Connecting the Programmer Hardware 7-2
Setting Up and Using the Programmer 7-12
Using the Programmer Display Screen 7-22
Starting a Patient Session 7-30
Programming Parameter Values 7-32
Checklist 7-35
Live Rhythm Monitor 7-42
Recording Live Waveforms 7-54
Printing Reports 7-61
Changes This Session 7-66
Ending a Patient Session 7-67
8. Conducting System Tests
System Tests Overview 8-2
Underlying Rhythm Test 8-3
Pacing Threshold Test 8-4
Lead Impedance Test 8-6
EGM Width Test 8-9
Manual EGM Amplitude Test 8-14
Test Charge / Dump 8-16
Data Collection Overview 10-2
Quick Look™ 10-4
Setting Up Data Collection 10-6
Counter Data 10-11
Episode Data 10-15
Patient Information 10-28
Battery and Lead Measurements 10-32
Device Status Indicators 10-36
Saving and Retrieving ICD Data 10-38
Induction 9-8
Table of Contents
11. Patient Alert™ Monitoring System
Patient Alert™ Setup 11-2
Instructions for the Patient 11-7
Instructions for Clinic Personnel 11-9
Patient Alert™ Event Log 11-10
12. Implant and Follow-up
Pre-Implant Information 12-2
Operative Set-Up 12-4
Surgical Approach 12-6
Sensing and Pacing Measurements 12-8
Defibrillation Testing 12-9
Placing the ICD 12-15
Replacing the ICD 12-16
Pre-Discharge Evaluation 12-18
Follow-Up Recommendations 12-19
A. Appendix A - Warnings and Precautions
General Warnings A-2
GEM®III VR System Reference Guide
v
Table of Contents
Sterilization, Storage, and Handling A-3
Implantation and ICD Programming A-4
Lead Evaluation and Lead Connection A-5
Follow-up Testing A-6
ICD Explant and Disposal A-7
Environmental and Medical Therapy Hazards A-7
Home and Occupational Environments A-9
Before implanting the ICD, it is strongly recommended that you:
■
Refer to the product literature packaged with the ICD for information
about prescribing the ICD.
■
Thoroughly read this manual and the technical manuals for the leads
used with the device.
■
Discuss the procedure and the ICD system with the patient and any
other interested parties, and provide them with any patient
information packaged with the ICD.
Education
Medtronic invites physicians to attend an education seminar on the
complete ICD system. The course includes indications for use, an
overview of ICD system functions, implant procedures, and
patient management.
Technical Support
Medtronic employs highly trained representatives and engineers located
throughout the world to serve you and, upon request, to provide training
to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to
provide technical consultation to product users. For medical consultation,
Medtronic can often refer product users to outside medical consultants
with appropriate expertise.
For more information, contact your local Medtronic representative, or call
or write Medtronic at the appropriate address or telephone number listed
on the back cover.
GEM®III VR System Reference Guide
vii
Introduction
References
Automatic defibrillation was invented and patented in 1953 by Dr. F.
Zacouto, who designed an external device that delivered a defibrillation
impulse to the heart upon detection of very rapid ECG activity in
combination with the absence of arterial pulsations.
The primary reference for background information is Zacouto FI, Guize LJ.
Fundamentals of Orthorhythmic Pacing. In: Luderitz B, ed.
Pacing Diagnostic and Therapeutic Tools.
New York: Springer-Verlag;
1976: 212-218.
See these additional references for more background information:
■
Singer I, Ed.
Implantable Cardioverter-Defibrillator
Futura Publishing Co. 1994.
■
Singer I, Barold SS, Camm AJ, Eds.
Nonpharmacological Therapy of
Arrhythmias for the 21st Century: The State of the Art
Futura Publishing Co. 1998.
■
Estes M, Manolis AS, Wang P, Eds.
Cardioverter-Defibrillators
■
Kroll MW, Lehmann MH, Eds.
. New York, NY: Marcel Dekker, Inc. 1994.
Therapy: The Engineering-Clinical Interface
Implantable
Implantable Cardioverter-Defibrillator
. Norwell, MA: Kluwer
Academic Publishers 1996.
Cardiac
. Armonk, NY:
. Armonk, NY:
viii
Notice
This software is provided as an informational tool for the end user. The
user is responsible for accurate input of patient information into the
software. Medtronic makes no representation as to the accuracy or
completeness of the data input into the software. MEDTRONIC SHALL
NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTIAL OR
CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH
RESULTS FROM THE USE OF THE INFORMATION PROVIDED IN THE
SOFTWARE.
GEM®III VR System Reference Guide
Abbreviations and Acronyms
ATPAntitachycardia Pacing
BOLBeginning of Life
bpmbeats per minute
CNIDCombined (VT and VF) Number of Intervals to Detect
QRS Snapshotb, and episode summaries (typically, only
episode summaries are provided for NSTs)
1000 intervals before interrogation, latest VF, and latest VT
- Non-sustained tachy episodes and SVTs
- Percentage paced/sensed rhythm
- Single PVCs and PVC runs
- Rate stabilization pulses and runs
Counts for each VF, FVT, VT Therapy:
- Delivered
- Successful
- Unsuccessful
- Intervention
Total number of aborted shocks
Up to 10 log entries: text and date for the first time an alert
is triggered between clearing the log entries.
b
GEM®III VR System Reference Guide
1-3
Quick Reference
Primary Capabilities
Other features
Direct replacement for DF-1 / IS-1 systems, including all Medtronic “C,”
“Cx,” and “D” ICDs
Medtronic icon-based user interface
Full-size report printing
Quick Look™ programming summary with Quick Links™ to stored
episodes and observations
Non-invasive EP study functions: ventricular inductions / manual
therapies
Smart Auto Cap Formation extends the formation timer after
therapeutic charging
Auto-Resume feature automatically resumes detection after an induction
Checklist feature provides streamlined, customizable navigation through
sets of tasks for follow-up and implant visits
1-4
GEM®III VR System Reference Guide
Physical Characteristics
Volume39.2 cc
Mass77.4 g
H x W x D
Surface area of device can69 cm
a
Quick Reference
Physical Characteristics
69.3 mm x 50.8 mm x 13.5 mm
2
Radiopaque ID
Materials in contact with
human tissue
b
c
PJL
Titanium / polyurethane / silicone rubber
BatteryLithium silver vanadium oxide
Connector configurationOne IS-1 connector for pacing and sensing,
Two DF-1 connectors for high voltage therapy,
Active Can® electrode
SVC
(HVX)
IS-1 V
RV
DF-1
(HVB)
Can (HVA)
a
This measurement refers to the thickness (depth) of the can; grommets may protrude slightly
beyond can surface.
b
Engineering series number follows the radiopaque code.
c
These materials have been successfully tested for the ability to avoid biological
incompatibility. The device does not produce an injurious temperature in the surrounding
tissue.
GEM®III VR System Reference Guide
1-5
Quick Reference
Electrical Characteristics
Electrical Characteristics
Longevity
Table 1-2 below provides longevity estimates based on accelerated
battery discharge data and device modeling with 60 ppm (min-1) pacing
rate, 2 V pacing pulse amplitude, 0.6 ms pacing pulse width, and
30 J delivered therapy energy
Table 1-2. GEM®III VR Projected Longevity in Years with 2 V Pacing Pulse
Percent
pacing
0%Semi-Annual8.98.9
15%Semi-Annual8.68.7
50%Semi-Annual8.28.5
100%Semi-Annual7.68.1
Maximum Energy
Charging Frequency
Quarterly7.87.8
Quarterly7.57.6
Quarterly7.27.4
Quarterly6.87.2
1
.
Amplitude and 0.6 ms Pacing Pulse Width
500 ohm pacing
impedance
900 ohm pacing
impedance
1-6
Table 1-3 below provides longevity estimates based on accelerated
battery discharge data and device modeling with 60 ppm (min-1) pacing
rate, 3 V pacing pulse amplitude, 0.4 ms pacing pulse width, and
30 J delivered therapy energy
Table 1-3. GEM®III VR Projected Longevity in Years with 3 V Pacing Pulse
Percent
pacing
0%Semi-Annual8.98.9
15%Semi-Annual8.48.5
50%Semi-Annual7.58.0
100%Semi-Annual6.57.4
1
EGM pre-storage feature set to Off.
Maximum Energy
Charging Frequency
Quarterly7.87.8
Quarterly7.47.5
Quarterly6.77.0
Quarterly6.06.6
GEM®III VR System Reference Guide
1
.
Amplitude and 0.4 ms Pacing Pulse Width
500 ohm pacing
impedance
900 ohm pacing
impedance
Quick Reference
Electrical Characteristics
The following factors result in decreased longevity:
■
an increase in pacing rate, pacing amplitude or pulse width, the ratio
of bradycardia paced to sensed events, or the charging frequency,
■
a decrease in pacing impedance, or
■
using the pre-onset EGM storage feature or Holter telemetry.
To maximize ICD longevity:
■
Optimize pacing rate and output parameters for the patient.
■
Set the delivered energy value for the first programmed high voltage
therapy to a value lower than maximum energy.
■
Set the automatic capacitor formation interval to a longer value
(unless a shorter interval is required to maintain clinically acceptable
charge times).
■
Use the pre-onset EGM storage feature and Holter telemetry only for
short periods of time.
Battery Voltage
Battery voltage appears on the programmer display and on printed
reports. See “Preparing the ICD for Implant” on page 12-5, and
“Replacement Indicators” on page 12-20 for more information.
Table 1-4. Battery Voltage Indicators
Elective Replacement (ERI)
a
≤ 2.55 V
Pacing mode and rate (magnet and non-magnet)as programmed
End of Life (EOL)
b
≤ 2.40 V
Pacing mode and rate (magnet and non-magnet)as programmed
a
The minimum time between ERI and End of Life is three months, assuming 100% VVI pacing
at 60 ppm (min
b
When the battery reaches EOL, immediate replacement is recommended. (See also “Charge
Circuit Timeout” on page 5-4.)
-1
), 3 V, 0.4 ms; 500 Ω pacing load; and six 30 J charges.
GEM®III VR System Reference Guide
1-7
Quick Reference
Electrical Characteristics
Charge Time
The high voltage capacitors charge more rapidly just following a capacitor
formation, when they are fully formed. Between formations, the capacitors
gradually deform, resulting in longer charge times toward the end of the
formation interval. See page 8-18 for information on managing charge
time. See page 8-17 for instructions on using the Test Charge feature to
measure charge time. Table 1-5 provides typical full energy charge times
at BOL and ERI based on the time interval since the last capacitor
formation. A full energy charge is any charge to full energy (either via a
capacitor formation or therapy delivery).
Table 1-5. Typical Full Energy (30 J Delivered) Charge Times
Capacitors
fully formed
At Beginning of Life (BOL) 6 seconds≈ 1 second
Near Elective Replacement
Indicator (ERI)
a
Typical values for stated conditions, actual charge time performance may vary.
14 seconds≈ 1 seconds
1 month after
capacitor
formation
longer than fully
formed
longer than fully
formed
a
6 months after
capacitor
formation
≈ 2 seconds
longer than fully
formed
≈ 5 seconds
longer than fully
formed
1-8
GEM®III VR System Reference Guide
Quick Reference
Electrical Characteristics
High Voltage Therapy Energy
Table 1-6 compares delivered energy levels to energy levels stored by the
device.
Note: The stored energy of the device is derived from the peak capacitor
voltage and is always greater than the energy delivered by the device.
Table 1-6. Comparing Programmed (Actual Delivereda) and Storedb Energy Levels
Programmed
(Actual Delivered)
Energy (J)
0.40.5.099111.94
0.60.8.1310122.16
0.81.0.1711132.38
1.01.2.2212142.59
1.21.5.2613152.81
1.41.7.3014163.02
1.62.0.3515183.24
1.82.2.3916193.46
22.4.4318213.89
33.6.6520244.32
44.7.8622264.75
55.91.0824285.18
67.11.3026315.62
78.31.5128336.05
89.41.7330356.48
a
Energy and voltage delivered at connector block into a 75 ohm load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at BOL, with fully formed capacitors.
Stored
Energy (J)
Charge
Timec (sec)
Programmed
(Actual Delivered)
Energy (J)
Stored
Energy (J)
Charge
Timec (sec)
GEM®III VR System Reference Guide
1-9
Quick Reference
Magnet Application
Magnet Application
Applying a magnet closes a reed switch within the ICD (see Table 1-7 for
magnet effects). When the magnet is removed, the ICD returns to its
programmed operations.
Pacing Modeas programmed
Pacing Rate and Intervalas programmed
VF, VT, and FVT Detection
and Therapy
Patient Alert™ Audible Tones
a
The programming head contains a magnet which can suspend detection. However, if the
Resume option has been enabled via the programmer, detection will remain enabled.
b
The alert sounds only if Patient Alert™ is programmed On for any alert except POR. A POR
alert causes an alarm every 20 hours, even if Patient Alert is programmed Off.
Table 1-7. Effects of Magnet Application
suspended
b
(30 seconds or less)
continuous tone (Alarm OK)
on/off intermittent tone (seek follow-up)
high/low dual tone (urgent follow-up)
a
1-10
GEM®III VR System Reference Guide
Emergency Settings
The table below lists the values that can be selected for the ICD
emergency therapy parameters. Default values are highlighted in
italicized
page 1-24.
Table 1-8. Emergency Therapy Parameters
ParameterSelectable Values (and
Defibrillation
Energy (J)10, 11, . . ., 16,
PathwayAX→B
Cardioversion
Energy (J)0.4, 0.6, . . ., 1.8,
PathwayAX→B
Fixed Burst
Pacing Interval (ms)100, 110, . . .,
Pulse Amplitude (V)8
Pulse Width (ms)1.6
text. The Notes column lists letters than correspond to notes on
18, 20, . . .,
2, 3, . . ., 16
18, 20, . . .,
360, 370, . . ., 600
30
30
350
,
default value
Quick Reference
Emergency Settings
bold,
)Notes
VVI Pacing
Pacing ModeVVI
Lower Rate (ppm, min-1)70
Vent. Rate StabilizationOff
HysteresisOff
SensitivityAs programmed
Pulse Amplitude (V)8 V
Pulse Width (ms)1.6 ms
Pace Blanking (ms)320 ms
Post Shock Pace Blanking (ms)320 msA.
GEM®III VR System Reference Guide
1-11
Quick Reference
Programmable Parameters Overview
Programmable Parameters Overview
The tables that follow (from “Detection Parameters” on page 1-13 to
“Patient Information Parameters” on page 1-20) list the values that can be
selected for all ICD programmable parameters. As applicable, these
tables also indicate the Medtronic nominal value, the POR value, and the
shipped value for each parameter.
Medtronic nominal values – are suggested settings for parameters
under nominal conditions. Medtronic nominal values, if applicable, are
indicated in the Programmable Values column with
POR values – are the values to which the ICD resets parameters after a
power-on reset (POR). If a POR occurs, the ICD sounds a high-urgency
Patient Alert™ tone every twenty hours. See the “Power-On Reset
Recovery Procedure” on page 10-37.
Shipped values – are the programmed parameter values when the ICD
is shipped from Medtronic.
The Notes column in each table lists letters that correspond to notes on
page 1-24. These include exceptions or clarifications that relate to the
values listed in the programmable parameter tables.
ALL THERAPIES IN A ZONE EXHAUSTED FOR AN EPISODE ALERT
Alert Enable - Urgency Off,
On-Low
, On-HighOffOff
On-Low
, On-HighOffOff
On-Low
, On-HighOffOff
On-Low
, On-HighOffOff
GEM®III VR System Reference Guide
W.
W.
W.
W.
1-19
Quick Reference
Patient Information Parameters
Patient Information Parameters
Table 1-15. Description of Patient Information
Information FieldDescription and Action
PatientEnter the patient name or ID number (up to 30 characters).
IDEnter the patient ID.
Date of BirthSelect patient’s date of birth.
ICD Number
(Not selectable)
Lead 1 . . .
Lead 2 . . .
Lead 3 . . .
ImplantEnter measured values from implant. Using the displayed submenu, enter new
NotesEnter notes about patient or other information.
History (2 fields)Select medical history from list box. If necessary, you can indicate that more
EF, onSelect the Ejection Fraction from a table of values. Then enter the date in the
PhysicianSelect or add the name of the physician.
PhoneEnter the phone of the physician.
HospitalSelect or add the hospital information.
Last Update
(Not selectable)
Displays serial number for implanted ICD after interrogation.
Enter detailed information for leads 1, 2, and 3:
•
Model
•
Position:
• Serial Number
• Manufacturer
• Implant date
information from the keyboard and make list box selections.
information is located in the Notes field.
following field.
Displays the date of the last Patient Information update.
pick from list of positions, or enter position
1-20
GEM®III VR System Reference Guide
EP Studies Parameters
Table 1-16. EP Studies Parameters (Sheet 1 of 2)
ParameterSelectable Values (and
INDUCTIONS
T-Shock™ Induction Parameters
#S12, 3, . . .,
S1S1 (ms)350, 360, . . .,
410, 420, . . ., 1760
Delay (ms)50, 60, . . .,
320, 330, . . ., 600
Energy (J)0.4,
Pathway
Waveform
Pulse Amplitude8 V (fixed)
Pulse Width1.6 ms (fixed)
EnableEnabled,
Resume at Deliver
2, 3, . . ., 16,
18, 20, . . ., 30
AX→B
Monophasic
Enabled
0.6
, . . ., 1.8,
, B→AX
, Disabled
8
400
310
,
, Biphasic
Disabled
,
default value
Quick Reference
EP Studies Parameters
)Notes
V.
U.
50 Hz Burst Induction ParametersV.
Interval (ms)20 ms (fixed)
Pulse Amplitude (V)1, 2, . . ., 6,
Pulse Width (ms)0.03, 0.06,
Resume at Burst
Manual Burst Induction ParametersV.
Interval (ms)100, 110, . . .,
Pulse Amplitude (V)1, 2, . . ., 5, 6, 8
Pulse Width (ms)0.03, 0.06,
Resume at Burst
0.1, 0.2, . . .,
Enabled
0.1, 0.2, . . .,
0.6, 0.7, . . ., 1.6
Enabled
8
1.6
, Disabled
600
0.5
,
, Disabled
GEM®III VR System Reference Guide
1-21
Quick Reference
EP Studies Parameters
Table 1-16. EP Studies Parameters (Sheet 2 of 2)
ParameterSelectable Values (and
PES Induction Parameters
#S11, 2, . . ., 8,
S1S1 (ms)100, 110, . . .,
S1S2 (ms)Off, 100, 110, . . .,
S2S3 (ms)
S3S4 (ms)
Pulse Amplitude (V)1, 2, . . ., 5, 6, 8
Pulse Width (ms)0.03, 0.06,
Resume at Deliver
MANUAL THERAPY PARAMETERS
In general, each manual therapy uses the same parameter values as the automatic therapy. See “Therapy
Parameters” beginning on page 1-14.
9, 10, . . ., 15
600
610, 620, . . ., 1760
410, 420, . . ., 600
Off
, 100, 110, . . ., 600V.
Off
, 100, 110, . . ., 600V.
0.1, 0.2, . . .,
0.6, 0.7, . . ., 1.6
Enabled
0.5
, Disabled
,
400
,
default value
,
)Notes
U.
V.
1-22
GEM®III VR System Reference Guide
Quick Reference
Fixed Parameters
Fixed Parameters
Table 1-17. Fixed Parameters
ParameterValueNotes
SENSING
Blanking Periods
After a sensed vent. event120 ms
After charging ends300 ms
After HV therapy520 ms
HIGH VOLTAGE THERAPIES
Maximum charging period30 seconds
WaveformBiphasic
Tilt50%
VF Synchronization PeriodsR.
After post-charge blanking500 ms
After a sensed event500 ms
Synchronization Refractory Periods
After V. sense during CV synch.200 ms
After charge during DF synch.400 msP.
Post-Therapy
Escape interval after CV/Defib1200 ms
Suspension of VT Detection
after Defibrillation therapy
EP STUDIES
T-Shock™ S1 pacing outputs1.6 ms at 8 VV.
HARDWARE PARAMETERS
Ventricular rate limit
(protective feature)
Input protection 100 kΩ minimum
17 V. events
171 ppm (min
-1
)S.
GEM®III VR System Reference Guide
1-23
Quick Reference
Notes
Notes
A.
Post-shock Pace Blanking is reprogrammed by Emergency VVI pacing,
but is not displayed on the Emergency Screen.
B.
The VF Initial NID cannot be programmed to a value greater than 30/40 if
VT Detection is enabled.
C.
The measured intervals are truncated to a 10 ms multiple (e.g., 457 ms
becomes 450 ms). The device uses this truncated interval value when
applying the programmed criteria and calculating interval averages.
D.
With a 40 ms sine2 waveform. When using the Cenelec waveform, the
rated sensing threshold value will be 1.5 times the rated sine
sensing threshold.
E.
This setting applies to all sensing for both tachyarrhythmia detection and
bradycardia pacing operations.
F.
Nominal value is blank until detection is enabled.
G.
EGM width is measured on the intracardiac EGM programmed as
Channel 2.
H.
The shortest pacing escape interval must exceed all programmed
Detection Intervals by at least 60 ms.
I.
These specifications apply to both automatic and manual therapies,
except as noted.
J.
For automatic therapy 3, 4, 5, or 6, energy must be at least 10 joules.
K.
Delivered energy based on a biphasic pulse into a 75 ohm load.
L.
Applies only to the first VF therapy that is programmed On.
M.
Smart Mode is only available for Therapies 1, 2, 3, and 4.
N.
Pacing mode must be VVI, and the hysteresis rate must be less than the
pacing rate.
O.
The ICD is shipped with Automatic Capacitor Formation Interval set to
Off. However, when VF Detection is set to On for the first time, the
Automatic Capacitor Formation Interval is automatically set to 6 month.
P.
When Reconfirm VF is active, a refractory period equal to 60 ms plus the
greatest programmed detection interval follows the post-charge blanking
period and all paced events.
Q.
Both channels appear on-screen as real-time telemetered signals.
R.
When Reconfirm VF is active, the VF synchronization (reconfirmation)
periods are 60 ms plus the greatest programmed detection interval; and
the synchronization escape period is equal to the current lower rate (i.e.,
sensor or programmed lower rate) escape period.
S.
Does not apply during therapies or programmed high rates.
T.
The time stamp on episode records and other stored data is determined
by the device’s date/time clock.
U.
Delivered in the VVI mode.
2
1-24
GEM®III VR System Reference Guide
Quick Reference
V.
Delivered in the VOO mode.
W.
After a POR, all programmed Patient Alert conditions are disabled. The
Reset condition itself activates a high-urgency alert that sounds every
twenty hours
X.
For manual ramp therapy, 6 is the nominal number of pulses, and 97% is
the nominal value for % of R-R interval.
until cleared.
Notes
GEM®III VR System Reference Guide
1-25
Quick Reference
Notes
1-26
GEM®III VR System Reference Guide
Emergency Therapy
2
2
Overview 2-2
Effect on System Operation 2-3
On-Screen and Display
Panel Buttons 2-3
Emergency Defibrillation 2-4
Emergency Cardioversion 2-5
Emergency Fixed Burst Pacing 2-6
Emergency VVI Pacing 2-7
GEM®III VR System Reference Guide
2-1
Emergency Therapy
Overview
Overview
The ICD provides the following Emergency therapies:
■
Defibrillation therapy
■
Cardioversion therapy
■
Fixed Burst pacing therapy
■
Maximum output VVI pacing
To deliver an Emergency therapy, the programming head must be
positioned over the ICD.
Defibrillation is the default Emergency therapy, and 30 joules is the default
delivered energy. When you select
[Emergency] and [DELIVER], the ICD
charges and delivers a biphasic 30 joule shock along the AX>B pathway.
If you select a different defibrillation energy, the programmer resets the
energy to 30 joules the next time you select
[Emergency]. Emergency
Cardioversion and Emergency Fixed Burst values remain as selected for
the duration of the session.
2-2
Aborting the Therapy – As a safety precaution, the programmer also
displays an
[ABORT] button which immediately terminates any Emergency
therapy in progress.
To return to other programming functions from an Emergency screen,
select
[Exit Emergency].
GEM®III VR System Reference Guide
Effect on System Operation
The ICD suspends its automatic detection features when Emergency
defibrillation, cardioversion, or fixed burst pacing therapies are delivered.
Detection is not suspended during Emergency VVI Pacing. Removing the
programming head or pressing
[Resume] re-enables detection.
On-Screen and Display Panel Buttons
The on-screen [Emergency] button and the red mechanical Emergency
button by the programmer display panel are equivalent at all times.
Emergency Therapy
Effect on System Operation
Red Emergency Button
The on-screen
[DELIVER] button and the yellow-on-blue mechanical
Functions the same as on-screen
[Emergency]
at all times.
Deliver button by the programmer display panel are equivalent during
Emergency operations only. The mechanical Deliver button operates only
during Emergency operations.
Yellow-on-Blue
Deliver Button
Functions the same as on-screen
[DELIVER], but only during
Emergency functions.
Temporary Parameter Values
Emergency tachyarrhythmia therapies use
the permanently programmed parameters of the ICD.
in effect until you begin the temporary operation by selecting
After the tachyarrhythmia therapy is complete, the ICD reverts to its
permanently programmed values.
test values
that do not change
1
Test values are not
[DELIVER].
1
Delivery of Emergency VVI Pacing permanently changes the programmed bradycardia
pacing values to those used during emergency pacing (see page 2-7).
GEM®III VR System Reference Guide
2-3
Emergency Therapy
Emergency Defibrillation
Emergency Defibrillation
To Deliver Emergency 30 Joule Defibrillation
♦
1.Position the programming head over the ICD.
2.Select [Emergency].
3.Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly
positioned and select
To Change the Defibrillation Energy and Deliver
♦
1.From the Emergency Therapy screen, select Energy.
2.Select a new energy value.
3.Select [DELIVER].
[Retry] or [Cancel].
Select to Change
Emergency Defibrillation
Select to Deliver
Emergency Defibrillation
2-4
GEM®III VR System Reference Guide
Energy
Figure 2-1. Emergency Defibrillation Screen
Emergency Cardioversion
To Deliver Emergency Cardioversion
♦
1.Position the programming head over the ICD.
2.Select [Emergency].
3.Select [Cardioversion].
4.Accept the cardioversion energy shown on the screen, or select
Energy and select a new value from the window.
5.Select [DELIVER].
If delivery is not confirmed, verify that the programming head is properly
positioned and select
Emergency Therapy
Emergency Cardioversion
[Retry] or [Cancel].
Emergency Cardioversion
Select to Change
Energy
Select to Deliver
Emergency Cardioversion
Figure 2-2. Emergency Cardioversion Screen
GEM®III VR System Reference Guide
2-5
Emergency Therapy
Emergency Fixed Burst Pacing
Emergency Fixed Burst Pacing
To Deliver Emergency Fixed Burst Pacing
♦
1.Position the programming head over the ICD.
2.Select [Emergency].
3.Select [Fixed Burst].
4.Accept the pacing interval shown on the screen, or select Interval for
a new interval value.
Select to Change
Burst Interval
Press and Hold to Deliver
Emergency Fixed
Burst Pacing
5.Select
If delivery is not confirmed, the programmer displays an error window.
Verify that the programming head is properly positioned. Select
the window and reselect
[BURST Press and Hold].
[OK] from
[BURST Press and Hold].
Figure 2-3. Emergency Fixed Burst Pacing Screen
2-6
GEM®III VR System Reference Guide
Emergency VVI Pacing
To Deliver Emergency VVI Pacing
♦
1.Position the programming head over the ICD.
2.Select [Emergency].
3.Select [VVI Pacing].
4.Select [PROGRAM]. A successful programming sets the ICD to the
following maximum output bradycardia pacing values.
Pacing ModeVVI
Lower Rate70 ppm (70 min-1)
Amplitude8 V
Pulse Width1.6 ms
Pace Blanking320 ms
Post Shock Pace Blanking320 ms
HysteresisOff
Ventricular Rate StabilizationOff
a
Emergency Therapy
Emergency VVI Pacing
a
Reprogrammed by Emergency VVI operation, but not displayed on
Emergency screen.
If programming is not confirmed, verify that the programming head is
properly positioned and select
[Retry] or [Cancel].
GEM®III VR System Reference Guide
2-7
Emergency Therapy
Emergency VVI Pacing
Emergency VVI Pacing
Parameters (fixed)
Select to Program
Emergency VVI Pacing
Figure 2-4. Emergency VVI Pacing Screen
2-8
GEM®III VR System Reference Guide
Overview
3
3
System Description 3-2
Detection and Therapy
Overview 3-4
GEM®III VR System Reference Guide
3-1
Overview
System Description
System Description
The GEM®III VR Model 7231 Implantable Cardioverter Defibrillator (ICD)
System is an implantable medical device system that automatically
detects and treats episodes of ventricular fibrillation (VF), ventricular
tachycardia (VT), fast ventricular tachycardia (FVT), and bradyarrhythmia.
When a cardiac arrhythmia is detected, the implantable device delivers
defibrillation, cardioversion, antitachycardia pacing, or bradycardia pacing
therapy.
The VF Detection circuitry monitors the cardiac cycle length to identify a
VF episode. Upon detection of VF, the ICD provides a biphasic
defibrillation shock of up to 30 joules delivered energy. If the VF episode
persists, up to five more individually programmed defibrillation shocks can
be delivered.
The VT Detection circuitry monitors the cardiac cycle length to identify the
presence of a tachycardia. Upon detection of VT, the ICD delivers its first
programmed VT therapy to restore the patient’s normal cardiac rhythm. If
the VT episode persists, up to five more individually programmed VT
therapies can be delivered. Therapy selections include three methods of
antitachycardia pacing and synchronized cardioversion.
3-2
FVT detection can be programmed through the VT Detection algorithm or
through the VF Detection algorithm. As with VF and VT, up to six FVT
therapies can be independently programmed. Therapy selections include
three methods of antitachycardia pacing and synchronized cardioversion.
The ICD provides rate responsive single chamber bradycardia pacing to
optimize hemodynamics. An internal piezoelectric sensor enables the ICD
to increase and decrease its pacing rate in response to the patient’s
detected physical activity.
The ICD is intended for implantation with transvenous or epicardial
defibrillation leads. The lead system should consist of bipolar or paired
unipolar pacing/sensing leads in the ventricle, and one or two high voltage
cardioversion/ defibrillation electrodes. (The Active Can
serves as one high voltage electrode.) Sensing of cardiac activity and the
delivery of pacing stimuli occur between the ventricular pacing/sensing
electrodes.
The physician can use the ICD to conduct noninvasive electrophysiologic
studies. The physician can also manually deliver any of the ICD therapies
to manage an induced or spontaneous tachyarrhythmia.
GEM®III VR System Reference Guide
®
device case also
Overview
System Description
A programmable Patient Alert™ monitoring feature notifies the patient by
audible tones if certain conditions (low battery voltage, high lead
impedance, etc.), selected by the physician, occur. The patient can then
respond based on instructions provided by the physician.
The ICD provides the physician with real-time information about its
detection and therapy parameters and status. The ICD also provides
accumulated data on its operation, including stored electrograms, records
of tachyarrhythmia episodes detected and treated, bradycardia
interventions, and the efficacy of therapy. All of this information can be
printed and retained in the patient’s file, saved in the programmer to read
back at a later time, or saved in electronic format on a floppy diskette.
GEM®III VR System Reference Guide
3-3
Overview
Detection and Therapy Overview
Detection and Therapy Overview
Implanted sensing electrodes provide a ventricular intracardiac
electrogram to the ICD. The ICD measures the time between events to
determine V-V intervals, and calculates the ventricular rate from these
intervals.
When a V-V interval falls within the programmed VF, VT, or FVT detection
zone, it increments a VF or VT event counter. If either counter reaches the
Number of Intervals to Detect (NID) programmed by the physician, the
ICD detects the arrhythmia and delivers the programmed therapy.
Monitor for next
R-wave
Normal sinus rhythm
Detect depolarization and
Measure the R-R interval
Classify the interval and update
event counters
Detect an arrhythmia episode
No
Ye s
Deliver therapy
Monitor for redetection and
termination
Redetection
After a therapy delivery, if the VF or VT event counter reaches the
programmed “Number of Intervals to Redetect” (RNID), the episode is
redetected and the next programmed therapy is delivered. This
progression continues until the ICD confirms termination of the episode
(page 4-27) or delivers all the therapies programmed for the detected
tachyarrhythmia.
3-4
GEM®III VR System Reference Guide
Overview
Detection and Therapy Overview
Therapy
The ICD provides five ventricular therapies: Burst pacing; Ramp pacing;
Ramp+ pacing; synchronized cardioversion; and defibrillation. Each VF
episode can be treated with up to six defibrillation shocks. Each VT or FVT
episode can be treated with up to six therapies of antitachycardia pacing
and/or cardioversion. An FVT therapy set must end with at least one
cardioversion therapy.
Suspending Detection and Therapy
Automatic Suspension of Detection
Detection is temporarily suspended during:
■
Programming head application (unless [Resume] is programmed) or
other magnet application. Detection resumes automatically when the
head or magnet is removed.
■
Temporary programmer operations such as manual and emergency
therapies delivered through the programmer. Detection remains
suspended until the programming head is removed or
programmed.
[Resume] is
Note: If the Resume at Deliver feature is disabled
and
detection was
suspended before a temporary induction is performed, then detection
remains suspended.
■
Automatic daily lead impedance and battery voltage measurements
(see page 10-33). Detection resumes when the measurement is
complete.
■
Automatic tachyarrhythmia therapy deliveries, including charging
periods for high voltage therapy and automatic capacitor formations.
Detection resumes automatically when the therapy or charging
period is complete.
VT Detection is temporarily suspended during the first 17 ventricular
events (paced and sensed) after an automatic ventricular defibrillation
therapy.
GEM®III VR System Reference Guide
3-5
Overview
Detection and Therapy Overview
Manually Suspending Detection and Therapy
ICD detection and therapy capabilities can be temporarily suspended by:
■
positioning the programming head, the Patient Alert™ patient
magnet, or another magnet over the ICD.
■
executing a [Suspend] command from the programmer.
The bradycardia pacing and manually-initiated therapy capabilities of the
ICD are still functional. Suspending detection does not change the
programmed settings of the ICD.
Removing the programming head or magnet from its position over the ICD
restores the ICD to normal operation. If automatic detection is enabled,
the
[Resume] button causes the ICD to resume automatic detection even
while the programming head is in position over the ICD.
Disabling Detection and Therapies
Automatically Disabling Detection and Therapies
The ICD turns off VT Detection if a power-on reset (POR) occurs. The
programmable parameter tables starting on page 1-13 list the POR
values for all programmable parameters.
3-6
The ICD automatically turns off all VT and VF automatic therapies if three
consecutive 30-second charges have failed to reach the programmed
energy level (see “Charge Circuit Inactive” on page 5-4).
Manually Disabling Detection and Therapies
To turn off the automatic detection and therapy functions of the ICD,
program the functions
1.Select the Params icon and choose the Detection option.
2.Set the Enable parameters to Off.
3.Position the programming head and select [PROGRAM].
To re-enable automatic detection and therapy, reprogramming is required.
GEM®III VR System Reference Guide
Off via the programmer.
Tachyarrhythmia
Detection
4
4
&
Sensing 4-2
VF Detection 4-5
VT Detection 4-8
Combined Count Detection 4-11
VT Stability Option 4-13
EGM Width Criterion 4-15
FVT Detection 4-21
Termination and Redetection 4-26
GEM®III VR System Reference Guide
4-1
Tachyarrhythmia Detection
Sensing
Sensing
Proper sensing is essential for the safe and effective use of the ICD.
To provide appropriate sensing, the ICD uses auto-adjusting sensing
thresholds after paced and sensed events (see page 4-3).
Programming for Appropriate Sensing
The programmable sensitivity threshold defines the minimum electrical
amplitude that the ICD recognizes as a ventricular event.
A combination of high pacing pulse width and/or amplitude with a low
sensitivity threshold may cause inappropriate sensing after pacing.
Programming a lower pulse width, lower amplitude, longer pace blanking,
or a higher threshold may eliminate this inappropriate sensing.
Ventricular Sensing
A ventricular sensitivity threshold of 0.3 mV is recommended to maximize
the probability that VF will be detected.
■
Always verify that the ICD is sensing and detecting VF properly. You
may need to reposition or replace the ventricular sensing lead.
■
A value greater than 0.6 mV is not recommended except for testing
and troubleshooting, as it may lead to undersensing
or underdetection.
■
Carefully evaluate the potential for increased susceptibility to EMI
and oversensing before changing the ventricular sensitivity threshold
to its most sensitive setting of 0.15 mV.
■
If you change the ventricular sensitivity threshold, evaluate for proper
sensing and detection by inducing VF and allowing the ICD to
automatically detect and treat the arrhythmia.
■
If the ICD is not sensing or detecting properly, program all automatic
therapies to
continuously for life-threatening arrhythmias until you program
automatic therapies
Off and evaluate the ICD system. Monitor the patient
On again.
4-2
GEM®III VR System Reference Guide
Tachyarrhythmia Detection
Sensing
Auto-Adjusting Sensitivity Thresholds
The ICD automatically adjusts its sensitivity threshold following certain
paced and sensed events (Figure 4-1) to reduce the incidence of T-wave
sensing. The auto-adjusting thresholds increase dramatically (Table 4-1),
and then gradually return to their programmed settings. Note that the
thresholds may already differ from their programmed values, having been
adjusted by the preceding sensed or paced event.
The exponential decay continues through a subsequent ventricular pacing pulse and its
blanking period.
c
Up to a maximum of 1.8 mV.
d
Until the preceding 450 ms decay, or the programmed threshold, is restored.
Increased to 75% of EGM peak.
450 ms decay constant
b
Increased to 4.5x the programmed value.
450 ms decay constant.
a
c
d
Ventricular EGM
Programmed
Sensitivity Threshold
Programmed
Post-Pace Blanking
Figure 4-1. Auto-Adjusting Sensitivity Threshold
4.5x
Changing Threshold
Post-PacePost-Sense
8x
V
P
V
P
V
S
GEM®III VR System Reference Guide
V
P
V
S
4-3
Tachyarrhythmia Detection
Sensing
Blanking Periods
During a blanking period the ICD does not sense electrical signals.
Blanking periods avoid sensing ICD outputs, post-pacing polarization,
T-waves, and multiple sensing of the same event. The blanking periods
following paced events are longer than those following sensed events to
avoid sensing the depolarization signal on the electrodes.
Table 4-2. Programmable Blanking Periods
Blanking after ventricular bradycardia pace150 – 440 ms
Pace blanking during high voltage therapy150 – 440 ms
Blanking after antitachycardia pace150 – 440 ms
Table 4-3. Non-Programmable Blanking Periods
Blanking after sensed ventricular event120 ms
Blanking after the end of a cardioversion or defibrillation
charging period
Blanking after delivered cardioversion or defibrillation therapy520 ms
300 ms
Refractory Periods
4-4
Synchronization refractory periods help prevent delivery of pacing or
shock pulses during the ventricular vulnerable period. (Refer to “High
Voltage Synchronization” on page 5-5 for more detail.)
Note: Refractory periods do not affect tachyarrhythmia detection.
Ventricular refractory period after charging ends
during defibrillation or cardioversion synchronization
Ventricular refractory period after ventricular sense
during cardioversion synchronization
GEM®III VR System Reference Guide
Table 4-4. Refractory Periods
400 ms
200 ms
VF Detection
Tachyarrhythmia Detection
VF Detection
The ICD detects VF by applying the VF Interval Criterion. If the ICD
senses the programmed number of VF intervals, it detects a VF episode
and delivers a VF therapy.
VF Detection details begin on page 4-7.
Table 4-5. VF Detection Programmable Parameters
VF EnableOn or Off
Sensitivity (mV)
VF Interval (Rate)V-V intervals shorter than the programmed interval
VF Initial NIDNumber of Intervals to Detect:
VF Redetect NIDNumber of Intervals to Redetect:
a
The programmed sensitivity value applies to ventricular tachyarrhythmia detection and
bradycardia pacing.
a
Minimum level of electrical signal that registers as a
sensed ventricular event.
are counted as VF events.
(Rates are shown in parentheses.)
number of VF events the ICD must count to detect a
VF episode.
number of VF events the ICD must count to redetect
VF after a therapy.
Programming VF Detection
To ensure proper VF detection, you should not program the VF Interval
below 300 ms.
Programming the ventricular sensitivity to 0.3 mV is recommended. A
value greater than 0.6 mV is not recommended except for testing and
troubleshooting. Also, carefully evaluate the possibility of increased
susceptibility to EMI and oversensing before changing the sensitivity to its
minimum (most sensitive) setting of 0.15 mV.
You can accelerate redetection by programming the Redetect NID lower
than the Initial NID.
When VF Detection is enabled for the first time (i.e., at implant) the ICD
automatically changes the Automatic Capacitor Formation Interval from
Off to 6 month.
GEM®III VR System Reference Guide
4-5
Tachyarrhythmia Detection
VF Detection
♦
VF Detection cannot be programmed Off unless VT Detection is also
programmed
Off.
To Program VF Detection
1.Select the Params icon and choose the Detection option. The
programmer displays the Detection screen (shown below).
2.Tu r n
3.Position the programming head and select [PROGRAM].
VF Enable to On and select VF Detection parameters.
Figure 4-2. Detection Screen Example
4-6
GEM®III VR System Reference Guide
Tachyarrhythmia Detection
VF Detection
How VF Detection Works
VF Interval Criterion
When a V-V interval is shorter than the programmed VF Interval, the ICD
marks that event as a VF event and tallies it in memory. If the number of
VF events reaches the programmed
episode and delivers the first programmed VF therapy. Following the
therapy, if the number of VF events reaches the
redetects VF and delivers the next programmed VF therapy. (See also
“Combined Count Detection” on page 4-11.) VF NIDs are defined as a
number of VF events within the
last sixteen intervals). All NID values are 75% of the detection window:
12/16, 18/24, etc.
VF Intervals (< 320 ms)
VF Initial NID, the ICD detects a VF
VF Redetect NID, the ICD
VF detection window
(e.g., twelve of the
A VF episode is declared
when the VF counter
reaches 12
(of the last 16 events).
V
S
VFOn12/169/12320 ms
VTOn128400 ms
Sensitivity: 0.3 (mV)
V
S
EnableInitialRedetectInterval
V
S
VF Event Counter:
V
FSFSFSFSFSFSFSFSFSFSF
S
123456789101112
Figure 4-3. VF Detection Example
GEM®III VR System Reference Guide
F
S
D
200 ms
|
4-7
Tachyarrhythmia Detection
VT Detection
VT Detection
The ICD detects VT by applying the VT Interval Criterion. If the ICD
senses the programmed number of VT intervals, it detects a VT episode
and delivers a VT therapy.
VT Detection details begin on page 4-10.
Table 4-6. VT Detection Programmable Parameters
VT EnableOn or Off
Sensitivity (mV)
VT Interval (Rate)V-V intervals shorter than the programmed interval
VT Initial NIDNumber of Intervals to Detect:
VT Redetect NIDNumber of Intervals to Redetect:
a
The programmed Ventricular Sensitivity value applies to ventricular tachyarrhythmia
detection and bradycardia pacing.
a
Minimum level of electrical signal that registers as a
sensed event.
are counted as VT events.
(Rates are shown in parentheses.)
number of VT events the ICD must count to detect a
VT episode.
number of VT events the ICD must count to redetect
VT after a therapy.
4-8
Programming VT Detection
To ensure proper VT detection, program the VT Interval to a value that is at
least 40 ms longer than the patient‘s tachycardia interval, and at least
40 ms longer than the programmed
Programming the ventricular sensitivity to 0.3 mV is recommended. (See
“Programming for Appropriate Sensing” on page 4-2).
You can accelerate redetection by programming the Redetect NID lower
than the Initial NID.
To ensure VF Detection backup during VT episodes, VT Detection cannot
be
On unless VF Detection is also On.
GEM®III VR System Reference Guide
VF Interval.
Tachyarrhythmia Detection
To Program VT Detection
♦
1.Select the Params icon and choose the Detection option. The
programmer displays the Detection screen (shown below).
VT Detection
2.Tu r n
3.Position the programming head and select [PROGRAM].
VT Enable to On and select VT Detection parameters.
Figure 4-4. Detection Screen Example
GEM®III VR System Reference Guide
4-9
Tachyarrhythmia Detection
VT Detection
How VT Detection Works
VT Interval Criterion
When the V-V interval is shorter than the programmed VT Interval, the ICD
marks that event as a VT event and tallies it in memory. (See also “VT
Stability Option” on page 4-13). If the number of VT events reaches the
programmed
first programmed VT therapy (See also “EGM Width Criterion” on
page 4-15).
VT Initial NID, the ICD detects a VT episode and delivers the
After the therapy, if the number of VT events reaches the
the ICD redetects VT and delivers the next programmed VT therapy. (See
also “Combined Count Detection” on page 4-11.) The ICD counts
consecutive VT events. The event counter resets to zero whenever an
interval fails to satisfy the Interval criterion, or the VT Stability option, if it
is
On. The VT counter remains at its current value if an interval is shorter
than the programmed
VT intervals > 320 and < 400 ms
V
S
VFOn12/169/12320 ms
VTOn128400 ms
Sensitivity (mV): 0.3
V
S
EnableInitialRedetectInterval
V
S
VT Counter Value:
VSV
S
T
TSTSTSTSTSTSTSTSTST
S
123456789101112
VT Redetect NID,
VF Interval.
A VT episode is declared
when the VT counter
reaches 12.
T
|
D
S
200 ms
A VT event increases the VT counter by one.
A normal sensed beat (≥ 400 ms) clears the
counter to zero.
A VF event suspends, but doesn’t clear, the
VT count.
4-10
Figure 4-5. VT Detection Example
GEM®III VR System Reference Guide
Combined Count Detection
Combined Count detection helps the ICD promptly detect or redetect a
ventricular tachyarrhythmia when the interval is fluctuating between the
VF and VT zones. The ICD tracks the combined number of events
registered on the VF and VT counters, and if this sum reaches the
Combined Number of Intervals to Detect (CNID), VF or VT is detected.
Combined Count Detection also applies to redetected episodes.
Combined Count detection is automatic (not programmable) when VT
Detection is enabled, and cannot be disabled.
How Combined Count Detection Works
Combined Count Detection automatically applies a Combined Number of
Intervals to Detect (CNID) when the VF event counter reaches six. The
ICD calculates the CNID as follows:
Initial CNID = VF Initial NID x 7/6 (rounded down)
Redetect CNID = VF Redetect NID x 7/6 (rounded down)
Tachyarrhythmia Detection
Combined Count Detection
For example, if the VF Initial NID is programmed to 18, the Initial CNID is
21. If the VF Redetect NID is 12, the Redetect CNID is 14.
Combined Count detection is fulfilled when the sum of the VF and VT
event counters equals or exceeds the CNID. The ICD then reviews the
eight intervals preceding detection:
■
If
any
of the eight was in the VF zone, the episode is classified as VF.
■
If
all
eight were outside the VF zone, the episode is classified as VT.
GEM®III VR System Reference Guide
4-11
Tachyarrhythmia Detection
Combined Count Detection
Surface ECG
Marker Channel™
VF Event Count
VT Event Count
Without Combined Count Detection, the
arrhythmia would be detected after 18 beats
15
VF = 12
10
VF
Counter
5
VT CounterVT Counter
Figure 4-6. Combined Count Detection Example
T
T
S
S
12
F
FSF
12 3 45
FSF
S
S
S
VF NID = 12/16, VT NID = 12, CNID = 14
With Combined Count Detection, the arrhythmia
is detected after 14 beats
15
10
VF
VT = 12
15105
Counter
5
T
S
S
67
3
VF = 12
T
S
4
FSFSF
89
D
10
FSF
Combined Count (VT+VF) = 14
VF = 6
VT = 12
15105
4-12
GEM®III VR System Reference Guide
VT Stability Option
VT Stability is an optional detection criterion that allows the ICD to
withhold VT detection for rapid rhythms (in the VT detection zone) with
irregular intervals.
VT Stability (ms)Off, or the maximum acceptable variation among
Programming VT Stability Option
Choosing a small VT Stability value (such as 30 ms) may not allow for the
normal VT interval variation. Thus, smaller VT Stability values may
decrease the sensitivity of the ICD to detect VT.
The VT Stability option compares measured intervals that have been
shortened to multiples of 10 ms, permitting a rounding error of up to
+10 ms from the programmed VT Stability value.
To Program VT Stability
♦
1.Select the Params icon and choose the Detection option. The
3.Position the programming head and select [PROGRAM].
VT Stability.
GEM®III VR System Reference Guide
4-13
Tachyarrhythmia Detection
VT Stability Option
How the VT Stability Option Works
When the VT Stability option is On, an interval in the VT zone is
considered
previous intervals is greater than the programmed VT Stability interval.
The ICD does not apply the VT Stability option until the VT event count
reaches three. When an interval in the VT zone fails to meet the VT
Stability option, it is marked as a normal sensed event and the VT event
count resets to zero.
The VT Stability option must be re-satisfied during redetection if it
is enabled.
unstable
if the difference between it and any of the three
VT Stability is not applied until VT count reaches three.
Stable events: Each differs from the three before it by
50 ms or less, and they are marked as VT events.
T
T
T
S
5
4
1
5
0
0
V
S
S
4
4
4
7
0
0
2 301232345
T
T
S
4
8
0
S
S
4
5
4
0
0
0
TST
V
S
4
4
7
8
0
0
T
S
4
7
0
TST
S
4
4
7
7
0
0
01
200 ms
Unstable event: Varies from previous interval
by 60 ms. VT counter resets to zero.
S
4-14
Figure 4-7. VT Stability Option
GEM®III VR System Reference Guide
EGM Width Criterion
The optional EGM Width Criterion can improve the specificity of VT
detection by identifying inappropriately detected supraventricular
arrhythmias (such as sinus tachycardia and atrial fibrillation or flutter). The
ICD measures the QRS complexes during a high rate episode, and
detects the episode as VT only if a sufficient number of the QRS
complexes are wide.
For more detail on the EGM width criterion, see page 4-18.
EGM Width
Width Threshold (ms)
Slew Threshold (mV/s)
Note: The ICD provides three diagnostic tools to evaluate and verify the
EGM Width parameter settings:
–The EGM Width test (page 8-9) helps you find the most effective
–QRS Snapshots (page 10-25) show how the parameters
A sensed event is designated wide if its QRS
complex equals or exceeds this width.
Defines the beginning and end of each sensed
QRS complex, using the rate of amplitude change
in the EGM 2 waveform.
parameter settings by measuring the patient’s QRS complexes
during sinus rhythm.
settings are being applied during automatic detection.
monitoring of the ICD’s ability to discriminate between VT and
SVTs.
GEM®III VR System Reference Guide
4-15
Tachyarrhythmia Detection
EGM Width Criterion
Programming the EGM Width Criterion
#
Cautions: The EGM Width criterion should not be used for patients
with fast or poorly tolerated VTs.
The EGM Width criterion is less likely to be effective in patients with
narrow complex VTs, with bundle branch block, or whose QRS
complexes are affected by antiarrhythmic medications.
When the EGM Width Criterion is programmed to
measures and stores QRS widths during VT episodes, but the criterion is
not applied for VT detection.
Select the slew threshold independently of the width threshold to establish
proper QRS width measurements during sinus rhythm. Recommended
initial threshold settings are:
–the lowest slew threshold that provides predictable, consistent
width measurements (variation of 12 ms or less) during sinus
rhythm.
–the lowest width threshold that exceeds all the width
measurements taken during sinus rhythm.
Higher slew thresholds tend to result in narrower and possibly more
consistent QRS width measurements. Lower slew thresholds tend to
provide wider and possibly more variable QRS width measurements (see
the drawing on page 4-20).
A larger width threshold increases the specificity of VT detection, but may
also decrease the sensitivity. While evaluating patients with a concomitant
pacemaker, measure the intrinsic QRS widths during normal sinus
rhythm.
Passive, the ICD
4-16
The width threshold and slew threshold parameters can also be
programmed from the EGM Width Test screen (when the criterion is set
to
On or Passive).
GEM®III VR System Reference Guide
Tachyarrhythmia Detection
To Program the EGM Width Criterion
♦
1.Select the Params icon and choose the Detection option. The
programmer displays the Detection screen.
EGM Width Criterion
EGM Width Criterion
Parameters
2.Set
3.Position the programming head and select [PROGRAM].
EGM Width to On or Passive and select values for the Width
Threshold and Slew Threshold.
Figure 4-8. EGM Width Criterion Parameters
GEM®III VR System Reference Guide
4-17
Tachyarrhythmia Detection
EGM Width Criterion
How the EGM Width Criterion Works
The EGM Width Criterion is based on the premise that QRS complexes of
ventricular origin are generally wider than those of supraventricular origin.
When a fast ventricular rhythm fulfills the other programmed VT detection
criteria and the EGM Width Criterion is programmed
the eight preceding QRS complexes, as sensed on the programmed
EGM 2 electrode source (see Figure 4-10). If fewer than six of these eight
complexes in the “lookback window” are wide, the ICD does not identify
the episode as VT.
The EGM Width Criterion is only applied on initial detection of a VT
episode. It does not apply to VF detection, FVT detection, or redetection
of VT after therapy.
Measuring a QRS Complex
The ICD identifies the start and end points of a sensed QRS complex
using the slew threshold parameter. Slew rate is the
millivolts per second) of the sensed EGM amplitude.
On, the ICD reviews
rate of change
(in
4-18
Low rate of
amplitude change
High rate of
amplitude change
Higher slew thresholds give narrower QRS widths. Lower
slew thresholds give wider QRS widths. Select the lowest
slew threshold that provides predictable, stable QRS width
measurements of the patient’s normal sinus rhythm.
Figure 4-9. Effect of Slew Threshold on QRS Width (Example)
The ICD measures the slew rate every 4 milliseconds, and marks the start
of the QRS at the earliest point where the slew rate first reaches the
programmed threshold (Figure 4-9). The end of the QRS occurs at the
last point where slew rate exceeds the programmed slew threshold. The
QRS width is then calculated as the time in milliseconds between the start
and end of the QRS.
GEM®III VR System Reference Guide
48 mV/s slew threshold results in 104 ms QRS width
90 mV/s slew threshold results in 84 ms QRS width
104 ms QRS width
48 mV/s slew threshold
90 mV/s slew threshold
84 ms QRS width
Bipolar EGM
Tachyarrhythmia Detection
EGM Width Criterion
Classifying the QRS Complex
If the width of the QRS is less than the programmed width threshold, it is
designated as
or if it is less than 16 milliseconds (i.e., too narrow to represent a normal
sinus rhythm), it is designated as
Far Field EGM
Narrow. If the width equals or exceeds the width threshold,
Wide.
QRS measurements are made from the programmed
EGM source for EGM channel 2
V
S
VT Counter Value:
QRS Width in ms:
V
S
T
T
S
S
2
1
T
T
S
S
3
4
80 6896100 88 96 104 88
EGM Width:ON
Width Threshold (ms):88
Slew Threshold (mV/s):30
VF FVT VT
Detection Status:ONOFFON
Interval (ms):320500
Initial NID:18/2412
Figure 4-10. Example of EGM Width Criterion Met
Classifying the Episode
When a ventricular high rate fulfills the Initial (or Combined) NID, the ICD
“looks back” at the eight complexes preceding, but not including, the event
that fulfilled detection.
■
If six or more of these eight QRSs are wide, the EGM Width criterion
is satisfied and the ICD detects the episode as VT (Figure 4-10).
■
If three or more are narrow, the criterion is not satisfied and the VT
counter resets to zero. (Figure 4-11: the criterion is not met because
the slew threshold is set too high.)
T
T
T
S
S
5
6
T
S
S
7
8
T
T
S
S
10
9
(8 events not including NID event)
84
When the VT NID is reached, the
device reviews the recent QRS
measurements.
Six of the eight events were
measured “wide” (≥ 88 ms), and the
episode is detected as VT.
T
T
|
D
S
12
11
200 ms
GEM®III VR System Reference Guide
4-19
Tachyarrhythmia Detection
EGM Width Criterion
VT beats are measured narrower than the width threshold.
Far Field EGM
Bipolar EGM
V
S
VT Counter Value:
VT Counter Value:
QRS Width in ms:
QRS Width in ms:
EGM Width:ON
Width Threshold (ms):88
Slew Threshold (mV/s):48
Detection Status:ONOFFON
Interval (ms):320500
Initial NID:18/2412
T
V
S
T
S
1
T
S
S
2
3
76 7684928480 88 84
VF FVT VT
T
S
4
With the slew threshold set to 48 mV/s, several
T
T
T
T
T
S
S
S
S
5
7
6
8
TST
S
10
9
(8 events not including NID event)
76
V
S
S
11 (12)0
200 ms
When the episode fails to meet the
EGM Width criterion, the VT counter
Slew = 48 mV/s
84 ms width
resets to zero.
4-20
Figure 4-11. Effect of Slew Threshold on Rhythm Classification
GEM®III VR System Reference Guide
FVT Detection
Tachyarrhythmia Detection
FVT Detection
The ICD can be programmed to further classify a tachycardia as Fast
Ventricular Tachycardia (FVT). The ICD provides an FVT episode counter
and a separate therapy regimen for FVT episodes.
FVT Detection details begin on page 4-23.
Table 4-9. FVT Detection Programmable Parameters
FVT Enablevia VF, via VT, or Off
Ventricular Sensitivity
a
(mV)
FVT Interval (Rate)V-V intervals between this value and the
a
The programmed sensitivity value applies to tachyarrhythmia detection and
bradycardia pacing.
Minimum level of electrical signal that registers as a
sensed ventricular event.
programmed VF Interval are marked as FVT
events. (Rates are shown in parentheses.)
Programming FVT Detection
To ensure VF Detection backup, FVT Detection cannot be On unless VF
Detection is also
On.
For FVT Detection
zone. For FVT Detection
via VF, the FVT Interval is programmed within the VF
via VT, the FVT Interval is within the VT zone. This
choice should depend on the patient’s VF and VT cycle lengths. After
determining a reliably sensitive VF Interval, consider the following:
■
If the patient exhibits a clinical VT cycle length in the VF zone, select
via VF to ensure reliable detection of VF. (VT Detection need not be
enabled at all.)
■
If the patient exhibits two clinical VTs, both outside the VF zone,
select
via VT to prevent misclassification of VF, and to offer a separate
therapy regimen for each VT.
■
If the patient has only one clinical VT, outside the VF zone, select VF
and VT Detection only, with
FVT Enable set to Off.
Select the VF, VT, and FVT Intervals at least 40 ms apart to allow for
normal variations in the patient’s interval.
FVT Detection uses the VF event counter and the programmed VF NID if
set to
via VF, or the VT counter and VT NID if set to via VT.
GEM®III VR System Reference Guide
4-21
Tachyarrhythmia Detection
FVT Detection
♦
To Program FVT Detection
1.Select the Params icon and choose the Detection option. The
programmer displays the Detection screen.
2.Set
3.Position the programming head and select [PROGRAM].
FVT Enable to via VT or via VF. Select the FVT Interval and
Ventricular Sensitivity values.
Figure 4-12. Programming FVT Detection
4-22
GEM®III VR System Reference Guide
Tachyarrhythmia Detection
FVT Detection
How FVT Detection Works
FVT Detection Via VF
When you set FVT Detection to via VF, the ICD detects FVT episodes
using the VF event counter.
A V-V interval within the FVT zone is marked as an “FVT via VF” event and
tallied on the more sensitive VF event counter. When the VF NID is
reached, the ICD reviews the last eight intervals.
■
If
any
of the last eight intervals is in the VF zone, it detects the
episode as VF.
■
If
all
of the last eight intervals are outside the VF zone, it detects the
episode as FVT (Figure 4-13).
FVT Detection Via VT
When you set FVT Detection to via VT, the ICD detects FVT episodes
using the VT event counter.
A V-V interval within the FVT zone is marked as an “FVT via VT” event and
tallied on the more specific, consecutive VT event counter. The VT
Stability option must also be satisfied if it is enabled. When the VT NID is
reached, the ICD reviews the last eight intervals.
■
If
any
of the last eight intervals is in the VF or FVT zones, it detects
the episode as FVT.
■
If
all
of the last eight intervals are outside the FVT and VF zones, it
detects the episode as VT.
GEM®III VR System Reference Guide
4-23
Tachyarrhythmia Detection
FVT Detection
VF NID reached. Since none of the last eight intervals is
a VF interval, the ICD classifies the episode as FVT and
Combined Count Detection of FVT
Combined Count Detection (page 4-11) also applies to the FVT zone. If
the VF event counter reaches six, and the sum of the VF and VT event
counters equals or exceeds the CNID, the ICD “looks back” at the eight
intervals preceding detection:
■
If any of the eight was in the VF zone, the episode is detected as VF.
■
If none were in the VF zone but one or more was in the FVT zone, the
episode is detected as FVT.
■
If all eight were outside the VF and FVT zones, the episode is
classified as VT.
will deliver an FVT therapy.
4-24
V
S
V
S
V
S
Enab
le
V
S
VF Counter:
V
T
S
F
T
T
.
.
F
F
T
T
.
F
T
.
.
F
F
123 4 567 89101112
LOOKBACK WINDOW
(8 INTERVALS BEFORE NID)
Initial
Redetect
VFOn12/169/12320 ms
FVTvia VF260 ms
VTOff
Figure 4-13. FVT Detection Via VF Example
GEM®III VR System Reference Guide
T
T
T
.
.
F
F
Inter
T
.
.
.
F
F
v
al
T
.
F
200 ms
T
|
F
Blanking
Depolarization
Blanking
Depolarization
VF Detection Zone
120 ms
120 ms
Tachyarrhythmia Detection
FVT Detection
Zone Merging After Detection
For FVT via VF detection, the FVT zone merges with the VF zone after
detection or redetection of VF. For FVT via VT detection, the FVT zone
merges with the VT zone after detection or redetection of FVT or VF. In
both instances, the merged-zone configuration remains until episode
termination. The merged zone uses the event counting and therapies for
the faster arrhythmia, to ensure that the ICD delivers sufficiently
aggressive therapies during an extended or highly variable
tachyarrhythmia episode. (See also “Progressive Episode Therapies” on
page 5-33.)
Before initial detection: three detection zones
VT
Interval
VT
Interval
Lower Rate
Period
Lower Rate
Period
FVT Detection Zone
FVT
Interval
Interval
VT Detection Zone
VF
After detection of VF: two zones for redetection
VF Detection ZoneVT Detection Zone
VF
Interval
Blanking
Depolarization
Blanking
Depolarization
VF Detection Zone
120 ms
VF Detection Zone
120 ms
Figure 4-14. Zone Merging (FVT Detection Via VF)
Before initial detection: three detection zones
FVT Detection Zone
VF IntervalFVT
VT Detection Zone
Interval
After detection of VF or FVT: two zones for redetection
FVT Detection Zone
VF
Interval
Figure 4-15. Zone Merging (FVT Detection Via VT)
GEM®III VR System Reference Guide
Lower Rate
Period
VT
Interval
Lower Rate
Period
VT
Interval
4-25
Tachyarrhythmia Detection
Termination and Redetection
Termination and Redetection
Outcome Monitoring
After delivering an automatic therapy, the ICD monitors the cardiac
interval for three possible outcomes:
■
Termination of the episode.
■
Redetection of the original ventricular tachyarrhythmia.
■
Redetection of a different ventricular arrhythmia, including
1
VT acceleration.
Further automatic ventricular therapies are not delivered unless the
patient‘s cardiac rhythm fulfills the redetection requirements for a
ventricular tachyarrhythmia.
The EGM Width Criterion is not applied during redetection. However, the
VT Stability criterion must be continually re-satisfied throughout
redetection if it is enabled.
Programming Considerations
You can accelerate redetection by programming Redetect NIDs lower
than the Initial NIDs.
Both the VF Redetect NID and the VT Redetect NID take effect during any
type of ventricular arrhythmia episode.
If you program the
page 5-33), a slower arrhythmia (e.g., VT) may receive therapies for a
faster arrhythmia (e.g., VF) during redetection.
Progressive Episode Therapies option On (see
4-26
1
VT Detection is temporarily suspended for 17 events, paced or sensed, following
defibrillation therapy.
GEM®III VR System Reference Guide
Tachyarrhythmia Detection
Termination and Redetection
Therapy Efficacy
A therapy is considered successful if arrhythmia termination occurs prior
to any redetection. The therapy is considered unsuccessful if redetection
occurs before termination.
If all programmed therapies are unsuccessful, the ICD suspends them
until either:
■
the episode terminates, or
■
Detection is manually reset (by programming Detection Enable Off
and then On again) during the episode.
Episode Termination
Episode termination occurs when eight consecutive V-V intervals are
greater than or equal to the VT detection interval.
1
Once the ICD detects
an arrhythmia, it considers the episode as ongoing until it detects episode
termination. Any subsequent detection after termination marks the start
of a new episode.
After antitachycardia pacing therapy, the counter for episode termination
begins with the first ventricular cycle. After cardioversion or defibrillation,
the counter for episode termination begins with the second ventricular
event following the shock. (Due to the extended post shock blanking, this
event may be the
third
event on the electrogram.)
A rhythm that does not satisfy the termination definition is considered part
of an ongoing episode, but the ICD does not deliver the next programmed
therapy until the redetection criteria are fulfilled.
Redetection
The ICD redetects an arrhythmia if the VF or VT event counter reaches its
Redetect NID, or if the combined VF and VT event counters reach the
Redetect CNID (see “Combined Count Detection” on page 4-11). The
ICD then delivers the next programmed therapy for the current arrhythmia,
and resumes monitoring for the outcome of that therapy.
1
VF Interval when VT Detection is programmed Off.
GEM®III VR System Reference Guide
4-27
Tachyarrhythmia Detection
Termination and Redetection
If the ICD redetects VF or an accelerated VT after an antitachycardia
pacing sequence delivery, it skips the subsequent pacing therapy
sequences for the duration of this episode and proceeds to deliver the
next therapy programmed for the current arrhythmia.
Figure 4-16 illustrates redetection of VT after cardioversion therapy.
VT Acceleration
The ICD classifies a VT as accelerated when it redetects during a VT
episode and the average of the four intervals before redetection has
decreased by 60 ms or more, compared to the average of the four
intervals before the initial VT detection.
to evaluate future VT accelerations under the 60 ms rule above.
If VT acceleration occurs after the delivery of an antitachycardia pacing
sequence, the ICD skips the subsequent sequences of the pacing therapy
for the duration of this episode and delivers the next programmed
VT therapy.
1
The current redetection is used
4-28
1
Or before the most recent “accelerated” VT redetection.
GEM®III VR System Reference Guide
0.6 joule cardioversion shock.
Not sensed in the blanking period
after cardioversion.
Tachyarrhythmia Detection
Termination and Redetection
Sensed, measured as 660 ms, and marked
as a normal sensed event.
C
T
|
D
VRV
E
C
S
D
V
T
S
3
3
0
T
S
S
3
3
3
3
0
0
After eight consecutive intervals within the VT zone, the VT
Redetect NID is fulfilled and the ICD redetects the VT episode.
EnableInitialRedetectInterval
VFOn12/169/12320 ms
VTOn128400 ms
VT Therapy: 123456
Therapy Status:OnOnOnOnOnOn
Therapy Type:CVCVCVCVCVCV
Energy (J):0.61030303030
Figure 4-16. Redetection of VT After Therapy
T
T
S
3
3
0
TSTST
3
3
3
3
0
0
S
3
3
0
The therapy was unsuccessful:
the ICD redetected VT before
episode termination (eight
consecutive beats outside the
detection zones).
T
|
D
S
3
3
0
200 ms
GEM®III VR System Reference Guide
4-29
Tachyarrhythmia Detection
Termination and Redetection
4-30
GEM®III VR System Reference Guide
Tachyarrhythmia
Therapies
5
5
High Voltage Therapy Overview 5-2
Defibrillation 5-9
Cardioversion 5-15
Antitachycardia Pacing
Therapy Overview 5-20
Programming ATP Therapies 5-22
Burst Pacing 5-24
Ramp Pacing 5-26
Ramp+ Pacing 5-28
Smart Mode 5-30
Progressive Episode
Therapies 5-33
GEM®III VR System Reference Guide
5-1
Tachyarrhythmia Therapies
High Voltage Therapy Overview
High Voltage Therapy Overview
Defibrillation and cardioversion are the two high voltage therapies
provided by the ICD. Up to six defibrillation shocks may be programmed
to treat a detected episode of VF. Cardioversion therapy can be
programmed as any of the six VT or FVT therapies.
When an arrhythmia is detected and defibrillation or cardioversion is the
next programmed therapy, the ICD charges the high voltage capacitors to
the programmed energy. When the programmed energy is reached, the
ICD initiates a synchronization sequence and attempts to synchronize the
pulse to the leading edge of a sensed ventricular event.
Defibrillation and cardioversion therapies can be delivered via the
following high voltage electrodes:
■
CAN (HVA) – device can
■
RV (HVB) – RV coil
■
SVC (HVX) – optional supplemental electrode; e.g. SVC coil
Each defibrillation or cardioversion therapy has separately programmable
energy and pathway polarity (see “High Voltage Therapy Parameters” on
page 5-7). The ICD automatically regulates the pulse width to obtain
50% tilt.
5-2
Comparison of Defibrillation and
Cardioversion
A defibrillation shock is delivered for VF; a cardioversion shock is
delivered for VT or FVT.
When the ICD delivers a defibrillation therapy, it attempts to synchronize
with a sensed R-wave, but does not require such synchronization.
However, a cardioversion therapy requires synchronization to a sensed
R-wave for delivery to occur.
GEM®III VR System Reference Guide
Tachyarrhythmia Therapies
High Voltage Therapy Overview
Committed vs. Non-Committed Therapies – The ICD can be
programmed to reconfirm VF before delivering the first programmed
defibrillation therapy. Reconfirmation is attempted only for the initial
detection of VF, and only if the Reconfirm VF feature is programmed
Subsequent VF therapies within the same episode are always
On.
committed
when the capacitors reach their programmed energy, the shock will be
delivered. Cardioversion is a
non-committed
therapy: if the ICD cannot
synchronize to a ventricular sensed event, the therapy aborts.
Table 5-1 summarizes the differences between defibrillation
and cardioversion.
Table 5-1. Comparison of Defibrillation and Cardioversion
DefibrillationCardioversion
Episode detectedVFVT or FVT
Reconfirm arrhythmia First therapy: optional
Later therapies: committed
SynchronizationNot required unless
Synchronization
successful
Synchronization
unsuccessful
Reconfirm VF is active
Delivers high voltage pulse Delivers high voltage pulse
Delivers asynchronous
defibrillation pulse at end of
synchronization sequence
(but if Reconfirm VF is
active, the therapy aborts)
All therapies non-committed
(require synchronization)
Required
Abort CV pulse at end of
synchronization sequence
:
GEM®III VR System Reference Guide
5-3
Tachyarrhythmia Therapies
High Voltage Therapy Overview
Charging Period
w
Warning: If a Charge Circuit Timeout (see below) occurs, please inform
a Medtronic representative. Immediate replacement is recommended.
The length of time to charge the ICD high voltage capacitors to the
programmed energy (charge time) increases with an increase in
programmed energy, battery depletion, or time since the last capacitor
formation. Table 1-5 on page 1-8 shows typical charge times to maximum
programmed energy at different battery voltages and levels of capacitor
formation.
During the capacitor charging period, the ICD:
■
operates in its programmed pacing mode,
■
applies the programmable Post Shock Pacing parameters (see
page 6-8), and
■
may lose the telemetry link to the programmer due to electrical noise.
After the charging period completes, the ICD starts a 300 ms blanking
period.
Charge Circuit Timeout
If a single charging period reaches 30 seconds, the ICD terminates
charging, aborts the therapy attempt (but retains the charge on the
capacitors), and resumes detection using the programmed detection
criteria. The charge circuit is still active. The ICD records the
Circuit Timeout Device Status Indicator, which is later sent to the
programmer via telemetry. If a
Charge Circuit Timeout indicator is
displayed, please inform a Medtronic representative and replace the
ICD immediately.
Charge
5-4
Charge Circuit Inactive
If three consecutive charging periods have each reached 30 seconds, the
ICD terminates charging, aborts the therapy attempt, and disables the
automatic tachyarrhythmia therapies and manual operations, except for
Emergency VVI pacing. The charge circuit is inactive. The ICD records
the
Charge Circuit Inactive Device Status Indicator, which is sent to the
programmer via telemetry. If a
displayed, do not try to reprogram the ICD. Please inform a Medtronic
representative and replace the ICD immediately.
GEM®III VR System Reference Guide
Charge Circuit Inactive indicator is
Tachyarrhythmia Therapies
High Voltage Therapy Overview
High Voltage Synchronization
The synchronization sequence is designed to confirm the continued
presence of the arrhythmia and synchronize the high voltage pulse to a
sensed R-wave. In general:
■
committed defibrillation therapies are timed to the first non-refractory
sensed event if possible, or delivered asynchronously if they cannot
be synchronized (see page 5-11).
■
non-committed defibrillation therapies (i.e. when Reconfirm VF is
active) and cardioversion therapies are timed to a non-refractory
arrhythmic
synchronized (see page 5-13 and page 5-17).
The ICD high voltage pulses are synchronized to events sensed on the
implanted electrodes. Surface ECGs exhibit morphologies and timing that
do not necessarily coincide exactly with those taking place at the
electrode sites.
Note: For any delivered high voltage therapy, there is a possibility of
inducing an atrial arrythmia.
sensed event if possible, or aborted if they cannot be
GEM®III VR System Reference Guide
5-5
Tachyarrhythmia Therapies
High Voltage Therapy Overview
After A Delivered High Voltage Therapy
Immediately after a delivered high voltage therapy, the ICD starts:
■
a post shock blanking period of 520 ms, and
■
one VVI pacing cycle at 50 ppm (escape interval of 1200 ms).
After the first ventricular event, the programmed bradycardia pacing mode
resumes, using the Post Shock Pacing parameters (see page 6-8).
The ICD monitors for an outcome from the delivered therapy: either
episode termination or redetection. After an automatic defibrillation
therapy, VT detection is temporarily suspended for 17 events, or until
episode termination is detected, whichever occurs first. This feature is
designed to reject transient VTs that may follow the high voltage therapy.
After any manual therapy delivered via the programmer (including
emergency HV therapies), automatic detection is suspended for as long
as the programming head remains over the device, or until
[Resume]
is selected.
After An Aborted High Voltage Therapy
5-6
Following an aborted defibrillation or cardioversion therapy, or an aborted
charging period, the ICD reverts immediately to its programmed
bradycardia pacing settings.
The ICD resumes monitoring the cardiac cycle for arrhythmias, using the
programmed redetection NIDs, after the next paced or sensed ventricular
event. If the ICD redetects the same arrhythmia after an aborted therapy,
it attempts to synchronize and deliver the same therapy. However, if the
ICD detects episode termination, it resumes normal detection.
Note: If a high voltage therapy aborts, leaving the energy stored on the
capacitors, the delivered energy of a subsequent cardioversion or
defibrillation therapy could be higher than the programmed value.
GEM®III VR System Reference Guide
Tachyarrhythmia Therapies
High Voltage Therapy Overview
High Voltage Therapy Parameters
Energy
The ICD can be programmed to deliver up to 30 joules1, which
corresponds to a maximum stored energy of 35 joules. This value is
derived from the peak capacitor voltage, which is always greater than the
energy delivered by the device. Table 1-6 on page 1-9 compares all of the
programmable delivered energy levels to their corresponding stored
energy levels. The defibrillation or cardioversion energy level is
programmed independently for each therapy.
Waveform
The waveform of the cardioversion or defibrillation pulse is biphasic.
The biphasic pulse consists of two phases. In the first phase, current is
delivered between the high-voltage electrodes until the pulse decays to
50% of its original voltage. The ICD then truncates the pulse and reverses
the current pathway for the second phase by transposing the cathode and
anode. After the second phase of the pulse has decayed to 50% of its
original voltage, the ICD truncates the pulse.
Second phase
truncated after 50%
pulse decay
Biphasic pulse
inverted after 50%
pulse decay
50% Tilt
Figure 5-1. Biphasic Waveform for Cardioversion and Defibrillation
Tilt
The ICD high voltage pulses are tilt controlled: the output pulse is
truncated when it has decayed by a percentage of its initial value, or “tilt.”
The ICD uses a fixed tilt value of 50%.
Figure 5-1 shows how the ICD delivers biphasic shocks across a single
pathway with 50% tilt. The biphasic pulse is inverted to a negative voltage
and then truncated to zero after a 50% decay of its second phase.
Negative voltage indicates that the current pathway has been reversed.
1
Delivered energy of biphasic waveform into 75 ohms.
GEM®III VR System Reference Guide
5-7
Tachyarrhythmia Therapies
High Voltage Therapy Overview
Pathway Polarity
(anode → cathode)
Pathway Polarity
Pathway polarity assigns the implanted electrodes to serve as the anode
and cathode in delivering the high voltage pulse. For example, an ICD
implanted with two electrodes, CAN (
programmed to deliver current from CAN to RV (
(
B>AX).
Table 5-2. Modes and Polarities for Cardioversion and Defibrillation
Two HV Electrodes:
CAN and RV
Options
CAN → RV = AX>B
or
RV → CAN = B>AX
HVA) and RV (HVB), can be
AX>B) or from RV to CAN
Three HV Electrodes:
CAN, RV, and SVC
CAN, SVC → RV = AX>B
or
RV → CAN, SVC = B>AX
5-8
GEM®III VR System Reference Guide
Defibrillation
Tachyarrhythmia Therapies
Defibrillation
Defibrillation shocks are independently programmable for energy and
pathway polarity. When Reconfirm VF is enabled, the first defibrillation
therapy delivered during an episode is non-committed.
Table 5-3. Defibrillation Programmable Parameters
VF Therapy Status
Energy (J)
Pathway
Reconfirm VF After Initial Charge
Progressive Episode Therapies
a
Any of the six VF therapies can be programmed Off. The ICD retains the settings but skips
the “Off” therapy and immediately delivers the next available therapy.
On or Off.
0.4 – 30 (see page 5-7)
AX>B or B>AX (see page 5-8)
Yes or No (see page 5-13)
On or Off (see page 5-33)
a
Programming Defibrillation Therapies
No more than one of the six VF Therapies should be programmed to less
than 30 joules.
You can program the ICD to reconfirm a VF episode before delivering the
first
programmed defibrillation therapy only. Subsequent therapies, and
re-applications of an aborted first therapy, are always committed.
If the Progressive Episode Therapies feature is enabled, the ICD may
deliver a defibrillation therapy at a higher energy level than was
programmed. This ensures that each therapy delivered during an ongoing
episode is at least as aggressive as the previous therapy (see
“Progressive Episode Therapies” on page 5-33).
GEM®III VR System Reference Guide
5-9
Tachyarrhythmia Therapies
Defibrillation
♦
To Program Defibrillation Therapies
1.Select the Params icon, and choose the Therapies menu option.
2.Select [VF] (see Figure 5-2).
3.Enable VF Therapies by setting VF Therapy Status parameters to On.
4.Set Energy and Pathway parameters for each enabled VF Therapy.
5.Enable VF reconfirmation by setting Reconfirm VF after initial charge
to Yes.
6.Select [PROGRAM].
VF Therapy Screen
Selected
Defibrillation
Parameters
Shared Therapy
Parameters Screen
Selected
Progressive Episode
Therapies Parameter
Figure 5-2. Programmable Defibrillation Parameters
7.To program the Progressive Episode Therapies feature, select
[Shared] (see Figure 5-3).
8.Set Progressive Episode Therapies to On or Off.
9.Select [PROGRAM].
5-10
Figure 5-3. Shared Therapy Parameters Screen
GEM®III VR System Reference Guide
Tachyarrhythmia Therapies
Defibrillation
Committed Defibrillation Synchronization
Note: A defibrillation therapy is
defibrillation shock delivered during an episode
always
committed, unless it is the first
and
the Reconfirm VF
feature is enabled, (see page 5-13).
A defibrillation therapy is synchronized to a sensed R-wave if possible.
Committed defibrillation uses a synchronization escape period of 500 ms
to identify the R-wave for delivering the shock.
At the
first non-refractory sensed event
, the shock is delivered
(Figure 5-4).
However, if the 500 ms synchronization escape period expires
without a
non-refractory event, the shock is delivered asynchronously (Figure 5-5).
The synchronization interval begins at the end of the 300 ms post-charge
blanking period, with an effective refractory period of 100 ms. If the ICD
senses a ventricular event during this refractory period, it starts a new
synchronization interval. This subsequent synchronization interval begins
with a 120 ms blanking period and no additional refractory period.
When Reconfirm VF is active, the first defibrillation shock is synchronized
to a non-refractory
arrhythmic
event (see page 5-13).
At the first non-refractory
event, the shock is
delivered.
After the charge blanking
period, the synchronization
interval of 500 ms begins with
an effective refractory period of
100 ms.
Refractory
Synchronization
Figure 5-4. Defibrillation Synchronized to an Arrhythmic Event
Blanking
C
E
300 ms
400 ms
500 ms
C
D
520 ms
200 ms
An extended blanking
period of 520 ms begins
after the shock.
GEM®III VR System Reference Guide
5-11
Tachyarrhythmia Therapies
Defibrillation
This refractory sensed
event restarts the
synchronization interval.
Blanking
Refractory
Synchronization
Figure 5-5. Defibrillation Delivered After Failing to Synchronize
These two low-amplitude
events go unsensed.
C
E
300 ms120 ms520 ms
400 ms
V
R
500 ms
500 ms
C
D
The synchronization sequence times
out and the device delivers the
defibrillation shock.
An extended blanking
period of 520 ms begins
200 ms
after the shock.
5-12
GEM®III VR System Reference Guide
Tachyarrhythmia Therapies
Defibrillation
Non-Committed Defibrillation Synchronization
When the Reconfirm VF feature is enabled, the first defibrillation shock
delivered during an episode is synchronized to a non-refractory
arrhythmic event, but it aborts if the VF terminates spontaneously into
bradycardia or sinus rhythm.
The ICD uses a sequence of reconfirmation “windows,” each lasting
60 ms plus the programmed
VT Interval
1
to identify an ongoing
tachyarrhythmia. The ICD defines a non-refractory sensed ventricular
event within this window as
arrhythmic
for reconfirmation purposes.
Synchronization begins with a 300 ms post-charge blanking period. After
this blanking period the ICD starts a reconfirmation window and a
refractory period of up to the same duration (
VT Interval
1
plus 60 ms).
If the next ventricular event occurs after the reconfirmation window, it is
classified as
normal
, and the ICD starts a new reconfirmation window. If
the event occurs during the reconfirmation window, it also falls within the
refractory period, and is therefore classified as
refractory
. If this occurs,
the refractory period is terminated, and the ICD starts a new
reconfirmation window.
Subsequent ventricular events during synchronization start new
reconfirmation intervals. A sensed event starts a 120 ms post-sense
blanking period and a reconfirmation window. Paced events start a
programmed post-shock
a refractory period of up to the same duration (
Pace Blank period, a reconfirmation window, and
VT Interval
1
plus 60 ms).
VF is reconfirmed and the ICD delivers the defibrillation therapy
synchronized with the second arrhythmic ventricular event. However, if
four normal ventricular intervals occur without reconfirming VF, the ICD
aborts the shock (see Figure 5-7).
Bradycardia Pacing During Reconfirmation – Each reconfirmation
interval also begins an escape period in the VVI mode. If the escape
period times out, the ICD paces the ventricle and continues to monitor for
reconfirmation of VF. VF cannot be reconfirmed on an event that
immediately follows a paced event, because a refractory period of up to
60 ms plus the programmed
VT Interval
1
follows a paced event (see
Figure 5-7).
1
Or VF Interval, if VT Detection is Off
GEM®III VR System Reference Guide
5-13
Tachyarrhythmia Therapies
Defibrillation
At the second arrhythmic event, the shock is delivered
and an extended brady escape period begins.
The first event after
charge end is refractory.
C
E
300 ms
Blanking
Refractory
Reconfirmation
a
The refractory period can last as long as the
VT Interval plus 60 ms, but will terminate with the
first ventricular event that falls within it.
b
VT Interval plus 60 ms
V
R
120 ms
460 ms
460 ms
a
b
460 ms
Figure 5-6. Defibrillation Delivered After Reconfirming VF
The VF has terminated
spontaneously after
it was detected.
A paced event occurs, followed by the
programmable pace blanking period
and a refractory period of 460 ms
(TDI plus 60 ms).
If necessary, the device delivers
V
C
S
D
120 ms 520 ms
b
1200 ms brady escape interval
a VVI pacing pulse.
After four normal intervals, the
defibrillation therapy aborts.
V
P
240 ms
200 ms
C
E
Blanking
Refractory
300 ms120 ms240 ms
a
Reconfirmation
a
The refractory period can last as long as
the TDI plus 60 ms but will terminate with
V
R
460 ms
460 ms
460 ms
b
b
the first ventricular event that falls within it.
b
TDI plus 60 ms
Figure 5-7. Defibrillation Aborts After Failure to Reconfirm
5-14
GEM®III VR System Reference Guide
V
S
V
P
V
S
120 ms120 ms120 ms
b
460 ms
460 ms
b
460 ms
b
460 ms
Each event generates a new 460 ms reconfirmation interval,
but the VF is never reconfirmed during any of them.
V
S
b
200 ms
Cardioversion
Tachyarrhythmia Therapies
Cardioversion
Table 5-4. Cardioversion Programmable Parameters
VT (or FVT) Therapy Status
Therapy Type
Energy (J)
Pathway
Progressive Episode Therapies
a
Any of the six VT or six FVT therapies can be programmed Off. The ICD retains the settings,
but skips the “off” therapy and immediately delivers the next available therapy.
On or Off
CV
0.4 – 30 (see page 5-7)
AX>B or B>AX (see page 5-8)
On or Off (see page 5-33)
a
Programming Cardioversion Therapies
If VT or FVT Therapy is programmed On, VF Therapy must also be On.
At least one therapy for VT and one therapy for FVT should be
programmed to 30 joule cardioversion.
The programmer requires that if FVT therapies are On, then at least one
must be programmed to cardioversion (any energy).
If the Progressive Episode Therapies feature is enabled, the ICD may
deliver a cardioversion therapy at a higher energy level than was
programmed. This ensures that each therapy delivered during an ongoing
episode is at least as aggressive as the previous therapy (see
“Progressive Episode Therapies” on page 5-33).
GEM®III VR System Reference Guide
5-15
Tachyarrhythmia Therapies
Cardioversion
♦
To Program Cardioversion Therapies
1.Select the Params icon, and choose the Therapies menu option.
2.Select either the [FVT] or [VT] (see Figure 5-8).
3.Set Therapy Type to CV (Therapy Status also turns On).
4.Set Energy and Pathway parameters for each cardioversion therapy.
5.Select [PROGRAM].
VT Therapy Screen
Selected
Cardioversion
Parameters
Figure 5-8. Programmable Cardioversion Parameters
5-16
GEM®III VR System Reference Guide
Tachyarrhythmia Therapies
Cardioversion
6.To program the Progressive Episode Therapies feature, select
[Shared] (see Figure 5-9).
7.Choose a setting for Progressive Episode Therapies.
8.Select [PROGRAM].
Shared Therapy Parameters
Screen Selected
Progressive Episode
Therapies Parameter
Figure 5-9. Shared Therapy Parameters Screen
Cardioversion Synchronization
In general, a cardioversion therapy is synchronized to a non-refractory
arrhythmic event, but it aborts if the tachycardia
terminates spontaneously.
Cardioversion uses a sequence of synchronization “windows,” each
lasting 60 ms plus the programmed
VT Interval
tachyarrhythmia. The ICD defines a non-refractory sensed ventricular
event within this window as
arrhythmic
for synchronization purposes. The
first synchronization window begins after the 300 ms post-charge
blanking period, with an effective refractory period of 100 ms.
1
, to identify an ongoing
The ICD delivers the cardioversion therapy on a ventricular event that:
■
is non-refractory and arrhythmic, and
■
is not the first arrhythmic event during synchronization.
If a synchronization window expires without a ventricular event, the shock
aborts at the next event or escape period timeout (see Figure 5-10). The
shock also aborts if three consecutive ventricular events are sensed
during refractory periods.
1
Or VF Interval, if VT Detection is Off.
GEM®III VR System Reference Guide
5-17
Tachyarrhythmia Therapies
Cardioversion
This sensed beat generates a second
synchronization window of 460 ms.
Any other refractory or arrhythmic ventricular event begins a new
synchronization window, with a 120 ms post-sense blanking period and a
200 ms refractory period.
Bradycardia Pacing During Synchronization – Each synchronization
window begins an escape period in the VVI mode. If the escape period
times out, the ICD aborts the cardioversion therapy and paces, using its
standard output settings. The escape period is the current V-V pacing rate
(i.e., sensor or programmed lower rate) interval, up to a maximum of
1760 ms.
The next ventricular event comes after the synchronization window has
timed out, and the cardioversion therapy aborts.
V
R
200 ms
460 ms (TDI plus 60)
Blanking
Refractory
Synchronization
C
E
300 ms120 ms120 ms
400 ms
Figure 5-10. Cardioversion Aborts After Failure to Reconfirm
V
S
200 ms
A post-sense or post-pace
blanking period begins when the
shock aborts. The programmed
pacing mode also resumes.
5-18
GEM®III VR System Reference Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.