Medtronic 9466 Technical Manual

9466 Tachy Patient Magnet
Description and intended use
The Model 9466 Patient Magnet is a bluecoated, ring-shaped magnet. It is used with certain Medtronic Implantable Cardioverter Defibrillators (ICDs) that provide a programmable Patient Alert™ monitoring feature.
If any Patient Alert condition (low battery voltage, lead impedance out of range, etc.) that is programmed On has occurred, these ICDs emit an audible Patient Alert™ tone. By placing the magnet over the ICD, the clinician or the patient can replay the Patient Alert™ tone.
The Patient Alert condition does not need to be present at the time of the replay. The tone indicates whether a programmed condition has occurred since the last device interrogation.
Warnings
The ICD suspends its tachyarrhythmia detection and therapy operations while the magnet is in place. Removing the magnet restores the ICD to its programmed operation. The magnet does not affect the ICD’s bradycardia pacing operations.
To ensure tachyarrhythmia therapy if necessary, the patient should not carry, store, or leave the magnet positioned over the ICD.
The patient should be careful to avoid sources of electromagnetic interference (EMI) while applying the magnet.
Directions for use
The Model 9466 Patient Magnet can be used whenever a Patient Alert™ tone has been announced. The patient should call the follow-up clinician and use the magnet to replay the tone over the telephone.
1. Call the follow-up clinician, using a conventional (not cellular) telephone.
2. Hold the face of the magnet directly over the ICD (Figure 1).
Figure 1. 9466 Magnet Application
3. After about a second, the ICD sounds one of three tones.
If more than one Patient Alert™ condition has occurred since the last interrogation, the highest urgency tone is sounded.
Continuous tone No programmed Patient Aler t™
Intermittent on/off tone A Low-Urgency condition has occurred
Dual high/low tone A High-Urgency condition has occurred
4. Move the magnet away from the ICD once the tone begins to sound.
5. The audible tone continues for at least two seconds (maximum of 30 seconds). The magnet can be applied again to retrigger the tone if necessary.
6. If a Patient Alert™ condition has occurred, the audible tone sounds automatically each day at the programmed time until the ICD is interrogated with a Medtronic programmer. At the time of interrogation, the clinician can retrieve a detailed report of the alert condition.
7. Refer to the applicable ICD technical manual for additional information.
condition has occurred
Storage and handling
The magnet could damage some electronic devices if kept too close.
Keep the magnet at least 15 cm (six inches) from electronic devices and recordings: VCRs, televisions, and videotapes; bank and credit cards, cordless and cellular telephones, computers, diskettes, calculators, etc.
Keep the magnet at least 5 cm (two inches) from watches and clocks.
If soiled, the magnet can be wiped clean with a soft cloth or a sponge, or washed with a non-abrasive cleanser. The magnet is not damaged by being submerged in water.
Technical specifications
Shape Ring
Size Approx. 75 mm (3") diameter x 16 mm
Materials Ferrous alloys coated with epoxy
Minimum field strength 90 gauss measured 40 mm (1.5") from
(5/8") thick
magnet surface
Technical support
Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users. For more information, contact your local Medtronic representative.
DISCLAIMER OF WARRANTY
ALTHOUGH THE MODEL 9466 PATIENT MAGNET HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY FOR ANY MEDICAL EXPENSES OR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, FAILURE OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
World Headquarters
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-514-4879