Medtronic 7275 Reference Guide

GEM®III DR 7275
Dual Chamber Implantable Cardioverter Defibrillator
System Reference Guide
1
GEM® III DR System Reference Guide
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician.
A guide to the operation and programming of the 7275
®
GEM Implantable Cardioverter Defibrillator
III DR Dual Chamber
GEM, Marker Channel, Decision Channel, PR Logic, T-Shock, Quick Look, QuickLink, Patient Alert, Active Can, and Flashback are trademarks of Medtronic, Inc.
Table of Contents
Table of Contents
Introduction vii Abbreviations and Acronyms ix
1. Quick Reference
Primary Capabilities 1-2 Physical Characteristics 1-5 Electrical Characteristics 1-6 Magnet Application 1-10 Emergency Settings 1-11 Programmable Parameters Overview 1-12 Detection Parameters 1-13 Therapy Parameters 1-15 Bradycardia Pacing Parameters 1-17 Automatic Capacitor Formation Parameters 1-19 Data Collection Parameters 1-20 Patient Alert™ Monitoring System Parameters 1-21 Patient Information Parameters 1-23 EP Studies Parameters 1-24 Fixed Parameters 1-26 Notes 1-28
2. Emergency Therapy
Overview 2-2 Effect on System Operation 2-3 On-Screen and Display Panel Buttons 2-3 Emergency Defibrillation 2-4 Emergency Cardioversion 2-5 Emergency Fixed Burst Pacing 2-6 Emergency VVI Pacing 2-7
3. Overview
System Description 3-2 Detection and Therapy Overview 3-4 Disabling Detection and Therapies 3-9
4. Tachyarrhythmia Detection
Sensing 4-2 VF Detection 4-7 VT Detection 4-10 Combined Count Detection 4-13 VT Stability Option 4-15
GEM®III DR System Reference Guide
iii
Table of Contents
FVT Detection 4-17 Dual Chamber SVT Criteria 4-22 Termination and Redetection 4-35
5. Tachyarrhythmia Therapies
High Voltage Therapy Overview 5-2 Defibrillation 5-9 Cardioversion 5-17 Antitachycardia Pacing Therapy Overview 5-23 Programming ATP Therapies 5-25 Burst Pacing 5-27 Ramp Pacing 5-29 Ramp+ Pacing 5-31 Smart Mode 5-33 Progressive Episode Therapies 5-36
6. Bradycardia Pacing
Pacing Mode 6-2 Bradycardia Pacing Parameters 6-9 Post Shock Pacing Parameters 6-22 Mode Switch 6-25 Rate Responsive Pacing 6-29 Rate Adaptive AV 6-35 Non-Competitive Atrial Pacing 6-38 Single Chamber Hysteresis 6-42 Ventricular Rate Stabilization 6-44 PVC Response 6-47 PMT Intervention 6-49 Ventricular Safety Pacing 6-51
7. Programmer and Software
Connecting the Programmer Hardware 7-2 Setting Up and Using the Programmer 7-12 Using the Programmer Display Screen 7-22 Starting a Patient Session 7-30 Programming Parameter Values 7-32 Checklist 7-35 Live Rhythm Monitor 7-42 Recording Live Waveforms 7-54 Printing Reports 7-62 Changes This Session 7-67 Ending a Patient Session 7-68
iv
GEM®III DR System Reference Guide
8. Conducting System Tests
System Tests Overview 8-2 Underlying Rhythm Test 8-3 Pacing Threshold Test 8-4 Lead Impedance Test 8-6 Manual EGM Amplitude Test 8-9 Test Charge / Dump 8-11 Capacitor Formation 8-13
9. EP Studies
EP Studies Overview 9-2 Delivering Inductions 9-6 T-Shock™ 50 Hz Burst Induction 9-9 Manual Burst Induction 9-10 PES Induction 9-11 Delivering Manual Therapies 9-12 Manual Defibrillation Therapy 9-13 Manual Cardioversion Therapy 9-14 Manual ATP (Antitachycardia Pacing) Therapies 9-14
10. Data Collection
Data Collection Overview 10-2 Quick Look™ 10-4 Setting Up Data Collection 10-6 Counter Data 10-11 Episode Data 10-15 Patient Information 10-28 Battery and Lead Measurements 10-32 Device Status Indicators 10-36 Saving and Retrieving ICD Data 10-38
Induction 9-8
Table of Contents
11. Patient Alert™ Monitoring System
Patient Alert™ Monitoring System 11-2 Instructions for the Patient 11-7 Instructions for Clinic Personnel 11-9 Patient Alert™ Event Log 11-10
12. Implant and Follow-up
Pre-Implant Information 12-2 Operative Set-Up 12-4 Surgical Approach 12-6 Sensing and Pacing Measurements 12-8 Defibrillation Testing 12-9
GEM®III DR System Reference Guide
v
Table of Contents
Placing the ICD 12-15 Replacing the ICD 12-16 Pre-Discharge Evaluation 12-17 Follow-Up Recommendations 12-18
A. Appendix A - Warnings and Precautions
General Warnings A-2 Sterilization, Storage, and Handling A-3 Implantation and ICD Programming A-4 Lead Evaluation and Lead Connection A-5 Follow-up Testing A-6 ICD Explant and Disposal A-7 Environmental and Medical Therapy Hazards A-7 Home and Occupational Environments A-10
B. Appendix B - Parameter Interlocks
Bradycardia Pacing Parameter Interlocks B-2 Detection Parameter Interlocks B-6 Resolving Tachy / Brady Parameter Interlocks B-8
G. Glossary
I. Index
vi
GEM®III DR System Reference Guide

Introduction

Introduction
Before implanting the ICD, it is strongly recommended that you:
Refer to the product literature packaged with the ICD for information about prescribing the ICD.
Thoroughly read this manual and the technical manuals for the leads used with the device.
Discuss the procedure and the ICD system with the patient and any other interested parties, and provide them with any patient information packaged with the ICD.

Education

Medtronic invites physicians to attend an education seminar on the complete ICD system. The course includes indications for use, an overview of ICD system functions, implant procedures, and patient management.

Technical Support

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
GEM®III DR System Reference Guide
vii
Introduction

References

Automatic defibrillation was invented and patented in 1953 by Dr. F. Zacouto, who designed an external device that delivered a defibrillation impulse to the heart upon detection of very rapid ECG activity in combination with the absence of arterial pulsations.
The primary reference for background information is Zacouto FI, Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B, ed.
Pacing Diagnostic and Therapeutic Tools.
New York: Springer-Verlag;
1976: 212-218.
See these additional references for more background information:
Singer I, Ed.
Implantable Cardioverter-Defibrillator
Futura Publishing Co. 1994.
Singer I, Barold SS, Camm AJ, Eds.
Nonpharmacological Therapy of
Arrhythmias for the 21st Century: The State of the Art
Futura Publishing Co. 1998.
Estes M, Manolis AS, Wang P, Eds.
Cardioverter-Defibrillators
Kroll MW, Lehmann MH, Eds.
. New York, NY: Marcel Dekker, Inc. 1994.
Therapy: The Engineering-Clinical Interface
Implantable
Implantable Cardioverter-Defibrillator
. Norwell, MA: Kluwer
Academic Publishers 1996.
Cardiac
. Armonk, NY:
. Armonk, NY:
viii

Notice

This software is provided as an informational tool for the end user. The user is responsible for accurate input of patient information into the software. Medtronic makes no representation as to the accuracy or completeness of the data input into the software. MEDTRONIC SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTIAL OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE INFORMATION PROVIDED IN THE SOFTWARE.
GEM®III DR System Reference Guide

Abbreviations and Acronyms

A- Atrial
AF Atrial Fibrillation
AFib/AFlutter Atrial Fibrillation and/or Atrial Flutter
ARP Atrial Refractory Period
ATP Antitachycardia Pacing
AVP Atrial Vulnerable Period
BOL Beginning of Life
bpm beats per minute
CNID Combined (VT and VF) Number of Intervals to Detect
CV Cardioversion
DF / Defib Defibrillation
ECG Electrocardiogram
EGM Electrogram
EOL End of Life
ERI Elective Replacement Indicator
FDI Fibrillation Detection Interval
FTI Fast Ventricular Tachycardia Detection Interval
FVT Fast Ventricular Tachycardia
ICD Implantable Cardioverter-Defibrillator
J joules
-1
min
ms milliseconds
mV millivolts
NCAP Non-Competitive Atrial Pacing
NID Number of Intervals to Detect
PAC Premature Atrial Contraction
PAV Paced A-V Delay
PES Programmed Electrical Stimulation
PMT Pacemaker-Mediated Tachycardia
POR Power On Reset
PP P-P interval; an atrial interval
ppm paces or pulses per minute
reciprocal minutes; for example, pacing pulses per minute
Abbreviations and Acronyms
GEM®III DR System Reference Guide
ix
Abbreviations and Acronyms
PR P-R interval; an interval between a P-wave and the
PVAB Post Ventricular Atrial Blanking period
PVARP Post Ventricular Atrial Refractory Period
PVC Premature Ventricular Contraction
RAAV Rate Adaptive A-V delay
RCNID Combined (VT and VF) Number of Intervals to Redetect
RNID Number of Intervals to Redetect
RP R-P interval; an interval between an R-wave and the
RR R-R interval; a ventricular interval
SAV Sensed A-V Delay
ST / Sinus Tach Sinus Tachycardia
SVT Supraventricular Tachycardia
TDI Tachycardia Detection Interval
V volts
V- Ventricular
VF Ventricular Fibrillation
VF NID VF Number of Intervals to Detect
VF RNID VF Number of Intervals to Redetect
VRS Ventricular Rate Stabilization
VSP Ventricular Safety Pacing
VT Ventricular Tachycardia
VT NID VT Number of Intervals to Detect
VT RNID VT Number of Intervals to Redetect
subsequent R-wave
subsequent P-wave
x
GEM®III DR System Reference Guide
Quick Reference
1
1
Primary Capabilities 1-2
Physical Characteristics 1-5
Electrical Characteristics 1-6
Magnet Application 1-10
Emergency Settings 1-11
Programmable Parameters Overview 1-12
Detection Parameters 1-13
Therapy Parameters 1-15
Bradycardia Pacing Parameters 1-17
Automatic Capacitor Formation Parameters 1-19
Data Collection Parameters 1-20
Patient Alert™ Monitoring System Parameters 1-21
Patient Information Parameters 1-23
EP Studies Parameters 1-24
Fixed Parameters 1-26
Notes 1-28
GEM®III DR System Reference Guide
1-1
Quick Reference

Primary Capabilities

Primary Capabilities

Pacing Management

Dual chamber rate responsive bradycardia pacing up to 150 ppm (min
Rate Adaptive AV, Sensor Varied PVARP, PMT Intervention, Non-Competitive Atrial Pacing, Ventricular Safety Pacing
Ventricular Rate Stabilization after short ventricular intervals
DDIR Mode Switch during paroxysmal atrial episodes
-1
): DDDR, DDD, DDIR, DDI, VVIR, VVI, AAIR, AAI modes;

Dual Chamber VF/VT/FVT Detection

Ventricular fibrillation and ventricular tachycardia detection (VF, VT, and FVT)
Single chamber VT Stability criterion
PR Logic™ Pattern and Rate Analysis identifies Atrial Fibrillation/Atrial Flutter, Sinus Tachycardia, and Other 1:1 SVTs
Adjustable 1:1 VT-ST Boundary for patients with slow conduction

Tachyarrhythmia Therapy

Up to six (each) automatic VF, VT, and FVT therapies
Ventricular defibrillation and cardioversion, 30 joules delivered (35 joules stored) with 50% tilt
1-2
Burst, Ramp, and Ramp+ antitachycardia pacing
Smart Mode disables ineffective ATP therapies
Progressive Episode Therapies ensures increasingly aggressive therapy sequencing
Dual chamber synchronization of high voltage therapies avoids the atrium’s vulnerable period
GEM®III DR System Reference Guide
Quick Reference
Primary Capabilities

Patient Management

Audible Patient Alert™ monitoring system
Automatic, painless pacing and HV lead impedance testing
Lead performance trends report and display
Sensing integrity counter
Patient information that can be entered, viewed and printed via the programmer
Two simultaneous real time EGMs: atrial / ventricular / far-field
Two channels stored EGM: atrial / ventricular / far-field
Dual chamber Marker Channel™ telemetry
Supplementary Decision Channel annotations in stored EGM and on real time recordings that indicate detection algorithm status.
256 Kbyte data storage, including two EGMs, Markers, and Flashback™ Memory of A-A and V-V intervals
Save-to-Disk and Read-from-Disk patient data features
Table 1-1. GEM®III DR ICD Data Collection Summary
Tachy episodes
SVT/NST episodesa50 SVT/NST episodes: text, stored EGM, intervals, and
Brady episodes 53 mode switch episodes: episode text
EGM capacity 12 minutes (2 channels) or 22 minutes (1 channel)
EGM sources Ten options: atrial / ventricular / far-field
EGM options Store before onset; Store during charging
Flashback™ Memory
Detection counters Lifetime total, Since cleared, and Since last session
a
150 VF/VT/FVT episodes: text, stored EGM, intervals
episode summaries (usually only episode summaries are provided for NSTs)
2000 intervals (containing both A-A and V-V): before latest VF, before latest VT, and before interrogation
GEM®III DR System Reference Guide
1-3
Quick Reference
Primary Capabilities
Table 1-1. GEM®III DR ICD Data Collection Summary
Episode counters Episodes:
- VF, FVT, and VT
- Atrial Fibrillation / Atrial Flutter episodes
- Sinus Tach episodes
- Other 1:1 SVTs episodes
- Non-sustained tachy episodes
- Mode switch episodes Percentage Pacing:
- AS-VS, AS-VP, AP-VS, AP-VP percentages Additional Counters:
- Single PVCs and PVC runs
- Rate stabilization pulses and runs
Therapy Efficacy Counters
Patient Alert™ Events
a
All tachy, NST, and SVT episode data are stored in the same block of memory.
Counts for each VF, FVT, VT Therapy:
- Delivered
- Successful
- Unsuccessful
- Intervention Total number of aborted shocks
Up to 10 log entries: text and date for the first time an alert is triggered between interrogations.

Other features

1-4
Direct replacement for DF-1 / IS-1 systems, including all Medtronic “C,” “Cx,” and “D” ICDs
Medtronic icon-based user interface
Full-size report printing
Quick Look™ programming summary with Quick Links™ to stored episodes and observations
Non-invasive EP study functions: atrial inductions, ventricular inductions / manual therapies
Smart Auto Cap Formation extends the formation timer after therapeutic charging
Auto-Resume feature automatically resumes detection after an induction
Checklist feature provides streamlined, customizable navigation through sets of tasks for follow-up and implant visits
GEM®III DR System Reference Guide

Physical Characteristics

Volume 39.7 cc
Mass 78.1 g
H x W x D
Surface area of device can 69.0 cm
a
Quick Reference
Physical Characteristics
70.1 mm x 50.8 mm x 13.5 mm
2
Radiopaque ID
Materials in contact with human tissue
b
c
PJM
Titanium / polyurethane / silicone rubber
Battery Lithium silver vanadium oxide
Connector configuration Two IS-1 connectors for pacing and sensing,
Two DF-1 connectors for high voltage therapy, Active Can® electrode
A
IS-1
V
SVC
(HVX)
RV
(HVB)
DF-1
Can (HVA)
a
This measurement refers to the thickness (depth) of the can; grommets may protrude slightly beyond can surface.
b
Engineering series number follows the radiopaque code.
c
These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue.
GEM®III DR System Reference Guide
1-5
Quick Reference

Electrical Characteristics

Electrical Characteristics

Longevity

Table 1-2 below provides longevity estimates based on accelerated battery discharge data and device modeling with 60 ppm (min-1) pacing
rate, 2 V pacing pulse amplitude, 0.6 ms pacing pulse width, and 30 J delivered therapy energy
Table 1-2. GEM®III DR Projected Longevity in Years with 2 V Pacing Pulse Amplitude
and 0.6 ms Pacing Pulse Width
Percent pacing
0% Semi-Annual 7.0 7.0 7.0 7.0
15% Semi-Annual 6.7 6.9 6.8 7.0
50% Semi-Annual 6.1 6.6 6.5 6.8
100% Semi-Annual 5.5 6.2 6.1 6.5
Maximum Energy Charging Frequency
Quarterly 6.3 6.3 6.3 6.3
Quarterly 6.0 6.2 6.1 6.3
Quarterly 5.6 6.0 5.9 6.1
Quarterly 5.1 5.7 5.5 6.0
1
.
500 ohm pacing
impedance
DDD Pacing VVI Pacing DDD Pacing VVI Pacing
900 ohm pacing
impedance
1-6
GEM®III DR System Reference Guide
Quick Reference
Electrical Characteristics
Table 1-3 below provides longevity estimates based on accelerated battery discharge data and device modeling with 60 ppm (min-1) pacing
rate, 3 V pacing pulse amplitude, 0.4 ms pacing pulse width, and 30 J delivered therapy energy
Table 1-3. GEM®III DR Projected Longevity in Years with 3 V Pacing Pulse Amplitude
and 0.4 ms Pacing Pulse Width
Percent pacing
0% Semi-Annual 7.0 7.0 7.0 7.0
15% Semi-Annual 6.4 6.7 6.6 6.9
50% Semi-Annual 5.5 6.2 6.0 6.5
100% Semi-Annual 4.5 5.5 5.3 6.1
Maximum Energy Charging Frequency
DDD Pacing VVI Pacing DDD Pacing VVI Pacing
Quarterly 6.3 6.3 6.3 6.3
Quarterly 5.8 6.0 6.0 6.2
Quarterly 5.0 5.6 5.5 5.9
Quarterly 4.2 5.1 4.9 5.5
1
.
500 ohm pacing
impedance
900 ohm pacing
impedance
The following factors result in decreased longevity:
an increase in pacing rate, pacing amplitude or pulse width, the ratio of bradycardia paced to sensed events, or the charging frequency,
a decrease in pacing impedance, or
using the pre-onset EGM storage feature or Holter telemetry
To maximize ICD longevity:
Optimize pacing rate and output parameters for the patient.
Set the delivered energy value for the first programmed high voltage therapy to a value lower than maximum energy.
Set the automatic capacitor formation interval to a longer value (unless a shorter interval is required to maintain clinically acceptable charge times).
Use the pre-onset EGM storage feature and Holter telemetry only for short periods of time.
1
EGM pre-storage feature set to Off.
GEM®III DR System Reference Guide
1-7
Quick Reference
Electrical Characteristics

Battery Voltage

Battery voltage appears on the programmer display and on printed reports. See “Preparing the ICD for Implant” on page 12-5, and “Replacement Indicators” on page 12-19 for more information.
Table 1-4. Battery Voltage Indicators
Elective Replacement (ERI)
Pacing mode and rate (magnet and non-magnet) as programmed
End of Life (EOL)
Pacing mode and rate (magnet and non-magnet) as programmed
a
The minimum time between ERI and End of Life is three months, assuming 100% DDD pacing at 60 ppm (min
b
When the battery reaches EOL, immediate replacement is recommended. (See also “Charge Circuit Timeout” on page 5-4.)
b
-1
a
), 3 V, 0.4 ms; 500 pacing load; and six 30 J charges.
2.55 V
2.40 V

Charge Time

The high voltage capacitors charge more rapidly just following a capacitor formation, when they are fully formed. Between formations, the capacitors gradually deform, resulting in longer charge times toward the end of the formation interval. See page 8-13 for information on managing charge time. See page 8-12 for instructions on using the Test Charge feature to measure charge time. Table 1-5 provides typical full energy charge times at BOL and ERI based on the time interval since the last capacitor formation.
Table 1-5. Typical Full Energy (30 J Delivered) Charge Times
Capacitors
fully formed
At Beginning of Life (BOL) 7 seconds 1 second
Near Elective Replacement Indicator (ERI)
a
Typical values for stated conditions, actual charge time performance may vary.
14 seconds 3 seconds
3 months after
capacitor
formation
longer than fully formed
longer than fully formed
a
6 months after
capacitor formation
2 seconds longer than fully formed
5 seconds longer than fully formed
1-8
GEM®III DR System Reference Guide
Quick Reference
Electrical Characteristics

High Voltage Therapy Energy

Table 1-5 compares delivered energy levels to energy levels stored by the device.
Note: The stored energy of the device is derived from the peak capacitor voltage and is always greater than the energy delivered by the device.
Table 1-6. Comparing Programmed (Actual Delivereda) and Storedb Energy Levels
Programmed
(Actual Delivered)
Energy (J)
0.4 0.5 .09 9 11 1.94
0.6 0.8 .13 10 12 2.15
0.8 1.0 .17 11 13 2.37
1.0 1.2 .22 12 14 2.58
1.2 1.5 .26 13 15 2.80
1.4 1.7 .30 14 16 3.01
1.6 2.0 .34 15 18 3.23
1.8 2.2 .39 16 19 3.44
2 2.4 .43 18 21 3.87
3 3.6 .65 20 24 4.30
4 4.7 .86 22 26 4.73
5 5.9 1.08 24 28 5.16
6 7.1 1.29 26 31 5.59
7 8.3 1.51 28 33 6.02
8 9.4 1.72 30 35 6.45
a
Energy and voltage delivered at connector block into a 75 ohm load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at BOL, with fully formed capacitors.
Stored
Energy (J)
Charge
Timec (sec)
Programmed
(Actual Delivered)
Energy (J)
Stored
Energy (J)
Charge
Timec (sec)
GEM®III DR System Reference Guide
1-9
Quick Reference

Magnet Application

Magnet Application
Applying a magnet closes a reed switch within the ICD (see Table 1-6 for magnet effects). When the magnet is removed, the ICD returns to its programmed operations.
Pacing Mode as programmed
Pacing Rate and Interval as programmed
VF, VT, and FVT Detection and Therapy
Patient Alert™ Audible Tones
a
The programming head contains a magnet which can suspend detection. However, if the Resume option has been enabled via the programmer, detection will remain enabled.
b
The alert sounds only if Patient Alert™ is programmed On for any alert except POR. A POR alert causes an alarm every 20 hours, even if Patient Alert is programmed Off.
Table 1-7. Effects of Magnet Application
suspended
b
(30 seconds or less) continuous tone (Alarm OK) on/off intermittent tone (seek follow-up) high/low dual tone (urgent follow-up)
a
1-10
GEM®III DR System Reference Guide

Emergency Settings

The table below lists the values that can be selected for the ICD emergency therapy parameters. Default values are highlighted in
bold, italicized
notes on page 1-28.
Table 1-8. Emergency Therapy Parameters
Parameter Selectable Values (and
Defibrillation
Energy (J) 10, 11, . . ., 16,
18, 20, . . .,
Pathway AXB
Cardioversion
Energy (J) 0.4, 0.6, . . ., 1.8,
2, 3, . . ., 16 18, 20, . . .,
Pathway AXB
Fixed Burst
Pacing Interval (ms) 100, 110, . . .,
360, 370, . . ., 600
Pulse Amplitude (V) 8
Pulse Width (ms) 1.6
text. The Notes column lists letters than correspond to
30
30
350
,
default value
Quick Reference
Emergency Settings
) Notes
VVI Pacing
Pacing Mode VVI
Lower Rate (ppm, min-1)70
Vent. Rate Stabilization Off
Hysteresis Off
V. Sensitivity/A.Sensitivity As programmed
Pulse Amplitude (V) 6 V
Pulse Width (ms) 1.6 ms
Pace Blanking (ms) 240 ms
Post Shock Pace Blanking (ms) 240 ms A.
GEM®III DR System Reference Guide
1-11
Quick Reference

Programmable Parameters Overview

Programmable Parameters Overview
The tables that follow (from “Detection Parameters” on page 1-13 to “Patient Information Parameters” on page 1-23) list the values that can be selected for all ICD programmable parameters. As applicable, these tables also indicate the Medtronic nominal value, the POR value, and the shipped value for each parameter.
Medtronic nominal values – are suggested settings for parameters under nominal conditions. Medtronic nominal values, if applicable, are indicated in the Programmable Values column with
POR values – are the values to which the ICD resets parameters after a power-on reset (POR). If a POR occurs, the ICD sounds a high-urgency Patient Alert™ tone every twenty hours. See the “Power-On Reset Recovery Procedure” on page 10-37.
Shipped values – are the programmed parameter values when the ICD is shipped from Medtronic.
The Notes column in each table lists letters that correspond to notes on page 1-28. These include exceptions or clarifications that relate to the values listed in the programmable parameter tables.
bold, italicized
text.
1-12
GEM®III DR System Reference Guide

Detection Parameters

Table 1-9. Detection Parameters (Sheet 1 of 2)
Parameter Programmable Values
VF DETECTION PARAMETERS
Detection Enable
Interval (ms) 240, 250, . . .,
Initial NID 12/16,
Redetect NID 6/8, 9/12,
V. Sensitivity (mV) 0.15,
A. Sensitivity (mV) 0.15,
(and
Medtronic Nominal
On
, Off On Off
320
,
330, 340, . . ., 400
18/24
12/16
0.3
, 0.45, 0.6,
0.3
, 0.45, 0.6,
, 24/32,
, 18/24,
30/40, 45/60, 60/80, 75/100, 90/120, 105/140, 120/160
21/28, 24/32, 27/36, 30/40
0.9, 1.2
0.9, 1.2, 1.5, 2.1
Quick Reference
Detection Parameters
)
POR
Value
320 320 C.
18/24 18/24 B.
12/16 12/16
0.3 0.3 D, E.
0.3 0.3 D, E.
Shipped
Value
Notes
FVT DETECTION PARAMETERS
Detection Enable
Interval (ms) 200, 210, . . ., 600 --- --- C, F.
V. Sensitivity (mV) 0.15,
A. Sensitivity (mV) 0.15,
Off
, via VF, via VT Off Off
0.3
0.9, 1.2
0.9, 1.2, 1.5, 2.1
, 0.45, 0.6,
0.3
, 0.45, 0.6,
0.3 0.3 D, E.
0.3 0.3 D, E.
GEM®III DR System Reference Guide
1-13
Quick Reference
Detection Parameters
Table 1-9. Detection Parameters (Sheet 2 of 2)
Parameter Programmable Values
VT DETECTION PARAMETERS
Detection Enable On,
Interval (ms) 280, 290, . . .,
Initial NID 12, 16, . . ., 52,
Redetect NID 4, 8, 12, . . ., 52 12 12
V. Sensitivity (mV) 0.15,
A.Sensitivity (mV) 0.15,
VT Stability Criterion Parameters
VT Stability (ms)
(and
Medtronic Nominal
)
Off
400
,
410, 420, . . ., 600
76, 100
0.3
0.9, 1.2
0.9, 1.2, 1.5, 2.1
Off
, 0.45, 0.6,
0.3
, 0.45, 0.6,
, 30, 40, . . ., 100 Off Off C.
POR
Value
Off Off
400 400 C.
16 16
0.3 0.3 D, E.
0.3 0.3 D, E.
Shipped
Value
Notes
DUAL CHAMBER SVT DETECTION
AFib / AFlutter
Sinus Tach
1:1 VT-ST Boundary (%) 35, 50, 66, 75, 85 50 50
Other 1:1 SVTs On,
SVT Limit (ms) 240, 250, . . .,
On
, Off Off Off H.
On
, Off Off Off H.
Off
320
330, 340, . . ., 600
,
Off Off
320 320 C, F.
I.
1-14
GEM®III DR System Reference Guide

Therapy Parameters

Table 1-10. Therapy Parameters (Sheet 1 of 2)
Parameter Programmable Values
VF THERAPY PARAMETERS
Therapy Status
Energy (J) 0.4, 0.6, . . ., 1.8,
Pathway
Reconfirm VF
VT and FVT THERAPY PARAMETERS
Cardioversion Therapy Parameters
Therapy Status
Energy (J) 0.4, 0.6, . . ., 1.8,
Pathway
(and
Medtronic Nominal
On
, Off On (all) On (all)
2, 3, . . ., 16, 18, 20, . . .,
AX→B
Yes
On
2, 3, . . ., 16, 18, 20, . . .,
AX→B
30
, BAX AXB (all) AXB (all) J.
, No Yes Yes M.
, Off --- ---
30
, BAX --- --- J.
Quick Reference
Therapy Parameters
)
POR
Value
30 (all) 30 (all) J, K, L.
--- --- J, L.
Shipped
Value
Notes
Burst Pacing Therapy Parameters
Therapy Status
Initial # Pulses 1, 2, . . ., 6, . . ., 15 --- --- J.
R-S1 Interval (% of R-R)
Pacing Interval Decrement (ms) 0,
# Sequences 1, 2, 3, . . ., 10 --- ---
Smart Mode On,
On
, Off --- ---
50, 53, 56, 59, 63, 66, . . ., 84, 88, 91, 94, 97
10
, . . ., 40 --- ---
Off
--- --- J.
--- --- N.
GEM®III DR System Reference Guide
1-15
Quick Reference
Therapy Parameters
Table 1-10. Therapy Parameters (Sheet 2 of 2)
Parameter Programmable Values
Ramp Pacing Therapy Parameters
Therapy Status
Initial # Pulses 1, 2, . . ., 8, . . ., 15 --- --- J, AB.
R-S1 Interval (% of R-R)
Pacing Interval Decrement (ms) 0, 10, . . ., 40 --- --- J.
# Sequences 1, 2, 3, . . ., 10 --- ---
Smart Mode On,
Ramp+ Pacing Therapy Parameters
Therapy Status
Initial # Pulses 1, 2, 3, . . ., 15 --- --- J.
R-S1 Interval (% of R-R)
S1-S2 Interval (% of R-R)
S2-Sn Interval (% of R-R)
# Sequences 1, 2, . . ., 5, . . ., 10 --- ---
Smart Mode On,
(and
Medtronic Nominal
On
, Off --- ---
50, 53, 56, 59, 63, 66, . . ., 84, 88,91, 94, 97
)
Off
On
, Off --- ---
50, 53, 56, 59, 63, 66, . . ., 75, . . ., 84, 88, 91, 94, 97
50, 53, 56, 59, 63, 66, 69, . . ., 84, 88, 91, 94, 97
50, 53, 56, 59, 63,66, . . ., 84, 88, 91, 94, 97
Off
POR
Value
--- --- J, AB.
--- --- N.
--- --- J.
--- --- J.
--- --- J.
--- --- N.
Shipped
Value
Notes
SHARED PARAMETERS FOR TACHYARRHYTHMIA THERAPIES
Shared by all Automatic ATP Therapies J.
V. Pulse Width (ms) 0.03, 0.06,
V. Amplitude (V) 1, 2, . . ., 6,
V. Pace Blank (ms) 150, 160, . . .,
ATP Minimum Interval (ms) 150, 160, . . .,
Shared by all VF, VT, and FVT Therapies
Progressive Episode Therapies Enable
1-16
GEM®III DR System Reference Guide
0.1, 0.2, . . .,
On,
Off
1.6
8
240
, . . ., 440 240 240
200
, . . ., 400 200 200
1.6 1.6
88
Off Off

Bradycardia Pacing Parameters

Bradycardia Pacing Parameters
Table 1-11. Bradycardia Pacing Parameters (Sheet 1 of 3)
Parameter Programmable Values
Pacing Mode
Lower Rate (ppm, min-1) 34, 35, 40, . . ., 60,
Upper Tracking Rate (ppm, min-1) 80, 85, . . ., 105,
Upper Sensor Rate (ppm, min-1)
A. Sensitivity (mV) 0.15,
A. Pulse Width (ms) 0.03, 0.06,
A. Pulse Amplitude (V) 1, 2, 3, . . ., 6 4 3
A.Pace Blanking (ms) 200, 210, . . .,
V. Sensitivity (mV) 0.15,
V. Pulse Width (ms) 0.03, 0.06,
V. Pulse Amplitude (V) 1, 2,
V. Pace Blanking (ms) 150, 160, . . .,
(and
Medtronic Nominal
DDDR
, DDD, DDIR, DDI, AAIR,
AAI, VVIR, VVI
70, 75, . . ., 105, 109, 115, 120
109, 115, 146, 150
80, 85, . . ., 105, 109, 115, 146, 150
0.9, 1.2, 1.5, 2.1
0.1, 0.2, . . .,
0.5, 0.6, . . ., 1.6
120
, 125, 130, 136, 140,
120
, 125, 130, 136, 140,
0.3
, 0.45, 0.6,
0.4
,
240
0.3
0.9, 1.2
0.1, 0.2, . . .,
0.5, 0.6, . . . , 1.6
, 0.45, 0.6,
0.4
,
3
, . . ., 6 6 3
200
210, 220, . . ., 440
)
, 250 240 240
,
POR
Value
VVI DDD
65 60 G.
120 120 G.
120 120 G.
0.3 0.3 D, E.
0.4 0.4
0.3 0.3 D, E.
1.6 0.4
240 200
Quick Reference
Shipped
Value
Notes
Post-Shock Pacing Parameters
A. Pulse Width (ms) 0.03, 0.06,
A. Pulse Amplitude (V) 1, 2, 3, 4, 5, 6 4 4
A. Pace Blanking (ms) 200, 210, . . .,
V. Pulse Width (ms) 0.03, 0.06,
V. Pulse Amplitude (V) 1, 2, 3, 4, 5,
V. Pace Blanking (ms) 150, 160, . . .,
0.1, 0.2, . . .,
1.6
240
0.1, 0.2, . . .,
1.6
6
200
210, 220, . . ., 440
, 250 240 240
,
1.6 1.6
1.6 1.6
66
240 200
GEM®III DR System Reference Guide
1-17
Quick Reference
Bradycardia Pacing Parameters
Table 1-11. Bradycardia Pacing Parameters (Sheet 2 of 3)
Parameter Programmable Values
Intervals
PAV (ms) 30, 40, . . .,
SAV (ms) 30, 40, . . .,
PVARP (ms) Varied,
A. Refractory (ms) 150, 160, . . .,
PVAB (ms) 100, 110, . . .,
Mode Switch Parameters
Enable On,
A. Detect Rate (bpm, min-1) 120, 125, 130, 136, 140, 146, 150,
Rate Response Pacing Parameters
Rate Response 1, 2, . . ., 7,
Activity Threshold Low, Medium Low,
Activity Acceleration (sec.) 15,
Activity Deceleration (min.) Exercise, 2.5, 5, 10 5 5
(and
Medtronic Nominal
180
190, 200, . . ., 350
150
160, 170, . . ., 350
150, 160, . . ., 320, 330, . . ., 500
310
310
320, 330, . . ., 500
150
160, 170, . . ., 350
,
,
)
,
,
,
Off
154, 158, 167, 171,
8, 9, 10
Medium
Medium High, High
,
30
, 60 30 30
176
POR
Value
180 180
150 150
310 310 O.
310 310
150 150 O.
Off Off
176 176
77
High Medium
Shipped
Value
Notes
Rate Adaptive AV Parameters
Enable On,
Start Rate (bpm, min-1) 50, 55, 60, . . ., 105, 109,
Stop Rate (bpm, min-1) 50, 55, . . ., 105, 109,
Minimum PAV (ms) 30, 40, . . .,
Minimum SAV (ms) 30, 40, . . .,
1-18
GEM®III DR System Reference Guide
Off
115, 120, 125, 130, 136, 140, 146, 150, 154, 158, 167, 171, 176, 182
115,
120
150, 154, 158, 167, 171, 176, 182
150, 160, . . ., 350
120, 130, . . ., 350
, 125, 130, 136, 140, 146,
140
,
110
,
Off Off
60 60
120 120
140 140
110 110
Table 1-11. Bradycardia Pacing Parameters (Sheet 3 of 3)
Parameter Programmable Values
Non-Competitive Atrial Pacing
Enable
Single Chamber Hysteresis
Enable-Rate (bpm, min-1)
Ventricular Rate Stabilization Parameters
Enable On,
Interval Increment (ms) 50, 60, . . .,
Minimum Interval (ms)
PMT Intervention Parameters
Enable On,
PVC Response Parameters
Enable
Ventricular Safety Pacing Parameters
Enable
(and
Medtronic Nominal
On
, Off On On
Off
, 34, 40, 50, 60 , 70, 80 Off Off P.
Off
150
,
160, 170, . . ., 400
500
, 550, . . ., 900 500 500
Off
On
, Off On On
On
, Off On On
Quick Reference
Automatic Capacitor Formation Parameters
)
POR
Value
Off Off
150 150
Off Off
Shipped
Value
Notes
Automatic Capacitor Formation Parameters
Table 1-12. Automatic Capacitor Formation Parameters
Parameter Programmable Values
Automatic Capacitor Formation Interval (months)
(and
Medtronic Nominal
Off, 1, 2, . . .,
6
)
GEM®III DR System Reference Guide
POR
Value
6 Off Q.
Shipped
Value
Notes
1-19
Quick Reference

Data Collection Parameters

Data Collection Parameters
Table 1-13. Data Collection Setup
Parameter Programmable Values
EGM Source Parameters
EGM Channel 1 Source HVA to HVB, HVA to Vring,
EGM Channel 1 Range (mV) ± 4, ± 8, ± 16 ± 8 ± 8
EGM Channel 2 Source HVA to HVB, HVA to Vring,
EGM Channel 2 Range (mV) ± 4, ± 8, ± 16 ± 8 ± 8
Store EGM Channel 1?
Store EGM Channel 2 Yes (fixed) --- ---
Store EGM during charging?
Store EGM before tachycardia starts?
(and
Medtronic Nominal
HVA to Aring, Vtip to HVB, Vtip to Vring, Vtip to Atip, Atip to HVB,
Vtip to HVB,
Yes
, No Yes Yes
Yes
, No Yes Yes
Ye s ,
No
Atip to Aring
Vtip
to
)
Vring
POR
Value
Atip to Aring Atip to Aring S.
Vtip to Vring Vtip to Vring S.
No No
Shipped
Value
Notes
Device Date/Time
Premature Event Recording Parameters
Premature Event Threshold (%)
Holter Telemetry Parameters
Duration (hours)
1-20
GEM®III DR System Reference Guide
(enter time and date) --- --- X.
56, 59, 62, 66, 69, . . ., 84, 88, 91, 94, 97
Off
, 0.5, 1, 2, 4,
8, 16, 24, 36, 46
69 69
Off Off
Quick Reference
Patient Alert™ Monitoring System Parameters
Patient Alert™ Monitoring System Parameters
Table 1-14. Patient Alert™ Setup (Sheet 1 of 2)
Parameter Programmable Values
ALERT TIME
Hour 0, 1, 2, . . ., 23 --- ---
Minute 0, 10, . . ., 50 --- ---
LEAD IMPEDANCE OUT OF RANGE ALERT
Atrial Pacing Lead Impedance Alert AA.
Enable On, Off Off Off
Minimum Threshold (ohms) 200 Ω (fixed) --- ---
Maximum Threshold (ohms) 2000 Ω (fixed) --- ---
Ventricular Pacing Lead Impedance Alert AA.
Enable On, Off Off Off
Minimum Threshold (ohms) 200 Ω (fixed) --- ---
Maximum Threshold (ohms) 2000 Ω (fixed) --- ---
Defibrillation Lead Impedance Alert AA.
Enable On, Off Off Off
Minimum Threshold (ohms) 10 Ω (fixed) --- ---
Maximum Threshold (ohms) 200 Ω (fixed) --- ---
Alert Urgency
(and
Medtronic Nominal
Low
, High --- ---
)
POR
Value
Shipped
Value
Notes
AA.
GEM®III DR System Reference Guide
1-21
Quick Reference
Patient Alert™ Monitoring System Parameters
Table 1-14. Patient Alert™ Setup (Sheet 2 of 2)
Parameter Programmable Values
LOW BATTERY VOLTAGE ALERT
Alert Enable - Urgency Off,
Battery Voltage Threshold (V) 2.55 (ERI) --- ---
(and
Medtronic Nominal
On-Low
, On-High Off Off
)
POR
Value
Shipped
Value
Notes
AA.
EXCESSIVE CHARGE TIME ALERT
Alert Enable - Urgency Off,
Charge Time Threshold (seconds) 15, 16, 17, 18, . . ., 25 --- ---
NUMBER OF SHOCKS DELIVERED IN AN EPISODE ALERT
Alert Enable - Urgency Off,
Number of Shocks Threshold 1, 2, 3, 4, 5, 6 --- ---
ALL THERAPIES IN A ZONE EXHAUSTED FOR AN EPISODE ALERT
Alert Enable - Urgency Off,
On-Low
, On-High Off Off
On-Low
, On-High Off Off
On-Low
, On-High Off Off
AA.
AA.
AA.
1-22
GEM®III DR System Reference Guide
Quick Reference

Patient Information Parameters

Patient Information Parameters
Table 1-15. Description of Patient Information
Information Field Description and Action
Patient Enter the patient name or ID number (up to 30 characters).
ID Enter the patient ID.
Date of Birth Select patient’s date of birth.
ICD Number (Not selectable)
Lead 1 . . . Lead 2 . . . Lead 3 . . .
Implant Enter measured values from implant. Using the displayed submenu, enter new
Notes Enter notes about patient or other information.
History (2 fields) Select medical history from list box. If necessary, you can indicate that more
EF, on Select the Ejection Fraction from a table of values. Then enter the date in the
Physician Select or add the name of the physician.
Phone Enter the phone of the physician.
Hospital Select or add the hospital information.
Last Update (Not selectable)
Displays serial number for implanted ICD after interrogation.
Enter detailed information for leads 1, 2, and 3:
Model
Position:
pick from list of positions, or enter position
• Serial Number
• Manufacturer
• Implant date
information from the keyboard and make list box selections.
information is located in the Notes field.
following field.
Displays the date of the last Patient Information update.
GEM®III DR System Reference Guide
1-23
Quick Reference

EP Studies Parameters

EP Studies Parameters
Table 1-16. EP Studies Parameters (Sheet 1 of 2)
Parameter Selectable Values (and
INDUCTIONS
T-Shock™ Induction Parameters
#S1 2, 3, . . .,
S1S1 (ms) 350, 360, . . .,
410, 420, . . ., 1760
Delay (ms) 50, 60, . . .,
320, 330, . . ., 600
Energy (J) 0.4,
Pathway
Waveform
Pulse Amplitude 6 V (fixed)
Pulse Width 1.6 ms (fixed)
Enable Enabled,
Resume at Deliver
2, 3, . . ., 16, 18, 20, . . ., 30
AX→B
Monophasic
Enabled
0.6
, . . ., 1.8,
, BAX
, Disabled
8
400
310
,
, Biphasic
Disabled
,
default value
) Notes
Z.
Z.
50 Hz Burst Induction Parameters Z.
Interval (ms) 20 ms (fixed)
Pulse Amplitude (V) 1, 2, . . ., 6,
Pulse Width (ms) 0.03, 0.06,
Resume at Burst
Manual Burst Induction Parameters
Chamber
Interval (ms) 100, 110, . . .,
Pulse Amplitude (V) 1, 2, . . ., 5, 6, 8
Pulse Width (ms) 0.03, 0.06,
Resume at Burst
1-24
GEM®III DR System Reference Guide
0.1, 0.2, . . .,
Enabled
Ventricle
0.1, 0.2, . . .,
0.6, 0.7, . . ., 1.6
Enabled
8
1.6
, Disabled
, Atrium Z.
600
0.5
,
, Disabled
Quick Reference
EP Studies Parameters
Table 1-16. EP Studies Parameters (Sheet 2 of 2)
Parameter Selectable Values (and
PES Induction Parameters
Chamber
#S1 1, 2, . . ., 8,
S1S1 (ms) 100, 110, . . .,
S1S2 (ms) Off, 100, 110, . . .,
S2S3 (ms)
S3S4 (ms)
Pulse Amplitude (V) 1, 2, . . ., 5, 6, 8
Pulse Width (ms) 0.03, 0.06,
Resume at Deliver
MANUAL THERAPY PARAMETERS
In general, each manual therapy uses the same parameter values as the automatic therapy. See “Therapy Parameters” beginning on page 1-15.
Ventricle
9, 10, . . ., 15
, Atrium
600
610, 620, . . ., 1760
410, 420, . . ., 600
Off
, 100, 110, . . ., 600 Z.
Off
, 100, 110, . . ., 600 Z.
0.1, 0.2, . . .,
0.6, 0.7, . . ., 1.6
Enabled
0.5
, Disabled
,
400
,
default value
,
) Notes
Y.
Z.
GEM®III DR System Reference Guide
1-25
Quick Reference

Fixed Parameters

Fixed Parameters
Table 1-17. Fixed Parameters (Sheet 1 of 2)
Parameter Value Notes
SENSING
Atrial Blanking Periods
After a sensed atrial event 100 ms
After a paced vent. event 30 ms
After charging ends 300 ms
After HV therapy 520 ms
Ventricular Blanking Periods
After a sensed vent. event 120 ms
After a paced atrial event 30 ms
After charging ends 300 ms
After HV therapy 520 ms
Atrial Refractory Periods
Non-Competitive Atrial Pacing Interval 300 ms
BRADYCARDIA PACING
Ventricular Safety Pacing Intervals 110 ms
PVC Response (PVARP extension)
PMT Intervention (PVARP extension)
1-26
GEM®III DR System Reference Guide
70 ms
T.
Extended to 400 ms U.
Extended to 400 ms
Quick Reference
Fixed Parameters
Table 1-17. Fixed Parameters (Sheet 2 of 2)
Parameter Value Notes
HIGH VOLTAGE THERAPIES
Maximum charging period 30 seconds
Waveform Biphasic
Tilt 50%
VF Synchronization Periods V.
After post-charge blanking 500 ms
After a sensed event 500 ms
Synchronization Refractory Periods
After V. sense during CV synch. 200 ms
After charge during DF synch. 400 ms R.
Atrial Vulnerable Period 250 ms
Post-Therapy
Escape interval after CV/Defib 1200 ms
Suspension of VT Detection after Defibrillation therapy
17 V. events
EP STUDIES
T-Shock™ S1 pacing outputs 1.6 ms at 6 V Z.
HARDWARE PARAMETERS
Atrial rate limit (protective feature)
Ventricular rate limit (protective feature)
Input protection 100 k minimum
171 ppm (min-1)W.
-1
171 ppm (min
)W.
GEM®III DR System Reference Guide
1-27
Quick Reference

Notes

Notes
A.
Post-shock V. Pace Blanking is reprogrammed by Emergency VVI pacing, but is not displayed on the Emergency Screen.
B.
The VF Initial NID cannot be programmed to a value greater than 30/40 if VT Detection is enabled.
C.
The measured intervals are truncated to a 10 ms multiple (e.g., 457 ms becomes 450 ms). The device uses this truncated interval value when applying the programmed criteria and calculating interval averages.
D.
With a 40 ms sine2 waveform (ventricular sensitivity) or a 20 ms sine waveform (atrial sensitivity). When using the Cenelec waveform, the rated sensing threshold value will be 1.5 times the rated sine sensing threshold.
E.
This setting applies to all sensing in this chamber, for both tachyarrhythmia detection and bradycardia pacing operations.
F.
Nominal value is blank until detection is enabled.
G.
The shortest pacing escape interval must exceed all programmed Detection Intervals by at least 60 ms. In dual chamber modes, the shortest VA interval must equal or exceed any programmed Detection Interval.
H.
The ICD is shipped with the Sinus Tach and A.Fib / A.Flutter criteria Off. However, when VT Detection is programmed On for the first time, these criteria are automatically enabled.
I.
Double tachycardia (i.e. “VF/FVT/VT plus SVT”) detection is automatically enabled when any Dual Chamber SVT criterion is enabled.
J.
These specifications apply to both automatic and manual therapies, except as noted.
K.
For automatic therapy 3, 4, 5, or 6, energy must be at least 10 joules.
L.
Delivered energy based on a biphasic pulse into a 75 ohm load.
M.
Applies only to the first VF therapy that is programmed On.
N.
Smart Mode is only available for Therapies 1, 2, 3, and 4.
O.
PVAB is the minimum value that PVARP is shortened to, under sensor-varied PVARP operation. Atrial events that fall within the PVAB are ignored by the Mode Switch, NCAP, PVC Response, and PMT Intervention features.
P.
Pacing mode must be VVI or AAI, and the hysteresis rate must be less than the pacing rate.
Q.
The ICD is shipped with Automatic Capacitor Formation Interval set to Off. However, when VF Detection is set to On for the first time, the Automatic Capacitor Formation Interval is automatically set to 6 month.
R.
When Reconfirm VF is active, a refractory period equal to 60 ms plus the greatest programmed detection interval follows the post-charge blanking period and all paced events.
2
2
1-28
GEM®III DR System Reference Guide
Quick Reference
S.
Both channels appear on-screen as real-time telemetered signals.
T.
The shorter VSP interval takes effect when the pacing rate exceeds [60,000 divided by 2x (ventricular pace blanking plus 110)] per minute.
U.
PVARP is only extended to 400 ms if the current PVARP (either the programmed PVARP value or the current Sensor-Varied PVARP value) is less than 400 ms.
V.
When Reconfirm VF is active, the VF synchronization (reconfirmation) periods are 60 ms plus the greatest programmed detection interval; and the synchronization escape period is equal to the current lower rate (i.e., sensor, Mode Switch, or programmed lower rate) escape period.
W.
Does not apply during therapies, programmed high rates, or ventricular safety pacing.
X.
The time stamp on episode records and other stored data is determined by the device’s date/time clock.
Y.
Delivered in the VVI or AAI mode (as applicable)
Z.
Delivered in the VOO or AOO mode (as applicable)
AA.
After a POR, all programmed Patient Alert conditions are disabled. The Reset condition itself activates a high-urgency alert that sounds every
twenty hours
AB.
For manual ramp therapy, 6 is the nominal number of pulses, and 97% is the nominal value for % of R-R interval.
until cleared.
Notes
GEM®III DR System Reference Guide
1-29
Quick Reference
Notes
1-30
GEM®III DR System Reference Guide
Emergency Therapy
2
2
Overview 2-2
Effect on System Operation 2-3
On-Screen and Display Panel Buttons 2-3
Emergency Defibrillation 2-4
Emergency Cardioversion 2-5
Emergency Fixed Burst Pacing 2-6
Emergency VVI Pacing 2-7
GEM®III DR System Reference Guide
2-1
Emergency Therapy

Overview

Overview
The ICD provides the following Emergency therapies:
Defibrillation therapy
Cardioversion therapy
Fixed Burst pacing therapy
Maximum output VVI pacing
To deliver an Emergency therapy, the programming head must be positioned over the ICD.
Defibrillation is the default Emergency therapy, and 30 joules is the default delivered energy. When you select
[Emergency] and [DELIVER], the ICD
charges and delivers a biphasic 30 joule shock along the AX>B pathway.
If you select a different defibrillation energy, the programmer resets the energy to 30 joules the next time you select
[Emergency]. Emergency
Cardioversion and Emergency Fixed Burst values remain as selected for the duration of the session.
2-2
Aborting the Therapy – As a safety precaution, the programmer also
displays an
[ABORT] button which immediately terminates any Emergency
therapy in progress.
To return to other programming functions from an Emergency screen, select
[Exit Emergency].
GEM®III DR System Reference Guide

Effect on System Operation

The ICD suspends its automatic detection features when Emergency defibrillation, cardioversion, or fixed burst pacing therapies are delivered. Detection is not suspended during Emergency VVI Pacing. Removing the programming head or pressing
[Resume] re-enables detection.
On-Screen and Display Panel Buttons
The on-screen [Emergency] button and the red mechanical Emergency button by the programmer display panel are equivalent at all times.
Emergency Therapy
Effect on System Operation
Red Emergency Button
The on-screen
[DELIVER] button and the yellow-on-blue mechanical
Functions the same as on-screen
[Emergency]
at all times.
Deliver button by the programmer display panel are equivalent during Emergency operations only. The mechanical Deliver button operates only during Emergency operations.
Yellow-on-Blue
Deliver Button
Functions the same as on-screen
[DELIVER], but only during
Emergency functions.

Temporary Parameter Values

Emergency tachyarrhythmia therapies use the permanently programmed parameters of the ICD. in effect until you begin the temporary operation by selecting After the tachyarrhythmia therapy is complete, the ICD reverts to its permanently programmed values.
test values
that do not change
1
Test values are not
[DELIVER].
1
Delivery of Emergency VVI Pacing permanently changes the programmed bradycardia pacing values to those used during emergency pacing (see page 2-7).
GEM®III DR System Reference Guide
2-3
Emergency Therapy
Emergency Defibrillation
Emergency Defibrillation
How to Deliver Emergency 30 Joule Defibrillation
1. Position the programming head over the ICD.
2. Select [Emergency].
3. Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly positioned and select
How to Change the Defibrillation Energy and Deliver
1. From the Emergency Therapy screen, select Energy.
2. Select a new energy value.
3. Select [DELIVER].
[Retry] or [Cancel].
Select to Change
Emergency Defibrillation
Select to Deliver
Emergency Defibrillation
2-4
GEM®III DR System Reference Guide
Energy
Figure 2-1. Emergency Defibrillation Screen

Emergency Cardioversion

How to Deliver Emergency Cardioversion
1. Position the programming head over the ICD.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the cardioversion energy shown on the screen, or select
Energy and select a new value from the window.
5. Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly positioned and select
Emergency Therapy
Emergency Cardioversion
[Retry] or [Cancel].
Emergency Cardioversion
Select to Change
Energy
Select to Deliver
Emergency Cardioversion
Figure 2-2. Emergency Cardioversion Screen
GEM®III DR System Reference Guide
2-5
Emergency Therapy

Emergency Fixed Burst Pacing

Emergency Fixed Burst Pacing
How to Deliver Emergency Fixed Burst Pacing
1. Position the programming head over the ICD.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval shown on the screen, or select Interval for
a new interval value.
Select to Change
Burst Interval
Press and Hold to Deliver
Emergency Fixed
Burst Pacing
5. Select
If delivery is not confirmed, the programmer displays an error window. Verify that the programming head is properly positioned. Select the window and reselect
[BURST Press and Hold].
[OK] from
[BURST Press and Hold].
Figure 2-3. Emergency Fixed Burst Pacing Screen
2-6
GEM®III DR System Reference Guide

Emergency VVI Pacing

How to Deliver Emergency VVI Pacing
1. Position the programming head over the ICD.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A successful programming sets the ICD to the
following maximum output bradycardia pacing values.
Pacing Mode VVI
Lower Rate 70 ppm (70 min-1)
V. Amplitude 6 V
V. Width 1.6 ms
V. Pace Blanking 240 ms
V. Post Shock Pace Blanking 240 ms
Hysteresis Off
Ventricular Rate Stabilization Off
a
Emergency Therapy
Emergency VVI Pacing
a
Reprogrammed by Emergency VVI operation, but not displayed on Emergency screen.
If programming is not confirmed, verify that the programming head is properly positioned and select
[Retry] or [Cancel].
GEM®III DR System Reference Guide
2-7
Emergency Therapy
Emergency VVI Pacing
Emergency VVI Pacing
Parameters (fixed)
Select to Program
Emergency VVI Pacing
Figure 2-4. Emergency VVI Pacing Screen
2-8
GEM®III DR System Reference Guide
Overview
3
3
System Description 3-2
Detection and Therapy Overview 3-4
Disabling Detection and Therapies 3-9
GEM®III DR System Reference Guide
3-1
Overview

System Description

System Description
The GEM®III DR Model 7275 Dual Chamber Implantable Cardioverter Defibrillator (ICD) System is an implantable medical device system that automatically detects and treats episodes of ventricular fibrillation (VF), ventricular tachycardia (VT), fast ventricular tachycardia (FVT), and bradyarrhythmia. When a cardiac arrhythmia is detected, the implantable device delivers defibrillation, cardioversion, antitachycardia pacing, or bradycardia pacing therapy.
The VF Detection circuitry monitors the cardiac cycle length to identify a VF episode. Upon detection of VF, the ICD provides a biphasic defibrillation shock of up to 30 joules delivered energy. If the VF episode persists, up to five more individually programmed defibrillation shocks can be delivered.
The VT Detection circuitry monitors the cardiac cycle length to identify the presence of a tachycardia. Upon detection of VT, the ICD delivers its first programmed VT therapy to restore the patient’s normal cardiac rhythm. If the VT episode persists, up to five more individually programmed VT therapies can be delivered. Therapy selections include three methods of antitachycardia pacing and synchronized cardioversion.
3-2
FVT detection can be programmed through the VT Detection algorithm or through the VF Detection algorithm. As with VF and VT, up to six FVT therapies can be independently programmed. Therapy selections include three methods of antitachycardia pacing and synchronized cardioversion.
The ICD can be programmed to discriminate between true ventricular arrhythmias and rapidly-conducted supraventricular tachyarrhythmias (SVTs), using PR Logic™ Pattern and Rate Analysis. The ICD also can detect an unrelated ventricular arrhythmia occurring simultaneously with an SVT, so that recognition of an SVT does not impede ventricular detection.
The ICD provides dual chamber bradycardia pacing to optimize hemodynamics. An internal piezoelectric sensor enables the ICD to increase and decrease its pacing rate in response to the patient’s detected physical activity.
GEM®III DR System Reference Guide
Overview
System Description
The ICD is intended for implantation with transvenous or epicardial defibrillation leads. The lead system should consist of bipolar or paired unipolar pacing/sensing leads in each chamber of the heart, and one or two high voltage cardioversion/ defibrillation electrodes. (The Active Can device case also serves as one high voltage electrode.) Sensing of cardiac activity and the delivery of pacing stimuli occur between the pacing/sensing electrodes in each chamber.
The physician can use the ICD to conduct noninvasive electrophysiologic studies. The physician can also manually deliver any of the ICD therapies to manage an induced or spontaneous tachyarrhythmia.
A programmable Patient Alert™ monitoring feature notifies the patient by audible tones if certain conditions (low battery voltage, high lead impedance, etc.), selected by the physician, occur. The patient can then respond based on instructions provided by the physician.
The ICD provides the physician with real-time information about its detection and therapy parameters and status. The ICD also provides accumulated data on its operation, including stored electrograms, records of tachyarrhythmia episodes detected and treated, bradycardia interventions, and the efficacy of therapy. All of this information can be printed and retained in the patient’s file, saved in the programmer to read back at a later time, or saved in electronic format on a floppy diskette.
®
GEM®III DR System Reference Guide
3-3
Overview

Detection and Therapy Overview

Detection and Therapy Overview
Implanted sensing electrodes provide atrial and ventricular intracardiac electrograms to the ICD. The ICD measures the time between events to determine A-A and V-V intervals, and calculates the atrial and ventricular rates from these intervals.
The ICD can operate in:
Single chamber detection (with all Dual Chamber SVT Criteria programmed using only the ventricular rate, or
Dual chamber detection (with any Dual Chamber SVT Criterion programmed using both the rate measurements and PR Logic™ Pattern and Rate Analysis results.
Figure 3-1 represents the detection paths with Dual Chamber SVT Criteria programmed chamber detection).
Single Chamber Detection – When a V-V interval falls within the programmed VF, VT, or FVT detection zone, it increments a VF or VT event counter. If either counter reaches the “Number of Intervals to Detect” (NID) programmed by the physician, the ICD detects the arrhythmia and delivers the programmed therapy.
Off), in which the ICD detects ventricular arrhythmias
On), in which the ICD detects ventricular arrhythmias
Off (single chamber detection) and On (dual
3-4
GEM®III DR System Reference Guide
Overview
Detection and Therapy Overview
Dual Chamber Detection – If the number of VF or VT intervals reaches the programmed NID, the ICD applies PR Logic™ Pattern and Rate Analysis to classify the rhythm as either:
a ventricular tachyarrhythmia,
a rapidly conducted SVT (page 4-28),
or a double tachycardia (page 4-33).
PR Logic™ evaluates the rates, patterns, and relative timing of atrial and ventricular activation to identify conducted SVTs (page 4-28). The ICD delivers ventricular therapy for ventricular or double tachycardias, but withholds therapy for rapidly conducted SVTs.
Ventricular Event
Update A-V Pattern and Intervals
Ventricular Tachyarrhythmias Detected by
No
One or More Dual-Chamber SVT Criteria On?
Single-Chamber Detection
Ye s
No/Redetection
Ye s
Ye s
No
Deliver
Ventricular
Therapy
Median Ventricular Interval
Less than SVT Limit?
VT/VF or Fast VT during AF/AFlutter?
Dual-Chamber SVT Criteria Met?
Ye s
No
No
Ye s
Figure 3-1. GEM®III DR Detection and Therapy Operation
GEM®III DR System Reference Guide
3-5
Overview
Detection and Therapy Overview

Redetection

After a therapy delivery, if the VF or VT event counter reaches the programmed Number of Intervals to Redetect (RNID), the episode is redetected and the next programmed therapy is delivered. This progression continues until the ICD confirms termination of the episode (page 4-36) or delivers all the therapies programmed for the detected tachyarrhythmia. The ICD does not apply the dual chamber detection and PR Logic™ algorithms during redetection.

Therapy

The ICD provides five ventricular therapies: Burst pacing; Ramp pacing; Ramp+ pacing; synchronized cardioversion; and defibrillation. Each VF episode can be treated with up to six defibrillation shocks. Each VT or FVT episode can be treated with up to six therapies of antitachycardia pacing and/or cardioversion. An FVT therapy set must end with at least one cardioversion therapy.
3-6
GEM®III DR System Reference Guide
Detection and Therapy Overview

Suspending Detection and Therapy

Automatic Suspension of Detection
Detection is temporarily suspended during:
Programming head application (unless [Resume] is programmed) or other magnet application. Detection resumes automatically when the head or magnet is removed.
Temporary programmer operations such as manual and emergency therapies delivered through the programmer. Detection remains suspended until the programming head is removed or is programmed.
[Resume]
Overview
Note: If the Resume at Deliver feature is disabled
and
detection was suspended before a temporary induction is performed, then detection remains suspended.
Automatic daily lead impedance measurements (see page 8-8). Detection resumes when the measurement is complete.
Automatic tachyarrhythmia therapy deliveries, including charging periods for high voltage therapy and automatic capacitor formations. Detection resumes automatically when the therapy or charging period is complete.
VT Detection is temporarily suspended during the first 17 ventricular events (paced and sensed) after an automatic ventricular defibrillation therapy.
GEM®III DR System Reference Guide
3-7
Overview
Detection and Therapy Overview
Manually Suspending Detection and Therapy
ICD detection and therapy capabilities can be temporarily suspended by:
positioning the programming head, the Patient Alert™ patient magnet, or another magnet over the ICD
executing a [Suspend] command from the programmer
The bradycardia pacing and manually-initiated therapy capabilities of the ICD are still functional. Suspending detection does not change the
programmed settings of the ICD.
Removing the programming head or magnet from its position over the ICD restores the ICD to normal operation. If automatic detection is enabled, the
[Resume] button causes the ICD to resume automatic detection even
while the programming head is in position over the ICD.
3-8
GEM®III DR System Reference Guide

Disabling Detection and Therapies

Automatically Disabling Detection and Therapies
The ICD turns off VT Detection if a power-on reset (POR) occurs. The programmable parameter tables starting on page 1-13 list the POR values for all programmable parameters.
The ICD automatically turns off all VT and VF automatic therapies if three consecutive 30-second charges have failed to reach the programmed energy level (see “Charge Circuit Inactive” on page 5-4).
Manually Disabling Detection and Therapies
To turn off the automatic detection and therapy functions of the ICD, program the functions
1. Select the Params icon and choose the Detection option.
2. Set the Enable parameters to Off.
3. Position the programming head and select [PROGRAM].
Off via the programmer.
Overview
Disabling Detection and Therapies
To re-enable automatic detection and therapy, reprogramming is required.
GEM®III DR System Reference Guide
3-9
Overview
Disabling Detection and Therapies
3-10
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
4
4
&
Sensing 4-2
VF Detection 4-7
VT Detection 4-10
Combined Count Detection 4-13
VT Stability Option 4-15
FVT Detection 4-17
Dual Chamber SVT Criteria 4-22
Termination and Redetection 4-35
GEM®III DR System Reference Guide
4-1
Tachyarrhythmia Detection

Sensing

Sensing
The ICD provides bipolar sensing in both the atrium and ventricle via the sensing electrodes of the implanted atrial and ventricular leads. The ICD
senses in both chambers, regardless of the programmed bradycardia pacing mode.
Proper sensing is essential for the safe and effective use of the ICD. To provide appropriate sensing, the ICD uses:
Auto-adjusting atrial and ventricular sensing thresholds (see page 4-4),
Short (30 ms) cross-chamber blanking after paced events, and
No cross-chamber blanking after sensed events.

Programming for Appropriate Sensing

Separately programmable sensitivity thresholds define the minimum electrical amplitude that the ICD recognizes as an atrial or ventricular event.
4-2
A combination of high pacing pulse width and/or amplitude with a low sensitivity threshold may cause inappropriate sensing after pacing. Programming a lower pulse width, lower amplitude, longer pace blanking, or a higher threshold may eliminate this inappropriate sensing.
Ventricular Sensing
A ventricular sensitivity threshold of 0.3 mV is recommended to maximize the probability that VF will be detected.
Always verify that the ICD is sensing and detecting VF properly. You may need to reposition or replace the ventricular sensing lead.
A value greater than 0.6 mV is not recommended except for testing and troubleshooting, as it may lead to undersensing or underdetection.
When the ICD is programmed to an atrial pacing mode, make sure that it does not sense atrial pacing pulses in the ventricle.
Carefully evaluate the potential for increased susceptibility to EMI and oversensing before changing the ventricular sensitivity threshold to its most sensitive setting of 0.15 mV.
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
Sensing
If you change the ventricular sensitivity threshold, evaluate for proper sensing and detection by inducing VF and allowing the ICD to automatically detect and treat the arrhythmia.
If the ICD is not sensing or detecting properly, program all automatic therapies to
Off and evaluate the ICD system. Monitor the patient
continuously for life-threatening arrhythmias until you program automatic therapies
On again.
Atrial Sensing
The ICD provides a nominal atrial sensitivity threshold of 0.3 mV to maximize the effectiveness of PR Logic™ Pattern and Rate Analysis and atrial pacing operations. Select an atrial threshold that does not sense far-field R-waves, but provides reliable sensing during SVTs.
Any Medtronic bipolar IS-1 compatible atrial lead may be used. Atrial leads with minimal tip to ring spacing may reduce far-field R-wave sensing.
A less sensitive atrial threshold could lead to undersensing of P-waves or inappropriate ventricular detection. Verify that the ICD senses properly in the atrium, both during sinus rhythm and during SVTs.
Verify that the ICD does not sense far-field R-waves by viewing the atrial Marker annotations.
Carefully evaluate the potential for increased susceptibility to EMI and oversensing before changing the atrial sensitivity threshold to its most sensitive setting of 0.15 mV.
You may need to reposition or replace the atrial sensing lead if none of the programmable atrial thresholds provides both reliable SVT sensing and exclusion of far-field R-wave sensing.
GEM®III DR System Reference Guide
4-3
Tachyarrhythmia Detection
Sensing

Auto-Adjusting Sensitivity Thresholds

The ICD automatically adjusts its sensitivity thresholds following certain paced and sensed events (Figure 4-1) to reduce the incidence of T-wave sensing and cross-chamber sensing. The auto-adjusting thresholds increase dramatically (Table 4-1), and then gradually return to their programmed settings. Note that the thresholds may already differ from their programmed values, having been adjusted by the preceding sensed or paced event.
Table 4-1. Auto-Adjusting Sensitivity Threshold Values
Atrial Sensitivity Ventricular Sensitivity
Post-atrial sense
Post-atrial pace
Post-ventricular sense
Post-ventricular pace
Increased to 75% of EGM peak. 200 ms decay constant.
None Increased by 0.45 mV.
None Increased to 75% of EGM peak.
Increased by 4x the programmed value.
b,e
Immediate return after 60 ms.
a
Up to a maximum of 8x the programmed value.
b
If the programmed value exceeds 0.3 mV (ventricular) or 1.2 mV (atrial), the threshold is not adjusted.
c
Until the preceding 450 ms decay, or the programmed threshold, is restored.
d
The exponential decay continues through a subsequent ventricular pacing pulse and its blanking period.
e
Up to a maximum of 1.8 mV.
a
None
60 ms decay constant.
450 ms decay constant
Increased to 4.5x the programmed value. 450 ms decay constant.
b
c
a
d
e
c
4-4
Ventricular EGM
8x
4.5x
Programmed
Sensitivity Threshold
Programmed
Post-Pace Blanking
Figure 4-1. Auto-Adjusting Sensitivity Threshold (Ventricular Shown)
GEM®III DR System Reference Guide
Changing Threshold
Post-Pace Post-Sense
A S
V P
A S
V P
V S
A S
A S
V P
V S
Tachyarrhythmia Detection
Sensing

Blanking Periods

During a blanking period the ICD does not sense electrical signals. Blanking periods avoid sensing ICD outputs, post-pacing polarization, T-waves, and multiple sensing of the same event. The blanking periods following paced events are longer than those following sensed events to avoid sensing the depolarization signal on the electrodes.
Notes:
To enhance sensing and detection during tachyarrhythmias, the ICD does not cross-blank (blank sensing in the opposite chamber) after a sensed event.
During the programmable “Post-Ventricular Atrial Blanking” (PVAB) period, the atrial sensing circuit is active. PVAB is described briefly on page 4-6, and in more detail on page 6-18.
Table 4-2. Programmable Blanking Periods
Atrial blanking after atrial pace 200 – 250 ms
Atrial pace blanking during high voltage therapy 200 – 250 ms
Ventricular blanking after ventricular bradycardia pace 150 – 440 ms
Ventricular pace blanking during high voltage therapy 150 – 440 ms
Ventricular blanking after antitachycardia pace 150 – 440 ms
Table 4-3. Non-Programmable Blanking Periods
Cross-chamber blanking after atrial or ventricular pacing pulse
Atrial blanking after sensed atrial event 100 ms
Ventricular blanking after sensed ventricular event 120 ms
Atrial and ventricular blanking after the end of a cardioversion or defibrillation charging period
Atrial and ventricular blanking after delivered cardioversion or defibrillation therapy
30 ms
300 ms
520 ms
GEM®III DR System Reference Guide
4-5
Tachyarrhythmia Detection
Sensing

Refractory Periods

Pacing refractory periods prevent certain pacing timing intervals from being started by inappropriate sensed signals such as far-field R-waves or electrical noise. Synchronization refractory periods help prevent delivery of pacing or shock pulses during the atrial and ventricular vulnerable periods. (Refer to “Non-Competitive Atrial Pacing” on page 6-38 and “High Voltage Synchronization” on page 5-5 for more detail.)
Note: Refractory periods do not affect tachyarrhythmia detection.
Table 4-4. Refractory Periods
Post-Ventricular Atrial Refractory Period (PVARP) 150 – 500 ms
Post-Ventricular Atrial Blanking (PVAB) 100 – 350 ms
Atrial Refractory Period (ARP) 150 – 500 ms
Ventricular refractory period after charging ends during defibrillation or cardioversion synchronization
Ventricular refractory period after ventricular sense during cardioversion synchronization
or Sensor Varied
400 ms
200 ms
4-6
In dual chamber pacing modes, the Post-Ventricular Atrial Refractory Period (PVARP) controls how the device responds to retrograde
P-waves. Refer to Chapter 6 for more information on PVARP and the PVC Response and PMT Intervention features.
The Post-Ventricular Atrial Blanking (PVAB) period acts as the minimum PVARP value. Sensed atrial events that fall within the PVAB period are ignored by the Mode Switch, PVC Response, PMT Intervention, and NCAP features, but are displayed via Marker Channel™ annotation and are used for Dual Chamber SVT Criteria. Refer to page 6-18 for more information on PVAB.
In the AAI and AAIR modes, the Atrial Refractory Period (ARP) prevents inappropriate atrial inhibition due to sensed far-field R-waves or electrical noise. The first portion of the ARP is a blanking period that disables atrial sensing.
GEM®III DR System Reference Guide

VF Detection

Tachyarrhythmia Detection
VF Detection
The ICD detects VF as either a primary rhythm or as part of a double tachycardia (VF plus SVT).
VF Detection details begin on page 4-9.
Table 4-5. VF Detection Programmable Parameters
VF Enable On or Off
Ventricular Sensitivity
a
(mV)
VF Interval (Rate) V-V intervals shorter than the programmed interval
VF Initial NID Number of Intervals to Detect:
VF Redetect NID Number of Intervals to Redetect:
a
The programmed Ventricular Sensitivity value applies to ventricular tachyarrhythmia detection and bradycardia pacing.
Minimum level of electrical signal that registers as a sensed ventricular event.
are counted as VF events. (Rates are shown in parentheses.)
number of VF events the ICD must count to detect a VF episode.
number of VF events the ICD must count to redetect VF after a therapy.

Programming VF Detection

To ensure proper VF detection, you should not program the VF Interval below 300 ms.
Programming the ventricular sensitivity to 0.3 mV is recommended. A value greater than 0.6 mV is not recommended except for testing and troubleshooting. Also, carefully evaluate the possibility of increased susceptibility to EMI and oversensing before changing the sensitivity to its minimum (most sensitive) setting of 0.15 mV.
The ICD can be programmed to exclude rapidly conducted SVTs from VF Detection. See “Dual Chamber SVT Criteria” on page 4-22.
You can accelerate redetection by programming the Redetect NID lower than the Initial NID.
When VF Detection is enabled for the first time (i.e., at implant) the ICD automatically changes the Automatic Capacitor Formation Interval from
Off to 6 month.
GEM®III DR System Reference Guide
4-7
Tachyarrhythmia Detection
VF Detection
VF Detection cannot be programmed Off unless VT Detection is also programmed
Off.
How to Program VF Detection
1. Select the Params icon and choose the Detection option. The
programmer displays the Detection screen (shown below).
2. Tu r n
3. Position the programming head and select [PROGRAM].
VF Enable to On and select VF Detection parameters.
Figure 4-2. Detection Screen Example
4-8
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
VF Detection

How VF Detection Works

If all Dual Chamber SVT Criteria are Off, the ICD detects VF based on the ventricular rate alone. If any Dual Chamber SVT Criterion is also examines the atrial rhythm to exclude SVTs from inappropriate detection as VF (page 4-22) and to identify a possible double tachycardia (VF plus SVT) (page 4-33).
VF Interval Criterion
When a V-V interval is shorter than the programmed VF Interval, the ICD marks that event as a VF event and tallies it in memory. If the number of VF events reaches the programmed episode and delivers the first programmed VF therapy. Following the therapy, if the number of VF events reaches the redetects VF and delivers the next programmed VF therapy. (See also “Combined Count Detection” on page 4-13.) VF NIDs are defined as a number of VF events within the last sixteen intervals). All NID values are 75% of the detection window: 12/16, 18/24, etc.
VF Initial NID, the ICD detects a VF
VF Redetect NID, the ICD
VF detection window
(e.g., twelve of the
On, the ICD
A S
V S
VF On 12/16 9/12 320 ms
VT On 12 8 400 ms
Dual Chamber SVT Criteria: All Off
Ventricular Sensitivity: 0.3 (mV)
Enab
A S
V S
le
A S
V S
VF Event Counter:
Initial
V S
VF Intervals (< 320 ms)
A R
F
F
F
F
F
S
S
F
S
S
S
S
123456789101112
Redetect
Inter
Figure 4-3. VF Detection Example
GEM®III DR System Reference Guide
A R
A VF episode is declared when the VF counter reaches 12 (of the last 16 events).
F
F
F
F
S
S
v
al
F
F
S
|
S
S
D
200 ms
4-9
Tachyarrhythmia Detection

VT Detection

VT Detection
The ICD detects VT as either a primary rhythm or as part of a double tachycardia (VT plus SVT).
VT Detection details begin on page 4-12.
Table 4-6. VT Detection Programmable Parameters
VT Enable On or Off
Ventricular Sensitivity
a
(mV)
VT Interval (Rate) V-V intervals shorter than the programmed interval
VT Initial NID Number of Intervals to Detect:
VT Redetect NID Number of Intervals to Redetect:
a
The programmed Ventricular Sensitivity value applies to ventricular tachyarrhythmia detection and bradycardia pacing.
Minimum level of electrical signal that registers as a sensed ventricular event.
are counted as VT events. (Rates are shown in parentheses.)
number of VT events the ICD must count to detect a VT episode.
number of VT events the ICD must count to redetect VT after a therapy.
4-10

Programming VT Detection

To ensure proper VT detection, program the VT Interval to a value that is at least 40 ms longer than the patient’s tachycardia interval, and at least 40 ms longer than the programmed
Programming the ventricular sensitivity to 0.3 mV is recommended. (See “Programming for Appropriate Sensing” on page 4-2).
The ICD can be programmed to exclude rapidly conducted SVTs from VT Detection. See “Dual Chamber SVT Criteria” on page 4-22.
You can accelerate redetection by programming the Redetect NID lower than the Initial NID.
When VT Detection is enabled for the first time,
Sinus Tach dual chamber SVT detection criteria are automatically
enabled.
GEM®III DR System Reference Guide
VF Interval.
AFib/AFlutter and
Tachyarrhythmia Detection
To ensure VF Detection backup during VT episodes, VT Detection cannot be
On unless VF Detection is also On.
How to Program VT Detection
1. Select the Params icon and choose the Detection option. The programmer displays the Detection screen (shown below).
VT Detection
2. Tu r n
3. Position the programming head and select [PROGRAM].
VT Enable to On and select VT Detection parameters.
Figure 4-4. Detection Screen Example
GEM®III DR System Reference Guide
4-11
Tachyarrhythmia Detection
VT Detection

How VT Detection Works

If all Dual Chamber SVT Criteria are Off, the ICD detects VT based on the ventricular rate alone. If any Dual Chamber SVT Criterion is also examines the atrial rhythm to exclude SVTs from inappropriate detection as VT (page 4-22) and to identify a possible double tachycardia (VT plus SVT) (page 4-33).
VT Interval Criterion
When the V-V interval is shorter than the programmed VT Interval, the ICD marks that event as a VT event and tallies it in memory. (See also “VT Stability Option” on page 4-15.) If the number of VT events reaches the programmed first programmed VT therapy. After the therapy, if the number of VT events reaches the programmed VT therapy. (See also “Combined Count Detection” on page 4-13.) The ICD counts consecutive VT events. The event counter resets to zero whenever an interval fails to satisfy the Interval criterion, or the VT Stability option, if it is value if an interval is shorter than the programmed
VT Initial NID, the ICD detects a VT episode and delivers the
VT Redetect NID, the ICD redetects VT and delivers the next
On. The VT counter remains at its current
VF Interval.
On, the ICD
A S
VF On 12/16 9/12 320 ms
VT On 12 8 400 ms
Dual Chamber SVT Criteria: All Off
Ventricular Sensitivity (mV): 0.3
A S
V S
Enab
A S
V S
le
A S
V S
V S
VT Counter Value:
Initial
Redetect
A R
V S
A R
T
T
S
S
123456789101112
Figure 4-5. VT Detection Example
VT intervals > 320 and < 400 ms
4-12
GEM®III DR System Reference Guide
T S
Inter
A R
A R
A VT episode is declared when the VT counter reaches 12.
T
T
S
S
v
al
T
S
S
S
T
T
T
T
T
T
S
S
S
200 ms
|
D
A VT event increases the VT counter by one.
A normal sensed beat ( 400 ms) clears the counter to zero.
A VF event suspends, but doesn’t clear, the VT count.

Combined Count Detection

Combined Count detection helps the ICD promptly detect or redetect a ventricular tachyarrhythmia when the interval is fluctuating between the VF and VT zones. The ICD tracks the combined number of events registered on the VF and VT counters, and if this sum reaches the Combined Number of Intervals to Detect (CNID), VF or VT is detected. Combined Count Detection also applies to redetected episodes.
Combined Count detection is automatic (not programmable) when VT Detection is enabled, and cannot be disabled.

How Combined Count Detection Works

Combined Count Detection automatically applies a Combined Number of Intervals to Detect (CNID) when the VF event counter reaches six. The ICD calculates the CNID as follows:
Initial CNID = VF Initial NID x 7/6 (rounded down) Redetect CNID = VF Redetect NID x 7/6 (rounded down)
Tachyarrhythmia Detection
Combined Count Detection
For example, if the VF Initial NID is programmed to 18, the Initial CNID is
21. If the VF Redetect NID is 12, the Redetect CNID is 14.
Combined Count detection is fulfilled when the sum of the VF and VT event counters equals or exceeds the CNID. The ICD then reviews the eight intervals preceding detection:
If
any
of the eight was in the VF zone, the episode is classified as VF.
If
all
eight were outside the VF zone, the episode is classified as VT.
GEM®III DR System Reference Guide
4-13
Tachyarrhythmia Detection
Combined Count Detection
Surface ECG
Marker Channel™
A S
A R
A R
A R
A R
VF Event Count VT Event Count
Without Combined Count Detection, the arrhythmia would be detected after 18 beats
15
VF = 12
10
VF
Counter
5
VT Counter VT Counter
Figure 4-6. Combined Count Detection Example
T
T
S
S
FSF
F
FSF
S
S
S
12 3 45
12
VF NID = 12/16, VT NID = 12, CNID = 14
With Combined Count Detection, the arrhythmia is detected after 14 beats
15
10
VF
VT = 12
Counter
15105
5
T S
3
VF = 12
T
FSF
S
67
FSFSF
S
89
4
Combined Count (VT+VF) = 14
VF = 6
D
10
VT = 12
15105
4-14
GEM®III DR System Reference Guide

VT Stability Option

VT Stability is an optional ventricular SVT criterion that allows the ICD to withhold VT detection for rapid rhythms (in the VT detection zone) with irregular intervals.
VT Stability (ms) Off, or the maximum acceptable variation among

Programming VT Stability Option

Choosing a small VT Stability value (such as 30 ms) may not allow for the normal VT interval variation. Thus, smaller VT Stability values may decrease the sensitivity of the ICD to detect VT.
The VT Stability option compares measured intervals that have been shortened to multiples of 10 ms, permitting a rounding error of up to +10 ms from the programmed VT Stability value.
How to Program VT Stability
1. Select the Params icon and choose the Detection option. The
Tachyarrhythmia Detection
VT Stability Option
Table 4-7. VT Stability Option Programmable Parameter
measured intervals
programmer displays the Detection screen.
2. Select a value for
3. Position the programming head and select [PROGRAM].
VT Stability.
GEM®III DR System Reference Guide
4-15
Tachyarrhythmia Detection
VT Stability Option

How the VT Stability Option Works

When the VT Stability option is On, an interval in the VT zone is considered previous intervals is greater than the programmed VT Stability interval.
The ICD does not apply the VT Stability option until the VT event count reaches three. When an interval in the VT zone fails to meet the VT Stability option, it is marked as a normal sensed event and the VT event count resets to zero.
The VT Stability option must be re-satisfied during redetection if it is enabled.
unstable
if the difference between it and any of the three
VT Stability is not applied until VT count reaches three.
A S
V S
A S
V S
A S
V S
VT Counter Value:
Unstable event: Varies from previous interval
by 70 ms. Event is marked as a normal sensed
A
A
S
S
T
V
S
S
4 8 0
1
beat and VT counter resets to zero.
Enab
le
VF On 12/16 9/12 320 ms
VT On 12 8 540 ms
Dual Chamber SVT Criteria: All Off
Ventricular Sensitivity (mV): 0.3 VT Stability (ms): 50
Stable events: Each differs from the three before it by
50 ms or less, and they are marked as VT events.
A
ASA
T S
4 7 0
A
S
T S
5 1 0
A
S
S
V
T
S
S
4
4
4
7
0
0
A
A
S
S
T
T
S
S
4
4
4
8
0
0
2 30123 2345
Unstable event: Varies from previous interval
by 60 ms. VT counter resets to zero.
Initial
Redetect
A
A
S
S
V S
5
4
0
8
0
0
01
S
T
T
S
S
4 7 0
Inter
A
A
S
S
T
T
S
S
4
4 7 0
v
4
7
7
0
0
200 ms
al
T S
4-16
Figure 4-7. VT Stability Option
GEM®III DR System Reference Guide

FVT Detection

Tachyarrhythmia Detection
FVT Detection
The ICD can be programmed to further classify a tachycardia as Fast Ventricular Tachycardia (FVT). The ICD detects FVT as a primary rhythm or as part of a double tachycardia (FVT plus SVT
)
. It provides an FVT
episode counter and a separate therapy regimen for FVT episodes.
FVT Detection details begin on page 4-19.
Table 4-8. FVT Detection Programmable Parameters
FVT Enable via VF, via VT, or Off
Ventricular Sensitivity
a
(mV)
FVT Interval (Rate) V-V intervals between this value and the
a
The programmed sensitivity value applies to tachyarrhythmia detection and bradycardia pacing.
Minimum level of electrical signal that registers as a sensed ventricular event.
programmed VF Interval are marked as FVT events. (Rates are shown in parentheses.)

Programming FVT Detection

To ensure VF Detection backup, FVT Detection cannot be On unless VF Detection is also
For FVT Detection zone. For FVT Detection
On.
via VF, the FVT Interval is programmed within the VF
via VT, the FVT Interval is within the VT zone. This
choice should depend on the patient’s VF and VT cycle lengths. After determining a reliably sensitive VF Interval, consider the following:
If the patient exhibits a clinical VT cycle length in the VF zone, select
via VF to ensure reliable detection of VF. (VT Detection need not be
enabled at all.)
If the patient exhibits two clinical VTs, both outside the VF zone, select
via VT to prevent misclassification of VF, and to offer a separate
therapy regimen for each VT.
If the patient has only one clinical VT, outside the VF zone, select VF and VT Detection only, with
FVT Enable set to Off.
Select the VF, VT, and FVT Intervals at least 40 ms apart to allow for normal variations in the patient’s interval.
The ICD can be programmed to exclude rapidly conducted SVTs from FVT Detection. See “Dual Chamber SVT Criteria” on page 4-22.
GEM®III DR System Reference Guide
4-17
Tachyarrhythmia Detection
FVT Detection
FVT Detection uses the VF event counter and the programmed VF NID if set to
via VF, or the VT counter and VT NID if set to via VT.
How to Program FVT Detection
1. Select the Params icon and choose the Detection option. The
programmer displays the Detection screen.
2. Set
3. Position the programming head and select [PROGRAM].
FVT Enable to via VT or via VF. Select the FVT Interval and
Ventricular Sensitivity values.
Figure 4-8. Programming FVT Detection
4-18
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
FVT Detection

How FVT Detection Works

If all Dual Chamber SVT Criteria are Off, the ICD detects FVT based on the ventricular rate alone. If any Dual Chamber SVT Criteria is also examines the atrial rhythm to exclude SVTs from inappropriate detection as FVT (page 4-22) and to identify a possible double tachycardia (FVT plus SVT) (page 4-33).
FVT Detection Via VF
When you set FVT Detection to via VF, the ICD detects FVT episodes using the VF event counter.
A V-V interval within the FVT zone is marked as an “FVT via VF” event and tallied on the more sensitive VF event counter. When the VF NID is reached, the ICD reviews the last eight intervals.
If
any
of the last eight intervals is in the VF zone, it detects the
episode as VF.
If
all
of the last eight intervals are outside the VF zone, it detects the
episode as FVT (Figure 4-9).
On, the ICD
FVT Detection Via VT
When you set FVT Detection to via VT, the ICD detects FVT episodes using the VT event counter.
A V-V interval within the FVT zone is marked as an “FVT via VT” event and tallied on the more specific, consecutive VT event counter. The VT Stability option must also be satisfied if it is enabled. When the VT NID is reached, the ICD reviews the last eight intervals.
If
any
of the last eight intervals is in the VF or FVT zones, it detects
the episode as FVT.
If
all
of the last eight intervals are outside the FVT and VF zones, it
detects the episode as VT.
GEM®III DR System Reference Guide
4-19
Tachyarrhythmia Detection
FVT Detection
VF NID reached. Since none of the last eight intervals is a VF interval, the ICD classifies the episode as FVT and
Combined Count Detection of FVT
Combined Count Detection (page 4-13) also applies to the FVT zone. If the VF event counter reaches six, and the sum of the VF and VT event counters equals or exceeds the CNID, the ICD “looks back” at the eight intervals preceding detection:
If any of the eight was in the VF zone, the episode is detected as VF.
If none was in the VF zone but one or more was in the FVT zone, the episode is detected as FVT.
If all eight were outside the VF and FVT zones, the episode is classified as VT.
will deliver an FVT therapy.
4-20
V S
V S
V S
V S
VF Counter:
V
T
S
F
T
T
.
.
F
F
T
T
.
F
T
.
.
F
F
1234567 89101112
(8 INTERVALS BEFORE NID)
Enab
le
Initial
Redetect
VF On 12/16 9/12 320 ms
FVT via VF 260 ms
VT Off
Figure 4-9. FVT Detection Via VF Example
GEM®III DR System Reference Guide
T
.
.
F
.
.
F
F
T
T
LOOKBACK WINDOW
Inter
v
al
T
.
F
200 ms
T
|
F
T
.
F
Blanking
Depolarization
Blanking
Depolarization
VF Detection Zone
120 ms
120 ms
Tachyarrhythmia Detection
FVT Detection
Zone Merging After Detection
For FVT via VF detection, the FVT zone merges with the VF zone after detection or redetection of VF. For FVT via VT detection, the FVT zone merges with the VT zone after detection or redetection of FVT or VF. In both instances, the merged-zone configuration remains until episode termination. The merged zone uses the event counting and therapies for the faster arrhythmia, to ensure that the ICD delivers sufficiently aggressive therapies during an extended or highly variable tachyarrhythmia episode. (See also “Progressive Episode Therapies” on page 5-36.)
Before initial detection: three detection zones
VF
Interval
VT Detection Zone
FVT Detection Zone
FVT
Interval
After detection of VF: two zones for redetection
VF Detection Zone VT Detection Zone
VF
Interval
VT
Interval
VT
Interval
Lower Rate
Period
Lower Rate
Period
Blanking
Depolarization
Blanking
Depolarization
VF Detection Zone
120 ms
VF Detection Zone
120 ms
Figure 4-10. Zone Merging (FVT Detection Via VF)
Before initial detection: three detection zones
FVT Detection Zone
VF Interval FVT
VT Detection Zone
Interval
After detection of VF or FVT: two zones for redetection
FVT Detection Zone
VF
Interval
Figure 4-11. Zone Merging (FVT Detection Via VT)
GEM®III DR System Reference Guide
Lower Rate
Period
VT
Interval
Lower Rate
Period
VT
Interval
4-21
Tachyarrhythmia Detection

Dual Chamber SVT Criteria

Dual Chamber SVT Criteria
The programmable Dual Chamber SVT Criteria are designed to withhold inappropriate ventricular detection during episodes of rapidly conducted supraventricular tachycardia (SVTs). The ICD analyzes the activation patterns and timing in both chambers, using PR Logic™ Pattern and Rate Analysis to identify different SVTs. You can specify a range of ventricular rates over which the ICD applies the SVT Criteria.
See page 4-28 for more details on the Dual Chamber SVT Criteria and PR Logic™ Pattern and Rate Analysis.
Table 4-9. Dual Chamber SVT Criteria Programmable Parameters
AFib/AFlutter Identifies rapidly conducted atrial fibrillation, atrial
Sinus Tach Identifies sinus tachycardia (see page 4-34).
1:1 VT-ST Boundary Threshold between the Retrograde and Antegrade
Other 1:1 SVTs Identifies other 1:1 SVTs where the atrial and
SVT Limit Defines the minimum ventricular cycle length for the
flutter, or atrial tachycardia (see page 4-34).
zones used by A-V Pattern Analysis (see page 4-29). Allows customization of Sinus Tach criterion for patients with slow 1:1 conduction.
ventricular activation are roughly simultaneous (see page 4-34).
ICD to apply the Dual Chamber SVT Criteria.
4-22
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
Dual Chamber SVT Criteria

Programming Dual Chamber SVT Criteria

#
Cautions:
Before enabling the Other 1:1 SVTs criterion, ensure that the atrial lead has matured. This criterion could inappropriately withhold therapy if atrial sensing is compromised by an unstable or dislodged atrial lead.
In patients who exhibit slow 1:1 retrograde conduction during VF or VT, use caution when programming the
Other 1:1 SVTs criterion. This
criterion could inappropriately withhold VF/VT therapy in such patients. (See Figure 4-15 on page 4-29).
When VT Detection is enabled for the first time,
Sinus Tach dual chamber SVT detection criteria are automatically
AFib/AFlutter and
enabled.
When enabling the
Sinus Tach criterion in patients who exhibit slow 1:1
retrograde conduction during VF or VT, or slow antegrade conduction during SVTs, make sure to set the
1:1 VT-ST Boundary parameter
appropriately. See “Customizing Detection for Patients with Slow Conduction” on page 4-25.
The nominal atrial sensitivity threshold is 0.3 mV. However, the threshold must not permit far-field R-wave sensing. See “Programming for Appropriate Sensing” on page 4-2.
To ensure therapy for hemodynamically compromising rates regardless of origin, the ICD always delivers therapy when the median ventricular cycle length is less than the programmed
SVT Limit (nominally 320 ms) if VT,
VF, or FVT detection criteria are satisfied.
If you program an
SVT Limit greater than the VF Interval, the Dual
Chamber SVT Criteria are effectively disabled for VF detection.
GEM®III DR System Reference Guide
4-23
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
How to Program the Dual Chamber SVT Criteria
1. Select the Params icon and choose the Detection option.
2. Set the AFib/AFlutter, Sinus Tach, or Other 1:1 SVTs parameters,
as appropriate.
3. Select an
4. Select an atrial Sensitivity value.
5. Position the programming head and select [PROGRAM].
Shading (blue) indicates the cycle lengths where
Dual Chamber SVT criteria are applied
SVT Limit value.
Atrial sensitivity value
4-24
GEM®III DR System Reference Guide
Dual Chamber SVT Criteria
Figure 4-12. Dual Chamber SVT Parameters
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
Customizing Detection for Patients with Slow Conduction
Slow antegrade conduction during an SVT or slow 1:1 retrograde conduction during VT can adversely affect the accuracy of the and Other 1:1 SVTs criteria. The ICD provides the programmable 1:1 VT-ST
Boundary parameter to customize the Sinus Tach criterion for patients that
exhibit slow conduction.
The
1:1 VT-ST Boundary parameter represents an A-V interval that is a
percentage of a V-V interval. This value forms the division between the Retrograde and Antegrade zones used by ICD pattern analysis operations (see “Pattern Analysis of A-V and V-A Intervals” on page 4-29).
If slow conduction causes antegrade events to occur in the Retrograde Zone, or retrograde events to occur in the Antegrade Zone, changing the
1:1 VT-ST Boundary parameter can increase the size of the appropriate
zone. Figure 4-13 below demonstrates the possible settings for the
VT-ST Boundary.
Sinus Tach
1:1
#
Caution: The Other 1:1 SVTs criterion is
Boundary. However, slow conduction can adversely affect its operation by
not
affected by the 1:1 VT-ST
causing atrial events to occur within junctional zones. Use caution when enabling the
Other 1:1 SVTs criterion in patients with slow antegrade or
retrograde conduction.
1:1 VT-ST Boundary
V-event V-event
Junctional
Zone
50 ms
Retrograde zone
(before boundary)
settings
50%66%75%85%
35%
Antegrade zone (after boundary)
Junctional
Zone
80 ms
Figure 4-13. Possible 1:1 VT-ST Boundary Settings
GEM®III DR System Reference Guide
4-25
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
How to Set the 1:1 VT-ST Boundary
#
Caution: The 1:1 VT-ST Boundary parameter should
50% if the patient exhibits slow 1:1 retrograde conduction during a VT
(long V-A intervals), or slow antegrade conduction during an SVT (long A-V intervals). Incorrect programming of this parameter may result in inappropriate therapies or underdetection of tachyarrhythmias.
1. Select the
2. Select the Additional Settings... parameter.
3. Select the 1:1 VT-ST Boundary... parameter.
Params icon and choose the Detection option.
only
be changed from
4-26
Figure 4-14. Programming the 1:1 VT-ST Boundary
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
Note: Before 1:1 VT-ST Boundary can be selected, Sinus Tach must be set to
On.
4. Select 1:1 VT-ST Boundary, and set the 1:1 VT-ST Boundary
parameter as follows:
a. If the patient exhibits long V-A intervals during VT with 1:1
retrograde conduction, select
35%.
b. If the patient exhibits long A-V intervals during an SVT, select a
value that exceeds the A-V / V-V ratio observed during stored SVT episodes (see “Atrial Interval Display Options” on page 10-22).
5. Select
[OK] in the 1:1 VT-ST Boundary and Additional Options
dialog boxes.
6. Position the programming head and select
[PROGRAM].
GEM®III DR System Reference Guide
4-27
Tachyarrhythmia Detection
Dual Chamber SVT Criteria

How the Dual Chamber SVT Criteria Work

Elements of PR Logic™ Pattern and Rate Analysis 4-29 Double Tachycardia Detection 4-33 Dual Chamber SVT Criteria Requirements 4-34
When a high ventricular rate fulfills the VF, VT, or FVT detection criteria (NID or CNID – see page 4-9), and the V-V median interval exceeds the
SVT Limit selected by the physician, the ICD applies the Dual
1
equals or
Chamber SVT Criteria.
The Dual Chamber SVT Criteria use PR Logic™ Pattern and Rate Analysis to evaluate the episode. PR Logic™ analyzes the patterns, rates, and timing of activation in both chambers to identify the high ventricular rate as either a double tachycardia; one of the programmed classes of SVT; or a ventricular arrhythmia.
If PR Logic™ identifies a dual tachyarrhythmia (VF, VT or FVT in the presence of an SVT – see page 4-33), the ICD delivers therapy for the ventricular arrhythmia.
If any Dual Chamber SVT Criterion is met, PR Logic™ identifies a rapidly conducted SVT. The ICD withholds therapy and continues monitoring the rate, patterns, and timing on each ventricular event.
If the Dual Chamber SVT Criteria are not met, PR Logic™ identifies a true ventricular arrhythmia (not related to a conducted SVT), and delivers VF, VT, or FVT therapy as programmed.
Dual Chamber SVT Criteria are applied only during initial detection.
4-28
1
Median of the 12 most recent ventricular interval values. (Note that the rhythm is detected and treated as VF, VT, or FVT if the V-V median is less than the
GEM®III DR System Reference Guide
SVT Limit.)
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
Elements of PR Logic™ Pattern and Rate Analysis
The ICD uses multiple criteria to evaluate and classify rhythms as a particular type of SVT. PR Logic™ analyzes the following aspects of atrial and ventricular activation to identify SVTs:
Pattern of A-V and V-A intervals
Rate – atrial and ventricular
AF evidence
Far-field R-wave sensing
A:V dissociation
V-V regularity
Pattern Analysis of A-V and V-A Intervals
The ICD uses pattern analysis to identify sinus tachycardia, atrial flutter, and other 1:1 SVTs. Within each V-V interval, the ICD categorizes the atrial activation according to the number of intervening atrial events and the zones in which those atrial events occur. (See Figure 4-15 below.) From this information, the ICD assigns pattern codes to the intervals, and interprets the pattern codes to identify different SVTs.
V-event V-event
Atrial events here
indicate:
Programmed
1:1 VT-ST Boundary
(set to 50%)
Junctional Zone
50 ms
PAC, PVC, junctional rhythm, atrial fibrillation, or atrial flutter
Retrograde Zone Antegrade (Normal)
Zone
Retrograde conduction Normal conduction
(sinus rhythm, sinus
tachycardia)
Figure 4-15. Zones Used During A-V Pattern Analysis
GEM®III DR System Reference Guide
Junctional Zone
80 ms
PAC, PVC, junctional rhythm, atrial fibrillation, or atrial flutter
4-29
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
AF Evidence
AF Evidence is provided by a counter that accrues evidence to help identify atrial fibrillation, or to help detect a double tachycardia (VF, VT or FVT in the presence of an SVT).
For each ventricular event, if the ICD identifies A-V pattern information for a high atrial rate, timing consistent with atrial tachyarrhythmia (see Figure 4-15) and greater than 1:1 conduction, the ICD adds one to the AF Evidence counter. If the A-V pattern is inconclusive, inconsistent, or if the ICD detects far-field R-wave sensing, the AF counter is unchanged. If no atrial event occurs within the current V-V interval or a consistent 1:1 pattern is present, the ICD subtracts one from the counter.
The AF Evidence criterion is satisfied when the AF Evidence count is greater than or equal to 6 (up to a maximum value of 10). Once the criterion is met, it remains satisfied for as long as the AF Evidence count is greater than or equal to 5.
4-30
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
Far-Field R-Wave Sensing
The ICD identifies far-field R-wave oversensing in the atrium in order to exclude far-field R-waves from the process of SVT classification.
If there are two atrial events within a ventricular interval, one of the atrial events may actually be a far-field R-wave. The ICD identifies a sensed far-field R-wave if it detects both:
a short-long pattern of A-A intervals, and
a short A-V interval (< 60 ms) or a short V-A interval (< 160 ms)
The ICD uses far-field R-wave sensing for the Sinus Tach and A.Fib/A.Flutter criteria as follows:
To identify Sinus Tach with far-field R-wave sensing, the ICD requires a far-field R-wave on at least 4 of the most recent 12 ventricular intervals.
To identify A.Fib/A.Flutter with far-field R-wave sensing, the ICD requires a far-field R-wave on at least 10 of the most recent 12 ventricular intervals.
VA
A
1
AA
2
Far-field R-wave
A
AA
1
V
A
2
AV
V
Figure 4-16. Intervals Evaluated in Far-Field R-Wave Determination
A:V Dissociation
A:V Dissociation provides cumulative evidence that there is no direct relationship between sensed atrial and ventricular events. The ICD identifies a rhythm as A:V dissociated if at least 4 of the most recent 8 ventricular intervals exhibit either:
no atrial events in the ventricular interval, or
an A-V interval that differs from the average of the previous eight A-V intervals by more than 40 ms.
GEM®III DR System Reference Guide
4-31
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
The ICD uses this count to help identify a double tachycardia (VF, VT or FVT in the presence of an SVT).
Ventricular Cycle Length Regularity
The ICD uses ventricular cycle length regularity or irregularity to evaluate atrial fibrillation, double tachycardia (VF, VT or FVT in the presence of an SVT), and other 1:1 SVTs.
The ICD continuously measures the regularity of the ventricular cycle lengths. The regularity count indicates how often the two most commonly occurring intervals (of at least 240 ms) occurred among the last 18 ventricular intervals.
For example, the figure below illustrates the 18 most recent intervals. The two most frequently occurring intervals are 330 ms (five intervals) and 320 ms (three intervals). Together these two account for eight of the 18 most recent intervals, for a regularity count of 44%.
5
3
1
00 0 0 0 0000000
0
240
260 280 300 320 340 360 380 400 420 440 460
2 2
1 1 1 11
4-32
For a double tachycardia to be detected, the regularity count must be at least 75%.
To withhold detection under the atrial fibrillation rule of the
AFib/AFlutter criterion, the regularity count must be 50% or less.
Atrial fibrillation with regularity between 50% and 75% is identified by the atrial flutter rule, but the count is not required.
To withhold detection for Other 1:1 SVTs, the regularity count must be at least 25%.
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
Double Tachycardia Detection
If PR Logic™ Pattern and Rate Analysis identifies double tachycardia, the ICD delivers the programmed ventricular therapies.
The ICD detects VF plus SVT or FVT via VF plus SVT if all of the following occur:
an atrial tachyarrhythmia is identified (page 4-30),
ventricular detection via the Interval or Combined Count criterion,
V-V median1≥ SVT Limit, and
the rhythm is A:V dissociated (page 4-31).
The ICD detects VT plus SVT or FVT via VT plus SVT if all of the above occur and:
the ventricular cycle length is very regular (regularity of at least 75% – see page 4-32).
1
Median of the 12 most recent ventricular interval values. (Note that the rhythm is detected and treated as VF, VT, or FVT if the V-V median is less than the
SVT Limit.)
GEM®III DR System Reference Guide
4-33
Tachyarrhythmia Detection
Dual Chamber SVT Criteria
Dual Chamber SVT Criteria Requirements
The AFib/AFlutter criterion consists of two independent rules: the atrial fibrillation rule and the atrial flutter rule. If either rule is satisfied, the
AFib/AFlutter criterion is met and the ICD withholds ventricular detection
and therapy.
The atrial fibrillation rule requires that:
the AF Evidence counter (page 4-30) indicates atrial fibrillation,
exclusive of far-field R-wave sensing (page 4-31);
the A-A median is 94% or less of the V-V median; and
the ventricular cycle length is not regular (regularity of 50% or
less, see page 4-32).
The atrial flutter rule is satisfied if A-V pattern information indicates atrial flutter, exclusive of far-field R-wave sensing.
The
Sinus Tach criterion is met if A-V pattern analysis indicates a 1:1 sinus
tachycardia (with atrial events primarily in the Antegrade Zone shown in Figure 4-15 on page 4-29), exclusive of far-field R-wave sensing. The size and timing of the Antegrade Zone is dependent on the parameter (see page 4-25).
1:1 VT-ST Boundary
4-34
The
Sinus Tach criterion also recognizes and withholds inappropriate
detection for a 1:1 sinus tachycardia when far-field R-wave oversensing consistently occurs.
The
Other 1:1 SVTs criterion is satisfied when A-V pattern information
indicates a 1:1 SVT in which the atria and ventricles are activated at approximately the same time; i.e., a junctional tachycardia (consistent atrial sensing in a junctional zone; see Figure 4-15 on page 4-29).
GEM®III DR System Reference Guide

Termination and Redetection

Outcome Monitoring

Tachyarrhythmia Detection
Termination and Redetection
After delivering an automatic therapy, the ICD monitors the cardiac interval for three possible outcomes:
Termination of the episode.
Redetection of the original ventricular tachyarrhythmia.
Redetection of a different ventricular arrhythmia, including
1
VT acceleration.
Further automatic ventricular therapies are not delivered unless the patient‘s cardiac rhythm fulfills the redetection requirements for a ventricular tachyarrhythmia.
The Dual Chamber SVT Criteria are not applied during redetection. However, the VT Stability criterion must be continually re-satisfied throughout redetection if it is enabled.

Programming Considerations

You can accelerate redetection by programming Redetect NIDs lower than the Initial NIDs.
Both the VF Redetect NID and the VT Redetect NID take effect during any type of ventricular arrhythmia episode.
If you program the page 5-36), a slower arrhythmia (e.g., VT) may receive therapies for a faster arrhythmia (e.g., VF) during redetection.
Progressive Episode Therapies option On (see
1
VT Detection is temporarily suspended for 17 events, paced or sensed, following defibrillation therapy.
GEM®III DR System Reference Guide
4-35
Tachyarrhythmia Detection
Termination and Redetection
Therapy Efficacy
A therapy is considered successful if arrhythmia termination occurs prior to any redetection. The therapy is considered unsuccessful if redetection occurs before termination.
If all programmed therapies are unsuccessful, the ICD suspends them until either:
the episode terminates, or
Detection is manually reset (by programming Detection Enable Off and then On again) during the episode.

Episode Termination

Episode termination occurs when eight consecutive V-V intervals are greater than or equal to the VT detection interval.
1
Once the ICD detects an arrhythmia, it considers the episode as ongoing until it detects episode termination. Any subsequent detection after termination marks the start of a new episode.
After antitachycardia pacing therapy, the counter for episode termination begins with the first ventricular cycle. After cardioversion or defibrillation, the counter for episode termination begins with the second ventricular event following the shock. (Due to the extended post shock blanking, this event may be the
third
event on the electrogram.)
A rhythm that does not satisfy the termination definition is considered part of an ongoing episode, but the ICD does not deliver the next programmed therapy until the redetection criteria are fulfilled.

Redetection

The ICD redetects an arrhythmia if the VF or VT event counter reaches its Redetect NID, or if the combined VF and VT event counters reach the Redetect CNID (see “Combined Count Detection” on page 4-13). The ICD then delivers the next programmed therapy for the current arrhythmia, and resumes monitoring for the outcome of that therapy.
4-36
1
VF Interval when VT Detection is programmed Off.
GEM®III DR System Reference Guide
Tachyarrhythmia Detection
Termination and Redetection
If the ICD redetects VF or an accelerated VT after an antitachycardia pacing sequence delivery, it skips the subsequent pacing therapy sequences for the duration of this episode and proceeds to deliver the next therapy programmed for the current arrhythmia.
Figure 4-17 illustrates redetection of VT after cardioversion therapy.
VT Acceleration
The ICD classifies a VT as accelerated when it redetects during a VT episode and the average of the four intervals before redetection has decreased by 60 ms or more, compared to the average of the four intervals before the initial VT detection. to evaluate future VT accelerations under the 60 ms rule above.
If VT acceleration occurs after the delivery of an antitachycardia pacing sequence, the ICD skips the subsequent sequences of the pacing therapy for the duration of this episode and delivers the next programmed VT therapy.
1
The current redetection is used
1
Or before the most recent “accelerated” VT redetection.
GEM®III DR System Reference Guide
4-37
Tachyarrhythmia Detection
Termination and Redetection
Not sensed in the blanking period after cardioversion.
cardioversion shock.
0.6 joule
VRV
A R
C
S
D
A R
C
T
|
E
D
A S
V
T
S
S
3 3 0
After eight consecutive intervals within the VT zone, the VT Redetect NID is fulfilled and the ICD redetects the VT episode.
Enable Initial Redetect Interval
VF On 12/16 9/12 320 ms
VT On 12 8 400 ms
VT Therapy: 1 2 3 4 5 6
Therapy Status: On On On On On On
Therapy Type: CV CV CV CV CV CV
Energy (J): 0.6 10 30 30 30 30
Sensed, measured as 660 ms, and marked as a normal sensed event.
A S
T
T
T
S
S
3
3 3 0
3
3
3
0
0
S
3 3 0
TSTST
A R
T
|
D
S
3
3 3 0
3
3
3
0
0
200 ms
The therapy was unsuccessful: the ICD redetected VT before episode termination (eight consecutive beats outside the detection zones).
4-38
Figure 4-17. Redetection of VT After Therapy
GEM®III DR System Reference Guide
Tachyarrhythmia Therapies
5
5
High Voltage Therapy Overview 5-2
Defibrillation 5-9
Cardioversion 5-17
Antitachycardia Pacing Therapy Overview 5-23
Programming ATP Therapies 5-25
Burst Pacing 5-27
Ramp Pacing 5-29
Ramp+ Pacing 5-31
Smart Mode 5-33
Progressive Episode Therapies 5-36
GEM®III DR System Reference Guide
5-1
Tachyarrhythmia Therapies

High Voltage Therapy Overview

High Voltage Therapy Overview
Defibrillation and cardioversion are the two high voltage therapies provided by the ICD. Up to six defibrillation shocks may be programmed to treat a detected episode of VF. Cardioversion therapy can be programmed as any of the six VT or FVT therapies.
When an arrhythmia is detected and defibrillation or cardioversion is the next programmed therapy, the ICD charges the high voltage capacitors to the programmed energy. When the programmed energy is reached, the ICD initiates a synchronization sequence and attempts to synchronize the pulse to the leading edge of a sensed ventricular event.
Defibrillation and cardioversion therapies can be delivered via the following high voltage electrodes:
CAN (HVA) – device can
RV (HVB) – RV coil
SVC (HVX) – optional supplemental electrode; e.g. SVC coil
Each defibrillation or cardioversion therapy has separately programmable energy and pathway polarity (see “High Voltage Therapy Parameters” on page 5-7). The ICD automatically regulates the pulse width to obtain 50% tilt.
5-2
Comparison of Defibrillation and Cardioversion
A defibrillation shock is delivered for VF; a cardioversion shock is delivered for VT or FVT.
When the ICD delivers a defibrillation therapy, it attempts to synchronize with a sensed R-wave, but does not require such synchronization. However, a cardioversion therapy requires synchronization to a sensed R-wave for delivery to occur.
GEM®III DR System Reference Guide
Loading...