Medtronic and Medtronic logo are trademarks of Medtronic.
AdaptiveStim™, Sequentia™, SoftStart/Stop™, and Vanta™ are
Importer
trademarks of a Medtronic company.
Android™ is a trademark of Google LLC.
For USA audiences only
The BLUETOOTH
®
word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc., and any use of such marks by
Medtronic is under license.
®
Wi-Fi
and Wi-Fi Alliance® are registered trademarks of Wi-Fi Alliance.
Information available for the system:
The information for prescribers manual provides information about contraindications, warnings, precautions, adverse events, sterilization, patient selection,
individualization of treatment, and component disposal. For customers in Japan, the appropriate package insert provides information about safety,
contraindications, warnings, precautions, and adverse events.
The indications sheet provides information about indications and related information. For customers in Japan, the appropriate package insert provides
information about indications.
The system eligibility and battery longevity manual describes programming considerations and provides battery longevity information to aid in the appropriate
neurostimulator selection.
MRI guidelines provide information about any MRI conditions and MRI-specic contraindications, warnings, and precautions for MRI scans with the
neurostimulation system.
Product manuals, such as programming guides, recharging guides, and implant manuals provide device descriptions, package contents, device
specications, product-specic warnings and precautions, and instructions for use.
The clinical summary provides information about the clinical study results for the neurostimulation system.
Refer to the literature provided by the clinician tablet manufacturer for information regarding wireless use.
Refer to the communicator technical manual for warnings, precautions, device description, package contents, device specications, instructions for use, and
maintenance information.
A71200 English 2022-01-15
2
A71200 English 2022-01-15
3
Table of Contents
Description of the Programming System ������������ 7
Model A71200 Vanta / Sequentia LT
clinician programmer application
The Medtronic Model A71200 Vanta / Sequentia LT clinician
programmer application (app) is intended for use by clinicians
in the programming of the Medtronic Model 977006 Vanta and
Model 977005 Sequentia LT implantable neurostimulators for pain
therapy.
Intended purpose
The clinician programmer application is intended for use by clinicians
in the programming of Medtronic neurostimulators for pain therapy.
Other components of the
programming system
The clinician programmer app is intended for use with the following
components.
Model CT900 Clinician Tablet
The clinician programmer app is installed on the Model CT900
Clinician Tablet with Android-based operating system.
Model A901 Communication Manager
application
The Model A901 Communication Manager application is an
application on the clinician tablet that manages the telemetry
communications for the clinician tablet.
Model A902 Patient Data Service application
The Model A902 Patient Data Service application is an application
on the clinician tablet that manages the storage of patient session
data on the clinician tablet. The Patient Data Service app can be
used to access or delete patient session data for any patients whose
Medtronic devices have been programmed using any clinician
programmer app on the clinician tablet.
Model 8880T2 Communicator
The Model 8880T2 Communicator is a handheld device used in
conjunction with the clinician tablet and clinician programmer app to
communicate with Medtronic neurostimulators.
A71200 English 2022-01-15
7
Description of the Programming System
Programmable Settings
The clinician programmer app is used to enter, review, and modify
programmable settings in the neurostimulator. These programmable
settings dene the stimulation therapy delivered to the patient.
Pulse width, amplitude, and electrode conguration are assigned
at the program level: that is, each program within a group can have
different values. Rate is assigned at the group level, and the rate of
each program within a group can be set as a fraction of the group
rate.
Stimulation parameters
Stimulation is delivered in the form of electrical pulses. Stimulation
parameters dene the attributes of these pulses and can be adjusted
■Pulse width: Duration of each pulse in microseconds (μs).
■Rate: Frequency of pulses in hertz (Hz).
■Amplitude: Strength of a pulse in milliamps (mA).
Groups and programs
A program is a specic combination of pulse width, rate, and
amplitude settings acting on a specic electrode conguration. The
program denes the stimulation pulses that will be delivered for
therapy. The Model 977005 Sequentia LT neurostimulator supports
one program.
For the Model 977006 Vanta neurostimulator, multiple programs can
be combined into groups.
When using more than one program, the pulses are delivered
sequentially—rst a pulse from one program, then a pulse from the
next program, and so on. Pulses from different programs are never
delivered simultaneously. Each group and its associated programs
can be used to provide therapy for specic areas of coverage or
specic patient activities.
Optional features
Optional features are available for adjusting the stimulation to
the patient's needs. Some features are not available for the
Model 977005 Sequentia LT neurostimulator.
■Adjustment: Allows the patient to increase or decrease
amplitude for all programs in the active group at the same time.
■AdaptiveStim Technology: Adjusts stimulation automatically
when the patient changes position.
■Cycling: Cycles stimulation on and off at clinician-determined
intervals.
■Electrode Redistribution: Assigns a lower or a higher
percentage of the program amplitude to an individual electrode
in a program.
■Equalize: Sets program amplitude to zero and recongures
the active electrodes to provide equal distribution of amplitude.
■IntelliStim: A semi-automated feature for selecting electrodes
to deliver therapy. IntelliStim is a scanning process in which
common combinations of electrodes are scanned across the
lead using the selected amplitude.
■Move: Sets program amplitude to zero and allows a congured
electrode combination to be moved up, down, or across the
lead or leads.
■Patient Limits: Identies the stimulation parameters that
the patient can adjust and sets the upper limits for those
parameters.
A71200 English 2022-01-15
8
Programmable Settings
■SoftStart/Stop: Slowly increases the amplitude when
stimulation is turned on and slowly decreases the amplitude
when stimulation is turned off.
Data Security and Network
Connectivity
Data security
The clinician programming system uses and stores data about the
patient's health and implanted medical device. This data is protected
by application-level controls and encryption provided by the clinician
tablet. The clinician programming system does not provide data
protection for data exported to another destination. Exported data
should be handled in accordance with your facility’s security policy for
data handling and storage.
Medtronic recommends that you always save exported data to the
default reports location on the clinician tablet.
Network connectivity
Network connectivity is required for initial app registration and for
installation of Medtronic app updates and communicator rmware
updates. Network connectivity is not required for neurostimulator
programming. To protect your clinician programming system,
Medtronic recommends you implement the following security
measures:
■Secure your clinician tablet by disabling network connectivity
during any programming session.
■Use a managed, trusted Wi-Fi connection when network
connectivity is needed.
■Connect the clinician tablet to the network periodically to check
for notications on available updates.
Note: Connecting the clinician tablet to a network that includes other
equipment could result in unforeseen risks to patients, operators, or
third parties. Changes to your network such as adding, disconnecting,
A71200 English 2022-01-15
9
Data Security and Network Connectivity
and upgrading equipment; upgrading or installing software; or
changing network congurations could also introduce additional risks.
Analyze, evaluate, and control any identied risks.
If you suspect a cybersecurity event has occurred, stop using the app
(if possible) and contact your IT Security team or Medtronic Technical
Services to document and respond to the suspected incident.
If your clinician tablet is lost or stolen, contact Medtronic using the
contact information listed on the back cover of this manual.
Installing application updates
For assistance with downloading, registering, or installing the
application, contact Medtronic using the information on the back
cover of this manual. In the U.S., call 1-800-707-0933 for Medtronic
Technical Services.
Medtronic periodically updates the therapy app and will not install
updates without notifying you. Network connectivity is required to
update the app. When notied that an app update is available, follow
the instructions provided by Medtronic to install the update.
Installing communicator updates
When notied to install a communicator rmware update, connect the
communicator to the clinician tablet using the USB connector cable.
Follow the instructions provided by Medtronic to install the update.
Returning the clinician tablet
If you need to return the clinician tablet for disposal or replacement,
contact Medtronic using the contact information listed on the back
cover of this manual. Instructions will be provided for preparing the
clinician tablet for return.
General Warnings and Cautions
If a serious incident related to the app occurs, immediately report the
incident to Medtronic and the applicable competent authority.
MRI warning for scan eligibility
Warning: Always program the following component information
accurately:
■Neurostimulator and lead model numbers
■Neurostimulator implant location and lead tip location
■Presence of any abandoned leads, extensions, and pocket
adaptors
If this information is not up-to-date or is entered incorrectly, MRI scan
eligibility data will be inaccurate, and the patient is at risk for one of
the following:
■The patient is allowed to have an MRI scan inappropriate for
the implanted components, which could cause tissue heating,
resulting in tissue damage or serious patient injury.
■The patient is unnecessarily restricted from having an MRI
scan.
Refer to the MRI Guidelines for Medtronic Neurostimulation Systems for Chronic Pain instructions for use manual for MRI scan eligibility
information, MRI scan conditions, and MRI-specic warnings and
cautions for conducting an MRI scan.
Sterile eld warning for
programming components
Warning: To use the programming components (external
and nonsterile) in a sterile eld, place a sterile barrier between the
patient and the programming components to prevent infection. Do
A71200 English 2022-01-15
10
General Warnings and Cautions
not sterilize any of the programming components. Sterilization may
damage the programming components.
EMI caution for telemetry signal
disruption
Caution: Electromagnetic interference (EMI) can disrupt
programming and telemetry communication. If EMI interference
is suspected, such as from radio frequency identication (RFID)
equipment, move away from the likely source of EMI and try again.
Cautions related to clinician
programming and possible
interactions with other devices
Refer to the Information for Prescribers booklet for cautions related
to clinician programming and possible interactions with other devices
(such as cardiac devices).
Setting up the Programming
System
Clinician tablet with
clinician programmer app
Power button
The connector cable is required for pairing
the communicator to the clinician tablet.
After pairing, the communicator and the clinician tablet can
communicate using BLUETOOTH wireless technology.
For wireless communication, make sure the clinician tablet
Figure 1. Components of the programming system.
remains within range.
Preparing the clinician tablet
Refer to the Model CT900 Clinician Tablet quick start guide for
instructions on how to turn on the tablet and to complete initial setup.
Check the battery level of the clinician tablet.
■The tablet's battery level is shown on the tablet status bar
(uppermost row on the tablet screen).
■Recharge the tablet if the battery level is low.
USB
connector cableCommunicatorNeurostimulator
Power button
Proximal Telemetry
Hold communicator directly
over neurostimulator.
BLUETOOTH
A71200 English 2022-01-15
11
Setting up the Programming System
Finding and opening the app
1. Navigate to the Apps on the clinician tablet.
2. Find the Vanta app icon.
Note: Vanta app is used to program both the Vanta and
Sequentia LT neurostimulators.
3. Tap the Vanta app icon to open the app.
Pairing the communicator to the
clinician tablet
Refer to the Model 8880T2 Communicator technical manual for
additional instructions, if needed. Only one communicator can be
paired with a clinician tablet at a time.
1. Connect the USB connector cable to the communicator and
the clinician tablet.
2. Turn on the communicator (slide the power button down, then
release).
3. A message appears on the clinician tablet, which asks
permission to open the Communication Manager when the
USB device is connected:
a. Select to use the USB device by default. This selection
prevents this message from appearing again.
b. Select OK.
4. On the initial app screen, tap CONNECT. Pairing begins and
the communicator displays a green LED light between the
tablet icon and the communicator icon.
Once the communicator is paired, you can disconnect the USB
connector cable. The communicator and clinician tablet communicate
using BLUETOOTH wireless technology. For wireless communication,
make sure the clinician tablet remains within range.
Note: If you have trouble pairing, either with the USB connector cable
or BLUETOOTH, restart the tablet and try again.
Initiating communication with the
neurostimulator
1. Hold the communicator directly over the neurostimulator.
Notes:
■The target symbol (
indicates the location of the internal antenna. The target
should be facing and centered over the neurostimulator.
■After making the initial connection, you can make future
connections at a greater distance for 12 hours following
the initial connection, or until the communicator
connects to a different neurostimulator.
2. On the Search for Device screen, tap FIND DEVICE.
Or
Press the communicate button (
3. Select the serial number of the neurostimulator you intend to
congure, and tap CONNECT.
4. After connecting to the neurostimulator, the CURRENT DEVICE STATUS screen will display.
From this screen, you can select workow options under
the SELECT FLOW heading. Refer to the "Neurostimulator
Workows" section on page 16 for workow details.
) on the back of the communicator
) on the communicator.
A71200 English 2022-01-15
12
Setting up the Programming System
Notes:
■If you have problems establishing communication, refer to
"Communicator and clinician tablet troubleshooting" on page
53.
■Do not leave the clinician tablet unattended during an active
programming session.
■When a congured neurostimulator is interrogated, a lead
connectivity test is performed to check for impedance issues.
The patient may feel a stimulation change during the test.
Overview of the Clinician
Programmer App
Using Demo mode
Use Demo mode to explore the clinician programmer app without
interrogating or updating a neurostimulator.
When the app is in Demo mode, it is not communicating with a real
neurostimulator and does not require the use of a communicator. Any
data displayed in Demo mode is not actual.
Demo mode can be used for training and demonstration purposes
and to familiarize yourself with the app interface before starting an
actual programming session.
1. On the initial Vanta/Sequentia LT app screen, tap DEMO.
2. Select a device, and tap CONNECT.
3. Navigate through the workow. One workow at a time can be
explored in a Demo session.
4. To exit Demo mode, go to the Summary screen and tap EXIT WORKFLOW.
Exiting the workow will return you to the initial app screen
where you can select DEMO again and select another device
to explore.
A71200 English 2022-01-15
Navigating through workows
The screens typically used during a programming session are
grouped into workows. The workows vary depending on the device
model and device status (new versus congured). The workows may
also vary depending on neurostimulator features.
13
Overview of the Clinician Programmer App
After you start a workow, a workow navigator appears at the top
of the screen. For example, the Followup workow for the Vanta
neurostimulator contains these screens:
Device
AdaptiveStim
Lead Select Tip Location Programs Impedance
Diaries Reports Summary
■To move from screen to screen, swipe left or right.
■Alternatively, you can tap on the screen name in the workow
navigator to go directly to that screen.
–If the workow navigator extends beyond the width of
the screen, swipe the workow navigator left or right to
show the screen names that are partially hidden.
■The current screen is underlined and highlighted.
■If input is required in a workow screen, you will not be able to
advance to the next screen until you enter the required input.
–The app will prompt you for the required input.
–Orange asterisks indicate required elds.
Using the Side Menu
Screens not in the current workow can be accessed from the Side
Menu. For example, the Patient Info screen is in the new device
workow, but not in the Followup workow. To change patient data in
a congured device, you need to access the Patient Info screen from
the Side Menu. To access the Side Menu, tap the Side Menu button
(
) in the top-left corner on the action bar.
On-screen help
Help information is available in the clinician programmer app for most
workow screens. View the help information to learn how the app
interface works and how to complete tasks on the screen.
■To display help information, tap the Help button (
right corner on the action bar.
■Tap the Help button (or tap anywhere on the screen) to close
the help screen.
) in the top-
Buttons and indicators on the action
bar
An action bar appears at the top of all screens in all workows. Table
1 describes the buttons and indicators that may appear on the action
bar.
Table 1. Buttons and indicators.
Button / IndicatorDescription
Side Menu button: Tap to access
screens outside of the workow.
Back button: Tap to close a screen you
accessed from the Side Menu and return
to the previous workow screen.
Demo mode indicator: Appears in the
action bar when in Demo mode.
Stimulation toggle: Tap to turn
stimulation on or off. Green (switch to the
right) indicates that stimulation is on.
Appears in the action bar when at least
one program exists.
A71200 English 2022-01-15
14
Overview of the Clinician Programmer App
Button / IndicatorDescription
AdaptiveStim indicator: Applicable
to neurostimulators with AdaptiveStim
Technology (977006 Vanta).
Appears in the action bar when
AdaptiveStim Technology is enabled.
MRI Status indicator: Appears in the
action bar when stimulation is off and
MRI Mode has been activated.
The MRI Status indicator does not
provide MRI scan eligibility information.
Help button: Tap to display help
information for the current screen.
Tap the Help button (or tap anywhere on
the screen) to close the help screen.
Go to Summary screen button: Tap to
go directly to the Summary screen. You
can exit the workow from the Summary
screen to properly end the programming
session.
When the app is placed in the
background
When you navigate away from the Vanta clinician programmer app
while in a programming session, the app is placed in the background
and is represented as a oating widget (Vanta Session in Progress)
to remind you that you are still in a session.
■Press and drag the oating widget to move it on the tablet
screen.
■Tap on the oating widget to return to the app and resume the
session.
Overview of the Communicator
The Model 8880T2 Communicator is a nonsterile component used in
conjunction with the clinician tablet and clinician programmer app to
communicate with Medtronic neurostimulators.
Warning: To use the communicator in a sterile eld, place
a sterile barrier between the patient and the communicator
to prevent infection. Do not sterilize any of the programming
components. Sterilization may damage the programming
components.
The communicator is handheld and battery-operated. Communication
between the communicator and the clinician tablet can occur
wirelessly using BLUETOOTH technology or wired using the USB
connector cable.
For information on the Model 8880T2 Communicator, including
description, specications, components, instructions for use,
maintenance, and troubleshooting, refer to the Model 8880T2
Communicator technical manual.
Using the USB connector cable
The following situations require the use of the USB connector cable:
■First-time pairing of any communicator with a clinician tablet.
■Firmware updates to the communicator.
■Environments where multiple devices are using BLUETOOTH
wireless technology and thereby creating interference.
■Environments where BLUETOOTH wireless technology is
prohibited.
A71200 English 2022-01-15
15
Overview of the Communicator
Communicator icons on the tablet
status bar
Table 2 describes the icons that appear in the tablet status bar
(uppermost row on the tablet screen) indicating the status of the
connection between the communicator and the clinician tablet.
Table 2. Communicator icons in the tablet status bar.
Communicator IconDescription
The communicator is in the process of
connecting to the clinician tablet.
The communicator is communicating
with the clinician tablet via the USB
connector cable.
The communicator requires the USB
connector cable to communicate with the
clinician tablet.
The communicator is communicating
with the clinician tablet using
BLUETOOTH wireless technology.
Neurostimulator Workows
When an implantable neurostimulator (INS) is interrogated, the
clinician programmer app determines if the INS is new or if it has
been previously congured. Based on the status of the INS, the
clinician programmer app automatically presents the applicable
workow options.
■Start Usage > Implant Device workow – go to page 17.
Use when a new INS is being congured (programmed) for the
rst time.
■Followup workow – go to page 19.
Use at followup sessions with the patient.
■View MRI workow – go to page 20.
Use to determine MRI scan eligibility and prepare a patient's
neurostimulation system for an MRI scan.
Notes:
■To display on-screen help information, tap the Help button (
■The screens typically used during a programming session are
shown in the workow navigator. Additional screens not in the
current workow can be accessed from the Side Menu (
■Unsaved information may be lost if the application is improperly
terminated during a programming session. To end a session,
go to the Summary screen and tap the EXIT WORKFLOW
button. Ending a session generates a complete session record
that is stored in a database on the clinician tablet.
■Some functions, such as multiple groups and programs,
and AdaptiveStim, are not available with the Sequentia LT
neurostimulator.
).
).
A71200 English 2022-01-15
16
Neurostimulator Workows
Start Usage > Implant Device
workow
You can initiate communication with the INS in the package before
it is moved into the sterile eld. Tasks such as checking the battery
status and entering device and patient information can be performed
without removing the INS from the package.
1. On the CURRENT DEVICE STATUS screen, conrm that the
device information corresponds with the intended INS.
2. Tap START USAGE.
This action takes the neurostimulator out of shelf state and
sets the In Use Date/Implant Date elds to the current date.
■The In Use Date and Implant Date are the same date.
■The In Use Date is shown on the CURRENT DEVICE
STATUS screen, and the Implant Date is shown on the
About System screen.
■The About System screen is available after an implant
location is selected on the DEVICE screen.
■You can change the implant date on the About System
screen, but this change will not impact the clock inside
the neurostimulator that determines ERI and EOS.
3. Under IMPLANT DEVICE, tap START.
4. Use Table 3 to guide you through the screens and tasks in the
workow.
A71200 English 2022-01-15
17
Neurostimulator Workows
Table 3. Implant Device workow.
ScreenTasksMore information
DeviceCheck the battery status of the neurostimulator.Page 21
Select the implant location.Page 21
Patient InfoEnter patient information.Page 21
Pain MapTap the PAINT button, then use your nger to highlight the area on the human gure where the
patient feels pain. To remove highlight from an area on the gure, tap the ERASE button and then
tap the highlighted area.
Lead SelectSelect the implanted leads. Select a lead from the scrollable drop-down list, and then drag and drop
it at the appropriate electrode numbers. Tap UPDATE to save the selected lead.
Ensure leads are properly connected to the neurostimulator, and then check connectivity by tapping
the CHECK CONNECTIVITY button.
Tip LocationIndicate whether extensions and pocket adaptors are implanted. Indicate whether abandoned leads
remain implanted. Select the tip location for each lead.
Programs (and
subscreens)
Create programs in groups. Tap a group to make it active. A highlighted box around a group indicates
the group is active. Tap the plus sign to add a program within a group.
Program subscreen: Assign the electrode conguration for the lead(s). Assign pulse width, rate,
and amplitude.
Lead Manipulation subscreen: Use to represent the relative positioning of the implanted leads as
viewed from the patient’s back.
Energy subscreen: Set parameters for SoftStart/Stop and cycling. Estimate battery life based
on various congurations and hours per day of group use with the Estimated Battery Longevity
feature.
Patient Access subscreen: Assign the stimulation parameters that the patient can adjust.Page 31
ImpedanceMeasure electrode impedance.
These measurements verify the integrity of lead, extension, and connector pathways. They may
provide information about lead problems (such as lead breakage, short circuit, open circuit).
SummaryConrm therapy settings.
End the session properly by tapping the EXIT WORKFLOW button.
Page 21
Page 21
Page 22
Page 22
Page 23
Page 23
Page 29
Page 30
Page 32
--
A71200 English 2022-01-15
18
Neurostimulator Workows
Followup workow
1. On the CURRENT DEVICE STATUS screen, conrm that the
patient and device information corresponds with the intended
patient and INS.
2. Under FOLLOWUP, tap START.
3. Use Table 4 to guide you through the screens and tasks in the
workow.
Table 4. Followup workow.
ScreenTasksMore information
DeviceCheck the battery status of the neurostimulator.Page 21
Lead SelectReview the information, and make any updates. Updates may delete programmed settings.Page 21
Tip LocationReview the information, and make any updates.Page 22
Programs (and
subscreens)
ImpedanceMeasure electrode impedance.
AdaptiveStimCongure the settings for AdaptiveStim Technology (Vanta Model 977006 only).Page 38
DiariesSelect a diary to view from the drop-down list. Diaries provide information on therapy usage.Page 48
ReportsSelect a report to view, download, or delete.Page 49
SummaryConrm therapy settings.
Review the groups, programs, and settings, and make any updates. Conrm that the active group is
the intended group.
Program subscreen: Assign the electrode conguration for the lead(s). Assign pulse width, rate,
and amplitude.
Lead Manipulation subscreen (only applicable if more than one lead): Use to represent the
relative positioning of the implanted leads as viewed from the patient’s back.
Energy subscreen: Set parameters for SoftStart/Stop and cycling. Estimate battery life based
on various congurations and hours per day of group use with the Estimated Battery Longevity
feature.
Patient Access subscreen: Assign the stimulation parameters that the patient can adjust.Page 31
These measurements verify the integrity of lead, extension, and connector pathways. They may
provide information about lead problems (such as lead breakage, short circuit, open circuit).
End the session properly by tapping the EXIT WORKFLOW button.
Page 23
Page 23
Page 29
Page 30
Page 32
--
A71200 English 2022-01-15
19
Neurostimulator Workows
View MRI workow
1. On the CURRENT DEVICE STATUS screen, conrm that the
patient and device information corresponds with the intended
patient and INS.
2. Tap VIEW MRI > START.
3. Use Table 5 to guide you through the screens and tasks in the
workow.
Table 5. View MRI workow.
ScreenTasksMore information
DeviceConrm the neurostimulator information and implant location.Page 36
Patient InfoConrm the patient information.Page 36
Lead SelectConrm the implanted lead information.Page 36
Tip LocationConrm the conguration and tip location information.Page 36
MRIReview the MRI scan eligibility.
Determine whether the patient is eligible for the intended MRI scan.
If patient is eligible, turn stimulation off to prepare the patient’s neurostimulation system for the scan.
ReportsPrepare the MRI Report. The MRI Report contains information intended for MRI clinicians about the
MRI scan eligibility of the implanted system.
SummaryConrm that stimulation is off and that MRI status shows the device is ready for an MRI scan.
End the session properly by tapping the EXIT WORKFLOW button.
Page 36
Page 36
--
A71200 English 2022-01-15
20
Neurostimulator Workows
Working with Device and Patient
Information
Checking battery status of the
neurostimulator
The current battery level for the neurostimulator, along with the
estimated battery life remaining, appears on the Current Device Status and Device screens. The estimated battery life remaining is
shown next to the Estimated Battery Remaining (Based on Last 7
Days) eld.
The current battery level also appears on the Summary screen. For
information on battery service life for implanted neurostimulators,
refer to "Elective Replacement Indicator (ERI) message" on page
53 and "End of Service (EOS) message" on page 53.
Entering implant location
The implant location for implantable neurostimulators is entered
on the Device screen. The Device screen is accessed through the
workow navigator.
■Select the implant location from the scrollable drop-down list.
Note: The implant location impacts the MRI scan eligibility shown on
the MRI screen.
Entering patient information
Patient information such as patient name and diagnosis is entered on
the Patient Info screen. The Patient Info screen is accessed through
the workow navigator or by tapping Patient Info on the Side Menu
(
) when the screen is not part of a workow.
■Enter the applicable patient information. Orange asterisks
indicate the required elds.
Mapping patient pain
The Pain Map screen is used to highlight areas of the body where
the patient feels pain. There are two options for updating the patient's
pain map: PAINT and ERASE. Only one can be selected at a time.
The Pain Map screen is accessed through the workow navigator or
by tapping Pain Map on the Side Menu (
part of a workow.
■Select the PAINT option and tap the area on the human gure
to highlight where the patient feels pain.
■Use the ERASE option to remove highlighting.
■Tap the ROTATE button to rotate the human gure from front to
back and from back to front.
■Tap UPDATE and swipe left to go to the next screen in the
workow.
) when the screen is not
Entering lead information
The Lead Select screen is used to identify which lead models
are implanted, associate the lead electrode numbering to the
neurostimulator electrode numbering, and check connectivity of the
lead contacts or extension contacts with the neurostimulator contacts.
The Lead Select screen is accessed through the workow navigator.
■To assign a lead, select the lead model from the scrollable
drop-down lists. Press and hold, then drag the lead image to
the appropriate electrode numbers of the device.
A71200 English 2022-01-15
21
Working with Device and Patient Information
■To delete a lead, tap the lead image, and then tap the X that
appears.
■To change a lead, you must rst delete the current lead, and
then assign a new lead.
Notes:
■Make sure the electrode numbering on the screen matches the
electrode numbering of the implanted leads. If necessary, you
can swap congured leads using the drag-and-drop feature.
■On Medtronic surgical leads, the blue box on the lead arm
represents the marker band on the actual lead. For more
information about surgical leads, refer to the lead implant
manual. The marker band indicates the lead arm for electrodes
0-7.
■The assigned leads impact the MRI scan eligibility shown on
the MRI screen.
Testing lead connectivity
Once the implanted lead(s) are assigned on the Lead Select screen,
you can check the connectivity of the lead contacts and/or extension
contacts with the neurostimulator contacts. Prior to running the
test, ensure the leads or extensions are properly connected to the
neurostimulator.
■Tap the CHECK CONNECTIVITY button (
disabled, tap UPDATE on the Lead Select screen to save the
lead data and enable the button.)
–Green electrodes indicate that the lead contacts and/
or extension contacts have a good connection to the
neurostimulator contacts.
–Red electrodes (with an “X”) indicate a bad connection.
). (If the button is
–If there are red electrodes, gently reconnect the leads or
extensions for a better connection to the neurostimulator
contacts. Then repeat the connectivity check.
Entering extensions, pocket
adaptors, abandoned leads, or tip
location information
The Tip Location screen is accessed through the workow navigator.
■Check the Yes or No checkbox to indicate whether abandoned
leads remain implanted.
■Check the Yes or No checkbox to indicate whether extensions
are implanted. If the No checkbox is selected for extensions,
the No checkbox for pocket adaptors is automatically selected.
■Check the Yes or No checkbox to indicate whether pocket
adaptors are implanted.
■Select the tip location for each lead from the scrollable drop-
down lists.
Note: The abandoned lead, extension, pocket adaptor, and tip
location information impacts the MRI scan eligibility shown on the MRI
screen.
A71200 English 2022-01-15
22
Working with Device and Patient Information
Adding uoro images
The uoro screen is accessed using the View Fluoro button at the
bottom-left of the Tip Location screen. This feature is not available
for the Model 977005 Sequentia LT neurostimulator.
Note: The View Fluoro button also appears at the bottom-left of the
Lead Manipulation subscreen (which is accessed from the Program
subscreen).
■Use the Add Photo button to add existing uoro images saved
in the tablet or take a photo of the uoro image displayed on
the monitor. The app can store up to four images.
■Annotate images with the Text button. Tap where you want the
text to appear on the image and enter the text. Edit existing
text by tapping the Text button and then tapping the text you
want to edit on the screen.
■Tap the Trash button (
uoro image exists in the uoro screen, swipe left or right to
select the image you want to delete.
) to delete an image. If more than one
Working with Groups and
Programs
Programs screen
The Programs screen is used to create, copy, move, and delete
programs within groups and to dene the active group. The
Programs screen is accessed through the workow navigator.
Note: For patients implanted with Sequentia LT, only one program is
available for programming.
■Only one group can be active at a time. A highlighted box
around a group indicates the active group.
■To make another group active, tap that group name.
■To copy or move a program, press and hold, then drag that
program to the intended location, and then select MOVE or
COPY from the message that appears.
■To delete a program, press and hold, then drag that program to
the Trash icon (
■To create a new program, tap the add symbol (+) on the
desired program. This will launch the Program subscreen,
where you can assign the program settings.
■To view the program settings for an existing program, rst
make the group active, and then tap the program to view the
Program subscreen.
Note: The AdaptiveStim indicator (
if AdaptiveStim Technology is enabled for the group.
)�
) appears next to the group name
A71200 English 2022-01-15
Program subscreen
The Program subscreen is used to assign, review, and change
program settings for programs in the active group. The Program
23
Working with Groups and Programs
subscreen is accessed by tapping a program on the Programs
screen.
■A line is shown under the program number that is currently
displayed on the Program subscreen.
■To move from program to program within the active group,
swipe left or right on the screen. Alternatively, you can tap a
program number.
■Use the Program subscreen to assign the electrode
conguration for the leads and to set pulse width, rate, and
amplitude. Refer to "Adjusting Stimulation Parameters" on
page 24.
■From the Program subscreen, you can access the following
subscreens:
–Lead Manipulation: For more information, refer to
"Relative Lead Positioning" on page 29.
–Energy: Refer to "Customizing SoftStart/Stop and
Cycling" on page 30. For estimating battery life based
on various congurations and hours per day of group
use, refer to "Estimating Battery Longevity" on page
31.
–Patient Access: For more information, refer to
"Enabling Patient Access" on page 31.
Turning stimulation on or off
The Stimulation toggle is found on the action bar at the top-right
corner of the screen. The toggle is available when at least one
program exists. When no programs exist, the area in the action bar
reads “Stimulation Not Setup”.
■Tap the Stimulation toggle to turn stimulation on (
(
) at any time.
) or off
Adjusting Stimulation
Parameters
Assigning electrode conguration
Electrode conguration for the leads is assigned on the Program
subscreen.
Electrode conguration must include at least one negative (–) and
one positive (+) electrode. You cannot set amplitude for the program
until you have assigned at least one negative (–) and one positive (+)
electrode to the lead.
■Tap an electrode to change its polarity to negative (–), positive
(+), or off. Tap Update when nished.
■To delete an electrode, press and hold, then drag that
electrode to the Trash icon (
■To delete all assigned electrodes, tap the Select All button
(
), and then tap the Trash icon ( ).
■To move all electrodes at once, tap the Select All button (
and then press and drag the electrodes to a new location on a
lead. This action will set the amplitude to 0.0 mA.
■To move an individual electrode, you must rst delete that
electrode and then assign a new electrode.
Note: Adding or deleting an electrode after the amplitude is assigned
will reset the amplitude to 0.0 mA for the affected program.
).
),
A71200 English 2022-01-15
24
Adjusting Stimulation Parameters
Assigning amplitude, pulse width,
and program rate
Amplitude, pulse width, and program rate are assigned on the
Program subscreen.
During a programming session, the selected values for amplitude,
pulse width, and program rate are automatically programmed into the
neurostimulator. Once amplitude is set and stimulation is turned on,
the neurostimulator changes stimulation to the new setting every time
a stimulation parameter is changed.
Caution: To prevent possible uncomfortable or
unexpected stimulation (jolting or shocking sensation) during a
programming session, adjust stimulation parameters in small
increments above the perception threshold (the parameter
values at which the patient rst perceives a sensation of
stimulation).
Caution: Changing the pulse width or rate in a program
may affect AdaptiveStim therapy. Test each of the AdaptiveStim
postures in the group for appropriate stimulation after making
pulse width or rate changes. Refer to "Testing AdaptiveStim
settings" on page 42.
Assigning amplitude
Amplitude is the strength of the pulse in milliamps (mA). The
electrode conguration for the leads must be assigned before you can
set amplitude.
1. In the Program subscreen, tap AMPLITUDE to access the
amplitude control.
2. Select the amplitude using any of these controls:
■Scrollable drop-down list.
■Dial (press and drag the dot on the dial).
■Back arrows (
increase.
Note: Amplitude increases in increments. When amplitude is
decreased, amplitude decreases immediately to the new value.
3. Use these buttons, as needed:
■Tap the Stop button (
increasing to the new value.
■Tap the Jump button (
without incremental increases.
■Tap the Amplitude to zero button (
set amplitude to 0.0 mA.
) to decrease and forward arrows ( ) to
) to stop stimulation from
) to jump to the new value
) to immediately
Assigning pulse width
Pulse width is the duration of the pulses in microseconds (μs).
Caution: To prevent possible uncomfortable or
unexpected stimulation (jolting or shocking sensation),
decrease the amplitude to the perception threshold (the
amplitude at which the patient rst perceives a sensation of
stimulation) before changing the pulse width. After changing
the pulse width, slowly increase the amplitude.
1. In the Program subscreen, tap PULSE WIDTH to access the
pulse width control.
2. Select the pulse width using any of these controls:
■Scrollable drop-down list.
■Dial (press and drag the dot on the dial).
■Back arrows (
increase.
) to decrease and forward arrows ( ) to
A71200 English 2022-01-15
25
Adjusting Stimulation Parameters
3. If stimulation is on, the Stop and Jump buttons become
available when the pulse width is increased. Use these
buttons, as needed.
■Tap the Stop button (
increasing to the new value.
■Tap the Jump button (
without incremental increases.
) to stop stimulation from
) to jump to the new value
Assigning program rate and group rate
Rate is the frequency of pulses per second, in hertz (Hz). For the
Vanta neurostimulator, which supports more than one program in a
group, the group rate is the initial rate for each program in that group.
A program rate is a fractioned value of the group rate.
To set a group rate
1. In the Program subscreen, tap PROGRAM RATE to access
the rate control.
2. Select the group rate using any of these controls:
■Scrollable drop-down list.
■Dial (press and drag the dot on the dial).
■Back arrows (
increase.
To set a program rate
A program rate is a fractioned value of the group rate.
The rate ratios available: 1/1, 1/2, 1/3, 1/4, and so on, up to 1/20.
1. If you need to set a program rate, tap the pop-up scrollable list
next to PROGRAM RATE in the rate control.
) to decrease and forward arrows ( ) to
2. Select a program rate. This rate will only be applied to the
selected program.
Program rate example:
■The group rate is set at 100 Hz.
■1/2: 50 Hz is selected as the program rate for that
program in the group.
■This program will now deliver pulses at 50 pulses per
second. Meanwhile, the other programs set with the
group rate will still deliver 100 pulses per second.
3. Tap CLOSE to close the control.
4. Reassess limits assigned in the Patient Access subscreen
after modifying stimulation parameters. Refer to "Limits" on
page 31.
Using electrode redistribution (assigning
percentage of amplitude for individual
electrodes)
In most programming sessions, the value selected for the amplitude is
distributed evenly among the assigned electrodes. Use the electrode
redistribution feature when you want to assign a lower or a higher
percentage of the amplitude to an individual electrode in the program.
Note: This feature is not available for the Model 977005 Sequentia LT
neurostimulator.
On the Program subscreen, the distribution of the amplitude among
the electrodes is shown as percentages for the positive (+) electrodes
and as a value for the negative (–) electrodes.
Note: The Device Electrode Redistribution feature must be
activated before you can use the feature. The feature only needs to
be activated once. It remains active across programming sessions.
A71200 English 2022-01-15
26
Adjusting Stimulation Parameters
1. To activate the Device Electrode Redistribution feature:
a. Tap the Display All (
subscreen. The Display All subscreen can only be
accessed from an active program with at least one
electrode pair.
b. In the Display All subscreen, check the status of the
DEVICE ELECTRODE REDISTRIBUTION toggle. If the
toggle is green and the switch is to the right (
feature is activated. If the toggle is blue and the switch
is to the left, tap the toggle to activate the feature.
c. Tap the CLOSE button to return to the Program
subscreen.
2. In the Program subscreen, tap the electrode you intend to
adjust.
Notes:
■To maintain the overall balance of energy between
the polarities, the electrodes that will be impacted by
the redistribution will have the same polarity as the
electrode that you selected to adjust.
■Electrode redistribution will automatically rebalance
the delivered stimulation amplitudes. After making
adjustments, review the electrode contribution
percentages.
3. Select the percentage of the amplitude that you want to assign
to the electrode.
Notes:
■Electrode amplitude increases in increments.
■If the neurostimulator is unable to deliver the amplitude
value in combination with the values set for the
pulse width and rate, an alert will notify you that
) button in the Program
), the
stimulation output is below the programmed amplitude.
Refer to "Out of regulation (OOR) message" in the
Troubleshooting section on page 53.
4. Repeat if you want to adjust another electrode.
5. Tap CLOSE to close the control.
Equalizing amplitude across electrodes
Use the Equalize feature to evenly distribute amplitude across the
assigned electrodes. You can equalize amplitude for a single program
or for all programs with unequal distribution.
Note: This feature is not available for the Model 977005 Sequentia LT
neurostimulator.
To equalize a single program
1. In the Program subscreen, tap the Equalize (
2. Tap YES in the conrmation pop-up to proceed.
3. Use the AMPLITUDE control to select an amplitude.
To equalize all programs with unequal distribution
1. In the Program subscreen, tap the Display All (
The Display All subscreen can only be accessed from an
active program with at least one electrode pair.
2. In the Display All subscreen, tap the DEVICE ELECTRODE REDISTRIBUTION toggle.
3. Tap DISABLE in the conrmation pop-up.
This action equalizes amplitude across all programs with
unequal distribution and reduces the amplitude to zero for
those programs. It also disables the electrode redistribution
and IntelliStim features.
) button.
) button.
A71200 English 2022-01-15
27
Adjusting Stimulation Parameters
4. Tap the CLOSE button to return to the Program subscreen.
Using IntelliStim
IntelliStim is a scanning process in which common combinations of
electrodes are scanned across the lead using the selected amplitude.
IntelliStim can help nd which electrodes to select to deliver therapy.
Note: This feature is not available for the Model 977005 Sequentia LT
neurostimulator.
When IntelliStim begins, the lead is automatically populated with a
negative (–) and positive (+) electrode pair:
■Electrode number 0 is assigned a negative (–) polarity.
■Electrode number 1 is assigned a positive (+) polarity.
You can either start the scanning process from the 0 and 1 electrodes,
or you can move the negative and positive electrode pair to another
position on the lead and start the scanning process from there.
To use IntelliStim, you must rst activate the Device Electrode
Redistribution feature. The feature only needs to be activated once.
It remains active across programming sessions.
1. To activate the Device Electrode Redistribution feature:
a. In the Program subscreen, tap the Display All button.
The Display All subscreen can only be accessed from
an active program with at least one electrode pair.
b. In the Display All subscreen, check the status of the
DEVICE ELECTRODE REDISTRIBUTION toggle. If the
toggle is green and the switch is to the right (
feature is activated. If the toggle is blue and the switch
is to the left, tap the toggle to activate the feature.
c. Tap the CLOSE button to return to the Program
subscreen.
), the
2. In the Program subscreen, tap the INTELLISTIM button (
and follow the on-screen instructions.
3. Position the electrode pair on the lead where you want to start
the scanning process.
■Tap the Forward button (
pair down the lead.
■Tap the Back button (
up the lead.
■Tap the Jump Forward button (
electrode pair to the top of the lead or electrode column
(if a surgical lead) on the right.
■Tap the Jump Back button (
pair to the top of the lead or electrode column (if a
surgical lead) on the left.
■Or move the electrode pair directly to another location
on the lead by tapping a target electrode on the lead.
(This option becomes unavailable once the Play button
is pressed for the rst time.)
4. Use the Amplitude control to set the amplitude.
■If the neurostimulator is unable to deliver the amplitude
value in combination with the values set for the
pulse width and rate, an alert will notify you that
stimulation output is below the programmed amplitude.
Refer to "Out of regulation (OOR) message" in the
Troubleshooting section on page 53.
5. Tap the Play and Pause buttons to start and pause the
scanning process.
■The X2 button is the default speed. Tap the X1 button
for a slower speed than the default speed.
) to advance the electrode
) to advance the electrode pair
) to move the
) to move the electrode
)
A71200 English 2022-01-15
28
Adjusting Stimulation Parameters
6. Ask the patient to provide their feedback during the scanning
process to assess which electrodes to select to deliver therapy.
7. Tap DONE if you nd an effective electrode combination.
8. If desired, use IntelliStim to create another program in the
same group for the other lead or other electrode column (if
programming a surgical lead).
Relative Lead Positioning
Use the Lead Manipulation subscreen (accessed from the Program
subscreen) to represent the relative positioning of the implanted leads
as viewed from the patient’s back. The subscreen is only applicable if
more than one lead is congured.
■To swap the relative position of the leads (for example, to
switch the position of lead #2 with lead #1), press and hold,
then drag a lead to the location of the other lead.
Repositioning the leads relative to each other on the Lead
Manipulation subscreen updates how the leads are shown on the
Programs screen and Program subscreen.
The uoro screen can be accessed using the View Fluoro button at
the bottom-left of the Lead Manipulation subscreen (it also appears
at the bottom-left of the Tip Location screen). Refer to "Adding uoro
images" on page 23 for more information.
A71200 English 2022-01-15
29
Relative Lead Positioning
Customizing SoftStart/Stop and
Cycling
Use the Energy subscreen (accessed from the Program subscreen)
to customize the optional device settings for the SoftStart/Stop and
Cycling features. These features are designed to increase patient
comfort and ease of use.
Turning cycling on or off
Cycling turns stimulation on and off at clinician-determined intervals.
Due to a carryover effect, the patient may continue to experience
symptom suppression during the cycling off time. The default setting
for the cycling feature is “off.”
1. In the Energy subscreen, tap the CYCLING toggle to turn
cycling on (
) or off ().
Changing SoftStart/Stop settings
SoftStart/Stop slowly increases the amplitude when stimulation is
turned on and slowly decreases the amplitude when stimulation is
turned off. The slow ramping may feel more comfortable to sensitive
patients.
SoftStart/Stop is assigned at the device level: that is, it is either on
or off for all groups. The SoftStart/Stop feature defaults to “on” with a
duration of 4 seconds.
■In the Energy subscreen, use the SOFTSTART/STOP drop-
down list to select a duration or to turn SoftStart/Stop off.
If AdaptiveStim Technology and SoftStart/Stop are both enabled
and the amplitude increases or decreases because of a change in
position:
■The amplitude will slowly increase using the ramping
programmed for SoftStart/Stop.
■The amplitude will decrease immediately without ramping.
2. If cycling is on, select the durations for the ON TIME and OFF TIME intervals.
3. Tap the group name toggle to turn cycling on (
(
) for a specic group. The setting affects all programs in
the group.
4. Tap UPDATE to apply the changes.
Note: Use the BATTERY LONGEVITY function in the Energy
subscreen to see how cycling affects battery longevity.
) or off
A71200 English 2022-01-15
30
Customizing SoftStart/Stop and Cycling
Estimating Battery Longevity
Battery longevity for the INS depends on the therapy conguration
and usage. The Estimated Battery Longevity feature is a calculator
that allows you to estimate battery longevity based on the current
program and group settings, along with the number of hours per day
the patient might use each group.
Note: This feature estimates battery longevity based on expected
usage. To see a battery life estimate based on previous actual usage
(the last seven days), refer to the Estimated Battery Remaining (Based on Last 7 Days) eld on the Device screen�
To estimate battery longevity using the Estimated Battery Longevity
feature:
1. In the Energy subscreen (accessed from the Program
subscreen), tap the ESTIMATE button.
2. For each group, press and drag the dot on the slider to set
hours per day of group usage.
■Use the GROUP drop-down lists to view the programs
within each group.
■Group usage can be set to a total of 24 hours among all
groups.
3. Tap ESTIMATE.
4. Review the estimates. Estimates are shown for three different
impedance ranges.
5. Tap CLOSE to return to the Energy subscreen.
Note: Battery longevity should be re-estimated if changes are made
to programming.
Enabling Patient Access
Use the Patient Access subscreen (accessed from the Program
subscreen) to assign patient controls, if appropriate for the patient.
There are two patient controls on the Patient Access subscreen:
Limits and Adjustment.
Note: The Adjustment patient control is not available for the
Model 977005 Sequentia LT neurostimulator.
Adjustment
Enabling the Adjustment patient control on the Patient Access
subscreen allows the patient to increase or decrease amplitude for all
programs in the active group at the same time.
■Under the ADJUSTMENT heading, tap the ALL PROGRAMS TOGETHER toggle to enable (
patient control.
Limits
The Limits patient control on the Patient Access subscreen
provides the ability to enable or disable patient control of stimulation
parameters (amplitude, pulse width, group rate). When patient control
is enabled for a stimulation parameter, the patient can adjust that
stimulation parameter using the patient programmer.
If patient control is enabled, the upper patient limit for the stimulation
parameter must be assigned. The lower patient limit is set to the
lowest possible value by default and is not customizable. Patient
control of stimulation parameters is assigned at the group level.
Note: By default, patient control of amplitude is enabled, and patient
control of pulse width and group rate is disabled.
1. Next to the LIMITS heading, make sure that the group you
intend to congure is shown.
) or disable () the
A71200 English 2022-01-15
31
Estimating Battery Longevity
2. Tap the toggle next to the stimulation parameter name to
enable (
parameter.
3. For enabled stimulation parameters, set the upper patient limit
using any of these controls:
■Scrollable drop-down list.
■Dial (press and drag the dot on the dial).
4. After the patient controls are dened, tap UPDATE.
Amplitude
■The lowest allowable option for the upper patient limit is
determined by the highest amplitude setting in the group's
existing programs.
■When AdaptiveStim Technology is enabled for a group, patient
control is automatically enabled for amplitude and cannot be
disabled for that group.
Pulse width
■The lowest allowable option for the upper patient limit is
determined by the highest pulse width setting in the group's
existing programs.
■When AdaptiveStim Technology is enabled for a group, patient
control is not available for pulse width.
Group rate
■The lowest allowable option for the upper patient limit is
determined by the current rate setting for the group.
Effect of subsequent parameter changes
■During a subsequent programming session, if a parameter is
increased above the upper patient limit, the upper patient limit
automatically changes to that new value.
) or disable () patient control of the stimulation
Checking System Performance
The measurement functions on the Impedance screen assist in
identifying problems with components of the implanted system.
Measurements and diagnostic data obtained from the clinician
programmer app are intended to aid in your clinical assessment.
However, as with any electronic system, internal and external factors
can inuence neurostimulator measurements. For example, changes
in lead position can affect the stimulation current or the impedance
measurement. If you obtain a reading that seems inconsistent with
your observations, repeat the measurement.
Note: Measure impedance at the beginning of each programming
session. These measurements verify the integrity of lead, extension,
and connector pathways. They may provide information about lead
problems (such as lead breakage, short circuit, open circuit). For
example, measurements that show a signicant increase in electrode
impedance can indicate a fractured lead conductor or a loose
setscrew. Conversely, a signicant decrease in electrode impedance
can indicate shorted conductors or a break in lead insulation.
Measurements taken at the beginning of the session may be useful in
interpreting diagnostic data collected since the previous session.
Testing impedance
The Impedance screen is accessed through the workow navigator.
Note: The Impedance screen can be accessed in the Side Menu
(
) when it is not part of a workow.
1. On the Impedance screen, tap the MEASURE ALL IMPEDANCES button (
2. Tap the START button on the pop-up screen.
3. After the test is complete, review the impedance results.
).
A71200 English 2022-01-15
32
Checking System Performance
Notes:
■If all impedance values are within range, the message "All
impedances are within normal range" appears in the GOOD
box.
■If any impedance value is out of normal ranges (too low or too
high), the results appear in either the AVOID box or the DO NOT USE box. Refer to "Troubleshooting Electrode Impedance
Results" on page 33.
■The impedance value shown for each electrode is measured
with respect to the selected reference electrode. Use the
REFERENCE ELECTRODE scrollable drop-down list to select
a different reference electrode, if applicable.
■Once an impedance test is performed, results can be viewed
at any time during the programming session by returning to
the Impedance screen. You can also view or print the Session
Report, which shows the impedance measurements.
Troubleshooting Electrode
Impedance Results
The AVOID and DO NOT USE boxes on the Impedance screen
identify where low impedance or high impedance was measured in
the system. Refer to Table 6 for more information about impedance
results indicators.
Table 6. Impedance results indicators.
IconDescription
GOOD (green electrodes with a check mark): Indicates
that the impedance measurement value for the electrode
is within the normal range (300 - 4,000 ohms).
AVOID (orange electrodes with an exclamation point):
Possible open circuit:
• Using these electrodes may increase energy
consumption or impact therapy efcacy.
• Indicates that a high impedance (4,000 - 40,000 ohms)
has been detected between electrodes. The Avoid icon
is displayed and adjacent to the number of the electrodes
where high impedance has been detected.
Possible short circuit:
• Indicates that a low impedance (under 300 ohms) has
been detected between electrodes.
• The electrode combinations where possible shorts are
detected are listed under the text. The electrodes in these
combinations will also be listed in the Avoid list.
DO NOT USE (red electrodes with an "X"): Indicates that
there is a strong likelihood of an open circuit (over 40,000
ohms) in the system.
Refer to the troubleshooting details for high impedance and low
impedance on the following pages. For further assistance with
A71200 English 2022-01-15
33
Troubleshooting Electrode Impedance Results
troubleshooting electrode impedance results, contact Medtronic using
the contact information listed on the back cover of this manual.
Intra-operative troubleshooting
High impedance troubleshooting
High impedance suggests a possible open circuit in the system. High
impedance could be caused by a broken wire, loose connection, or
highly resistive tissue.
High impedance: >4,000 ohms
1. Analyze electrode impedance measurement results. An outlier
measurement may indicate an issue with a setscrew or wire.
Ensure that all leads and extensions are fully inserted and
setscrews are properly tightened and measure impedance
again.
For more information about measuring electrode impedance,
refer to “Testing impedance” on page 32.
2. If a trend towards normal impedance range is not observed,
disconnect the lead from the extension (if applicable) and
the neurostimulator. Reconnect to the Wireless External
Neurostimulator (WENS) and use the A71300 Stimulation
Trialing clinician programmer application to test the impedance
of the lead alone.
■Refer to the A71300 Stimulation Trialing Clinician Programmer Application programming guide for
instructions for measuring impedance with the WENS.
■If the impedance value is between 300 and 4,000 ohms,
then the lead is intact.
■If the impedance value is greater than 4,000 ohms,
visually or uoroscopically inspect the lead for damage.
3. If lead is intact, reconnect the lead to the extension (if
applicable) and use the WENS to test the impedance of the
lead and extension together.
■If the extension impedance value is between 300 and
4,000 ohms, the extension is intact. If the extension
impedance value exceeds 4,000 ohms, visually or
uoroscopically inspect the extension and extension
connection for damage.
■If two extensions are being used, switch the extensions
with one another and measure impedance again to see
if the out-of-range impedance value is specic to one of
the extensions. If the out-of-range impedance value is
specic to one of the extensions, consider replacing that
extension.
4. Reconnect the neurostimulator to the lead and extension (if
applicable). Ensure that the neurostimulator is in the pocket
and retest the system. If the impedance value remains high,
visually or uoroscopically inspect the connection between
the neurostimulator and the lead (or the connection between
the neurostimulator and the extension, if applicable) and the
neurostimulator connector block for damage.
■If the neurostimulator impedance value is between 300
and 4,000 ohms, the neurostimulator is intact.
■If the neurostimulator impedance value exceeds
4,000 ohms, visually or uoroscopically inspect the
neurostimulator and neurostimulator-to-lead (or
extension, if applicable) connection for damage.
■If two neurostimulators are being used, swap the
neurostimulators with one another and measure
impedance again to see if the out-of-range impedance
value is specic to one of the neurostimulators. If
the out-of-range impedance value is specic to one
A71200 English 2022-01-15
34
Troubleshooting Electrode Impedance Results
of the neurostimulators, consider replacing that
neurostimulator.
5. If the previous steps fail to reveal the underlying issue, assess
whether it is possible to achieve therapy without using the
electrodes with high impedance.
6. If troubleshooting measures fail to resolve the high impedance
values, consider substituting new system components as
needed.
Low impedance troubleshooting
Low impedance suggests a possible short circuit in the system that
can lead to inadequate therapy.
Low impedance: <300 ohms
Avoid using electrodes with a low impedance. Assess whether it
is possible to achieve therapy without using the electrodes with
low impedance. If the low-impedance electrodes are necessary for
programming, revision surgery may be needed.
Post-implant troubleshooting
Caution: When troubleshooting issues related to
impedance measurements, follow these instructions correctly
and completely to identify and resolve the issue.
■Mistaking an undamaged component as damaged could
result in unnecessary surgical revision.
■Failure to identify a damaged component could result in
unresolved intermittent or loss of stimulation.
High impedance troubleshooting
High impedance suggests a possible open circuit in the system. High
impedance could be caused by a broken wire, loose connection, or
highly resistive tissue.
High impedance: >4,000 ohms
Follow these steps to troubleshoot high impedance values:
1. If impedance values are high (>4,000 ohms), consider these
troubleshooting options:
■Assess whether therapy can be achieved using other
electrodes.
■Check for incorrect information. For example, check the Lead Select screen to ensure the electrode numbering
on the screen matches the electrode numbering of the
implanted lead(s).
■Consider using x-ray to inspect the system components
for improper connections or damage.
2. If troubleshooting measures fail to resolve the high impedance
values, consider replacing system components as needed.
Low impedance troubleshooting
Low impedance suggests a possible short circuit in the system that
can lead to inadequate therapy.
Low impedance: <300 ohms
Avoid using electrodes with a low impedance. Assess whether it
is possible to achieve therapy without using the electrodes with
low impedance. If the low-impedance electrodes are necessary for
programming, revision surgery may be needed.
A71200 English 2022-01-15
35
Troubleshooting Electrode Impedance Results
Preparing System for an MRI
Scan
Refer to the MRI Guidelines for Medtronic Neurostimulation Systems
for Chronic Pain instructions for use manual for the MRI conditions
and MRI-specic warnings and cautions for conducting an MRI scan.
Always obtain the latest MRI guidelines. Refer to the contact
information at the back of this manual, or go to www.medtronic.
com/mri.
When the MRI Mode feature is activated, stimulation is turned off.
Stimulation must be off during the MRI scan.
To check MRI scan eligibility and
enter MRI Mode
The information in the View MRI workow impacts the MRI scan
eligibility shown on the MRI screen.
1. On the CURRENT DEVICE STATUS screen, conrm that the
patient and device information corresponds with the intended
patient and INS.
2. Tap VIEW MRI > START.
3. You are now in the View MRI workow.
Note: You can also access the MRI screen on the Side Menu
(
) when not part of a workow.
4. On the Device screen, conrm that the following information is
complete and accurate:
■Neurostimulator model number
■Neurostimulator implant location
5. On the Patient Info screen, conrm that the patient information
is complete and accurate.
6. On the Lead Select screen, conrm that the lead model
numbers are accurate.
7. On the Tip Location screen, conrm that the information for
the following components is complete and accurate:
■Presence of abandoned leads, extensions, and pocket
adaptors
■Lead tip location
8. On the MRI screen, review the MRI scan eligibility. You will
see one of several possible eligibility outcomes, with icons and
symbols. Explanations of the outcomes and their icons are
described in Table 7.
Do not proceed if eligibility cannot be determined, or if the
system is designated as MR Unsafe. Contact your Medtronic
representative, if needed.
Table 7. MRI eligibility icons
Icons shownExplanation
Full-body scan eligible — The implanted
neurostimulation system is eligible to
have MRI scans of any part of the body,
including a head scan, using specic
conditions. The MRI clinician must consult
the MRI guidelines for those conditions.
Head scan eligible with transmit/
receive head coil — The implanted
neurostimulation system is eligible for
MRI scans of the head only using an RF
transmit/receive head coil and under other
specic conditions. The MRI clinician
must consult the MRI guidelines for those
conditions.
A71200 English 2022-01-15
36
Preparing System for an MRI Scan
The neurostimulation system MRI
scan eligibility cannot be conrmed —
The MRI clinician must consult the MRI
guidelines to determine how to proceed or
contact Medtronic technical support.
MR Unsafe: Patients implanted with
a neurostimulation system that is
designated as MR Unsafe on the MRI
screen are ineligible for MRI scans.
Therefore, activating MRI Mode is not
applicable for MR Unsafe systems.
9. To prepare the device for an MRI scan, set the MRI MODE
toggle to Activated (
).
10. Conrm that the message “Device ready for MRI scan”
appears above the MRI MODE toggle and the MRI status
indicator (
) displays in the action bar.
11. An MRI report can be generated in the Reports screen
by selecting MRI Report from the drop-down list. Refer to
"Working with Reports" on page 49 for details about viewing
and printing reports.
Note: If you cannot print your reports, you can manually
complete the MRI Eligibility Form in the MRI Guidelines
for Medtronic Neurostimulation Systems for Chronic Pain
instructions for use manual.
12. Inform the patient of the following:
■When the MRI Mode feature is activated, stimulation is
turned off. Stimulation must be off during the MRI scan.
■Because stimulation is off, pain symptoms may return.
■Bring the most up-to-date patient ID card to all MRI
appointments.
■Bring the patient programmer to all MRI appointments.
■After the MRI scan is complete and the patient is outside
of the magnet room, stimulation can be turned back
on using the patient programmer. Or, if appropriate,
the patient can revisit the clinician after the MRI scan
to have stimulation turned back on with the clinician
programmer app.
13. Give the MRI Report to the patient or send the report to the
MRI center.
The MRI clinician uses the MRI Report to conrm the
neurostimulator model, session date, patient information, and
MRI scan eligibility of the patient’s implanted system.
To deactivate MRI Mode
There are two different ways to deactivate MRI Mode and restart
stimulation:
1. Start a workow with the INS. If the MRI status indicator (
displays in the action bar, stimulation will be off.
2. Tap the Stimulation toggle in the action bar to turn stimulation
on and deactivate MRI Mode.
3. A pop-up message will appear that says turning stimulation on
will deactivate MRI Mode. Tap STIM ON.
Or
1. From the MRI screen, tap the Deactivated/Activated toggle.
2. A pop-up message will appear that says deactivating
MRI Mode will not turn on stimulation automatically. Tap
DEACTIVATE.
3. Tap the Stimulation toggle in the action bar to turn stimulation
on.
)
A71200 English 2022-01-15
37
Preparing System for an MRI Scan
AdaptiveStim Technology
The Model 977006 Vanta neurostimulator with AdaptiveStim
Technology can be programmed to adjust amplitude automatically
when the patient changes body position. Amplitude settings can be
assigned to the following positions:
■Upright
■Mobile
■Reclining
■Lying Front
■Lying Back
■Lying Right
■Lying Left
When AdaptiveStim Technology is turned on for a patient, the
neurostimulator automatically adjusts the amplitude based on the
patient's body position and the clinician-programmed settings. If the
patient makes an amplitude adjustment with the patient programmer,
the neurostimulator associates the new amplitude with the patient's
body position as long as the patient stays in that position for at
least three minutes. The next time the patient enters that position,
the neurostimulator will return to the new amplitude. In this way,
AdaptiveStim Technology can respond to patient feedback.
Note: During a programming session, AdaptiveStim settings are not
active.
To accommodate the individual needs of the patient, the following
AdaptiveStim settings can be adjusted by the clinician:
■Position angles: The range or area of a position.
■Transition times: The amount of time that must elapse after
moving into a new position (and remaining in the new position)
before the amplitude will change.
■Mobility rate: The intensity of movement that is considered a
change from the stationary position of Upright to the moving
position of Mobile.
Caution regarding cervical location
Caution: The use of AdaptiveStim Technology associated with
position changes in the cervical location has not been shown to be
safe and effective. The use of AdaptiveStim Technology associated
with position changes for locations outside of the thoracic or lumbar
spine may result in unintended stimulation.
Caution regarding healing period
Caution: Do not congure AdaptiveStim Technology during
the implant procedure. A four-week healing period is recommended
after implanting the neurostimulator to allow time for the implanted
components to stabilize. If the tissue around the neurostimulator is
not sufciently healed, movement of the neurostimulator is possible
and may affect orientation.
Conguring AdaptiveStim
When you congure AdaptiveStim Technology with the setup wizard,
AdaptiveStim Technology is enabled for all programmed groups
and every position, including Mobile, no matter what settings were
previously assigned to those groups.
When a group is AdaptiveStim-enabled, the AdaptiveStim indicator (
appears next to the group name on the Programs screen, and patient
access for pulse width is not available for that group:
Note: Amplitude limit is automatically enabled when AdaptiveStim
Technology is congured for a group and cannot be disabled for that
group.
)
A71200 English 2022-01-15
38
AdaptiveStim Technology
1. Go to the Programs screen, and tap a group name to activate
the group you want to congure.
2. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Congure.
),
2. Select AdaptiveStim in the workow, and tap CONFIGURE to
start the setup wizard.
3. Follow the on-screen instructions for all applicable positions.
■On the orientation screen, have the patient get in the
position and tap NEXT to complete the orientation.
Note: You can tap SKIP to bypass certain positions.
Orientation still occurs for the position and a position
amplitude of zero is assigned to that position.
■On the programming screen, tap the program number
you want to congure, and then set the position
amplitude using the amplitude control. Refer to
"Assigning amplitude" on page 25.
■If applicable, tap another program number, and then set the position amplitude. When nished, tap NEXT to go
to the next position.
4. A conrmation screen will appear when AdaptiveStim Technology has been congured for the group. Tap CLOSE to
go to the AdaptiveStim screen.
Reconguring an AdaptiveStimenabled group
If you want to recongure an AdaptiveStim-enabled group, complete
these steps to delete the existing AdaptiveStim settings and to
reassign position amplitudes for that group. You also have the option
to reorient patient positions. To reorient only, refer to "Reorienting
patient positions" on page 43.
1. Go to the Programs screen, and tap a group name to activate
the group you want to recongure.
3. Tap Start Over.
4. (Optional) Tap the Reorient checkbox if you also want to
reorient the neurostimulator to the patient positions.
5. Tap Yes to proceed.
6. Follow the on-screen instructions to reorient the
neurostimulator to the patient positions (if applicable) and to
reassign position amplitudes for the programs in the active
group.
7. When the wizard is nished, a conrmation screen displays
stating that you have successfully recongured AdaptiveStim
Technology for the group.
8. Tap CLOSE to go to the AdaptiveStim screen.
What you can do on the
AdaptiveStim screen
The AdaptiveStim screen displays when an AdaptiveStim-enabled
group is the active group and you select AdaptiveStim in the
workow. The AdaptiveStim screen can also be accessed by tapping
AdaptiveStim on the Side Menu (
Refer to Table 8 on page 40 for information about the tasks you can
complete on the AdaptiveStim screen.
Note: Some tasks may not be available until AdaptiveStim
Technology has been congured.
) when not part of a workow.
A71200 English 2022-01-15
39
AdaptiveStim Technology
Table 8. Tasks on the AdaptiveStim screen.
Tasks on the AdaptiveStim screenMore information
Review position amplitudes for all AdaptiveStim-enabled groups.
Use this feature to view a summary of the position amplitudes for all AdaptiveStim-enabled groups and programs.
Test AdaptiveStim settings.
Use this feature to test AdaptiveStim settings during a programming session.
Adjust position amplitudes for individual positions.
Use this feature to adjust position amplitudes for individual positions in the active group.
Check patient positions.
Use this feature to determine if the neurostimulator correctly detects the patient positions.
Reorient patient positions.
Use this feature to reorient the neurostimulator to the patient positions without changing any group settings.
Recongure an AdaptiveStim-enabled group.
Use this feature to delete all existing AdaptiveStim settings for a group and to reassign position amplitudes. You also
have the option to reorient patient positions.
Disable AdaptiveStim Technology for a group.
Use this feature to clear all existing AdaptiveStim settings for a group. All other AdaptiveStim-enabled groups are not
impacted.
Enable or disable AdaptiveStim Technology for the patient.
If the feature is enabled, AdaptiveStim Technology and all assigned settings will be on when the programming session
ends.
If disabled, AdaptiveStim Technology and all assigned settings will be off when the programming session ends.
However, the patient can turn on AdaptiveStim Technology using the patient programmer.
Turn the Position Diary on or off.
The Position Diary feature enables the Position Trend chart to be populated in the Diaries screen. The Position
Trend chart shows the time the patient spent in each position during the periods between the last four programming
sessions.
Setting stability time
Stability time is the amount of time that the patient needs to be in a position before that new intensity for that position
will be saved.
Page 42
Page 42
Page 42
Page 43
Page 43
Page 39
Page 43
Page 43
Page 43
Page 44
A71200 English 2022-01-15
40
AdaptiveStim Technology
Tasks on the AdaptiveStim screenMore information
Adjust position angles.
Position angle is the range or area of a position and is used to ensure the neurostimulator properly recognizes the
Upright position from the Lying positions.
Adjust transition times.
Transition time is the amount of time that must elapse between changes in position before the amplitude will change.
Adjust mobility rate.
Mobility rate is the intensity of movement that is considered a change from the stationary position of Upright to the
moving position of Mobile.
Turn off mobility detection.
Turning off mobility detection disables the Mobile position for AdaptiveStim Technology.
Troubleshoot AdaptiveStim Technology.
Table 9 lists potential issues that can occur with AdaptiveStim Technology, as well as possible solutions for resolving
those issues.
Page 44
Page 44
Page 45
Page 45
Page 45
A71200 English 2022-01-15
41
AdaptiveStim Technology
Reviewing position amplitudes for
all AdaptiveStim-enabled groups
A summary of the position amplitudes for all AdaptiveStim-enabled
groups and programs can be viewed from the AdaptiveStim screen.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Overview.
Adjusting position amplitudes for
individual positions
Complete these steps to adjust position amplitudes for individual
positions in a group.
),
1. Go to the Programs screen, and tap a group name to activate
the group you want to adjust.
2. Review the position amplitudes assigned to the AdaptiveStimenabled groups. If all groups and programs are not visible, you
may need to swipe left or right to see all of the data.
3. Tap CLOSE to return to the AdaptiveStim screen.
Testing AdaptiveStim settings
During a programming session, AdaptiveStim settings are suspended.
The Test AdaptiveStim feature can be used to test AdaptiveStim
settings in session and to obtain patient feedback. There must be at
least two different position amplitudes dened in a program to run the
test.
1. Go to the Programs screen, and tap a group name to activate
the group you want to test.
2. Go to the AdaptiveStim screen, and tap the TEST ADAPTIVESTIM button (
3. Follow the on-screen instructions to test the AdaptiveStim
settings and obtain patient feedback.
4. Tap STOP to end the test.
).
2. Go to the AdaptiveStim screen, and tap the position you want
to adjust.
On the screen, you see the current programs in the group.
Within each program box, the program amplitude is displayed
in white and the position amplitude is displayed in green. A
white box around a program identies the active program.
3. Tap the program number you want to adjust.
4. Use the amplitude control to adjust the program amplitude
(the stimulation the patient is receiving). Refer to "Assigning
amplitude" on page 25.
The white dot represents the program amplitude. As you make
amplitude changes, the white dot moves to reect the current
amplitude.
Note: You can also drag the white dot to assign amplitude.
Stimulation ramps up to the amplitude value.
5. When you nd an acceptable amplitude, tap SET POSITION AMPLITUDE to assign the stimulation the patient is receiving
as the position amplitude.
The blue dot represents the position amplitude.
6. When nished making adjustments, tap CLOSE to return to the
AdaptiveStim screen.
A71200 English 2022-01-15
42
AdaptiveStim Technology
Checking patient positions
Complete these steps to determine if the neurostimulator correctly
detects the patient positions.
1. Place the patient in the position you intend to check.
2. Go to the AdaptiveStim screen, and tap the CHECK POSITION button (
3. Conrm that the position shown on the AdaptiveStim screen
corresponds to the patient's body position.
).
Reorienting patient positions
Complete these steps to reorient the neurostimulator to the patient
positions without changing any group settings.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Congure.
2. Tap Reorient Only.
3. Tap Yes to clear the current orientation.
4. Follow the on-screen instructions to reorient the
neurostimulator to the patient positions.
5. (Optional) Tap SKIP if you want to bypass a position.
6. When all positions have been reoriented, a conrmation screen
displays.
7. Tap CLOSE to go to the AdaptiveStim screen.
Disabling AdaptiveStim Technology
for a group
Complete these steps to clear all existing AdaptiveStim settings for a
group. All other AdaptiveStim-enabled groups will not be impacted.
1. Go to the Programs screen, and tap a group name to activate
the group.
2. Go to the AdaptiveStim screen, and tap the on/off toggle next
to the Settings button (
3. Tap UPDATE, and then conrm the disable notication.
) to enable or disable AdaptiveStim.
Enabling or disabling AdaptiveStim
Technology for the patient
),
You can choose to either enable or disable AdaptiveStim Technology
for the patient. If enabled, AdaptiveStim Technology and all assigned
settings will be on when the programming session ends. If disabled,
AdaptiveStim Technology and all assigned settings will be off when
the programming session ends. However, the patient can turn on
AdaptiveStim Technology using the patient programmer.
1. Go to the AdaptiveStim screen, and tap the RESUME on/off
toggle to enable or disable AdaptiveStim.
2. Tap UPDATE to apply the changes.
Turning the Position Diary on or off
The Position Diary feature enables the Position Trend chart to be
populated in the Diaries screen. The Position Trend chart shows the
time the patient spent in each position during the periods between the
last four programming sessions. The chart is accessed through the
Diaries screen.
A71200 English 2022-01-15
43
AdaptiveStim Technology
AdaptiveStim must be congured for position data to be tracked.
Refer to "Conguring AdaptiveStim" on page 38 for instructions. By
default, the Position Diary is turned on when AdaptiveStim Technology
is initially congured or recongured, or when patient positions are
reoriented. If the Position Diary is turned off, from that point, position
data is not saved. If the Position Diary is on and AdaptiveStim
Technology is turned off, position data continues to be tracked.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Settings.
2. On the Options tab under POSITION DIARY, tap the on/off
toggle. Green (toggle to the right) indicates that the Position
Diary is on.
3. Tap UPDATE to apply the changes.
4. Tap CLOSE to return to the AdaptiveStim screen.
),
from the Lying positions. By default, the Upright position angle is set
to Small and the Lying position angle is set to XX-Large. All Lying
positions (Back, Front, Left, Right) use the same position angle value.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Settings.
2. Tap the Angles tab.
3. On the Angles diagram, tap the position name for the angle
you want to adjust (for example, Upright, Lying Back, or
Lying Front).
4. Press and drag one of the delimiter dots to set a larger or
smaller angle.
5. Make other adjustments to the position angles as needed.
6. Tap UPDATE to apply the changes.
),
Setting stability time
The Stability Time setting is the amount of time that the patient needs
to be in a position (after having set a new intensity value with the
patient programmer) before the intensity value for the position is
saved to the neurostimulator. The default setting is 0 seconds.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Settings.
2. Tap the STABILITY TIME drop-down list and select a duration.
3. Tap UPDATE to apply the changes.
4. Tap CLOSE to return to the AdaptiveStim screen.
),
Adjusting position angles
Position angle is the range or area of a position and is used to ensure
that the neurostimulator properly recognizes the Upright position
A71200 English 2022-01-15
7. Tap CLOSE to return to the AdaptiveStim screen.
Note: To test the changes, refer to "Testing AdaptiveStim settings" on
page 42.
Adjusting transition times
Transition time is the amount of time that must elapse between
changes in position before the amplitude will change. You can adjust
transition times for each position change.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Settings.
2. Tap the Transitions tab.
3. Swipe up and down on the screen to scroll and nd the
position change you want to adjust.
4. Tap to open the scrollable drop-down list and select a duration.
44
AdaptiveStim Technology
),
5. Make adjustments to other transition times as needed.
6. Tap UPDATE to apply the changes.
7. Tap CLOSE to return to the AdaptiveStim screen.
Note: To test the changes, refer to "Testing AdaptiveStim settings" on
page 42.
Turning off mobility detection
Turning off mobility detection disables the Mobile position for
AdaptiveStim Technology. When mobility detection is off, the Mobile
position button is dimmed in the AdaptiveStim screen.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Settings.
),
Adjusting mobility rate
Mobility rate is the intensity of movement that is considered a change
from the stationary position of Upright to the moving position of
Mobile. The mobility rate is only accessible if the Mobile posture is
enabled.
1. Go to the AdaptiveStim screen, tap the Settings button (
and then tap Settings.
2. Tap the Mobility Rate tab.
3. Conrm that the on/off toggle is in the on position. Green
indicates that the Mobile posture is enabled.
4. Press and drag the Mobility Rate dot on the slider.
■Low (L): Takes less motion for the neurostimulator to
detect the Mobile position.
■High (H): Takes more motion for the neurostimulator to
detect the Mobile position.
5. Tap UPDATE to apply the changes.
6. Tap CLOSE to return to the AdaptiveStim screen.
Note: To test the change in mobility rate, refer to "Testing
AdaptiveStim settings" on page 42. To check patient positions, refer
to "Checking patient positions" on page 43.
),
2. Tap the Mobility Rate tab.
3. Tap the on/off toggle. Gray indicates that the Mobile position is
disabled.
4. Tap UPDATE to apply the changes.
5. Tap CLOSE to return to the AdaptiveStim screen.
Troubleshooting AdaptiveStim
Technology
Refer to Table 9 on page 46 for a list of potential issues that can
occur with AdaptiveStim Technology, as well as possible solutions for
resolving those issues.
A71200 English 2022-01-15
45
AdaptiveStim Technology
Table 9. Troubleshooting AdaptiveStim Technology.
ProblemPossible Solutions
The delay between a position
change and the amplitude
change is too slow or too fast for
the patient.
The amplitude changes too often
for the patient.
The amplitude changes to the
amplitude for the Mobile position
while the patient is in the Upright
position but not mobile.
The amplitude does not change
to the amplitude for Mobile when
the patient is moving.
The patient does not like that the
amplitude changes to the Mobile
amplitude when starting to walk
from the Upright position.
Neurostimulator may have
rotated or ipped within the
pocket.
Patient feels the amplitude for
the Upright position when lying
down.
Patient is not getting
adequate therapy in a position
programmed with AdaptiveStim.
■ Use the Transitions setting to increase or decrease the delay between position changes and amplitude
changes.
■ Use the Test AdaptiveStim feature to obtain patient feedback and make adjustments as needed.
■ Use the Angles setting to increase the angle for the position that is changing too often.
Or
■ Use the Transitions setting to increase the delay between position change and amplitude change.
■ Use the Mobility Rate setting to increase the intensity of movement that is required to change to the
amplitude for the Mobile position.
■ Have the patient remain still in the Upright position for at least one minute.
■ Use the Check Position feature to conrm the patient’s position. If “Mobile” is shown instead of
“Upright” for the patient’s current position, increase the mobility rate.
■ Use the Mobility Rate setting to decrease the intensity of movement that is required to change to the
amplitude for the Mobile position.
■ Have the patient walk down the hall for at least one minute.
■ Use the Check Position feature to conrm the patient’s position. If “Upright” is shown instead of
“Mobile” for the patient’s current position, decrease the mobility rate.
■ Turn off the Mobility Rate feature.
■ When the Mobility Rate feature is turned off, the Mobile position is no longer enabled for AdaptiveStim
Technology.
■ Reorient the neurostimulator to the patient positions.
■ Use the Angles setting to increase the angle for the Lying positions.
■ Review position amplitudes, and check that the position amplitude is not set to 0 mA.
A71200 English 2022-01-15
46
AdaptiveStim Technology
ProblemPossible Solutions
Patient unable to adjust their
stimulation rate.
■ Create multiple identical groups by copying the current AdaptiveStim-enabled group. Adjust the
stimulation rate for each group to provide the patient with different stimulation rates with AdaptiveStim
enabled. Provide the patient guidance on how to switch groups.
A71200 English 2022-01-15
47
AdaptiveStim Technology
Viewing Diaries
The charts accessed from the Diaries screen provide overviews of
how the patient is using the neurostimulator. Not all types of diaries
are supported by the Model 977005 Sequentia LT neurostimulator.
Stimulation Usage
The Stimulation Usage chart shows the patient's stimulation usage.
The chart displays the percentage of time per day that stimulation is
on over a 30-day period. The metrics provide the average percentage
of stimulation usage over the last 30 days and the percentage of time
stimulation was on since the last programming session.
Note: If 30 days of stimulation usage data is not available, the
calculations are based on the available number of days.
Group Usage
The Group Usage chart displays the percentage of time that each
group was used since the last programming session.
Adjustments
The Adjustments chart is only applicable to neurostimulators with
AdaptiveStim Technology. The chart shows the number of amplitude
adjustments made by the patient since the last programming session.
Data is shown for each position and AdaptiveStim-enabled group.
Composite
The Composite chart is only applicable to neurostimulators with
AdaptiveStim Technology. The chart shows the number of hours of
stimulation usage for time spent reclining, mobile, upright, and lying
within the last 30 days.
■Tap the toggle to display the graph for a position on the chart.
Multiple positions can be selected to show the position graphs
together. Green indicates that the position has been selected.
■The background graph that is displayed (before any position
graphs are added) represents stimulation usage.
■Select a start time and end time if you want to see usage
during a specic time period.
Position Trend
The Position Trend chart is only applicable to neurostimulators with
AdaptiveStim Technology. The chart shows the time the patient spent
in each position during the periods between the last four programming
sessions. The default view shows all positions and the amount of time
in percentages.
■Select a position from the scrollable drop-down list if you want
to view a specic position only.
■Select Time if you want to see the average amount of time per
day in hours and minutes.
A71200 English 2022-01-15
Resting Trend
The Resting Trend chart is only applicable to neurostimulators with
AdaptiveStim Technology. The chart shows the average number of
position changes from one lying position to another during the periods
between the last four programming sessions.
48
Viewing Diaries
Working with Reports
The following report types are available from the Reports screen:
■Session Report: Contains information about the settings
programmed into the neurostimulator during a programming
session as well as therapy usage and system status
information.
■MRI Report: Contains information intended for MRI clinicians
about the MRI scan eligibility of the implanted system for the
selected patient.
■Medtronic Data Report: Contains information about the
Medtronic data collected from the neurostimulator. This
report is used by Medtronic personnel when troubleshooting.
To compile a Medtronic Data Report, refer to "Compiling a
Medtronic Data Report" on page 52.
The Reports screen is accessed through the workow navigator or by
tapping Reports on the Side Menu (
You can also tap REPORTS on the initial app screen to view reports
for a patient without having to start a session.
Note: Any report generated in Demo mode is labeled DEMO MODE
at the top of the report.
Using the Reports screen
1. Select a report type from the drop-down list. The default is
Session Report.
) when not part of a workow.
5. If applicable, tap the checkbox to remove the device serial
number from the report, and then tap CONFIRM.
Table 10. Report buttons.
ButtonDescription
View button – to view the report in PDF format.
This action takes you out of the clinician programmer
app. The app runs in the background and your
session data remains intact.
Share button – to send the PDF report to another
destination.
This action takes you out of the clinician programmer
app. The app runs in the background and your
session data remains intact.
Download button – to download and save the PDF
report on the clinician tablet.
This action saves a PDF le in the default reports
folder on the clinician tablet.
Delete button – to delete the selected patient
session record from the clinician tablet.
When a session record is deleted, the data no longer
exists on the clinician tablet. However, deleting a
session record does not delete any existing PDF les
created from that record and stored on the clinician
tablet.
2. (Optional) Use the lter options to nd the patient data you
want.
3. Tap the desired patient session record in the scrollable list. The
most recent patient session record appears at the top.
4. Tap one of the buttons shown in Table 10.
A71200 English 2022-01-15
49
Working with Reports
Patient Data Service app
The Patient Data Service app manages the storage of patient session
data in a locally stored database on the clinician tablet.
The Patient Data Service app can be used to access or delete patient
session data for any patients whose Medtronic devices have been
programmed using any clinician programmer app on the clinician
tablet.
The Patient Data Service app can also be used to adjust (or disable)
the PDF auto-delete feature, which automatically deletes PDF les
from the default reports folder on the clinician tablet after a designated
amount of time.
Finding and opening the app
1. Navigate to the Apps on the clinician tablet.
2. Find the icon for the Patient Data Service app.
3. Tap the icon to open the app.
Table 11. Patient Data Service buttons.
ButtonDescription
Import button – to import the selected session
record into the Reports screen in the clinician
programmer app.
Refer to "Using the Reports screen" on page 49.
Delete button – to delete the selected patient
session record from the clinician tablet.
When a session record is deleted, the data no longer
exists on the clinician tablet. However, deleting a
session record does not delete any existing PDF les
created from that record and stored on the clinician
tablet.
Deleting multiple patient session
records at once
1. In the Patient Data Service app, use the lter options to nd
the patient data you want.
Accessing or deleting a patient
session record
1. In the Patient Data Service app, use the lter options to nd
the patient data you want.
2. Tap the desired patient session record in the scrollable list. The
most recent patient session record appears at the top.
3. Tap one of the buttons shown in Table 11.
A71200 English 2022-01-15
2. Tap and hold on any record in the results list to activate the
multiple record selection feature.
3. Tap the applicable checkboxes to select all records you want to
delete from the scrollable list.
4. Tap the Trash icon (
records to be deleted is shown next to the icon.
Note: To cancel without deleting records, tap the back arrow in
the upper left corner.
50
Patient Data Service app
) to delete the records. The number of
Setting the PDF auto-delete feature
Use the auto-delete feature to automatically delete PDF les stored
on the clinician tablet based on a timeframe you select. The autodelete feature is enabled by default with a preset amount of time.
This feature only deletes PDF les from the default reports folder on
the clinician tablet. It does not affect patient session data stored in the
database.
1. In the Patient Data Service app, tap the Settings button (
2. Select PDF Auto Delete.
3. Select the appropriate auto-delete settings, and then tap
UPDATE.
).
Viewing System Information
The About System screen is accessed from the Side Menu ().
This screen provides the following information:
■System information (such as model numbers, serial numbers,
rmware numbers, version numbers) for the neurostimulator,
Includes the information for the controller (patient
programmer), and communicator that most recently
communicated with the neurostimulator or was paired with the
clinician tablet.
■Trademarks and licenses information.
This screen provides the following capabilities:
■Ability to change the device date and time for the
neurostimulator.
■Ability to change the neurostimulator implant date.
■Ability to generate a Medtronic Data Report.
Viewing information about system
components
On the About System screen, tap the tab for the component
information you want to view.
A71200 English 2022-01-15
Viewing information about
trademarks and licenses
On the About System screen, tap the Licenses tab, and then tap the
button for the information you want to view.
51
Viewing System Information
Changing device date/time or
implant date information
The Device tab on the About System screen shows the device date,
device time, and implant date. You can change the date and time
information if needed.
■Tap the date that you want to change. Use the calendar
selector to select a new date, and then tap outside the
calendar.
■Tap the time that you want to change. Use the time selector to
select a new time, and then tap outside the time.
Compiling a Medtronic Data Report
The Medtronic Data Report tab on the About System screen is
used to compile a report containing information about the Medtronic
data collected from the neurostimulator.
■Tap the COMPILE REPORT button (
Medtronic Data Report. You cannot cancel the action after it is
initiated.
To access previously compiled Medtronic Data Reports, refer to
"Working with Reports" on page 49.
) to compile the
Viewing Alerts
The clinician programmer app displays alerts to inform you of specic
events or changes in the system. There are three levels of alerts in
the clinician programmer app: warning, caution, and information.
When alerts are present, an ALERTS button displays on the screen.
The button shows the number of alerts that need to be acknowledged.
Read and follow the instructions given in the alert.
Working with alerts
■To view alerts, tap the ALERTS button in the lower left corner
of the screen. Each alert appears in a separate pop-up box.
–Swipe left or right to view individual alerts.
■To close an alert without dismissing it, tap the X (or anywhere
outside of the alert pop-up box).
■To dismiss an alert, tap the DISMISS button. The alert will be
marked as a dismissed alert.
■When all alerts are dismissed, the ALERTS button reads "NO
ALERTS". Those alerts, however, now exist as dismissed
alerts and can still be viewed.
■To view dismissed alerts, tap the NO ALERTS button, and then
tap the SHOW DISMISSED button.
■To restore a dismissed alert, tap the RESTORE button.
A71200 English 2022-01-15
52
Viewing Alerts
Troubleshooting
Approach troubleshooting conservatively. Prior to performing
invasive procedures, ensure that all noninvasive solutions have been
considered. If additional assistance is needed, contact Medtronic
using the contact information listed on the back cover of this manual.
The clinician programmer app displays messages to inform you of
issues in the system. Follow the instructions given in the message.
Data validation error message
During initial interrogation, the app performs a data validation check
to ensure that the data in the neurostimulator is valid and can be
understood. If data is found to be invalid, the app displays a data
validation error message and provides the option to clear the invalid
data.
Elective Replacement Indicator (ERI)
message
When an implanted neurostimulator reaches its recommended
replacement date, a message displays. Refer to the System Eligibility and Battery Longevity reference manual for information about
neurostimulator battery longevity.
End of Service (EOS) message
When an implanted neurostimulator reaches its end of service (EOS),
a message displays. At EOS, therapy is not available and the device
must be replaced. Refer to "Estimating Battery Longevity" on page
31 for information about neurostimulator battery longevity.
Note: Impedance can be measured at EOS. Refer to "Testing
impedance" on page 32 for more information.
Power on reset (POR) message
If a power on reset (POR) message displays at initial interrogation,
this indicates that one or more POR events occurred since the last
programming session. If a power on reset message displays while in
a current programming session, the session will end and any unsaved
data will be lost.
Out of regulation (OOR) message
An out of regulation (OOR) message indicates that the delivered
amplitude may be lower than the programmed amplitude for the
indicated programs because of low impedance. If an out of regulation
situation occurs, decrease settings or reduce the number of
electrodes for the indicated programs.
A71200 English 2022-01-15
Using impedance measurements for
troubleshooting
Refer to "Checking System Performance" on page 32.
Communicator and clinician tablet
troubleshooting
Refer to Table 12 on page 55 for a list of potential issues that can
occur with the communicator or the clinician tablet, as well as possible
solutions for resolving those issues.
To reset a neurostimulator
If a neurostimulator persistently cannot be found and communication
to the neurostimulator has been lost, use the following steps to reset
the neurostimulator.
53
Troubleshooting
1. Open the application.
2. Tap the Settings button (
) in the top-right corner of the
CONNECT screen, then tap About.
3. Tap the Device tab.
4. After reading the on-screen text, tap the RESET
NEUROSTIMULATOR(S) button, then conrm the action.
After sending the reset command, the communicator will beep.
Note: Therapy is automatically stopped after resetting the
neurostimulator. To resume therapy, refer to "Initiating communication
with the neurostimulator" on page 12.
Contact the appropriate Medtronic representative listed on the back
cover of this manual if additional assistance is needed.
A71200 English 2022-01-15
54
Troubleshooting
Table 12. Communicator and clinician tablet troubleshooting.
ProblemPossible Solutions
The communicator cannot
communicate with the
neurostimulator.
The communicator cannot pair
with the clinician tablet.
The communicator cannot
communicate wirelessly with
the clinician tablet during a
programming session.
■ The communicator may not be positioned correctly over the neurostimulator.
– Position the target symbol (
facing the neurostimulator.
■ The communicator may be too far away from the neurostimulator.
– Move the communicator closer to the neurostimulator.
■ Metal surfaces can interfere with communication between the communicator and the neurostimulator.
– If the communicator is on a metal table or a metal tray, move the communicator to a nonmetal
surface.
■ There could be radio-frequency (RF) interference.
– Move the communicator closer to the neurostimulator.
■ The communicator may have interrogated a different device.
– Place the communicator directly over the neurostimulator (use a sterile barrier, if applicable) and
tap FIND DEVICE on the clinician programmer app.
BLUETOOTH communication requires an initial pairing with a USB cable. Refer to “Pairing the
communicator to the clinician tablet” on page 12 if this initial pairing has not been completed.
When attempting to connect through BLUETOOTH wireless technology:
■ Make sure the communicator is turned on and within range of the clinician tablet. See the 8880T2
Technical Manual for details.
■ Make sure that the BLUETOOTH wireless technology is enabled on the clinician tablet.
Refer to "Communicator icons on the tablet status bar" on page 16 for what the communicator icons
mean when displayed in the tablet status bar.
Possible reasons:
■ You are in an environment where multiple devices are using BLUETOOTH wireless technology and
thereby creating interference.
■ You have moved into an environment where the BLUETOOTH wireless technology is prohibited.
Possible solutions:
■ Use the USB connector cable to connect the communicator to the clinician tablet.
■ Make sure that the BLUETOOTH wireless technology is enabled on the clinician tablet.
Refer to "Communicator icons on the tablet status bar" on page 16 for what the communicator icons
mean when displayed in the tablet status bar.
) on the back of the communicator so that it is centered over and
A71200 English 2022-01-15
55
Troubleshooting
ProblemPossible Solutions
The communicator battery level
is low.
The clinician tablet, the clinician
programmer app, or the
communicator is unresponsive.
The communicator is damaged
or overheats.
If the communicator requires repair, is damaged, or is nonfunctional and a replacement is needed, contact Medtronic using the contact
information listed on the back cover of this manual.
■ Open the battery case and replace the batteries.
– If needed, refer to the instructions for replacing the communicator batteries in the Model 8880T2
Communicator technical manual.
■ After replacing the communicator batteries, the clinician programmer app reconnects with the
communicator using the BLUETOOTH wireless technology and resumes the session.
■ Turn off the power for the clinician tablet or the communicator, then turn the power on.
■ If you continue to have communication problems or cannot use the app to program, contact Medtronic
using the contact information listed on the back cover of this manual.
Use a different communicator:
■ Ensure that the damaged communicator is turned off before using another communicator.
■ If using a replacement communicator prior to starting a programming session, use the USB
connector cable and pair the replacement communicator to the clinician tablet.
– Refer to "Pairing the communicator to the clinician tablet" on page 12.
Or
– If replacing the communicator while in a programming session, use the USB connector cable to
connect the communicator to the clinician tablet and continue with the programming session.
A71200 English 2022-01-15
56
Troubleshooting
Patient programmer messages
The patient programmer displays text and iconic error and
informational messages. Messages provide a service code and tell
the patients to contact their clinician. These service codes and their
troubleshooting procedures are displayed in Table 13. If additional
assistance is needed, contact Medtronic using the contact information
listed on the back cover of this manual.
Table 13. Service codes on the patient programmer
indicating the need to contact a clinician.
Service
code
201 or 204Elective Replacement Indicator (ERI)
302End of Service (EOS)
303Update App
Explanation and Possible Solutions
One of these service codes will display when an
implanted neurostimulator reaches its recommended
replacement date.
The implanted neurostimulator has reached its end
of service, a message displays. At EOS, therapy
is not available and the neurostimulator must be
replaced.
A “System Update Needed” alert will display
on the patient programmer app if the app is not
compatible with the neurostimulator. Conrm that
the patient has the correct programmer model for
their neurostimulator. The patient or clinician should
connect to a Wi-Fi network to update the patient
programmer app. When the update is complete, WiFi capability can be turned off again.
Service
Explanation and Possible Solutions
code
306Remove Neurostimulator
This message will appear if the user attempts to
remove a paired neurostimulator from the About
screen in the patient programmer app. A paired
neurostimulator is required to use the patient
programmer app.
307Remove Communicator
This message will appear if the user attempts to
remove a paired communicator from the “About”
screen in the patient programmer app. A paired
communicator is required to use the patient
programmer app with the neurostimulator.
311Unexpected Command Response
Patient will see an “Error Found” alert. Contact
Medtronic with the service code.
312 Unexpected Device Status
Patient will see an “Error Found” alert. Contact
Medtronic with the service code.
318Invalid Therapy Control Settings
Patient will see an “Update Settings” alert after
invalid settings are detected. Resolve this issue
by updating the neurostimulator settings with the
clinician programmer app.
319No Therapy
A “No Therapy” alert will display on the patient
programmer app when the neurostimulator has not
been programmed yet with the clinician programmer
app.
A71200 English 2022-01-15
57
Troubleshooting
Service
Explanation and Possible Solutions
code
322Firmware not Running
A “No Therapy” alert will display on the patient
programmer app when a rmware issue has been
detected. Contact Medtronic with the service code.
323Unexpected Therapy Off
A “Therapy Off” alert will display on the patient
programmer app and the neurostimulator will
turn stimulation off when settings are too high.
Reprogram the neurostimulator with reduced settings
(fewer electrodes and programs; lower amplitude,
pulse width, and rate).
324Unexpected Telemetry Error
Patient will see an “Error Found” alert. Contact
Medtronic with the service code.
A71200 English 2022-01-15
58
Troubleshooting
A71200 English 2022-01-15
59
Manufacturer
Medtronic, Inc.
710 Medtronic Parkway
Minneapolis, MN 55432
USA
www.medtronic.com
Tel. +1-763-505-5000
REP
EC
Authorized Representative
in the European Community