StealthStation™ S8 System Quick Reference Guide
Description
This quick reference guide provides an overview of the StealthStation™ S8 system. For detailed system information, including all warnings and precautions, refer to the system manual (9735573). For specific software instructions, refer to the software’s instructions for use.
System carts
StealthStation™ S8 System Quick Reference Guide
Main cart monitor 8 Docking release lever
1
Camera lift arm 9
2
Camera 0 Front storage bin
3
Camera lower arm -
4
Camera cart monitor = Caster locks
5
Work surface q Cord wraps
6
Storage drawer w Main cart lower arm
7
Work surface that slides into a horizontal position
Main cart CD/DVD drive and keyboard storage
Setting up and starting the system
Separating the carts (optional)
Make sure that the system is on a level surface before attempting to separate the carts.
1. Disconnect and store any loose cables.
2. While standing behind the camera cart, pull up on one or both of the docking release levers located under the handle on both sides of the camera cart.
1:
1 Cart docking release lever
2 9735731-G02 Rev. 1
StealthStation™ S8 System Quick Reference Guide
3. Separate the carts with a gentle tug.
Positioning the main cart monitor
1. Grasp the sides of the monitor, and pull the monitor toward the front of the cart to release it from its stowed state.
2. Move the monitor to the desired position.
Connecting and starting the system
1. If you are using the camera cart in addition to the main cart, connect the carts with the cart-to-cart cable before you start the system.
Unwrap the cart-to-cart cable on the camera cart and connect it to the cart cable connections located on the camera cart I/O panel and on the main cart communication I/O panel.
2. Unwrap the power cord from the main cart and plug it into an electrical outlet.
When positioning the main cart, make sure to maintain easy access to the power outlet in case it becomes necessary to unplug the system quickly.
3. Press and hold the LED power switch located on either the main cart deck or the camera cart deck for one second.
When the carts are connected by the cart-to-cart cable, pressing the power button on either cart powers up both carts. The system powers up and the login screen displays when all system bootup diagnostics are complete.
Positioning the camera
1. Pull the camera handle away from the plastic clip to release the camera from its stowed state.
2. Move the camera to the desired position.
Connecting either the optical or the EM hardware to the main cart
For optical localization, connect the optical hardware
1. If you are using a wired (active) patient reference frame, connect it to port A on the main cart localization I/O panel.
2. If you are using a microscope bracket or ultrasound tracker, connect the cable to port B on the main cart localization I/O panel.
3. If you are using a wired (active) instrument, connect it to port C on the main cart localization I/O panel.
4. Connect the footswitch to the footswitch port on the main cart I/O panel.
For electromagnetic localization, connect the EM hardware
1. Connect the EM instrument interface to the EM instrument interface port on the main cart localization I/O panel.
9735731-G02 Rev. 1 3