Medtronic 8637 User Manual

SynchroMed® II
Programmable pumps
Implant manual
  Rx only
8637
2003
Refer to the appropriate product for symbols that apply.
Open here
Do not reuse
Do not resterilize
2
STERILIZE
Do not use if package is damaged
Sterilized using ethylene oxide
-XX °C
-XX °F
EO
Consult instructions for use
Use by
Date of manufacture
Manufacturer
XX °C
Temperature limitation
XXX °F
Keep away from magnets
Serial number
STERILE
Conformité Européenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (NB
0123).
For USA audiences only
8637 2016-06-01  English 3
EC
Authorized representative in the European community
REP
4 English  8637 2016-06-01
Medtronic® and SynchroMed® are trademarks of Medtronic, Inc., registered in the U.S. and other countries.
®
Prialt
is a registered trademark of Azur Pharma International Limited.
8637 2016-06-01  English 5
6 English  8637 2016-06-01

Table of contents

Description 9
Package contents 10
Patient identification card 10
Device specifications 11
Device longevity 14
Flow rate accuracy 15
Measurement error 15 Fluid volume 15 Environmental conditions 16
Declaration of Conformity 18
Instructions for use 19
Preparing for pump implant 19 Sterile procedure 20 Emptying the pump 20 Preparing to fill the pump 21 Filling the pump 21 Priming the pump before implant (if applicable) 22 Replacing an implanted pump 22 Preparing the pump pocket 23
Refer to the indications, drug stability, and emergency procedures reference manual for indications and related information.
Refer to the appropriate information for prescribers booklet for contraindications, warnings, precautions, adverse events summary, individualization of treatment, patient selection, use in specific populations, and component disposal.
Refer to the appropriate drug labeling for indications, contraindications, warnings, precautions, dosage and administration information, and screening procedures.
! USA
Refer to the clinical summary booklet for information on the clinical study
results of the infusion system and adverse events summary.
8637 2016-06-01  English 7
Implanting the pump 23 Programming the pump 24 Updating the patient record 26 Monitoring the patient 26 Refilling the pump or accessing the catheter access port 26
Technical support 27
8 English  8637 2016-06-01

Description

The implantable Medtronic Model 8637 SynchroMed II programmable pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a Model 8637 SynchroMed II pump and a catheter.
The catheter connects to the pump catheter port. The pump is anchored in the pump pocket using the suture loops on the outside of the pump (Figure 1).
Reservoir fill port
Figure 1. Pump exterior view.
The drug is stored in the pump reservoir (Figure 2). Per a programmed prescription, the drug moves from the pump reservoir, through the internal pump tubing, catheter port, and catheter, to the infusion site. The catheter access port (CAP) allows injection of drug directly into the implanted catheter for drug administration and diagnostic purposes. Drug injected into the CAP bypasses the pump mechanism and goes directly through the catheter port into the implanted catheter to the infusion site. The CAP allows entry of a 24­gauge noncoring needle to prevent accidental injection during refill procedures (which use the 22-gauge noncoring needle supplied in the refill kit).
The manufacturer and model code recorded on a radiopaque identifier are visible using standard x-ray procedures.
Suture loop
Catheter access port
Catheter port
8637 2016-06-01  English 9
Reservoir fill port
Radiopaque
identifier
Catheter access port
Pump reservoir
Figure 2. Pump interior view.
Internal pump tubing

Package contents

Pump
Needle, 22-gauge (black sheath)
Needle, 24-gauge (purple sheath)
Product literature
Registration form

Patient identification card

! USA
Warranty card
Patient identification card
A patient identification card is packaged with this device. Advise the patient to carry the identification card at all times.
! USA
The patient identification card packaged with the device is temporary; a permanent
card is mailed to the patient when Medtronic receives the registration form.
! USA
The implant registration form registers the device warranties and creates a record of
the device in the Medtronic implant data system.
10 English  8637 2016-06-01

Device specifications

Table 1. Shipping and operating values for the Model 8637 SynchroMed II pump
Shipping Operating
Fluid in reservoir Sterile water — Shipping flow rate 0.006 mL/day — Infusion modes Simple continuous Single bolus
Alarms
Critical alarmNon-critical alarm
Disabled Disabled
Priming bolus Bridge bolus Simple continuous Flex Minimum rate Stopped pump
Enabled with an interval programmed
Enabled with an interval programmed
8637 2016-06-01  English 11
Table 2. Device specifications for the Model 8637 SynchroMed II pump
8637-20 8637-40
Pump
Thickness (including septum)Weight (empty/full)Displacement volumeDiameter (including CAP)
19.5 mm 165/185 g 91 mL
87.5 mm
26.0 mm 175/215 g 121 mL
87.5 mm
Pump reservoir
VolumeResidual volumeFill volume at shipping
20.0 mL
1.4 mL
17.5 mL
40.0 mL
1.4 mL
37.5 mL
Internal pump tubing  Volume
b
0.140 mL 0.140 mL
Reservoir fill port  Septum puncture life 500 punctures 500 punctures
Catheter access port  Septum puncture life 500 punctures 500 punctures
Flow rate
Maximum programmableMinimum programmable
Stopped pump maximum
c
c
24 mL/day
0.048 mL/day
0.030 mL/day
24 mL/day
0.048 mL/day
0.030 mL/day
 leakage Bacterial retentive filter
 Pore size 0.22 μm (micron) 0.22 μm (micron) Power source
BatteryLongevity
Lithium hybrid cathode Rate dependent
(Figure 3)
Lithium hybrid cathode Rate dependent
(Figure 3)
Transmitter
Carrier frequencyOutput level (at 300 m)
175 kHz
-48 dBuV/m
175 kHz
-48 dBuV/m
a
12 English  8637 2016-06-01
Table 2. Device specifications for the Model 8637 SynchroMed II pump
(continued)
a
8637-20 8637-40
Radiopaque identifier NGP NGV Reservoir pressure 20.68 kPa to 34.75 kPa 20.68 kPa to 34.75
a
All measurements are approximate.
b
When priming the pump before implant, use a priming bolus of 0.300 mL to fill the internal pump tubing with drug before connecting the catheter and implanting the pump (back table prime). The total prime volume is greater than the internal pump tubing volume to ensure that the entire fluid path in the pump is flushed and primed before implant.
c
Actual limits depend on pump calibration constant and selected infusion mode.
Table 3. Material of components in the Model 8637 SynchroMed II sterile package
Component Material Material contacts
human tissue
kPa
Material contacts
drug
Pump
Exterior Titanium Yes NoReservoir Titanium No YesReservoir valve Titanium No YesTubing Silicone rubber No YesReservoir fill port
septumCatheter access
port septum
Silicone rubber Yes Yes
Silicone rubber Yes Yes
Catheter port Titanium Yes YesBacterial
retentive filter
Polyvinylidene fluoride
No Yes
Suture loops Titanium Yes NoPropellant Inert gas No No
Needles Stainless steel Yes Yes
8637 2016-06-01  English 13

Device longevity

Device longevity is a function of flow rate. Flow rates affect the battery voltage and motor revolutions (Figure 3).
10
9
8
7
6
5
4
Typical Device Longevity (years)
3
2
1
0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5
Flow Rate (mL/day)
Figure 3. Typical device longevity based on flow rate.
Device longevity is the calculated number of service months remaining based on actual usage rates. when the pump nears the end of its service life (EOS). At ERI, the pump continues to operate within specifications. The ERI thresholds allow the pump to operate for a minimum of 90 days, at rates up to 1.5 mL/day, between ERI activation and EOS (Figure 4). When activated, ERI is date stamped and displayed by the programmer after interrogating the pump. The EOS activation indicates the pump has reached the end of its service life. At EOS, the pump stops, but telemetry is available until the pump battery is depleted.
1
An elective replacement indicator (ERI) message displays on the programmer
1  
Device longevity sources include bat tery life (voltage), device life (year s), and motor life (revolutions).
14 English  8637 2016-06-01
ERI
EOS
Device Longevity (months)
Time Since Implant 90 Days
Figure 4. ERI and EOS.

Flow rate accuracy

The flow rate accuracy of the pump is within ±14.5% of the programmed flow rate at
0.048-24 mL/day, 37 °C, 50% reservoir volume, and 300 meters above sea level. Measurement error, fluid volume, and changes in environmental conditions (eg, body temperature and atmospheric pressure) all affect the flow rate. The effects of these changes on flow rate are cumulative if the conditions exist simultaneously.

Measurement error

The apparent flow rate based on clinical measurements can vary due to measurement error (eg, syringe measurement accuracy, human error, and the volume of fluid in the extension tubing and filter).

Fluid volume

The flow rate of the pump varies slightly with the volume of fluid in the pump reservoir. The pump flow rate decreases as the reservoir volume approaches 1 mL. The pump flow rate decreases rapidly and then stops as the reservoir volume decreases from 1 mL to 0 mL. Therefore, the pump should be refilled prior to reaching 1 mL or less. Typically, the flow rate
8637 2016-06-01  English 15
decreases by about 4% as the volume is reduced from the half-full volume to a volume of 1 mL. The usable volume is the reservoir volume minus 1 mL (Figure 5).
130%
120%
110%
100%
90%
Flow Rate Accuracy
80%
70%
1 mL 20 or 40 mL
Usable Volume
Volume in Reservoir (mL)
Figure 5. Flow rate accuracy as a function of fluid volume in reservoir.

Environmental conditions

Body temperature
The flow rate of the pump varies with body temperature. The flow rate increases as the temperature increases above 37 °C and decreases as the temperature decreases below 37 °C (Figure 6).
16 English  8637 2016-06-01
120%
115%
110%
105%
100%
33 34 35 36 37 38 39 40 41
95%
Flow Rate Accuracy
90%
85%
80%
Temperature (Celsius)
Figure 6. Flow rate accuracy as a function of temperature (typical effect).
Atmospheric pressure
Patients living or traveling (eg, airline flights, mountain climbing) at altitudes above sea level are exposed to lower atmospheric pressures. Within days of exposure to the lower pressures, the flow rate of the pump can increase and then stabilize at the higher flow rate. In circumstances where a potential increase in flow rate may pose a risk to a patient, reprogramming the infusion prescription offsets this higher flow rate (Figure 7).
In rare instances, exposure to the lower atmospheric pressure can cause the pump to deliver more than 14.5% of the programmed flow rate while the patient is exposed to the lower pressure. Consider changes in drug concentrations or changes to pump programming for patients exposed to lower pressures.
8637 2016-06-01  English 17
120%
115%
110%
105%
100%
95%
Flow Rate Accuracy
90%
85%
80%
0
500 1000 1500 2000 2500 3000 3500 4000 4500
Altitude (meters)
Figure 7. Flow rate accuracy as a function of altitude (typical effect).

Declaration of Conformity

Medtronic declares that this product is in conformity with the essential requirements of Directive 90/385/EEC on Active Implantable Medical Devices.
For additional information, contact the appropriate Medtronic representative listed on the inside back cover of this manual.
18 English  8637 2016-06-01

Instructions for use

Implanting physicians should be experienced in pump and catheter implant procedures and should be thoroughly familiar with all product labeling.
Cautions:
Do not implant a pump that was dropped onto a hard surface or shows signs of
damage. Implanting a pump that has been dropped or damaged can result in lack of intended therapy, and require additional surgery to replace the pump.
Do not implant the pump unless pump operation has been confirmed. Failure to
confirm pump operation before implant can result in additional surgery to replace the pump.
Do not prematurely activate the pump reservoir valve. Activation of the pump
reservoir valve seals the pump reservoir valve closed. Unusual resistance or the inability to inject the entire fill volume may indicate activation of the pump reservoir valve. If the valve closes, a portion of the reservoir contents must be delivered or removed before filling can be completed. Procedural delays can occur. To prevent activation of the pump reservoir valve during emptying and filling procedures:
completely aspirate all contents of the pump reservoir before filling.
do not allow air into the pump reservoir through an open needle in the
septum or an unclamped extension. do not exceed the maximum reservoir volume indicated in the pump
labeling.
To use the nonsterile clinician programmer in a sterile field, place a sterile barrier
between the patient and the programming head to prevent infection. Do not sterilize any part of the clinician programmer. Sterilization may damage the programmer.

Preparing for pump implant

1. Assemble equipment and supplies.
Sterile items
The pump package containing the pump, 22-gauge noncoring needle (for filling the
pump), and 24-gauge noncoring needle (for flushing the catheter access port) Empty 20-mL syringes (for emptying the pump)
0.22-µm (micron) filter
Syringe containing prescribed fluid (volume not to exceed the reservoir volume of
the pump) 10-mL syringe with 1–2 mL of sterile, preservative-free saline (for flushing the
catheter access port)
8637 2016-06-01  English 19
Nonsterile items
Medtronic clinician programmer
2. Before opening the shelf package, use the clinician programmer to interrogate the
pump and verify pump battery status and current settings. a. Confirm that there are no active alarm events.
Note: If the pump is still in Shelf Mode, audible alarms are disabled. The pump
must be interrogated to determine if an alarm has been activated.
b. Confirm that the pump calibration constant displayed on the screen matches the
calibration constant printed on the shelf package.
Warning: The calibration constant displayed on the programmer screen
after reading the pump status must match the calibration constant printed on the shelf package. If calibration constants differ, contact the appropriate Medtronic representative listed on the inside back cover of this manual. Using an incorrect calibration constant can result in a clinically significant or fatal drug underdose or overdose.
3. Attach a "FOR YOUR RECORDS" label (enclosed in the shelf package) to the
patient's record. This label displays the pump model number, reservoir size, calibration constant, and serial number.
Note: Updating the pump with the new parameters can be performed at this time or
after the implant procedure. Refer to "Programming the pump" on page 24 for instructions.

Sterile procedure

1. Open the sterile pump package and remove the pump.
2. Remove the protective cap from the catheter port (a small amount of water might be
present in the protective cap).

Emptying the pump

1. Assemble the 22-gauge noncoring needle and the empty syringe.
2. Insert the needle into the reservoir fill port until the needle touches the metal needle
stop.
3. Withdraw the sterile water from the pump into the empty syringe (the pump is shipped
nearly full).
Note: If the volume of fluid in the pump reservoir exceeds the volume of the syringe
used for emptying, remove the filled syringe and needle. Attach an empty syringe and needle, and repeat until the pump reservoir is empty.
4. Empty the pump reservoir until air bubbles no longer appear in the syringe, ensuring
all water and air is removed from the pump reservoir.
5. Remove the syringe and needle from the reservoir fill port.
20 English  8637 2016-06-01

Preparing to fill the pump

1.
If using Prialt
labeling for instructions for use of this drug with the pump.
2. For all indicated drugs refer to Table 4 to determine the fill method.
Notes:
A change in concentration is not recommended at the time of replacement.
The pump reservoir capacity is 20 mL or 40 mL. Because some sterile water
remains in the pump reservoir, the final concentration of drug varies based on the fill method.
Table 4. Expected concentration of drug in pump reservoir based on fill method
Pump reservoir
capacity
8637-20 93% 98% 99% 8637-40 97% 99% 100%
3. If you are rinsing the pump reservoir before filling, rinse and discard the appropriate
volume based on the fill method shown in Table 4.
1
(preservative-free ziconotide sterile solution), refer to the drug
Filling without
rinsing
Rinsing with 3 mL of drug
Rinsing with
10 mL of drug

Filling the pump

1. Attach the filter to the syringe containing the prescribed fluid.
2. Attach the needle to the syringe containing the prescribed fluid and filter, and purge
the air from the fluid pathway.
3. Read the actual fill volume in the syringe.
4. Insert the needle into the reservoir fill port, and inject the prescribed fluid slowly into
the pump reservoir.
5. If the reservoir valve is activated before the pump is filled completely, discontinue
injection, remove the needle from the reservoir fill port, and return to "Emptying the pump" on page 20, step 4.
6. When filling is complete, remove the needle from the reservoir fill port.
7. Flush the catheter access port using a 24-gauge noncoring needle and a syringe filled
with 1 to 2 mL of saline (or a heparinized solution for vascular applications, if not contraindicated).
a. Gently insert the needle into the catheter access port until the needle touches the
metal needle stop. b. Inject fluid into the catheter access port until fluid is observed at the catheter port. c. Remove the needle from the catheter access port.
1  
Prialt is a brand name drug for zicon otide. Prialt is listed here for USA audiences only.
8637 2016-06-01  English 21

Priming the pump before implant (if applicable)

Warning: If the pump is being primed before implant, allow the prime to finish before
connecting the catheter to the pump. If the prime is not finished, drug can be bolused into the catheter and can result in a clinically significant or fatal overdose.
Note: Refer to the clinician programmer guide for information on how to program the priming
bolus.
For a full system implant or a pump replacement with aspirated catheter
To reduce the risk of overdose, consider using a priming bolus of 0.300 mL to fill the
1.
internal pump tubing with drug pump (back table prime).
2. If implanting a new pump, proceed to "Preparing the pump pocket" on page 23.
3. If replacing an implanted pump, proceed to "Replacing an implanted pump" on
page 22.
For a pump replacement with non-aspirated catheter
1.
Use a priming bolus of 0.300 mL to fill the internal pump tubing with drug before connecting the catheter and implanting the pump (back table prime).
Warning: If this is a pump replacement and the catheter has not been replaced
and has not been aspirated, the internal pump tubing should be primed connecting the catheter and implanting the pump. Do not program a priming bolus after the catheter has been connected to the implanted pump. Programming a priming bolus in this situation can result in a clinically significant or fatal overdose.
2. Proceed to "Replacing an implanted pump" on page 22.
before connecting the catheter and implanting the
before

Replacing an implanted pump

1. If applicable, remove the suture at the catheter connector.
2. Disconnect the implanted pump from the implanted catheter. To avoid damage to the
pump connector, leave the connector attached to the catheter.
Note: If you are replacing a SynchroMed EL Model 8626, 8626L, 8627, or 8627L
Implantable Pump, interrogate the replaced pump for catheter volume information that may be stored in the pump. If you are replacing a SynchroMed II Model 8637 Implantable Pump, interrogate the replaced pump for catheter volume information. Enter the catheter volume information into the clinician programmer.
3. If not replacing the catheter, slowly aspirate 1 to 2 mL of fluid from the catheter using a
1-mL tuberculin syringe. Leave the syringe in place to avoid CSF loss. Aspirating directly from the catheter clears the catheter of drug and confirms catheter patency.
Note: Conditions might exist under which the catheter is not patent or is not aspirated.
If the catheter is not patent it must be replaced. Refer to the appropriate catheter implant manual for catheter replacement instructions.
22 English  8637 2016-06-01
Warning: During vascular applications, do not aspirate blood through the
catheter access port or catheter. Blood sampling or aspiration through the catheter access port is contraindicated in vascular applications. Residual blood from aspiration or blood sampling can occlude the catheter or pump and inhibit drug delivery, resulting in a loss of or change in therapy, which may lead to a return of underlying symptoms, drug withdrawal symptoms, or a clinically significant or fatal drug underdose, and require surgical revision or replacement.
4. Proceed to "Implanting the pump" on page 23.

Preparing the pump pocket

Prepare the subcutaneous pocket using an incision in the lower abdomen. Ensure that the subcutaneous pump pocket allows the pump to be implanted within 2.5 cm
from the surface of the skin and in an area where sutures will not be directly over the reservoir fill port or catheter access port.
Caution: Select a location in the lower abdomen that is:
away from bony structures (eg, 3 to 4 cm) to minimize discomfort at the pump
site.
away from areas of restriction or pressure to minimize the potential for skin
erosion and patient discomfort.
away from existing scar tissue.
For programmable pumps, select a location that is also:
a minimum of 20 cm away from another programmable device to minimize
telemetry interference and incorrect or incomplete programming.
in an area accessible to the patient for proper operation of a patient control device
(if applicable). In the pediatric population, care must be taken to select an appropriate location by
taking into consideration:
available body mass.
presence of ostomies.
growth and development.

Implanting the pump

1. Connect the implanted catheter to the pump according to the catheter implant manual
instructions.
2. Place the filled pump into the prepared pocket.
8637 2016-06-01  English 23
Cautions:
Implant the pump no more than 2.5 cm from the surface of the skin in order
to maintain access to the reservoir and catheter access ports. Implantation of the pump is contraindicated if the pump cannot be implanted 2.5 cm or less from the surface of the skin.
Place the pump in the prepared pocket so:
the reservoir fill port is anteriorly oriented and the reservoir fill port and
catheter access port will be easy to access after implant. no sutures to the skin will be directly over the reservoir fill port or the
catheter access port. the catheter is not kinked or twisted and is secured well away from the
pump ports.
Improper component placement can result in inaccessible pump ports, inadequate drug delivery, component damage, or procedural delays, and require surgical revision or replacement.
3. Suture the pump in the subcutaneous pocket using the following steps:
a. Suture first to the fascia in the bottom of the subcutaneous pocket. b. Use these two sutures and the lower suture loops on the pump to draw the pump
into the pocket. c. Tie the sutures. d. Suture the remaining two loops at the top of the pump pocket. e. Tie the sutures, securing the pump into the pocket.
4. Irrigate the pump pocket.
5. Close the incisions per normal procedure and apply dressing.

Programming the pump

For a full system implant or a pump replacement with aspirated catheter
Enter the following into the clinician programmer: patient information, catheter model
1.
number, implanted catheter length (in centimeters), drug name and concentration, and the volume of prescribed fluid placed in the pump reservoir at implant.
Note: If you are replacing a SynchroMed EL Model 8626, 8626L, 8627, or 8627L
Implantable Pump, interrogate the replaced pump for catheter volume information that may be stored in the pump. If you are replacing a SynchroMed II Model 8637 Implantable Pump, interrogate the replaced pump for catheter volume information. Enter the catheter volume information into the clinician programmer.
Warning: Use the catheter length recorded at implant or catheter revision when
calculating catheter volume. The actual implanted catheter length and catheter model number are required to accurately calculate catheter volume.
value does not exist that can be used as a substitute for this knowledge.
An inaccurate calculation of the catheter volume can result in a clinically significant or fatal drug underdose or overdose.
A universal
24 English  8637 2016-06-01
2. Program the implanted pump to deliver a priming bolus. Refer to the clinician
programmer guide for information on how to calculate and program the priming bolus.
If the pump was primed before implant, enter the catheter volume as the total
a.
prime volume.
Warning: If the catheter is new or has been aspirated and the pump was
primed before implant, prime catheter and implanting the pump. volume in the total prime volume calculation. An inaccurate calculation of the total prime volume can result in a clinically significant or fatal overdose.
b.
If the pump was not primed before implant, enter the sum of the internal pump
tubing volume and the catheter volume as the total prime volume.
Note: Priming bolus default parameters have been carefully selected based on
extensive modeling and testing. To ensure optimal initiation of therapy,
only the catheter after connecting the
Do not include the internal pump tubing
modifications to these values are not recommended.
3. Set the Low Reservoir Alarm (to at least 1 mL).
4. Program the pump with new parameters. Note: Refer to the clinician programmer guide for instructions on programming the
pump.
5. Print out the patient's prescription and pump settings (pump status).
For a pump replacement with non-aspirated catheter
1.
Enter the following into the clinician programmer: patient information, catheter model number, implanted catheter length (in centimeters), drug name and concentration, and the volume of prescribed fluid placed in the pump reservoir at implant.
Note: If you are replacing a SynchroMed EL Model 8626, 8626L, 8627, or 8627L
Implantable Pump, interrogate the replaced pump for catheter volume information that may be stored in the pump. If you are replacing a SynchroMed II Model 8637 Implantable Pump, interrogate the replaced pump for catheter volume information. Enter the catheter volume information into the clinician programmer.
Warning: Use the catheter length recorded at implant or catheter revision when
calculating catheter volume. The actual implanted catheter length and catheter model number are required to accurately calculate catheter volume.
universal value does not exist that can be used as a substitute for this knowledge.
clinically significant or fatal drug underdose or overdose.
An inaccurate calculation of the catheter volume can result in a
2. Program the implanted pump to deliver the prescribed infusion.
A
8637 2016-06-01  English 25
Warning: If this is a pump replacement and the catheter has not been replaced
and has not been aspirated, the internal pump tubing should be primed connecting the catheter and implanting the pump. Do not program a priming bolus after the catheter has been connected to the implanted pump. Programming a priming bolus in this situation can result in a clinically significant or fatal overdose.
3. Set the Low Reservoir Alarm (to at least 1 mL).
4. Program the pump with new parameters. Note: Refer to the clinician programmer guide for instructions on programming the
pump.
5. Print out the patient's prescription and pump settings (pump status).
before

Updating the patient record

1. Place the prescription and pump settings (pump status) in the patient's records.
2. Determine the refill date from the printout.
3. Schedule a refill appointment.

Monitoring the patient

Medtronic recommends monitoring patients after any priming bolus procedure involving intrathecal therapy.
Opioids: Patients should be monitored with pulse oximetry for a minimum of 24 hours
in a facility equipped with emergency airway management, oxygen, naloxone for treatment of opioid overdose and other emergency services.
Baclofen: Patients should be monitored in a facility that provides experienced nursing
observation, with the ability and personnel for emergency airway management and ventilator support readily available. Patients should be monitored for a minimum of 8 hours or until they demonstrate stable neurological, respiratory and cardiac function.
Ziconotide: There are no labeling guidelines for patient monitoring after starting or
restarting ziconotide therapy. Published guidance recommends an overnight admission.
Based on the therapeutic index of the drug and the sensitivity of the patient, some individuals may need additional monitoring until the delivered drug reaches the intended concentration.

Refilling the pump or accessing the catheter access port

When refilling a Medtronic pump, use the appropriate Medtronic refill kit and associated refill kit instructions for use.
When accessing the catheter access port of a Medtronic pump, use the appropriate Medtronic CAP kit and associated CAP kit manuals and instructions for use.
26 English  8637 2016-06-01

Technical support

To obtain a copy of the refill kit or CAP kit instructions for use, or to receive additional technical support:
US only: Contact Medtronic Technical Services at 1-800-707-0933. Technical support
service is available 24 hours a day for clinicians managing patients with Medtronic implantable infusion pumps.
Outside of the US: Contact your local representative by using the phone numbers
listed on the last pages of this manual.
8637 2016-06-01  English 27
Contacts:
Asia:
Medtronic International Ltd. Tel. 02919-1300 Fax 02891-6830
Medtronic Asia Ltd. Tel. (02)-548-1148
Fax (02)-518-4786
Australia:
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde, NSW 2113 Australia Tel. +61-2-9857-9000 Fax +61-2-9878-5100 Toll-free 1-800-668-670
Austria:
Medtronic Österreich GmbH Tel. 01-240440 Fax 01-24044-100
Belgium:
Medtronic Belgium S.A. Tel. 02-456-0900 Fax 02-460-2667
Canada:
Medtronic of Canada Ltd. Tel. (1-905)-460-3800 Fax (1905)-826-6620
Czech Republic:
Medtronic Czechia s.r.o. Tel. 2-965-795-80 Fax 2-965-795-89
Denmark:
Medtronic Danmark A/S Tel. 45-32-48-18-00 Fax 45-32-48-18-01
Finland:
Medtronic Finland Oy/LTD Tel. (09)-755-2500 Fax (09)-755-25018
France:
Medtronic France S.A.S. Tel. 01-5538-1700 Fax 01-5538-1800
Germany:
Medtronic GmbH Tel. (02159)-81490 Fax (02159)-8149100
Greece:
Medtronic Hellas S.A. Tel. 210-67-79-099 Fax 210-67-79-399
Hungary:
Medtronic Hungária Kft. Tel. 1-889-06-00 Fax 1-889-06-99
Ireland:
Medtronic Ireland Ltd. Tel. (01)-890-6522 Fax (01)-890-7220
Italy:
Medtronic Italia SpA Tel. 02-241371 Fax 02-241381
Tel. 06-328141 Fax 06-3215812
Japan:
Medtronic Japan Tel. 03-6776-0017 Fax 03-6774-4645
Latin America:
Medtronic, Inc. Tel. (1305)-500-9328 Fax (1786)-709-4244
Norway:
Medtronic Norge AS Tel. 67-10-32-00 Fax 67-10-32-10
Poland:
Medtronic Poland Sp. z.o.o. Tel. (022)-465-69-00 Fax (022)-465-69-17
Portugal:
Medtronic Portugal, Lda. Tel. 21-724-5100 Fax 21-724-5199
Russia:
Medtronic Russia Tel. (8495) 580-7377 Fax (8495) 580-7378
Slovakia:
Medtronic Slovakia, o.z. Tel. 0268 206 911 Fax 0268 206 999
Spain:
Medtronic Ibérica, S.A. Tel. 91-625-0400 Fax 91-650-7410
Sweden:
Medtronic AB Tel. 08-568-585-00 Fax 08-568-585-01
Switzerland:
Medtronic (Schweiz) AG Tel. 031-868-0100 Fax 031-868-0199
The Netherlands:
Medtronic B.V. Tel. (045)-566-8000 Fax (045)-566-8668
Turkey:
Medtronic Turkey Tel. +90 216 636 1000 Fax +90 216 636 1008
U.K.:
Medtronic U.K. Ltd. Tel. 01923-212213 Fax 01923-241004
USA:
Medtronic, Inc. Tel. (1-763)-505-5000 Fax (1-763)-505-1000 Toll-free: (1-800)-328-0810
Manufacturer
Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA www.medtronic.com Tel. +1-763-505-5000 Fax +1-763-505-1000
REP
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10, 6422 PJ Heerlen, The Netherlands Tel. +31-45-566-8000 Fax +31-45-566-8668
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31, Case Postale 84 CH - 1131 Tolochenaz, Switzerland www.medtronic.eu Tel. +41-21-802-7000 Fax +41-21-802-7900
Asia-Pacific
Medtronic International Ltd. Suite 1106-11, 11/F, Tower 1, The Gateway, 25 Canton Road, Tsimshatsui, Kowloon, Hong Kong Tel. +852-2919-1300 Fax +852-2891-6830
Contacts for specific countries are listed inside this cover.
EC
*M221311A060*
© Medtronic, Inc. 2016 All Rights Reserved
M221311A060 Rev A
Loading...