Medtronic 7278 Reference Guide

MAXIMO™ DR 7278
Dual Chamber Implantable Cardioverter Defibrillator
Reference Manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
MaximoTM DR 7278 0
Reference Manual 0
A guide to the operation and programming of the Model 7278 MaximoTM DR Dual Chamber Implantable Cardioverter Defibrillator
The following are trademarks of Medtronic:
Active Can, Cardiac Compass, Checklist, Decision Channel, Flashback, GEM, Leadless ECG, Marker Channel, Maximo, Medtronic, Patient Alert, PR Logic, Quick Look, QuickLink, RapidRead, T-Shock

Table of contents

Introduction 11
Abbreviations and acronyms 13
Part I Quick overview
1 Quick reference 17
Physical characteristics 18
Magnet application 19
Longevity projections 19
Replacement indicators 22
Typical charge times 23
High voltage therapy energy 23
Stored data and diagnostics 25
New and enhanced features 27
2 The Maximo DR system 31
System overview 32
Indications and usage 35
Contraindications 35
Patient screening 35
3 Emergency therapy 37
Delivering emergency therapies 38
Part II Device implant and patient follow-up procedures
4 Implanting the ICD 47
Overview 48
Preparing for an implant 48
Replacing an ICD 50
Positioning the leads 51
Testing sensing and pacing thresholds 53
Connecting the leads to the ICD 54
Testing defibrillation operation and effectiveness 55
Positioning and securing the ICD 58
Maximo DR 7278 Reference Manual
6
Table of contents
Completing the implant procedure 59
5 Conducting a patient follow-up session 61
Patient follow-up guidelines 62
Verifying the status of the implanted system 62
Verifying accurate detection and appropriate therapy 63
Verifying effective bradycardia pacing 65
Part III Configuring the ICD for the patient
6 Detecting tachyarrhythmias 69
Detection overview 70
Setting up sensing 73
Detecting VF episodes 78
Detecting VT episodes 82
Detecting FVT episodes 88
Detecting tachyarrhythmia episodes with Combined Count 93
Monitoring episodes for termination or redetection 95
Enhancing detection with PR Logic criteria 98
Enhancing VT detection with the Stability criterion 108
Detecting double tachycardias 111
Detecting prolonged tachyarrhythmias with High Rate Timeout 112
Key terms 114
7 Treating tachyarrhythmia episodes 119
Treating VF with defibrillation 120
Treating VT and FVT with antitachycardia pacing 130
Treating VT and FVT with cardioversion 140
Optimizing therapy with Smart Mode and Progressive Episode Therapies 149
Key terms 153
8 Treating bradycardia 157
Providing basic pacing therapy 158
Dual-chamber pacing 164
Single chamber pacing 174
Enhancing pacing for optimal cardiac output 177
Maximo DR 7278 Reference Manual
Table of contents
Adjusting the pacing rate with Mode Switch 190
Preventing competitive atrial pacing 194
Detecting and preventing pacemaker-mediated tachycardia 199
Providing Ventricular Safety Pacing 203
Providing pacing after high voltage therapies 206
Key terms 208
9 Optimizing charge time and device longevity 213
Optimizing charge time 214
Optimizing device longevity 217
Key terms 218
Part IV Evaluating and managing patient treatment
10 Using the programmer 221
Setting up and using the programmer 222
Display screen features 223
Viewing and programming device parameters 230
Starting and ending patient sessions 234
Viewing live waveform traces 236
Recording live waveform strips 243
Saving and retrieving device data 245
Printing reports 249
Key terms 254
11 Using system evaluation tools 255
A summary of system evaluation tools 256
Taking a quick look at device activity 257
Using the Patient Alert feature 259
Streamlining follow-ups with Checklist 267
Key terms 270
7
12 Setting up and viewing collected data 271
A summary of data collection 272
Setting up data collection 273
Collecting lead performance data 278
Viewing the episode and therapy efficacy counters 279
Maximo DR 7278 Reference Manual
8
Table of contents
Viewing episode data 284
Viewing Flashback Memory 292
Viewing battery and lead status data 294
Viewing lead performance trends 296
Using Cardiac Compass to view long term clinical trends 298
Viewing and entering patient information 305
Automatic device status monitoring 308
Key terms 311
13 Testing the system 315
Testing overview 316
Evaluating the underlying rhythm 316
Measuring pacing thresholds 317
Measuring lead impedance 320
Measuring EGM Amplitude 322
Testing the device capacitors 324
Key terms 326
14 Conducting Electrophysiologic Studies 327
EP Study overview 328
Inducing VF with T-Shock 330
Inducing VF with 50 Hz Burst 334
Inducing an arrhythmia with Manual Burst 337
Inducing an arrhythmia with PES 340
Delivering a manual therapy 343
Key terms 346
15 Solving system problems 349
Overview 350
Solving sensing problems 351
Solving tachyarrhythmia detection problems 353
Solving tachyarrhythmia therapy problems 354
Solving bradycardia pacing problems 355
Responding to device status indicators 357
Key terms 358
Maximo DR 7278 Reference Manual
Part V Appendices
A Warning and precautions 363
General warnings 364
Storage and handling 364
Resterilization 365
Device implantation and ICD programming 365
Lead evaluation and lead connection 367
Follow-up testing 368
Explant and disposal 369
Medical therapy hazards 369
Home and occupational environments 371
BDevice parameters373
Emergency settings 374
Detection parameters 375
Therapy parameters 377
Bradycardia pacing parameters 379
System maintenance parameters 382
Data collection parameters 383
System test and EP study parameters 384
Fixed parameters 387
Patient information parameters 389
Programmer symbols 390
Parameter interlocks 392
Table of contents
9
Index 393
Maximo DR 7278 Reference Manual

Introduction

Using this manual
Before implanting the ICD, it is strongly recommended that you:
Refer to the product literature packaged with the ICD for information about prescribing the ICD.
Thoroughly read this manual and the technical manuals for the leads used with the device.
Discuss the procedure and the ICD system with the patient and any other interested parties, and provide them with any patient information packaged with the ICD.
Contacting technical support
Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
Introduction
11
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
Customer education
Medtronic invites physicians to attend an education seminar on the complete ICD system. The course includes indications for use, an overview of ICD system functions, implant procedures, and patient management.
Maximo DR 7278 Reference Manual
12
Introduction
References
Notice
The primary reference for background information is Zacouto FI, Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B, ed. Cardiac Pacing Diagnostic and Therapeutic Tools. New York: Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
Singer I, Ed. Implantable Cardioverter-Defibrillator. Armonk, NY: Futura Publishing Co. 1994.
Singer I, Barold SS, Camm AJ, Eds. Nonpharmacological Therapy of Arrhythmias for the 21st Century: The State of the Art. Armonk, NY: Futura Publishing Co. 1998.
Estes M, Manolis AS, Wang P, Eds. Implantable Cardioverter-Defibrillators. New York, NY: Marcel Dekker, Inc.
1994.
Kroll MW, Lehmann MH, Eds. Implantable Cardioverter-Defibrillator Therapy: The Engineering-Clinical Interface. Norwell, MA: Kluwer Academic Publishers 1996.
This software is provided as an informational tool for the end user. The user is responsible for accurate input of patient information into the software. Medtronic makes no representation as to the accuracy or completeness of the data input into the software. MEDTRONIC SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTIAL OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE INFORMATION PROVIDED IN THE SOFTWARE.
Maximo DR 7278 Reference Manual

Abbreviations and acronyms

A- Atrial
AF Atrial Fibrillation
AFib/AFlutter Atrial Fibrillation and/or Atrial Flutter
ARP Atrial Refractory Period
ATP Antitachycardia Pacing
AVP Atrial Vulnerable Period
BOL Beginning of Life
bpm beats per minute
CNID Combined (VT and VF) Number of Intervals to Detect
CV Cardioversion
DF/Defib Defibrillation
ECG Electrocardiogram
Introduction
13
EGM Electrogram
EOL End of Life
ERI Elective Replacement Indicator
FDI Fibrillation Detection Interval
FTI Fast Ventricular Tachycardia Detection Interval
FVT Fast Ventricular Tachycardia
ICD Implantable Cardioverter Defibrillator
J joules
-1
reciprocal minutes; for example, pacing pulses per minute
min
ms milliseconds
mV millivolts
NCAP Non-Competitive Atrial Pacing
NID Number of Intervals to Detect
NST Non-Sustained Tachycardia
Maximo DR 7278 Reference Manual
14
Introduction
PAC Premature Atrial Contraction
PAV Paced A-V Delay
PES Programmed Electrical Stimulation
PMT Pacemaker-Mediated Tachycardia
P-P an atrial interval
ppm paces or pulses per minute
P-R an interval between a P-wave and the subsequent R-wave
PVAB Post Ventricular Atrial Blanking period
PVARP Post Ventricular Atrial Refractory Period
PVC Premature Ventricular Contraction
RAAV Rate Adaptive A-V delay
RNID Number of Intervals to Redetect
R-P an interval between an R-wave and the subsequent P-wave
R-R a ventricular interval
SAV Sensed A-V Delay
ST/Sinus Tach Sinus Tachycardia
SVT Supraventricular Tachycardia
TDI Tachycardia Detection Interval
V volts
V- Ve n tr ic ul ar
VF Ventricular Fibrillation
VF NID VF Number of Intervals to Detect
VRS Ventricular Rate Stabilization
VSP Ventricular Safety Pacing
VT Ventricular Tachycardia
VT NID VT Number of Intervals to Detect
Maximo DR 7278 Reference Manual

Quick overview

Part I
Maximo DR 7278 Reference Manual
Physical characteristics 18
Magnet application 19
Longevity projections 19
Replacement indicators 22
Typical charge times 23
High voltage therapy energy 23
Stored data and diagnostics 25
New and enhanced features 27

Quick reference1

1
Maximo DR 7278 Reference Manual
18
Chapter 1

Physical characteristics

Physical characteristics
Table 1-1. ICD physical characteristicsa
Volume 39 cc
Mass 76 g
b
H x W x D
Surface area of device can 67 cm
Radiopaque IDc
Materials in contact with human tissue
d
68 mm x 51 mm x 15 mm
2
PRM
Titanium / polyurethane / silicone rubber
Battery Lithium silver vanadium oxide
Connectors Two IS-1 connectors for pacing and
sensing, Two DF-1 connectors for high voltage therapy, Active Can electrode (programmable)
Device Port Connector
Typ e
Software Name
SVC DF-1 HVX
A
V
SVC
RV
RV DF-1 HVB
Can n/a HVA, Can
V IS-1 bipolar
A IS-1 bipolar
Can
74lead.eps
Suture holes
78Suture.eps
a
Measurements are nominal values based on CAD (computer aided design) model measurements and are rounded to the nearest unit.
b
Grommets may protrude slightly beyond the can surface.
c
Engineering series number follows the radiopaque code.
d
These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue.
Maximo DR 7278 Reference Manual

Magnet application

Bringing a magnet close to the device triggers changes in device operation as shown in Table 1-2. When the magnet is removed, the device returns to its programmed operations.
Table 1-2. Effects of magnet application on the device
Pacing mode as programmed
Pacing rate and interval as programmed
VF, VT, and FVT detection suspended
Patient Alert audible tones (20 seconds or less)
a
Rate response adjustments are suspended while a Patient Alert tone sounds.
b
Detection resumes if telemetry is established and the application software is running, or it resumes after the application software has started.
c
The Test tone does not sound if “VF Detection/Therapy Off” is the only alert enabled.
Quick reference
Magnet application
a
b
with programmable alert(s) enabled:
continuous tone (Test)
on/off intermittent tone (seek follow-up)
high/low dual tone (urgent follow-up)
with programmable alerts disabled:
no tone
high/low dual tone (urgent follow-up)
c
19

Longevity projections

Longevity estimates are based on accelerated battery discharge data and device modeling with EGM pre-storage off, 60 ppm
-1
(min
) pacing rate, with:
2.5 V pacing pulse amplitude, 0.4 ms pacing pulse width, and 35 J delivered therapy energy (see Table 1-3)
3 V pacing pulse amplitude, 0.4 ms pacing pulse width, and 35 J delivered therapy energy (see Table 1-4)
This model assumes default automatic capacitor formation setting. As a guideline, each full energy charge decreases device longevity by approximately 31 days.
Maximo DR 7278 Reference Manual
20
Chapter 1
Longevity projections
Table 1-3. Projected longevity in years with 2.5 V pacing amplitude and
0.4 ms pulse width
a
EGM pre-storage
500 ohm
b
pacing impedance
900 ohm pacing impedance
Percent pacing
Maximum energy charging frequency
DDD VVI DDD VVI
0% Semi-Annual Off 8.5 8.5 8.5 8.5
On 8.3 8.3 8.3 8.3
Quarterly Off 7.1 7.1 7.1 7.1
On 7.0 7.0 7.0 7.0
15% Semi-Annual Off 8.1 8.4 8.3 8.5
On 7.9 8.2 8.1 8.3
Quarterly Off 6.9 7.1 7.0 7.1
On 6.7 6.9 6.8 7.0
50% Semi-Annual Off 7.4 8.0 7.9 8.3
On 7.2 7.8 7.7 8.1
Quarterly Off 6.4 6.8 6.7 7.0
On 6.2 6.6 6.5 6.8
100% Semi-Annual Off 6.6 7.5 7.3 8.0
On 6.4 7.3 7.2 7.8
Quarterly Off 5.7 6.4 6.3 6.8
On 5.6 6.3 6.2 6.6
a
Maximum energy charging frequency may include full energy therapy shocks or capacitor formations.
b
The data provided for programming EGM pre-storage on is based on a 6 month period (two 3-month follow-up intervals) over the life of the device. Additional use of EGM pre-storage reduces longevity by approximately 27% or 3 months per year.
Maximo DR 7278 Reference Manual
Quick reference
Longevity projections
Table 1-4. Projected longevity in years with 3 V pacing amplitude and
0.4 ms pulse width
Percent pacing
0% Semi-Annual Off 8.5 8.5 8.5 8.5
15% Semi-Annual Off 8.0 8.3 8.2 8.4
50% Semi-Annual Off 7.1 7.8 7.6 8.1
100% Semi-Annual Off 6.1 7.2 6.9 7.8
a
Maximum energy charging frequency may include full energy therapy shocks or capacitor formations.
b
The data provided for programming EGM pre-storage on is based on a 6 month period (two 3-month follow-up intervals) over the life of the device. Additional use of EGM pre-storage reduces longevity by approximately 27% or 3 months per year.
Maximum energy charging frequency
a
EGM pre-storage
500 ohm
b
pacing impedance
900 ohm pacing impedance
DDD VVI DDD VVI
On 8.3 8.3 8.3 8.3
Quarterly Off 7.1 7.1 7.1 7.1
On 7.0 7.0 7.0 7.0
On 7.8 8.1 8.0 8.2
Quarterly Off 6.8 7.0 6.9 7.1
On 6.6 6.9 6.8 6.9
On 6.9 7.6 7.4 7.9
Quarterly Off 6.1 6.7 6.5 6.9
On 6.0 6.5 6.4 6.7
On 5.9 7.0 6.8 7.6
Quarterly Off 5.4 6.2 6.0 6.6
On 5.2 6.0 5.9 6.5
21
Maximo DR 7278 Reference Manual
22
Chapter 1

Replacement indicators

Replacement indicators
Battery voltage and messages about replacement status appear on the programmer display and on printed reports. Table 1-5 lists the Elective Replacement Indicator (ERI) and the End of Life (EOL) conditions.
Table 1-5. Replacement indicators
Elective Replacement (ERI) 2.62 V
End of Life (EOL) 3 months after ERI
ERI date – The programmer displays the date when the battery reached ERI on the Quick Look and Battery and Lead Measurements screens.
Temporary voltage decrease – The battery voltage temporarily decreases following a high voltage charge. If a battery measurement is taken immediately after a high voltage charge, the ERI or EOL indicator may be displayed. However, this is a temporary status which will return to normal when the battery has recovered from the charge.
EOL indication – If the programmer indicates that the device is at EOL, replace the device immediately.
Post-ERI conditions – EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% DDD pacing at 60 ppm, 3 V, 0.4 ms; 500 pacing load; and six 35 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.
Maximo DR 7278 Reference Manual

Typical charge times

The most recent capacitor charge time appears on the programmer display and on printed reports and can be evaluated using the Charge/Dump test (see Table 1-6).
Ta b l e 1 - 6 . Ty p ic a la full energy charge times
At Beginning of Life (BOL) 7.0 seconds
At Elective Replacement (ERI) 8.9 seconds
a
These charge times are typical when the capacitors are fully formed.

High voltage therapy energy

The stored energy of the device is derived from the peak capacitor voltage and is always greater than the energy delivered by the device. Table 1-7 compares the programmed energy levels delivered by the device to the energy levels stored in the capacitors before delivery.
Quick reference
Typical charge times
23
Maximo DR 7278 Reference Manual
24
Chapter 1
High voltage therapy energy
Table 1-7. Comparing delivereda (programmed) and storedb energy levels
Energy (J) Charge
Delivered a/
Stored
b
Programmed
35 39 7.0 10 12 2.0
32 37 6.4 9 10.5 1.8
30 34 6.0 8 9.3 1.6
28 32 5.6 7 8.2 1.4
26 30 5.2 6 7.1 1.2
25 29 5.0 5 5.9 1.0
24 27 4.8 4 4.8 0.8
22 25 4.4 3 3.6 0.6
20 23 4.0 2 2.4 0.4
18 21 3.6 1.8 2.2 0.4
16 19 3.2 1.6 2.0 0.3
15 17 3.0 1.4 1.7 0.3
14 16 2.8 1.2 1.5 0.2
13 15 2.6 1.0 1.2 0.2
12 14 2.4 0.8 1.0 0.2
11 13 2.2 0.6 0.8 0.1
a
Energy delivered at connector block into a 75 ohm load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth of a second.
Time
c
(sec)
Delivered a/ Programmed
Energy (J) Charge
Timec (sec)
b
Stored
0.4 0.5 0.1
Maximo DR 7278 Reference Manual

Stored data and diagnostics

Table 1-8. Stored data and diagnostics
Episode data
Tachy episodes 150 VF/VT/FVT episodes: intervals, text, EGM
Quick reference
Stored data and diagnostics
25
EGM capacity for tachy episodes
14 minutes of dual-channel EGM, or 23.5 minutes of single-channel EGM
SVT/NST episodes 50 SVT/NST episodes: intervals, text, EGM (the
device does not usually store detailed episode records for NST episodes)
EGM capacity for SVT/NST episodes
2 minutes of dual-channel EGM, or 3.6 minutes of single-channel EGM
Brady episodes 53 mode switch episodes
EGM sources 9 options: atrial / ventricular / far-field
EGM options Store before onset; Store during charging
Flashback memory 2000 intervals (containing both A-A and V-V):
before latest VF, before latest VT, and before interrogation
Counter data
Detection counters Lifetime total, since cleared, and since last
session
Episode counters Episodes:
VF, FVT, and VT
Atrial Fibrillation / Atrial Flutter episodes
Sinus Tach episodes
Other 1:1 SVT episodes
NST episodes
Mode switch episodes
Percentage pacing:
AS-VS, AS-VP, AP-VS, AP-VP percentages
Additional counters:
Single PVCs and PVC runs
Rate stabilization pulses and runs
Maximo DR 7278 Reference Manual
26
Chapter 1
Stored data and diagnostics
Table 1-8. Stored data and diagnostics (continued)
Therapy efficacy counters
Other stored data
Counts for each VF, FVT, VT Therapy:
Delivered
Successful
Unsuccessful
Intervention (manually aborted)
Total number of aborted shocks
Patient Alert events Up to 10 log entries: text and date for the first
time an alert is triggered between interrogations
Battery and lead measurements
Battery voltage, last capacitor formation, last charge, lead impedance, EGM amplitude measurements, last high voltage therapy, and sensing integrity counter
Lead performance trends
14 days of daily measurements plus 80 weeks of weekly minimum and maximum measurements:
Lead impedance: atrial pacing, ventricular pacing, defibrillation pathway, and SVC lead (if used)
EGM amplitude: atrial (P-waves), ventricular (R-waves)
Cardiac Compass trends
14 months of measurement trends:
VT and VF episodes per day
High voltage therapies delivered per day
Ventricular rate during VT or VF
Episodes of non-sustained tachycardia per day
Heart rate variability
Total daily time in AF or AT
Ventricular rate during AF or AT
Percent pacing per day
Patient activity
Average day and night ventricular heart rate
Maximo DR 7278 Reference Manual

New and enhanced features

The following features are new or changed from the 7275 GEM III DR ICD.
Patient management
RapidRead telemetry – Communication between the device and
programmer is approximately 20 times faster than telemetry in previous Medtronic ICD devices. The magnitude of improvement depends on the amount and type of data that is interrogated. RapidRead telemetry is more reliable and has an increased range that makes placing the programming head easier.
Cardiac Compass trends report – This report displays up to 14 months of trend data related to tachyarrhythmia episodes, heart rate, and patient activity. See “Using Cardiac Compass to view long term clinical trends” on page 298.
Patient Alert – The alert duration when a magnet is applied to the device is now 20 seconds. The device also provides several new alerts:
SVC (HVX) lead impedance out of range
Active Can off without SVC
DOO/VOO mode programmed
VF Detection programmed off, or fewer than four VF therapies enabled for at least six hours
charge circuit timeout occurred
excessive charge time ERI
Quick reference
New and enhanced features
27
For more information, see “Using the Patient Alert feature” on page 259.
EGM amplitude trends – The device automatically measures R-wave and P-wave EGM amplitudes every day. These daily measurements are included in the data displayed on the Lead Performance Trends screen. See “Setting up data collection” on page 273.
Maximo DR 7278 Reference Manual
28
Chapter 1
New and enhanced features
EGM amplitude test – You can use the EGM Amplitude test to measure R-wave and P-wave EGM amplitudes. The results are reported on the EGM Amplitude Test screen. See “Measuring EGM Amplitude” on page 322.
Lead impedance measurement for SVC (HVX) – Along with other lead impedance measurements, the device provides an independent SVC (HVX) measurement to check the integrity of the supplementary high voltage electrode. See “Measuring lead impedance” on page 320.
Leadless ECG signal – If a supplementary high voltage electrode is placed in the SVC, the device provides the Leadless ECG signal through either the Can to SVC (HVX) or RV (HVB) to SVC (HVX) EGM source. See “Setting up data collection” on page 273.
Expanded pre-onset EGM storage – The device can now store up to 20 seconds of EGM before a tachycardia starts. See “Setting up data collection” on page 273.
Smart Auto Cap Formation – When the Auto Cap Formation Interval is set to Auto, the formation interval automatically adjusts to optimize device longevity and charge times. See “Smart Auto Cap” on page 216.
Ending a patient session – The device audits the programmed parameter settings when you end a patient session and alerts you if any of the settings are atypical. See “Starting and ending patient sessions” on page 234.
Tachyarrhythmia detection
VT Monitoring – VT detection can be set to Monitor, which allows
the device to detect and record VT episodes without delivering therapy or influencing VF detection. See “VT monitoring” on page 86.
High Rate Timeout – High Rate Timeout can turn off detection enhancements (Stability, PR Logic criteria) if a high rate episode is longer than a programmed duration. See “Details about High Rate Timeout” on page 113.
Maximo DR 7278 Reference Manual
Tachyarrhythmia therapy
Episode confirmation during and after charging – The device
continually monitors the ventricular rhythm during and after charging for cardioversion or defibrillation (when VF confirmation is active) to ensure the arrhythmia is present before delivering the high voltage shock. See “Confirming VF after initial detection” on page 126 and “Confirming VT or FVT after detection” on page 144.
Programmable Active Can – If a supplementary electrode is connected to the SVC (HVX) port, you can deselect the device Can as a high voltage electrode. See “Delivery pathway electrodes” on page 123.
Output – The device has a maximum delivered energy of 35 joules.
Bradycardia pacing
Quick reference
New and enhanced features
29
EP studies
Accelerometer-based rate response – The device uses an
accelerometer to provide rate responsive pacing.
Additional bradycardia pacing modes – The device provides asynchronous pacing in the DOO and VOO pacing modes and provides the ODO mode to disable pacing. See Chapter 8, “Treating bradycardia” on page 157.
Enhanced AT-style Mode Switch – Mode Switch episodes are detected using a combination of the median atrial rate and the AF evidence criterion. See “Details about Mode Switch” on page 192.
Defibrillation threshold testing support – The T-Shock and 50 Hz Burst induction screens allow you to monitor time between inductions, program ventricular sensing and VF therapy settings, adjust induction settings, select manual therapies, and retrieve episode records after therapy. See “How to perform defibrillation threshold testing” on page 57.
Maximo DR 7278 Reference Manual
30
Chapter 1
New and enhanced features
Backup VVI pacing during atrial inductions – You can choose to have the device deliver backup ventricular pacing during Manual Burst and PES inductions that are delivered to the atrium. See “EP Study overview” on page 328.
Maximo DR 7278 Reference Manual
System overview 32
Indications and usage 35
Contraindications 35
Patient screening 35

The Maximo DR system2

2
Maximo DR 7278 Reference Manual
32
Chapter 2

System overview

System overview
The Model 7278 Maximo DR Dual Chamber Implantable Cardioverter Defibrillator (ICD) system is an implantable medical device system that automatically detects and treats episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. The ICD system includes three major components:
ICD
The ICD senses the electrical activity of the patient’s heart via the sensing electrodes of the implanted atrial and ventricular leads. It then analyzes the heart rhythm based on selectable sensing and detection parameters. If the ICD detects a tachyarrhythmia, it delivers defibrillation, cardioversion, or antitachycardia pacing therapy to the patient’s heart. If the ICD identifies a bradyarrhythmia, it delivers bradycardia pacing therapy to the patient’s heart.
Leads
The ICD can be used with transvenous or epicardial defibrillation leads. The lead system should consist of bipolar or paired unipolar
1
pacing/sensing leads in each chamber of the heart and one or two high voltage cardioversion/ defibrillation electrodes. You can program the Active Can device case as a high voltage electrode. The pacing and sensing electrodes in each chamber sense cardiac activity and deliver pacing stimuli.
Programmer and software
The Medtronic programmer and 9978 application software allow you to perform the following tasks:
configure the detection, therapy, and bradycardia features for your patient
perform electrophysiological studies and system tests
monitor, display, or print patient cardiac activity information
view patient and device diagnostic data
1
With an appropriate unipolar to bipolar adapter kit.
Maximo DR 7278 Reference Manual
The Maximo DR devices and 9978 application software are compatible with the following programmer systems:
Medtronic CareLink Model 2090 programmer with a Model 2067 or 2067L programming head
Medtronic Model 9790C programmer with a Model 9767 or 9767L programming head
For information about:
indications, contraindications, lead compatibility, warnings and precautions, and patient selection, see the Maximo DR 7278 Implant Manual, which accompanies each device.
basic programmer and software desktop functions that are not included in Chapter 10, “Using the programmer” on page 221, see the manual accompanying the programmer.
installing the programming head, see the manual accompanying the programming head.
implanting leads, refer to the manuals accompanying the leads.
Detecting and treating tachyarrhythmias
The Maximo DR system
System overview
33
The ICD monitors the cardiac rhythm for short ventricular intervals that may indicate the presence of VF, VT, or FVT.
Upon detection of VF, the ICD delivers a biphasic defibrillation shock of up to 35 joules. If the VF episode persists, up to five more individually programmed defibrillation shocks can be delivered.
Upon detection of VT, the ICD delivers either a Ramp, Ramp+, or Burst antitachycardia pacing therapy or a biphasic cardioversion shock of up to 35 joules synchronized to a ventricular depolarization. If the VT episode persists, up to five more individually programmed VT therapies can be delivered. You can also program the ICD to monitor the VT episode without delivering therapy.
Upon detection of FVT, the ICD delivers either a Ramp, Ramp+, or Burst antitachycardia pacing therapy or a biphasic cardioversion shock of up to 35 joules synchronized to a ventricular depolarization. If the FVT episode persists, up to five more individually programmed FVT therapies can be delivered.
Maximo DR 7278 Reference Manual
34
Chapter 2
System overview
You can program the ICD to distinguish between true ventricular arrhythmias and rapidly conducted supraventricular tachycardia (SVT) and withhold therapy for SVT.
You can also program the ICD to detect a double tachycardia (an unrelated ventricular arrhythmia occurring simultaneously with an SVT), so that therapy is not withheld for a ventricular arrhythmia in the presence of an SVT.
Treating bradycardia
The ICD provides dual chamber rate responsive bradycardia pacing to optimize hemodynamics. An internal accelerometer senses the patient’s physical activity, allowing the ICD to increase and decrease the pacing rate in response to changes in the level of activity.
Monitoring for real-time and stored data
The ICD and programmer provide real-time information on
detection and therapy parameters and status during a patient session. The ICD also provides accumulated data on device operation, including stored electrograms, detected and treated tachyarrhythmia episodes, bradycardia interventions, and the efficacy of therapy. The Cardiac Compass report provides up to 14 months of clinically significant data, including arrhythmia episodes, therapies delivered, physical activity, heart rate, and bradycardia pacing activities.
All of this information can be printed and retained in the patient’s file or saved in electronic format on a floppy diskette.
Conducting electrophysiologic tests
You can use the system to conduct non-invasive electrophysiologic studies including manual delivery of any of the ICD therapies to manage an induced or spontaneous tachyarrhythmia.
Alerting the patient to system events
You can use the programmable Patient Alert monitoring feature to notify the patient with audible tones if certain conditions related to the leads, battery, charge time, and therapies occur. The patient can then respond based on your prescribed instructions.
Maximo DR 7278 Reference Manual

Indications and usage

The Model 7278 Maximo DR system is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias.

Contraindications

The Maximo DR system is contraindicated for
patients whose tachyarrhythmias may have transient or reversible causes, such as acute myocardial infarction, digitalis intoxication, drowning, electrocution, electrolyte imbalance, hypoxia, or sepsis.
patients with incessant VT or VF
patients who have a unipolar pacemaker
patients whose primary disorder is bradyarrhythmias or atrial arrhythmias
The Maximo DR system
Indications and usage
35

Patient screening

Prior to implant, patients should undergo a complete cardiac evaluation, including electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed tachyarrhythmia therapies are recommended during and after the implantation of the device.
Other optional screening procedures could include exercise stress testing to determine the patient’s maximum sinus rate, and cardiac catheterization to determine if there is a need for concomitant surgery and/or medical therapy.
Maximo DR 7278 Reference Manual

Emergency therapy3

Delivering emergency therapies 38
How to deliver emergency 35 joule defibrillation 40
How to deliver emergency cardioversion 41
How to deliver emergency fixed burst pacing 42
How to deliver emergency VVI pacing 44
3
Maximo DR 7278 Reference Manual
38
Chapter 3

Delivering emergency therapies

Delivering emergency therapies
The device provides the following emergency therapies:
defibrillation
cardioversion
fixed burst pacing
emergency VVI pacing
The default emergency therapy is 35 joule defibrillation. When you select [Emergency] and [DELIVER], the device charges and delivers a biphasic 35 joule shock along the AX>B pathway
The programmer resets the emergency defibrillation energy to 35 joules each time you select [Emergency]. Emergency cardioversion and fixed burst values remain as selected for the duration of the session.
To return to other programming functions from an Emergency screen, select [Exit Emergency].
Effect on system operation
1
.
The device suspends the automatic detection features when emergency defibrillation, cardioversion, or fixed burst pacing therapies are delivered. Detection is not suspended during emergency VVI pacing. Removing the programming head or pressing [Resume] turns detection on again.
Aborting an emergency therapy
As a safety precaution, the programmer also displays an [ABORT] button which immediately terminates any emergency therapy in progress.
1
If Active Can is turned off, the defibrillation is delivered between the HVX and HVB electrodes.
Maximo DR 7278 Reference Manual
Emergency therapy
Delivering emergency therapies
Mechanical Emergency buttons on the Model 9790C programmer
If you press the red mechanical [Emergency] button on the programmer display panel, the programmer displays the Emergency screen. The mechanical yellow-on-blue [Deliver] button activates the emergency therapy displayed on the programmer screen. This button functions only when the Emergency screen is displayed.
Mechanical Emergency VVI button on the CareLink Model 2090 programmer
If you press the red Emergency VVI button on the programmer display panel, the device initiates Emergency VVI pacing and the programmer displays the Emergency screen.
Temporary parameter values
39
Emergency tachyarrhythmia therapies use temporary values that
1
do not change the programmed parameters of the device.
These values are not in effect until you select [DELIVER]. After the tachyarrhythmia therapy is complete, the device reverts to its programmed values.
1
Delivery of Emergency VVI Pacing changes the programmed bradycardia pacing values to the emergency values (see page 43).
Maximo DR 7278 Reference Manual
40
Chapter 3
Delivering emergency therapies
Delivering an emergency defibrillation therapy
The default emergency therapy is a full-energy defibrillation. When you select [Emergency] and [DELIVER], the device charges and delivers a biphasic full-energy shock. The programmer resets the emergency defibrillation energy to its maximum value each time you select [Emergency] or select the [Defibrillation] option from an Emergency screen.
Parameters
Energy – Amount of energy delivered to the
heart by the therapy.
* Medtronic nominal setting
10, 11, ... 16 18, 20, 22, 24, 25, 26, 28, 30, 32, 35*J
Pathwaya – Direction the electrical current flows through the heart.
a
If Active Can is Off, the HVA (Can) electrode is not used as part of the high-energy delivery pathway.
How to deliver emergency 35 joule defibrillation
3
4
2
AX>B (fixed)
1. Position the programming head over the device.
2. Select [Emergency].
3. Accept the defibrillation energy shown on the screen, or select Energy and select a new value from the window.
4. Select [DELIVER]. If delivery is not confirmed,
verify that the programming head is properly positioned and select [Retry] or [Cancel].
Maximo DR 7278 Reference Manual
Delivering emergency therapies
Delivering an emergency cardioversion therapy
When you initiate an emergency cardioversion therapy, the device charges its capacitors to the selected energy and attempts to deliver therapy synchronized with a sensed tachyarrhythmia event. If the cardioversion therapy cannot be synchronized, it is aborted. See “Synchronizing cardioversion after charging” on page 145.
Parameters
Energy – Amount of energy delivered to the
heart by the therapy.
a
Pathway
through the heart.
a
If Active Can is Off, the HVA (Can) electrode is not used as part of the high-energy delivery pathway.
– Direction the electrical current flows
How to deliver emergency cardioversion
3
4
5
2
Emergency therapy
* Medtronic nominal setting
0.4, 0.6, ...1.8, 2, 3, ... 16 18, 20, 22, 24, 25, 26, 28, 30, 32, 35*J
AX>B (fixed)
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the cardioversion energy shown on the screen, or select Energy and select a new value from the window.
5. Select [DELIVER]. If delivery is not confirmed,
verify that the programming head is properly positioned and select [Retry] or [Cancel].
41
Maximo DR 7278 Reference Manual
42
Chapter 3
Delivering emergency therapies
Delivering emergency fixed burst pacing
Emergency fixed burst pacing delivers maximum output pacing pulses to the ventricle at a selectable interval. The therapy continues for as long as you keep the programmer stylus on the [BURST Press and Hold] button.
Parameters
How to deliver emergency fixed burst pacin
Interval – Time interval between pacing
pulses delivered during the fixed burst therapy.
V. Amplitude – Voltage of the ventricular pacing pulses delivered during the fixed burst therapy.
V. Pulse Width – Duration of the ventricular pacing pulses delivered during the fixed burst therapy.
How to deliver emergency fixed burst pacing
3
4
2
5
* Medtronic nominal setting
100, 110, ... 350* 360, 370, ... 600 ms
8V(fixed)
1.6 ms (fixed)
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval shown on the screen, or select Interval for a new interval value.
5. Select [BURST Press and Hold].
If delivery is not confirmed, the programmer displays an error window. Verify that the programming head is properly positioned. Select [OK] from the window and reselect [BURST Press and Hold].
Maximo DR 7278 Reference Manual
Enabling emergency VVI pacing
Emergency VVI pacing programs the device to deliver high-output ventricular pacing. You can initiate emergency VVI pacing from the Emergency screen or by pressing the red mechanical button on the programmer display panel and selecting VVI Pacing on the screen. To disable emergency VVI pacing, reprogram the bradycardia pacing parameters from the Parameters screen.
Emergency therapy
Delivering emergency therapies
43
Parameters
How to deliver emergency fixed burst pacin
Pacing Mode– NBG Codea for the pacing
VVI mode provided during emergency VVI pacing.
Lower Rate – Minimum pacing rate to
70 min
-1
maintain adequate heart rate during periods of inactivity.
V. Amplitude – Voltage of the ventricular
6V pacing pulses delivered during emergency VVI pacing.
V. Pulse Width – Duration of the
1.6 ms ventricular pacing pulses delivered during emergency VVI pacing.
V. Pace Blanking – Time interval during
240 ms which sensing is disabled after a pacing pulse.
Hysteresis – Enables tracking of intrinsic
Off heart rate below programmed Lower Rate to prevent pacing during extended periods of inactivity, such as when a patient is sleeping.
V. Rate Stabilization – Modifies the
Off pacing rate to eliminate the long pause that typically follows a premature ventricular contraction.
a
N–North American Society of Pacing and Electrophysiology (NASPE), B–British Pacing and Electrophysiology Group (BPEG), G–Generic Pacemaker Code
Maximo DR 7278 Reference Manual
44
Chapter 3
Delivering emergency therapies
How to deliver emergency VVI pacing
4
2
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A successful programming
3
sets the device to the following maximum output bradycardia pacing values.
Pacing Mode: VVI
Lower Rate: 70 ppm
-1
(70 min
V. Amplitude: 6 V
V. Width: 1.6 ms
V. Pace Blanking: 240 ms
Hysteresis: Off
V. Rate Stabilization: Off
)
If programming is not confirmed, verify that the programming head is properly positioned and select [Retry] or [Cancel].
Maximo DR 7278 Reference Manual
Part II

Device implant and patient follow-up procedures

Maximo DR 7278 Reference Manual

Implanting the ICD4

Overview 48
Preparing for an implant 48
Replacing an ICD 50
Positioning the leads 51
Testing sensing and pacing thresholds 53
Connecting the leads to the ICD 54
Testing defibrillation operation and effectiveness 55
Positioning and securing the ICD 58
Completing the implant procedure 59
4
Maximo DR 7278 Reference Manual
48
Chapter 4

Overview

Overview

Preparing for an implant

The tasks for implanting an ICD include
1. Preparing for an implant
2. Replacing an ICD
3. Positioning the leads
4. Testing sensing and pacing thresholds
5. Connecting the leads to the ICD
6. Testing defibrillation operation and effectiveness
7. Positioning and securing the ICD
8. Completing the implant procedure
These tasks are described in the sections that follow.
Warning: Keep a back-up external defibrillator available during the implant for transthoracic rescue when arrhythmias are induced.
Equipment for an implant
The equipment that is needed for an implant is as follows:
Medtronic CareLink Model 2090 programmer with a Model 2067 or 2067L programming head, or a Model 9790C programmer with a Model 9767 or 9767L programming head
Maximo Model 9978 software application
Model 2290 or 8090 Analyzer lead analysis device or equivalent pacing system analyzer
external defibrillator
Maximo DR 7278 Reference Manual
Sterile supplies for an implant
The sterile supplies that are needed for an implant are as follows:
implantable device and lead system components
programming head sleeve or programming head
analyzer cables
lead introducers appropriate for the lead system
extra stylets of appropriate length and shape
How to prepare for implanting
Set up the programmer and start the application
1. Set up the programmer as described in the instructions provided with the programmer.
2. Install the Maximo DR Model 9978 software on the programmer, if it is not already installed.
3. Place the programming head over the device and start the application. Select the device model or select [Auto identify].
Note: The programmer automatically interrogates the device when the application starts.
Implanting the ICD
Preparing for an implant
49
Preprogram the device
Before opening the sterile package, prepare the ICD for implant as follows:
1. Check the “use by” date printed on the package. Do not implant the device after the “use by” date because the battery’s longevity could be reduced.
2. Interrogate the ICD, and print a full summary report.
3. Confirm that the battery voltage is at least 3.0 V at room temperature.
a
If the device has been exposed to lower temperatures or has delivered a recent high voltage charge, the battery voltage will be temporarily lower.
4. Set up data collection parameters and the ICD internal clock (see page 275).
5. Perform a manual capacitor formation (see page 324).
6. Program the therapy and pacing parameters to values appropriate for the patient (see page 161). Ensure that all tachyarrhythmia detection is programmed Off (see page 72).
a
Use the Quick Look screen to verify the voltage, see page 257.
Maximo DR 7278 Reference Manual
50
Chapter 4

Replacing an ICD

Replacing an ICD
If you are replacing a previously implanted ICD, turn off ICD detection and therapies before explanting.
When implanting the ICD with a chronic lead system, perform the following evaluations to ensure appropriate detection and therapy:
Check the integrity of the chronic high voltage leads with a test shock, chest X-ray, and inspection.
Perform chronic pacing and sensing measurements.
Measure high voltage lead impedances.
Test defibrillation efficacy.
Confirm adequate sensing during VF.
Ensure proper fit of the lead connectors in the ICD connector block.
Notes:
To meet the implant requirements, it may be necessary to reposition or replace the chronic leads or to add a third high voltage electrode.
Any unused leads that remain implanted must be capped.
How to explant and replace an ICD
1. Program all tachyarrhythmia detection Off.
2. Dissect the leads and the ICD free from the surrounding tissues in the surgical pocket. Be careful not to nick or breach the lead insulation during the process of exposing the system.
3. Loosen each setscrew, and gently retract the lead from the connector block.
4. Remove the ICD from the surgical pocket.
5. If the connector pin of any implanted lead shows signs of pitting or corrosion, replace the implanted lead with a new lead. The damaged lead should be discarded and replaced to assure the integrity of the device system.
6. Measure sensing, pacing, and defibrillation efficacy using the replacement ICD.
7. Evaluate the defibrillation efficacy of the replacement system.
Maximo DR 7278 Reference Manual

Positioning the leads

Implant endocardial leads according to the supplied instructions, unless suitable chronic leads are already in place. Do not use any lead with this device without first verifying connector compatibility (refer to the Maximo DR 7278 Implant Manual). Transvenous or epicardial leads may be used. A bipolar atrial lead with closely spaced pacing and sensing electrodes is recommended.
Using transvenous leads
Use standard transvenous implant techniques to position the ventricular lead tip in the right ventricular apex and the atrial pacing lead tip high on the right atrial appendage.
Follow the general guidelines below for initial positioning of other transvenous leads (the final positions are determined by defibrillation efficacy tests):
SVC (HVX) lead: Place the lead tip high in the innominate vein, approximately 5 cm proximal to the right atrium (RA) and SVC junction.
SQ patch: Place the patch along the left mid-axillary, centered over the fourth-to-fifth intercostal space.
CS lead: Advance the lead tip to just under the left atrial appendage, if possible.
Implanting the ICD
Positioning the leads
51
If using a subclavian approach, position the lead laterally to avoid pinching the lead body between the clavicle and the first rib.
Warning: Pinching the lead can damage the lead conductor or insulation, which may cause unwanted high voltage therapies or result in the loss of sensing or pacing therapy.
Maximo DR 7278 Reference Manual
52
Chapter 4
Positioning the leads
Using epicardial leads
A variety of surgical approaches can be used to implant epicardial leads, including a limited left thoracotomy or median sternotomy. A typical placement may use an anterior right ventricular patch as the RV (HVB) and a posterolateral left ventricular patch as SVC (HVX).
Follow the general guidelines below for positioning epicardial leads:
If unipolar epicardial pacing leads are used, position the electrodes about 1 to 2 cm apart to reduce electromagnetic interference, and route the leads together with several loose twists.
Suture the smooth face of each patch lead against the epicardium or pericardium in locations that produce optimal defibrillation.
Place the patches so that they encompass the maximum amount of cardiac mass and they have approximately equal amounts of mass between them.
Ensure that the patches do not overlap and the electrode portions do not touch.
Avoid placing extra-pericardial patches over the phrenic nerve.
Surgical incisions
A single-incision submuscular or subcutaneous approach is recommended when the ICD is implanted in the pectoral region. Make the implant pocket about 1.5 times the size of the ICD.
Submuscular implant – An incision extending over the deltoid-pectoral groove typically provides access to the cephalic and subclavian veins as well as the implant pocket. Place the ICD sufficiently medial to the humeral head to avoid interference with shoulder motion.
Subcutaneous implant – A transverse incision typically permits isolation of the cephalic vein. Place the ICD far medially to keep the leads away from the axilla. Make sure that the upper edge of the ICD remains inferior to the incision.
Maximo DR 7278 Reference Manual

Testing sensing and pacing thresholds

Testing sensing and pacing thresholds
Sensing and pacing tests include the following measurements:
EGM amplitude
slew rate
pacing threshold
pacing lead impedance
Medtronic recommends that you use a Model 2290 or 8090 Analyzer lead analysis device to perform sensing and pacing measurements. If you use a Pacing System Analyzer (PSA), perform both atrial and ventricular measurements via the ventricular channel of the PSA.
Refer to the technical manual for the Analyzer you use to find details on performing sensing and pacing measurements.
Parameters
Measured sensing and pacing values must meet the following specific requirements at implant.
Implanting the ICD
53
Considerations
Table 4-1. Sensing and pacing values at implant
Measurement Acute Transvenous Leads Chronic Leads
R- wave amplitude 5 mV 3 mV
P- wave amplitude 2 mV 1 mV
Slew rate:
atrial 0.5 V/s 0.3 V/s
ventricular 0.75 V/s 0.5 V/s
a
Capture threshold
atrial 1.5 V 3.0 V
ventricular 1.0 V 3.0 V
a
At 0.5 ms pulse width
:
When measuring sensing and pacing values, measure between the tip (cathode) and ring or coil (anode) of each bipolar pacing/sensing lead.
Maximo DR 7278 Reference Manual
54
Chapter 4

Connecting the leads to the ICD

For unipolar epicardial pacing leads, either electrode can be the cathode; use the configuration that yields the lower pacing threshold.
Note: Do not measure the intracardiac EGM telemetered from the ICD to assess sensing.
Connecting the leads to the ICD
For more detailed information about lead/connector compatibility, see the Maximo DR 7278 Implant Manual, or contact Medtronic Technical Services at 1-800-723-4636.
Table 4-2. Lead connections
A
V
Can
SVC
RV
Device Port
SVC DF-1 HVX
RV DF-1 HVB
Can n/a HVA, Can
V IS-1 bipolar
74lead.eps
A IS-1 bipolar
Connector Type
Software Name
Warning: Loose lead connections may result in inappropriate
sensing and failure to deliver necessary arrhythmia therapy.
Caution: Use only the torque wrench supplied with the device. It is designed to prevent damage to the device from overtightening a setscrew.
For easier lead insertion, insert the ventricular IS-1 leg before the other legs.
Maximo DR 7278 Reference Manual

Testing defibrillation operation and effectiveness

How to connect the lead to the device
1
a
2
b
74SetScrew.eps
74LeadTIp.eps
Implanting the ICD
1. Insert the torque wrench into the appropriate setscrew.
a. If the port is obstructed, retract the
setscrew to clear it. Take care not to disengage the setscrew from the connector block.
b. Leave the torque wrench in the setscrew
until the lead is secure. This allows a pathway for venting trapped air when the lead is inserted.
2. Push the lead or plug into the connector port until the lead pin is clearly visible in the pin viewing area. No sealant is required, but sterile water may be used as a lubricant.
3. Tighten the setscrew by turning clockwise until the torque wrench clicks.
4. Tug gently on the lead to confirm a secure fit. Do not pull on the lead until all setscrews have been tightened.
5. Repeat these steps for each lead.
55
Testing defibrillation operation and effectiveness
Demonstrate reliable defibrillation effectiveness with the implanted lead system by using your preferred method to establish that a 10 J (minimum) safety margin exists.
Note: If the 10 J (minimum) safety margin cannot be ensured, see “Solving tachyarrhythmia therapy problems” on page 354.
High voltage implant values
Measured values must meet the following requirements at implant.
Table 4-3. High voltage therapy values at implant
Measurement Acute or Chronic Leads
V. Defib impedance
SVC (HVX) impedance (if applicable)
Defibrillation threshold
20 - 200 ohms
20 - 200 ohms
25 J
Maximo DR 7278 Reference Manual
56
Chapter 4
Testing defibrillation operation and effectiveness
Warning: Ensure that an external defibrillator is charged for a rescue shock.
How to prepare for defibrillation threshold testing
1. Place the programming head over the ICD, start a patient session, and interrogate the device, if you have not already done so.
2. Observe the Marker Channel telemetry annotations and the programmer ECG display to verify that the ICD is sensing properly.
3. Conduct a manual Lead Impedance Test lead connections. Perform this test with the ICD in the surgical pocket and keep the pocket very moist. If the impedance is out of range, perform one or more of the following tasks:
Recheck lead connections and electrode placement.
Repeat the measurement.
Inspect the bipolar EGM for abnormalities.
Measure the defibrillation impedance with a manual test shock.
4. Program the ICD to properly detect VF with an adequate safety margin (1.2 mV sensitivity).
a
See “Measuring lead impedance” on page 320.
a
to verify the defibrillation
Maximo DR 7278 Reference Manual
Testing defibrillation operation and effectiveness
How to perform defibrillation threshold testing
3
2
1
4
6
5
7
8
Implanting the ICD
1. Select Tests > EP Study.
2. Select either 50 Hz BURST or T-shock induction.
3. Select [Resume at BURST] or [Resume at DELIVER].
4. Select [Adjust Permanent...].
5. Program VF Enable On.
6. Program the automatic therapy energy settings. Therapies 2-6 should be set to the maximum energy.
7. Select [Program].
8. Select [Close].
57
Maximo DR 7278 Reference Manual
58
Chapter 4

Positioning and securing the ICD

11
9
12
13
10
14
9. If performing a T-Shock
induction, select the [Enable] checkbox.
10. Select [DELIVER], or [50 Hz BURST Press and Hold].
If necessary, you can abort an induction or therapy in progress by pressing [ABORT].
11. Observe the live rhythm monitor for proper post-shock sensing.
12. Use the [Adjust Permanent...] button to program the energy level.
13. Wait until the on-screen timer reaches 5 minutes, then repeat steps 9 through 12 as desired.
14. Select Params > Detection and program VF, FVT, and VT detection Off before closing.
Positioning and securing the ICD
Cautions: If no SVC electrode is implanted, the pin plug
provided with the device must be secured in the SVC port.
Program tachyarrhythmia detection Off before closing.
How to position and secure the device
Suture Hole Locations
Maximo DR 7278 Reference Manual
1. Ensure that each lead pin or plug is fully inserted into the connector block and that all setscrews are tight.
2. Coil any excess lead length beneath the device. Avoid kinks in the lead conductors.
3. Implant the device within 5 cm of the skin. This position optimizes the ambulatory monitoring operations.
4. Suture the device securely within the pocket to
78Suture.eps
minimize post-implant rotation and migration of the device. Use a surgical needle to penetrate the suture holes.

Completing the implant procedure

After implanting the device, X-ray the patient to verify the device and leads placement. To complete programming the device, select parameters that are appropriate for the patient.
How to complete programming the device
1. After closing the pocket, program detection On. Program ventricular tachyarrhythmia therapies On as desired.
2. Do not enable the Other 1:1 SVTs PR Logic detection criterion until the atrial lead has matured (approximately one month post implant).
3. Monitor the patient after the implant, and take X-rays as soon as possible to document and assess the location of the leads.
4. Program patient information. See “How to view and enter new patient information” on page 307.
5. Configure the Patient Alert feature. See “Using the Patient Alert feature” on page 259.
6. Set up data collection parameters. See “Setting up data collection” on page 273.
7. Interrogate the device after any spontaneous episodes to evaluate the detection and therapy parameter settings.
8. If the patient has not experienced spontaneous episodes, you may induce the clinical tachyarrhythmias using the non-invasive EP Study features to further assess the performance of the system. See Chapter 14, “Conducting Electrophysiologic Studies” on page 327.
9. Recheck pacing and sensing values, and adjust if necessary.
Implanting the ICD
Completing the implant procedure
59
Maximo DR 7278 Reference Manual
Conducting a patient follow-up
Patient follow-up guidelines 62
Verifying the status of the implanted system 62
Verifying accurate detection and appropriate therapy 63
Verifying effective bradycardia pacing 65
session
5
5
Maximo DR 7278 Reference Manual
62
Chapter 5

Patient follow-up guidelines

Patient follow-up guidelines
Schedule regular patient follow-up sessions to monitor the condition of the ICD and leads and to verify that the ICD is configured appropriately for your patient.
During the first few months after receiving a new device, the patient may require close monitoring. Schedule an office visit at least every three months.
The Quick Look screen, which is displayed after you interrogate the device, provides a good beginning for the follow-up review. Using this screen you can
verify that the device is functioning correctly.
review the clinical performance and long term trends.
print appropriate reports1 to compare the results to the patient’s history and to retain for future reference.
Note: The Checklist feature provides a standard list of tasks to perform at a complete follow-up visit. You can also customize your own checklists if you wish. See “Streamlining follow-ups with Checklist” on page 267 for more information.

Verifying the status of the implanted system

To verify that the ICD and leads are functioning correctly, review the following information from the Quick Look screen and perform follow-up tests as indicated:
Review the displayed battery voltage for comparison to the Elective Replacement Indicator value (see page 22). Remember that battery voltage may be low if high voltage charging has occurred within 24 hours.
Review the last full energy charge.
– For information about adjusting the capacitor formation
interval, see “Optimizing charge time” on page 214.
– If the programmer displays an Excessive Charge Time ERI,
the ICD should be replaced immediately.
1
See “Using Cardiac Compass to view long term clinical trends” on page 298 for information on this new report.
Maximo DR 7278 Reference Manual
Conducting a patient follow-up session

Verifying accurate detection and appropriate therapy

Review the defibrillation and pacing lead impedance values for inappropriate values or large changes since the last follow-up. See “Measuring lead impedance” on page 320.
Perform an EGM Amplitude test in each chamber for comparison to previous EGM Amplitude measurements. See “How to perform an EGM Amplitude test” on page 323.
To review longer term trends in sensing and impedance measurements, select the [>>] button from the lead impedance area of the Quick Look screen. The programmer displays a detailed history of automatic sensing and impedance measurements. See “Taking a quick look at device activity” on page 257.
Verifying accurate detection and appropriate therapy
To verify that the ICD is providing effective tachyarrhythmia detection and therapy, review the following information from the Quick Look screen and investigate as indicated:
Review Quick Look Observations that relate to patient history and device operation. To display more detailed information about any observation, select the observation and then select the [>>] button.
Review any Patient Alerts listed in the Observations of the Quick Look screen. For the most detailed information about Patient Alerts, select Patient Alert from the Data icon and select [Events].
Check stored episode records for appropriate sensing and detection of arrhythmias. See “Viewing episode data” on page 284.
Check stored SVT episode records for appropriate identification of SVTs.
63
Considerations
Review the following information before verifying detection and therapy.
Flashback memory – In addition to the episode text and stored electrograms, use Flashback memory and interval plots to help investigate the accuracy and specificity of ventricular detection.
Maximo DR 7278 Reference Manual
64
Chapter 5
Verifying accurate detection and appropriate therapy
Episode misidentification – If the episode records indicate that false detections have occurred, the Sensing Integrity counter may help in determining the prevalence of oversensing. For more information, see “Sensing integrity counter” on page 279.
If the ICD is oversensing, consider these programming options:
Increase the Pace Blanking value.
Increase the sensitivity threshold.
Caution: Do not re-program the ICD to decrease oversensing without assuring that appropriate sensing is maintained. See “Setting up sensing” on page 73.
If the episode records reveal that a stable monomorphic VT has been identified and treated as VF, consider these options to improve the detection accuracy:
Review the Interval Plot for the episode, and adjust VF Interval, if necessary. Use caution when reprogramming the VF Interval, because changes to this value can adversely affect VF detection.
Consider enabling FVT via VF detection. See “Detecting FVT episodes” on page 88.
If the SVT episode records include episodes of true VT, review the SVT episode record to identify the SVT detection criterion that withheld detection. Adjust the SVT detection criteria parameters as necessary. See “Enhancing detection with PR Logic criteria” on page 98, and “Enhancing VT detection with the Stability criterion” on page 108.
Maximo DR 7278 Reference Manual
Conducting a patient follow-up session

Verifying effective bradycardia pacing

Verifying effective bradycardia pacing
To verify that the ICD is sensing and pacing appropriately, review the following information from the Quick Look screen and investigate as indicated:
Confirm that the patient is receiving adequate cardiac support for daily living activities.
Review the pacing conduction history for comparison to the patient history. A sharp increase in the paced beats percentage may indicate a need for investigation and analysis.
Review the recorded Mode Switch episodes for comparison to the patient’s atrial arrhythmia history. A dramatic increase in frequency or duration of atrial episodes may indicate a need for investigation and analysis.
To display more detailed information about the Mode Switch episodes, perform these steps: select Episodes and Counters from the Data icon; select the Mode Switch episodes from the listed episode counters; then select the [Open Data] button.
Review the Cardiac Compass report for comparison to patient history (see page 298).
Conduct pacing threshold tests (see page 317) to verify that the programmed pacing outputs provide a sufficient safety margin.
65
Considerations
Review the following information before verifying bradycardia pacing.
Atrial Pacing – If the conduction history shows a predominance of atrial pacing despite a healthy sinus response, consider these options to decrease the atrial pacing burden:
Decrease the Lower Rate.
Decrease the rate response or increase the activity threshold.
Maximo DR 7278 Reference Manual
66
Chapter 5
Verifying effective bradycardia pacing
Ventricular Pacing – If the ventricle is predominantly paced and the patient exhibits adequate ventricular response, consider these options:
Decrease the Lower Rate.
Increase the AV delays.
Conduction History – If the reported percentages in the conduction history do not add up to 100, the percentages may be rounded. Frequent premature contractions or A:V dissociation may also be the cause. Consider any of the following options:
Program the pacing mode to DDD or DDDR to promote A-V synchrony. (If appropriate, enable Mode Switch to keep the benefits of DDIR pacing during atrial high rate episodes.)
Enable Ventricular Rate Stabilization to smooth the heart rate following premature ventricular beats.
Maximo DR 7278 Reference Manual
Part III

Configuring the ICD for the patient

Maximo DR 7278 Reference Manual

Detecting tachyarrhythmias6

Detection overview 70
Setting up sensing 73
Detecting VF episodes 78
Detecting VT episodes 82
Detecting FVT episodes 88
Detecting tachyarrhythmia episodes with Combined Count 93
Monitoring episodes for termination or redetection 95
Enhancing detection with PR Logic criteria 98
Enhancing VT detection with the Stability criterion 108
Detecting double tachycardias 111
6
Detecting prolonged tachyarrhythmias with High Rate Timeout 112
Key terms 114
Maximo DR 7278 Reference Manual
70
Chapter 6

Detection overview

Detection overview
The device detects ventricular tachyarrhythmias (VF, VT, and FVT) by comparing the time intervals between sensed ventricular events to a set of programmable detection intervals. If enough intervals occur that are shorter than the programmed intervals, the device detects a tachyarrhythmia, and responds automatically with a programmed therapy. After delivering the therapy, the device either redetects the arrhythmia and delivers the next programmed therapy or detects episode termination.
To avoid detecting rapidly conducted SVTs (for example, sinus tachycardia or atrial fibrillation) as ventricular tachyarrhythmias, the device provides several detection enhancements, including PR Logic and Stability detection criteria.
Figure 6-1 shows how all of these detection features interact during initial detection. During redetection, the device does not apply the PR Logic detection criteria.
Note: Detection functions can be turned off by programming the VF Enable, FVT Enable, and VT Enable parameters to Off. For an example, see “How to program VF detection” on page 80.
Maximo DR 7278 Reference Manual
Detecting tachyarrhythmias
Figure 6-1. How detection features interact during initial detection
71
Detection overview
No
Ye s
No
No / suspended by High Rate Timeout
Ye s
Update counts and pattern information
Has High Rate Timeout suspended detection
Does Stability reset the VT event count?
Are one or more PR Logic criteria on?
Is the median ventricular interval less than the SVT
Ventricular Event
Is the interval in the
VF, FVT, or VT detection zone?
Ye s
enhancements?
No
(VT and FVT via VT detection only)
No
Has a tachyarrhythmia event count
reached an NID?
Ye s
Ye s
Limit?
No
Ye s
Tachy
Episode
Detected
Ye s
No
Is there a double tachycardia in progress?
No
Are one or more PR Logic criteria withholding
detection?
Maximo DR 7278 Reference Manual
Ye s
72
Chapter 6
Detection overview
Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops classifying and counting tachyarrhythmia intervals. Sensing and bradycardia pacing remain active, and the programmed detection settings are not modified.
Detection is suspended
when the device senses the presence of a strong magnet. The programmer head contains a magnet which suspends detection, but once telemetry between the device and programmer is established, detection resumes.
while performing any of the manual system tests, including Underlying Rhythm, Pacing Threshold, Lead Impedance, EGM Amplitude, and Charge/Dump. Detection automatically resumes once the test is complete.
while performing a T-Shock, 50 Hz Burst, Manual Burst, or PES Induction. You can choose to have the device automatically resume detection after delivering the induction.
when you deliver a Manual or Emergency therapy. You can resume detection by selecting the [Resume] button or removing the programming head from the device.
when you select the on-screen [Suspend] button. You can resume detection by selecting the [Resume] button or by removing the programming head from the device.
during the automatic daily lead impedance measurements. Detection resumes when the measurements are complete.
while the device is delivering an automatic tachyarrhythmia therapy (including capacitor charging for defibrillation and cardioversion). However, the device does continue to confirm the detected episode during charging. Detection resumes when the therapy is complete.
Note: The device suspends VT detection (and Combined Count detection; see page 93) for 17 events following a defibrillation therapy delivered in response to a detected VF.
during charging for Automatic Capacitor Formation. Detection resumes when charging is complete.
1
1
If the defibrillation therapy is delivered as a result of a High Rate Timeout Therapy operation, VT detection is not suspended (see page 114).
Maximo DR 7278 Reference Manual

Setting up sensing

The device provides bipolar sensing in both the atrium and ventricle via the sensing electrodes of the implanted atrial and ventricular leads. You can adjust the sensitivity to intracardiac signals using independent atrial and ventricular sensitivity settings. These settings define the minimum electrical amplitude recognized by the device as an atrial or ventricular sensed event.
Proper sensing is essential for the safe and effective use of the device. To provide appropriate sensing, the device uses:
auto-adjusting atrial and ventricular sensing thresholds
short (30 ms) cross-chamber blanking after paced events
no cross-chamber blanking after sensed events
See details about sensing on page 76.
Parameters
V. Sensitivity (mV) – Minimum amplitude of
electrical signal that registers as a sensed ventricular event.
Detecting tachyarrhythmias
Setting up sensing
* Medtronic nominal setting
0.15, 0.3*, 0.45, 0.6,
0.9, 1.2
73
Considerations
A. Sensitivity (mV) – Minimum amplitude of
electrical signal that registers as a sensed atrial event.
0.15, 0.3*, 0.45, 0.6,
0.9, 1.2, 1.5, 2.1
Review the following information before programming sensing parameters.
Dual chamber sensing and bradycardia pacing modes – The device senses in both the atrium and the ventricle at all times, except when the programmed bradycardia pacing mode is DOO or VOO. When the pacing mode is programmed to DOO or VOO mode, there is no sensing in the ventricle. In order to program either DOO or VOO mode, you must first disable detection.
Sensitivity thresholds – The programmed atrial and ventricular sensitivity thresholds apply to all features related to sensing, including detection and bradycardia pacing.
Maximo DR 7278 Reference Manual
74
Chapter 6
Setting up sensing
Bradycardia pacing and sensing – A combination of high pacing pulse width or high amplitude with a low sensitivity threshold may cause inappropriate sensing across chambers or in the same chamber. Programming a lower pulse width, lower amplitude, longer pace blanking, or a higher sensitivity threshold may eliminate this inappropriate sensing.
Recommended ventricular sensitivity threshold – A ventr icular sensitivity threshold of 0.3 mV is recommended to maximize the probability of detecting VF and to limit the possibility of oversensing and cross-chamber sensing.
High ventricular sensitivity threshold – Setting V. Sensitivity to a value greater than 0.6 mV is not recommended except for testing. Doing this may cause undersensing, which can cause any of the following situations:
delayed or aborted cardioversion therapy
delayed defibrillation therapy (when VF confirmation is active)
asynchronous pacing
underdetection of tachyarrhythmias
Low ventricular sensitivity threshold – If you set V. Sensitivity to its most sensitive value of 0.15 mV, the device will be more susceptible to EMI, cross-chamber sensing, and oversensing.
Recommended atrial sensitivity threshold – An atrial sensitivity threshold of 0.3 mV is recommended to optimize the effectiveness of PR Logic detection criteria and atrial pacing operations, while limiting the possibility of oversensing and cross-chamber sensing.
High atrial sensitivity threshold – If you set the A. Sensitivity value too high, the device may not provide reliable sensing of P-waves during SVTs and sinus rhythm.
Low atrial sensitivity threshold – If you set the A. Sensitivity value to its most sensitive value of 0.15 mV, the device will be more susceptible to EMI, far-field R-wave sensing, and oversensing.
Maximo DR 7278 Reference Manual
Detecting tachyarrhythmias
Setting up sensing
Testing sensitivity after reprogramming – If you change the ventricular sensitivity threshold, evaluate for proper sensing and detection by inducing VF and allowing the device to automatically detect and treat the arrhythmia.
Atrial pacing and ventricular sensing – If you program the device to an atrial pacing mode, make sure that it does not sense atrial pacing pulses as ventricular events.
Sensing during VF – Always verify that the device senses properly during VF. If the device is not sensing or detecting properly, program detection and therapies off, and evaluate the system (making sure to monitor the patient for life-threatening arrhythmias until you enable detection and therapies again). You may need to reposition or replace the ventricular sensing lead to achieve proper sensing.
Atrial lead selection – Atrial leads with minimal tip-to-ring spacing may reduce far-field R-wave sensing.
Repositioning the atrial lead – You may need to reposition or replace the atrial sensing lead if reprogramming the atrial sensitivity threshold does not provide reliable atrial sensing during SVTs and sinus rhythm.
75
How to program sensitivity
1. Select Params > Detection.
2. Select the desired A. Sensitivity and V. Sensitivity parameters.
3. Select [PROGRAM].
2
3
1
Maximo DR 7278 Reference Manual
76
Chapter 6
Setting up sensing
Details about sensing
Auto-adjusting sensitivity thresholds
The device automatically adjusts the sensitivity thresholds after certain paced and sensed events to help reduce oversensing from T-waves, cross-chamber events, and pacing. Figure 6-2 shows how sensitivity thresholds are adjusted after different types of events.
Figure 6-2. Auto-adjusting sensitivity thresholds
A S
Sensitivity Threshold
4
V P
5
74Autoadjust.eps
Rectified and
Filtered A. EGM
Rectified and
Filtered V. EGM
Marker Channel
1 2
A S
3
A P
V S
V S
1 After an atrial sensed event, the atrial sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed value, decay constant: 200 ms).
2 After a ventricular sensed event, the ventricular sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed value, decay constant: 450 ms).
3 After an atrial paced event, the device does not adjust the atrial
sensitivity threshold. The ventricular sensitivity threshold increases by 0.45 mV (decay constant: 60 ms).
4 After a ventricular paced event, the atrial sensitivity threshold
increases to 4x the programmed value (maximum: 1.8 mV, immediate return after 60 ms).
5 After the ventricular pace blanking period is finished, the ventricular
threshold increases to 4.5x the programmed value (maximum:
1.8 mV, decay constant: 450 ms).
a
The exponential decay continues through a subsequent ventricular pacing pulse and its blanking period.
b
If the programmed sensitivity value exceeds 0.3 mV (ventricular) or 1.2 mV (atrial), the threshold is not adjusted.
Maximo DR 7278 Reference Manual
a
b
b
Detecting tachyarrhythmias
Setting up sensing
Blanking periods
During a blanking period, the device does not sense electrical signals. This helps prevent sensing of device pacing, cardioversion and defibrillation pulses, post-pacing depolarization, T-waves, and multiple sensing of the same event. The blanking periods following paced events are longer than those following sensed events to avoid sensing the atrial and ventricular depolarizations.
Notes:
To enhance sensing and detection during tachyarrhythmias, the device does not apply cross-chamber blanking (sense blanking in the opposite chamber) after a sensed event.
Atrial sensing is still active during the Post-Ventricular Atrial Blanking (PVAB) period (see “Post-Ventricular Atrial Blanking Period” on page 170).
Table 6-1 shows the duration of the fixed blanking periods. For information on programmable pace blanking periods, see page 158.
Table 6-1. Fixed blanking periods
Cross-chamber blanking after atrial or ventricular pacing pulse 30 ms
77
Atrial blanking after sensed atrial event 100 ms
Ventricular blanking after sensed ventricular event 120 ms
Atrial and ventricular blanking after delivered cardioversion or defibrillation therapy
Maximo DR 7278 Reference Manual
520 ms
78
Chapter 6

Detecting VF episodes

Detecting VF episodes
Refractory periods
During a refractory period, the device senses normally, but classifies sensed events as refractory and limits its response to these events. Pacing refractory periods prevent inappropriately sensed signals, such as far-field R-waves (ventricular events sensed in the atrium) or electrical noise, from triggering certain pacing timing intervals.
Synchronization refractory periods help prevent the device from delivering cardioversion and defibrillation therapies at inappropriate times. See “Synchronizing defibrillation without confirming VF” on page 125 and “Synchronizing cardioversion after charging” on page 145.
Note: Refractory periods do not affect tachyarrhythmia detection.
The device detects VF episodes by examining the cardiac rhythm for short ventricular intervals. If a predetermined number of intervals occurs that are short enough to be considered VF events, the device detects VF and delivers the first programmed VF therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing.
See details about VF detection on page 80.
Parameters
VF Detection Enable – Turns VF
detection on or off.
VF Interval (ms) – V-V intervals shorter than this value are counted as VF events.
VF Initial NID – Number of Intervals to Detect: number of VF events the device must count to detect a VF episode.
VF Redetect NID – Number of Intervals to Redetect: number of VF events the device must count to redetect a continuing VF after a therapy.
Maximo DR 7278 Reference Manual
* Medtronic nominal setting
On*, Off
240, 250, . . ., 320*, . . .,400
12/16, 18/24*, 24/32, 30/40, 45/60, 60/80, 75/100, 90/120, 105/140, 120/160
6/8, 9/12, 12/16*, 18/24, 21/28, 24/32, 27/36, 30/40
Considerations
Detecting tachyarrhythmias
Detecting VF episodes
Review the following information before programming VF Detection parameters.
VF Interval minimum setting – To ensure proper VF detection, you should not program the VF Interval less than 300 ms.
VF Interval maximum setting – Programming the VF Interval to a value greater than 350 ms may cause inappropriate detection of rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter, which is more sensitive than the consecutive VT event counter.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
79
Enabling VF detection – When VF Detection Enable is programmed On for the first time, the device
enables Automatic Capacitor Formation
starts recording Cardiac Compass data
starts recording lead performance trends (starting at 3:00 AM, by the device clock)
clears all brady pacing counters
VF detection and PR Logic criteria – You can program the device to exclude rapidly conducted SVTs from VF Detection by enabling the PR Logic detection criteria. Note that the SVT Limit must be programmed shorter than the VF Interval in order for the PR Logic criteria to affect VF detection. See “Enhancing detection with PR Logic criteria” on page 98.
Double tachycardia detection – When any PR Logic detection criteria are enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT). See “Detecting double tachycardias” on page 111.
Maximo DR 7278 Reference Manual
80
Chapter 6
Detecting VF episodes
Restrictions
How to program VF detection
2
Review the following information before programming VF Detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for bradycardia pacing and tachyarrhythmia detection. See “Parameter interlocks” on page 392.
VF detection backup – To ensure VF Detection backup during VT and FVT episodes, if VF Detection is off, both VT Detection and FVT Detection must also be off.
To program VF detection:
1. Select Params > Detection.
2. Select the desired values for VF Enable, VF Initial NID, VF Redetect NID, and VF Interval.
3. Select [PROGRAM].
Details about VF detection
The device detects VF by counting the number of VF events, which are V-V intervals shorter than the programmed VF Interval. On each event, the device counts the number of recent VF events. The number of recent events examined is called the VF detection window. The size of the VF detection window is the second number in the programmed VF NID (for example, 24 events if the VF Initial NID is 18/24).
Maximo DR 7278 Reference Manual
1
3
The threshold for detecting VF is the first number in the programmed VF NID (for example, 18 events if the VF Initial NID is 18/24). This threshold is always 75% of the VF detection window. That is, if 75% of the events in the VF detection window are VF events, the device detects a VF episode (see Figure 6-3).
After the device detects VF, it delivers the first programmed VF therapy. Following the therapy, if the number of VF events reaches the programmed VF Redetect NID, the device redetects VF and delivers the next programmed VF therapy.
Note: The device can also detect VF Episodes via the Combined Count detection criterion (see page 93).
Figure 6-3. Device detects VF
ECG
Detecting tachyarrhythmias
Detecting VF episodes
1 2 3
81
A S
A S
A R
A R
A R
A R
A R
Marker Channel
FSFSFSFSFSFSFSFSFSF
F
F
S
VSVSV
D
S
VF Event Count
V S
VSV
FSFSF
S
1 2 3 4 5 6 6 7 8 9 10 11 12 13 1415 16 17 18
FSFSF
V
S
S
S
S
VF Interval
200 ms
1 VF starts, and the device begins counting VF events (intervals less than the programmed
VF Interval).
2 A ventricular interval occurs outside the VF detection zone. The VF event count is not
incremented.
3 The VF event count reaches the programmed VF NID value of 18 events out of 24, and the
device detects VF.
74VFDetection.eps
Maximo DR 7278 Reference Manual
82
Chapter 6

Detecting VT episodes

Detecting VT episodes
Parameters
The device detects VT episodes by examining the cardiac rhythm for short ventricular intervals. If enough intervals occur that are short enough to be considered VT events (but are not VF or FVT events), the device detects VT and delivers the first programmed VT therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing.
You can program the device to detect and record VT episodes without treating them with VT therapies by setting VT Detection Enable to Monitor. If a patient’s VT episodes are well-tolerated, this feature allows you to collect data about these episodes without delivering therapy or affecting VF detection.
See details about VT detection on page 84.
* Medtronic nominal setting
VT Detection Enable – Turns VT detection on or off, or enables VT monitoring.
On, Off*, or Monitor
VT Interval (Rate) (ms) – V-V intervals shorter than this value are counted as VT events.
VT Initial NID – Number of Intervals to Detect: number of VT events the device must count to detect a VT episode.
VT Redetect NID – Number of Intervals to Redetect: number of VT events the device must count to redetect a continuing VT after a therapy.
Considerations
Review the following information before programming VT Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Maximo DR 7278 Reference Manual
280, 290, . . ., 400*, . . ., 600
12, 16*, . . ., 52, 76, 100
4, 8, 12*, . . ., 52
Detecting tachyarrhythmias
Detecting VT episodes
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
VT Detection Enable, AFib/AFlutter, and Sinus Tach – When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and Sinus Tach parameters are also automatically set to On.
VT detection and Combined Count detection – When VT Detection is On, the device applies the Combined Count detection criterion to help speed detection of rhythms that fluctuate between detection zones. Combined Count detection is disabled if VT Detection is set to Off or Monitor. See “Detecting tachyarrhythmia episodes with Combined Count” on page 93.
VT detection and rapidly conducted SVTs – You can program the device to exclude rapidly conducted SVTs from VT detection by enabling the PR Logic or Stability detection criteria. See “Enhancing VT detection with the Stability criterion” on page 108, and “Enhancing detection with PR Logic criteria” on page 98.
83
Restrictions
Double tachycardia detection – When any PR Logic detection
criteria are enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT; see page 111).
Review the following information before programming VT detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for bradycardia pacing and tachyarrhythmia detection. See “Parameter interlocks” on page 392.
VF detection backup – To ensure VF detection backup during VT and FVT episodes, if VF Detection is off, both VT Detection and FVT Detection must also be off.
Maximo DR 7278 Reference Manual
84
Chapter 6
Detecting VT episodes
How to program VT detection
2
Details about VT detection
To program VT detection:
1. Select Params > Detection.
2. Select the desired values for VT Enable, VT Initial NID, VT Redetect NID, and VT Interval.
3. Select [PROGRAM].
1
3
The device detects VT by counting the number of consecutive VT events. A VT event is a V-V interval shorter than the programmed VT Interval but greater than or equal to the VF Interval. If the number of consecutive VT events reaches the programmed VT Initial NID, the device detects VT (see Figure 6-4).
The VT event count resets to zero whenever an interval occurs that is greater than or equal to the programmed VT Interval. The count remains at the current value if an interval is shorter than the programmed VF Interval.
After the device detects VT, it delivers the first programmed VT therapy. Following the therapy, if the VT event counter reaches the VT Redetect NID, the device redetects VT and delivers the next programmed therapy.
Note: The device can also detect VT Episodes via the Combined Count detection criterion (see page 93).
Maximo DR 7278 Reference Manual
Figure 6-4. Device detects VT
ECG
132
Detecting tachyarrhythmias
Detecting VT episodes
85
A S
A S
ARARARARARARA
R
Marker Channel
VT Event Count
V S
VSV
S
V
T S
10123456
TSTSTSTSTST
S
S
VT Interval
ECG
ARARARARARARARARARARARA
R
Marker Channel
T D
V S
74VTDetection.eps
VT Event Count
TSTSTSTSTSTSTSTST
78910111213141516
S
VT Interval
200 ms
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval, but greater than or equal to the VF Interval).
2 A ventricular interval occurs outside VT detection zone. The VT event count resets to zero.
3 The VT event count reaches the programmed VT NID of 16 events, and the device
detects VT.
Maximo DR 7278 Reference Manual
86
Chapter 6
Detecting VT episodes
VT monitoring
You can program the device to record VT episodes without delivering VT therapy by setting VT Detection to Monitor. When VT monitoring is enabled, the device detects VT episodes but does not deliver VT therapy (see Figure 6-5). Instead, it records VT episodes, labeling them as “monitored,” and waits for episode termination to occur.
When VT Detection is set to Monitor, several detection operations work differently.
VT event counting – Before the device detects an episode, it counts VT events normally. However, once the VT Initial NID is reached, the device sets the VT event count to zero and suspends VT event counting for the rest of the episode.
VF and FVT detection – VF and FVT detection operate as if VT detection is off. Specifically, Combined Count detection is disabled, and FVT via VT detection is not selectable. If a monitored VT episode accelerates into the FVT or VF detection zone, the device applies the VF Initial NID to detect the new tachyarrhythmia. Once an episode is in progress, VT event counting doesn’t resume until the episode ends.
Caution: Programming the VF Interval greater than 350 ms may result in inappropriate detection of rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter, which is more sensitive than the consecutive VT event counter.
PR Logic and Stability criteria – Before the device detects a tachyarrhythmia episode, the PR Logic and Stability criteria, if turned on, are applied. If a monitored VT episode accelerates into the FVT or VF detection zone, the device continues to apply PR Logic criteria as initial VF or FVT detection begins. However, because the Stability feature does not affect VF detection or FVT via VF detection, it is not applied.
Episode termination – The device compares ventricular intervals to the VT Interval to identify when a VT monitored episode has ended. However, if a VF episode or FVT via VF episode occurs when VT monitoring is enabled, the device compares ventricular intervals to the VF Interval to identify episode termination.
Maximo DR 7278 Reference Manual
Figure 6-5. Device detects and monitors VT
ECG
Detecting tachyarrhythmias
Detecting VT episodes
2 31
87
A S
A
ARARA
R
A
R
R
ARARARARA
A R
R
Marker Channel
VT Event Count
V S
VSTSTSTST
1234
TSTSTSTDTSTST
S
13 14 15 16 0 0 0
S
74VTmonitor.eps
VT Interval
200 ms
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval but greater than or equal to the VF Interval).
2 The VT event count reaches the programmed VT NID of 16 events, and the device
detects VT.
3 After detecting the VT episode, the device resets the VT event count to zero and monitors
the episode until termination.
Maximo DR 7278 Reference Manual
88
Chapter 6

Detecting FVT episodes

Detecting FVT episodes
Parameters
The device detects episodes of Fast Ventricular Tachycardia (FVT) by examining the cardiac rhythm for short ventricular intervals. If enough intervals occur in the programmed FVT detection zone, the device detects FVT and delivers the first programmed FVT therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing. To make sure it delivers sufficiently aggressive therapies, the device can merge the programmed detection zones during redetection to increase sensitivity.
See details about FVT detection on page 90.
* Medtronic nominal setting
FVT Detection Enable – Enables FVT detection via the VF or the VT detection algorithm.
Off*, via VF, or via VT
Considerations
FVT Interval (Rate) (ms) – V-V
intervals between this value and the programmed VF Interval are marked as FVT events.
200, 210, . . ., 600
Review the following information before programming FVT Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
Maximo DR 7278 Reference Manual
Detecting tachyarrhythmias
Detecting FVT episodes
FVT detection enable – Your choice for an appropriate setting for FVT Detection should depend on the patient’s VF and VT cycle lengths. After determining a reliably sensitive VF Interval, consider the following suggestions:
If the patient presents with a clinical VT interval in the VF zone, select via VF to ensure reliable detection of VF. (VT Detection need not be enabled at all.)
If the patient presents with two clinical VTs, both outside the VF zone, select via VT to allow for correct classification of the faster VT and to offer a separate therapy regimen for each VT.
If the patient presents with only one clinical VT which is outside the VF zone, select VF and VT Detection only, and set FVT Enable to Off.
FVT detection and PR Logic criteria – You can program the device to exclude rapidly conducted SVTs from FVT Detection by enabling the PR Logic detection criteria. Note that the SVT Limit must be programmed shorter than the VF Interval for the PR Logic criteria to affect FVT via VF detection.
89
Restrictions
Double tachycardia detection – When any PR Logic detection
criteria is enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT, see page 111).
Review the following information before programming FVT Detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To ensure reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for bradycardia pacing and tachyarrhythmia detection. See “Parameter interlocks” on page 392.
VF detection backup – To ensure VF Detection backup during VT and FVT episodes, VT and FVT Detection cannot be on unless VF Detection is also on.
Maximo DR 7278 Reference Manual
90
Chapter 6
Detecting FVT episodes
How to program FVT detection
2
FVT detection – To ensure reliable ventricular tachyarrhythmia
detection, the programmer regulates the values available for the FVT parameter as follows:
VT Detection must be set to On if FVT Detection is set to via VT.
If FVT Detection is set to via VF, the FVT Interval must be programmed to a value shorter than the VF Interval.
If FVT Detection is set to via VT, the FVT Interval must be programmed to a value greater than the VF Interval and less than or equal to the VT Interval.
To program FVT detection:
1. Select Params > Detection.
2. Select the desired values for FVT Enable and FVT Interval.
3. Select [PROGRAM].
Details about FVT detection
You can program the device to detect FVT episodes via the VF or VT detection zone and NID.
When FVT Detection is set to via VF, a V-V interval within the FVT detection zone is marked as an “FVT via VF” event. When the VF NID is reached, the device reviews the last eight intervals:
If any of the last eight intervals are in the VF zone, it detects the episode as VF.
If all of the last eight intervals are outside the VF zone, it detects the episode as FVT (see Figure 6-6).
Maximo DR 7278 Reference Manual
1
3
When FVT Detection is set to via VT, a V-V interval within the FVT detection zone is marked as an “FVT via VT” event. When the VT NID is reached, the device reviews the last eight intervals:
If any of the last eight intervals are in the VF or FVT zones, it detects the episode as FVT.
If all of the last eight intervals are outside the FVT and VF zones, it detects the episode as VT.
Note: The device can also detect FVT episodes via the Combined Count detection criterion (see page 93).
Figure 6-6. Device detects FVT via VF
Detecting tachyarrhythmias
Detecting FVT episodes
91
1 2
3
ECG
A S
A S
ARARA
A R
ARARARARARARARARA
R
Marker Channel
VF Event Count
V S
V S
V
T
T
S
F
1 1 2345
TFTFTFT
S
TFT
F
TFTFTFTFTFV
F
13 14 15 16 17 18
VF and FVT Intervals
1 A fast ventricular tachycardia starts, and the first event falls into the FVT detection zone.
2 The second event of the FVT episode has an interval that falls into the VT zone. The VF
event count is not incremented.
3 The device detects FVT after the VF event count reaches the VF Initial NID.
R
S
74FvtDetect.eps
200 ms
Maximo DR 7278 Reference Manual
92
Chapter 6
Detecting FVT episodes
Figure 6-7. FVT zone merging
Zone merging after detection
To ensure the device delivers sufficiently aggressive therapies during an extended or highly variable tachyarrhythmia episode, the device merges detection zones during redetection in some instances, as shown in Figure 6-7. The merged zone configuration uses the event counting and therapies for the faster arrhythmia and remains in effect until episode termination.
FVT set to “via VF” FVT set to “via VT”
Before detection:
After VF detection:
After FVT detection:
VF
FVT
VT
VF and FVT zones merge, leaving a larger VF zone.
VF
FVT
VT
VF
FVT
VT
VT and FVT zones merge, leaving a larger FVT zone.
VF
FVT
VT
All zones remain unchanged. VT and FVT zones merge, leaving a
larger FVT zone.
VF
FVT
VT
Detection Intervals: VF Interval: 320 ms, FVT Interval: 280 ms / 360 ms , VT Interval: 400 ms
VF
FVT
VT
FVT1.eps
FVT2.eps
FVT3.eps
Maximo DR 7278 Reference Manual
Detecting tachyarrhythmias

Detecting tachyarrhythmia episodes with Combined Count

Detecting tachyarrhythmia episodes with Combined Count
Because the device counts VF and VT events separately, rhythms with variable cycle lengths can cause both event counts to increment during an episode. To prevent these rhythms from delaying detection, the device automatically enables the Combined Count detection criterion if both VF and VT detection are programmed On.
The Combined Count criterion compares the sum of the VF and VT event counts to the Combined Number of Intervals to Detect (CNID), which the device calculates automatically from the programmed VF NID values. If the CNID is met, the device reviews the recent intervals to determine if the episode should be treated as a VF, FVT, or VT episode. The Combined Count criterion applies during both initial detection and redetection.
Details about Combined Count detection
The Combined Count detection algorithm expedites detection or redetection of ventricular tachyarrhythmias with ventricular intervals that fluctuate between the VF and VT detection zones. When VT detection is on, the device applies Combined Count detection, which tracks the combined number of VT and VF events counted. If this sum reaches the Combined Number of Intervals to Detect (CNID), the device detects VF, FVT, or VT. Combined Count detection also applies to redetected episodes.
93
Note: Combined Count detection is off when VT detection is set to Monitor or Off.
If the VF event counter reaches six, the device automatically applies the Combined Number of Intervals to Detect (CNID). The CNID is calculated by multiplying the current VF NID (Initial or Redetect) by 7/6 and rounding down. Table 6-2 shows the CNID values that correspond to each VF NID value.
Table 6-2. CNID values for each initial or redetect VF NID value
VF NID CNID VF NID CNID
6/8 7 21/28 24
9/12 10 24/32 28
12/16 14 27/36 31
18/24 21 30/40 35
Maximo DR 7278 Reference Manual
94
Chapter 6
Detecting tachyarrhythmia episodes with Combined Count
Combined Count detection is fulfilled when the sum of the VF and VT event counts equals or exceeds the CNID. The device then reviews the last eight intervals and classifies the episode as
VF, if any of the last eight were in the VF zone.
FVT, if FVT Detection is enabled and none of the last eight was in the VF zone, but one or more was in the FVT zone.
VT, if all eight were outside the VF zone (and FVT zone, if FVT detection is enabled).
Figure 6-8. Device detects VF with the Combined Count criterion
1
2
3
ECG
A
ARARARARARARARARARARARA
S
ARARARARARARA
R
R
A
R
R
A
Marker Channel
V
VSF
VF Event Count
VT Event Count
Combined Count
F
S
S
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
1 2 3 4
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
FSFSFSFSF
S
FST
S
SFSFSFSFSFSFSFSFS
TST
F
FDV
S
S
S
VF and VT Interval
200 ms
1 A slow VF episode starts, with a ventricular cycle length that varies between the VF and VT
detection zones.
2 When a VT event occurs, the device increments the VT event count and the Combined
Count.
3 The device detects VF even though the VF event count hasn’t yet reached the VF Initial NID
(18/24 in this example). The Combined Count reaches the CNID value of 21 first.
74Combined.eps
Maximo DR 7278 Reference Manual
Detecting tachyarrhythmias

Monitoring episodes for termination or redetection

Monitoring episodes for termination or redetection
Once the device detects an arrhythmia, it considers the episode ongoing until it detects that the episode has ended. After delivering therapy, it monitors the ventricular rhythm using the programmed Redetect NIDs. If one of these NIDs is met, the device delivers the next programmed therapy for the detected arrhythmia.
See details about episode termination and redetection on
page 96.
Parameters
* Medtronic nominal setting
VF Redetect NID – Number of Intervals to Redetect: number of VF events the device must count to redetect a continuing VF after a therapy.
6/8, 9/12, 12/16*, 18/24, 21/28, 24/32, 27/36, 30/40
95
Considerations
VT Redetect NID – Number of
Intervals to Redetect: number of VT events the device must count to redetect a continuing VT after a therapy.
4, 8, 12*, . . . 52
Review the following information before programming redetection parameters.
Initial and Redetect NIDs – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
Maximo DR 7278 Reference Manual
96
Chapter 6
Monitoring episodes for termination or redetection
How to program redetection parameters
2
3
Details about episode termination and redetection
After a therapy is delivered, the device evaluates the ventricular rhythm to determine if the episode has terminated, is continuing, or has changed to a different arrhythmia.
1. Select Params > Detection.
2. Select the desired values for VT Redetect NID and VF Redetect NID.
3. Select [PROGRAM].
1
Episode termination
The device determines that the episode has terminated if one of the following conditions occurs:
eight consecutive ventricular intervals are greater than or equal to the programmed VT interval.
20 seconds elapse with no ventricular intervals shorter than the programmed VT interval.
After antitachycardia pacing therapy, the device begins evaluating intervals for episode termination on the first ventricular cycle. After cardioversion or defibrillation, the device begins evaluating intervals for episode termination on the second ventricular event. (Due to the extended post shock blanking, this event may be the third event on the electrogram.)
Note: Any subsequent detection after the end of the episode marks the start of a new episode.
1
VF interval if VT Detection is set to Off or Monitor, and the episode is a VF or an FVT via VF episode.
Maximo DR 7278 Reference Manual
1
1
Monitoring episodes for termination or redetection
Episode redetection
After the device detects a tachyarrhythmia episode and delivers a therapy, the device redetects an arrhythmia if the VF or VT event count reaches the Redetect NID or if the combined VF and VT event count reaches the Redetect CNID (see “Detecting tachyarrhythmia episodes with Combined Count” on page 93).
The device then delivers the next programmed therapy for the current arrhythmia and resumes monitoring for the outcome of that therapy. Figure 6-9 shows an example of redetection.
Figure 6-9. VT episode redetected after therapy
ECG
Detecting tachyarrhythmias
97
2 31
ARASASASASASARARARARARARARARARARARARA
Marker Channel
VT Event Count
VT Interval
T
TPT P
D
TPT
TPTPTPTPTPTSTSTSTSTSTSTSTSTSTSTST
P
016 0 1 2 3 4 5 6 7 8 9 10 11 12
1 A VT episode is detected, and the device delivers a Burst ATP therapy.
2 After therapy, the device continues to detect events in the VT zone.
3 When the VT event count reaches the VT Redetect NID, the device redetects the VT.
Notes:
The device suspends VT detection (and Combined Count detection) for 17 events following a defibrillation therapy delivered in response to a detected VF.
1
Suspending VT detection helps avoid detecting transient VTs that can follow high voltage therapies.
The PR Logic criteria are not applied during redetection. However, the Stability criterion may withhold detection or redetection of VT throughout an episode.
1
If the defibrillation therapy is delivered as a result of a High Rate Timeout Therapy operation, VT detection is not suspended (see page 114).
R
D
74Redetect.eps
200 ms
Maximo DR 7278 Reference Manual
98
Chapter 6

Enhancing detection with PR Logic criteria

VT acceleration
If the device redetects VT, it classifies the rhythm as accelerated if the average of the four intervals before redetection is at least 60 ms less than the average of the four intervals before initial VT detection. The most recent interval average is used to identify VT acceleration if VT is redetected again during the episode.
If the device redetects VF or an accelerated VT after an antitachycardia pacing sequence delivery, it skips the subsequent pacing therapy sequences for the duration of the episode and delivers the next therapy programmed for the current arrhythmia.
Enhancing detection with PR Logic criteria
The PR Logic detection criteria are designed to withhold inappropriate ventricular detection during episodes of rapidly conducted supraventricular tachycardia (SVT). The device analyzes the activation patterns and timing in both chambers using PR Logic pattern and rate analysis. This information helps identify evidence of atrial fibrillation, atrial flutter, sinus tachycardia, and other 1:1 SVTs. If this analysis indicates the presence of one or more of these rhythms, the device withholds detection.
For more information, see
“Details about PR Logic pattern and rate analysis” on page 102
“Details about the PR Logic detection criteria” on page 106
Parameters
Maximo DR 7278 Reference Manual
t
AFib/AFlutter – Identifies rapidly conducted
atrial fibrillation, atrial flutter, or atrial tachycardia.
Sinus Tach – Identifies sinus tachycardia. On
1:1 VT-ST Boundary (%)Threshold between
the retrograde and antegrade zones used by A-V pattern analysis. Allows customizing of the Sinus Tach criterion for patients with slow one-to-one conduction.
* Medtronic nominal setting
Ona, Off*
a
, Off*
35, 50*, 66, 75, 85
Detecting tachyarrhythmias
Enhancing detection with PR Logic criteria
* Medtronic nominal setting
99
Considerations
Other 1:1 SVTs – Identifies other one-to-one
SVTs where the atrial and ventricular activation are roughly simultaneous.
SVT Limit (ms) – Defines the minimum ventricular interval at which the device applies the PR Logic criteria.
a
When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and Sinus Tach parameters are also automatically set to On.
On, Off*
240, 250, . . ., 320*, 330, 340, . . ., 600
Review the following information before programming PR Logic parameters.
Cautions:
Before enabling the Other 1:1 SVTs criterion, ensure that the atrial lead has matured. This criterion could inappropriately withhold therapy if atrial sensing is compromised by an unstable or dislodged atrial lead.
Use caution when programming the Other 1:1 SVTs criterion in patients who exhibit slow 1:1 retrograde conduction during VF or VT. This criterion could inappropriately withhold VF/VT therapy in such patients. See “Pattern analysis of A-V and V-A intervals” on page 102.
Use caution when programming the 1:1 VT-ST Boundary. Incorrect programming of this parameter can result in inappropriate therapies or underdetection of tachyarrhythmias.
PR Logic criteria and double tachycardia detection – When any of the PR Logic criteria are enabled, the device also enables double tachycardia detection (VF, VT, or FVT in the presence of an SVT; see page 111).
Sinus Tach and 1:1 VT-ST Boundary – When enabling the Sinus Tach criterion in patients who exhibit slow 1:1 retrograde conduction during VF or VT or slow antegrade conduction during SVTs, make sure to set the 1:1 VT-ST Boundary parameter appropriately. See “Customizing PR Logic for patients with slow conduction” on page 100.
Maximo DR 7278 Reference Manual
100
Chapter 6
Enhancing detection with PR Logic criteria
VT Detection Enable, AFib/AFlutter and Sinus Tach – When you set VT Detection Enable to On or Monitor, the AFib/AFlutter and Sinus Tach parameters also automatically set to On.
SVT Limit – To ensure that therapy is delivered for hemodynamically compromising rates of any origin, the device always delivers therapy when the median ventricular interval is less than the programmed SVT Limit (nominally 320 ms) if VT, VF, or FVT detection criteria are satisfied.
VF Interval and SVT Limit – If you program an SVT Limit greater than the VF Interval, you are effectively disabling the PR Logic criteria for VF detection.
Customizing PR Logic for patients with slow conduction
Slow antegrade conduction during an SVT or slow 1:1 retrograde conduction during VT can adversely affect the accuracy of the Sinus Tach and Other 1:1 SVTs criteria. The 1:1 VT-ST Boundary parameter allows you to customize the Sinus Tach criterion for patients who exhibit slow conduction.
Note: Changing the 1:1 VT-ST Boundary parameter does not affect the operation of the Other 1:1 SVTs criterion. Use caution when enabling this criterion in patients who exhibit slow antegrade or retrograde conduction.
The 1:1 VT-ST Boundary parameter represents an A-V interval that is a percentage of a V-V interval. This value separates the retrograde and antegrade zones used by PR Logic pattern analysis operations. See “Pattern analysis of A-V and V-A intervals” on page 102.
If slow retrograde or antegrade conduction causes events to occur in the incorrect zone, you can use the 1:1 VT-ST Boundary parameter to increase the size of the appropriate zone. See Figure 6-10 on page 103.
If a patient exhibits slow antegrade or retrograde conduction and could benefit from a different 1:1 VT-ST Boundary setting, choose a new setting as follows:
If the patient exhibits long V-A intervals during VT with 1:1 retrograde conduction, select 35%.
Maximo DR 7278 Reference Manual
Loading...