Medtronic ADDR01 Reference Guide

Page 1
ADAPTA® / VERSA® / SENSIA® / RELIA™
A family of implantable pulse generators
Pacemaker Reference Guide
Reference Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Page 2
Page 3
ADAPTA® / VERSA® / SENSIA® / RELIA™
Reference Manual
A reference guide for Adapta/Versa/Sensia/Relia pacemakers
Page 4
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
Achieve, Adapta, Capture Management, Intrinsic, Marker Channel, Medtronic, MVP, Quick Look, QuickLink, Relia, Sensia, TherapyGuide, Versa
Page 5
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Contents

1 Introduction .......................................................... 9
1.1 Introduction ......................................................... 9
2 Pacing modes ....................................................... 11
2.1 Introduction ........................................................ 11
2.2 Rationale for mode selection ......................................... 13
2.3 MVP modes ........................................................ 13
2.4 DDDR mode ....................................................... 13
2.5 DDD mode ......................................................... 15
2.6 DDIR mode ........................................................ 16
2.7 DDI mode .......................................................... 17
2.8 DVIR mode ........................................................ 18
2.9 DVI mode .......................................................... 18
2.10 VDD mode ......................................................... 19
2.11 AAIR / ADIR modes ................................................. 20
2.12 AAI / ADI modes .................................................... 21
2.13 VVIR / VDIR modes ................................................. 22
2.14 VVI / VDI modes .................................................... 23
2.15 AAT / VVT modes ................................................... 24
2.16 DOOR / AOOR / VOOR modes ....................................... 25
2.17 DOO / AOO / VOO modes ............................................ 26
2.18 ODO / OAO / OVO modes ............................................ 27
3 Rate response ....................................................... 28
3.1 Introduction to rate responsive pacing .................................. 28
3.2 Preset rate response at implant ....................................... 29
3.3 Rate Profile Optimization operation .................................... 30
3.4 Individualizing Rate Profile Optimization ................................ 36
3.5 Activity sensor operation ............................................. 37
3.6 Manual control of Rate Profile Optimization ............................. 41
Reference Manual 5
Page 6
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
4 Pacemaker timing ................................................... 43
4.1 Rates ............................................................. 43
4.2 AV intervals ........................................................ 50
4.3 Rate Adaptive AV ................................................... 52
4.4 Search AV+ and diagnostic ........................................... 55
4.5 Blanking periods .................................................... 58
4.6 Refractory periods .................................................. 60
4.7 High rate atrial tracking .............................................. 68
5 Lead/cardiac tissue interface ......................................... 71
5.1 Implant Detection ................................................... 71
5.2 Automatic polarity configuration ....................................... 72
5.3 Lead Monitor ....................................................... 76
5.4 Lead impedance data ............................................... 78
5.5 Capture Management and diagnostic .................................. 79
5.6 Sensing Assurance and diagnostic .................................... 99
5.7 Manually selecting pacing parameters ................................ 103
5.8 Manually selecting sensing parameters ............................... 105
5.9 Transtelephonic follow-up features .................................... 108
6 Special therapy options ............................................. 111
6.1 Mode Switch and diagnostic ......................................... 111
6.2 Managed Ventricular Pacing (MVP) ................................... 117
6.3 Conducted AF Response ........................................... 122
6.4 Non-competitive atrial pacing ........................................ 123
6.5 PMT intervention ................................................... 125
6.6 PVC Response .................................................... 127
6.7 Ventricular Safety Pacing ............................................ 129
6.8 Sinus Preference .................................................. 130
6.9 Atrial Preference Pacing ............................................ 133
6.10 Rate Drop Response and diagnostic .................................. 137
6.11 Sleep Function .................................................... 143
6.12 Single Chamber Hysteresis .......................................... 145
7 Telemetry data ..................................................... 147
7.1 Establishing telemetry .............................................. 147
6 Reference Manual
Page 7
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
7.2 Parameter summary ................................................ 147
7.3 Patient information ................................................. 149
7.4 Using TherapyGuide to select parameter values ........................ 150
7.5 Battery and lead information ......................................... 151
7.6 Marker Channel telemetry ........................................... 153
7.7 Intracardiac electrograms ........................................... 154
7.8 Extended Telemetry ................................................ 156
8 Miscellaneous operations ........................................... 157
8.1 Magnet mode operation ............................................. 157
8.2 Temporary programming ............................................ 158
8.3 Electrical reset ..................................................... 159
8.4 Recommended Replacement Time (RRT/ERI) ......................... 161
8.5 Emergency pacing ................................................. 162
9 Diagnostics ........................................................ 163
9.1 Introduction to diagnostics .......................................... 163
9.2 Heart Rate Histograms ............................................. 166
9.3 AV Conduction Histograms .......................................... 168
9.4 Search AV+ Histogram ............................................. 170
9.5 Sensor Indicated Rate Profile ........................................ 171
9.6 High Rate Episodes ................................................ 173
9.7 Ventricular Rate Histogram During Atrial Arrhythmias .................... 179
9.8 Atrial Arrhythmia Trend .............................................. 180
9.9 Atrial Arrhythmia Durations .......................................... 181
9.10 Custom Rate Trend ................................................. 181
9.11 Key Parameter History .............................................. 183
10 Troubleshooting the pacing system .................................. 185
10.1 Troubleshooting the pacing system ................................... 185
10.2 Troubleshooting electrical problems .................................. 185
10.3 Troubleshooting hemodynamic problems .............................. 187
A Emergency settings ................................................ 189
A.1 Emergency settings ................................................ 189
Reference Manual 7
Page 8
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
B Telemetry and diagnostic values ..................................... 190
B.1 Magnet Mode operations ........................................... 190
B.2 Telemetry functions ................................................ 191
B.3 Automatic diagnostics .............................................. 193
B.4 Clinician-selectable diagnostics ...................................... 194
B.5 Cardiac event counters ............................................. 198
C Rate Response programming guidelines ............................. 199
C.1 Rate Response programming guidelines .............................. 199
D Patient counseling .................................................. 201
D.1 Patient counseling information ....................................... 201
Glossary ................................................................ 202
Index ................................................................... 209
8 Reference Manual
Page 9
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

1 Introduction

1.1 Introduction

The information in the Adapta/Versa/Sensia/Relia Pacemaker Reference Guide (PRG) applies to the following Adapta/Versa/Sensia/Relia pacemakers:
Adapta ADDR01/03/06
Adapta S ADDRS1
Adapta L ADDRL1
Adapta ADD01
Adapta ADVDD01
Adapta ADSR01/03/06
Versa VEDR01
Sensia SEDR01
Sensia L SEDRL1
Sensia SED01
Sensia SESR01
Sensia SES01
Relia REDR01
Relia RED01
Relia RESR01
Relia RES01
Relia REVDD01

1.1.1 How to use this guide

Product information about Adapta/Versa/Sensia/Relia pacemakers and the associated software for the programmer is presented in two separate guides.
The Pacemaker Reference Guide (PRG) provides detailed information on the pacemakers.
The Pacemaker Programming Guide (PPG) contains instructions on how to use the programmer and the programming software.
Reference Manual 9
Page 10
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
1.1.1.1 About the Pacemaker Reference Guide
The Pacemaker Reference Guide (PRG) describes in detail how the pacemakers operate and specifies the capabilities of the pacemakers. The PRG includes the following information:
Describes the pacing modes, rate response options, special therapy features, telemetry types, and data collection options. In some cases, guidelines are given on how to configure the pacemaker operation.
Contains troubleshooting information for electrical and hemodynamic problems.
Specifies parameter and data collection capabilities, longevity projections, and mechanical and electrical specifications.
Provides general warnings and cautions, potential interference sources, and general indications for pacing.
Contains a glossary of terms.
1.1.1.2 About the Pacemaker Programming Guide
The Pacemaker Programming Guide describes how to program Adapta/Versa/Sensia/Relia pacemakers using a programmer. The PPG presents the following information:
How to set up and configure the programmer.
How to start a patient session, use the various follow-up features during the session, and properly end the session.
How to view and print the patient’s ECG and EGM waveform traces.
How to configure the pacemaker to collect diagnostic data and how to retrieve and view this information.
How to measure stimulation thresholds and sensing levels.
How to use TherapyGuide to obtain suggested parameter values.
How to program parameter values and verify rate response parameters settings.
How to run EP Studies.
1.1.1.3 The Implant Manuals supplement these guides
For each pacemaker model in the Adapta/Versa/Sensia/Relia family, there is an implant manual. The Pacemaker Programming Guide and the Pacemaker Reference Guide do not specify which features apply to each individual pacemaker model. Refer to the applicable implant manual for specific capabilities of individual models.
10 Reference Manual
Page 11
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

2 Pacing modes

2.1 Introduction

2.1.1 Pacing mode selection

This chapter provides an introduction to pacemaker modes as an aid to pacing mode selection. The chapter is organized as follows:
Definition of basic pacing modes – The names for most of the pacing modes are defined on the 1991 ACC/AHA guidelines for pacemaker implantation.
Rationale for mode selection – In order to get pacing mode suggestions, the use of TherapyGuide is recommended. TherapyGuide is a programmer feature that suggests parameter settings based on a patient’s clinical conditions. For models which do not contain TherapyGuide, refer to the device implant manual for guidance in mode selection.
Mode descriptions – These descriptions provide indications and contraindications for modes available with the pacemaker and brief descriptions of how these modes operate.

2.1.2 NBG pacing codes

The pacemaker modes are defined in NBG Code.2 Each five-letter NBG code describes a specific type of operation for implantable pacemakers. For simplicity, this manual uses only the first three or four letters, such as DDD, DDIR, DVIR, and so forth. Figure 1 describes the first four letters of the NBG code.
1
1
Dreifus LS, Fisch C, Griffin JC, et al. Guidelines for implantation of cardiac pacemakers and antiarrhythmia devices. A report of the American College of Cardiology/American Heart Association Task Force on Assessment of Diagnostic and Therapeutic Cardiovascular Procedures (Committee on Pacemaker Implantation). Journal of the American College of Cardiology. 1991; 18: 1-13.
2
Bernstein A., et al., “The NASPE/BPEG Pacemaker Code,” PACE, 10(4), Jul-Aug 1987. (“NBG” stands for The North American Society of Pacing and Electrophysiology [NASPE] and the British Pacing and Electrophysiology Group [BPEG] Generic. NBG’s five-letter code supersedes the ICHD Code.
Reference Manual 11
Page 12
Chamber Paced
Chamber Sensed
Mode of Response
Programmable/Rate Response
DDDR
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
T = Triggered
I = Inhibited
D = Double (both)
O = None
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
P = Programmable
M = Multiprogrammable
C = Communicating
R = Rate Responsive
O = None
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 1. NBG pacing codes

2.1.3 Further information

The mode descriptions in this chapter provide only a basic overview of each mode. For further details on the rate response, timing, and therapy capabilities, refer to Section 3.1.1, “Rate response”, page 28, Chapter 4, “Pacemaker timing”, page 43, and Chapter 6, “Special therapy options”, page 111.
12 Reference Manual
Page 13
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

2.2 Rationale for mode selection

TherapyGuide offers a simple clinically-focused method for a clinician to obtain suggested parameter values. At implant or an early follow-up appointment, the clinician enters information about the patient’s clinical conditions. Based on those inputs the programmer suggests parameter settings. The suggestions are based on clinical studies, literature, current practice, and the consensus of physicians.
TherapyGuide does not replace a physician’s expert judgment. The physician is free to accept, reject, or modify any of the suggested parameter values.
For more information about TherapyGuide, refer to Section 7.4, “Using TherapyGuide to select parameter values”, page 150.
For each pacemaker model, TherapyGuide suggests a programmable mode. It bases the suggestion on clinical conditions such as the condition of the sinus node and the quality of AV conduction.
TherapyGuide offers a Rationale screen that shows the basis for each setting of pacing modes and of other parameters. To access the screen, perform the following steps:
1. Interrogate the pacemaker (before or after implant).
2. Select the Params icon. On the Therapy Parameters screen, select the [TherapyGuide] button to open the TherapyGuide window.
3. Select the [Rationale…] button to open the Rationale window.
4. Select [Close] twice to return to the Therapy Parameters screen.
Note: It is not necessary to do any parameter programming at this time. Refer to the Adapta/Versa/Sensia/Relia Pacemaker Programming Guide for instructions on programming parameters using TherapyGuide.

2.3 MVP modes

Two MVP modes are available: AAIR<=>DDDR and AAI<=>DDD.
Note: For information about AAIR<=>DDDR and AAI<=>DDD modes, refer to Section 6.2, “Managed Ventricular Pacing (MVP)”, page 117.

2.4 DDDR mode

Note: For information about the AAIR<=>DDDR mode, refer to Section 6.2, “Managed
Ventricular Pacing (MVP)”, page 117.
Reference Manual 13
Page 14
V P
V S
V S
V P
A P
A S
A P
200 ms
A P
A S
V P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
In the DDDR mode, the pacemaker tracks the faster of the intrinsic atrial rate or the sensor-indicated rate. If the intrinsic rate is faster, the DDDR mode provides atrial synchronous pacing; otherwise, AV sequential pacing occurs at the sensor-indicated rate.
Rate limits for atrial tracking (Upper Tracking Rate)3 and sensor tracking (Upper Sensor Rate) are separately programmable.
The AV intervals that follow sensed atrial events (SAV) and paced atrial events (PAV) are separately programmable, and they can be programmed to shorten with increasing rates (Rate Adaptive AV) or to change with intrinsic conduction times (Search AV+).
A nonrefractory sensed event in either chamber inhibits pacing in that chamber. A ventricular nonrefractory sensed event in the VA interval that is not preceded by an atrial sense (AS or AR) is a pacemaker-defined PVC and starts a new VA interval.

Figure 2. Example of DDDR mode operation

1 Sensor-indicated interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms PVARP = 280 ms Sensor-indicated Rate =
SAV Interval = 170 ms
90 bpm (667 ms)
3
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser value. Refer to Chapter 4, “Pacemaker timing”, page 43 for more information on TARP.
14 Reference Manual
Page 15
V P
V S
V S
A P
A S
A P
A P
200 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

2.5 DDD mode

Note: For information about the AAI<=>DDD mode, refer to Section 6.2, “Managed
Ventricular Pacing (MVP)”, page 117.
The DDD mode provides atrial synchronous pacing in the presence of intrinsic atrial activity; otherwise, AV sequential pacing occurs at the Lower Rate.
Each atrial paced or nonrefractory atrial sensed event starts an AV interval and a lower rate interval. The AV intervals that follow sensed atrial events (SAV) and paced atrial events (PAV) are separately programmable, and the SAV may be optionally programmed to shorten with increasing rate (Rate Adaptive AV) or to change with intrinsic conduction times (Search AV+).
A ventricular paced event may track an atrial sensed event up to the programmed Upper Tracking Rate.
A ventricular nonrefractory sensed event in the VA interval that is not preceded by an atrial sense (AS or AR) is a pacemaker-defined PVC and starts a new VA interval.

Figure 3. Example of DDD mode operation

4
1 Lower rate interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms
SAV Interval = 170 ms
4
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser value.
Reference Manual 15
Page 16
V P
A P
A S
A P
A P
V P
V P
V P
A P
A P
200 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

2.6 DDIR mode

The DDIR mode provides dual chamber, sensor-driven, atrioventricular (AV) sequential pacing for heart rate variation without atrial tracking.
Atrial pacing occurs at the sensor-indicated rate. If it is not inhibited, ventricular pacing occurs at the end of the PAV interval.
The AV intervals that follow paced atrial events (PAV) are separately programmable, and they can be programmed to shorten with increasing rates (Rate Adaptive AV) or to change with intrinsic conduction times (Search AV+).
An atrial event sensed outside the PVARP will inhibit a scheduled atrial stimulus but will not start an AV interval. That is, ventricular paced events after such sensed atrial events occur at the sensor-indicated rate. The following ventriculoatrial (VA) interval may be extended slightly to avoid an increasing atrial paced rate.
A ventricular nonrefractory sensed event in the VA interval starts a new VA interval.

Figure 4. Example of DDIR mode operation

1 Sensor-indicated interval 2 Sensor-indicated VA interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms Sensor-indicated Rate = 90 bpm (667 ms)
16 Reference Manual
Page 17
V P
A P
A S
A P
A P
V P
V P
200 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

2.7 DDI mode

The DDI mode provides dual chamber atrioventricular (AV) sequential pacing with atrial sensing but without atrial tracking.
Atrial pacing occurs at the Lower Rate. If it is not inhibited, ventricular pacing occurs at the end of the PAV interval.
The AV intervals that follow paced atrial events (PAV) are separately programmable, and they can be programmed to change with intrinsic conduction times (Search AV+).
An atrial event sensed outside the PVARP will inhibit a scheduled atrial stimulus but will not start an AV interval. Ventricular paced events after such sensed atrial events occur at the Lower Rate.
A ventricular nonrefractory sensed event in the ventriculoatrial (VA) interval starts a new VA interval.

Figure 5. Example of DDI mode operation

1 Lower Rate interval 2 Lower Rate VA interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms
Reference Manual 17
Page 18
V P
V S
V S
A P
200 ms
A P
A P
A P
V P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

2.8 DVIR mode

The DVIR mode provides AV sequential pacing at the sensor-indicated rate unless inhibited by ventricular sensed events.
Atrial pacing occurs at the sensor-indicated rate. If it is not inhibited, ventricular pacing occurs at the end of the PAV interval.
The AV intervals that follow paced atrial events (PAV) are separately programmable, and they can be programmed to shorten with increasing rates (Rate Adaptive AV) or to change with intrinsic conduction times (Search AV+).
The DVIR mode ignores intrinsic atrial events. Sensing occurs only in the ventricle. A ventricular nonrefractory sensed event during the ventriculoatrial (VA) interval starts a new VA interval.

Figure 6. Example of DVIR mode operation

1 Sensor-indicated interval 2 Sensor-indicated VA Interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms Sensor-indicated Rate = 90 bpm (667 ms)

2.9 DVI mode

The DVI mode provides dual chamber AV sequential pacing without atrial sensing/tracking.
Atrial pacing occurs at the Lower Rate. If it is not inhibited, ventricular pacing occurs at
18 Reference Manual
the end of the PAV interval.
Page 19
V P
V S
V S
A P
200 ms
A P
A P
V P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
The AV intervals that follow paced atrial events (PAV) are separately programmable, and they can be programmed to change with intrinsic conduction times (Search AV+).
Sensing occurs only in the ventricle, and intrinsic atrial events are ignored. A ventricular nonrefractory sensed event during the VA interval starts a new ventriculoatrial (VA) interval.

Figure 7. Example of DVI mode operation

1 Lower Rate interval 2 Lower Rate VA interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms

2.10 VDD mode

The VDD mode provides atrial synchronous pacing (or VVI pacing at the Lower Rate). The ventricle is paced synchronously up to the programmed Upper Tracking Rate.5 Sensing occurs in both the atrium and ventricle, but pacing occurs only in the ventricle.
To promote atrial synchronous pacing at slow rates, a sensed atrial event occurring near the end of the Lower Rate interval will be followed by the programmed SAV interval. The result is an extension of the ventricular lower rate.
The AV intervals that follow sensed atrial events (SAV) are separately programmable, and they can be programmed to shorten with increasing rates (Rate Adaptive AV) or to change with intrinsic conduction times (Search AV+).
5
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser value.
Reference Manual 19
Page 20
200 ms
V P
A S
A S
V P
V P
A S
A S
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
A ventricular nonrefractory sensed event in the V-V interval that is not preceded by an atrial sense (AS or AR) is a pacemaker-defined PVC, and it starts a new V-V interval.

Figure 8. Example of VDD mode operation

1 Lower Rate interval 2 SAV interval
Parameters:
Lower Rate = 60 bpm (1000 ms) SAV Interval = 200 ms Upper Tracking Rate = 120 bpm (500 ms) PVARP = 250 ms

2.11 AAIR / ADIR modes

Note: For information about the AAIR<=>DDDR mode, refer to Section 6.2, “Managed
Ventricular Pacing (MVP)”, page 117.
The AAIR mode provides atrial-based rate responsive pacing in patients with intact AV conduction. Sensing and pacing occur only in the atrium. In the absence of sensed events, the chamber is paced at the sensor-indicated rate.
The ADIR mode operates the same as the AAIR mode except that events sensed in the ventricle are recorded by the diagnostics. When used in conjunction with Marker Channel recordings and concurrent ECG, this mode may be used to observe the conducted ventricular rhythm without affecting atrial pacing.
Note: In the AAIR and ADIR modes, atrial refractory sensed events do not restart the Upper Sensor Rate interval.
20 Reference Manual
Page 21
A P
A S
A P
A P
A R
200 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 9. Example of AAIR mode operation

1 Sensor-indicated interval
Parameters:
Sensor-indicated Rate = 75 bpm (800 ms) Atrial Refractory Period = 250 ms Upper Sensor Rate = 100 bpm (600 ms)

2.12 AAI / ADI modes

Note: For information about the AAI<=>DDD mode, refer to Section 6.2, “Managed
Ventricular Pacing (MVP)”, page 117.
The AAI mode provides single chamber inhibited atrial pacing. Sensing and pacing occur only in the atrium. Pacing occurs at the programmed Pacing Rate unless inhibited by sensed events.
The ADI mode operates the same as the AAI mode except that events sensed in the ventricle are recorded by the diagnostics. When used in conjunction with Marker Channel recordings and concurrent ECG, this mode may be used to observe the conducted ventricular rhythm without affecting atrial pacing.
Reference Manual 21
Page 22
A P
A S
A P
A P
A R
200 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 10. Example of AAI mode operation

1 Pacing Rate interval
Parameters:
Pacing Rate = 75 bpm (800 ms) Atrial Refractory Period = 250 ms

2.13 VVIR / VDIR modes

The VVIR mode provides ventricular rate responsive pacing in patients for whom atrial-based pacing is deemed unnecessary or inappropriate. In the absence of sensed events, the ventricle is paced at the sensor-indicated rate.
The VDIR mode operates the same as the VVIR mode except that events sensed in the atrium are recorded by the diagnostics. When used in conjunction with Marker Channel recordings and concurrent ECG, this mode may be used to observe the underlying atrial rhythm without affecting ventricular pacing.
Note: In the VVIR and VDIR modes, ventricular refractory sensed events restart the Upper Sensor Rate interval.
22 Reference Manual
Page 23
V P
200 ms
V P
V P
V R
V P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 11. Example of VVIR mode operation

1 Sensor-indicated interval 2 Upper Sensor Rate interval
Parameters:
Lower Rate = 60 bpm (1000 ms) Upper Sensor Rate = 120 bpm (500 ms) Sensor-indicated Rate = 90 bpm (667 ms) Ventricular Refractory Period = 300 ms

2.14 VVI / VDI modes

The VVI mode provides single chamber inhibited pacing at the programmed Pacing Rate unless inhibited by sensed events. Sensing occurs only in the ventricle.
The VDI mode operates the same as the VVI mode except that events sensed in the atrium are recorded by the diagnostics. When used in conjunction with Marker Channel recordings and concurrent ECG, this mode may be used to observe the underlying atrial rhythm without affecting ventricular pacing.
Reference Manual 23
Page 24
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 12. Example of VVI mode operation

1 Pacing Rate interval
Parameters:
Pacing Rate = 60 bpm (1000 ms) Ventricular Refractory Period = 300 ms

2.15 AAT / VVT modes

In the AAT and VVT modes, pacing occurs at the programmed Lower Rate, but a nonrefractory sensed event triggers an immediate pacing output (rather than inhibiting such output). With the exception that pacing outputs occur when events are sensed, the triggered modes operate identically to the corresponding inhibited modes.
Note: Programmed triggered pacing will not occur faster than 300 ms (200 bpm) from the previous paced event. Temporary programmed triggered pacing is not limited to 300 ms (200 bpm).
24 Reference Manual
Page 25
200 ms
V P
V P
T P
V R
V P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 13. Example of VVT mode operation

1 Pacing Rate interval
Parameters:
Pacing Rate = 60 bpm (1000 ms) Ventricular Refractory Period = 300 ms

2.16 DOOR / AOOR / VOOR modes

The DOOR, AOOR, and VOOR modes operate as follows:
The DOOR mode provides asynchronous AV sequential pacing at the sensor-indicated rate. Intrinsic events are ignored.
The AOOR and VOOR modes provide single chamber pacing at the sensor-indicated rate. Intrinsic events are ignored.
Reference Manual 25
Page 26
V P
200 ms
A P
V P
A P
V P
A P
A P
V P
A P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 14. Example of DOOR mode operation

1 Sensor-indicated interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms Sensor-indicated Rate = 90 bpm (667 ms)

2.17 DOO / AOO / VOO modes

Caution: Sensing is disabled in DOO / AOO / VOO modes.
The DOO, AOO, and VOO modes operate as follows:
The DOO mode provides AV sequential pacing at the programmed Lower Rate with no inhibition by intrinsic events.
The AOO and VOO modes provide pacing at the programmed Lower Rate with no inhibition by intrinsic events in the applicable chamber.
In addition to being directly programmable, the DOO mode is the Magnet mode of the corresponding dual chamber modes, except for the VDD mode, which is the VOO mode. AOO and VOO modes are the Magnet modes of the corresponding atrial and ventricular single chamber modes, respectively.
26 Reference Manual
Page 27
V P
200 ms
A P
V P
A P
V P
A P
A P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 15. Example of DOO mode operation

1 Lower Rate interval
Parameters:
Lower Rate = 60 bpm (1000 ms) PAV Interval = 200 ms

2.18 ODO / OAO / OVO modes

Warning: ODO / OAO/ OVO modes disable pacing. Never program these modes for
pacemaker-dependent patients. For pacemaker-dependent patients, use the programmer’s inhibit function for brief interruption of outputs.
In the ODO, OAO, and OVO modes, sensing occurs in the designated chamber or chambers. When used in conjunction with Marker Channel telemetry and concurrent ECG, these modes may be used to observe underlying rhythms.
Blanking periods in these modes are automatically minimized to maximize the sensing window or windows. Thus, Marker Channel telemetry may display sense markers for cardiac events (for example, far-field R waves) that otherwise would not appear due to longer blanking.
No timing intervals or refractory periods are used.
Reference Manual 27
Page 28
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

3 Rate response

3.1 Introduction to rate responsive pacing

3.1.1 Rate response

The pacemaker may provide appropriate rate response for patients who require cardiac pacing support at both submaximal and maximal rates. Submaximal rates are moderate pacing rates near the Activities of Daily Living Rate (ADL Rate) obtained during typical daily activities, such as walking or daily chores. Maximal rates are rates (at or near the Upper Sensor Rate) obtained during vigorous activities. To achieve appropriate rate response, the pacemaker provides activity sensor-driven pacing with rate response control in both the ADL rate range and the exertion rate range.
The pacemaker provides appropriate rate response by employing the following operations:
Three programmable rates control the submaximal and maximal rate ranges: Lower Rate, ADL Rate (Activities of Daily Living Rate), and Upper Sensor Rate. The ADL Rate is equivalent to the average target rate that the patient achieves for moderate activities.
Independent control of rate response is provided in both the ADL and exertion rate ranges.

3.1.2 Automatic features

For models in a rate responsive mode, the pacemaker automatically enables rate response after implant and automatically adjusts rate response, if necessary, once each day.
During the first 30 minutes after implant, pacing occurs at the implanted mode but without rate response. 30 minutes after implant, rate response operation is enabled.
Once each day, rate response is assessed and adjusted, if necessary, in the ADL and exertion rate ranges. The assessment is based on comparing the pacemaker’s historical sensor-indicated rate profiles against a clinician prescribed target rate profile of the patient. If the rate profiles differ, rate response is adjusted slightly in the appropriate rate range, and the assessment is repeated again the next day.
28 Reference Manual
Page 29
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

3.1.3 For further information

Refer to Section 3.3, “Rate Profile Optimization operation”, page 30 for information on how the pacemaker optimizes rate response.

3.2 Preset rate response at implant

3.2.1 Overview

Pacemakers shipped in rate responsive modes operate in a non-rate-responsive mode until implant detection is completed, which is typically 30 minutes after implant. Thereafter, the pacemakers automatically enable rate responsive pacing. Consequently, no programming is required for rate response operation.

3.2.2 Three pacing rate controls

If customization of rate response is desired, three pacing rates are provided to control the ADL and exertion rate ranges:
Lower Rate defines the slowest rate at which pacing occurs in the absence of a sinus rate or physical activity.
ADL Rate (Activities of Daily Living Rate) is the approximate rate that the patient’s heart is expected to reach during moderate exercise.
Upper Sensor Rate provides the upper limit for the sensor-driven rate during vigorous exercise.
Refer to Section 4.1, “Rates”, page 43 for additional considerations when selecting pacemaker rates.

3.2.3 Starting rate response immediately

In situations where the clinician wishes to start rate responsive pacing before the 30-minute implant detection period is completed, perform the following steps:
1. After the device is implanted, program Implant Detection to “Off/Complete.”
2. Configure pace and sense lead polarities and Lead Monitor.
3. Verify that Rate Profile Optimization is On.
4. Verify that the parameter values for Lower Rate, ADL Rate, and Upper Sensor Rate are appropriate.
5. Verify that the parameter values for ADL Response, Exertion Response, Activity Threshold, Activity Acceleration, and Activity Deceleration are appropriate.
Reference Manual 29
Page 30
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

3.2.4 For further information

Refer to Section 3.3, “Rate Profile Optimization operation”, page 30 and Section 3.4, “Individualizing Rate Profile Optimization”, page 36.

3.3 Rate Profile Optimization operation

3.3.1 Overview

When Rate Profile Optimization is programmed On, the pacemaker can adapt ADL and exertion rate response levels once each day by comparing the patient’s current sensor rate profiles against a target rate profile. This feature is intended to provide automatic and independent monitoring of rate response at both moderate rates for daily patient activities, such as walking and daily chores, and at exertion rates for vigorous patient activities.
Optimization can be individualized to the patient’s activity levels. Refer to Section 3.4, “Individualizing Rate Profile Optimization”, page 36.
Optimization can also operate in the background when a non-rate responsive mode is programmed. This can provide appropriate rate response to patient activity if rate response is needed later or for certain therapy features, such as mode switching to a non-atrial tracking rate responsive mode.

3.3.2 Rate control in the ADL and exertion rate ranges

The pacemaker maintains linear rate control between the activity sensor signal and the sensor-indicated rate from the Lower Rate to the ADL Rate. Refer to Section 3.5.2, “How Activity Threshold influences rate”, page 38. It maintains independent linear rate control in the exertion rate range. Optimization controls how rapidly and to what level the sensor-indicated rate increases and decreases in these two rate ranges. The three programmable rate controls [Lower Rate, ADL Rate (Activities of Daily Living Rate), and Upper Sensor Rate] define the rate ranges (see Figure 16).
Moderate pacing rates are achieved during typical daily patient activities. These rates (in the ADL rate range) are at or near the ADL Rate.
Exertion pacing rates are achieved during vigorous activities. These rates (in the exertion rate range) are at or near the Upper Sensor Rate
Figure 16 shows a graph of sensor-indicated rate as a function of increasing activity. The sensor-indicated rate curve has two slopes. The first slope, which controls how aggressively the pacing rate increases from the Lower Rate to the ADL Rate, is determined by the programmed ADL Response parameter. The second slope, which controls how aggressively the pacing rate approaches the Upper Sensor Rate, is determined by the programmed Exertion Response parameter.
30 Reference Manual
Page 31
Upper Sensor Rate
ADL Rate
Increasing activity
Lower Rate
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
When you program new values for rates or Rate Profile Optimization, immediate changes occur. The new values are predictions based on automatic diagnostic data and the selected Rate Profile Optimization settings. The pacemaker continues to adjust Rate Response over time.
Note: If the patient does not have any data in the Sensor Indicated Rate Profile diagnostic, optimization does not adjust immediately when these parameters are programmed. 24 hours of diagnostic data are required.

Figure 16. A sensor-indicated rate curve

1 ADL rate range 2 Exertion rate range

3.3.3 Optimization using rate profiles

Optimization of rate response occurs independently in both the ADL rate range and the exertion rate range. The sensor-indicated rate curve is assessed daily based on the following rate profile data:
Sensor rate profile – An actual rate versus time distribution of the patient’s averaged sensor-indicated rates. Once each day, the pacemaker collects a daily sensor rate profile and cumulates the data into a long-term average. Both the daily and long-term rate profiles are assessed each day to determine if adjustments to rate response are required. The long-term sensor rate profile is automatically stored in the Sensor Indicated Rate Profile diagnostic.
Target rate profile – A programmable rate versus time distribution of the patient’s desired rates. The ADL Response and Exertion Response parameters define the percentage of time that the sensor-indicated rate stays in the ADL rate range and in the exertion rate range, respectively.
Reference Manual 31
Page 32
30%
20%
10%
Lower Rate
ADL Rate
Upper Sensor Rate
Percentage of time
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Figure 17 shows a typical rate profile (either a sensor rate profile or a target rate profile).

Figure 17. Example of a rate profile

1 ADL rate range 2 Exertion rate range

3.3.4 Daily optimization of rate response

Once each day, the pacemaker evaluates the percentage of time the sensor rate is in the ADL and exertion rate ranges by comparing the daily and long-term sensor rate profiles against the target rate profile. This operation follows the sequence shown in Figure 18.
The pacemaker calculates the sensor indicated rate based on the activity sensor signal.
From the actual sensor indicated rate values, it generates a daily sensor rate profile. It also merges that data into a long-term sensor rate profile.
It compares the target rate profile to the daily and long-term sensor rate profiles. Refer to Figure 19 and Figure 20 for details.
If the sensor rate profiles match the target rate profile or if the daily and long-term sensor rate profiles contradict each other, no rate response adjustments occur.
Otherwise it makes an automatic adjustment to rate response, which affects the calculation of the sensor-indicated rate in either or both of the rate ranges.
This sequence repeats each day.
32 Reference Manual
Page 33
CompareCompare
Daily sensor rate
profile
Long-term sensor
rate profile
Adjust ADL rate
response as needed
based on comparison
Adjust exertion rate
response as needed
based on comparison
Sensor signal processing
Target rate profile
Rate calculation
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
As a result of this operation, the pacemaker automatically adjusts rate response in the ADL and exertion ranges, if necessary, based on the following criteria:
If the sensor rate profiles show a higher percentage of time spent pacing than the target rate profile, rate response for the pertinent rate range is set to be less responsive. Conversely, if a lower percentage of time spent pacing is profiled than targeted for, rate response is set to be more responsive.
If the sensor rate profiles match the target rate profile or the daily and long-term sensor rate profiles contradict each other, no rate response adjustments occur.

Figure 18. Daily operation of Rate Profile Optimization

The goal of this operation is to keep the patient’s sensor rate profiles equivalent to the target rate profile. This is shown in two examples.
Reference Manual 33
Page 34
Percentage of time
Increasing rate
Rate response is made more aggressive
Increasing rate Increasing rate
= Target rate profile
= Sensor rate profile = New rate response curve
= Old rate response curve
Rate response is made less aggressive
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
In Figure 19, a comparison of the sensor rate profile and target rate profile shows that pacing in the ADL rate range occurs for a larger percentage of time than was targeted. In the sensor rate curve, rate response is adjusted to be less aggressive in this range.
The same comparison shows that pacing in the exertion rate range occurs for a smaller percentage of time than was targeted. In the sensor rate curve, rate response is adjusted to be more aggressive in this range.

Figure 19. Result of comparing rate profiles: first example

1 ADL rate range 2 Exertion rate range
The example in Figure 20 is the opposite of the one in Figure 19. Lower than targeted pacing in the ADL rate range results in a rate response adjustment to make rate response more aggressive in this range. Higher than targeted pacing in the exertion rate range results in rate response that is less aggressive in this range.
34 Reference Manual
Page 35
Increasing activity
= Target rate profile
= Sensor rate profile
= Old rate response curve
= New rate response curve
Increasing rate
Percentage of time
Rate response is made more aggressive
Rate response is made less aggressive
Increasing rate
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 20. Result of comparing rate profiles: second example

1 ADL rate range 2 Exertion rate range
Note: Two additional cases are possible:
Lower than targeted pacing in both the ADL and exertion rate range. Rate response is adjusted to be more aggressive in both ranges.
Higher than targeted pacing in both the ADL and exertion rate range. Rate response is adjusted to be less aggressive in both ranges.

3.3.5 Adaptations in Optimization operation

The pacemaker adapts rate response more rapidly for the first ten days after Optimization is first activated post-implant or after certain rate response parameters are manually reprogrammed (e.g., Lower Rate, ADL Rate, Upper Sensor Rate, ADL Response, or Exertion Response). The intent is to quickly match rate response to the target rate profile prescribed by the parameter changes.
When you program new values for rates or Rate Profile Optimization, immediate changes occur. The new values are predictions based on automatic diagnostic data and the selected Rate Profile Optimization settings. The pacemaker continues to adjust Rate Response over time.
Note: If the patient does not have any data in the Sensor Indicated Rate Profile diagnostic, optimization does not adjust immediately when these parameters are programmed.
Reference Manual 35
Page 36
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Optimization is skipped on any day that a device interrogation or parameter programming occurs.

3.4 Individualizing Rate Profile Optimization

3.4.1 Overview

The clinician can prescribe a target rate profile using the ADL Response and Exertion Response parameters to match the patient’s life-style or activity levels. The programmable ADL Response parameter alters the targeted rate distribution in the ADL rate range, while the Exertion Response parameter alters the rate distribution in the exertion rate range.

3.4.2 ADL rate profiles

The nominal setting for the ADL Response parameter is “3.” Programming a higher number redefines the target rate profile to spend more time pacing at or above the ADL Rate, thereby increasing rate responsiveness in the ADL rate range. Programming a lower number redefines the rate profile to spend less time pacing at or above the ADL Rate, thereby decreasing rate responsiveness.

3.4.3 Exertion rate profiles

The nominal setting for the Exertion Response parameter is “3.” Programming a higher number redefines the target rate profile to spend more time pacing near the Upper Sensor Rate, thereby increasing rate responsiveness in the exertion rate range. Programming a lower number redefines the rate profile to spend less time pacing near the Upper Sensor Rate, thereby decreasing rate responsiveness.

3.4.4 Programming guidelines

If it is necessary to adjust rate response from the nominal setting, first verify that the three rate controls are appropriate for the patient. Refer to Section 3.2.2, “Three pacing rate controls”, page 29.
If these rate control settings are appropriate, the ADL Response and/or Exertion Response settings can then be adjusted based on the guidelines in Table 1.
36 Reference Manual
Page 37
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Table 1. ADL Response and Exertion Response guidelines

Rate region Patient Select these settings
ADL Response
Lower Rate to ADL Rate Reached ADL Rate too quickly Lower number (less rate
response)
Reached ADL Rate too slowly Higher number (more rate
response)
Exertion Response
ADL Rate to Upper Sensor Rate Reached Upper Sensor Rate
too quickly Reached Upper Sensor Rate
too slowly
a
If a higher Exertion Response setting has not produced the desired rate response, increase the ADL Response setting.
Lower number (less rate response)
Higher number (more rate response)
a
For more detailed programming guidelines, refer to Table 29, page 199 and Table 30, page 200, which list the targeted time spent pacing for the five ADL Response and Exertion Response settings.

3.5 Activity sensor operation

3.5.1 Overview

Activity sensor based pacing is controlled by the following programmable parameters:
Activity Threshold determines the minimum intensity of detected physical activity to which the pacemaker responds.
Activity Acceleration and Activity Deceleration times control how rapidly the pacing rate changes in response to increased or decreased activity. One programmable Activity Deceleration setting, “Exercise,” provides an extended deceleration period following prolonged exercise.
Note that Activity Threshold, Activity Acceleration, and Activity Deceleration are automatically set to shipping settings 30 minutes after implant or can be manually programmed.
Reference Manual 37
Page 38
Activity Sensor
Ouput
Time
Activity Threshold = Medium/ Low
High
High
Med/High
Med/High
Med/Low
Med/Low
Low
Low
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

3.5.2 How Activity Threshold influences rate

A transducer, bonded to the pacemaker circuitry, is deflected by physical motion. The activity sensor converts detected motion into electrical signals. The programmed Activity Threshold screens out activity signals below the selected setting. Detected sensor signals vary from patient to patient. Only sensor signals whose amplitude exceeds the programmed Activity Threshold (as shown in Figure 21) are used in computing the sensor-indicated rate. The lower the Activity Threshold, the smaller the signal required to influence the sensor-indicated rate.

Figure 21. Activity sensor signal (threshold set to medium/low)

3.5.3 Evaluating the Activity Threshold setting

Activities such as walking increase the pacing rate; sitting results in pacing at or near the programmed Lower Rate. Use Table 2 as a guide for selecting an appropriate setting.

Table 2. Activity Threshold guidelines

Programmable set­tings Typical rate performance
Low Responds to most body activity, including minimal exertion. Medium/Low Limited response to minimal exertion; responds to moderate or greater
exertion.
38 Reference Manual
Page 39
210 4 53
15 s 30 s 60 s
Upper Sensor Rate
Lower Rate
Time (min)
Activity Acceleration programmable settings
Rate range
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Table 2. Activity Threshold guidelines (continued)
Programmable set­tings Typical rate performance
Medium/High Limited response to moderate body movements and exertion. High Responds to only vigorous body movements and exertion.

3.5.4 How Activity Acceleration and Deceleration influence rate

Activity Acceleration and Activity Deceleration times control how rapidly the pacing rate changes in response to increased or decreased physical activity. One programmable Activity Deceleration setting, “Exercise,” provides an extended deceleration period following prolonged exercise.
Activity Acceleration time is the time required to achieve approximately 90% of the difference between the current rate and a higher steady-state rate consistent with the current level of activity. Figure 22 shows a graphic representation of the acceleration curves at the onset of strenuous exercise.
Activity Deceleration time is the time required to achieve approximately 90% of the difference between the current rate and a lower steady-state rate consistent with the current level of activity. Figure 23 shows a graphic representation of the deceleration curves at an abrupt cessation of strenuous exercise.

Figure 22. Activity Acceleration curves

Reference Manual 39
Page 40
6543210 7 8 9 10
2.5 min 5 min 10 min
Upper Sensor Rate
Lower Rate
Time (min)
Activity Deceleration programmable settings
Rate range
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 23. Activity Deceleration curves

3.5.5 Exercise Deceleration operation

Activity Deceleration programmed to “Exercise” extends the rate slowing period following an exercise episode, providing up to 20 minutes of rate deceleration. When it is programmed on, the pacemaker uses activity sensor data to detect periods of vigorous, prolonged exercise. At the end of such an exercise period, the pacemaker uses a longer deceleration curve for the central portion of the programmed rate range. The actual deceleration rate is determined dynamically based on the intensity and duration of exercise and the new level of activity. Figure 24 shows the composite deceleration curve that applies after the abrupt cessation of sustained exercise.
40 Reference Manual
Page 41
121086420 14 16 18 20 22
Upper Sensor Rate
Lower Rate
Time (min)
Rate range
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 24. Exercise Deceleration

1 5 min deceleration curve 2 Begins exercise deceleration
3 Ends exercise deceleration 4 5 min deceleration curve

3.6 Manual control of Rate Profile Optimization

3.6.1 Overview

As an alternative to automatic Rate Profile Optimization, a programmer assisted Exercise test can be used to manually set rate response for the ADL and exertion rate ranges. The Exercise test is used to immediately set rate response to certain levels. Rate response parameters remain set to their programmed values if Optimization is Off. When Optimization is On, it can adjust these parameters once each day.

3.6.2 Evaluate and program rate response

The Exercise test is used to evaluate the patient’s rate response and allow the programmer to customize two rate response control parameters:
ADL Setpoint (Activities of Daily Living Setpoint) determines the minimum sensor response to pace at the ADL Rate, which falls within the ADL rate range.
UR Setpoint (Upper Rate Setpoint) determines the minimum sensor response to pace at the Upper Sensor Rate, which is at the upper limit of the exertion rate range.
Note: The programmed ADL Setpoint setting must be less than the UR Setpoint setting.
Reference Manual 41
Page 42
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Refer to the Adapta/Versa/Sensia/Relia Pacemaker Programming Guide for programming instructions.
42 Reference Manual
Page 43
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4 Pacemaker timing

4.1 Rates

4.1.1 Overview

The following programmable rates control timing in the pacemaker:
Normal operating rates:
– Lower Rate
– ADL Rate
– Upper Tracking Rate
– Upper Sensor Rate
Other operating rates:
– Sleep Rate (for Sleep function)
– Hysteresis Rate (for single chamber demand and triggered modes)
– Sinus Preference Zone (for Sinus Preference)
– Intervention Rate (for Rate Drop Response)
– Overdrive Rate (for Post Mode Switch Overdrive Pacing)
– Maximum Rate (for Conducted AF Response)
– Maximum Rate (for Atrial Preference Pacing)
Additionally, rates calculated by the pacemaker are used for some operations. These are:
Sensor-indicated rate
Mean atrial rate
The other operating rates are described in Chapter 6, “Special therapy options”, page 111 along with the functions that use them. The normal rates are described in this chapter.

4.1.2 A-A and V-V timing

A-A timing – In all modes that pace the atrium, the pacemaker times from atrial event to atrial
event (A-A timing). This timing method mimics a natural sinus rhythm, producing A-A intervals that are nearly equal, except when timing is interrupted by one of the following:
PACs in DDIR and DDI modes
PVCs in DDDR, DDD, DDIR, DDI modes (PVC Response operation)
Reference Manual 43
Page 44
200 ms
P
P
V
DDD
A
S
S
P
S
P
P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
A ventricular sensed event during the VA interval in the DVIR and DVI modes
An atrial refractory sensed event that triggers an NCAP extension
VA intervals vary due to adjustments by A-A timing operations in order to achieve sensor-indicated or lower rate operation in the presence of varying AV conduction.
V-V timing – In modes that do not pace the atrium (e.g., VDD or VDIR) or single chamber ventricular modes, the pacemaker times from ventricular event to ventricular event (V-V timing).

4.1.3 Lower Rate

The programmed Lower Rate defines the slowest rate at which pacing occurs during a mode’s basic operation. In rate responsive modes, in the absence of sensor-detected activity, the sensor-indicated rate is equal to the programmed Lower Rate.

Figure 25. Example of Lower Rate operation

1 Lower Rate interval
Parameters: Lower Rate = 60 bpm (1000 ms) PVARP = 300 ms
PAV Interval = 200 ms Ventricular Refractory Period =
240 ms
SAV Interval = 180 ms
44 Reference Manual
Page 45
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.1.4 Operating lower rate

Under certain circumstances, the programmed Lower Rate may be overridden by an operating lower rate that is higher or lower than the programmed value. The following rates may become the operating lower rate:
Switching from and back to atrial tracking mode (for Mode Switch)
Conducted AF Response determined rate
Sinus Preference Zone (for Sinus Preference)
Sleep Rate (for Sleep function)
Intervention Rate (for Rate Drop Response)
Hysteresis Rate (for single chamber modes)
Threshold Margin Test rate of 100 bpm
Magnet Mode rate of 85 bpm
Recommended Replacement Time (RRT/ERI) rate of 65 bpm
Overdrive Rate (for Post Mode Switch Overdrive Pacing function)
Rate determined by Atrial Preference Pacing
Rate determined by Capture Management (ACM and VCM)
Sensor indicated rate

4.1.5 Selecting a Lower Rate

Program the Lower Rate to maintain adequate heart rates during periods of inactivity or during pauses in atrial rhythms when the pacemaker is operating in the DDDR, DDD, VDD, AAIR, ADIR, AAI, and ADI modes.
Note: In the VDD mode, atrial tracking near the Lower Rate may result in V-V intervals that exceed the Lower Rate interval. This is normal operation.
Lower Rates from 120 to 130 bpm are intended for pediatric patients. Lower Rates below 50 bpm and above 100 bpm are primarily intended for diagnostic purposes.
Reference Manual 45
Page 46
200 ms
P
V
DDDR
A
S
SS P
PPP
Sensor Sensor Sensor
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.1.6 Sensor-indicated rate

The sensor-indicated rate is the basic pacing rate when the pacemaker is operating in a rate responsive mode (DDDR, DDIR, DVIR, DOOR, VVIR, VDIR, VOOR, AAIR, ADIR, or AOOR). It is determined by the pacemaker based on the sensor-detected level of patient activity and the programmed rate response parameters. The sensor-indicated rate will never be greater than the Upper Sensor Rate or less than the Lower Rate.

Figure 26. Example of sensor-indicated rate operation

1 Sensor-Indicated interval
Parameters: Sensor-Indicated Rate =
PVARP = 300 ms
90 bpm (667 ms) PAV Interval = 200 ms Ventricular Refractory Period =
220 ms
SAV Interval = 190 ms
In rate responsive modes, the sensor-indicated rate tracks the activity sensor, which is detected by the transducer sensor’s frequency and amplitude.
In dual chamber rate responsive modes, the sensor-indicated interval is the AS-AP or AP-AP interval.
In single chamber rate responsive modes, the sensor-indicated interval is the A-A or V-V interval. In these modes, sensor-indicated rate intervals start with a sensed or paced event in the chamber being paced.
46 Reference Manual
Page 47
200 ms
S
P
DDD
A
V
S
P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.1.7 Sensor indicated rate effect on other intervals

The sensor-indicated rate is used to determine the values of certain other timing intervals. These intervals are:
Rate adaptive paced AV (PAV) interval
Sensor-varied PVARP (even in non-rate responsive DDD and VDD modes)
PVARP extension (sensor-corroboration before PMT intervention)

4.1.8 ADL Rate

The ADL Rate (Activities of Daily Living Rate) is the target rate which the patient’s heart rate is expected to reach during moderate exercise.

4.1.9 Upper Tracking Rate

The programmable Upper Tracking Rate is the maximum rate at which the ventricle may be paced in response to sensed atrial events when the pacemaker is operating in the DDDR, DDD, and VDD modes. Sensed atrial events below the Upper Tracking Rate will be tracked at a 1:1 ratio, but sensed events above the Upper Tracking Rate will result in pacemaker Wenckebach (for example, 6:5, 4:3, 3:2, or 2:1 block). The Upper Tracking Rate usually should be programmed to a value less than the 2:1 block rate. Refer to Section 4.7, “High rate atrial tracking”, page 68 for details.

Figure 27. Example of Upper Tracking Rate (Wenckebach) operation

1 Upper Tracking Rate
Reference Manual 47
Page 48
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Parameters:
Sensor-indicated Rate = 75 bpm (800 ms)
Upper Tracking Rate = 100 bpm (600 ms)
SAV Interval = 200 ms

4.1.10 Upper Sensor Rate

In rate responsive modes, the programmable Upper Sensor Rate provides the upper limit for the sensor-indicated rate during physical activity, particularly during vigorous exercise. In the DDDR mode, the Upper Sensor Rate may be higher than, lower than, or the same as the Upper Tracking Rate.

4.1.11 Programming considerations and restrictions

ADL Rate – It is recommended that the ADL Rate be at least 10 bpm less than the Upper
Sensor Rate or 20 bpm greater than the Lower Rate. However, programming the ADL Rate above or below these limits is permitted.
Upper rates – Programming a combination of high Upper Sensor Rate and Upper Tracking Rate and a long refractory period may result in a shorter “sensing window.” Loss of sensing in such cases could result in competitive pacing (unless Non-Competitive Atrial Pacing is programmed On). See Section 6.4, “Non-competitive atrial pacing”, page 123 for more information.
Programming the Upper Tracking Rate to a value greater than the Upper Sensor Rate permits the atrial rhythm to be tracked to a rate higher than the sensor-driven rate.
The Upper Sensor Rate and/or Upper Tracking Rate must be greater than the Lower Rate. The Upper Sensor Rate must be greater than the ADL Rate.

4.1.12 Rate limit

An internal circuit, independent of the pacing timers, limits single chamber atrial or ventricular pacing rates to 200 bpm (±20 bpm) for most single component failures. For dual chamber modes, atrial and ventricular rates are limited independently to 200 bpm (±20 bpm). The rate limit is automatically disabled during temporary pacing in the AAI, ADI, AAT, AOO, VVI, VDI, VVT, and VOO modes to allow high rate pacing for diagnostic or therapeutic purposes.
Note: When the Upper Tracking Rate is programmed to 190, 200, or 210, the circuit limit is 227 bpm (±17 bpm).
48 Reference Manual
Page 49
60
80
100
120
140
160
180
200
0 5 10 15 20 25 30 35
Time (seconds)
Rate (min
-1
)
MAR
Intrinsic rate
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.1.13 Possible atrial competition at high rates

At high sensor-driven rates when the pacemaker is operating in the DDDR and DDIR modes, sensor-driven pacing may approximate the intrinsic atrial rate, with some intrinsic atrial events falling into the PVARP. This could result in asynchronous pacing with the potential for competitive atrial pacing. Consider the potential for asynchronous pacing at high rates before selecting an Upper Sensor Rate, especially for patients known to be susceptible to induction of atrial tachyarrhythmias. Weigh the benefits of high rate sensor-driven pacing against the potential for competitive pacing.
Note: Use of the Rate Adaptive AV feature and sensor-varied or automatic PVARP can reduce the likelihood of the type of asynchronous pacing described above. When the pacemaker is operating in the DDDR mode, Sinus Preference and NCAP can also be considered.

4.1.14 Mean atrial rate

The mean atrial rate (MAR) is a running average of the atrial rate for use by the Rate Adaptive AV and automatic PVARP features. The average uses all A-A intervals (except AS-AP or AR-AP intervals). In order to respond quickly to rapidly increasing atrial rates, the average gives preference to shorter A-A intervals over longer intervals when calculating the MAR. Figure 28 shows how the MAR tracks an increasing intrinsic atrial rate.

Figure 28. Increasing mean atrial rate

Reference Manual 49
Page 50
P
DDD
P P
A
V
P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.2 AV intervals

4.2.1 Overview

In dual chamber modes, the AV intervals determine the time between the occurrence of an atrial event and the scheduled delivery of a ventricular stimulus. Separate AV intervals for paced and sensed atrial events are available. The lengths of these intervals may be programmed to fixed values or (optionally) rate adaptive or therapeutically determined.
Paced AV Interval (PAV) – PAV follows an atrial pace when the pacemaker is operating in the DDDR, DDD, DDIR, DDI, DVIR, DVI, DOOR, and DOO modes. The PAV interval duration may differ from the programmed value due to one of the following operations:
Rate Adaptive AV
Search AV+
Ventricular Safety Pacing
Non-Competitive Atrial Pacing

Figure 29. Example of PAV interval operation

1 PAV interval 2 PAV
Sensed AV Interval (SAV) – SAV follows an atrial sensed event when the pacemaker is operating in an atrial synchronous pacing mode (DDDR, DDD, and VDD). The SAV interval duration may differ from the programmed value due to one of the following operations:
Rate Adaptive AV
Automatic PVARP
Search AV+
Wenckebach
50 Reference Manual
Page 51
S
P P
DDD
A
V
S
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
For Wenckebach operation, the SAV is extended to avoid violation of the Upper Tracking Rate or the total atrial refractory period while tracking a fast intrinsic atrial rate.

Figure 30. Example of SAV Interval operation

1 SAV interval 2 SAV

4.2.2 Selecting PAV and SAV

Using MVP or Search AV+ should eliminate the need to manually adjust the AV intervals for most patients. It is recommended that MVP or Search AV+ be used to reduce pacing in the right ventricle.
However, when programming AV intervals in patients with third degree block, the general hemodynamic goal is to assure that, to the extent possible, left-atrial systole is completed before left-ventricular systole begins. To achieve this, the AV interval durations may be adjusted independently of each other.
To accommodate the difference in interatrial conduction times, the SAV usually should be programmed to a shorter duration than the PAV, typically 30 to 50 ms shorter. If an SAV greater than the PAV is selected, the programmer notes that this is not usual, but the selected values may be programmed if clinically warranted.
When the SAV is longer than the PAV, a V pace following an atrial sense will always occur after the full SAV, even when the sensor-indicated rate or Lower Rate interval expires first.
In certain patients, short AV intervals may be used as a prophylaxis for AV nodal or accessory pathway reentrant tachycardias in dual chamber modes.
Reference Manual 51
Page 52
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Long PAV intervals (greater than or equal to 250 ms) should be used with caution. If intrinsic ventricular events occur and are not sensed, a long PAV may result in pacing into the ventricle’s relative refractory period, including the T wave, or loss of AV synchrony, which may precipitate retrograde activation of the atria with corresponding hemodynamic consequences and symptoms. Long PAV intervals may also result from some Search AV+ settings (see Section 4.4, “Search AV+ and diagnostic”, page 55 and Section 4.5.3, “Ventricular blanking”, page 59.

4.3 Rate Adaptive AV

4.3.1 Overview

In the normal heart, AV conduction times tend to shorten as the heart rate increases and to lengthen as the heart rate decreases. The Rate Adaptive AV (RAAV) feature, available when the pacemaker is operating in the DDDR, DDD, DDIR, DVIR, DOOR, and VDD modes, mimics this physiologic response. When RAAV is programmed On, the pacemaker shortens AV intervals for atrial rates within a programmed rate range. This feature provides increased opportunity for atrial sensing, as follows:
Shortened SAV intervals increase the tracking range at fast atrial rates by shortening the Total Atrial Refractory Period (TARP) and increasing the 2:1 block rate. Refer to Section 4.6.9, “Total Atrial Refractory Period (TARP)”, page 64 and Section 4.7, “High rate atrial tracking”, page 68 for more information.
Shortened PAV intervals lengthen the atrial sensing window of the VA interval at higher sensor-driven rates.
Note: RAAV will not shorten PAV intervals to less than 30 ms or shorten SAV intervals to less than 10 ms.

4.3.2 Programming for Rate Adaptive AV

For RAAV operation, the SAV and PAV are programmed (as applicable) to the values desired for low rates. Three additional programmable parameters control how AV intervals are adjusted at higher rates:
Start Rate – RAAV operation of shortening SAV and PAV intervals begins at this rate.
Stop Rate – The shortest SAV and PAV occur at this rate and at all higher rates, up to the
upper rate limits.
Maximum Offset – The maximum amount of time (in ms) by which the SAV and PAV intervals can be shortened.
52 Reference Manual
Page 53
Programmed
Start rate
PAV
PAV
SAV
SAV
Rate (min-1)
AV Interval (ms)
Stop rate
MinimumRate Adaptive
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
The PAV minus the Maximum Offset gives the shortest PAV interval at the Stop Rate (e.g., 200 ms - 100 ms = 100 ms). Subtracting the Maximum Offset from the SAV gives the shortest SAV interval (e.g., 170 ms - 100 ms = 70 ms).
Figure 31 shows how the SAV and PAV intervals are linearly shortened as the rate increases from below the Start Rate to above the Stop Rate.

Figure 31. Rate Adaptive AV operation (DDDR Mode)

1 Shortest PAV (PAV minus Maximum Offset) 2 Shortest SAV (SAV minus Maximum Offset)
Parameters: Programmed SAV =
170 ms Programmed PAV =
200 ms
Start Rate = 80 bpm Maximum Offset =
−30 ms
Stop Rate = 150 bpm
Reference Manual 53
Page 54
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.3.3 RAAV operations

Shortening of the AV interval(s) occurs when the appropriate rate exceeds the programmed Start Rate, as follows:
SAV – The mean atrial rate determines SAV adjustments. Because of how the mean atrial rate is calculated:
SAV adjustments will lag during rapid increases or decreases in intrinsic atrial rates.
The SAV is not adjusted for isolated events (PACs).
AS-AP or AR-AP intervals may affect the SAV value since these intervals are not used in the mean atrial rate calculation.
PAV – The sensor-indicated rate determines PAV adjustments.
The approximate difference between programmed SAV and PAV is maintained as the SAV and PAV intervals are adjusted.

4.3.4 Programming considerations and restrictions

Search AV+ – RAAV can be enabled while Search AV+ is enabled. Search AV+ will operate
using the AV intervals determined by the RAAV rather than the programmed AV intervals.

4.3.5 RAAV and sick sinus syndrome

If RAAV is activated for a sick sinus syndrome patient who has AV conduction, consider the following:
The rate at which AV conduction is lost should not be too low (i.e., below 90 bpm).
Review of the AV Conduction Histogram diagnostic data may aid in appropriate programming of Start Rate and Stop Rate to maintain AV conduction as long as possible.
54 Reference Manual
Page 55
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.4 Search AV+ and diagnostic

4.4.1 Overview

The Search AV+ feature is intended to promote intrinsic ventricular activation in patients with intact or intermittent AV conduction and prevent inappropriate therapy in patients without conduction. Search AV+ is available when the pacemaker is programmed to the DDDR, DDD, DDIR, DDI, DVIR, DVI, or VDD mode. The pacemaker searches for the patient’s intrinsic AV conduction time and adjusts the SAV and PAV intervals either longer or shorter to promote intrinsic activation of the ventricles. When Rate Adaptive AV is active, the pacemaker also adjusts the SAV and PAV intervals relative to the rate adaptive values. If the pacemaker does not observe intrinsic ventricular activation during its periodic searches over the course of a week, it turns off the Search AV+ feature.

4.4.2 Programming to Search AV+

Programming Search AV+ to “On” requires setting the Max Increase to AV parameter. This parameter defines the maximum amount of time (in ms) by which the operating SAV and PAV intervals can be lengthened to allow ventricular sensing to occur. The operating SAV and PAV intervals will adapt to the observed conduction time, but will not exceed the Max Increase to AV parameter.

4.4.3 Search AV+ operation

The pacemaker attempts to keep intrinsic conducted events in an “AV delay window” that precedes scheduled paced events. The AV delay window is set to promote intrinsic conduction to the ventricles, but end early enough to avoid fusion or pseudo-fusion beats if pacing is necessary (see Figure 32).
Reference Manual 55
Page 56
200 ms
PAV is now 212 ms
PAV is now 204 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 32. Search AV+ operation

1 Previous 16 AV events, 8 or more VS events are within 15 ms of the scheduled VP, thus PAV and
SAV are extended by 62 ms to promote intrinsic conduction.
2 Previous 16 AV events, 8 or more VS events occur 55 ms before the scheduled VP, thus PAV and
SAV are shortened by 8 ms to limit long AV delays if ventricular pacing is needed.
Parameters: DDDR SAV = 120 ms
Lower Rate = 60 bpm PAV = 150 ms Sensor-Indicated Rate =
90 bpm
Max. Increase to AV= 170 ms
To determine when intrinsic conducted events occur, the pacemaker assesses the 16 most recent AV conduction sequences that start with a nonrefractory atrial sense (when the pacemaker is operating in the DDDR, DDD, and VDD modes) or an atrial pace (when the pacemaker is operating in the DDDR, DDD, DDIR, DDI, DVIR, and DVI modes) and end with a ventricular pace or a nonrefractory ventricular sense.
Search criteria of AV conduction times – The measured AV conduction times are classified as on time, too short or too long.
Too long means 8 or more of the last 16 ventricular sensed events occurred within 15 ms of the scheduled ventricular pace, or 8 or more of the last 16 ventricular events were paced events.
Too short means 8 or more of the last 16 ventricular sensed events occurred more than 55 ms before the scheduled ventricular pace.
Adjustment of SAV and PAV intervals – If AV conduction times are classified as too long, the pacemaker lengthens the operating SAV and PAV intervals by 62 ms for the next 16 pacing cycles to promote intrinsic conduction. The maximum that the SAV and PAV can be lengthened is limited by the Search AV+ Maximum Increase to AV parameter.
56 Reference Manual
Page 57
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
If the previous 16 AV intervals are classified as too short, the pacemaker shortens the operating SAV and PAV intervals by 8 ms for the next 16 pacing cycles. The maximum that SAV and PAV can be shortened is limited by the programmed SAV and PAV values or the RAAV Maximum Offset parameter, if RAAV is On.

4.4.4 Suspension of Search AV+ operation

Search AV+ promotes conduction in patients with intrinsic conduction and prevents inappropriate therapy for patients without intrinsic conduction. If AV conduction is not found, Search AV+ suspends operation for progressively longer periods: 15 minutes, 30 minutes, 1, 2, 4, 8, and 16 hours. If AV conduction is not found following 10 consecutive 16-hour suspensions (approximate duration, 1 week), the device automatically turns Search AV+ to Off.

4.4.5 Programming considerations and restrictions

Both Automatic PVARP and Rate Adaptive AV can shorten the AV intervals at higher rates and potentially lead to ventricular pacing.
Automatic PVARP – When automatic PVARP is active and Search AV+ is set to On, the pacemaker will ignore conduction times that are the result of automatic PVARP shortening of the SAV interval.
Rate Adaptive AV – RAAV can be enabled while Search AV+ is enabled. Search AV+ will operate using the RAAV-determined AV intervals rather than the programmed AV intervals.
MVP modes – Search AV+ is not pertinent and cannot be enabled if the pacemaker is programmed to an MVP mode (AAIR<=>DDDR or AAI<=>DDD).

4.4.6 Recording AV interval adaptations

AV interval diagnostics record data about AV operations for the Search AV+ feature.
4.4.6.1 Automatic Search AV+ Histogram
Programming Search AV+ parameters automatically initiates recording of data by the Search AV+ Histogram diagnostic. This histogram shows the percentage of A-VS, VS from Search, and A-VP intervals versus rate. A histogram can be displayed or printed from the Data icon.
Reference Manual 57
Page 58
P
P P
DDD
A
V
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
4.4.6.2 Clearing AV interval data
AV interval data is normally cleared by the pacemaker one hour after a programming session.
However, you can select the option to clear data immediately. Be sure to save the session data or print the episode report before ending the patient session.
4.4.6.3 Search AV+ and compromised ventricular function
Consider turning Search AV+ off if intrinsic ventricular activation is not desired.

4.5 Blanking periods

Blanking periods disable sensing for a programmable or nonprogrammable interval. Signals that are blanked may originate in either chamber or from outside sources such as noise from muscle movement.

Figure 33. Example of dual chamber blanking operation

1 Nonprogrammable Atrial Blanking 2 Programmable Post-Ventricular Atrial
Blanking
3 Programmable Ventricular Blanking 4 Nonprogrammable Ventricular Blanking
Note: Black bars indicate blanking periods.

4.5.1 Nonprogrammable blanking periods

Immediately following a sensed or paced event in either chamber, sensing for that chamber is blanked for a nonprogrammable period that may typically vary from 50 to 100 ms. The actual duration of the blanking period is determined dynamically by the pacemaker, based on the strength and duration of the signal. Dynamic blanking prevents sensing the same signal twice, while minimizing total blanking time.
58 Reference Manual
Page 59
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.5.2 Post-Ventricular Atrial Blanking

The programmable Post-Ventricular Atrial Blanking (PVAB) period, used when the pacemaker is operating in the DDDR, DDD, DDIR, DDI, VDD, VDIR, and VDI modes, prevents sensing of ventricular paced events or far-field R waves on the atrial lead. Any ventricular event (paced or sensed) starts the PVAB, which is also the first portion of the Post-Ventricular Atrial Refractory period (PVARP). The PVAB is limited to values equal to or less than the programmed PVARP, except in VDIR and VDI modes where PVARP does not apply.
Note: PVAB is programmed to a value less than or equal to PVARP.

4.5.3 Ventricular blanking

The programmable Ventricular Blanking period, which follows an atrial pacing stimulus when the pacemaker is operating in the DDDR, DDD, DDIR, DDI, DVIR, and DVI modes, prevents ventricular inhibition or ventricular safety pacing due to sensing of the atrial stimulus on the ventricular lead (crosstalk). The Ventricular Blanking period also applies to the ADIR and ADI modes to prevent sensing of the atrial stimulation.
Long blanking periods (36 ms or greater) increase the possibility of unsensed ventricular events.
Long blanking periods used in conjunction with long PAV intervals (250 ms or greater) may result in pacing into the T wave when intrinsic ventricular events are blanked and not sensed. PAV values (200 ms or less) should reduce the possibility of T wave pacing.
Long PAV intervals may also result from some Search AV+ operation (See Section 4.4, “Search AV+ and diagnostic”, page 55). To minimize the possibility of undersensing intrinsic events, Search AV+ reduces ventricular blanking to 20 ms unless Ventricular Safety Pacing is observed.

4.5.4 Single chamber atrial blanking

The programmable single chamber atrial blanking period, used when the pacemaker is operating in the AAIR, ADIR, AAI, ADI, and AAT mode, prevents sensing of far-field R waves. It is started by a paced, sensed, or refractory sensed atrial event.
Note: Atrial Blanking must be programmed at least 50 ms less than the Atrial Refractory Period.
Reference Manual 59
Page 60
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.6 Refractory periods

4.6.1 Overview

A refractory period is an interval during which an intrinsic event sensed on a particular lead channel cannot start certain timing intervals. Each refractory period begins with a blanking period, during which no sensing occurs. During the unblanked portion of a refractory period, sensing occurs, but sensed events may not directly affect timing operations. Refractory periods are intended to prevent certain timing intervals from being started by inappropriate signals such as retrograde P waves, far-field R waves, or electrical noise.
Though they may not start timing intervals, refractory sensed events are monitored by the pacemaker, and they affect the operation of PVC Response, Mode Switch, Rate Adaptive AV operation, automatic PVARP, Non-Competitive Atrial Pacing, and other features for which the periodicity or number of sensed events are pertinent. Refractory sensed events are included on Marker Channel recordings.

4.6.2 Post-Ventricular Atrial Refractory Period

The Post-Ventricular Atrial Refractory Period (PVARP) follows a paced, sensed, or refractory sensed ventricular event when the pacemaker is operating in the DDDR, DDD, DDIR, DDI, and VDD modes. It is intended primarily to prevent the sensing of retrograde P waves that might promote Pacemaker-Mediated Tachycardias (PMTs) in atrial tracking modes. When the pacemaker is operating in the DDIR and DDI modes, PVARP prevents atrial inhibition from retrograde P waves.
The first portion of the PVARP is the programmable Post-Ventricular Atrial Blanking period (PVAB). During the remainder of the PVARP, intrinsic atrial events may be sensed as refractory sensed events (AR) and identified on Marker Channel recordings, but they do not affect stimulus timing.
When the pacemaker is operating in the DDDR, DDD, and VDD modes, an SAV is not started.
When the pacemaker is operating in the DDDR, DDD, DDIR, and DDI modes, the scheduled atrial pace is not inhibited.
60 Reference Manual
Page 61
P
DDD
A
V
P P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 34. Example of PVARP operation

1 PVARP
The duration of the PVARP may be selected as follows:
The PVARP should be programmed to a value greater than the patient’s ventriculoatrial (VA) retrograde time when retrograde conduction is present.
Excessively long PVARPs may induce 2:1 block at high intrinsic rates when the pacemaker is operating in an atrial tracking mode (DDDR, DDD, or VDD).
To reduce the 2:1 block point, PVARP can be set to vary based on the sensor-indicated rate (sensor-varied PVARP) or the mean atrial rate (automatic PVARP).

4.6.3 Sensor-varied PVARP

When sensor-varied PVARP is programmed, the pacemaker determines a value for the PVARP based on the sensor-indicated rate. The intended purpose of the sensor-varied PVARP depends upon the mode:
When the pacemaker is operating in the DDDR, DDD, and VDD modes, sensor-varied PVARP is intended to do the following:
– Enhance protection against PMT at lower rates by providing longer PVARPs at low
sensor-indicated rates.
– Allow tracking of higher atrial rates (that is, provide a higher 2:1 block rate) by
shortening the PVARP at high sensor-indicated rates.
When the pacemaker is operating in the DDIR mode, the sensor-varied PVARP is intended to promote AV synchrony by preventing inhibition of atrial pacing by an atrial sense early in the VA interval. It also reduces the likelihood of competitive atrial pacing at high sensor-indicated rates.
Reference Manual 61
Page 62
400 ms 300 ms 400 ms
200 ms
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 35. Sensor-varied PVARP operation (DDDR Mode)

1 Upper Sensor Rate 2 Lower Rate
3 PAV 4 PVARP

4.6.4 Determining sensor-varied PVARP

The pacemaker determines the duration of the sensor-varied PVARP as follows:
When the pacemaker is operating in the DDDR, DDD, and VDD modes, the sensor-varied PVARP is limited to 400 ms at low rates and the programmed PVAB at high rates (as shown in Figure 35).
When the pacemaker is operating in the DDIR mode, PVARP is the pacing interval, minus the PAV interval, minus 300 ms. At high rates, PVARP is limited by the programmed PVAB.
When the pacemaker is operating in the DDDR, DDD, DDIR, and VDD modes, the sensor-varied PVARP is automatically adjusted to maintain a 300 ms sensing window (as shown in Figure 35).
62 Reference Manual
Page 63
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.6.5 Automatic PVARP

When automatic PVARP is programmed, the pacemaker determines a value for the PVARP based on the mean atrial rate (which is an average of all A-A intervals except those starting with an atrial sense or atrial refractory sense and ending with an atrial pace). When the pacemaker is operating in the DDDR, DDD, and VDD modes, automatic PVARP is intended to provide a higher 2:1 block rate by shortening the PVARP and SAV (if necessary) at higher tracking rates and protect against PMTs at lower rates by providing a longer PVARP.

4.6.6 Determining automatic PVARP

The pacemaker determines the duration of the automatic PVARP as follows:
After every four pacing cycles, a 2:1 block rate is calculated that is 30 bpm above the current mean atrial rate.
PVARP is then adjusted so the Total Atrial Refractory Period equals the calculated 2:1 block rate. The programmable Minimum PVARP parameter controls the minimum value that PVARP can be shortened to.
If the minimum PVARP value is reached and the 2:1 block rate is still too low, the SAV interval can be shortened to increase the 2:1 block rate. The minimum SAV that can be set is the rate adaptive SAV value (i.e., the programmed SAV value minus the RAAV Maximum Offset value).
The minimum adjustable 2:1 block rate is 100 bpm. The maximum adjustable 2:1 block rate is the Upper Tracking Rate plus 35 bpm. If Mode Switch is On, the maximum 2:1 block rate can be the Detect Rate if this rate is less than the Upper Tracking Rate calculation.

4.6.7 Programming restrictions for automatic PVARP

Rate Drop Response – Automatic PVARP is not available when Rate Drop Response is
programmed On.

4.6.8 Spontaneous PVARP extension

The programmed PVARP duration, the sensor-varied PVARP, and the automatic PVARP may be overridden by the PVC Response and PMT Intervention features, as follows:
When the PVC Response feature is programmed On and a pacemaker-defined PVC occurs, the PVARP is forced to 400 ms for one cycle if a lesser value is in effect.
When PMT Intervention is programmed On and a pacemaker-defined PMT is detected, the PVARP is forced to 400 ms for one cycle after the ninth paced ventricular event of the PMT.
Reference Manual 63
Page 64
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Refer to Section 6.5, “PMT intervention”, page 125 and Section 6.6, “PVC Response”, page 127 for further details on the PMT Intervention and PVC Response features and their interactions with PVARP.

4.6.9 Total Atrial Refractory Period (TARP)

In dual chamber modes that sense in the atrium, the Total Atrial Refractory Period (TARP) is the sum of two intervals, as follows:
AV Interval – The AV interval begins with an atrial event and ends with a ventricular event. The first portion is a nonprogrammable blanking period. Its complete duration is determined as follows:
When the pacemaker is operating in the DDDR, DDD, and VDD modes, the PAV or SAV interval is the AV interval.
When the pacemaker is operating in the DDIR and DDI modes, the AV interval starts with the first atrial sensed event in the VA interval or with an atrial pacing stimulus; it ends when the PAV expires, even when ventricular pacing is inhibited.
Post-Ventricular Atrial Refractory Period (PVARP) – The PVARP is described on Section 4.6.2.

Figure 36. Total Atrial Refractory Period

1 TARP 2 SAV + PVARP
3 SAV 4 PVARP
During atrial tracking, TARP = SAV + PVARP, and its duration determines the rate at which 2:1 block occurs. Refer to Section 4.7, “High rate atrial tracking”, page 68 for more information.
64 Reference Manual
Page 65
P
DDD
A
V
P P
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.6.10 Ventricular Refractory Period

The programmable Ventricular Refractory Period (VRP) follows paced, sensed, and refractory sensed ventricular events (including PVCs) in all dual chamber and ventricular modes that sense in the ventricle. The VRP is intended to prevent sensing of the T wave or a PVC. The first portion of the VRP is a nonprogrammable blanking period. A ventricular refractory sensed event affects pacemaker timing as follows:
Ventricular blanking and refractory periods restart in all modes.
When the pacemaker is operating in the DDDR, DDD, and VDD modes, the upper tracking rate interval, PVARP, and PVAB also restart.
When the pacemaker is operating in the VVIR and VDIR modes, the upper sensor rate interval restarts.
Note: In dual chamber modes, the VRP should be programmed shorter than the PVARP.

Figure 37. Example of Ventricular Refractory Period operation

1 VRP
In dual chamber modes, a ventricular refractory sensed event does not affect a scheduled sensor-driven or lower rate atrial output. Thus, a sensor-driven atrial output pulse will initiate a PAV with a ventricular output pulse following, unless inhibited.

4.6.11 Atrial Refractory Period (single chamber)

The programmable Atrial Refractory Period (ARP) follows paced, sensed, and refractory sensed atrial events. The ARP is used in the AAIR, ADIR, AAI, ADI, and AAT modes. It is intended to prevent inhibition due to far-field R wave sensing. The first portion of the ARP is a programmable blanking period. The ARP should be programmed to a value long enough (180 ms or greater) to prevent far-field R wave sensing but short enough to ensure atrial sensing up to the programmed Upper Sensor Rate.
Reference Manual 65
Page 66
DDDR
A
V
P P P
P R R R P P
200 ms
Sensor Sensor Sensor
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
If the pacemaker is programmed to an MVP mode (AAIR<=>DDDR or AAI<=>DDD) and is operating in AAIR or AAI mode, the ARP is automatically adjusted to 75% of the cardiac cycle length, up to a maximum of 600 ms.

4.6.12 Noise reversion

When sensing occurs during the Atrial Refractory Period (ARP) or Ventricular Refractory Period (VRP), the refractory period (and its blanking period) are restarted. The operation associated with continuous refractory sensing in the ARP or VRP is called noise reversion. Multiple restarts of the ARP or VRP (continuous noise reversion) do not inhibit scheduled pacing. Pacemaker behavior during continuous noise reversion is as follows:
Pacing occurs at the sensor-indicated rate for all rate responsive modes (except VVIR and VDIR).
Pacing occurs at the programmed Lower Rate for all non-rate responsive modes (including VVIR and VDIR).
On the ECG, noise reversion may be difficult to distinguish from loss of sensing, but Marker Channel recordings will show refractory sense markers when noise reversion occurs.

Figure 38. Example of noise reversion in DDDR at sensor-indicated rate.

1 Sensor-indicated interval
Parameters: Lower Rate = 60 bpm (1000 ms
± 17 ms)
Upper Sensor Rate = 120 bpm (500 ms ± 17 ms)
PAV Interval = 200 ms Ventricular Refractory Period =
240 ms
PVARP = 300 ms PVAB = 200 ms
66 Reference Manual
Page 67
200 ms
Sensor P
V
R R R R P
VVIR
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Note: In DDDR mode, the sensor-indicated rate occurs anywhere between the Lower Rate and the Upper Sensor Rate, depending on patient activity.
Note: If an atrial refractory sensed event occurs, the pacemaker does not restart the refractory period. However, an atrial refractory sensed event will start a short blanking period of 50 to 100 ms depending on the signal strength and duration of the atrial event.

Figure 39. Example of noise reversion in VVIR at lower rate.

1 Lower Rate
Parameters: Lower Rate = 60 bpm (1000 ms
±17 ms)
Ventricular Refractory Period = 240 ms
Upper Sensor Rate = 120 bpm (500 ms ±17 ms)

4.6.13 Preventing noise sensing

Noise reversion may be caused by electromagnetic interference (EMI), myopotentials, excessively high output settings, or low sensitivity settings. When it has been identified, noise reversion usually can be reduced or eliminated by one of the following actions:
Remove patient from EMI environment.
Reprogram sensitivity to a less sensitive setting (higher numerical value) or program Sensing Assurance to On, to monitor and if necessary, adjust the sensitivity value. Refer to Section 5.6, “Sensing Assurance and diagnostic”, page 99.
Reprogram sensing polarity to bipolar polarity (if available).
Reduce the amplitude and/or pulse width in the same or opposite chamber.
Reference Manual 67
Page 68
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Program Capture Management to Adaptive to monitor capture thresholds, and, if necessary, adjust amplitude and pulse width values. Refer to Section 5.5, “Capture Management and diagnostic”, page 79.

4.7 High rate atrial tracking

4.7.1 Overview

When the pacemaker is operating in the DDDR, DDD, and VDD modes, the fastest atrial rate the pacemaker can track is determined by the Total Atrial Refractory Period (TARP), which is the sum of the SAV and the PVARP. Pacemaker behavior at high atrial rates in these modes is determined by the relationship between the TARP and the interval corresponding to the Upper Tracking Rate. In the DDDR mode, the interval corresponding to the Upper Sensor Rate also must be considered.

4.7.2 2:1 block

When the intrinsic atrial interval is shorter than the TARP, some atrial events will fall in the PVARP and not be tracked. At the rate where this first occurs, ventricular tracking occurs only on alternate beats, and 2:1 block ensues. When the pacemaker is operating in the DDD and VDD modes, the ventricular pacing rate drops precipitously.
When sensor-varied PVARP or automatic PVARP is selected, the 2:1 block rate may occur at a higher rate during activity due to shortening of the PVARP and the SAV (automatic PVARP only), thus increasing atrial tracking.
When Rate Adaptive AV operation is selected, the SAV shortens at high atrial rates, shortening the TARP and raising the 2:1 block rate.
When the 2:1 block rate is less than the Upper Tracking Rate, the Upper Tracking Rate cannot be achieved.
When the pacemaker is operating in DDDR mode, pacing at the sensor-indicated rate may prevent a precipitous rate drop at the 2:1 block point when activity is present.
For patients with a documented propensity for prolonged or sustained atrial fibrillation or flutter, the clinician can select Upper Tracking Rate, SAV, and PVARP values that induce 2:1 block at a desired rate (2:1 block rate = 60,000/TARP). Alternatives for controlling rates in these patients include use of the Mode Switch feature and DDIR mode pacing.
When the pacemaker is operating in the DDDR mode, atrial competition may occur if the Upper Sensor Rate exceeds the 2:1 block rate.
68 Reference Manual
Page 69
S
P
DDDR
A
V
S
P
S
P
S
P P
R
200 ms
P
Sensor
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

4.7.3 Pacemaker Wenckebach

When the 2:1 block rate exceeds the programmed Upper Tracking Rate, pacemaker Wenckebach may occur. When the intrinsic rate exceeds the Upper Tracking Rate, a pacing stimulus at the expiration of the SAV would violate the upper tracking rate. The pacemaker therefore extends the SAV until the upper tracking rate interval expires. Subsequent SAVs require greater extension, until an atrial event falls in the PVARP and is not tracked.
When the pacemaker is operating in the DDDR, DDD, and VDD modes, the result is normally a fixed ratio between atrial and ventricular rates (3:2, 4:3, and so forth).
When the pacemaker is operating in the DDDR mode, the pacemaker Wenckebach rate may be smoothed by sensor-driven ventricular pacing, thereby overriding the fixed ratio.
Figure 40 shows how pacemaker Wenckebach operation occurs in the DDDR, DDD, or VDD modes.

Figure 40. Example of pacemaker Wenckebach operation

1 Upper Tracking Rate interval 2 SAV interval
Parameters: Sensor-Indicated Rate =
PVARP = 300 ms
90 bpm (667 ms) PAV Interval = 230 ms Upper Tracking Rate = 100 bpm
(600 ms)
SAV Interval = 200 ms

4.7.4 High rate operation in the DDDR mode

Table 3 summarizes how the total atrial refractory period (TARP), the Upper Tracking Rate (UTR) interval, and the Upper Sensor Rate (USR) interval may interact at high atrial rates when the pacemaker is operating in the DDDR mode.
Reference Manual 69
Page 70
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Table 3. Upper rates interaction with TARP

Relationship Between TARP and Upper Rate Intervals
TARP > both USR and UTR intervals
USR interval > TARP > UTR interval
USR interval > UTR interval > TARP
UTR interval > both USR interval and TARP
a
Unless the Non-Competitive Atrial Pacing is On, see Section 6.4, “Non-competitive atrial pacing”, page 123.
Wenckebach Before
2:1 Block
no no yes
no no no
yes yes no
yes yes yes
Achieve Upper Track-
ing Rate
Potential Atrial Com-
petition
a
a
70 Reference Manual
Page 71
Implant 30 minutes
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

5 Lead/cardiac tissue interface

5.1 Implant Detection

5.1.1 Overview

Implant Detection is a 30 min period, beginning at lead connection, during which the pacemaker verifies that each lead has been connected by measuring lead impedance. After 30 min of continuous lead connection, the pacemaker completes Implant Detection and activates the following features (see Figure 41):
Operating polarity (automatic configuration occurs during Implant Detection)
MVP operations including conduction checks and mode changes
Adaptive sensitivity settings (Sensing Assurance)
Rate responsive pacing, including adaptive rate profile optimization (Rate Profile Optimization)
Adaptive ventricular output settings for threshold management (Capture Management)
Diagnostic data collection

Figure 41. Implant Detection period

1 Lead connection verified 2 Leads connected 3 Lead polarities confirmed, Rate Response enabled, Adaptive features activated, and Data
collection activated
Implant Detection is available in all pacing modes and is turned on at shipment.
Note: Search AV+ initializes 60 min after Implant Detection is complete.
Reference Manual 71
Page 72
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Note: Automatic polarity configuration does not occur during Implant Detection if the Lead Monitor parameter is programmed to Monitor Only before implant. Automatic configuration will take place if Lead Monitor is programmed to Adaptive before implant or remains set to its shipping setting of Configure.

5.1.2 Verifying lead connection during Implant Detection

At the time of lead connection, the pacemaker begins verifying that each lead is present by measuring the impedance of each pacing pulse. When a pace is delivered, the pacemaker determines if the impedance is within the acceptable range, which is programmable between 200 to 40006 Ω for both bipolar and unipolar configurations.
High impedance paces cause Implant Detection to reset.
Low impedance paces (and continuous sensing) are considered acceptable for the purpose of determining lead connection and configuring polarity.

5.2 Automatic polarity configuration

5.2.1 Overview

During Implant Detection, bipolar pacemakers automatically configure pacing and sensing polarities through the Lead Monitor feature. Bipolar pacemakers are shipped with the Atrial and Ventricular Lead Monitor set to Configure, enabling automatic polarity determination shortly after lead connection.
Unipolar-only pacemakers are configured to unipolar pacing and sensing polarity at the time of manufacture and remain unipolar during the operational life of the pacemaker.
6
The acceptable maximum impedance limit for a valid lead (bipolar or unipolar) can be changed by programming the Notify If > (Greater Than) parameter, which is part of the Lead Monitor and is found under the programmed lead polarities. The minimum impedance limit of 200 Ω is nonprogrammable.
72 Reference Manual
Page 73
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

5.2.2 Measuring lead impedance during configuration

Bipolar pacemakers, using either bipolar or unipolar leads, automatically configure pacing and sensing polarities by measuring the impedance of each pace during the configuration period. (Lead Monitor must be set to Configure or Adaptive. See Section 5.3, “Lead Monitor”, page 76.) Impedance measurement during configuration is as follows:
The pacemaker issues a bipolar pace and immediately checks the pace for high impedance.
– If the pace is within range, it is considered an acceptable bipolar pace.
– If high impedance is found, the pacemaker immediately follows the bipolar pace with
a backup unipolar pace.
If a unipolar backup pace is issued, the pacemaker checks it for high impedance.
– If the pace is within range, it is considered an acceptable unipolar pace.
– If high impedance is found, the pacemaker assumes a lead is not attached and
restarts Implant Detection.
Notes:
During polarity configuration, the pacemaker also detects low impedance paces but does not follow them with unipolar backup paces. To avoid possible loss-of-capture due to continuous low impedance pacing, the pacemaker sets unipolar polarity when 3 of 16 paces are detected as low impedance during the first phase of configuration (see “Initial Configuration Phase”, page 74).
If the pacemaker assumes a lead is not present (a high impedance unipolar pace is detected) in one chamber of a dual chamber pacemaker, Implant Detection will restart.

5.2.3 How polarities are automatically configured

Atrial and ventricular lead polarities are configured independently in dual chamber bipolar models, with the exception of pacemakers in the VDD Series, which have fixed bipolar atrial sensing only.
Operating pacing and sensing polarities for bipolar pacemakers are configured automatically in two phases during Implant Detection. (See Figure 42.) During these phases the pacemaker continues to measure impedance as described above.
Reference Manual 73
Page 74
Implant
Leads connected
Implant detection
Initial configuration
phase
Confirmation phase
Lead polarities
configured
Lead polarities
confirmed
30 minutes5 minutes
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 42. Automatic configuration of bipolar models

Initial Configuration Phase – The pacemaker sets lead polarities five minutes after lead
connection unless a high impedance unipolar pace has reset Implant Detection, a prevalence of low impedance bipolar paces has set unipolar pacing, or if continuous sensing has occurred.
The pacemaker sets initial polarities as follows:
The pacemaker delivers three asynchronous paces at magnet rate.
If at least two of the three paces are determined to be bipolar, the pacemaker remains set to bipolar polarity (with backup unipolar paces) for the Confirmation Phase.
If at least two of the three paces are determined to be unipolar, the pacemaker sets polarity to unipolar. Unipolar polarity becomes the operating polarity for pacing and sensing, and no Confirmation Phase is required.
Note: If one of the asynchronous paces in a given chamber is a high impedance unipolar pace, the pacemaker restarts Implant Detection in both chambers. Polarity configuration restarts only in the affected chamber, however.
Confirmation Phase – Twenty-five minutes after initial configuration, the pacemaker confirms final operating polarity for leads configured bipolar in the Initial Configuration Phase. During this 25-minute Confirmation Phase, the pacemaker measures the lead for high and low impedance paces. If 8 of 16 paces (or the programmed number of paces out of sixteen) are out of range, the lead will automatically be reconfigured to unipolar polarity.
At the end of twenty-five minutes, operating polarity for leads detected as bipolar during the Initial Configuration Phase is determined as follows:
74 Reference Manual
The pacemaker delivers three asynchronous paces at magnet rate.
Page 75
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
If at least two of the three paces are determined to be bipolar, the pacemaker sets operating polarity to bipolar (or Adaptive operation if Lead Monitor was programmed to Adaptive prior to implant). See Section 5.3, “Lead Monitor”, page 76.
If at least two of the three paces are determined to be unipolar, the pacemaker sets operating pace and sense polarities to unipolar.
Note: If a high impedance unipolar pace occurs in a given chamber at any time during the 25-minute Confirmation Phase, the pacemaker restarts Implant Detection in both chambers. Polarity configuration restarts only in the affected chamber, however.
Leads configured unipolar during the Initial Configuration Phase continue to operate in the unipolar configuration. If bipolar polarity switches to unipolar during the Confirmation Phase, the operating polarity remains unipolar with no additional confirmation through asynchronous pacing.
Warning: If, at implant, the setscrews, both tip and ring, for a 3.2 mm connector pacemaker are not properly engaged and all electrical contacts are not sealed, leakage between the tip and ring contacts may occur. Such leakage may cause the pacemaker to falsely identify a unipolar lead as bipolar, resulting in a loss of output. The same result could occur for all bipolar models if the electrical contacts were not properly sealed when using lead extenders or adaptors.

5.2.4 When automatic configuration is complete

Thirty minutes after lead connection, Implant Detection and automatic configuration are complete. If, after this time period, the leads are detached and lead polarity type is changed, automatic configuration and Implant Detection do not restart automatically at reinsertion. The clinician must reprogram Implant Detection to On/Restart, which automatically resets Lead Monitor to Configure.
The Lead Monitor feature (see Section 5.3, “Lead Monitor”, page 76) monitors and reports on lead stability when Implant Detection and automatic configuration are complete. It is automatically set by the pacemaker as follows:
Bipolar models are shipped with Lead Monitor set to Configure. However, after Implant Detection ends, Lead Monitor is automatically set to Monitor Only for bipolar and unipolar configurations.
Bipolar models with Lead Monitor programmed to Adaptive before implant are set to Adaptive if polarity was determined to be bipolar. If polarity for these models was determined to be unipolar, they are set to Monitor Only.
Unipolar models are shipped with Lead Monitor set to Monitor Only and continue to operate at that setting both during and after Implant Detection.
Reference Manual 75
Page 76
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

5.2.5 Manually setting polarities

To manually program atrial or ventricular lead polarities at implant, the clinician first must “turn off” the Configure setting under Lead Monitor by choosing the Monitor Only setting instead. Implant Detection still provides the 30-minute detection period, followed by automatic feature and diagnostics activation, when the pacemaker is programmed manually.

5.2.6 Programming interactions

Programming Implant Detection to Off /Complete before completion of the 30-minute automatic polarity configuration period requires the clinician to manually program Lead Monitor and polarities. The pacemaker’s automatic features are activated when Implant Detection is turned off.
Manually programming Implant Detection to On/Restart restarts lead detection, polarity configuration (Lead Monitor is set to Configure), and automatic feature activation.
Initiating a programming session at any time during Implant Detection causes Implant Detection to be restarted.
If a dual chamber pacemaker is programmed to a single chamber mode, only one lead is configured. If the mode is reprogrammed to a dual chamber mode, the clinician will need to change the Lead Monitor from Configure to Monitor Only or Adaptive for the unconfigured lead, and then program pace and sense polarity for it manually.

5.3 Lead Monitor

5.3.1 Overview

The Lead Monitor feature measures lead impedances during the life of the pacemaker. When programmed to do so, Lead Monitor also enables the pacemaker to switch bipolar pacing and sensing to unipolar when bipolar lead integrity is in doubt. It also controls automatic configuration of lead polarities at implant. Lead Monitor is available in all pacing modes.
Caution: If the Lead Monitor detects out-of-range lead impedance, investigate lead integrity more thoroughly.

5.3.2 How lead monitoring works

The Lead Monitor feature monitors lead impedance of paced chambers, as defined by the mode, by measuring the impedance of each pacing pulse to see if it falls within the programmed impedance range for a stable lead.
76 Reference Manual
Page 77
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
The three programmable values under Atrial or Ventricular Lead Monitor are as follows:
Configure – provides automatic configuration of polarity (see Section 5.2.3, “How polarities are automatically configured”, page 73).
Adaptive
– monitors bipolar paces for high impedance and provides unipolar backup paces
when high impedance is detected.
– switches pacing and sensing polarity from bipolar to unipolar when the pacemaker
detects a prevalence of high or low impedance paces (see Monitor Sensitivity parameter in Table 4). The Lead Monitor setting changes to Monitor Only when polarity switches.
– provides automatic polarity configuration when selected prior to implant.
Monitor Only – monitors either unipolar or bipolar paces to determine if they are out of range but does not switch polarity when an out-of-range lead is indicated.
Lead Monitor is activated at lead connection, and it is automatically set to its operating value of Monitor Only or Adaptive at the end of Implant Detection. See Section 5.2.4, “When automatic configuration is complete”, page 75.
When Lead Monitor is set to Adaptive or Monitor Only and a lead is determined to be out of range, the pacemaker issues a lead warning that appears on the programmer screen at the next interrogation.
Lead Monitor programmable parameters – The programmable parameters for the Lead Monitor feature are shown below.

Table 4. Programmable parameters for Lead Monitor

General Parameters Meaning
Atrial Lead Monitor Monitors lead impedance in the atrium; option to provide unipolar
backup paces and to switch from bipolar to unipolar polarity for an out-of-range lead; provides automatic configuration of polarity at implant.
Ventricular Lead Monitor Monitors lead impedance in the ventricle; option to provide unipolar
backup paces and to switch from bipolar to unipolar polarity for an out-of-range lead; provides automatic configuration of polarity at implant.
Notify If < (Less Than) Nonprogrammable minimum boundary for acceptable atrial and ven-
tricular bipolar lead impedance. Fixed at 200 Ω.
Notify If > (Greater Than) Maximum boundary for acceptable atrial and ventricular bipolar lead
impedance.
Monitor Sensitivity Number of high or low impedance paces out of 16 that define an
out-of-range lead on each channel.
Reference Manual 77
Page 78
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Lead Monitor should not be programmed to Adaptive for patients with implantable defibrillators. When a prevalence of out-of-range lead impedance paces is detected, the monitor automatically reprograms the selected lead(s) to unipolar polarity. Pacing in the unipolar configuration may cause the defibrillator either to provoke inappropriate therapy or to withhold appropriate therapy.

5.4 Lead impedance data

Lead impedance data is recorded automatically.

5.4.1 Automatic Lead Impedance (Chronic Lead Trend)

The automatic Lead Impedance diagnostic data is based on measurements taken every three hours for each chamber that is being paced. The maximum, average, and minimum lead impedances are recorded every seven days for the most recent 14 months.
The following data is continuously updated:
Initial impedance (recorded at implant or when Chronic Lead Trend is cleared)
Lifetime minimum impedance (recorded since implant)
Lifetime maximum impedance (recorded since implant)
High impedance paces
Low impedance paces
If the maximum number of high impedance paces or low impedance paces is reached, that value will remain until the data is cleared. All other diagnostic data collection will continue.
Note: To avoid high output pacing, Chronic Lead Trend data collection can be programmed to Off. The outputs are increased to make lead impedance measurements every 3 hours.

5.4.2 Clearing Lead Impedance data

Automatic (Chronic) Lead Impedance Trend data is retained by the pacemaker unless you use the Clear Data function on the programmer. Note that the data should be cleared only when a lead is replaced. Be sure to save the session data or print the trend report before ending the patient session.

5.4.3 For further information

Refer to the Adapta/Versa/Sensia/Relia Pacemaker Programming Guide for information about collecting and displaying lead impedance trends.
78 Reference Manual
Page 79
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

5.5 Capture Management and diagnostic

5.5.1 Overview

When Capture Management is enabled, the pacemaker automatically monitors pacing thresholds at periodic intervals. Once the threshold is determined, the pacemaker determines a target output based on the programmable safety margin and programmable minimum amplitude.
If programmed to Adaptive, the pacemaker reprograms outputs toward the target.
If programmed to Monitor Only, the pacemaker does not reprogram outputs.
Caution: Epicardial leads have not been determined appropriate for use with the Ventricular Capture Management feature. Program Ventricular Capture Management to Off if implanting an epicardial lead.
Note: The pacemaker enables Capture Management once Implant Detection is completed.
Note: In the event of partial or complete lead dislodgment, Capture Management may not
prevent loss-of-capture.
5.5.1.1 Ventricular Capture Management (VCM)
At programmable intervals, the pacemaker performs a ventricular pacing threshold search to determine the ventricular threshold, which is the combination of minimum amplitude and minimum pulse width that consistently results in capture of the ventricular myocardium.
5.5.1.2 Atrial Capture Management (ACM)
At programmable intervals, the pacemaker performs an atrial pacing threshold search to determine the atrial amplitude threshold, which is the minimum amplitude that consistently results in capture of the atrial myocardium.
5.5.1.3 Initiating the pacing threshold search
Scheduling the search – A pacing threshold search is initiated according to the schedule
programmed by the clinician. The Capture Test Frequency parameter under Capture Management allows the clinician to schedule the search at fixed time intervals. The clinician can also program the Day At Rest value that allows the pacemaker to determine when to run the search.
Reference Manual 79
Page 80
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
If a pacing threshold search is scheduled to occur once per Day At Rest, the pacemaker tries to initiate the first pacing threshold search 12 hours after Implant Detection is complete. It then runs subsequent searches at 24-hour intervals from the time of the last successfully completed search.
If a search cannot be completed, the pacemaker retries after 30 minutes.
5.5.1.4 Programmable Capture Management parameters
Capture Management can be programmed to Adaptive, Monitor Only, or Off. When Capture Management is programmed to Adaptive, the parameters in Table 5 are used to control Capture Management operation. They can also be programmed for diagnostic use when Capture Management is programmed to Monitor Only.

Table 5. Programmable parameters for Capture Management

General parameters Meaning
Amplitude Margin The safety margin applied to the pacing threshold search results for
Amplitude. Minimum Adapted Amplitude The lower limit to which the operating Amplitude can be adapted. Capture Test Frequency Determines how often the pacing threshold search will be initiated. V. Acute Phase Days Remain-
ing
A. Acute Phase Days Remaining
V. Sensing During Search The polarity used for ventricular sensing during ventricular pacing
Time in days during which output settings can be adapted both
upward and downward, but not below the permanently programmed
ventricular outputs.
Time in days during which output settings can be adapted both
upward and downward, but not below the permanently programmed
atrial outputs.
threshold searches. See page 86.

5.5.2 Ventricular Capture Management (VCM)

Checking for stable rhythm – Before a pacing threshold search can be initiated, the
pacemaker determines if the patient is pacing or sensing at a low rate. A low rate is desirable during the pacing threshold search to reduce the risk of competition from forced pacing with
80 Reference Manual
Page 81
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
fast intrinsic rhythms. To make the determination, the pacemaker looks for intrinsic or rate-related events indicating:
Out of 8 measured V-V intervals no more than 2 are faster than:
– 100 bpm if the upper sensor rate and upper tracking rate are ≥135 bpm
– 95 bpm if the upper sensor rate and upper tracking rate are ≥125 bpm
– 90 bpm if the upper sensor rate and upper tracking rate are <125 bpm
the sensor rate, checked at the end of the 8 intervals, is at or below the ADL rate.
in dual chamber modes, at least 1 valid AV interval (AS-VS, AS-VP, AP-VS, AP-VP) occurred during the 8 measured intervals.
The pacemaker also looks for automatic feature interaction indicating:
Rate Drop Response is not in an intervention state.
Mode Switch is not changing between a tracking and a nontracking mode.
Battery measurements
Atrial Preference Pacing
Sleep function
If the stable rhythm check is successful, the pacing threshold search is initiated. If any of the criteria for the stable rhythm check is not met, the pacemaker defers the pacing threshold search until the next scheduled search period. See “Scheduling the search”, page 79.

5.5.3 The ventricular pacing threshold search

The pacemaker performs the pacing threshold search at a given Amplitude and Pulse Width setting through a series of support cycles and test paces. Each series has three sets of support cycles, with each set followed by a test pace and an automatic backup pace (see Figure 43).
The support cycles are pacing cycles at the programmed Amplitude and Pulse Width that may or may not include ventricular paced events. The pacing threshold search begins with the support cycles.
A test pace follows each set of support cycles and is delivered at a test Amplitude or Pulse Width. Amplitude and Pulse Width settings above the threshold cause capture of the myocardium; settings below the threshold result in loss-of-capture.
A backup pace automatically follows each test pace regardless of capture or loss-of-capture for that pace. It is delivered 110 ms after the test pace at the programmed Amplitude and a 1.0 ms Pulse Width setting.
Reference Manual 81
Page 82
S S S T
B
S S S T
B
S S S T
B
S = Support cycles T = Test Paces B = Backup Paces
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 43. Pacing threshold search

The pacemaker may use one to three of the test paces in a series to determine if a particular Amplitude or Pulse Width is above or below the patient’s stimulation threshold.
When the first of the three test paces indicates capture (CAP), or the last two test paces indicate capture following loss-of-capture (LOC) on the first pace, the series is determined to be above the threshold. See Table 6.
When two of the three test paces indicate loss-of-capture, the series is determined to be below the threshold. See Table 6.

Table 6. Above/below threshold determination

Series of 3 test paces Results
a
CAP
Above threshold LOCb, CAP, CAP Above threshold LOC, LOC Below threshold LOC, CAP, LOC Below threshold
a
CAP = capture
b
LOC = loss-of-capture
Not all test paces qualify for use in the determination series. If a test pace meets abort criteria (see page 86), the pace is ignored, the Amplitude and Pulse Width remain the same, and the next test pace is used in the determination series instead. For example, if a ventricular intrinsic sense inhibits the test pace during the test cycle, that intended pace will be ignored and not evaluated for capture.
If a support pace meets abort criteria, e.g., a ventricular safety pace occurs on a support pace, the pace is counted in the support/test cycle, but causes the next test pace to be ignored. The test pace following the ignored test pace remains at the Amplitude and Pulse Width and is used in the determination series.
Modifying Amplitude and Pulse Width during the search – When modifying first Amplitude and then Pulse Width during the pacing threshold search, the pacemaker is
82 Reference Manual
Page 83
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
looking for two points that lie on the strength duration curve. These points define the boundary between settings that capture the myocardium and those that do not.
Amplitude modification operates as follows:
1. The test Amplitude is set at the last Amplitude result from the previous pacing threshold search or at 0.75 V if no previous search has been done. Pulse Width is set at 0.4 ms.
2. The pacemaker initiates a series of support cycles and test paces to determine whether the Amplitude setting is above or below the patient’s stimulation threshold.
3. If the test Amplitude is above the patient’s threshold, the pacemaker will reduce it by one setting and repeat the test series. The process of reducing the Amplitude setting and retesting continues until a point below the patient’s threshold is found, indicating loss-of-capture.
4. The pacemaker then increments and retests the Amplitude one setting at a time until the setting is above the patient’s stimulation threshold for three consecutive test series, indicating capture is recovered. The setting at which capture is recovered is called the amplitude threshold.
Pulse Width modification operates as follows:
1. The test Pulse Width is set at the last Pulse Width setting from the previous search or at 0.21 ms if no previous search has been done. Amplitude is set at two times the amplitude threshold determined during the Amplitude search. The upper Pulse Width limit for the test pulse is 0.4 ms.
2. The pacemaker performs the actions detailed in Step 2 through Step 4 for Amplitude modification above, but for Pulse Width. The Pulse Width setting that is in operation when capture is recovered at two times the amplitude threshold is designated the pulse width threshold.
If the clinician uses the in-office Capture Management test, the programmer uses the amplitude threshold and pulse width threshold to construct a strength duration curve. (See Figure 44).
Reference Manual 83
Page 84
Amplitude (V)
1.00.4
13
2
4
3
1
2
4
Loss-of-capture
Starting Amplitude
Amplitude reduced by one setting (2)
Amplitude threshold (4)
Pulse Width reduced by one setting (2)
Pulse Width threshold (4)
Starting Pulse Width (at 2 times the amplitude threshold)
Loss-of-capture
Pulse width (ms)
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 44. Modifying Amplitude and Pulse Width

Pacing therapy during the search – During a pacing threshold search, the pacemaker
must provide ventricular test paces (which may affect normal pacing operation). To ensure ventricular pacing, if necessary, the pacemaker may adapt timing in both tracking and nontracking modes.
Note: If an MVP mode is programmed (AAIR<=>DDDR or AAI<=>DDD), Capture Management sets the pacing mode to DDDR or DDD and suspends MVP mode changes for the duration of the pacing threshold search.
When operating in tracking modes:
The pacemaker shortens the AV interval for each support cycle and test pace based on calculations using the shortest AV interval measured during the stable rhythm check.
The test pace is always followed by a backup pace after 110 ms.
PVARP is fixed at 350 ms for support cycles, but PVARP, PVAB, and the ventricular refractory period are timed from the backup pace on test paces.
Ventricular Safety Pacing is disabled on test pace cycles.
84 Reference Manual
Page 85
110 ms
Atrial pace
Support AV Support AV Support AV (-)15 ms
Support pace Support pace
Atrial pace
Test pace
Atrial pace
Backup pace
110 ms
Support pace
Test pace
Backup pace
Support pace
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 45. Rate modulation in DDD/tracking modes

When operating in nontracking modes:
The lower rate is not changed on support cycles.
The lower rate on test paces is set to the fastest V-V interval seen on any support cycle or seen during the stable rhythm check plus 15 bpm or minus 150 ms, whichever results in a faster rate.

Figure 46. Rate modulation in nontracking modes

Because the pacing threshold search operates as the highest priority feature in control of pacing cycle parameters, the following pacing features are disabled during the search:
Sensing Assurance
Rate Drop Response detection
AV modulation (Search AV+, Rate Adaptive AV)
NCAP operation
Lead Monitor
Auto PVARP and sensor-varied PVARP operations
Ventricular Safety Pacing on test paces
Atrial Preference Pacing
Sinus Preference
Reference Manual 85
Page 86
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
When the pacing threshold search aborts – Sometimes the pacing threshold search cannot be performed as scheduled. The pacemaker ensures that false events will not influence the determination of capture and loss-of-capture during the search by aborting the search immediately when the following occur:
The sensor rate is greater than the ADL rate.
Mode Switch occurs.
The pacemaker allows the following to occur several times before aborting the search:
Upper tracking rate extension on a test pace.
Nonrefractory ventricular senses in a tracking mode.
V-V rate greater than 90 to 100 bpm on support cycles, depending on the value of the upper sensor rate and upper tracking rate (see “Checking for stable rhythm”, page 80).
Ventricular Safety Pacing during the support cycles.
Consecutive ventricular refractory events.
Other conditions can also cause the pacing threshold search to abort:
Pacemaker RRT/ERI or low battery is detected.
Noise reversion is detected.
Initiation of a programming or transtelephonic session.
Capture is not determined during the entire search, indicating possible high thresholds.
Loss-of-capture does not occur during the entire search, indicating a possible lead problem or undetected intrinsic events.
When a pacing threshold search cannot be completed, the pacemaker will automatically initiate another search within 30 minutes (or within 15 minutes, if the Capture Test Frequency parameter is programmed for every 15 minutes). See “Scheduling the search”, page 79. If four search attempts abort during one test period, however, the pacing threshold test is suspended until the next test period.
When the pacing threshold search aborts, a message indicating why the search could not be performed is stored in the Capture Management Detail diagnostic (if enabled).
Preventing undersensing of ventricular evoked response events during the search –
To minimize the likelihood of undersensing of ventricular evoked response events during a pacing threshold search, the clinician can program the V. Sensing During Search parameter to Adaptive, if the Ventricular Sense Polarity parameter is set to Bipolar. Then, when an out-of-range measurement (see Figure 48) occurs, the pacing threshold search will be automatically repeated, using the Unipolar sensing setting.
86 Reference Manual
Page 87
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
If the second search is successful, the out-of-range measurement is ignored. If the second search results in another out-of-range measurement, outputs will remain at 5 V and 1 ms.
If the V. Sensing During Search parameter is set to Bipolar or Unipolar, ventricular sensing during the pacing threshold search will be done at that setting.
5.5.3.1 Automatic ventricular output adaptation
Capture Management can be programmed (Adaptive setting) to provide automatic adaptation of ventricular Amplitude and Pulse Width based on pacing threshold search results. Following each search, the pacemaker creates a target output by applying a programmable safety margin (Amplitude Margin parameter) to the amplitude threshold determined during the search. The pacemaker’s calculation for the target is always rounded up to the next programmable setting. The pacemaker then adapts outputs toward this target.
Adaptation can take place only within an output range that is defined by a programmable lower limit (Minimum Adapted Amplitude parameter) and the upper threshold limit of 5.0 V and 1.0 ms. The minimum pulse width for ventricular capture management is 0.4 ms.
Amplitude is adapted only when a pacing threshold search is successful; otherwise, it remains as programmed.
If the operating Amplitude is below the target, it is immediately adapted to the target. (See Figure 47.)
Reference Manual 87
Page 88
P= Pulse width (ms)
A=Amplitude (V)
Output Range
2
1
.2 .4 .6 .8 1.0
3
5
4
P
A
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 47. Adapting outputs upward to the target

1 Target (when 2.0X margins applied) 2 Operating Amplitude and Pulse Width 3 The bottom line indicates the Programmed Minimum Adapted Amplitude
If the operating Amplitude is above the target, the pacemaker adapts the Amplitude downward one programmable setting per successful pacing threshold search. If the target is below the programmed minimum output limit, the adaptation stops at the minimum limit.
A High Threshold Condition warning is issued if the amplitude threshold is greater than
2.5 V. The pacemaker responds by adapting to an Amplitude of 5.0 V and Pulse Width of 1.0 ms.
If the amplitude threshold multiplied by the safety margin indicates an amplitude target greater than 5.0 V, the pacemaker responds by adapting to the highest possible Capture Management settings, an Amplitude of 5.0 V and Pulse Width of 1.0 ms. (See Figure 48.)
88 Reference Manual
Page 89
1.2
2
1
.2 .4 .6 .8 1.0
3
5
4
6
Amplitude (V)
Pulse width (ms)
Output Range
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 48. Limiting high thresholds

1 Target (when 3X margin applied) 2 Output adapted to 5.0 V, 1.0 ms limit 3 Operating Amplitude and Pulse Width
5.5.3.2 Considerations when programming parameters
In a small percentage of patients, the following conditions can influence thresholds measured by Capture Management and can lead to possible symptoms:
Lead fixation – With poor lead fixation, modulations in pacing timing and rate can influence thresholds.
Reference Manual 89
4 Amplitude threshold 5 The bottom line indicates the Programmed
Minimum Adapted Amplitude
Page 90
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Intrinsic event rejection – In rare instances, combinations of morphology and rhythm may result in a low threshold measurement. This may occur if the pacing threshold search is unable to differentiate between myocardial contractions caused by the pacing pulse and those caused by physiologic means.
Evoked response detection – In rare instances, the pacemaker may not detect the electronic waveform created by the contracting myocardium immediately following a pacing pulse. In such instances, a high threshold measurement may result. See page 86
The Capture Management Threshold Test and manual threshold measurements provide data that can help in programming the parameters that control outputs (see Section 5.5.1.4, “Programmable Capture Management parameters”, page 80).
5.5.3.3 Programming interactions
Warning: Capture Management will not program ventricular outputs above 5.0 V or 1.0 ms.
If the patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually.
The pacemaker must be programmed to a mode that permits pacing and sensing in the ventricle (but not VVT mode) in order to use the Ventricular Capture Management feature.

5.5.4 Atrial Capture Management (ACM)

Checking for stable rhythm – Before a pacing threshold search can be initiated, the
pacemaker determines if the patient is pacing or sensing at a low rate. A low rate is desirable during the pacing threshold search to reduce the risk of competition from forced pacing with fast intrinsic rhythms. A Pacing Threshold Search (PTS) is performed when a stable atrial rhythm is observed for eight pacing cycles and the sensor rate is less than the ADL rate.
ACM does not operate during Mode Switch episodes.
ACM operates in DDDR and DDD modes.
The sensed atrial rate must not be faster than 87 bpm.
The paced atrial rate must be slower than 90 bpm.
7
If the device is programmed to an MVP mode (AAIR<=>DDDR or AAI<=>DDD), the pacemaker changes the mode to DDDR or DDD during ACM operations.
90 Reference Manual
7
Page 91
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
5.5.4.1 The atrial pacing threshold search
Conducting the search – The pacemaker performs the pacing threshold search through a
series of support cycles followed by a test pace applied at a slightly faster rate.
Beginning at one setting below the last measured value (or at 0.75 V if no previous search has been done), the amplitude is reduced in one setting decrements at 0.4 ms pulse width until loss-of-capture is detected. When two of the three test paces indicate loss-of-capture, the series is determined to be below the threshold.
The pacemaker then increments and retests the amplitude one setting at a time until the setting is above the patient’s stimulation threshold for three consecutive test series, indicating capture is recovered. The setting at which capture is recovered is called the amplitude threshold. The threshold criteria are shown in Table 7.

Table 7. Above/below threshold determination for a single test series

Series of up to 3 test paces Results
CAP, CAP LOCb, CAP, CAP Above threshold LOC, LOC Below threshold CAP, LOC, CAP Above threshold LOC, CAP, LOC Below threshold
a
CAP = capture
b
LOC = loss-of-capture
a
Above threshold
Not all test paces qualify for use in the determination series. If sensed events surrounding the test pace meet abort criteria, the pace is ignored, the Amplitude remains the same, and the next test pace is used in the determination series instead. If a support cycle meets abort criteria, the previous set of support cycles is discarded, and a new set of support cycles is started.
Atrial Capture Management (ACM) observes the timing of sensed P and R waves (not evoked response) to evaluate capture.
ACM automatically selects one of two methods for evaluating atrial capture, based on the patient’s rhythm at the time of the PTS. If the patient has a stable sinus rhythm, the device selects the Atrial Chamber Reset (ACR) method. Otherwise, the device selects the AV Conduction (AVC) method.
Reference Manual 91
Page 92
AS
VP
VP VP VP VP VP VP
VP VP VPVS or VP VP
AS
AS
AS
AS AP
AP
ARAS
AS
AS
AS
AS
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
5.5.4.2 Atrial Chamber Reset (ACR) Method
Atrial Chamber Reset runs during stable sinus rhythm. It evaluates capture by observing the response of the intrinsic rhythm to the atrial test pace. If the test pace does not capture, the sinus node is not reset, and an atrial refractory sensed event (AR) is observed after the test pace. If no AR is observed within the AV interval, ACR concludes that the test pace captured the myocardium. (See Figure 49.)

Figure 49. Atrial Chamber Reset test method

1 test 2 CAP
3 LOC
5.5.4.3 Atrial-Ventricular Conduction (AVC) Method
Atrial-Ventricular Conduction method runs when stable 1:1 AV conduction is observed with atrial pacing. The atrial pacing rate is increased by 15 bpm (but no faster than 101 bpm) and the AV interval is lengthened to try to achieve a stable AP-VS rhythm.
AVC evaluates capture by observing the conducted ventricular response to the atrial test pace. Each atrial test pace is followed by a backup pace at programmed amplitude and a
1.0 ms pulse width to maintain rhythm stability during the test. If a conducted VS event is observed at approximately the expected AP-VS interval following the atrial test pace, AVC concludes that the test pace captured the myocardium. (See Figure 50.)
92 Reference Manual
Page 93
70 ms
AP AP
VS VS
VP
AP-VS
Scheduled
VP
Overdrive rate interval Overdrive rate interval -
prematurity
Last support AP-VS interval
85 ms
Capture
Loss of Capture
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 50. Atrial-Ventricular Conduction test method

1 Test AP 2 Backup AP
5.5.4.4 Pacing therapy during the search
Because the pacing threshold search operates as the highest priority feature in control of pacing cycle parameters, the following pacing features are disabled during the search:
Sensing Assurance
Rate Drop Response detection
AV modulation (Search AV+, Rate Adaptive AV)
Blanked Flutter Search
Lead Monitor
Lower rate modulation (APP, Sinus Preference, Conducted AF Response)
Reference Manual 93
3 (Expected VS from Test AP) 4 (Expected VS from Backup AP)
Page 94
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Note: If an MVP mode is programmed (AAIR<=>DDDR or AAI<=>DDD), the capture management sets the pacing mode to DDDR or DDD and suspends MVP mode changes for the duration of the pacing threshold search.
When the pacing threshold search aborts – Sometimes the pacing threshold search cannot be performed as scheduled. The pacemaker ensures that false events do not influence the determination of capture and loss-of-capture during the search by aborting the search immediately when the following occur:
Fast atrial rate (>87 bpm with ACR, >100 bpm with AVC)
Ventricular pacing (AVC)
Undesirable cardiac events (PVCs, PACs)
Mode Switch occurs.
The pacemaker allows the following to occur several times before aborting the search:
fast or variable intrinsic rate or ectopic events
slow intrinsic rate8 and a lack of AV conduction
Variable AV conduction time
Unexpected sensed events before or after the test pace
Ventricular Safety Pacing
Other conditions can also cause the pacing threshold search to abort:
Pacemaker RRT/ERI or low battery is detected.
Initiation of a programming or transtelephonic session occurs.
Capture is not determined during the entire search, indicating possible high thresholds.
Loss-of-capture does not occur during the entire search, indicating undetected intrinsic events.
5.5.4.5 Automatic atrial threshold adaptation
Capture Management can be programmed (Adaptive setting) to provide automatic adaptation of atrial amplitude based on pacing threshold search results.
8
Turning the Sleep Function to On may allow ACM to run with slower intrinsic rates.
94 Reference Manual
Page 95
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
The pacemaker applies the programmable amplitude safety margin to the amplitude threshold value measured at a 0.4 ms pulse width to determine the target amplitude. If the operating amplitude is above the target, the pacemaker adapts the amplitude down toward the target in one-step decrements. If the operating amplitude is below the target, the amplitude is immediately adapted to the target.
Outputs are not adapted below the programmable minimum amplitude. During the programmable Acute Phase following implant detection, outputs are not adapted below the programmed outputs or the default shipping settings (3.5 V, 0.4 ms).
A High Threshold Warning is issued if the amplitude threshold is greater than 2.5 V. The pacemaker responds by adapting to an Amplitude of 5.0 V and Pulse Width of 1.0 ms.
A High Threshold warning is issued if the target amplitude is greater than 5.0 V and outputs are adapted to 5.0 V and 1.0 ms (See Figure 51).
Reference Manual 95
Page 96
1.2
2
1
.2 .4 .6 .8 1.0
3
5
4
6
Amplitude (V)
Pulse width (ms)
Output Range
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 51. Limiting high atrial thresholds

1 Target (when 3X margin applied) 2 Output adapted to 5.0 V, 1.0 ms limit 3 Operating Amplitude and Pulse Width
5.5.4.6 Programming interactions
Warning: Capture Management will not program atrial outputs above 5.0 V or 1.0 ms. If the
patient needs a pacing output higher than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually.
96 Reference Manual
4 Amplitude Threshold 5 The bottom line indicates the Programmed
Minimum Adapted Amplitude
Page 97
0
1
2
3
4
5
11/16 12/16 01/15 02/14 03/16 04/15 05/15 06/14
Average to max. threshold
Amplitude (V @ 0.4ms)
Date (Month/Day)
Monitor only "What if"
Max. adapted
Threshold > 2.5V
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
The pacemaker must be programmed to DDDR, DDD, AAIR<=>DDDR, or AAI<=>DDD mode in order to use the Atrial Capture Management feature.

5.5.5 Recording Capture Management data

Programming Capture Management also initiates automatic data collection. In addition, you may collect detailed information about every Capture Management search.
5.5.5.1 Summary Capture Management data
The automatic Capture Management trend records pacing threshold history. Every 7 days, the pacemaker collects the following Capture Management data automatically for 14 months:
Maximum Threshold
Average Threshold
Maximum Amplitude
Type of Capture Management search (Adaptive or Monitor Only)
Figure 52 shows a typical summary Capture Management trend.

Figure 52. Summary Capture Management data

Reference Manual 97
Page 98
0
1
2
3
4
5
09/04/0308/03/03
Amplitude (V)
Date
Threshold
Monitor only "What if"
Adapted
Aborted search
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
5.5.5.2 Detail Capture Management data
Capture Management Detail data provides a record of the pacing threshold, pulse width, and amplitude in the selected chamber. For the most recent 668 ventricular or 359 atrial Capture Management searches, the pacemaker records the following information for the selected chamber:
Date and Time
Threshold Amplitude
Threshold Pulse Width (VCM only)
Method of Search (ACM only)
Adapted or Monitor Only (suggested) Amplitude (with programmed margin applied)
Adapted or Monitor Only (suggested) Ventricular Pulse Width
A beat-to-beat trend for the most recent Capture Management search is available. Recording stops if more than 1000 events occur during the search. Clinicians have the option to record an EGM (atrial, ventricular, or summed EGM). Figure 53 shows an example of a detailed Capture Management trend.

Figure 53. Detail Capture Management trend

The data collected can be displayed or printed from the Data icon.
5.5.5.3 Clearing Capture Management data
Automatic Capture Management Trend data can be cleared from the Clear Data screen by selecting the option to clear lead impedance trend data for the respective chamber. However, the clinician-selected Capture Management Detail data is cleared automatically one hour after the end of a programming session.
98 Reference Manual
Page 99
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™
Also, you can select the option to clear data immediately. Be sure to save the session data or print the report before ending the patient session.
5.5.5.4 For further information
Refer to the Adapta/Versa/Sensia/Relia Pacemaker Programming Guide for information about collecting and displaying Capture Management data.

5.6 Sensing Assurance and diagnostic

5.6.1 Overview

The Sensing Assurance feature automatically adjusts atrial and ventricular sensitivities within defined limits. At the completion of Implant Detection, the pacemaker enables Sensing Assurance and begins monitoring the peak amplitude of sensed signals. In response to monitoring, the pacemaker automatically increases or decreases Sensitivity to maintain an adequate sensing margin with respect to the patient’s sensed P and R waves.

5.6.2 Monitoring sensitivity thresholds

The pacemaker monitors each nonrefractory sensed event (AS or VS) by measuring the ratio of the peak amplitude of the P or R wave to the Sensitivity setting. The pacemaker then compares the measured sensing margin to a target sensing margin.
Sensing Assurance provides the following target sensing margins:
Atrial bipolar sensing: 4:1 ratio to 5.6:1 ratio
Atrial unipolar sensing: 2.8:1 ratio to 4:1 ratio
Ventricular (unipolar or bipolar) sensing: 2.8:1 ratio to 4:1 ratio
Reference Manual 99
Page 100
2.8 mV
4 mV
1 mV
Atrial Unipolar Sensing
Medtronic ADAPTA® / VERSA® / SENSIA® / RELIA™

Figure 54. Maintaining a sensing margin of 3:1

1 Target sensing margin 2 Sensitivity setting

5.6.3 Qualifying sensed events

When monitoring nonrefractory sensed events, the pacemaker checks each event to see if it qualifies for use in determining Sensitivity threshold adjustments. Events are disqualified when:
The pacemaker is monitoring a high level of continuous interference
A second event occurs in either chamber within 40 ms after a sensed event

5.6.4 Adjusting sensing thresholds

To adjust sensing thresholds, the pacemaker keeps a record of many sensed events. Non-PVC, nonrefractory senses that fall below the target sensing margin (low events) are assigned a negative value, and those that are above the target sensing margin (high events) are assigned a positive value. When the accumulated value of the events exceeds the upper or lower limits of a counter, the pacemaker will adjust the sensitivity by one setting:
Many low events indicate an adjustment of one setting to a more sensitive (smaller numerical) setting.
Many high events indicate an adjustment of one setting to a less sensitive (larger numerical) setting.
At least 17 low events are required to cause an adjustment to the next more sensitive setting, and 36 high events are required to cause an adjustment to the next less sensitive setting.
Adjustments occur more gradually if a mixture of low and high events are occurring or if paced events are intermingled with sensed events. If fewer than 60% of events are high (or low), or if the pace to sense ratio is greater than 5:1, no sensitivity adjustment is made.
100 Reference Manual
Loading...