Medtronic 9735666 Instructions for Use

Page 1
StealthStation™ S8 Planning Station System Manual
Read this manual completely before using this device.
Page 2
No part of this documentation may be reproduced in any form or by any means or used to make any derivative work (such as translation, transformation, or adaptation) without written permission from Medtronic Navigation, Inc.
Page 3
1.Introduction
Intended use/indications for use 3 Contraindications 3 Side effects 3 Material composition 3 Description of the StealthStation System installation 4 Use profile 4 Potential complications 4 Clinical benefits 4 Conventions 4 Warnings and precautions 5
Warnings 5
Precautions 5 Content of this manual 6 Related documents 6 Reporting 6 Contact information 7
Telephone 7
Mailing address 7
Medtronic E.C. authorized representative 7
Australia 7
World Wide Web 7
E-mail 8
Planning Station 3
2.System overview
How the system works 9 System components 9
Computer 9
Touchscreen monitor 10
Keyboard and mouse 10
3.System Operation
Computer Input/Output panel 11
Computer I/O panel connections 11 System security 14
StealthStation™ S8 Planning Station security functionality 14
Best practices to maintain security of the StealthStation
What to do if you suspect the StealthStation™ S8 Planning Station has been compromised 14 Network connection information 15
Connecting to a wired network 15
Configuring user accounts and changing passwords 16 System set up 18
Starting the system 18
Launching the software 18 System shutdown 19
Exiting the software 19
Shutting down the system 19 Cleaning the planning station 19 Storing the planning station 19
S8 Planning Station 14
4.System specifications
1
Page 4
System specifications 21 System electromagnetic emissions and immunity declarations 22 Electromagnetic emissions and immunity declarations 22
Warnings and precautions 22
5.Troubleshooting
Component connections 25
Running the self-test tool 25
System power 25
No power to system 25
Monitor 26
The monitor screen is blank 26 The monitor screen displays ‘no video detected’ 26
The display on the monitor is distorted 26 Mouse and keyboard 26 Recommended maintenance 27
Replacing the power cord and cables 27 System disposal 27
6.Symbols
Explanation of symbols on package labeling 29
2
Page 5

1 Introduction

Intended use/indications for use

The StealthStation™ Planning Station is intended as a planning tool to aid in preparation for surgery. The planning station can also be used to send and receive patient exams across the hospital network. The StealthStation viewing of DICOM datasets and pre-planning for the use of navigation may be appropriate.

Contraindications

Medical conditions which contraindicate the use of a Medtronic computer-assisted surgery system and its associated applications include any medical conditions which may contraindicate the medical procedure itself.

Side effects

The Medtronic computer assisted surgery system and its associated applications do not independently produce any side effects. When used in conjunction with navigated instruments for surgical procedures, the associated side effects are the same as those for the procedure itself. See the procedure pocket guides for detailed lists of the side effects associated with these procedures.
Planning Station is indicated for any medical condition in which

Material composition

Material contained in product that can cause an allergic reaction: nickel.
For additional materials of concern information such as REACH, CA Prop 65, or other product stewardship programs, go to www.medtronic.com/productstewardship.

Description of the StealthStation™ Planning Station

The StealthStation™ Planning Station is intended to provide a consistent experience between preoperative planning and navigation with the StealthStation access to planning features in the StealthStation users the flexibility to plan for surgical procedures at any time. In addition, the planning station can be used to send and receive patient images across the hospital network. Surgical plans created on the StealthStation
The StealthStation™ Planning Station is comprised of a computer with an operating system, a monitor with a stand, a keyboard, and a mouse.
Planning Station can be transferred to the StealthStation™ navigation system.
software on a desktop computer, which gives
system. The planning station provides
3
Page 6
Introduction

System installation

System installation
The StealthStation™ S8 Planning Station is installed by qualified Medtronic personnel only. If you have any questions about your system installation, contact Medtronic Navigation

Use profile

The intended user profile for the StealthStation™ system is operating room staff, including surgeons, nurses, surgical technologists, and hospital biomedical engineers.
The intended patient population for the system is patients for whom stereotactic image guided surgery is appropriate.

Potential complications

Potential complications related to the use of a Medtronic computer-assisted surgery system and its associated applications are the same as the potential complications for the procedures in which the system is used. See the procedure pocket guides for detailed lists of the potential complications associated with these procedures.

Clinical benefits

.
Clinical benefits related to the use of a Medtronic computer-assisted surgery system and its associated applications are the same as the clinical benefits for the procedures in which the system is used. See the procedure pocket guides for detailed lists of the clinical benefits associated with these procedures.

Conventions

This document employs the following conventions:
Warnings are preceded by the word Warning. Failure to observe a warning may result in physical
Precautions are preceded by the word Caution. Failure to observe a precaution could result in
References to buttons that appear on the system display are enclosed in square brackets. For
References to menu options that appear on the system display are printed in bold letters. For
injury to the patient or operator. Pay special attention to these items.
damaged equipment, forfeited time or effort, or the need to abort use of the system.
example:
Click [Edit...].
example:
Choose Clear from the list.
4
Page 7
Instructions to click an object on the screen means to place the pointer over the object using the system mouse, and press and release the left mouse button. Instructions to tap an object on the screen means to tap the screen with a finger, a gloved finger, or a capacitive stylus. Click and tap are used interchangeably.

Warnings and precautions

Warnings
Warning: The system hardware and software should be used only by qualified medical
professionals who are trained in performing surgery and familiar with computer-assisted surgery systems.
Warning: Inspect all visible system components for damage before use. Do not use any component
that is visibly damaged.
Warning: To avoid the risk of electric shock, this equipment must only be connected to a grounded
power outlet.
Introduction
Warnings and precautions
Precautions
Warning: Avoid dripping any fluids into any enclosure on the StealthStation
Shut down the system, disconnect the power, and allow the system to dry if you suspect fluids may have entered any part of the system.
Warning: The system is not suitable for use in the presence of a flammable, anesthetic mixture with
air, oxygen, or nitrous oxide.
Warning: Do not modify the StealthStation
Medtronic Navigation, Inc. Prohibited modifications include but are not limited to: altering, repairing, or replacing system components; altering software provided by Medtronic; updating the operating system or drivers; and installing any software that is not provided by Medtronic, including anti-virus software. Secure the system to prevent unauthorized modifications. Unauthorized modifications could render the system unsafe or ineffective for its intended use.
Warning: Follow system component maintenance and care instructions.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Caution: The system mouse and keyboard are not designed for sterilization and may be damaged
if sterilization is attempted.
Caution: System components are fragile. Use care when handling system components.
S8 Planning Station without prior written consent from
S8 Planning Station.
5
Page 8
Introduction

Content of this manual

Content of this manual
This system manual is intended as a reference document for biomedical engineers or other qualified personnel who require familiarity with and details about the StealthStation This manual is not a software usage manual. For instructions on using specific software, refer to the instructions for use provided with the software.

Related documents

For software instructions, refer to the specific software’s instructions for use.

Reporting

If a serious incident occurs in relation to the use of this Medtronic computer-assisted surgery system, report it to Medtronic Navigation. If a serious incident occurs in the European Union, also report it to the competent authority in the Member State where the incident occurred.
S8 Planning Station.
6
Page 9

Contact information

Telephone
Technical support worldwide +1 720 890 3160
Technical support in the USA +1 800 595 9709
General +1 720 890 3200
Fax +1 720 890 3500
Mailing address
Medtronic Navigation, Inc.
826 Coal Creek Circle
Louisville, CO, U.S.A.
80027
Introduction
Contact information
Medtronic E.C. authorized representative
Medtronic B.V.
Earl Bakkenstraat 10
6422 PJ Heerlen
The Netherlands
Tel. +31 45 566 80 00
Australia
Medtronic Australasia Pty Ltd
2 Alma Road
Macquarie Park, NSW 2113
Australia
Tel. 1800 668 670 (within Australia)
World Wide Web
www.medtronic.com
7
Page 10
Introduction
Contact information
E-mail
E-mail product enhancement requests to: dl.navsuggestions@medtronic.com
8
Page 11

2 System overview

How the system works

The StealthStation™ S8 Planning Station is a desktop workstation comprised of a computer that is compatible with the StealthStation planning station may be used as a temporary server for assembling and transferring patient data before and after a procedure.
Patient images may be transferred to the planning station from a CT scan, MR scan, or other compatible DICOM imaging modality. The software reconstructs the image data to display the patient anatomy from a variety of perspectives (axial, sagittal, coronal, oblique) and 3-dimensional renderings. The planning station provides surgeons the flexibility to plan for surgical procedures at any time and without utilizing a full surgical StealthStation™ system. Surgical plans created on the StealthStation or to a PACS.
S8 Planning Station can be transferred to the full surgical StealthStation™ S8 system

System components

S8 software, a touchscreen monitor, keyboard, and mouse. The
Figure 1: StealthStation™ S8 Planning Station components
Computer
The StealthStation™ S8 Planning Station desktop computer is compatible with the StealthStation™ S8 software. The computer includes a CD/DVD drive, USB ports, and an Ethernet port for data input and output.
9
Page 12
System overview
System components
Touchscreen monitor
The touchscreen monitor is a high-resolution, flat panel computer display with built-in speakers. The capacitive touchscreen interface can be controlled by a finger or gloved finger.
Caution: Do not use a hard object or a sharp object to control the touchscreen interface because
the object may damage the screen.
Keyboard and mouse
A wired USB keyboard and a wired mouse are provided with the StealthStation™ S8 Planning Station. They are used to control the system, and they can be plugged into any USB port on the planning station computer.
Warning: Only use the Medtronic supplied keyboard and mouse with the StealthStation
Planning Station. Connecting incompatible items could render the system unsafe or ineffective for its intended use.
S8
10
Page 13

3 System Operation

Computer Input/Output panel

The planning station computer includes input/output panels with external connection ports for various input and output devices. The panels are on the front and rear of the computer and under the touchscreen monitor.
Warning: Connect only items that are specified as part of the StealthStation
or compatible with the system. Connecting incompatible items could render the system unsafe or ineffective for its intended use.
Computer I/O panel connections
1 CD/DVD-ROM drive 2 CD/DVD-ROM eject button
Figure 2: Front I/O panel
S8 Planning Station
3 Power button 3 2 super-speed USB 3.0 ports for loading patient scans onto the system
11
Page 14
System Operation
Computer Input/Output panel
Figure 3: Rear I/O panel
1 USB 2.0 ports for the keyboard, the mouse, and the touch interface on the
touchscreen monitor
2 Display port: Disabled 3 Display port: Disabled 4 Display port: Disabled 5 Display port: Disabled 6 USB-C: Connects the computer to the touchscreen monitor via a USB-C to HDMI
adapter
7 Ethernet: Connects the system to the site Local Area Network (LAN) 8 Ethernet: Disabled 9 Power cord
Note: Do not use disabled ports 2 (Display port), 3(Display port), 4(Display port), 5(Display
port) and
8(Ethernet).
12
Page 15
Figure 4: I/O panel under the touchscreen monitor
1 Power 2 HDMI: Connects the touchscreen monitor to the computer 3 DVI-I: Disabled
System Operation
Computer Input/Output panel
4 Display port in: Disabled 5 Audio in: Disabled 6 Audio out: Disabled 7 VGA port: Disabled 8 USB Type B: Connects the touchscreen monitor to the computer 9 Functional ground
Note: Ports
S8 Planning Station or compatible with the planning station.
The user assumes responsibility for verifying the output quality of any attached devices that were not supplied for this use by Medtronic.
3 through 7 are not used for any item that is specified as part of the StealthStation
13
Page 16
System Operation

System security

System security
Good security practices are needed to protect both patient data and the integrity of any network­connected product. The StealthStation facilitate management of security. These features work in conjunction with the security practices of hospitals and clinics to provide safe and secure operation of the StealthStation and to protect attached networks and devices.
S8 Planning Station includes features and functionality that
S8 Planning Station
StealthStation™ S8 Planning Station security functionality
The following features and functions help to protect the integrity of the system, the data it contains, and any networks or other devices the system may be connected to:
• Access control: The StealthStation ability to include password protection, for different system user types. User accounts may also be integrated with hospital network privileges. Controlling access to the StealthStation Planning Station has the ability to prevent unauthorized user access to the system, the data contained on the system, and other devices that may be networked with the system.
• Encryption: The StealthStation medical images and surgical plan information when the system is not in use.
• Antivirus: With the antivirus product that is installed on the system by Medtronic, users can update virus definitions and run an on-demand, on-disk antivirus scan to identify potential threats.
S8 Planning Station has specified user accounts, with the
S8 Planning Station encrypts directories containing patient
S8
Best practices to maintain security of the StealthStation™ S8 Planning Station
Maintain good physical controls over the StealthStation™ S8 Planning Station. Keeping the system in a secure location and environment reduces the probability of unauthorized access and unauthorized modifications to the system.
Only connect the StealthStation
Only install Medtronic supplied updates and patches for security purposes.
S8 Planning Station to managed, secure networks.
What to do if you suspect the StealthStation™ S8 Planning Station has been compromised
If you suspect that the system has been compromised by a security threat, follow these steps:
1. Shut down the system.
2. Disconnect the StealthStation
3. Contact Medtronic Navigation
potential security threats.
S8 Planning Station from any networks or other devices.
technical support for further assistance in troubleshooting
14
Page 17

Network connection information

A network connection is provided on the StealthStation™ S8 Planning Station. The network connection is used to transfer data, including patient images and surgical plans, to and from the system. The network connection may also provide the ability for Medtronic Navigation support to monitor and troubleshoot remotely, if this capability has been authorized and configured. The user network must be 10/100/1000 BASE-T to function with the StealthStation Station.
System Operation
Network connection information
technical
S8 Planning
For network connection, the StealthStation the background that is always checking for DICOM data being pushed to the system. If this application is not running, no data will be accepted. If there is a network failure, a CD/DVD-ROM or a USB drive may be used to transfer images to or from the system. Navigated procedures may be delayed while alternate media is used for image transfer, or navigation may be aborted if alternate image transfer media is unavailable.
Connecting the StealthStation could result in previously unidentified risks to patients, operators, or third parties. Users should identify, analyze, evaluate, and control these risks. Subsequent changes to the network could reduce the functionality of the system or introduce new risks, and therefore require additional analysis. The following activities are examples of changes that could change the risk profile and require analysis:
Changing the network configuration
Connecting additional items to the network
Disconnecting items from the network
Upgrading equipment connected to the network
See “Configuring antivirus scan settings” on page 16 to configure antivirus scan settings and control identified risks.
Connecting to a wired network
S8 Planning Station includes an application running in
S8 Planning Station to an IT network that includes other equipment
Follow these steps to connect the StealthStation™ S8 Planning Station to a network:
1. Plug in an Ethernet cable.
2. Follow these steps to open the StealthStation Configuration tool.
a. At the login screen, select Other and then log in using the Stealth Admin (stealthadmin)
username and password.
b. Double-tap StealthStation Configuration.
c. Enter the Stealth Admin password and tap [OK].
3. Select the Network Settings tab.
4. Tap [Configure Network Settings].
5. Select DHCP or Static.
If you choose DHCP, tap [OK]. You do not need to fill in any information.
15
Page 18
System Operation
Network connection information
If you choose Static, fill in Static IP, Netmask, Default Gateway, and Primary DNS. Optionally,
fill in Secondary DNS and Tertiary DNS. These fields are limited to numbers and periods only. Ta p [OK].
Current Network Configuration information displays.
Configuring user accounts and changing passwords
The StealthStation™ S8 Planning Station is configured with 3 accounts: Stealth Admin, Stealth User, and StealthViz User. Each of these 3 accounts has a default password, which can be changed. For assistance with the default passwords for these accounts, contact Medtronic Navigation support.
Follow these steps to configure accounts for users to log in to the system.
1. At the login screen, select Other then log in using the Stealth Admin (stealthadmin) username
and password.
2. Double-tap StealthStation Configuration.
3. Enter the Stealth Admin password and tap [OK].
technical
4. Tap the System Settings tab.
5. Tap [Change Date and Time], and then use the drop-down menus to change the date and the
time zone. To change the time, type in the time.
Set the time before connecting to Active Directory so that the time will be associated with the Active Directory server. The time needs to be within 5 minutes of the World Clock time and the Active Directory server time.
6. Tap the Authentication Settings tab.
7. Choose either Local Authentication or Active Directory Authentication.
If you choose Local Authentication, then there are 3 password options.
– New Stealth Admin Password
– New Stealth User Password
– New StealthViz User Password
To change a password, enter the password and tap [Reset].
If you choose Active Directory Authentication, obtain information from the hospital IT
administrator to fill in all fields, and then tap [Join Domain].
– If joining the domain is successful, tap [OK]. Then tap [Change Password] to change
one of the 3 passwords.
– If joining the domain fails, tap [Show Details] to display additional information that you
can provide when you call Medtronic Navigation
technical support.
16
Configuring antivirus scan settings
An antivirus product is installed on the StealthStation™ S8 Planning Station. The default setting is to scan the root directory (/). To select a different directory to scan or to exclude directories or files from antivirus scans, follow these instructions.
Page 19
System Operation
Network connection information
Note: For instructions on updating antivirus definitions, see “Updating antivirus definitions” on page
17.
1. At the login screen, select Other then log in using the Stealth Admin username and password.
2. Double-tap StealthStation Configuration.
3. Enter the Stealth Admin password and tap [OK].
4. Tap the Antivirus Settings tab.
5. Choose the directory to scan.
6. Choose directories or files to exclude from scans.
7. Tap [Configure].
If the Enter Directory to Scan field is empty or if any directories or files are invalid, an error message appears.
Updating antivirus definitions
An antivirus product is installed on the StealthStation™ S8 Planning Station. Follow hospital IT policies regarding how frequently to update the antivirus definitions and scan the system according to these instructions:
1. At the login screen, select Other and then log in using the Stealth Admin username and
password.
2. Double-tap StealthStation Self-Test.
3. Make sure that the StealthStation
system is connected to the Internet.
4. Tap the Antivirus Status tab.
To update antivirus definitions, tap [Update Antivirus Definitions].
To scan the system, tap [Scan].
The default setting is to scan the root directory (/). To select a different directory to scan or to exclude directories or files from antivirus scans, see “Configuring antivirus scan settings” on page 16.
– If antivirus definitions have never been updated, a message displays in the Last
Antivirus Definition Update field, and the [Scan] button is disabled. To enable scanning, update antivirus definitions.
– If the scan fails, tap [Details] to display additional information that you can provide when
you call Medtronic Navigation
technical support.
17
Page 20
System Operation

System set up

Remote Support
Remote Support is a digital service that allows Medtronic to provide features such as: remote diagnostics, system configuration backup and restoration, predictive part failure reporting, remote troubleshooting support using a real-time-screen-share feature, and remote software install. Remote Support requires an active, secured Internet connection between the Covered Product and Medtronic for connectivity. Medtronic Remote Support is securely enabled by the Medtronic Pathway Connect device on-site to Medtronic Pathway Connect
443. The StealthStation
server when the following conditions are met:
System is powered on.
System is connected to the Internet via Ethernet or Wi-Fi.
server, Medtronic’s digital service cloud solution. All communication originates from the
System will attempt to communicate with Medtronic Pathway Connect™
server as an outbound HTTPS connection on port
The screen-share function requires the person operating the StealthStation screen-share session by physically opening the Remote Support application in the StealthStation Configuration tool with the assistance of Medtronic Technical Support. When a session is initiated, the Medtronic Pathway Connect Technical Support agent and the system for the duration of the session using certificate-based authentication.Contact Medtronic Technical Support for additional information about enabling Remote Support on your system.
System set up
Warning: Prevent fluid from entering any part of the StealthStation™ S8 Planning Station. If you
suspect fluid has entered any part of the unit, turn the system off immediately using the steps in “System shutdown” on page 19, disconnect from electrical power, and allow adequate drying time before powering the system back on.
Starting the system
1. Press the power button on the monitor.
2. Press and hold the LED power switch located on the front of the computer.
The power switch illuminates blue when the system is on.
system to initiate a
server creates a point-to-point connection between the Medtronic
Launching the software
Warning: Before using the system, verify monitor functionality by ensuring that the monitor responds
to touch and displays the expected screen output.
If the touchscreen is not functioning properly, use alternative input methods (mouse and keyboard). Call Medtronic Navigation
1. Type in your username and password to log in to the system (if required at your facility).
2. Reference the software pocket guide for instructions on launching the software.
18
technical support to report touchscreen functionality issues.
Page 21

System shutdown

Exiting the software
Reference the software pocket guide for instructions on exiting the software.
Shutting down the system
After use, shut down the system. Follow these steps to shut down the system.
1. At the login screen, tap the gear icon in the upper right corner of the screen to access the drop-
down menu.
2. Tap [Shut Down].
The computer will fully shut down.
3. Press the power button on the monitor to shut down the monitor.
Emergency shutdown
System Operation
System shutdown
If you need to shut down the system quickly in an emergency, or if you are not able to exit the software, follow this procedure:
1. Press and hold the blue LED power button on the computer until the computer shuts down.
2. Press the power button on the monitor to shut down the monitor.

Cleaning the planning station

For instructions on cleaning the system, refer to the StealthStation™ S8 cleaning instructions (9735727).

Storing the planning station

When the planning station is not in use, store the planning station in an environment within the following temperature and humidity ranges:
Storage temperature 15° to 30°C (59° to 86°F)
Relative humidity 20% to 80% Non-condensing
19
Page 22
System Operation
Storing the planning station
20
Page 23

4 System specifications

System specifications

The specifications listed apply to system operation under typical conditions.
Tab le 1: StealthStation™ S8 Planning Station specifications
Operating temperature 18° to 30°C (64° to 86°F)
Operating pressure 101.3 kPa to 69.7 kPa
Operating altitude -50 m (-164 ft) minimum
3000 m (9842 ft) maximum
Shipping temperature -29° to 60°C (-20° to 140°F)
Storage temperature 15° to 30°C (59° to 86°F)
Input voltage (computer and monitor)
Input frequency (computer and monitor)
Maximum power (computer)
Maximum power (monitor) 60 W
Typical power dissipation (computer)
Typical power dissipation (monitor)
Recommended power supply cord
Recommended overcurrent protection
Relative humidity 20% to 80% Non-condensing
Monitor dimensions 389 mm high by 648 mm wide by 63 mm deep (15.3 in high by
~100–~230 VAC
50–60 Hz
920 VA
200 W
50 W
#18AWG (0.75mm
15A
25.5 in wide by 2.5 in deep)
2
) minimum
Monitor stand dimensions 29.2 cm wide by 35.6 cm high by 24.1 cm deep (11.5 in wide by 14 in high
by 9.5 in deep)
Monitor weight 10.93 kg (24 lbs)
Monitor display Resolution = 2560 x 1440 pixels, 60 Hz
21
Page 24
System specifications

System electromagnetic emissions and immunity declarations

Computer footprint 22.9 cm wide by 34.3 cm deep by 21.6 cm high (9 in wide by 13.5 in deep
Computer weight approximately 9.0 kg (19.8 lb)
by 8.5 in high)
System electromagnetic emissions and immunity declarations
Tab le 2: Guidance and Manufacturer's Declaration - Cables, Transducers, and Accessories
The listed I/O cables, transducers, and accessories have been determined by Medtronic to be compliant with the
emissions and immunity requirements.
Medtronic Part
Number
System Equipment
M704430B667 Computer NA NA
M704430B336 Monitor NA NA
M704430B013 Keyboard NA NA
M704430B235 Mouse NA NA
Cables
Description Cable Length Shielded (Y/N)
Generic Ethernet cable < 3.0 m (9 ft 8 in) Yes
M704430B035 HDMI cable 2.4 m (8 ft) Yes
M010929C001 USB-C to HDMI video converter 93 mm (3.7 in) No
M704430B036 USB cable for touchscreen 2.0 m (6 ft 6 in) Yes
M704430B237 USB cable for keyboard 1.8 m (6 ft) Yes

Electromagnetic emissions and immunity declarations

Note: The StealthStation™ S8 Planning Station has undergone Electromagnetic Compatibility (EMC)
testing and, in expected use conditions, is unlikely to impact other systems or devices.
Note: The StealthStation
for a Class B digital device, pursuant to Part 15 of the FCC Rules. This device may not cause harmful interference and this device must accept any interference received, including interference that may cause undesired operation.
Warnings and precautions
Warning: RF emitters typically found in an operating room can affect the StealthStation™ S8
Planning Station.
S8 Planning Station has been tested and found to comply with the limits
22
Page 25
System specifications
Electromagnetic emissions and immunity declarations
Warning: RF interference, including RFID or other low frequency technologies, could hamper
operation of the StealthStation™ S8 Planning Station or the operation of other nearby electrical devices. If you suspect either of these conditions, move the conflicting equipment farther apart, separate the equipment with an RF barrier, or discontinue use of the system.
Caution: The StealthStation
S8 Planning Station needs special precautions regarding EMC (electromagnetic compatibility) and needs to be installed and put into service according to the EMC information provided in table 2 in order to remain compliant.
Caution: The use of accessories and cables other than those specified, with the exception of cables
sold by Medtronic Navigation increased emissions or decreased immunity of the StealthStation
as replacement parts for internal components, may result in
S8 Planning Station.
23
Page 26
System specifications
Electromagnetic emissions and immunity declarations
24
Page 27

5 Troubleshooting

Component connections

System malfunctions are sometimes the result of loose, misconnected, or disconnected cables. Do not disconnect any cables unless instructed to do so by a Medtronic Navigation representative.
Running the self-test tool
Follow these steps to test whether system components are properly connected and configured.
Note: The StealthStation
components listed in this section may not be included in your system configuration. Therefore, “Unavailable” status for certain components may be normal in some system configurations.
1. Open the StealthStation Self-Test tool.
a. At the login screen, select Other and then log in using the Stealth Admin username and
password
b. Double-tap StealthStation Self-Test.
2. Tap the Software Information tab to view the status of operating system services and user
input devices such as the keyboard, mouse, and touchscreen.
Green “Connected” status indicates proper connection.
Red status indicates a connection problem. If a user input device has red status, take the following actions:
Check cable connections.
technical support
S8 system is available in multiple configurations. All of the system
Plug the device in to a different port.

System power

No power to system
Is the system plugged in to an electrical power outlet? If not, plug the power cord into an electrical power outlet.
Is the system plugged into a power outlet that supplies power? Test the outlet with a multimeter or a wall power circuit tester. If there is no power, use an electrical power outlet that is supplying power.
25
Page 28
Troubleshooting

Monitor

Check all system connections. Check cables for crimps or damage. Check connector housing and pins for bent or broken components. Contact Medtronic technical support to replace any damaged cables. If necessary, power down the system and re-connect all external cables and devices (monitor, keyboard, and mouse) and turn the system power back on.
Monitor
The monitor screen is blank
Note: The monitor enters sleep mode after periods of inactivity. Move the mouse to make sure that
the monitor is not in sleep mode.
The system is off or the monitor power is off. Verify that both the system and monitor are receiving power. The monitor is turned on and receiving a signal if you can hear a hum emanating from the back of the monitor.
The monitor screen displays ‘no video detected’
Adjust the screen display settings with the monitor’s on-screen display button.
1. Press the on-screen display button located below the on-screen display symbol on the right side
of the monitor.
Figure 5: On-screen display symbol
2. Use the up/down display buttons located next to the on-screen display button to move through
the display menu.
3. Press the on-screen display button to choose Video Input Select.
4. Use the up/down display buttons to move through the Video input select menu.
5. Press the on-screen display button to select HDMI Mode.
The display on the monitor is distorted
The contrast settings are incorrect in the software. Adjust the Level (brightness) and Width (contrast) settings for images using the on-screen controls in the application software.
If adjusting Level and Width does not resolve the problem, shut down and reboot the system.
If the problem persists, call Medtronic Navigation
for technical support.

Mouse and keyboard

The mouse and keyboard are optional input devices. Make sure that the mouse or keyboard cable is properly plugged in on both ends and is connected to a USB port on the system. If a mouse or a keyboard is connected to a USB port on the system, and the system is not responding to input from the device, follow these troubleshooting instructions.
26
Page 29
If the mouse buttons are not operational, the mouse may have been disconnected at start up. Reconnect the mouse to the system computer and reboot the system.
Turn the system power off, wait 10 seconds, and turn the system power back on.
If the cursor does not move on the screen, the computer is not responding. Reboot the system.
If the keyboard is not operational, the keyboard may have been disconnected at start up.
Reconnect the keyboard to the system computer and reboot the system.

Recommended maintenance

Because the StealthStation™ S8 Planning Station contains no user-repairable parts, the interior of the system is normally inaccessible. However, it may occasionally be necessary for a qualified Medtronic service person to remove panels and access interior components.
Troubleshooting
Recommended maintenance
The StealthStation which is consumed during operation.
The StealthStation and annual inspection and testing.
Please contact Medtronic Navigation, Inc. (see page 7) to schedule a full maintenance and system check appointment.
Inspection should include:
Inspect cables for damage, cuts, or connector wear.
Inspect the following items for damage, wear, or unexpected behavior: monitor and computer.
Testing should include:
Verify full functionality of keyboard, mouse, computer, monitor, and fans.
Verify full functionality of the software including image load, image settings, and file export.
Verify system electrical safety.
S8 Planning Station contains no user-serviceable components and no material
S8 Planning Station and its associated components require cleaning as needed
Replacing the power cord and cables
If the power cord or a cable is damaged, contact Medtronic Navigation™ technical support to obtain a power cord or schedule a cable replacement.

System disposal

Warning: Some system components contain batteries. Do not recharge or disassemble batteries
that have been removed from system components. Observe local regulations concerning battery disposal.
Do not dispose of the StealthStation waste stream. Dispose of this product according to local regulations. See http://recycling.medtronic.com for instructions on proper disposal of this product.
S8 Planning Station or components in the unsorted municipal
27
Page 30
Troubleshooting
System disposal
28
Page 31

6 Symbols

Explanation of symbols on package labeling

The following symbols may appear on system equipment, on system packaging, or in this document.
NRTL/MET-Cus Mark indicates compliance to IEC 62368-1 covering electrical safety requirement for the U.S.
This device complies with the China Compulsory Certification.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician. Caution
Consult instructions for use.
Use by date
Batch code
Serial number
Catalog number
Do not sterilize.
Manufacturer
Authorized representative in the European Community
Date of manufacture
Keep dry.
Do not use if package is damaged.
Fragile, handle with care.
29
Page 32
Symbols
Explanation of symbols on package labeling
Keep upright.
Do not stack.
Power on. Connect to main power.
Power off. Disconnect from main power.
Power on for part of the system
Power off for part of the system
Functional ground
-29 C
-20 F
15 C 59 F
60 C 140 F
30 C 86 F
Storage temperature between 15°C and 30°C (59°F and 86°F)
Shipping temperature between -29°C and 60°C (-20°F and 140°F)
Atmospheric pressure limitation between 101.3 kPa and 69.7 kPa. with an elevation range of 0 m to 3000 m
Humidity limitation between 15% and 90% for shipping
Humidity limitation between 20% and 80% for system operation and storage
Do not disassemble.
USB 2.0 port
Super-speed USB 3.0 port
Protect from heat and radioactive sources.
30
Do not dispose of this product in the unsorted municipal waste stream. Dispose of this product according to local regulations. See http://recycling.medtronic.com for instructions on proper disposal of this product.
Chinese Standard (SJ/T11364-2006) Logo: Electronic Information Products Pollution Control symbol. (The “e” in the logo signifies none of the restricted substances are present above permitted levels)
China RoHS compliant. Environmental protection use period of 50 years. Environmental protection use period of 5 years.
Page 33
Medical device.
Importer.
Symbols
Explanation of symbols on package labeling
31
Page 34
Page 35
Page 36
Medtronic Navigation, Inc. 826 Coal Creek Circle Louisville, Colorado 80027 USA Main 720 890 3200 Fax 720 890 3500 Technical Support U.S. 1 800 595 9709 International 1 720 890 3160 www.medtronic.com www.manuals.medtronic.com
©2021 Medtronic, Inc
All Rights Reserved
9735904