Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
InSync Model 8040
0
Device Reference Guide
Reference information about the InSync Model 8040 atrial synchronous
biventricular pacing device
0
The following list includes trademarks or registered trademarks of Medtronic in the
United States and possibly in other countries. All other trademarks are the property of
their respective owners.
InSync, Marker Channel, Medtronic
Required Physician Training
In order to implant a Medtronic biventricular pacing system, the
physician is required to:
1. Thoroughly read this manual, and all associated device
and/or lead technical manuals.
2. Provide a copy of the patient manual to the patient and
discuss it with him or her and any other interested parties.
3. Be trained on the following topics:
■
Indications for use
■
Device operation to ensure therapy delivery
■
Measuring and managing biventricular thresholds
■
Assembly and use of LV lead implant tools
■
Placement of the LV lead
■
Patient management and system follow-up
Prior to implanting the system, Medtronic will certify that
physicians received training.
5
Required Physician Training
How to Use This Guide
This guide provides comprehensive information about the InSync
Model 8040 and its programmable and follow-up functions.
Note: For specific programming procedures, refer to the InSync
Model 8040 Device Programming Guide.
Organization of this guide divides information into two parts:
Part I - Understanding Device Operation – This part describes
operation of the device system, which consists of the InSync
device and the connected pacing leads. Included is information on
pacing modes, rate response and special therapy options,
diagnostic reporting features, and troubleshooting. Part II expands
on the information provided in the device technical manual, which
covers information needed primarily at the time of device
implantation.
InSync Model 8040 Device Reference Guide
6
How to Use This Guide
Part II - Reference Information – This part includes
quick-reference information organized by topic or feature. Various
specifications, parameter values, and feature options are
presented in tabular format.
InSync Model 8040 Device Reference Guide
Contents
Required Physician Training5
How to Use This Guide5
1Pacing modes13
Introduction14
Mode Pertinency Tables16
Indications and Usage18
Contraindications18
DDDR Mode19
DDD Mode20
DDIR Mode21
DDI Mode22
DVIR Mode23
DVI Mode24
VDD Mode25
VVIR / VDIR Modes26
VVI / VDI Modes27
Other Available Modes28
AAI / ADI Modes29
AAT / VVT Modes30
DOOR / AOOR / VOOR Modes31
DOO / AOO / VOO Modes32
ODO / OAO / OVO Modes33
2Rate response35
Introduction to Rate Responsive Pacing36
Activity Threshold38
Activity Rate Response40
Acceleration and Deceleration Times42
Rate Response Optimization44
3Device timing49
Rates50
AV Intervals58
InSync Model 8040 Device Reference Guide
8
Contents
Rate Adaptive AV61
Blanking Periods64
Refractory Periods 66
High Rate Atrial Tracking74
4Lead/cardiac tissue interface77
Selecting Pacing Parameters78
Selecting Sensing Parameters83
Monitoring Lead Stability86
Transtelephonic Capture Verification with TMT89
5Special therapy options93
Mode Switch and Diagnostic94
Non-Competitive Atrial Pacing 99
PMT Intervention102
PVC Response104
Ventricular Safety Pacing107
Rate Drop Response108
Sleep Function113
Single Chamber Hysteresis115
6Telemetry data 117
Parameter Summary 118
Battery and Lead Information119
Marker Channel Telemetry 121
Intracardiac Electrogram 122
Extended Telemetry125
7Miscellaneous operations127
Magnet Mode Operation128
Temporary Programming130
Electrical Reset131
Elective Replacement Indicator133
Emergency Pacing 134
8Diagnostics 135
Introduction to Diagnostics136
Event Summary138
InSync Model 8040 Device Reference Guide
Rate Histogram140
AV Conduction Histogram142
High Rate Episode144
Rate Versus Time 148
9Troubleshooting the device system 151
Troubleshooting Strategy 152
Troubleshooting Electrical Problems153
Troubleshooting Hemodynamic Problems155
Handling, Storage, and Resterilization157
Device Longevity158
Replacing the Device159
Patient Information and Service160
A Device description165
Basic Description166
Lead Compatibility166
Radiopaque Code167
Physical Dimensions167
Connector Dimensions 168
Contents
9
B Preset parameter settings169
Shipping Parameters 170
Nominal Settings172
Electrical Reset Settings174
Emergency Settings176
C Device programming recommendations 177
Device Programming Recommendations178
D Parameter values and restrictions 181
Programmable Modes and Parameters182
Programming Requirements and Restrictions186
Nonprogrammable Parameters189
Temporary Modes and Parameters 189
Timing Reference190
ETelemetry and diagnostic values 195
Magnet Mode Operation 196
InSync Model 8040 Device Reference Guide
10
Contents
Telemetry Functions197
Diagnostic Options198
Event Summaries200
FLongevity projections201
Longevity Projections (Normal Operating Life)202
Prolonged Service Period203
Elective Replacement Indicator (ERI)204
Battery Specifications204
G Warnings, precautions, and EMI 205
Special Notice206
Warnings207
Precautions212
Potential Complications217
Environmental and Medical Therapy Hazards218
Home and Job Environment Interference 221
H Clinical Studies225
Clinical Studies226
IGlossary227
Index 233
InSync Model 8040 Device Reference Guide
Understanding device operation
Part I
Introduction14
Mode Pertinency Tables 16
Indications and Usage18
Contraindications18
DDDR Mode19
DDD Mode20
DDIR Mode21
DDI Mode22
DVIR Mode23
DVI Mode 24
VDD Mode 25
VVIR / VDIR Modes26
Pacing modes
1
1
VVI / VDI Modes27
Other Available Modes28
AAI / ADI Modes29
AAT / VVT Modes 30
DOOR / AOOR / VOOR Modes 31
DOO / AOO / VOO Modes 32
ODO / OAO / OVO Modes 33
14
Chapter 1
Introduction
Introduction
Pacing Mode Selection
Delivery of Cardiac resynchronization therapy requires that
patients receive biventricular pacing for each cardiac cycle.
Special consideration is required in programming device
parameters for the InSync system to provide continuous
biventricular pacing.
This chapter provides an introduction to pacing modes as an aid
to mode selection. The chapter is organized as follows:
■
Mode Pertinency Tables – These tables show the features
and parameters that apply to each commonly used pacing
mode.
■
Mode Descriptions – Brief descriptions of how the available
modes operate.
Note: The biventricular pacing capability of the InSync device
does not affect the event timing operation of the pacing modes
described in this chapter. The two ventricular leads are connected
to the device in a parallel configuration that provides simultaneous
pacing at both ventricular stimulation sites.
Warning: The atrial only pacing modes available with the InSync
device do not provide cardiac resynchronization for heart failure
patients.
InSync Model 8040 Device Reference Guide
NBG Pacing Codes
CHAMBER PACED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
CHAMBER SENSED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
MODE OF RESPONSE
T = Triggered
I = Inhibited
D = Double (Both)
O = None
PROGRAMMABLE/RATE
RESPONSE
P = Programmable
M = Multiprogrammable
C = Communicating
R = Rate Responsive
O = None
DDDR
Pacing modes
Introduction
Pacing modes are defined in NBG Code.1 Each five-letter NBG
code describes a specific type of operation for implantable
devices. For simplicity, this manual uses only the first three or four
letters, such as DDD, DDIR, DVIR, and so forth. Figure 1-1
describes the first four letters of the NBG code.
These pacing modes are used for conventional pacing. Currently
there are no established pacing codes for biventricular pacing.
15
Further Information
Figure 1-1. NBG Pacing Codes
For biventricular pacing, a subset of these modes applies, e.g.,
VDD, DDD.
The mode descriptions in this chapter provide only a basic
overview of each mode. For further details on rate response,
timing, and therapy capabilities refer to “Rate response” on
page 35, “Device timing” on page 49, and “Special therapy
options” on page 93.
1
Bernstein A., et al., “The NASPE/BPEG Pacemaker Code,” PACE, 10(4),
Jul-Aug 1987. (“NBG” stands for “The North American Society of Pacing and
Electrophysiology [NASPE] and the British Pacing and Electrophysiology Group
[BPEG] Generic.” NBG’s five-letter code supersedes the ICHD Code.
InSync Model 8040 Device Reference Guide
16
Chapter 1
Mode Pertinency Tables
a
✓✓✓ ✓✓
✓✓✓✓ ✓
Pacing Operation Parameters
✓ ✓ ✓ ✓✓✓ ✓ ✓ ✓✓✓✓ ✓ ✓✓✓ ✓
Lower Rate
✓✓✓
Upper Tracking
Rate
Table 1-1 and Table 1-2 show which pacing parameters and features apply to each pacing mode
as indicated by black check marks. Asynchronous modes are not shown in these tables.
Mode Pertinency Tables
InSync Model 8040 Device Reference Guide
Table 1-1. Pacing Parameters Available For Each Mode
Pacing Parameter DDDR DDD DDIR DDI DVIR DVI VDD VVIR VDIR VVI VDI VVT AAIR ADIR AAI ADI AAT
✓✓✓✓✓✓✓
✓ ✓ ✓ ✓✓✓
✓✓✓
✓✓✓✓✓
Upper Activity Rate
Paced AV Interval
Sensed AV Interval
Rate Adaptive AV
✓ ✓✓✓✓
b
PVARP
✓ ✓✓✓✓✓✓
PVAB
Atrial Refractory
✓ ✓ ✓ ✓✓✓ ✓ ✓ ✓✓✓✓
✓ ✓ ✓ ✓✓✓✓✓
Period
Atrial Blanking
Ventricular
Refractory Period
Ventricular Blanking
(after AP)
See “Device timing” on page 49 for operational descriptions of these timing parameters.bSensor-varied PVARP available in the DDDR, DDD, DDIR, and VDD modes.
a
Pacing modes
17
Mode Pertinency Tables
✓✓ ✓✓ ✓ ✓
b
a
Special Therapy Options
Rate Response Therapy Options
✓✓
✓✓✓
✓
FeatureDDDR DDD DDIR DDI DVIR DVI VDD VVIR VDIR VVI VDI VVT AAIR ADIR AAI ADI AAT
Table 1- 2. Features Available For Each Mode
Managing Atrial
Rhythm
Mode Switch
Non-Competitive Atrial
Pacing
Managing Ventricular Rhythm
✓✓✓
✓✓✓✓✓
PMT Intervention
PVC Response
✓ ✓ ✓✓✓✓
Ventricular Safety Pacing
Special Pacing Operations
Rate Drop Response
✓ ✓ ✓✓✓✓✓ ✓ ✓✓✓✓ ✓ ✓✓ ✓✓
✓✓✓✓✓✓✓
✓✓✓✓✓✓✓
✓✓✓✓✓✓✓
✓✓✓✓✓✓✓
✓✓✓✓✓✓✓
Single Chamber
Hysteresis
Sleep Function
Activity Rate Response
Activity Threshold
Activity Acceleration
Activity Deceleration
Rate Response
Optimization
InSync Model 8040 Device Reference Guide
See “Special therapy options” on page 93 for operational descriptions of special therapy options.bSee “Rate response” on page 35 for operational descriptions of rate response features.
a
18
Chapter 1
Indications and Usage
Indications and Usage
Contraindications
The InSync Model 8040 is indicated for the reduction of the
symptoms of moderate to severe heart failure (NYHA Functional
Class III or IV) in those patients who remain symptomatic despite
stable, optimal medical therapy (as defined in the clinical trials
section), and have a left ventricular ejection fraction ≤ 35% and a
QRS duration ≥ 130 ms.
Asynchronous pacing is contraindicated in the presence (or
likelihood) of competitive paced and intrinsic rhythms.
Unipolar pacing is contraindicated in patients with an implanted
defibrillator or cardioverter-defibrillator (ICD) because it may
cause unwanted delivery or inhibition of defibrillator or ICD
therapy.
In the DDDR mode, the device tracks the faster of the intrinsic
atrial rate or the sensor-indicated rate. If the intrinsic rate is faster,
the DDDR mode provides atrial synchronous pacing; otherwise,
AV sequential pacing occurs at the sensor-indicated rate.
■
Rate limits for atrial tracking (Upper Tracking Rate)1 and
sensor tracking (Upper Activity Rate) are separately
programmable.
■
The AV intervals that follow sensed atrial events (SAV) and
paced atrial events (PAV) are separately programmable, and
they can be programmed to shorten with increasing rates
(Rate Adaptive AV).
■
A nonrefractory sensed event in either chamber inhibits
pacing in that chamber. A ventricular nonrefractory sensed
event in the VA interval that is not preceded by an atrial sense
(AS or AR) is a device-defined PVC, and starts a new VA
interval.
This mode may be appropriate for heart failure patients as it
provides both AV synchrony and cardiac resynchronization
therapy.
19
Rate responsiveness has not been evaluated in this patient
population.
Figure 1-2. Example of DDDR Mode Operation
1
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
The DDD mode provides atrial synchronous pacing in the
presence of intrinsic atrial activity; otherwise, AV sequential pacing
occurs at the Lower Rate.
■
Each atrial paced or nonrefractory atrial sensed event starts
an AV interval and a Lower Rate interval. The AV intervals that
follow sensed atrial events (SAV) and paced atrial events
(PAV) are separately programmable, and the SAV may be
optionally programmed to shorten with increasing rate (Rate
Adaptive AV).
■
Ventricular paced events may track atrial sensed events up to
the programmed Upper Tracking Rate.
■
A ventricular nonrefractory sensed event in the VA interval that
1
is not preceded by an atrial sense (AS or AR) is a
device-defined PVC, and starts a new VA interval.
This mode is appropriate for heart failure patients as it provides
both AV synchrony and cardiac resynchronization therapy.
Figure 1-3. Example of DDD Mode Operation
1
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
The DDIR mode provides dual chamber, sensor-driven,
atrioventricular (AV) sequential pacing for heart rate variation
without atrial tracking.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval. That
is, ventricular paced events after such sensed atrial events
occur at the sensor-indicated rate.
■
A ventricular nonrefractory sensed event in the VA interval
starts a new VA interval.
DDIR mode should not be permanently programmed in heart
failure patients with normal sinus rhythm. The device will switch to
DDIR/DDI modes when a mode switch occurs. Mode switch may
be appropriate for patients with a history of atrial arrhythmias.
The DDI mode provides dual chamber atrioventricular (AV)
sequential pacing with atrial sensing but without atrial tracking.
■
Atrial pacing occurs at the Lower Rate, with ventricular pacing
at the end of the PAV interval unless inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval.
Ventricular paced events after such sensed atrial events occur
at the Lower Rate.
■
A ventricular nonrefractory sensed event in the ventriculoatrial
(VA) interval starts a new VA interval.
DDI mode should not be permanently programmed in heart failure
patients with normal sinus rhythm. The device will switch to
DDIR/DDI modes when a mode switch occurs. Mode switch may
be appropriate for patients with a history of atrial arrhythmias.
The DVIR mode provides AV sequential pacing at the
sensor-indicated rate unless inhibited by ventricular sensed
events.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
The DVIR mode ignores intrinsic atrial events. Sensing occurs
only in the ventricle. A ventricular nonrefractory sensed event
during the ventriculoatrial (VA) interval starts a new VA
interval.
DVIR mode is not appropriate for heart failure patients with normal
sinus rhythm.
The DVI mode provides dual chamber AV sequential pacing
without atrial sensing/tracking.
■
Atrial pacing occurs at the Lower Rate, with ventricular pacing
at the end of the PAV interval unless inhibited.
■
Sensing occurs only in the ventricle, and intrinsic atrial events
are ignored. A ventricular nonrefractory sensed event during
the VA interval starts a new ventriculoatrial (VA) interval.
DVI mode is not appropriate for heart failure patients with normal
sinus rhythm.
The VDD mode provides atrial synchronous pacing (or VVI pacing
at the Lower Rate). The ventricle is paced synchronously up to the
programmed Upper Tracking Rate.
1
Sensing occurs in both the
atrium and ventricle, but pacing occurs only in the ventricle.
■
To promote atrial synchronous pacing at slow rates, a sensed
atrial event occurring near the end of the Lower Rate interval
will be followed by the programmed maximum SAV interval.
The result is an extension of the ventricular lower rate.
■
A ventricular nonrefractory sensed event in the V-V interval
that is not preceded by an atrial sense (AS or AR) is a
device-defined PVC, and it starts a new V-V interval.
This mode is appropriate for heart failure patients as it provides
both AV synchrony and cardiac resynchronization therapy.
25
Figure 1-8. Example of VDD Operation
1
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
Sensor-indicated Rate = 90 ppm (667 ms)Ventricular Refractory Period = 300 ms
Sensor-indicated
Interval
Sensor-indicated
Interval
Sensor-indicated
Interval
Upper Activity
Rate Interval
Chapter 1
VVIR / VDIR Modes
VVIR / VDIR Modes
The VVIR mode provides ventricular rate responsive pacing in
patients for whom atrial-based pacing is deemed unnecessary or
inappropriate. In the absence of sensed events, the ventricle is
paced at the sensor-indicated rate.
The VDIR mode operates the same as the VVIR mode except that
events sensed in the atrium are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the underlying
atrial rhythm without affecting ventricular pacing.
Note: In the VVIR and VDIR modes, ventricular refractory sensed
events restart the Upper Activity Rate interval.
VVIR/VDIR modes are generally not appropriate for heart failure
patients with normal sinus rhythm. In these modes, patients may
not receive cardiac resynchronization therapy.
Figure 1-9. Example of VVIR Mode Operation
InSync Model 8040 Device Reference Guide
VVI / VDI Modes
Pacing Rate Interval
Parameters:
Pacing Rate = 60 ppm (1000 ms)
Ventricular Refractory Period = 300 ms
Pacing Rate Interval
The VVI mode provides single chamber inhibited pacing at the
programmed Lower Rate unless inhibited by sensed events.
Sensing occurs only in the ventricle.
The VDI mode operates the same as the VVI mode except that
events sensed in the atrium are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the underlying
atrial rhythm without affecting ventricular pacing.
VVI/VDI modes are generally not appropriate for heart failure
patients with normal sinus rhythm. In these modes, patients may
not receive cardiac resynchronization therapy.
Warning: Atrial only pacing modes do not provide cardiac
resynchronization.
The AAIR mode provides atrial-based rate responsive pacing in
patients with intact AV conduction. Sensing and pacing occur only
in the atrium. In the absence of sensed events, the chamber is
paced at the sensor-indicated rate.
The ADIR mode operates the same as the AAIR mode except that
events sensed in the ventricle are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the conducted
ventricular rhythm without affecting atrial pacing.
Note: In the AAIR and ADIR modes, atrial refractory sensed
events do not restart the Upper Activity Rate interval.
AAIR/ADIR modes are generally not appropriate for heart failure
patients with normal sinus rhythm. In these modes, patients may
not receive cardiac resynchronization therapy.
Figure 1-11. Example of AAIR Mode Operation
InSync Model 8040 Device Reference Guide
AAI / ADI Modes
Pacing Rate Interval
Parameters:
Pacing Rate = 75ppm (800 ms)
Pacing Rate Interval
The AAI mode provides single chamber inhibited atrial pacing.
Sensing and pacing occur only in the atrium. Pacing occurs at the
programmed Lower Rate unless inhibited by sensed events.
The ADI mode operates the same as the AAI mode except that
events sensed in the ventricle are recorded by the diagnostics.
When used in conjunction with Marker Channel recordings and
concurrent ECG, this mode may be used to observe the conducted
ventricular rhythm without affecting atrial pacing.
AAI/ADI modes are generally not appropriate for heart failure
patients with normal sinus rhythm. In these modes, patients may
not receive cardiac resynchronization therapy.
Pacing modes
AAI / ADI Modes
29
Figure 1-12. Example of AAI Mode Operation
InSync Model 8040 Device Reference Guide
30
Pacing Rate Interval
Parameters:
Pacing Rate = 60 ppm (1000 ms)
Ventricular Refractory Period = 300 ms
Pacing Rate Interval
Chapter 1
AAT / VVT Modes
AAT / VVT Modes
Pacing occurs at the programmed rate, but a nonrefractory
sensed event triggers an immediate pacing output (rather than
inhibiting such output). Except that pacing outputs occur when
events are sensed, the triggered modes operate identically to the
corresponding inhibited modes.
Note: Permanently triggered pacing will not occur faster than
300 ms (200 ppm) from the previous paced event. Temporary
programmed triggered pacing is not limited to 300 ms (200 ppm).
AAT/VVT modes are generally not appropriate for heart failure
patients with normal sinus rhythm. In these modes, patients may
not receive cardiac resynchronization therapy.
Besides being directly programmable, the DOO mode is the
magnet mode of the corresponding dual chamber modes, except
for the VDD mode, which is the VOO mode. AOO and VOO modes
are the magnet modes of the corresponding atrial and ventricular
single chamber modes, respectively.
In general, these modes should not be used in heart failure
patients.
The DOO mode provides AV sequential pacing at the
programmed Lower Rate with no inhibition by intrinsic events.
The AOO and VOO modes provide pacing at the programmed
Pacing Rate with no inhibition by intrinsic events in the
applicable chamber.
Figure 1-15. Example of DOO Mode Operation
InSync Model 8040 Device Reference Guide
ODO / OAO / OVO Modes
Warning: The ODO, OVO, and OAO modes should never be
permanently programmed for pacemaker-dependent patients.
■
For such patients, the programmer’s Inhibit function may be
used for brief interruption of output.
■
Use of a magnet with these modes causes OOO operation,
i.e., no sensing or pacing. Do not leave the patient in any of
these modes without constant monitoring and supervision.
In the ODO, OAO, and OVO modes, sensing occurs in the
designated chamber(s). When used in conjunction with Marker
Channel telemetry and concurrent ECG, these modes may be
used to observe underlying rhythms.
■
Blanking periods in these modes are automatically minimized
to maximize the sensing window(s). Thus, Marker Channel
telemetry may display sense markers for cardiac events (for
example, far-field R waves) that otherwise would not appear
due to longer blanking.
■
No timing intervals or refractory periods are used.
Pacing modes
ODO / OAO / OVO Modes
33
These modes should not be permanently programmed for heart
failure patients, as no therapy is provided.
InSync Model 8040 Device Reference Guide
Introduction to Rate Responsive Pacing36
Activity Threshold38
Activity Rate Response40
Acceleration and Deceleration Times42
Rate Response Optimization44
Rate response
2
2
36
Chapter 2
Introduction to Rate Responsive Pacing
Introduction to Rate Responsive Pacing
Overview
Rate responsive modes are available for those heart failure
patients who may develop a need for rate responsive pacing. Rate
responsive pacing was not studied in this patient population.
These modes should not be programmed unless the patient needs
this type of support.
Activity-based rate responsive pacing varies the pacing rate in
response to the patient’s detected physical activity. The device
offers the following rate responsive modes:
■
Dual chamber modes: DDDR, DDIR, DVIR, DOOR
■
Single chamber modes: VVIR, VDIR, VOOR, AAIR, ADIR,
AOOR
All of the rate response features discussed in this section apply to
all of these modes, except for Rate Response Optimization. Refer
to “Parameter values and restrictions” on page 181 for specific
capabilities.
Sensor-Indicated Rate
In rate responsive modes, pacing occurs at the sensor-indicated
rate unless inhibited by sensed events. The pacing rate may vary
as the patient’s activity level changes. The sensor-indicated rate is
determined from the activity sensor signal and the following
programmable parameters:
■
Lower Rate
■
Upper Activity Rate
■
Activity Threshold
■
Activity Rate Response
■
Acceleration
■
Deceleration
The Lower Rate and Upper Activity Rate respectively control the
minimum and maximum sensor-indicated pacing rates. The other
parameters are described in the sections that follow.
InSync Model 8040 Device Reference Guide
Typical Rate Response Settings
The nominal rate response parameters are adequate for many
patients (Activity Threshold = Medium, Activity Rate
Response = 7, Acceleration = 0.5 minutes, and Deceleration =
5 minutes). For heart failure patients, rate response should only be
programmed when necessary.
■
For most patients, the device may be programmed to operate
at or near the programmed Lower Rate when the patient is
lying, sitting, or standing. If the patient has an elevated pacing
rate at rest, Activity Threshold may need to be programmed to
a higher setting.
■
When the patient is walking at a moderate pace, the pacing
rate will typically increase to about 90 ppm, and up to 120 ppm
during more brisk exercise. If the patient has minimal rate
response during exercise, Activity Threshold may need to be
programmed to a lower setting.
■
A simple programmer-directed exercise test may be used to
tailor rate response settings to a patient’s needs (refer to
InSync Model 8040 Device Programming Guide).
■
The Rate Histogram or Rate versus Time diagnostic may be
used to validate programmed rate response settings.
Rate response
Introduction to Rate Responsive Pacing
37
InSync Model 8040 Device Reference Guide
38
Time
Activity Threshold = Medium
Activity
Sensor
Output
High
MHigh
Med
MLow
Low
Low
MLow
Med
MHigh
High
Settings
Chapter 2
Activity Threshold
Activity Threshold
Overview
The programmable Activity Threshold determines the minimum
intensity of detected physical activity to which the device
responds.
How Activity Threshold Influences Rate
A piezoelectric crystal, bonded to the inside of the titanium device
shield, is deflected by activity-induced pressure waves within the
body. The sensor converts these pressure waves into electrical
signals. The programmed Activity Threshold screens out activity
signals below the selected setting. Detected sensor signals will
vary from patient to patient due to body structure, placement of
device, and so forth. Only sensor signals whose amplitude
exceeds the programmed Activity Threshold are used in
computing the sensor-indicated pacing rate. The lower the Activity
Threshold, the smaller the signal required to influence the pacing
rate, as shown in Figure 2-1.
Figure 2-1. Activity Sensor Signal with Threshold Set to Medium
InSync Model 8040 Device Reference Guide
Evaluating the Activity Threshold Setting
Marker Channel telemetry may be used to record or display the
activity sensor signal for evaluation. Few (or no) sensor detect
markers should appear when the patient is sitting quietly.
Typical Rate Performance
The Medium setting for Activity Threshold usually provides
satisfactory rate response while minimizing response to vibration
sources when the patient is inactive. Walking increases pacing
rate; sitting results in pacing at or near the Lower Rate. Use the
table below as a guide for selecting an appropriate setting.
Note: External pressure (such as lying prone) coupled with a Low
Activity Threshold setting may cause the pacing rate to increase.
Tab le 2-1. Activity Threshold Guidelines
Programmable
Settings
Typical Rate Performance
Rate response
Activity Threshold
39
LowResponds to most body activity, including
Medium/LowLimited response to minimal exertion;
MediumResponds to moderate and vigorous body
Medium/HighLimited response to moderate body
HighResponds to only vigorous body movements
minimal exertion.
responds to moderate or greater exertion.
movements and exertion.
movements and exertion.
and exertion.
InSync Model 8040 Device Reference Guide
40
Chapter 2
Activity Rate Response
Activity Rate Response
Overview
Basic Operation
The Activity Rate Response slope, in conjunction with the Lower
Rate and Upper Activity Rate, establishes the steady-state pacing
rate for a given level of detected activity (indicated by activity
signals that exceed Activity Threshold).
■
Ten Activity Rate Response slopes are available, with the
most responsive setting (10) providing the greatest
beat-to-beat rate change for a given change in detected
activity.
■
In general, more conditioned patients have greater cardiac
reserves, and they may require a lower programmed Activity
Rate Response setting.
The higher Activity Rate Response settings result in a higher
sensor-indicated rate for a given level of detected activity, as
follows:
■
All Activity Rate Response slopes are linear and extend from
the Lower Rate to the Upper Activity Rate.
■
The Upper Activity Rate can be attained with any Activity Rate
Response slope.
■
When the activity level stabilizes, the sensor-indicated rate will
stabilize.
Determining the Steady-State Pacing Rate
For any Activity Rate Response slope, the steady-state rate
corresponding to a given level of activity depends on Lower Rate
(LR) and Upper Activity Rate (UAR).
Figure 2-2 shows the Activity Rate Response slopes for two sets
of rate limits, first for an elderly patient and second for a younger
patient. For a given Activity Rate Response slope (for example,
slope 7), both patients achieve their Upper Activity Rates for the
same level of sustained sensor-detected activity, but the rates are
quite different. Use the programmed rate limits and Activity Rate
Response to match the rate prescription to the patient’s needs.
InSync Model 8040 Device Reference Guide
elderly patient
younger patient
Pacing Rate (ppm)
Pacing Rate (ppm)
Increasing activity
Increasing activity
Rate response
Activity Rate Response
41
Figure 2-2. Activity Rate Response Settings
InSync Model 8040 Device Reference Guide
42
Time (Minutes)
Rate Range
Lower
Rate
Upper
Activity
Rate
Activity Acceleration
Programmable Settings
0.25 Minutes
0.5 Minutes
1.0 Minutes
Chapter 2
Acceleration and Deceleration Times
Acceleration and Deceleration Times
Overview
Programmable Acceleration and Deceleration times control how
rapidly the pacing rate changes in response to increased or
decreased physical activity. One programmable Deceleration
option, Exercise Deceleration, provides an extended deceleration
period following prolonged exercise.
Rate responsive pacing was not studied in this patient population.
Consider this when deciding how to program acceleration time.
Acceleration Operation
Acceleration time is the time required to achieve approximately
90% of the difference between the current rate and a higher
steady-state rate consistent with the current level of activity.
Figure 2-3 shows a graphic representation of the Acceleration
curves at the onset of strenuous exercise.
InSync Model 8040 Device Reference Guide
Figure 2-3. Activity Acceleration Curves
Deceleration Operation
Time (Minutes)
Rate Range
Lower
Rate
Upper
Activity
Rate
Activity Deceleration
Programmable Settings
2.5 Minutes
5 Minutes
10 Minutes
Deceleration time is the time required to achieve approximately
90% of the difference between the current rate and a lower
steady-state rate consistent with the current level of activity.
Figure 2-4 shows a graphic representation of the Deceleration
curves at an abrupt cessation of strenuous exercise.
Rate response
Acceleration and Deceleration Times
43
Figure 2-4. Activity Deceleration Curves
Exercise Deceleration Operation
Exercise Deceleration extends the rate slowing period following
an exercise episode, providing up to 20 minutes of rate
deceleration. When it is programmed On, the device uses activity
sensor data to detect periods of vigorous, prolonged exercise. At
the end of such an exercise period, the device uses a longer
deceleration rate (longer deceleration time) for the central portion
of the programmed rate range. The actual deceleration rate is
determined dynamically based on the intensity and duration of
exercise and the new level of activity.
Figure 2-5 shows the composite deceleration curve that applies
after the abrupt cessation of sustained exercise.
InSync Model 8040 Device Reference Guide
44
Time (Minutes)
Rate Range
Lower
Rate
Upper
Activity
Rate
5 Minute Deceleration Curve
Begins Exercise Deceleration
Ends Exercise
Deceleration
5 Minute
Deceleration
Curve
Chapter 2
Rate Response Optimization
.
Figure 2-5. Exercise Deceleration
Rate Response Optimization
Overview
Rate Response Optimization is a programmable On/Off feature
that adapts rate responsive pacing to the patient’s changing
pacing needs by automatically reprogramming the Activity Rate
Response setting.
InSync Model 8040 Device Reference Guide
Caution: Rate Response Optimization may not be appropriate for
patients where it is necessary to limit the rate response to
exercise. Rate response pacing has not been studied in this
patient population.
■
■
The clinician programs the Patient Lifestyle parameter based
on the patient’s usual level of activity or work: Active,
Moderately Active, or Less Active. The Moderately Active
setting should work for most patients.
The adjustment of Activity Rate Response is based on a
running average of the patient’s activity.
■
For dramatic changes in patient activity, Activity Rate
Response adjustment may occur in as few as three days. (An
example might be an acute illness that limits the patient’s
activity.)
■
For more gradual changes in patient activity, the adjustment
may require two weeks or more. (An example might be the
period immediately following implant when cardiovascular
training effects sometimes occur.)
When reprogramming occurs, the Event Summary reports the
number of adjustments the next time the device is interrogated.
Rate Response Optimization Operation
Activity Rate Response slope adjustment is based on the average
number of daily exercise episodes. The device cumulates the
number of exercise episodes into a running average. Once each
day, the running average is compared with minimum and
maximum exercise episode values appropriate for the
programmed Patient Lifestyle.
■
If the average is above the maximum, the device adjusts the
programmed Activity Rate Response slope to the next lower
(less aggressive) setting.
■
If the average is below the minimum, the device adjusts the
programmed Activity Rate Response slope to the next higher
(more aggressive) setting.
Rate response
Rate Response Optimization
45
Programming Considerations
When programming Rate Response Optimization On, the clinician
should consider how the programmed Patient Lifestyle and
Activity Threshold settings may affect the actual rate response
observed for a given patient. The clinician should also consider
how adjustment of the Activity Rate Response slope affects
achieving the programmed Upper Activity Rate.
■
The Patient Lifestyle parameter determines how easily the
number of daily exercise episodes are achieved for the
patient's activity level. Programming the Less Active setting for
a normally active patient may result in no change or a change
to a lower (less aggressive) setting of the Activity Rate
Response parameter. Conversely, programming the Active
setting for a sedentary or minimally active patient may result in
InSync Model 8040 Device Reference Guide
46
Chapter 2
Rate Response Optimization
a change to a higher (more aggressive) setting. Select the
appropriate setting with consideration of the device patient’s
activity level against the population as a whole. For example,
a 65-year-old heart failure patient may be considered to be
“less active” in relation to the general population.
■
The Activity Threshold parameter determines the minimum
intensity of detected physical activity to which the device
responds by pacing. Sensor driven pacing contributes to the
number of daily exercise episodes. Programming a less
sensitive setting (e.g., Medium/High) for a normally active
patient may contribute to fewer daily exercise episodes and
result in a change to a higher (more aggressive) setting of the
Activity Rate Response parameter.
■
The programmed Upper Activity Rate can be achieved with
any of the programmed Activity Rate Response slopes. If the
device reprograms the Activity Rate Response parameter, the
level of exertion to achieve the Upper Activity Rate will
change. Lower (less aggressive) settings will require more
exertion, whereas higher (more aggressive) settings will
require less exertion.
Note: Extended periods of patient inactivity (e.g., bed rest due to
illness) may cause the device to adjust the Activity Rate Response
parameter to a higher (more aggressive) setting. This may be
important for heart failure patients.
InSync Model 8040 Device Reference Guide
Reporting Rate Response Parameter Changes
When Rate Response Optimization is active, the Activity Rate
Response parameter may be changed from the setting last
programmed. When the device is interrogated, the Event
Summary will report the total number of adjustments.
Recording Rate Response Optimization Events
Rate response adjustment may be recorded using the Rate
Response Optimization diagnostic. When this diagnostic is
selected, it records the time and the new parameter setting
whenever the Activity Rate Response is reprogrammed. Two
recording methods are available:
■
The Frozen method documents up to 120 changes.
■
The Rolling method documents the most recent 120 changes,
overwriting older data, if necessary.
Because 120 changes between programming sessions are highly
unlikely, the two recording methods will generally perform
identically.
Figure 2-6 shows how the Rate Response Optimization report is
presented on the programmer. The report can be printed out or
displayed on the programmer.
Rate response
Rate Response Optimization
47
Figure 2-6. Example of Rate Response Optimization Diagnostic
InSync Model 8040 Device Reference Guide
48
Chapter 2
Rate Response Optimization
Possible Loss of Diagnostic Data
Certain programmer functions erase stored diagnostic data.
Review or print optimization data before performing any of the
following functions:
■
■
■
Note: The Rate Response Optimization diagnostic is disabled
and all episode data erased if the Lead Monitor automatically
enables the Lead Trend diagnostic. See “Monitoring Lead
Stability” on page 86 for more information on this feature.
Diagnostic data collection is suspended if the Elective
Replacement Indicator is set, but data already collected are
retained.
programming another diagnostic,
using the programmer “Clear Data” command, or
initiating an activity threshold test or exercise test.
InSync Model 8040 Device Reference Guide
Rates50
AV Intervals58
Rate Adaptive AV 61
Blanking Periods64
Refractory Periods66
High Rate Atrial Tracking74
Device timing
3
3
50
Chapter 3
Rates
Rates
Overview
The following programmable rates control timing in the device:
■
Normal operating rates:
– Lower Rate
– Upper Tracking Rate
– Upper Activity Rate
■
Other operating rates:
– Sleep Rate (for Sleep Function)
– Hysteresis Rate (for single chamber demand modes)
– Intervention Rate (for Rate Drop Response)
Additionally, rates calculated by the device are used for some
operations. These are:
■
sensor-indicated rate
■
mean atrial rate
The other operating rates are described in “Special therapy
options” on page 93 along with the functions that use them. The
normal rates are described in this chapter.
InSync Model 8040 Device Reference Guide
A-A and V-V Timing
Device timing
Rates
A-A Timing – In all modes that pace the atrium, the device times
from atrial event to atrial event (A-A timing). This timing method
mimics a natural sinus rhythm, producing A-A intervals that are
nearly equal, except when timing is interrupted by:
■
PACs in the DDIR and DDI modes,
■
PVCs in the DDDR, DDD, DDIR, and DDI modes (PVC
Response operation),
■
A ventricular sensed event during the VA interval in the DVIR
and DVI modes, or
■
An atrial refractory sensed event that triggers an NCAP
extension.
VA intervals vary due to adjustments by A-A timing operations in
order to achieve sensor-indicated or lower rate operation in the
presence of varying AV conduction.
V-V Timing – In modes that do not pace the atrium (e.g., VDD or
VDIR) or single chamber ventricular modes, the device times from
ventricular event to ventricular event (V-V timing).
SAV Interval = 180 msVentricular Refractory Period = 240 ms
Chapter 3
Rates
Lower Rate
The programmed Lower Rate defines the slowest rate at which
pacing occurs during a mode’s basic operation. In rate responsive
modes, in the absence of sensor-detected activity, the
sensor-indicated rate is equal to the programmed Lower Rate.
Figure 3-1. Example of Lower Rate Operation
InSync Model 8040 Device Reference Guide
Operating Lower Rate
Under certain circumstances, the programmed Lower Rate may
be overridden by an operating lower rate that is higher or lower
than the programmed value. The following rates may become the
operating lower rate:
■
■
■
■
■
■
■
Selecting a Lower Rate
Program the Lower Rate to maintain adequate heart rates during
periods of inactivity or during pauses in atrial rhythms in the
DDDR, DDD, VDD, AAIR, ADIR, AAI, and ADI modes.
Device timing
Rates
Switching from and back to atrial tracking mode (for Mode
Switch)
Sleep rate (for Sleep function)
Intervention rate (for Rate Drop Response)
Hysteresis rate (for single chamber modes)
Threshold margin test rate of 100 ppm
Magnet mode rate of 85 ppm
Elective replacement indicator rate of 65 ppm
53
Note: In the VDD mode, atrial tracking near the Lower Rate may
result in V-V intervals that exceed the Lower Rate interval. This is
normal operation.
InSync Model 8040 Device Reference Guide
54
Activity
Activity
Parameters:
Sensor-Indicated Rate = 90 ppm (667 ms)
PAV Interval = 200 msPVARP = 300 ms
SAV Interval = 190 msVentricular Refractory Period = 220 ms
Sensor-Indicated Interval
Sensor-Indicated Interval
Chapter 3
Rates
Sensor-Indicated Rate
The sensor-indicated rate is the basic pacing rate in all rate
responsive modes (DDDR, DDIR, DVIR, DOOR, VVIR, VDIR,
VOOR, AAIR, ADIR, and AOOR). It is determined by the device
based on the sensor-detected level of patient activity and the
programmed rate response parameters. The sensor-indicated
rate will never be greater than the Upper Activity Rate or less than
the Lower Rate.
Figure 3-2. Example of Sensor-Indicated Rate Operation
In rate responsive modes, the sensor-indicated rate tracks the
activity signal detected by the piezoelectric crystal sensor’s
frequency and amplitude.
■
In dual chamber rate responsive modes, the sensor-indicated
interval is the AS-AP or AP-AP interval.
■
In single chamber rate responsive modes, the
sensor-indicated interval is the A-A or V-V interval. In these
modes, sensor-indicated rate intervals start with a sensed or
paced event in the chamber being paced.
InSync Model 8040 Device Reference Guide
Sensor Indicated Rate Effect on Other Intervals
Upper Tracking Rate
Parameters:
Sensor-indicated Rate =
75 ppm (800 ms)
Upper Tracking Rate =
100 ppm (600 ms)
SAV Interval = 200 ms
The sensor-indicated rate is used to determine the values of
certain other timing intervals. These intervals are:
■
Rate adaptive paced AV (PAV) interval
■
Sensor-varied PVARP (even in nonrate responsive DDD and
VDD modes)
Upper Tracking Rate
For heart failure patients, the upper tracking rate should be set to
a value above the patient’s expected maximum intrinsic rate to
insure that cardiac resynchronization therapy is provided
throughout the range of patient activity. “Device Programming
Recommendations” on page 178.
The programmable Upper Tracking Rate is the maximum rate at
which the ventricle may be paced in response to sensed atrial
events in the DDDR, DDD, and VDD modes. Sensed atrial events
below the Upper Tracking Rate will be tracked at a 1:1 ratio, but
sensed events above the Upper Tracking Rate result in
pacemaker Wenckebach by gradual extension of the SAV
(causing for example 6:5, 4:3, 3:2, or 2:1 block). The Upper
Tracking Rate usually should be programmed to a value less than
the 2:1 block rate. Refer to “High Rate Atrial Tracking” later in this
chapter for details.
Device timing
Rates
55
Figure 3-3. Example of Upper Tracking Rate (Wenckebach) Operation
InSync Model 8040 Device Reference Guide
56
Chapter 3
Rates
Note: In situations where the patient is sensitive to variations in
the AV interval, special consideration should be given to selecting
a value for Upper Tracking Rate that will provide protection against
inappropriately high ventricular rates while reducing the variations
in the SAV that result from Wenckebach operation (see
“Pacemaker Wenckebach” on page 75).
Upper Activity Rate
In rate responsive modes, the programmable Upper Activity Rate
provides the upper limit for the sensor-indicated rate during
physical activity, particularly during vigorous exercise. In the
DDDR mode, the Upper Activity Rate may be higher than, lower
than, or the same as the Upper Tracking Rate.
Programming Considerations and Restrictions
Programming a combination of high Upper Activity Rate and
Upper Tracking Rate and a long refractory period may result in a
shorter “sensing window.” Loss of sensing in such cases could
result in competitive pacing (unless Non-Competitive Atrial Pacing
is programmed On). See “Non-Competitive Atrial Pacing” on
page 99 for more information.
Programming the Upper Tracking Rate to a value greater than the
Upper Activity Rate permits the atrial rhythm to be tracked to a rate
higher than the sensor-driven rate.
The Upper Activity Rate and/or Upper Tracking Rate must be
greater than the Lower Rate.
Rate Limit
An internal circuit, independent of the pacing timers, limits single
chamber atrial or ventricular pacing rates to 200 ppm for most
single component failures. For dual chamber modes, atrial and
ventricular rates are limited independently to 200 ppm. The rate
limit is automatically disabled during temporary pacing in the AAI,
ADI, AAT, AOO, VVI, VDI, VVT, and VOO modes to allow high rate
pacing for diagnostic or therapeutic purposes.
InSync Model 8040 Device Reference Guide
Limiting Response to Environmental Vibration
In a rate responsive mode, the device may respond to strong
environmental vibrations (such as from heavy machinery, extreme
acoustic sources, prolonged device palpation, exercise
equipment, and so forth) by pacing at elevated rates. (Such rates
are always limited by the programmed Upper Activity Rate.) To
eliminate this response, move the patient away from the vibration
source. If this is not possible, reprogram the Upper Activity Rate
setting to a value that is appropriate to the patient’s condition or
adjust the Activity Threshold to a less sensitive setting.
Possible Atrial Competition at High Rates
At high sensor-driven rates in the DDDR and DDIR modes,
sensor-driven pacing may approximate the intrinsic atrial rate, with
some intrinsic atrial events falling in the PVARP. This could result
in asynchronous pacing with the potential for competitive atrial
pacing. Consider the potential for asynchronous pacing at high
rates before selecting an Upper Activity Rate, especially for
patients known to be susceptible to induction of atrial
tachyarrhythmias. Weigh the benefits of high rate sensor-driven
pacing against the potential for competitive pacing.
Device timing
Rates
57
Mean Atrial Rate
Note: Use of the Rate Adaptive AV feature, sensor-varied PVARP,
and Non-Competitive Atrial Pacing (DDDR mode only) can reduce
the likelihood of the type of asynchronous pacing described
above.
The mean atrial rate (MAR) is a running average of the atrial rate
for use by the Rate Adaptive AV and automatic PVARP features.
The average uses all A-A intervals (except those starting with an
atrial sense or atrial refractory sense and ending with an atrial
pace). In order to respond quickly to rapidly increasing atrial rates,
the average gives preference to shorter A-A intervals over longer
intervals when calculating the MAR. Figure 3-4 shows how the
MAR tracks an increasing intrinsic atrial rate.
InSync Model 8040 Device Reference Guide
58
Atrial Rate Increasing by 2 bpm/beat
Time (Seconds)
Rate (bpm)
MAR
Intrinsic Rate
Chapter 3
AV Intervals
AV Intervals
Figure 3-4. Mean Atrial Rate
Overview
In dual chamber modes, the AV intervals determine the time
between the occurrence of an atrial event and the scheduled
delivery of a ventricular stimulus. Separate AV intervals for paced
and sensed atrial events are available. The lengths of these
intervals may be programmed to fixed values or (optionally) rate
adaptive.
■
Paced AV Interval (PAV) – PAV follows an atrial pace in the
DDDR, DDD, DDIR, DDI, DVIR, DVI, DOOR, and DOO
modes. The PAV interval duration may differ from the
InSync Model 8040 Device Reference Guide
programmed value due to:
– Rate Adaptive AV operation,
– Ventricular Safety Pacing, or
– Non-Competitive Atrial Pacing therapy.
Device timing
PAV
Interval
PAV
Interval
PAVPAV
AV Intervals
Figure 3-5. Example of PAV Interval Operation
■
Sensed AV Interval (SAV) – SAV follows an atrial sensed
event in atrial synchronous pacing modes (DDDR, DDD, and
VDD). The SAV interval duration may differ from the
programmed value due to:
– Rate Adaptive AV operation or
– Wenckebach operation, where the SAV is extended to
avoid violation of the Upper Tracking Rate or the total atrial
refractory period while tracking a fast intrinsic atrial rate.
59
InSync Model 8040 Device Reference Guide
60
SAV
Interval
SAV
Interval
SAVSAV
Chapter 3
AV Intervals
Selecting PAV and SAV
Figure 3-6. Example of SAV Interval Operation
When programming AV intervals, the general hemodynamic goal
is to assure that, to the extent possible, left-atrial systole is
completed before left-ventricular systole begins. To achieve this,
the AV interval durations may be adjusted independently of each
other.
■
To accommodate the difference in interatrial conduction times,
the SAV usually should be programmed to a shorter duration
than the PAV, typically 30 to 50 ms shorter. (If an SAV greater
than the PAV is selected, the programmer notes that this is not
usual, but the selected values may be programmed if clinically
warranted.)
■
When the SAV is longer than the PAV, a V pace following an
atrial sense will always occur after the full SAV, even when the
sensor-indicated rate or Lower Rate interval expires first.
■
In certain patients, short AV intervals may be used as a
prophylaxis for AV nodal or accessory pathway reentrant
tachycardias in dual chamber modes.
■
For consistent ventricular pacing, the programmed setting for
PAV and SAV must be less than the patient’s intrinsic PR
interval.
InSync Model 8040 Device Reference Guide
■
Rate Adaptive AV
Overview
For heart failure patients, rate adaptive AV can be used to ensure
pre-stimulation of the patient’s ventricles at higher rates.
In the normal heart, AV conduction times tend to shorten as the
heart rate increases and lengthen as the heart rate decreases.
The Rate Adaptive AV (RAAV) feature, available in the DDDR,
DDD, DDIR, DVIR, DOOR, and VDD modes, mimics this
physiologic response. When RAAV is programmed On, the device
shortens AV intervals for atrial rates within a programmed rate
range. This feature provides increased opportunity for atrial
sensing, as follows:
■
■
Device timing
Rate Adaptive AV
Long PAV intervals (≥ 250 ms) should not be used, as they
may not allow for optimal biventricular pacing.
Shortened SAV intervals increase the tracking range at fast
atrial rates by shortening the total atrial refractory period
(TARP) and increasing the 2:1 block rate. Refer to “Total Atrial
Refractory Period (TARP)” and “High Rate Atrial Tracking”
later in this chapter for more information.
Shortened PAV intervals lengthen the atrial sensing window of
the VA interval at higher sensor-driven rates.
61
Note: RAAV will not shorten AV intervals to less than 30 ms.
Programming for Rate Adaptive AV
For RAAV operation, the SAV and PAV are programmed (as
applicable) to the values desired for low rates. Three additional
programmable parameters control how AV intervals are adjusted
at higher rates:
Start Rate – RAAV operation of shortening AV intervals begins at
this rate.
Stop Rate – The shortest SAV and PAV occur at this rate and at
all higher rates, up to the upper rate limits.
InSync Model 8040 Device Reference Guide
62
Rate (ppm)
AV Interval (ms)
Parameters:
Programmed SAV = 170 msStart Rate = 80 ppmMinimum PAV = 100 ms
Programmed PAV = 200 msStop Rate = 150 ppmMinimum SAV = 70 ms
Start RateStop Rate
Programmed PAV
Programmed SAV
Minimum SAV
Minimum PAV
R
a
t
e
A
d
a
p
t
i
v
e
P
A
V
R
a
t
e
A
d
a
p
t
i
v
e
S
A
V
Chapter 3
Rate Adaptive AV
Minimum AV Interval – A minimum value is selected for either
the SAV or PAV, depending on the pacing mode:
■
Minimum Sensed AV Interval - The shortest allowable SAV,
used at or above the Stop Rate, is programmed in the DDDR,
DDD, and VDD modes.
■
Minimum Paced AV Interval - The shortest allowable PAV,
used at or above the Stop Rate, is programmed in the DDIR,
DVIR, and DOOR modes. In the DDDR mode, the value is
automatically determined by the programmer.
Figure 3-7 shows how the SAV and PAV intervals are linearly
shortened as the rate increases from below the Start Rate to
above the Stop Rate.
.
Figure 3-7. Rate Adaptive AV Operation (DDDR Mode)
InSync Model 8040 Device Reference Guide
RAAV Operations
Shortening of the AV interval(s) occurs when the appropriate rate
exceeds the programmed Start Rate, as follows:
■
SAV – The mean atrial rate determines SAV adjustments.
Because of how the mean atrial rate is calculated:
– SAV adjustments will lag during rapid increases or
decreases in intrinsic atrial rates.
– The SAV is not adjusted for isolated events (PACs).
– Atrial sensed or atrial refractory sensed events may affect
the SAV if the next atrial event is not paced since these A-A
events are not used in the mean atrial rate.
■
PAV – The sensor-indicated rate determines PAV
adjustments.
The approximate difference between programmed SAV and PAV
is maintained as the SAV and PAV intervals are adjusted.
Programming Considerations and Restrictions
Device timing
Rate Adaptive AV
63
RAAV subordinate parameters – The Start Rate must be less
than the Stop Rate. The Min Sensed AV and Min Paced AV
intervals must be less than the programmed SAV and PAV,
respectively. The Min Paced AV interval is set to a pending value
for DDDR or DDD mode switching.
InSync Model 8040 Device Reference Guide
64
Note: Black bars indicate blanking
periods.
1 1.Nonprogrammable Atrial
Blanking
2 2.Programmable
Post-Ventricular Atrial Blanking
3 3.Programmable Ventricular
Blanking
4 4.Nonprogrammable Ventricular
Blanking
Chapter 3
Blanking Periods
Using RAAV with Mode Switch
Blanking Periods
RAAV operation is enabled when Mode Switch is selected, RAAV
parameter choices are affected as follows:
■
Start Rate must be less than or equal to 70 ppm.
■
Stop Rate is forced to 150 ppm by the programmer.
■
Minimum SAV is forced to 30 ms.
■
Minimum PAV is forced anywhere from 30 to 100 ms for the
DDDR and DDD modes.
Blanking periods disable sensing for a programmable or
nonprogrammable interval. Signals that are blanked may originate
in either chamber or from outside sources such as noise from
muscle movement.
Figure 3-8. Example of Dual Chamber Blanking Operation
Nonprogrammable Blanking Periods
Immediately following a sensed or paced event in either chamber,
sensing for that chamber is blanked for a nonprogrammable
period that may vary from 50 to 100 ms. The actual duration of the
blanking period is determined dynamically by the device, based on
the strength and duration of the signal. Dynamic blanking prevents
sensing the same signal twice, while minimizing total blanking
time.
InSync Model 8040 Device Reference Guide
Post-Ventricular Atrial Blanking
The programmable Post-Ventricular Atrial Blanking (PVAB)
period, used in the DDDR, DDD, DDIR, DDI, and VDD modes,
prevents sensing of ventricular paced events or far-field R waves
on the atrial lead. Any ventricular event (paced or sensed) starts
the PVAB, which is also the first portion of the Post-Ventricular
Atrial Refractory period (PVARP). PVAB is limited to values equal
to or less than the programmed PVARP, except in the VDIR and
VDI modes since PVARP does not apply to these modes.
Note: PVAB is programmed to a value less than or equal to
PVARP.
Ventricular Blanking
The programmable Ventricular Blanking period, which follows an
atrial pacing stimulus in the DDDR, DDD, DDIR, DDI, DVIR, and
DVI modes, prevents ventricular inhibition or ventricular safety
pacing due to sensing of the atrial stimulus on the ventricular lead
(crosstalk). The Ventricular Blanking period also applies to the
ADIR and ADI modes to prevent sensing of the atrial pacing.
■
Long blanking periods (36 ms or greater) increase the
possibility of unsensed ventricular events.
■
Long blanking periods used in conjunction with long PAV
intervals (250 ms or greater) may result in pacing into the
T wave when intrinsic ventricular events are blanked and not
sensed. PAV values (200 ms or less) should reduce the
possibility of T wave pacing.
Device timing
Blanking Periods
65
InSync Model 8040 Device Reference Guide
66
Chapter 3
Refractory Periods
Single Chamber Atrial Blanking
The programmable single chamber atrial blanking period, used in
the AAIR, ADIR, AAI, ADI, and AAT mode, prevents sensing of
far-field R waves. It is started by a paced, sensed, or refractory
sensed atrial event.
Note: Atrial Blanking must be programmed at least 50 ms less
than the Atrial Refractory Period.
Refractory Periods
Overview
A refractory period is an interval during which an intrinsic event
sensed on a particular lead channel cannot start certain timing
intervals. Each refractory period begins with a blanking period,
during which no sensing occurs. During the unblanked portion of
a refractory period, sensing occurs, but sensed events may not
directly affect timing operations. Refractory periods are intended
to prevent certain timing intervals from being started by
inappropriate signals such as retrograde P waves, far-field
R waves, or electrical noise.
Though they may not start timing intervals, refractory sensed
events are monitored by the device, and they affect the operation
of PVC Response, Mode Switch, Rate Adaptive AV operation,
Non-Competitive Atrial Pacing, and other features for which the
periodicity or number of sensed events are pertinent. Refractory
sensed events are included on Marker Channel recordings and
Marker Channel Diagrams.
Post-Ventricular Atrial Refractory Period
The Post-Ventricular Atrial Refractory Period (PVARP) follows a
paced, sensed, or refractory sensed ventricular event in the
DDDR, DDD, DDIR, DDI, and VDD modes. It is intended primarily
to prevent the sensing of retrograde P waves that might promote
Pacemaker-Mediated Tachycardias (PMTs) in atrial tracking
modes. In the DDIR and DDI modes, PVARP prevents atrial
inhibition from retrograde P waves.
InSync Model 8040 Device Reference Guide
Device timing
PVARP
Refractory Periods
The first portion of the PVARP is the programmable
Post-Ventricular Atrial Blanking period (PVAB). During the
remainder of the PVARP, intrinsic atrial events may be sensed as
refractory sensed events (AR) and identified on Marker Channel
recordings and Marker Channel Diagrams, but they do not affect
stimulus timing. That is:
■
In the DDDR, DDD, and VDD modes, an SAV is not started.
■
In the DDDR, DDD, DDIR, and DDI modes, the scheduled
atrial pace is not inhibited.
67
Figure 3-9. Example of PVARP Operation
The duration of the PVARP may be selected as follows:
■
■
■
Sensor-Varied PVARP
When sensor-varied PVARP is programmed on, the device
determines a value for the PVARP based on the sensor-indicated
rate. The intended purpose of the sensor-varied PVARP depends
upon the mode:
■
The PVARP should be programmed to a value greater than
the patient’s retrograde (VA) time when retrograde conduction
is present.
Excessively long PVARPs may induce 2:1 block at high
intrinsic rates in atrial tracking modes (DDDR, DDD, and
VDD).
To reduce the 2:1 block point, PVARP can be set to vary based
on the sensor-indicated rate.
In the DDDR, DDD, and VDD modes, sensor-varied PVARP is
intended to do the following:
InSync Model 8040 Device Reference Guide
68
Upper
Activity
Rate
Lower
Rate
PAVPVARP
PAV
PVARP
Chapter 3
Refractory Periods
– Enhance protection against PMT at lower rates by
providing longer PVARPs at low sensor-indicated rates.
– Allow tracking of higher atrial rates (that is, provide a higher
2:1 block rate) by shortening the PVARP at high
sensor-indicated rates.
■
In the DDIR mode, the sensor-varied PVARP is intended to
promote AV synchrony by preventing inhibition of atrial pacing
by an atrial sense early in the VA interval. It also reduces the
likelihood of competitive atrial pacing at high sensor-indicated
rates.
Note: When Rate Drop Response is on in the DDD mode,
sensor-varied PVARP is not available.
The device determines the duration of the sensor-varied PVARP
as follows:
■
In the DDDR, DDD, and VDD modes, the sensor-varied
PVARP is limited to 400 ms at low rates and the programmed
PVAB at high rates (as shown in Figure 3-10).
■
In the DDIR mode, the sensor-varied PVARP is approximately
400 ms at low rates and the programmed PVAB at high rates.
■
In the DDDR, DDD, DDIR, and VDD modes, the sensor-varied
PVARP is automatically adjusted to maintain a 300-ms
sensing window (as shown in Figure 3-10).
Spontaneous PVARP Extension
The programmed PVARP duration and the sensor-varied PVARP
may be overridden by the PVC Response and PMT Intervention
features, as follows:
■
When the PVC Response feature is programmed On and a
device-defined PVC occurs, the PVARP is forced to 400 ms
for one cycle if a lesser value is in effect.
■
When PMT Intervention is programmed On and a
device-defined PMT is detected, the PVARP is forced to
400 ms for one cycle after the ninth paced ventricular event of
the PMT.
Device timing
Refractory Periods
69
Refer to “PVC Response” on page 104 and “PMT Intervention” on
page 102 for further details of the PVC Response and PMT
Intervention features and their interactions with PVARP.
InSync Model 8040 Device Reference Guide
70
TARPTARP
SAV + PVARP
SAV
PVARP
PVARP
SAV
Chapter 3
Refractory Periods
Total Atrial Refractory Period (TARP)
In dual chamber modes that sense in the atrium, the Total Atrial
Refractory Period (TARP) is the sum of two intervals, as follows:
■
AV Interval – The AV interval begins with an atrial event and
ends with a ventricular event. The first portion is a
nonprogrammable blanking period. Its complete duration is
determined as follows:
– In the DDDR, DDD, and VDD modes, the PAV or SAV
– In the DDIR and DDI modes, the AV interval starts with the
■
Post-Ventricular Atrial Refractory Period (PVARP) –The
PVARP is described earlier in this chapter.
interval is the AV interval.
first atrial sensed event in the VA interval or with an atrial
pacing stimulus; it ends when the PAV expires, even when
ventricular pacing is inhibited.
Figure 3-11. Total Atrial Refractory Period
During atrial tracking, TARP = SAV + PVARP, and its duration
determines the rate at which 2:1 block occurs. Refer to “High Rate
Atrial Tracking” on page 74 for more information.
InSync Model 8040 Device Reference Guide
Ventricular Refractory Period
VRP
The programmable Ventricular Refractory Period (VRP) follows
paced, sensed, and refractory sensed ventricular events
(including PVCs) in all dual chamber and ventricular modes that
sense in the ventricle. The VRP is intended to prevent sensing of
the T wave or a PVC. The first portion of the VRP is a
nonprogrammable blanking period. A ventricular refractory
sensed event affects device timing as follows:
■
Ventricular blanking and refractory periods restart in all
modes.
■
In the DDDR, DDD, and VDD modes, the Upper Tracking Rate
interval, PVARP, and PVAB also restart.
■
In the VVIR and VDIR modes, the Upper Activity Rate interval
restarts.
Note: In dual chamber modes, the VRP should be programmed
shorter than the PVARP.
Device timing
Refractory Periods
71
Figure 3-12. Example of Ventricular Refractory Period Operation
In dual chamber modes, a ventricular refractory sensed event
does not affect a scheduled sensor-driven or Lower Rate atrial
output. Thus, a sensor-driven atrial output pulse will initiate a PAV
with a ventricular output pulse following, unless inhibited.
InSync Model 8040 Device Reference Guide
72
Chapter 3
Refractory Periods
Atrial Refractory Period (Single Chamber)
The programmable Atrial Refractory Period (ARP) follows paced,
sensed, and refractory sensed atrial events. The ARP is used in
the AAIR, ADIR, AAI, ADI, and AAT modes. It is intended to
prevent inhibition due to far-field R wave sensing. The first portion
of the ARP is a programmable blanking period. The ARP should
be programmed to a value long enough (150 ms or greater) to
prevent far-field R wave sensing but short enough to ensure atrial
sensing up to the programmed Upper Activity Rate.
Noise Reversion
When sensing occurs during the Atrial Refractory Period (ARP) or
Ventricular Refractory Period (VRP), the refractory period (and its
blanking period) are restarted. The operation associated with
continuous refractory sensing in the ARP or VRP is called noise
reversion. Multiple restarts of the ARP or VRP (continuous noise
reversion) do not inhibit scheduled pacing. Device behavior during
continuous noise reversion is as follows:
■
Pacing occurs at the sensor-indicated rate for all rate
responsive modes (except VVIR and VDIR).
■
Pacing occurs at the Lower Rate for all nonrate-responsive
modes.
Figure 3-13. Example of Noise Reversion in DDDR at Sensor-Indicated Rate.
Device timing
73
Figure 3-14. Example of Noise Reversion in VVIR at Lower Rate.
InSync Model 8040 Device Reference Guide
74
Chapter 3
High Rate Atrial Tracking
Preventing Noise Sensing
On the ECG, noise reversion may be difficult to distinguish from
loss of sensing, but Marker Channel recordings will show
refractory sense markers when noise reversion occurs.
Note: Atrial sensing during the PVARP, or refractory period
following an atrial paced or sensed event in the DDDR, DDD,
DDIR, DDI, or VDD modes does not restart the refractory period.
An atrial refractory sensed event, however, will start a short
blanking period of 50 to 100 ms depending on the signal strength
and duration of the atrial event.
Noise reversion may be caused by electromagnetic interference
(EMI), myopotentials, excessively high output settings, or low
sensitivity settings. When it has been identified, noise reversion
usually can be reduced or eliminated by one of the following
actions:
■
Reprogram sensitivity to a less sensitive (higher numerical
value) setting.
■
Reprogram sensing to bipolar polarity (if available).
■
Reduce the amplitude and/or pulse width in the same or
opposite chamber.
■
Remove patient from EMI environment.
High Rate Atrial Tracking
Overview
In the DDDR, DDD, and VDD modes, the fastest atrial rate the
device can track is determined by the total atrial refractory period
(TARP), which is the sum of the SAV and the PVARP. Device
behavior at high atrial rates in these modes is determined by the
relationship between the TARP and the interval corresponding to
the Upper Tracking Rate. In the DDDR mode, the interval
corresponding to the Upper Activity Rate also must be considered.
For heart failure patients, 2:1 block and Wenckebach operation
should be avoided to prevent loss of cardiac resynchronization
therapy.
InSync Model 8040 Device Reference Guide
2:1 Block
Device timing
High Rate Atrial Tracking
When the intrinsic atrial interval is shorter than the TARP, some
atrial events will fall in the PVARP and not be tracked. At the rate
where this first occurs, ventricular tracking occurs only on
alternate beats, and 2:1 block ensues. In the DDD and VDD
modes, the ventricular pacing rate drops precipitously.
■
When sensor-varied PVARP is selected, the 2:1 block rate
may occur at a higher rate during activity due to shortening of
the PVARP, thus increasing atrial tracking.
■
When Rate Adaptive AV operation is selected, the SAV
shortens at high atrial rates, shortening the TARP and raising
the 2:1 block rate.
■
When the 2:1 block rate is less than the Upper Tracking Rate,
the Upper Tracking Rate cannot be achieved.
■
In the DDDR mode, pacing at the sensor-indicated rate may
prevent a precipitous rate drop at the 2:1 block point when
activity is present.
■
For patients with a documented propensity for prolonged or
sustained atrial fibrillation or flutter, the clinician can select
Upper Tracking Rate, SAV, and PVARP values that induce
2:1 block at a desired rate (2:1 block rate = 60,000/TARP).
Alternatives for controlling rates in these patients include use
of the Mode Switch feature and DDIR mode pacing.
■
In the DDDR mode, atrial competition may occur if Upper
Activity Rate exceeds the 2:1 block rate.
75
Pacemaker Wenckebach
When the 2:1 block rate exceeds the programmed Upper Tracking
Rate, pacemaker Wenckebach may occur. When the intrinsic rate
exceeds the Upper Tracking Rate, a pacing stimulus at the
expiration of the SAV would violate the Upper Tracking Rate. The
device therefore extends the SAV until the Upper Tracking Rate
interval expires. Subsequent SAVs require greater extension, until
an atrial event falls in the PVARP and is not tracked.
■
In the DDDR, DDD, and VDD modes, the result normally is a
fixed ratio between atrial and ventricular rates (3:2, 4:3, and so
forth).
■
In the DDDR mode, the pacemaker Wenckebach rate may be
smoothed by sensor-driven ventricular pacing, thereby
overriding the fixed ratio.
InSync Model 8040 Device Reference Guide
76
Upper Tracking
Rate Interval
Parameters:
Sensor-Indicated Rate = 90 ppm (667 ms)PVARP = 300 ms
The following example shows how pacemaker Wenckebach
operation occurs in the DDDR, DDD, or VDD modes.
Figure 3-15. Example of Pacemaker Wenckebach Operation
High Rate Operation in the DDDR Mode
Table 3-1 summarizes how the total atrial refractory period
(TARP), the Upper Tracking Rate (UTR) interval, and the Upper
Activity Rate (UAR) interval may interact at high atrial rates in the
DDDR mode.
Tabl e 3-1. Upper Rates Interaction With TARP
Relationship Between
TARP and Upper Rate
Intervals
InSync Model 8040 Device Reference Guide
TARP > both UAR and
UAR interval > TARP >
UTR interval > both UAR
a
Unless Non-Competitive Atrial Pacing is On; see “Non-Competitive Atrial Pacing”
on page 99.
UTR intervals
UTR interval
UAR interval > UTR
interval > TARP
interval and TARP
Wenckebach
Before 2:1
Block
Achieve
Upper
Tracking Rate
nonoyes
nonono
yesyesno
yesyesyes
Potential
Atrial
Competition
a
a
Lead/cardiac tissue interface
Selecting Pacing Parameters78
Selecting Sensing Parameters83
Monitoring Lead Stability86
Transtelephonic Capture Verification with TMT89
4
4
78
Chapter 4
Selecting Pacing Parameters
Selecting Pacing Parameters
Overview
Pacing outputs (amplitude and pulse width) should be set high
enough to guarantee reliable capture in the atrium and reliable
biventricular capture in the ventricles, but not so high as to
unnecessarily deplete the device battery. Whether selecting
pacing outputs at implant or follow-up, the same considerations
apply:
■
Select pacing polarity for leads
■
Determine pacing thresholds
■
Select appropriate outputs (pulse width and amplitude)
Selecting Pacing Polarity
If bipolar leads are used, Pacing Polarity for the atrial output and
the ventricular outputs can be programmed to bipolar or unipolar.
■
Bipolar Pacing Polarity – The lead tip is the active electrode;
the lead ring is the common electrode. Bipolar pacing is less
likely to produce muscle stimulation, but it produces smaller
pacing artifacts on the ECG.
■
Unipolar Pacing Polarity – The lead tip is the active electrode;
the device case is the common electrode. Unipolar pacing
produces large pacing artifacts that aid in ECG interpretation.
However, it is more likely to cause muscle stimulation,
especially at high pacing amplitudes.
Warning: No pacing output results when a unipolar lead is
present and Pacing Polarity is programmed to bipolar. (An
exception to this is the use of both a unipolar lead and a bipolar
lead for biventricular pacing in the ventricles. The outputs for this
ventricular unipolar/bipolar lead combination can be programmed
to unipolar. See the following section “Ventricular Polarity
Configurations”.
Ventricular Polarity Configurations
Since the programmed setting for polarity (Pacing Polarity or
Sensing Polarity) is common to both ventricular leads (not
independently programmable), special consideration should be
given to the lead polarity configurations.
InSync Model 8040 Device Reference Guide
Two bipolar leads – Ventricular polarity can be programmed to
either Bipolar or Unipolar. Both leads will function according to the
programmed polarity setting.
Two unipolar leads – Ventricular polarity must be programmed
to Unipolar. Both leads in this system can function only as unipolar
leads. The Bipolar setting results in no pacing or sensing.
One bipolar lead and one unipolar lead – Ventricular polarity
can be programmed to either Bipolar or Unipolar.
■
With the Unipolar setting, both leads operate as unipolar
leads.
■
With the Bipolar setting, the ring electrode on the bipolar lead
acts as the common electrode (current return path) for both
leads. This configuration thus results in “shared-ring” bipolar
pacing and sensing.
Bipolar Pacing Polarity Confirmation
Before programming from unipolar to bipolar pacing, the
programmer verifies the presence of a bipolar lead by testing lead
impedance for each lead. Testing is done under magnet operation
for four seconds at an Amplitude of 5.0 V and a Pulse Width of
1.0 ms if the permanent settings are at or below this level. If
permanent settings are above these values, the measurement will
be made at the permanent settings.
■
If bipolar lead impedance is between 200 ohms and
3000 ohms, a bipolar lead is assumed to be present.
Note: Impedance for the ventricular two-lead system is
measured across the parallel combination of both leads.
■
If bipolar lead impedance is outside this range, a unipolar lead
is assumed to be present. The programmer warns that the test
failed, and pacing polarity remains set to unipolar. This
interlock feature may be overridden to force lead pacing
polarity to bipolar.
Lead/cardiac tissue interface
Selecting Pacing Parameters
79
Warning: If the clinician overrides the bipolar lead verification
routine and programs bipolar polarity when a unipolar lead is
connected, no pacing output results.
InSync Model 8040 Device Reference Guide
80
Chapter 4
Selecting Pacing Parameters
Determining Stimulation Thresholds at Implant
At implant, use a Medtronic Pacing System Analyzer (PSA) to
determine threshold values for capture. Refer to the PSA manual
for detailed instructions.
Verifying Stimulation Thresholds at Follow-up
Medtronic programmers provide both automatic and manual
threshold tests for determining the patient’s stimulation threshold
at follow-up. Stimulation threshold resolution is determined by the
available increments of amplitude and pulse width. Included with
the automatic threshold test on the programmer is a
strength-duration graph that shows voltage safety margins.
For a detailed description of the threshold tests and test
procedures, refer to the InSync Model 8040 Device Programming
Guide.
Measuring Stimulation Thresholds in the Ventricles
Special consideration should be given to measuring stimulation
thresholds in the ventricles.
Output adjustment affects both leads – Adjusting amplitude
and pulse width values during threshold measurement changes
the output energy to both ventricular leads.
When the two ventricular leads have different thresholds –
When the two leads do not present the same energy requirement
to maintain capture, it is important to determine the thresholds for
the lead presenting the higher energy requirement. This is the
biventricular threshold, the point above which both leads will
maintain capture.
■
Measuring the biventricular threshold — During threshold
measurement as device output is decreased in a step-by-step
fashion, capture will be lost first by the higher-threshold lead,
while the other lead continues to pace. For this reason, it is
necessary to watch for changes in depolarization waveform
morphology (rather than the complete loss of pacing capture)
to identify the biventricular threshold point.
InSync Model 8040 Device Reference Guide
Lead/cardiac tissue interface
Output Settings (example):
Loss of Capture
by One Lead
3.0 V
0.20 ms 0.15 ms 0.12 ms
Selecting Pacing Parameters
■
Recognizing loss of biventricular capture — The loss of
biventricular capture in cases where the two ventricular leads
exhibit different thresholds may cause a widening of the
depolarization waveform as shown in the example below.
Since such variables as lead placement and depolarization
vectors can affect waveform shape, the actual morphology
change will vary with each patient.
Figure 4-1. Illustrative Example of Capture Loss by One Lead
When both leads have the same threshold – In cases where
there is biventricular capture and both ventricular leads have the
same threshold, there is only a single threshold point indicated by
complete loss of ventricular capture.
81
Caution: The presence of a single ventricular threshold point also
can indicate that biventricular capture was not present at the start
of threshold measurement (only one lead is maintaining capture).
In such cases, it may be advisable to conduct the threshold test
again, starting at higher amplitude and pulse width test values to
increase the likelihood that both leads are capturing at the start of
the test.
Achieving reliable biventricular capture – When the two
ventricular leads do not present the same energy requirement to
maintain capture, amplitude and pulse width must be set to
provide an adequate safety margin above the thresholds for the
lead presenting the higher energy requirement. Without consistent
capture by both ventricular leads, the benefits of biventricular
pacing are lost.
InSync Model 8040 Device Reference Guide
82
Chapter 4
Selecting Pacing Parameters
Selecting Output Parameters
For Further Information
Generally, to provide an adequate safety margin, select a pacing
voltage twice the chronic stimulation threshold voltage for a given
pulse width. For most patients, pacing outputs are the major
contributor to battery depletion.
■
To maximize battery longevity, select the lowest amplitudes
and pulse widths that provide at least a 2:1 voltage safety
margin.
■
To assure reliable biventricular capture in the ventricles, the
voltage safety margin should be based on the threshold of the
lead exhibiting the higher energy requirement.
■
When amplitudes greater than 2.5 V are required during the
maturation phase of the lead(s), threshold(s) and output
setting(s) should be carefully reevaluated at the first follow-up.
Note: High output pacing at 7.5 V may affect ECG or
intracardiac electrogram (EGM) waveform quality and
potentially cause crosstalk or self-inhibition.
For further information on device longevity under various pacing
scenarios refer to “Longevity Projections (Normal Operating Life)”
on page 202. For information on amplitude and pulse width
parameter settings refer to “Programmable Modes and
Parameters” on page 182. For information on crosstalk or
self-inhibition refer to “Precautions” on page 212.
InSync Model 8040 Device Reference Guide
Selecting Sensing Parameters
Overview
Sensitivity determines the minimum intracardiac signal that the
device can detect when intrinsic atrial or ventricular events occur.
Whether selecting sensing parameters at implant or verifying
sensing at follow-up, the same considerations apply:
■
Select sensing polarity for leads
■
Determine sensing thresholds
■
Select appropriate sensitivity settings
Selecting Sensing Polarity
Atrial and ventricular sensing polarities can be programmed for
each chamber when used with bipolar leads.
■
Bipolar Sensing Polarity – The lead tip and the lead ring
electrode are the poles of the sensing circuit. Because bipolar
sensing is more localized, it reduces the likelihood of sensing
myopotentials and electromagnetic interference. It may also
permit sensitivity to be programmed to a more sensitive
setting.
■
Unipolar Sensing Polarity – The lead tip and the noninsulated
device case are the sensing electrodes. Unipolar sensing may
allow sensing of smaller intrinsic signals than does bipolar
sensing and therefore, can be selected when intrinsic cardiac
signals are difficult to detect with bipolar sensing. Oversensing
due to myopotentials is more common with unipolar sensing
than with bipolar sensing.
Lead/cardiac tissue interface
Selecting Sensing Parameters
83
Refer to “Ventricular Polarity Configurations” on page 78 for
information about the ventricular two-lead system.
Bipolar Sensing Polarity Confirmation
Before programming from unipolar to bipolar sensing, the
programmer verifies the presence of a functioning bipolar lead by
testing impedance for the lead.
operation for four seconds at an Amplitude of 5.0 V and a Pulse
Width of 1.0 ms if the permanent settings are at or below this level.
If permanent settings are above these values, the measurement
will be made at the permanent settings.
1
Testing is done under magnet
InSync Model 8040 Device Reference Guide
84
Chapter 4
Selecting Sensing Parameters
■
If bipolar lead impedance is between 200 ohms and
3000 ohms), a bipolar lead is assumed to be present.
■
If bipolar lead impedance is outside this range, a unipolar lead
is assumed to be present.
test failed, and sensing polarity remains set to unipolar. This
interlock feature may be overridden and lead sensing polarity
forced to bipolar.
Warning: Do not override for an implanted lead. The override
feature is intended only as a means to program the device for
bipolar polarity before lead connection.
Determining Sensing Threshold(s) at Implant
At implant, use Medtronic Pacing System Analyzer (PSA) to
determine sensing threshold values for the device. Refer to the
PSA manual for detailed instructions.
Before connecting a lead, measure the sensing potentials in the
unipolar and the bipolar configurations. Adequate intracardiac
signal should be present in both configurations to ensure proper
sensing in either.
1
1
The programmer warns that the
Verifying Sensing Threshold(s) at Follow-up
Intracardiac signal amplitudes decrease during the lead
maturation process. Medtronic programmers provide an automatic
sensitivity test that allows the follow-up clinician to verify a
patient’s sensitivity settings. The automatic test provides for atrial
or ventricular monitoring. The test provides the sensitivity setting
just above and below the point at which P waves or R waves are
sensed.
The cardiac signal presented to the device by the ventricular
two-lead system is a composite signal from the parallel
combination of both ventricular leads. The Sensing test treats this
signal as a single input with a measurable amplitude that can be
used to determine an appropriate setting for ventricular sensitivity.
Conducting the Sensing test for the ventricular two-lead system
does not require any special considerations.
1
Impedance for the ventricular two-lead system is measured across the parallel
combination of both leads.
InSync Model 8040 Device Reference Guide
Selecting Sensitivity Settings
Atrial and ventricular sensitivity are independently programmable.
In general, a 2:1 to 3:1 sensitivity safety margin (threshold
sensitivity value divided by 2 or 3) is adequate for newly implanted
or chronic leads. For example, an atrial sensitivity of 1.0 mV
should be satisfactory for intrinsic atrial signals between 2.0 mV
and 3.0 mV.
■
Always perform an atrial sensing test to determine the
appropriate atrial sensitivity setting.
■
Excessively sensitive (low) settings can cause some or all of
the following problems:
– oversensing due to electromagnetic interference (EMI),
myopotentials, T waves, or crosstalk
– undersensing due to overloading of the sensing circuit
– noise reversion operation
■
Atrial sensitivity with bipolar sensing polarity allows 0.18 mV,
0.25 mV, and 0.35 mV atrial sensitivity settings. To prevent
oversensing of muscle noise or electromagnetic interference,
unipolar sensing is limited to values no less than 0.5 mV.
■
Ventricular sensitivities 1.0 mV or 1.4 mV with wide atrial
pulse widths or high atrial amplitudes may result in Ventricular
Safety Pacing (if On) with some lead systems at high
sensor-driven pacing rates. Reprogramming Ventricular
Sensitivity to a less sensitive setting (higher numerical value)
is one option under such circumstances. Other options include
programming a longer Ventricular Blanking Period.
Lead/cardiac tissue interface
Selecting Sensing Parameters
85
For Further Information
In the PSA manual, refer to sensitivity threshold test procedures.
Refer to “Noise Reversion” on page 72 for a description of noise
reversion operation. Refer to “Programmable Modes and
Parameters” on page 182 for sensitivity parameter settings.
InSync Model 8040 Device Reference Guide
86
Chapter 4
Monitoring Lead Stability
Monitoring Lead Stability
Overview
The Lead Monitor feature periodically measures the pacing
impedance of each lead. It is intended for troubleshooting when an
intermittent lead problem is suspected. The Lead Monitor also has
an optional Polarity Switch feature that can automatically
reprogram polarity from bipolar to unipolar for each lead. It is
intended to support backup unipolar pacing and sensing in the
event of a bipolar lead failure.
The Lead Monitor is available in all pacing modes. It is not
recommended to use Lead Monitor with single chamber triggered
modes (VVT and AAT).
Caution: If the Lead Monitor detects out-of-range lead
impedance, investigate lead integrity more thoroughly.
Warnings:
■
■
Programming Lead Monitor with the optional polarity switch is
contraindicated for patients with implantable defibrillators.
When an out-of-range lead is detected, the monitor
automatically reprograms the selected leads to unipolar
polarity. Pacing in the unipolar configuration may either
provoke inappropriate therapy or withhold appropriate
therapy.
Do not program Lead Monitor on prior to implanting the device
in the patient. With no leads connected, lead impedance is
infinitely high and determined by the device to be out-of-range.
If the device has been programmed to switch polarity, the
resulting unipolar condition will not support pacing until the
device is placed in the pocket, thereby completing the circuit.
Note: The Lead Monitor is automatically disabled if the elective
replacement indicator (ERI) is tripped.
How the Monitor Works
Impedance for each active lead is measured every three hours.
The measurement requires up to three pacing cycles during which
Amplitude and Pulse Width change to 5.0 V and 1.0 ms,
respectively, if the permanent settings are at or below this level. If
permanent settings are above these values, the measurement will
be made at the permanent settings.
InSync Model 8040 Device Reference Guide
■
Sustained intrinsic activity may prevent measurement from
occurring.
■
If a lead impedance measurement falls outside the range of
200 to 3000 ohms, the Lead Trend diagnostic (described
below) is automatically activated.
Note: Impedance for the ventricular two-lead system is measured
across the parallel combination of both leads.
How the Polarity Switch Works
The polarity of the selected bipolar lead is automatically
reprogrammed by the device from bipolar to unipolar if the lead
impedance measurement is outside the range of 200 to 3000
ohms. Programmable options for Polarity Switching are:
■
atrium and ventricle (the atrial and ventricular polarity switch
to unipolar)
■
atrium (only the atrial polarity switches to unipolar)
■
ventricle (only the ventricular polarity switches to unipolar)
Recording a Lead Trend
Lead/cardiac tissue interface
Monitoring Lead Stability
87
The Lead Trend attempts to measure and log the impedance of
each active lead every three days. Lead impedance data
collection is the same as for the Lead Monitor.
■
The device can collect up to 720 days of lead impedance data
for single chamber configurations and 360 days of data for
dual chamber configurations.
■
Lead Trend is automatically activated when the Lead Monitor
detects an out-of-range lead.
Figure 4-2 shows how the Lead Trend report is presented on the
programmer. The report can be printed out or displayed on the
programmer.
InSync Model 8040 Device Reference Guide
88
Chapter 4
Monitoring Lead Stability
Overriding Other Diagnostics
Figure 4-2. Example of Lead Trend Diagnostic
If the Lead Trend is activated because the Lead Monitor detects
out-of-range lead impedance, any programmed diagnostic is
disabled, and its data are overwritten by Lead Trend data.
For Further Information
Refer to the InSync Model 8040 Device Programming Guide for
information on configuring the Lead Trend diagnostic and
collecting its data.
InSync Model 8040 Device Reference Guide
Lead/cardiac tissue interface
DDD Operation
Magnet Applied
Threshold Margin Test
Rate = 100 ppm
Asynchronous (DOO)
Pacing Rate = 85 ppm
Parameters:
Mode = DDDPAV = 180 ms
Lower Rate = 60 ppm
Pulse Widths Reduced
by 25%
PAV = 100 ms
Transtelephonic Capture Verification with TMT
Transtelephonic Capture Verification with TMT
Overview
During transtelephonic monitoring, the Threshold Margin Test
(TMT) provides a check for loss of capture at 25% reduced pulse
width(s). The TMT may indicate that loss of capture is possible but
cannot verify that safety margin is adequate.
Warning: Loss of capture during TMT indicates that the pacing
safety margin is inadequate. Have the patient come to the clinic as
soon as possible for threshold evaluation and reprogramming of
outputs for a 2:1 voltage safety margin. Imminent loss of
ventricular capture for a pacemaker-dependent patient may
constitute an emergency situation.
Dual Chamber TMT Operation
Applying the magnet over the device initiates a Threshold Margin
Test (TMT). As shown in Figure 4-3, the device delivers three
asynchronous AV sequential pulses at a rate of 100 ppm with a
paced AV interval of 100 ms. The first two sequences of pulses are
delivered at the programmed Pulse Widths. The third sequence is
delivered at a 25% reduction of the programmed Pulse Widths.
89
Figure 4-3. Dual Chamber TMT Operation
InSync Model 8040 Device Reference Guide
90
VVIR Operation
Magnet Applied
Threshold Margin Test
Rate = 100 ppm
Parameters:
Mode = VVIR
Pacing Rate = 60 ppm
Pulse Widths Reduced
by 25%
Chapter 4
Transtelephonic Capture Verification with TMT
Single Chamber and VDD Mode TMT Operation
Applying the magnet over the device initiates TMT. As illustrated
below for ventricular operation, the device delivers three
asynchronous pulses at a rate of 100 ppm. The first two pulses are
delivered at the programmed Pulse Width. The third pulse is
delivered at a 25% reduction of the programmed Pulse Width.
Figure 4-4. Single Chamber and VDD Mode TMT Operation
Enhanced Transtelephonic Monitoring
The Transtelephonic Monitor is a programmable On or Off feature
intended for use with remote device monitoring services.
Programming the feature Off does not affect conventional
transtelephonic monitoring.
When the Transtelephonic Monitor is programmed On, the device
delays the Threshold Margin Test for five seconds upon magnet
application to enhance communication with transtelephonic
equipment. If the pacing polarity is permanently programmed to
bipolar, it is temporarily set to unipolar to provide improved ECG
artifact detection for the remote monitoring equipment. The
programmed polarity is restored when the magnet is removed.
When Transtelephonic Monitor is programmed Off, the Threshold
Margin Test is not delayed upon magnet application, and
conventional transtelephonic monitoring can occur.
InSync Model 8040 Device Reference Guide
Warning: Programming Transtelephonic Monitor On is
contraindicated for patients with implantable defibrillators. When it
is programmed On, the pacing polarity is temporarily set to
unipolar when the magnet is applied. Pacing in the unipolar
configuration may cause the defibrillator to provoke inappropriate
therapy or withhold appropriate therapy.
For Further Information
Refer to the InSync Model 8040 Device Programming Guide for
programming instructions regarding Transtelephonic Monitor.
Lead/cardiac tissue interface
Transtelephonic Capture Verification with TMT
91
InSync Model 8040 Device Reference Guide
Special therapy options
Mode Switch and Diagnostic94
Non-Competitive Atrial Pacing99
PMT Intervention102
PVC Response104
Ventricular Safety Pacing107
Rate Drop Response108
Sleep Function113
Single Chamber Hysteresis115
5
5
94
Chapter 5
Mode Switch and Diagnostic
Mode Switch and Diagnostic
Overview
Mode Switch is a programmable On or Off feature designed to
prevent the tracking of paroxysmal atrial tachycardias in the
DDDR, DDD, and VDD modes. It has a programmable Detect
Rate for when to mode switch. Mode Switch is not recommended
for patients known to have chronic refractory atrial
tachyarrhythmias: atrial tachycardia, atrial fibrillation, or atrial
flutter.
When the device detects an atrial tachyarrhythmia, it switches
from the programmed atrial tracking mode to a non-atrial tracking
mode until the atrial tachycardia ceases, see Figure 5-1.
Mode switch should be used only in heart failure patients with a
history of atrial tachycardias as it affects other device parameters
associated with delivery of cardiac resynchronization. See “Device
Programming Recommendations” on page 178.
Atrial Tracking ModeNon-Atrial Tracking Mode
DDDRDDIR
DDDDDIR
VDDVVIR
Figure 5-1. Mode Switching Modes
How the Device Defines Atrial Tachycardia
The device defines an atrial tachycardia based on the mean atrial
rate and the programmable Detect Rate:
Mean atrial rate (MAR) – A running average calculated from all
A-A intervals (except AS-AP and AR-AP). To respond quickly to
the onset of atrial tachycardia, the MAR increases rapidly by giving
preference to shorter A-A intervals. For longer A-A intervals, the
MAR decreases gradually to ensure that atrial tachycardia has
stopped.
InSync Model 8040 Device Reference Guide
Detect Rate – The rate at which device-defined atrial tachycardia
Ensuing atrial
tachycardia, interval
is about 190 ms
Mode switch from
DDDR to DDIR
Ventricular interval
at 500 ms
Ventricular
interval increases
by 40 ms each VP
starts. Note that ventricular tracking is limited by the Upper
Tracking Rate or the total atrial refractory period even when the
atrial rate rises above the Detect Rate.
When the MAR exceeds the Detect Rate, atrial tachycardia is
considered in progress. When either the MAR drops below the
Upper Tracking Rate or five consecutive atrial paces occur, the
atrial tachycardia is considered to have ceased.
Mode Switch Operation
When the device detects an atrial tachycardia in the DDDR, DDD,
or VDD mode, it switches to the appropriate non-atrial tracking
mode in approximately 15 seconds. To avoid an abrupt drop in the
ventricular rate, it smoothly reduces the pacing rate from the atrial
synchronous rate to the sensor-indicated rate in approximately
2.5 minutes, see Figure 5-2.
After the rate transition is completed, the device continues
sensor-driven pacing in the ventricle, operating in the non-atrial
tracking mode until the atrial tachycardia ceases.
Special therapy options
Mode Switch and Diagnostic
95
Figure 5-2. Mode Switching to Non-Atrial Tracking Mode
When the device determines that the atrial tachycardia has
ceased, it begins to switch back to the programmed atrial tracking
mode.
InSync Model 8040 Device Reference Guide
96
Chapter 5
Mode Switch and Diagnostic
Mode Switching Interruption
DDDR and DDD Modes – Abrupt changes in ventricular rate are
avoided by smoothly varying the pacing rate until the A-A interval
corresponds to the mean atrial rate. At that point, the device
switches to the programmed mode.
VDD Mode – The device immediately switches back to the
programmed mode.
The typical mode switching sequence may be interrupted by either
of these two occurrences:
■
the atrial tachycardia episode ceases before the device
completes the rate transition to the appropriate non-atrial
tracking mode or
■
the atrial tachycardia episode ceases briefly but resumes
before the atrial tracking mode is restored.
In either case, the device responds by adjusting the rate transition
in the appropriate direction. The criteria for switching to the atrial
tracking mode are unaffected.
Preset Parameter Settings
When Mode Switch is programmed On, several parameters are
automatically set to the following values:
■
Upper Activity Rate is set equal to the Upper Tracking Rate.
■
Post-Ventricular Atrial Blanking and Rate Adaptive AV are set
to values based on the programmed Upper Tracking Rate.
■
Post-Ventricular Atrial Refractory Period is set to a value
based on the programmed Detection Rate.
■
Rate Drop Response is set to Off.
The values selected are necessary for proper mode switching
operations.
InSync Model 8040 Device Reference Guide
Recording Mode Switch Episodes
Mode switch episodes are always counted, and the total is
presented in the Event Summary report. (See “Event Summary”
on page 138). Additional details for up to 14 mode switch episodes
can be recorded by the programmable Mode Switch Episode
diagnostic. Two recording methods are available:
■
The Frozen method collects detailed data for the first
14 episodes detected.
■
The Rolling method collects detailed data for the most recent
14 episodes, overwriting older data if necessary.
For each of the 14 mode switch episodes, the device records the
following data:
■
date and time of occurrence
■
episode duration (in minutes)
■
approximate atrial rate (Tachy Rate)
■
rates of 24 events associated with the mode switch episode,
as follows:
– the last 8 events preceding the mode switch,
– the first 8 events of mode switch operation, and
– the first 8 events after return to permanent mode.
Special therapy options
Mode Switch and Diagnostic
97
Figure 5-3 shows how mode switch episodes are summarized on
the programmer. Figure 5-4 shows how the programmer presents
detailed data for a selected episode. All reports can be printed out
or displayed on the programmer.
InSync Model 8040 Device Reference Guide
98
Chapter 5
Mode Switch and Diagnostic
Figure 5-3. Example of Mode Switch Episodes Diagnostic
Figure 5-4. Example of Detailed Data for a Selected Mode Switch
Episode (Diagnostic)
InSync Model 8040 Device Reference Guide
Erasing Mode Switch Episode Data
Certain programmer functions erase stored Mode Switch Episode
data. Review or print episode data before performing any of the
following functions:
■
programming another diagnostic,
■
using the programmer “Clear Data” command, or
■
initiating an activity threshold test or exercise test.
Note: The Mode Switch Episode diagnostic is disabled and all
episode data erased if the Lead Monitor enables the Lead Trend
diagnostic. See “Monitoring Lead Stability” in Chapter 4 for more
information on this feature. Episode data collection is suspended
if the elective replacement indicator is set, but data already
collected are retained.
For Further Information
Refer to the InSync Model 8040 Device Programming Guide for
information on configuring the Mode Switch Episode diagnostic
and collecting its data.
Special therapy options
Non-Competitive Atrial Pacing
99
Non-Competitive Atrial Pacing
Overview
Non-Competitive Atrial Pacing (NCAP) is intended to prevent
triggering of atrial tachycardias by an atrial pacing stimulus that
falls within the atrium’s relative refractory period. This feature may
be programmed On or Off in the DDDR mode only.
Note: NCAP should be programmed On for heart failure patients.
How NCAP Affects Atrial Timing
When NCAP is programmed On, a refractory sensed atrial event
falling in the PVARP starts a 300 ms NCAP period, during which
no atrial pacing may occur:
■
If a sensor-driven or lower rate pacing stimulus is scheduled
to occur during the NCAP period, the VA interval is extended
until the NCAP period expires.
InSync Model 8040 Device Reference Guide
100
Pace Atrium safely,
no capture
Relative refractory
period, atrial pace
may induce atrial
tachycardia
Pace Atrium safely,
capture
Chapter 5
Non-Competitive Atrial Pacing
■
If no pacing stimulus is scheduled to occur during the NCAP
period, timing is unaffected; pacing occurs at the end of the VA
interval unless inhibited.
■
An atrial refractory sensed event occurring during the NCAP
period starts a new NCAP period.
Figure 5-5. The Atrium’s Relative Refractory Period
How NCAP Affects Ventricular Timing
When an atrial pacing stimulus is delayed by the NCAP operation,
the device attempts to maintain a stable ventricular rate by
shortening the PAV interval that follows. It will not, however,
shorten the PAV interval to less than 80 ms. When a relatively high
Lower Rate and long PVARP are programmed, NCAP operation
may result in ventricular pacing slightly below the Lower Rate.
InSync Model 8040 Device Reference Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.