Medtronic 9010000820 Instructions for Use

Disassembly/Assembly M708348B212E Rev. B
Disassembly/Assembly M708348B212 Rev. B Instructions for
SET SCREW DRIVER,
P/N 9010000820
Instructions for SET SCREW DRIVER, P/N 9010000820
2014-01-21
1800 Pyramid Place Memphis, TN 38132 Telephone 800 933 2635 (In U.S.A.) 901 396 3133 (Outside U.S.A.) Fax 901 396 0356
ENGLISH
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR SET SCREW DRIVER, P/N 9010000820
DISASSEMBLY INSTRUCTIONS
Refer to the section entitled “Reprocessing – General Considerations” in the Medtronic Reusable Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is disassembled, continue with step 3 in the section entitled “Manual Cleaning Instructions-Cleaning Instructions-Dedicated Cleaning Area.”
Note: Reassemble the instrument after sterilization.
DISASSEMBLY
1. Pull back on the metallic T-handle while pressing both buttons down on the metallic T-handle.
3. Pull back on the metallic T-handle to release the inner shaft from the outer shaft.
SET SCREW DRIVER,
P/N 9010000820
1800 Pyramid Place Memphis, TN 38132 Telephone 800 933 2635 (In U.S.A.) 901 396 3133 (Outside U.S.A.) Fax 901 396 0356
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR SET SCREW DRIVER, P/N 9010000820
ENGLISH
2014-01-21
1. Pull back on the metallic T-handle while pressing both buttons down on the metallic T-handle.
DISASSEMBLY INSTRUCTIONS
Refer to the section entitled “Reprocessing – General Considerations in the Medtronic Reusable Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is disassembled, continue with step 3 in the section entitled Manual Cleaning Instructions-Cleaning Instructions-Dedicated Cleaning Area.
Note: Reassemble the instrument after sterilization.
DISASSEMBLY
3. Pull back on the metallic T-handle to release the inner shaft from the outer shaft.
2. Release the metallic T-handle and buttons.
ASSEMBLY
4. Pull the inner shaft out to remove.
1. Insert the inner shaft into the outer shaft.
3. Release the metallic T-handle and buttons.
2. While pulling back on the metallic T-handle and pressing both buttons down on the metallic T-handle, press the inner shaft downwards into the outer shaft.
4. Pull back on the metallic T-handle to ensure that the inner shaft is fully seated and does not protrude from the outer shaft when actuated.
©2014 MEDTRONIC SOFAMOR DANEK USA, Inc. All rights reserved.