time Water Temperature Detergent Type
Pre-wash 2 minutes (02:00) Cold tap water Not applicable
Wash 5 minutes (05:00) 43°C (set point) Alkaline detergent with pH 8.0 - 10.5
Rinse 1 minute (01:00) Hot tap water Not applicable
Thermal Rinse 1 minute (01:00) 90°C Not applicable
Puried Water Rinse 1 minute 30 seconds (01:30) 66°C (set point) Not applicable
Medtronic Powered Surgical Solutions
4620 North Beach Street Fort Worth, TX 76137 USA medtronic.com +1 800 643 2773
EC REP
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands +31 45 566 8000
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH- 1131 Tolochenaz Switzerland +41 21 802 7000
Australia
Medtronic Australasia Pty Ltd 5 Alma Road Macquarie Park, NSW 2113 Australia
Canada
Medtronic of Canada Ltd 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada +1 905 460 3800
Technical manuals medtronic.com/manuals
© 2016 Medtronic M971858A001 A 2016-12
Medtronic Powered Surgical Solutions
4620 North Beach Street Fort Worth, TX 76137 USA medtronic.com +1 800 643 2773
EC REP
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands +31 45 566 8000
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH- 1131 Tolochenaz Switzerland +41 21 802 7000
Australia
Medtronic Australasia Pty Ltd 5 Alma Road Macquarie Park, NSW 2113 Australia
Canada
Medtronic of Canada Ltd 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada +1 905 460 3800
Technical manuals medtronic.com/manuals
© 2016 Medtronic M971858A001 Rev A 2016-12
Loading...