Software Model 9970 4.0
Software Model 9971 4.0
Software Model 9974 4.0
Vision™ 4.0.1 Programmer Software
ICD System Reference Guide Supplement
Page 2
Page 3
Vision 4.0.1
ICD Software Supplement
A guide to using supplemental
software for programming the
7275 GEM III
III VR, and 7276 GEM III AT
Implantable Cardioverter
Defibrillators
®
DR, 7231 GEM
Page 4
GEM, Marker Channel, Decision Channel, PR Logic, T-Shock, Quick Look, Quick Links, Active Can,
and Flashback are trademarks of Medtronic, Inc.
Page 5
Table of Contents
Table of Contents
About This Software Supplement v
New Software Capabilities vi
1. Preferences Update
Preferences 1-2
2. Waveform Trace Enhancements
Adjusting and Configuring Waveforms 2-2
I. Index
Vision 4.0.1 ICD Software Supplementiii
Page 6
Table of Contents
ivVision 4.0.1 ICD Software Supplement
Page 7
About This Software Supplement
This programming guide supplement covers new Medtronic Vision 4.0.1
features and updates for software Models 9970, 9971, and 9974. This
guide supplements information provided in the following technical
literature:
■
Model 7275 GEM III DR System Reference Guide
■
Model 7231 GEM III VR System Reference Guide
■
Model 7276 GEM III AT System Reference Guide
About This Software Supplement
Vision 4.0.1 ICD Software Supplementv
Page 8
New Software Capabilities
New Software Capabilities
This software update includes new features, as well as enhancements to
existing features, as follows:
FeatureModels
Preferences features7275 GEM III DR, 7231 GEM III VR,
Waveform trace enhancements7275 GEM III DR, 7231 GEM III VR,
7276 GEM III AT
7276 GEM III AT
viVision 4.0.1 ICD Software Supplement
Page 9
Preferences Update1
Preferences 1-2
1
Vision 4.0.1 ICD Software Supplement1-1
Page 10
Preferences Update
Preferences
Preferences
A new menu item in the session icon allows you to access options for
printing reports. Selections are retained between sessions.
Select printing preferences
from this list.
Selecting the
Preferences... in the session icon displays the following
options:
■
Printing
■
Reports
Printing Preferences
Printing Preferences let you select the number of copies, default printer,
and whether to print now or later. Printing Preferences are retained
between sessions. The options are:
■
Pop up these options when any Print button is selected displays print
options when any Print button is selected (default)
■
Number of copies (default is 1)
■
Printer
–Full Size
–Programmer (default)
■
Print report
Now (default)
–
–Later from Print Queue
1-2Vision 4.0.1 ICD Software Supplement
Page 11
Preferences Update
For one step printing (excluding custom reports) from any screen, press
[Print...] once, using the following procedure:
Procedure for one-step printing
♦
1.Select the Session icon.
2.Select the Preferences... option.
3.Verify that Printing is selected on the list under Index.
4.Verify that Pop up these options when any Print Button is selected is
not selected.
5.Select the Printer.
Preferences
6.Select
7.Select [Ok].
Now under Print Report.
Reports Preferences
Selecting Print Initial Report after first interrogation allows you to
automatically print Quick Look and Episode Lists reports at the start of a
session, following the first interrogation. You can also print Histograms
from some of the other applications. Initial Report preference is retained
between sessions.
Vision 4.0.1 ICD Software Supplement1-3
Page 12
Preferences Update
Preferences
1-4Vision 4.0.1 ICD Software Supplement
Page 13
Waveform Trace
Enhancements
2
2
Adjusting and Configuring
Waveforms 2-2
Vision 4.0.1 ICD Software Supplement2-1
Page 14
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
Adjusting and Configuring Waveforms
The programmer Adjust window has been modified, and a waveform
adjustment button bar has been added to each waveform for greater
flexibility in adjusting waveform options.
The Adjust Window
Options available for adjusting the waveform traces using the Adjust
window are as follows:
ClippingSelect this checkbox to truncate the tops and bottoms of
ECG FilterSelect this checkbox to turn the ECG filter on or off. A
Show ArtifactsSelect this checkbox to enable or disable the pacing
Sweep speedSelect the Sweep Speed field to display the sweep speed
NormalizeSelect this button to equalize the trace spacing and adjust
OKSelect this button to close the Adjust window.
waveform traces at a 22 mm boundary. A checkmark (✓)
indicates that the filter is turned on. Default setting is (✓).
checkmark (✓) indicates that the filter is turned on. The
filter may improve clarity of the ECG in the presence of
interference. Default setting is (✓).
artifact enhancement function. A checkmark (✓) indicates
that the function is enabled.
options (12.5, 25, 50, and 100 mm/sec). To change the
sweep speed, select the desired option.
the size of each trace to the default setting.
2-2Vision 4.0.1 ICD Software Supplement
Page 15
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
To Open the Adjust Window
♦
1.Select the [Adjust...] button near the top of the tool palette. This
expands the live rhythm display window (if it is not already
expanded) and displays the Adjust window.
Vision 4.0.1 ICD Software Supplement2-3
Page 16
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
To Adjust Clipping
♦
1.Select theClipping checkbox in the Adjust window.
[Clipping] button
Selecting the
to be truncated at the 22 mm boundary, with a flat line drawn at the
boundary wherever a waveform is clipped.
To Select the ECG Filter
♦
1.Select the ECG Filter checkbox to turn the ECG filter on or off. A
checkmark (✓) indicates that the filter is turned on.
Show Artifact checkbox
The ECG filter on may improve clarity of both the displayed and printed
ECG in the presence of interference. The filter affects the ECG detection
bandwidth as follows:
■
Filter Off (no checkmark) – Bandwidth is 0.05 to 100 Hz
■
Filter On (checkmark) – Bandwidth is 0.5 to 40 Hz
Clipping checkbox causes EGM and ECG waveforms
ECG Filter checkbox
2-4Vision 4.0.1 ICD Software Supplement
Page 17
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
Enabling Artifact Enhancement
♦
Select the Show Artifacts checkbox to enable or disable the pacing artifact
enhancement function. A checkmark (✓) indicates that the function
is enabled.
The two trace illustrations below show how the ECG trace appears with
and without this feature enabled.
Artifact Enhancement Enabled (Checkmark)
Artifact Enhancement Disabled (No Checkmark)
Selecting a Different Sweep Speed
From the Adjust window, you can set the trace sweep speed to one of four
settings: 12.5, 25, 50, and 100 mm/sec.
To Change the Trace Display Sweep Speed
♦
➤ Select the Sweep Speed field in the Adjust window and select the
desired speed.
Sweep Speed Field
Menu of Choices
To Normalize the Rhythm Monitor
♦
➤ Select the [Normalize] button in the Adjust window. This refreshes the
Rhythm Monitor screen and restores each trace to its original default
size, with equal vertical spacing between the traces.
Vision 4.0.1 ICD Software Supplement2-5
Page 18
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
The Waveform Adjustment Button Bar
A waveform adjustment button bar has been added to each waveform for
greater flexibility in adjusting waveform options. To select a button, touch
it with the stylus.
Waveform Adjustment Buttons
Normalize button
Increase size – press and release to increase the size of the
waveform.
Normalize – refreshes the waveform trace and restores it to its
original default size.
Decrease size – press and release to decrease the size of the
waveform.
Waveform source – use to select waveform source.
Waveform print selection – Select to enable waveform for printing.
Up to two waveforms can be enabled.
2-6Vision 4.0.1 ICD Software Supplement
Page 19
♦
[Normalize] button
[Size] buttons
♦
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
To Adjust the Size or Amplitude of a Waveform
1.Select or in the waveform adjustment button bar for the trace
you wish to adjust.
2.Press the appropriate button to increase or decrease waveform size.
Each button may be pressed up to six times for EGM waveforms,
and up to 4 times for ECG waveforms.
To Normalize a Waveform
➤ To return the waveform to its default size, press the [Normalize] button.
Selecting a Trace Source
You can change the order in which traces appear in the monitor window.
One method is by changing the “source” of one or more of the traces as
described below. Also, you can use the stylus to “drag” a trace to the
desired location as described on page 2-9.
Note: The EGM 1 and EGM 2 sources reflect the programmed electrogram
sources, which are selectable from the
To Change the Source of a Trace
♦
1.Select the [Source] button from the waveform adjustment button bar.
[Source] button
■
Traces may be superimposed on one another. In this example
above, both the ECG Lead II and Marker Annotations make up
the first trace. Note that the name of only one source appears at
the left edge of the screen.
Vision 4.0.1 ICD Software Supplement2-7
Data - Collection Setup screen.
Page 20
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
■
2.Select the source field for the trace you want to change.
Source field
3.From the drop-down menu of Source options, select the
desired source.
Selecting a source option causes the source of the trace presently
displayed to change positions with the trace source you selected.
To display the name of the other source, “tap” the trace name at
the left edge of the screen. It will change to show the source of
the other trace.
Color Coding the Traces
To Change the Color of a Trace
♦
1.Open the Adjust... window. Select thecolor field in the waveform
trace area.
2.Select the color field for the trace you want to change.
Color Field
3.Select the desired color from the drop-down menu of color options.
2-8Vision 4.0.1 ICD Software Supplement
Page 21
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
Selecting waveforms for Chart Recorder Printing
At any time during a patient session, you can initiate a continuous, live
recording of the patient’s ECG, EGM, and Marker Channel traces on the
programmer’s chart recorder.
By default, the topmost ECG and the topmost EGM waveforms on the
programmer display are enabled for printing (below).
To Select Waveforms for Chart Recorder Printing
♦
1.Select the print button next to a print-enabled waveform that you do
not want to print.
Printing disabled for
this waveform
Printing enabled for
this waveform
This disables printing for that waveform. All other waveforms are then
available for selection.
2.Select the blank button next to the waveform you want to print.
The blank button will then change to a print button, indicating that the
waveform is enabled for printing.
Arranging the Traces
In addition to ordering the traces by selecting the source as described on
page 2-7, you can use the stylus to “drag” a trace to the desired position.
The procedure below shows how to move the Marker Annotation trace
from its current position to a position over the ECG Lead II trace.
Vision 4.0.1 ICD Software Supplement2-9
Page 22
Waveform Trace Enhancements
Adjusting and Configuring Waveforms
To Move a Trace Using the Stylus
♦
1.Press and hold the stylus against the name of the trace you want
to move.
2.Without lifting the stylus, drag the box now appearing around the
trace name and waveform adjustment button bar to the desired
location.
Note: To change the position of a superimposed trace by dragging it,
the source that you want to move must show at the left edge of
the screen.
3.When you have the box positioned where you want the trace to
appear, lift the stylus. If you are positioning the trace over another,
the trace will “snap” into position.
To equalize the spacing between the traces in the new arrangement,
select the
Medtronic Japan
Solid Square West Tower 6F,
580 Horikawa-cho, Saiwai-ku,
Kawasaki, Kanagawa 210-0913
Japan
Tel. 81-44-540-6112
Fax 81-44-540-6200
Australia
Medtronic Australasia Pty. Ltd.
Unit 4/446 Victoria Road
Gladesville NSW 2111
Australia
Tel. 61-2-9879-5999
Fax 61-2-9879-5100
Asia
Medtronic International Ltd.
Suite 1602 16/F, Manulife Plaza
The Lee Gardens, 33 Hysan Avenue
Causeway Bay
Hong Kong
Tel. 852-2891-4068
Fax 852-2591-0313
Americas
Latin America Headquarters
Medtronic, Inc.
7000 Central Avenue NE
Minneapolis, MN 55432-3576
USA
Tel. 763-514-4000
Fax 763-514-4879
Canada
Medtronic of Canada Ltd.
6733 Kitimat Road
Mississauga, Ontario L5N 1W3
Tel. 905-826-6020
Fax 905-826-6620
Toll-free in Canada: 1-800-268-5346
United States
World Headquarters
Medtronic, Inc.
710 Medtronic Parkway NE
Minneapolis, MN 55432-5604
USA
Internet: www.medtronic.com
Tel. 763-514-4000
Fax 763-514-4879
Medtronic USA, Inc.
Toll-free in the USA: 1-800-723-4636
(24-hour consultation for physicians
and medical professionals)