Disassembly/Assembly M708348B212E Rev. B
Disassembly/Assembly M708348B212 Rev. B
Instructions for
SET SCREW DRIVER,
P/N 9010000820
Instructions for
SET SCREW DRIVER, P/N 9010000820
2014-01-21
Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place
Memphis, TN 38132
Telephone 800 933 2635 (In U.S.A.)
901 396 3133 (Outside U.S.A.)
Fax 901 396 0356
ENGLISH
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR SET SCREW DRIVER, P/N 9010000820
DISASSEMBLY INSTRUCTIONS
Refer to the section entitled “Reprocessing – General Considerations” in the Medtronic
Reusable Instruments Instructions for Use prior to reprocessing this instrument. After
the instrument is disassembled, continue with step 3 in the section entitled “Manual
Cleaning Instructions-Cleaning Instructions-Dedicated Cleaning Area.”
Note: Reassemble the instrument after sterilization.
DISASSEMBLY
1. Pull back on the metallic T-handle while
pressing both buttons down on the metallic
T-handle.
3. Pull back on the metallic T-handle to release the
inner shaft from the outer shaft.

SET SCREW DRIVER,
P/N 9010000820
Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place
Memphis, TN 38132
Telephone 800 933 2635 (In U.S.A.)
901 396 3133 (Outside U.S.A.)
Fax 901 396 0356
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR SET SCREW DRIVER, P/N 9010000820
ENGLISH
2014-01-21
1. Pull back on the metallic T-handle while
pressing both buttons down on the metallic
T-handle.
DISASSEMBLY INSTRUCTIONS
Refer to the section entitled “Reprocessing – General Considerations” in the Medtronic
Reusable Instruments Instructions for Use prior to reprocessing this instrument. After
the instrument is disassembled, continue with step 3 in the section entitled “Manual
Cleaning Instructions-Cleaning Instructions-Dedicated Cleaning Area.”
Note: Reassemble the instrument after sterilization.
DISASSEMBLY
3. Pull back on the metallic T-handle to release the
inner shaft from the outer shaft.
2. Release the metallic T-handle and buttons.
ASSEMBLY
4. Pull the inner shaft out to remove.
1. Insert the inner shaft into the outer shaft.
3. Release the metallic T-handle and buttons.
2. While pulling back on the metallic T-handle
and pressing both buttons down on the
metallic T-handle, press the inner shaft
downwards into the outer shaft.
4. Pull back on the metallic T-handle to ensure
that the inner shaft is fully seated and does
not protrude from the outer shaft when
actuated.
©2014 MEDTRONIC SOFAMOR DANEK USA, Inc. All rights reserved.