Medtronic 7230B Reference Guide

Reference Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician or properly licensed practitioner.
MARQUIS®VR 7230Cx, 7230B, 7230E
Single Chamber Implantable Cardioverter Defibrillator (VVE–VVIR)
MARQUIS VR 7230Cx, 7230B, 7230E
Reference Manual 0
A guide to the operation and programming of the Model 7230Cx, B, and E Marquis VR Single Chamber Implantable Cardioverter Defibrillator
0
The following are trademarks of Medtronic:
Active Can, Cardiac Compass, Checklist, Decision Channel, Flashback, GEM, Leadless ECG, Marker Channel, MARQUIS, Patient Alert, Quick Look, QuickLink, RapidRead, T-Shock, Wavelet
Table of contents
Introduction xv
Using this manual xv Contacting technical support xv Customer education xv References xvi Notice xvi
Abbreviations and acronyms xvii
Part I Quick overview
1 Quick reference 3
Physical characteristics 4
Magnet application 8
Longevity projections 9
Replacement indicators 12
Typical charge times 13
High voltage therapy energy 13
Stored data and diagnostics 14
New and enhanced features 17
Lead connection header options 17 Patient management 17 Tachyarrhythmia detection 19 Tachyarrhythmia therapy 20 Bradycardia pacing 20 EP studies 21
2 The Marquis VR system 23
System overview 24
Detecting and treating tachyarrhythmias 25 Treating bradycardia 26 Monitoring for real-time and stored data 26 Conducting electrophysiologic tests 26 Alerting the patient to system events 26
Indications and usage 27
Contraindications 27
Marquis VR 7230Cx, B, and E Reference Manual
iv
Table of contents
Patient screening 27
3 Emergency therapy 29
Delivering emergency therapies 30
Effect on system operation 30 Aborting an emergency therapy 30 On-screen and display panel buttons 31 Temporary parameter values 31 How to deliver emergency 30 joule defibrillation 32 How to deliver emergency cardioversion 32 How to deliver emergency fixed burst pacing 33 How to deliver emergency VVI pacing 33
Part II Device implant and patient follow-up procedures
4 Implanting the ICD 37
Overview 38
Preparing for an implant 38
Equipment for an implant 38 Sterile supplies 39 How to prepare for implanting 39
Replacing an ICD 40
How to explant and replace an ICD 40
Positioning the leads 41
Using transvenous leads 41 Using epicardial leads 41 Surgical incisions 42
Testing sensing and pacing thresholds 43
Parameters 43 Considerations 43
Connecting the leads to the ICD 44
How to connect the lead to the device 47
Testing defibrillation operation and effectiveness 47
High voltage implant values 48 Binary search protocol 48 How to prepare for defibrillation threshold testing 49 How to perform defibrillation threshold testing 50
Positioning and securing the ICD 52
How to position and secure the device 52
Marquis VR 7230Cx, B, and E Reference Manual
Completing the implant procedure 52
How to complete programming the device 53
5 Conducting a patient follow-up session 55
Patient follow-up guidelines 56
Verifying the status of the implanted system 56
Verifying accurate detection and appropriate therapy 57
Considerations 57
Verifying effective bradycardia pacing 58
Considerations 59
Part III Configuring the ICD for the patient
6 Detecting tachyarrhythmias 63
Detection overview 64
Suspending tachyarrhythmia detection 66
Setting up sensing 67
Parameters 67 Considerations 67 How to program sensitivity 69 Details about sensing 69
Detecting VF episodes 71
Parameters 72 Considerations 72 Restrictions 73 How to program VF detection 74 Details about VF detection 74
Detecting VT episodes 76
Parameters 76 Considerations 76 Restrictions 77 How to program VT detection 78 Details about VT detection 78
Detecting FVT episodes 82
Parameters 82 Considerations 82 Restrictions 83 How to program FVT detection 84 Details about FVT detection 84
Table of contents
v
Marquis VR 7230Cx, B, and E Reference Manual
vi
Table of contents
Detecting tachyarrhythmia episodes with Combined Count 87
Details about Combined Count detection 87
Monitoring episodes for termination or redetection 90
Parameters 90 Considerations 90 How to program redetection parameters 91 Details about episode termination and redetection 91
Enhancing detection with Wavelet 94
Parameters 94 Considerations 95 How to program Wavelet 98 Details about Wavelet 98 Details about Auto Collection 101
Enhancing VT detection with the Onset criterion 103
Parameters 103 Considerations 103 How to program Onset 104 Details about Onset 104
Enhancing VT detection with the Stability criterion 108
Parameters 108 Considerations 108 How to program Stability 109 Details about Stability 109
Detecting prolonged tachyarrhythmias with High Rate Timeout 111
Parameters 111 Considerations 111 How to program High Rate Timeout 112 Details about High Rate Timeout 112
Key ter ms 113
7 Treating tachyarrhythmia episodes 119
Treating VF with defibrillation 120
Parameters 120 Considerations 121 Restrictions 122 How to program VF therapies 122 Details about VF therapy 122
Treating VT and FVT with antitachycardia pacing 130
Parameters for all ATP therapies 130
Marquis VR 7230Cx, B, and E Reference Manual
Table of contents
Parameters for Burst therapy 131 Parameters for Ramp therapy 131 Parameters for Ramp+ therapy 132 Considerations 132 Restrictions 133 How to program ATP therapies 133 Details about ATP therapies 134
Treating VT and FVT with cardioversion 140
Parameters 140 Considerations 141 Restrictions 142 How to program cardioversion therapies 142 Details about cardioversion therapy 143
Optimizing therapy with Smart Mode and Progressive Episode Therapies 149
Parameters 149 Considerations 149 Restrictions 150 How to program Smart Mode 150 Details about Smart Mode 151 How to program Progressive Episode Therapies 152 Details about Progressive Episode Therapies 152
Key ter ms 153
vii
8 Treating bradycardia 157
Providing basic pacing therapy 158
Parameters 158 Considerations 159 Restrictions 160 How to program bradycardia pacing parameters 160 Details about basic bradycardia pacing 161
Enhancing pacing for optimal cardiac output 162
Parameters 162 Considerations 163 Restrictions 163 How to program Rate Response 164 Details about Rate Responsive Pacing 165 How to program Hysteresis 169 Details about Hysteresis 169
Marquis VR 7230Cx, B, and E Reference Manual
viii
Table of contents
How to program Ventricular Rate Stabilization 171 Details about Ventricular Rate Stabilization 171
Providing pacing after high voltage therapies 173
Parameters 173 Considerations 173 How to program pacing after high voltage therapies 174 Details about Post Shock Pacing parameters 174
Key ter ms 174
9 Optimizing charge time and device longevity 177
Optimizing charge time 178
Parameters 178 Considerations 178 How to evaluate charging performance 179 Details about managing charge time 179
Optimizing device longevity 181
Considerations 181
Key ter ms 182
Part IV Evaluating and managing patient treatment
10 Using the programmer 185
Connecting the programmer hardware 186
Using an external printer 187
To view a list of supported printers 188 Materials you will need 188 To connect the printer to your programmer 190
Using the programming head 192
During an episode in progress 192 During marker transmissions 192 How capacitor charging affects the light array 193 Alternative program and interrogate buttons 193
Display screen features 193
Programmer status bar display 194 Live Rhythm Monitor window 195 Task area 196 Command bar 198 Tool palette 198 Buttons 200
Marquis VR 7230Cx, B, and E Reference Manual
Setting programmer preferences 200
How to set the programmer time and date 201 How to set audio preferences 201 How to select a different language 202 How to check the software version number 203 How to set printing preferences 203 How to set reports preferences 204
Starting and ending patient sessions 204
How to start a patient session 205 How to end a patient session 206 Automatic interrogation 206
Viewing live waveforms 207
Parameters 207 How to use the Adjust window 208 How to use the waveform adjustment button bar 209 Details about the live rhythm monitor 209
Recording live waveforms 213
Printing a report while recording live ECG 214
Saving and retrieving device data 215
Considerations 215 How to save ICD data to a disk 216 How to read ICD data from a disk 216 Saving data to a disk 218 Data file names 218 Reading device data from diskette 218
Printing reports 219
How to print a report 220 Print Queue 220
Key ter ms 221
Table of contents
ix
11 Using system evaluation tools 223
A summary of system evaluation tools 224
Automatic daily measurements 224
Taking a quick look at device activity 225
How to use Quick Look 225 Quick Look observations 226
Using the Patient Alert feature 227
Parameters 227 Considerations 229
Marquis VR 7230Cx, B, and E Reference Manual
x
Table of contents
How to program the Patient Alert feature 232 Instructing the patient 232 Viewing the Patient Alert Events 233 How to view the Patient Alert events 234
Streamlining follow-ups with Checklist 235
How to select and use a checklist 236 How to create, edit, and delete a checklist 237
Key ter ms 238
12 Setting up and viewing collected data 239
A summary of data collection 240
Setting up data collection 241
Parameters 241 Considerations 242 How to set up data collection 243 Details about data collection parameters 244 How to clear collected data 247
Collecting lead performance data 247
Daily lead impedance and EGM amplitude measurements 247 Sensing integrity counter 249
Viewing the episode and therapy efficacy counters 249
How to view and print counter data 250 Details about the episode and therapy efficacy counters 251 Therapy counters 253
Viewing episode data 254
How to view episode data 255 Details about episode data 256 How to view an Interval Plot 258 How to view an EGM Strip 259 How to display the episode text information 261 How to display QRS Snapshot data 262
Viewing Flashback Memory 263
How to view the Flashback Memory 264
Viewing battery and lead status data 265
How to view battery and lead status data 267
Viewing lead performance trends 267
How to view lead performance trend graphs 268
Using Cardiac Compass to view long term clinical trends 269
Marquis VR 7230Cx, B, and E Reference Manual
How to print a Cardiac Compass report 271 Details on Cardiac Compass trend data 271
Viewing and entering patient information 275
How to view and enter new patient information 277 Displaying and printing patient information 277
Automatic device status monitoring 278
Device status indicator warnings 278 How to respond to an electrical reset 279
Key ter ms 280
13 Testing the system 285
Testing overview 286
Evaluating the underlying rhythm 287
Considerations 287 How to Perform an Underlying Rhythm test 287
Measuring pacing thresholds 288
Parameters 289 Considerations 289 How to Perform a Pacing Threshold Test 290
Testing the Wavelet criterion 291
Parameters 291 Considerations 291 Restrictions 293 How to evaluate the current template with the Wavelet test 293 How to collect a template with the Wavelet test 294
Measuring lead impedance 295
Considerations 295 How to perform a Lead impedance test 295 Details about the Lead Impedance test 296
Measuring EGM Amplitude 297
Parameters 297 Considerations 297 Restrictions 298 How to perform an EGM Amplitude test 298
Testing the device capacitors 299
Considerations 299 How to perform a Charge/Dump test 300
Key ter ms 301
Table of contents
xi
Marquis VR 7230Cx, B, and E Reference Manual
xii
Table of contents
14 Conducting Electrophysiologic Studies 303
EP Study overview 304
Inducing VF with T-Shock 306
Parameters 306 Considerations 307 Restrictions 307 How to deliver a T-Shock induction 308 Details about T-Shock induction 308
Inducing VF with 50 Hz Burst 309
Parameters 309 Considerations 310 How to deliver a 50 Hz Burst induction 311 Details about 50 Hz Burst induction 312
Inducing an arrhythmia with Manual Burst 312
Parameters 312 Considerations 312 How to deliver a Manual Burst induction 313 Details about Manual Burst induction 314
Inducing an arrhythmia with PES 314
Parameters 314 Considerations 315 How to deliver a PES induction 315 Details about PES induction 316
Delivering a manual therapy 316
Parameters for manual defibrillation and cardioversion 316 Parameters for manual ATP therapies 316 Considerations 318 How to deliver a manual therapy 318 Details about manual therapies 318
Key ter ms 319
15 Solving system problems 321
Overview 322
Solving sensing problems 323
Solving tachyarrhythmia detection problems 324
Solving tachyarrhythmia therapy problems 325
Solving bradycardia pacing problems 326
Responding to device status indicators 327
Marquis VR 7230Cx, B, and E Reference Manual
Key ter ms 328
Part V Appendices
A Warnings and precautions 333
General warnings 334
Storage and handling 334
Resterilization 335
Device implantation and ICD programming 335
Lead evaluation and lead connection 337
Follow-up testing 338
Explant and disposal 339
Medical therapy hazards 339
Home and occupational environments 341
BDevice Parameters343
Emergency settings 344
Detection parameters 345
Therapy parameters 347
Bradycardia pacing parameters 349
System maintenance parameters 350
Data collection parameters 352
System test and EP study parameters 353
Fixed parameters 356
Patient information parameters 357
Programmer symbols 358
Parameter Interactions Window 359
Parameter interlocks 360
Table of contents
xiii
Index 361
Marquis VR 7230Cx, B, and E Reference Manual
xiv
Table of contents
Marquis VR 7230Cx, B, and E Reference Manual

Introduction

Using this manual

Before implanting the ICD, it is strongly recommended that you:
Refer to the product literature packaged with the ICD for information about prescribing the ICD.
Thoroughly read this manual and the technical manuals for the leads used with the device.
Discuss the procedure and the ICD system with the patient and any other interested parties, and provide them with any patient information packaged with the ICD.

Contacting technical support

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
Introduction
xv
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise.
For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.

Customer education

Medtronic invites physicians to attend an education seminar on the complete ICD system. The course includes indications for use, an overview of ICD system functions, implant procedures, and patient management.
Marquis VR 7230Cx, B, and E Reference Manual
xvi
Introduction

References

Notice

The primary reference for background information is Zacouto FI, Guize LJ. Fundamentals of Orthorhythmic Pacing. In: Luderitz B, ed. Cardiac Pacing Diagnostic and Therapeutic Tools. New York: Springer-Verlag; 1976: 212-218.
See these additional references for more background information:
Singer I, Ed. Implantable Cardioverter-Defibrillator. Armonk, NY: Futura Publishing Co. 1994.
Singer I, Barold SS, Camm AJ, Eds. Nonpharmacological Therapy of Arrhythmias for the 21st Century: The State of the Art. Armonk, NY: Futura Publishing Co. 1998.
Estes M, Manolis AS, Wang P, Eds. Implantable Cardioverter-Defibrillators. New York, NY: Marcel Dekker, Inc.
1994.
Kroll MW, Lehmann MH, Eds. Implantable Cardioverter-Defibrillator Therapy: The Engineering-Clinical Interface. Norwell, MA: Kluwer Academic Publishers 1996.
This software is provided as an informational tool for the end user. The user is responsible for accurate input of patient information into the software. Medtronic makes no representation as to the accuracy or completeness of the data input into the software. MEDTRONIC SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTIAL OR CONSEQUENTIAL DAMAGES TO ANY THIRD PARTY WHICH RESULTS FROM THE USE OF THE INFORMATION PROVIDED IN THE SOFTWARE.
Marquis VR 7230Cx, B, and E Reference Manual

Abbreviations and acronyms

ATP Antitachycardia Pacing
BOL Beginning of Life
bpm beats per minute
CNID Combined (VT and VF) Number of Intervals to Detect
CV Cardioversion
DF/Defib Defibrillation
ECG Electrocardiogram
EGM Electrogram
EOL End of Life
ERI Elective Replacement Indicator
FDI Fibrillation Detection Interval
FTI Fast Ventricular Tachycardia Detection Interval
Abbreviations and acronyms
xvii
FVT Fast Ventricular Tachycardia
ICD Implantable Cardioverter Defibrillator
J joules
-1
reciprocal minutes; for example, pacing pulses per minute
min
ms milliseconds
mV millivolts
NID Number of Intervals to Detect
NST Non-sustained Tachycardia
PES Programmed Electrical Stimulation
ppm paces or pulses per minute
PVC Premature Ventricular Contraction
RNID Number of Intervals to Redetect
R-R a ventricular interval
SVT Supraventricular Tachycardia
Marquis VR 7230Cx, B, and E Reference Manual
xviii
Abbreviations and acronyms
TDI Tachycardia Detection Interval
V volts
V- Ventricular
VF Ventricular Fibrillation
VF NID VF Number of Intervals to Detect
VRS Ventricular Rate Stabilization
VT Ventricular Tachycardia
VT NID VT Number of Intervals to Detect
Marquis VR 7230Cx, B, and E Reference Manual

Quick overview

Part I
Marquis VR 7230Cx, B, and E Reference Manual
Physical characteristics 4
Magnet application 8
Longevity projections 9
Replacement indicators 12
Typical charge times 13
High voltage therapy energy 13
Stored data and diagnostics 14
New and enhanced features 17

Quick reference1

1
Marquis VR 7230Cx, B, and E Reference Manual
4
Chapter 1

Physical characteristics

Physical characteristics
Table 1-1. ICD physical characteristics
Connector type Cx
a
B or E
Volume 36 cc 43 cc
Mass 75 g 80 g
H x W x D
b
Surface area of device can 66 cm
Radiopaque ID
c
68 mm x 51 mm x 14 mm
2
74 mm x 51 mm x 14 mm
2
63 cm
PKD B-type connector PLW
E-type connector PLY
Materials in contact with human tissue
d
Titanium / polyurethane / silicone rubber
Battery Lithium silver vanadium oxide
a
Measurements are nominal values based on CAD (computer aided design) model measurements and are rounded to the nearest unit.
b
Grommets may protrude slightly beyond the can surface.
c
Engineering series number follows the radiopaque code.
d
These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue.
Marquis VR 7230Cx, B, and E Reference Manual
Table 1-2. Cx-type connector characteristics
General description
One IS-1 connector for pacing and sensing, Two DF-1 connectors for high voltage therapy, Active Can electrode (programmable).
Figure 1-1. Cx-type connector
2
Device Port
SVC DF-1 HVX
RV DF-1 HVB
Can n/a HVA, Can
V IS-1 bipolar
1
Quick reference
Physical characteristics
Connector Typ e
Software Name
5
5
1 suture holes
2 SVC port (DF-1)
3RV port (DF-1)
4 programmable Active Can
5V. port (IS-1)
3
4
Marquis VR 7230Cx, B, and E Reference Manual
6
Chapter 1
Physical characteristics
Table 1-3. B-type connector characteristics
General description
Port Connector Type
HVB 6.5 mm HVB
Device
Three 6.5 mm unipolar high voltage ports and one
3.2 mm low profile bipolar pace/sense (IS-1 compatible) port.
a
The HVA and HVX ports are electronically connected and are treated as the HVX electrode. For more information see “B- and E-type connector pathways” on page 123.
HVA
HVX
a
a
6.5 mm HVX
6.5 mm HVX
Can n/a HVA, Can
P/S bipolar pace sense
3.2 mm bipolar (IS-1 compatible)
Figure 1-2. B-type connector
Software Name
7
2
6
5
3
4
1 suture holes
2 HVB port (6.5 mm)
3 HVA port (6.5 mm)
4 programmable Active Can
5 HVX port (6.5 mm)
6/7 P+/S and P-/S port (3.2 mm)
1
Marquis VR 7230Cx, B, and E Reference Manual
Quick reference
Physical characteristics
Table 1-4. E-type connector characteristics
General description
Port
P+/S port 5.0 mm
Device
Two 6.5 mm unipolar high voltage ports and two 5 mm unipolar pace/sense ports.
HVA port 6.5 mm HVX
HVB port 6.5 mm HVB
Can n/a HVA, Can
P-/S port 5.0 mm
a
The HVA port of the E-type connector can be used as the HVX electrode when Active Can is programmed off, see “B- and E-type connector pathways” on page 123.
Figure 1-3. E-type connector
Connector Typ e
Software Name
a
7
6
5
2
3
4
1 suture holes
2 P+/S port (5.0 mm)
3 HVA port (6.5 mm)
4 programmable Active Can
5 HVB port (6.5 mm)
6 P-/S port (5.0 mm)
1
Marquis VR 7230Cx, B, and E Reference Manual
8
Chapter 1

Magnet application

Magnet application
Bringing a magnet close to the device triggers changes in device operation as shown in Table 1-5. When the magnet is removed, the device returns to its programmed operations.
Table 1-5. Effects of magnet application on the device
Pacing mode as programmed
Pacing rate and interval as programmed
VF, VT, and FVT detection suspended
Patient Alert audible tones (20 seconds or less)
with programmable alert(s) enabled:
continuous tone (Test)
on/off intermittent tone (seek follow-up)
high/low dual tone (urgent follow-up)
with programmable alerts disabled:
no tone
high/low dual tone (urgent follow-up)
a
Rate response adjustments are suspended while a Patient Alert tone sounds.
b
Detection resumes if telemetry is established and the application software is running, or it resumes after the application software has started.
c
The Test tone does not sound if “VF Detection/Therapy Off” is the only alert enabled.
a
b
c
Marquis VR 7230Cx, B, and E Reference Manual

Longevity projections

Longevity estimates are based on accelerated battery discharge data and device modeling with EGM pre-storage off, 60 ppm
-1
(min
) pacing rate, with
2.5 V pacing pulse amplitude, 0.4 ms pacing pulse width, and 30 J delivered therapy energy (see Table 1-6)
3 V pacing pulse amplitude, 0.4 ms pacing pulse width, and 30 J delivered therapy energy (see Table 1-7)
This model assumes default automatic capacitor formation setting. As a guideline, each full energy charge decreases device longevity by approximately 33 days.
Device longevity is affected by how certain features are programmed, such as EGM pre-storage. For more information, see “Optimizing device longevity” on page 181.
Considerations for using EGM pre-storage
When the EGM pre-storage feature is programmed off, the device starts to store EGM following the third tachyarrhythmia event and also provides up to 20 seconds of information before the onset of the tachyarrhythmia, including:
VV intervals
Marker Channel
interval plot Flashback
Quick reference
Longevity projections
9
When the EGM pre-storage feature is programmed on, the device also collects up to 20 seconds of EGM information before the onset of the arrhythmia.
In a patient who uniformly repeats the same onset mechanisms, the greatest clinical benefit of pre-onset EGM storage is achieved after a few episodes are captured.To maximize the effectiveness of the EGM pre-storage feature and optimize device longevity, consider these programming options:
Turn pre-storage on to capture possible changes in the onset mechanism following significant clinical adjustments, for example, device implant, medication changes, and surgical procedures.
Turn pre-storage off once you have successfully captured the information of interest.
Marquis VR 7230Cx, B, and E Reference Manual
10
Chapter 1
Longevity projections
a
b
Table 1-6. Projected longevity in years with 2.5 V pacing amplitude and
0.4 ms pulse width
Percent pacing
Maximum energy charging frequency
a
EGM pre-storage
500 ohm
b
impedance
pacing
900 ohm
pacing
impedance
0% Semi-Annual Off 10.4 10.4
On 10.0 10.0
Quarterly Off 8.7 8.7
On 8.5 8.5
15% Semi-Annual Off 10.0 10.2
On 9.9 10.0
Quarterly Off 8.5 8.6
On 8.3 8.4
50% Semi-Annual Off 9.5 10.0
On 9.3 9.7
Quarterly Off 8.1 8.4
On 8.0 8.2
100% Semi-Annual Off 8.9 9.5
On 8.6 9.3
Quarterly Off 7.6 8.1
On 7.5 8.0
Maximum energy charging frequency may include full energy therapy shocks or capacitor formations.
The data provided for programming EGM pre-storage on is based on a 6 month period (two 3-month follow-up intervals) over the life of the device. Additional use of EGM pre-storage reduces longevity by approximately 25% or 3 months per year.
Marquis VR 7230Cx, B, and E Reference Manual
Quick reference
Longevity projections
Table 1-7. Projected longevity in years with 3V pacing amplitude and 0.4 ms pulse width
Maximum Percent pacing
energy charging
frequency
a
EGM pre-storage
0% Semi-Annual Off 10.4 10.4
On 10.0 10.0
Quarterly Off 8.7 8.7
On 8.5 8.5
15% Semi-Annual Off 10.0 10.1
On 9.8 10.0
Quarterly Off 8.5 8.5
On 8.3 8.3
50% Semi-Annual Off 9.3 9.7
On 9.0 9.5
Quarterly Off 7.9 8.3
On 7.8 8.0
100% Semi-Annual Off 8.4 9.2
On 8.2 9.0
Quarterly Off 7.3 7.9
On 7.1 7.7
a
Maximum energy charging frequency may include full energy therapy shocks or capacitor formations.
b
The data provided for programming EGM pre-storage on is based on a 6 month period (two 3-month follow-up intervals) over the life of the device. Additional use of EGM pre-storage reduces longevity by approximately 25% or 3 months per year.
500 ohm
b
impedance
pacing
900 ohm
pacing
impedance
11
Marquis VR 7230Cx, B, and E Reference Manual
12
Chapter 1

Replacement indicators

Replacement indicators
Battery voltage and messages about replacement status appear on the programmer display and on printed reports. Table 1-8 lists the Elective Replacement Indicator (ERI) and the End of Life (EOL) conditions.
Table 1-8. Replacement indicators
Elective Replacement (ERI) 2.62 V
End of Life (EOL) 3 months after ERI
ERI date – The programmer displays the date when the battery reached ERI on the Quick Look and Battery and Lead Measurements screens.
Temporary voltage decrease – The battery voltage temporarily decreases following a high voltage charge. If a battery measurement is taken immediately after a high voltage charge, the ERI or EOL indicator may be displayed. However, this is a temporary status which will return to normal when the battery has recovered from the charge.
EOL indication – If the programmer indicates that the device is at EOL, replace the device immediately.
Post-ERI conditions – EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 ppm (min
0.4 ms; 500 pacing load; and six 30 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.
Marquis VR 7230Cx, B, and E Reference Manual
-1
), 3 V,

Typical charge times

The most recent capacitor charge time appears on the programmer display and on printed reports and can be evaluated using the Charge/Dump test (see Table 1-9).
Ta b l e 1 - 9 . T yp i c a la full energy charge times
At Beginning of Life (BOL) 5.9 seconds
At Elective Replacement (ERI) 7.7 seconds
a
These charge times are typical when the capacitors are fully formed.

High voltage therapy energy

The stored energy of the device is derived from the peak capacitor voltage and is always greater than the energy delivered by the device. Table 1-10 compares the programmed energy levels delivered by the device to the energy levels stored in the capacitors before delivery.
Quick reference
Typical charge times
13
Marquis VR 7230Cx, B, and E Reference Manual
14
Chapter 1

Stored data and diagnostics

Table 1-10. Comparing delivereda (programmed) and storedb energy levels
Energy (J) Charge
Delivered a/
Stored
b
Programmed
30 35 5.9 8 9.4 1.6
28 33 5.6 7 8.3 1.4
26 31 5.2 6 7.1 1.2
24 28 4.7 5 5.9 1.0
22 26 4.4 4 4.7 0.8
20 24 4.0 3 3.6 0.6
18 21 3.5 2 2.4 0.4
16 19 3.2 1.8 2.2 0.4
15 18 3.0 1.6 2.0 0.3
14 16 2.7 1.4 1.7 0.3
13 15 2.5 1.2 1.5 0.3
12 14 2.4 1.0 1.2 0.2
11 13 2.2 0.8 1.0 0.2
10 12 2.0 0.6 0.8 0.1
9 11 1.9 0.4 0.5 0.1
a
Energy delivered at connector block into a 75 ohm load.
b
Energy stored at end of charge on capacitor.
c
Typical charge time at Beginning of Life (BOL) with fully formed capacitors, rounded to the nearest tenth of a second.
Time
c
(sec)
Delivered a/ Programmed
Energy (J) Charge
Timec (sec)
b
Stored
Stored data and diagnostics
Table 1-11. Stored data and diagnostics
Episode data
Tachy episodes 150 VF/VT/FVT episodes: intervals, text, EGM,
EGM capacity for tachy episodes
Marquis VR 7230Cx, B, and E Reference Manual
QRS Snapshot
14 minutesb of dual-channel EGM, or
23.5 minutes
a
c
of single-channel EGM
Quick reference
Stored data and diagnostics
Table 1-11. Stored data and diagnostics (continued)
SVT/NST episodes 50 SVT/NST episodes: intervals, text, EGM,
QRS Snapshot store detailed episode records for NST episodes)
EGM capacity for SVT/NST episodes
2 minutes
3.6 minutes
EGM sources Six options: ventricular / far-field
EGM options Store before onset; Store during charging
a
(the device does not usually
d
of dual-channel EGM, or
e
of single-channel EGM
15
Flashback memory 2000 intervals (V-V): before latest VF, before
latest VT, and before interrogation
Counter data
Detection counters Lifetime total, since cleared, and since last
session
Episode counters Episodes:
VF, FVT, and VT
NST episodes and SVTs
Percentage pacing:
VS and VP percentages
Additional counters:
Single PVCs and PVC runs
Rate stabilization pulses and runs
Therapy efficacy counters
Counts for each VF, FVT, VT Therapy:
Delivered
Successful
Unsuccessful
Intervention (manually aborted)
Total number of aborted shocks
Other stored data
Patient Alert events Up to 10 log entries: text and date for the first
time an alert is triggered between interrogations
Battery and lead measurements
Battery voltage, last capacitor formation, last charge, lead impedance, EGM amplitude measurements, last high voltage therapy, and sensing integrity counter
Marquis VR 7230Cx, B, and E Reference Manual
16
Chapter 1
Stored data and diagnostics
Table 1-11. Stored data and diagnostics (continued)
Lead performance trends
14 days of daily measurements plus 80 weeks of weekly minimum and maximum measurements:
Lead impedance: ventricular pacing, defibrillation pathway, and SVC lead (if used)
Ventricular EGM amplitude (R-waves)
Cardiac Compass trends
14 months of measurement trends:
VT and VF episodes per day
High voltage therapies delivered per day
Ventricular rate during VT or VF
Periods of non-sustained tachycardia per day
Heart rate variability
Percent of time pacing is active
Patient activity
Average day and night ventricular heart rate
a
When Wavelet is set to On or Monitor
b
13.5 minutes if Wavelet is set to On or Monitor
c
22 minutes if Wavelet is set to On or Monitor
d
1.6 minutes if Wavelet is set to On or Monitor.
e
2.75 minutes if Wavelet is set to On or Monitor.
Marquis VR 7230Cx, B, and E Reference Manual

New and enhanced features

The following features are new or changed from the 7231 GEM III VR ICD.
Lead connection header options
The Marquis VR supports three types of lead connection headers. The connection-types are Cx, B, and E.
Cx-type connector is normally used with one multipolar transvenous lead in the ventricle for sensing, pacing, and delivering therapies.
B-type connector is normally used with one multipolar transvenous lead in the ventricle for sensing and pacing and with two or three high voltages electrodes placed to deliver cardioversion/defibrillation therapies.
E -type connector is normally used with two 5mm unipolar myocardial pacing and sensing lead connectors and two
6.5mm high voltage lead connectors.
Regardless of the connector type used, you can program the Active Can feature so that the device Can serves as a second high voltage electrode and the HVX port high voltage electrode if desired. For more information about connection pathways, see “B- and E-type connector pathways” on page 123.
Quick reference
New and enhanced features
1
can accommodate a third
17
Patient management
RapidRead telemetry – Communication between the device and
programmer is approximately 20 times faster than telemetry in previous Medtronic ICD devices. The magnitude of improvement depends on the amount and type of data that is interrogated. RapidRead telemetry is more reliable and has an increased range that makes placing the programming head easier.
Cardiac Compass trends report – This report displays up to 14 months of trend data related to tachyarrhythmia episodes, heart rate, and patient activity. See “Using Cardiac Compass to view long term clinical trends” on page 269.
1
The HVA port of the E-type connector serves as the HVX electrode.
Marquis VR 7230Cx, B, and E Reference Manual
18
Chapter 1
New and enhanced features
Patient Alert – The alert duration when a magnet is applied to the device is now 20 seconds. The device also provides several new alerts:
SVC (HVX) lead impedance out of range
Active Can off without SVC
VOO mode programmed
VF Detection programmed off, or fewer than four VF therapies enabled for at least six hours
charge circuit timeout occurred
excessive charge time ERI
For more information, see “Using the Patient Alert feature” on page 227.
EGM amplitude trends – The device automatically measures R-wave EGM amplitudes every day. These daily measurements are included in the data displayed on the Lead Performance Trends screen. See “Collecting lead performance data” on page 247.
1
EGM Amplitude test – You can use the EGM Amplitude test to measure R-wave EGM amplitudes. The results are reported on the EGM Amplitude test screen. See “Measuring EGM Amplitude” on page 297.
Lead impedance measurements for SVC (HVX)
other lead impedance measurements, the device provides an independent SVC (HVX) measurement to check the integrity of the supplementary high voltage electrode. See “Measuring lead impedance” on page 295.
Leadless ECG signal – If a supplementary high voltage electrode is placed in the SVC, the device provides the Leadless ECG signal through either the Can to SVC (HVB) to SVC (HVX) EGM source. See “Setting up data collection” on page 241.
1
SVC refers to the HVX electrode, for pathway information see “B- and E-type connector pathways” on page 123.
2
The HVA port of the E-type connector serves as the HVX electrode.
Marquis VR 7230Cx, B, and E Reference Manual
2
Along with
1
(HVX) or RV
Quick reference
New and enhanced features
Expanded pre-onset EGM storage – The device can now store up to 20 seconds of EGM before a tachycardia starts. See “Setting up data collection” on page 241.
Smart Auto Cap Formation – When the Auto Cap Formation Interval is set to Auto, the formation interval automatically adjusts to optimize device longevity and charge times. See “Smart Auto Cap” on page 180.
Ending a patient session – The device audits the programmed parameter settings when you end a patient session and alerts you if any of the settings are atypical. See “Starting and ending patient sessions” on page 204.
Wavelet test – The Wavelet test allows you to manually collect and assess the template used by the Wavelet criterion. See “Testing the Wavelet criterion” on page 291.
Auto Collection for Wavelet – The Wavelet criterion includes the option to have the device automatically collect and maintain the template used to distinguish between SVT and ventricular tachyarrhythmia. See “Details about Auto Collection” on page 101.
19
QRS Snapshot data – When the Wavelet criterion is programmed to On or Monitor, the device stores QRS Snapshot data with VT, VF, FVT and SVT episode records. This data includes graphical representations of up to eight QRS complexes, along with match scores and event classifications as “Match” or “No Match.”
Tachyarrhythmia detection
Wavelet Dynamic Discrimination criterion – The Wavelet
criterion is designed to prevent detection of rapidly conducted SVTs as ventricular tachyarrhythmias. It compares the shape of each QRS complex during a fast ventricular rate to a stored template. If enough events occur that match this template, Wavelet withholds detection. See “Enhancing detection with Wavelet” on page 94.
Marquis VR 7230Cx, B, and E Reference Manual
20
Chapter 1
New and enhanced features
Tachyarrhythmia therapy
VT Monitoring – VT detection can be set to Monitor, which allows
the device to detect and record VT episodes without delivering therapy or influencing VF detection. See “Monitoring episodes for termination or redetection” on page 90.
Onset criterion – The Onset criterion is designed to prevent detection of sinus tachycardia as VT by requiring that a rapid increase in ventricular rate occurs before VT events can be classified. See “Details about Onset” on page 104.
High Rate Timeout – High Rate Timeout can turn off detection enhancements (Wavelet, Onset, or Stability) if a high rate episode is longer than a programmed duration. See “Details about High Rate Timeout” on page 112.
Episode confirmation during and after charging – The device continually monitors the ventricular rhythm during and after charging for cardioversion or defibrillation (when VF confirmation is active) to ensure the arrhythmia is present before delivering the high voltage shock. See “Confirming VF after initial detection” on page 126 and “Confirming VT or FVT after detection” on page 145.
Programmable Active Can – If a supplementary electrode is connected to the SVC (HVX) port, you can deselect the device Can as a high voltage electrode. For more information about connection pathways, see “Delivery pathway electrodes” on page 123.
Bradycardia pacing
Accelerometer-based rate response – The device uses an
accelerometer to provide rate responsive pacing.
Additional bradycardia pacing modes – The device provides asynchronous pacing in the VOO pacing mode, and provides the OVO mode to disable pacing. See Chapter 8, "Treating bradycardia" on page 157.
Marquis VR 7230Cx, B, and E Reference Manual
EP studies
Quick reference
New and enhanced features
Defibrillation threshold testing support – The T-Shock and 50 Hz Burst induction screens allow you to monitor time between inductions, program ventricular sensing and VF therapy settings, adjust induction settings, select manual therapies, and retrieve episode records after therapy. See “How to perform defibrillation threshold testing” on page 50.
21
Marquis VR 7230Cx, B, and E Reference Manual
22
Chapter 1
New and enhanced features
Marquis VR 7230Cx, B, and E Reference Manual
System overview 24
Indications and usage 27
Contraindications 27
Patient screening 27

The Marquis VR system2

2
Marquis VR 7230Cx, B, and E Reference Manual
24
Chapter 2

System overview

System overview
The 7230 Marquis VR Single Chamber Implantable Cardioverter Defibrillator (ICD) system is an implantable medical device system that automatically detects and treats episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. The ICD system includes three major components:
ICD
The ICD senses the electrical activity of the patient’s heart via the sensing electrodes of the implanted ventricular leads. It then analyzes the heart rhythm based on selectable sensing and detection parameters. If the ICD detects a tachyarrhythmia, it delivers defibrillation, cardioversion, or antitachycardia pacing therapy to the patient’s heart. If the ICD identifies a bradyarrhythmia, it delivers bradycardia pacing therapy to the patient’s heart.
Leads
The ICD can be used with transvenous or epicardial defibrillation leads. The lead system should consist of a bipolar pacing/sensing lead (or paired unipolar
1
pacing/sensing leads) in the ventricle and one or two high voltage cardioversion/ defibrillation electrodes. You can program the Active Can device case as a high voltage electrode. The pacing and sensing electrodes sense cardiac activity and deliver pacing stimuli.
Programmer and software
2
The Medtronic 9790C
programmer and 9967 application
software allow you to
configure the detection, therapy, and bradycardia features
for your patient
perform electrophysiological studies and system tests
monitor, display, or print patient cardiac activity information
1
With an appropriate unipolar to bipolar adapter kit.
2
With the model 9767 or 9767L programming head
Marquis VR 7230Cx, B, and E Reference Manual
For information about:
indications, contraindications, lead compatibility, warnings and precautions, and patient selection, see the Marquis VR 7230 Implant Manual, which accompanies each device.
basic programmer and software desktop functions that are not included in Chapter 10, "Using the programmer" on page 185, see the manual accompanying the programmer.
installing the 9767 or 9767L programming head, see the manual accompanying the programming head.
implanting leads, refer to the manuals accompanying the leads.
Detecting and treating tachyarrhythmias
The ICD monitors the cardiac rhythm for short ventricular intervals that may indicate the presence of VF, VT, or FVT.
Upon detection of VF, the ICD delivers a biphasic defibrillation shock of up to 30 joules. If the VF episode persists, up to five more individually programmed defibrillation shocks can be delivered.
Upon detection of VT, the ICD delivers either a Ramp, Ramp+, or Burst antitachycardia pacing therapy or a biphasic cardioversion shock of up to 30 joules synchronized to a ventricular depolarization. If the VT episode persists, up to five more individually programmed VT therapies can be delivered. You can also program the ICD to monitor the VT episode without delivering therapy.
Upon detection of FVT, the ICD delivers either a Ramp, Ramp+, or Burst antitachycardia pacing therapy, or a biphasic cardioversion shock of up to 30 joules synchronized to a ventricular depolarization. If the FVT episode persists, up to five more individually programmed FVT therapies can be delivered.
The Marquis VR system
System overview
25
You can use the Wavelet, Stability and Onset criteria to help the ICD distinguish between true ventricular arrhythmias and rapidly conducted supraventricular tachycardia (SVT) and withhold therapy for SVT.
Marquis VR 7230Cx, B, and E Reference Manual
26
Chapter 2
System overview
Treating bradycardia
The ICD provides rate responsive ventricular pacing to treat bradycardia. An internal accelerometer senses the patient’s physical activity, allowing the ICD to increase and decrease the pacing rate in response to changes in the level of activity.
Monitoring for real-time and stored data
The ICD and programmer provide real-time information on
detection and therapy parameters and status during a patient session. The ICD also provides accumulated data on device operation, including stored electrograms, detected and treated tachyarrhythmia episodes, bradycardia interventions, and the efficacy of therapy. The Cardiac Compass report provides up to 14 months of clinically significant data, including arrhythmia episodes, therapies delivered, physical activity, heart rate, and bradycardia pacing activities.
All of this information can be printed and retained in the patient’s file or saved in electronic format on a floppy diskette.
Conducting electrophysiologic tests
You can use the system to conduct non-invasive electrophysiologic studies including manual delivery of any of the ICD therapies to manage an induced or spontaneous tachyarrhythmia.
Alerting the patient to system events
You can use the programmable Patient Alert monitoring feature to notify the patient with audible tones if certain conditions related to the leads, battery, charge time and therapies occur. The patient can then respond based on your prescribed instructions.
Marquis VR 7230Cx, B, and E Reference Manual

Indications and usage

The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias.

Contraindications

The Marquis VR system is contraindicated for
patients whose tachyarrhythmias may have transient or reversible causes, such as acute myocardial infarction, digitalis intoxication, drowning, electrocution, electrolyte imbalance, hypoxia, or sepsis.
patients with incessant VT or VF
patient who have a unipolar pacemaker
patients whose primary disorder is bradyarrhythmias or atrial arrhythmias.
The Marquis VR system
Indications and usage
27

Patient screening

Prior to implant, patients should undergo a complete cardiac evaluation, including electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed tachyarrhythmia therapies are recommended during and after the implantation of the device.
Other optional screening procedures could include exercise stress testing to determine the patient’s maximum sinus rate, and cardiac catheterization to determine if there is a need for concomitant surgery and/or medical therapy.
Marquis VR 7230Cx, B, and E Reference Manual
28
Chapter 2
Patient screening
Marquis VR 7230Cx, B, and E Reference Manual

Emergency therapy3

Delivering emergency therapies 30
How to deliver emergency 30 joule defibrillation 32
How to deliver emergency cardioversion 32
How to deliver emergency fixed burst pacing 33
How to deliver emergency VVI pacing 33
3
Marquis VR 7230Cx, B, and E Reference Manual
30
Chapter 3

Delivering emergency therapies

Delivering emergency therapies
The device provides the following emergency therapies:
defibrillation
cardioversion
fixed burst pacing
emergency VVI pacing
The default emergency therapy is 30 joule defibrillation. When you select [Emergency] and [DELIVER], the device charges and delivers a biphasic 30 joule shock along the AX>B pathway.
The programmer resets the emergency defibrillation energy to 30 joules each time you select [Emergency]. Emergency cardioversion and fixed burst values remain as selected for the duration of the session.
To return to other programming functions from an Emergency screen, select [Exit Emergency].
Effect on system operation
1
The device suspends the automatic detection features when emergency defibrillation, cardioversion, or fixed burst pacing therapies are delivered. Detection is not suspended during emergency VVI pacing. Removing the programming head or pressing [Resume] turns detection on again.
Aborting an emergency therapy
As a safety precaution, the programmer also displays an [ABORT] button which immediately terminates any emergency therapy in progress.
1
If Active Can is turned off, the defibrillation is delivered between the HVX and HVB electrodes. . For more information, see “Delivery pathway electrodes” on page 123
Marquis VR 7230Cx, B, and E Reference Manual
On-screen and display panel buttons
The on-screen [Emergency] button and the red mechanical Emergency button by the programmer display panel function the same at all times.
Emergency therapy
Delivering emergency therapies
31
Red Emergency
The on-screen [DELIVER] button and the yellow-on-blue mechanical Deliver button by the programmer display panel function the same during emergency operations only. The mechanical Deliver button operates only during emergency operations.
Yellow-on-blue
Deliver Button
Temporary parameter values
Emergency tachyarrhythmia therapies use temporary values that do not change the programmed parameters of the device. values are not in effect until you select [DELIVER]. After the tachyarrhythmia therapy is complete, the device reverts to its programmed values.
Button
Functions the same as on-screen [Emergency] button.
Functions the same as on-screen [DELIVER], but only during Emergency functions.
1
These
1
Delivery of Emergency VVI Pacing changes the programmed bradycardia pacing values to the emergency values (see page 33).
Marquis VR 7230Cx, B, and E Reference Manual
32
Chapter 3
Delivering emergency therapies
How to deliver emergency 30 joule defibrillation
3
4
2
How to deliver emergency cardioversion
3
4
5
2
1. Position the programming head over the device.
2. Select [Emergency].
3. Accept the defibrillation energy shown on the screen, or select Energy and select a new value from the window.
4. Select [DELIVER]. If delivery is not confirmed,
verify that the programming head is properly positioned and select [Retry] or [Cancel].
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Cardioversion].
4. Accept the cardioversion energy shown on the screen, or select Energy and select a new value from the window.
5. Select [DELIVER]. If delivery is not confirmed,
verify that the programming head is properly positioned and select [Retry] or [Cancel].
Marquis VR 7230Cx, B, and E Reference Manual
How to deliver emergency fixed burst pacing
3
4
5
2
How to deliver emergency VVI pacing
3
4
2
Emergency therapy
Delivering emergency therapies
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [Fixed Burst].
4. Accept the pacing interval shown on the screen, or select Interval for a new interval value.
5. Select [BURST Press and Hold].
If delivery is not confirmed, the programmer displays an error window. Verify that the programming head is properly positioned. Select [OK] from the window and reselect [BURST Press and Hold].
1. Position the programming head over the device.
2. Select [Emergency].
3. Select [VVI Pacing].
4. Select [PROGRAM]. A successful programming sets the device to the following maximum output bradycardia pacing values.
Pacing Mode: VVI
Lower Rate: 70 ppm (70 min-1)
V. Amplitude: 8 V
V. Width: 1.6 ms
V. Pace Blanking: 320 ms
Hysteresis: Off
Vent r i cula r Rate Stabilization: Off
If programming is not confirmed, verify that the programming head is properly positioned and select [Retry] or [Cancel].
33
Marquis VR 7230Cx, B, and E Reference Manual
34
Chapter 3
Delivering emergency therapies
Marquis VR 7230Cx, B, and E Reference Manual
Part II

Device implant and patient follow-up procedures

Marquis VR 7230Cx, B, and E Reference Manual

Implanting the ICD4

Overview 38
Preparing for an implant 38
Replacing an ICD 40
Positioning the leads 41
Testing sensing and pacing thresholds 43
Connecting the leads to the ICD 44
Testing defibrillation operation and effectiveness 47
Positioning and securing the ICD 52
Completing the implant procedure 52
4
Marquis VR 7230Cx, B, and E Reference Manual
38
Chapter 4

Overview

Overview

Preparing for an implant

The tasks for implanting an ICD include
1. Preparing for an implant
2. Replacing an ICD
3. Positioning the leads
4. Testing sensing and pacing thresholds
5. Connecting the leads to the ICD
6. Testing defibrillation operation and effectiveness
7. Positioning and securing the ICD
8. Completing the implant procedure
These tasks are described in the sections that follow.
Warning: Keep a back-up external defibrillator available during the implant for transthoracic rescue when arrhythmias are induced.
Equipment for an implant
Model 9790C programmer and Model 9767 or 9767L programming head
9967 software application
8090 Analyzer lead analysis device or equivalent pacing system analyzer
external defibrillator
5358 Defibrillator Implant Support Device and software application (optional)
Marquis VR 7230Cx, B, and E Reference Manual
Sterile supplies
Marquis VR ICD and lead system components
Programming head sleeve or programming head
Analyzer cables
Lead introducers appropriate for the lead system
Extra stylets of appropriate length and shape
How to prepare for implanting
Set up the implant support instrument
When using an implant support instrument such as the 5358 Defibrillator Implant Support Device:
1. Calibrate any monitoring or recording equipment while recording the EGM and marker outputs of the support instrument.
2. Verify the high energy output of the support instrument by delivering a high energy defibrillation shock into the test load.
Set up the programmer and start the application
1. Set up the programmer as described in the instructions provided with the programmer.
2. Install the Marquis VR Model 9967 software on the programmer, if it is not already installed.
3. Place the programming head over the device and start the application. Select the device model or select [Auto identify].
Note: The programmer automatically interrogates the device when the application starts.
Implanting the ICD
Preparing for an implant
39
Preprogram the device
Before opening the sterile package, prepare the ICD for implant as follows:
1. Check the “use by” date printed on the package. Do not implant the device after the “use by” date because the battery’s longevity could be reduced.
2. Interrogate the ICD, and print a full summary report.
3. Confirm that the battery voltage is at least 3.0 V at room temperature (see page 225).
4. Perform a manual capacitor formation (see page 299).
5. Set up data collection parameters and the ICD internal clock (see page 243).
6. Program the therapy and pacing parameters to values appropriate for the patient (see page 160). Ensure that all tachyarrhythmia detection is programmed Off (see page 64).
Marquis VR 7230Cx, B, and E Reference Manual
40
Chapter 4

Replacing an ICD

Replacing an ICD
If you are replacing a previously implanted ICD, turn off ICD detection and therapies before explanting.
When implanting the ICD with a chronic lead system, perform the following evaluations to ensure appropriate detection and therapy:
Check the integrity of the chronic high voltage leads with a test shock, chest X-ray, and inspection.
Perform chronic pacing and sensing measurements.
Measure high voltage lead impedances.
Test defibrillation efficacy.
Confirm adequate sensing during VF.
Ensure proper fit of the lead connectors in the ICD connector block.
Notes:
To meet the implant requirements, it may be necessary to reposition or replace the chronic leads or to add a third high voltage electrode.
Any unused leads that remain implanted must be capped.
How to explant and replace an ICD
1. Program all tachyarrhythmia detection Off.
2. Dissect the leads and the ICD free from the surrounding tissues in the surgical pocket. Be careful not to nick or breach the lead insulation during the process of exposing the system.
3. Loosen each setscrew, and gently retract the lead from the connector block.
4. Remove the ICD from the surgical pocket.
5. If the connector pin of any implanted lead shows signs of pitting or corrosion, replace the implanted lead with a new lead. The damaged lead should be discarded and replaced to assure the integrity of the device system.
6. Measure sensing, pacing, and defibrillation efficacy using the replacement ICD or an implant support instrument.
7. Evaluate the defibrillation efficacy of the replacement system.
Marquis VR 7230Cx, B, and E Reference Manual

Positioning the leads

Implant endocardial leads according to the supplied instructions, unless suitable chronic leads are already in place. Do not use any lead with this device without first verifying connector compatibility (refer to the Marquis VR 7230Cx, 7230B, 7230E Implant Manual). Transvenous or epicardial leads may be used.
Using transvenous leads
Use standard transvenous implant techniques to position the ventricular lead tip in the right ventricular apex.
Follow the general guidelines below for initial positioning of other transvenous leads (the final positions are determined by defibrillation efficacy tests):
SVC (HVX) lead: Place the lead tip high in the innominate vein, approximately 5 cm proximal to the right atrium (RA) and SVC junction.
SQ patch: Place the patch along the left mid-axillary, centered over the fourth-to-fifth intercostal space.
CS lead: Advance the lead tip to just under the left atrial appendage, if possible.
Implanting the ICD
Positioning the leads
41
If using a subclavian approach, position the lead laterally to avoid pinching the lead body between the clavicle and the first rib.
Using epicardial leads
A variety of surgical approaches can be used to implant epicardial leads, including a limited left thoracotomy or median sternotomy. A typical placement may use an anterior right ventricular patch as the RV (HVB) and a posterolateral left ventricular patch as SVC (HVX).
Warning: Pinching the lead can damage the lead conductor or insulation, which may cause unwanted high voltage therapies or result in the loss of sensing or pacing therapy.
Marquis VR 7230Cx, B, and E Reference Manual
42
Chapter 4
Positioning the leads
Surgical incisions
Follow the general guidelines below for positioning epicardial leads:
If unipolar epicardial pacing leads are used, position the electrodes about 1 to 2 cm apart to reduce electromagnetic interference, and route the leads together with several loose twists.
Suture the smooth face of each patch lead against the epicardium or pericardium in locations that produce optimal defibrillation.
Place the patches so that they encompass the maximum amount of cardiac mass and they have approximately equal amounts of mass between them.
Ensure that the patches do not overlap and the electrode portions do not touch.
Avoid placing extra-pericardial patches over the phrenic nerve.
A single-incision submuscular or subcutaneous approach is recommended when the ICD is implanted in the pectoral region. Make the implant pocket about 1.5 times the size of the ICD.
Submuscular implant – An incision extending over the deltoid-pectoral groove typically provides access to the cephalic and subclavian veins as well as the implant pocket. Place the ICD sufficiently medial to the humeral head to avoid interference with shoulder motion.
Subcutaneous implant – A transverse incision typically permits isolation of the cephalic vein. Place the ICD far medially to keep the leads away from the axilla. Make sure that the upper edge of the ICD remains inferior to the incision.
Marquis VR 7230Cx, B, and E Reference Manual

Testing sensing and pacing thresholds

Testing sensing and pacing thresholds
Sensing and pacing tests include the following measurements:
EGM amplitude
slew rate
pacing threshold
pacing lead impedance
Medtronic recommends that you use an 8090 Analyzer lead analysis device to perform sensing and pacing measurements.
Refer to the technical manual for the analyzer you use to find details on performing sensing and pacing measurements.
Parameters
Measured sensing and pacing values must meet the following specific requirements at implant.
Table 4-1. Sensing and pacing values at implant
Measurement Acute Transvenous Leads Chronic Leads
R- wave amplitude 5 mV 3 mV
Ventricular slew rate: 0.75 V/s 0.5 V/s
Ventricular capture threshold:
a
At 0.5 ms pulse width
a
1.0 V 3.0 V
Implanting the ICD
43
Considerations
When measuring sensing and pacing values, measure between the tip (cathode) and ring or coil (anode) of each bipolar pacing/sensing lead.
For unipolar epicardial pacing leads, either electrode can be the cathode; use the configuration that yields the lower pacing threshold.
Note: Do not measure the intracardiac EGM telemetered from the ICD to assess sensing.
Marquis VR 7230Cx, B, and E Reference Manual
44
Chapter 4

Connecting the leads to the ICD

Connecting the leads to the ICD
For more detailed information about lead/connector compatibility, see the Marquis VR 7230Cx, 7230B, 7230E Implant Manual, or contact Medtronic Technical Services at 1-800-723-4636.
Warning: Loose lead connections may result in inappropriate sensing and failure to deliver necessary arrhythmia therapy.
Caution: Use only the torque wrench supplied with the device. It is designed to prevent damage to the device from overtightening a setscrew.
Table 4-2. Cx-type connector characteristics
General description
One IS-1 connector for pacing and sensing, Two DF-1 connectors for high voltage therapy, Active Can electrode (programmable)
Device Port
SVC DF-1 HVX
RV DF-1 HVB
Can n/a HVA, Can
V IS-1 bipolar
Connector Typ e
Software Name
Figure 4-1. Cx-type connector
5
1 suture holes
2 SVC port (DF-1)
3 RV port (DF-1)
4 Programmable Active Can
5 V. port (IS-1)
Marquis VR 7230Cx, B, and E Reference Manual
1
2
3
4
Implanting the ICD
Connecting the leads to the ICD
Table 4-3. B-type connector characteristics
General description
Port Connector Type
HVB 6.5 mm HVB
Device
Three 6.5mm unipolar high voltage ports and one
3.2mm low profile bipolar pace/sense (IS-1 compatible) port.
a
The HVA and HVX ports are electronically connected and are treated as the HVX electrode. For more information see “B- and E-type connector pathways” on page 123.
HVA
HVX
a
a
6.5 mm HVX
6.5 mm HVX
Can n/a HVA, Can
P/S bipolar pace sense
3.2 mm bipolar (IS-1 compatible)
Figure 4-2. B-type connector
Software Name
45
7
2
6
5
3
4
1 suture holes
2 HVB port (6.5 mm)
3 HVA port (6.5 mm)
4 Programmable Active Can
5 HVX port (6.5 mm)
6/7 P+/S and P-/S port (3.2 mm)
1
Marquis VR 7230Cx, B, and E Reference Manual
46
Chapter 4
Connecting the leads to the ICD
Table 4-4. E-type connector characteristics
General description
Two 6.5mm unipolar high voltage ports and two 5mm unipolar pace/sense ports.
a
The HVA port of the E-type connector can be used as the HVX electrode when Active Can is programmed off, see “B- and E-type connector pathways” on page 123.
Figure 4-3. E-type connector
Device Port
Connector Type
Software Name
P+/S port 5.0 mm
HVA port 6.5 mm HVX
a
HVB port 6.5 mm HVB
Can n/a HVA, Can
P-/S port 5.0 mm
6
5
2
3
4
1 suture holes
2 P+/S port (5.0 mm)
3 HVA port (6.5 mm)
4 Programmable Active Can
5 HVB port (6.5 mm)
6 P-/S port (5.0 mm)
For easier lead insertion, insert the ventricular IS-1 leg before the other legs.
1
Marquis VR 7230Cx, B, and E Reference Manual

Testing defibrillation operation and effectiveness

How to connect the lead to the device
1
a
2
7230Cx connector shown in example
b
30SetScrew.eps
30LeadTIp.eps
Implanting the ICD
1. Insert the torque wrench into the appropriate setscrew.
a. If the port is obstructed, retract the
setscrew to clear it. Take care not to disengage the setscrew from the connector block.
b. Leave the torque wrench in the setscrew
until the lead is secure. This allows a pathway for venting trapped air when the lead is inserted.
2. Push the lead or plug into the connector port until the lead pin is clearly visible in the pin viewing area. No sealant is required, but sterile water may be used as a lubricant.
3. Tighten the setscrew by turning clockwise until the torque wrench clicks.
4. Tug gently on the lead to confirm a secure fit. Do not pull on the lead until all setscrews have been tightened.
5. Repeat these steps for each lead.
47
Testing defibrillation operation and effectiveness
To demonstrate reliable defibrillation effectiveness with the implanted lead system, use the ICD to complete one of the following tests:
Terminate two consecutively induced VF episodes using a delivered energy of 20 joules or less.
Use the binary search procedure to establish a defibrillation threshold (DFT) of 18 joules or less.
Note: An external defibrillation implant support instrument can also be used to test defibrillation effectiveness. For instructions, see the applicable documentation for the specific support instrument.
Marquis VR 7230Cx, B, and E Reference Manual
48
Chapter 4
Testing defibrillation operation and effectiveness
High voltage implant values
Binary search protocol
Measured values must meet the following requirements at implant.
Table 4-5. High voltage therapy values at implant
Measurement Acute or Chronic Leads
V. Defib impedance
SVC (HVX) impedance (if applicable)
Defibrillation threshold
Warning: Ensure that an external defibrillator is charged for a rescue shock.
20 - 200 ohms
20 - 200 ohms
20 J (two consecutive) or 18 J (binary search)
By testing defibrillation efficacy at successively lower energy levels, the binary search protocol provides an accurate threshold measurement. However, it generally requires more inductions and more time. Also, cumulative charging of the ICD capacitor affects the device longevity.
For most reliable defibrillation efficacy testing, allow at least five minutes between VF inductions. The ICD software provides an on-screen timer to record elapsed time since the last induction.
If a two-electrode system fails to meet the implant criterion, consider implanting a third electrode and connecting it to the SVC (HVX) port. You can also evaluate the efficacy of ventricular tachycardia therapies after successfully completing ventricular defibrillation testing.
1
The timer appears on the 50 Hz Burst and T-Shock induction screens.
Marquis VR 7230Cx, B, and E Reference Manual
1
Figure 4-4. Binary search protocol
Implanting the ICD
Testing defibrillation operation and effectiveness
Start
49
First induced episode
Second induced episode
Third induced episode
Defibrillation Threshold
Success Failure FailureSuccess
3 J
Success Failure Success Failure Success Failure
<=3 J 6 J 9 J 15 J 18 J
Success Failure
6 J
9 J 15 J
Implant
Criterion Met
12 J
12 J
How to prepare for defibrillation threshold testing
1. Place the programming head over the ICD, start a patient session, and interrogate the device, if you have not already done so.
2. Observe the Marker Channel telemetry annotations and the programmer ECG display to verify that the ICD is sensing properly.
3. Conduct a manual Lead Impedance Test lead connections. Perform this test with the ICD in the surgical pocket and keep the pocket very moist. If the impedance is out of range, perform one or more of the following tasks:
Recheck lead connections and electrode placement.
Repeat the measurement.
Inspect the bipolar EGM for abnormalities.
Measure the defibrillation impedance with a manual test shock.
4. Program the ICD or support instrument to properly detect VF with an adequate safety margin (1.2 mV sensitivity).
a
See “Measuring lead impedance” on page 295.
18 J
Success
Reprogram polarity, reposition the lead, or try a different lead
a
Try to meet
2 at 20 J
Failure
to verify the defibrillation
74binary.eps
Marquis VR 7230Cx, B, and E Reference Manual
50
Chapter 4
Testing defibrillation operation and effectiveness
How to perform defibrillation threshold testing
3
2
4
6
5
1. Select Tests > EP Study.
2. Select either 50 Hz BURST or T-shock induction.
3. Select [Resume at BURST] or [Resume at DELIVER].
4. Select [Adjust Permanent...].
1
5. Program VF Enable On.
6. Program the automatic therapy energy settings. Therapies 2-6 should be set to the maximum energy.
7. Select [Program].
7
8
8. Select [Close].
Marquis VR 7230Cx, B, and E Reference Manual
13
11
12
Implanting the ICD
51
Testing defibrillation operation and effectiveness
9. If performing a T-Shock induction, select the [Enable] checkbox.
10. Select [DELIVER], or [50 Hz BURST Press and Hold].
If necessary, you can abort an
9
10
14
induction or therapy in progress by pressing [ABORT].
11. Observe the live rhythm monitor for proper post-shock sensing.
12. If using the binary search protocol, use the [Adjust Permanent...] button to program the next appropriate energy level (see Figure 4-4).
13. Wait until the on-screen timer reaches 5 minutes, then repeat steps 9 through 12 as needed.
14. Select Params > Detection and program VF, FVT, and VT detection Off before closing.
Marquis VR 7230Cx, B, and E Reference Manual
52
Chapter 4

Positioning and securing the ICD

Positioning and securing the ICD
Cautions:
If no SVC electrode is implanted, the pin plug provided with the device must be secured in the SVC port.
Program tachyarrhythmia detection Off before closing.
How to position and secure the device
Suture Hole Locations
1. Ensure that each lead pin or plug is fully inserted into the connector block and that all setscrews are tight.
2. Coil any excess lead length beneath the device. Avoid kinks in the lead conductors.
3. Implant the device within 5 cm of the skin. This position optimizes the ambulatory monitoring operations.
4. Suture the device securely within the pocket to
30Suture.eps
minimize post-implant rotation and migration of the device. Use a surgical needle to penetrate the suture holes. See Figure 4-2 and Figure 4-3 for suture hole locations on the B- and E-type connectors.

Completing the implant procedure

After implanting the device, X-ray the patient to verify the device and leads placement. To complete programming the device, select parameters that are appropriate for the patient.
Marquis VR 7230Cx, B, and E Reference Manual
How to complete programming the device
1. After closing the pocket, program detection On. Program ventricular tachyarrhythmia therapies On as desired.
2. If external equipment was used to conduct the defibrillation efficacy tests, perform a final VF induction, and allow the implanted system to detect and treat the arrhythmia.
3. Monitor the patient after the implant, and take X-rays as soon as possible to document and assess the location of the leads.
4. Program patient information. See “How to view and enter new patient information” on page 277.
5. Configure the Patient Alert feature. See “Using the Patient Alert feature” on page 227.
6. Set up data collection parameters. See “Setting up data collection” on page 241.
7. Interrogate the device after any spontaneous episodes to evaluate the detection and therapy parameter settings.
8. If the patient has not experienced spontaneous episodes, you may induce the clinical tachyarrhythmias using the non-invasive EP Study features to further assess the performance of the system. See Chapter 14, "Conducting Electrophysiologic Studies" on page 303.
9. Recheck pacing and sensing values, and adjust if necessary.
Implanting the ICD
Completing the implant procedure
53
Marquis VR 7230Cx, B, and E Reference Manual
54
Chapter 4
Completing the implant procedure
Marquis VR 7230Cx, B, and E Reference Manual
Conducting a patient follow-up
Patient follow-up guidelines 56
Verifying the status of the implanted system 56
Verifying accurate detection and appropriate therapy 57
Verifying effective bradycardia pacing 58
session
5
5
Marquis VR 7230Cx, B, and E Reference Manual
56
Chapter 5

Patient follow-up guidelines

Patient follow-up guidelines
Schedule regular patient follow-up sessions to monitor the condition of the ICD and leads and to verify that the ICD is configured appropriately for your patient.
During the first few months after receiving a new device, the patient may require close monitoring. Schedule an office visit at least every three months.
The Quick Look screen, which is displayed after you interrogate the device, provides a good beginning for the follow-up review. Using this screen you can
verify that the device is functioning correctly.
review the clinical performance and long term trends.
print appropriate reports1 to compare the results to the patient’s history and to retain for future reference.
Note: The Checklist feature provides a standard list of tasks to perform at a complete follow-up visit. You can also customize your own checklists if you wish. See “Streamlining follow-ups with Checklist” on page 235 for more information.

Verifying the status of the implanted system

To verify that the ICD and leads are functioning correctly, review the following information from the Quick Look screen and perform follow-up tests as indicated:
Review the displayed battery voltage for comparison to the Elective Replacement Indicator value (see page 12). Remember that battery voltage may be low if high voltage charging has occurred within 24 hours.
Review the last full energy charge.
– For information about adjusting the capacitor formation
interval, see “Optimizing charge time” on page 178.
– If the programmer displays an Excessive Charge Time ERI,
the ICD should be replaced immediately.
1
See “Using Cardiac Compass to view long term clinical trends” on page 269 for information on this new report.
Marquis VR 7230Cx, B, and E Reference Manual
Conducting a patient follow-up session

Verifying accurate detection and appropriate therapy

Review the defibrillation and pacing lead impedance values for inappropriate values or large changes since the last follow-up. See “Measuring lead impedance” on page 295.
Perform an EGM Amplitude test for comparison to previous EGM Amplitude measurements. See “How to perform an EGM Amplitude test” on page 298.
To review longer term trends in sensing and impedance measurements, select the [>>] button from the lead impedance area of the Quick Look screen. The programmer displays a detailed history of automatic sensing and impedance measurements. See “Taking a quick look at device activity” on page 225.
Verifying accurate detection and appropriate therapy
To verify that the ICD is providing effective tachyarrhythmia detection and therapy, review the following information from the Quick Look screen and investigate as indicated:
Review Quick Look Observations that relate to patient history and device operation. To display more detailed information about any observation, select the observation and then select the [>>] button.
Review any Patient alerts listed in the Observations of the Quick Look screen. For the most detailed information about Patient Alerts, select Patient Alert from the Data icon and select [Events].
Check stored episode records for appropriate sensing and detection of arrhythmias. See “Viewing episode data” on page 254.
Check stored SVT episode records for appropriate identification of SVTs.
57
Considerations
Review the following information before verifying detection and therapy.
Flashback memory – In addition to the episode text and stored electrograms, use Flashback memory and interval plots to help investigate the accuracy and specificity of ventricular detection.
Marquis VR 7230Cx, B, and E Reference Manual
58
Chapter 5

Verifying effective bradycardia pacing

Episode misidentification – If the episode records indicate that false detections have occurred, the Sensing Integrity counter may help in determining the prevalence of oversensing. For more information, see “Sensing integrity counter” on page 249.
If the ICD is oversensing, consider these programming options:
Increase the Pace Blanking value.
Increase the sensitivity threshold.
Caution: Do not re-program the ICD to decrease oversensing without assuring that appropriate sensing is maintained, See “Setting up sensing” on page 67.
If the episode records reveal that a stable monomorphic VT has been identified and treated as VF, consider these options to improve the detection accuracy:
Review the Interval Plot for the episode, and adjust VF Interval, if necessary. Use caution when reprogramming the VF Interval, because changes to this value can adversely affect VF detection.
Consider enabling FVT via VF detection. See “Detecting FVT episodes” on page 82.
If the SVT episode records include episodes of true VT, review the SVT episode record to identify the SVT detection criterion that withheld detection. Adjust the SVT detection criteria parameters as necessary. See “Enhancing detection with Wavelet” on page 94, “Enhancing VT detection with the Onset criterion” on page 103, and“Enhancing VT detection with the Stability criterion” on page 108.
Verifying effective bradycardia pacing
To verify that the ICD is sensing and pacing appropriately, review the following information from the Quick Look screen and investigate as indicated:
Confirm that the patient is receiving adequate cardiac support for daily living activities.
Review the pacing history for comparison to the patient history. A sharp increase in the paced beats percentage may indicate a need for investigation and analysis.
Marquis VR 7230Cx, B, and E Reference Manual
Considerations
Conducting a patient follow-up session
Verifying effective bradycardia pacing
Review the Cardiac Compass report for comparison to patient history (see page 269).
Conduct a pacing threshold test (see page 288) to verify that the programmed pacing parameters provide a sufficient safety margin.
Review the following information before verifying bradycardia pacing.
Ventricular Pacing – If the ventricle is predominantly paced and the patient exhibits adequate ventricular response, consider decreasing the Lower Rate.
% Pacing – Because the percentages for this counter are rounded to the nearest unit, they may not add up to 100%.
59
Marquis VR 7230Cx, B, and E Reference Manual
60
Chapter 5
Verifying effective bradycardia pacing
Marquis VR 7230Cx, B, and E Reference Manual
Part III

Configuring the ICD for the patient

Marquis VR 7230Cx, B, and E Reference Manual

Detecting tachyarrhythmias6

Detection overview 64
Setting up sensing 67
Detecting VF episodes 71
Detecting VT episodes 76
Detecting FVT episodes 82
Detecting tachyarrhythmia episodes with Combined Count 87
Monitoring episodes for termination or redetection 90
Enhancing detection with Wavelet 94
Enhancing VT detection with the Onset criterion 103
Enhancing VT detection with the Stability criterion 108
6
Detecting prolonged tachyarrhythmias with High Rate Timeout 111
Key ter ms 113
Marquis VR 7230Cx, B, and E Reference Manual
64
Chapter 6

Detection overview

Detection overview
The device detects ventricular tachyarrhythmias (VF, VT, and FVT) by comparing the time intervals between sensed ventricular events to a set of programmable detection intervals. If enough intervals occur that are shorter than the programmed intervals, the device detects a tachyarrhythmia, and responds automatically with a programmed therapy. After delivering the therapy, the device either redetects the arrhythmia and delivers the next programmed therapy or detects episode termination.
To avoid detecting rapidly conducted SVTs (for example, sinus tachycardia or atrial fibrillation) as ventricular tachyarrhythmias, the device provides several detection enhancements, including the Wavelet Dynamic Discrimination criterion, the Onset criterion, and the Stability criterion.
Figure 6-1 shows how all of these detection features interact during initial detection. During redetection, the device does not apply the Wavelet and Onset detection criteria.
Note: Detection functions can be turned off by programming the VF Enable, FVT Enable, and VT Enable parameters to Off. For an example, see “How to program VF detection” on page 74.
Marquis VR 7230Cx, B, and E Reference Manual
Detecting tachyarrhythmias
Figure 6-1. How detection features interact during initial detection
65
Detection overview
No
No
Ye s
No
No / suspended by
High Rate Timeout
Update counts and pattern information
Has High Rate Timeout suspended detection
Is the Onset criterion allowing VT event counting? (VT and
Does Stability reset the VT event count?
Ventricular Event
Is the interval in the
VF, FVT, or VT detection zone?
Ye s
enhancements?
No
FVT via VT detection only)
Ye s
(VT and FVT via VT detection only)
No
Has a tachyarrhythmia event count
reached an NID?
Ye s
Is Wavelet on?
Ye s
Ye s
Tachy
Episode
Detected
Ye s
No
Is the median ventricular interval less than the
SVT Limit?
No
Is Wavelet withholding detection?
Ye s
Marquis VR 7230Cx, B, and E Reference Manual
66
Chapter 6
Detection overview
Suspending tachyarrhythmia detection
When detection is suspended, the device temporarily stops classifying and counting tachyarrhythmia intervals. Sensing and bradycardia pacing remain active, and the programmed detection settings are not modified.
Detection is suspended
when the device senses the presence of a strong magnet. The programmer head contains a magnet which suspends detection, but once telemetry between the device and programmer is established, detection resumes.
while performing any of the manual system tests, including Underlying Rhythm, Pacing Threshold, Lead Impedance, EGM Amplitude, Wavelet, and Charge/Dump. Detection automatically resumes once the test is complete.
while performing a T-Shock, 50 Hz Burst, Manual Burst, or PES Induction. You can choose to have the device automatically resume detection after delivering the induction.
when you deliver a Manual or Emergency therapy. You can resume detection by selecting the [Resume] button or removing the programming head from the device.
when you select the on-screen [Suspend] button. You can resume detection by selecting the [Resume] button or by removing the programming head from the device.
during the automatic daily lead impedance measurements. Detection resumes when the measurements are complete.
while the device is delivering an automatic tachyarrhythmia therapy (including capacitor charging for defibrillation and cardioversion). However, the device does continue to confirm the detected episode during charging. Detection resumes when the therapy is complete.
Note: The device suspends VT detection (and Combined Count detection; see page 87) for 17 events following a defibrillation therapy delivered in response to a detected VF.
during charging for Automatic Capacitor Formation. Detection resumes when charging is complete.
1
1
If the defibrillation therapy is delivered as a result of a High Rate Timeout Therapy operation, VT detection is not suspended (see page 113).
Marquis VR 7230Cx, B, and E Reference Manual

Setting up sensing

The device provides bipolar sensing in the ventricle via the sensing electrodes of the implanted ventricular lead. You can adjust the sensitivity to intracardiac signals using the ventricular sensitivity setting. This setting defines the minimum electrical amplitude recognized by the device as a ventricular sensed event.
Proper sensing is essential for the safe and effective use of the device. To provide appropriate sensing, the device uses an auto-adjusting ventricular sensing threshold.
See details about sensing on page 69.
Parameters
V. Sensitivity (mV) – Minimum amplitude of
electrical signal that registers as a sensed ventricular event.
Considerations
Detecting tachyarrhythmias
Setting up sensing
* Medtronic nominal setting
0.15, 0.3*, 0.45, 0.6,
0.9, 1.2
67
Review the following information before programming sensing parameters.
Sensitivity thresholds – The programmed ventricular sensitivity threshold applies to all features related to sensing, including detection and bradycardia pacing.
Bradycardia pacing and sensing – A combination of high pacing pulse width or high amplitude with a low sensitivity threshold may cause inappropriate sensing. Programming a lower pulse width, lower amplitude, longer pace blanking, or a higher sensitivity threshold may eliminate this inappropriate sensing.
Recommended ventricular sensitivity threshold – A ventricular sensitivity threshold of 0.3 mV is recommended to maximize the probability of detecting VF and to limit the possibility of oversensing.
Marquis VR 7230Cx, B, and E Reference Manual
68
Chapter 6
Setting up sensing
High ventricular sensitivity threshold – Setting V. Sensitivity to a value greater than 0.6 mV is not recommended except for testing. Doing this may cause undersensing, which can cause any of the following situations:
delayed or aborted cardioversion therapy
delayed defibrillation therapy (when VF confirmation is active)
asynchronous pacing
underdetection of tachyarrhythmias.
Low ventricular sensitivity threshold – If you set V. Sensitivity to its most sensitive value of 0.15 mV, the device will be more susceptible to EMI and oversensing.
Testing sensitivity after reprogramming – If you change the ventricular sensitivity threshold, evaluate for proper sensing and detection by inducing VF and allowing the device to automatically detect and treat the arrhythmia.
Sensing during VF – Always verify that the device senses properly during VF. If the device is not sensing or detecting properly, program detection and therapies off, and evaluate the system (making sure to monitor the patient for life-threatening arrhythmias until you enable detection and therapies again). You may need to reposition or replace the ventricular sensing lead to achieve proper sensing.
Marquis VR 7230Cx, B, and E Reference Manual
How to program sensitivity
Details about sensing
Auto-adjusting sensitivity thresholds
Detecting tachyarrhythmias
69
Setting up sensing
1. Select Params > Detection
2. Select the desired V. Sensitivity parameter value.
3. Select [PROGRAM].
1
2
3
The device automatically adjusts the sensitivity thresholds after certain paced and sensed events to help reduce oversensing from T-waves and pacing. Figure 6-2 shows how sensitivity thresholds are adjusted after different types of events.
Marquis VR 7230Cx, B, and E Reference Manual
70
Chapter 6
Setting up sensing
Figure 6-2. Auto-adjusting sensitivity threshold
Sensitivity threshold
1 2
Rectified and Filtered
V. E G M
Marker Channel
30Autoadjust.eps
V S
V P
1 After a ventricular sensed event, the ventricular sensitivity threshold
increases to 75% of the EGM peak (maximum: 8x the programmed value, decay constant: 450 ms).
a
2 After the ventricular pace blanking period is finished, the ventricular
threshold increases to 4.5x the programmed value (maximum:
1.8 mV, decay constant: 450 ms).
a
The exponential decay continues through a subsequent ventricular pacing pulse and its blanking period.
Marquis VR 7230Cx, B, and E Reference Manual
Detecting tachyarrhythmias

Detecting VF episodes

Blanking periods
During a blanking period, the device does not sense electrical signals. This helps prevent sensing of device pacing, cardioversion and defibrillation pulses, post-pacing depolarization, T-waves, and multiple sensing of the same event. The blanking periods following paced events are longer than those following sensed events to avoid sensing ventricular depolarizations.
Table 6-1 shows the duration of the fixed blanking periods. For information on programmable pace blanking periods (see page 158).
Table 6-1. Fixed blanking periods
Ventricular blanking after sensed ventricular event 120 ms
71
Ventricular blanking after delivered cardioversion or defibrillation therapy
Refractory periods
During a refractory period, the device senses normally, but classifies sensed events as refractory and limits its response to these events. Refractory periods are used during synchronization to help prevent the device from delivering cardioversion and defibrillation therapies at inappropriate times. See “Synchronizing defibrillation without confirming VF” on page 125 and “Synchronizing cardioversion after charging” on page 146.
Note: Refractory periods do not affect tachyarrhythmia detection.
Detecting VF episodes
The device detects VF episodes by examining the cardiac rhythm for short ventricular intervals. If a predetermined number of intervals occurs that are short enough to be considered VF events, the device detects VF and delivers the first programmed VF therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing.
See details about VF detection on page 74.
520 ms
Marquis VR 7230Cx, B, and E Reference Manual
72
Chapter 6
Detecting VF episodes
Parameters
VF Detection Enable – Turns VF
detection on or off.
* Medtronic nominal setting
On*, Off
Considerations
VF Interval (ms) – V-V intervals shorter
than this value are counted as VF events.
VF Initial NID – Number of Intervals to Detect: number of VF events the device must count to detect a VF episode.
VF Redetect NID – Number of Intervals to Redetect: number of VF events the device must count to redetect a continuing VF after a therapy.
240, 250, . . ., 320*, . . .,400
12/16, 18/24*, 24/32, 30/40, 45/60, 60/80, 75/100, 90/120, 105/140, 120/160
6/8, 9/12, 12/16*, 18/24, 21/28, 24/32, 27/36, 30/40
Review the following information before programming VF detection parameters.
VF Interval minimum setting – To ensure proper VF detection, you should not program the VF Interval less than 300 ms.
VF Interval maximum setting – Programming the VF Interval to a value greater than 350 ms may cause inappropriate detection of rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter, which is more sensitive than the consecutive VT event counter.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
Enabling VF detection – When VF Detection Enable is programmed On for the first time, the device
enables Automatic Capacitor Formation
Marquis VR 7230Cx, B, and E Reference Manual
Restrictions
Detecting tachyarrhythmias
Detecting VF episodes
starts recording Cardiac Compass data
starts recording lead performance trends (starting at 3:00 AM, by the device clock)
clears all brady pacing counters
VF Detection and Wavelet – You can program the device to exclude rapidly conducted SVTs from VF detection by enabling the Wavelet Dynamic Discrimination criterion. Note that the SVT Limit must be programmed shorter than the VF Interval in order for Wavelet to affect VF detection. See “Enhancing detection with Wavelet” on page 94.
Review the following information before programming VF detection parameters.
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for bradycardia pacing and tachyarrhythmia detection. See “Parameter interlocks” on page 360.
73
VF detection backup – To ensure VF detection backup during VT and FVT episodes, if VF Detection is off, both VT Detection and FVT Detection must also be off.
Marquis VR 7230Cx, B, and E Reference Manual
74
Chapter 6
Detecting VF episodes
How to program VF detection
2
Details about VF detection
To program VF detection:
1. Select Params > Detection
2. Select the desired values for VF Enable, VF Initial NID, VF Redetect NID and VF Interval.
3. Select [PROGRAM].
1
3
The device detects VF by counting the number of VF events, which are V-V intervals shorter than the programmed VF Interval. On each event, the device counts the number of recent VF events. The number of recent events examined is called the VF detection window. The size of the VF detection window is the second number in the programmed VF NID (for example, 24 events if the VF Initial NID is 18/24).
The threshold for detecting VF is the first number in the programmed VF NID (for example, 18 events if the VF Initial NID is 18/24). This threshold is always 75% of the VF detection window. That is, if 75% of the events in the VF detection window are VF events, the device detects a VF episode (see Figure 6-3).
After the device detects VF, it delivers the first programmed VF therapy. Following the therapy, if the number of VF events reaches the programmed VF Redetect NID, the device redetects VF and delivers the next programmed VF therapy.
Note: The device can also detect VF Episodes via the Combined Count detection criterion (see page 87).
Marquis VR 7230Cx, B, and E Reference Manual
Figure 6-3. Device detects VF
ECG
Marker Channel
Detecting tachyarrhythmias
Detecting VF episodes
1 2 3
75
FSFSFSFSFSFSFSFSFSF
F
F
S
VSVSV
D
S
VF Event Count
V S
V S
FSFSF
V S
1 2 3 4 5 6 6 7 8 9 10 11 12 13 14 15 16 17 18
FSFSF
V
S
S
S
S
VF Interval
200 ms
1 VF starts, and the device begins counting VF events (intervals less than the programmed
VF Interval).
2 A ventricular interval occurs outside the VF detection zone. The VF event count is not
incremented.
3 The VF event count reaches the programmed VF NID value of 18 events out of 24, and the
device detects VF.
30VFDetection.eps
Marquis VR 7230Cx, B, and E Reference Manual
76
Chapter 6

Detecting VT episodes

Detecting VT episodes
Parameters
The device detects VT episodes by examining the cardiac rhythm for short ventricular intervals. If enough intervals occur that are short enough to be considered VT events (but are not VF or FVT events), the device detects VT and delivers the first programmed VT therapy. After therapy, the device continues to evaluate the ventricular rhythm to determine if the episode is ongoing.
You can program the device to detect and record VT episodes without treating them with VT therapies by setting VT Detection Enable to Monitor. If a patient’s VT episodes are well-tolerated, this feature allows you to collect data about these episodes without delivering therapy or affecting VF detection.
See details about VT detection on page 78.
* Medtronic nominal setting
Considerations
VT Detection Enable – Turns VT
detection on or off, or enables VT monitoring.
VT Interval (Rate) (ms) – V-V intervals shorter than this value are counted as VT events.
VT Initial NID – Number of Intervals to Detect: number of VT events the device must count to detect a VT episode.
VT Redetect NID – Number of Intervals to Redetect: number of VT events the device must count to redetect a continuing VT after a therapy.
On, Off*, or Monitor
280, 290, . . ., 400*, . . ., 600
12, 16*, . . ., 52, 76, 100
4, 8, 12*, . . ., 52
Review the following information before programming VT detection parameters.
VF, FVT, and VT Intervals – To allow for normal variations in the patient’s tachycardia interval, you should program the VF, FVT, and VT intervals at least 40 ms apart.
Marquis VR 7230Cx, B, and E Reference Manual
Restrictions
Detecting tachyarrhythmias
Detecting VT episodes
Episode redetection – You can expedite redetection by programming the VF and VT Redetect NIDs lower than the Initial NIDs.
VT detection and Combined Count detection – When VT Detection is On, the device applies the Combined Count detection criterion to help speed detection of rhythms that fluctuate between detection zones. Combined Count detection is disabled if VT Detection is set to Off or Monitor. See “Detecting tachyarrhythmia episodes with Combined Count” on page 87.
VT detection and rapidly conducted SVTs – You can program the device to exclude rapidly conducted SVTs from VT detection by enabling the Wavelet, Onset, or Stability detection criteria. See “Enhancing detection with Wavelet” on page 94, “Enhancing VT detection with the Onset criterion” on page 103, and “Enhancing VT detection with the Stability criterion” on page 108.
Review the following information before programming VT detection parameters.
77
Tachyarrhythmia detection and bradycardia pacing – To e n s u r e reliable ventricular tachyarrhythmia detection, the programmer regulates the values available for bradycardia pacing and tachyarrhythmia detection. See “Parameter interlocks” on page 360.
VF detection backup – To ensure VF detection backup during VT and FVT episodes, if VF Detection is off, both VT Detection and FVT Detection must also be off.
Marquis VR 7230Cx, B, and E Reference Manual
78
Chapter 6
Detecting VT episodes
How to program VT detection
2
Details about VT detection
To program VT detection:
1. Select Params > Detection
2. Select the desired values for VT Enable, VT Initial NID, VT Redetect NID and VT Interval.
3. Select [PROGRAM].
1
3
The device detects VT by counting the number of consecutive VT events. A VT event is a V-V interval shorter than the programmed VT Interval but greater than or equal to the VF Interval. If the number of consecutive VT events reaches the programmed VT Initial NID, the device detects VT (see Figure 6-4).
The VT event count resets to zero whenever an interval occurs that is greater than or equal to the programmed VT Interval. The count remains at the current value if an interval is shorter than the programmed VF Interval.
After the device detects VT, it delivers the first programmed VT therapy. Following the therapy, if the VT event counter reaches the VT Redetect NID, the device redetects VT and delivers the next programmed therapy.
Note: The device can also detect VT Episodes via the Combined Count detection criterion (see page 87).
Marquis VR 7230Cx, B, and E Reference Manual
Figure 6-4. Device detects VT
ECG
Marker Channel
132
Detecting tachyarrhythmias
Detecting VT episodes
79
VT Event Count
V S
VSV
S
V
T S
10123456
TSTSTSTSTST
S
S
VT Interval
ECG
Marker Channel
T D
V S
30VTDetection.eps
VT Event Count
TSTSTSTSTSTSTSTST
78910111213141516
S
VT Interval
200 ms
1 VT starts, and the device begins counting VT events (intervals less than the programmed
VT Interval, but greater than or equal to than the VF Interval).
2 A ventricular interval occurs outside VT detection zone. The VT event count resets to zero.
3 The VT event count reaches the programmed VT NID of 16 events, and the device detects
VT.
Marquis VR 7230Cx, B, and E Reference Manual
80
Chapter 6
Detecting VT episodes
VT monitoring
You can program the device to record VT episodes without delivering VT therapy by setting VT Detection to Monitor. When VT monitoring is enabled, the device detects VT episodes but does not deliver VT therapy (see Figure 6-5). Instead, it records VT episodes, labeling them as “monitored,” and waits for episode termination to occur.
When VT Detection is set to Monitor, several detection operations work differently.
VT event counting – Before the device detects an episode, it counts VT events normally. However, once the VT Initial NID is reached, the device sets the VT event count to zero and suspends VT event counting for the rest of the episode.
VF and FVT detection – VF and FVT detection operate as if VT detection is off. Specifically, Combined Count detection is disabled, and FVT via VT detection is not selectable. If a monitored VT episode accelerates into the FVT or VF detection zone, the device applies the VF Initial NID to detect the new tachyarrhythmia. Once an episode is in progress, VT event counting doesn’t resume until the episode ends.
Caution: Programming the VF Interval greater than 350 ms may result in inappropriate detection of rapidly conducted atrial fibrillation as VF or FVT via VF. Intervals shorter than the VF Interval are counted using the VF event counter, which is more sensitive than the consecutive VT event counter.
Wavelet, Onset, and Stability criteria – Before the device detects a tachyarrhythmia episode, the Wavelet, Onset, and Stability criteria, if turned on, are applied. If a monitored VT episode accelerates into the FVT or VF detection zone, the device continues to apply Wavelet as initial VF or FVT detection begins. However, since Stability and Onset do not affect either VF detection or FVT via VF detection, they are not applied.
Episode termination – The device compares ventricular intervals to the VT Interval to identify when a VT monitored episode has ended. However, if a VF episode or FVT via VF episode occurs when VT monitoring is enabled, the device compares ventricular intervals to the VF Interval to identify episode termination.
Marquis VR 7230Cx, B, and E Reference Manual
Loading...