Dual chamber implantable cardioverter defibrillator with
cardiac resynchronization therapy including sequential
biventricular pacing
™
7304
Reference Manual
Caution: Federal Law (USA) restricts this device to sale
by or on the order of a physician (or properly licensed
practitioner).
INSYNC MAXIMO
™
7304
Reference Manual
A guide to the operation and programming of the Model 7304
InSync Maximo dual chamber implantable cardioverter defibrillator
with cardiac resynchronization therapy
Before implanting the device, it is strongly recommended that you:
Manual conventions
Throughout this document, the word “device” refers to the
implanted InSync Maximo device.
Introduction
Refer to the product literature packaged with the device for
•
information about prescribing the device.
Thoroughly read this manual and the technical manuals for
•
the leads and the implant tools used with the device.
Discuss the procedure and the device system with the
•
patient and any other interested parties, and provide them
with any patient information packaged with the device.
11
Technical support
The
nominal value for that parameter.
On-screen buttons are shown with the name of the button
surrounded by brackets: [Button Name].
Medtronic employs highly trained representatives and engineers
located throughout the world to serve you and, upon request,
to provide training to qualified hospital personnel in the use of
Medtronic products.
In addition, Medtronic maintains a professional staff of consultants
to provide technical consultation to product users. For medical
consultation, Medtronic can often refer product users to outside
medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative,
or call or write Medtronic at the appropriate address or telephone
number listed on the back cover.
symbol in parameter tables indicates the Medtronic
INSYNC MAXIMO™7304Reference Manual
12
Introduction
References
Notice
The primary reference for background information is Zacouto FI,
Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B,
Cardiac Pacing Diagnostic and Therapeutic Tools.
ed.
Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
Estes M, Manolis AS, Wang P, Eds.
•
Cardioverter-Defibrillators
Inc. 1994.
Kroll MW, Lehmann MH, Eds.
•
Cardioverter-Defibrillator Therapy: The Engineering-Clinical
Interface
Singer I, Ed.
•
NY: Futura Publishing Co. 1994.
Singer I, Barold SS, Camm AJ, Eds.
•
Therapy of Arrhythmias for the 21st Century: The State of
the Art
. Norwell, MA: Kluwer Academic Publishers 1996.
Implantable Cardioverter-Defibrillator
. Armonk, NY: Futura Publishing Co. 1998.
. New York, NY: Marcel Dekker,
Implantable
Implantable
Nonpharmacological
New York:
. Armonk,
This software is provided as an informational tool for the end
user. The user is responsible for accurate input of patient
information into the software. Medtronic makes no representation
as to the accuracy or completeness of the data input into the
software. Medtronic SHALL NOT BE LIABLE FOR ANY DIRECT,
INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES TO
ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE
INFORMATION PROVIDED IN THE SOFTWARE.
sensing, one IS-1 connector for
pacing, two DF-1 connectors for
high-voltage therapy, Active Can
electrode (programmable)
a
Measurements are nominal values based on CAD (computer aided design)
model measurements and are rounded to the nearest unit.
b
Volume with connector holes unplugged.
c
Grommets may protrude slightly beyond the can surface.
d
Engineering series number follows the radiopaque code.
e
These materials have been successfully tested and verified for biocompatibility.
The device does not produce an injurious temperature in the surrounding tissue.
Table 1-2. Lead connections
Device portConnector typeSoftware name
SVC (HVX)DF-1SVC
RV (HVB)DF-1RVcoil
Can
RVIS-1 bipolarRVtip and RVring
INSYNC MAXIMO™7304Reference Manual
—
Can
1Quick reference
2
5
4
6
3
1
Physical characteristics
Table 1-2. Lead connections (continued)
Device portConnector typeSoftware name
LVIS-1 bipolarLVtip and LVring
AIS-1 bipolarAtip and Aring
Figure 1-1. Lead connections
1 SVC (HVX)
2 RV (HVB)
3 Can (HVA)
4 LV
5 RV
6 A
17
Figure 1-2. Suture holes
INSYNC MAXIMO™7304Reference Manual
18
Chapter 1
Magnet application
1.2Magnet application
Bringing a magnet close to the device triggers changes in device
operation as shown in Table 1-3. When the magnet is removed,
the device returns to its programmed operations.
For information on demonstrating Patient Alert tones to the
patient, see Section 11.3.4, “Instructing the patient”, page 260.
Table 1-3. Effects of magnet application on the device
Pacing modeas programmed
Pacing rate and intervalas programmed
VF, VT, and FVT detectionsuspended
Patient Alert audible tones
(20 s or less)
a
b
c
a
b
high/low dual tone (high urgency
•
alert occurred)
on/off intermittent tone (lower
•
urgency alert occurred)
continuous test tone (no alert
•
occurred)
no tone (alerts are disabled)
•
Rate response adjustments are suspended while a Patient Alert tone sounds.
Detection resumes if telemetry is established and the application software is
running, or detection resumes after the application software has started.
Or “VF Detection/Therapy Off” is the only alert enabled.
c
1.3Projected longevity
Longevity estimates are based on accelerated battery discharge
data and device modeling with EGM pre-storage off (see
Table 1-4).
INSYNC MAXIMO™7304Reference Manual
1Quick reference
Projected longevity
Table 1-4. Projected longevity in years with pacing outputs programmed
to the specified amplitude and 0.4 ms pulse width, DDD mode, 100%
biventricular pacing, specified percentage of atrial pacing at 60 ppm
with the remainder at 70 bpm atrial tracking, semi-annual full-energy
charges.
Amplitudes:
A/RV; LV
2.5 V; 3.0 V
3.0 V; 4.0 V
a
a
% atrial
pacing
0%5.86.26.5
25%5.66.16.4
50%
100%5.25.86.1
0%5.0
25%4.85.3
50%4.65.25.6
100%4.45.05.4
Semi-annual full-energy charges may include therapy shocks or capacitor
formations. Additional full-energy charges due to therapy shocks, device testing,
or capacitor formation reduces device longevity by approximately 23 days (0.06
years). Additional 30 J charges due to therapy shocks or device testing reduces
device longevity by approximately 21 days (0.06 years).
500 Ù
pacing
impedance
5.5
700 Ù
pacing
impedance
6.06.3
5.5
900 Ù
pacing
impedance
5.8
5.7
19
The longevity of the ICD is dependent on several factors. The
following factors result in decreased longevity:
An increase in pacing rate, pacing amplitude or pulse
•
width; the ratio of paced to sensed events; or the charging
frequency
A decrease in pacing impedance
•
Using the pre-onset EGM storage feature or Holter telemetry
•
Considerations for using EGM pre-storage – When the EGM
pre-storage feature is programmed Off, the device starts to store
EGM following the third tachyarrhythmia event and also provides
up to 20 s of information before the onset of the tachyarrhythmia,
including:
AA and VV intervals
•
Marker Channel
•
Interval Plot Flashback
•
INSYNC MAXIMO™7304Reference Manual
20
Chapter 1
Replacement indicators
When the EGM pre-storage feature is programmed On, the
device also collects up to 20 s of EGM information before the
onset of the arrhythmia.
In a patient who uniformly repeats the same onset mechanisms,
the greatest clinical benefit of pre-onset EGM storage is achieved
after a few episodes are captured. To maximize the effectiveness
of the EGM pre-storage feature and optimize device longevity,
consider these programming options:
•
•
Turn pre-storage On to capture possible changes in the
onset mechanism following significant clinical adjustments,
for example, device implant, medication changes, and
surgical procedures.
Turn pre-storage Off once you have successfully captured
the information of interest.
1.4Replacement indicators
Battery voltage and messages about replacement status appear
on the programmer display and on printed reports. Table 1-5 lists
the Elective Replacement Indicator (ERI) and the End of Life
(EOL) conditions.
Table 1-5. Replacement indicators
Elective Replacement (ERI)≤ 2.62 V
End of Life (EOL)3 months after ERI
ERI date – The Quick Look and Battery and Lead Measurements
screens display the date when the battery reached ERI.
Temporary voltage decrease – The battery voltage temporarily
decreases following a high-voltage charge. If a battery
measurement is taken immediately after a high-voltage charge,
the ERI or EOL indicator may be displayed. However, this is a
temporary status that will return to normal when the battery has
recovered from the charge.
EOL indication – If the programmer indicates that the device is
at EOL, replace the device immediately.
INSYNC MAXIMO™7304Reference Manual
Post-ERI conditions – EOL device status is defined as 3 months
following an ERI indication assuming the following post-ERI
conditions: 0% DDD atrial pacing, 100% DDD RV and LV pacing
at 60 ppm, 3 V atrial and RV pacing amplitude, 4 V LV pacing
amplitude, 0.4 ms pulse width; 500 Ù pacing load; and six 35 J
charges. EOL may be indicated before the end of 3 months if the
device exceeds these conditions.
1.5Typical charge times
The most recent capacitor charge time appears on the
programmer display and on printed reports and can be evaluated
using the Charge/Dump test (see Table 1-6).
Table 1-6. Typicalafull-energy charge times
At Beginning of Life (BOL)
At Elective Replacement (ERI)
a
These charge times are typical when the capacitors are fully formed.
7.1 s
9.0 s
1Quick reference
Typical charge times
21
1.6High-voltage therapy energy
The stored energy of the device is derived from the peak
capacitor voltage and is always greater than the energy delivered
by the device. Table 1-7 compares the programmed energy
levels delivered by the device to the energy levels stored in the
capacitors before delivery.
Table 1-7. Comparison of delivered (programmed) and stored energy levels
Energy ( J)
Delivereda/
ProgrammedStored
35397.1
32376.5
30346.1
2832
26305.2
25295.0
24274.8
22254.4
b
Charge
Timec( s)
5.7
Delivereda/
ProgrammedStored
Energy ( J)
10122.0
910.51.8
89.31.6
7
67.11.2
5
44.80.8
33.60.6
INSYNC MAXIMO™7304Reference Manual
8.21.4
5.91.0
b
Charge
Timec( s)
22
Chapter 1
Stored data and diagnostics
Table 1-7. Comparison of delivered (programmed) and stored energy levels (continued)
Energy ( J)
Delivereda/
ProgrammedStored
20234.0
18213.6
16193.2
15173.0
14162.8
13152.6
12142.4
11132.2
b
Charge
Timec( s)
Delivereda/
ProgrammedStored
Energy ( J)
Charge
b
Timec( s)
22.40.4
1.82.20.4
1.62.00.3
1.41.70.3
1.21.50.2
1.01.20.2
0.81.00.2
0.60.80.1
0.40.50.1
a
Energy delivered at connector block into a 75 Ù load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth
Detection countersLifetime total, since cleared, and since last session
Episode countersEpisodes:
Therapy efficacy countersCounts for each VF, FVT, VT Therapy:
Other stored data
Patient Alert — System eventsUp to 10 log entries: text and date for the first time an alert
Battery and lead measurementsBattery voltage, last capacitor formation, last charge,
Lead performance trends14 days of daily measurements plus 80 weeks of weekly
intervals, markers, and counter data text
VF, FVT, and VT
•
Atrial Fibrillation/Atrial Flutter episodes
•
Sinus Tach episodes
•
Other 1:1 SVT episodes
•
NST episodes
•
Mode switch episodes
•
Percentage pacing:
AS-VS, AS-VP, AP-VS, AP-VP percentages
•
Additional counters:
Single PVCs and PVC runs
•
Rate stabilization pulses and runs
•
Delivered
•
Successful
•
Unsuccessful
•
Intervention (manually aborted)
•
Total number of aborted shocks
is triggered between interrogations
lead impedance, EGM amplitude measurements, last
high-voltage therapy, and sensing integrity counter
minimum and maximum measurements:
Lead impedance: atrial pacing, RV pacing, LV pacing,
•
defibrillation pathway, and SVC lead (if used)
atrial EGM amplitude: (P-waves)
•
INSYNC MAXIMO™7304Reference Manual
24
Chapter 1
New and enhanced features
Table 1-8. Stored data and diagnostics (continued)
Rate Histograms
Cardiac Compass trends14 months of daily measurements:
Atrial rate
•
Ventricular rate
•
Ventricular rate during AT/AF
•
VT and VF episodes per day
•
one or more high-voltage therapy delivered
•
ventricular rate during VT or VF
•
episodes of non-sustained tachycardia per day
•
heart rate variability
•
total daily time in AT/AF
•
ventricular rate during AT/AF
•
percent pacing per day
•
patient activity
•
average day and night ventricular heart rate
•
1.8New and enhanced features
The following features are new or changed from the 7289 InSync
II Marquis ICD.
1.8.1Tachyarrhythmia therapy
Output – The InSync Maximo has a maximum delivered energy
of 35 J.
1.8.2Cardiac resynchronization therapy
Sequential biventricular pacing – The ventricular sequence
and V-V pace delay are programmable to support improved
hemodynamics.
Pacing outputs for RV and LV are independently programmable.
INSYNC MAXIMO™7304Reference Manual
2The InSync Maximo system
2.1 System overview26
2.2 Indications and usage29
2.3 Contraindications30
2.4 Patient screening30
2
26
Chapter 2
System overview
2.1System overview
The Model 7304 InSync Maximo dual chamber implantable
cardioverter defibrillator with cardiac resynchronization therapy
(CRT ICD), including sequential biventricular pacing, is an
implantable medical device system that includes 3 major
components:
CRT ICD device
•
leads
•
programmer, software, and accessories
•
Each is described in detail below.
Device – The device senses the electrical activity of the patient’s
heart using the sensing electrodes of the implanted atrial and
right ventricular leads. It then analyzes the heart rhythm based
on selectable sensing and detection parameters. The device
provides the following functions:
simultaneous or sequential biventricular pacing for cardiac
•
resynchronization
automatic detection and treatment of ventricular
•
tachyarrhythmias (ventricular fibrillation, ventricular
tachycardia, and fast ventricular tachycardia) with
defibrillation, cardioversion, and antitachycardia pacing
therapies
single or dual chamber pacing for patients requiring rate
•
support
diagnostics and monitors that evaluate the system and
•
assist in patient care
Leads – The device can be used with transvenous or epicardial
defibrillation leads. The lead system should consist of bipolar
or paired unipolar pacing/sensing leads in the right atrium and
right ventricle of the heart, a pacing lead for the left ventricle,
and 1 or 2 high-voltage cardioversion/defibrillation electrodes. In
addition to the lead system, the Active Can acts as one of the
high-voltage electrodes. The device delivers pacing and cardiac
resynchronization therapy via the atrial (A), right ventricular (RV),
and left ventricular (LV) leads. The device senses using the atrial
and RV leads. Cardioversion/defibrillation therapy is delivered
with 2 lead-based high-voltage electrodes, or with the Active Can
and 1 or 2 lead-based high-voltage electrodes.
INSYNC MAXIMO™7304Reference Manual
2The InSync Maximo system
System overview
Programmer and software – The Medtronic CareLink
Programmer, Model 2090,1and Model 9998 software application
allow you to perform the following functions:
configure the cardiac resynchronization, arrhythmia detection
•
and therapy, and bradycardia features for your patient
perform electrophysiological studies and system tests
•
monitor, display, or print patient cardiac activity information
•
For information about:
indications, contraindications, warnings and precautions,
•
see the implant manual that accompanies the device.
basic programmer and software desktop functions
•
that are not included in Chapter 10, “Using the
programmer”, page 217, see the manual accompanying
the programmer.
installing the Model 2067 or Model 2067L programming
•
head, see the manuals accompanying the programming
heads.
implanting leads and using implant tools, refer to the
•
manuals accompanying the leads and implant tools.
27
2.1.1Cardiac resynchronization
To improve cardiac output in patients with ventricular dysynchrony,
the device provides biventricular pacing. The device paces either
the right ventricle or both ventricles as programmed, unless
pacing is inhibited by a sensed event in the RV.
Ventricular pacing sequence, V-V pace delay and LV pacing
•
vector are programmable.
Optional CRT features promote sustained resynchronization
•
pacing that could be interrupted during episodes of
accelerated AV conduction, atrial rate excursions, PVCs, or
atrial arrhythmia.
Pacing amplitudes and pulse widths are selected
•
independently for each ventricle.
1
With the model 2067 programming head
INSYNC MAXIMO™7304Reference Manual
28
Chapter 2
System overview
2.1.2Detecting and treating tachyarrhythmias
The device monitors the cardiac rhythm for short ventricular
intervals that may indicate the presence of VF, VT, or FVT.
Upon detection of VF, the device delivers a biphasic
•
defibrillation shock of up to 35 J. If the VF episode persists,
up to 5 more individually programmed defibrillation shocks
can be delivered.
Upon detection of VT, the device delivers either a Ramp,
•
Ramp+, or Burst antitachycardia pacing therapy or a
biphasic cardioversion shock of up to 35 J synchronized to
a ventricular depolarization. If the VT episode persists, up
to 5 more individually programmed VT therapies can be
delivered. You can also program the device to monitor the
VT episode without delivering therapy.
Upon detection of FVT, the device delivers either a Ramp,
•
Ramp+, or Burst antitachycardia pacing therapy or a
biphasic cardioversion shock of up to 35 J synchronized to
a ventricular depolarization. If the FVT episode persists,
up to 5 more individually programmed FVT therapies can
be delivered.
You can program the device to distinguish between true
ventricular arrhythmias and rapidly conducted supraventricular
tachyarrhythmias (SVT) and withhold therapy for SVT.
SVT discrimination includes the capability to detect a double
tachycardia (an unrelated ventricular arrhythmia occurring
simultaneously with an SVT) so that therapy is not withheld for a
ventricular arrhythmia in the presence of an SVT.
2.1.3Treating bradycardia
The device provides rate responsive pacing to optimize
hemodynamics. An internal accelerometer senses the patient’s
physical activity, allowing the device to increase and decrease the
pacing rate in response to changes in the level of activity.
INSYNC MAXIMO™7304Reference Manual
2.1.4Monitoring for real-time and stored data
The device and programmer provide real-time information on
detection and therapy parameters and status during a patient
session. The device also provides accumulated data on
device operation, including ventricular sensing episodes, stored
electrograms, detected and treated tachyarrhythmia episodes,
bradycardia interventions, and the efficacy of therapy. The
Cardiac Compass report provides up to 14 months of clinically
significant data, including physical activity, heart rate, percent
pacing, arrhythmia episodes, and therapies delivered.
All of this information can be printed and retained in the patient’s
file or saved in electronic format on a floppy diskette.
2.1.5Conducting electrophysiologic tests
You can use the system to conduct non-invasive electrophysiologic
studies, including manual delivery of any of the device therapies
to manage an induced or spontaneous tachyarrhythmia.
2.1.6Alerting the patient to system events
2The InSync Maximo system
Indications and usage
29
You can use the programmable Patient Alert monitoring feature to
notify the patient with audible tones if certain conditions related to
the leads, battery, charge time, and therapies occur. The patient
can then respond based on your prescribed instructions.
2.2Indications and usage
The InSync Maximo is indicated for ventricular antitachycardia
pacing and ventricular defibrillation for automated treatment of life
threatening ventricular arrhythmias. The system is also indicated
for the reduction of the symptoms of moderate to severe heart
failure (NYHA Functional Class III or IV) in those patients who
remain symptomatic despite stable, optimal medical therapy and
have a left ventricular ejection fraction ≤ 35% and a prolonged
QRS duration.
INSYNC MAXIMO™7304Reference Manual
30
Chapter 2
Contraindications
2.3Contraindications
2.4Patient screening
Do not use the InSync Maximo system in:
Patients whose ventricular tachyarrythmias may have
•
transient or reversible causes, such as:
– acute myocardial infarction
– digitalis intoxication
– drowning
– electric shock
– electrolyte imbalance
– hypoxia
– sepsis
Patients with incessant VT or VF
•
Patients who have a unipolar pacemaker.
•
Before implant, patients should undergo a complete cardiac
evaluation, including electrophysiologic testing. Also,
electrophysiologic evaluation and testing of the safety and efficacy
of the proposed tachyarrhythmia therapies are recommended
during and after device implant.
Other optional screening procedures could include exercise
stress testing to determine the patient’s maximum sinus rate,
and cardiac catheterization to determine if there is a need for
concomitant surgery and/or medical therapy.
INSYNC MAXIMO™7304Reference Manual
3Emergency therapy
3.1 Delivering emergency therapies32
3.2 Delivering an emergency defibrillation therapy33
3.3 Delivering an emergency cardioversion therapy34
3.4 Delivering emergency fixed burst pacing34
3.5 Enabling emergency VVI pacing35
3
32
Chapter 3
Delivering emergency therapies
3.1Delivering emergency therapies
The device provides the following emergency therapies:
defibrillation
•
cardioversion
•
fixed burst pacing
•
emergency VVI pacing
•
The default emergency therapy is 35 J defibrillation. When you
select [Emergency] and [DELIVER], the device charges and
delivers a biphasic 35 J shock along the AX>B pathway.
The programmer resets the emergency defibrillation energy to
35 J each time you select [Emergency]. Emergency cardioversion
and fixed burst values remain as selected for the duration of the
session.
To return to other programming functions from an Emergency
screen, select [Exit Emergency].
1
3.1.1Effect on system operation
The device suspends the automatic detection features when
emergency defibrillation, cardioversion, or fixed burst pacing
therapies are delivered. Detection is not suspended during
emergency VVI pacing. Removing the programming head or
pressing [Resume] turns detection on again.
3.1.2Aborting an emergency therapy
As a safety precaution, the programmer also displays an [ABORT]
button, which immediately terminates any emergency therapy
in progress.
3.1.3Mechanical Emergency VVI button on the Medtronic CareLink
Model 2090 programmer
If you press the red Emergency VVI button on the programmer
display panel, the device initiates Emergency VVI pacing and the
programmer displays the Emergency screen.
1
If Active Can is turned off, the defibrillation is delivered between the HVX
and HVB electrodes.
INSYNC MAXIMO™7304Reference Manual
3.1.4Temporary parameter values
2
4
3
3Emergency therapy
Delivering an emergency defibrillation therapy
33
Emergency tachyarrhythmia therapies use temporary values
that do not change the programmed parameters of the device.
These values are not in effect until you select [DELIVER]. After
the tachyarrhythmia therapy is complete, the device reverts to
its programmed values.
3.2Delivering an emergency defibrillation therapy
The default emergency therapy is a full-energy defibrillation.
When you select [Emergency] and [DELIVER], the device charges
and delivers a biphasic full-energy shock. The programmer sets
the emergency defibrillation energy to its maximum value each
time you select [Emergency] or select the [Defibrillation] option
from an Emergency screen.
3.2.1How to deliver emergency defibrillation
1. Position the programming head
over the device.
2. Select [Emergency].
3. Accept the defibrillation energy
shown on the screen, or select
a new Energy value.
4. Select [DELIVER].
If delivery is not confirmed,
verify that the programming
head is properly positioned, and
select [Retry] or [Cancel].
2
2
Delivery of Emergency VVI Pacing changes the programmed bradycardia
pacing values to the emergency values (see Section 3.5, page 35).
INSYNC MAXIMO™7304Reference Manual
34
2
5
4
3
Chapter 3
Delivering an emergency cardioversion therapy
3.3Delivering an emergency cardioversion therapy
When you initiate an emergency cardioversion therapy, the device
charges its capacitors to the selected energy and attempts to
deliver therapy synchronized with a sensed tachyarrhythmia
event. If the cardioversion therapy cannot be synchronized, it is
aborted. See Section 7.3.5.6, page 136.
3.3.1How to deliver emergency cardioversion
1. Position the programming head
over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the defibrillation energy
shown on the screen, or select
a new Energy value.
5. Select [DELIVER].
If delivery is not confirmed,
verify that the programming
head is properly positioned and
select [Retry] or [Cancel].
3.4Delivering emergency fixed burst pacing
Emergency fixed burst pacing delivers maximum output pacing
pulses to the ventricle at a selectable interval. The therapy
continues for as long as you keep the programmer stylus on the
[BURST Press and Hold] button.
INSYNC MAXIMO™7304Reference Manual
Enabling emergency VVI pacing
3
4
2
5
3.4.1How to deliver emergency fixed burst pacing
1. Position the programming head
over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval
shown on the screen, or select
a new interval value.
5. Select [BURST Press and
Hold].
If delivery is not confirmed,
the programmer displays an
error window. Verify that the
programming head is properly
positioned. Select [OK] from the
window and reselect [BURST
Press and Hold].
3.5Enabling emergency VVI pacing
3Emergency therapy
35
Emergency VVI pacing programs the device to deliver high-output
ventricular pacing. You can initiate emergency VVI pacing
from the Emergency screen or by pressing the red mechanical
button on the Medtronic CareLink Model 2090 programmer
display panel. To disable emergency VVI pacing, reprogram the
bradycardia pacing parameters from the Parameters screen.
INSYNC MAXIMO™7304Reference Manual
36
3
4
2
Chapter 3
Enabling emergency VVI pacing
3.5.1How to deliver emergency VVI pacing
1. Position the programming head
over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A
successful programming sets
the device to the following
maximum output bradycardia
pacing values:
Pacing Mode: VVI
•
Lower Rate: 70 ppm
•
RV. Amplitude: 6 V
•
RV. Pulse Width: 1.6 ms
•
V. Pace Blanking: 240 ms
•
V. Pacing: RV
•
V. Sense Response: Off
•
Ventricular Rate
•
Stabilization: Off
Cond AF Response: Off
•
If programming is not confirmed,
verify that the programming
head is properly positioned and
select [Retry] or [Cancel].
INSYNC MAXIMO™7304Reference Manual
Device implant and patient follow-up
procedures
Part II
4Implanting the device
4.1 Overview40
4.2 Preparing for an implant40
4.3 Replacing a device42
4.4 Surgical approach43
4.5 Sensing and pacing measurements47
4.6 Connecting the leads to the device47
4.7 Testing defibrillation operation and effectiveness51
4.8 Positioning and securing the device54
4.9 Completing the implant procedure54
4
40
Chapter 4
Overview
4.1Overview
The tasks for implanting a device include:
1. Preparing for an implant
2. Replacing a device
3. Surgical approach
4. Sensing and pacing measurements
5. Connecting the leads to the device
6. Testing defibrillation operation and effectiveness
7. Positioning and securing the device
8. Completing the implant procedure
These tasks are described in the sections that follow.
4.2Preparing for an implant
Warning: Keep a backup external defibrillator available
during the implant for transthoracic rescue when arrhythmias
are induced.
Warning: Do not permit the patient to contact grounded
equipment that could produce hazardous leakage current
during implant. Resulting arrhythmia induction could result
in the patient’s death.
The device is intended for implant with Medtronic transvenous or
epicardial leads. No claims of safety and efficacy can be made
with regard to other acutely or chronically implanted lead systems
that are not manufactured by Medtronic.
4.2.1Equipment for an implant
Medtronic CareLink Programmer Model 2090 and Model
•
2067or 2067L programming head
Model 9998 software application
•
Model 8090 Analyzer lead analysis device or equivalent
•
pacing system analyzer
external defibrillator
•
INSYNC MAXIMO™7304Reference Manual
Model 5358 Defibrillator Implant Support Device and
•
software application (optional)
4.2.2Sterile supplies
InSync Maximo and lead system components
•
Programming head sleeve or programming head
•
Analyzer cables
•
Lead introducers or delivery systems appropriate for the
•
lead system
Extra stylets of appropriate length and shape
•
4.2.3How to set up the implant support instrument
When using an implant support instrument such as the Model 5358
Defibrillator Implant Support device:
1. Calibrate any monitoring or recording equipment while recording
the EGM and marker outputs of the support instrument.
2. Verify the high-energy output of the support instrument by delivering
a high-energy defibrillation shock into the test load.
4Implanting the device
Preparing for an implant
41
4.2.4How to set up the programmer and start the application
1. Set up the programmer as described in the instructions provided
with the programmer.
2. Install the InSync Maximo Model 9998 software on the programmer
if it is not already installed.
3. Place the programming head over the device, and start the
application. Select the device model or select [Auto identify].
Note: The programmer automatically interrogates the device when
the application starts.
INSYNC MAXIMO™7304Reference Manual
42
Chapter 4
Replacing a device
4.2.5How to preprogram the device
Before opening the sterile package, prepare the device for implant as
follows:
1. Check the “Use by” date printed on the package. Do not implant
the device after the “Use by” date because the battery’s longevity
can be reduced.
2. Interrogate the device and print a full summary report.
Note: If the programmer reports that an electrical reset occurred,
do not implant the device. Contact a Medtronic representative.
3. Confirm that the battery voltage is at least 3.0 V at room
temperature.
If the device temperature is lower than room temperature or has
delivered a recent high-voltage charge, the battery voltage will be
temporarily lower.
4. Set up data collection parameters and the device internal clock
(see Section 12.9.3, “How to set up data collection”, page 305).
5. Perform a manual capacitor formation (see “Manual capacitor
formation”, page 325).
6. Program the therapy and pacing parameters to values appropriate
for the patient. Ensure that all tachyarrhythmia detection is
programmed Off (see Section 8.1.4, “How to program CRT and
bradycardia pacing parameters”, page 151).
Note: Do not enable VF Detection until the device has been
successfully implanted.
1
4.3Replacing a device
If you are replacing a previously implanted device, turn off device
detection and therapies before explanting.
When implanting the device with a chronic lead system, perform
the following evaluations to ensure appropriate detection and
therapy:
Check the integrity of the chronic high-voltage leads with a
•
test shock, chest x-ray, and inspection.
Perform chronic pacing and sensing measurements.
•
Measure high-voltage lead impedances.
•
Test defibrillation efficacy.
•
1
Use the Quick Look screen to verify the voltage; see Section 11.2.2, “Quick
Look observations”, page 253.
INSYNC MAXIMO™7304Reference Manual
Confirm adequate sensing during VF.
•
Ensure proper fit of the lead connectors in the device
•
connector block.
Notes:
To meet the implant requirements, it may be necessary to
•
reposition or replace the chronic leads or to add a third
high-voltage electrode.
Any unused leads that remain implanted must be capped.
•
4.3.1How to explant and replace a device
1. Program all tachyarrhythmia detection Off.
2. Dissect the leads and the device free from the surrounding tissues
in the surgical pocket. Be careful not to nick or breach the lead
insulation during the process of exposing the system.
3. Loosen each setscrew, and gently retract the lead from the
connector block.
4. Remove the device from the surgical pocket.
5. If the connector pin of any implanted lead shows signs of pitting
or corrosion, replace the implanted lead with a new lead. The
damaged lead should be discarded and replaced to assure the
integrity of the device system.
6. Measure sensing, pacing, and defibrillation efficacy using the
replacement device or an implant support instrument.
4Implanting the device
Surgical approach
43
4.4Surgical approach
The InSync Maximo system typically requires a left ventricular
lead implant in the coronary sinus and positioned in a cardiac
vein. For this reason, physicians should expect that implanting an
InSync Maximo system will require more time than implanting a
traditional ICD or pacemaker system.
4.4.1Implanting the device
In most cases the device is implanted in the pectoral region,
either submuscularly or subcutaneously, using transvenous leads
for pacing, sensing, and high-voltage therapies.
INSYNC MAXIMO™7304Reference Manual
44
Chapter 4
Surgical approach
4.4.2Lead configurations
Follow these single-incision guidelines when implanting the
device in the pectoral region. The pocket should be about 1.5
times the size of the device.
Submuscular approach: A single incision that extends over
•
the delta-pectoral groove can provide access to the cephalic
and subclavian veins and access to form the implant pocket.
Place the device sufficiently medial to the humeral head to
avoid interference with shoulder motion.
Subcutaneous approach: The subcutaneous approach can
•
be similar to a pacemaker implant, using a single transverse
incision long enough to permit isolation of the cephalic and
subclavian veins. Place the device quite far medially to keep
the leads away from the axilla, and keep the upper edge of
the device inferior to the incision.
The device is typically implanted with the following leads:
one transvenous lead for the left ventricle (LV) for pacing
•
one bipolar transvenous lead in the right ventricle (RV) for
•
sensing, pacing, and cardioversion/defibrillation therapies
one bipolar transvenous lead in the atrium (A) for sensing
•
and pacing
The InSync Maximo is an Active Can system in which the device
case serves as a high-voltage electrode. The device case, the RV
coil, and the SVC coil form a 3 electrode system for simultaneous
mode defibrillation and cardioversion. High-voltage therapies are
delivered between the RV coil and the combined device case
and SVC coil.
The device may be implanted with other transvenous leads. Refer
to Section 4.4.4, “Using other transvenous leads”, page 46, for
more information.
Notes:
Medtronic 3.2 mm low-profile leads are not directly
•
compatible with the device IS-1 connector block.
For questions about lead and device compatibility, please
•
contact your Medtronic representative.
INSYNC MAXIMO™7304Reference Manual
4.4.3Implanting the lead
Note: Do not implant the LV, atrial, and RV leads in the same
venous access site. Medtronic recommends implanting the LV
lead in the subclavian vein, and the atrial and RV leads in the
cephalic vein.
If inserting using a subclavian approach, position the lead laterally
to avoid pinching the lead body between the clavicle and the first
rib. Pinching the lead may eventually cause conductor fracture,
damage to the insulation, or other damage to the lead. This may
result in complications such as loss of detection, loss of pacing
therapies, or loss of cardioversion/defibrillation therapy.
Certain anatomical abnormalities, such as thoracic outlet
syndrome, may also cause pinching and subsequent fracture
of the lead. If significant resistance is encountered during a
subclavian stick implant procedure, do not force passage of the
lead by adjusting the patient’s posture (by raising the arm or
putting a towel behind the person’s back, for example). Use an
alternate venous entry site instead.
4.4.3.1 Implanting the ventricular leads
4Implanting the device
Surgical approach
45
Warning: Backup pacing should be readily available during
implant. Use of the delivery system or leads may cause
heart block.
Note: Due to the variability of cardiac venous systems, the
venous anatomy should be assessed before implanting the LV
lead to determine an optimal LV lead position. Before placing a
lead in the coronary sinus, obtain a venogram.
When implanting a LV or RV lead, follow the directions for use in
the respective lead technical manuals.
Final lead positioning should attempt to optimize both cardiac
resynchronization and defibrillation threshold.
When pacing at 10 V using an external pacing device, test
for extracardiac stimulation from the LV lead. If extracardiac
stimulation is present, consider repositioning the lead.
4.4.3.2 Implanting the right atrial lead
Implant the atrial lead in the cephalic vein, if possible. Follow the
directions for use in the lead technical manual.
INSYNC MAXIMO™7304Reference Manual
46
Chapter 4
Surgical approach
4.4.3.3 Lead positioning
For the initial lead configuration the following positions are
recommended:
LV tip electrode – Position as far laterally into the coronary
venous system as possible (deflection of the LV EGM signal as
late as possible in the QRS complex).
RV tip electrode – Position as far apically as possible (deflection
of the RV EGM signal on the RV electrode should start early in
the QRS complex). Defibrillation threshold, atrial far-field sensing
and RV-LV timing must be considered when choosing the optimal
location.
Atrial tip electrode – The atrial tip electrode should be positioned
close to the SA node or on the lateral wall to minimize the
possibility of far-field R-wave oversensing. Lead placement in
these positions potentially provides optimal atrial sensing while
reducing far-field R-wave sensing.
4.4.4Using other transvenous leads
Follow the general guidelines below for initial positioning of
other transvenous leads (the final positions are determined by
defibrillation efficacy tests):
For an SVC lead, place the lead tip high in the innominate
•
vein, approximately 5 cm proximal to the atrium (A) and SVC
junction.
For an SQ patch lead, place the patch along the left
•
midaxillary line, centered over the fourth-to-fifth intercostal
space.
4.4.5Epicardial lead systems
The following guidelines should be observed:
The distance between the RVring electrode and the RVtip
•
electrode should not exceed 1 cm.
The RVtip electrode should not lie in a direct path between
•
epicardial patches.
The RVtip/RVring and Atip/Aring intrinsic, post-pace, and
•
post shock electrograms should be thoroughly evaluated for
oversensing, undersensing, cross-talk, and noise.
INSYNC MAXIMO™7304Reference Manual
4Implanting the device
Sensing and pacing measurements
The pacing thresholds and electrogram characteristics
•
should meet the values indicated in Table 4-1.
4.5Sensing and pacing measurements
Sensing and pacing measurements include EGM amplitudes,
slew rates, capture thresholds, and pacing lead impedances.
Medtronic recommends using the Model 8090 Analyzer for
sensing and pacing measurements. Refer to the Analyzer
technical manual for detailed procedures on performing these
measurements.
Do not use the intracardiac EGM telemetered from the device
to assess sensing.
Table 4-1 lists the acceptable implant values for acute and
chronically implanted lead systems.
Table 4-1. Acceptable implant values
Measurement required
EGM amplitude (during NSR)
Slew rate
Capture threshold (0.5 ms pulse
a
width)
Typical pacing lead impedance
a
Verify adequate pacing threshold margins at implant and at each followup visit.
b
The measured pacing lead impedance is a reflection of measuring equipment and lead technology.
c
Please refer to specific lead technical manuals for acceptable impedance values.
Acute
A
≥ 2 mV
≥
0.5 V/s
≤ 1.5 V
b
250 - 1000 Ùc(all)
RV
≥ 5 mV
≥
0.75 V/s
≤ 1.0 V
LV
≥ 3 mV
—
≤ 3.0 V
Chronic
A
≥ 1 mV
≥
0.3 V/s
≤ 3.0 V
RV
≥ 3 mV
≥
0.5 V/s
≤ 3.0 V
47
LV
≥ 1 mV
—
≤ 4.0 V
4.6Connecting the leads to the device
For more detailed information about lead/connector compatibility,
see the
Medtronic representative.
InSync Maximo 7304 Implant Manual
INSYNC MAXIMO™7304Reference Manual
, or contact your
48
2
5
4
6
3
1
Chapter 4
Connecting the leads to the device
Table 4-2. Lead connections
Device portConnector typeSoftware name
SVC (HVX)DF-1SVC
RV (HVB)DF-1RVcoil
Can
RVIS-1 bipolarRVtip and RVring
LVIS-1 bipolarLVtip and LVring
AIS-1 bipolarAtip and Aring
Figure 4-1. Lead connections
—
Can
1 SVC (HVX)
2 RV (HVB)
3 Can (HVA)
4 LV
5 RV
6 A
Caution: Loose lead connections may result in inappropriate
sensing and failure to deliver necessary arrhythmia therapy.
INSYNC MAXIMO™7304Reference Manual
4Implanting the device
Connecting the leads to the device
Cautions:
To ensure appropriate sensing and detection, insert the
•
right ventricular lead IS-1 connector into the RV port and
insert the right ventricular lead DF-1 connector into the
RV (HVB) port.
If no SVC electrode is implanted, the pin plug provided
•
with the device must be secured in the SVC (HVX) port.
Use only the torque wrench supplied with the device.
•
It is designed to prevent damage to the device from
overtightening a setscrew.
Do not reuse the torque wrench after this implant.
•
Lead and Active Can electrodes in electrical contact
•
with each other during a high-voltage therapy could
cause current to bypass the heart, possibly damaging
the device and leads. While the device is connected to
the leads, make sure that no therapeutic electrodes,
stylets, or guide wires are touching or connected by an
accessory low impedance conductive pathway. Move
any objects that contain conductive materials (e.g., an
implanted guide wire) well away from all electrodes
before a high-voltage shock is delivered.
49
Notes:
For easier lead insertion, insert the LV IS-1 lead before the
•
other leads.
If a chronic lead system uses a different size or number of
•
leads, insert the lead connectors into appropriate adaptors.
See the appropriate lead technical manuals for more details.
Upon initial lead connection to the connector block, LVtip,
•
RVtip, and RVcoil all need to be connected within the
connector block to provide biventricular pacing.
INSYNC MAXIMO™7304Reference Manual
50
a
b
1
2
Chapter 4
Connecting the leads to the device
4.6.1How to connect the lead to the device
1. Insert the torque wrench into
the appropriate setscrew.
a. If the port is obstructed by
the setscrew, retract the
setscrew to clear it. Take
care not to disengage the
setscrew from the connector
block.
b. Leave the torque wrench in
the setscrew until the lead
is secure. This allows a
pathway for venting trapped
air when the lead is inserted.
2. Push the lead or plug into the
connector port until the lead
pin is clearly visible in the pin
viewing area. No sealant is
required, but sterile water may
be used as a lubricant.
3. Tighten the setscrew by turning
clockwise until the torque
wrench clicks.
4. Tug gently on the lead to
confirm a secure fit. Do not tug
on the lead until all setscrews
have been tightened.
5. Repeat these steps for each
connector port.
4.6.2Post pace oversensing testing
Changes in the programmed ventricular pacing configuration (RV,
RV→LV, or LV→RV) may evoke slightly different EGMs. To ensure
proper sensing regardless of the pacing configuration, or in the
INSYNC MAXIMO™7304Reference Manual
event of lead dislodgment or failure, the sensing behavior during
pacing for all configurations should be evaluated as follows:
1. Program RV Pacing.
2. Print a 10 to 30 s test strip.
3. Check for post pace oversensing.
Repeat Step 1 through Step 3 with ventricular pacing programmed
to RV→LV and LV→RV.
4Implanting the device
Testing defibrillation operation and effectiveness
If oversensing occurs during one of the pacing configurations,
repositioning of the RV lead may help alleviate the oversensing.
4.7Testing defibrillation operation and effectiveness
Demonstrate reliable defibrillation effectiveness with the implanted
lead system by using your preferred method to establish that a
10 J (minimum) safety margin exists.
Note: If the 10 J (minimum) safety margin cannot be ensured,
see Section 15.5, “Solving tachyarrhythmia therapy problems”,
page 350.
4.7.1High-voltage implant values
Measured values must meet the following requirements at implant.
Table 4-3. High-voltage therapy values at implant
MeasurementAcute or Chronic Leads
V. Defib impedance
SVC (HVX) impedance (if applicable)
Defibrillation threshold≤ 25 J
20 – 200 Ù
20 – 200 Ù
51
Warning: Ensure that an external defibrillator is charged for
a rescue shock.
INSYNC MAXIMO™7304Reference Manual
52
Chapter 4
Testing defibrillation operation and effectiveness
4.7.2How to prepare for defibrillation threshold testing
1. Place the programming head over the device, start a patient
session, and interrogate the device, if you have not already done so.
2. Observe the Marker Channel telemetry annotations and the
programmer ECG display to verify that the device is sensing
properly.
3. Conduct a manual Lead Impedance Test2to verify the defibrillation
lead connections. Perform this test with the device in the surgical
pocket, and keep the pocket very moist. If the impedance is out of
range, perform one or more of the following tasks:
Recheck lead connections and electrode placement.
•
Repeat the measurement.
•
Inspect the bipolar EGM for abnormalities.
•
Measure the defibrillation impedance with a manual test shock.
•
4. Program the device or support instrument to properly detect VF
with an adequate safety margin (1.2 mV sensitivity).
2
See Section 13.4, “Measuring lead impedance”, page 320 .
INSYNC MAXIMO™7304Reference Manual
Testing defibrillation operation and effectiveness
1
2
3
4
5
6
7
8
9
10
11
12
13
14
4.7.3How to perform defibrillation threshold testing
4Implanting the device
1. Select Tests > EP Study.
2. Select either 50 Hz BURST or
T-shock induction.
3. Select [Resume at BURST] or
[Resume at DELIVER].
4. Select [Adjust Permanent…].
5. Program VF Enable On.
6. Program the automatic therapy
energy settings. Therapies 2-6
should be set to the maximum
energy.
7. Select [Program].
8. Select [Close].
9. If performing a T-Shock
induction, select the [Enable]
checkbox.
10. Select [DELIVER], or [50 Hz
BURST Press and Hold].
If necessary, you can abort an
induction or therapy in progress
by pressing [ABORT].
11. Observe the live rhythm monitor
for proper post-shock sensing.
12. Use the [Adjust Permanent…]
button to program the next
appropriate energy level.
13. Wait until the on-screen timer
reaches 5 min, then repeat
Step 9 through Step 12 as
needed.
14. Select Params > Detection
and program VF, FVT, and VT
detection Off before closing.
53
INSYNC MAXIMO™7304Reference Manual
54
Chapter 4
Positioning and securing the device
4.8Positioning and securing the device
Cautions:
If no SVC electrode is implanted, the pin plug provided
•
with the device must be secured in the SVC (HVX) port.
Program tachyarrhythmia detection Off before closing.
•
4.8.1How to position and secure the device
1. Ensure that each lead pin or
plug is fully inserted into the
connector block and that all
setscrews are tight.
2. Coil any excess lead length
beneath the device. Avoid kinks
in the lead conductors.
3. Implant the device within
5 cm of the skin and with the
Medtronic label facing the skin
for improved Patient Alert
volume. This position optimizes
the ambulatory monitoring
operations.
4. Use nonabsorbable sutures to
secure the device within the
pocket to minimize post-implant
rotation and migration of the
device. Use a surgical needle to
penetrate the suture holes.
4.9Completing the implant procedure
After implanting the device, x-ray the patient to verify device and
lead placement. To complete programming the device, select
parameters that are appropriate for the patient.
INSYNC MAXIMO™7304Reference Manual
Completing the implant procedure
4.9.1How to complete programming the device
1. After closing the pocket, program detection On. Program ventricular
tachyarrhythmia therapies On as desired.
2. Do not enable the Other 1:1 SVTs PR Logic detection criterion until
the atrial lead has matured (approximately 1 month post implant).
3. If external equipment was used to conduct the defibrillation efficacy
tests, perform a final VF induction, and allow the implanted system
to detect and treat the arrhythmia.
4. Monitor the patient after the implant, and take x-rays as soon as
possible to document and assess the location of the leads.
5. Program patient information. See Section 12.10, “Viewing and
entering patient information”, page 308.
6. Configure the Patient Alert feature. See Section 11.3, “Using the
Patient Alert feature”, page 254.
7. Set up data collection parameters. See Section 12.9, “Setting up
data collection”, page 302.
8. Interrogate the device after any spontaneous episodes to evaluate
the detection and therapy parameter settings.
9. If the patient has not experienced spontaneous episodes, you
may induce the clinical tachyarrhythmias using the non-invasive
EP Study features to further assess the performance of the
system. See Chapter 14, “Conducting Electrophysiologic Studies”,
page 327.
10. Recheck pacing and sensing values, and adjust if necessary. See
Section 13.3, “Measuring pacing thresholds”, page 317.
4Implanting the device
55
INSYNC MAXIMO™7304Reference Manual
5Conducting a patient follow-up session
5.1 Patient follow-up guidelines58
5.2 Verifying the status of the implanted system59
5.5 Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias62
5
58
Chapter 5
Patient follow-up guidelines
5.1Patient follow-up guidelines
Schedule regular patient follow-up sessions to monitor the
condition of the device and leads and to verify that the device is
configured appropriately for your patient.
During the first few months after receiving a new device, the
patient may require close monitoring. Schedule an office visit
at least every 3 months.
The Quick Look screen, which is displayed after you interrogate
the device, provides a good beginning for the follow-up review.
Using this screen you can perform the following tasks:
Verify that the device is functioning correctly.
•
Verify that the device is delivering biventricular pacing most
•
or all of the time.
Review the clinical performance and long term trends.
•
Print appropriate reports to compare the results to the
•
patient’s history and to retain for future reference.
For details about reviewing clinical trend data, see
Section 12.4, “Viewing the Heart Failure Management
•
report”, page 278
Section 12.5, “Viewing clinical trends in the Cardiac
•
Compass report”, page 282
Note: The Checklist feature displays a list of standard follow-up
tasks on the programmer screen for easy reference during the
follow-up session. You can also customize your own checklists
if you wish. See Section 11.4, “Streamlining follow-ups with
Checklist”, page 262, for more information.
INSYNC MAXIMO™7304Reference Manual
5Conducting a patient follow-up session
Verifying the status of the implanted system
5.2Verifying the status of the implanted system
To verify that the device and leads are functioning correctly,
review the following information from the Quick Look screen and
perform followup tests as indicated:
Review the displayed battery voltage for comparison to the
•
Elective Replacement Indicator value (see Section 1.4,
“Replacement indicators”, page 20). Remember that battery
voltage may be low if high-voltage charging has occurred
within 24 hours.
Review the last full-energy charge.
•
– For information about adjusting the capacitor formation
interval, see Section 9.1, “Optimizing charge time”,
page 210.
– If the programmer displays an Excessive Charge Time
ERI, the device should be replaced immediately.
Review the defibrillation and pacing lead impedance values
•
for inappropriate values or large changes since the last
followup. See Section 13.4, “Measuring lead impedance”,
page 320.
Perform an EGM Amplitude test in the RV and atrium for
•
comparison to previous EGM amplitude measurements.
See Section 13.5.5, “How to perform an EGM Amplitude
test”, page 324.
To review longer term trends in sensing and impedance
•
measurements, select the [>>] button from the lead
performance area of the Quick Look screen. The
programmer displays a detailed history of automatic sensing
and impedance measurements. See Section 11.2, “Taking a
quick look at device activity”, page 253.
To verify that the device is providing cardiac resynchronization
therapy appropriately, review the following information from the
initial interrogation report or Quick Look screen and investigate as
indicated:
Confirm that the patient is receiving adequate cardiac
•
support for daily living activities.
INSYNC MAXIMO™7304Reference Manual
60
Chapter 5
Verifying effective basic pacing
5.3.1Considerations
Review Quick Look Observations related to ventricular
•
pacing percentage or ventricular sensing episodes.
Check the stored Ventricular Sensing Episode records for
•
appropriate interventions and continuity of resynchronization
pacing. See Section 12.7, “Viewing episode data”, page 291.
Check the rate histograms for more information on atrial and
•
ventricular pacing in general, and ventricular rates during
AT/AF episodes.
Review the Heart Failure Management and Cardiac
•
Compass reports for comparison to patient history (see
Section 12.4, “Viewing the Heart Failure Management
report”, page 278).
Review the following information before assessing
resynchronization therapy.
Undersensing, far-field oversensing, or loss of capture
•
are basic pacing issues that can affect delivery of cardiac
resynchronization therapy. These situations can often
be resolved with basic programming changes and then
monitored for further occurrences.
If CRT delivery is lost during AF or during atrial rate
•
excursions, consider the following programming options:
– Perform basic programming changes to increase the
Upper Tracking Rate or decrease the total atrial refractory
period (for information on total atrial refractory period,
see Section 8.2.4, “Programming considerations for atrial
rates”, page 162).
– Enable Ventricular Sense Response, Atrial Tracking
Recovery, or Conducted AF Response to promote more
continuous delivery of CRT.
5.4Verifying effective basic pacing
To verify that the device is sensing and pacing appropriately,
review the following information from the Initial Interrogation
Report, and Quick Look screen, and investigate as indicated:
Confirm that the patient is receiving adequate cardiac
•
support for daily living activities.
INSYNC MAXIMO™7304Reference Manual
•
•
•
•
5.4.1Considerations
Review the following information before verifying bradycardia
pacing.
5Conducting a patient follow-up session
Verifying effective basic pacing
Review the pacing conduction history and rate histograms
for comparison to the patient history. A sharp increase
in the atrial pacing percentage may indicate a need for
investigation and analysis.
Review the recorded Mode Switch episodes for comparison
to the patient’s atrial arrhythmia history (see Section 12.7,
“Viewing episode data”, page 291). A dramatic increase in
frequency or duration of atrial episodes may indicate a need
for investigation and analysis.
Review the Cardiac Compass report for comparison to
patient history (see Section 12.5, “Viewing clinical trends in
the Cardiac Compass report”, page 282).
Conduct pacing threshold tests (see Section 13.3,
“Measuring pacing thresholds”, page 317) to verify that
the programmed pacing outputs provide a sufficient safety
margin.
61
Atrial pacing – If the conduction history or rate histograms show
a predominance of atrial pacing despite a healthy sinus response,
consider these options to decrease the atrial pacing burden:
Decrease the Lower Rate.
•
Decrease the rate response or increase the activity
•
threshold.
Conduction history – If the reported percentages in the
conduction history do not add up to 100, the percentages may
be rounded.
INSYNC MAXIMO™7304Reference Manual
62
Chapter 5
Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias
5.5Verifying accurate detection and appropriate therapy
for ventricular tachyarrhythmias
To verify that the device is providing effective tachyarrhythmia
detection and therapy, review the following information from the
Quick Look screen and investigate as indicated:
Review Quick Look Observations that relate to patient
•
history and device operation. To display more detailed
information about any observation, select the observation
and then select the [>>] button.
Review any Patient Alerts listed in the Observations of the
•
Quick Look screen. For the most detailed information about
Patient Alerts, select Patient Alert from the Data icon and
select [Events].
Check stored episode records for appropriate sensing
•
and detection of arrhythmias. See Section 12.7, “Viewing
episode data”, page 291.
Check stored SVT episode records for appropriate
•
identification of SVTs.
5.5.1Considerations
Review the following information before verifying detection and
therapy.
Flashback memory – In addition to the episode text and stored
electrograms, use Flashback memory and interval plots to help
investigate the accuracy and specificity of ventricular detection.
Episode misidentification – If the episode records indicate that
false detections have occurred, the Sensing Integrity counter may
help in determining the prevalence of oversensing. For more
information, see Section 12.2.4.3, “Sensing integrity counter”,
page 274.
If the device is oversensing, consider these programming options:
Increase the Pace Blanking value.
•
Increase the sensitivity threshold.
•
Caution: Do not reprogram the device to decrease
oversensing without assuring that appropriate sensing is
maintained. See Section 6.2, “Setting up sensing”, page 71.
INSYNC MAXIMO™7304Reference Manual
5Conducting a patient follow-up session
Verifying accurate detection and appropriate therapy for ventricular tachyarrhythmias
If the episode records reveal that a stable monomorphic VT has
been identified and treated as VF, consider these options to
improve the detection accuracy:
Review the Interval Plot for the episode, and adjust the VF
•
Interval, if necessary. Use caution when reprogramming the
VF Interval, because changes to this value can adversely
affect VF detection.
Consider enabling FVT via VF detection. See Section 6.5,
•
“Detecting FVT episodes”, page 85.
If the SVT episode records include episodes of true VT, review
the SVT episode record to identify the SVT detection criterion that
withheld detection. Adjust the SVT detection criteria parameters
as necessary. See Section 6.8, “Enhancing detection with PR
Logic criteria”, page 96, and Section 6.9, “Enhancing VT detection
with the Stability criterion”, page 104.
63
INSYNC MAXIMO™7304Reference Manual
Configuring the device for the patient
Part III
6Detecting tachyarrhythmias
6.1 Detection overview68
6.2 Setting up sensing71
6.3 Detecting VF episodes76
6.4 Detecting VT episodes79
6.5 Detecting FVT episodes85
6.6 Detecting tachyarrhythmia episodes with Combined Count90
6.7 Monitoring episodes for termination or redetection92
6.8 Enhancing detection with PR Logic criteria96
6.9 Enhancing VT detection with the Stability criterion104
6.10 Detecting double tachycardias106
6.11 Detecting prolonged tachyarrhythmias with High Rate Timeout107
6
68
Chapter 6
Detection overview
6.1Detection overview
The device detects ventricular tachyarrhythmias (VF, VT, and
FVT) by comparing the time intervals between sensed ventricular
events to a set of programmable detection intervals. If enough
intervals occur that are shorter than the programmed intervals,
the device detects a tachyarrhythmia and responds automatically
with a programmed therapy. After delivering the therapy, the
device either redetects the arrhythmia and delivers the next
programmed therapy or detects episode termination.
To avoid detecting rapidly conducted SVTs (for example, sinus
tachycardia or atrial fibrillation) as ventricular tachyarrhythmias,
the device provides several detection enhancements, including
PR Logic and Stability detection criteria.
All of the detection features interact during initial detection (see
Figure 6-1). During redetection, the device does not apply the PR
Logic detection criteria.
Note: Detection functions can be turned off by programming the
VF Enable, FVT Enable, and VT Enable parameters to Off. For
an example, see Section 6.3.4, “How to program VF detection”,
page 78.
INSYNC MAXIMO™7304Reference Manual
6Detecting tachyarrhythmias
Ventricular Event
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
Yes
Yes
Update counts and pattern information
Is the interval in the VF, FVT, or VT detection zone?
Has High Rate Timeout suspended detection enhancements?
Does Stability reset the VT event count? (VT and FVT via VT
detection only)
Has a tachyarrhythmia event count reached an NID?
Are one or more PR Logic criteria on?
No/suspended by High Rate
Timeout
Tachy
Episode
Detected
Is the median ventricular interval less than the SVT Limit?
Is there a double tachycardia in progress?
Are one or more PR Logic criteria withholding detection?
Figure 6-1. How detection features interact during initial detection
69
Detection overview
6.1.1Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops
classifying and counting tachyarrhythmia intervals. Sensing
and programmed pacing remain active, and the programmed
detection settings are not modified.
INSYNC MAXIMO™7304Reference Manual
70
Chapter 6
Detection overview
Detection is suspended:
when the device senses the presence of a strong magnet.
•
The programmer head contains a magnet which suspends
detection, but once telemetry between the device and
programmer is established, detection resumes.
while performing any of the manual system tests, including
•
Underlying Rhythm, Pacing Threshold, Lead Impedance,
EGM Amplitude, and Charge/Dump. Detection automatically
resumes once the test is complete.
while performing a T-Shock, 50 Hz Burst, Manual Burst,
•
or PES Induction. You can choose to have the device
automatically resume detection after delivering the induction.
when you deliver a Manual or Emergency therapy. You
•
can resume detection by selecting the [Resume] button or
removing the programming head from the device.
when you select the on-screen [Suspend] button. You can
•
resume detection by selecting the [Resume] button or by
removing the programming head from the device.
during the automatic daily lead impedance measurements.
•
Detection resumes when the measurements are complete.
while the device is delivering an automatic tachyarrhythmia
•
therapy (including capacitor charging for defibrillation and
cardioversion). However, the device does continue to
confirm the detected episode during charging. Detection
resumes when the therapy is complete.
Note: The device suspends VT detection and Combined
Count detection (see Section 6.6, “Detecting tachyarrhythmia
episodes with Combined Count”, page 90) for 17 events
following a defibrillation therapy delivered in response to
a detected VF.
during charging for Automatic Capacitor Formation.
•
Detection resumes when charging is complete.
1
1
If the defibrillation therapy is delivered as a result of either a High Rate
Timeout Therapy operation or Progressive Episode Therapies operation,
VT detection is not suspended (see Section 6.11.4.1, “High Rate Timeout
Therapy”, page 109).
INSYNC MAXIMO™7304Reference Manual
6.2Setting up sensing
The device provides bipolar sensing in both the atrium and right
ventricle using the sensing electrodes of the implanted atrial
and RV leads. You can adjust the sensitivity to intracardiac
signals using independent atrial and ventricular sensitivity
settings. These settings define the minimum electrical amplitude
recognized by the device as an atrial or ventricular sensed event.
Proper sensing is essential for the safe and effective use of the
device. To provide appropriate sensing, the device uses:
auto-adjusting atrial and ventricular sensing thresholds
•
short (30 ms) cross-chamber blanking after paced events
•
no cross-chamber blanking after sensed events
•
See Section 6.2.4, “Details about sensing”, page 74.
6.2.1Parameters
6Detecting tachyarrhythmias
Setting up sensing
71
V. Sensitivity – Minimum amplitude of
electrical signal that registers as a sensed
ventricular event.
A. Sensitivity – Minimum amplitude of
electrical signal that registers as a sensed
atrial event.
6.2.2Considerations
Review the following information before programming sensing
parameters.
Dual chamber sensing and pacing modes – The device senses
in both the atrium and the right ventricle at all times, except when
the programmed bradycardia pacing mode is DOO. When the
pacing mode is programmed to DOO or VOO mode, there is no
sensing in the ventricle.
mode, you must first disable detection.
Sensitivity thresholds – The programmed atrial and ventricular
sensitivity thresholds apply to all features related to sensing,
including detection and bradycardia pacing.
2
Atrial sensing occurs in VOO mode.
0.15; 0.3
0.6; 0.9; 1.2 mV
0.15; 0.3
0.6; 0.9; 1.2; 1.5;
2.1 mV
2
In order to program either DOO or VOO
; 0.45;
; 0.45;
INSYNC MAXIMO™7304Reference Manual
72
Chapter 6
Setting up sensing
Bradycardia pacing and sensing – A combination of high
pacing pulse width or high amplitude with a low sensitivity
threshold may cause inappropriate sensing across chambers or
in the same chamber. Programming a lower pulse width, lower
amplitude, longer pace blanking, or a higher sensitivity threshold
may eliminate this inappropriate sensing.
Recommended ventricular sensitivity threshold – A
ventricular sensitivity threshold of 0.3 mV is recommended to
maximize the probability of detecting VF and to limit the possibility
of oversensing and cross-chamber sensing.
High ventricular sensitivity threshold – Setting V. Sensitivity
to a value greater than 0.6 mV is not recommended except for
testing. Doing this may cause undersensing, which can cause
any of the following situations:
delayed or aborted cardioversion therapy
•
delayed defibrillation therapy (when VF confirmation is
•
active)
asynchronous pacing
•
underdetection of tachyarrhythmias
•
Low ventricular sensitivity threshold – If you set V. Sensitivity
to its most sensitive value of 0.15 mV, the device will be more
susceptible to EMI, cross-chamber sensing, and oversensing.
Recommended atrial sensitivity threshold – An atrial sensitivity
threshold of 0.3 mV is recommended to optimize the effectiveness
of PR Logic detection criteria and atrial pacing operations, while
limiting the possibility of oversensing and cross-chamber sensing.
High atrial sensitivity threshold – If you set the A. Sensitivity
value too high, the device may not provide reliable sensing of
P-waves during SVTs and sinus rhythm.
Low atrial sensitivity threshold – If you set the A. Sensitivity
value to its most sensitive value of 0.15 mV, the device will
be more susceptible to EMI, far-field R-wave sensing, and
oversensing.
Testing sensitivity after reprogramming – If you change the
ventricular sensitivity threshold, evaluate for proper sensing and
detection by inducing VF and allowing the device to automatically
detect and treat the arrhythmia.
INSYNC MAXIMO™7304Reference Manual
Atrial pacing and ventricular sensing – If you program the
1
2
3
device to an atrial pacing mode, make sure that it does not sense
atrial pacing pulses as ventricular events.
Sensing during VF – Always verify that the device senses
properly during VF. If the device is not sensing or detecting
properly, program detection and therapies off, and evaluate the
system (making sure to monitor the patient for life-threatening
arrhythmias until you enable detection and therapies again). You
may need to reposition or replace the ventricular sensing lead to
achieve proper sensing.
Atrial lead selection – Atrial leads with minimal tip-to-ring
spacing may reduce far-field R-wave sensing.
Repositioning the atrial lead – You may need to reposition
or replace the atrial sensing lead if reprogramming the atrial
sensitivity threshold does not provide reliable atrial sensing during
SVTs and sinus rhythm.
6.2.3How to program sensitivity
6Detecting tachyarrhythmias
Setting up sensing
1. Select Params > Detection.
2. Select the desired A. Sensitivity
and V. Sensitivity parameters.
3. Select [PROGRAM].
73
INSYNC MAXIMO™7304Reference Manual
74
A
P
V
S
A
S
V
S
A
S
V
P
Sensitivity Threshold
Rectified and
Filtered A. EGM
Rectified and
Filtered RV EGM
Marker Channel
1
2
3
4
5
Chapter 6
Setting up sensing
6.2.4Details about sensing
6.2.4.1 Auto-adjusting sensitivity thresholds
The device automatically adjusts the sensitivity thresholds after
certain paced and sensed events to help reduce oversensing
from T-waves, cross-chamber events, and pacing. You can see
how sensitivity thresholds are adjusted after different types of
events (see Figure 6-2).
Figure 6-2. Auto-adjusting sensitivity thresholds
1 After an atrial sensed event, the atrial sensitivity threshold increases
to 75% of the EGM peak (maximum: 8x the programmed value,
decay constant: 200 ms).
2 After a ventricular sensed event, the ventricular sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed
value, decay constant: 450 ms).
3 After an atrial paced event, the device does not adjust the atrial
sensitivity threshold. The ventricular sensitivity threshold increases
by 0.45 mV (decay constant: 60 ms).
4 After a ventricular paced event, the atrial sensitivity threshold
increases to 4x the programmed value (maximum: 1.8 mV,
immediate return after 60 ms).
5 After the ventricular pace blanking period is finished, the ventricular
threshold increases to 4.5x the programmed value (maximum:
1.8 mV, decay constant: 450 ms).
3
The exponential decay continues through a subsequent ventricular pacing
pulse and its blanking period.
INSYNC MAXIMO™7304Reference Manual
3
4
4
6Detecting tachyarrhythmias
Setting up sensing
6.2.4.2 Blanking periods
During a blanking period, the device does not sense
electrical signals. This helps prevent sensing of device
pacing, cardioversion and defibrillation pulses, post-pacing
depolarization, T-waves, and multiple sensing of the same event.
The blanking periods following paced events are longer than
those following sensed events to avoid sensing the atrial and
ventricular depolarizations.
Notes:
To enhance sensing and detection during tachyarrhythmias,
•
the device does not apply cross-chamber blanking (blank
sensing in the opposite chamber) after a sensed event.
Atrial sensing is still active during the Post-Ventricular
•
Atrial Blanking (PVAB) period (see Section 8.2.3.5,
“Post-Ventricular Atrial Refractory Period”, page 160).
The duration of the fixed blanking periods is shown in Table 6-1.
For information on programmable pace blanking periods, see
Section 8.1, “Providing basic pacing therapy”, page 146.
Table 6-1. Duration of fixed blanking periods
Cross-chamber blanking after atrial or ventricular
pacing pulse
Atrial blanking after sensed atrial event
Ventricular blanking after sensed ventricular event
Atrial and ventricular blanking after delivered
cardioversion or defibrillation therapy
30 ms
100 ms
120 ms
520 ms
75
6.2.4.3 Refractory periods
During a refractory period, the device senses normally, but
classifies sensed events as refractory and limits its response to
these events. Pacing refractory periods prevent inappropriately
sensed signals, such as far-field R-waves (ventricular events
sensed in the atrium) or electrical noise, from triggering certain
pacing timing intervals.
4
If the programmed sensitivity value exceeds 0.3 mV (ventricular) or 1.2 mV
(atrial), the threshold is not adjusted.
INSYNC MAXIMO™7304Reference Manual
76
Chapter 6
Detecting VF episodes
Synchronization refractory periods help prevent the device
from delivering cardioversion and defibrillation therapies at
inappropriate times. See Section 7.1.5.5, “Synchronizing
defibrillation without confirming VF”, page 116, and
Section 7.3.5.6, “Synchronizing cardioversion after charging”,
page 136.
Note: Refractory periods do not affect tachyarrhythmia detection.
6.3Detecting VF episodes
The device detects VF episodes by examining the cardiac rhythm
for short ventricular intervals. If a predetermined number of
intervals occurs that are short enough to be considered VF
events, the device detects VF and delivers the first programmed
VF therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing.
See Section 6.3.5, “Details about VF detection”, page 78.
6.3.1Parameters
VF Detection Enable – Turns VF
detection on or off.
VF Interval (Rate) – V-V intervals
shorter than this value are counted as
VF events.
VF Initial NID – Number of Intervals to
Detect: number of VF events the device
must count to detect a VF episode.
VF Redetect NID – Number of Intervals
to Redetect: number of VF events
the device must count to redetect a
continuing VF after a therapy.
6.3.2Considerations
Review the following information before programming VF
Detection parameters.
VF Interval minimum setting – To ensure proper VF detection,
you should not program the VF Interval to a value less than
300 ms.
VF Interval maximum setting – Programming the VF Interval to
a value greater than 350 ms may cause inappropriate detection of
rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals
shorter than the VF Interval are counted using the VF event
counter, which is more sensitive than the consecutive VT event
counter.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the Initial
NIDs.
Enabling VF detection – When VF Detection Enable is
programmed On for the first time, the device:
enables Automatic Capacitor Formation
•
starts recording Cardiac Compass data and Heart Failure
•
Management data
starts recording lead performance trends (starting at
•
3:00 AM by the device clock)
clears all brady pacing counters
•
VF detection and PR Logic criteria – You can program the
device to exclude rapidly conducted SVTs from VF Detection by
enabling the PR Logic detection criteria. Note that the SVT Limit
must be programmed shorter than the VF Interval in order for
the PR Logic criteria to affect VF detection. See Section 6.8,
“Enhancing detection with PR Logic criteria”, page 96.
77
6.3.3Restrictions
Double tachycardia detection – When any PR Logic detection
criteria are enabled, the device also enables double tachycardia
detection (VF, VT, or FVT in the presence of an SVT). See
Section 6.10, “Detecting double tachycardias”, page 106.
Review the following information before programming VF
Detection parameters.
Tachyarrhythmia detection and pacing – To ensure reliable
ventricular tachyarrhythmia detection, the programmer regulates
the values available for pacing and tachyarrhythmia detection.
See Section B.11, “Parameter interlocks”, page 388.
INSYNC MAXIMO™7304Reference Manual
78
1
2
3
Chapter 6
Detecting VF episodes
VF detection backup – To ensure VF Detection backup during
VT and FVT episodes, if VF Detection is off, both VT Detection
and FVT Detection must also be off.
6.3.4How to program VF detection
1. Select Params > Detection.
2. Select the desired values for
VF Enable, VF Initial NID, VF
Redetect NID, and VF Interval.
3. Select [PROGRAM].
6.3.5Details about VF detection
The device detects VF by counting the number of VF events,
which are V-V intervals shorter than the programmed VF Interval.
On each event, the device counts the number of recent VF
events. The number of recent events that are examined is called
the VF detection window. The size of the VF detection window is
the second number in the programmed VF NID (for example, 24
events if the VF Initial NID is 18/24).
The threshold for detecting VF is the first number in the
programmed VF NID (for example, 18 events if the VF Initial
NID is 18/24). This threshold is always 75% of the VF detection
window. That is, if 75% of the events in the VF detection window
are VF events, the device detects a VF episode (see Figure 6-3).
After the device detects VF, it delivers the first programmed
VF therapy. Following the therapy, if the number of VF events
reaches the programmed VF Redetect NID, the device redetects
INSYNC MAXIMO™7304Reference Manual
VF and delivers the next programmed VF therapy.
Note: The device can also detect VF Episodes using the
Combined Count detection criterion (see Section 6.6, page 90).
Figure 6-3. Device detects VF
A
S
A
S
V
S
V
S
V
S
V
S
V
S
V
S
V
S
A
R
A
R
A
R
A
R
A
R
1 2 3 4 5 66 7 8 9 10 11 12 13 1415 16 17 18
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
D
200 ms
2
3
1
VF Event Count
ECG
Marker Channel
VF Interval
1 VF starts, and the device begins counting VF events (intervals less
than the programmed VF Interval).
2 A ventricular interval occurs outside the VF detection zone. The VF
event count is not incremented.
3 The VF event count reaches the programmed VF NID value of 18
events out of 24, and the device detects VF.
6Detecting tachyarrhythmias
Detecting VT episodes
79
6.4Detecting VT episodes
The device detects VT episodes by examining the cardiac rhythm
for short ventricular intervals. If enough intervals occur that are
short enough to be considered VT events (but are not VF or FVT
events), the device detects VT and delivers the first programmed
VT therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing.
You can program the device to detect and record VT episodes
without treating them with VT therapies by setting VT Detection
Enable to Monitor. If a patient’s VT episodes are well-tolerated,
this feature allows you to collect data about these episodes
without delivering therapy or affecting VF detection.
See Section 6.4.5, “Details about VT detection”, page 81.
INSYNC MAXIMO™7304Reference Manual
80
Chapter 6
Detecting VT episodes
6.4.1Parameters
VT Detection Enable – Turns VT
detection on or off, or enables VT
monitoring.
VT Interval (Rate) – V-V intervals
shorter than this value are counted as
VT events.
VT Initial NID – Number of Intervals
to Detect: number of VT events the
device must count to detect a VT
episode.
VT Redetect NID – Number of
Intervals to Redetect: number of
VT events the device must count
to redetect a continuing VT after a
therapy.
6.4.2Considerations
Review the following information before programming VT
Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the Initial
NIDs.
On; Off
280; 290 … 400
12; 16
4; 8; 12
; or Monitor
…52; 76; 100
… 52
… 600 ms
VT Detection Enable, AFib/AFlutter, and Sinus Tach – When
you set VT Detection Enable to On or Monitor, the AFib/AFlutter
and Sinus Tach parameters are also automatically set to On.
VT detection and Combined Count detection – When VT
Detection is On, the device applies the Combined Count detection
criterion to help speed detection of rhythms that fluctuate between
detection zones. Combined Count detection is disabled if VT
Detection is set to Off or Monitor. See Section 6.6, “Detecting
tachyarrhythmia episodes with Combined Count”, page 90.
VT detection and rapidly conducted SVTs – You can program
the device to exclude rapidly conducted SVTs from VT detection
by enabling the PR Logic or Stability detection criteria. See
Section 6.9, “Enhancing VT detection with the Stability criterion”,
page 104, and Section 6.8, “Enhancing detection with PR Logic
criteria”, page 96.
INSYNC MAXIMO™7304Reference Manual
Double tachycardia detection – When any PR Logic detection
1
2
3
criteria are enabled, the device also enables double tachycardia
detection (VF, VT, or FVT in the presence of an SVT; see
Section 6.10, page 106).
6.4.3Restrictions
Review the following information before programming VT
detection parameters.
Tachyarrhythmia detection and pacing – To ensure reliable
ventricular tachyarrhythmia detection, the programmer regulates
the values available for pacing and tachyarrhythmia detection.
See Section B.11, “Parameter interlocks”, page 388.
VF detection backup – To ensure VF detection backup during
VT and FVT episodes, if VF Detection is off, both VT Detection
and FVT Detection must also be off.
6.4.4How to program VT detection
6Detecting tachyarrhythmias
Detecting VT episodes
1. Select Params > Detection.
2. Select the desired values for
VT Enable, VT Initial NID, VT
Redetect NID, and VT Interval.
3. Select [PROGRAM].
81
6.4.5Details about VT detection
The device detects VT by counting the number of consecutive VT
events. A VT event is a V-V interval shorter than the programmed
VT Interval but greater than or equal to the VF Interval. If the
number of consecutive VT events reaches the programmed VT
Initial NID, the device detects VT (see Figure 6-4).
INSYNC MAXIMO™7304Reference Manual
82
Chapter 6
Detecting VT episodes
The VT event count resets to zero whenever an interval occurs
that is greater than or equal to the programmed VT Interval. The
count remains at the current value if an interval is shorter than
the programmed VF Interval.
After the device detects VT, it delivers the first programmed VT
therapy. Following the therapy, if the VT event counter reaches
the VT Redetect NID, the device redetects VT and delivers the
next programmed therapy.
Note: The device can also detect VT Episodes using the
Combined Count detection criterion (see Section 6.6, page 90).
INSYNC MAXIMO™7304Reference Manual
Figure 6-4. Device detects VT
A
S
A
S
V
S
V
S
V
S
V
S
A
R
A
R
A
R
A
R
A
R
A
R
A
R
10 12 34 56
T
S
T
S
T
S
T
S
T
S
T
S
T
S
V
S
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
A
R
78 910 11 12 13 14 15 16
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
D
2
1
3
VT Event Count
ECG
Marker Channel
VT Interval
VT Event Count
ECG
Marker Channel
VT Interval
6Detecting tachyarrhythmias
Detecting VT episodes
83
1 VT starts, and the device begins counting VT events (intervals less
than the programmed VT Interval but greater than or equal to the
VF Interval).
2 A ventricular interval occurs outside VT detection zone. The VT event
count resets to zero.
3 The VT event count reaches the programmed VT NID of 16 events,
and the device detects VT.
6.4.5.1 VT monitoring
You can program the device to record VT episodes without
delivering VT therapy by setting VT Detection to Monitor. When
VT monitoring is enabled, the device detects VT episodes but
does not deliver VT therapy (see Figure 6-5). Instead, it records
VT episodes, labeling them as “monitored,” and waits for episode
termination to occur.
When VT Detection is set to Monitor, several detection operations
work differently.
INSYNC MAXIMO™7304Reference Manual
84
Chapter 6
Detecting VT episodes
VT event counting – Before the device detects an episode, it
counts VT events normally. However, once the VT Initial NID is
reached, the device sets the VT event count to zero and suspends
VT event counting for the rest of the episode.
VF and FVT detection – VF and FVT detection operate as if
VT detection is off. Specifically, Combined Count detection
is disabled, and FVT via VT detection is not selectable. If a
monitored VT episode accelerates into the FVT or VF detection
zone, the device applies the VF Initial NID to detect the new
tachyarrhythmia. Once an episode is in progress, VT event
counting doesn’t resume until the episode ends.
Caution: Programming the VF Interval greater than 350 ms
may result in inappropriate detection of rapidly conducted
atrial fibrillation as VF or FVT via VF. Intervals shorter than
the VF Interval are counted using the VF event counter,
which is more sensitive than the consecutive VT event
counter.
PR Logic and Stability criteria – Before the device detects a
tachyarrhythmia episode, the PR Logic and Stability criteria, if
turned on, are applied. If a monitored VT episode accelerates
into the FVT or VF detection zone, the device continues to apply
PR Logic criteria as initial VF or FVT detection begins. However,
because the Stability feature does not affect VF detection or FVT
via VF detection, it is not applied.
Episode termination – The device compares ventricular intervals
to the VT Interval to identify when a VT monitored episode has
ended. However, if a VF episode or FVT via VF episode occurs
when VT monitoring is enabled, the device compares ventricular
intervals to the VF Interval to identify episode termination.
INSYNC MAXIMO™7304Reference Manual
Figure 6-5. Device detects and monitors VT
200 ms
1234
A
S
A
R
A RARA
R
V
S
V ST ST ST ST
S
13141516000
A
R
A
R
A RA RA RA RA
R
T ST ST ST DT ST ST
S
2
3
1
VT Event Count
ECG
Marker Channel
VT Interval
1 VT starts, and the device begins counting VT events (intervals less
than the programmed VT Interval but greater than or equal to the
VF Interval).
2 The VT event count reaches the programmed VT NID of 16 events,
and the device detects VT.
3 After detecting the VT episode, the device resets the VT event count
to zero and monitors the episode until termination.
6Detecting tachyarrhythmias
Detecting FVT episodes
85
6.5Detecting FVT episodes
The device detects episodes of Fast Ventricular Tachycardia
(FVT) by examining the cardiac rhythm for short ventricular
intervals. If enough intervals occur in the programmed FVT
detection zone, the device detects FVT and delivers the first
programmed FVT therapy. After therapy, the device continues
to evaluate the ventricular rhythm to determine if the episode
is ongoing. To make sure it delivers sufficiently aggressive
therapies, the device can merge the programmed detection zones
during redetection to increase sensitivity.
See Section 6.5.5, “Details about FVT detection”, page 88.
INSYNC MAXIMO™7304Reference Manual
86
Chapter 6
Detecting FVT episodes
6.5.1Parameters
FVT Detection Enable – Enables FVT
detection via the VF or the VT detection
algorithm.
FVT Interval (Rate) – V-V intervals between
this value and the programmed VF Interval are
marked as FVT events.
6.5.2Considerations
Review the following information before programming FVT
Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the Initial
NIDs.
FVT detection enable – Your choice for an appropriate setting
for FVT Detection should depend on the patient’s VF and VT
cycle lengths. After determining a reliably sensitive VF Interval,
consider the following suggestions:
•
; via VF; or via
Off
VT
200; 210 … 600 ms
If the patient presents with a clinical VT interval in the VF
zone, select via VF to ensure reliable detection of VF. (VT
Detection need not be enabled at all.)
If the patient presents with 2 clinical VTs, both outside the
•
VF zone, select via VT to allow for correct classification of
the faster VT and to offer a separate therapy regimen for
each VT.
If the patient presents with only 1 clinical VT, which is outside
•
the VF zone, select VF and VT Detection only, and set FVT
Enable to Off.
FVT detection and PR Logic criteria – You can program the
device to exclude rapidly conducted SVTs from FVT Detection
by enabling the PR Logic detection criteria. Note that the SVT
Limit must be programmed shorter than the VF Interval for the PR
Logic criteria to affect FVT via VF detection.
INSYNC MAXIMO™7304Reference Manual
6.5.3Restrictions
6Detecting tachyarrhythmias
Detecting FVT episodes
Double tachycardia detection – When any PR Logic detection
criterion is enabled, the device also enables double tachycardia
detection (VF, VT, or FVT in the presence of an SVT, see
Section 6.10, page 106).
Review the following information before programming FVT
Detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To
ensure reliable ventricular tachyarrhythmia detection, the
programmer regulates the values available for bradycardia pacing
and tachyarrhythmia detection. See Section B.11, “Parameter
interlocks”, page 388.
VF detection backup – To ensure VF Detection backup during
VT and FVT episodes, VT and FVT Detection cannot be on
unless VF Detection is also on.
FVT detection – To ensure reliable ventricular tachyarrhythmia
detection, the programmer regulates the values available for the
FVT parameter as follows:
VT Detection must be set to On if FVT Detection is set to
•
via VT.
87
If FVT Detection is set to via VF, the FVT Interval must be
•
programmed to a value shorter than the VF Interval.
If FVT Detection is set to via VT, the FVT Interval must be
•
programmed to a value greater than the VF Interval and less
than or equal to the VT Interval.
INSYNC MAXIMO™7304Reference Manual
88
1
2
3
Chapter 6
Detecting FVT episodes
6.5.4How to program FVT detection
6.5.5Details about FVT detection
You can program the device to detect FVT episodes via the VF
or VT detection zone and NID.
1. Select Params > Detection.
2. Select the desired values for
FVT Enable and FVT Interval.
3. Select [PROGRAM].
When FVT Detection is set to via VF, a V-V interval within the FVT
detection zone is marked as an “FVT via VF” event. When the VF
NID is reached, the device reviews the last 8 intervals:
If any of the last 8 intervals are in the VF zone, it detects
•
the episode as VF.
If all of the last 8 intervals are outside the VF zone, it detects
•
the episode as FVT (see Figure 6-6).
When FVT Detection is set to via VT, a V-V interval within the
FVT detection zone is marked as an “FVT via VT” event. When
the VT NID is reached, the device reviews the last 8 intervals:
If any of the last 8 intervals are in the VF or FVT zones, it
•
detects the episode as FVT.
If all of the last 8 intervals are outside the FVT and VF zones,
•
it detects the episode as VT.
Note: The device can also detect FVT episodes via the Combined
INSYNC MAXIMO™7304Reference Manual
Count detection criterion (see Section 6.6, page 90).
Figure 6-6. Device detects FVT via VF
200 ms
13 14 15 16 17 18
A RA RA RA RA RA RA RA RA
R
T FT
F
T
F
T
F
T
F
T
F
T FV
S
112345
A
S
A
S
A RA RA
R
A
R
V
S
V
S
V
S
T
S
T
F
T
F
T
F
T
F
T F
1
2
3
VF Event Count
ECG
Marker Channel
VF and FVT Intervals
1 A fast ventricular tachycardia starts, and the first event falls into the
FVT detection zone.
2 The second event of the FVT episode has an interval that falls into
the VT zone. The VF event count is not incremented.
3 The device detects FVT after the VF event count reaches the VF
Initial NID.
6Detecting tachyarrhythmias
Detecting FVT episodes
89
6.5.5.1 Zone merging after detection
To ensure that the device delivers sufficiently aggressive
therapies during an extended or highly variable tachyarrhythmia
episode, the device merges detection zones during redetection
in some instances, as shown in Figure 6-7. The merged zone
configuration uses the event counting and therapies for the faster
arrhythmia and remains in effect until episode termination.
INSYNC MAXIMO™7304Reference Manual
90
Before detection
After VF detection
After FVT detection
FVT set to “via VF”
VF and FVT zones merge, leaving a
larger VF zone.
All zones remain unchanged.
FVT set to “via VT”
VT and FVT zones merge, leaving a
larger FVT zone.
VT and FVT zones merge, leaving a
larger FVT zone.
Detection Intervals: VF Interval: 320 ms, FVT Interval: 280 ms / 360 ms, VT
Interval: 400 ms
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
VF
FVT
VT
Chapter 6
Detecting tachyarrhythmia episodes with Combined Count
Figure 6-7. FVT zone merging
6.6Detecting tachyarrhythmia episodes with Combined
Count
INSYNC MAXIMO™7304Reference Manual
Because the device counts VF and VT events separately,
rhythms with variable cycle lengths can cause both event counts
to increment during an episode. To prevent these rhythms
from delaying detection, the device automatically enables the
Combined Count detection criterion if both VF and VT detection
are programmed On.
The Combined Count criterion compares the sum of the VF
and VT event counts to the Combined Number of Intervals to
Detect (CNID), which the device calculates automatically from
the programmed VF NID values. If the CNID is met, the device
reviews the recent intervals to determine if the episode should
be treated as a VF, FVT, or VT episode. The Combined Count
criterion applies during both initial detection and redetection.
Detecting tachyarrhythmia episodes with Combined Count
6.6.1Details about Combined Count detection
The Combined Count detection algorithm expedites detection
or redetection of ventricular tachyarrhythmias with ventricular
intervals that fluctuate between the VF and VT detection zones.
When VT detection is on, the device applies Combined Count
detection, which tracks the combined number of VT and VF
events counted. If this sum reaches the Combined Number of
Intervals to Detect (CNID), the device detects VF, FVT, or VT.
Combined Count detection also applies to redetected episodes.
Note: Combined Count detection is off when VT detection is set
to Monitor or Off.
If the VF event counter reaches 6, the device automatically
applies the Combined Number of Intervals to Detect (CNID). The
CNID is calculated by multiplying the current VF NID (Initial or
Redetect) by 7/6 and rounding down. Table 6-2 shows the CNID
values that correspond to each VF NID value.
Table 6-2. CNID values for each initial or redetect VF NID value
VF NIDCNIDVF NIDCNID
6/8
9/12
12/16
18/24
7
10
14
21
6Detecting tachyarrhythmias
21/28
24/32
27/36
30/40
24
28
31
35
91
Combined Count detection is fulfilled when the sum of the VF and
VT event counts equals or exceeds the CNID. The device then
reviews the last 8 intervals and classifies the episode as one of
the following:
VF, if any of the last 8 were in the VF zone
•
FVT, if FVT Detection is enabled and none of the last 8 was
•
in the VF zone, but one or more was in the FVT zone
VT, if all 8 were outside the VF zone (and FVT zone, if FVT
Monitoring episodes for termination or redetection
Figure 6-8. Device detects VF with the Combined Count criterion
1 A slow VF episode starts with a ventricular cycle length that varies
between the VF and VT detection zones.
2 When a VT event occurs, the device increments the VT event count
and the Combined Count.
3 The device detects VF even though the VF event count hasn’t yet
reached the VF Initial NID (18/24 in this example). The Combined
Count reaches the CNID value of 21 first.
6.7Monitoring episodes for termination or redetection
Once the device detects an arrhythmia, it considers the episode
INSYNC MAXIMO™7304Reference Manual
ongoing until it detects that the episode has ended. After
delivering therapy, it monitors the ventricular rhythm using the
programmed Redetect NIDs. If one of these NIDs is met, the
device delivers the next programmed therapy for the detected
arrhythmia.
See Section 6.7.4, “Details about episode termination and
redetection”, page 93.
6.7.1Parameters
1
2
3
6Detecting tachyarrhythmias
Monitoring episodes for termination or redetection
93
VF Redetect NID – Number of Intervals
to Redetect: number of VF events
the device must count to redetect a
continuing VF after a therapy.
VT Redetect NID – Number of Intervals
to Redetect: number of VT events
the device must count to redetect a
continuing VT after a therapy.
6.7.2Considerations
Review the following information before programming redetection
parameters.
Initial and Redetect NIDs – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the Initial
NIDs.
6.7.3How to program redetection parameters
6/8; 9/12; 12/16
21/28; 24/32; 27/36; 30/40
4; 8; 12
1. Select Params > Detection.
2. Select the desired values for VT
Redetect NID and VF Redetect
NID.
3. Select [PROGRAM].
… 52
; 18/24;
6.7.4Details about episode termination and redetection
After a therapy is delivered, the device evaluates the ventricular
rhythm to determine if the episode has terminated, is continuing,
or has changed to a different arrhythmia.
INSYNC MAXIMO™7304Reference Manual
94
Chapter 6
Monitoring episodes for termination or redetection
6.7.4.1 Episode termination
The device determines that the episode has terminated if one of
the following conditions occurs:
8 consecutive ventricular intervals are greater than or equal
•
to the programmed VT interval.
20 s elapse with no ventricular intervals shorter than the
•
programmed VT interval.
After antitachycardia pacing therapy, the device begins evaluating
intervals for episode termination on the first ventricular cycle.
After cardioversion or defibrillation, the device begins evaluating
intervals for episode termination on the second ventricular event.
(Due to the extended post shock blanking, this event may be the
third event on the electrogram.)
Note: Any subsequent detection after the end of the episode
marks the start of a new episode.
6.7.4.2 Episode redetection
After the device detects a tachyarrhythmia episode and delivers
a therapy, the device redetects an arrhythmia if the VF or VT
event count reaches the Redetect NID or if the combined VF and
VT event count reaches the Redetect CNID (see Section 6.6,
“Detecting tachyarrhythmia episodes with Combined Count”,
page 90).
5
5
The device then delivers the next programmed therapy for the
current arrhythmia and resumes monitoring for the outcome of
that therapy (see Figure 6-9).
5
VF interval if VT Detection is set to Off or Monitor and the episode is a VF
or an FVT via VF episode.
INSYNC MAXIMO™7304Reference Manual
Monitoring episodes for termination or redetection
A RA SA SA SA SA SA RA RA RA RA RA RA RA RA RA RA RA RA
R
T
P
T PT
P
T PT DT PT PT PT PT PT ST ST ST ST ST ST ST ST ST ST ST
D
200 ms
016 012345678910 1112
3
2
1
VT Event Count
ECG
Marker Channel
VT Interval
Figure 6-9. VT episode redetected after therapy
1 A VT episode is detected, and the device delivers a Burst ATP therapy.
2 After therapy, the device continues to detect events in the VT zone.
3 When the VT event count reaches the VT Redetect NID, the device
redetects the VT.
Notes:
The device suspends VT detection (and Combined Count
•
detection) for 17 events following a defibrillation therapy
delivered in response to a detected VF.
detection helps avoid detecting transient VTs that can follow
high-voltage therapies.
The PR Logic criteria are not applied during redetection.
•
However, the Stability criterion may withhold detection or
redetection of VT (or FVT via VT) throughout an episode.
6Detecting tachyarrhythmias
6
Suspending VT
95
6.7.4.3 VT acceleration
If the device redetects VT, it classifies the rhythm as accelerated
if the average of the 4 intervals before redetection is at least
60 ms less than the average of the 4 intervals before initial VT
detection. The most recent interval average is used to identify VT
acceleration if VT is redetected again during the episode.
6
If the defibrillation therapy is delivered as a result of either a High Rate Timeout
Therapy operation or Progressive Episode Therapies operation, VT detection
is not suspended (see Section 6.11.4.1, page 109).
INSYNC MAXIMO™7304Reference Manual
96
Chapter 6
Enhancing detection with PR Logic criteria
If the device redetects VF or an accelerated VT after an
antitachycardia pacing sequence delivery, it skips the subsequent
pacing therapy sequences for the duration of the episode and
delivers the next therapy programmed for the current arrhythmia.
6.8Enhancing detection with PR Logic criteria
The PR Logic detection criteria are designed to withhold
inappropriate ventricular detection during episodes of rapidly
conducted supraventricular tachycardia (SVT). The device
analyzes the activation patterns and timing in both chambers
using PR Logic pattern and rate analysis. This information
helps identify evidence of atrial fibrillation, atrial flutter, sinus
tachycardia, and other 1:1 SVTs. If this analysis indicates the
presence of one or more of these rhythms, the device withholds
detection.
For more information, see
Section 6.8.5, “Details about PR Logic pattern and rate
•
analysis”, page 99
Section 6.8.6, “Details about the PR Logic detection
•
criteria”, page 103
6.8.1Parameters
AFib/AFlutter – Identifies rapidly conducted
atrial fibrillation, atrial flutter, or atrial
tachycardia
Sinus Tach – Identifies sinus tachycardia
1:1 VT-ST Boundary – Allows customizing
of the Sinus Tach criterion for patients with
slow one-to-one conduction by identifying
the threshold between the retrograde and
antegrade zones used by A-V pattern
analysis.
Other 1:1 SVTs – Identifies other one-to-one
SVTs where the atrial and ventricular
activation are roughly simultaneous
SVT Limit – Defines the minimum ventricular
interval at which the device applies the PR
Logic criteria
a
When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and
Sinus Tach parameters are also automatically set to On.
INSYNC MAXIMO™7304Reference Manual
Ona; Off
Ona; Off
35; 50
On; Off
240; 250 … 320
330; 340 … 600 ms
; 66; 75; 85%
;
6.8.2Considerations
Review the following information before programming PR Logic
parameters.
PR Logic criteria and double tachycardia detection – When
any of the PR Logic criteria are enabled, the device also enables
double tachycardia detection (VF, VT, or FVT in the presence of
an SVT; see Section 6.10.1, page 106).
6Detecting tachyarrhythmias
Enhancing detection with PR Logic criteria
Cautions:
Before enabling the Other 1:1 SVTs criterion, ensure
•
that the atrial lead has matured. This criterion could
inappropriately withhold therapy if atrial sensing is
compromised by an unstable or dislodged atrial lead.
Use caution when programming the Other 1:1 SVTs
•
criterion in patients who exhibit slow 1:1 retrograde
conduction during VF or VT. This criterion could
inappropriately withhold VF/VT therapy in such patients.
See Section 6.8.5.1, “Pattern analysis of A-V and V-A
intervals”, page 100.
Use caution when programming the 1:1 VT-ST
•
Boundary. Incorrect programming of this parameter can
result in inappropriate therapies or underdetection of
tachyarrhythmias.
97
Sinus Tach and 1:1 VT-ST Boundary – When enabling the
Sinus Tach criterion in patients who exhibit slow 1:1 retrograde
conduction during VF or VT or slow antegrade conduction during
SVTs, make sure to set the 1:1 VT-ST Boundary parameter
appropriately. See Section 6.8.2.1, “Customizing PR Logic for
patients with slow conduction”, page 98.
VT Detection Enable, AFib/AFlutter and Sinus Tach – When
you set VT Detection Enable to On or Monitor, the AFib/AFlutter
and Sinus Tach parameters also automatically set to On.
SVT Limit – To ensure that therapy is delivered for
hemodynamically compromising rates of any origin, the device
always delivers therapy when the median ventricular interval is
less than the programmed SVT Limit (nominally 320 ms) if VT,
VF, or FVT detection criteria are satisfied.
VF Interval and SVT Limit – If you program an SVT Limit greater
than the VF Interval, you are effectively disabling the PR Logic
criteria for VF detection.
INSYNC MAXIMO™7304Reference Manual
98
Chapter 6
Enhancing detection with PR Logic criteria
6.8.2.1 Customizing PR Logic for patients with slow
conduction
Slow antegrade conduction during an SVT or slow 1:1 retrograde
conduction during VT can adversely affect the accuracy of the
Sinus Tach and Other 1:1 SVTs criteria. The 1:1 VT-ST Boundary
parameter allows you to customize the Sinus Tach criterion for
patients who exhibit slow conduction.
Note: Changing the 1:1 VT-ST Boundary parameter does not
affect the operation of the Other 1:1 SVTs criterion. Use caution
when enabling this criterion in patients who exhibit slow antegrade
or retrograde conduction.
The 1:1 VT-ST Boundary parameter represents an A-V interval
that is a percentage of a V-V interval. This value separates
the retrograde and antegrade zones used by PR Logic pattern
analysis operations. See Section 6.8.5.1, “Pattern analysis of A-V
and V-A intervals”, page 100.
If slow retrograde or antegrade conduction causes events to
occur in the incorrect zone, you can use the 1:1 VT-ST Boundary
parameter to increase the size of the appropriate zone. See
Figure 6-10, page 100.
If a patient exhibits slow antegrade or retrograde conduction and
could benefit from a different 1:1 VT-ST Boundary setting, choose
a new setting as follows:
If the patient exhibits long V-A intervals during VT with 1:1
•
retrograde conduction, select 35%.
If the patient exhibits long A-V intervals during an SVT,
•
select a value that exceeds the A-V/V-V ratio observed
during stored VT/VF or SVT episodes. See Section 12.7.3.1,
page 297.
6.8.3Restrictions
Review the following information before programming PR Logic
parameters.
Sinus Tach and 1:1 VT-ST Boundary – The Sinus Tach criterion
must be on before 1:1 VT-ST Boundary can be selected.
VT Detection Enable and SVT Limit – The SVT Limit must be
shorter than the VT Interval (or VF Interval if VT Detection is off).
INSYNC MAXIMO™7304Reference Manual
6Detecting tachyarrhythmias
1
2
3
4
5
6
Enhancing detection with PR Logic criteria
6.8.4How to program the PR Logic detection criteria
1. Select Params > Detection.
2. Select the desired values for
AFib/AFlutter, Sinus Tach,
Other 1:1 SVTs, and SVT Limit.
3. If you need to adjust the 1:1
VT-ST Boundary for your
patient, select Additional
Settings….
4. Select the desired value for the
1:1 VT-ST Boundary.
5. Select [OK].
6. Select [PROGRAM].
99
6.8.5Details about PR Logic pattern and rate analysis
PR Logic pattern and rate analysis is based on the following
aspects of atrial and ventricular activation:
A-V and V-A interval patterns
•
atrial and ventricular rate
•
AF evidence
•
far-field R-wave sensing
•
A:V dissociation
•
V-V regularity
•
The information collected by PR Logic pattern and rate analysis is
used by the PR Logic detection criteria to identify the presence of
SVTs and withhold detection.
INSYNC MAXIMO™7304Reference Manual
100
85%
75%
66%
35%
1
2
3
4
5
6
Chapter 6
Enhancing detection with PR Logic criteria
6.8.5.1 Pattern analysis of A-V and V-A intervals
The device uses pattern analysis to identify sinus tachycardia,
atrial flutter, and other 1:1 SVTs. Within each V-V interval, the
device categorizes the atrial rhythm according to the number
of intervening atrial events and the zones in which those atrial
events occur (see Figure 6-10). From this information, the device
assigns pattern codes to the intervals, and interprets the pattern
codes to identify SVTs.
Figure 6-10. Zones used during A-V pattern analysis
1 Junctional Zone: 50 ms. Atrial events in the first junctional zone
indicate PAC, PVC, junctional rhythms, atrial fibrillation, or atrial flutter.
2 Retrograde Zone. Atrial events in the retrograde zone indicate
retrograde conduction of ventricular events.
3 The 1:1 VT-ST Boundary separates the retrograde and antegrade
zones. The nominal value of this boundary is 50%.
4 Antegrade (Normal) Zone. Atrial events in the antegrade zone
INSYNC MAXIMO™7304Reference Manual
indicate normal conduction (sinus rhythm, sinus tachycardia).
5 Junctional Zone: 80 ms. Atrial events in the second junctional zone
indicate PAC, PVC, junctional rhythms, atrial fibrillation, or atrial flutter.
6 The 1:1 VT-ST Boundary can be programmed to different values,
changing the relative size of the antegrade and retrograde zones.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.