Using this manualxv
Contacting technical supportxv
Customer educationxv
Referencesxvi
Notice xvi
Abbreviations and acronymsxvii
Part IQuick overview
1Quick reference3
Physical characteristics 4
Magnet application 8
Longevity projections9
Replacement indicators 12
Typical charge times13
High voltage therapy energy 13
Stored data and diagnostics14
New and enhanced features17
Lead connection header options17
Patient management17
Tachyarrhythmia detection 19
Tachyarrhythmia therapy 20
Bradycardia pacing 20
EP studies21
2The Marquis VR system23
System overview 24
Detecting and treating tachyarrhythmias25
Treating bradycardia26
Monitoring for real-time and stored data26
Conducting electrophysiologic tests26
Alerting the patient to system events 26
Indications and usage 27
Contraindications27
Marquis VR 7230Cx, B, and E Reference Manual
iv
Table of contents
Patient screening27
3Emergency therapy29
Delivering emergency therapies 30
Effect on system operation30
Aborting an emergency therapy30
On-screen and display panel buttons31
Temporary parameter values31
How to deliver emergency 30 joule defibrillation32
How to deliver emergency cardioversion32
How to deliver emergency fixed burst pacing33
How to deliver emergency VVI pacing 33
Part IIDevice implant and patient follow-up procedures
4Implanting the ICD 37
Overview 38
Preparing for an implant38
Equipment for an implant38
Sterile supplies 39
How to prepare for implanting39
Replacing an ICD 40
How to explant and replace an ICD40
Positioning the leads41
Using transvenous leads41
Using epicardial leads41
Surgical incisions 42
Testing sensing and pacing thresholds43
Parameters43
Considerations43
Connecting the leads to the ICD44
How to connect the lead to the device 47
Testing defibrillation operation and effectiveness47
High voltage implant values48
Binary search protocol48
How to prepare for defibrillation threshold testing49
How to perform defibrillation threshold testing50
Positioning and securing the ICD 52
How to position and secure the device 52
Marquis VR 7230Cx, B, and E Reference Manual
Completing the implant procedure52
How to complete programming the device 53
5Conducting a patient follow-up session 55
Patient follow-up guidelines56
Verifying the status of the implanted system 56
Verifying accurate detection and appropriate therapy 57
Considerations57
Verifying effective bradycardia pacing58
Considerations59
Part IIIConfiguring the ICD for the patient
6Detecting tachyarrhythmias 63
Detection overview 64
Suspending tachyarrhythmia detection66
Setting up sensing 67
Parameters 67
Considerations67
How to program sensitivity69
Details about sensing 69
Detecting VF episodes 71
Parameters 72
Considerations72
Restrictions 73
How to program VF detection 74
Details about VF detection 74
Detecting VT episodes 76
Parameters 76
Considerations76
Restrictions 77
How to program VT detection 78
Details about VT detection 78
Detecting FVT episodes82
Parameters 82
Considerations82
Restrictions 83
How to program FVT detection 84
Details about FVT detection 84
Table of contents
v
Marquis VR 7230Cx, B, and E Reference Manual
vi
Table of contents
Detecting tachyarrhythmia episodes with Combined Count87
Details about Combined Count detection87
Monitoring episodes for termination or redetection 90
Parameters 90
Considerations90
How to program redetection parameters 91
Details about episode termination and redetection91
Enhancing detection with Wavelet 94
Parameters94
Considerations95
How to program Wavelet98
Details about Wavelet 98
Details about Auto Collection 101
Enhancing VT detection with the Onset criterion103
Parameters 103
Considerations103
How to program Onset 104
Details about Onset104
Enhancing VT detection with the Stability criterion108
Parameters 108
Considerations108
How to program Stability 109
Details about Stability 109
Detecting prolonged tachyarrhythmias with High Rate Timeout111
Parameters 111
Considerations111
How to program High Rate Timeout 112
Details about High Rate Timeout112
Key ter ms113
7Treating tachyarrhythmia episodes119
Treating VF with defibrillation 120
Parameters120
Considerations 121
Restrictions 122
How to program VF therapies122
Details about VF therapy 122
Treating VT and FVT with antitachycardia pacing 130
Parameters for all ATP therapies 130
Marquis VR 7230Cx, B, and E Reference Manual
Table of contents
Parameters for Burst therapy 131
Parameters for Ramp therapy 131
Parameters for Ramp+ therapy 132
Considerations 132
Restrictions 133
How to program ATP therapies133
Details about ATP therapies 134
Treating VT and FVT with cardioversion 140
Parameters 140
Considerations 141
Restrictions 142
How to program cardioversion therapies142
Details about cardioversion therapy 143
Optimizing therapy with Smart Mode and Progressive Episode
Therapies 149
Parameters 149
Considerations149
Restrictions 150
How to program Smart Mode 150
Details about Smart Mode151
How to program Progressive Episode Therapies 152
Details about Progressive Episode Therapies152
Key ter ms153
vii
8Treating bradycardia157
Providing basic pacing therapy 158
Parameters158
Considerations159
Restrictions 160
How to program bradycardia pacing parameters 160
Details about basic bradycardia pacing161
Enhancing pacing for optimal cardiac output162
Parameters162
Considerations163
Restrictions 163
How to program Rate Response164
Details about Rate Responsive Pacing 165
How to program Hysteresis169
Details about Hysteresis 169
Marquis VR 7230Cx, B, and E Reference Manual
viii
Table of contents
How to program Ventricular Rate Stabilization171
Details about Ventricular Rate Stabilization171
Providing pacing after high voltage therapies173
Parameters173
Considerations173
How to program pacing after high voltage therapies 174
Details about Post Shock Pacing parameters174
Key ter ms174
9Optimizing charge time and device longevity 177
Optimizing charge time178
Parameters178
Considerations178
How to evaluate charging performance179
Details about managing charge time179
Optimizing device longevity181
Considerations181
Key ter ms182
Part IVEvaluating and managing patient treatment
10Using the programmer 185
Connecting the programmer hardware 186
Using an external printer187
To view a list of supported printers188
Materials you will need 188
To connect the printer to your programmer 190
Using the programming head 192
During an episode in progress192
During marker transmissions192
How capacitor charging affects the light array 193
Alternative program and interrogate buttons 193
Display screen features 193
Programmer status bar display 194
Live Rhythm Monitor window 195
Task area196
Command bar 198
Tool palette 198
Buttons 200
Marquis VR 7230Cx, B, and E Reference Manual
Setting programmer preferences200
How to set the programmer time and date 201
How to set audio preferences201
How to select a different language 202
How to check the software version number 203
How to set printing preferences 203
How to set reports preferences204
Starting and ending patient sessions 204
How to start a patient session205
How to end a patient session 206
Automatic interrogation206
Viewing live waveforms 207
Parameters207
How to use the Adjust window 208
How to use the waveform adjustment button bar209
Details about the live rhythm monitor 209
Recording live waveforms 213
Printing a report while recording live ECG214
Saving and retrieving device data215
Considerations215
How to save ICD data to a disk216
How to read ICD data from a disk 216
Saving data to a disk218
Data file names218
Reading device data from diskette 218
Printing reports219
How to print a report220
Print Queue 220
Key ter ms221
Table of contents
ix
11Using system evaluation tools 223
A summary of system evaluation tools224
Automatic daily measurements 224
Taking a quick look at device activity225
How to use Quick Look225
Quick Look observations 226
Using the Patient Alert feature 227
Parameters227
Considerations229
Marquis VR 7230Cx, B, and E Reference Manual
x
Table of contents
How to program the Patient Alert feature232
Instructing the patient 232
Viewing the Patient Alert Events 233
How to view the Patient Alert events234
Streamlining follow-ups with Checklist 235
How to select and use a checklist236
How to create, edit, and delete a checklist237
Key ter ms238
12Setting up and viewing collected data239
A summary of data collection240
Setting up data collection 241
Parameters241
Considerations242
How to set up data collection243
Details about data collection parameters244
How to clear collected data247
Collecting lead performance data247
Daily lead impedance and EGM amplitude measurements247
Sensing integrity counter249
Viewing the episode and therapy efficacy counters249
How to view and print counter data250
Details about the episode and therapy efficacy counters251
Therapy counters253
Viewing episode data254
How to view episode data255
Details about episode data256
How to view an Interval Plot258
How to view an EGM Strip259
How to display the episode text information 261
How to display QRS Snapshot data 262
Viewing Flashback Memory 263
How to view the Flashback Memory264
Viewing battery and lead status data 265
How to view battery and lead status data267
Viewing lead performance trends267
How to view lead performance trend graphs 268
Using Cardiac Compass to view long term clinical trends269
Marquis VR 7230Cx, B, and E Reference Manual
How to print a Cardiac Compass report271
Details on Cardiac Compass trend data271
Viewing and entering patient information275
How to view and enter new patient information 277
Displaying and printing patient information277
Automatic device status monitoring 278
Device status indicator warnings278
How to respond to an electrical reset279
Key ter ms280
13Testing the system285
Testing overview286
Evaluating the underlying rhythm 287
Considerations287
How to Perform an Underlying Rhythm test287
Measuring pacing thresholds288
Parameters289
Considerations289
How to Perform a Pacing Threshold Test 290
Testing the Wavelet criterion 291
Parameters291
Considerations291
Restrictions 293
How to evaluate the current template with the Wavelet test293
How to collect a template with the Wavelet test 294
Measuring lead impedance 295
Considerations295
How to perform a Lead impedance test 295
Details about the Lead Impedance test 296
Measuring EGM Amplitude 297
Parameters297
Considerations297
Restrictions 298
How to perform an EGM Amplitude test298
Testing the device capacitors 299
Considerations299
How to perform a Charge/Dump test 300
Key ter ms301
Table of contents
xi
Marquis VR 7230Cx, B, and E Reference Manual
xii
Table of contents
14Conducting Electrophysiologic Studies 303
EP Study overview304
Inducing VF with T-Shock306
Parameters306
Considerations307
Restrictions 307
How to deliver a T-Shock induction308
Details about T-Shock induction 308
Inducing VF with 50 Hz Burst 309
Parameters309
Considerations310
How to deliver a 50 Hz Burst induction311
Details about 50 Hz Burst induction312
Inducing an arrhythmia with Manual Burst312
Parameters312
Considerations312
How to deliver a Manual Burst induction 313
Details about Manual Burst induction314
Inducing an arrhythmia with PES 314
Parameters314
Considerations315
How to deliver a PES induction315
Details about PES induction316
Delivering a manual therapy316
Parameters for manual defibrillation and cardioversion 316
Parameters for manual ATP therapies 316
Considerations318
How to deliver a manual therapy 318
Details about manual therapies 318
Key ter ms319
15Solving system problems321
Overview 322
Solving sensing problems323
Solving tachyarrhythmia detection problems 324
Solving tachyarrhythmia therapy problems325
Solving bradycardia pacing problems326
Responding to device status indicators327
Marquis VR 7230Cx, B, and E Reference Manual
Key ter ms328
Part VAppendices
AWarnings and precautions 333
General warnings 334
Storage and handling334
Resterilization335
Device implantation and ICD programming 335
Lead evaluation and lead connection337
Follow-up testing338
Explant and disposal 339
Medical therapy hazards339
Home and occupational environments341
BDevice Parameters343
Emergency settings 344
Detection parameters 345
Therapy parameters 347
Bradycardia pacing parameters 349
System maintenance parameters 350
Data collection parameters 352
System test and EP study parameters 353
Fixed parameters356
Patient information parameters357
Programmer symbols358
Parameter Interactions Window359
Parameter interlocks360
Table of contents
xiii
Index 361
Marquis VR 7230Cx, B, and E Reference Manual
xiv
Table of contents
Marquis VR 7230Cx, B, and E Reference Manual
Introduction
Using this manual
Before implanting the ICD, it is strongly recommended that you:
■
Refer to the product literature packaged with the ICD for
information about prescribing the ICD.
■
Thoroughly read this manual and the technical manuals for the
leads used with the device.
■
Discuss the procedure and the ICD system with the patient
and any other interested parties, and provide them with any
patient information packaged with the ICD.
Contacting technical support
Medtronic employs highly trained representatives and engineers
located throughout the world to serve you and, upon request, to
provide training to qualified hospital personnel in the use of
Medtronic products.
Introduction
xv
In addition, Medtronic maintains a professional staff of consultants
to provide technical consultation to product users. For medical
consultation, Medtronic can often refer product users to outside
medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative,
or call or write Medtronic at the appropriate address or telephone
number listed on the back cover.
Customer education
Medtronic invites physicians to attend an education seminar on
the complete ICD system. The course includes indications for use,
an overview of ICD system functions, implant procedures, and
patient management.
Marquis VR 7230Cx, B, and E Reference Manual
xvi
Introduction
References
Notice
The primary reference for background information is Zacouto FI,
Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B,
ed. Cardiac Pacing Diagnostic and Therapeutic Tools. New York:
Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
Singer I, Barold SS, Camm AJ, Eds. Nonpharmacological
Therapy of Arrhythmias for the 21st Century: The State of the
Art. Armonk, NY: Futura Publishing Co. 1998.
■
Estes M, Manolis AS, Wang P, Eds. Implantable
Cardioverter-Defibrillators. New York, NY: Marcel Dekker, Inc.
This software is provided as an informational tool for the end user.
The user is responsible for accurate input of patient information
into the software. Medtronic makes no representation as to the
accuracy or completeness of the data input into the software.
MEDTRONIC SHALL NOT BE LIABLE FOR ANY DIRECT,
INDIRECT, INCIDENTIAL OR CONSEQUENTIAL DAMAGES TO
ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE
INFORMATION PROVIDED IN THE SOFTWARE.
Marquis VR 7230Cx, B, and E Reference Manual
Abbreviations and acronyms
ATP Antitachycardia Pacing
BOL Beginning of Life
bpm beats per minute
CNID Combined (VT and VF) Number of Intervals to Detect
CV Cardioversion
DF/Defib Defibrillation
ECG Electrocardiogram
EGM Electrogram
EOL End of Life
ERI Elective Replacement Indicator
FDI Fibrillation Detection Interval
FTI Fast Ventricular Tachycardia Detection Interval
Abbreviations and acronyms
xvii
FVT Fast Ventricular Tachycardia
ICD Implantable Cardioverter Defibrillator
J joules
-1
reciprocal minutes; for example, pacing pulses per minute
min
ms milliseconds
mV millivolts
NID Number of Intervals to Detect
NST Non-sustained Tachycardia
PES Programmed Electrical Stimulation
ppm paces or pulses per minute
PVC Premature Ventricular Contraction
RNID Number of Intervals to Redetect
R-R a ventricular interval
SVT Supraventricular Tachycardia
Marquis VR 7230Cx, B, and E Reference Manual
xviii
Abbreviations and acronyms
TDI Tachycardia Detection Interval
V volts
V- Ventricular
VF Ventricular Fibrillation
VF NID VF Number of Intervals to Detect
VRS Ventricular Rate Stabilization
VT Ventricular Tachycardia
VT NID VT Number of Intervals to Detect
Marquis VR 7230Cx, B, and E Reference Manual
Quick overview
Part I
Marquis VR 7230Cx, B, and E Reference Manual
Physical characteristics 4
Magnet application 8
Longevity projections 9
Replacement indicators 12
Typical charge times 13
High voltage therapy energy 13
Stored data and diagnostics 14
New and enhanced features 17
Quick reference1
1
Marquis VR 7230Cx, B, and E Reference Manual
4
Chapter 1
Physical characteristics
Physical characteristics
Table 1-1. ICD physical characteristics
Connector typeCx
a
B or E
Volume 36 cc43 cc
Mass 75 g80 g
H x W x D
b
Surface area of device can 66 cm
Radiopaque ID
c
68 mm x 51 mm
x 14 mm
2
74 mm x 51 mm
x 14 mm
2
63 cm
PKDB-type connector PLW
E-type connector PLY
Materials in contact with
human tissue
d
Titanium / polyurethane / silicone rubber
BatteryLithium silver vanadium oxide
a
Measurements are nominal values based on CAD (computer aided design)
model measurements and are rounded to the nearest unit.
b
Grommets may protrude slightly beyond the can surface.
c
Engineering series number follows the radiopaque code.
d
These materials have been successfully tested for the ability to avoid biological
incompatibility. The device does not produce an injurious temperature in the
surrounding tissue.
Marquis VR 7230Cx, B, and E Reference Manual
Table 1-2. Cx-type connector characteristics
General description
One IS-1 connector for pacing and
sensing, Two DF-1 connectors for
high voltage therapy, Active Can
electrode (programmable).
Figure 1-1. Cx-type connector
2
Device
Port
SVCDF-1HVX
RVDF-1HVB
Cann/aHVA, Can
VIS-1 bipolar
1
Quick reference
Physical characteristics
Connector
Typ e
Software
Name
5
5
1 suture holes
2 SVC port (DF-1)
3RV port (DF-1)
4 programmable Active Can
5V. port (IS-1)
3
4
Marquis VR 7230Cx, B, and E Reference Manual
6
Chapter 1
Physical characteristics
Table 1-3. B-type connector characteristics
General description
PortConnector Type
HVB6.5 mmHVB
Device
Three 6.5 mm unipolar high
voltage ports and one
3.2 mm low profile bipolar
pace/sense (IS-1
compatible) port.
a
The HVA and HVX ports are electronically connected and are treated as the HVX
electrode. For more information see “B- and E-type connector pathways” on
page 123.
HVA
HVX
a
a
6.5 mmHVX
6.5 mmHVX
Cann/aHVA, Can
P/S bipolar
pace sense
3.2 mm bipolar
(IS-1 compatible)
Figure 1-2. B-type connector
Software
Name
7
2
6
5
3
4
1 suture holes
2 HVB port (6.5 mm)
3 HVA port (6.5 mm)
4 programmable Active Can
5 HVX port (6.5 mm)
6/7 P+/S and P-/S port (3.2 mm)
1
Marquis VR 7230Cx, B, and E Reference Manual
Quick reference
Physical characteristics
Table 1-4. E-type connector characteristics
General description
Port
P+/S port5.0 mm
Device
Two 6.5 mm unipolar high voltage
ports and two 5 mm unipolar
pace/sense ports.
HVA port6.5 mmHVX
HVB port6.5 mmHVB
Cann/aHVA, Can
P-/S port5.0 mm
a
The HVA port of the E-type connector can be used as the HVX electrode when
Active Can is programmed off, see “B- and E-type connector pathways” on
page 123.
Figure 1-3. E-type connector
Connector
Typ e
Software
Name
a
7
6
5
2
3
4
1 suture holes
2 P+/S port (5.0 mm)
3 HVA port (6.5 mm)
4 programmable Active Can
5 HVB port (6.5 mm)
6 P-/S port (5.0 mm)
1
Marquis VR 7230Cx, B, and E Reference Manual
8
Chapter 1
Magnet application
Magnet application
Bringing a magnet close to the device triggers changes in device
operation as shown in Table 1-5. When the magnet is removed,
the device returns to its programmed operations.
Table 1-5. Effects of magnet application on the device
Pacing modeas programmed
Pacing rate and intervalas programmed
VF, VT, and FVT detection suspended
Patient Alert audible tones
(20 seconds or less)
with programmable alert(s) enabled:
■
continuous tone (Test)
■
on/off intermittent tone (seek follow-up)
■
high/low dual tone (urgent follow-up)
with programmable alerts disabled:
■
no tone
■
high/low dual tone (urgent follow-up)
a
Rate response adjustments are suspended while a Patient Alert tone sounds.
b
Detection resumes if telemetry is established and the application software is
running, or it resumes after the application software has started.
c
The Test tone does not sound if “VF Detection/Therapy Off” is the only alert
enabled.
a
b
c
Marquis VR 7230Cx, B, and E Reference Manual
Longevity projections
Longevity estimates are based on accelerated battery discharge
data and device modeling with EGM pre-storage off, 60 ppm
-1
(min
) pacing rate, with
■
2.5 V pacing pulse amplitude, 0.4 ms pacing pulse width, and
30 J delivered therapy energy (see Table 1-6)
■
3 V pacing pulse amplitude, 0.4 ms pacing pulse width, and
30 J delivered therapy energy (see Table 1-7)
This model assumes default automatic capacitor formation
setting. As a guideline, each full energy charge decreases device
longevity by approximately 33 days.
Device longevity is affected by how certain features are
programmed, such as EGM pre-storage. For more information,
see “Optimizing device longevity” on page 181.
Considerations for using EGM pre-storage
When the EGM pre-storage feature is programmed off, the device
starts to store EGM following the third tachyarrhythmia event and
also provides up to 20 seconds of information before the onset of
the tachyarrhythmia, including:
■
VV intervals
■
Marker Channel
■
interval plot Flashback
Quick reference
Longevity projections
9
When the EGM pre-storage feature is programmed on, the device
also collects up to 20 seconds of EGM information before the
onset of the arrhythmia.
In a patient who uniformly repeats the same onset mechanisms,
the greatest clinical benefit of pre-onset EGM storage is achieved
after a few episodes are captured.To maximize the effectiveness
of the EGM pre-storage feature and optimize device longevity,
consider these programming options:
■
Turn pre-storage on to capture possible changes in the onset
mechanism following significant clinical adjustments, for
example, device implant, medication changes, and surgical
procedures.
■
Turn pre-storage off once you have successfully captured the
information of interest.
Marquis VR 7230Cx, B, and E Reference Manual
10
Chapter 1
Longevity projections
a
b
Table 1-6. Projected longevity in years with 2.5 V pacing amplitude and
0.4 ms pulse width
Percent
pacing
Maximum
energy charging
frequency
a
EGM
pre-storage
500 ohm
b
impedance
pacing
900 ohm
pacing
impedance
0%Semi-AnnualOff10.410.4
On10.010.0
QuarterlyOff8.78.7
On8.58.5
15%Semi-AnnualOff10.010.2
On9.910.0
QuarterlyOff8.58.6
On8.38.4
50%Semi-AnnualOff9.510.0
On9.39.7
QuarterlyOff8.18.4
On8.08.2
100%Semi-AnnualOff8.99.5
On8.69.3
QuarterlyOff7.68.1
On7.58.0
Maximum energy charging frequency may include full energy therapy shocks or
capacitor formations.
The data provided for programming EGM pre-storage on is based on a 6 month
period (two 3-month follow-up intervals) over the life of the device. Additional use
of EGM pre-storage reduces longevity by approximately 25% or 3 months per
year.
Marquis VR 7230Cx, B, and E Reference Manual
Quick reference
Longevity projections
Table 1-7. Projected longevity in years with 3V pacing amplitude and 0.4
ms pulse width
Maximum
Percent
pacing
energy charging
frequency
a
EGM
pre-storage
0%Semi-AnnualOff10.410.4
On10.010.0
QuarterlyOff8.78.7
On8.58.5
15%Semi-AnnualOff10.010.1
On9.810.0
QuarterlyOff8.58.5
On8.38.3
50%Semi-AnnualOff9.39.7
On9.09.5
QuarterlyOff7.98.3
On7.88.0
100%Semi-AnnualOff8.49.2
On8.29.0
QuarterlyOff7.37.9
On7.17.7
a
Maximum energy charging frequency may include full energy therapy shocks or
capacitor formations.
b
The data provided for programming EGM pre-storage on is based on a 6 month
period (two 3-month follow-up intervals) over the life of the device. Additional use
of EGM pre-storage reduces longevity by approximately 25% or 3 months per
year.
500 ohm
b
impedance
pacing
900 ohm
pacing
impedance
11
Marquis VR 7230Cx, B, and E Reference Manual
12
Chapter 1
Replacement indicators
Replacement indicators
Battery voltage and messages about replacement status appear
on the programmer display and on printed reports. Table 1-8 lists
the Elective Replacement Indicator (ERI) and the End of Life
(EOL) conditions.
Table 1-8. Replacement indicators
Elective Replacement (ERI)≤ 2.62 V
End of Life (EOL)3 months after ERI
ERI date – The programmer displays the date when the battery
reached ERI on the Quick Look and Battery and Lead
Measurements screens.
Temporary voltage decrease – The battery voltage temporarily
decreases following a high voltage charge. If a battery
measurement is taken immediately after a high voltage charge,
the ERI or EOL indicator may be displayed. However, this is a
temporary status which will return to normal when the battery has
recovered from the charge.
EOL indication – If the programmer indicates that the device is at
EOL, replace the device immediately.
Post-ERI conditions – EOL device status is defined as three
months following an ERI indication assuming the following
post-ERI conditions: 100% VVI pacing at 60 ppm (min
0.4 ms; 500 Ω pacing load; and six 30 J charges. EOL may be
indicated before the end of three months if the device exceeds
these conditions.
Marquis VR 7230Cx, B, and E Reference Manual
-1
), 3 V,
Typical charge times
The most recent capacitor charge time appears on the
programmer display and on printed reports and can be evaluated
using the Charge/Dump test (see Table 1-9).
Ta b l e 1 - 9 . T yp i c a la full energy charge times
At Beginning of Life (BOL)5.9 seconds
At Elective Replacement (ERI)7.7 seconds
a
These charge times are typical when the capacitors are fully formed.
High voltage therapy energy
The stored energy of the device is derived from the peak capacitor
voltage and is always greater than the energy delivered by the
device. Table 1-10 compares the programmed energy levels
delivered by the device to the energy levels stored in the
capacitors before delivery.
Quick reference
Typical charge times
13
Marquis VR 7230Cx, B, and E Reference Manual
14
Chapter 1
Stored data and diagnostics
Table 1-10. Comparing delivereda (programmed) and storedb energy levels
Energy (J)Charge
Delivered a/
Stored
b
Programmed
30355.989.41.6
28335.678.31.4
26315.267.11.2
24284.755.91.0
22264.444.70.8
20244.033.60.6
18213.522.40.4
16193.21.82.20.4
15183.01.62.00.3
14162.71.41.70.3
13152.51.21.50.3
12142.41.01.20.2
11132.20.81.00.2
10122.00.60.80.1
9111.90.40.50.1
a
Energy delivered at connector block into a 75 ohm load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth of a
second.
QRS Snapshot
store detailed episode records for NST
episodes)
EGM capacity for
SVT/NST episodes
2 minutes
3.6 minutes
EGM sourcesSix options: ventricular / far-field
EGM optionsStore before onset; Store during charging
a
(the device does not usually
d
of dual-channel EGM, or
e
of single-channel EGM
15
Flashback memory 2000 intervals (V-V): before latest VF, before
latest VT, and before interrogation
Counter data
Detection countersLifetime total, since cleared, and since last
session
Episode countersEpisodes:
■
VF, FVT, and VT
■
NST episodes and SVTs
Percentage pacing:
■
VS and VP percentages
Additional counters:
■
Single PVCs and PVC runs
■
Rate stabilization pulses and runs
Therapy efficacy
counters
Counts for each VF, FVT, VT Therapy:
■
Delivered
■
Successful
■
Unsuccessful
■
Intervention (manually aborted)
Total number of aborted shocks
Other stored data
Patient Alert eventsUp to 10 log entries: text and date for the first
time an alert is triggered between interrogations
Battery and lead
measurements
Battery voltage, last capacitor formation, last
charge, lead impedance, EGM amplitude
measurements, last high voltage therapy, and
sensing integrity counter
Marquis VR 7230Cx, B, and E Reference Manual
16
Chapter 1
Stored data and diagnostics
Table 1-11. Stored data and diagnostics (continued)
Lead performance
trends
14 days of daily measurements plus 80 weeks
of weekly minimum and maximum
measurements:
■
Lead impedance: ventricular pacing,
defibrillation pathway, and SVC lead (if
used)
■
Ventricular EGM amplitude (R-waves)
Cardiac Compass
trends
14 months of measurement trends:
■
VT and VF episodes per day
■
High voltage therapies delivered per day
■
Ventricular rate during VT or VF
■
Periods of non-sustained tachycardia per
day
■
Heart rate variability
■
Percent of time pacing is active
■
Patient activity
■
Average day and night ventricular heart
rate
a
When Wavelet is set to On or Monitor
b
13.5 minutes if Wavelet is set to On or Monitor
c
22 minutes if Wavelet is set to On or Monitor
d
1.6 minutes if Wavelet is set to On or Monitor.
e
2.75 minutes if Wavelet is set to On or Monitor.
Marquis VR 7230Cx, B, and E Reference Manual
New and enhanced features
The following features are new or changed from the
7231 GEM III VR ICD.
Lead connection header options
The Marquis VR supports three types of lead connection headers.
The connection-types are Cx, B, and E.
■
Cx-type connector is normally used with one multipolar
transvenous lead in the ventricle for sensing, pacing, and
delivering therapies.
■
B-type connector is normally used with one multipolar
transvenous lead in the ventricle for sensing and pacing and
with two or three high voltages electrodes placed to deliver
cardioversion/defibrillation therapies.
■
E -type connector is normally used with two 5mm unipolar
myocardial pacing and sensing lead connectors and two
6.5mm high voltage lead connectors.
Regardless of the connector type used, you can program the
Active Can feature so that the device Can serves as a second high
voltage electrode and the HVX port
high voltage electrode if desired. For more information about
connection pathways, see “B- and E-type connector pathways” on
page 123.
Quick reference
New and enhanced features
1
can accommodate a third
17
Patient management
RapidRead telemetry – Communication between the device and
programmer is approximately 20 times faster than telemetry in
previous Medtronic ICD devices. The magnitude of improvement
depends on the amount and type of data that is interrogated.
RapidRead telemetry is more reliable and has an increased range
that makes placing the programming head easier.
Cardiac Compass trends report – This report displays up to
14 months of trend data related to tachyarrhythmia episodes,
heart rate, and patient activity. See “Using Cardiac Compass to
view long term clinical trends” on page 269.
1
The HVA port of the E-type connector serves as the HVX electrode.
Marquis VR 7230Cx, B, and E Reference Manual
18
Chapter 1
New and enhanced features
Patient Alert – The alert duration when a magnet is applied to the
device is now 20 seconds. The device also provides several new
alerts:
■
SVC (HVX) lead impedance out of range
■
Active Can off without SVC
■
VOO mode programmed
■
VF Detection programmed off, or fewer than four VF therapies
enabled for at least six hours
■
charge circuit timeout occurred
■
excessive charge time ERI
For more information, see “Using the Patient Alert feature” on
page 227.
EGM amplitude trends – The device automatically measures
R-wave EGM amplitudes every day. These daily measurements
are included in the data displayed on the Lead Performance
Trends screen. See “Collecting lead performance data” on
page 247.
1
EGM Amplitude test – You can use the EGM Amplitude test to
measure R-wave EGM amplitudes. The results are reported on
the EGM Amplitude test screen. See “Measuring EGM Amplitude”
on page 297.
Lead impedance measurements for SVC (HVX)
other lead impedance measurements, the device provides an
independent SVC (HVX) measurement to check the integrity of
the supplementary high voltage electrode. See “Measuring lead
impedance” on page 295.
Leadless ECG signal – If a supplementary high voltage
electrode is placed in the SVC, the device provides the
Leadless ECG signal through either the Can to SVC
(HVB) to SVC (HVX) EGM source. See “Setting up data collection”
on page 241.
1
SVC refers to the HVX electrode, for pathway information see “B- and E-type
connector pathways” on page 123.
2
The HVA port of the E-type connector serves as the HVX electrode.
Marquis VR 7230Cx, B, and E Reference Manual
2
– Along with
1
(HVX) or RV
Quick reference
New and enhanced features
Expanded pre-onset EGM storage – The device can now store
up to 20 seconds of EGM before a tachycardia starts. See “Setting
up data collection” on page 241.
Smart Auto Cap Formation – When the Auto Cap Formation
Interval is set to Auto, the formation interval automatically adjusts
to optimize device longevity and charge times. See “Smart Auto
Cap” on page 180.
Ending a patient session – The device audits the programmed
parameter settings when you end a patient session and alerts you
if any of the settings are atypical. See “Starting and ending patient
sessions” on page 204.
Wavelet test – The Wavelet test allows you to manually collect
and assess the template used by the Wavelet criterion. See
“Testing the Wavelet criterion” on page 291.
Auto Collection for Wavelet – The Wavelet criterion includes the
option to have the device automatically collect and maintain the
template used to distinguish between SVT and ventricular
tachyarrhythmia. See “Details about Auto Collection” on page 101.
19
QRS Snapshot data – When the Wavelet criterion is
programmed to On or Monitor, the device stores QRS Snapshot
data with VT, VF, FVT and SVT episode records. This data
includes graphical representations of up to eight QRS complexes,
along with match scores and event classifications as “Match” or
“No Match.”
Tachyarrhythmia detection
Wavelet Dynamic Discrimination criterion – The Wavelet
criterion is designed to prevent detection of rapidly conducted
SVTs as ventricular tachyarrhythmias. It compares the shape of
each QRS complex during a fast ventricular rate to a stored
template. If enough events occur that match this template, Wavelet
withholds detection. See “Enhancing detection with Wavelet” on
page 94.
Marquis VR 7230Cx, B, and E Reference Manual
20
Chapter 1
New and enhanced features
Tachyarrhythmia therapy
VT Monitoring – VT detection can be set to Monitor, which allows
the device to detect and record VT episodes without delivering
therapy or influencing VF detection. See “Monitoring episodes for
termination or redetection” on page 90.
Onset criterion – The Onset criterion is designed to prevent
detection of sinus tachycardia as VT by requiring that a rapid
increase in ventricular rate occurs before VT events can be
classified. See “Details about Onset” on page 104.
High Rate Timeout – High Rate Timeout can turn off detection
enhancements (Wavelet, Onset, or Stability) if a high rate episode
is longer than a programmed duration. See “Details about High
Rate Timeout” on page 112.
Episode confirmation during and after charging – The device
continually monitors the ventricular rhythm during and after
charging for cardioversion or defibrillation (when VF confirmation
is active) to ensure the arrhythmia is present before delivering the
high voltage shock. See “Confirming VF after initial detection” on
page 126 and “Confirming VT or FVT after detection” on
page 145.
Programmable Active Can – If a supplementary electrode is
connected to the SVC (HVX) port, you can deselect the device
Can as a high voltage electrode. For more information about
connection pathways, see “Delivery pathway electrodes” on
page 123.
Bradycardia pacing
Accelerometer-based rate response – The device uses an
accelerometer to provide rate responsive pacing.
Additional bradycardia pacing modes – The device provides
asynchronous pacing in the VOO pacing mode, and provides the
OVO mode to disable pacing. See Chapter 8, "Treating
bradycardia" on page 157.
Marquis VR 7230Cx, B, and E Reference Manual
EP studies
Quick reference
New and enhanced features
Defibrillation threshold testing support – The T-Shock and
50 Hz Burst induction screens allow you to monitor time between
inductions, program ventricular sensing and VF therapy settings,
adjust induction settings, select manual therapies, and retrieve
episode records after therapy. See “How to perform defibrillation
threshold testing” on page 50.
21
Marquis VR 7230Cx, B, and E Reference Manual
22
Chapter 1
New and enhanced features
Marquis VR 7230Cx, B, and E Reference Manual
System overview 24
Indications and usage 27
Contraindications 27
Patient screening 27
The Marquis VR system2
2
Marquis VR 7230Cx, B, and E Reference Manual
24
Chapter 2
System overview
System overview
The 7230 Marquis VR Single Chamber Implantable Cardioverter
Defibrillator (ICD) system is an implantable medical device system
that automatically detects and treats episodes of ventricular
fibrillation, ventricular tachycardia, fast ventricular tachycardia,
and bradyarrhythmia. The ICD system includes three major
components:
■
ICD
The ICD senses the electrical activity of the patient’s heart via
the sensing electrodes of the implanted ventricular leads. It
then analyzes the heart rhythm based on selectable sensing
and detection parameters. If the ICD detects a
tachyarrhythmia, it delivers defibrillation, cardioversion, or
antitachycardia pacing therapy to the patient’s heart. If the ICD
identifies a bradyarrhythmia, it delivers bradycardia pacing
therapy to the patient’s heart.
■
Leads
The ICD can be used with transvenous or epicardial
defibrillation leads. The lead system should consist of a
bipolar pacing/sensing lead (or paired unipolar
1
pacing/sensing leads) in the ventricle and one or two high
voltage cardioversion/ defibrillation electrodes. You can
program the Active Can device case as a high voltage
electrode. The pacing and sensing electrodes sense cardiac
activity and deliver pacing stimuli.
■
Programmer and software
2
The Medtronic 9790C
programmer and 9967 application
software allow you to
■
configure the detection, therapy, and bradycardia features
for your patient
■
perform electrophysiological studies and system tests
■
monitor, display, or print patient cardiac activity information
1
With an appropriate unipolar to bipolar adapter kit.
2
With the model 9767 or 9767L programming head
Marquis VR 7230Cx, B, and E Reference Manual
For information about:
■
indications, contraindications, lead compatibility, warnings and
precautions, and patient selection, see the Marquis VR 7230 Implant Manual, which accompanies each device.
■
basic programmer and software desktop functions that are not
included in Chapter 10, "Using the programmer" on page 185,
see the manual accompanying the programmer.
■
installing the 9767 or 9767L programming head, see the
manual accompanying the programming head.
■
implanting leads, refer to the manuals accompanying the
leads.
Detecting and treating tachyarrhythmias
The ICD monitors the cardiac rhythm for short ventricular intervals
that may indicate the presence of VF, VT, or FVT.
■
Upon detection of VF, the ICD delivers a biphasic defibrillation
shock of up to 30 joules. If the VF episode persists, up to five
more individually programmed defibrillation shocks can be
delivered.
■
Upon detection of VT, the ICD delivers either a Ramp, Ramp+,
or Burst antitachycardia pacing therapy or a biphasic
cardioversion shock of up to 30 joules synchronized to a
ventricular depolarization. If the VT episode persists, up to five
more individually programmed VT therapies can be delivered.
You can also program the ICD to monitor the VT episode
without delivering therapy.
■
Upon detection of FVT, the ICD delivers either a Ramp,
Ramp+, or Burst antitachycardia pacing therapy, or a biphasic
cardioversion shock of up to 30 joules synchronized to a
ventricular depolarization. If the FVT episode persists, up to
five more individually programmed FVT therapies can be
delivered.
The Marquis VR system
System overview
25
You can use the Wavelet, Stability and Onset criteria to help the
ICD distinguish between true ventricular arrhythmias and rapidly
conducted supraventricular tachycardia (SVT) and withhold
therapy for SVT.
Marquis VR 7230Cx, B, and E Reference Manual
26
Chapter 2
System overview
Treating bradycardia
The ICD provides rate responsive ventricular pacing to treat
bradycardia. An internal accelerometer senses the patient’s
physical activity, allowing the ICD to increase and decrease the
pacing rate in response to changes in the level of activity.
Monitoring for real-time and stored data
The ICD and programmer provide real-time information on
detection and therapy parameters and status during a patient
session. The ICD also provides accumulated data on device
operation, including stored electrograms, detected and treated
tachyarrhythmia episodes, bradycardia interventions, and the
efficacy of therapy. The Cardiac Compass report provides up to
14 months of clinically significant data, including arrhythmia
episodes, therapies delivered, physical activity, heart rate, and
bradycardia pacing activities.
All of this information can be printed and retained in the patient’s
file or saved in electronic format on a floppy diskette.
Conducting electrophysiologic tests
You can use the system to conduct non-invasive
electrophysiologic studies including manual delivery of any of the
ICD therapies to manage an induced or spontaneous
tachyarrhythmia.
Alerting the patient to system events
You can use the programmable Patient Alert monitoring feature to
notify the patient with audible tones if certain conditions related to
the leads, battery, charge time and therapies occur. The patient
can then respond based on your prescribed instructions.
Marquis VR 7230Cx, B, and E Reference Manual
Indications and usage
The implantable cardioverter defibrillator is intended to provide
ventricular antitachycardia pacing and ventricular defibrillation for
automated treatment of life threatening ventricular arrhythmias.
Contraindications
The Marquis VR system is contraindicated for
■
patients whose tachyarrhythmias may have transient or
reversible causes, such as acute myocardial infarction,
digitalis intoxication, drowning, electrocution, electrolyte
imbalance, hypoxia, or sepsis.
■
patients with incessant VT or VF
■
patient who have a unipolar pacemaker
■
patients whose primary disorder is bradyarrhythmias or atrial
arrhythmias.
The Marquis VR system
Indications and usage
27
Patient screening
Prior to implant, patients should undergo a complete cardiac
evaluation, including electrophysiologic testing. Also,
electrophysiologic evaluation and testing of the safety and efficacy
of the proposed tachyarrhythmia therapies are recommended
during and after the implantation of the device.
Other optional screening procedures could include exercise stress
testing to determine the patient’s maximum sinus rate, and cardiac
catheterization to determine if there is a need for concomitant
surgery and/or medical therapy.
Marquis VR 7230Cx, B, and E Reference Manual
28
Chapter 2
Patient screening
Marquis VR 7230Cx, B, and E Reference Manual
Emergency therapy3
Delivering emergency therapies 30
How to deliver emergency 30 joule defibrillation 32
How to deliver emergency cardioversion 32
How to deliver emergency fixed burst pacing 33
How to deliver emergency VVI pacing 33
3
Marquis VR 7230Cx, B, and E Reference Manual
30
Chapter 3
Delivering emergency therapies
Delivering emergency therapies
The device provides the following emergency therapies:
■
defibrillation
■
cardioversion
■
fixed burst pacing
■
emergency VVI pacing
The default emergency therapy is 30 joule defibrillation. When you
select [Emergency] and [DELIVER], the device charges and
delivers a biphasic 30 joule shock along the AX>B pathway.
The programmer resets the emergency defibrillation energy to
30 joules each time you select [Emergency]. Emergency
cardioversion and fixed burst values remain as selected for the
duration of the session.
To return to other programming functions from an Emergency
screen, select [Exit Emergency].
Effect on system operation
1
The device suspends the automatic detection features when
emergency defibrillation, cardioversion, or fixed burst pacing
therapies are delivered. Detection is not suspended during
emergency VVI pacing. Removing the programming head or
pressing [Resume] turns detection on again.
Aborting an emergency therapy
As a safety precaution, the programmer also displays an [ABORT]
button which immediately terminates any emergency therapy in
progress.
1
If Active Can is turned off, the defibrillation is delivered between the HVX and
HVB electrodes. . For more information, see “Delivery pathway electrodes” on
page 123
Marquis VR 7230Cx, B, and E Reference Manual
On-screen and display panel buttons
The on-screen [Emergency] button and the red mechanical
Emergency button by the programmer display panel function the
same at all times.
Emergency therapy
Delivering emergency therapies
31
Red Emergency
The on-screen [DELIVER] button and the yellow-on-blue
mechanical Deliver button by the programmer display panel
function the same during emergency operations only. The
mechanical Deliver button operates only during emergency
operations.
Yellow-on-blue
Deliver Button
Temporary parameter values
Emergency tachyarrhythmia therapies use temporary values that
do not change the programmed parameters of the device.
values are not in effect until you select [DELIVER]. After the
tachyarrhythmia therapy is complete, the device reverts to its
programmed values.
Button
Functions the same as
on-screen [Emergency] button.
Functions the same as on-screen
[DELIVER], but only during
Emergency functions.
1
These
1
Delivery of Emergency VVI Pacing changes the programmed bradycardia
pacing values to the emergency values (see page 33).
Marquis VR 7230Cx, B, and E Reference Manual
32
Chapter 3
Delivering emergency therapies
How to deliver emergency 30 joule defibrillation
3
4
2
How to deliver emergency cardioversion
3
4
5
2
1. Position the programming
head over the device.
2. Select [Emergency].
3. Accept the defibrillation
energy shown on the screen,
or select Energy and select a
new value from the window.
4. Select [DELIVER].
If delivery is not confirmed,
verify that the programming
head is properly positioned
and select [Retry] or [Cancel].
1. Position the programming
head over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the cardioversion
energy shown on the screen,
or select Energy and select a
new value from the window.
5. Select [DELIVER].
If delivery is not confirmed,
verify that the programming
head is properly positioned
and select [Retry] or [Cancel].
Marquis VR 7230Cx, B, and E Reference Manual
How to deliver emergency fixed burst pacing
3
4
5
2
How to deliver emergency VVI pacing
3
4
2
Emergency therapy
Delivering emergency therapies
1. Position the programming
head over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval
shown on the screen, or
select Interval for a new
interval value.
5. Select [BURST Press and
Hold].
If delivery is not confirmed,
the programmer displays
an error window. Verify that
the programming head is
properly positioned. Select
[OK] from the window and
reselect [BURST Press and
Hold].
1. Position the programming
head over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A
successful programming
sets the device to the
following maximum output
bradycardia pacing values.
■
Pacing Mode: VVI
■
Lower Rate: 70 ppm
(70 min-1)
■
V. Amplitude: 8 V
■
V. Width: 1.6 ms
■
V. Pace Blanking: 320 ms
■
Hysteresis: Off
■
Vent r i cula r Rate
Stabilization: Off
If programming is not
confirmed, verify that the
programming head is
properly positioned and
select [Retry] or [Cancel].
33
Marquis VR 7230Cx, B, and E Reference Manual
34
Chapter 3
Delivering emergency therapies
Marquis VR 7230Cx, B, and E Reference Manual
Part II
Device implant and patient follow-up
procedures
Marquis VR 7230Cx, B, and E Reference Manual
Implanting the ICD4
Overview 38
Preparing for an implant 38
Replacing an ICD 40
Positioning the leads 41
Testing sensing and pacing thresholds 43
Connecting the leads to the ICD 44
Testing defibrillation operation and effectiveness 47
Positioning and securing the ICD 52
Completing the implant procedure 52
4
Marquis VR 7230Cx, B, and E Reference Manual
38
Chapter 4
Overview
Overview
Preparing for an implant
The tasks for implanting an ICD include
1. Preparing for an implant
2. Replacing an ICD
3. Positioning the leads
4. Testing sensing and pacing thresholds
5. Connecting the leads to the ICD
6. Testing defibrillation operation and effectiveness
7. Positioning and securing the ICD
8. Completing the implant procedure
These tasks are described in the sections that follow.
Warning: Keep a back-up external defibrillator available
during the implant for transthoracic rescue when arrhythmias
are induced.
Equipment for an implant
■
Model 9790C programmer and Model 9767 or 9767L
programming head
■
9967 software application
■
8090 Analyzer lead analysis device or equivalent pacing
system analyzer
■
external defibrillator
■
5358 Defibrillator Implant Support Device and software
application (optional)
Marquis VR 7230Cx, B, and E Reference Manual
Sterile supplies
■
Marquis VR ICD and lead system components
■
Programming head sleeve or programming head
■
Analyzer cables
■
Lead introducers appropriate for the lead system
■
Extra stylets of appropriate length and shape
How to prepare for implanting
Set up the implant support instrument
When using an implant support instrument such as the 5358
Defibrillator Implant Support Device:
1. Calibrate any monitoring or recording equipment while recording
the EGM and marker outputs of the support instrument.
2. Verify the high energy output of the support instrument by delivering
a high energy defibrillation shock into the test load.
Set up the programmer and start the application
1. Set up the programmer as described in the instructions provided
with the programmer.
2. Install the Marquis VR Model 9967 software on the programmer, if it
is not already installed.
3. Place the programming head over the device and start the
application. Select the device model or select [Auto identify].
Note: The programmer automatically interrogates the device when
the application starts.
Implanting the ICD
Preparing for an implant
39
Preprogram the device
Before opening the sterile package, prepare the ICD for implant as
follows:
1. Check the “use by” date printed on the package. Do not implant the
device after the “use by” date because the battery’s longevity could
be reduced.
2. Interrogate the ICD, and print a full summary report.
3. Confirm that the battery voltage is at least 3.0 V at room
temperature (see page 225).
4. Perform a manual capacitor formation (see page 299).
5. Set up data collection parameters and the ICD internal clock (see
page 243).
6. Program the therapy and pacing parameters to values appropriate
for the patient (see page 160). Ensure that all tachyarrhythmia
detection is programmed Off (see page 64).
Marquis VR 7230Cx, B, and E Reference Manual
40
Chapter 4
Replacing an ICD
Replacing an ICD
If you are replacing a previously implanted ICD, turn off ICD
detection and therapies before explanting.
When implanting the ICD with a chronic lead system, perform the
following evaluations to ensure appropriate detection and therapy:
■
Check the integrity of the chronic high voltage leads with a test
shock, chest X-ray, and inspection.
■
Perform chronic pacing and sensing measurements.
■
Measure high voltage lead impedances.
■
Test defibrillation efficacy.
■
Confirm adequate sensing during VF.
■
Ensure proper fit of the lead connectors in the ICD connector
block.
Notes:
■
To meet the implant requirements, it may be necessary to
reposition or replace the chronic leads or to add a third high
voltage electrode.
■
Any unused leads that remain implanted must be capped.
How to explant and replace an ICD
1. Program all tachyarrhythmia detection Off.
2. Dissect the leads and the ICD free from the surrounding tissues in
the surgical pocket. Be careful not to nick or breach the lead
insulation during the process of exposing the system.
3. Loosen each setscrew, and gently retract the lead from the
connector block.
4. Remove the ICD from the surgical pocket.
5. If the connector pin of any implanted lead shows signs of pitting or
corrosion, replace the implanted lead with a new lead. The
damaged lead should be discarded and replaced to assure the
integrity of the device system.
6. Measure sensing, pacing, and defibrillation efficacy using the
replacement ICD or an implant support instrument.
7. Evaluate the defibrillation efficacy of the replacement system.
Marquis VR 7230Cx, B, and E Reference Manual
Positioning the leads
Implant endocardial leads according to the supplied instructions,
unless suitable chronic leads are already in place. Do not use any
lead with this device without first verifying connector compatibility
(refer to the Marquis VR 7230Cx, 7230B, 7230E Implant Manual).
Transvenous or epicardial leads may be used.
Using transvenous leads
Use standard transvenous implant techniques to position the
ventricular lead tip in the right ventricular apex.
Follow the general guidelines below for initial positioning of other
transvenous leads (the final positions are determined by
defibrillation efficacy tests):
■
SVC (HVX) lead: Place the lead tip high in the innominate
vein, approximately 5 cm proximal to the right atrium (RA) and
SVC junction.
■
SQ patch: Place the patch along the left mid-axillary, centered
over the fourth-to-fifth intercostal space.
■
CS lead: Advance the lead tip to just under the left atrial
appendage, if possible.
Implanting the ICD
Positioning the leads
41
If using a subclavian approach, position the lead laterally to avoid
pinching the lead body between the clavicle and the first rib.
Using epicardial leads
A variety of surgical approaches can be used to implant epicardial
leads, including a limited left thoracotomy or median sternotomy.
A typical placement may use an anterior right ventricular patch as
the RV (HVB) and a posterolateral left ventricular patch as SVC
(HVX).
Warning: Pinching the lead can damage the lead conductor
or insulation, which may cause unwanted high voltage
therapies or result in the loss of sensing or pacing therapy.
Marquis VR 7230Cx, B, and E Reference Manual
42
Chapter 4
Positioning the leads
Surgical incisions
Follow the general guidelines below for positioning epicardial
leads:
■
If unipolar epicardial pacing leads are used, position the
electrodes about 1 to 2 cm apart to reduce electromagnetic
interference, and route the leads together with several loose
twists.
■
Suture the smooth face of each patch lead against the
epicardium or pericardium in locations that produce optimal
defibrillation.
■
Place the patches so that they encompass the maximum
amount of cardiac mass and they have approximately equal
amounts of mass between them.
■
Ensure that the patches do not overlap and the electrode
portions do not touch.
■
Avoid placing extra-pericardial patches over the phrenic nerve.
A single-incision submuscular or subcutaneous approach is
recommended when the ICD is implanted in the pectoral region.
Make the implant pocket about 1.5 times the size of the ICD.
Submuscular implant – An incision extending over the
deltoid-pectoral groove typically provides access to the cephalic
and subclavian veins as well as the implant pocket. Place the ICD
sufficiently medial to the humeral head to avoid interference with
shoulder motion.
Subcutaneous implant – A transverse incision typically permits
isolation of the cephalic vein. Place the ICD far medially to keep
the leads away from the axilla. Make sure that the upper edge of
the ICD remains inferior to the incision.
Marquis VR 7230Cx, B, and E Reference Manual
Testing sensing and pacing thresholds
Testing sensing and pacing thresholds
Sensing and pacing tests include the following measurements:
■
EGM amplitude
■
slew rate
■
pacing threshold
■
pacing lead impedance
Medtronic recommends that you use an 8090 Analyzer lead
analysis device to perform sensing and pacing measurements.
Refer to the technical manual for the analyzer you use to find
details on performing sensing and pacing measurements.
Parameters
Measured sensing and pacing values must meet the following
specific requirements at implant.
Table 4-1. Sensing and pacing values at implant
MeasurementAcute Transvenous Leads Chronic Leads
R- wave amplitude ≥ 5 mV≥ 3 mV
Ventricular slew rate: ≥ 0.75 V/s≥ 0.5 V/s
Ventricular capture
threshold:
a
At 0.5 ms pulse width
a
≤ 1.0 V≤ 3.0 V
Implanting the ICD
43
Considerations
When measuring sensing and pacing values, measure between
the tip (cathode) and ring or coil (anode) of each bipolar
pacing/sensing lead.
For unipolar epicardial pacing leads, either electrode can be the
cathode; use the configuration that yields the lower pacing
threshold.
Note: Do not measure the intracardiac EGM telemetered from the
ICD to assess sensing.
Marquis VR 7230Cx, B, and E Reference Manual
44
Chapter 4
Connecting the leads to the ICD
Connecting the leads to the ICD
For more detailed information about lead/connector compatibility,
see the Marquis VR 7230Cx, 7230B, 7230E Implant Manual, or
contact Medtronic Technical Services at 1-800-723-4636.
Warning: Loose lead connections may result in inappropriate
sensing and failure to deliver necessary arrhythmia therapy.
Caution: Use only the torque wrench supplied with the
device. It is designed to prevent damage to the device from
overtightening a setscrew.
Table 4-2. Cx-type connector characteristics
General description
One IS-1 connector for pacing and
sensing, Two DF-1 connectors for
high voltage therapy, Active Can
electrode (programmable)
The HVA and HVX ports are electronically connected and are treated as the HVX
electrode. For more information see “B- and E-type connector pathways” on
page 123.
HVA
HVX
a
a
6.5 mmHVX
6.5 mmHVX
Cann/aHVA, Can
P/S bipolar
pace sense
3.2 mm bipolar
(IS-1 compatible)
Figure 4-2. B-type connector
Software
Name
45
7
2
6
5
3
4
1 suture holes
2 HVB port (6.5 mm)
3 HVA port (6.5 mm)
4 Programmable Active Can
5 HVX port (6.5 mm)
6/7 P+/S and P-/S port (3.2 mm)
1
Marquis VR 7230Cx, B, and E Reference Manual
46
Chapter 4
Connecting the leads to the ICD
Table 4-4. E-type connector characteristics
General description
Two 6.5mm unipolar high voltage
ports and two 5mm unipolar
pace/sense ports.
a
The HVA port of the E-type connector can be used as the HVX electrode when
Active Can is programmed off, see “B- and E-type connector pathways” on
page 123.
Figure 4-3. E-type connector
Device
Port
Connector
Type
Software
Name
P+/S port5.0 mm
HVA port6.5 mmHVX
a
HVB port6.5 mmHVB
Cann/aHVA, Can
P-/S port5.0 mm
6
5
2
3
4
1 suture holes
2 P+/S port (5.0 mm)
3 HVA port (6.5 mm)
4 Programmable Active Can
5 HVB port (6.5 mm)
6 P-/S port (5.0 mm)
For easier lead insertion, insert the ventricular IS-1 leg before the
other legs.
1
Marquis VR 7230Cx, B, and E Reference Manual
Testing defibrillation operation and effectiveness
How to connect the lead to the device
1
a
2
7230Cx connector shown in example
b
30SetScrew.eps
30LeadTIp.eps
Implanting the ICD
1. Insert the torque wrench into the
appropriate setscrew.
a. If the port is obstructed, retract the
setscrew to clear it. Take care not to
disengage the setscrew from the
connector block.
b. Leave the torque wrench in the setscrew
until the lead is secure. This allows a
pathway for venting trapped air when the
lead is inserted.
2. Push the lead or plug into the connector
port until the lead pin is clearly visible in the
pin viewing area. No sealant is required, but
sterile water may be used as a lubricant.
3. Tighten the setscrew by turning clockwise
until the torque wrench clicks.
4. Tug gently on the lead to confirm a secure
fit. Do not pull on the lead until all setscrews
have been tightened.
5. Repeat these steps for each lead.
47
Testing defibrillation operation and effectiveness
To demonstrate reliable defibrillation effectiveness with the
implanted lead system, use the ICD to complete one of the
following tests:
■
Terminate two consecutively induced VF episodes using a
delivered energy of 20 joules or less.
■
Use the binary search procedure to establish a defibrillation
threshold (DFT) of 18 joules or less.
Note: An external defibrillation implant support instrument can
also be used to test defibrillation effectiveness. For instructions,
see the applicable documentation for the specific support
instrument.
Marquis VR 7230Cx, B, and E Reference Manual
48
Chapter 4
Testing defibrillation operation and effectiveness
High voltage implant values
Binary search protocol
Measured values must meet the following requirements at implant.
Table 4-5. High voltage therapy values at implant
MeasurementAcute or Chronic Leads
V. Defib impedance
SVC (HVX) impedance (if applicable)
Defibrillation threshold
Warning: Ensure that an external defibrillator is charged for a
rescue shock.
By testing defibrillation efficacy at successively lower energy
levels, the binary search protocol provides an accurate threshold
measurement. However, it generally requires more inductions and
more time. Also, cumulative charging of the ICD capacitor affects
the device longevity.
For most reliable defibrillation efficacy testing, allow at least five
minutes between VF inductions. The ICD software provides an
on-screen timer to record elapsed time since the last induction.
If a two-electrode system fails to meet the implant criterion,
consider implanting a third electrode and connecting it to the SVC
(HVX) port. You can also evaluate the efficacy of ventricular
tachycardia therapies after successfully completing ventricular
defibrillation testing.
1
The timer appears on the 50 Hz Burst and T-Shock induction screens.
Marquis VR 7230Cx, B, and E Reference Manual
1
Figure 4-4. Binary search protocol
Implanting the ICD
Testing defibrillation operation and effectiveness
Start
49
First induced episode
Second induced episode
Third induced episode
Defibrillation
Threshold
SuccessFailureFailureSuccess
3 J
SuccessFailureSuccessFailureSuccessFailure
<=3 J6 J9 J15 J18 J
SuccessFailure
6 J
9 J15 J
Implant
Criterion Met
12 J
12 J
How to prepare for defibrillation threshold testing
1. Place the programming head over the ICD, start a patient session,
and interrogate the device, if you have not already done so.
2. Observe the Marker Channel telemetry annotations and the
programmer ECG display to verify that the ICD is sensing properly.
3. Conduct a manual Lead Impedance Test
lead connections. Perform this test with the ICD in the surgical
pocket and keep the pocket very moist. If the impedance is out of
range, perform one or more of the following tasks:
■
Recheck lead connections and electrode placement.
■
Repeat the measurement.
■
Inspect the bipolar EGM for abnormalities.
■
Measure the defibrillation impedance with a manual test shock.
4. Program the ICD or support instrument to properly detect VF with
an adequate safety margin (1.2 mV sensitivity).
a
See “Measuring lead impedance” on page 295.
18 J
Success
Reprogram polarity, reposition the
lead, or try a different lead
a
Try to meet
2 at 20 J
Failure
to verify the defibrillation
74binary.eps
Marquis VR 7230Cx, B, and E Reference Manual
50
Chapter 4
Testing defibrillation operation and effectiveness
How to perform defibrillation threshold testing
3
2
4
6
5
1. Select Tests > EP Study.
2. Select either 50 Hz BURST or
T-shock induction.
3. Select [Resume at BURST] or
[Resume at DELIVER].
4. Select [Adjust Permanent...].
1
5. Program VF Enable On.
6. Program the automatic therapy
energy settings. Therapies 2-6
should be set to the maximum
energy.
7. Select [Program].
7
8
8. Select [Close].
Marquis VR 7230Cx, B, and E Reference Manual
13
11
12
Implanting the ICD
51
Testing defibrillation operation and effectiveness
9. If performing a T-Shock
induction, select the [Enable]
checkbox.
10. Select [DELIVER], or [50 Hz
BURST Press and Hold].
If necessary, you can abort an
9
10
14
induction or therapy in
progress by pressing [ABORT].
11. Observe the live rhythm
monitor for proper post-shock
sensing.
12. If using the binary search
protocol, use the [Adjust
Permanent...] button to
program the next appropriate
energy level (see Figure 4-4).
13. Wait until the on-screen timer
reaches 5 minutes, then repeat
steps 9 through 12 as needed.
14. Select Params > Detection and
program VF, FVT, and VT
detection Off before closing.
Marquis VR 7230Cx, B, and E Reference Manual
52
Chapter 4
Positioning and securing the ICD
Positioning and securing the ICD
Cautions:
■
If no SVC electrode is implanted, the pin plug provided
with the device must be secured in the SVC port.
■
Program tachyarrhythmia detection Off before closing.
How to position and secure the device
Suture Hole Locations
1. Ensure that each lead pin or plug is fully
inserted into the connector block and that all
setscrews are tight.
2. Coil any excess lead length beneath the device.
Avoid kinks in the lead conductors.
3. Implant the device within 5 cm of the skin. This
position optimizes the ambulatory monitoring
operations.
4. Suture the device securely within the pocket to
30Suture.eps
minimize post-implant rotation and migration of
the device. Use a surgical needle to penetrate
the suture holes. See Figure 4-2 and Figure 4-3
for suture hole locations on the B- and E-type
connectors.
Completing the implant procedure
After implanting the device, X-ray the patient to verify the device
and leads placement. To complete programming the device, select
parameters that are appropriate for the patient.
Marquis VR 7230Cx, B, and E Reference Manual
How to complete programming the device
1. After closing the pocket, program detection On. Program ventricular
tachyarrhythmia therapies On as desired.
2. If external equipment was used to conduct the defibrillation efficacy
tests, perform a final VF induction, and allow the implanted system to
detect and treat the arrhythmia.
3. Monitor the patient after the implant, and take X-rays as soon as
possible to document and assess the location of the leads.
4. Program patient information. See “How to view and enter new patient
information” on page 277.
5. Configure the Patient Alert feature. See “Using the Patient Alert
feature” on page 227.
6. Set up data collection parameters. See “Setting up data collection” on
page 241.
7. Interrogate the device after any spontaneous episodes to evaluate
the detection and therapy parameter settings.
8. If the patient has not experienced spontaneous episodes, you may
induce the clinical tachyarrhythmias using the non-invasive EP Study
features to further assess the performance of the system. See
Chapter 14, "Conducting Electrophysiologic Studies" on page 303.
9. Recheck pacing and sensing values, and adjust if necessary.
Implanting the ICD
Completing the implant procedure
53
Marquis VR 7230Cx, B, and E Reference Manual
54
Chapter 4
Completing the implant procedure
Marquis VR 7230Cx, B, and E Reference Manual
Conducting a patient follow-up
Patient follow-up guidelines 56
Verifying the status of the implanted system 56
Verifying accurate detection and appropriate therapy 57
Verifying effective bradycardia pacing 58
session
5
5
Marquis VR 7230Cx, B, and E Reference Manual
56
Chapter 5
Patient follow-up guidelines
Patient follow-up guidelines
Schedule regular patient follow-up sessions to monitor the
condition of the ICD and leads and to verify that the ICD is
configured appropriately for your patient.
During the first few months after receiving a new device, the
patient may require close monitoring. Schedule an office visit at
least every three months.
The Quick Look screen, which is displayed after you interrogate
the device, provides a good beginning for the follow-up review.
Using this screen you can
■
verify that the device is functioning correctly.
■
review the clinical performance and long term trends.
■
print appropriate reports1 to compare the results to the
patient’s history and to retain for future reference.
Note: The Checklist feature provides a standard list of tasks to
perform at a complete follow-up visit. You can also customize your
own checklists if you wish. See “Streamlining follow-ups with
Checklist” on page 235 for more information.
Verifying the status of the implanted system
To verify that the ICD and leads are functioning correctly, review
the following information from the Quick Look screen and perform
follow-up tests as indicated:
■
Review the displayed battery voltage for comparison to the
Elective Replacement Indicator value (see page 12).
Remember that battery voltage may be low if high voltage
charging has occurred within 24 hours.
■
Review the last full energy charge.
– For information about adjusting the capacitor formation
interval, see “Optimizing charge time” on page 178.
– If the programmer displays an Excessive Charge Time ERI,
the ICD should be replaced immediately.
1
See “Using Cardiac Compass to view long term clinical trends” on page 269 for
information on this new report.
Marquis VR 7230Cx, B, and E Reference Manual
Conducting a patient follow-up session
Verifying accurate detection and appropriate therapy
■
Review the defibrillation and pacing lead impedance values for
inappropriate values or large changes since the last follow-up.
See “Measuring lead impedance” on page 295.
■
Perform an EGM Amplitude test for comparison to previous
EGM Amplitude measurements. See “How to perform an EGM
Amplitude test” on page 298.
■
To review longer term trends in sensing and impedance
measurements, select the [>>] button from the lead
impedance area of the Quick Look screen. The programmer
displays a detailed history of automatic sensing and
impedance measurements. See “Taking a quick look at device
activity” on page 225.
Verifying accurate detection and appropriate therapy
To verify that the ICD is providing effective tachyarrhythmia
detection and therapy, review the following information from the
Quick Look screen and investigate as indicated:
■
Review Quick Look Observations that relate to patient history
and device operation. To display more detailed information
about any observation, select the observation and then select
the [>>] button.
■
Review any Patient alerts listed in the Observations of the
Quick Look screen. For the most detailed information about
Patient Alerts, select Patient Alert from the Data icon and
select [Events].
■
Check stored episode records for appropriate sensing and
detection of arrhythmias. See “Viewing episode data” on
page 254.
■
Check stored SVT episode records for appropriate
identification of SVTs.
57
Considerations
Review the following information before verifying detection and
therapy.
Flashback memory – In addition to the episode text and stored
electrograms, use Flashback memory and interval plots to help
investigate the accuracy and specificity of ventricular detection.
Marquis VR 7230Cx, B, and E Reference Manual
58
Chapter 5
Verifying effective bradycardia pacing
Episode misidentification – If the episode records indicate that
false detections have occurred, the Sensing Integrity counter may
help in determining the prevalence of oversensing. For more
information, see “Sensing integrity counter” on page 249.
If the ICD is oversensing, consider these programming options:
■
Increase the Pace Blanking value.
■
Increase the sensitivity threshold.
Caution: Do not re-program the ICD to decrease oversensing
without assuring that appropriate sensing is maintained, See
“Setting up sensing” on page 67.
If the episode records reveal that a stable monomorphic VT has
been identified and treated as VF, consider these options to
improve the detection accuracy:
■
Review the Interval Plot for the episode, and adjust
VF Interval, if necessary. Use caution when reprogramming
the VF Interval, because changes to this value can adversely
affect VF detection.
■
Consider enabling FVT via VF detection. See “Detecting FVT
episodes” on page 82.
If the SVT episode records include episodes of true VT, review the
SVT episode record to identify the SVT detection criterion that
withheld detection. Adjust the SVT detection criteria parameters
as necessary. See “Enhancing detection with Wavelet” on
page 94, “Enhancing VT detection with the Onset criterion” on
page 103, and“Enhancing VT detection with the Stability criterion”
on page 108.
Verifying effective bradycardia pacing
To verify that the ICD is sensing and pacing appropriately, review
the following information from the Quick Look screen and
investigate as indicated:
■
Confirm that the patient is receiving adequate cardiac support
for daily living activities.
■
Review the pacing history for comparison to the patient
history. A sharp increase in the paced beats percentage may
indicate a need for investigation and analysis.
Marquis VR 7230Cx, B, and E Reference Manual
Considerations
Conducting a patient follow-up session
Verifying effective bradycardia pacing
■
Review the Cardiac Compass report for comparison to patient
history (see page 269).
■
Conduct a pacing threshold test (see page 288) to verify that
the programmed pacing parameters provide a sufficient safety
margin.
Review the following information before verifying bradycardia
pacing.
Ventricular Pacing – If the ventricle is predominantly paced and
the patient exhibits adequate ventricular response, consider
decreasing the Lower Rate.
% Pacing – Because the percentages for this counter are
rounded to the nearest unit, they may not add up to 100%.
59
Marquis VR 7230Cx, B, and E Reference Manual
60
Chapter 5
Verifying effective bradycardia pacing
Marquis VR 7230Cx, B, and E Reference Manual
Part III
Configuring the ICD for the patient
Marquis VR 7230Cx, B, and E Reference Manual
Detecting tachyarrhythmias6
Detection overview 64
Setting up sensing 67
Detecting VF episodes 71
Detecting VT episodes 76
Detecting FVT episodes 82
Detecting tachyarrhythmia episodes with Combined Count 87
Monitoring episodes for termination or redetection 90
Enhancing detection with Wavelet 94
Enhancing VT detection with the Onset criterion 103
Enhancing VT detection with the Stability criterion 108
6
Detecting prolonged tachyarrhythmias with High Rate Timeout 111
Key ter ms 113
Marquis VR 7230Cx, B, and E Reference Manual
64
Chapter 6
Detection overview
Detection overview
The device detects ventricular tachyarrhythmias (VF, VT, and FVT)
by comparing the time intervals between sensed ventricular
events to a set of programmable detection intervals. If enough
intervals occur that are shorter than the programmed intervals, the
device detects a tachyarrhythmia, and responds automatically with
a programmed therapy. After delivering the therapy, the device
either redetects the arrhythmia and delivers the next programmed
therapy or detects episode termination.
To avoid detecting rapidly conducted SVTs (for example, sinus
tachycardia or atrial fibrillation) as ventricular tachyarrhythmias,
the device provides several detection enhancements, including
the Wavelet Dynamic Discrimination criterion, the Onset criterion,
and the Stability criterion.
Figure 6-1 shows how all of these detection features interact
during initial detection. During redetection, the device does not
apply the Wavelet and Onset detection criteria.
Note: Detection functions can be turned off by programming the
VF Enable, FVT Enable, and VT Enable parameters to Off. For an
example, see “How to program VF detection” on page 74.
Marquis VR 7230Cx, B, and E Reference Manual
Detecting tachyarrhythmias
Figure 6-1. How detection features interact during initial detection
65
Detection overview
No
No
Ye s
No
No / suspended by
High Rate Timeout
Update counts and pattern information
Has High Rate Timeout suspended detection
Is the Onset criterion allowing VT event counting? (VT and
Does Stability reset the VT event count?
Ventricular Event
Is the interval in the
VF, FVT, or VT detection zone?
Ye s
enhancements?
No
FVT via VT detection only)
Ye s
(VT and FVT via VT detection only)
No
Has a tachyarrhythmia event count
reached an NID?
Ye s
Is Wavelet on?
Ye s
Ye s
Tachy
Episode
Detected
Ye s
No
Is the median ventricular interval less than the
SVT Limit?
No
Is Wavelet withholding detection?
Ye s
Marquis VR 7230Cx, B, and E Reference Manual
66
Chapter 6
Detection overview
Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops
classifying and counting tachyarrhythmia intervals. Sensing and
bradycardia pacing remain active, and the programmed detection
settings are not modified.
Detection is suspended
■
when the device senses the presence of a strong magnet. The
programmer head contains a magnet which suspends
detection, but once telemetry between the device and
programmer is established, detection resumes.
■
while performing any of the manual system tests, including
Underlying Rhythm, Pacing Threshold, Lead Impedance,
EGM Amplitude, Wavelet, and Charge/Dump. Detection
automatically resumes once the test is complete.
■
while performing a T-Shock, 50 Hz Burst, Manual Burst, or
PES Induction. You can choose to have the device
automatically resume detection after delivering the induction.
■
when you deliver a Manual or Emergency therapy. You can
resume detection by selecting the [Resume] button or
removing the programming head from the device.
■
when you select the on-screen [Suspend] button. You can
resume detection by selecting the [Resume] button or by
removing the programming head from the device.
■
during the automatic daily lead impedance measurements.
Detection resumes when the measurements are complete.
■
while the device is delivering an automatic tachyarrhythmia
therapy (including capacitor charging for defibrillation and
cardioversion). However, the device does continue to confirm
the detected episode during charging. Detection resumes
when the therapy is complete.
Note: The device suspends VT detection (and Combined
Count detection; see page 87) for 17 events following a
defibrillation therapy delivered in response to a detected VF.
■
during charging for Automatic Capacitor Formation. Detection
resumes when charging is complete.
1
1
If the defibrillation therapy is delivered as a result of a High Rate Timeout
Therapy operation, VT detection is not suspended (see page 113).
Marquis VR 7230Cx, B, and E Reference Manual
Setting up sensing
The device provides bipolar sensing in the ventricle via the
sensing electrodes of the implanted ventricular lead. You can
adjust the sensitivity to intracardiac signals using the ventricular
sensitivity setting. This setting defines the minimum electrical
amplitude recognized by the device as a ventricular sensed event.
Proper sensing is essential for the safe and effective use of the
device. To provide appropriate sensing, the device uses an
auto-adjusting ventricular sensing threshold.
See details about sensing on page 69.
Parameters
V. Sensitivity (mV) – Minimum amplitude of
electrical signal that registers as a sensed
ventricular event.
Considerations
Detecting tachyarrhythmias
Setting up sensing
* Medtronic nominal setting
0.15, 0.3*, 0.45, 0.6,
0.9, 1.2
67
Review the following information before programming sensing
parameters.
Sensitivity thresholds – The programmed ventricular sensitivity
threshold applies to all features related to sensing, including
detection and bradycardia pacing.
Bradycardia pacing and sensing – A combination of high
pacing pulse width or high amplitude with a low sensitivity
threshold may cause inappropriate sensing. Programming a lower
pulse width, lower amplitude, longer pace blanking, or a higher
sensitivity threshold may eliminate this inappropriate sensing.
Recommended ventricular sensitivity threshold – A ventricular
sensitivity threshold of 0.3 mV is recommended to maximize the
probability of detecting VF and to limit the possibility of
oversensing.
Marquis VR 7230Cx, B, and E Reference Manual
68
Chapter 6
Setting up sensing
High ventricular sensitivity threshold – Setting V. Sensitivity to
a value greater than 0.6 mV is not recommended except for
testing. Doing this may cause undersensing, which can cause any
of the following situations:
■
delayed or aborted cardioversion therapy
■
delayed defibrillation therapy (when VF confirmation is active)
■
asynchronous pacing
■
underdetection of tachyarrhythmias.
Low ventricular sensitivity threshold – If you set V. Sensitivity
to its most sensitive value of 0.15 mV, the device will be more
susceptible to EMI and oversensing.
Testing sensitivity after reprogramming – If you change the
ventricular sensitivity threshold, evaluate for proper sensing and
detection by inducing VF and allowing the device to automatically
detect and treat the arrhythmia.
Sensing during VF – Always verify that the device senses
properly during VF. If the device is not sensing or detecting
properly, program detection and therapies off, and evaluate the
system (making sure to monitor the patient for life-threatening
arrhythmias until you enable detection and therapies again). You
may need to reposition or replace the ventricular sensing lead to
achieve proper sensing.
Marquis VR 7230Cx, B, and E Reference Manual
How to program sensitivity
Details about sensing
Auto-adjusting sensitivity thresholds
Detecting tachyarrhythmias
69
Setting up sensing
1. Select Params > Detection
2. Select the desired
V. Sensitivity parameter
value.
3. Select [PROGRAM].
1
2
3
The device automatically adjusts the sensitivity thresholds after
certain paced and sensed events to help reduce oversensing from
T-waves and pacing. Figure 6-2 shows how sensitivity thresholds
are adjusted after different types of events.
Marquis VR 7230Cx, B, and E Reference Manual
70
Chapter 6
Setting up sensing
Figure 6-2. Auto-adjusting sensitivity threshold
Sensitivity threshold
12
Rectified and Filtered
V. E G M
Marker Channel
30Autoadjust.eps
V
S
V
P
1 After a ventricular sensed event, the ventricular sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed
value, decay constant: 450 ms).
a
2 After the ventricular pace blanking period is finished, the ventricular
threshold increases to 4.5x the programmed value (maximum:
1.8 mV, decay constant: 450 ms).
a
The exponential decay continues through a subsequent ventricular pacing pulse
and its blanking period.
Marquis VR 7230Cx, B, and E Reference Manual
Detecting tachyarrhythmias
Detecting VF episodes
Blanking periods
During a blanking period, the device does not sense electrical
signals. This helps prevent sensing of device pacing,
cardioversion and defibrillation pulses, post-pacing depolarization,
T-waves, and multiple sensing of the same event. The blanking
periods following paced events are longer than those following
sensed events to avoid sensing ventricular depolarizations.
Table 6-1 shows the duration of the fixed blanking periods. For
information on programmable pace blanking periods (see
page 158).
Table 6-1. Fixed blanking periods
Ventricular blanking after sensed ventricular event120 ms
71
Ventricular blanking after delivered
cardioversion or defibrillation therapy
Refractory periods
During a refractory period, the device senses normally, but
classifies sensed events as refractory and limits its response to
these events. Refractory periods are used during synchronization
to help prevent the device from delivering cardioversion and
defibrillation therapies at inappropriate times. See “Synchronizing
defibrillation without confirming VF” on page 125 and
“Synchronizing cardioversion after charging” on page 146.
Note: Refractory periods do not affect tachyarrhythmia detection.
Detecting VF episodes
The device detects VF episodes by examining the cardiac rhythm
for short ventricular intervals. If a predetermined number of
intervals occurs that are short enough to be considered VF events,
the device detects VF and delivers the first programmed VF
therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing.
See details about VF detection on page 74.
520 ms
Marquis VR 7230Cx, B, and E Reference Manual
72
Chapter 6
Detecting VF episodes
Parameters
VF Detection Enable – Turns VF
detection on or off.
* Medtronic nominal setting
On*, Off
Considerations
VF Interval (ms) – V-V intervals shorter
than this value are counted as VF events.
VF Initial NID – Number of Intervals to
Detect: number of VF events the device
must count to detect a VF episode.
VF Redetect NID – Number of Intervals
to Redetect: number of VF events the
device must count to redetect a
continuing VF after a therapy.
Review the following information before programming
VF detection parameters.
VF Interval minimum setting – To ensure proper VF detection,
you should not program the VF Interval less than 300 ms.
VF Interval maximum setting – Programming the VF Interval to
a value greater than 350 ms may cause inappropriate detection of
rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals
shorter than the VF Interval are counted using the VF event
counter, which is more sensitive than the consecutive VT event
counter.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the
Initial NIDs.
Enabling VF detection – When VF Detection Enable is
programmed On for the first time, the device
■
enables Automatic Capacitor Formation
Marquis VR 7230Cx, B, and E Reference Manual
Restrictions
Detecting tachyarrhythmias
Detecting VF episodes
■
starts recording Cardiac Compass data
■
starts recording lead performance trends (starting at
3:00 AM, by the device clock)
■
clears all brady pacing counters
VF Detection and Wavelet – You can program the device to
exclude rapidly conducted SVTs from VF detection by enabling the
Wavelet Dynamic Discrimination criterion. Note that the SVT Limit
must be programmed shorter than the VF Interval in order for
Wavelet to affect VF detection. See “Enhancing detection with
Wavelet” on page 94.
Review the following information before programming
VF detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e
reliable ventricular tachyarrhythmia detection, the programmer
regulates the values available for bradycardia pacing and
tachyarrhythmia detection. See “Parameter interlocks” on
page 360.
73
VF detection backup – To ensure VF detection backup during
VT and FVT episodes, if VF Detection is off, both VT Detection
and FVT Detection must also be off.
Marquis VR 7230Cx, B, and E Reference Manual
74
Chapter 6
Detecting VF episodes
How to program VF detection
2
Details about VF detection
To program VF detection:
1. Select Params > Detection
2. Select the desired values for
VF Enable, VF Initial NID,
VF Redetect NID and
VF Interval.
3. Select [PROGRAM].
1
3
The device detects VF by counting the number of VF events, which
are V-V intervals shorter than the programmed VF Interval. On
each event, the device counts the number of recent VF events.
The number of recent events examined is called the VF detection
window. The size of the VF detection window is the second
number in the programmed VF NID (for example, 24 events if the
VF Initial NID is 18/24).
The threshold for detecting VF is the first number in the
programmed VF NID (for example, 18 events if the VF Initial NID
is 18/24). This threshold is always 75% of the VF detection
window. That is, if 75% of the events in the VF detection window
are VF events, the device detects a VF episode (see Figure 6-3).
After the device detects VF, it delivers the first programmed VF
therapy. Following the therapy, if the number of VF events reaches
the programmed VF Redetect NID, the device redetects VF and
delivers the next programmed VF therapy.
Note: The device can also detect VF Episodes via the Combined
Count detection criterion (see page 87).
Marquis VR 7230Cx, B, and E Reference Manual
Figure 6-3. Device detects VF
ECG
Marker Channel
Detecting tachyarrhythmias
Detecting VF episodes
123
75
FSFSFSFSFSFSFSFSFSF
F
F
S
VSVSV
D
S
VF Event Count
V
S
V
S
FSFSF
V
S
1 2 3 4 5 66 7 8 9 10 11 12 13 14 15 16 17 18
FSFSF
V
S
S
S
S
VF Interval
200 ms
1 VF starts, and the device begins counting VF events (intervals less than the programmed
VF Interval).
2 A ventricular interval occurs outside the VF detection zone. The VF event count is not
incremented.
3 The VF event count reaches the programmed VF NID value of 18 events out of 24, and the
device detects VF.
30VFDetection.eps
Marquis VR 7230Cx, B, and E Reference Manual
76
Chapter 6
Detecting VT episodes
Detecting VT episodes
Parameters
The device detects VT episodes by examining the cardiac rhythm
for short ventricular intervals. If enough intervals occur that are
short enough to be considered VT events (but are not VF or FVT
events), the device detects VT and delivers the first programmed
VT therapy. After therapy, the device continues to evaluate the
ventricular rhythm to determine if the episode is ongoing.
You can program the device to detect and record VT episodes
without treating them with VT therapies by setting VT Detection
Enable to Monitor. If a patient’s VT episodes are well-tolerated,
this feature allows you to collect data about these episodes without
delivering therapy or affecting VF detection.
See details about VT detection on page 78.
* Medtronic nominal setting
Considerations
VT Detection Enable – Turns VT
detection on or off, or enables VT
monitoring.
VT Interval (Rate) (ms) – V-V intervals
shorter than this value are counted as
VT events.
VT Initial NID – Number of Intervals to
Detect: number of VT events the device
must count to detect a VT episode.
VT Redetect NID – Number of Intervals
to Redetect: number of VT events the
device must count to redetect a
continuing VT after a therapy.
On, Off*, or Monitor
280, 290, . . ., 400*, . . ., 600
12, 16*, . . ., 52,
76, 100
4, 8, 12*, . . ., 52
Review the following information before programming
VT detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the
patient’s tachycardia interval, you should program the VF, FVT,
and VT intervals at least 40 ms apart.
Marquis VR 7230Cx, B, and E Reference Manual
Restrictions
Detecting tachyarrhythmias
Detecting VT episodes
Episode redetection – You can expedite redetection by
programming the VF and VT Redetect NIDs lower than the
Initial NIDs.
VT detection and Combined Count detection – When
VT Detection is On, the device applies the Combined Count
detection criterion to help speed detection of rhythms that
fluctuate between detection zones. Combined Count detection is
disabled if VT Detection is set to Off or Monitor. See “Detecting
tachyarrhythmia episodes with Combined Count” on page 87.
VT detection and rapidly conducted SVTs – You can program
the device to exclude rapidly conducted SVTs from VT detection
by enabling the Wavelet, Onset, or Stability detection criteria. See
“Enhancing detection with Wavelet” on page 94, “Enhancing VT
detection with the Onset criterion” on page 103, and “Enhancing
VT detection with the Stability criterion” on page 108.
Review the following information before programming
VT detection parameters.
77
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e
reliable ventricular tachyarrhythmia detection, the programmer
regulates the values available for bradycardia pacing and
tachyarrhythmia detection. See “Parameter interlocks” on
page 360.
VF detection backup – To ensure VF detection backup during
VT and FVT episodes, if VF Detection is off, both VT Detection
and FVT Detection must also be off.
Marquis VR 7230Cx, B, and E Reference Manual
78
Chapter 6
Detecting VT episodes
How to program VT detection
2
Details about VT detection
To program VT detection:
1. Select Params > Detection
2. Select the desired values for
VT Enable, VT Initial NID,
VT Redetect NID and
VT Interval.
3. Select [PROGRAM].
1
3
The device detects VT by counting the number of consecutive VT
events. A VT event is a V-V interval shorter than the programmed
VT Interval but greater than or equal to the VF Interval. If the
number of consecutive VT events reaches the programmed VT
Initial NID, the device detects VT (see Figure 6-4).
The VT event count resets to zero whenever an interval occurs
that is greater than or equal to the programmed VT Interval. The
count remains at the current value if an interval is shorter than the
programmed VF Interval.
After the device detects VT, it delivers the first programmed VT
therapy. Following the therapy, if the VT event counter reaches the
VT Redetect NID, the device redetects VT and delivers the next
programmed therapy.
Note: The device can also detect VT Episodes via the Combined
Count detection criterion (see page 87).
Marquis VR 7230Cx, B, and E Reference Manual
Figure 6-4. Device detects VT
ECG
Marker Channel
132
Detecting tachyarrhythmias
Detecting VT episodes
79
VT Event Count
V
S
VSV
S
V
T
S
10123456
TSTSTSTSTST
S
S
VT Interval
ECG
Marker Channel
T
D
V
S
30VTDetection.eps
VT Event Count
TSTSTSTSTSTSTSTST
78910111213141516
S
VT Interval
200 ms
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval, but greater than or equal to than the VF Interval).
2 A ventricular interval occurs outside VT detection zone. The VT event count resets to zero.
3 The VT event count reaches the programmed VT NID of 16 events, and the device detects
VT.
Marquis VR 7230Cx, B, and E Reference Manual
80
Chapter 6
Detecting VT episodes
VT monitoring
You can program the device to record VT episodes without
delivering VT therapy by setting VT Detection to Monitor. When VT
monitoring is enabled, the device detects VT episodes but does
not deliver VT therapy (see Figure 6-5). Instead, it records VT
episodes, labeling them as “monitored,” and waits for episode
termination to occur.
When VT Detection is set to Monitor, several detection operations
work differently.
VT event counting – Before the device detects an episode, it
counts VT events normally. However, once the VT Initial NID is
reached, the device sets the VT event count to zero and suspends
VT event counting for the rest of the episode.
VF and FVT detection – VF and FVT detection operate as if VT
detection is off. Specifically, Combined Count detection is
disabled, and FVT via VT detection is not selectable. If a
monitored VT episode accelerates into the FVT or VF detection
zone, the device applies the VF Initial NID to detect the new
tachyarrhythmia. Once an episode is in progress, VT event
counting doesn’t resume until the episode ends.
Caution: Programming the VF Interval greater than 350 ms
may result in inappropriate detection of rapidly conducted
atrial fibrillation as VF or FVT via VF. Intervals shorter than the
VF Interval are counted using the VF event counter, which is
more sensitive than the consecutive VT event counter.
Wavelet, Onset, and Stability criteria – Before the device
detects a tachyarrhythmia episode, the Wavelet, Onset, and
Stability criteria, if turned on, are applied. If a monitored VT
episode accelerates into the FVT or VF detection zone, the device
continues to apply Wavelet as initial VF or FVT detection begins.
However, since Stability and Onset do not affect either VF
detection or FVT via VF detection, they are not applied.
Episode termination – The device compares ventricular
intervals to the VT Interval to identify when a VT monitored
episode has ended. However, if a VF episode or FVT via VF
episode occurs when VT monitoring is enabled, the device
compares ventricular intervals to the VF Interval to identify
episode termination.
Marquis VR 7230Cx, B, and E Reference Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.