Medtronic 8225275 Instructions for Use

MONOPOLAR BALL-TIP STIMULATING PROBE
0123
0123
MONOPOLAR BALL-TIP STIMULATING PROBE
Ball Tip Probe and Handle, 1.0mm 1/EA REF 8225275 Ball Tip Probe and Handle, 2.3mm 1/EA REF 8225276 Ball Tip Probe Only, 1.0mm 10/BX REF 8225277 Ball Tip Probe Only, 2.3mm 10/BX REF 8225278
PRODUCT INFORMATION AND INSTRUCTIONS
DEVICE DESC RIPTI ON
e Monopolar Ball-Tip Stimulating Probe is connected to a 2 meter lead ending with a connector. e uoroplastic insulation of the medical stainless steel wire terminates ush with the tip of the wire. is construction facilitates direct nerve stimulation while minimizing cerebrospinal uid shunting of the electrical stimulus. e nal 20 mm of the probe can easily be bent to provide optimal contact between the probe tip and nerve within the connes of the surgical eld.
is device is intended for use as an intraoperative motor nerve stimulator with the Medtronic Xomed Nerve Integrity Monitor (NIM®) or other EMG monitors.
INDIC ATIONS
is device is indicated for use for intraoperative motor nerve location.
CONTRA INDIC ATIONS
None known.
CAUTIO NS
• DO NOT RESTERILIZE. Resterilization can damage the tip insulation of the probe.
• To ensure delivery of current to intended tissues, use only with EMG monitor(s) equipped with active audio and/or visual current delivered feedback systems.
DIREC TION S FOR USE
1. Insert the large end of the probe into the handle assembly. Insert rmly to prevent the probe from dislodging and falling from the handle during the procedure.
2. Plug the connector end of the handle assembly into the cathode (-) connection on the NIM® Patient Interface, or other EMG monitor patient interface.
3. Place a subdermal EMG needle electrode (not included) in the desired location to serve as the ground, or anode (+) for the stimulator.
4. Verify prior to surgical use that the probe delivers current at the intended setting as evidenced by the EMG monitor’s audio and/or visual current delivered feedback system.
Ball Tip Probe and Handle, 1.0mm 1/EA REF 8225275 Ball Tip Probe and Handle, 2.3mm 1/EA REF 8225276 Ball Tip Probe Only, 1.0mm 10/BX REF 8225277 Ball Tip Probe Only, 2.3mm 10/BX REF 8225278
PRODUCT INFORMATION AND INSTRUCTIONS
DEVICE DESC RIPTI ON
e Monopolar Ball-Tip Stimulating Probe is connected to a 2 meter lead ending with a connector. e uoroplastic insulation of the medical stainless steel wire terminates ush with the tip of the wire. is construction facilitates direct nerve stimulation while minimizing cerebrospinal uid shunting of the electrical stimulus. e nal 20 mm of the probe can easily be bent to provide optimal contact between the probe tip and nerve within the connes of the surgical eld.
is device is intended for use as an intraoperative motor nerve stimulator with the Medtronic Xomed Nerve Integrity Monitor (NIM®) or other EMG monitors.
INDIC ATIONS
is device is indicated for use for intraoperative motor nerve location.
CONTRA INDIC ATIONS
None known.
CAUTIO NS
• DO NOT RESTERILIZE. Resterilization can damage the tip insulation of the probe.
• To ensure delivery of current to intended tissues, use only with EMG monitor(s) equipped with active audio and/or visual current delivered feedback systems.
DIREC TION S FOR USE
1. Insert the large end of the probe into the handle assembly. Insert rmly to prevent the probe from dislodging and falling from the handle during the procedure.
2. Plug the connector end of the handle assembly into the cathode (-) connection on the NIM® Patient Interface, or other EMG monitor patient interface.
3. Place a subdermal EMG needle electrode (not included) in the desired location to serve as the ground, or anode (+) for the stimulator.
4. Verify prior to surgical use that the probe delivers current at the intended setting as evidenced by the EMG monitor’s audio and/or visual current delivered feedback system.
STERI LITY
is product is provided STERILE and is intended for SINGLE PATIENT USE ONLY. DO NOT RESTERILIZE THIS PRODUCT. MEDTRONIC XOMED ASSUMES NO LIABILITY FOR PRODUCTS WHICH HAVE BEEN RESTERILIZED BY HEALTH CARE FACILITIES.
Medtronic Xomed, Inc. 6743 Southpoint Drive North Jacksonville, FL 32216 800-874-5797 or 904-296-9600 www.medtronicENT.com
® is a registered mark of Medtronic Xomed, Inc.
68E3267 D 02/11
Rx Only
STERI LITY
is product is provided STERILE and is intended for SINGLE PATIENT USE ONLY. DO NOT RESTERILIZE THIS PRODUCT. MEDTRONIC XOMED ASSUMES NO LIABILITY FOR PRODUCTS WHICH HAVE BEEN RESTERILIZED BY HEALTH CARE FACILITIES.
Medtronic Xomed, Inc. 6743 Southpoint Drive North Jacksonville, FL 32216 800-874-5797 or 904-296-9600 www.medtronicENT.com
® is a registered mark of Medtronic Xomed, Inc.
68E3267 D 02/11
Rx Only