Medtronic 9736349 Instructions for Use

StealthStation FlexENT™
Customer Repair Manual
StealthStation FlexENT™
Customer Repair Manual
D00447579, Rev. A
2021-04
Copyright © 2021 Medtronic Navigation, Inc. All rights reserved.
No part of this documentation may be reproduced in any form or by any means or used to make any derivative work
(such as translation, transformation, or adaptation) without written permission from Medtronic Navigation, Inc.
The information contained in this document is accurate at time of publication. Medtronic Navigation, Inc. reserves the
right to revise this documentation and to make any changes in content from time to time without obligation on the part of
Medtronic Navigation, Inc. to provide notification of such revision or change.
Medtronic Navigation, Inc. provides this documentation without warranty, term, or condition of any kind, either implied or
expressed, including, but not limited to, the implied warranties, terms or conditions of merchantability, satisfactory
quality, and fitness for a particular purpose. Medtronic Navigation, Inc. may make improvements or changes in the
product(s) and/or the program(s) described in this documentation at any time.
O-arm™, StealthStation™ S7 and StealthStation™ S8, and StealthStation FlexENT™ are registered trademarks of
Medtronic Navigation, Inc. All other product or service names are the property of their respective owners.
Refer to manuals.medtronic.com for the current version of this manual.
Table of Contents
Replacement Part Numbers 7
Accessory Power Cable 11
Bifurcated Cable 17
Desk Stand 25
Dual Monitor Mount 31
External Power Cable 37
Flat Emitter Bin 43
IPC Nexus Mount 47
IV Bracket 51
Lower Shelf with Bin 53
Power Strip 57
Push Handle 61
Shelf 65
Shelf with Rails 69
Side Emitter Mount 73
Tap Strip 79
Cable diagram 83
9736349 |D00447579, Rev. A 5 of 84

Replacement Part Numbers

9736349 | D00447579, Rev. A 7 of 84
Replacement Part Numbers
9736349 | D00447579, Rev. A 8 of 84
Replacement Part Numbers
9736349 | D00447579, Rev. A 9 of 84

Accessory Power Cable

Warning: Do not modify this equipment, including system, hardware, or accessories, without authorization of the manufacturer.
Caution: Always order replacement parts fromMedtronic. Only Medtronic replacements have been tested with the system and can be guaranteed to perform as expected. To order replacement parts, contact Medtronic technical support.
Power down and disconnect the accessory.
Follow proper shutdown procedure before performing
service on the cart.
Caution: Removing power from the StealthStation FlexENT™ system before it has been properly shut down can corrupt the system's software causing loss of data and functionality. Do not remove power from the system until the unit has been properly shut down.
9736349 | D00447579, Rev. A 11 of 84
Accessory Power Cable
9736349 | D00447579, Rev. A 12 of 84
Accessory Power Cable
Disconnect all accessory cables from the power strip.
Remove the replacement part from the shipping
container. Inspect the part for damage.
Contact Medtronic technical services if the part is
damaged or incomplete.
Refer to the "Cable diagram" on page83 to check
cable connections.
Caution: Cables must be connected as shown in the
"Cable diagram" on page83. Incorrect cable
connections can cause unexpected equipment performance. Contact Medtronic Technical Service for additional information.
9736349 | D00447579, Rev. A 13 of 84
Accessory Power Cable
Make sure that the circuit breaker is on.
I = On
O = Off
Connect the accessory power cable to the appropriate
accessory on the cart.
9736349 | D00447579, Rev. A 14 of 84
Accessory Power Cable
Power on the All-in-One computer and all accessories.
Make sure that all systems function as expected.
Verify that the accessory with the new power cable
functions as expected.
9736349 | D00447579, Rev. A 15 of 84

Bifurcated Cable

9736318—100-200V
9736319—220-230V
Warning: Do not modify this equipment, including system, hardware, or accessories, without authorization of the manufacturer.
Caution: Always order replacement parts fromMedtronic. Only Medtronic replacements have been tested with the system and can be guaranteed to perform as expected. To order replacement parts, contact Medtronic technical support.
Follow proper shutdown procedure before performing
service on the cart.
Caution: Removing power from the StealthStation FlexENT™ system before it has been properly shut down can corrupt the system's software causing loss of data and functionality. Do not remove power from the system until the unit has been properly shut down.
9736349 | D00447579, Rev. A 17 of 84
Bifurcated Cable
Retain hardware for installation.
9736349 | D00447579, Rev. A 18 of 84
Bifurcated Cable
9736349 | D00447579, Rev. A 19 of 84
Bifurcated Cable
Remove the replacement part from the shipping
container. Inspect the part for damage.
Contact Medtronic technical services if the part is
damaged or incomplete.
9736349 | D00447579, Rev. A 20 of 84
Bifurcated Cable
9736349 | D00447579, Rev. A 21 of 84
Bifurcated Cable
Caution: To avoid potential power loss to the All-in­One computer, only connect the bifurcated power cable to provide power to the computer. Accessory power cables cannot protect the All-in-One computer if the FlexENT™ cart loses power.
9736349 | D00447579, Rev. A 22 of 84
Make sure that the circuit breaker is on.
I = On
Bifurcated Cable
O = Off
Refer to the "Cable diagram" on page83 to check
cable connections.
Caution: Cables must be connected as shown in the
"Cable diagram" on page83. Incorrect cable
connections can cause unexpected equipment performance. Contact Medtronic Technical Service for additional information.
9736349 | D00447579, Rev. A 23 of 84
Bifurcated Cable
Power on the All-in-One computer and all accessories.
Make sure that all systems function as expected.
9736349 | D00447579, Rev. A 24 of 84

Desk Stand

Warning: Do not modify this equipment, including system, hardware, or accessories, without authorization of the manufacturer.
Caution: Always order replacement parts fromMedtronic. Only Medtronic replacements have been tested with the system and can be guaranteed to perform as expected. To order replacement parts, contact Medtronic technical support.
Follow proper shutdown procedure before performing
service on the cart.
Caution: Removing power from the StealthStation FlexENT™ system before it has been properly shut down can corrupt the system's software causing loss of data and functionality. Do not remove power from the system until the unit has been properly shut down.
9736349 | D00447579, Rev. A 25 of 84
Desk Stand
Disconnect other cables as needed.
Caution: Remove all sharp objects and tools from the bench before putting the monitor screen on the bench.
Remove the replacement part from the shipping
container. Inspect the part for damage.
Contact Medtronic technical services if the part is
damaged or incomplete.
9736349 | D00447579, Rev. A 26 of 84