Medtronic 7304 Reference Guide

INSYNC MAXIMO
Dual chamber implantable cardioverter defibrillator with cardiac resynchronization therapy including sequential biventricular pacing
7304
Reference Manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
INSYNC MAXIMO
7304
Reference Manual
A guide to the operation and programming of the Model 7304 InSync Maximo dual chamber implantable cardioverter defibrillator with cardiac resynchronization therapy
The following are trademarks of Medtronic:
Active Can, Cardiac Compass, Checklist, Decision Channel, Flashback, InSync, InSync II Marquis, InSync Sentry, InSync Maximo, Leadless, Marker Channel, Marquis, Maximo, Medtronic, Medtronic CareLink, PR Logic, Patient Alert, Quick Look, QuickLink, T-Shock
Contents
Introduction 11
Part I Quick overview 13
1 Quick reference 15
1.1 Physical characteristics 16
1.2 Magnet application 18
1.3 Projected longevity 18
1.4 Replacement indicators 20
1.5 Typical charge times 21
1.6 High-voltage therapy energy 21
1.7 Stored data and diagnostics 22
1.8 New and enhanced features 24
2 The InSync Maximo system 25
2.1 System overview 26
2.2 Indications and usage 29
2.3 Contraindications 30
2.4 Patient screening 30
3 Emergency therapy 31
3.1 Delivering emergency therapies 32
3.2 Delivering an emergency defibrillation therapy 33
3.3 Delivering an emergency cardioversion therapy 34
3.4 Delivering emergency fixed burst pacing 34
3.5 Enabling emergency VVI pacing 35
Part II Device implant and patient follow-up
procedures 37
4 Implanting the device 39
4.1 Overview 40
4.2 Preparing for an implant 40
4.3 Replacing a device 42
4.4 Surgical approach 43
INSYNC MAXIMO™7304 Reference Manual
6
Contents
4.5 Sensing and pacing measurements 47
4.6 Connecting the leads to the device 47
4.7 Testing defibrillation operation and effectiveness 51
4.8 Positioning and securing the device 54
4.9 Completing the implant procedure 54
5 Conducting a patient follow-up session 57
5.1 Patient follow-up guidelines 58
5.2 Verifying the status of the implanted system 59
5.3 Verifying effective cardiac resynchronization therapy 59
5.4 Verifying effective basic pacing 60
5.5 Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias 62
Part III Configuring the device for the patient 65
6 Detecting tachyarrhythmias 67
6.1 Detection overview 68
6.2 Setting up sensing 71
6.3 Detecting VF episodes 76
6.4 Detecting VT episodes 79
6.5 Detecting FVT episodes 85
6.6 Detecting tachyarrhythmia episodes with Combined Count 90
6.7 Monitoring episodes for termination or redetection 92
6.8 Enhancing detection with PR Logic criteria 96
6.9 Enhancing VT detection with the Stability criterion 104
6.10 Detecting double tachycardias 106
6.11 Detecting prolonged tachyarrhythmias with High Rate Timeout 107
7 Treating tachyarrhythmia episodes 111
7.1 Treating VF with defibrillation 112
7.2 Treating VT and FVT with antitachycardia pacing 121
7.3 Treating VT and FVT with cardioversion 131
INSYNC MAXIMO™7304 Reference Manual
Contents
7.4 Optimizing therapy with Smart Mode and Progressive Episode Therapies 139
8 Treating ventricular dysynchrony and
bradycardia 145
8.1 Providing basic pacing therapy 146
8.2 Dual chamber pacing 154
8.3 Single chamber pacing 165
8.4 Promoting continuous CRT delivery 169
8.5 Rate adjustments to optimize cardiac output 179
8.6 Managing atrial tracking to optimize A-V synchrony 191
8.7 Providing Ventricular Safety Pacing 204
9 Optimizing charge time and device longevity 209
9.1 Optimizing charge time 210
9.2 Optimizing device longevity 212
Part IV Evaluating and managing patient treatment 215
10 Using the programmer 217
10.1 Setting up and using the programmer 218
10.2 Display screen features 219
10.3 Viewing and programming device parameters 224
10.4 Starting and ending patient sessions 229
10.5 Viewing live waveform traces 231
10.6 Recording live waveform strips 238
10.7 Saving and retrieving device data 240
10.8 Printing reports 244
11 Using system evaluation tools 251
11.1 A summary of system evaluation tools 252
11.2 Taking a quick look at device activity 253
11.3 Using the Patient Alert feature 254
11.4 Streamlining follow-ups with Checklist 262
12 Setting up and viewing collected data 267
12.1 A summary of data collection 268
12.2 Viewing battery and lead status data 269
12.3 Printing rate histograms 274
7
INSYNC MAXIMO™7304 Reference Manual
8
Contents
12.4 Viewing the Heart Failure Management report 278
12.5 Viewing clinical trends in the Cardiac Compass report 282
12.6 Viewing the episode and therapy efficacy counters 288
12.7 Viewing episode data 291
12.8 Viewing Flashback Memory 301
12.9 Setting up data collection 302
12.10 Viewing and entering patient information 308
12.11 Automatic device status monitoring 311
13 Testing the system 315
13.1 Testing overview 316
13.2 Evaluating the underlying rhythm 316
13.3 Measuring pacing thresholds 317
13.4 Measuring lead impedance 320
13.5 Measuring EGM amplitude 322
13.6 Testing the device capacitors 324
14 Conducting Electrophysiologic Studies 327
14.1 EP Study overview 328
14.2 Inducing VF with T-Shock 329
14.3 Inducing VF with 50 Hz Burst 333
14.4 Inducing an arrhythmia with Manual Burst 335
14.5 Inducing an arrhythmia with PES 338
14.6 Delivering a manual therapy 340
15 Solving system problems 345
15.1 Overview 346
15.2 Solving cardiac resynchronization therapy problems 346
15.3 Solving sensing problems 347
15.4 Solving tachyarrhythmia detection problems 349
15.5 Solving tachyarrhythmia therapy problems 350
15.6 Solving bradycardia pacing problems 351
15.7 Responding to device status indicators 352
INSYNC MAXIMO™7304 Reference Manual
Appendices 355
A Warnings and precautions 357
A.1 General warnings 358
A.2 Handling and storage instructions 358
A.3 Resterilization 359
A.4 Device operation 360
A.5 Lead evaluation and lead connection 361
A.6 Follow-up testing 363
A.7 Explant and disposal 363
A.8 Medical therapy hazards 364
A.9 Home and occupational environments 366
B Device parameters 369
B.1 Emergency settings 370
B.2 Detection parameters 371
B.3 Therapy parameters 372
B.4 Pacing parameters 375
B.5 System maintenance parameters 378
B.6 Data collection parameters 380
B.7 System test and EP study parameters 381
B.8 Fixed parameters 384
B.9 Patient information parameters 385
B.10 Programmer symbols 386
B.11 Parameter interlocks 388
C Device programming recommendations 389
C.1 Device programming recommendations 390
C.2 V-V Delay programming recommendations 392
Contents
9
Glossary 397
Index 407
INSYNC MAXIMO™7304 Reference Manual
Introduction
Using this manual
Before implanting the device, it is strongly recommended that you:
Manual conventions
Throughout this document, the word “device” refers to the implanted InSync Maximo device.
Introduction
Refer to the product literature packaged with the device for
information about prescribing the device.
Thoroughly read this manual and the technical manuals for
the leads and the implant tools used with the device.
Discuss the procedure and the device system with the
patient and any other interested parties, and provide them with any patient information packaged with the device.
11
Technical support
The nominal value for that parameter.
On-screen buttons are shown with the name of the button surrounded by brackets: [Button Name].
Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
symbol in parameter tables indicates the Medtronic
INSYNC MAXIMO™7304 Reference Manual
12
Introduction
References
Notice
The primary reference for background information is Zacouto FI, Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B,
Cardiac Pacing Diagnostic and Therapeutic Tools.
ed. Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
Estes M, Manolis AS, Wang P, Eds.
Cardioverter-Defibrillators
Inc. 1994.
Kroll MW, Lehmann MH, Eds.
Cardioverter-Defibrillator Therapy: The Engineering-Clinical Interface
Singer I, Ed.
NY: Futura Publishing Co. 1994.
Singer I, Barold SS, Camm AJ, Eds.
Therapy of Arrhythmias for the 21st Century: The State of the Art
. Norwell, MA: Kluwer Academic Publishers 1996.
Implantable Cardioverter-Defibrillator
. Armonk, NY: Futura Publishing Co. 1998.
. New York, NY: Marcel Dekker,
Implantable
Implantable
Nonpharmacological
New York:
. Armonk,
This software is provided as an informational tool for the end user. The user is responsible for accurate input of patient information into the software. Medtronic makes no representation as to the accuracy or completeness of the data input into the software. Medtronic SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE INFORMATION PROVIDED IN THE SOFTWARE.
INSYNC MAXIMO™7304 Reference Manual
Quick overview
Part I
1Quick reference
1.1 Physical characteristics 16
1.2 Magnet application 18
1.3 Projected longevity 18
1.4 Replacement indicators 20
1.5 Typical charge times 21
1.6 High-voltage therapy energy 21
1.7 Stored data and diagnostics 22
1.8 New and enhanced features 24
1
16
Chapter 1
Physical characteristics
1.1 Physical characteristics
Table 1-1. Device physical characteristics
Volume
Mass
H x W x D
c
Surface area of device
40 cm
78 g
73 mm x 51 mm x 15 mm
67 cm
a
3 b
2
can
Radiopaque ID
Materials in contact with human tissue
d
e
PRL
Titanium / polyurethane /silicone rubber / silicone adhesive
Battery Lithium silver vanadium oxide
Connectors Two IS-1 connectors for pacing and
sensing, one IS-1 connector for pacing, two DF-1 connectors for high-voltage therapy, Active Can electrode (programmable)
a
Measurements are nominal values based on CAD (computer aided design) model measurements and are rounded to the nearest unit.
b
Volume with connector holes unplugged.
c
Grommets may protrude slightly beyond the can surface.
d
Engineering series number follows the radiopaque code.
e
These materials have been successfully tested and verified for biocompatibility. The device does not produce an injurious temperature in the surrounding tissue.
Table 1-2. Lead connections
Device port Connector type Software name
SVC (HVX) DF-1 SVC
RV (HVB) DF-1 RVcoil
Can
RV IS-1 bipolar RVtip and RVring
INSYNC MAXIMO™7304 Reference Manual
Can
1Quick reference
2
5
4
6
3
1
Physical characteristics
Table 1-2. Lead connections (continued)
Device port Connector type Software name
LV IS-1 bipolar LVtip and LVring
A IS-1 bipolar Atip and Aring
Figure 1-1. Lead connections
1 SVC (HVX) 2 RV (HVB) 3 Can (HVA) 4 LV 5 RV 6 A
17
Figure 1-2. Suture holes
INSYNC MAXIMO™7304 Reference Manual
18
Chapter 1
Magnet application
1.2 Magnet application
Bringing a magnet close to the device triggers changes in device operation as shown in Table 1-3. When the magnet is removed, the device returns to its programmed operations.
For information on demonstrating Patient Alert tones to the patient, see Section 11.3.4, “Instructing the patient”, page 260.
Table 1-3. Effects of magnet application on the device
Pacing mode as programmed
Pacing rate and interval as programmed
VF, VT, and FVT detection suspended
Patient Alert audible tones (20 s or less)
a
b
c
a
b
high/low dual tone (high urgency
alert occurred) on/off intermittent tone (lower
urgency alert occurred) continuous test tone (no alert
occurred) no tone (alerts are disabled)
Rate response adjustments are suspended while a Patient Alert tone sounds.
Detection resumes if telemetry is established and the application software is running, or detection resumes after the application software has started.
Or “VF Detection/Therapy Off” is the only alert enabled.
c
1.3 Projected longevity
Longevity estimates are based on accelerated battery discharge data and device modeling with EGM pre-storage off (see Table 1-4).
INSYNC MAXIMO™7304 Reference Manual
1Quick reference
Projected longevity
Table 1-4. Projected longevity in years with pacing outputs programmed to the specified amplitude and 0.4 ms pulse width, DDD mode, 100% biventricular pacing, specified percentage of atrial pacing at 60 ppm with the remainder at 70 bpm atrial tracking, semi-annual full-energy charges.
Amplitudes: A/RV; LV
2.5 V; 3.0 V
3.0 V; 4.0 V
a
a
% atrial pacing
0% 5.8 6.2 6.5
25% 5.6 6.1 6.4
50%
100% 5.2 5.8 6.1
0% 5.0
25% 4.8 5.3
50% 4.6 5.2 5.6
100% 4.4 5.0 5.4
Semi-annual full-energy charges may include therapy shocks or capacitor formations. Additional full-energy charges due to therapy shocks, device testing, or capacitor formation reduces device longevity by approximately 23 days (0.06 years). Additional 30 J charges due to therapy shocks or device testing reduces device longevity by approximately 21 days (0.06 years).
500 Ù pacing impedance
5.5
700 Ù pacing impedance
6.0 6.3
5.5
900 Ù pacing impedance
5.8
5.7
19
The longevity of the ICD is dependent on several factors. The following factors result in decreased longevity:
An increase in pacing rate, pacing amplitude or pulse
width; the ratio of paced to sensed events; or the charging frequency
A decrease in pacing impedance
Using the pre-onset EGM storage feature or Holter telemetry
Considerations for using EGM pre-storage – When the EGM pre-storage feature is programmed Off, the device starts to store EGM following the third tachyarrhythmia event and also provides up to 20 s of information before the onset of the tachyarrhythmia, including:
AA and VV intervals
Marker Channel
Interval Plot Flashback
INSYNC MAXIMO™7304 Reference Manual
20
Chapter 1
Replacement indicators
When the EGM pre-storage feature is programmed On, the device also collects up to 20 s of EGM information before the onset of the arrhythmia.
In a patient who uniformly repeats the same onset mechanisms, the greatest clinical benefit of pre-onset EGM storage is achieved after a few episodes are captured. To maximize the effectiveness of the EGM pre-storage feature and optimize device longevity, consider these programming options:
Turn pre-storage On to capture possible changes in the onset mechanism following significant clinical adjustments, for example, device implant, medication changes, and surgical procedures.
Turn pre-storage Off once you have successfully captured the information of interest.
1.4 Replacement indicators
Battery voltage and messages about replacement status appear on the programmer display and on printed reports. Table 1-5 lists the Elective Replacement Indicator (ERI) and the End of Life (EOL) conditions.
Table 1-5. Replacement indicators
Elective Replacement (ERI) 2.62 V
End of Life (EOL) 3 months after ERI
ERI date – The Quick Look and Battery and Lead Measurements screens display the date when the battery reached ERI.
Temporary voltage decrease – The battery voltage temporarily decreases following a high-voltage charge. If a battery measurement is taken immediately after a high-voltage charge, the ERI or EOL indicator may be displayed. However, this is a temporary status that will return to normal when the battery has recovered from the charge.
EOL indication – If the programmer indicates that the device is at EOL, replace the device immediately.
INSYNC MAXIMO™7304 Reference Manual
Post-ERI conditions – EOL device status is defined as 3 months following an ERI indication assuming the following post-ERI conditions: 0% DDD atrial pacing, 100% DDD RV and LV pacing at 60 ppm, 3 V atrial and RV pacing amplitude, 4 V LV pacing amplitude, 0.4 ms pulse width; 500 Ù pacing load; and six 35 J charges. EOL may be indicated before the end of 3 months if the device exceeds these conditions.
1.5 Typical charge times
The most recent capacitor charge time appears on the programmer display and on printed reports and can be evaluated using the Charge/Dump test (see Table 1-6).
Table 1-6. Typicalafull-energy charge times
At Beginning of Life (BOL)
At Elective Replacement (ERI)
a
These charge times are typical when the capacitors are fully formed.
7.1 s
9.0 s
1Quick reference
Typical charge times
21
1.6 High-voltage therapy energy
The stored energy of the device is derived from the peak capacitor voltage and is always greater than the energy delivered by the device. Table 1-7 compares the programmed energy levels delivered by the device to the energy levels stored in the capacitors before delivery.
Table 1-7. Comparison of delivered (programmed) and stored energy levels
Energy ( J)
Delivereda/ Programmed Stored
35 39 7.1
32 37 6.5
30 34 6.1
28 32
26 30 5.2
25 29 5.0
24 27 4.8
22 25 4.4
b
Charge Timec( s)
5.7
Delivereda/ Programmed Stored
Energy ( J)
10 12 2.0
9 10.5 1.8
8 9.3 1.6
7
6 7.1 1.2
5
4 4.8 0.8
3 3.6 0.6
INSYNC MAXIMO™7304 Reference Manual
8.2 1.4
5.9 1.0
b
Charge Timec( s)
22
Chapter 1
Stored data and diagnostics
Table 1-7. Comparison of delivered (programmed) and stored energy levels (continued)
Energy ( J)
Delivereda/ Programmed Stored
20 23 4.0
18 21 3.6
16 19 3.2
15 17 3.0
14 16 2.8
13 15 2.6
12 14 2.4
11 13 2.2
b
Charge Timec( s)
Delivereda/ Programmed Stored
Energy ( J)
Charge
b
Timec( s)
2 2.4 0.4
1.8 2.2 0.4
1.6 2.0 0.3
1.4 1.7 0.3
1.2 1.5 0.2
1.0 1.2 0.2
0.8 1.0 0.2
0.6 0.8 0.1
0.4 0.5 0.1
a
Energy delivered at connector block into a 75 Ù load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth
of a second.
1.7 Stored data and diagnostics
Table 1-8. Stored data and diagnostics
Episode data
Tachy episodes 150 VF/VT/FVT episodes: intervals, text, EGM
EGM capacity for tachy episodes 14 min of dual-channel EGM or 23.5 min of single-channel
SVT/NST episodes 50 SVT/NST episodes: intervals, text, EGM (the device
EGM capacity for SVT/NST episodes
EGM sources 14 options: atrial/ RV / LV / far-field
EGM options Store before onset; Store during charging
Flashback memory 2000 intervals (containing both A-A and V-V): before latest
Mode switch episodes 53 mode switch episodes (fastest, longest, first and 50
EGM
does not usually store detailed episode records for NST episodes)
2 min of dual-channel EGM or 3.6 min of single-channel EGM
VF, before latest VT, and before interrogation
latest): episode text
INSYNC MAXIMO™7304 Reference Manual
Table 1-8. Stored data and diagnostics (continued)
1Quick reference
Stored data and diagnostics
23
Ventricular sensing episodes 9 ventricular sensing episodes (longest, first, and 7 latest):
Counter data
Detection counters Lifetime total, since cleared, and since last session
Episode counters Episodes:
Therapy efficacy counters Counts for each VF, FVT, VT Therapy:
Other stored data
Patient Alert — System events Up to 10 log entries: text and date for the first time an alert
Battery and lead measurements Battery voltage, last capacitor formation, last charge,
Lead performance trends 14 days of daily measurements plus 80 weeks of weekly
intervals, markers, and counter data text
VF, FVT, and VT
Atrial Fibrillation/Atrial Flutter episodes
Sinus Tach episodes
Other 1:1 SVT episodes
NST episodes
Mode switch episodes
Percentage pacing:
AS-VS, AS-VP, AP-VS, AP-VP percentages
Additional counters:
Single PVCs and PVC runs
Rate stabilization pulses and runs
Delivered
Successful
Unsuccessful
Intervention (manually aborted)
Total number of aborted shocks
is triggered between interrogations
lead impedance, EGM amplitude measurements, last high-voltage therapy, and sensing integrity counter
minimum and maximum measurements:
Lead impedance: atrial pacing, RV pacing, LV pacing,
defibrillation pathway, and SVC lead (if used) atrial EGM amplitude: (P-waves)
INSYNC MAXIMO™7304 Reference Manual
24
Chapter 1
New and enhanced features
Table 1-8. Stored data and diagnostics (continued)
Rate Histograms
Cardiac Compass trends 14 months of daily measurements:
Atrial rate
Ventricular rate
Ventricular rate during AT/AF
VT and VF episodes per day
one or more high-voltage therapy delivered
ventricular rate during VT or VF
episodes of non-sustained tachycardia per day
heart rate variability
total daily time in AT/AF
ventricular rate during AT/AF
percent pacing per day
patient activity
average day and night ventricular heart rate
1.8 New and enhanced features
The following features are new or changed from the 7289 InSync II Marquis ICD.
1.8.1 Tachyarrhythmia therapy
Output – The InSync Maximo has a maximum delivered energy
of 35 J.
1.8.2 Cardiac resynchronization therapy
Sequential biventricular pacing – The ventricular sequence
and V-V pace delay are programmable to support improved hemodynamics.
Pacing outputs for RV and LV are independently programmable.
INSYNC MAXIMO™7304 Reference Manual
2The InSync Maximo system
2.1 System overview 26
2.2 Indications and usage 29
2.3 Contraindications 30
2.4 Patient screening 30
2
26
Chapter 2
System overview
2.1 System overview
The Model 7304 InSync Maximo dual chamber implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT ICD), including sequential biventricular pacing, is an implantable medical device system that includes 3 major components:
CRT ICD device
leads
programmer, software, and accessories
Each is described in detail below.
Device – The device senses the electrical activity of the patient’s heart using the sensing electrodes of the implanted atrial and right ventricular leads. It then analyzes the heart rhythm based on selectable sensing and detection parameters. The device provides the following functions:
simultaneous or sequential biventricular pacing for cardiac
resynchronization
automatic detection and treatment of ventricular
tachyarrhythmias (ventricular fibrillation, ventricular tachycardia, and fast ventricular tachycardia) with defibrillation, cardioversion, and antitachycardia pacing therapies
single or dual chamber pacing for patients requiring rate
support
diagnostics and monitors that evaluate the system and
assist in patient care
Leads – The device can be used with transvenous or epicardial defibrillation leads. The lead system should consist of bipolar or paired unipolar pacing/sensing leads in the right atrium and right ventricle of the heart, a pacing lead for the left ventricle, and 1 or 2 high-voltage cardioversion/defibrillation electrodes. In addition to the lead system, the Active Can acts as one of the high-voltage electrodes. The device delivers pacing and cardiac resynchronization therapy via the atrial (A), right ventricular (RV), and left ventricular (LV) leads. The device senses using the atrial and RV leads. Cardioversion/defibrillation therapy is delivered with 2 lead-based high-voltage electrodes, or with the Active Can and 1 or 2 lead-based high-voltage electrodes.
INSYNC MAXIMO™7304 Reference Manual
2The InSync Maximo system
System overview
Programmer and software – The Medtronic CareLink Programmer, Model 2090,1and Model 9998 software application allow you to perform the following functions:
configure the cardiac resynchronization, arrhythmia detection
and therapy, and bradycardia features for your patient
perform electrophysiological studies and system tests
monitor, display, or print patient cardiac activity information
For information about:
indications, contraindications, warnings and precautions,
see the implant manual that accompanies the device.
basic programmer and software desktop functions
that are not included in Chapter 10, “Using the programmer”, page 217, see the manual accompanying the programmer.
installing the Model 2067 or Model 2067L programming
head, see the manuals accompanying the programming heads.
implanting leads and using implant tools, refer to the
manuals accompanying the leads and implant tools.
27
2.1.1 Cardiac resynchronization
To improve cardiac output in patients with ventricular dysynchrony, the device provides biventricular pacing. The device paces either the right ventricle or both ventricles as programmed, unless pacing is inhibited by a sensed event in the RV.
Ventricular pacing sequence, V-V pace delay and LV pacing
vector are programmable.
Optional CRT features promote sustained resynchronization
pacing that could be interrupted during episodes of accelerated AV conduction, atrial rate excursions, PVCs, or atrial arrhythmia.
Pacing amplitudes and pulse widths are selected
independently for each ventricle.
1
With the model 2067 programming head
INSYNC MAXIMO™7304 Reference Manual
28
Chapter 2
System overview
2.1.2 Detecting and treating tachyarrhythmias
The device monitors the cardiac rhythm for short ventricular intervals that may indicate the presence of VF, VT, or FVT.
Upon detection of VF, the device delivers a biphasic
defibrillation shock of up to 35 J. If the VF episode persists, up to 5 more individually programmed defibrillation shocks can be delivered.
Upon detection of VT, the device delivers either a Ramp,
Ramp+, or Burst antitachycardia pacing therapy or a biphasic cardioversion shock of up to 35 J synchronized to a ventricular depolarization. If the VT episode persists, up to 5 more individually programmed VT therapies can be delivered. You can also program the device to monitor the VT episode without delivering therapy.
Upon detection of FVT, the device delivers either a Ramp,
Ramp+, or Burst antitachycardia pacing therapy or a biphasic cardioversion shock of up to 35 J synchronized to a ventricular depolarization. If the FVT episode persists, up to 5 more individually programmed FVT therapies can be delivered.
You can program the device to distinguish between true ventricular arrhythmias and rapidly conducted supraventricular tachyarrhythmias (SVT) and withhold therapy for SVT.
SVT discrimination includes the capability to detect a double tachycardia (an unrelated ventricular arrhythmia occurring simultaneously with an SVT) so that therapy is not withheld for a ventricular arrhythmia in the presence of an SVT.
2.1.3 Treating bradycardia
The device provides rate responsive pacing to optimize hemodynamics. An internal accelerometer senses the patient’s physical activity, allowing the device to increase and decrease the pacing rate in response to changes in the level of activity.
INSYNC MAXIMO™7304 Reference Manual
2.1.4 Monitoring for real-time and stored data
The device and programmer provide real-time information on detection and therapy parameters and status during a patient session. The device also provides accumulated data on device operation, including ventricular sensing episodes, stored electrograms, detected and treated tachyarrhythmia episodes, bradycardia interventions, and the efficacy of therapy. The Cardiac Compass report provides up to 14 months of clinically significant data, including physical activity, heart rate, percent pacing, arrhythmia episodes, and therapies delivered.
All of this information can be printed and retained in the patient’s file or saved in electronic format on a floppy diskette.
2.1.5 Conducting electrophysiologic tests
You can use the system to conduct non-invasive electrophysiologic studies, including manual delivery of any of the device therapies to manage an induced or spontaneous tachyarrhythmia.
2.1.6 Alerting the patient to system events
2The InSync Maximo system
Indications and usage
29
You can use the programmable Patient Alert monitoring feature to notify the patient with audible tones if certain conditions related to the leads, battery, charge time, and therapies occur. The patient can then respond based on your prescribed instructions.
2.2 Indications and usage
The InSync Maximo is indicated for ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The system is also indicated for the reduction of the symptoms of moderate to severe heart failure (NYHA Functional Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy and have a left ventricular ejection fraction 35% and a prolonged QRS duration.
INSYNC MAXIMO™7304 Reference Manual
30
Chapter 2
Contraindications
2.3 Contraindications
2.4 Patient screening
Do not use the InSync Maximo system in:
Patients whose ventricular tachyarrythmias may have
transient or reversible causes, such as:
– acute myocardial infarction
– digitalis intoxication
– drowning
– electric shock
– electrolyte imbalance
– hypoxia
– sepsis
Patients with incessant VT or VF
Patients who have a unipolar pacemaker.
Before implant, patients should undergo a complete cardiac evaluation, including electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed tachyarrhythmia therapies are recommended during and after device implant.
Other optional screening procedures could include exercise stress testing to determine the patient’s maximum sinus rate, and cardiac catheterization to determine if there is a need for concomitant surgery and/or medical therapy.
INSYNC MAXIMO™7304 Reference Manual
3Emergency therapy
3.1 Delivering emergency therapies 32
3.2 Delivering an emergency defibrillation therapy 33
3.3 Delivering an emergency cardioversion therapy 34
3.4 Delivering emergency fixed burst pacing 34
3.5 Enabling emergency VVI pacing 35
3
32
Chapter 3
Delivering emergency therapies
3.1 Delivering emergency therapies
The device provides the following emergency therapies:
defibrillation
cardioversion
fixed burst pacing
emergency VVI pacing
The default emergency therapy is 35 J defibrillation. When you select [Emergency] and [DELIVER], the device charges and delivers a biphasic 35 J shock along the AX>B pathway.
The programmer resets the emergency defibrillation energy to 35 J each time you select [Emergency]. Emergency cardioversion and fixed burst values remain as selected for the duration of the session.
To return to other programming functions from an Emergency screen, select [Exit Emergency].
1
3.1.1 Effect on system operation
The device suspends the automatic detection features when emergency defibrillation, cardioversion, or fixed burst pacing therapies are delivered. Detection is not suspended during emergency VVI pacing. Removing the programming head or pressing [Resume] turns detection on again.
3.1.2 Aborting an emergency therapy
As a safety precaution, the programmer also displays an [ABORT] button, which immediately terminates any emergency therapy in progress.
3.1.3 Mechanical Emergency VVI button on the Medtronic CareLink Model 2090 programmer
If you press the red Emergency VVI button on the programmer display panel, the device initiates Emergency VVI pacing and the programmer displays the Emergency screen.
1
If Active Can is turned off, the defibrillation is delivered between the HVX and HVB electrodes.
INSYNC MAXIMO™7304 Reference Manual
3.1.4 Temporary parameter values
2
4
3
3Emergency therapy
Delivering an emergency defibrillation therapy
33
Emergency tachyarrhythmia therapies use temporary values that do not change the programmed parameters of the device. These values are not in effect until you select [DELIVER]. After the tachyarrhythmia therapy is complete, the device reverts to its programmed values.
3.2 Delivering an emergency defibrillation therapy
The default emergency therapy is a full-energy defibrillation. When you select [Emergency] and [DELIVER], the device charges and delivers a biphasic full-energy shock. The programmer sets the emergency defibrillation energy to its maximum value each time you select [Emergency] or select the [Defibrillation] option from an Emergency screen.
3.2.1 How to deliver emergency defibrillation
1. Position the programming head over the device.
2. Select [Emergency].
3. Accept the defibrillation energy shown on the screen, or select a new Energy value.
4. Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly positioned, and select [Retry] or [Cancel].
2
2
Delivery of Emergency VVI Pacing changes the programmed bradycardia pacing values to the emergency values (see Section 3.5, page 35).
INSYNC MAXIMO™7304 Reference Manual
34
2
5
4
3
Chapter 3
Delivering an emergency cardioversion therapy
3.3 Delivering an emergency cardioversion therapy
When you initiate an emergency cardioversion therapy, the device charges its capacitors to the selected energy and attempts to deliver therapy synchronized with a sensed tachyarrhythmia event. If the cardioversion therapy cannot be synchronized, it is aborted. See Section 7.3.5.6, page 136.
3.3.1 How to deliver emergency cardioversion
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the defibrillation energy shown on the screen, or select a new Energy value.
5. Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly positioned and select [Retry] or [Cancel].
3.4 Delivering emergency fixed burst pacing
Emergency fixed burst pacing delivers maximum output pacing pulses to the ventricle at a selectable interval. The therapy continues for as long as you keep the programmer stylus on the [BURST Press and Hold] button.
INSYNC MAXIMO™7304 Reference Manual
Enabling emergency VVI pacing
3
4
2
5
3.4.1 How to deliver emergency fixed burst pacing
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval shown on the screen, or select a new interval value.
5. Select [BURST Press and Hold].
If delivery is not confirmed, the programmer displays an error window. Verify that the programming head is properly positioned. Select [OK] from the window and reselect [BURST Press and Hold].
3.5 Enabling emergency VVI pacing
3Emergency therapy
35
Emergency VVI pacing programs the device to deliver high-output ventricular pacing. You can initiate emergency VVI pacing from the Emergency screen or by pressing the red mechanical button on the Medtronic CareLink Model 2090 programmer display panel. To disable emergency VVI pacing, reprogram the bradycardia pacing parameters from the Parameters screen.
INSYNC MAXIMO™7304 Reference Manual
36
3
4
2
Chapter 3
Enabling emergency VVI pacing
3.5.1 How to deliver emergency VVI pacing
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A successful programming sets the device to the following maximum output bradycardia pacing values:
Pacing Mode: VVI
Lower Rate: 70 ppm
RV. Amplitude: 6 V
RV. Pulse Width: 1.6 ms
V. Pace Blanking: 240 ms
V. Pacing: RV
V. Sense Response: Off
Ventricular Rate
Stabilization: Off Cond AF Response: Off
If programming is not confirmed, verify that the programming head is properly positioned and select [Retry] or [Cancel].
INSYNC MAXIMO™7304 Reference Manual
Device implant and patient follow-up procedures
Part II
4Implanting the device
4.1 Overview 40
4.2 Preparing for an implant 40
4.3 Replacing a device 42
4.4 Surgical approach 43
4.5 Sensing and pacing measurements 47
4.6 Connecting the leads to the device 47
4.7 Testing defibrillation operation and effectiveness 51
4.8 Positioning and securing the device 54
4.9 Completing the implant procedure 54
4
40
Chapter 4
Overview
4.1 Overview
The tasks for implanting a device include:
1. Preparing for an implant
2. Replacing a device
3. Surgical approach
4. Sensing and pacing measurements
5. Connecting the leads to the device
6. Testing defibrillation operation and effectiveness
7. Positioning and securing the device
8. Completing the implant procedure
These tasks are described in the sections that follow.
4.2 Preparing for an implant
Warning: Keep a backup external defibrillator available
during the implant for transthoracic rescue when arrhythmias are induced.
Warning: Do not permit the patient to contact grounded equipment that could produce hazardous leakage current during implant. Resulting arrhythmia induction could result in the patient’s death.
The device is intended for implant with Medtronic transvenous or epicardial leads. No claims of safety and efficacy can be made with regard to other acutely or chronically implanted lead systems that are not manufactured by Medtronic.
4.2.1 Equipment for an implant
Medtronic CareLink Programmer Model 2090 and Model
2067or 2067L programming head
Model 9998 software application
Model 8090 Analyzer lead analysis device or equivalent
pacing system analyzer
external defibrillator
INSYNC MAXIMO™7304 Reference Manual
Model 5358 Defibrillator Implant Support Device and
software application (optional)
4.2.2 Sterile supplies
InSync Maximo and lead system components
Programming head sleeve or programming head
Analyzer cables
Lead introducers or delivery systems appropriate for the
lead system
Extra stylets of appropriate length and shape
4.2.3 How to set up the implant support instrument
When using an implant support instrument such as the Model 5358 Defibrillator Implant Support device:
1. Calibrate any monitoring or recording equipment while recording the EGM and marker outputs of the support instrument.
2. Verify the high-energy output of the support instrument by delivering a high-energy defibrillation shock into the test load.
4Implanting the device
Preparing for an implant
41
4.2.4 How to set up the programmer and start the application
1. Set up the programmer as described in the instructions provided with the programmer.
2. Install the InSync Maximo Model 9998 software on the programmer if it is not already installed.
3. Place the programming head over the device, and start the application. Select the device model or select [Auto identify].
Note: The programmer automatically interrogates the device when the application starts.
INSYNC MAXIMO™7304 Reference Manual
42
Chapter 4
Replacing a device
4.2.5 How to preprogram the device
Before opening the sterile package, prepare the device for implant as follows:
1. Check the “Use by” date printed on the package. Do not implant the device after the “Use by” date because the battery’s longevity can be reduced.
2. Interrogate the device and print a full summary report.
Note: If the programmer reports that an electrical reset occurred, do not implant the device. Contact a Medtronic representative.
3. Confirm that the battery voltage is at least 3.0 V at room temperature.
If the device temperature is lower than room temperature or has delivered a recent high-voltage charge, the battery voltage will be temporarily lower.
4. Set up data collection parameters and the device internal clock (see Section 12.9.3, “How to set up data collection”, page 305).
5. Perform a manual capacitor formation (see “Manual capacitor formation”, page 325).
6. Program the therapy and pacing parameters to values appropriate for the patient. Ensure that all tachyarrhythmia detection is programmed Off (see Section 8.1.4, “How to program CRT and bradycardia pacing parameters”, page 151).
Note: Do not enable VF Detection until the device has been successfully implanted.
1
4.3 Replacing a device
If you are replacing a previously implanted device, turn off device detection and therapies before explanting.
When implanting the device with a chronic lead system, perform the following evaluations to ensure appropriate detection and therapy:
Check the integrity of the chronic high-voltage leads with a
test shock, chest x-ray, and inspection.
Perform chronic pacing and sensing measurements.
Measure high-voltage lead impedances.
Test defibrillation efficacy.
1
Use the Quick Look screen to verify the voltage; see Section 11.2.2, “Quick Look observations”, page 253.
INSYNC MAXIMO™7304 Reference Manual
Confirm adequate sensing during VF.
Ensure proper fit of the lead connectors in the device
connector block.
Notes:
To meet the implant requirements, it may be necessary to
reposition or replace the chronic leads or to add a third high-voltage electrode.
Any unused leads that remain implanted must be capped.
4.3.1 How to explant and replace a device
1. Program all tachyarrhythmia detection Off.
2. Dissect the leads and the device free from the surrounding tissues in the surgical pocket. Be careful not to nick or breach the lead insulation during the process of exposing the system.
3. Loosen each setscrew, and gently retract the lead from the connector block.
4. Remove the device from the surgical pocket.
5. If the connector pin of any implanted lead shows signs of pitting or corrosion, replace the implanted lead with a new lead. The damaged lead should be discarded and replaced to assure the integrity of the device system.
6. Measure sensing, pacing, and defibrillation efficacy using the replacement device or an implant support instrument.
4Implanting the device
Surgical approach
43
4.4 Surgical approach
The InSync Maximo system typically requires a left ventricular lead implant in the coronary sinus and positioned in a cardiac vein. For this reason, physicians should expect that implanting an InSync Maximo system will require more time than implanting a traditional ICD or pacemaker system.
4.4.1 Implanting the device
In most cases the device is implanted in the pectoral region, either submuscularly or subcutaneously, using transvenous leads for pacing, sensing, and high-voltage therapies.
INSYNC MAXIMO™7304 Reference Manual
44
Chapter 4
Surgical approach
4.4.2 Lead configurations
Follow these single-incision guidelines when implanting the device in the pectoral region. The pocket should be about 1.5 times the size of the device.
Submuscular approach: A single incision that extends over
the delta-pectoral groove can provide access to the cephalic and subclavian veins and access to form the implant pocket. Place the device sufficiently medial to the humeral head to avoid interference with shoulder motion.
Subcutaneous approach: The subcutaneous approach can
be similar to a pacemaker implant, using a single transverse incision long enough to permit isolation of the cephalic and subclavian veins. Place the device quite far medially to keep the leads away from the axilla, and keep the upper edge of the device inferior to the incision.
The device is typically implanted with the following leads:
one transvenous lead for the left ventricle (LV) for pacing
one bipolar transvenous lead in the right ventricle (RV) for
sensing, pacing, and cardioversion/defibrillation therapies
one bipolar transvenous lead in the atrium (A) for sensing
and pacing
The InSync Maximo is an Active Can system in which the device case serves as a high-voltage electrode. The device case, the RV coil, and the SVC coil form a 3 electrode system for simultaneous mode defibrillation and cardioversion. High-voltage therapies are delivered between the RV coil and the combined device case and SVC coil.
The device may be implanted with other transvenous leads. Refer to Section 4.4.4, “Using other transvenous leads”, page 46, for more information.
Notes:
Medtronic 3.2 mm low-profile leads are not directly
compatible with the device IS-1 connector block.
For questions about lead and device compatibility, please
contact your Medtronic representative.
INSYNC MAXIMO™7304 Reference Manual
4.4.3 Implanting the lead
Note: Do not implant the LV, atrial, and RV leads in the same
venous access site. Medtronic recommends implanting the LV lead in the subclavian vein, and the atrial and RV leads in the cephalic vein.
If inserting using a subclavian approach, position the lead laterally to avoid pinching the lead body between the clavicle and the first rib. Pinching the lead may eventually cause conductor fracture, damage to the insulation, or other damage to the lead. This may result in complications such as loss of detection, loss of pacing therapies, or loss of cardioversion/defibrillation therapy.
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also cause pinching and subsequent fracture of the lead. If significant resistance is encountered during a subclavian stick implant procedure, do not force passage of the lead by adjusting the patient’s posture (by raising the arm or putting a towel behind the person’s back, for example). Use an alternate venous entry site instead.
4.4.3.1 Implanting the ventricular leads
4Implanting the device
Surgical approach
45
Warning: Backup pacing should be readily available during implant. Use of the delivery system or leads may cause heart block.
Note: Due to the variability of cardiac venous systems, the venous anatomy should be assessed before implanting the LV lead to determine an optimal LV lead position. Before placing a lead in the coronary sinus, obtain a venogram.
When implanting a LV or RV lead, follow the directions for use in the respective lead technical manuals.
Final lead positioning should attempt to optimize both cardiac resynchronization and defibrillation threshold.
When pacing at 10 V using an external pacing device, test for extracardiac stimulation from the LV lead. If extracardiac stimulation is present, consider repositioning the lead.
4.4.3.2 Implanting the right atrial lead
Implant the atrial lead in the cephalic vein, if possible. Follow the directions for use in the lead technical manual.
INSYNC MAXIMO™7304 Reference Manual
46
Chapter 4
Surgical approach
4.4.3.3 Lead positioning
For the initial lead configuration the following positions are recommended:
LV tip electrode – Position as far laterally into the coronary venous system as possible (deflection of the LV EGM signal as late as possible in the QRS complex).
RV tip electrode – Position as far apically as possible (deflection of the RV EGM signal on the RV electrode should start early in the QRS complex). Defibrillation threshold, atrial far-field sensing and RV-LV timing must be considered when choosing the optimal location.
Atrial tip electrode – The atrial tip electrode should be positioned close to the SA node or on the lateral wall to minimize the possibility of far-field R-wave oversensing. Lead placement in these positions potentially provides optimal atrial sensing while reducing far-field R-wave sensing.
4.4.4 Using other transvenous leads
Follow the general guidelines below for initial positioning of other transvenous leads (the final positions are determined by defibrillation efficacy tests):
For an SVC lead, place the lead tip high in the innominate
vein, approximately 5 cm proximal to the atrium (A) and SVC junction.
For an SQ patch lead, place the patch along the left
midaxillary line, centered over the fourth-to-fifth intercostal space.
4.4.5 Epicardial lead systems
The following guidelines should be observed:
The distance between the RVring electrode and the RVtip
electrode should not exceed 1 cm.
The RVtip electrode should not lie in a direct path between
epicardial patches.
The RVtip/RVring and Atip/Aring intrinsic, post-pace, and
post shock electrograms should be thoroughly evaluated for oversensing, undersensing, cross-talk, and noise.
INSYNC MAXIMO™7304 Reference Manual
4Implanting the device
Sensing and pacing measurements
The pacing thresholds and electrogram characteristics
should meet the values indicated in Table 4-1.
4.5 Sensing and pacing measurements
Sensing and pacing measurements include EGM amplitudes, slew rates, capture thresholds, and pacing lead impedances. Medtronic recommends using the Model 8090 Analyzer for sensing and pacing measurements. Refer to the Analyzer technical manual for detailed procedures on performing these measurements.
Do not use the intracardiac EGM telemetered from the device to assess sensing.
Table 4-1 lists the acceptable implant values for acute and chronically implanted lead systems.
Table 4-1. Acceptable implant values
Measurement required
EGM amplitude (during NSR)
Slew rate
Capture threshold (0.5 ms pulse
a
width)
Typical pacing lead impedance
a
Verify adequate pacing threshold margins at implant and at each followup visit.
b
The measured pacing lead impedance is a reflection of measuring equipment and lead technology.
c
Please refer to specific lead technical manuals for acceptable impedance values.
Acute
A
2 mV
0.5 V/s
1.5 V
b
250 - 1000 Ùc(all)
RV
5 mV
0.75 V/s
1.0 V
LV
3 mV
3.0 V
Chronic
A
1 mV
0.3 V/s
3.0 V
RV
3 mV
0.5 V/s
3.0 V
47
LV
1 mV
4.0 V
4.6 Connecting the leads to the device
For more detailed information about lead/connector compatibility, see the Medtronic representative.
InSync Maximo 7304 Implant Manual
INSYNC MAXIMO™7304 Reference Manual
, or contact your
48
2
5
4
6
3
1
Chapter 4
Connecting the leads to the device
Table 4-2. Lead connections
Device port Connector type Software name
SVC (HVX) DF-1 SVC
RV (HVB) DF-1 RVcoil
Can
RV IS-1 bipolar RVtip and RVring
LV IS-1 bipolar LVtip and LVring
A IS-1 bipolar Atip and Aring
Figure 4-1. Lead connections
Can
1 SVC (HVX) 2 RV (HVB) 3 Can (HVA) 4 LV 5 RV 6 A
Caution: Loose lead connections may result in inappropriate sensing and failure to deliver necessary arrhythmia therapy.
INSYNC MAXIMO™7304 Reference Manual
4Implanting the device
Connecting the leads to the device
Cautions:
To ensure appropriate sensing and detection, insert the
right ventricular lead IS-1 connector into the RV port and insert the right ventricular lead DF-1 connector into the RV (HVB) port.
If no SVC electrode is implanted, the pin plug provided
with the device must be secured in the SVC (HVX) port.
Use only the torque wrench supplied with the device.
It is designed to prevent damage to the device from overtightening a setscrew.
Do not reuse the torque wrench after this implant.
Lead and Active Can electrodes in electrical contact
with each other during a high-voltage therapy could cause current to bypass the heart, possibly damaging the device and leads. While the device is connected to the leads, make sure that no therapeutic electrodes, stylets, or guide wires are touching or connected by an accessory low impedance conductive pathway. Move any objects that contain conductive materials (e.g., an implanted guide wire) well away from all electrodes before a high-voltage shock is delivered.
49
Notes:
For easier lead insertion, insert the LV IS-1 lead before the
other leads.
If a chronic lead system uses a different size or number of
leads, insert the lead connectors into appropriate adaptors. See the appropriate lead technical manuals for more details.
Upon initial lead connection to the connector block, LVtip,
RVtip, and RVcoil all need to be connected within the connector block to provide biventricular pacing.
INSYNC MAXIMO™7304 Reference Manual
50
a
b
1
2
Chapter 4
Connecting the leads to the device
4.6.1 How to connect the lead to the device
1. Insert the torque wrench into the appropriate setscrew.
a. If the port is obstructed by
the setscrew, retract the setscrew to clear it. Take care not to disengage the setscrew from the connector block.
b. Leave the torque wrench in
the setscrew until the lead is secure. This allows a pathway for venting trapped air when the lead is inserted.
2. Push the lead or plug into the connector port until the lead pin is clearly visible in the pin viewing area. No sealant is required, but sterile water may be used as a lubricant.
3. Tighten the setscrew by turning clockwise until the torque wrench clicks.
4. Tug gently on the lead to confirm a secure fit. Do not tug on the lead until all setscrews have been tightened.
5. Repeat these steps for each connector port.
4.6.2 Post pace oversensing testing
Changes in the programmed ventricular pacing configuration (RV, RVLV, or LVRV) may evoke slightly different EGMs. To ensure proper sensing regardless of the pacing configuration, or in the
INSYNC MAXIMO™7304 Reference Manual
event of lead dislodgment or failure, the sensing behavior during pacing for all configurations should be evaluated as follows:
1. Program RV Pacing.
2. Print a 10 to 30 s test strip.
3. Check for post pace oversensing.
Repeat Step 1 through Step 3 with ventricular pacing programmed to RVLV and LVRV.
4Implanting the device
Testing defibrillation operation and effectiveness
If oversensing occurs during one of the pacing configurations, repositioning of the RV lead may help alleviate the oversensing.
4.7 Testing defibrillation operation and effectiveness
Demonstrate reliable defibrillation effectiveness with the implanted lead system by using your preferred method to establish that a 10 J (minimum) safety margin exists.
Note: If the 10 J (minimum) safety margin cannot be ensured, see Section 15.5, “Solving tachyarrhythmia therapy problems”, page 350.
4.7.1 High-voltage implant values
Measured values must meet the following requirements at implant.
Table 4-3. High-voltage therapy values at implant
Measurement Acute or Chronic Leads
V. Defib impedance
SVC (HVX) impedance (if applicable)
Defibrillation threshold 25 J
20 – 200 Ù
20 – 200 Ù
51
Warning: Ensure that an external defibrillator is charged for a rescue shock.
INSYNC MAXIMO™7304 Reference Manual
52
Chapter 4
Testing defibrillation operation and effectiveness
4.7.2 How to prepare for defibrillation threshold testing
1. Place the programming head over the device, start a patient session, and interrogate the device, if you have not already done so.
2. Observe the Marker Channel telemetry annotations and the programmer ECG display to verify that the device is sensing properly.
3. Conduct a manual Lead Impedance Test2to verify the defibrillation lead connections. Perform this test with the device in the surgical pocket, and keep the pocket very moist. If the impedance is out of range, perform one or more of the following tasks:
Recheck lead connections and electrode placement.
Repeat the measurement.
Inspect the bipolar EGM for abnormalities.
Measure the defibrillation impedance with a manual test shock.
4. Program the device or support instrument to properly detect VF with an adequate safety margin (1.2 mV sensitivity).
2
See Section 13.4, “Measuring lead impedance”, page 320 .
INSYNC MAXIMO™7304 Reference Manual
Testing defibrillation operation and effectiveness
1
2
3
4
5
6
7
8
9
10
11
12
13
14
4.7.3 How to perform defibrillation threshold testing
4Implanting the device
1. Select Tests > EP Study.
2. Select either 50 Hz BURST or T-shock induction.
3. Select [Resume at BURST] or [Resume at DELIVER].
4. Select [Adjust Permanent…].
5. Program VF Enable On.
6. Program the automatic therapy energy settings. Therapies 2-6 should be set to the maximum energy.
7. Select [Program].
8. Select [Close].
9. If performing a T-Shock induction, select the [Enable] checkbox.
10. Select [DELIVER], or [50 Hz BURST Press and Hold].
If necessary, you can abort an induction or therapy in progress by pressing [ABORT].
11. Observe the live rhythm monitor for proper post-shock sensing.
12. Use the [Adjust Permanent…] button to program the next appropriate energy level.
13. Wait until the on-screen timer reaches 5 min, then repeat Step 9 through Step 12 as needed.
14. Select Params > Detection and program VF, FVT, and VT detection Off before closing.
53
INSYNC MAXIMO™7304 Reference Manual
54
Chapter 4
Positioning and securing the device
4.8 Positioning and securing the device
Cautions:
If no SVC electrode is implanted, the pin plug provided
with the device must be secured in the SVC (HVX) port.
Program tachyarrhythmia detection Off before closing.
4.8.1 How to position and secure the device
1. Ensure that each lead pin or plug is fully inserted into the connector block and that all setscrews are tight.
2. Coil any excess lead length beneath the device. Avoid kinks in the lead conductors.
3. Implant the device within 5 cm of the skin and with the Medtronic label facing the skin for improved Patient Alert volume. This position optimizes the ambulatory monitoring operations.
4. Use nonabsorbable sutures to secure the device within the pocket to minimize post-implant rotation and migration of the device. Use a surgical needle to penetrate the suture holes.
4.9 Completing the implant procedure
After implanting the device, x-ray the patient to verify device and lead placement. To complete programming the device, select parameters that are appropriate for the patient.
INSYNC MAXIMO™7304 Reference Manual
Completing the implant procedure
4.9.1 How to complete programming the device
1. After closing the pocket, program detection On. Program ventricular tachyarrhythmia therapies On as desired.
2. Do not enable the Other 1:1 SVTs PR Logic detection criterion until the atrial lead has matured (approximately 1 month post implant).
3. If external equipment was used to conduct the defibrillation efficacy tests, perform a final VF induction, and allow the implanted system to detect and treat the arrhythmia.
4. Monitor the patient after the implant, and take x-rays as soon as possible to document and assess the location of the leads.
5. Program patient information. See Section 12.10, “Viewing and entering patient information”, page 308.
6. Configure the Patient Alert feature. See Section 11.3, “Using the Patient Alert feature”, page 254.
7. Set up data collection parameters. See Section 12.9, “Setting up data collection”, page 302.
8. Interrogate the device after any spontaneous episodes to evaluate the detection and therapy parameter settings.
9. If the patient has not experienced spontaneous episodes, you may induce the clinical tachyarrhythmias using the non-invasive EP Study features to further assess the performance of the system. See Chapter 14, “Conducting Electrophysiologic Studies”, page 327.
10. Recheck pacing and sensing values, and adjust if necessary. See Section 13.3, “Measuring pacing thresholds”, page 317.
4Implanting the device
55
INSYNC MAXIMO™7304 Reference Manual
5Conducting a patient follow-up session
5.1 Patient follow-up guidelines 58
5.2 Verifying the status of the implanted system 59
5.3 Verifying effective cardiac resynchronization therapy 59
5.4 Verifying effective basic pacing 60
5.5 Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias 62
5
58
Chapter 5
Patient follow-up guidelines
5.1 Patient follow-up guidelines
Schedule regular patient follow-up sessions to monitor the condition of the device and leads and to verify that the device is configured appropriately for your patient.
During the first few months after receiving a new device, the patient may require close monitoring. Schedule an office visit at least every 3 months.
The Quick Look screen, which is displayed after you interrogate the device, provides a good beginning for the follow-up review. Using this screen you can perform the following tasks:
Verify that the device is functioning correctly.
Verify that the device is delivering biventricular pacing most
or all of the time.
Review the clinical performance and long term trends.
Print appropriate reports to compare the results to the
patient’s history and to retain for future reference.
For details about reviewing clinical trend data, see
Section 12.4, “Viewing the Heart Failure Management
report”, page 278
Section 12.5, “Viewing clinical trends in the Cardiac
Compass report”, page 282
Note: The Checklist feature displays a list of standard follow-up tasks on the programmer screen for easy reference during the follow-up session. You can also customize your own checklists if you wish. See Section 11.4, “Streamlining follow-ups with Checklist”, page 262, for more information.
INSYNC MAXIMO™7304 Reference Manual
5Conducting a patient follow-up session
Verifying the status of the implanted system
5.2 Verifying the status of the implanted system
To verify that the device and leads are functioning correctly, review the following information from the Quick Look screen and perform followup tests as indicated:
Review the displayed battery voltage for comparison to the
Elective Replacement Indicator value (see Section 1.4, “Replacement indicators”, page 20). Remember that battery voltage may be low if high-voltage charging has occurred within 24 hours.
Review the last full-energy charge.
– For information about adjusting the capacitor formation
interval, see Section 9.1, “Optimizing charge time”, page 210.
– If the programmer displays an Excessive Charge Time
ERI, the device should be replaced immediately.
Review the defibrillation and pacing lead impedance values
for inappropriate values or large changes since the last followup. See Section 13.4, “Measuring lead impedance”, page 320.
Perform an EGM Amplitude test in the RV and atrium for
comparison to previous EGM amplitude measurements. See Section 13.5.5, “How to perform an EGM Amplitude test”, page 324.
To review longer term trends in sensing and impedance
measurements, select the [>>] button from the lead performance area of the Quick Look screen. The programmer displays a detailed history of automatic sensing and impedance measurements. See Section 11.2, “Taking a quick look at device activity”, page 253.
59
5.3 Verifying effective cardiac resynchronization therapy
To verify that the device is providing cardiac resynchronization therapy appropriately, review the following information from the initial interrogation report or Quick Look screen and investigate as indicated:
Confirm that the patient is receiving adequate cardiac
support for daily living activities.
INSYNC MAXIMO™7304 Reference Manual
60
Chapter 5
Verifying effective basic pacing
5.3.1 Considerations
Review Quick Look Observations related to ventricular
pacing percentage or ventricular sensing episodes.
Check the stored Ventricular Sensing Episode records for
appropriate interventions and continuity of resynchronization pacing. See Section 12.7, “Viewing episode data”, page 291.
Check the rate histograms for more information on atrial and
ventricular pacing in general, and ventricular rates during AT/AF episodes.
Review the Heart Failure Management and Cardiac
Compass reports for comparison to patient history (see Section 12.4, “Viewing the Heart Failure Management report”, page 278).
Review the following information before assessing resynchronization therapy.
Undersensing, far-field oversensing, or loss of capture
are basic pacing issues that can affect delivery of cardiac resynchronization therapy. These situations can often be resolved with basic programming changes and then monitored for further occurrences.
If CRT delivery is lost during AF or during atrial rate
excursions, consider the following programming options:
– Perform basic programming changes to increase the
Upper Tracking Rate or decrease the total atrial refractory period (for information on total atrial refractory period, see Section 8.2.4, “Programming considerations for atrial rates”, page 162).
– Enable Ventricular Sense Response, Atrial Tracking
Recovery, or Conducted AF Response to promote more continuous delivery of CRT.
5.4 Verifying effective basic pacing
To verify that the device is sensing and pacing appropriately, review the following information from the Initial Interrogation Report, and Quick Look screen, and investigate as indicated:
Confirm that the patient is receiving adequate cardiac
support for daily living activities.
INSYNC MAXIMO™7304 Reference Manual
5.4.1 Considerations
Review the following information before verifying bradycardia pacing.
5Conducting a patient follow-up session
Verifying effective basic pacing
Review the pacing conduction history and rate histograms for comparison to the patient history. A sharp increase in the atrial pacing percentage may indicate a need for investigation and analysis.
Review the recorded Mode Switch episodes for comparison to the patient’s atrial arrhythmia history (see Section 12.7, “Viewing episode data”, page 291). A dramatic increase in frequency or duration of atrial episodes may indicate a need for investigation and analysis.
Review the Cardiac Compass report for comparison to patient history (see Section 12.5, “Viewing clinical trends in the Cardiac Compass report”, page 282).
Conduct pacing threshold tests (see Section 13.3, “Measuring pacing thresholds”, page 317) to verify that the programmed pacing outputs provide a sufficient safety margin.
61
Atrial pacing – If the conduction history or rate histograms show a predominance of atrial pacing despite a healthy sinus response, consider these options to decrease the atrial pacing burden:
Decrease the Lower Rate.
Decrease the rate response or increase the activity
threshold.
Conduction history – If the reported percentages in the conduction history do not add up to 100, the percentages may be rounded.
INSYNC MAXIMO™7304 Reference Manual
62
Chapter 5
Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias
5.5 Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias
To verify that the device is providing effective tachyarrhythmia detection and therapy, review the following information from the Quick Look screen and investigate as indicated:
Review Quick Look Observations that relate to patient
history and device operation. To display more detailed information about any observation, select the observation and then select the [>>] button.
Review any Patient Alerts listed in the Observations of the
Quick Look screen. For the most detailed information about Patient Alerts, select Patient Alert from the Data icon and select [Events].
Check stored episode records for appropriate sensing
and detection of arrhythmias. See Section 12.7, “Viewing episode data”, page 291.
Check stored SVT episode records for appropriate
identification of SVTs.
5.5.1 Considerations
Review the following information before verifying detection and therapy.
Flashback memory – In addition to the episode text and stored electrograms, use Flashback memory and interval plots to help investigate the accuracy and specificity of ventricular detection.
Episode misidentification – If the episode records indicate that false detections have occurred, the Sensing Integrity counter may help in determining the prevalence of oversensing. For more information, see Section 12.2.4.3, “Sensing integrity counter”, page 274.
If the device is oversensing, consider these programming options:
Increase the Pace Blanking value.
Increase the sensitivity threshold.
Caution: Do not reprogram the device to decrease oversensing without assuring that appropriate sensing is maintained. See Section 6.2, “Setting up sensing”, page 71.
INSYNC MAXIMO™7304 Reference Manual
5Conducting a patient follow-up session
Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias
If the episode records reveal that a stable monomorphic VT has been identified and treated as VF, consider these options to improve the detection accuracy:
Review the Interval Plot for the episode, and adjust the VF
Interval, if necessary. Use caution when reprogramming the VF Interval, because changes to this value can adversely affect VF detection.
Consider enabling FVT via VF detection. See Section 6.5,
“Detecting FVT episodes”, page 85.
If the SVT episode records include episodes of true VT, review the SVT episode record to identify the SVT detection criterion that withheld detection. Adjust the SVT detection criteria parameters as necessary. See Section 6.8, “Enhancing detection with PR Logic criteria”, page 96, and Section 6.9, “Enhancing VT detection with the Stability criterion”, page 104.
63
INSYNC MAXIMO™7304 Reference Manual
Configuring the device for the patient
Part III
6Detecting tachyarrhythmias
6.1 Detection overview 68
6.2 Setting up sensing 71
6.3 Detecting VF episodes 76
6.4 Detecting VT episodes 79
6.5 Detecting FVT episodes 85
6.6 Detecting tachyarrhythmia episodes with Combined Count 90
6.7 Monitoring episodes for termination or redetection 92
6.8 Enhancing detection with PR Logic criteria 96
6.9 Enhancing VT detection with the Stability criterion 104
6.10 Detecting double tachycardias 106
6.11 Detecting prolonged tachyarrhythmias with High Rate Timeout 107
6
68
Chapter 6
Detection overview
6.1 Detection overview
The device detects ventricular tachyarrhythmias (VF, VT, and FVT) by comparing the time intervals between sensed ventricular events to a set of programmable detection intervals. If enough intervals occur that are shorter than the programmed intervals, the device detects a tachyarrhythmia and responds automatically with a programmed therapy. After delivering the therapy, the device either redetects the arrhythmia and delivers the next programmed therapy or detects episode termination.
To avoid detecting rapidly conducted SVTs (for example, sinus tachycardia or atrial fibrillation) as ventricular tachyarrhythmias, the device provides several detection enhancements, including PR Logic and Stability detection criteria.
All of the detection features interact during initial detection (see Figure 6-1). During redetection, the device does not apply the PR Logic detection criteria.
Note: Detection functions can be turned off by programming the VF Enable, FVT Enable, and VT Enable parameters to Off. For an example, see Section 6.3.4, “How to program VF detection”, page 78.
INSYNC MAXIMO™7304 Reference Manual
6Detecting tachyarrhythmias
Ventricular Event
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
Yes
Yes
Update counts and pattern information
Is the interval in the VF, FVT, or VT detection zone?
Has High Rate Timeout suspended detection enhancements?
Does Stability reset the VT event count? (VT and FVT via VT
detection only)
Has a tachyarrhythmia event count reached an NID?
Are one or more PR Logic criteria on?
No/suspended by High Rate
Timeout
Tachy
Episode
Detected
Is the median ventricular interval less than the SVT Limit?
Is there a double tachycardia in progress?
Are one or more PR Logic criteria withholding detection?
Figure 6-1. How detection features interact during initial detection
69
Detection overview
6.1.1 Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops classifying and counting tachyarrhythmia intervals. Sensing and programmed pacing remain active, and the programmed detection settings are not modified.
INSYNC MAXIMO™7304 Reference Manual
70
Chapter 6
Detection overview
Detection is suspended:
when the device senses the presence of a strong magnet.
The programmer head contains a magnet which suspends detection, but once telemetry between the device and programmer is established, detection resumes.
while performing any of the manual system tests, including
Underlying Rhythm, Pacing Threshold, Lead Impedance, EGM Amplitude, and Charge/Dump. Detection automatically resumes once the test is complete.
while performing a T-Shock, 50 Hz Burst, Manual Burst,
or PES Induction. You can choose to have the device automatically resume detection after delivering the induction.
when you deliver a Manual or Emergency therapy. You
can resume detection by selecting the [Resume] button or removing the programming head from the device.
when you select the on-screen [Suspend] button. You can
resume detection by selecting the [Resume] button or by removing the programming head from the device.
during the automatic daily lead impedance measurements.
Detection resumes when the measurements are complete.
while the device is delivering an automatic tachyarrhythmia
therapy (including capacitor charging for defibrillation and cardioversion). However, the device does continue to confirm the detected episode during charging. Detection resumes when the therapy is complete.
Note: The device suspends VT detection and Combined Count detection (see Section 6.6, “Detecting tachyarrhythmia episodes with Combined Count”, page 90) for 17 events following a defibrillation therapy delivered in response to a detected VF.
during charging for Automatic Capacitor Formation.
Detection resumes when charging is complete.
1
1
If the defibrillation therapy is delivered as a result of either a High Rate Timeout Therapy operation or Progressive Episode Therapies operation, VT detection is not suspended (see Section 6.11.4.1, “High Rate Timeout Therapy”, page 109).
INSYNC MAXIMO™7304 Reference Manual
6.2 Setting up sensing
The device provides bipolar sensing in both the atrium and right ventricle using the sensing electrodes of the implanted atrial and RV leads. You can adjust the sensitivity to intracardiac signals using independent atrial and ventricular sensitivity settings. These settings define the minimum electrical amplitude recognized by the device as an atrial or ventricular sensed event.
Proper sensing is essential for the safe and effective use of the device. To provide appropriate sensing, the device uses:
auto-adjusting atrial and ventricular sensing thresholds
short (30 ms) cross-chamber blanking after paced events
no cross-chamber blanking after sensed events
See Section 6.2.4, “Details about sensing”, page 74.
6.2.1 Parameters
6Detecting tachyarrhythmias
Setting up sensing
71
V. Sensitivity – Minimum amplitude of electrical signal that registers as a sensed ventricular event.
A. Sensitivity – Minimum amplitude of electrical signal that registers as a sensed atrial event.
6.2.2 Considerations
Review the following information before programming sensing parameters.
Dual chamber sensing and pacing modes – The device senses in both the atrium and the right ventricle at all times, except when the programmed bradycardia pacing mode is DOO. When the pacing mode is programmed to DOO or VOO mode, there is no sensing in the ventricle. mode, you must first disable detection.
Sensitivity thresholds – The programmed atrial and ventricular sensitivity thresholds apply to all features related to sensing, including detection and bradycardia pacing.
2
Atrial sensing occurs in VOO mode.
0.15; 0.3
0.6; 0.9; 1.2 mV
0.15; 0.3
0.6; 0.9; 1.2; 1.5;
2.1 mV
2
In order to program either DOO or VOO
; 0.45;
; 0.45;
INSYNC MAXIMO™7304 Reference Manual
72
Chapter 6
Setting up sensing
Bradycardia pacing and sensing – A combination of high pacing pulse width or high amplitude with a low sensitivity threshold may cause inappropriate sensing across chambers or in the same chamber. Programming a lower pulse width, lower amplitude, longer pace blanking, or a higher sensitivity threshold may eliminate this inappropriate sensing.
Recommended ventricular sensitivity threshold – A ventricular sensitivity threshold of 0.3 mV is recommended to maximize the probability of detecting VF and to limit the possibility of oversensing and cross-chamber sensing.
High ventricular sensitivity threshold – Setting V. Sensitivity to a value greater than 0.6 mV is not recommended except for testing. Doing this may cause undersensing, which can cause any of the following situations:
delayed or aborted cardioversion therapy
delayed defibrillation therapy (when VF confirmation is
active)
asynchronous pacing
underdetection of tachyarrhythmias
Low ventricular sensitivity threshold – If you set V. Sensitivity to its most sensitive value of 0.15 mV, the device will be more susceptible to EMI, cross-chamber sensing, and oversensing.
Recommended atrial sensitivity threshold – An atrial sensitivity threshold of 0.3 mV is recommended to optimize the effectiveness of PR Logic detection criteria and atrial pacing operations, while limiting the possibility of oversensing and cross-chamber sensing.
High atrial sensitivity threshold – If you set the A. Sensitivity value too high, the device may not provide reliable sensing of P-waves during SVTs and sinus rhythm.
Low atrial sensitivity threshold – If you set the A. Sensitivity value to its most sensitive value of 0.15 mV, the device will be more susceptible to EMI, far-field R-wave sensing, and oversensing.
Testing sensitivity after reprogramming – If you change the ventricular sensitivity threshold, evaluate for proper sensing and detection by inducing VF and allowing the device to automatically detect and treat the arrhythmia.
INSYNC MAXIMO™7304 Reference Manual
Atrial pacing and ventricular sensing – If you program the
1
2
3
device to an atrial pacing mode, make sure that it does not sense atrial pacing pulses as ventricular events.
Sensing during VF – Always verify that the device senses properly during VF. If the device is not sensing or detecting properly, program detection and therapies off, and evaluate the system (making sure to monitor the patient for life-threatening arrhythmias until you enable detection and therapies again). You may need to reposition or replace the ventricular sensing lead to achieve proper sensing.
Atrial lead selection – Atrial leads with minimal tip-to-ring spacing may reduce far-field R-wave sensing.
Repositioning the atrial lead – You may need to reposition or replace the atrial sensing lead if reprogramming the atrial sensitivity threshold does not provide reliable atrial sensing during SVTs and sinus rhythm.
6.2.3 How to program sensitivity
6Detecting tachyarrhythmias
Setting up sensing
1. Select Params > Detection.
2. Select the desired A. Sensitivity and V. Sensitivity parameters.
3. Select [PROGRAM].
73
INSYNC MAXIMO™7304 Reference Manual
74
A P
V S
A S
V S
A S
V P
Sensitivity Threshold
Rectified and
Filtered A. EGM
Rectified and
Filtered RV EGM
Marker Channel
1
2
3
4
5
Chapter 6
Setting up sensing
6.2.4 Details about sensing
6.2.4.1 Auto-adjusting sensitivity thresholds
The device automatically adjusts the sensitivity thresholds after certain paced and sensed events to help reduce oversensing from T-waves, cross-chamber events, and pacing. You can see how sensitivity thresholds are adjusted after different types of events (see Figure 6-2).
Figure 6-2. Auto-adjusting sensitivity thresholds
1 After an atrial sensed event, the atrial sensitivity threshold increases
to 75% of the EGM peak (maximum: 8x the programmed value, decay constant: 200 ms).
2 After a ventricular sensed event, the ventricular sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed value, decay constant: 450 ms).
3 After an atrial paced event, the device does not adjust the atrial
sensitivity threshold. The ventricular sensitivity threshold increases by 0.45 mV (decay constant: 60 ms).
4 After a ventricular paced event, the atrial sensitivity threshold
increases to 4x the programmed value (maximum: 1.8 mV, immediate return after 60 ms).
5 After the ventricular pace blanking period is finished, the ventricular
threshold increases to 4.5x the programmed value (maximum:
1.8 mV, decay constant: 450 ms).
3
The exponential decay continues through a subsequent ventricular pacing pulse and its blanking period.
INSYNC MAXIMO™7304 Reference Manual
3
4
4
6Detecting tachyarrhythmias
Setting up sensing
6.2.4.2 Blanking periods
During a blanking period, the device does not sense electrical signals. This helps prevent sensing of device pacing, cardioversion and defibrillation pulses, post-pacing depolarization, T-waves, and multiple sensing of the same event. The blanking periods following paced events are longer than those following sensed events to avoid sensing the atrial and ventricular depolarizations.
Notes:
To enhance sensing and detection during tachyarrhythmias,
the device does not apply cross-chamber blanking (blank sensing in the opposite chamber) after a sensed event.
Atrial sensing is still active during the Post-Ventricular
Atrial Blanking (PVAB) period (see Section 8.2.3.5, “Post-Ventricular Atrial Refractory Period”, page 160).
The duration of the fixed blanking periods is shown in Table 6-1. For information on programmable pace blanking periods, see Section 8.1, “Providing basic pacing therapy”, page 146.
Table 6-1. Duration of fixed blanking periods
Cross-chamber blanking after atrial or ventricular pacing pulse
Atrial blanking after sensed atrial event
Ventricular blanking after sensed ventricular event
Atrial and ventricular blanking after delivered cardioversion or defibrillation therapy
30 ms
100 ms
120 ms
520 ms
75
6.2.4.3 Refractory periods
During a refractory period, the device senses normally, but classifies sensed events as refractory and limits its response to these events. Pacing refractory periods prevent inappropriately sensed signals, such as far-field R-waves (ventricular events sensed in the atrium) or electrical noise, from triggering certain pacing timing intervals.
4
If the programmed sensitivity value exceeds 0.3 mV (ventricular) or 1.2 mV (atrial), the threshold is not adjusted.
INSYNC MAXIMO™7304 Reference Manual
76
Chapter 6
Detecting VF episodes
Synchronization refractory periods help prevent the device from delivering cardioversion and defibrillation therapies at inappropriate times. See Section 7.1.5.5, “Synchronizing defibrillation without confirming VF”, page 116, and Section 7.3.5.6, “Synchronizing cardioversion after charging”, page 136.
Note: Refractory periods do not affect tachyarrhythmia detection.
6.3 Detecting VF episodes
The device detects VF episodes by examining the cardiac rhythm for short ventricular intervals. If a predetermined number of intervals occurs that are short enough to be considered VF events, the device detects VF and delivers the first programmed VF therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing.
See Section 6.3.5, “Details about VF detection”, page 78.
6.3.1 Parameters
VF Detection Enable – Turns VF
detection on or off.
VF Interval (Rate) – V-V intervals shorter than this value are counted as VF events.
VF Initial NID – Number of Intervals to Detect: number of VF events the device must count to detect a VF episode.
VF Redetect NID – Number of Intervals to Redetect: number of VF events the device must count to redetect a continuing VF after a therapy.
6.3.2 Considerations
Review the following information before programming VF Detection parameters.
VF Interval minimum setting – To ensure proper VF detection, you should not program the VF Interval to a value less than 300 ms.
INSYNC MAXIMO™7304 Reference Manual
; Off
On
240; 250 … 320 400 ms
12/16; 18/24 30/40; 45/60; 60/80; 75/100; 90/120; 105/140; 120/160
6/8; 9/12; 12/16 21/28; 24/32; 27/36; 30/40
; 24/32;
; 18/24;
6Detecting tachyarrhythmias
Detecting VF episodes
VF Interval maximum setting – Programming the VF Interval to a value greater than 350 ms may cause inappropriate detection of rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter, which is more sensitive than the consecutive VT event counter.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
Enabling VF detection – When VF Detection Enable is programmed On for the first time, the device:
enables Automatic Capacitor Formation
starts recording Cardiac Compass data and Heart Failure
Management data
starts recording lead performance trends (starting at
3:00 AM by the device clock)
clears all brady pacing counters
VF detection and PR Logic criteria – You can program the device to exclude rapidly conducted SVTs from VF Detection by enabling the PR Logic detection criteria. Note that the SVT Limit must be programmed shorter than the VF Interval in order for the PR Logic criteria to affect VF detection. See Section 6.8, “Enhancing detection with PR Logic criteria”, page 96.
77
6.3.3 Restrictions
Double tachycardia detection – When any PR Logic detection
criteria are enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT). See Section 6.10, “Detecting double tachycardias”, page 106.
Review the following information before programming VF Detection parameters.
Tachyarrhythmia detection and pacing – To ensure reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for pacing and tachyarrhythmia detection. See Section B.11, “Parameter interlocks”, page 388.
INSYNC MAXIMO™7304 Reference Manual
78
1
2
3
Chapter 6
Detecting VF episodes
VF detection backup – To ensure VF Detection backup during VT and FVT episodes, if VF Detection is off, both VT Detection and FVT Detection must also be off.
6.3.4 How to program VF detection
1. Select Params > Detection.
2. Select the desired values for VF Enable, VF Initial NID, VF Redetect NID, and VF Interval.
3. Select [PROGRAM].
6.3.5 Details about VF detection
The device detects VF by counting the number of VF events, which are V-V intervals shorter than the programmed VF Interval. On each event, the device counts the number of recent VF events. The number of recent events that are examined is called the VF detection window. The size of the VF detection window is the second number in the programmed VF NID (for example, 24 events if the VF Initial NID is 18/24).
The threshold for detecting VF is the first number in the programmed VF NID (for example, 18 events if the VF Initial NID is 18/24). This threshold is always 75% of the VF detection window. That is, if 75% of the events in the VF detection window are VF events, the device detects a VF episode (see Figure 6-3).
After the device detects VF, it delivers the first programmed VF therapy. Following the therapy, if the number of VF events reaches the programmed VF Redetect NID, the device redetects
INSYNC MAXIMO™7304 Reference Manual
VF and delivers the next programmed VF therapy.
Note: The device can also detect VF Episodes using the Combined Count detection criterion (see Section 6.6, page 90).
Figure 6-3. Device detects VF
A S
A S
V S
V S
V S
V S
V S
V S
V S
A R
A R
A R
A R
A R
1 2 3 4 5 6 6 7 8 9 10 11 12 13 1415 16 17 18
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F S
F D
200 ms
2
3
1
VF Event Count
ECG
Marker Channel
VF Interval
1 VF starts, and the device begins counting VF events (intervals less
than the programmed VF Interval).
2 A ventricular interval occurs outside the VF detection zone. The VF
event count is not incremented.
3 The VF event count reaches the programmed VF NID value of 18
events out of 24, and the device detects VF.
6Detecting tachyarrhythmias
Detecting VT episodes
79
6.4 Detecting VT episodes
The device detects VT episodes by examining the cardiac rhythm for short ventricular intervals. If enough intervals occur that are short enough to be considered VT events (but are not VF or FVT events), the device detects VT and delivers the first programmed VT therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing.
You can program the device to detect and record VT episodes without treating them with VT therapies by setting VT Detection Enable to Monitor. If a patient’s VT episodes are well-tolerated, this feature allows you to collect data about these episodes without delivering therapy or affecting VF detection.
See Section 6.4.5, “Details about VT detection”, page 81.
INSYNC MAXIMO™7304 Reference Manual
80
Chapter 6
Detecting VT episodes
6.4.1 Parameters
VT Detection Enable – Turns VT
detection on or off, or enables VT monitoring.
VT Interval (Rate) – V-V intervals shorter than this value are counted as VT events.
VT Initial NID – Number of Intervals to Detect: number of VT events the device must count to detect a VT episode.
VT Redetect NID – Number of Intervals to Redetect: number of VT events the device must count to redetect a continuing VT after a therapy.
6.4.2 Considerations
Review the following information before programming VT Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
On; Off
280; 290 … 400
12; 16
4; 8; 12
; or Monitor
…52; 76; 100
… 52
… 600 ms
VT Detection Enable, AFib/AFlutter, and Sinus Tach – When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and Sinus Tach parameters are also automatically set to On.
VT detection and Combined Count detection – When VT Detection is On, the device applies the Combined Count detection criterion to help speed detection of rhythms that fluctuate between detection zones. Combined Count detection is disabled if VT Detection is set to Off or Monitor. See Section 6.6, “Detecting tachyarrhythmia episodes with Combined Count”, page 90.
VT detection and rapidly conducted SVTs – You can program the device to exclude rapidly conducted SVTs from VT detection by enabling the PR Logic or Stability detection criteria. See Section 6.9, “Enhancing VT detection with the Stability criterion”, page 104, and Section 6.8, “Enhancing detection with PR Logic criteria”, page 96.
INSYNC MAXIMO™7304 Reference Manual
Double tachycardia detection – When any PR Logic detection
1
2
3
criteria are enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT; see Section 6.10, page 106).
6.4.3 Restrictions
Review the following information before programming VT detection parameters.
Tachyarrhythmia detection and pacing – To ensure reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for pacing and tachyarrhythmia detection. See Section B.11, “Parameter interlocks”, page 388.
VF detection backup – To ensure VF detection backup during VT and FVT episodes, if VF Detection is off, both VT Detection and FVT Detection must also be off.
6.4.4 How to program VT detection
6Detecting tachyarrhythmias
Detecting VT episodes
1. Select Params > Detection.
2. Select the desired values for VT Enable, VT Initial NID, VT Redetect NID, and VT Interval.
3. Select [PROGRAM].
81
6.4.5 Details about VT detection
The device detects VT by counting the number of consecutive VT events. A VT event is a V-V interval shorter than the programmed VT Interval but greater than or equal to the VF Interval. If the number of consecutive VT events reaches the programmed VT Initial NID, the device detects VT (see Figure 6-4).
INSYNC MAXIMO™7304 Reference Manual
82
Chapter 6
Detecting VT episodes
The VT event count resets to zero whenever an interval occurs that is greater than or equal to the programmed VT Interval. The count remains at the current value if an interval is shorter than the programmed VF Interval.
After the device detects VT, it delivers the first programmed VT therapy. Following the therapy, if the VT event counter reaches the VT Redetect NID, the device redetects VT and delivers the next programmed therapy.
Note: The device can also detect VT Episodes using the Combined Count detection criterion (see Section 6.6, page 90).
INSYNC MAXIMO™7304 Reference Manual
Figure 6-4. Device detects VT
A S
A S
V S
V S
V S
V S
A R
A R
A R
A R
A R
A R
A R
1 0 1 2 3 4 5 6
T S
T S
T S
T S
T S
T S
T S
V S
A R
A R
A R
A R
A R
A R
A R
A R
A R
A R
A R
A R
7 8 9 10 11 12 13 14 15 16
T S
T S
T S
T S
T S
T S
T S
T S
T S
T D
2
1
3
VT Event Count
ECG
Marker Channel
VT Interval
VT Event Count
ECG
Marker Channel
VT Interval
6Detecting tachyarrhythmias
Detecting VT episodes
83
1 VT starts, and the device begins counting VT events (intervals less
than the programmed VT Interval but greater than or equal to the VF Interval).
2 A ventricular interval occurs outside VT detection zone. The VT event
count resets to zero.
3 The VT event count reaches the programmed VT NID of 16 events,
and the device detects VT.
6.4.5.1 VT monitoring
You can program the device to record VT episodes without delivering VT therapy by setting VT Detection to Monitor. When VT monitoring is enabled, the device detects VT episodes but does not deliver VT therapy (see Figure 6-5). Instead, it records VT episodes, labeling them as “monitored,” and waits for episode termination to occur.
When VT Detection is set to Monitor, several detection operations work differently.
INSYNC MAXIMO™7304 Reference Manual
84
Chapter 6
Detecting VT episodes
VT event counting – Before the device detects an episode, it counts VT events normally. However, once the VT Initial NID is reached, the device sets the VT event count to zero and suspends VT event counting for the rest of the episode.
VF and FVT detection – VF and FVT detection operate as if VT detection is off. Specifically, Combined Count detection is disabled, and FVT via VT detection is not selectable. If a monitored VT episode accelerates into the FVT or VF detection zone, the device applies the VF Initial NID to detect the new tachyarrhythmia. Once an episode is in progress, VT event counting doesn’t resume until the episode ends.
Caution: Programming the VF Interval greater than 350 ms may result in inappropriate detection of rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter, which is more sensitive than the consecutive VT event counter.
PR Logic and Stability criteria – Before the device detects a tachyarrhythmia episode, the PR Logic and Stability criteria, if turned on, are applied. If a monitored VT episode accelerates into the FVT or VF detection zone, the device continues to apply PR Logic criteria as initial VF or FVT detection begins. However, because the Stability feature does not affect VF detection or FVT via VF detection, it is not applied.
Episode termination – The device compares ventricular intervals to the VT Interval to identify when a VT monitored episode has ended. However, if a VF episode or FVT via VF episode occurs when VT monitoring is enabled, the device compares ventricular intervals to the VF Interval to identify episode termination.
INSYNC MAXIMO™7304 Reference Manual
Figure 6-5. Device detects and monitors VT
200 ms
1 2 3 4
A S
A R
A RA RA
R
V S
V ST ST ST ST
S
13 14 15 16 0 0 0
A R
A R
A RA RA RA RA
R
T ST ST ST DT ST ST
S
2
3
1
VT Event Count
ECG
Marker Channel
VT Interval
1 VT starts, and the device begins counting VT events (intervals less
than the programmed VT Interval but greater than or equal to the VF Interval).
2 The VT event count reaches the programmed VT NID of 16 events,
and the device detects VT.
3 After detecting the VT episode, the device resets the VT event count
to zero and monitors the episode until termination.
6Detecting tachyarrhythmias
Detecting FVT episodes
85
6.5 Detecting FVT episodes
The device detects episodes of Fast Ventricular Tachycardia (FVT) by examining the cardiac rhythm for short ventricular intervals. If enough intervals occur in the programmed FVT detection zone, the device detects FVT and delivers the first programmed FVT therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing. To make sure it delivers sufficiently aggressive therapies, the device can merge the programmed detection zones during redetection to increase sensitivity.
See Section 6.5.5, “Details about FVT detection”, page 88.
INSYNC MAXIMO™7304 Reference Manual
86
Chapter 6
Detecting FVT episodes
6.5.1 Parameters
FVT Detection Enable – Enables FVT
detection via the VF or the VT detection algorithm.
FVT Interval (Rate) – V-V intervals between this value and the programmed VF Interval are marked as FVT events.
6.5.2 Considerations
Review the following information before programming FVT Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
FVT detection enable – Your choice for an appropriate setting for FVT Detection should depend on the patient’s VF and VT cycle lengths. After determining a reliably sensitive VF Interval, consider the following suggestions:
; via VF; or via
Off VT
200; 210 … 600 ms
If the patient presents with a clinical VT interval in the VF zone, select via VF to ensure reliable detection of VF. (VT Detection need not be enabled at all.)
If the patient presents with 2 clinical VTs, both outside the
VF zone, select via VT to allow for correct classification of the faster VT and to offer a separate therapy regimen for each VT.
If the patient presents with only 1 clinical VT, which is outside
the VF zone, select VF and VT Detection only, and set FVT Enable to Off.
FVT detection and PR Logic criteria – You can program the device to exclude rapidly conducted SVTs from FVT Detection by enabling the PR Logic detection criteria. Note that the SVT Limit must be programmed shorter than the VF Interval for the PR Logic criteria to affect FVT via VF detection.
INSYNC MAXIMO™7304 Reference Manual
6.5.3 Restrictions
6Detecting tachyarrhythmias
Detecting FVT episodes
Double tachycardia detection – When any PR Logic detection criterion is enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT, see Section 6.10, page 106).
Review the following information before programming FVT Detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To ensure reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for bradycardia pacing and tachyarrhythmia detection. See Section B.11, “Parameter interlocks”, page 388.
VF detection backup – To ensure VF Detection backup during VT and FVT episodes, VT and FVT Detection cannot be on unless VF Detection is also on.
FVT detection – To ensure reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for the FVT parameter as follows:
VT Detection must be set to On if FVT Detection is set to
via VT.
87
If FVT Detection is set to via VF, the FVT Interval must be
programmed to a value shorter than the VF Interval.
If FVT Detection is set to via VT, the FVT Interval must be
programmed to a value greater than the VF Interval and less than or equal to the VT Interval.
INSYNC MAXIMO™7304 Reference Manual
88
1
2
3
Chapter 6
Detecting FVT episodes
6.5.4 How to program FVT detection
6.5.5 Details about FVT detection
You can program the device to detect FVT episodes via the VF or VT detection zone and NID.
1. Select Params > Detection.
2. Select the desired values for FVT Enable and FVT Interval.
3. Select [PROGRAM].
When FVT Detection is set to via VF, a V-V interval within the FVT detection zone is marked as an “FVT via VF” event. When the VF NID is reached, the device reviews the last 8 intervals:
If any of the last 8 intervals are in the VF zone, it detects
the episode as VF.
If all of the last 8 intervals are outside the VF zone, it detects
the episode as FVT (see Figure 6-6).
When FVT Detection is set to via VT, a V-V interval within the FVT detection zone is marked as an “FVT via VT” event. When the VT NID is reached, the device reviews the last 8 intervals:
If any of the last 8 intervals are in the VF or FVT zones, it
detects the episode as FVT.
If all of the last 8 intervals are outside the FVT and VF zones,
it detects the episode as VT.
Note: The device can also detect FVT episodes via the Combined
INSYNC MAXIMO™7304 Reference Manual
Count detection criterion (see Section 6.6, page 90).
Figure 6-6. Device detects FVT via VF
200 ms
13 14 15 16 17 18
A RA RA RA RA RA RA RA RA
R
T FT
F
T F
T F
T F
T F
T FV
S
1 1 2 3 4 5
A S
A S
A RA RA
R
A R
V S
V S
V S
T S
T F
T F
T F
T F
T F
1
2
3
VF Event Count
ECG
Marker Channel
VF and FVT Intervals
1 A fast ventricular tachycardia starts, and the first event falls into the
FVT detection zone.
2 The second event of the FVT episode has an interval that falls into
the VT zone. The VF event count is not incremented.
3 The device detects FVT after the VF event count reaches the VF
Initial NID.
6Detecting tachyarrhythmias
Detecting FVT episodes
89
6.5.5.1 Zone merging after detection
To ensure that the device delivers sufficiently aggressive therapies during an extended or highly variable tachyarrhythmia episode, the device merges detection zones during redetection in some instances, as shown in Figure 6-7. The merged zone configuration uses the event counting and therapies for the faster arrhythmia and remains in effect until episode termination.
INSYNC MAXIMO™7304 Reference Manual
90
Before detection
After VF detection
After FVT detection
FVT set to “via VF”
VF and FVT zones merge, leaving a larger VF zone.
All zones remain unchanged.
FVT set to “via VT”
VT and FVT zones merge, leaving a larger FVT zone.
VT and FVT zones merge, leaving a larger FVT zone.
Detection Intervals: VF Interval: 320 ms, FVT Interval: 280 ms / 360 ms, VT Interval: 400 ms
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
Chapter 6
Detecting tachyarrhythmia episodes with Combined Count
Figure 6-7. FVT zone merging
6.6 Detecting tachyarrhythmia episodes with Combined Count
INSYNC MAXIMO™7304 Reference Manual
Because the device counts VF and VT events separately, rhythms with variable cycle lengths can cause both event counts to increment during an episode. To prevent these rhythms from delaying detection, the device automatically enables the Combined Count detection criterion if both VF and VT detection are programmed On.
The Combined Count criterion compares the sum of the VF and VT event counts to the Combined Number of Intervals to Detect (CNID), which the device calculates automatically from the programmed VF NID values. If the CNID is met, the device reviews the recent intervals to determine if the episode should be treated as a VF, FVT, or VT episode. The Combined Count criterion applies during both initial detection and redetection.
Detecting tachyarrhythmia episodes with Combined Count
6.6.1 Details about Combined Count detection
The Combined Count detection algorithm expedites detection or redetection of ventricular tachyarrhythmias with ventricular intervals that fluctuate between the VF and VT detection zones. When VT detection is on, the device applies Combined Count detection, which tracks the combined number of VT and VF events counted. If this sum reaches the Combined Number of Intervals to Detect (CNID), the device detects VF, FVT, or VT. Combined Count detection also applies to redetected episodes.
Note: Combined Count detection is off when VT detection is set to Monitor or Off.
If the VF event counter reaches 6, the device automatically applies the Combined Number of Intervals to Detect (CNID). The CNID is calculated by multiplying the current VF NID (Initial or Redetect) by 7/6 and rounding down. Table 6-2 shows the CNID values that correspond to each VF NID value.
Table 6-2. CNID values for each initial or redetect VF NID value
VF NID CNID VF NID CNID
6/8
9/12
12/16
18/24
7
10
14
21
6Detecting tachyarrhythmias
21/28
24/32
27/36
30/40
24
28
31
35
91
Combined Count detection is fulfilled when the sum of the VF and VT event counts equals or exceeds the CNID. The device then reviews the last 8 intervals and classifies the episode as one of the following:
VF, if any of the last 8 were in the VF zone
FVT, if FVT Detection is enabled and none of the last 8 was
in the VF zone, but one or more was in the FVT zone
VT, if all 8 were outside the VF zone (and FVT zone, if FVT
detection is enabled)
INSYNC MAXIMO™7304 Reference Manual
92
200 ms
A SA RA RA RA RA RA RA RA RA RA RA RA
R
A RA RA RA RA RA RA RA
R
A R
V SV SF
S
F S
F SF SF SF SF SF ST SF SF SF SF SF SF SF SF
S
F S
T ST
S
F DV
S
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
1 2 3 4
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
VF Event Count
VT Event Count
Combined Count
VF and VT Interval
ECG
Marker Channel
1
2
3
Chapter 6
Monitoring episodes for termination or redetection
Figure 6-8. Device detects VF with the Combined Count criterion
1 A slow VF episode starts with a ventricular cycle length that varies
between the VF and VT detection zones.
2 When a VT event occurs, the device increments the VT event count
and the Combined Count.
3 The device detects VF even though the VF event count hasn’t yet
reached the VF Initial NID (18/24 in this example). The Combined Count reaches the CNID value of 21 first.
6.7 Monitoring episodes for termination or redetection
Once the device detects an arrhythmia, it considers the episode
INSYNC MAXIMO™7304 Reference Manual
ongoing until it detects that the episode has ended. After delivering therapy, it monitors the ventricular rhythm using the programmed Redetect NIDs. If one of these NIDs is met, the device delivers the next programmed therapy for the detected arrhythmia.
See Section 6.7.4, “Details about episode termination and redetection”, page 93.
6.7.1 Parameters
1
2
3
6Detecting tachyarrhythmias
Monitoring episodes for termination or redetection
93
VF Redetect NID – Number of Intervals to Redetect: number of VF events the device must count to redetect a continuing VF after a therapy.
VT Redetect NID – Number of Intervals to Redetect: number of VT events the device must count to redetect a continuing VT after a therapy.
6.7.2 Considerations
Review the following information before programming redetection parameters.
Initial and Redetect NIDs – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
6.7.3 How to program redetection parameters
6/8; 9/12; 12/16 21/28; 24/32; 27/36; 30/40
4; 8; 12
1. Select Params > Detection.
2. Select the desired values for VT Redetect NID and VF Redetect NID.
3. Select [PROGRAM].
… 52
; 18/24;
6.7.4 Details about episode termination and redetection
After a therapy is delivered, the device evaluates the ventricular rhythm to determine if the episode has terminated, is continuing, or has changed to a different arrhythmia.
INSYNC MAXIMO™7304 Reference Manual
94
Chapter 6
Monitoring episodes for termination or redetection
6.7.4.1 Episode termination
The device determines that the episode has terminated if one of the following conditions occurs:
8 consecutive ventricular intervals are greater than or equal
to the programmed VT interval.
20 s elapse with no ventricular intervals shorter than the
programmed VT interval.
After antitachycardia pacing therapy, the device begins evaluating intervals for episode termination on the first ventricular cycle. After cardioversion or defibrillation, the device begins evaluating intervals for episode termination on the second ventricular event. (Due to the extended post shock blanking, this event may be the
third event on the electrogram.)
Note: Any subsequent detection after the end of the episode
marks the start of a new episode.
6.7.4.2 Episode redetection
After the device detects a tachyarrhythmia episode and delivers a therapy, the device redetects an arrhythmia if the VF or VT event count reaches the Redetect NID or if the combined VF and VT event count reaches the Redetect CNID (see Section 6.6, “Detecting tachyarrhythmia episodes with Combined Count”, page 90).
5
5
The device then delivers the next programmed therapy for the current arrhythmia and resumes monitoring for the outcome of that therapy (see Figure 6-9).
5
VF interval if VT Detection is set to Off or Monitor and the episode is a VF or an FVT via VF episode.
INSYNC MAXIMO™7304 Reference Manual
Monitoring episodes for termination or redetection
A RA SA SA SA SA SA RA RA RA RA RA RA RA RA RA RA RA RA
R
T P
T PT
P
T PT DT PT PT PT PT PT ST ST ST ST ST ST ST ST ST ST ST
D
200 ms
016 0 1 2 3 4 5 6 7 8 9 10 11 12
3
2
1
VT Event Count
ECG
Marker Channel
VT Interval
Figure 6-9. VT episode redetected after therapy
1 A VT episode is detected, and the device delivers a Burst ATP therapy. 2 After therapy, the device continues to detect events in the VT zone. 3 When the VT event count reaches the VT Redetect NID, the device
redetects the VT.
Notes:
The device suspends VT detection (and Combined Count
detection) for 17 events following a defibrillation therapy delivered in response to a detected VF. detection helps avoid detecting transient VTs that can follow high-voltage therapies.
The PR Logic criteria are not applied during redetection.
However, the Stability criterion may withhold detection or redetection of VT (or FVT via VT) throughout an episode.
6Detecting tachyarrhythmias
6
Suspending VT
95
6.7.4.3 VT acceleration
If the device redetects VT, it classifies the rhythm as accelerated if the average of the 4 intervals before redetection is at least 60 ms less than the average of the 4 intervals before initial VT detection. The most recent interval average is used to identify VT acceleration if VT is redetected again during the episode.
6
If the defibrillation therapy is delivered as a result of either a High Rate Timeout Therapy operation or Progressive Episode Therapies operation, VT detection is not suspended (see Section 6.11.4.1, page 109).
INSYNC MAXIMO™7304 Reference Manual
96
Chapter 6
Enhancing detection with PR Logic criteria
If the device redetects VF or an accelerated VT after an antitachycardia pacing sequence delivery, it skips the subsequent pacing therapy sequences for the duration of the episode and delivers the next therapy programmed for the current arrhythmia.
6.8 Enhancing detection with PR Logic criteria
The PR Logic detection criteria are designed to withhold inappropriate ventricular detection during episodes of rapidly conducted supraventricular tachycardia (SVT). The device analyzes the activation patterns and timing in both chambers using PR Logic pattern and rate analysis. This information helps identify evidence of atrial fibrillation, atrial flutter, sinus tachycardia, and other 1:1 SVTs. If this analysis indicates the presence of one or more of these rhythms, the device withholds detection.
For more information, see
Section 6.8.5, “Details about PR Logic pattern and rate
analysis”, page 99
Section 6.8.6, “Details about the PR Logic detection
criteria”, page 103
6.8.1 Parameters
AFib/AFlutter – Identifies rapidly conducted
atrial fibrillation, atrial flutter, or atrial tachycardia
Sinus Tach – Identifies sinus tachycardia
1:1 VT-ST Boundary – Allows customizing
of the Sinus Tach criterion for patients with slow one-to-one conduction by identifying the threshold between the retrograde and antegrade zones used by A-V pattern analysis.
Other 1:1 SVTs – Identifies other one-to-one SVTs where the atrial and ventricular activation are roughly simultaneous
SVT Limit – Defines the minimum ventricular interval at which the device applies the PR Logic criteria
a
When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and Sinus Tach parameters are also automatically set to On.
INSYNC MAXIMO™7304 Reference Manual
Ona; Off
Ona; Off
35; 50
On; Off
240; 250 … 320 330; 340 … 600 ms
; 66; 75; 85%
;
6.8.2 Considerations
Review the following information before programming PR Logic parameters.
PR Logic criteria and double tachycardia detection – When any of the PR Logic criteria are enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT; see Section 6.10.1, page 106).
6Detecting tachyarrhythmias
Enhancing detection with PR Logic criteria
Cautions:
Before enabling the Other 1:1 SVTs criterion, ensure
that the atrial lead has matured. This criterion could inappropriately withhold therapy if atrial sensing is compromised by an unstable or dislodged atrial lead.
Use caution when programming the Other 1:1 SVTs
criterion in patients who exhibit slow 1:1 retrograde conduction during VF or VT. This criterion could inappropriately withhold VF/VT therapy in such patients. See Section 6.8.5.1, “Pattern analysis of A-V and V-A intervals”, page 100.
Use caution when programming the 1:1 VT-ST
Boundary. Incorrect programming of this parameter can result in inappropriate therapies or underdetection of tachyarrhythmias.
97
Sinus Tach and 1:1 VT-ST Boundary – When enabling the Sinus Tach criterion in patients who exhibit slow 1:1 retrograde conduction during VF or VT or slow antegrade conduction during SVTs, make sure to set the 1:1 VT-ST Boundary parameter appropriately. See Section 6.8.2.1, “Customizing PR Logic for patients with slow conduction”, page 98.
VT Detection Enable, AFib/AFlutter and Sinus Tach – When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and Sinus Tach parameters also automatically set to On.
SVT Limit – To ensure that therapy is delivered for hemodynamically compromising rates of any origin, the device always delivers therapy when the median ventricular interval is less than the programmed SVT Limit (nominally 320 ms) if VT, VF, or FVT detection criteria are satisfied.
VF Interval and SVT Limit – If you program an SVT Limit greater than the VF Interval, you are effectively disabling the PR Logic criteria for VF detection.
INSYNC MAXIMO™7304 Reference Manual
98
Chapter 6
Enhancing detection with PR Logic criteria
6.8.2.1 Customizing PR Logic for patients with slow conduction
Slow antegrade conduction during an SVT or slow 1:1 retrograde conduction during VT can adversely affect the accuracy of the Sinus Tach and Other 1:1 SVTs criteria. The 1:1 VT-ST Boundary parameter allows you to customize the Sinus Tach criterion for patients who exhibit slow conduction.
Note: Changing the 1:1 VT-ST Boundary parameter does not affect the operation of the Other 1:1 SVTs criterion. Use caution when enabling this criterion in patients who exhibit slow antegrade or retrograde conduction.
The 1:1 VT-ST Boundary parameter represents an A-V interval that is a percentage of a V-V interval. This value separates the retrograde and antegrade zones used by PR Logic pattern analysis operations. See Section 6.8.5.1, “Pattern analysis of A-V and V-A intervals”, page 100.
If slow retrograde or antegrade conduction causes events to occur in the incorrect zone, you can use the 1:1 VT-ST Boundary parameter to increase the size of the appropriate zone. See Figure 6-10, page 100.
If a patient exhibits slow antegrade or retrograde conduction and could benefit from a different 1:1 VT-ST Boundary setting, choose a new setting as follows:
If the patient exhibits long V-A intervals during VT with 1:1
retrograde conduction, select 35%.
If the patient exhibits long A-V intervals during an SVT,
select a value that exceeds the A-V/V-V ratio observed during stored VT/VF or SVT episodes. See Section 12.7.3.1, page 297.
6.8.3 Restrictions
Review the following information before programming PR Logic parameters.
Sinus Tach and 1:1 VT-ST Boundary – The Sinus Tach criterion must be on before 1:1 VT-ST Boundary can be selected.
VT Detection Enable and SVT Limit – The SVT Limit must be shorter than the VT Interval (or VF Interval if VT Detection is off).
INSYNC MAXIMO™7304 Reference Manual
6Detecting tachyarrhythmias
1
2
3
4
5
6
Enhancing detection with PR Logic criteria
6.8.4 How to program the PR Logic detection criteria
1. Select Params > Detection.
2. Select the desired values for AFib/AFlutter, Sinus Tach, Other 1:1 SVTs, and SVT Limit.
3. If you need to adjust the 1:1 VT-ST Boundary for your patient, select Additional Settings….
4. Select the desired value for the 1:1 VT-ST Boundary.
5. Select [OK].
6. Select [PROGRAM].
99
6.8.5 Details about PR Logic pattern and rate analysis
PR Logic pattern and rate analysis is based on the following aspects of atrial and ventricular activation:
A-V and V-A interval patterns
atrial and ventricular rate
AF evidence
far-field R-wave sensing
A:V dissociation
V-V regularity
The information collected by PR Logic pattern and rate analysis is used by the PR Logic detection criteria to identify the presence of SVTs and withhold detection.
INSYNC MAXIMO™7304 Reference Manual
100
85%
75%
66%
35%
1
2
3
4
5
6
Chapter 6
Enhancing detection with PR Logic criteria
6.8.5.1 Pattern analysis of A-V and V-A intervals
The device uses pattern analysis to identify sinus tachycardia, atrial flutter, and other 1:1 SVTs. Within each V-V interval, the device categorizes the atrial rhythm according to the number of intervening atrial events and the zones in which those atrial events occur (see Figure 6-10). From this information, the device assigns pattern codes to the intervals, and interprets the pattern codes to identify SVTs.
Figure 6-10. Zones used during A-V pattern analysis
1 Junctional Zone: 50 ms. Atrial events in the first junctional zone
indicate PAC, PVC, junctional rhythms, atrial fibrillation, or atrial flutter.
2 Retrograde Zone. Atrial events in the retrograde zone indicate
retrograde conduction of ventricular events.
3 The 1:1 VT-ST Boundary separates the retrograde and antegrade
zones. The nominal value of this boundary is 50%.
4 Antegrade (Normal) Zone. Atrial events in the antegrade zone
INSYNC MAXIMO™7304 Reference Manual
indicate normal conduction (sinus rhythm, sinus tachycardia).
5 Junctional Zone: 80 ms. Atrial events in the second junctional zone
indicate PAC, PVC, junctional rhythms, atrial fibrillation, or atrial flutter.
6 The 1:1 VT-ST Boundary can be programmed to different values,
changing the relative size of the antegrade and retrograde zones.
Loading...