Medtronic 97713 User Manual

RestoreAdvanced® SureScan® MRI
Rechargeable neurostimulator
Implant manual
  Rx only
97713
Explanation of symbols on product or package labeling
Refer to the appropriate product for symbols that apply.
Open here
Do not use if package is damaged
Do not reuse
Do not resterilize
2
STERILIZE
Sterilized using ethylene oxide
-XX °C
-XX °F
EC
EO
Consult instructions for use
Date of manufacture
Manufacturer
Use by
XX °C
Temperature limitation
XXX °F
Serial number
Conformité Européenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (NB
0123).
Authorized Representative in the European Community
REP
STERILE
For USA audiences only
97713 2016-06-01  English 3
Medtronic®, GroupAdjust®, MyStim®, RestoreAdvanced®, SoftStart/Stop®, and SureScan are trademarks of Medtronic, Inc., registered in the U.S. and other countries.
®
4 English  97713 2016-06-01

Table of contents

Description 7
Neurostimulation systems with SureScan MRI Technology 7
Package contents 7
Patient identification card 7
Device specifications 7
Declaration of conformity 10
Implanted components and MRI scans 11
Implant criteria for full-body MRI scan eligibility 11 When changing components 11 When explanting components 12
Instructions for use 12
Charging the neurostimulator battery 12 Verifying neurostimulator operation 12 Connecting the extension or lead to the neurostimulator 13 Implanting the neurostimulator 15
Refer to the indications sheet for indications and related information.
Refer to the appropriate information for prescribers booklet for contraindications, warnings, precautions, adverse events summary, individualization of treatment, patient selection, use in specific populations, resterilization, and component disposal.
Refer to the MRI Guidelines for Medtronic Neurostimulation Systems for Chronic Pain instructions for use manual for the MRI conditions and MRI-specific warnings and precautions for conducting an MRI scan.
Refer to the System Eligibility, Battery Longevity, Specifications reference manual for neurostimulator selection, battery longevity calculations and specific neurostimulator specifications.
Refer to the clinical summary booklet for information on the clinical study
results of the neurostimulation system and individualization of treatment.
97713 2016-06-01  English 5
Checking system integrity 17 Completing the implant procedure 17
6 English  97713 2016-06-01

Description

The Medtronic RestoreAdvanced SureScan MRI Model 97713 Neurostimulator is part of a neurostimulation system for pain therapy.

Neurostimulation systems with SureScan MRI Technology

When a Medtronic neurostimulation system with SureScan MRI Technology is implanted as directed (see "Implant criteria for full-body MRI scan eligibility" on page 11), a patient's full body may be eligible for MRI scans under specific conditions, ie, any part of the patient's anatomy can be scanned when specific conditions are met.
For the MRI conditions and MRI-specific warnings and precautions for conducting an MRI scan, refer to the MRI guidelines for Medtronic neurostimulation systems for chronic pain instructions for use manual.

Package contents

Neurostimulator
Torque wrench
Product literature
Warranty card
Registration form

Patient identification card

Patient identification card
A patient identification card is packaged with this device. Advise the patient to carry the most up-to-date identification card at all times and to bring the card to all MRI appointments.
The patient identification card packaged with the device is temporary; a permanent
card will be mailed to the patient when Medtronic receives the registration form. The implant registration form registers the device warranties and creates a record of the
device in Medtronic’s implant data system.

Device specifications

The neurostimulator is a multi-programmable, rechargeable device that delivers stimulation through 1 or more leads. The stimulation settings are stored in programs. A program is a specific combination of pulse width, rate, and amplitude settings acting on a specific electrode combination (up to 16 electrodes per program). Up to four programs can be combined into a MyStim group. When using more than one program, the pulses are delivered sequentially—first a pulse from one program, then a pulse from the next program.
97713 2016-06-01  English 7
Pulse width, amplitude, and electrode polarity for each program within the group can have different values. Rate, rate limits, ramping, and cycling for each program within the group have the same values.
Table 1. Operating values for the RestoreAdvanced SureScan MRI Model 97713
neurostimulator
a
Programmable parameter Operating range and resolution
Number of defined MyStim groups
Number of programs per MyStim group
Electrode configuration 2 to 16 electrodes as anode (+), cathode (-), or
Amplitude
Amplitude – upper patient limit Tracking limit: programmed value +0 to +4 V (0.5
Amplitude – lower patient limit Custom limit: 0 V to the programmed value (same
b
1 to 26
b
1 to 4
Off
0 to 10.5 V with 0.05 V or 0.1 V resolution
c
V resolution) Custom limit: programmed value up to 10.5 V
(same resolution as amplitude)
resolution as amplitude) Pulse width 60 to 450 µs (10 µs resolution) Pulse width – upper patient limit Tracking limit: programmed value +0 to +150 µs
(30 µs resolution)
Custom limit: programmed value up to 450 µs
(10 µs resolution) Pulse width – lower patient limit Custom limit: 60 µs to the programmed value
Rate 2 to 130 Hz (1 Hz resolution to 10 Hz; 5 Hz
Rate – upper patient limit Tracking limit: programmed value +0 to +50 Hz
Rate – lower patient limit Custom limit: 2 Hz to the programmed value (1 Hz
(10 µs resolution)
resolution from 10 Hz to 130 Hz)
(10 Hz resolution)
resolution to 10 Hz; 5 Hz resolution from 10 Hz to
d
130 Hz)
d
SoftStart/Stop Off, On: 1, 2, 4, or 8 sec ramp duration Cycling Off, On: 0.1 sec to 30 min (resolution: 0.1 sec
from 0.1 sec to 1 sec; 1 sec from 1 sec to 1 min;
1 min from 1 min to 30 min)
8 English  97713 2016-06-01
Table 1. Operating values for the RestoreAdvanced SureScan MRI Model 97713
neurostimulator
a
(continued)
Programmable parameter Operating range and resolution
Scheduled therapy Off, On: 1 to 8 events per 24 h period GroupAdjust Off, On
a
All values are approximate.
b
No more than 32 programs may be defined within the 26 MyStim groups.
c
Only 0.1 V resolution possible from 10.0 V to 10.5 V amplitude.
d
Rate limited to 85 Hz when 3 programs/group are active and to 65 Hz when 4 programs/group are active.
Table 2. Physical characteristics of the RestoreAdvanced SureScan MRI Model 97713
neurostimulator
a
Description Value
Connector type Octapolar, in-line 2.8 mm (0.110 in) spacing Height 65.0 mm (2.6 in) Length 49.0 mm (1.9 in) Thickness
case 15.0 mm (0.6 in)
connector 15.0 mm (0.6 in) Weight 72.0 g (2.5 oz) Volume
39.0 cm
3
(2.38 in3) Battery life 9 years
Power source
b
Lithium ion rechargeable battery
Temperature limitation –18 °C to +52 °C (0 °F to +126 °F)
Serial number model designator
c
Radiopaque identification (ID) code
a
All measurements are approximate.
b
The neurostimulator is not shipped with a full battery charge. Refer to the charging system user manual for neurostimulator charging instructions.
c
The serial number is the model designator followed by a number. The clinician programmer displays the entire serial number beginning with the model designator.
d
The radiopaque ID code is located in the connector block; NMA indicates that the neurostimulator has SureScan MRI Technology. This radiopaque ID code is for confirming, if needed, that a SureScan MRI
NMC
d
NMA
97713 2016-06-01  English 9
Table 2. Physical characteristics of the RestoreAdvanced SureScan MRI Model 97713
neurostimulator
a
Description Value
neurostimulator is implanted and is not to be used for concluding that the entire neurostimulation system is full-body MRI scan eligible.
Table 3. Material of components in the RestoreAdvanced SureScan MRI Model 97713
package
Components Material Material contacts
human tissue
Neurostimulator
Case Titanium Yes Connector block Polyurethane, silicone rubber,
silicone medical adhesive
Yes
Recharge coil Polysulfone Yes Grommets, seals Silicone rubber Yes Setscrews Titanium alloy Yes Adhesive Silicone medical adhesive Yes
Torque wrench
Handle Polyetherimide Yes Shaft Stainless steel Yes

Declaration of conformity

Medtronic declares that this product is in conformity with the essential requirements of Directive 90/385/EEC on Active Implantable Medical Devices.
For additional information, contact the appropriate Medtronic representative listed on the inside back cover of this manual.
10 English  97713 2016-06-01

Implanted components and MRI scans

Implant criteria for full-body MRI scan eligibility

Caution: To allow a patient full-body MRI scan eligibility under specific conditions,
implant a Medtronic neurostimulation system with SureScan MRI Technology as follows:
Use only SureScan MRI neurostimulation system components (eg, leads and
neurostimulators). Note: Extension model numbers in the 3708 series (eg, 37081, 37082, 37083) are
not full-body MRI scan eligible. Implant the neurostimulator in the buttocks, abdomen, or flank (ie, the lateral and
posterior region between the ribs and pelvis). Place the lead tip(s) in the spinal epidural space.
Explant any previously abandoned pain leads or extensions that may be in the
patient (ie, leads or extensions, or portions of, that are not connected to a neurostimulator).
Note: Confirm MRI compatibility of any other implanted medical devices. Other implanted medical devices may limit or restrict MRI scans.
Enter all component model number and implant location information using the
clinician programmer.
If the above implant criteria are not met, the patient will not have a neurostimulation system with full-body MRI scan eligibility. MRI scan eligibility will be restricted.
For the MRI conditions and MRI-specific warnings and precautions for conducting an MRI scan, refer to the MRI guidelines for Medtronic neurostimulation systems for chronic pain instructions for use manual. MR scans performed under different conditions can result in patient injury or damage to the implantable device.

When changing components

Warning: Before explanting and replacing an existing neurostimulator, print a report
from the existing neurostimulator that shows implanted and any abandoned component information. When adding, changing, or removing neurostimulators, leads, extensions, and accessories, always program up-to-date component model numbers, implant locations, and any abandoned component information to re-establish MRI­scan eligibility.
If this information is not updated or is entered incorrectly, MRI scan-type eligibility data will be inaccurate, and the patient is at risk for one of the following:
The patient is allowed to have an MRI scan inappropriate for the implanted
components, which could cause tissue heating, resulting in tissue damage or serious patient injury.
The patient is unnecessarily restricted from having an MRI scan.
977132016-06-01  English11

When explanting components

Caution: If permanently explanting a neurostimulator, be sure to also explant all leads,
extensions, and accessories. Abandoned components may prevent the patient from being allowed MRI scans in the future due to concerns of lead electrode heating that can result in tissue damage.

Instructions for use

Implanting physicians should be experienced in epidural-access procedures and should be thoroughly familiar with all product labeling.
Warning: DO NOT use the recharger on an unhealed wound. The recharging system is not sterile, and contact with the wound can cause an infection.
Caution: Advise patients to charge the neurostimulator when a Low battery ( ) screen is displayed on the patient programmer or recharger; this prevents the battery from overdischarging. If the neurostimulator battery is allowed to overdischarge, the patient cannot charge the neurostimulator; however, the clinician may be able to restore the battery function using the Physician Recharge Mode on the recharger (refer to the troubleshooting section of the software manual).
Allowing the neurostimulator battery to overdischarge will permanently affect the neurostimulator in one of the following ways:
Battery function is restored; however, charging sessions may be more frequent
because battery capacity has been reduced. Battery function is not restored and the neurostimulator must be surgically
replaced. Battery function is not restored when:
the neurostimulator battery is permanently damaged.
the neurostimulator battery has been overdischarged and restored twice
before. The third time the battery is overdischarged, the neurostimulator will reach end of service. Surgery is required to replace the neurostimulator.
Cautions:
When using sharp instruments near the neurostimulator, be extremely careful to
avoid nicking or damaging the case, the insulation, or the connector block. Damaging the neurostimulator may require surgical replacement.
Do not use saline or other ionic fluids at connections, which could result in a short
circuit.

Charging the neurostimulator battery

Charge the neurostimulator battery before opening the package. For charging instructions, refer to the charging system user manual.

Verifying neurostimulator operation

Before opening the sterile neurostimulator package, verify that the neurostimulator is operable by using the clinician programmer to interrogate the neurostimulator and read the
12 English  97713 2016-06-01
neurostimulator battery charge level. (Refer to the software manual for instructions on how to read the battery charge level.)
Caution: Do not implant a neurostimulator if it was dropped onto a hard surface from a height of 30 cm (12 in) or more, because the neurostimulator may be damaged and fail to operate properly.
Note: The neurostimulator pocket may be flushed with an antibiotic solution; do not submerge the neurostimulator in fluid.

Connecting the extension or lead to the neurostimulator

Caution: Before connecting components, wipe off any body fluids and dry all
connections. Fluids in the connection may result in stimulation at the connection site, intermittent stimulation, or loss of stimulation.
1. Wipe the extension or lead connector pins with sterile gauze. If necessary, use sterile (United States Pharmacopeia [USP]) water or a nonionic antibiotic solution.
2. Make sure the connector block receptacles are dry and clean.
3. Insert the appropriate extension or lead connector pins into the appropriate
neurostimulator socket until they are seated fully within the connector block (Figure 1).
Notes:
During insertion, some resistance is typical.
To retract the setscrews, insert the torque wrench into the self-sealing grommet
and rotate the setscrews counterclockwise; however, do not remove the setscrews from the connector block.
Caution: Do not insert the extension or lead connector into the connector block if the setscrews are not sufficiently retracted. Unretracted setscrews may damage the extension or lead and prevent the extension or lead from fully seating into the connector block.
97713 2016-06-01  English 13
Socket II (Electrodes 8-15)
Socket I (Electrodes 0-7)
Extension or Lead 1
Figure 1. Insert the extension or lead connector pins fully into the neurostimulator.
Note:
Insert a connector plug (from an accessory kit) into any unused neurostimulator
socket.
4.
For each extension, lead, or plug, fully insert the torque wrench into each self-sealing grommet of the connector block and tighten each setscrew (Figure 2).
Cautions:
Be sure the torque wrench is fully inserted into the self-sealing grommet. If
the torque wrench is not fully inserted, the setscrew may be damaged, resulting in intermittent or loss of stimulation.
Before tightening setscrews, ensure that the extension or lead connector
pins are inserted into the connector block to prevent damaging the lead or extension.
Verify that each leaf of the self-sealing grommet is closed after the torque
wrench is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the patient may experience shocking, burning, or irritation at the neurostimulator implant location, or intermittent stimulation, or loss of stimulation.
Extension or Lead 2
14 English  97713 2016-06-01
Figure 2. Tightening the setscrews in the self-sealing grommet.

Implanting the neurostimulator

Warning
: Do not enclose a neurostimulator with SureScan MRI Technology in a cover, such as those used to mitigate a titanium allergy. A neurostimulator cover may cause significant heating of the lead electrodes during an MRI scan, placing the patient at risk for tissue heating, resulting in tissue damage or serious patient injury.
Warning
: Do not tie ligatures directly around the lead body. Ligatures around the lead body can damage the lead body or conductor wires, resulting in a loss of therapy. During an MRI scan, the patient could be at risk of tissue heating, resulting in tissue damage or serious patient injury.
Caution
: To prevent device inversion, do not make the neurostimulator pocket any larger than necessary to fit the neurostimulator and excess lead or extension. Device inversion may result in component damage, lead dislodgement, skin erosion, or stimulation at the implant site, requiring repeat surgery to restore therapy.
Note:
For full-body MRI scan eligibility, confirm that the subcutaneous pocket for the
neurostimulator has been created in the buttocks, abdomen, or flank. (See the caution in "Implant criteria for full-body MRI scan eligibility" on page 11.)
1.
With the Medtronic logo facing outward, away from muscle tissue, rotate the neurostimulator counterclockwise to coil the excess lead or extension (Figure 3).
Caution
: Do not twist or kink the lead or extension bodies when rotating the neurostimulator and coiling the excess lead or extension. Twisting or kinking of the components creates a torsional load that may increase the risk of unwanted movement or damage to the neurostimulation system components.
97713 2016-06-01  English 15
Figure 3. Rotate counterclockwise to coil excess lead or extension length.
2.
Insert the neurostimulator and excess lead or extension coils into the subcutaneous pocket with the Medtronic logo facing outward, away from muscle tissue. Ensure that the leads or extensions are not twisted or bent sharply.
Cautions:
Ensure that the neurostimulator is placed no deeper than 1 cm (0.4 in) below
the skin and is parallel to the skin. If the neurostimulator is too deep or is not parallel to the skin, recharge may be inefficient or unsuccessful.
Position the neurostimulator with the white surface facing outward. If
implanted with the white surface facing inward, the neurostimulator cannot be charged.
Do not coil excess extensions or leads in front of the neurostimulator. Wrap
excess extensions or leads around the perimeter (Figure 4) or behind the neurostimulator to help minimize potential damage during neurostimulator replacement surgery, help minimize potential kinking of the extension or lead, and minimize interference with telemetry and recharge operation.
Figure 4. Wrap excess extensions or leads around the perimeter (or behind) the
16 English  97713 2016-06-01
neurostimulator.
3.
Use the suture holes in the connector block to secure the neurostimulator to the muscle fascia with nonabsorbable silk.
Notes:
Secure the neurostimulator in the pocket to minimize movement or migration of
the neurostimulator. Suturing the neurostimulator also may prevent movement of the neurostimulator
from torque and other forces during an MRI scan.

Checking system integrity

1.
To ensure that you have properly connected each extension or lead to the neurostimulator, use the clinician programmer to confirm the integrity of the connected system.
Caution:
To use the nonsterile clinician programmer in a sterile field, place a sterile barrier between the patient and the programming head to prevent infection. Do not sterilize any part of the clinician programmer. Sterilization may damage the programmer.
Note:
The neurostimulator should be in the pocket during system interrogation for
integrity to ensure proper readings. If the system integrity test results are not acceptable, refer to "Connecting the
extension or lead to the neurostimulator" on page 13.
2.
Program the basic stimulation parameters and check the battery status; if applicable, check the electrode impedances to rule out a short or open circuit.
3.
Complete the stimulation assessment form.

Completing the implant procedure

1.
Close and dress all incisions.
2.
Ensure that a patient control device and a completed patient identification card are given to the patient.
3.
Complete the device tracking and patient registration paperwork and return the documents to Medtronic.
97713 2016-06-01  English 17
Contacts:
Asia:
Medtronic International Ltd. Tel. 02919-1300 Fax 02891-6830
Medtronic Asia Ltd. Tel. (02)-548-1148
Fax (02)-518-4786
Australia:
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde, NSW 2113 Australia Tel. +61-2-9857-9000 Fax +61-2-9878-5100 Toll-free 1-800-668-670
Austria:
Medtronic Österreich GmbH Tel. 01-240440 Fax 01-24044-100
Belgium:
Medtronic Belgium S.A. Tel. 02-456-0900 Fax 02-460-2667
Canada:
Medtronic of Canada Ltd. Tel. (1-905)-460-3800 Fax (1905)-826-6620
Czech Republic:
Medtronic Czechia s.r.o. Tel. 2-965-795-80 Fax 2-965-795-89
Denmark:
Medtronic Danmark A/S Tel. 45-32-48-18-00 Fax 45-32-48-18-01
Finland:
Medtronic Finland Oy/LTD Tel. (09)-755-2500 Fax (09)-755-25018
France:
Medtronic France S.A.S. Tel. 01-5538-1700 Fax 01-5538-1800
Germany:
Medtronic GmbH Tel. (02159)-81490 Fax (02159)-8149100
Greece:
Medtronic Hellas S.A. Tel. 210-67-79-099 Fax 210-67-79-399
Hungary:
Medtronic Hungária Kft. Tel. 1-889-06-00 Fax 1-889-06-99
Ireland:
Medtronic Ireland Ltd. Tel. (01)-890-6522 Fax (01)-890-7220
Italy:
Medtronic Italia SpA Tel. 02-241371 Fax 02-241381
Tel. 06-328141 Fax 06-3215812
Japan:
Medtronic Japan Tel. 03-6776-0017 Fax 03-6774-4645
Latin America:
Medtronic, Inc. Tel. (1305)-500-9328 Fax (1786)-709-4244
Norway:
Medtronic Norge AS Tel. 67-10-32-00 Fax 67-10-32-10
Poland:
Medtronic Poland Sp. z.o.o. Tel. (022)-465-69-00 Fax (022)-465-69-17
Portugal:
Medtronic Portugal, Lda. Tel. 21-724-5100 Fax 21-724-5199
Russia:
Medtronic Russia Tel. (8495) 580-7377 Fax (8495) 580-7378
Slovakia:
Medtronic Slovakia, o.z. Tel. 0268 206 911 Fax 0268 206 999
Spain:
Medtronic Ibérica, S.A. Tel. 91-625-0400 Fax 91-650-7410
Sweden:
Medtronic AB Tel. 08-568-585-00 Fax 08-568-585-01
Switzerland:
Medtronic (Schweiz) AG Tel. 031-868-0100 Fax 031-868-0199
The Netherlands:
Medtronic B.V. Tel. (045)-566-8000 Fax (045)-566-8668
Turkey:
Medtronic Turkey Tel. +90 216 636 1000 Fax +90 216 636 1008
U.K.:
Medtronic U.K. Ltd. Tel. 01923-212213 Fax 01923-241004
USA:
Medtronic, Inc. Tel. (1-763)-505-5000 Fax (1-763)-505-1000 Toll-free: (1-800)-328-0810
Manufacturer
Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA www.medtronic.com Tel. +1-763-505-5000 Fax +1-763-505-1000
REP
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10, 6422 PJ Heerlen, The Netherlands Tel. +31-45-566-8000 Fax +31-45-566-8668
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31, Case Postale 84 CH - 1131 Tolochenaz, Switzerland www.medtronic.eu Tel. +41-21-802-7000 Fax +41-21-802-7900
Asia-Pacific
Medtronic International Ltd. Suite 1106-11, 11/F, Tower 1, The Gateway, 25 Canton Road, Tsimshatsui, Kowloon, Hong Kong Tel. +852-2919-1300 Fax +852-2891-6830
Contacts for specific countries are listed inside this cover.
EC
*M940099A007*
© Medtronic, Inc. 2016 All Rights Reserved
M940099A007 Rev A
Loading...