Explanation of symbols on product or package labeling
Refer to the appropriate product for symbols that apply.
Conformité Europ éenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (NB 0123) and R&TTE Directive 1999/5/EC.
Warning
Consult instructions for use
Do not reuse
Do not resterilize
2
STERILIZE
Date of manufacture
Manufacturer
Open here
Use by
Serial number
Medtronic Confidential
STERILE
EC
! USA
XXX °F
-XX °C
-XX °F
M946871A001
Sterilized using ethylene oxide
EO
Authorized Representative in the European Community
REP
For USA audiences only
XX °C
Temperature limitation
97715 2012-06 English 1
2012-06
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential
Medtronic® and SoftStart/Stop® are trademarks of Medtronic, Inc., registered in the U.S. and other countries.
AdaptiveStim™ and Intellis™ are trademarks of Medtronic, Inc.
2 English 97715 2012-06
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN

Table of contents

Description 5
Package contents 5
Patient identification card 5
Device specifications 5
Declaration of conformity 8
Instructions for use 9
Charging the neurostimulator battery 9 Verifying neurostimulator operation 9 Connecting the extension or lead to the neurostimulator 10 Implanting the neurostimulator 11 Checking system integrity 12 Completing the implant procedure 12
ImplantManual.xsl - IPGTemplate.fm
Medtronic Confidential
Template version: 08-02-2011
Refer to the indications sheet for indications and related information.
Refer to the appropriate information for prescribers booklet for contraindications, warnings, precautions, adverse events summary, individualization of treatment, patient selection, use in specific populations, resterilization, and component disposal.
Refer to System Eligibility, Battery Longevity, Specifications reference manual for neurostimulator selection, battery longevity calculations and specific neurostimulator specifications.
Refer to the clinical summary booklet for information on the clinical study
results of the neurostimulation system and individualization of treatment.
97715 2012-06 English 3
M946871A001
2012-06
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Medtronic Confidential
Template version: 08-02-2011
4 English 97715 2012-06
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential

Description

The Medtronic Intellis Model 97715 Neurostimulator is part of a neurostimulation system for pain therapy.

Package contents

Neurostimulator
Torque wrench
Pocket sizer
Bore plugs (2)
Product literature
Registration form

Patient identification card

Warranty card
Patient identification card
A patient identification card is packaged with this device. Advise the patient to carry the identification card at all times.
! USA
The patient identi fication card packaged with the de vice is temporary; a permanent card
will be mailed to the patient when Medtronic receives the registration form. The implant registra tion form registers the device warranties and creates a record of the device
in Medtronic’s implant data system.

Device specifications

The neurostimulator is a multi-programmable, rechargeable device that delivers stimulation through 1 or more leads. The stimulation settings are stored in programs to target pain areas. A program is a specific combination of pulse width, rate, and intensity settings acting on a specific electrode combination (up to 16 electrodes per program). Up to 4 pain areas can be targeted by programs. When stimulating more than one pain area, the pulses are delivered sequentially—first a pulse from one program, then a pulse from the next program.
Date, pulse width, intensity, cycling, and electrode polarity for each program within the group can have different values. Rate, rate limits, pulse width and intensity limits, and ramping for each program within the group have the same values.
97715 2012-06 English 5
M946871A001
2012-06
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential
Table 1. Operating values for the Intellis Model 97715 Neurostimulator
Programmable parameter
Operating values and ranges
a
Number of defined groups 1-3 (optional) Number of programs per pain
area
1-3
Number of programs 12 Number of pain areas 1-4 Electrode configuration 2-16 electrodes as anode, cathode, or Off Maximum intensity per
electrode
0-25.5 mA (0.1 mA increment)
Program intensity 0-100 mA Intensity – limits Enabled or disabled at maximum of 25.5 mA per electrode Pulse width 60-1000 µs (10-µs increment)
Pulse width – limits
Master rate
a
b
Enabled or disabled at maximum of 1000 Hz
10-1200 Hz (increment: 1 Hz from 10-30 Hz, 5 Hz from 30-250 Hz, 10 Hz from 250-500 Hz, 20 Hz from 500-1000 Hz,
50 Hz from 1000-1200 Hz) Rate ratio A fraction of the master rate (1/1, 1/2, 1/3, 1/4, 1/5) Rate limits
Enabled or disabled (at maximum of 1200)
a
SoftStart/Stop Off, On: 1, 2, 4, or 8 second ramp duration
c
Cycling
AdaptiveStim
a
Rate availability depends on how many pain areas are defined. For example, the maximum rate available in one defined pain area is 1200 Hz. The maximum rate available if two pain areas are defined is 600 Hz in each of those two pain areas.
b
Pulse width limits are not available when AdaptiveStim is enabled.
c
Cycling is not available for pain areas with AdaptiveStim enabled.
Off, On: 0.1 s-30 min (resolution: 0.1 s from 0.1-1 s, 1 s from 1
s-1 min, 1 min from 1-30 min)
Off, On: 7 positions
6 English 97715 2012-06
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Medtronic Confidential
Template version: 08-02-2011
Table 2. Physical characteristics of the Intellis Model 97715 Neurostimulator
Description Value
Connector type Octapolar, in-line 2.8-mm (0.110-in) spacing Height 57.0 mm (2.1 in) Width 47.3 mm (1.9 in) Thickness
case 6.3 mm (0.2 in)
connector 8.9 mm (0.4 in) Weight 29.9 g (1.1 oz) Volume
13.7 cm
3
Battery life 9 years before ERI Power source Lithium ion rechargeable battery Storage temperature –18 °C to +52 °C (0 °F to +126 °F)
Serial number model designator
b
Radiopaque identification (ID) code
a
All measurements are approximate.
b
The serial number is the model designator followed by a number. The clinician programmer displays the entire serial number beginning with the model designator.
NME
NME
a
M946871A001
97715 2012-06 English 7
2012-06
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential
Table 3. Material of components in the Intellis Model 97715 package
Components
Material Material contacts
human tissue
Neurostimulator
Case Titanium Yes Connector block Titanium, polysulfone, silicone
rubber, silicone medical adhesive
Yes
Grommets, seals Silicone rubber Yes Setscrews Titanium alloy Yes Adhesive Silicone medical adhesive Yes Pocket sizer Polypropylene Yes Bore plug Silicone rubber; stainless steel Yes
Torque wrench
Handle Polyetherimide Yes Shaft
Stainless steel Yes

Declaration of conformity

Medtronic declares that t his product is in conformity with the es sential requirements of Directive 1999/5/EC on Radio and Telecommunications Terminal Equipment and Directive 90/385/EEC on Active Implantable Medical Devices.
For additional info rmation, contact the appropriate M edtronic representative listed on the in side back cover of this manual.
8 English 97715 2012-06
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential

Instructions for use

Implanting physicians should be experienced in epidural-access procedures and should be thoroughly familiar with all product labeling.
w
Warning: DO NOT use the recharger on an unhealed wound. The recharging system is not sterile, and contact with the wound can cause an infection.
#
Caution: If the neurostimulator is not being used for an extended period of time, recommend that your patient charge the neurostimulator at least once per year. If the battery is discharged, stimulation will stop and the neurostimulator may not communicate with the controller.
#
Caution: Advise patients to charge the neurostimulator when a low battery message is displayed on the controller in order to maintain uninterrupted therapy from the neurostimulator. If the battery is discharged, stimulation will stop and the neurostimulator may not communicate with the controller.
Note: The patient will be able to use the controller and recharger to charge a discharged battery without causing damage to the battery or the neurostimulator.
#
Cautions:
When using sharp instruments near the neurostimulator, be extremely careful to avoid nicking or damaging the case, the insulation, or the connector block. Damaging the neurostimulator may require surgical replacement.
Do not use saline or other ionic fluids at connections, which could result in a short circuit.

Charging the neurostimulator battery

Check the battery level of the neurostimulator before opening the package, and recharge the neurostimulator if the battery is low. For charging instructions, refer to the recharging system user manual. If the patient will be sent home with stimulation on, charge the neurostimulator in the package before implant.

Verifying neurostimulator operation

Before opening the sterile neurostimulator packa ge, verify that the neurostimulato r is operable by using the clinician programmer to interrogate the neurostimulator and read the neurostimulator battery charge level. (Refer to the programming guide for instructions on how to read the battery charge level.)
#
Caution: Do not implant a neurostimulator if it was dropped onto a hard surface from a height of 30 cm (12 in) or more, because the neurostimulator may be damaged and fail to operate properly.
Note: The neurostimulator pocket may be flushed with an antibiotic solution; do not submerge the neurostimulator in fluid.
M946871A001
97715 2012-06 English 9
2012-06
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential

Connecting the extension or lead to the neurostimulator

#
Caution: Before connecting components, wipe off any body fluids and dry all
connections. Fluids in the connection may result in stimulation at the connection site,
intermittent stimulation, or loss of stimulation.
1. Wipe the extension or lead connector pins with sterile gauze. If necessary, use sterile (United States Pharmacopeia [USP]) water or a nonionic antibiotic solution.
2. Make sure the connector block receptacles are dry and clean.
3. Insert the appropriate extension or lead connector pins into the appropriate
neurostimulator socket until they are seated fully within the connector block (Figure 1).
Notes:
During insertion, some resistance is typical.
To retract the setscrews, insert the torque wrench into the self-sealing grommet and
rotate the setscrews counterclockwise; however, do not remove the setscrews from the connector block.
#
Caution: Do not insert the extension or lead connector into the connector block if the setscrews are not sufficiently retracted. If the setscrews are not retracted, the setscrews may damage the extension or lead and the extension or lead will not be seated fully into the connector block.
Socket I (Electrodes 0-7)
Socket II (Electrodes 8-15)
Extension or Lead 1
Extension or Lead 2
Figure 1. Insert the extension or lead connector pins fully into the neurostimulator.
10 English 97715 2012-06
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN
Note: Insert a connector plug (from an accessory kit) into any unused neurostimulator socket.
4. For each extension, lead, or plug, fully insert the torque wrench (packaged with the
rechargeable neurostimulation system) into each self-sealing grommet of the connector block and tighten each setscrew (Figure 2).
#
Cautions:
Be sure the torque wrench is fully inserted into the self-sealing grommet. If the torque wrench is not fully inserted, the setscrew may be damaged, resulting in intermittent or loss of stimulation.
Before tightening setscrews, ensure that the extension or lead connector pins are inserted into the connector block to prevent damaging the lead or extension.
Verify that each leaf of the self-sealing grommet is closed after the torque wrench is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the patient may experience shocking, burning, or irritation at the neurostimulator implant location, or intermittent stimulation, or loss of stimulation.
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential
Figure 2. Tightening the setscrews in the self-sealing grommet.

Implanting the neurostimulator

1. Place the neurostimulator into the subcutaneous pocket using the pocket sizer if desired
and ensure that the extension or lead is not bent sharply. The neurostimulator can be implanted and charged with either side facing outward. for optimal recharging, place the neurostimulator within 1-2 cm (0.8 in.) of the skin surface.
#
Cautions:
Ensure that the neurostimulator is placed no deeper than 3 cm (1.2 in) below the skin and is parallel to the skin. If the neurostimulator is too deep or is not parallel to the skin, recharge may be inefficient or unsuccessful.
97715 2012-06 English 11
M946871A001
2012-06
Filename Date Time UC200xxxxxx EN
Do not coil excess extension in front of the neurostimulator. Wrap excess extension or leads around the perimeter (Figure 3) or behind the neurostimulator to help minimize potential damage during neurostimulator replacement surgery, help minimize potential kinking of the extension or lead, and minimize interference with telemetry and recharge operation.
Figure 3. Wrap excess extension around the perimeter of the neurostimulator.
2. Use the suture holes in the connector block to secure the neurostimulator to the muscle fascia with nonabsorbable silk.
Note: Secure the neurostimulator in the pocket to minimize movement or migration of the neurostimulator.
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
Medtronic Confidential

Checking system integrity

The connections of the extensions and leads to the neurostimulator can be checked using the clinician programmer. Refer to the programming guide for detailed programming instructions.
w
Warning: To use the nonsterile programmer system components in a sterile field, place a sterile barrier between the patient and system components to prevent infection. Do not sterilize any components of the programmer system. Sterilization may damage the components.
1. To ensure proper connection of each extension or lead to the neurostimulator, use the clinician programmer to perform the lead insertion check.
2. If the lead insertion check results are not acceptable, refer to "Connecting the extension or lead to the neurostimulator".

Completing the implant procedure

1. Close and dress all incisions.
2. Turn off stimulation before sending your patient home.
3. Ensure that a patient control device is given to the patient.
12 English 97715 2012-06
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Template version: 08-02-2011
4. Complete the device tracking and patient registration paperwork and return the
documents to Medtronic.
Medtronic Confidential
M946871A001
97715 2012-06 English 13
2012-06
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Medtronic Confidential
Template version: 08-02-2011
Contacts:
Asia:
Medtronic International Ltd. Tel. 02891-4068 Fax 02591-0313
Medtronic Asia Ltd. Tel. (02)-548-1148
Fax (02)-518-4786
Australia:
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde, NSW 2113 Australia Tel. +61-2-9857-9000 Fax +61-2-9878-5100 Toll free 1-800-668-6700
Austria:
Medtronic Österreich GmbH Tel. 01-240440 Fax 01-24044-100
Belgium:
Medtronic Belgium S.A. Tel. 02-456-0900 Fax 02-460-2667
Canada:
Medtronic of Canada Ltd. Tel. (1-905)-460-3800 Fax (1905)-826-6620
Czech Republic:
Medtronic Czechia s.r.o. Tel. 2-965-795-80 Fax 2-965-795-89
Denmark:
Medtronic Danmark A/S Tel. 45-32-48-18-00 Fax 45-32-48-18-01
Finland:
Medtronic Finland Oy/LTD Tel. (09)-755-2500 Fax (09)-755-25018
France:
Medtronic France S.A.S. Tel. 01-5538-1700 Fax 01-5538-1800
Germany:
Medtronic GmbH Tel. (02159)-81490 Fax (02159)-8149100
Greece:
Medtronic Hellas S.A. Tel. 210-67-79-099 Fax 210-67-79-399
Hungary:
Medtronic Hungária Kft. Tel. 1-889-06-00 Fax 1-889-06-99
Ireland:
Medtronic Ireland Ltd. Tel. (01)-890-6522 Fax (01)-890-7220
Italy:
Medtronic Italia SpA Tel. 02-241371 Fax 02-241381
Tel. 06-328141 Fax 06-3215812
Japan:
Medtronic Japan Tel. 3-6430-2016 Fax 3-6430-7110
Latin America:
Medtronic, Inc. Tel. (1305)-500-9328 Fax (1786)-709-4244
Norway:
Medtronic Norge AS Tel. 067-10-32-00 Fax 067-10-32-10
Poland:
Medtronic Poland Sp. z.o.o. Tel. (022)-465-69-00 Fax (022)-465-69-17
Portugal:
Medtronic Portugal, Lda. Tel. 21-724-5100 Fax 21-724-5199
Russia:
Medtronic Russia Tel. (8495) 580-7377 Fax (8495) 580-7378
Slovakia
Medtronic Slovakia, o.z. Tel. 0268 206 911 Fax 0268 206 999
Spain:
Medtronic Ibérica, S.A. Tel. 91-625-0400 Fax 91-650-7410
Sweden:
Medtronic AB Tel. 08-568-585-00 Fax 08-568-585-01
2012-06
M946871A001
Filename Date Time UC200xxxxxx EN
ImplantManual.xsl - IPGTemplate.fm
Medtronic Confidential
Template version: 08-02-2011
Switzerland:
Medtronic (Schweiz) AG Tel. 031-868-0100 Fax 031-868-0199
The Netherlands:
Medtronic B.V. Tel. (045)-566-8000 Fax (045)-566-8668
U.K.:
Medtronic U.K. Ltd. Tel. 01923-212213 Fax 01923-241004
USA:
Medtronic, Inc. Tel. (1763)-505-5000 Fax (1763)-505-1000 Toll-free: (1-800)-328-0810
M946871A001
2012-06
Filename Date Time UC200xxxxxx EN
Manufacturer
Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. www.medtronic.com Tel. +1-763-505-5000 Fax +1-763-505-1000
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10, 6422 PJ Heerlen, The Netherlands Tel. +31-45-566-8000 Fax +31-45-566-8668
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31, Case Postale 84 CH-1131 Tolochenaz, Switzerland www.medtronic.eu Tel. +41-21-802-7000 Fax +41-21-802-7900
Asia-Pacific
Medtronic International Ltd. Suite 1602 16/F, Manulife Plaza, The Lee Gardens, 33 Hysan Avenue, Causeway Bay, Hong Kong Tel. +852-2891-4068 Fax +852-2591-0313