Medtronic 7288 Reference Guide

INTRINSIC™/INTRINSIC
30
7288/7287
Dual Chamber Implantable Cardioverter Defibrillator (VVE-DDDR)
Reference Manual
INTRINSIC™/INTRINSIC
30
7288/7287
Reference Manual
A guide to the operation and programming of the Intrinsic Model 7288 and Intrinsic 30 Model 7287 Dual Chamber Implantable Cardioverter Defibrillators
The following are trademarks of Medtronic:
Active Can, Cardiac Compass, Checklist, Flashback, Intrinsic, Leadless ECG, MVP, Marker Channel, Medtronic, Medtronic CareLink, PR Logic, Patient Alert, Quick Look, QuickLink
Table of Contents
Introduction 11
Part I Quick overview 15
1 Quick reference 17
1.1 New to Intrinsic 18
1.2 Physical characteristics 18
1.3 Magnet application 20
1.4 Longevity projections 21
1.5 Replacement indicators 23
1.6 Typical charge times 23
1.7 High voltage therapy energy 24
1.8 Stored data and diagnostics 26
2 The Intrinsic system 29
2.1 System overview 30
2.2 Indications and usage 33
2.3 Contraindications 33
2.4 Patient screening 33
3 Emergency therapy 35
3.1 Delivering emergency therapies 36
3.2 Delivering an emergency defibrillation therapy 37
3.3 Delivering an emergency cardioversion therapy 38
3.4 Delivering emergency fixed burst pacing 39
3.5 Enabling emergency VVI pacing 40
Part II Device implant and patient follow-up
procedures 43
4 Implanting the device 45
4.1 Overview 46
4.2 Preparing for an implant 46
4.3 Replacing a device 48
4.4 Positioning the leads 49
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
6
Table of Contents
Part III Configuring the device for the patient 67
4.5 Testing sensing and pacing thresholds 51
4.6 Connecting the leads to the device 52
4.7 Testing defibrillation operation and effectiveness 54
4.8 Positioning and securing the device 58
4.9 Completing the implant procedure 58
5 Conducting a patient follow-up session 61
5.1 Patient follow-up guidelines 62
5.2 Verifying the status of the implanted system 62
5.3 Verifying accurate detection and appropriate therapy 63
5.4 Verifying effective bradycardia pacing 64
6 Detecting tachyarrhythmias 69
6.1 Detection overview 70
6.2 Setting up sensing 73
6.3 Detecting VF episodes 78
6.4 Detecting VT episodes 81
6.5 Detecting FVT episodes 87
6.6 Detecting tachyarrhythmia episodes with Combined Count 92
6.7 Monitoring episodes for termination or redetection 94
6.8 Enhancing detection with PR Logic criteria 98
6.9 Enhancing VT detection with the Stability criterion 106
6.10 Detecting double tachycardias 108
6.11 Detecting prolonged tachyarrhythmias with High Rate Timeout 109
7 Treating tachyarrhythmia episodes 113
7.1 Treating VF with defibrillation 114
7.2 Treating VT and FVT with antitachycardia pacing 123
7.3 Treating VT and FVT with cardioversion 133
7.4 Optimizing therapy with Smart Mode and Progressive Episode Therapies 141
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
Table of Contents
8 Treating bradycardia 147
8.1 Providing basic pacing therapy 148
8.2 Dual chamber pacing 152
8.3 Single chamber pacing 162
8.4 MVP (Managed Ventricular Pacing) 165
8.5 Enhancing pacing for optimal cardiac output 170
8.6 Adjusting the pacing rate with Mode Switch 184
8.7 Preventing competitive atrial pacing 186
8.8 Detecting and preventing pacemaker-mediated tachycardia 192
8.9 Providing Ventricular Safety Pacing 196
8.10 Providing pacing after high voltage therapies 200
9 Optimizing charge time and device longevity 203
9.1 Optimizing charge time 204
9.2 Optimizing device longevity 207
Part IV Evaluating and managing patient treatment 209
7
10 Using the programmer 211
10.1 Setting up and using the programmer 212
10.2 Display screen features 213
10.3 Viewing and programming device parameters 218
10.4 Starting and ending patient sessions 222
10.5 Viewing live waveform traces 225
10.6 Recording live waveform strips 232
10.7 Saving and retrieving device data 234
10.8 Printing reports 238
11 Using system evaluation tools 245
11.1 A summary of system evaluation tools 246
11.2 Taking a quick look at device activity 247
11.3 Using the Patient Alert feature 249
11.4 Streamlining follow-ups with Checklist 256
12 Setting up and viewing collected data 261
12.1 A summary of data collection 262
12.2 Setting up data collection 263
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
8
Table of Contents
12.3 Collecting lead performance data 267
12.4 Viewing the episode and therapy efficacy counters 269
12.5 Viewing episode data 272
12.6 Viewing Flashback Memory 280
12.7 Viewing battery and lead status data 281
12.8 Viewing lead performance trends 283
12.9 Using Cardiac Compass to view long ter m clinical trends 285
12.10 Viewing and entering patient information 292
12.11 Automatic device status monitoring 295
13 Testing the system 297
13.1 Testing overview 298
13.2 Evaluating the underlying rhythm 298
13.3 Measuring pacing thresholds 299
13.4 Measuring lead impedance 302
13.5 Measuring EGM amplitude 303
13.6 Testing the device capacitors 305
14 Conducting Electrophysiologic Studies 307
14.1 EP Study overview 308
14.2 Inducing VF with T-Shock 309
14.3 Inducing VF with 50 Hz Burst 313
14.4 Inducing an arrhythmia with Manual Burst 315
14.5 Inducing an arrhythmia with PES 318
14.6 Delivering a manual therapy 320
15 Solving system problems 325
15.1 Overview 326
15.2 Solving sensing problems 326
15.3 Solving tachyarrhythmia detection problems 328
15.4 Solving tachyarrhythmia therapy problems 329
15.5 Solving bradycardia pacing problems 330
15.6 Responding to device status indicators 331
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
Appendices 333
A Warnings and precautions 335
A.1 General warnings 336
A.2 Storage and handling 336
A.3 Resterilization 337
A.4 Device operation 338
A.5 Lead evaluation and lead connection 339
A.6 Follow-up testing 340
A.7 Explant and disposal 340
A.8 Medical therapy hazards 341
A.9 Home and occupational environments 343
B Device parameters 345
B.1 Emergency settings 346
B.2 Detection parameters 347
B.3 Therapy parameters 348
B.4 Pacing parameters 350
B.5 System maintenance parameters 353
B.6 Data collection parameters 354
B.7 System test and EP study parameters 355
B.8 Fixed parameters 357
B.9 Patient information parameters 359
B.10 Programmer symbols 360
B.11 Parameter interlocks 361
Table of Contents
9
Glossary 363
Index 371
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
Introduction
About this manual
Introduction
This manual describes the operation and intended use of the Intrinsic Model 7288 and Intrinsic 30 Model 7287 systems.
In these systems, both devices provide ventricular tachyarrhythmia detection and therapy and a full range of dual-chamber bradycardia pacing modes and associated features. Unless otherwise noted, all information in this manual applies to both devices.
The full delivered energy is different for Intrinsic Model 7288 devices and Intrinsic 30 Model 7287 devices. Intrinsic devices can be programmed to deliver a full energy of up to 35 J. Intrinsic 30 devices can be programmed to deliver up to 30 J. Throughout the manual, the term "full-energy" refers to either 35 J or 30 J, depending on the device model.
Programmer hardware and screen images
11
The screen image examples in this document show the Medtronic CareLink Model 2090 programmer screen. Wherever possible, these screen images show the application for an Intrinsic Model 7288 device.
The information provided in this manual about using the programmer assumes the Medtronic CareLink Model 2090 Programmer is used. For information about using the Model 9790C Programmer, see the
Manual
Manual conventions
Throughout this document, the word “device” refers in general to both Intrinsic and Intrinsic 30 devices.
The nominal value for that parameter.
On-screen buttons are shown with the name of the button surrounded by brackets: [Button Name].
.
symbol in parameter tables indicates the Medtronic
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
9790/9790C Programmer Instruction
12
Introduction
Additional literature
Technical support
Before implanting the device, it is strongly recommended that you take the following actions:
Refer to the product literature packaged with the device for
information about prescribing the device.
Thoroughly read the technical manuals for the leads used
with the device.
Discuss the procedure and the device with the patient and
any other interested parties, and provide them with any patient information packaged with the device.
Medtronic employs highly trained representatives and engineers located throughout the world to ser ve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropr iate expertise.
For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
Customer education
Medtronic invites physicians to attend an educational seminar on the device. The course describes indications for use, system functions, implant procedures, and patient management.
References
The primary reference for background information is Zacouto FI, Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B,
Cardiac Pacing Diagnostic and Therapeutic Tools.
ed. Springer-Verlag; 1976: 212-218.
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
New York:
Notice
Introduction
See these additional references for more background information:
Estes M, Manolis AS, Wang P, Eds.
Cardioverter-Defibrillators
Inc. 1994.
Kroll MW, Lehmann MH, Eds.
Cardioverter-Defibrillator Therapy: The Engineering-Clinical Interface
Singer I, Ed.
NY: Futura Publishing Co. 1994.
Singer I, Barold SS, Camm AJ, Eds.
Therapy of Arrhythmias for the 21st Century: The State of the Art
This software is provided as an informational tool for the end user. The user is responsible for accurate input of patient information into the software. Medtronic makes no representation as to the accuracy or completeness of the data input into the software. Medtronic SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE INFORMATION PROVIDED IN THE SOFTWARE.
. Norwell, MA: Kluwer Academic Publishers 1996.
Implantable Cardioverter-Defibrillator
. Armonk, NY: Futura Publishing Co. 1998.
. New York, NY: Marcel Dekker,
Implantable
Implantable
. Armonk,
Nonpharmacological
13
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
Quick overview
Part I
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
1Quick reference
1.1 New to Intrinsic 18
1.2 Physical characteristics 18
1.3 Magnet application 20
1.4 Longevity projections 21
1.5 Replacement indicators 23
1.6 Typical charge times 23
1.7 High voltage therapy energy 24
1.8 Stored data and diagnostics 26
1
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
18
Chapter 1
New to Intrinsic
1.1 New to Intrinsic
MVP (Managed Ventricular Pacing) – The MVP modes promote
intrinsic conduction by reducing unnecessary right ventricular pacing. The MVP modes are indicated by AAIR<=>DDDR and AAI<=>DDD on the mode selection screen. These modes provide atrial-based pacing with ventricular backup. For loss of AV conduction, the device switches to DDDR or DDD mode. Periodic checks are performed, and if AV conduction resumes, the device switches back to AAIR or AAI mode.
1.2 Physical characteristics
Table 1-1. Device physical characteristics
Volume
3
Intrinsic
Intrinsic 30
Mass
Intrinsic
Intrinsic 30
H x W x D
Surface area of device can
Radiopaque ID
Materials in contact with human tissue
b
Intrinsic
Intrinsic 30
Intrinsic
Intrinsic 30
c
Intrinsic PUB
Intrinsic 30 PUA
d
38 cm
3
36 cm
76 g
75 g
68 mm x 51 mm x 15 mm
68 mm x 51 mm x 14 mm
2
67 cm
2
66 cm
Titanium/polyurethane/silicone rubber
a
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
1Quick reference
A
V
1
2
3
4
5
Physical characteristics
Table 1-1. Device physical characteristicsa(continued)
Battery Lithium silver vanadium oxide
Connectors Two IS-1 connectors for pacing and
a
Measurements are nominal values based on CAD (computer aided design) model measurements and are rounded to the nearest unit.
b
Grommets may protrude slightly beyond the can surface.
c
Engineering series number follows the radiopaque code.
d
These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue.
Table 1-2. Lead connections
Device port Connector type Software name
SVC DF-1 HVX
RV DF-1 HVB
Can n/a HVA; Can
V IS-1 bipolar
A IS-1 bipolar
sensing, Two DF-1 connectors for high voltage therapy, Active Can electrode (programmable)
19
Figure 1-1. Lead connections
1 DF-1 connector port, SVC (HVX) 2 DF-1 connector port, RV (HVB) 3 Device Active Can electrode, Can (HVA) 4 IS-1 connector port, V 5 IS-1 connector port, A
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
20
Chapter 1
Magnet application
Figure 1-2. Suture holes
1.3 Magnet application
Bringing a magnet close to the device triggers changes in device operation as shown in Table 1-3. When the magnet is removed, the device returns to its programmed operations.
Table 1-3. Effects of magnet application on the device
Pacing mode as programmed
Pacing rate and interval as programmed
VF, VT, and FVT detection suspended
Patient Alert audible tones (20 s or less)
a
b
c
a
b
with programmable alerts enabled:
continuous tone (Test)
on/off intermittent tone (seek
follow-up) high/low dual tone (urgent
follow-up)
with programmable alerts disabled:
no tone
high/low dual tone (urgent
follow-up)
Rate response adjustments are suspended while a Patient Alert tone sounds.
Detection resumes if telemetry is established and the application software is running, or it resumes after the application software has started.
The Test tone does not sound if “VF Detection/Therapy Off” is the only alert enabled.
c
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
1Quick reference
Longevity projections
1.4 Longevity projections
1.4.1 Intrinsic Model 7288 longevity projections
Longevity estimates are based on accelerated battery discharge data and device modeling with EGM pre–storage off and a 60 ppm pacing rate. The estimates provided in Table 1-4 apply to a delivered therapy energy of 35 J and pacing pulses having a
0.4 ms pulse width and either a 2.5 V or 3 V amplitude.
This model assumes default automatic capacitor formation setting. As a guideline, each full-energy charge decreases device longevity by approximately 31 days.
Table 1-4. Intrinsic Model 7288: Projected longevity in years with 0.4 ms pulse width and 60 ppm pacing rate
Maximum
Pacing
DDD, 0%
energy charging frequency
a
Semi-annual Off
Pre-arrhythmia EGM storage
On
Quarterly Off
On
DDD, 50%
Semi-annual Off
On
Quarterly Off
On
AAI<=>DDD (MVP mode), 50% atrial, 5% ventricular
Semi-annual Off
On
Quarterly Off
On
DDD, 100%
Semi-annual Off
On
Quarterly Off
On
a
Maximum energy charging frequency may include full-energy therapy shocks or capacitor formations.
b
The data provided for programming Pre-arrhythmia EGM on is based on a 6 month period (two 3 month
follow-up intervals) over the life of the device. Additional use of Pre-arrhythmia EGM reduces longevity by approximately 27% or 3 months per year.
500pacing impedance
b
2.5 V 3.0 V 2.5 V 3.0 V
8.5 8.5 8.5 8.5
8.3 8.3 8.3 8.3
7.1 7.1 7.1 7.1
7.0 7.0 7.0 7.0
7.4 7.1 7.9 7.6
7.2 6.9
6.4 6.1 6.7 6.5
6.2 6.0 6.5 6.4
7.6 7.4 7.9 7.8
7.4 7.2
6.5 6.4 6.7 6.6
6.4 6.2 6.6 6.5
6.6 6.1 7.3 6.9
6.4 5.9 7.2 6.8
5.7
5.4 6.3 6.0
5.6 5.2 6.2 5.9
900pacing impedance
7.7
7.7
7.4
7.6
21
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
22
Chapter 1
Longevity projections
1.4.2 Intrinsic 30 Model 7287 longevity projections
Longevity estimates are based on accelerated battery discharge data and device modeling with EGM pre–storage off and a 60 ppm pacing rate. The estimates provided in Table 1-5 apply to a delivered therapy energy of 30 J and pacing pulses having a
0.4 ms pulse width and either a 2.5 V or 3 V amplitude.
This model assumes default automatic capacitor formation setting. As a guideline, each full-energy charge decreases device longevity by approximately 24 days.
Table 1-5. Intrinsic 30 Model 7287: Projected longevity in years with 0.4 ms pulse width and 60 ppm pacing rate
Maximum
Pacing
DDD, 0%
energy charging frequency
a
Semi-annual Off
Pre-arrhythmia EGM storage
On
Quarterly Off
On
DDD, 50%
Semi-annual Off
On
Quarterly Off
On
AAI<=>DDD (MVP mode), 50% atrial, 5% ventricular
Semi-annual Off
On
Quarterly Off
On
DDD, 100%
Semi-annual Off
On
Quarterly Off
On
a
Maximum energy charging frequency may include full-energy therapy shocks or capacitor formations.
b
The data provided for programming Pre-arrhythmia EGM on is based on a 6 month period (two 3 month
follow-up intervals) over the life of the device. Additional use of Pre-arrhythmia EGM reduces longevity by approximately 25% or 3 months per year.
500pacing impedance
b
2.5 V 3.0 V 2.5 V 3.0 V
8.6 8.6 8.6 8.6
8.5 8.5 8.5 8.5
7.5 7.5 7.5 7.5
7.3 7.3 7.3 7.3
7.5
7.2 8.0 7.8
7.4 7.0 7.9 7.6
6.6 6.4 7.0 6.8
6.5 6.2 6.8 6.7
7.8 7.6 8.1 7.9
7.6 7.4 7.9 7.8
6.8 6.7 7.0 6.9
6.7 6.5 6.9 6.8
6.7 6.2
6.5 6.0 7.3 7.0
6.0
5.5
5.8 5.4 6.4 6.1
900pacing impedance
7.5
7.1
6.6 6.3
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
1.5 Replacement indicators
Battery voltage and messages about replacement status appear on the programmer display and on printed reports. Table 1-6 lists the Elective Replacement Indicator (ERI) and the End of Life (EOL) conditions.
Table 1-6. Replacement indicators
Elective Replacement (ERI) 2.62 V
End of Life (EOL) 3 months after ERI
ERI date – The programmer displays the date when the battery reached ERI on the Quick Look and Battery and Lead Measurements screens.
Temporary voltage decrease – The battery voltage temporarily decreases following a high voltage charge. If a battery measurement is taken immediately after a high voltage charge, ERI or EOL indicator may be displayed. However, this is a temporary status which will return to normal when the battery has recovered from the charge.
EOL indication – If the programmer indicates that the device is at EOL, replace the device immediately.
1Quick reference
Replacement indicators
23
Post-ERI conditions – EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% DDD pacing at 60 ppm, 3 V, 0.4 ms;
pacing load; and six full-energy charges. EOL may be
500 indicated before the end of three months if the device exceeds these conditions.
1.6 Typical charge times
The most recent capacitor charge time appears on the programmer display and on printed reports and can be evaluated using the Charge/Dump test (see Table 1-7 and Table 1-8).
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
24
Chapter 1
High voltage therapy energy
Table 1-7. Intrinsic Model 7288: typicalafull energy charge times
At Beginning of Life (BOL)
At Elective Replacement (ERI)
a
These charge times are typical when the capacitors are fully formed.
Table 1-8. Intrinsic 30 Model 7287: typicalafull energy charge times
At Beginning of Life (BOL)
At Elective Replacement (ERI)
a
These charge times are typical when the capacitors are fully formed.
1.7 High voltage therapy energy
The stored energy of the device is derived from the peak capacitor voltage and is always greater than the energy delivered by the device. Table 1-9 compares the programmed energy levels delivered by the Intrinsic to the energy levels stored in the capacitors before delivery. Table 1-10 makes the same comparison for the Intrinsic 30 device.
7.0 s
8.9 s
5.9 s
7.5 s
Table 1-9. Intrinsic Model 7288: delivered (programmed) and stored energy levels
Energy ( J)
Delivereda/ Programmed Stored
35 39 7.0
32 37 6.4
30 34 6.0
28 32 5.6
26 30 5.2
25 29 5.0
24 27 4.8
22 25 4.4
20 23 4.0
18 21 3.6
16 19 3.2
15 17 3.0
14 16 2.8
13 15 2.6
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
b
Charge Timec( s)
Delivereda/ Programmed Stored
Energy ( J)
b
10 12 2.0
9 10.5 1.8
8 9.3 1.6
7
6 7.1 1.2
5
4 4.8 0.8
3 3.6 0.6
2 2.4 0.4
1.8 2.2 0.4
1.6 2.0 0.3
1.4 1.7 0.3
1.2 1.5 0.2
1.0 1.2 0.2
8.2 1.4
5.9 1.0
Charge Timec( s)
1Quick reference
High voltage therapy energy
Table 1-9. Intrinsic Model 7288: delivered (programmed) and stored energy levels (continued)
25
Energy ( J)
Delivereda/ Programmed Stored
12 14 2.4
11 13 2.2
b
Charge Timec( s)
Delivereda/ Programmed Stored
Energy ( J)
Charge
b
Timec( s)
0.8 1.0 0.2
0.6 0.8 0.1
0.4 0.5 0.1
a
Energy delivered at connector block into a 75
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth
of a second.
load.
Table 1-10. Intrinsic 30 Model 7287: delivered (programmed) and stored energy levels
Energy ( J)
Delivereda/ Programmed Stored
b
Charge Timec( s)
Delivereda/ Programmed Stored
30 35 5.9
28 33 5.6
26 31 5.2
24 28 4.7
22 26 4.4
20 24 4.0
18 21 3.5
16 19 3.2
15 18 3.0
14 16 2.7
13 15 2.5
12 14 2.4
11 13 2.2
10 12 2.0
9 11 1.9
a
Energy delivered at connector block into a 75
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth of a second.
load.
Energy ( J)
Charge
b
Timec( s)
8 9.4 1.6
7
8.3 1.4
6 7.1 1.2
5
5.9 1.0
4 4.7 0.8
3 3.6 0.6
2 2.4 0.4
1.8 2.2 0.4
1.6 2.0 0.3
1.4 1.7 0.3
1.2 1.5 0.3
1.0 1.2 0.2
0.8 1.0 0.2
0.6 0.8 0.1
0.4 0.5 0.1
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
26
Chapter 1
Stored data and diagnostics
1.8 Stored data and diagnostics
Table 1-11. Stored data and diagnostics
Episode data
Tachy episodes 150 VF/VT/FVT episodes: intervals, text, EGM
EGM capacity for tachy episodes 14 minutes of dual-channel EGM, or 23.5 minutes of
SVT/NST episodes 50 SVT/NST episodes: intervals, text, EGM (the device
EGM capacity for SVT/NST episodes
Brady episodes 53 mode switch episodes
EGM sources 9 options: atrial/ventricular/far-field
EGM options Store before onset; Store during charging
Flashback memory 2000 intervals (containing both A-A and V-V): before latest
Counter data
Detection counters Lifetime total, since cleared, and since last session
Episode counters Episodes:
Therapy efficacy counters Counts for each VF, FVT, VT Therapy:
Other stored data
Patient Alert events Up to 10 log entries: text and date for the first time an alert
single-channel EGM
does not usually store detailed episode records for NST episodes)
2 minutes of dual-channel EGM, or 3.6 minutes of single-channel EGM
VF, before latest VT, and before interrogation
Percentage pacing:
Additional counters:
Total number of aborted shocks
is triggered between interrogations
VF, FVT, and VT Atrial Fibrillation/Atrial Flutter episodes Sinus Tach episodes Other 1:1 SVT episodes NST episodes Mode switch episodes
AS-VS, AS-VP, AP-VS, AP-VP percentages
Single PVCs and PVC runs Rate stabilization pulses and runs
Delivered Successful Unsuccessful Intervention (manually aborted)
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
Table 1-11. Stored data and diagnostics (continued)
1Quick reference
Stored data and diagnostics
27
Battery and lead measurements Battery voltage, last capacitor formation, last charge, lead
Lead performance trends 14 days of daily measurements plus 80 weeks of weekly
Cardiac Compass trends 14 months of measurement trends:
impedance, EGM amplitude measurements, last high voltage therapy, and sensing integrity counter
minimum and maximum measurements:
Lead impedance: atrial pacing, ventricular pacing,
defibrillation pathway, and SVC lead (if used) EGM amplitude: atrial (P-waves), ventricular
(R-waves)
VT and VF episodes per day
High voltage therapies delivered per day
Ventricular rate during VT or VF
Episodes of non-sustained tachycardia per day
Heart rate variability
Total daily time in AF or AT
Ventricular rate during AF or AT
Percent pacing per day
Patient activity
Average day and night ventricular heart rate
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
2The Intrinsic system
2.1 System overview 30
2.2 Indications and usage 33
2.3 Contraindications 33
2.4 Patient screening 33
2
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
30
Chapter 2
System overview
2.1 System overview
The Intrinsic and Intrinsic 30 Dual Chamber Implantable Cardioverter Defibrillator (ICD) systems are implantable medical device systems that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. The device system includes three major components:
ICD
The ICD senses the electrical activity of the patient’s heart via the sensing electrodes of the implanted atrial and ventricular leads. It then analyzes the heart rhythm based on selectable sensing and detection parameters. If the device detects a tachyarrhythmia, it delivers defibrillation, cardioversion, or antitachycardia pacing therapy to the patient’s heart. If the device identifies a bradyarrhythmia, it delivers bradycardia pacing therapy to the patient’s heart.
Leads
The ICD can be used with transvenous or epicardial defibrillation leads. The lead system should consist of bipolar or paired unipolar1pacing/sensing leads in each chamber of the hear t and one or two high voltage cardioversion/defibrillation electrodes. You can program the Active Can device case as a high voltage electrode. The pacing and sensing electrodes in each chamber sense cardiac activity and deliver pacing stimuli.
Programmer and software
The Medtronic programmer and application software (Model 9997 for Intrinsic and Model 9996 for Intrinsic 30) allow you to perform the following tasks:
– configure the detection, therapy, and bradycardia features
for your patient
– perform electrophysiological studies and system tests
– monitor, display, or print patient cardiac activity
information
– view patient and device diagnostic data
1
With an appropriate unipolar to bipolar adapter kit.
INTRINSIC™/INTRINSIC™30 7288/7287 Reference Manual
Loading...
+ 360 hidden pages