Detecting and treating tachyarrhythmias 21
Treating bradycardia 22
Monitoring for real-time and stored data 22
Conducting electrophysiologic tests 22
Alerting the patient to system events 22
Indications and usage 23
Contraindications 23
Patient screening 23
Marquis DR 7274 Reference Manual
iv
Table of contents
3Emergency therapy 25
Delivering emergency therapies 26
Effect on system operation 26
Aborting an emergency therapy 26
On-screen and display panel buttons 27
Temporary parameter values 27
How to deliver emergency 30 joule defibrillation 28
How to deliver emergency cardioversion 28
How to deliver emergency fixed burst pacing 29
How to deliver emergency VVI pacing 30
Part IIDevice implant and patient follow-up procedures
4Implanting the ICD 33
Overview 34
Preparing for an implant 34
Equipment for an implant 34
Sterile supplies 35
How to prepare for implanting 35
Replacing an ICD 36
How to explant and replace an ICD 37
Positioning the leads 37
Using transvenous leads 37
Using epicardial leads 38
Surgical incisions 39
Testing sensing and pacing thresholds 39
Parameters 40
Considerations 40
Connecting the leads to the ICD 40
How to connect the lead to the device 41
Testing defibrillation operation and effectiveness 42
High voltage implant values 42
Binary search protocol 43
How to prepare for defibrillation threshold testing 44
How to perform defibrillation threshold testing 45
Positioning and securing the ICD 46
How to position and secure the device 46
Completing the implant procedure 47
Marquis DR 7274 Reference Manual
How to complete programming the device 47
5Conducting a patient follow-up session 49
Patient follow-up guidelines 50
Verifying the status of the implanted system 50
Verifying accurate detection and appropriate therapy 51
Considerations 51
Verifying effective bradycardia pacing 52
Considerations 53
Part IIIConfiguring the ICD for the patient
6Detecting tachyarrhythmias 57
Detection overview 58
Suspending tachyarrhythmia detection 60
Setting up sensing 61
Parameters 61
Considerations 61
How to program sensitivity 63
Details about sensing 64
Detecting VF episodes 66
Parameters 66
Considerations 67
Restrictions 68
How to program VF detection 68
Details about VF detection 68
Detecting VT episodes 70
Parameters 70
Considerations 70
Restrictions 71
How to program VT detection 72
Details about VT detection 72
Detecting FVT episodes 76
Parameters 76
Considerations 76
Restrictions 77
How to program FVT detection 78
Details about FVT detection 78
Detecting tachyarrhythmia episodes with Combined Count 81
Table of contents
v
Marquis DR 7274 Reference Manual
vi
Table of contents
Details about Combined Count detection 81
Monitoring episodes for termination or redetection 84
Parameters 84
Considerations 84
How to program redetection parameters 85
Details about episode termination and redetection 85
Enhancing detection with PR Logic criteria 88
Parameters 88
Considerations 89
Restrictions 90
How to program the PR Logic detection criteria 91
Details about PR Logic pattern and rate analysis 91
Details about the PR Logic detection criteria 95
Enhancing VT detection with the Stability criterion 97
Parameters 97
Considerations 97
How to program Stability 98
Details about Stability 98
Detecting double tachycardias 100
Details about double tachycardia detection 100
Detecting prolonged tachyarrhythmias with High Rate Timeout 101
Parameters 101
Considerations 101
How to program High Rate Timeout 102
Details about High Rate Timeout 102
Key terms 103
7Treating tachyarrhythmia episodes 109
Treating VF with defibrillation 110
Parameters 110
Considerations 111
Restrictions 112
How to program VF therapies 112
Details about VF therapy 112
Treating VT and FVT with antitachycardia pacing 120
Parameters for all ATP therapies 120
Parameters for Burst therapy 121
Parameters for Ramp therapy 121
Parameters for Ramp+ therapy 122
Marquis DR 7274 Reference Manual
Table of contents
Considerations 122
Restrictions 123
How to program ATP therapies 123
Details about ATP therapies 124
Treating VT and FVT with cardioversion 130
Parameters 130
Considerations 131
Restrictions 132
How to program cardioversion therapies 132
Details about cardioversion therapy 133
Optimizing therapy with Smart Mode and Progressive Episode
Therapies 139
Parameters 139
Considerations 139
Restrictions 140
How to program Smart Mode 140
Details about Smart Mode 141
How to program Progressive Episode Therapies 142
Details about Progressive Episode Therapies 142
Key terms 143
vii
8Treating bradycardia 147
Providing basic pacing therapy 148
Parameters 148
Considerations 149
Restrictions 151
How to program bradycardia pacing parameters 151
Details about basic bradycardia pacing 152
Dual-chamber pacing 154
Parameters 154
Considerations 156
Details about dual-chamber pacing 156
Programming considerations for atrial rates 161
Single chamber pacing 164
Parameters 164
Considerations 164
Details about single chamber pacing 165
Enhancing pacing for optimal cardiac output 167
Parameters 167
Marquis DR 7274 Reference Manual
viii
Table of contents
Considerations 169
Restrictions 170
How to program Rate Response 170
Details about Rate Responsive Pacing 171
How to program Rate Adaptive AV 175
Details about Rate Adaptive AV 175
How to program Single Chamber Hysteresis 176
Details about Single Chamber Hysteresis 177
How to program Ventricular Rate Stabilization 178
Details about Ventricular Rate Stabilization 178
Adjusting the pacing rate with Mode Switch 180
Parameters 180
Considerations 180
Restrictions 181
How to program Mode Switch 182
Details about Mode Switch 182
Preventing competitive atrial pacing 184
Parameters 184
Considerations 184
How to program Non-Competitive Atrial Pacing 185
Details about Non-Competitive Atrial Pacing 185
Detecting and preventing pacemaker-mediated tachycardia 189
Parameters 189
How to Program PVC Response and PMT Intervention 190
Details about PVC Response 190
Details about PMT Intervention 191
Providing Ventricular Safety Pacing 193
Parameters 193
Restrictions 193
How to program Ventricular Safety Pacing 194
Details about Ventricular Safety Pacing 194
Providing pacing after high voltage therapies 196
Parameters 196
Considerations 196
How to program pacing after high voltage therapies 197
Details about Post Shock Pacing parameters 197
Key terms 198
Marquis DR 7274 Reference Manual
9Optimizing charge time and device longevity 203
Optimizing charge time 204
Parameters 204
Considerations 204
How to evaluate charging performance 205
Details about managing charge time 205
Optimizing device longevity 207
Considerations 207
Key terms 208
Part IVEvaluating and managing patient treatment
10Using the programmer 211
Connecting the programmer hardware 212
Using an external printer 213
To view a list of supported printers214
Materials you will need 214
To connect the printer to your programmer 216
Using the programming head 218
During an episode in progress 218
During marker transmissions 218
How capacitor charging affects the light array 219
Alternative program and interrogate buttons 219
Display screen features 219
Programmer status bar display 220
Live Rhythm Monitor window 221
Task area 222
Command bar 224
Tool palette 224
Buttons 226
Setting programmer preferences 226
How to set the programmer time and date 227
How to set audio preferences 227
How to select a different language 228
How to check the software version number 229
How to set printing preferences 229
How to set reports preferences 230
Starting and ending patient sessions 230
Table of contents
ix
Marquis DR 7274 Reference Manual
x
Table of contents
How to start a patient session 231
How to end a patient session 232
Automatic interrogation 232
Viewing live waveforms 233
Parameters 233
How to use the Adjust window 234
How to use the waveform adjustment button bar 235
Details about the live rhythm monitor 235
Recording live waveforms 240
Printing a report while recording live ECG 241
Saving and retrieving device data 242
Considerations 242
How to save ICD data to a disk 243
How to read ICD data from a disk 243
Saving data to a disk 245
Data file names 245
Reading device data from diskette 245
Printing reports 246
How to print a report 247
Print Queue 247
Key terms 248
11Using system evaluation tools 249
A summary of system evaluation tools 250
Automatic daily measurements 250
Taking a quick look at device activity 251
How to use Quick Look 251
Quick Look observations 252
Using the Patient Alert feature 253
Parameters 253
Considerations 255
How to program the Patient Alert feature 258
Instructing the patient 258
Viewing the Patient Alert events 259
How to view the Patient Alert events 260
Streamlining follow-ups with Checklist 261
How to select and use a checklist 262
How to create, edit, and delete a checklist 263
Marquis DR 7274 Reference Manual
Table of contents
Key terms 264
12Setting up and viewing collected data 265
A summary of data collection 266
Setting up data collection 267
Parameters 267
Considerations 268
How to set up data collection 269
Details about data collection parameters 269
How to clear collected data 273
Collecting lead performance data 274
Daily lead impedance and EGM amplitude measurements 274
Sensing integrity counter 275
Viewing the episode and therapy efficacy counters 275
How to view and print counter data 276
Details about the episode and therapy efficacy counters 277
Therapy counters 279
Viewing episode data 280
How to view episode data 281
Details about episode data 282
How to view an Interval Plot 284
How to view an EGM Strip 285
How to display the episode text information 287
Viewing Flashback Memory 288
How to view the Flashback Memory 289
Viewing battery and lead status data 290
How to view battery and lead status data 292
Viewing lead performance trends 292
How to view lead performance trend graphs 293
Using Cardiac Compass to view long term clinical trends 294
How to print a Cardiac Compass report 296
Details on Cardiac Compass trend data 296
Viewing and entering patient information 301
How to view and enter new patient information 303
Displaying and printing patient information 303
Automatic device status monitoring 304
Device status indicator warnings 304
How to respond to an electrical reset 305
xi
Marquis DR 7274 Reference Manual
xii
Table of contents
13Testing the system 311
14Conducting Electrophysiologic Studies 323
Key terms 306
Testing overview 312
Evaluating the underlying rhythm 312
Considerations 312
How to Perform an Underlying Rhythm test 313
Measuring pacing thresholds 313
Parameters 314
Considerations 315
How to perform a Pacing Threshold Test 315
Measuring lead impedance 316
Considerations 316
How to perform a Lead Impedance test 316
Details about the Lead Impedance test 317
Measuring EGM Amplitude 318
Parameters 318
Considerations 318
Restrictions 319
How to perform an EGM Amplitude test 320
Testing the device capacitors 320
Considerations 320
How to perform a Charge/Dump test 321
Key terms 322
EP Study overview 324
Inducing VF with T-Shock 326
Parameters 326
Considerations 327
Restrictions 327
How to deliver a T-Shock induction 328
Details about T-Shock
induction 328
Inducing VF with 50 Hz Burst 329
Parameters 329
Considerations 330
How to deliver a 50 Hz Burst induction 331
Details about 50 Hz Burst induction 332
Inducing an arrhythmia with Manual Burst 332
Marquis DR 7274 Reference Manual
Parameters 332
Considerations 333
How to deliver a Manual Burst induction 334
Details about Manual Burst induction 334
Inducing an arrhythmia with PES 335
Parameters 335
Considerations 336
How to deliver a PES induction 337
Details about PES induction 337
Delivering a manual therapy 338
Parameters for manual defibrillation and cardioversion 338
Parameters for manual ATP therapies 338
Considerations 340
How to deliver a manual therapy340
Details about manual therapies 340
Key terms 341
15Solving system problems 343
Overview 344
Solving sensing problems 345
Solving tachyarrhythmia detection problems 347
Solving tachyarrhythmia therapy problems 348
Solving bradycardia pacing problems 349
Responding to device status indicators 351
Key terms 352
Table of contents
xiii
Part VAppendices
AWarnings and precautions 357
General warnings 358
Storage and handling 358
Resterilization 359
Device implantation and ICD programming 359
Lead evaluation and lead connection 361
Follow-up testing 362
Explant and disposal 363
Medical therapy hazards 363
Marquis DR 7274 Reference Manual
xiv
Table of contents
Home and occupational environments 365
BDevice Parameters367
Emergency settings 368
Detection parameters 369
Therapy parameters 371
Bradycardia pacing parameters 373
System maintenance parameters 376
Data collection parameters 377
System test and EP study parameters 378
Fixed parameters 381
Patient information parameters 383
Programmer symbols 384
Parameter Interactions Window 385
Parameter interlocks 386
Index 387
Marquis DR 7274 Reference Manual
Introduction
Using this manual
Before implanting the ICD, it is strongly recommended that you:
■
Refer to the product literature packaged with the ICD for
information about prescribing the ICD.
■
Thoroughly read this manual and the technical manuals for the
leads used with the device.
■
Discuss the procedure and the ICD system with the patient
and any other interested parties, and provide them with any
patient information packaged with the ICD.
Contacting technical support
Medtronic employs highly trained representatives and engineers
located throughout the world to serve you and, upon request, to
provide training to qualified hospital personnel in the use of
Medtronic products.
Introduction
xv
In addition, Medtronic maintains a professional staff of consultants
to provide technical consultation to product users. For medical
consultation, Medtronic can often refer product users to outside
medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative,
or call or write Medtronic at the appropriate address or telephone
number listed on the back cover.
Customer education
Medtronic invites physicians to attend an education seminar on
the complete ICD system. The course includes indications for use,
an overview of ICD system functions, implant procedures, and
patient management.
Marquis DR 7274 Reference Manual
xvi
Introduction
References
Notice
The primary reference for background information is Zacouto FI,
Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B,
ed. Cardiac Pacing Diagnostic and Therapeutic Tools. New York:
Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
Singer I, Barold SS, Camm AJ, Eds. Nonpharmacological
Therapy of Arrhythmias for the 21st Century: The State of the
Art. Armonk, NY: Futura Publishing Co. 1998.
■
Estes M, Manolis AS, Wang P, Eds. Implantable
Cardioverter-Defibrillators. New York, NY: Marcel Dekker, Inc.
This software is provided as an informational tool for the end user.
The user is responsible for accurate input of patient information
into the software. Medtronic makes no representation as to the
accuracy or completeness of the data input into the software.
MEDTRONIC SHALL NOT BE LIABLE FOR ANY DIRECT,
INDIRECT, INCIDENTIAL OR CONSEQUENTIAL DAMAGES TO
ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE
INFORMATION PROVIDED IN THE SOFTWARE.
Marquis DR 7274 Reference Manual
Abbreviations and acronyms
Atrial
A-
Atrial Fibrillation
AF
Abbreviations and acronyms
xvii
AFib/AFlutter
Atrial Refractory Period
ARP
Antitachycardia Pacing
ATP
Atrial Vulnerable Period
AVP
Beginning of Life
BOL
beats per minute
bpm
Combined (VT and VF) Number of Intervals to Detect
CNID
Cardioversion
CV
DF/Defib
Electrocardiogram
ECG
Electrogram
EGM
End of Life
EOL
Elective Replacement Indicator
ERI
Fibrillation Detection Interval
FDI
Fast Ventricular Tachycardia Detection Interval
FTI
Fast Ventricular Tachycardia
FVT
Atrial Fibrillation and/or Atrial Flutter
Defibrillation
Implantable Cardioverter Defibrillator
ICD
joules
J
-1
reciprocal minutes; for example, pacing pulses per minute
min
milliseconds
ms
millivolts
mV
NCAP
NID
NST
Non-Competitive Atrial Pacing
Number of Intervals to Detect
Non-sustained Tachycardia
Marquis DR 7274 Reference Manual
xviii
Abbreviations and acronyms
Premature Atrial Contraction
PAC
Paced A-V Delay
PAV
Programmed Electrical Stimulation
PES
Pacemaker-Mediated Tachycardia
PMT
an atrial interval
P-P
paces or pulses per minute
ppm
an interval between a P-wave and the subsequent R-wave
P-R
Post Ventricular Atrial Blanking period
PVAB
PVARP
PVC
RAAV
RNID
R-P
R-R
SAV
ST/Sinus Tach
SVT
TDI
V
V-
VF
VF NID
VRS
VSP
Post Ventricular Atrial Refractory Period
Premature Ventricular Contraction
Rate Adaptive A-V delay
Number of Intervals to Redetect
an interval between an R-wave and the subsequent P-wave
a ventricular interval
Sensed A-V Delay
Sinus Tachycardia
Supraventricular Tachycardia
Tachycardia Detection Interval
volts
Ventricular
Ventricular Fibrillation
VF Number of Intervals to Detect
Ventricular Rate Stabilization
Ventricular Safety Pacing
Ventricular Tachycardia
VT
VT NID
Marquis DR 7274 Reference Manual
VT Number of Intervals to Detect
Quick overview
Part I
Marquis DR 7274 Reference Manual
Physical characteristics 4
Magnet application 5
Longevity projections 6
Replacement indicators 8
Typical charge times 10
High voltage therapy energy 10
Stored data and diagnostics 11
New and enhanced features 14
Quick reference
1
1
Marquis DR 7274 Reference Manual
4
Chapter 1
Physical characteristics
Physical characteristics
Table 1-1. ICD physical characteristicsa
Volu m e 36 cc
Mass 75 g
b
H x W x D
Surface area of device can 66 cm
Radiopaque IDc
Materials in contact with
human tissue
d
68.3 mm x 50.8 mm x 13.9 mm
2
PKC
Titanium / polyurethane / silicone rubber
BatteryLithium silver vanadium oxide
Connectors Two IS-1 connectors for pacing and
sensing, Two DF-1 connectors for high
voltage therapy, Active Can electrode
(programmable)
Device Port Connector
Typ e
Software
Name
SVCDF-1HVX
A
V
SVC
RV
RVDF-1HVB
Cann/aHVA, Can
VIS-1 bipolar
AIS-1 bipolar
Can
74lead.eps
Suture holes
74Suture.eps
a
Measurements are nominal values based on CAD (computer aided design)
model measurements and are rounded to the nearest unit.
b
Grommets may protrude slightly beyond the can surface.
c
Engineering series number follows the radiopaque code.
d
These materials have been successfully tested for the ability to avoid biological
incompatibility. The device does not produce an injurious temperature in the
surrounding tissue.
Marquis DR 7274 Reference Manual
Magnet application
Bringing a magnet close to the device triggers changes in device
operation as shown in Table 1-2. When the magnet is removed,
the device returns to its programmed operations.
Table 1-2. Effects of magnet application on the device
Pacing modeas programmed
Pacing rate and intervalas programmed
VF, VT, and FVT detection suspended
Quick reference
5
Magnet application
a
b
Patient Alert audible tones
(20 seconds or less)
with programmable alert(s) enabled:
■
continuous tone (Test)
■
on/off intermittent tone (seek follow-up)
■
high/low dual tone (urgent follow-up)
c
with programmable alerts disabled:
■
no tone
■
high/low dual tone (urgent follow-up)
a
Rate response adjustments are suspended while a Patient Alert tone sounds.
b
Detection resumes if telemetry is established and the application software is
running, or it resumes after the application software has started.
c
The Test tone does not sound if “VF Detection/Therapy Off” is the only alert
enabled.
Marquis DR 7274 Reference Manual
6
Chapter 1
Longevity projections
Longevity projections
Longevity estimates are based on accelerated battery discharge
data and device modeling at 60 ppm (min
■
2.5 V pacing pulse amplitude, 0.4 ms pacing pulse width, and
-1
) pacing rate, with:
30 J delivered therapy energy (see Table 1-3)
■
3 V pacing pulse amplitude, 0.4 ms pacing pulse width, and
30 J delivered therapy energy (see Table 1-4)
This model assumes default automatic capacitor formation
setting. As a guideline, each full energy charge decreases device
longevity by approximately 24 days.
Device longevity is affected by how certain features are
programmed, such as EGM pre-storage. For more information,
see “Optimizing device longevity” on page 207.
Considerations for using EGM pre-storage
When the EGM pre-storage feature is programmed off, the device
starts to store EGM following the third tachyarrhythmia event and
also provides up to 20 seconds of information before the onset of
the tachyarrhythmia, including:
■
AA and VV intervals
■
Marker Channel
■
interval plot Flashback
When the EGM pre-storage feature is programmed on, the device
also collects up to 20 seconds of EGM information before the
onset of the arrhythmia.
In a patient who uniformly repeats the same onset mechanisms,
the greatest clinical benefit of pre-onset EGM storage is achieved
after a few episodes are captured.To maximize the effectiveness
of the EGM pre-storage feature and optimize device longevity,
consider these programming options:
■
Turn pre-storage on to capture possible changes in the onset
mechanism following significant clinical adjustments, for
example, device implant, medication changes, and surgical
procedures.
■
Turn pre-storage off once you have successfully captured the
information of interest.
Marquis DR 7274 Reference Manual
Quick reference
Longevity projections
Table 1-3. Projected longevity in years with 2.5V pacing amplitude and
0.4 ms pulse width
Maximum
energy
Percent
pacing
charging
frequency
EGM
a
pre-storage
0%Semi-Annual Off
On
QuarterlyOff
On
15%Semi-Annual Off
On
QuarterlyOff
On
50%Semi-Annual Off
On
QuarterlyOff
On
100%Semi-Annual Off
On
QuarterlyOff
On
a
Maximum energy charging frequency may include full energy therapy shocks or
capacitor formations.
b
The data provided for programming EGM pre-storage on is based on a 6 month
period (two 3-month follow-up intervals) over the life of the device. Additional use
of EGM pre-storage reduces longevity by approximately 25% or 3 months per
year.
500 ohm
pacing
impedance
b
DDDVVIDDDVVI
8.68.68.68.6
8.58.58.58.5
7.57.57.57.5
7.37.37.37.3
8.38.58.58.6
8.18.48.38.5
7.27.47.37.5
7.07.37.17.3
7.58.28.08.5
7.48.07.98.2
6.67.17.07.3
6.57.06.87.1
6.77.77.58.1
6.57.57.38.0
6.06.86.67.1
5.86.56.46.9
900 ohm
pacing
impedance
7
Marquis DR 7274 Reference Manual
8
Chapter 1
Replacement indicators
Table 1-4. Projected longevity in years with 3V pacing amplitude and
0.4 ms pulse width
500 ohm
pacing
impedance
b
DDDVVIDDDVVI
Percent
pacing
Maximum
energy
charging
frequency
EGM
a
pre-storage
0%Semi-Annual Off8.68.68.68.6
On 8.58.58.58.5
QuarterlyOff7.57.57.57.5
On7.37.37.37.3
15%Semi-Annual Off8.18.58.48.6
On8.08.38.28.4
QuarterlyOff7.17.37.37.5
On6.97.27.17.3
50%Semi-Annual Off7.28.07.88.3
On7.07.87.68.1
QuarterlyOff6.47.06.87.2
On6.26.86.77.0
100%Semi-Annual Off6.27.37.18.0
On6.07.17.07.7
QuarterlyOff5.56.56.36.9
On5.46.36.16.8
a
Maximum energy charging frequency may include full energy therapy shocks or
capacitor formations.
b
The data provided for programming ECG pre-storage on is based on a 6 month
period (two 3-month follow-up intervals) over the life of the device. Additional use
of EGM pre-storage reduces longevity by approximately 25% or 3 months per
year.
900 ohm
pacing
impedance
Replacement indicators
Battery voltage and messages about replacement status appear
on the programmer display and on printed reports. Table 1-5 lists
the Elective Replacement Indicator (ERI) and the End of Life
(EOL) conditions.
Marquis DR 7274 Reference Manual
Quick reference
Replacement indicators
Table 1-5. Replacement indicators
Elective Replacement (ERI)≤ 2.62 V
End of Life (EOL)3 months after ERI
ERI date – The programmer displays the date when the battery
reached ERI on the Quick Look and Battery and Lead
Measurements screens.
Temporary voltage decrease – The battery voltage temporarily
decreases following a high voltage charge. If a battery
measurement is taken immediately after a high voltage charge,
the ERI or EOL indicator may be displayed. However, this is a
temporary status which will return to normal when the battery has
recovered from the charge.
EOL indication – If the programmer indicates that the device is at
EOL, replace the device immediately.
Post-ERI conditions – EOL device status is defined as three
months following an ERI indication assuming the following
post-ERI conditions: 100% DDD pacing at 60 ppm (min
-1
), 3 V,
0.4 ms; 500 Ω pacing load; and six 30 J charges. EOL may be
indicated before the end of three months if the device exceeds
these conditions.
9
Marquis DR 7274 Reference Manual
10
Chapter 1
Typical charge times
Typical charge times
The most recent capacitor charge time appears on the
programmer display and on printed reports and can be evaluated
using the Charge/Dump test (see Table 1-6).
Ta bl e 1 -6 . Ty p ic ala full energy charge times
At Beginning of Life (BOL)5.9 seconds
At Elective Replacement (ERI)7.5 seconds
a
These charge times are typical when the capacitors are fully formed.
High voltage therapy energy
The stored energy of the device is derived from the peak capacitor
voltage and is always greater than the energy delivered by the
device. Table 1-7 compares the programmed energy levels
delivered by the device to the energy levels stored in the
capacitors before delivery.
Marquis DR 7274 Reference Manual
Quick reference
Stored data and diagnostics
Table 1-7. Comparing delivereda (programmed) and storedb energy levels
Energy (J)
a
Delivered
/
ProgrammedStored
b
Charge
c
Time
(sec)
Delivered a/
ProgrammedStored
30355.989.41.6
28335.678.31.4
26315.267.11.2
24284.755.91.0
22264.444.70.8
20244.033.60.6
18213.522.40.4
16193.21.82.20.4
15183.01.62.00.3
14162.71.41.70.3
13152.51.21.50.3
12142.41.01.20.2
11132.20.81.00.2
10122.00.60.80.1
9111.90.40.50.1
a
Energy delivered at connector block into a 75 ohm load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth of a
second.
EGM optionsStore before onset; Store during charging
2 minutes of dual-channel EGM, or 3.6 minutes
of single-channel EGM
Flashback memory 2000 intervals (containing both A-A and V-V):
before latest VF, before latest VT, and before
interrogation
Counter data
Detection countersLifetime total, since cleared, and since last
session
Episode countersEpisodes:
■
VF, FVT, and VT
■
Atrial Fibrillation / Atrial Flutter episodes
■
Sinus Tach episodes
■
Other 1:1 SVT episodes
■
NST episodes
■
Mode switch episodes
Percentage pacing:
■
AS-VS, AS-VP, AP-VS, AP-VP percentages
Additional counters:
■
Single PVCs and PVC runs
■
Rate stabilization pulses and runs
Therapy efficacy
counters
Counts for each VF, FVT, VT Therapy:
■
Delivered
■
Successful
■
Unsuccessful
■
Intervention (manually aborted)
Total number of aborted shocks
Other stored data
Patient Alert eventsUp to 10 log entries: text and date for the first
Battery and lead
measurements
Marquis DR 7274 Reference Manual
time an alert is triggered between interrogations
Battery voltage, last capacitor formation, last
charge, lead impedance, EGM amplitude
measurements, last high voltage therapy, and
sensing integrity counter
Stored data and diagnostics
Table 1-8. Stored data and diagnostics (continued)
Lead performance
trends
Cardiac Compass
trends
14 days of daily measurements plus 80 weeks
of weekly minimum and maximum
measurements:
■
Lead impedance: atrial pacing, ventricular
pacing, defibrillation pathway, and SVC
lead (if used)
The following features are new or changed from the
7275 GEM III DR ICD.
Patient management
RapidRead telemetry – Communication between the device and
programmer is approximately 20 times faster than telemetry in
previous Medtronic ICD devices. The magnitude of improvement
depends on the amount and type of data that is interrogated.
RapidRead telemetry is more reliable and has an increased range
that makes placing the programming head easier.
Cardiac Compass trends report – This report displays up to
14 months of trend data related to tachyarrhythmia episodes,
heart rate, and patient activity. See “Using Cardiac Compass to
view long term clinical trends” on page 294.
Patient Alert – The alert duration when a magnet is applied to the
device is now 20 seconds. The device also provides several new
alerts:
■
SVC (HVX) lead impedance out of range
■
Active Can off without SVC
■
DOO/VOO mode programmed
■
VF Detection programmed off, or fewer than four VF therapies
enabled for at least six hours
■
charge circuit timeout occurred
■
excessive charge time ERI
For more information, see “Using the Patient Alert feature” on
page 253.
EGM amplitude trends – The device automatically measures
R-wave and P-wave EGM amplitudes every day. These daily
measurements are included in the data displayed on the Lead
Performance Trends screen. See “Setting up data collection” on
page 267.
Marquis DR 7274 Reference Manual
Quick reference
New and enhanced features
EGM amplitude test – You can use the EGM Amplitude test to
measure R-wave and P-wave EGM amplitudes. The results are
reported on the EGM Amplitude Test screen. See “Measuring
EGM Amplitude” on page 318.
Lead impedance measurement for SVC (HVX) – Along with
other lead impedance measurements, the device provides an
independent SVC (HVX) measurement to check the integrity of
the supplementary high voltage electrode. See “Measuring lead
impedance” on page 316.
Leadless ECG signal – If a supplementary high voltage
electrode is placed in the SVC, the device provides the
Leadless ECG signal through either the Can to SVC (HVX) or RV
(HVB) to SVC (HVX) EGM source. See “Setting up data collection”
on page 267.
Expanded pre-onset EGM storage – The device can now store
up to 20 seconds of EGM before a tachycardia starts. See “Setting
up data collection” on page 267.
15
Smart Auto Cap Formation – When the Auto Cap Formation
Interval is set to Auto, the formation interval automatically adjusts
to optimize device longevity and charge times. See “Smart Auto
Cap” on page 206.
Ending a patient session – The device audits the programmed
parameter settings when you end a patient session and alerts you
if any of the settings are atypical. See “Starting and ending patient
sessions” on page 230.
Tachyarrhythmia detection
VT Monitoring – VT detection can be set to Monitor, which allows
the device to detect and record VT episodes without delivering
therapy or influencing VF detection. See “Monitoring episodes for
termination or redetection” on page 84.
Marquis DR 7274 Reference Manual
16
Chapter 1
New and enhanced features
Tachyarrhythmia therapy
Bradycardia pacing
High Rate Timeout – High Rate Timeout can turn off detection
enhancements (Stability, PR Logic criteria) if a high rate episode
is longer than a programmed duration. See “Details about High
Rate Timeout” on page 102.
Episode confirmation during and after charging – The device
continually monitors the ventricular rhythm during and after
charging for cardioversion or defibrillation (when VF confirmation
is active) to ensure the arrhythmia is present before delivering the
high voltage shock. See “Confirming VF after initial detection” on
page 115 and “Confirming VT or FVT after detection” on
page 134.
Programmable Active Can – If a supplementary electrode is
connected to the SVC (HVX) port, you can deselect the device
Can as a high voltage electrode. See “Delivery pathway
electrodes” on page 113.
Accelerometer-based rate response – The device uses an
accelerometer to provide rate responsive pacing.
Additional bradycardia pacing modes – The device provides
asynchronous pacing in the DOO and VOO pacing modes and
provides the ODO mode to disable pacing. See Chapter 8,
"Treating bradycardia" on page 147.
Enhanced AT-style Mode Switch – Mode Switch episodes are
detected using a combination of the median atrial rate and the AF
evidence criterion. See “Details about Mode Switch” on page 182.
Marquis DR 7274 Reference Manual
EP studies
Quick reference
New and enhanced features
Defibrillation threshold testing support – The T-Shock and
50 Hz Burst induction screens allow you to monitor time between
inductions, program ventricular sensing and VF therapy settings,
adjust induction settings, select manual therapies, and retrieve
episode records after therapy. See “How to perform defibrillation
threshold testing” on page 45.
Backup VVI pacing during atrial inductions – You can choose
to have the device deliver backup ventricular pacing during Manual
Burst and PES inductions that are delivered to the atrium. See “EP
Study overview” on page 324.
17
Marquis DR 7274 Reference Manual
18
Chapter 1
New and enhanced features
Marquis DR 7274 Reference Manual
System overview 20
Indications and usage 23
Contraindications 23
Patient screening 23
The Marquis DR system
2
2
Marquis DR 7274 Reference Manual
20
Chapter 2
System overview
System overview
The 7274 Marquis DR Dual Chamber Implantable Cardioverter
Defibrillator (ICD) system is an implantable medical device system
that automatically detects and treats episodes of ventricular
fibrillation, ventricular tachycardia, fast ventricular tachycardia,
and bradyarrhythmia. The ICD system includes three major
components:
■
ICD
The ICD senses the electrical activity of the patient’s heart via
the sensing electrodes of the implanted atrial and ventricular
leads. It then analyzes the heart rhythm based on selectable
sensing and detection parameters. If the ICD detects a
tachyarrhythmia, it delivers defibrillation, cardioversion, or
antitachycardia pacing therapy to the patient’s heart. If the ICD
identifies a bradyarrhythmia, it delivers bradycardia pacing
therapy to the patient’s heart.
■
Leads
The ICD can be used with transvenous or epicardial
defibrillation leads. The lead system should consist of bipolar
or paired unipolar
1
pacing/sensing leads in each chamber of
the heart and one or two high voltage cardioversion/
defibrillation electrodes. You can program the Active Can
device case as a high voltage electrode. The pacing and
sensing electrodes in each chamber sense cardiac activity
and deliver pacing stimuli.
■
Programmer and software
The Medtronic 9790C
2
programmer and 9966 application
software allow you to
■
configure the detection, therapy, and bradycardia features
for your patient
■
perform electrophysiological studies and system tests
■
monitor, display, or print patient cardiac activity information
1
With an appropriate unipolar to bipolar adapter kit.
2
With the model 9767 or 9767L programming head
Marquis DR 7274 Reference Manual
For information about:
■
indications, contraindications, lead compatibility, warnings and
precautions, and patient selection, see the Marquis DR 7274 Implant Manual, which accompanies each device.
■
basic programmer and software desktop functions that are not
included in Chapter 10, "Using the programmer" on page 211,
see the manual accompanying the programmer.
■
installing the 9767 or 9767L programming head, see the
manual accompanying the programming head.
■
implanting leads, refer to the manuals accompanying the
leads.
Detecting and treating tachyarrhythmias
The ICD monitors the cardiac rhythm for short ventricular intervals
that may indicate the presence of VF, VT, or FVT.
■
Upon detection of VF, the ICD delivers a biphasic defibrillation
shock of up to 30 joules. If the VF episode persists, up to five
more individually programmed defibrillation shocks can be
delivered.
■
Upon detection of VT, the ICD delivers either a Ramp, Ramp+,
or Burst antitachycardia pacing therapy or a biphasic
cardioversion shock of up to 30 joules synchronized to a
ventricular depolarization. If the VT episode persists, up to five
more individually programmed VT therapies can be delivered.
You can also program the ICD to monitor the VT episode
without delivering therapy.
■
Upon detection of FVT, the ICD delivers either a Ramp,
Ramp+, or Burst antitachycardia pacing therapy or a biphasic
cardioversion shock of up to 30 joules synchronized to a
ventricular depolarization. If the FVT episode persists, up to
five more individually programmed FVT therapies can be
delivered.
The Marquis DR system
System overview
21
You can program the ICD to distinguish between true ventricular
arrhythmias and rapidly conducted supraventricular tachycardia
(SVT) and withhold therapy for SVT.
You can also program the ICD to detect a double tachycardia (an
unrelated ventricular arrhythmia occurring simultaneously with an
SVT), so that therapy is not withheld for a ventricular arrhythmia in
the presence of an SVT.
Marquis DR 7274 Reference Manual
22
Chapter 2
System overview
Treating bradycardia
The ICD provides dual chamber rate responsive bradycardia
pacing to optimize hemodynamics. An internal accelerometer
senses the patient’s physical activity, allowing the ICD to increase
and decrease the pacing rate in response to changes in the level
of activity.
Monitoring for real-time and stored data
The ICD and programmer provide real-time information on
detection and therapy parameters and status during a patient
session. The ICD also provides accumulated data on device
operation, including stored electrograms, detected and treated
tachyarrhythmia episodes, bradycardia interventions, and the
efficacy of therapy. The Cardiac Compass report provides up to 14
months of clinically significant data, including arrhythmia
episodes, therapies delivered, physical activity, heart rate, and
bradycardia pacing activities.
All of this information can be printed and retained in the patient’s
file or saved in electronic format on a floppy diskette.
Conducting electrophysiologic tests
You can use the system to conduct non-invasive
electrophysiologic studies including manual delivery of any of the
ICD therapies to manage an induced or spontaneous
tachyarrhythmia.
Alerting the patient to system events
You can use the programmable Patient Alert monitoring feature to
notify the patient with audible tones if certain conditions related to
the leads, battery, charge time, and therapies occur. The patient
can then respond based on your prescribed instructions.
Marquis DR 7274 Reference Manual
Indications and usage
The implantable cardioverter defibrillator is intended to provide
ventricular antitachycardia pacing and ventricular defibrillation for
automated treatment of life threatening ventricular arrhythmias.
Contraindications
The Marquis DR systemis contraindicated for
■
patients whose tachyarrhythmias may have transient or
reversible causes, such as acute myocardial infarction,
digitalis intoxication, drowning, electrocution, electrolyte
imbalance, hypoxia, or sepsis.
■
patients with incessant VT or VF
■
patient who have a unipolar pacemaker
■
patients whose primary disorder is bradyarrhythmias or atrial
arrhythmias
The Marquis DR system
Indications and usage
23
Patient screening
Other optional screening procedures could include exercise stress
testing to determine the patient’s maximum sinus rate, and cardiac
catheterization to determine if there is a need for concomitant
surgery and/or medical therapy.
Marquis DR 7274 Reference Manual
24
Chapter 2
Patient screening
Marquis DR 7274 Reference Manual
Emergency therapy
Delivering emergency therapies 26
How to deliver emergency 30 joule defibrillation 28
How to deliver emergency cardioversion 28
How to deliver emergency fixed burst pacing 29
How to deliver emergency VVI pacing 30
3
3
Marquis DR 7274 Reference Manual
26
Chapter 3
Delivering emergency therapies
Delivering emergency therapies
The device provides the following emergency therapies:
■
defibrillation
■
cardioversion
■
fixed burst pacing
■
emergency VVI pacing
The default emergency therapy is 30 joule defibrillation. When you
select [Emergency] and [DELIVER], the device charges and
delivers a biphasic 30 joule shock along the AX>B pathway
The programmer resets the emergency defibrillation energy to
30 joules each time you select [Emergency]. Emergency
cardioversion and fixed burst values remain as selected for the
duration of the session.
To return to other programming functions from an Emergency
screen, select [Exit Emergency].
Effect on system operation
1
.
The device suspends the automatic detection features when
emergency defibrillation, cardioversion, or fixed burst pacing
therapies are delivered. Detection is not suspended during
emergency VVI pacing. Removing the programming head or
pressing [Resume] turns detection on again.
Aborting an emergency therapy
As a safety precaution, the programmer also displays an [ABORT]
button which immediately terminates any emergency therapy in
progress.
1
If Active Can is turned off, the defibrillation is delivered between the HVX and
HVB electrodes.
Marquis DR 7274 Reference Manual
On-screen and display panel buttons
The on-screen [Emergency] button and the red mechanical
Emergency button by the programmer display panel function the
same at all times.
Red Emergency
Button
The on-screen [DELIVER] button and the yellow-on-blue
mechanical Deliver button by the programmer display panel
function the same during emergency operations only. The
mechanical Deliver button operates only during emergency
operations.
Emergency therapy
Delivering emergency therapies
Functions the same as
on-screen [Emergency] button.
27
Yellow-on-blue
Deliver Button
Temporary parameter values
Emergency tachyarrhythmia therapies use temporary values that
do not change the programmed parameters of the device.
values are not in effect until you select [DELIVER]. After the
tachyarrhythmia therapy is complete, the device reverts to its
programmed values.
Functions the same as on-screen
[DELIVER], but only during
Emergency functions.
1
These
1
Delivery of Emergency VVI Pacing changes the programmed bradycardia
pacing values to the emergency values (see page 30).
Marquis DR 7274 Reference Manual
28
Chapter 3
Delivering emergency therapies
How to deliver emergency 30 joule defibrillation
3
4
2
How to deliver emergency cardioversion
3
4
5
2
1. Position the programming
head over the device.
2. Select [Emergency].
3. Accept the defibrillation
energy shown on the screen,
or select Energy and select a
new value from the window.
4. Select [DELIVER].
If delivery is not confirmed,
verify that the programming
head is properly positioned
and select [Retry] or [Cancel].
1. Position the programming
head over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the cardioversion
energy shown on the screen,
or select Energy and select a
new value from the window.
5. Select [DELIVER].
If delivery is not confirmed,
verify that the programming
head is properly positioned
and select [Retry] or [Cancel].
Marquis DR 7274 Reference Manual
How to deliver emergency fixed burst pacing
3
4
5
2
Emergency therapy
Delivering emergency therapies
1. Position the programming
head over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval
shown on the screen, or
select Interval for a new
interval value.
5. Select [BURST Press and
Hold].
If delivery is not confirmed,
the programmer displays
an error window. Verify that
the programming head is
properly positioned. Select
[OK] from the window and
reselect [BURST Press and
Hold].
29
Marquis DR 7274 Reference Manual
30
Chapter 3
Delivering emergency therapies
How to deliver emergency VVI pacing
4
2
1. Position the programming
head over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A
successful programming
3
sets the device to the
following maximum output
bradycardia pacing values.
■
Pacing Mode: VVI
■
Lower Rate: 70 ppm
(70 min-1)
■
V. Amplitude: 6 V
■
V. Width: 1.6 ms
■
V. Pace Blanking: 240 ms
■
Hysteresis: Off
■
Ventricular Rate
Stabilization: Off
If programming is not
confirmed, verify that the
programming head is properly
positioned and select [Retry]
or [Cancel].
Marquis DR 7274 Reference Manual
Part II
Device implant and patient follow-up
procedures
Marquis DR 7274 Reference Manual
Implanting the ICD
Overview 34
Preparing for an implant 34
Replacing an ICD 36
Positioning the leads 37
Testing sensing and pacing thresholds 39
Connecting the leads to the ICD 40
Testing defibrillation operation and effectiveness 42
Positioning and securing the ICD 46
Completing the implant procedure 47
4
4
Marquis DR 7274 Reference Manual
34
Chapter 4
Overview
Overview
Preparing for an implant
The tasks for implanting an ICD include
1. Preparing for an implant
2. Replacing an ICD
3. Positioning the leads
4. Testing sensing and pacing thresholds
5. Connecting the leads to the ICD
6. Testing defibrillation operation and effectiveness
7. Positioning and securing the ICD
8. Completing the implant procedure
These tasks are described in the sections that follow.
Warning: Keep a back-up external defibrillator available
during the implant for transthoracic rescue when arrhythmias
are induced.
Equipment for an implant
■
Model 9790c programmer and Model 9767 or 9767L
programming head
■
9966 software application
■
8090 Analyzer lead analysis device or equivalent pacing
system analyzer
■
external defibrillator
■
5358 Defibrillator Implant Support Device and software
application (optional)
Marquis DR 7274 Reference Manual
Sterile supplies
■
Marquis DR ICD and lead system components
■
Programming head sleeve or programming head
■
Analyzer cables
■
Lead introducers appropriate for the lead system
■
Extra stylets of appropriate length and shape
How to prepare for implanting
Set up the implant support instrument
When using an implant support instrument such as the 5358
Defibrillator Implant Support Device:
1. Calibrate any monitoring or recording equipment while recording
the EGM and marker outputs of the support instrument.
2. Verify the high energy output of the support instrument by delivering
a high energy defibrillation shock into the test load.
Set up the programmer and start the application
1. Set up the programmer as described in the instructions provided
with the programmer.
2. Install the Marquis DR Model 9966 software on the programmer, if it
is not already installed.
3. Place the programming head over the device and start the
application. Select the device model or select [Auto identify].
Note: The programmer automatically interrogates the device when
the application starts.
Implanting the ICD
Preparing for an implant
35
Marquis DR 7274 Reference Manual
36
Chapter 4
Replacing an ICD
Preprogram the device
Before opening the sterile package, prepare the ICD for implant as
follows:
1. Check the “use by” date printed on the package. Do not implant the
device after the “use by” date because the battery’s longevity could
be reduced.
2. Interrogate the ICD, and print a full summary report.
3. Confirm that the battery voltage is at least 3.0 V at room
temperature.
If the device has been exposed to lower temperatures or has
delivered a recent high voltage charge, the battery voltage will be
temporarily lower.
4. Set up data collection parameters and the ICD internal clock (see
page 269).
5. Perform a manual capacitor formation (see page 320).
6. Program the therapy and pacing parameters to values appropriate
for the patient (see page 151). Ensure that all tachyarrhythmia
detection is programmed Off (see page 58).
a
Use the Quick Look screen to verify the voltage, see page 251.
a
Replacing an ICD
If you are replacing a previously implanted ICD, turn off ICD
detection and therapies before explanting.
When implanting the ICD with a chronic lead system, perform the
following evaluations to ensure appropriate detection and therapy:
■
Check the integrity of the chronic high voltage leads with a test
shock, chest X-ray, and inspection.
■
Perform chronic pacing and sensing measurements.
■
Measure high voltage lead impedances.
■
Test defibrillation efficacy.
■
Confirm adequate sensing during VF.
■
Ensure proper fit of the lead connectors in the ICD connector
block.
Marquis DR 7274 Reference Manual
Notes:
■
To meet the implant requirements, it may be necessary to
reposition or replace the chronic leads or to add a third high
voltage electrode.
■
Any unused leads that remain implanted must be capped.
How to explant and replace an ICD
1. Program all tachyarrhythmia detection Off.
2. Dissect the leads and the ICD free from the surrounding tissues in
the surgical pocket. Be careful not to nick or breach the lead
insulation during the process of exposing the system.
3. Loosen each setscrew, and gently retract the lead from the
connector block.
4. Remove the ICD from the surgical pocket.
5. If the connector pin of any implanted lead shows signs of pitting or
corrosion, replace the implanted lead with a new lead. The
damaged lead should be discarded and replaced to assure the
integrity of the device system.
6. Measure sensing, pacing, and defibrillation efficacy using the
replacement ICD or an implant support instrument.
7. Evaluate the defibrillation efficacy of the replacement system.
Implanting the ICD
Positioning the leads
37
Positioning the leads
Implant endocardial leads according to the supplied instructions,
unless suitable chronic leads are already in place. Do not use any
lead with this device without first verifying connector compatibility
(refer to the Marquis DR 7274 Implant Manual). Transvenous or
epicardial leads may be used. A bipolar atrial lead with closely
spaced pacing and sensing electrodes is recommended.
Using transvenous leads
Use standard transvenous implant techniques to position the
ventricular lead tip in the right ventricular apex and the atrial
pacing lead tip high on the right atrial appendage.
Follow the general guidelines below for initial positioning of other
transvenous leads (the final positions are determined by
defibrillation efficacy tests):
Marquis DR 7274 Reference Manual
38
Chapter 4
Positioning the leads
Using epicardial leads
■
SVC (HVX) lead: Place the lead tip high in the innominate
vein, approximately 5 cm proximal to the right atrium (RA) and
SVC junction.
■
SQ patch: Place the patch along the left mid-axillary, centered
over the fourth-to-fifth intercostal space.
■
CS lead: Advance the lead tip to just under the left atrial
appendage, if possible.
If using a subclavian approach, position the lead laterally to avoid
pinching the lead body between the clavicle and the first rib.
Warning: Pinching the lead can damage the lead conductor
or insulation, which may cause unwanted high voltage
therapies or result in the loss of sensing or pacing therapy.
A variety of surgical approaches can be used to implant epicardial
leads, including a limited left thoracotomy or median sternotomy.
A typical placement may use an anterior right ventricular patch as
the RV (HVB) and a posterolateral left ventricular patch as SVC
(HVX).
Follow the general guidelines below for positioning epicardial
leads:
■
If unipolar epicardial pacing leads are used, position the
electrodes about 1 to 2 cm apart to reduce electromagnetic
interference, and route the leads together with several loose
twists.
■
Suture the smooth face of each patch lead against the
epicardium or pericardium in locations that produce optimal
defibrillation.
■
Place the patches so that they encompass the maximum
amount of cardiac mass and they have approximately equal
amounts of mass between them.
■
Ensure that the patches do not overlap and the electrode
portions do not touch.
■
Avoid placing extra-pericardial patches over the phrenic nerve.
Marquis DR 7274 Reference Manual
Testing sensing and pacing thresholds
Surgical incisions
A single-incision submuscular or subcutaneous approach is
recommended when the ICD is implanted in the pectoral region.
Make the implant pocket about 1.5 times the size of the ICD.
Submuscular implant – An incision extending over the
deltoid-pectoral groove typically provides access to the cephalic
and subclavian veins as well as the implant pocket. Place the ICD
sufficiently medial to the humeral head to avoid interference with
shoulder motion.
Subcutaneous implant – A transverse incision typically permits
isolation of the cephalic vein. Place the ICD far medially to keep
the leads away from the axilla. Make sure that the upper edge of
the ICD remains inferior to the incision.
Testing sensing and pacing thresholds
Sensing and pacing tests include the following measurements:
■
EGM amplitude
■
slew rate
■
pacing threshold
■
pacing lead impedance
Implanting the ICD
39
Medtronic recommends that you use an 8090 Analyzer lead
analysis device to perform sensing and pacing measurements. If
you use a Pacing System Analyzer (PSA), perform both atrial and
ventricular measurements via the ventricular channel of the PSA.
Refer to the technical manual for the Analyzer you use to find
details on performing sensing and pacing measurements.
Marquis DR 7274 Reference Manual
40
Chapter 4
Connecting the leads to the ICD
Parameters
Considerations
Measured sensing and pacing values must meet the following
specific requirements at implant.
Table 4-1. Sensing and pacing values at implant
MeasurementAcute Transvenous Leads Chronic Leads
R- wave amplitude ≥ 5 mV≥ 3 mV
P- wave amplitude≥ 2 mV≥ 1 mV
Slew rate:
atrial≥ 0.5 V/s≥ 0.3 V/s
ventricular≥ 0.75 V/s≥ 0.5 V/s
a
Capture threshold
atrial≤ 1.5 V≤ 3.0 V
ventricular≤ 1.0 V≤ 3.0 V
a
At 0.5 ms pulse width
:
When measuring sensing and pacing values, measure between
the tip (cathode) and ring or coil (anode) of each bipolar
pacing/sensing lead.
For unipolar epicardial pacing leads, either electrode can be the
cathode; use the configuration that yields the lower pacing
threshold.
Note: Do not measure the intracardiac EGM telemetered from the
ICD to assess sensing.
Connecting the leads to the ICD
For more detailed information about lead/connector compatibility,
see the Marquis DR 7274 Implant Manual, or contact Medtronic
Technical Services at 1-800-723-4636.
Marquis DR 7274 Reference Manual
Implanting the ICD
Connecting the leads to the ICD
Table 4-2. Lead connections
Device
Port
Connector
Type
Software Name
SVCDF-1HVX
A
V
SVC
RV
RVDF-1HVB
Cann/aHVA, Can
VIS-1 bipolar
Can
74lead.eps
AIS-1 bipolar
Warning: Loose lead connections may result in inappropriate
sensing and failure to deliver necessary arrhythmia therapy.
Caution: Use only the torque wrench supplied with the
device. It is designed to prevent damage to the device from
overtightening a setscrew.
For easier lead insertion, insert the ventricular IS-1 leg before
the other legs.
41
How to connect the lead to the device
1
a
2
b
74SetScrew.eps
74LeadTIp.eps
1. Insert the torque wrench into the
appropriate setscrew.
a. If the port is obstructed, retract the
setscrew to clear it. Take care not to
disengage the setscrew from the
connector block.
b. Leave the torque wrench in the setscrew
until the lead is secure. This allows a
pathway for venting trapped air when the
lead is inserted.
2. Push the lead or plug into the connector
port until the lead pin is clearly visible in the
pin viewing area. No sealant is required, but
sterile water may be used as a lubricant.
3. Tighten the setscrew by turning clockwise
until the torque wrench clicks.
4. Tug gently on the lead to confirm a secure
fit. Do not pull on the lead until all setscrews
have been tightened.
5. Repeat these steps for each lead.
Marquis DR 7274 Reference Manual
42
Chapter 4
Testing defibrillation operation and effectiveness
Testing defibrillation operation and effectiveness
To demonstrate reliable defibrillation effectiveness with the
implanted lead system, use the ICD to complete one of the
following tests:
■
Terminate two consecutively induced VF episodes using a
delivered energy of 20 joules or less.
■
Use the binary search procedure to establish a defibrillation
threshold (DFT) of 18 joules or less.
Note: An external defibrillation implant support instrument can
also be used to test defibrillation effectiveness. For instructions,
see the applicable documentation for the specific support
instrument.
High voltage implant values
Measured values must meet the following requirements at implant.
Warning: Ensure that an external defibrillator is charged for a
rescue shock.
Marquis DR 7274 Reference Manual
Binary search protocol
By testing defibrillation efficacy at successively lower energy
levels, the binary search protocol provides an accurate threshold
measurement. However, it generally requires more inductions and
more time. Also, cumulative charging of the ICD capacitor affects
the device longevity.
For most reliable defibrillation efficacy testing, allow at least five
minutes between VF inductions. The ICD software provides an
on-screen timer to record elapsed time since the last induction.
If a two-electrode system fails to meet the implant criterion,
consider implanting a third electrode and connecting it to the SVC
(HVX) port. You can also evaluate the efficacy of ventricular
tachycardia and atrial arrhythmia therapies after successfully
completing ventricular defibrillation testing.
Figure 4-1. Binary search protocol
Implanting the ICD
Testing defibrillation operation and effectiveness
Start
43
1
First induced episode
Second induced episode
Third induced episode
Defibrillation
Threshold
SuccessFailure
SuccessFailureFailureSuccess
3 J
SuccessFailureSuccessFailureSuccessFailu re
<=3 J6 J9 J15 J18 J
1
The timer appears on the 50 Hz Burst and T-Shock induction screens.
6 J
9 J15 J
Implant
Criterion Met
12 J
12 J
Success
Reprogram polarity, reposition the
lead, or try a different lead
Tr y t o me et
2 at 20 J
Failure
18 J
74binary.eps
Marquis DR 7274 Reference Manual
44
Chapter 4
Testing defibrillation operation and effectiveness
How to prepare for defibrillation threshold testing
1. Place the programming head over the ICD, start a patient session,
and interrogate the device, if you have not already done so.
2. Observe the Marker Channel telemetry annotations and the
programmer ECG display to verify that the ICD is sensing properly.
3. Conduct a manual Lead Impedance Test
lead connections. Perform this test with the ICD in the surgical
pocket and keep the pocket very moist. If the impedance is out of
range, perform one or more of the following tasks:
■
Recheck lead connections and electrode placement.
■
Repeat the measurement.
■
Inspect the bipolar EGM for abnormalities.
■
Measure the defibrillation impedance with a manual test shock.
4. Program the ICD or support instrument to properly detect VF with
an adequate safety margin (1.2 mV sensitivity).
a
See “Measuring lead impedance” on page 316.
a
to verify the defibrillation
Marquis DR 7274 Reference Manual
Testing defibrillation operation and effectiveness
How to perform defibrillation threshold testing
3
2
1
4
6
5
7
8
Implanting the ICD
1. Select Tests > EP Study.
2. Select either 50 Hz BURST or
T-shock induction.
3. Select [Resume at BURST] or
[Resume at DELIVER].
4. Select [Adjust Permanent...].
5. Program VF Enable On.
6. Program the automatic therapy
energy settings. Therapies 2-6
should be set to the maximum
energy.
7. Select [Program].
8. Select [Close].
45
Marquis DR 7274 Reference Manual
46
Chapter 4
Positioning and securing the ICD
11
9
12
13
10
14
9. If performing a T-Shock
induction, select the [Enable]
checkbox.
10. Select [DELIVER], or [50 Hz
BURST Press and Hold].
If necessary, you can abort an
induction or therapy in progress
by pressing [ABORT].
11. Observe the live rhythm monitor
for proper post-shock sensing.
12. If using the binary search
protocol, use the [Adjust
Permanent...] button to program
the next appropriate energy level
(see Figure 4-1).
13. Wait until the on-screen timer
reaches 5 minutes, then repeat
steps 9 through 12 as needed.
14. Select Params > Detection and
program VF, FVT, and VT
detection Off before closing.
Positioning and securing the ICD
Cautions: If no SVC electrode is implanted, the pin plug
provided with the device must be secured in the SVC port.
Program tachyarrhythmia detection Off before closing.
How to position and secure the device
Suture Hole Locations
Marquis DR 7274 Reference Manual
1. Ensure that each lead pin or plug is fully
inserted into the connector block and that all
setscrews are tight.
2. Coil any excess lead length beneath the device.
Avoid kinks in the lead conductors.
3. Implant the device within 5 cm of the skin. This
position optimizes the ambulatory monitoring
operations.
4. Suture the device securely within the pocket to
74Suture.eps
minimize post-implant rotation and migration of
the device. Use a surgical needle to penetrate
the suture holes.
Completing the implant procedure
After implanting the device, X-ray the patient to verify the device
and leads placement. To complete programming the device, select
parameters that are appropriate for the patient.
How to complete programming the device
1. After closing the pocket, program detection On. Program ventricular
tachyarrhythmia therapies On as desired.
2. Do not enable the Other 1:1 SVTs PR Logic detection criterion until
the atrial lead has matured (approximately one month post implant).
3. If external equipment was used to conduct the defibrillation efficacy
tests, perform a final VF induction, and allow the implanted system to
detect and treat the arrhythmia.
4. Monitor the patient after the implant, and take X-rays as soon as
possible to document and assess the location of the leads.
5. Program patient information. See “How to view and enter new patient
information” on page 303.
6. Configure the Patient Alert feature. See “Using the Patient Alert
feature” on page 253.
7. Set up data collection parameters. See “Setting up data collection” on
page 267.
8. Interrogate the device after any spontaneous episodes to evaluate
the detection and therapy parameter settings.
9. If the patient has not experienced spontaneous episodes, you may
induce the clinical tachyarrhythmias using the non-invasive EP Study
features to further assess the performance of the system. See
Chapter 14, "Conducting Electrophysiologic Studies" on page 323.
10. Recheck pacing and sensing values, and adjust if necessary.
Implanting the ICD
Completing the implant procedure
47
Marquis DR 7274 Reference Manual
48
Chapter 4
Completing the implant procedure
Marquis DR 7274 Reference Manual
Conducting a patient follow-up
Patient follow-up guidelines 50
Verifying the status of the implanted system 50
Verifying accurate detection and appropriate therapy 51
Verifying effective bradycardia pacing 52
session
5
5
Marquis DR 7274 Reference Manual
50
Chapter 5
Patient follow-up guidelines
Patient follow-up guidelines
Schedule regular patient follow-up sessions to monitor the
condition of the ICD and leads and to verify that the ICD is
configured appropriately for your patient.
During the first few months after receiving a new device, the
patient may require close monitoring. Schedule an office visit at
least every three months.
The Quick Look screen, which is displayed after you interrogate
the device, provides a good beginning for the follow-up review.
Using this screen you can
■
verify that the device is functioning correctly.
■
review the clinical performance and long term trends.
■
print appropriate reports1 to compare the results to the
patient’s history and to retain for future reference.
Note: The Checklist feature provides a standard list of tasks to
perform at a complete follow-up visit. You can also customize your
own checklists if you wish. See “Streamlining follow-ups with
Checklist” on page 261 for more information.
Verifying the status of the implanted system
To verify that the ICD and leads are functioning correctly, review
the following information from the Quick Look screen and perform
follow-up tests as indicated:
■
Review the displayed battery voltage for comparison to the
Elective Replacement Indicator value (see page 8).
Remember that battery voltage may be low if high voltage
charging has occurred within 24 hours.
■
Review the last full energy charge.
– For information about adjusting the capacitor formation
interval, see “Optimizing charge time” on page 204.
– If the programmer displays an Excessive Charge Time ERI,
the ICD should be replaced immediately.
1
See “Using Cardiac Compass to view long term clinical trends” on page 294 for
information on this new report.
Marquis DR 7274 Reference Manual
Conducting a patient follow-up session
Verifying accurate detection and appropriate therapy
■
Review the defibrillation and pacing lead impedance values for
inappropriate values or large changes since the last follow-up.
See “Measuring lead impedance” on page 316.
■
Perform an EGM Amplitude test in each chamber for
comparison to previous EGM Amplitude measurements. See
“How to perform an EGM Amplitude test” on page 320.
■
To review longer term trends in sensing and impedance
measurements, select the [>>] button from the lead
impedance area of the Quick Look screen. The programmer
displays a detailed history of automatic sensing and
impedance measurements. See “Taking a quick look at device
activity” on page 251.
Verifying accurate detection and appropriate therapy
To verify that the ICD is providing effective tachyarrhythmia
detection and therapy, review the following information from the
Quick Look screen and investigate as indicated:
■
Review Quick Look Observations that relate to patient history
and device operation. To display more detailed information
about any observation, select the observation and then select
the [>>] button.
■
Review any Patient alerts listed in the Observations of the
Quick Look screen. For the most detailed information about
Patient Alerts, select Patient Alert from the Data icon and
select [Events].
■
Check stored episode records for appropriate sensing and
detection of arrhythmias. See “Viewing episode data” on
page 280.
■
Check stored SVT episode records for appropriate
identification of SVTs.
51
Considerations
Review the following information before verifying detection and
therapy.
Flashback memory – In addition to the episode text and stored
electrograms, use Flashback memory and interval plots to help
investigate the accuracy and specificity of ventricular detection.
Marquis DR 7274 Reference Manual
52
Chapter 5
Verifying effective bradycardia pacing
Episode misidentification – If the episode records indicate that
false detections have occurred, the Sensing Integrity counter may
help in determining the prevalence of oversensing. For more
information, see “Sensing integrity counter” on page 275.
If the ICD is oversensing, consider these programming options:
■
Increase the Pace Blanking value.
■
Increase the sensitivity threshold.
Caution: Do not re-program the ICD to decrease oversensing
without assuring that appropriate sensing is maintained. See
“Setting up sensing” on page 61.
If the episode records reveal that a stable monomorphic VT has
been identified and treated as VF, consider these options to
improve the detection accuracy:
■
Review the Interval Plot for the episode, and adjust
VF Interval, if necessary. Use caution when reprogramming
the VF Interval, because changes to this value can adversely
affect VF detection.
■
Consider enabling FVT via VF detection. See “Detecting FVT
episodes” on page 76.
If the SVT episode records include episodes of true VT, review the
SVT episode record to identify the SVT detection criterion that
withheld detection. Adjust the SVT detection criteria parameters
as necessary. See “Enhancing detection with PR Logic criteria” on
page 88, and “Enhancing VT detection with the Stability criterion”
on page 97.
Verifying effective bradycardia pacing
To verify that the ICD is sensing and pacing appropriately, review
the following information from the Quick Look screen and
investigate as indicated:
■
Confirm that the patient is receiving adequate cardiac support
for daily living activities.
■
Review the pacing conduction history for comparison to the
patient history. A sharp increase in the paced beats
percentage may indicate a need for investigation and analysis.
Marquis DR 7274 Reference Manual
Considerations
Conducting a patient follow-up session
Verifying effective bradycardia pacing
■
Review the recorded Mode Switch episodes for comparison to
the patient’s atrial arrhythmia history. A dramatic increase in
frequency or duration of atrial episodes may indicate a need
for investigation and analysis.
To display more detailed information about the Mode Switch
episodes, perform these steps: select Episodes and Counters
from the Data icon; select the Mode Switch episodes from the
listed episode counters; then select the [Open Data] button.
■
Review the Cardiac Compass report for comparison to patient
history (see page 294).
■
Conduct pacing threshold tests (see page 313) to verify that
the programmed pacing outputs provide a sufficient safety
margin.
Review the following information before verifying bradycardia
pacing.
Atrial Pacing – If the conduction history shows a predominance
of atrial pacing despite a healthy sinus response, consider these
options to decrease the atrial pacing burden:
■
Decrease the Lower Rate.
■
Decrease the rate response or increase the activity threshold.
53
Ventricular Pacing – If the ventricle is predominantly paced and
the patient exhibits adequate ventricular response, consider these
options:
■
Decrease the Lower Rate.
■
Increase the AV delays.
Conduction History – If the reported percentages in the
conduction history do not add up to 100, the percentages may be
rounded. Frequent premature contractions or A:V dissociation
may also be the cause. Consider any of the following options:
■
Program the pacing mode to DDD or DDDR to promote A-V
synchrony. (If appropriate, enable Mode Switch to keep the
benefits of DDIR pacing during atrial high rate episodes.)
■
Enable Ventricular Rate Stabilization to smooth the heart rate
following premature ventricular beats.
Marquis DR 7274 Reference Manual
54
Chapter 5
Verifying effective bradycardia pacing
Marquis DR 7274 Reference Manual
Part III
Configuring the ICD for the patient
Marquis DR 7274 Reference Manual
Detecting tachyarrhythmias
Detection overview 58
Setting up sensing 61
Detecting VF episodes 66
Detecting VT episodes 70
Detecting FVT episodes 76
Detecting tachyarrhythmia episodes with Combined Count 81
Monitoring episodes for termination or redetection 84
Enhancing detection with PR Logic criteria 88
Enhancing VT detection with the Stability criterion 97
Detecting double tachycardias 100
6
6
Detecting prolonged tachyarrhythmias with High Rate Timeout 101
Key terms 103
Marquis DR 7274 Reference Manual
58
Chapter 6
Detection overview
Detection overview
The device detects ventricular tachyarrhythmias (VF, VT, and FVT)
by comparing the time intervals between sensed ventricular
events to a set of programmable detection intervals. If enough
intervals occur that are shorter than the programmed intervals, the
device detects a tachyarrhythmia, and responds automatically with
a programmed therapy. After delivering the therapy, the device
either redetects the arrhythmia and delivers the next programmed
therapy or detects episode termination.
To avoid detecting rapidly conducted SVTs (for example, sinus
tachycardia or atrial fibrillation) as ventricular tachyarrhythmias,
the device provides several detection enhancements, including
PR Logic and Stability detection criteria.
Figure 6-1 shows how all of these detection features interact
during initial detection. During redetection, the device does not
apply the PR Logic detection criteria.
Note: Detection functions can be turned off by programming the
VF Enable, FVT Enable, and VT Enable parameters to Off. For an
example, see “How to program VF detection” on page 68.
Marquis DR 7274 Reference Manual
Detecting tachyarrhythmias
Figure 6-1. How detection features interact during initial detection
59
Detection overview
No
Ye s
No
No / suspended by
High Rate Timeout
Ye s
Update counts and pattern information
Ventricular Event
Is the interval in the
VF, FVT, or VT detection zone?
Ye s
Has High Rate Timeout suspended detection
enhancements?
No
Does Stability reset the VT event count?
(VT and FVT via VT detection only)
No
Has a tachyarrhythmia event count
reached an NID?
Ye s
Are one or more PR Logic criteria on?
Ye s
Is the median ventricular interval less than the SVT
Limit?
No
Ye s
Tachy
Episode
Detected
Ye s
No
Is there a double tachycardia in progress?
No
Are one or more PR Logic criteria withholding
detection?
Marquis DR 7274 Reference Manual
Ye s
60
Chapter 6
Detection overview
Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops
classifying and counting tachyarrhythmia intervals. Sensing and
bradycardia pacing remain active, and the programmed detection
settings are not modified.
Detection is suspended
■
when the device senses the presence of a strong magnet. The
programmer head contains a magnet which suspends
detection, but once telemetry between the device and
programmer is established, detection resumes.
■
while performing any of the manual system tests, including
Underlying Rhythm, Pacing Threshold, Lead Impedance,
EGM Amplitude, and Charge/Dump. Detection automatically
resumes once the test is complete.
■
while performing a T-Shock, 50 Hz Burst, Manual Burst, or
PES Induction. You can choose to have the device
automatically resume detection after delivering the induction.
■
when you deliver a Manual or Emergency therapy. You can
resume detection by selecting the [Resume] button or
removing the programming head from the device.
■
when you select the on-screen [Suspend] button. You can
resume detection by selecting the [Resume] button or by
removing the programming head from the device.
■
during the automatic daily lead impedance measurements.
Detection resumes when the measurements are complete.
■
while the device is delivering an automatic tachyarrhythmia
therapy (including capacitor charging for defibrillation and
cardioversion). However, the device does continue to confirm
the detected episode during charging. Detection resumes
when the therapy is complete.
Note: The device suspends VT detection (and Combined
Count detection; see page 81) for 17 events following a
defibrillation therapy delivered in response to a detected VF.
■
during charging for Automatic Capacitor Formation. Detection
resumes when charging is complete.
1
If the defibrillation therapy is delivered as a result of a High Rate Timeout
Therapy operation, VT detection is not suspended (see page 103).
Marquis DR 7274 Reference Manual
1
Setting up sensing
The device provides bipolar sensing in both the atrium and
ventricle via the sensing electrodes of the implanted atrial and
ventricular leads. You can adjust the sensitivity to intracardiac
signals using independent atrial and ventricular sensitivity
settings. These settings define the minimum electrical amplitude
recognized by the device as an atrial or ventricular sensed event.
Proper sensing is essential for the safe and effective use of the
device. To provide appropriate sensing, the device uses:
■
auto-adjusting atrial and ventricular sensing thresholds
■
short (30 ms) cross-chamber blanking after paced events
■
no cross-chamber blanking after sensed events
See details about sensing on page 64.
Parameters
V. Sensitivity (mV) – Minimum amplitude of
electrical signal that registers as a sensed
ventricular event.
Detecting tachyarrhythmias
Setting up sensing
* Medtronic nominal setting
0.15, 0.3*, 0.45, 0.6,
0.9, 1.2
61
Considerations
A. Sensitivity (mV) – Minimum amplitude of
electrical signal that registers as a sensed
atrial event.
0.15, 0.3*, 0.45, 0.6,
0.9, 1.2, 1.5, 2.1
Review the following information before programming sensing
parameters.
Dual chamber sensing and bradycardia pacing modes – The
device senses in both the atrium and the ventricle at all times,
except when the programmed bradycardia pacing mode is DOO or
VOO. When the pacing mode is programmed to DOO or VOO
mode, there is no sensing in the ventricle. In order to program
either DOO or VOO mode, you must first disable detection.
Sensitivity thresholds – The programmed atrial and ventricular
sensitivity thresholds apply to all features related to sensing,
including detection and bradycardia pacing.
Marquis DR 7274 Reference Manual
62
Chapter 6
Setting up sensing
Bradycardia pacing and sensing – A combination of high
pacing pulse width or high amplitude with a low sensitivity
threshold may cause inappropriate sensing across chambers or in
the same chamber. Programming a lower pulse width, lower
amplitude, longer pace blanking, or a higher sensitivity threshold
may eliminate this inappropriate sensing.
Recommended ventricular sensitivity threshold – A ventricular
sensitivity threshold of 0.3 mV is recommended to maximize the
probability of detecting VF and to limit the possibility of
oversensing and cross-chamber sensing.
High ventricular sensitivity threshold – Setting V. Sensitivity to
a value greater than 0.6 mV is not recommended except for
testing. Doing this may cause undersensing, which can cause any
of the following situations:
■
delayed or aborted cardioversion therapy
■
delayed defibrillation therapy (when VF confirmation is active)
■
asynchronous pacing
■
underdetection of tachyarrhythmias
Low ventricular sensitivity threshold – If you set V. Sensitivity
to its most sensitive value of 0.15 mV, the device will be more
susceptible to EMI, cross-chamber sensing, and oversensing.
Recommended atrial sensitivity threshold – An atrial
sensitivity threshold of 0.3 mV is recommended to optimize the
effectiveness of PR Logic detection criteria and atrial pacing
operations, while limiting the possibility of oversensing and
cross-chamber sensing.
High atrial sensitivity threshold – If you set the A. Sensitivity
value too high, the device may not provide reliable sensing of
P-waves during SVTs and sinus rhythm.
Low atrial sensitivity threshold – If you set the A. Sensitivity
value to its most sensitive value of 0.15 mV, the device will be more
susceptible to EMI, far-field R-wave sensing, and oversensing.
Marquis DR 7274 Reference Manual
Detecting tachyarrhythmias
Setting up sensing
Testing sensitivity after reprogramming – If you change the
ventricular sensitivity threshold, evaluate for proper sensing and
detection by inducing VF and allowing the device to automatically
detect and treat the arrhythmia.
Atrial pacing and ventricular sensing – If you program the
device to an atrial pacing mode, make sure that it does not sense
atrial pacing pulses as ventricular events.
Sensing during VF – Always verify that the device senses
properly during VF. If the device is not sensing or detecting
properly, program detection and therapies off, and evaluate the
system (making sure to monitor the patient for life-threatening
arrhythmias until you enable detection and therapies again). You
may need to reposition or replace the ventricular sensing lead to
achieve proper sensing.
Atrial lead selection – Atrial leads with minimal tip-to-ring
spacing may reduce far-field R-wave sensing.
Repositioning the atrial lead – You may need to reposition or
replace the atrial sensing lead if reprogramming the atrial
sensitivity threshold does not provide reliable atrial sensing during
SVTs and sinus rhythm.
63
How to program sensitivity
1. Select Params > Detection.
2. Select the desired A. Sensitivity
and V. Sensitivity parameters.
3. Select [PROGRAM].
2
3
1
Marquis DR 7274 Reference Manual
64
Chapter 6
Setting up sensing
Details about sensing
Auto-adjusting sensitivity thresholds
The device automatically adjusts the sensitivity thresholds after
certain paced and sensed events to help reduce oversensing from
T-waves, cross-chamber events, and pacing. Figure 6-2 shows
how sensitivity thresholds are adjusted after different types of
events.
Figure 6-2. Auto-adjusting sensitivity thresholds
A
S
Sensitivity
Threshold
4
V
P
5
74Autoadjust.eps
Rectified and
Filtered A. EGM
Rectified and
Filtered V. EGM
Marker Channel
1 2
A
S
3
A
P
V
S
V
S
1 After an atrial sensed event, the atrial sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed
value, decay constant: 200 ms).
2 After a ventricular sensed event, the ventricular sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed
value, decay constant: 450 ms).
3 After an atrial paced event, the device does not adjust the atrial
sensitivity threshold. The ventricular sensitivity threshold increases
by 0.45 mV (decay constant: 60 ms).
4 After a ventricular paced event, the atrial sensitivity threshold
increases to 4x the programmed value (maximum: 1.8 mV,
immediate return after 60 ms).
5 After the ventricular pace blanking period is finished, the ventricular
threshold increases to 4.5x the programmed value (maximum:
1.8 mV, decay constant: 450 ms).
a
The exponential decay continues through a subsequent ventricular pacing pulse
and its blanking period.
b
If the programmed sensitivity value exceeds 0.3 mV (ventricular) or 1.2 mV
(atrial), the threshold is not adjusted.
Marquis DR 7274 Reference Manual
a
b
b
Detecting tachyarrhythmias
Setting up sensing
Blanking periods
During a blanking period, the device does not sense electrical
signals. This helps prevent sensing of device pacing,
cardioversion and defibrillation pulses, post-pacing depolarization,
T-waves, and multiple sensing of the same event. The blanking
periods following paced events are longer than those following
sensed events to avoid sensing the atrial and ventricular
depolarizations.
Notes:
■
To enhance sensing and detection during tachyarrhythmias,
the device does not apply cross-chamber blanking (blank
sensing in the opposite chamber) after a sensed event.
■
Atrial sensing is still active during the Post-Ventricular Atrial
Blanking (PVAB) period (see “Post-Ventricular Atrial Blanking
Period” on page 160).
Table 6-1 shows the duration of the fixed blanking periods. For
information on programmable pace blanking periods, see
page 148.
Table 6-1. Fixed blanking periods
Cross-chamber blanking after atrial or ventricular
pacing pulse
30 ms
65
Atrial blanking after sensed atrial event100 ms
Ventricular blanking after sensed ventricular event120 ms
Atrial and ventricular blanking after delivered
cardioversion or defibrillation therapy
Marquis DR 7274 Reference Manual
520 ms
66
Chapter 6
Detecting VF episodes
Detecting VF episodes
Refractory periods
During a refractory period, the device senses normally, but
classifies sensed events as refractory and limits its response to
these events. Pacing refractory periods prevent inappropriately
sensed signals, such as far-field R-waves (ventricular events
sensed in the atrium) or electrical noise, from triggering certain
pacing timing intervals.
Synchronization refractory periods help prevent the device from
delivering cardioversion and defibrillation therapies at
inappropriate times. See “Synchronizing defibrillation without
confirming VF” on page 114 and “Synchronizing cardioversion
after charging” on page 135.
Note: Refractory periods do not affect tachyarrhythmia detection.
The device detects VF episodes by examining the cardiac rhythm
for short ventricular intervals. If a predetermined number of
intervals occurs that are short enough to be considered VF events,
the device detects VF and delivers the first programmed VF
therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing.
See details about VF detection on page 68.
Parameters
VF Detection Enable – Turns VF
detection on or off.
VF Interval (ms) – V-V intervals shorter
than this value are counted as VF events.
VF Initial NID – Number of Intervals to
Detect: number of VF events the device
must count to detect a VF episode.
VF Redetect NID – Number of Intervals
to Redetect: number of VF events the
device must count to redetect a
continuing VF after a therapy.
Review the following information before programming
VF Detection parameters.
VF Interval minimum setting – To ensure proper VF detection,
you should not program the VF Interval less than 300 ms.
VF Interval maximum setting – Programming the VF Interval to
a value greater than 350 ms may cause inappropriate detection of
rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals
shorter than the VF Interval are counted using the VF event
counter, which is more sensitive than the consecutive VT event
counter.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the
Initial NIDs.
67
Enabling VF detection – When VF Detection Enable is
programmed On for the first time, the device
■
enables Automatic Capacitor Formation
■
starts recording Cardiac Compass data
■
starts recording lead performance trends (starting at
3:00 AM, by the device clock)
■
clears all brady pacing counters
VF detection and PR Logic criteria – You can program the
device to exclude rapidly conducted SVTs from VF Detection by
enabling the PR Logic detection criteria. Note that the SVT Limit
must be programmed shorter than the VF Interval in order for the
PR Logic criteria to affect VF detection. See “Enhancing detection
with PR Logic criteria” on page 88.
Double tachycardia detection – When any PR Logic detection
criteria are enabled, the device also enables double tachycardia
detection (VF, VT, or FVT in the presence of an SVT). See
“Detecting double tachycardias” on page 100.
Marquis DR 7274 Reference Manual
68
Chapter 6
Detecting VF episodes
Restrictions
How to program VF detection
2
Review the following information before programming
VF Detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e
reliable ventricular tachyarrhythmia detection, the programmer
regulates the values available for bradycardia pacing and
tachyarrhythmia detection. See “Parameter interlocks” on
page 386.
VF detection backup – To ensure VF Detection backup during
VT and FVT episodes, if VF Detection is off, both VT Detection
and FVT Detection must also be off.
To program VF detection:
1. Select Params > Detection.
2. Select the desired values for
VF Enable, VF Initial NID,
VF Redetect NID, and
VF Interval.
3. Select [PROGRAM].
Details about VF detection
The device detects VF by counting the number of VF events, which
are V-V intervals shorter than the programmed VF Interval. On
each event, the device counts the number of recent VF events.
The number of recent events examined is called the VF detection
window. The size of the VF detection window is the second
number in the programmed VF NID (for example, 24 events if the
VF Initial NID is 18/24).
Marquis DR 7274 Reference Manual
1
3
The threshold for detecting VF is the first number in the
programmed VF NID (for example, 18 events if the VF Initial NID
is 18/24). This threshold is always 75% of the VF detection
window. That is, if 75% of the events in the VF detection window
are VF events, the device detects a VF episode (see Figure 6-3).
After the device detects VF, it delivers the first programmed VF
therapy. Following the therapy, if the number of VF events reaches
the programmed VF Redetect NID, the device redetects VF and
delivers the next programmed VF therapy.
Note: The device can also detect VF Episodes via the Combined
Count detection criterion (see page 81).
Figure 6-3. Device detects VF
ECG
Detecting tachyarrhythmias
Detecting VF episodes
123
69
A
S
A
S
A
R
A
R
A
R
A
R
A
R
Marker Channel
FSFSFSFSFSFSFSFSFSF
F
F
S
VSVSV
D
S
VF Event Count
V
S
VSV
FSFSF
S
1 2 3 4 5 66 7 8 9 10 11 12 13 14 15 16 17 18
FSFSF
V
S
S
S
S
VF Interval
200 ms
1 VF starts, and the device begins counting VF events (intervals less than the programmed
VF Interval).
2 A ventricular interval occurs outside the VF detection zone. The VF event count is not
incremented.
3 The VF event count reaches the programmed VF NID value of 18 events out of 24, and the
device detects VF.
74VFDetection.eps
Marquis DR 7274 Reference Manual
70
Chapter 6
Detecting VT episodes
Detecting VT episodes
Parameters
The device detects VT episodes by examining the cardiac rhythm
for short ventricular intervals. If enough intervals occur that are
short enough to be considered VT events (but are not VF or FVT
events), the device detects VT and delivers the first programmed
VT therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing.
You can program the device to detect and record VT episodes
without treating them with VT therapies by setting VT Detection
Enable to Monitor. If a patient’s VT episodes are well-tolerated,
this feature allows you to collect data about these episodes without
delivering therapy or affecting VF detection.
See details about VT detection on page 72.
* Medtronic nominal setting
VT Detection Enable – Turns VT
detection on or off, or enables VT
monitoring.
On, Off*, or Monitor
VT Interval (Rate) (ms) – V-V intervals
shorter than this value are counted as
VT events.
VT Initial NID – Number of Intervals to
Detect: number of VT events the device
must count to detect a VT episode.
VT Redetect NID – Number of Intervals
to Redetect: number of VT events the
device must count to redetect a
continuing VT after a therapy.
Considerations
Review the following information before programming
VT Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Marquis DR 7274 Reference Manual
280, 290, . . ., 400*, . . ., 600
12, 16*, . . ., 52,
76, 100
4, 8, 12*, . . ., 52
Detecting tachyarrhythmias
Detecting VT episodes
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the
Initial NIDs.
VT Detection Enable, AFib/AFlutter, and Sinus Tach – When
you set VT Detection Enable to On or Monitor, the AFib/AFlutter
and Sinus Tach parameters are also automatically set to On.
VT detection and Combined Count detection – When
VT Detection is On, the device applies the Combined Count
detection criterion to help speed detection of rhythms that
fluctuate between detection zones. Combined Count detection is
disabled if VT Detection is set to Off or Monitor. See “Detecting
tachyarrhythmia episodes with Combined Count” on page 81.
VT detection and rapidly conducted SVTs – You can program
the device to exclude rapidly conducted SVTs from VT detection
by enabling the PR Logic or Stability detection criteria. See
“Enhancing VT detection with the Stability criterion” on page 97,
and “Enhancing detection with PR Logic criteria” on page 88.
71
Restrictions
Double tachycardia detection – When any PR Logic detection
criteria are enabled, the device also enables double tachycardia
detection (VF, VT, or FVT in the presence of an SVT; see
page 100).
Review the following information before programming
VT detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e
reliable ventricular tachyarrhythmia detection, the programmer
regulates the values available for bradycardia pacing and
tachyarrhythmia detection. See “Parameter interlocks” on
page 386.
VF detection backup – To ensure VF detection backup during
VT and FVT episodes, if VF Detection is off, both VT Detection
and FVT Detection must also be off.
Marquis DR 7274 Reference Manual
72
Chapter 6
Detecting VT episodes
How to program VT detection
2
Details about VT detection
To program VT detection:
1. Select Params > Detection.
2. Select the desired values for
VT Enable, VT Initial NID,
VT Redetect NID, and
VT Interval.
3. Select [PROGRAM].
1
3
The device detects VT by counting the number of consecutive VT
events. A VT event is a V-V interval shorter than the programmed
VT Interval but greater than or equal to the VF Interval. If the
number of consecutive VT events reaches the programmed VT
Initial NID, the device detects VT (see Figure 6-4).
The VT event count resets to zero whenever an interval occurs
that is greater than or equal to the programmed VT Interval. The
count remains at the current value if an interval is shorter than the
programmed VF Interval.
After the device detects VT, it delivers the first programmed VT
therapy. Following the therapy, if the VT event counter reaches the
VT Redetect NID, the device redetects VT and delivers the next
programmed therapy.
Note: The device can also detect VT Episodes via the Combined
Count detection criterion (see page 81).
Marquis DR 7274 Reference Manual
Figure 6-4. Device detects VT
ECG
132
Detecting tachyarrhythmias
Detecting VT episodes
73
A
S
A
S
ARARARARARARA
R
Marker Channel
VT Event Count
V
S
VSV
S
V
T
S
10123456
TSTSTSTSTST
S
S
VT Interval
ECG
ARARARARARARARARARARARA
R
Marker Channel
T
D
V
S
74VTDetection.eps
VT Event Count
TSTSTSTSTSTSTSTST
78910111213141516
S
VT Interval
200 ms
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval, but greater than or equal to the VF Interval).
2 A ventricular interval occurs outside VT detection zone. The VT event count resets to zero.
3 The VT event count reaches the programmed VT NID of 16 events, and the device detects
VT.
Marquis DR 7274 Reference Manual
74
Chapter 6
Detecting VT episodes
VT monitoring
You can program the device to record VT episodes without
delivering VT therapy by setting VT Detection to Monitor. When VT
monitoring is enabled, the device detects VT episodes but does
not deliver VT therapy (see Figure 6-5). Instead, it records VT
episodes, labeling them as “monitored,” and waits for episode
termination to occur.
When VT Detection is set to Monitor, several detection operations
work differently.
VT event counting – Before the device detects an episode, it
counts VT events normally. However, once the VT Initial NID is
reached, the device sets the VT event count to zero and suspends
VT event counting for the rest of the episode.
VF and FVT detection – VF and FVT detection operate as if VT
detection is off. Specifically, Combined Count detection is
disabled, and FVT via VT detection is not selectable. If a
monitored VT episode accelerates into the FVT or VF detection
zone, the device applies the VF Initial NID to detect the new
tachyarrhythmia. Once an episode is in progress, VT event
counting doesn’t resume until the episode ends.
Caution: Programming the VF Interval greater than 350 ms
may result in inappropriate detection of rapidly conducted
atrial fibrillation as VF or FVT via VF. Intervals shorter than the
VF Interval are counted using the VF event counter, which is
more sensitive than the consecutive VT event counter.
PR Logic and Stability criteria – Before the device detects a
tachyarrhythmia episode, the PR Logic and Stability criteria, if
turned on, are applied. If a monitored VT episode accelerates into
the FVT or VF detection zone, the device continues to apply
PR Logic criteria as initial VF or FVT detection begins. However,
because the Stability feature does not affect VF detection or FVT
via VF detection, it is not applied.
Episode termination – The device compares ventricular
intervals to the VT Interval to identify when a VT monitored
episode has ended. However, if a VF episode or FVT via VF
episode occurs when VT monitoring is enabled, the device
compares ventricular intervals to the VF Interval to identify
episode termination.
Marquis DR 7274 Reference Manual
Figure 6-5. Device detects and monitors VT
ECG
Detecting tachyarrhythmias
Detecting VT episodes
231
75
A
S
A
ARARA
R
A
R
R
ARARARARA
A
R
R
Marker Channel
VT Event Count
V
S
VSTSTSTST
1234
TSTSTSTDTSTST
S
13 141516000
S
74VTmonitor.eps
VT Interval
200 ms
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval but greater than or equal to the VF Interval).
2 The VT event count reaches the programmed VT NID of 16 events, and the device detects
VT.
3 After detecting the VT episode, the device resets the VT event count to zero and monitors
the episode until termination.
Marquis DR 7274 Reference Manual
76
Chapter 6
Detecting FVT episodes
Detecting FVT episodes
Parameters
The device detects episodes of Fast Ventricular Tachycardia (FVT)
by examining the cardiac rhythm for short ventricular intervals. If
enough intervals occur in the programmed FVT detection zone,
the device detects FVT and delivers the first programmed FVT
therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing. To make
sure it delivers sufficiently aggressive therapies, the device can
merge the programmed detection zones during redetection to
increase sensitivity.
See details about FVT detection on page 78.
* Medtronic nominal setting
FVT Detection Enable – Enables FVT
detection via the VF or the VT
detection algorithm.
Off*, via VF, or via VT
Considerations
FVT Interval (Rate) (ms) – V-V
intervals between this value and the
programmed VF Interval are marked
as FVT events.
200, 210, . . ., 600
Review the following information before programming
FVT Detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the
Initial NIDs.
Marquis DR 7274 Reference Manual
Detecting tachyarrhythmias
Detecting FVT episodes
FVT detection enable – Your choice for an appropriate setting for
FVT Detection should depend on the patient’s VF and VT cycle
lengths. After determining a reliably sensitive VF Interval, consider
the following suggestions:
■
If the patient presents with a clinical VT interval in the VF
zone, select via VF to ensure reliable detection of VF. (VT
Detection need not be enabled at all.)
■
If the patient presents with two clinical VTs, both outside
the VF zone, select via VT to allow for correct classification
of the faster VT and to offer a separate therapy regimen for
each VT.
■
If the patient presents with only one clinical VT which is
outside the VF zone, select VF and VT Detection only, and
set FVT Enable to Off.
FVT detection and PR Logic criteria – You can program the
device to exclude rapidly conducted SVTs from FVT Detection by
enabling the PR Logic detection criteria. Note that the SVT Limit
must be programmed shorter than the VF Interval for the PR Logic
criteria to affect FVT via VF detection.
77
Restrictions
Double tachycardia detection – When any PR Logic detection
criteria is enabled, the device also enables double tachycardia
detection (VF, VT, or FVT in the presence of an SVT, see
page 100).
Review the following information before programming
FVT Detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To
ensure reliable ventricular tachyarrhythmia detection, the
programmer regulates the values available for bradycardia pacing
and tachyarrhythmia detection. See “Parameter interlocks” on
page 386.
VF detection backup – To ensure VF Detection backup during
VT and FVT episodes, VT and FVT Detection cannot be on unless
VF Detection is also on.
Marquis DR 7274 Reference Manual
78
Chapter 6
Detecting FVT episodes
How to program FVT detection
2
FVT detection – To ensure reliable ventricular tachyarrhythmia
detection, the programmer regulates the values available for the
FVT parameter as follows:
■
VT Detection must be set to On if FVT Detection is set to
via VT.
■
If FVT Detection is set to via VF, the FVT Interval must be
programmed to a value shorter than the VF Interval.
■
If FVT Detection is set to via VT, the FVT Interval must be
programmed to a value greater than the VF Interval and less
than or equal to the VT Interval.
To program FVT detection:
1. Select Params > Detection.
2. Select the desired values for
FVT Enable and FVT Interval.
3. Select [PROGRAM].
Details about FVT detection
You can program the device to detect FVT episodes via the VF or
VT detection zone and NID.
When FVT Detection is set to via VF, a V-V interval within the FVT
detection zone is marked as an “FVT via VF” event. When the
VF NID is reached, the device reviews the last eight intervals:
■
If any of the last eight intervals are in the VF zone, it detects
the episode as VF.
■
If all of the last eight intervals are outside the VF zone, it
detects the episode as FVT (see Figure 6-6).
Marquis DR 7274 Reference Manual
1
3
When FVT Detection is set to via VT, a V-V interval within the FVT
detection zone is marked as an “FVT via VT” event. When the VT
NID is reached, the device reviews the last eight intervals:
■
If any of the last eight intervals are in the VF or FVT zones, it
detects the episode as FVT.
■
If all of the last eight intervals are outside the FVT and VF
zones, it detects the episode as VT.
Note: The device can also detect FVT episodes via the Combined
Count detection criterion (see page 81).
Figure 6-6. Device detects FVT via VF
Detecting tachyarrhythmias
Detecting FVT episodes
79
12
3
ECG
A
S
A
S
ARARA
A
R
ARARARARARARARARA
R
Marker Channel
VF Event Count
V
S
V
S
V
T
T
S
F
1 12345
TFTFTFT
S
TFT
F
TFTFTFTFTFV
F
13 14 15 16 17 18
VF and FVT Intervals
1 A fast ventricular tachycardia starts, and the first event falls into the FVT detection zone.
2 The second event of the FVT episode has an interval that falls into the VT zone. The VF
event count is not incremented.
3 The device detects FVT after the VF event count reaches the VF Initial NID.
R
S
74FvtDetect.eps
200 ms
Marquis DR 7274 Reference Manual
80
Chapter 6
Detecting FVT episodes
Figure 6-7. FVT zone merging
Zone merging after detection
To ensure the device delivers sufficiently aggressive therapies
during an extended or highly variable tachyarrhythmia episode,
the device merges detection zones during redetection in some
instances, as shown in Figure 6-7. The merged zone configuration
uses the event counting and therapies for the faster arrhythmia
and remains in effect until episode termination.
FVT set to “via VF”FVT set to “via VT”
Before
detection:
After VF
detection:
After FVT
detection:
VF
FVT
VT
VF and FVT zones merge, leaving
a larger VF zone.
VF
FVT
VT
VF
FVT
VT
VT and FVT zones merge, leaving a
larger FVT zone.
VF
FVT
VT
All zones remain unchanged.VT and FVT zones merge, leaving a
larger FVT zone.
VF
FVT
VT
Detection Intervals: VF Interval: 320 ms, FVT Interval: 280 ms / 360 ms , VT Interval: 400 ms
VF
FVT
VT
FVT1.eps
FVT2.eps
FVT3.eps
Marquis DR 7274 Reference Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.