StealthStation™ S8 cleaning instructions
General procedure
directly on the system. Disconnect from power and allow the system to dry if you suspect fluids may have entered any part of the unit.
1. Power down and unplug the system before performing cleaning or disinfection procedures.
2. Refer to Table 1 for compatible disinfectants and recommended chemistries. If a premoistened wipe is not available, use a clean wipe and moisten with the disinfectant. Do not over saturate the wipe so much that liquid can enter openings in the parts to be cleaned.
Table 1: Compatible cleaner and disinfectant chemistries
Disinfectant type Disinfectant level
Quaternary ammonium germicidal detergent solution Low level
Phenolic germicidal detergent solution Low level
Isopropyl alcohol (70%) Low level
Sodium hypochlorite (5.25% to 6.15%) Low level
3. Wipe the external surfaces with a clean wipe moistened with water or 70% isopropyl alcohol to remove disinfectant residue on the carts. This will help maintain the appearance of the carts.
Note: The StealthStation™ S8 system is available in several configurations. All of the
system components described in this document may not be included in your system configuration.
Cleaning the StealthStation™ S8 system carts and planning station
1. Power down and unplug the system before performing cleaning or disinfection procedures.
2. Wipe and clean the exterior surfaces of the main cart, camera cart, and planning station with a moistened low-level disinfectant wipe per the manufacturer’s instructions.
StealthStation™ S8 cleaning instructions
3. If disinfection is desired, make sure that the surface to be disinfected remains in contact with the disinfectant for the specified contact time as directed by the manufacturer.
- The contact time refers to the minimum amount of time the surface needs to
remain visibly wet.
- Use additional wipes, if needed, to maintain continuous contact time with
the disinfectant.
4. Examine the surfaces for visible soil. If soil is present, repeat cleaning.
5. Wipe the external surfaces with a clean wipe moistened with water or 70% isopropyl alcohol to remove disinfectant residue.
1: Figure 1: Examples of frequently touched surfaces on StealthStation™ S8 system carts
Main cart monitor
1
Work surface that slides into a horizontal
2
position
Main cart handles
3
Camera handle
4
Camera cart monitor
5
Camera cart drawer façade
6
2 9735727-G02 Revision 1
Camera cart handle
7
EM instrument interface handle
8
EM instrument interface cable
9
Side-mount emitter
0
Side-mount emitter cable
-
StealthStation™ S8 cleaning instructions
Cleaning the EM components
1. Power down and unplug the system before performing cleaning or disinfection procedures.
2. Unplug the EM instrument interface and the emitters from the main cart.
3. Unplug any instruments from the EM instrument interface ports.
4. Wipe and clean the external surfaces of the EM instrument interface, the side­mount emitter, and the flat emitter with a moistened low-level disinfectant wipe per the manufacturer’s instructions.
Warning: Do not allow liquid to enter the EM instrument interface ports.
5. Clean each EM cable by wrapping a low-level disinfectant wipe completely around the cable starting at the bend relief closest to the EM component. Pull the cable through the wipe in one direction finishing at the terminal connector.
Warning: Do not allow liquid to enter the interior surfaces of the connector.
6. If disinfection is desired, make sure that the surface to be disinfected remains in contact with the disinfectant for the specified contact time as directed by the manufacturer.
- The contact time refers to the minimum amount of time the surface needs to
remain visibly wet.
- Use additional wipes, if needed, to maintain continuous contact time with
the disinfectant.
7. Examine the surfaces for visible soil. If soil is present, repeat cleaning.
8. Wipe the external surfaces with a clean wipe moistened with water or 70% isopropyl alcohol to remove disinfectant residue.
9735727-G02 Revision 1 3
StealthStation™ S8 cleaning instructions
2: Figure 2: StealthStation™ S8 EM components
Cables
1
Side-mount emitter
2
Flat emitter
3
EM instrument interface handle
4
EM instrument interface housing
5
4 9735727-G02 Revision 1