To obtain information about a warranty, if any, contact Covidien Technical Services at
1.800.635.5267 or your local representative.
Purchase of this instrument confers no express or implied license under any Covidien patent
to use the instrument with any Ventilator System that is not manufactured or licensed by
Covidien.
Page 3
Copyright Information
Copyright 2011 Covidien. COVIDIEN, COVIDIEN with logo. The Covidien logo and
positive results for life are U.S. and internationally registered trademarks of Covidien
AG. Other brands are trademarks of a Covidien company. U.S. Patents 5,540,222;
5,596,984; 5,632,270; 5,664,560; 5,673,689; 5,909,731; 5,915,382; 5,934,274; 6,116,240;
6,142,150; 6,321,748; 6,467,478.
The information contained in this manual is the sole property of Covidien and may not
be duplicated without permission. This manual may be revised or replaced by Covidien
at any time and without notice. You should ensure that you have the most current
applicable version of this manual; if in doubt, contact the Technical Publications
Department of Covidien. While the information set forth herein is believed to be
accurate, it is not a substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals.
Covidien’s sole responsibility with respect to the ventilator, and its use, is as stated in
the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Covidien’s right to revise or
otherwise change or modify the equipment (including its software) described herein,
without notice. In the absence of an express, written agreement to the contrary,
Covidien has no obligation to furnish any such revisions, changes, or modifications to
the owner or user of the equipment (including its software) described herein.
Page 4
Preface
Warning
Caution
Warning
Definitions
This manual uses these special indicators to convey information of a
specific nature:
Indicates a condition that can endanger the patient or
the ventilator operator.
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular emphasis that make operation of
the ventilator more efficient or convenient.
Warnings, cautions, and notes
Please take the time to familiarize yourself with the following safety
considerations, special handling requirements, and regulations that
govern the use of the 700 Series Ventilator System.
ii
To avoid an electrical shock hazard while servicing the ventilator,
be sure to remove all power to the ventilator by disconnecting
the power source and turning off all ventilator power switches.
Page 5
Preface
Warning
Warning
Warning
Warning
To avoid a fire hazard, keep matches, lighted cigarettes, and all
other sources of ignition (e.g., flammable anesthetics and/or
heaters) away from the ventilator and oxygen hoses.
Do not use oxygen hoses that are worn, frayed, or contaminated
by combustible materials such as grease or oils. (Textiles, oils,
and other combustibles are easily ignited and burn with great
intensity in air enriched with oxygen.)
In case of fire or a burning smell, immediately disconnect the
ventilator from the oxygen supply and electrical power source.
Patients on life-support equipment should be appropriately
monitored by competent medical personnel and suitable
monitoring devices.
The 700 Series Ventilator is not intended to be a comprehensive
monitoring device and does not activate alarms for all types of
dangerous conditions for patients on life-support equipment.
Check the ventilator periodically as outlined in the service manual;
do not use if defective. Immediately replace parts that are broken,
missing, obviously worn, distorted, or contaminated.
An alternative source of ventilation should always be available
when using the 700 Series Ventilator System.
iii
Page 6
Preface
Warning
Warning
Warning
Caution
To ensure proper servicing and avoid the possibility of physical
injury, only qualified personnel should attempt to service or
make authorized modifications to the ventilator.
The user of this product shall have sole responsibility for any
ventilator malfunction due to operation or maintenance
performed by anyone not trained by Covidien staff.
For a thorough understanding of ventilator operations, be
sure to read the 700 Series Ventilator System Operator's Manual in its entirety before attempting to use the system.
Before activating any part of the ventilator, be sure to check
the equipment for proper operation and, if appropriate, run
the self-diagnostic short self test (SST) program described in
this manual.
iv
U.S. Federal law restricts this device to sale by or on the order
of a physician.
Warranty
The 700 Series Ventilator System is warranted against defects in
material and workmanship in accordance with Covidien Medical
Equipment Warranty for a period of one year from the time of sale.
To ensure the validity of the warranty, be sure to keep a maintenance
record.
Year of manufacture
The 700 Series Ventilator System’s year of manufacture is indicated
by the fifth and sixth digits of the serial number which is located at
the lower edge of the ventilator front panel.
Page 7
Preface
Manufacturer:
Covidien llc
15 Hampshire Street
Mansfield, MA 02048
USA
Authorized Representative:
Covidien Ireland Limited
IDA Business & Technology Park
Tull amore
Manufacturer
Electromagnetic susceptibility
The 700 Series Ventilator System complies with the requirements of
IEC 60601-1-2 (EMC Collateral Standard), which includes E-field
susceptibility and ESD requirements. However, even though the
device is compliant at the levels of immunity specified in the
standard, certain transmitting devices (cellular phones, walkietalkies, cordless phones, paging transmitters, etc.) emit radio
frequencies that could interrupt ventilator operation if located in a
range too close to the ventilator. It is difficult to determine when the
field strength of these devices becomes excessive. Practitioners
should be aware that radio frequency emissions are additive, and
that the ventilator must be located a sufficient distance from
transmitting devices to avoid interruption. Do not operate the
ventilator in a magnetic resonance imaging (MRI) environment.
The Alarm handling section of this manual describes possible
ventilator alarms and what to do if they occur. Consult with your
institution’s biomedical engineering department in case of
interrupted ventilator operation, and before relocating any life
support equipment.
v
Customer assistance
For further assistance contact your local Covidien representative.
Figure 2-3.Plugging the external battery into the ventilator ............................. 2-5
Figure 2-4.Disconnecting the external battery ................................................. 2-6
Figure 2-5.Connecting the ventilator power cord ............................................. 2-7
Figure 2-6.Storing the power cord on the ventilator ......................................... 2-7
Figure 2-7.Connecting the oxygen supply........................................................ 2-9
Figure 2-9.Installing the collector vial ............................................................ 2-13
Figure 2-10. Installing the flex arm.................................................................... 2-14
Figure 2-11. Shortening the flex arm................................................................. 2-15
Figure 2-12. Installing the humidifier ................................................................ 2-16
Figure 2-13. Locking and unlocking the cart’s front wheels .............................. 2-17
Figure 3-1.Turning the power switch on (the “I” position)................................. 3-2
Figure 5-1.Viewing patient data ....................................................................... 5-7
Figure 5-2.Volume bar graph (760 Ventilator only) .......................................... 5-8
Figure 6-1.Using the
Figure 6-2.Using the
Figure 7-1.Viewing active alarms ..................................................................... 7-2
More active alarms
Autoreset alarms
menu function to view autoreset alarms .... 6-7
menu function to view active alarms ....... 6-6
Page 16
This page is intentionally blank.
xiv
Page 17
Introduction
The 700 Series Ventilator System (including the 740 and 760
Ventilators) provides respiratory support for a wide range of
pediatric to adult patients for a wide variety of clinical conditions.
The ventilator’s mixing technique allows it to ventilate critically ill
patients at adjustable oxygen concentrations without the need for a
blender, compressor, or hospital-grade wall air.
The 700 Series Ventilator System can be mains- or battery-powered.
Each ventilator includes two microcontrollers: one for breath delivery
(which controls ventilation), and one for the user interface (which
monitors ventilator and patient data). Each microcontroller verifies that
the other is functioning properly. Using two independent
microcontrollers in this fashion prevents a single fault from causing a
simultaneous failure of controlling and monitoring functions.
The 700 Series Ventilator System supplies mandatory or spontaneous
breaths with a piston-based pneumatic system. Table 1-1 summarizes
the modes and breath types offered by the 740 and 760 Ventilators.
Mandatory breaths can be volume control ventilation (VCV, available
on 740 and 760 Ventilators) or pressure control ventilation (PCV,
available on the 760 Ventilator only). VCV delivers breaths to the
patient at a preset tidal volume, peak flow, waveform, and oxygen
concentration at a minimum respiratory rate. PCV delivers breaths to
Page 18
Introduction
the patient at a preset inspiratory pressure, I:E ratio or inspiratory
time, rise time factor (how quickly inspiratory pressure rises to
achieve the set inspiratory pressure), and oxygen concentration at a
minimum respiratory rate. A spontaneous breath allows the patient
inspiratory flows of up to 300 L/min, with or without pressure
support ventilation (PSV). On the 760 Ventilator, you can set the rise
time factor and exhalation flow sensitivity (that is, the point at
which the ventilator cycles from inspiration to exhalation) in PSV.
The ventilator begins apnea ventilation if no breath (patient- ,
ventilator-, or operator-initiated) is delivered within the selected
apnea interval. Apnea ventilation is active during all modes. On the
740 Ventilator, only VCV breaths are available in apnea ventilation.
On the 760 Ventilator, VCV or PCV breaths are available in apnea
ventilation.
The 760 Ventilator also offers the ability to perform respiratory
mechanics calculations and maneuvers as a standard feature using
the EXP PAUSE (to calculate auto-PEEP) and INSP PAUSE (to calculate
patient resistance and compliance) keys.
Table 1-1: Mode/breath type availability on 740/760 Ventilators
Mode/breath type740 Ventilator760 Ventilator
VCV breath type••
PCV breath type•
PSV breath type (support pressure
setting)
PSV (rise time factor and exhalation
sensitivity settings)
SIMV mode••
Apnea ventilation (VCV breath type)••
Apnea ventilation (choice of VCV or
PCV breath type)
Respiratory mechanics (EXP PAUSE
and INSP PAUSE)
1-2
••
•
•
•
Page 19
This manual tells you how to operate and perform simple
maintenance for the 700 Series Ventilator. Covidien recommends
that you become familiar with this manual and accompanying labels
before attempting to operate or maintain the ventilator. If you need
additional copies of this manual, contact your Covidien
representative.
To ensure optimum performance of the 700 Series Ventilator System,
Covidien recommends that a qualified service technician perform
periodic maintenance on the ventilator. For more information,
contact your Covidien representative.
1.1 Functional description
By pressing keys and turning the knob on the ventilator keyboard,
the operator gives initial instructions and data to the ventilator
(Figure 1-1). The user interface microcontroller processes this
information and stores it in the ventilator’s memory. The breath
delivery microcontroller uses this stored information to control and
monitor the flow of gas to and from the patient.
The 700 Series Ventilator uses a flow trigger to recognize patient
effort. The trigger monitors flow from the piston during exhalation.
When the patient inhales, patient circuit pressure drops very slightly
below end-expiratory pressure. At the same time, the piston moves
forward to deliver flow to the ventilator breathing circuit and
maintain the preset PEEP/CPAP level. The level of flow depends on
the patient’s effort. If this flow exceeds the user-set level, the
ventilator triggers. By design, the ventilator attempts to maintain
PEEP in the presence of a circuit leak. Since a leak drives the piston to
deliver flow to make up for pressure losses, a circuit leak can require
an increase in the flow trigger level to avoid autocycling.
Introduction
During exhalation, the ventilator’s piston retracts and draws air and
oxygen into the cylinder. The ventilator uses room air, which means
the ventilator can operate without a compressor or wall air source.
Room air enters the ventilator through a protected user-replaceable
air intake filter just inside the ventilator cabinet. This filter captures
airborne particles.
1-3
Page 20
Introduction
Oxygen from a cylinder or wall supply enters the ventilator through
a hose and oxygen fitting (the fitting is available in several versions).
Once inside the ventilator, the oxygen is regulated to a pressure the
ventilator can use, then mixed with air, according to the selected
%O
.
2
The flow-triggered piston/cylinder system and motor controller circuit control the flow of gas to the patient. On the 760 Ventilator in
PCV or PSV, the rate of flow is also determined by the preset rise time
factor. This system is designed with a minute gap (about the size of a
thin sheet of paper) between the piston and the cylinder wall.
This design eliminates the friction between the piston and cylinder,
allowing it to respond more rapidly than a “sealed” system.
A small amount of gas leaks through the gap between the piston
and cylinder. Ventilator software and a continuous forward motion
of the piston compensate for this leak.
The piston delivers the mixed air and oxygen through the inspiratory manifold system, and out to the patient. The oxygen concentration
and temperature of the delivered gas are monitored here, using a
galvanic oxygen sensor and a thermistor. The galvanic sensor
generates a voltage proportional to the partial pressure of oxygen,
from which the oxygen concentration is calculated. The ventilator
alarms if the monitored oxygen concentration is more than ten
percentage points above or below the % O
setting. The inspiratory
2
manifold system also includes a safety valve to relieve patient
pressure if necessary (for example, if the ventilator breathing circuit
is kinked or occluded).
The patient system includes the components external to the ventilator
that route gas between the ventilator and the patient. These
components include the inspiratory filter (which protects against
contamination between the ventilator and patient), a humidification
device, ventilator breathing circuit (the tubing through which the gas
travels), collector vial (which protects the exhalation system from
moisture in the exhaled gas, and can be emptied without losing circuit
PEEP), and an expiratory filter (which limits the bacteria in the
patient’s exhaled gas from escaping to room air or contaminating the
ventilator).
1-4
Page 21
Introduction
-
+
740 Ventilator
Filter
Keyboard
Regulator
Oxygen
Room air
Patient
7-00017
Inspiration
manifold
Gas intake
system
Inspiratory
filter
Humidification
device
Ventilator breathing circuit
(inspiratory limb)
Piston/cylinder
system
Exhalation/
PEEP/CPAP
system
Expiratory
filter
Collector
vial
Ventilator breathing circuit
(expiratory limb)
Figure 1-1 . Block diagram: 700 Series Ventilator function
The heated exhalation system monitors the flow of the patient’s
exhaled gas using a differential pressure transducer. The patient
exhales through the exhalation valve. During exhalation, the
PEEP/CPAP system maintains user-selected pressure in the ventilator
breathing circuit.
1-5
Page 22
Introduction
Throughout the respiratory cycle, pressure transducers monitor
inspiratory, expiratory, and atmospheric pressures. The temperatures
of the pneumatic compartment and inspiratory gas are also
monitored. Information from these transducers is continuously used
to update the calculations that control ventilation. (Appendix F
provides a diagram of the ventilator’s pneumatic system and
ventilator breathing circuit.)
Power to operate the ventilator comes from ac mains (wall) or
battery power. The power supply is designed to protect against
excessive voltages, temperatures, or current draws. A power cord
retainer prevents the cord from accidental disconnection.
The ventilator includes an internal battery, and accommodates an
optional external battery. Depending on the ventilator settings,
battery backup power can be supplied for up to 2½
hours using the
internal battery, and up to 7 hours using the external battery.
Both batteries are recharged during operation from ac power.
If both are installed, the external battery is used first when ac power
is not present. If the external battery is depleted or not installed, the
internal battery supplies power to the ventilator when ac power is
not available. The keyboard indicates the source of power and
battery charge level of the internal battery at all times.
Emergency modes: The ventilator declares a ventilator inoperative
(VENT INOP) condition if a hardware failure or critical software error
that could compromise safe ventilation occurs. In case of a ventilator
inoperative condition, the VENT INOP indicator lights and the
ventilator enters the safety valve open (SVO) state. To correct a
ventilator inoperative condition, the ventilator must be turned off,
then powered on again; at power-on, the operator must run
extended self-test (EST). The ventilator must pass EST before normal
ventilation can resume.
The safety valve allows the patient to breathe room air unassisted
when the ventilator is in the SVO state. The ventilator remains in the
SVO state until power-on self-test (POST) verifies that power levels to
the ventilator are acceptable and that the motor controller and
microcontrollers are functioning correctly, and until the user has
confirmed ventilator settings.
1-6
Page 23
Introduction
If the ventilator enters the SVO state and POST is not running, the
SAFETY VALVE OPEN indicator lights and a high-priority alarm
sounds. The ventilator enters the SVO state if it detects a hardware
or software failure that could compromise safe ventilation. In case of
a malfunction that prevents software from opening the safety valve,
there is also an analog circuit that opens the safety valve when
system pressure exceeds 115 cmH
O (113 hPa).
2
If the ventilator detects an occlusion or a continuous high inspiratory
pressure condition, it opens the safety and exhalation valves to vent
excess pressure, then shuts them and begins occlusion cycling mode.
In occlusion cycling mode the ventilator uses current settings except
for those summarized in Table 1-2. If the ventilator again detects an
occlusion or continuous high pressure condition, it again opens the
safety and exhalation valves then resumes occlusion cycling mode.
If the operator presses the alarm reset key or the ventilator does not
detect an occlusion or continuous high pressure condition, it reverts
to normal ventilation using the most recently accepted settings.
Table 1-2: Changes to current settings in occlusion cycling mode
SettingChange to setting
HIGH PRESSURE
alarm (VCV breath
type)
PEEP (all modes)Set to 0 cmH
% O
(all modes) Set to 100%
2
SPONT mode
(PSV breath type)
SUPPORT
PRESSURE
(PSV breath type)
SUPPORT
PRESSURE
(PCV breath type)
RISE TIME FACTOR
(PCV breath type)
Set to 30 cmH
Breaths are delivered at a rate of 12/min with an
inspiratory time of 2 seconds.
If less than 15 cmH
If 15 cmH
(no change).
If less than 15 cmH
If 15 cmH
(no change).
Set to 70%
O or above: the current setting is used
2
O or above: the current setting is used
2
O
2
O
2
O: set to 15 cmH2O.
2
O: set to 15 cmH2O.
2
1-7
Page 24
Introduction
SN
1.2 Symbols and labels
These symbols and labels appear on the 700 Series Ventilator System:
Power switch positions per IEC 601-1.”I” represents ON
position; “O” represents OFF position.
7-00421
Refer to manual per IEC 601-1. When this symbol appears
on product, it means “Refer to documentation for
7-00418
7-00416
7-00426
7-00414
information.”
Potential equalization point, per IEC 601-1
External battery connection
Circuit breaker
Serial number
ac current
7-00427
Type B equipment, per IEC 601-1
7-00415
Indicates the degree of protection provided by enclosure
(drip-proof).
7-00403
Signifies compliance with the Medical Device Directive, 93/
42/EEC
1-8
Page 25
CSA and NRTL (Nationally Recognized Testing Laboratory)
EXHAUST
TO
PATIENT
FROM
PATIENT
certification, granted by CSA
8-00417
The TUV Rheinland logo signifies TUV Rheinland Type Test
approval to Annex III of the Medical Device Directive
7-00420
Exhaust port connector
Inspiratory limb connector
Expiratory limb connector
Introduction
1-9
Page 26
Introduction
61006
Oxygen inlet port label
O
2
V
.
ma
Air intake label
61010
Cooling fan label
61009
General life support equipment warning label
61011
1-10
Page 27
740 Ventilator back panel label
Introduction
1-11
Page 28
Introduction
760 Ventilator back panel label
1-12
Page 29
1.3 Keyboard
Caution
Patient dataVentilator status
Ventilator settings
7-00048
To avoid damaging the keyboard, do not press on it with sharp
objects.
The keyboard (Figure 1-2 shows the 740 keyboard and Figure 1-3
shows the 760 keyboard) is grouped into three sections:
• VENTILATOR SETTINGS: Where you set breath delivery variables.
Introduction
• PATIENT DATA: Where you set alarm limits and view
mon
itored pressures, breath timing, and volumes.
• VENTILATOR STATUS: Where you see the alarm status and
operating condition of the ventilator.
Figure 1-2. 740 Ventilator System keyboard
the
1-13
Page 30
Introduction
Patient dataVentilator status
Ventilator settings
7-00123
Figure 1-3. 760 Ventilator System keyboard
1.3.1 VENTILATOR SETTINGS
The VENTILATOR SETTINGS section of the keyboard allows you to
select the ventilation mode, breath type, and settings. For more
detail on ventilation modes and breath delivery, see Appendix D.
To change the mode and settings, select the mode, then the breath
type, and then the ventilator settings. The keys flash during setup
and mode changes to ensure that you review all pertinent settings.
The keyboard is designed to minimize accidental or unintentional
changes.
Table 1-3 summarizes the functions of the keys, knob, and indicators
in the VENTILATOR SETTINGS section of the keyboard. Ventilator
settings are also limited by these breath delivery boundaries:
• I:E ratio 4:1 for PCV (760 Ventilator only), 3:1 for all other
breath types
• Inspiratory time = 0.2 to 8 seconds (excluding plateau)
• Expiratory time 0.2 seconds
1-14
Page 31
Introduction
• PEEP/CPAP + SUPPORT PRESSURE or INSPIRATORY PRESSURE
80 cmH
O (80 hPa)
2
NOTE:
Maximum SUPPORT PRESSURE is 70 cmH
INSPIRATORY PRESSURE is 80 cmH
2
O.
O, maximum
2
• SUPPORT PRESSURE or INSPIRATORY PRESSURE + PEEP/CPAP
< HIGH PRESSURE - 2 cmH
O (2 hPa)
2
• HIGH PRESSURE (in A/C and SIMV modes) > PEEP/CPAP +
7cmH
O (7 hPa)
2
• HIGH PRESSURE (in SPONT mode) > PEEP/CPAP + SUPPORT
PRESSURE + 2 cmH
O(2hPa)
2
• HIGH PRESSURE > LOW INSP PRESSURE
• Minute volume 50 L/min at an I:E ratio of 2:1
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Mode/breath type settings
A/CAssist/control modeVCV (volume control
ventilation) and PCV
(pressure control ventilation)
breath types.
(PCV available on 760
Ventilator only.)
SIMVSynchronous intermittent
mandatory ventilation mode
SPONTSpontaneous modePSV breath type
VCVVCV breath typeVCV available on 740 and
VCV, PCV (760 only), and PSV
(pressure support
ventilation) breath types.
760 Ventilators in A/C or
SIMV modes.
1-15
Page 32
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Mode/breath type settings (continued)
PCVPCV breath typePCV available on 760
Ventilator only in A/C or
SIMV modes.
PSVPSV breath typePSV available in SIMV or
SPONT modes.
Mandatory (VCV) settings
RESPIRATORY
RATE
TIDAL
VOLUME
PEAK FLOWMaximum flow of gas
PLATEAU (s)Length of inspiratory pause
Mandatory (PCV) settings (760 Ventilator only)
RESPIRATORY
RATE
The minimum number of
mandatory breaths the
patient receives per minute.
During apnea ventilation the
minimum RESPIRATORY RATE
setting is 6 /minute.
Volume delivered to the
patient during a mandatory
breath, compliancecompensated and corrected to
body temperature and
pressure, saturated (BTPS).
delivered during a mandatory
breath (BTPS). (Combined
with tidal volume, peak flow
defines the active portion of
inspiratory time.)
after a mandatory breath has
been delivered, during which
no gas is delivered.
The minimum number of
mandatory breaths the
patient receives per minute.
During apnea ventilation the
minimum RESPIRATORY RATE
setting is 6 /minute.
1 to 70 /minute
Accuracy:
(0.1 + 1%) /minute
40 to 2000 ml
Accuracy:
(10 ml + 10% of setting)
3 to 150 L/min
Accuracy:
(5 + 10% of setting) L/min
0.0 to 2.0 second
Accuracy: 0.05 second
1 to 70 /minute
Accuracy:
(0.1 + 1%) /minute
1-16
Page 33
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Pressure above PEEP during
the inspiratory phase of a PCV
breath.
You can use the MENU key to
select inspiratory time (TI) or
I:E ratio as the breath timing
setting for a PCV breath. You
can change the selected
breath timing setting (T
or I:E
I
ratio), but the setting remains
constant when you change
the respiratory rate in PCV.
Selecting I:E ratio makes the
set ratio of inspiratory time to
expiratory time for a PCV
breath.
The time for inspiratory
pressure to rise from 0 to 95%
of the target pressure level
during a PCV breath.
A setting of 100 = a 100-ms
rise time, and a setting of
5 = 80% of the inspiratory
time or 2500 ms, whichever is
less.
When you adjust this setting,
the message window shows
the actual time (in seconds) to
reach 95% of target pressure.
5 to 80 cmH
O (5 to 80 hPa)
2
Accuracy:
(3 + 2.5% of setting) cmH
Inspiratory time (T
):
I
0.2 to 8 seconds
Accuracy: 0.05 second
I:E ratio: ≤ 4:1
Accuracy: (0.1 + 2%)
5 to 100
O
2
NOTE:
To help determine the correct setting for inspiratory time, during
PCV the message window displays peak inspiratory flow, end
inspiratory flow, and end exhalation flow in L/min.
1-17
Page 34
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Spontaneous (PSV) settings
SUPPORT
PRESSURE
RISE TIME
FACTOR
(760 only)
Pressure above PEEP
maintained during
spontaneous inspiration.
Support pressure is
terminated when inspiratory
flow falls to 25% of peak
inspiratory flow, or to the
exhalation sensitivity setting
(760 only), or 10 L/min or 25%
of peak flow, whichever is
lower (740 only). Maximum
inspiratory time is 3.5 seconds
for adults, and 2.5 seconds for
pediatric patients.
The time for inspiratory
pressure to rise from 0 to 95%
of the target pressure level
during a PSV breath.
A setting of 100 = a 100-ms
rise time, and a setting of
5 = 80% of the inspiratory
time or 1500 ms (when adult
ventilator breathing circuit is
selected) or 600 ms (when
pediatric circuit is selected),
whichever is less.
When you adjust this setting,
the message window shows
the actual time (in seconds) to
reach 95% of target pressure.
0 to 70 cmH
O (0 to 70 hPa)
2
Accuracy:
(3 + 2.5% of setting) cmH
5 to 100
O
2
1-18
Page 35
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Spontaneous (PSV) settings (continued)
EXH
SENSITIVITY
(760 only)
The percent of peak expiratory
flow at which the ventilator
cycles from inspiration to
exhalation for spontaneous
breaths. The flow at which the
ventilator cycles from
inspiration to exhalation for
PSV breaths. Exhalation begins
when the inspiratory flow is
less than the set value.
To help set EXH SENSITIVITY
appropriately, the peak
inspiratory flow and end
inspiratory flow are displayed
in the message window in PSV.
(To ensure accurate estimates
of these flows, it is important
to run SST so that tubing
compliance calculations are
correct.)
Common settings
PEEP/CPAPPositive end expiratory
pressure/continuous positive
airway pressure. Minimum
pressure maintained during
inspiratory and expiratory
phases.
TRIGGER
SENSITIVITY
(L/min)
Inspiratory flow required to
trigger the ventilator to
deliver a breath.
1 to 80%
0 to 35 cmH
Accuracy:
(2 cmH
O + 4% of setting)
2
1 to 20 L/min
O (0 to 35 hPa)
2
% O
2
Percentage of inspired oxygen
of the gas delivered to the
21 to 100%
Accuracy: 3% full scale
patient.
NOTE:
It may take several minutes for the oxygen percentage
to stabilize.
1-19
Page 36
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Other keys, knobs, and indicators
APNEA
PARAMS key
Allows you to select VCV or PCV (760 Ventilator only) apnea
ventilation, and the apnea ventilation settings. The apnea
interval is adjustable from 10 to 60 seconds. Apnea
ventilation is available in all modes. For apnea parameter
ranges, see mandatory VCV settings and mandatory PCV
(760 Ventilator only) settings.
MENUAllows you to view active and reset alarms, run SST and EST,
adjust certain settings (including alarm volume, PCV timing,
volume bar graph display, endotracheal tube size, and date
and time), access oxygen sensor functions (calibrate, enable
or disable % O
alarm limits, and enable or disable display of
2
oxygen sensor reading), enter standby mode, view battery
information, display software revision, display service
information, and enter EasyNeb nebulizer functions.
(Chapter 6 tells you how to use the menu function.)
100% O
Switches the % O2 to 100% for 2 minutes, then returns to
2
the current % O2 setting. The 2-minute interval restarts
every time you press 100% O
. Once the 100% O2 has
2
started, you can press CLEAR to stop the maneuver (unless
you have entered a MENU key function or selected a
setting).
MANUAL
INSP
Delivers one mandatory breath to the patient according to
the current mandatory settings (in A/C or SIMV) or the
current apnea parameters (in SPONT). You can deliver a
MANUAL INSP at any time during the exhalation phase of a
breath as long as the exhaled flow is less than 30% of the
peak exhaled flow, except during apnea ventilation.
1-20
Page 37
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Other keys, knobs, and indicators (continued)
EXP PAUSE
(760 only)
Allows you to measure the patient’s auto-PEEP. An EXP PAUSE
maneuver causes the ventilator to close the exhalation valve
at the end of the expiratory phase, and not deliver the next
mandatory breath. At the end of the maneuver, the message
window shows the calculated value for auto-PEEP (expiratory
pressure at the beginning of the maneuver minus expiratory
pressure at the end of the maneuver) and total PEEP for
30 seconds.
The message window shows the end expiratory flow in L/min
at the beginning of each breath. If there is expiratory flow
when the ventilator delivers the next breath, auto-PEEP is
present.
The EXP PAUSE continues as long as you hold down the key,
and should last only until expiratory pressure stabilizes. An
EXP PAUSE maneuver ends when you release the key, the
patient initiates a breath, an alarm occurs, the expiratory
phase (including the maneuver) lasts more than 20 seconds,
or the ventilator detects a leak.
Auto-PEEP:
Range: 1 to 35 cmH
Accuracy: (1 cmH
O.
2
O + 3% of reading).
2
The EXP PAUSE maneuver is unavailable when the
RESPIRATORY RATE setting is less than 3 /minute.
1-21
Page 38
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Other keys, knobs, and indicators (continued)
INSP PAUSE
(760 only)
Allows you to measure the patient’s compliance and
resistance. An extended inspiratory pause also allows you to
expand the patient’s lungs for up to 10 seconds.
Pressing INSP PAUSE momentarily causes the ventilator to
wait until the end of the inspiratory phase of the current or
next mandatory breath (in SPONT mode, the ventilator
delivers a mandatory breath using the MANUAL INSP key
according to the apnea settings), stop breath delivery, and
keep the exhalation valve closed. The INSP PAUSE continues
until the ventilator detects a stable plateau pressure or
2 seconds have elapsed. An INSP PAUSE maneuver ends when
a stable plateau is reached or an alarm occurs.
Pressing INSP PAUSE for 2 or more seconds after the pause
begins causes the ventilator to deliver an inspiratory pause for
as long as you hold down the key. An extended INSP PAUSE
maneuver ends when you release the key or 10 seconds have
elapsed.
You can press CLEAR or release the INSP PAUSE key at any
time to cancel an INSP PAUSE maneuver, and alarms cancel
the maneuver.
At the end of the breath, the message window shows the
calculated value for compliance and resistance (if the
mandatory breath was a VCV breath) or compliance (if the
mandatory breath was a PCV breath) for 30 seconds.
Compliance:
Range: 1 to 150 mL/H
Accuracy: (1 mL/cmH
O.
2
O + 20% of reading).
2
Resistance:
Range: 0 to 150 cmH2O/L/second.
Accuracy: (3 cmH
O/L/second + 20% of reading).
2
CLEARPressing CLEAR before accepting a setting cancels the
proposed setting. Pressing CLEAR does not cancel accepted
settings.
Pressing CLEAR twice returns the ventilator to its previous
state (unless you have entered a MENU key function that
requires you to press CLEAR more times).
Pressing CLEAR during a 100% O
maneuver cancels the
2
maneuver.
1-22
Page 39
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicatorSpecifies...Range
Other keys, knobs, and indicators (continued)
ACCEPTMakes changes to settings effective. If you don’t press
ACCEPT within 30 seconds of proposing a new setting, the
user interface returns to its previous state.
KnobAdjusts the value of a setting or selects a menu option.
A setting value that flashes means that the knob is linked to
that setting. Turning the knob clockwise increases the value,
and turning the knob counterclockwise decreases the value.
CURRENTLights when the ventilator is operating according to the
displayed settings, or during apnea ventilation. (There is one
indicator for mandatory breaths, and one for spontaneous
breaths.)
PROPOSEDLights when you propose a mode or breath type, or you are
setting apnea parameters. Once a proposed setting is
accepted, it becomes effective at the next breath.
APNEA
PARAMS
indicator
Message
window
Lights when apnea ventilation is active.
Lights with PROPOSED indicator when you are setting apnea
parameters, and both indicators turn off once apnea
parameters are accepted.
Shows up to four lines of information (20 characters per
line).
First line: Reserved for the highest-priority active or
autoreset alarm. On the 760 Ventilator only, if no alarm is
active and the display of the oxygen sensor reading is
enabled, the % O
is displayed here. Otherwise, if VCV is the
2
current or proposed breath type and there are no active or
autoreset alarms, shows the selected VCV flow pattern (ramp
or square).
Second line: Information about the menu function or
settings, alarm silence time remaining, or current date and
time. During normal ventilation, shows “Flow (L/min).”
Third and fourth lines: Reserved for other messages. For
every breath type, peak and end inspiratory flows are
displayed on the third line, and end expiratory flow is
displayed on the fourth line (except that inspiratory flow is
not displayed during VCV breaths or VCV apnea ventilation).
1-23
Page 40
Introduction
1.3.2 PATIENT DATA
The PATIENT DATA section of the keyboard allows you to view the
pressure, breath timing, and volume of the patient’s breath. You can
also view the alarm settings. A lighted key indicates that a
measurement is selected, and its value appears in the display
window. Values are continuously displayed and updated during
ventilation.
Table 1-4 summarizes the functions of the keys and indicators in the
PATIENT DATA section of the keyboard.
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Pressure
MEAN
PRESSURE
PEAK
PRESSURE
PLATEAU
PRESSURE
(760 only)
PEEP/CPAP
(760 only)
FunctionRange
Shows the calculated value of
ventilator breathing circuit
pressure over an entire
respiratory cycle. Updated at the
beginning of each breath.
Shows the maximum pressure
measured during inspiration.
Updated at the beginning of
each expiratory phase. (Default
pressure display.)
Shows the pressure measured at
the end of the plateau period of
a mandatory inspiration
(whether the inspiration is in a
regular VCV breath or is part of
an inspiratory pause maneuver).
Updated at the beginning of
each expiratory phase. The
PRESSURE display shows a blank
if the ventilator does not detect
a stable plateau pressure.
Shows the pressure measured at
the expiratory limb before any
inspiratory effort. Updated at
the beginning of each
inspiratory phase.
0 to 99 cmH2O
(0 to 99 hPa)
Accuracy:
(3 + 4% of reading) cmH
0 to 140 cmH
O
2
(0 to 140 hPa)
Accuracy:
(3 + 4% of reading) cmH
0 to 140 cmH
O
2
(0 to 140 hPa)
Accuracy:
(3 + 4% of reading) cmH
0 to 140 cmH2O
(0 to 140 hPa)
Accuracy:
(3 + 4% of reading) cmH
O
2
O
2
O
2
O
2
1-24
Page 41
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Breath timing
RATE
(/min)
I:E RATIOShows the ratio of measured
INSP
TIME(s)
(760 only)
Shows the calculated value of
the total respiratory rate, based
on the previous 60 seconds or 8
breaths (whichever interval is
shorter). Updated at the
beginning of each breath.
(Default breath timing display.)
The calculation is reset (and
display is blank) when
ventilation starts, when apnea
ventilation starts or autoresets,
when you change the mode,
breath type, or RESPIRATORY
RATE setting, and when you
press the alarm reset key.
inspiratory time to measured
expiratory time. Updated at the
beginning of each breath.
The measured inspiratory time,
including breaths that are
truncated due to a HIGH
PRESSURE alarm. Updated at the
beginning of each expiratory
phase.
FunctionRange
1 to 199 /minute
Accuracy:
(0.1 +1% of reading)/minute
1:99.9 to 9.9:1
Accuracy:
(0.1 + 2%)
0.1 to 9.90 seconds
Accuracy:
0.05 seconds
1-25
Page 42
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Volume
EXHALED
VOLUME
(ml)
TOTAL
MINUTE
VOLUME
(L)
DELIVERED
VOLUME
(ml)
(760 only)
FunctionRange
Shows the patient’s measured
expiratory tidal volume
averaged over the last 5 breaths
(for A/C VCV breaths, ventilatorinitiated PCV breaths, and PCV
apnea breaths) or for the justcompleted breath (for all other
breaths). Corrected to BTPS and
compliance-compensated.
Updated at the beginning of
each inspiration. (Default
volume display.)
The calculation is reset when
ventilation starts, when apnea
ventilation starts or autoresets,
when you change the mode or
breath type, and when you press
the alarm reset key.
Shows the patient’s measured
expiratory minute volume, based
on the previous 60 seconds or 8
breaths (whichever interval is
shorter). Updated at the
beginning of each breath.
The calculation is reset when
ventilation starts, when apnea
ventilation starts or autoresets,
when you change the mode or
breath type, and when you press
the alarm reset key.
Shows the measured inspiratory
tidal volume for the justcompleted PCV or PSV breath.
Corrected to BTPS and
compliance-compensated.
Updated at the beginning of
each inspiration for PCV and PSV
breath types.
0 to 9 L
Accuracy:
(10 ml + 10% of reading)
0 to 99 L
Accuracy:
(10 ml + 10% of reading)
0 to 3000 ml
Accuracy:
(10 ml + 10% of reading)
1-26
Page 43
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
FunctionRange
Volume (continued)
SPONT
MINUTE
VOLUME
(L)
(760 only)
Shows the patient’s measured
expiratory minute volume for all
spontaneous breaths, based on
the previous 60 seconds or 8
breaths (whichever interval is
shorter). Updated at the
beginning of each breath.
The calculation is reset when
ventilation starts, when apnea
ventilation starts or autoresets,
when you change the mode or
breath type, and when you press
the alarm reset key.
Alarm settings
HIGH RATE An active alarm indicates that
measured respiratory rate is
higher than the alarm setting.
HIGH
TIDAL
VOLUME
An active alarm indicates that
exhaled volume for three out of
four consecutive breaths was
above the alarm setting.
LOW INSP
PRESSURE
An active alarm indicates that
monitored circuit pressure is
below the alarm setting at the
end of inspiration. Inactive in for
any spontaneous breath.
LOW
TIDAL
VOLUME
An active alarm indicates that
delivered volume for three out
of four consecutive breaths were
below the alarm setting. (If this
alarm is set to 0 ml and breath
type is PCV, an active alarm
indicates that delivered volume
is less than 3 ml for three out of
four consecutive breaths.)
0 to 99 L
Accuracy:
(10 ml + 10% of reading)
3 to 100 /minute
Accuracy:
(0.1 +1% of setting)/minute
20 to 6000 ml
Accuracy:
(10 ml + 10% of setting)
3 to 60 cmH
O (3 to 60 hPa)
2
Accuracy:
(1 + 3% of setting)
0 to 2000 ml
Accuracy:
(10 ml + 10% of setting)
1-27
Page 44
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
FunctionRange
Alarm settings (continued)
HIGH
PRESSURE
An active alarm indicates that
two consecutive breaths were
truncated because circuit
pressure reached the alarm
setting.
LOW
MINUTE
VOLUME
An active alarm indicates that
monitored minute volume is less
than the alarm setting, based on
an eight-breath running average
or the previous minute,
whichever is less.
Other indicators
Pressure
bar graph
Shows real-time pressures in
centimeters of water (cmH2O) or
hectopascals (hPa).
LEDs show the current HIGH
PRESSURE alarm setting and the
peak pressure of the last breath
during exhalation.
Volume
bar graph
(760 only)
Shows real-time exhaled volume
in milliliters (ml). Volumes are
compliance-compensated and
corrected to BTPS.
The active scale is determined by
the HIGH TIDAL VOLUME alarm
setting. LEDs show the current
HIGH TIDAL VOLUME and LOW
TIDAL VOLUME alarm settings.
You can use the MENU key to
enable or disable the volume bar
graph. During exhalation, LEDs
show the maximum exhaled
volume of the last breath.
MANDLights at the start of each breath
to indicate a ventilator- or
operator-initiated (time or
manually triggered) mandatory
breath is being delivered.
10 to 90 cmH
O
2
(10 to 90 hPa)
Accuracy:
(1 + 3% of setting)
0 to 50 L
Accuracy:
(10 ml + 10% of setting)
-10 to 90 cmH
(-10to90hPa)
Resolution: 1 cmH
If HIGH TIDAL VOLUME
setting < 500 ml: 0 to 500 ml
Resolution: 5 ml
If HIGH TIDAL VOLUME
setting
500 ml: 0 to 2000 ml
>
Resolution: 20 ml
Not applicable
O
2
O (1 hPa)
2
1-28
Page 45
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Other indicators (continued)
ASSISTLights at the start of each breath
to indicate a patient-initiated
mandatory (flow triggered)
breath is being delivered.
SPONTLights at the start of each breath
to indicate a patient-initiated
spontaneous (flow triggered)
breath is being delivered.
FunctionRange
Not applicable
Not applicable
1.3.3 VENTILATOR STATUS
The VENTILATOR STATUS section of the keyboard shows the operating
condition of the ventilator, and is continuously updated during
operation. Table 1-5 summarizes the functions of the keys and
indicators in the VENTILATOR STATUS section of the keyboard.
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/
indicator
ALARMRed
ColorFunction
Flashes when a high-priority alarm is active.
(high priority)
A repeating sequence of three, then two
beeps sounds. Lights steadily when a highpriority alarm has been autoreset.
CAUTIONYellow
(medium
priority)
NORMALGreenLights when no alarm condition is present.
Flashes when a medium-priority alarm is
active. A repeating sequence of three beeps
sounds. Lights steadily when a mediumpriority alarm has been autoreset.
1-29
Page 46
Introduction
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/
indicator
VENT INOPRed
(high priority)
SAFETY
VALVE
Red
(high priority)
OPEN
ON AC/
GreenLights when the ventilator is running on ac
BATTERY
CHARGING
ColorFunction
Lights to indicate that the ventilator
is inoperative, and the ventilator safety
valve is open. A qualified service technician
must run and pass the extended self-test
(EST) before normal ventilation can resume.
If the condition that caused the safety valve
to open no longer exists, and the VENT
INOP indicator is off, press the alarm reset
key to resume ventilation.
Lights when the ventilator’s safety valve
and exhalation valve open and only room
air is available to the patient.
Can indicate that the ventilator is inoperative,
or there is an occlusion in the ventilator
breathing circuit. If possible, the message
window shows the alarm that triggered the
safety valve open condition and how much
time has elapsed since the last breath was
triggered.
power and the battery is charging.
ON
INTERNAL
YellowFlashes when the ventilator is running on
the internal battery.
BATTERY
ON
EXTERNAL
YellowFlashes when the ventilator is running on
the external battery.
BATTERY
INTERNAL
BATTERY
GreenShows the relative charge level of the
internal battery.
LEVEL
YellowAlarm silence: Silences the alarm sound for
2 minutes from the most recent key press.
7-00423
1-30
Page 47
Introduction
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/
indicator
7-00424
ColorFunction
Not applicableAlarm reset: Reestablishes all alarm
indicators, cancels the alarm silence period,
and resets the patient data displays. If the
condition that caused the alarm still exists,
the alarm reactivates. Cancels apnea
ventilation, if active. Reestablishes previous
settings and ventilation resumes, unless the
ventilator is inoperative.
1-31
Page 48
This page is intentionally blank.
1-32
Page 49
Setting up the ventilator
Warning
This section tells you how to set up the ventilator, including:
• Connecting and using the internal and external batteries
• Connecting the electrical supply
• Connecting the oxygen supply
• Connecting the ventilator breathing circuit
• Installing the collector vial
• Installing the flex arm
• Installing the humidifier
• Using the ventilator cart
• To avoid tipping or damaging the ventilator, do not stack
oth
er equipment on the ventilator. 700 Series Ventilators
are designed to be mounted on either a cart or a shelf by
lified service technician. When lifting the ventilator, li
qua
m the base, and use assistance and appropriate safety
• To avoid the possibility of injury to the patient and to
ensure proper ventilator operation, do not atta
de
sp
• To minimize the increased risk of fire due to an oxygenenriched environment, do not use the ventilat
hype
• To avoid raising the oxygen concentration of room air, use
the ventilator in an adequately ventilated room.
Do not obstruct the cooling fan.
ch any
vice to the port labeled “EXHAUST” unless the device
ecifically authorized by Covidien.
or in a
rbaric chamber.
is
Figure 2-1. Lifting the ventilator
2.1 Connecting and using internal and external batteries
A Covidien battery must always be installed in the ventilator.
Without a battery, the ventilator is not protected against low
or lost ac power. Do not use the ventilator unless a battery
with at least minimal charge is installed.
2-2
Page 51
Setting up the ventilator
7-00054
Internal
battery level
NOTE:
If the ventilator has been stored for an extended period, allow
it to acclimate to its environment before turning it on. This
helps ensure that the ventilator powers up correctly.
Every 700 Series Ventilator System includes an internal battery as a
backup power supply that comes standard with the ventilator.
The internal battery can provide up to 2
½ hours of backup power,
depending on ventilator settings and battery charge level. The
ventilator operates on backup power when ac power is lost or drops
below minimum. The internal battery’s charge level is continuously
indicated on the keyboard (see Figure 2-2). The ventilator alarms
when it determines that only 5 minutes of power remain using the
current settings.
Figure 2-2. Internal battery charge indicator
The optional external battery supplies up to 7 additional hours of
backup power in case ac power is lost or falls below minimum. (See
Appendix B for external battery ordering information.) If installed,
the external battery is the first source of backup power. The ventilator
uses the internal battery if the external battery is depleted or is not
connected.
NOTE:
When the ventilator switches off because battery charge is
inadequate, turn off the power switch to prevent the battery
from being fully depleted. Connect the ventilator to ac power
as soon as possible.
2-3
Page 52
Setting up the ventilator
Caution
The ventilator charges the internal and external batteries during ac
power operation, or in standby mode when attached to ac power.
(See Appendix C for battery specifications.) When the ventilator is
operating on battery, you can use the MENU key to check the
estimated operational time remaining until recharge is required for
the internal and external batteries. (Chapter 6 tells you how to use
the MENU key.) The ventilator must operate for 2 minutes before it
can estimate battery time remaining.
You can disconnect and connect the external battery during normal
operation. When operating the ventilator on battery power, you can
install the external battery on the cart for easy transport. Follow these
steps to connect the external battery:
1.Mount the external battery on the base of the cart (or othe
table location).
sui
r
2.Plug the external battery’s cable into the connector (Figure 2-3).
3.Figure 2-4 shows you how to disconnect the external battery.
To ensure that the batteries retain their charge, store the ventilator
in standby mode, with its power switch on, and connected to ac
power. If the ventilator is not stored in this manner, check battery
charge levels before using the ventilator.
If you plan to store the ventilator for more than 6 months,
remove batteries before storage. Replace the internal battery
before using the ventilator again. A qualified service technician
must replace the battery according to the instructions in the 700 Series Ventilator System Service Manual.
If you turn on the ventilator after it has been unplugged for an
extended period, the LOW EXT BATTERY and LOW INT BATTERY
alarms may become active. If so, recharge the internal battery by
leaving the ventilator plugged in and turned on in standby mode
(with no patient connected) for up to 8 hours.
If the LOW EXT BATTERY, LOW INT BATTERY, or BAT NOT CHARGING
alarm is still active after 8 hours, replace the batteries. If you turn off
the ventilator when the battery is absent or inadequately charged, a
LOSS OF POWER alarm sounds for at least 2 minutes.
2-4
Page 53
Setting up the ventilator
7-00055
External
battery
connector
External
battery
NOTE:
The ON BATTERY indicator flashes when the ventilator is
operating on battery power.
2-5
Figure 2-3. Plugging the external battery into the ventilator
Page 54
Setting up the ventilator
Warning
7-00051
Figure 2-4. Disconnecting the external battery
2.2 Connecting the electrical supply
• To avoid electrical shock hazard, connect the ventilator
power cord into a grounded ac power outlet. If the integrit
of
the ac ground is in question, operate the ventilator fr
e internal or external battery.
th
• If used in the U.S., connect the ventilator to an ac
mar
ked “Hospital Only” or “Hospital Grade” to ensure
grounding reliability.
y
om
receptacle
2-6
Every 700 Series Ventilator System is supplied with a power cord.
A power cord retainer covers the connector and socket to protect
against liquid spills or accidental disconnection.
When the ventilator is ready for operation on ac power, connect the
power cord to ac power (see Figure 2-5).
Page 55
Setting up the ventilator
7-00056
Power cord retainer
Power cord
To ac
power
7-00057
NOTE:
During operation on ac power, the power cord retainer must
always be in place.
Figure 2-5. Connecting the ventilator power cord
When the power cord is not in use, you can wrap the power cord
around the brackets on the back of the ventilator for convenient
storage (see Figure 2-6).
2-7
Figure 2-6. Storing the power cord on the ventilator
Page 56
Setting up the ventilator
Warning
Warning
Caution
2.3 Connecting the oxygen supply
• To ensure proper oxygen concentration, do not obstruct the
vent
ilator’s air intake.
• To ensure adequate oxygen delivery to the patient, us
vidien-supplied oxygen hoses only. Use of other oxygen
Co
hoses could result in inadequate or inappropriate ox
pr
essures or leaks at the oxygen inlet.
• When using a cylinder oxygen supply, point the cylind
pr
essure relief device away from the vent
T
his helps avoid creating an oxygen-rich environmen
ithin the ventilator in the event that the cylind
w
gulator malfunctions.
re
The 700 Series Ventilator System can use oxygen from a cylinder or
wall supply. Follow these steps to connect the oxygen supply:
e
ygen
er's
ilator air intake.
t
er oxygen
1.Ensure that the oxygen supply pressure is between 40 and 90 ps
(
275 and 620 kPa) (50 psi (345 kPa) recommended).
Due to excessive restriction of certain hose assemblies (listed in
Table B-1), reduced FiO
pressures< 50 psi (345 kPa) are used. To maintain correct FiO
levels may result when oxygen inlet
2
2
levels, make sure that oxygen inlet pressure is 50 psi (345 kPa)
when using these hose assemblies.
2.Connect the oxygen supply to the oxygen inlet connector on the
side o
f the ventilator (see Figure 2-7). Make sure ventilator
con
figured with adapter(s) as required.
is
To prevent damage to the ventilator, ensure that the
connection to the oxygen supply is clean and unlubricated.
i
2-8
Page 57
Setting up the ventilator
7-00058
Oxygen inlet
connector
Oxygen hose (to
oxygen supply)
Air intake
NOTE:
Whenever a pressurized oxygen source is connected to the
ventilator, the oxygen regulator has a maximum bleed rate of
3 L/min, even when the ventilator is not in use. Always take this
bleed rate into account when calculating oxygen cylinder volume
and duration. This is especially important when using the
ventilator and cart while transporting patients who are being
ventilated at above-ambient FiO
levels.
2
Figure 2-7. Connecting the oxygen supply
2-9
Page 58
Setting up the ventilator
Warning
Caution
2.4 Connecting the ventilator breathing circuit
• To minimize the risk of bacterial contamination or
comp
onent damage, inspiratory and expiratory filters must
always be handled with care and connected to the
ventilator during use.
• Use of an external pneumatically powered nebulizer with
700 Series V
breathing circuit.
volumes, spirometry and spirometry alarms,
tidal
Fi
O
, and patient triggering. (However, the Covidien EasyNeb
2
Nebu
patient with
data.)
•
To minimize the risk of patient injury, use only ventilato
br
eathing circuits qualified for use in oxygen-enriched
environments with 700 Series Ventilators. Do not us
antistat
tight connection, only use connectors and tubes with ISOstandard cone and socket fittings (or use adapters to
connect barbed cuff fittings to ISO-standard fittings).
• Use ventilator breathing circuits identified by Covidien fo
us
e with 700 Series Ventilators or circuits that ensu
th
e maximum resistance values specified in Table C-5 ar
exceeded.
event the ventilator from ventilating as long as th
pr
vent
ilator passes SST.
entilator results in added flow to the ventil
This additional flow can af
lizer is designed to
deliver aerosol medications to the
out affecting ventilator performance or patien
ic or electrically conductive tubing. To ensure a leak-
Using a circuit with a higher resistance do
a
ator
fect delivered
delivered
t
r
e
r
re that
e not
es not
e
2-10
Adding accessories to the ventilator can increase system
resistance. Ensure that any changes to the recommended
ventilator circuit configurations do not exceed the values for
resistance provided in
Appendix C.
Page 59
Setting up the ventilator
NOTE:
• Covidien recommends that you run SST every 15 days,
between patients, and when you change the ventilator
breathing circuit. Covidien recognizes that the protocol for
running SST varies widely among healthcare institutions.
It is not possible for Covidien to specify or require specific
practices that will meet all needs, or to be responsible for
the effectiveness of those practices.
• The ventilator uses an oxygen sensor to trigger an alarm if
the delivered % O
the % O
setting. You can view the ventilator’s % O
2
is 10 percentage points above or below
2
2
measurement by using the User settings function of the
MENU key.
• Check the inspiratory and expiratory limbs of the ventilator
breathing circuit and the collector vial and in-line water
traps regularly for water buildup. Under certain conditions,
they can fill quickly. In case of a partial occlusion alarm,
check for water in the ventilator breathing circuit. In very
severe cases, such a water occlusion can potentially cause
the ventilator to enter occlusion cycling mode. Empty and
clean the collector vial and in-line water traps as necessary.
• Covidien recommends using pediatric circuits when
ventilating patients with 5-mm or smaller internal-diameter
artificial airways.
Figure 2-8 shows how to connect the ventilator breathing circuit,
including the inspiratory filter, humidifier (if used), inspiratory limb,
patient wye, expiratory limb, collector vial, and expiratory filter.
2-11
Page 60
Setting up the ventilator
7-00059
Ventilator
(to patient)
Inspiratory filter
(arrow on filter
indicates direction
of flow)
Humidifier
Inspiratory limb of the
ventilator breathing
circuit (install EasyNeb
between inspiratory
filter and patient wye)
Patient wye
Expiratory
limb of the
ventilator
breathing
circuit
Collector vial
Expiratory filter
(arrow on filter
indicates direction
of flow)
Ventilator
(from patient)
2-12
Figure 2-8 . Connecting the ventilator breathing circuit
Page 61
2.5 Installing the collector vial
7-00060
Bracket (on
cart or shelfmounting
plate)
Collector
vial
Figure 2-9 shows you how to slide the collector vial into the bracket
on the cart or shelf mounting plate.
Figure 2-9. Installing the collector vial
Setting up the ventilator
2.6 Installing the flex arm
Figure 2-10 shows you how to install the flex arm onto the threaded
socket on the ventilator.
2-13
Page 62
Setting up the ventilator
7-00061
Flex arm
NOTE:
If the flex arm is too large to fit into a tight space, you can
shorten the arm by removing the middle segment (Figure 2-11).
Figure 2-10. Installing the flex arm
2-14
Page 63
Setting up the ventilator
Remove
Remove
Remove
Figure 2-11. Shortening the flex arm
2.7 Installing the humidifier
Figure 2-12 shows you how to slide the humidifier into the bracket
on the cart or shelf mounting plate.
NOTE:
The humidifier must be powered independently of the
ventilator (regardless of whether the ventilator is running on
ac or battery power).
2-15
Page 64
Setting up the ventilator
7-00063
Bracket (on
cart or shelfmounting
plate)
Humidifier
Figure 2-12. Installing the humidifier
2-16
Page 65
2.8 Using the ventilator cart
Caution
Unlocked
Locked
7-00064
Unlocked
Ventilator
cart base
The ventilator cart is equipped with a storage area for the external
battery. You can lock and unlock the cart’s front wheels with your
foot, as shown in Figure 2-13.
To prevent equipment damage, pull, rather than push, the
ventilator over high thresholds when using the cart to transport
the ventilator. Ensure that the flex arm is positioned at the front
of the ventilator.
Setting up the ventilator
Figure 2-13. Locking and unlocking the cart’s front wheels
2-17
Page 66
2-18
This page is intentionally blank.
Page 67
Getting started
This section tells you how to:
• Power up the ventilator
• Select ventilator settings
• View and change alarm limits
• Enter and exit standby mode
NOTE:
Covidien recommends that before using the ventilator for the
first time, you wipe the ventilator exterior clean and sterilize its
components according to the instructions in Appendix A. Follow
your institution’s protocol for cleaning and sterilizing the
ventilator and its components.
3.1 Powering up the ventilator
NOTE:
To ensure battery charge(s), store the ventilator in standby
mode with the power cord plugged into the mains and the
power switch in the on (I) position whenever possible. The
power switch controls ac mains and battery power to the
ventilator.
Page 68
Getting started
Warning
7-00065
1.Turn the power switch on (see Figure 3-1). The ventilator
disp
lays this mess
P
OST running...
age:
NPB 7x0 S/W Rev x
PM due xxxxx hrs
“PM due” refers to the number of hours until a routine
preventive maintenance procedure is due, based on the
ventilator’s operational hours.
If a single beep is not audible during POST, the main audible
alarm may be malfunctioning. Remove the ventilator from use
and contact service.
NOTE:
• If the ventilator is running on ac but the battery is low, u
wer up it may begin ventilating using th
po
ttings and an ABNORMAL RESTART alarm may occur.
se
• Do not switch off ventilator power during POST
e last valid
.
pon
3-
2
Figure 3-1. Turning the power switch on (the “I” position)
Page 69
Getting started
Warning
Caution
2.Once the power-on self-test (POST) is complete (several
seconds), the ventilator displays the message:
Accept settings
to start ventilation
The last valid settings are proposed, and the appropriate
indicator lights to show the ventilator’s source of power.
Always let the ventilator run for 10 minutes before placing it
on a patient to allow heaters to warm up. Failure to warm up
the ventilator could result in failure of the SST flow sensor test,
and higher measured exhaled tidal volumes.
Covidien recommends running short self-test (SST) and testing
alarms to check proper operation of the ventilator before
ventilation begins or according to your institution’s protocol.
Chapter 4 tells you how to run SST. Appendix E tells you how
to test alarms.
NOTE:
If the ventilator detects low oxygen pressure, a low battery
level, or an occluded or missing air intake filter during POST,
POST continues uninterrupted. These alarms become active
when POST is complete, and you can override them by pressing
the alarm reset key.
3-3
Page 70
Getting started
A/C
SIMV
SPONT
7-00066
Press each flashing key
and turn the knob to
adjust.
3.2 Selecting ventilator settings
NOTE:
Each keypress should result in a “blip” sound; otherwise, the
main alarm speaker may be faulty. If a keypress does not result
in this sound, turn the ventilator off then on to run POST and
verify the proper functioning of the speaker.
1.Select the mode: To select a mode, press A/C, SIMV, or SPONT.
(Appendix D gives a detailed explanation of these ventilatio
mod
es.)
In assist/control (A/C) mode, you can only select volume
control ventilation (VCV) or pressure control
ventilation (PCV) settings (PCV is only available on the
760 Ventilator).
In synchronous intermittent mandatory ventilation
(SIMV) mode, you can select both VCV or PCV (760 only)
and pressure support ventilation (PSV) settings.
In spontaneous (SPONT) mode, you can only select PSV
settings.
n
2.Confirm or change settings: For every flashing key light, yo
3-
4
mus
t touch the key (and adjust the setting if necess
can apply the settings. You can confirm or change setting
you
in any orde
r.
Once you press a key, the proposed
setting flashes to indicate that the
knob is linked to the selected
setting. The setting’s LCD window
shows the setting and the message
window displays the current
setting, range, units, name of the
setting, and associated parameter.
Turn the knob to change the
setting or press another key to
leave the setting as is. (The setting
then stops flashing.)
u
ary) before
s
Page 71
Getting started
PEEP/CPAP
TRIGGER
SENSITIVITY
L/min
% O2
ACCEPT
7-00067
3.Press CLEAR to cancel the most recently changed setting and
return to the last valid setting. (This only works if you haven’t
pressed ACCEPT yet.) Pressing CLEAR after you’ve pressed
100% O
stops the 100% O2 maneuver. The ventilator will beep
2
if you attempt to change a setting to a value above or below
the setting range.
4.PEEP/CPAP, TRIGGER SENSITIVITY, and % O
: You
2
can change these settings in any mode. These keys
don’t flash when you select a new mode, and you
are not required to confirm these settings.
NOTE:
The O
% alarm is delayed for 30 seconds and 12 L of tidal
2
volume following a change in setting so that the oxygen
concentration can stabilize.
5.Apply the settings: Press ACCEPT.
NOTE:
If apnea ventilation is possible, the ventilator displays this
message when normal ventilation begins:
Review Ta = xx s
where Ta is the apnea interval and xx is the current setting in
seconds (from 10 to 60 s). Chapter 5 tells you how to adjust
apnea ventilation parameters. You are not required to review
apnea parameters before normal ventilation begins.
3-5
Page 72
Getting started
Warning
7-00066
7-00068
Ventilation begins according to the displayed settings and the
existing alarm limits. You should review all alarm limits and
change them if necessary. (Section 3.3 tells you how.) You can
change alarm limits or ventilator settings at any time during
normal ventilation.
In PSV it is impossible to set support pressure + PEEP to
deliberately cause a HIGH PRESSURE alarm. Although Covidien
does not recommend doing so, it is possible in VCV to set a tidal
volume that causes breath truncation and a HIGH PRESSURE
alarm. To avoid activating a HIGH PRESSURE alarm, select the
appropriate TIDAL VOLUME and HIGH PRESSURE alarm settings.
3.3 Viewing and changing alarm settings
You can view alarm settings at any time without affecting
ventilation by pressing the appropriate key. The ALARM SETTINGS
window displays the value of the alarm being viewed or changed.
The alarm setting flashes for up to 30 seconds, and you can change
an alarm setting any time its setting is flashing in the window.
All alarm settings are in effect during ventilation (except for
LOW INSP PRESSURE, which is not active in SPONT mode). You can
change alarm settings at any time during ventilation. You can only
change one alarm setting at a time.
3-
6
To change a setting: touch, turn, ACCEPT.
1.Touch the key: The key light turns
on, and the current alarm setting
flashes in the window for 30
seconds. The message window
shows the current setting, range,
and units of the alarm.
2.Turn the knob to adjust the alarm
setting (this only works when the
alarm setting is flashing in the
window). The ventilator will beep if
you attempt to select a setting
above or below the setting range.
Page 73
3.Press ACCEPT to apply the new
Warning
ACCEPT
7-00067
alarm setting
se
tting stops flashing a
sage win
mes
sage:
mes
. The new alarm
nd the
dow displays this
Setting(s) accepted
• Inadvertent extubation in which the endotracheal tube
remains
pr
connected to the breathing circuit could result
essurization above the LOW INSP PRESSURE alarm setting.
The LOW INSP PRESSURE and low volume alarms sh
t appropriately.
se
• To ensure patient safety, always set the LOW
VOLUME and
HIGH RATE alarms.
TIDAL
NOTE:
The O
O
and the low O
the set % O
% alarm limits are based on the set % O2. The high
2
% alarm limit is 10 percentage points above the set % O2,
2
% alarm limit is 10 percentage points below
2
. The measured % O2 is shown in the message
2
window if its display is enabled (using the Oxygen sensor
menu function, described in Chapter 6).
Getting started
in
ould be
3.4 Entering and exiting standby mode
Standby mode is a waiting state in which there is no ventilation, but
the ventilator maintains its settings and battery charge(s).
3-7
Page 74
Getting started
Warning
• To avoid raising the oxygen concentration of room air,
ensure that the ventilator is in an adequately ventilat
ro
om if connected to an oxygen supply while in stan
e. Disconnect the oxygen supply if you do not intend to
mod
use the ventilator immediately.
• To avoid patient injury due to lack of ventilatory support, do
not enter standby mode with a patient atta
ilator. You must confirm that no patient is attached
vent
ched to the
before entering standby mode.
NOTE:
To maintain battery charge(s), the ventilator must be plugged
into ac power with the power switch on while in standby
mode.
Follow these steps to enter standby mode:
ed
dby
3-
1.Turn on the ventilator. If ventilation has already begun,
mus
t turn off the ventilator, then turn it back on withou
st
arting ventilation. This allows you access to the Standby mode
you
t
menu function.
2.Press MENU.
3.Turn the knob to select standby mode.
4.Press ACCEPT. The ventilator displays this message:
Is pt disconnected?
ACCEPT to proceed
5.Confirm that no patient is attached by pressing ACCEPT. (If you
don’t press ACCEPT while this message is displayed, th
vent
ilator will not enter standby mode.)
8
e
Page 75
Getting started
6.This message tells you that the ventilator is in standby mode:
In standby mode
Clear to exit
The ON AC/BATTERY CHARGING, INTERNAL BATTERY LEVEL,
and SAFETY VALVE OPEN indicators turn on. All other key lights
and displays are off during standby mode. The only alarms that
can become active during standby mode are FAN FAILED ALERT,
BAT NOT CHARGING, LOSS AC POWER, and LOSS OF POWER.
Follow these steps to exit from standby mode and resume ventilation:
1.Press CLEAR.
2.The ventilator displays this message:
POST running...
3.After a few seconds, POST is complete and the ventilator begins
ventilation using the last valid settings.
3-9
Page 76
3-10
This page is intentionally blank.
Page 77
Self tests (SST and EST)
Warning
All 700 Series Ventilators include these built-in self tests:
• Short self test (SST) is a short (about 2 to 3 minutes) and
sequen
the ventilator
complia
• Extended self test (EST) is a thorough test of the operationa
integ
pne
• Before running SST or EST, you must disconnect the
• A fault identified in SST or EST indicates that the ventil
ce of tests that verifies proper ventilator op
breathing circuit for leaks, and measures the circui
nce. Power on self test (POST) is part of SST.
rity of the ventilator, both the electronics and
umatics. POST and SST are part of EST.
ventilator from the patient. Running SST or EST while the
ventilator is connected to the patient can injure the patient.
an associated com
or associated component should be repaired before th
ventilator is returned to service, unless it can be determined
with ce
patient,
hazards.
rtainty that the defect cannot create a hazard for
or add to the risks which ma
ponent is defective. A defective ventilator
y arise from other
eration, checks
simple
the
ator or
e
the
t
l
Page 78
Self tests (SST and EST)
Caution
To ensure accurate SST and EST operation, run SST and EST in
room air (% O
on for at least 10 minutes. (EST also requires an oxygen
supply.)
Table 4-1 describes SST and EST.
Test namePurposeWhen test is run
is 21%) after the ventilator has been powered
2
Table 4-1: 700 Series Ventilator self tests
Short self
test (SST)
Extended
self test
(EST)
An abbreviated version of EST to
be run primarily by the operator.
Makes detailed checks of the
pneumatics and electronics. It also
characterizes system leaks and
system/tubing compliance to
compensate during breath
delivery. POST is run as part of SST.
Thoroughly tests the operational
integrity of the ventilator, both
the electronics and pneumatics.
POST and SST are part of EST.
Before patient is
connected to the
ventilator or after
ventilator breathing
circuit or humidifier is
changed.
• When the ventilator
is serviced.
• As part of the
ventilator
performance
verification.
• Following a ventilator
inoperative.
NOTE:
The performance verification, described in the 700
Series Ventilator System Service Manual, is a more
thorough test of the ventilator to verify specifications
are met. The technician runs a partial or full
performance verification after servicing the ventilator
and at regular intervals.
4-
2
Page 79
4.1 Short self test (SST)
Table 4-2 summarizes the functions of SST.
NOTE:
• Covidien recommends that you run SST every 15 da
tween patients, and when you change the ventilat
be
eathing circuit. Covidien recognizes that the pr
br
ru
nning SST varies widely among healthcare institutions
It is
not possible for Covidien to specify or require specif
pr
actices that will meet all needs, or to be responsible fo
e effectiveness of those practices.
th
• When SST asks you to do something (for example, press
ACCEPT or block the patient wye), the ventilator will wa
indefinit
Follow these steps to run SST:
ely for your response.
Self tests (SST and EST)
ys,
or
otocol for
.
ic
r
it
1.Turn on the ventilator. If ventilation has already begun,
must
turn off the ventilator, then turn it back on withou
st
arting ventilation. This allows you access to the Self test menu
function.
If you select the Self test function during ventilation th
ow displays this message:
wind
N/A in ventilation
2.Press MENU, turn the knob to select Self test, then press ACCEPT.
The usual 30-second timeout for your input is disabled during
the entry to the Self test function.
3.The window displays this message:
Short self test
Turn knob to view
4.Turn the knob to select Short self test (you
Extended self test). The
Short self test
Vent warming 09:59
CLEAR to bypass
window displays this message:
can also select
you
t
e
4-3
Page 80
Self tests (SST and EST)
Warning
If the ventilator has already been running for more than
10 minutes before you select Short self test, this message is
bypassed (see step 5). If the ventilator was recently running and
is already warmed up, you can press CLEAR to override the
Vent warming message and begin SST. Covidien cannot
guarantee the accuracy of test results in this case, however.
5.The window displays this message:
6.Press ACCEPT to confirm that no patient is connected to the
ventilator. (Or press CLEAR to back up through the menu.)
The window displays this message:
7.Block the patient wye and then press ACCEPT.
8.The ventilator runs power-on self-test (POST) and displays this
message for a few seconds:
Is pt disconnected?
ACCEPT to proceed
Block wye
POST running...
4-
4
If a single beep is not audible during POST, the main audible
alarm may be malfunctioning. Remove the ventilator from use
and contact service.
NOTE:
• If the ventilator is running on ac but the battery is low, upon
power up it may begin ventilating using the last valid settings
and an ABNORMAL RESTART alarm may occur.
• During POST it is recommended that you not switch off
ventilator power.
9.The message window then displays this message:
humidification device type
Choose humidifier
where the humidification device type is the most recently
selected type, as follows:
• Dual heated wire: Humidifier with heated wire on expiratory
limb, or on both inspiratory and expiratory limbs
• No heated wire: Conventional humidifier without heated
wire circuit on the expiratory limb
Turn the knob to select the humidifier type attached, then
press ACCEPT.
Incorrectly specifying the humidifier type during SST can affect
the accuracy of spirometry calculations.
10. The message window then displays this message:
tubing type
Choose tubing type
where the tubing type is the most recently selected ventilator
breathing circuit type.
Turn the knob to select the ventilator breathing circuit type
(adult or pediatric tubing) attached to the ventilator, then
press ACCEPT.
Incorrectly specifying the ventilator breathing circuit type
during SST can cause an inappropriate sensitivity for the leak
test and occlusion alarm. Covidien recommends using pediatric
circuits when ventilating patients with 5-mm or smaller
internal-diameter artificial airways.
NOTE:
To ensure that compliance compensation functions correctly,
you must run SST with the circuit configured as intended for
use on the patient.
4-5
Page 82
Self tests (SST and EST)
Warning
11. The message window then displays this message:
Turn the knob to select the appropriate ET (endotracheal tube)
size, then press ACCEPT.
Specifying an ET tube size that is too large can cause premature
termination of breaths on very small pediatric patients. During
normal ventilation, you can use the MENU key’s User settings
function to change the ET tube size (see Chapter 6).
12. The ventilator automatically starts the test sequence. A message
indicat
cannot be interrupted. In most cases you don’t need
a
required. When a test passes, SST automatically starts the next
test.
NOTE:
The PRESSURE bar graph and the PRESSURE display window
show the real-time expiratory pressure throughout SST.
ET size: x.x mm
Choose ET size
es which test is being performed. Once started, a test
to do
nything unless a test result is fault or fail, or your action is
4-
6
Page 83
These keys are valid during SST:
ACCEPT
CLEAR
MANUAL
INSP
To confirm that a requested action is complete.
To repeat a test, return to the prompt at the start of a test,
or indicate an inaudible alarm (during the Main alarm
speaker test and Backup alarm test only).
To override an Incomplete or Fault test result.
To restart from the beginning of SST.
7-00424
To stop testing and skip to the end of SST.
Self tests (SST and EST)
7-00423
7-00072
NOTE:
If the tubing or humidifier type or ET tube size that
you’ve selected has changed from the previous time
SST was run, this key is temporarily disabled until the
Circuit comp test is complete.
4-7
Page 84
Self tests (SST and EST)
Display:
Disconnect I tubing
Reconnect I tubing
(Do not disconnect inspiratory filter.)
7-00069
Display:
Unblock wye
Block wye
7-00070
Display:
Disconnect E tubing
Reconnect E tubing
(Do not disconnect expiratory filter.)
7-00071
13. The I/E filter test prompts you to complete these actions (press
ACCEPT when each action is complete):
4-
8
Page 85
Self tests (SST and EST)
14. When the final test in the sequence is complete, this message is
displayed along with the overall SST result:
SST finished testing
Table 4-3 summarizes overall SST results and how to proceed in
each case.
15. Unblock the patient wye, then press ACCEPT. The ventilator
reruns POST.
Table 4-2: SST sequence of tests
TestFunctionComments
Is pt
disconnected?
Leak testChecks the ventilator
P
e Pcyl
compare
Auto-zero sol
test
Circuit comp
test
Asks for confirmation that a
patient is not connected to
the ventilator.
breathing circuit for leaks.
Displays expiratory pressure
on pressure bar graph.
Verifies proper function of
pressure transducers in
internal pneumatics and
ventilator breathing circuit.
Displays expiratory pressure
on pressure bar graph.
Verifies the proper function
(ability to open and close)
of the P
P
cyl
Determines the compliance
of the ventilator breathing
circuit. Displays expiratory
pressure on pressure bar
graph.
(expiratory) and
e
(cylinder) solenoids.
SST cannot proceed until
you confirm that a patient is
not connected.
Fault if not passed.
Overriding a fault could
cause improper compliance
compensation, inaccurate
tidal volume delivery,
or autocycling.
Fault if not passed.
Overriding a fault could
cause improper compliance
compensation,
or autocycling.
Failure if not passed.
Fault if not passed.
Overriding a fault could
cause improper compliance
compensation or inaccurate
tidal volume delivery.
4-9
Page 86
Self tests (SST and EST)
TestFunctionComments
Table 4-2: SST sequence of tests (continued)
Safety valve
test
Verifies that safety valve
relieves excess circuit
pressure. Displays
expiratory pressure on
pressure bar graph.
PEEP system
test
Verifies that the PEEP
system can generate and
maintain preset PEEP levels.
Displays expiratory pressure
on pressure bar graph.
Flow/O
2
sensor test
Checks the accuracy of the
exhalation flow sensor and
the oxygen sensor.
I/E filter testChecks the pressure drop
across the inspiratory and
expiratory limbs of the
entire patient system.
Checks the pressure drop
across the filters.
Failure if not passed.
Failure if not passed.
Failure if flow sensor test
does not pass.
Fault if oxygen sensor test
does not pass. Use the
MENU key’s Oxygen sensor
function to recalibrate the
oxygen sensor (see
Chapter 6).
Overriding a fault (oxygen
sensor test only) could result
in inaccurate % O
2
measurement.
Fault if not passed. This test
is optional: pressing CLEAR
at the initial prompt skips
this test.
Skipping this test or
overriding a fault could
result in inadequate
bacteria protection or
excessive resistance to
inspiration or exhalation.
4-
10
Page 87
Self tests (SST and EST)
Table 4-2: SST sequence of tests (continued)
TestFunctionComments
Heaters testVerifies correct operation of
the heaters in the
exhalation subsystem.
Main alarm
test
Backup alarm
test
Sounds alarm at three
volumes and verifies that
user reports hearing it.
Checks the operation of the
backup alarm circuit.
NOTE:
Faults detected during SST may require removing
the ventilator from clinical use immediately and
servicing it. A trained operator, however, must
evaluate the situation and determine whether the
ventilator can be used. Failures detected during SST
require rerunning SST (using the alarm reset key).
It is not possible to exit SST if SST fails. If SST
continues to fail, the ventilator must be removed
from service.
Fault if not passed.
Overriding a fault could
cause inaccurate monitoring
of exhaled volume or lead to
damage to the ventilator’s
flow sensor or exhalation
valve.
Failure if not passed.
If alarm is audible press
ACCEPT. If not, press CLEAR.
Fault if not passed. If alarm
is audible press ACCEPT.
If not, press CLEAR.
Overriding a fault could
cause no backup audible
alarm in case the main
audible alarm fails.
4-11
Page 88
Self tests (SST and EST)
Table 4-3: Overall SST results
If the SST
result is:
It means:Do this:
SST passedAll tests passed.Press ACCEPT to exit SST
or press the alarm reset
key to restart SST from
the beginning.
SST
incomplete
SST faultOne or more tests did not
All tests performed were
passed, but some were
skipped. The skipped tests
were passed on a previous
run.
pass. These tests might not
compromise the ventilator’s
ability to ventilate safely,
based on the operator’s
Press the alarm reset key
to restart SST from the
beginning or press
MANUAL INSP to override
this result and exit SST.*
Press the alarm reset key
to restart SST from the
beginning or press
MANUAL INSP to override
this result and exit SST.*
evaluation of the situation.
SST failedOne or more critical
problems were detected.
Press the alarm reset key to
rerun SST. It is not possible
to exit SST if SST fails. If SST
continues to fail, remove
the ventilator from use and
contact service.
4-
* If you press MANUAL INSP to exit SST, you will be prompted to confirm that you
wish to override the fault or incomplete test. Confirm by pressing MANUAL INSP
again.
12
Page 89
4.2 Extended self test (EST)
Warning
To run EST, you must have the hardware listed in Table 4-4.
Table 4-4: EST hardware requirements
Self tests (SST and EST)
Hardware
description
Ventilator
breathing circuit
Manufacturer/model or Covidien part number
To use as test circuit: G-061208-00 or equivalent (adult,
reusable, without heated wire)
NOTE:
To ensure that compliance compensation
functions correctly, the user must run EST or SST
with the circuit configured as intended for use
on the patient.
Stopper, wye
(no. 2)
Stopper,
inspiration port
(no. 3)
Oxygen source,
40 to 90 psi
(275 to 620 kPa)
Due to excessive restriction of Air Liquide™*, Australian, and
Dräger™* hose assemblies, reduced FiO2 levels can result if you use
oxygen inlet pressures < 50 psi (345 kPa). Make sure oxygen inlet
pressure is 50 psi (345 kPa) when using these hose assemblies, to
maintain correct FiO2 levels.
G-061574-00 or local supplier
G-061575-00 or local supplier
Local supplier
4-13
Page 90
Self tests (SST and EST)
NOTE:
Covidien recommends that you always run the full EST before
placing the ventilator into operation following service and as
part of the ventilator’s routine performance verification. For
more information on performance verification, see the
700 Series Ventilator System Service Manual.
Follow these steps to run EST:
1.Set up the ventilator as for normal operation, co
hu
midifier, if applicable, and leak-tight ventilator breathing
mplete with
circuit.
2.Turn on the ventilator. If ventilation has already begun,
mus
t turn off the ventilator, then turn it back on withou
st
arting ventilation. This allows you access to the Self test menu
you
t
function.
3.Press MENU, turn the knob to select Self test, then press ACCEPT.
4.Turn the knob to select Extended self test. The window displays
the messages shown in Table 4-5.
Table 4-5: EST setup messages
MessageDescription
Extended self
test Vent
warming xx:xx
CLEAR to bypass
This message is displayed (followed by a countdown) if
you just turned on the ventilator. The ventilator must
be on for 10 minutes before running EST so that
component temperatures can stabilize.
If the ventilator was recently running and is already
warmed up, you can press CLEAR to override the Vent warming message and begin SST. Covidien cannot
guarantee the accuracy of test results in this case,
however.
4-
14
Is pt
disconnected?
Block wyeInstall no. 2 stopper and then press ACCEPT.
Confirm that patient is disconnected by pressing
ACCEPT; or press CLEAR to return to service menu.
Page 91
Self tests (SST and EST)
Warning
Table 4-5: EST setup messages (continued)
MessageDescription
POST running...Wait a few seconds until POST is completed.
NOTE:
• If a single beep is not audible during POST,
the main audible alarm may be
malfunctioning. Contact service.
• During POST do not switch off ventilator
power.
humidification
device type
Choose
humidifier
Turn knob to select desired humidification device,
then press ACCEPT. Device types include: HME (heatmoisture exchanger or “artificial nose”), Dual heated
wire (humidifier with heated wires on both inspiratory
and expiratory limbs), or No heated wire (humidifier
without a heated wire on expiratory limb).
Incorrectly specifying the humidifier type during
EST can affect the accuracy of spirometry
calculations.
4-15
Page 92
Self tests (SST and EST)
Warning
Warning
MessageDescription
Table 4-5: EST setup messages (continued)
tubing type
Choose tubing
type
ET size: x.x mm
Choose ET size
Turn knob to select either Adult tubing or Pediatric
tubing and then press ACCEPT.
Incorrectly specifying the ventilator breathing
circuit type during EST can cause an
inappropriate sensitivity for the leak test and
occlusion alarm. Covidien recommends using
pediatric circuits when ventilating patients with
5-mm or smaller internal-diameter artificial
airways.
NOTE:
To ensure that compliance compensation
functions correctly, run EST with the circuit
configured as intended for use on the patient.
Turn knob to select appropriate ET (endotracheal
tube) size, and then press ACCEPT.
Specifying an ET tube size that is too large can
cause premature termination of breaths on very
small pediatric patients. During normal
ventilation, you can use the MENU key’s User settings function to change the ET tube size (see
Chapter 6).
4-
16
Page 93
Self tests (SST and EST)
5.The ventilator automatically starts the test sequence (Table 4-6).
Unless a test requires your intervention (and waits indefinitely
for your response), you don’t need to do anything until a test
result is fault or failed, or EST is complete. Use the keys listed in
Table 4-7 as needed.
Table 4-6: EST test sequence
Test n u m b e r
and name
1. DAC-ADC
loop test
2. Safety valve
test
3. Motor
sensor test
4. Zeroing of
P
o
solenoids
5. O
2
test
FunctionComments
A loopback test to check
digital-to-analog (D/A) and
analog-to-digital (A/D)
converters’ operation.
Verifies that safety valve
relieves excess circuit
pressure.
Checks operation of rotary
encoder.
Zeroes oxygen regulator
pressure transducer.
Disconnect and reconnect
oxygen supply, as directed.
Press ACCEPT to signify you
have performed each
requested action.
Checks oxygen solenoids
and oxygen sensor.
Failure if not passed. Not in
SST.
Failure if not passed. Also in
SST.
Failure if not passed. Not in
SST.
Fault if not passed. Not in
SST.
NOTE:
Overriding this fault
could result in
inaccurate oxygen
percentage delivery.
Fault if not passed. Not in
SST.
NOTE:
Overriding this fault
could result in
inaccurate oxygen
percentage delivery.
6. Flow sensor
test
4-17
Checks accuracy of
exhalation flow sensor.
Failure if not passed. Also in
SST.
Page 94
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
7. Leak testChecks ventilator breathing
circuit for leaks.
7. P
ePcyl
compare
(also part
of test 7)
Verifies proper functioning
of pressure transducers and
ventilator breathing circuit.
FunctionComments
Fault if not passed. Also in
SST.
NOTE:
Overriding this fault
could cause improper
compliance
compensation,
inaccurate tidal volume
delivery, or autocycling.
Fault if not passed. Also in
SST.
NOTE:
Overriding this fault
could cause improper
compliance
compensation,
inaccurate tidal volume
delivery, or autocycling.
Overriding this fault
could cause improper
compliance
compensation,
inaccurate tidal volume
delivery, or autocycling.
Fault if not passed. Also in
SST.
Page 95
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
9. I/E filter test Checks pressure drop across
inspiratory and expiratory
limbs of entire patient
system. Checks pressure
drop across filters.
Disconnect and reconnect
inspiration (I) tube from/to
inspiratory filter outlet, and
exhalation (E) tube from/to
expiratory filter inlet, as
directed. Block and unblock
wye with a no. 2 stopper, as
directed. Press ACCEPT to
signify you have performed
each requested action.
10. PEEP
system test
Verifies that PEEP system can
generate and maintain
preset PEEP levels within
either of two sets of limits. If
results are within the inner
set of limits, test passes the
first time. If the results are
outside of the outer set of
limits, test fails the first
time. If the results are
outside the inner limits but
within the outer limits, the
ventilator adjusts its
calibration table in
nonvolatile RAM (NVRAM)
and repeats the test to verify
the calibration. The test fails
if the calibration cannot be
verified after five attempts.
FunctionComments
Fault if not passed. Also in
SST. Mandatory in EST, but
optional in SST.
NOTE:
Overriding this fault
could result in
inadequate bacteria
protection or excessive
resistance to inspiration
or exhalation.
Failure if not passed. Also in
SST, but not identical.
11. Check
valve test
4-19
Checks check valves on inlet
and outlet ports of cylinder
and on exhalation limb of
ventilator breathing circuit.
Failure if not passed. Not in
SST.
Page 96
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
12. Piston leak
test
13. Lightsdisplays
test
14. Keys testVerifies that every row and
Checks piston leak against a
table of leak values stored
in NVRAM.
First remove inspiratory filter;
then block and unblock
inspiratory filter port as
directed, using no. 3 stopper.
Press ACCEPT to signify you
have performed each
requested action. Replace
inspiratory tubing at end of
test.
Verifies that lights and
displays function.
Observe each named display,
and press ACCEPT to verify
that it is lit. Press CLEAR if a
display is not lit.
column of key matrix can be
correctly read.
Press each named key (each
key’s LED will also be lit).
Press CLEAR if a key’s LED is
not lit.
FunctionComments
Failure if not passed. Not in
SST.
Failure if not passed. Not in
SST.
Failure if not passed. Not in
SST.
4-
20
15. Main alarm
test
Checks whether various
alarm sounds are audible (at
maximum volume).
Listen for each named
sound, then press ACCEPT to
verify that it is audible. Press
CLEAR if a sound is not
audible.
Failure if not passed. Also in
SST.
Page 97
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
16. Backup
alarm test
17. Fan testVerifies that main fan is
18. Heaters
test
Checks operation of backup
alarm circuit.
Press ACCEPT if alarm is
audible or press CLEAR if
alarm is not audible.
operating correctly by
reading fan status bits.
Verifies correct operation of
heaters in exhalation
system.
FunctionComments
Fault if not passed. Also in
SST.
NOTE:
Overriding this fault
could result in no
audible alarm if the
main speaker fails.
Failure if not passed. Not in
SST.
Fault if not passed. Also in
SST.
NOTE:
Overriding this fault
could result in
inaccurate monitoring
of exhaled volume or
damage to the
ventilator’s flow sensor
or exhalation valve.
At the end of each test, the test name and pass/fail/fault status
are displayed. When a test passes, EST automatically starts the
next test. You do not need to do anything unless a test result is
Fault or Fail, or your action is required. If a test does not pass,
you can still continue EST, although the overall EST result is not
passed until the error condition is corrected. Since information
on multiple errors can help troubleshooting, it can be useful to
complete EST even with errors.
4-21
Page 98
Self tests (SST and EST)
ACCEPT
CLEAR
MANUAL
INSP
Table 4-7: Key functions during EST
KeyFunction
To confirm that a requested action is complete.
Repeat a test or return to prompt at start of a test.
To override an Incomplete or Fault test result.
To restart from the beginning of SST.
7-00424
To stop testing and skip to the end of SST.
4-
22
7-00423
7-00072
NOTE:
If the tubing or humidifier type or ET tube size
that you’ve selected has changed from the
previous time SST was run, this key is temporarily
disabled until the Circuit comp test is complete.
6.Some tests display additional prompts, some of which require
your
response. These are listed in Table 4-8.
Table 4-8: Prompts during EST
PromptYour response
Disconnect O
supply
Reconnect O
supply
2
Disconnect ventilator from oxygen supply.
Press ACCEPT.
2
Reconnect ventilator to oxygen supply. Press ACCEPT.
Page 99
Self tests (SST and EST)
Table 4-8: Prompts during EST (continued)
PromptYour response
Disconnect I tubing Disconnect tubing from inspiratory filter outlet.
Leave inspiratory filter in place. Press ACCEPT.
I drop = x x.x at 100
or
I drop = x x.x at 40
Reconnect I tubing
This is the inspiratory filter pressure drop in cmH
O at
2
a flow of 100 L/min (adult tubing) or 40 L/min
(pediatric tubing). Reconnect tubing to inspiratory
filter outlet. Press ACCEPT.
Unblock wyeRemove stopper from patient wye. Press ACCEPT.
Block wyeInsert a stopper into patient wye. Press ACCEPT.
Disconnect E tubing Reconnect tubing to inspiratory filter outlet, and
disconnect tubing from expiratory filter inlet. Leave
expiratory filter in place. Press ACCEPT.
E drop = x.x at 100
or
E drop = x x.x at 40
Reconnect E tubing
This is the expiratory filter pressure drop in cmH
a flow of 100 L/min (adult tubing) or 40 L/min
(pediatric tubing). Press ACCEPT. Reconnect tubing to
expiratory filter outlet.
Reconnect I tubingRemove stopper from inspiratory filter outlet.
Reinstall inspiratory filter. Press ACCEPT.
Name of a section
of display
Upper MDW =
blocks?
Press ACCEPT to confirm that all LEDs/LCDs in named
section are on.
Press ACCEPT to confirm that all pixels are turned on
in upper half.
Lower MDW =
blocks?
Press ACCEPT to confirm that all pixels are turned on
in lower half.
Name of a keyPress named key.
High alarm soundConfirm that you hear alarm by pressing ACCEPT.
Medium alarm
Confirm that you hear alarm by pressing ACCEPT.
sound
4-23
Page 100
Self tests (SST and EST)
No soundConfirm that you do not hear alarm by pressing
Table 4-8: Prompts during EST (continued)
PromptYour response
ACCEPT.
Is backup alarm
ON?
Confirm that you hear alarm by pressing ACCEPT.
7.When the last test in the sequence is complete, the EST
comp
letion status is displayed (Table 4-9). Respond as indicate
If cert
ain tests do not pass, safe ventilation
romised, and the ventilator is rendered inoperative
comp
paired. This is called a ventilator failure. If certain othe
is re
tests do no
(subje
t pass, but ventilation might not be compromise
ct to a trained operator’s evaluation), this is called a
may be
ventilator fault.
NOTE:
Faults detected during EST may not require the ventilator to
be serviced or removed from use immediately. A trained
operator, however, must evaluate the situation and determine
whether the ventilator can be used. Failures detected during
EST require immediate servicing and removal of the ventilator
from clinical use.
d.
until it
r
d
4-
24
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.