Medtronic 700Series_Operators_Manual_EN_10066984E00 COULTER HmX Hematology Analyzer with Autoloader

Page 1
Operator’s Manual
Puritan Bennett
TM

700 Series Ventilator System

Page 2
To obtain information about a warranty, if any, contact Covidien Technical Services at
1.800.635.5267 or your local representative. Purchase of this instrument confers no express or implied license under any Covidien patent
to use the instrument with any Ventilator System that is not manufactured or licensed by Covidien.
Page 3
Copyright Information
Copyright 2011 Covidien. COVIDIEN, COVIDIEN with logo. The Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. U.S. Patents 5,540,222; 5,596,984; 5,632,270; 5,664,560; 5,673,689; 5,909,731; 5,915,382; 5,934,274; 6,116,240; 6,142,150; 6,321,748; 6,467,478.
The information contained in this manual is the sole property of Covidien and may not be duplicated without permission. This manual may be revised or replaced by Covidien at any time and without notice. You should ensure that you have the most current applicable version of this manual; if in doubt, contact the Technical Publications Department of Covidien. While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals. Covidien’s sole responsibility with respect to the ventilator, and its use, is as stated in the limited warranty provided.
Nothing in this manual shall limit or restrict in any way Covidien’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Covidien has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.
Page 4
Preface
Warning
Caution
Warning
Definitions
This manual uses these special indicators to convey information of a specific nature:
Indicates a condition that can endanger the patient or the ventilator operator.
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular emphasis that make operation of the ventilator more efficient or convenient.
Warnings, cautions, and notes
Please take the time to familiarize yourself with the following safety considerations, special handling requirements, and regulations that govern the use of the 700 Series Ventilator System.
ii
To avoid an electrical shock hazard while servicing the ventilator, be sure to remove all power to the ventilator by disconnecting the power source and turning off all ventilator power switches.
Page 5
Preface
Warning
Warning
Warning
Warning
To avoid a fire hazard, keep matches, lighted cigarettes, and all other sources of ignition (e.g., flammable anesthetics and/or heaters) away from the ventilator and oxygen hoses.
Do not use oxygen hoses that are worn, frayed, or contaminated by combustible materials such as grease or oils. (Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.)
In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen supply and electrical power source.
Patients on life-support equipment should be appropriately monitored by competent medical personnel and suitable monitoring devices.
The 700 Series Ventilator is not intended to be a comprehensive monitoring device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment.
Check the ventilator periodically as outlined in the service manual; do not use if defective. Immediately replace parts that are broken, missing, obviously worn, distorted, or contaminated.
An alternative source of ventilation should always be available when using the 700 Series Ventilator System.
iii
Page 6
Preface
Warning
Warning
Warning
Caution
To ensure proper servicing and avoid the possibility of physical injury, only qualified personnel should attempt to service or make authorized modifications to the ventilator.
The user of this product shall have sole responsibility for any ventilator malfunction due to operation or maintenance performed by anyone not trained by Covidien staff.
For a thorough understanding of ventilator operations, be sure to read the 700 Series Ventilator System Operator's Manual in its entirety before attempting to use the system.
Before activating any part of the ventilator, be sure to check the equipment for proper operation and, if appropriate, run the self-diagnostic short self test (SST) program described in this manual.
iv
U.S. Federal law restricts this device to sale by or on the order of a physician.
Warranty
The 700 Series Ventilator System is warranted against defects in material and workmanship in accordance with Covidien Medical Equipment Warranty for a period of one year from the time of sale. To ensure the validity of the warranty, be sure to keep a maintenance record.
Year of manufacture
The 700 Series Ventilator System’s year of manufacture is indicated by the fifth and sixth digits of the serial number which is located at the lower edge of the ventilator front panel.
Page 7
Preface
Manufacturer: Covidien llc 15 Hampshire Street Mansfield, MA 02048 USA
Authorized Representative: Covidien Ireland Limited IDA Business & Technology Park Tull amore
Manufacturer
Electromagnetic susceptibility
The 700 Series Ventilator System complies with the requirements of IEC 60601-1-2 (EMC Collateral Standard), which includes E-field susceptibility and ESD requirements. However, even though the device is compliant at the levels of immunity specified in the standard, certain transmitting devices (cellular phones, walkie­talkies, cordless phones, paging transmitters, etc.) emit radio frequencies that could interrupt ventilator operation if located in a range too close to the ventilator. It is difficult to determine when the field strength of these devices becomes excessive. Practitioners should be aware that radio frequency emissions are additive, and that the ventilator must be located a sufficient distance from transmitting devices to avoid interruption. Do not operate the ventilator in a magnetic resonance imaging (MRI) environment. The Alarm handling section of this manual describes possible ventilator alarms and what to do if they occur. Consult with your institution’s biomedical engineering department in case of interrupted ventilator operation, and before relocating any life support equipment.
v
Customer assistance
For further assistance contact your local Covidien representative.
Page 8
This page is intentionally blank.
vi
Page 9

Table of Contents

Table of Contents 2
1 Introduction
1.1 Functional description............................................................ 1-3
1.2 Symbols and labels ................................................................. 1-8
1.3 Keyboard............................................................................... 1-13
1.3.1 VENTILATOR SETTINGS ............................................. 1-14
1.3.2 PATIENT DATA ......................................................... 1-24
1.3.3 VENTILATOR STATUS ............................................... 1-29
2 Setting up the ventilator
2.1 Connecting and using internal and external batteries ........ 2-2
2.2 Connecting the electrical supply ........................................... 2-6
2.3 Connecting the oxygen supply .............................................. 2-8
2.4 Connecting the ventilator breathing circuit ....................... 2-10
2.5 Installing the collector vial................................................... 2-13
2.6 Installing the flex arm .......................................................... 2-13
2.7 Installing the humidifier ...................................................... 2-15
2.8 Using the ventilator cart ...................................................... 2-17
3 Getting started
3.1 Powering up the ventilator.................................................... 3-1
3.2 Selecting ventilator settings .................................................. 3-4
3.3 Viewing and changing alarm settings................................... 3-6
3.4 Entering and exiting standby mode...................................... 3-7
4 Self tests (SST and EST)
4.1 Short self test (SST) ................................................................. 4-3
4.2 Extended self test (EST) ........................................................ 4-13
5 Once ventilation begins
5.1 Changing settings: a quick review......................................... 5-1
5.1.1 Changing settings ....................................................... 5-1
5.1.2 Switching between VCV, PCV, and PSV ..................... 5-2
5.1.3 Changing the mode .................................................... 5-3
5.2 Viewing and changing alarm settings: a quick review ........ 5-3
5.3 Adjusting apnea parameters ................................................. 5-4
5.3.1 Adjusting the apnea interval ..................................... 5-5
vii
Page 10
Table of Contents
5.4 Viewing patient data.............................................................. 5-6
5.5 The 100% O
5.6 The EXP PAUSE and INSP PAUSE keys (760 only) ................ 5-10
6 The MENU key
6.1 More active alarms.................................................................. 6-6
6.2 Autoreset alarms..................................................................... 6-6
6.3 Self tests................................................................................... 6-7
6.4 User settings ........................................................................... 6-8
6.4.1 Endotracheal tube....................................................... 6-8
6.4.2 Humidifier type ........................................................... 6-8
6.4.3 Date and time set ........................................................ 6-9
6.4.4 Apnea interval (Ta) ................................................... 6-10
6.4.5 VCV flow pattern ...................................................... 6-10
6.4.6 Speaking valve setup ................................................ 6-11
6.4.7 Alarm volume ............................................................ 6-17
6.4.8 PCV timing setting .................................................... 6-17
6.4.9 Volume LED bar......................................................... 6-17
6.5 Oxygen sensor ....................................................................... 6-18
6.6 Standby mode ....................................................................... 6-20
6.7 Battery info .......................................................................... 6-20
6.8 Software revision .................................................................. 6-21
6.9 Service summary.................................................................... 6-21
6.10 Nebulizer ............................................................................... 6-21
and MANUAL INSP keys................................... 5-9
2
7 Alarm handling
7.1 Autoreset alarms..................................................................... 7-2
7.2 Alarm silence ........................................................................... 7-3
7.3 Alarm reset .............................................................................. 7-4
7.4 Clinical and technical alarms .................................................. 7-4
7.5 Power alarm .......................................................................... 7-20
7.5.1 Loss of AC Power ...................................................... 7-20
7.5.2 Loss of Power............................................................. 7-20
A Maintenance
A.1 Cleaning, disinfection, and sterilization ...............................A-2
A.1.1 Cleaning: general guidelines ......................................A-4
A.1.2 Disinfection and sterilization......................................A-5
viii
Page 11
A.2 Preventive maintenance ........................................................A-6
A.2.1 Daily or as required: inspiratory and expiratory bacteria
A.2.2 Daily or as required: collector vial ............................. A-9
A.2.3 Daily or as required: in-line water traps.................. A-10
A.2.4 As necessary: oxygen sensor calibration.................. A-10
A.2.5 Every 250 hours (or 1 month of use): cooling fan filter ... A-11 A.2.6 Every 1000 hours (or 3 months of use): air intake filter ... A-12
A.2.7 Every 2 years: device checks ..................................... A-13
A.2.8 Storage ......................................................................A-13
A.2.9 Repacking ..................................................................A-14
B Part numbers
C Specifications
C.1 Physical ................................................................................... C-2
C.2 Environmental ....................................................................... C-3
C.3 Power .................................................................................... C-5
C.4 Compliance and approvals .................................................... C-7
C.5 Technical ................................................................................ C-8
Table of Contents
filters .......................................................................... A-8
D Breath delivery
D.1 A/C mode ................................................................................ D-3
D.2 SPONT mode ..........................................................................D-4
D.3 SIMV mode .............................................................................D-5
D.3.1 Breath timing ..............................................................D-5
E Alarm testing
F Pneumatic schematic
GGlossary
ix
Page 12
This page is intentionally blank.
x
Page 13

List of Tables

Table 1-1. Mode/breath type availability on 740/760 Ventilators.................... 1-2
Table 1-2. Changes to current settings in occlusion cycling mode................. 1-7
Table 1-3. 700 Series Ventilator keyboards: VENTILATOR SETTINGS ...... 1-15
Table 1-4. 700 Series Ventilator keyboards: PATIENT DATA ...................... 1-24
Table 1-5. 700 Series Ventilator keyboards: VENTILATOR STATUS .......... 1-29
Table 4-1. 700 Series Ventilator self tests ...................................................... 4-2
Table 4-2. SST sequence of tests .................................................................. 4-9
Table 4-3. Overall SST results ..................................................................... 4-12
Table 4-4. EST hardware requirements ....................................................... 4-13
Table 4-5. EST setup messages .................................................................. 4-14
Table 4-6. EST test sequence ...................................................................... 4-17
Table 4-7. Key functions during EST ........................................................... 4-22
Table 4-8. Prompts during EST .................................................................... 4-22
Table 4-9. EST completion status ............................................................... 4-25
Table 5-1. Breath type availability .................................................................. 5-2
Table 6-1. Menu function summary ................................................................ 6-3
Table 7-1. Clinical alarms ............................................................................... 7-5
Table 7-2. Technical alarms ........................................................................ 7-14
Table A-1. Cleaning, disinfection, and sterilization ......................................... A-3
Table A-2. Preventive maintenance schedule ............................................... A-7
Table B-1. Ventilator accessories ................................................................... B-2
Table C-1. Physical specifications .................................................................. C-2
Table C-2. Environmental specifications......................................................... C-3
Table C-3. Power specifications...................................................................... C-5
Table C-4. Compliance and approvals............................................................ C-7
Table C-5. Technical specifications ................................................................ C-8
xi
Page 14
This page is intentionally blank.
xii
Page 15

List of Figures

Figure 1-2. 740 Ventilator System keyboard................................................... 1-13
Figure 1-3. 760 Ventilator System keyboard................................................... 1-14
Figure 2-1. Lifting the ventilator......................................................................... 2-2
Figure 2-2. Internal battery charge indicator ..................................................... 2-3
Figure 2-3. Plugging the external battery into the ventilator ............................. 2-5
Figure 2-4. Disconnecting the external battery ................................................. 2-6
Figure 2-5. Connecting the ventilator power cord ............................................. 2-7
Figure 2-6. Storing the power cord on the ventilator ......................................... 2-7
Figure 2-7. Connecting the oxygen supply........................................................ 2-9
Figure 2-9. Installing the collector vial ............................................................ 2-13
Figure 2-10. Installing the flex arm.................................................................... 2-14
Figure 2-11. Shortening the flex arm................................................................. 2-15
Figure 2-12. Installing the humidifier ................................................................ 2-16
Figure 2-13. Locking and unlocking the cart’s front wheels .............................. 2-17
Figure 3-1. Turning the power switch on (the “I” position)................................. 3-2
Figure 5-1. Viewing patient data ....................................................................... 5-7
Figure 5-2. Volume bar graph (760 Ventilator only) .......................................... 5-8
Figure 6-1. Using the Figure 6-2. Using the
Figure 7-1. Viewing active alarms ..................................................................... 7-2
More active alarms Autoreset alarms
menu function to view autoreset alarms .... 6-7
menu function to view active alarms ....... 6-6
Page 16
This page is intentionally blank.
xiv
Page 17

Introduction

The 700 Series Ventilator System (including the 740 and 760 Ventilators) provides respiratory support for a wide range of pediatric to adult patients for a wide variety of clinical conditions. The ventilator’s mixing technique allows it to ventilate critically ill patients at adjustable oxygen concentrations without the need for a blender, compressor, or hospital-grade wall air.
The 700 Series Ventilator System can be mains- or battery-powered. Each ventilator includes two microcontrollers: one for breath delivery (which controls ventilation), and one for the user interface (which monitors ventilator and patient data). Each microcontroller verifies that the other is functioning properly. Using two independent microcontrollers in this fashion prevents a single fault from causing a simultaneous failure of controlling and monitoring functions.
The 700 Series Ventilator System supplies mandatory or spontaneous breaths with a piston-based pneumatic system. Table 1-1 summarizes the modes and breath types offered by the 740 and 760 Ventilators. Mandatory breaths can be volume control ventilation (VCV, available on 740 and 760 Ventilators) or pressure control ventilation (PCV, available on the 760 Ventilator only). VCV delivers breaths to the patient at a preset tidal volume, peak flow, waveform, and oxygen concentration at a minimum respiratory rate. PCV delivers breaths to
Page 18
Introduction
the patient at a preset inspiratory pressure, I:E ratio or inspiratory time, rise time factor (how quickly inspiratory pressure rises to achieve the set inspiratory pressure), and oxygen concentration at a minimum respiratory rate. A spontaneous breath allows the patient inspiratory flows of up to 300 L/min, with or without pressure support ventilation (PSV). On the 760 Ventilator, you can set the rise time factor and exhalation flow sensitivity (that is, the point at which the ventilator cycles from inspiration to exhalation) in PSV.
The ventilator begins apnea ventilation if no breath (patient- , ventilator-, or operator-initiated) is delivered within the selected apnea interval. Apnea ventilation is active during all modes. On the 740 Ventilator, only VCV breaths are available in apnea ventilation. On the 760 Ventilator, VCV or PCV breaths are available in apnea ventilation.
The 760 Ventilator also offers the ability to perform respiratory mechanics calculations and maneuvers as a standard feature using the EXP PAUSE (to calculate auto-PEEP) and INSP PAUSE (to calculate patient resistance and compliance) keys.
Table 1-1: Mode/breath type availability on 740/760 Ventilators
Mode/breath type 740 Ventilator 760 Ventilator
VCV breath type
PCV breath type
PSV breath type (support pressure setting)
PSV (rise time factor and exhalation sensitivity settings)
SIMV mode
Apnea ventilation (VCV breath type)
Apnea ventilation (choice of VCV or PCV breath type)
Respiratory mechanics (EXP PAUSE and INSP PAUSE)
1-2
••
Page 19
This manual tells you how to operate and perform simple maintenance for the 700 Series Ventilator. Covidien recommends that you become familiar with this manual and accompanying labels before attempting to operate or maintain the ventilator. If you need additional copies of this manual, contact your Covidien representative.
To ensure optimum performance of the 700 Series Ventilator System, Covidien recommends that a qualified service technician perform periodic maintenance on the ventilator. For more information, contact your Covidien representative.

1.1 Functional description

By pressing keys and turning the knob on the ventilator keyboard, the operator gives initial instructions and data to the ventilator (Figure 1-1). The user interface microcontroller processes this information and stores it in the ventilator’s memory. The breath delivery microcontroller uses this stored information to control and monitor the flow of gas to and from the patient.
The 700 Series Ventilator uses a flow trigger to recognize patient effort. The trigger monitors flow from the piston during exhalation. When the patient inhales, patient circuit pressure drops very slightly below end-expiratory pressure. At the same time, the piston moves forward to deliver flow to the ventilator breathing circuit and maintain the preset PEEP/CPAP level. The level of flow depends on the patient’s effort. If this flow exceeds the user-set level, the ventilator triggers. By design, the ventilator attempts to maintain PEEP in the presence of a circuit leak. Since a leak drives the piston to deliver flow to make up for pressure losses, a circuit leak can require an increase in the flow trigger level to avoid autocycling.
Introduction
During exhalation, the ventilator’s piston retracts and draws air and oxygen into the cylinder. The ventilator uses room air, which means the ventilator can operate without a compressor or wall air source. Room air enters the ventilator through a protected user-replaceable air intake filter just inside the ventilator cabinet. This filter captures airborne particles.
1-3
Page 20
Introduction
Oxygen from a cylinder or wall supply enters the ventilator through a hose and oxygen fitting (the fitting is available in several versions). Once inside the ventilator, the oxygen is regulated to a pressure the ventilator can use, then mixed with air, according to the selected %O
.
2
The flow-triggered piston/cylinder system and motor controller circuit control the flow of gas to the patient. On the 760 Ventilator in PCV or PSV, the rate of flow is also determined by the preset rise time factor. This system is designed with a minute gap (about the size of a thin sheet of paper) between the piston and the cylinder wall. This design eliminates the friction between the piston and cylinder, allowing it to respond more rapidly than a “sealed” system.
A small amount of gas leaks through the gap between the piston and cylinder. Ventilator software and a continuous forward motion of the piston compensate for this leak.
The piston delivers the mixed air and oxygen through the inspiratory manifold system, and out to the patient. The oxygen concentration and temperature of the delivered gas are monitored here, using a galvanic oxygen sensor and a thermistor. The galvanic sensor generates a voltage proportional to the partial pressure of oxygen, from which the oxygen concentration is calculated. The ventilator alarms if the monitored oxygen concentration is more than ten percentage points above or below the % O
setting. The inspiratory
2
manifold system also includes a safety valve to relieve patient pressure if necessary (for example, if the ventilator breathing circuit is kinked or occluded).
The patient system includes the components external to the ventilator that route gas between the ventilator and the patient. These components include the inspiratory filter (which protects against contamination between the ventilator and patient), a humidification device, ventilator breathing circuit (the tubing through which the gas travels), collector vial (which protects the exhalation system from moisture in the exhaled gas, and can be emptied without losing circuit PEEP), and an expiratory filter (which limits the bacteria in the patient’s exhaled gas from escaping to room air or contaminating the ventilator).
1-4
Page 21
Introduction
-
+
740 Ventilator
Filter
Keyboard
Regulator
Oxygen
Room air
Patient
7-00017
Inspiration manifold
Gas intake system
Inspiratory filter
Humidification device
Ventilator breathing circuit (inspiratory limb)
Piston/cylinder
system
Exhalation/
PEEP/CPAP
system
Expiratory
filter
Collector
vial
Ventilator breathing circuit
(expiratory limb)
Figure 1-1 . Block diagram: 700 Series Ventilator function
The heated exhalation system monitors the flow of the patient’s exhaled gas using a differential pressure transducer. The patient exhales through the exhalation valve. During exhalation, the PEEP/CPAP system maintains user-selected pressure in the ventilator breathing circuit.
1-5
Page 22
Introduction
Throughout the respiratory cycle, pressure transducers monitor inspiratory, expiratory, and atmospheric pressures. The temperatures of the pneumatic compartment and inspiratory gas are also monitored. Information from these transducers is continuously used to update the calculations that control ventilation. (Appendix F provides a diagram of the ventilator’s pneumatic system and ventilator breathing circuit.)
Power to operate the ventilator comes from ac mains (wall) or battery power. The power supply is designed to protect against excessive voltages, temperatures, or current draws. A power cord retainer prevents the cord from accidental disconnection.
The ventilator includes an internal battery, and accommodates an optional external battery. Depending on the ventilator settings, battery backup power can be supplied for up to 2½
hours using the
internal battery, and up to 7 hours using the external battery. Both batteries are recharged during operation from ac power. If both are installed, the external battery is used first when ac power is not present. If the external battery is depleted or not installed, the internal battery supplies power to the ventilator when ac power is not available. The keyboard indicates the source of power and battery charge level of the internal battery at all times.
Emergency modes: The ventilator declares a ventilator inoperative (VENT INOP) condition if a hardware failure or critical software error
that could compromise safe ventilation occurs. In case of a ventilator inoperative condition, the VENT INOP indicator lights and the ventilator enters the safety valve open (SVO) state. To correct a ventilator inoperative condition, the ventilator must be turned off, then powered on again; at power-on, the operator must run extended self-test (EST). The ventilator must pass EST before normal ventilation can resume.
The safety valve allows the patient to breathe room air unassisted when the ventilator is in the SVO state. The ventilator remains in the SVO state until power-on self-test (POST) verifies that power levels to the ventilator are acceptable and that the motor controller and microcontrollers are functioning correctly, and until the user has confirmed ventilator settings.
1-6
Page 23
Introduction
If the ventilator enters the SVO state and POST is not running, the SAFETY VALVE OPEN indicator lights and a high-priority alarm sounds. The ventilator enters the SVO state if it detects a hardware or software failure that could compromise safe ventilation. In case of a malfunction that prevents software from opening the safety valve, there is also an analog circuit that opens the safety valve when system pressure exceeds 115 cmH
O (113 hPa).
2
If the ventilator detects an occlusion or a continuous high inspiratory pressure condition, it opens the safety and exhalation valves to vent excess pressure, then shuts them and begins occlusion cycling mode. In occlusion cycling mode the ventilator uses current settings except for those summarized in Table 1-2. If the ventilator again detects an occlusion or continuous high pressure condition, it again opens the safety and exhalation valves then resumes occlusion cycling mode. If the operator presses the alarm reset key or the ventilator does not detect an occlusion or continuous high pressure condition, it reverts to normal ventilation using the most recently accepted settings.
Table 1-2: Changes to current settings in occlusion cycling mode
Setting Change to setting
HIGH PRESSURE alarm (VCV breath type)
PEEP (all modes) Set to 0 cmH
% O
(all modes) Set to 100%
2
SPONT mode (PSV breath type)
SUPPORT PRESSURE (PSV breath type)
SUPPORT PRESSURE (PCV breath type)
RISE TIME FACTOR (PCV breath type)
Set to 30 cmH
Breaths are delivered at a rate of 12/min with an inspiratory time of 2 seconds.
If less than 15 cmH If 15 cmH
(no change).
If less than 15 cmH If 15 cmH
(no change).
Set to 70%
O or above: the current setting is used
2
O or above: the current setting is used
2
O
2
O
2
O: set to 15 cmH2O.
2
O: set to 15 cmH2O.
2
1-7
Page 24
Introduction
SN

1.2 Symbols and labels

These symbols and labels appear on the 700 Series Ventilator System:
Power switch positions per IEC 601-1.”I” represents ON position; “O” represents OFF position.
7-00421
Refer to manual per IEC 601-1. When this symbol appears on product, it means “Refer to documentation for
7-00418
7-00416
7-00426
7-00414
information.”
Potential equalization point, per IEC 601-1
External battery connection
Circuit breaker
Serial number
ac current
7-00427
Type B equipment, per IEC 601-1
7-00415
Indicates the degree of protection provided by enclosure (drip-proof).
7-00403
Signifies compliance with the Medical Device Directive, 93/ 42/EEC
1-8
Page 25
CSA and NRTL (Nationally Recognized Testing Laboratory)
EXHAUST
TO
PATIENT
FROM
PATIENT
certification, granted by CSA
8-00417
The TUV Rheinland logo signifies TUV Rheinland Type Test approval to Annex III of the Medical Device Directive
7-00420
Exhaust port connector
Inspiratory limb connector
Expiratory limb connector
Introduction
1-9
Page 26
Introduction
61006
Oxygen inlet port label
O
2
V
.
ma
Air intake label
61010
Cooling fan label
61009
General life support equipment warning label
61011
1-10
Page 27
740 Ventilator back panel label
Introduction
1-11
Page 28
Introduction
760 Ventilator back panel label
1-12
Page 29

1.3 Keyboard

Caution
Patient data Ventilator status
Ventilator settings
7-00048
To avoid damaging the keyboard, do not press on it with sharp objects.
The keyboard (Figure 1-2 shows the 740 keyboard and Figure 1-3 shows the 760 keyboard) is grouped into three sections:
• VENTILATOR SETTINGS: Where you set breath delivery variables.
Introduction
• PATIENT DATA: Where you set alarm limits and view mon
itored pressures, breath timing, and volumes.
• VENTILATOR STATUS: Where you see the alarm status and operating condition of the ventilator.
Figure 1-2. 740 Ventilator System keyboard
the
1-13
Page 30
Introduction
Patient data Ventilator status
Ventilator settings
7-00123
Figure 1-3. 760 Ventilator System keyboard

1.3.1 VENTILATOR SETTINGS

The VENTILATOR SETTINGS section of the keyboard allows you to select the ventilation mode, breath type, and settings. For more detail on ventilation modes and breath delivery, see Appendix D.
To change the mode and settings, select the mode, then the breath type, and then the ventilator settings. The keys flash during setup and mode changes to ensure that you review all pertinent settings. The keyboard is designed to minimize accidental or unintentional changes.
Table 1-3 summarizes the functions of the keys, knob, and indicators in the VENTILATOR SETTINGS section of the keyboard. Ventilator settings are also limited by these breath delivery boundaries:
• I:E ratio 4:1 for PCV (760 Ventilator only), 3:1 for all other breath types
• Inspiratory time = 0.2 to 8 seconds (excluding plateau)
• Expiratory time 0.2 seconds
1-14
Page 31
Introduction
• PEEP/CPAP + SUPPORT PRESSURE or INSPIRATORY PRESSURE 80 cmH
O (80 hPa)
2
NOTE:
Maximum SUPPORT PRESSURE is 70 cmH INSPIRATORY PRESSURE is 80 cmH
2
O.
O, maximum
2
• SUPPORT PRESSURE or INSPIRATORY PRESSURE + PEEP/CPAP < HIGH PRESSURE - 2 cmH
O (2 hPa)
2
• HIGH PRESSURE (in A/C and SIMV modes) > PEEP/CPAP + 7cmH
O (7 hPa)
2
• HIGH PRESSURE (in SPONT mode) > PEEP/CPAP + SUPPORT PRESSURE + 2 cmH
O(2hPa)
2
• HIGH PRESSURE > LOW INSP PRESSURE
• Minute volume 50 L/min at an I:E ratio of 2:1
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Mode/breath type settings
A/C Assist/control mode VCV (volume control
ventilation) and PCV (pressure control ventilation) breath types.
(PCV available on 760 Ventilator only.)
SIMV Synchronous intermittent
mandatory ventilation mode
SPONT Spontaneous mode PSV breath type
VCV VCV breath type VCV available on 740 and
VCV, PCV (760 only), and PSV (pressure support ventilation) breath types.
760 Ventilators in A/C or SIMV modes.
1-15
Page 32
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Mode/breath type settings (continued)
PCV PCV breath type PCV available on 760
Ventilator only in A/C or SIMV modes.
PSV PSV breath type PSV available in SIMV or
SPONT modes.
Mandatory (VCV) settings
RESPIRATORY RATE
TIDAL VOLUME
PEAK FLOW Maximum flow of gas
PLATEAU (s) Length of inspiratory pause
Mandatory (PCV) settings (760 Ventilator only)
RESPIRATORY RATE
The minimum number of mandatory breaths the patient receives per minute. During apnea ventilation the minimum RESPIRATORY RATE setting is 6 /minute.
Volume delivered to the patient during a mandatory breath, compliance­compensated and corrected to body temperature and pressure, saturated (BTPS).
delivered during a mandatory breath (BTPS). (Combined with tidal volume, peak flow defines the active portion of inspiratory time.)
after a mandatory breath has been delivered, during which no gas is delivered.
The minimum number of mandatory breaths the patient receives per minute. During apnea ventilation the minimum RESPIRATORY RATE setting is 6 /minute.
1 to 70 /minute Accuracy:
(0.1 + 1%) /minute
40 to 2000 ml Accuracy:
(10 ml + 10% of setting)
3 to 150 L/min Accuracy:
(5 + 10% of setting) L/min
0.0 to 2.0 second Accuracy: 0.05 second
1 to 70 /minute Accuracy:
(0.1 + 1%) /minute
1-16
Page 33
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Mandatory (PCV) settings (760 Ventilator only) (continued)
INSPIRATORY PRESSURE (760 only)
/I:E RATIO
T
I
(760 only)
RISE TIME FACTOR (760 only)
Pressure above PEEP during the inspiratory phase of a PCV breath.
You can use the MENU key to select inspiratory time (TI) or I:E ratio as the breath timing setting for a PCV breath. You can change the selected breath timing setting (T
or I:E
I
ratio), but the setting remains constant when you change the respiratory rate in PCV.
Selecting I:E ratio makes the set ratio of inspiratory time to expiratory time for a PCV breath.
The time for inspiratory pressure to rise from 0 to 95% of the target pressure level during a PCV breath.
A setting of 100 = a 100-ms rise time, and a setting of 5 = 80% of the inspiratory time or 2500 ms, whichever is less.
When you adjust this setting, the message window shows the actual time (in seconds) to reach 95% of target pressure.
5 to 80 cmH
O (5 to 80 hPa)
2
Accuracy: (3 + 2.5% of setting) cmH
Inspiratory time (T
):
I
0.2 to 8 seconds Accuracy: 0.05 second
I:E ratio: ≤ 4:1 Accuracy: (0.1 + 2%)
5 to 100
O
2
NOTE:
To help determine the correct setting for inspiratory time, during PCV the message window displays peak inspiratory flow, end inspiratory flow, and end exhalation flow in L/min.
1-17
Page 34
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Spontaneous (PSV) settings
SUPPORT PRESSURE
RISE TIME FACTOR (760 only)
Pressure above PEEP maintained during spontaneous inspiration. Support pressure is terminated when inspiratory flow falls to 25% of peak inspiratory flow, or to the exhalation sensitivity setting (760 only), or 10 L/min or 25% of peak flow, whichever is lower (740 only). Maximum inspiratory time is 3.5 seconds for adults, and 2.5 seconds for pediatric patients.
The time for inspiratory pressure to rise from 0 to 95% of the target pressure level during a PSV breath.
A setting of 100 = a 100-ms rise time, and a setting of 5 = 80% of the inspiratory time or 1500 ms (when adult ventilator breathing circuit is selected) or 600 ms (when pediatric circuit is selected), whichever is less.
When you adjust this setting, the message window shows the actual time (in seconds) to reach 95% of target pressure.
0 to 70 cmH
O (0 to 70 hPa)
2
Accuracy: (3 + 2.5% of setting) cmH
5 to 100
O
2
1-18
Page 35
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Spontaneous (PSV) settings (continued)
EXH SENSITIVITY (760 only)
The percent of peak expiratory flow at which the ventilator cycles from inspiration to exhalation for spontaneous breaths. The flow at which the ventilator cycles from inspiration to exhalation for PSV breaths. Exhalation begins when the inspiratory flow is less than the set value.
To help set EXH SENSITIVITY appropriately, the peak inspiratory flow and end inspiratory flow are displayed in the message window in PSV. (To ensure accurate estimates of these flows, it is important to run SST so that tubing compliance calculations are correct.)
Common settings
PEEP/CPAP Positive end expiratory
pressure/continuous positive airway pressure. Minimum pressure maintained during inspiratory and expiratory phases.
TRIGGER SENSITIVITY (L/min)
Inspiratory flow required to trigger the ventilator to deliver a breath.
1 to 80%
0 to 35 cmH Accuracy:
(2 cmH
O + 4% of setting)
2
1 to 20 L/min
O (0 to 35 hPa)
2
% O
2
Percentage of inspired oxygen of the gas delivered to the
21 to 100% Accuracy: 3% full scale
patient.
NOTE:
It may take several minutes for the oxygen percentage to stabilize.
1-19
Page 36
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Other keys, knobs, and indicators
APNEA PARAMS key
Allows you to select VCV or PCV (760 Ventilator only) apnea ventilation, and the apnea ventilation settings. The apnea interval is adjustable from 10 to 60 seconds. Apnea ventilation is available in all modes. For apnea parameter ranges, see mandatory VCV settings and mandatory PCV (760 Ventilator only) settings.
MENU Allows you to view active and reset alarms, run SST and EST,
adjust certain settings (including alarm volume, PCV timing, volume bar graph display, endotracheal tube size, and date and time), access oxygen sensor functions (calibrate, enable or disable % O
alarm limits, and enable or disable display of
2
oxygen sensor reading), enter standby mode, view battery information, display software revision, display service information, and enter EasyNeb nebulizer functions. (Chapter 6 tells you how to use the menu function.)
100% O
Switches the % O2 to 100% for 2 minutes, then returns to
2
the current % O2 setting. The 2-minute interval restarts every time you press 100% O
. Once the 100% O2 has
2
started, you can press CLEAR to stop the maneuver (unless you have entered a MENU key function or selected a setting).
MANUAL INSP
Delivers one mandatory breath to the patient according to the current mandatory settings (in A/C or SIMV) or the current apnea parameters (in SPONT). You can deliver a MANUAL INSP at any time during the exhalation phase of a breath as long as the exhaled flow is less than 30% of the peak exhaled flow, except during apnea ventilation.
1-20
Page 37
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Other keys, knobs, and indicators (continued)
EXP PAUSE (760 only)
Allows you to measure the patient’s auto-PEEP. An EXP PAUSE maneuver causes the ventilator to close the exhalation valve at the end of the expiratory phase, and not deliver the next mandatory breath. At the end of the maneuver, the message window shows the calculated value for auto-PEEP (expiratory pressure at the beginning of the maneuver minus expiratory pressure at the end of the maneuver) and total PEEP for 30 seconds.
The message window shows the end expiratory flow in L/min at the beginning of each breath. If there is expiratory flow when the ventilator delivers the next breath, auto-PEEP is present.
The EXP PAUSE continues as long as you hold down the key, and should last only until expiratory pressure stabilizes. An EXP PAUSE maneuver ends when you release the key, the patient initiates a breath, an alarm occurs, the expiratory phase (including the maneuver) lasts more than 20 seconds, or the ventilator detects a leak.
Auto-PEEP: Range: 1 to 35 cmH Accuracy: (1 cmH
O.
2
O + 3% of reading).
2
The EXP PAUSE maneuver is unavailable when the RESPIRATORY RATE setting is less than 3 /minute.
1-21
Page 38
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Other keys, knobs, and indicators (continued)
INSP PAUSE (760 only)
Allows you to measure the patient’s compliance and resistance. An extended inspiratory pause also allows you to expand the patient’s lungs for up to 10 seconds.
Pressing INSP PAUSE momentarily causes the ventilator to wait until the end of the inspiratory phase of the current or next mandatory breath (in SPONT mode, the ventilator delivers a mandatory breath using the MANUAL INSP key according to the apnea settings), stop breath delivery, and keep the exhalation valve closed. The INSP PAUSE continues until the ventilator detects a stable plateau pressure or 2 seconds have elapsed. An INSP PAUSE maneuver ends when a stable plateau is reached or an alarm occurs.
Pressing INSP PAUSE for 2 or more seconds after the pause begins causes the ventilator to deliver an inspiratory pause for as long as you hold down the key. An extended INSP PAUSE maneuver ends when you release the key or 10 seconds have elapsed.
You can press CLEAR or release the INSP PAUSE key at any time to cancel an INSP PAUSE maneuver, and alarms cancel the maneuver.
At the end of the breath, the message window shows the calculated value for compliance and resistance (if the mandatory breath was a VCV breath) or compliance (if the mandatory breath was a PCV breath) for 30 seconds.
Compliance: Range: 1 to 150 mL/H Accuracy: (1 mL/cmH
O.
2
O + 20% of reading).
2
Resistance: Range: 0 to 150 cmH2O/L/second. Accuracy: (3 cmH
O/L/second + 20% of reading).
2
CLEAR Pressing CLEAR before accepting a setting cancels the
proposed setting. Pressing CLEAR does not cancel accepted settings.
Pressing CLEAR twice returns the ventilator to its previous state (unless you have entered a MENU key function that requires you to press CLEAR more times).
Pressing CLEAR during a 100% O
maneuver cancels the
2
maneuver.
1-22
Page 39
Introduction
Table 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS
Key/indicator Specifies... Range
Other keys, knobs, and indicators (continued)
ACCEPT Makes changes to settings effective. If you don’t press
ACCEPT within 30 seconds of proposing a new setting, the user interface returns to its previous state.
Knob Adjusts the value of a setting or selects a menu option.
A setting value that flashes means that the knob is linked to that setting. Turning the knob clockwise increases the value, and turning the knob counterclockwise decreases the value.
CURRENT Lights when the ventilator is operating according to the
displayed settings, or during apnea ventilation. (There is one indicator for mandatory breaths, and one for spontaneous breaths.)
PROPOSED Lights when you propose a mode or breath type, or you are
setting apnea parameters. Once a proposed setting is accepted, it becomes effective at the next breath.
APNEA PARAMS indicator
Message window
Lights when apnea ventilation is active. Lights with PROPOSED indicator when you are setting apnea
parameters, and both indicators turn off once apnea parameters are accepted.
Shows up to four lines of information (20 characters per line).
First line: Reserved for the highest-priority active or autoreset alarm. On the 760 Ventilator only, if no alarm is active and the display of the oxygen sensor reading is enabled, the % O
is displayed here. Otherwise, if VCV is the
2
current or proposed breath type and there are no active or autoreset alarms, shows the selected VCV flow pattern (ramp or square).
Second line: Information about the menu function or settings, alarm silence time remaining, or current date and time. During normal ventilation, shows “Flow (L/min).”
Third and fourth lines: Reserved for other messages. For every breath type, peak and end inspiratory flows are displayed on the third line, and end expiratory flow is displayed on the fourth line (except that inspiratory flow is not displayed during VCV breaths or VCV apnea ventilation).
1-23
Page 40
Introduction

1.3.2 PATIENT DATA

The PATIENT DATA section of the keyboard allows you to view the pressure, breath timing, and volume of the patient’s breath. You can also view the alarm settings. A lighted key indicates that a measurement is selected, and its value appears in the display window. Values are continuously displayed and updated during ventilation.
Table 1-4 summarizes the functions of the keys and indicators in the PATIENT DATA section of the keyboard.
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Pressure
MEAN PRESSURE
PEAK PRESSURE
PLATEAU PRESSURE
(760 only)
PEEP/CPAP (760 only)
Function Range
Shows the calculated value of ventilator breathing circuit pressure over an entire respiratory cycle. Updated at the beginning of each breath.
Shows the maximum pressure measured during inspiration. Updated at the beginning of each expiratory phase. (Default pressure display.)
Shows the pressure measured at the end of the plateau period of a mandatory inspiration (whether the inspiration is in a regular VCV breath or is part of an inspiratory pause maneuver). Updated at the beginning of each expiratory phase. The PRESSURE display shows a blank if the ventilator does not detect a stable plateau pressure.
Shows the pressure measured at the expiratory limb before any inspiratory effort. Updated at the beginning of each inspiratory phase.
0 to 99 cmH2O (0 to 99 hPa)
Accuracy: (3 + 4% of reading) cmH
0 to 140 cmH
O
2
(0 to 140 hPa) Accuracy:
(3 + 4% of reading) cmH
0 to 140 cmH
O
2
(0 to 140 hPa) Accuracy:
(3 + 4% of reading) cmH
0 to 140 cmH2O (0 to 140 hPa)
Accuracy: (3 + 4% of reading) cmH
O
2
O
2
O
2
O
2
1-24
Page 41
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Breath timing
RATE (/min)
I:E RATIO Shows the ratio of measured
INSP TIME(s) (760 only)
Shows the calculated value of the total respiratory rate, based on the previous 60 seconds or 8 breaths (whichever interval is shorter). Updated at the beginning of each breath. (Default breath timing display.)
The calculation is reset (and display is blank) when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode, breath type, or RESPIRATORY RATE setting, and when you press the alarm reset key.
inspiratory time to measured expiratory time. Updated at the beginning of each breath.
The measured inspiratory time, including breaths that are truncated due to a HIGH PRESSURE alarm. Updated at the beginning of each expiratory phase.
Function Range
1 to 199 /minute Accuracy:
(0.1 +1% of reading)/minute
1:99.9 to 9.9:1 Accuracy:
(0.1 + 2%)
0.1 to 9.90 seconds Accuracy:
0.05 seconds
1-25
Page 42
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Volume
EXHALED VOLUME (ml)
TOTAL MINUTE VOLUME (L)
DELIVERED VOLUME (ml) (760 only)
Function Range
Shows the patient’s measured expiratory tidal volume averaged over the last 5 breaths (for A/C VCV breaths, ventilator­initiated PCV breaths, and PCV apnea breaths) or for the just­completed breath (for all other breaths). Corrected to BTPS and compliance-compensated. Updated at the beginning of each inspiration. (Default volume display.)
The calculation is reset when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode or breath type, and when you press the alarm reset key.
Shows the patient’s measured expiratory minute volume, based on the previous 60 seconds or 8 breaths (whichever interval is shorter). Updated at the beginning of each breath.
The calculation is reset when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode or breath type, and when you press the alarm reset key.
Shows the measured inspiratory tidal volume for the just­completed PCV or PSV breath. Corrected to BTPS and compliance-compensated. Updated at the beginning of each inspiration for PCV and PSV breath types.
0 to 9 L Accuracy:
(10 ml + 10% of reading)
0 to 99 L Accuracy:
(10 ml + 10% of reading)
0 to 3000 ml Accuracy:
(10 ml + 10% of reading)
1-26
Page 43
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Function Range
Volume (continued)
SPONT MINUTE VOLUME (L) (760 only)
Shows the patient’s measured expiratory minute volume for all spontaneous breaths, based on the previous 60 seconds or 8 breaths (whichever interval is shorter). Updated at the beginning of each breath.
The calculation is reset when ventilation starts, when apnea ventilation starts or autoresets, when you change the mode or breath type, and when you press the alarm reset key.
Alarm settings
HIGH RATE An active alarm indicates that
measured respiratory rate is higher than the alarm setting.
HIGH TIDAL VOLUME
An active alarm indicates that exhaled volume for three out of four consecutive breaths was above the alarm setting.
LOW INSP PRESSURE
An active alarm indicates that monitored circuit pressure is below the alarm setting at the end of inspiration. Inactive in for any spontaneous breath.
LOW TIDAL VOLUME
An active alarm indicates that delivered volume for three out of four consecutive breaths were below the alarm setting. (If this alarm is set to 0 ml and breath type is PCV, an active alarm indicates that delivered volume is less than 3 ml for three out of four consecutive breaths.)
0 to 99 L Accuracy:
(10 ml + 10% of reading)
3 to 100 /minute Accuracy:
(0.1 +1% of setting)/minute
20 to 6000 ml Accuracy:
(10 ml + 10% of setting)
3 to 60 cmH
O (3 to 60 hPa)
2
Accuracy: (1 + 3% of setting)
0 to 2000 ml Accuracy:
(10 ml + 10% of setting)
1-27
Page 44
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Function Range
Alarm settings (continued)
HIGH PRESSURE
An active alarm indicates that two consecutive breaths were truncated because circuit pressure reached the alarm setting.
LOW MINUTE VOLUME
An active alarm indicates that monitored minute volume is less than the alarm setting, based on an eight-breath running average or the previous minute, whichever is less.
Other indicators
Pressure bar graph
Shows real-time pressures in centimeters of water (cmH2O) or hectopascals (hPa).
LEDs show the current HIGH PRESSURE alarm setting and the peak pressure of the last breath during exhalation.
Volume bar graph (760 only)
Shows real-time exhaled volume in milliliters (ml). Volumes are compliance-compensated and corrected to BTPS.
The active scale is determined by the HIGH TIDAL VOLUME alarm setting. LEDs show the current HIGH TIDAL VOLUME and LOW TIDAL VOLUME alarm settings. You can use the MENU key to enable or disable the volume bar graph. During exhalation, LEDs show the maximum exhaled volume of the last breath.
MAND Lights at the start of each breath
to indicate a ventilator- or operator-initiated (time or manually triggered) mandatory breath is being delivered.
10 to 90 cmH
O
2
(10 to 90 hPa) Accuracy:
(1 + 3% of setting)
0 to 50 L Accuracy:
(10 ml + 10% of setting)
-10 to 90 cmH (-10to90hPa)
Resolution: 1 cmH
If HIGH TIDAL VOLUME setting < 500 ml: 0 to 500 ml
Resolution: 5 ml
If HIGH TIDAL VOLUME setting
500 ml: 0 to 2000 ml
> Resolution: 20 ml
Not applicable
O
2
O (1 hPa)
2
1-28
Page 45
Introduction
Table 1-4: 700 Series Ventilator keyboards: PATIENT DATA
Key/
indicator
Other indicators (continued)
ASSIST Lights at the start of each breath
to indicate a patient-initiated mandatory (flow triggered) breath is being delivered.
SPONT Lights at the start of each breath
to indicate a patient-initiated spontaneous (flow triggered) breath is being delivered.
Function Range
Not applicable
Not applicable

1.3.3 VENTILATOR STATUS

The VENTILATOR STATUS section of the keyboard shows the operating condition of the ventilator, and is continuously updated during operation. Table 1-5 summarizes the functions of the keys and indicators in the VENTILATOR STATUS section of the keyboard.
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/
indicator
ALARM Red
Color Function
Flashes when a high-priority alarm is active.
(high priority)
A repeating sequence of three, then two beeps sounds. Lights steadily when a high­priority alarm has been autoreset.
CAUTION Yellow
(medium priority)
NORMAL Green Lights when no alarm condition is present.
Flashes when a medium-priority alarm is active. A repeating sequence of three beeps sounds. Lights steadily when a medium­priority alarm has been autoreset.
1-29
Page 46
Introduction
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/
indicator
VENT INOP Red
(high priority)
SAFETY VALVE
Red (high priority)
OPEN
ON AC/
Green Lights when the ventilator is running on ac BATTERY CHARGING
Color Function
Lights to indicate that the ventilator is inoperative, and the ventilator safety valve is open. A qualified service technician must run and pass the extended self-test (EST) before normal ventilation can resume.
If the condition that caused the safety valve to open no longer exists, and the VENT INOP indicator is off, press the alarm reset key to resume ventilation.
Lights when the ventilator’s safety valve and exhalation valve open and only room air is available to the patient.
Can indicate that the ventilator is inoperative, or there is an occlusion in the ventilator breathing circuit. If possible, the message window shows the alarm that triggered the safety valve open condition and how much time has elapsed since the last breath was triggered.
power and the battery is charging.
ON INTERNAL
Yellow Flashes when the ventilator is running on
the internal battery.
BATTERY
ON EXTERNAL
Yellow Flashes when the ventilator is running on
the external battery.
BATTERY
INTERNAL BATTERY
Green Shows the relative charge level of the
internal battery.
LEVEL
Yellow Alarm silence: Silences the alarm sound for
2 minutes from the most recent key press.
7-00423
1-30
Page 47
Introduction
Table 1-5: 700 Series Ventilator keyboards: VENTILATOR STATUS
Key/
indicator
7-00424
Color Function
Not applicable Alarm reset: Reestablishes all alarm
indicators, cancels the alarm silence period, and resets the patient data displays. If the condition that caused the alarm still exists, the alarm reactivates. Cancels apnea ventilation, if active. Reestablishes previous settings and ventilation resumes, unless the ventilator is inoperative.
1-31
Page 48
This page is intentionally blank.
1-32
Page 49

Setting up the ventilator

Warning
This section tells you how to set up the ventilator, including:
• Connecting and using the internal and external batteries
• Connecting the electrical supply
• Connecting the oxygen supply
• Connecting the ventilator breathing circuit
• Installing the collector vial
• Installing the flex arm
• Installing the humidifier
• Using the ventilator cart
• To avoid tipping or damaging the ventilator, do not stack oth
er equipment on the ventilator. 700 Series Ventilators
are designed to be mounted on either a cart or a shelf by
lified service technician. When lifting the ventilator, li
qua
m the base, and use assistance and appropriate safety
fro precautions. Figure 2-1 shows proper lifting technique.
a ft
Page 50
Setting up the ventilator
Caution
Warning
Use two people to lift
7-00053
Lift ventilator from base only
• To avoid the possibility of injury to the patient and to ensure proper ventilator operation, do not atta de sp
• To minimize the increased risk of fire due to an oxygen­enriched environment, do not use the ventilat hype
• To avoid raising the oxygen concentration of room air, use the ventilator in an adequately ventilated room.
Do not obstruct the cooling fan.
ch any
vice to the port labeled “EXHAUST” unless the device
ecifically authorized by Covidien.
or in a
rbaric chamber.
is
Figure 2-1. Lifting the ventilator

2.1 Connecting and using internal and external batteries

A Covidien battery must always be installed in the ventilator. Without a battery, the ventilator is not protected against low or lost ac power. Do not use the ventilator unless a battery with at least minimal charge is installed.
2-2
Page 51
Setting up the ventilator
7-00054
Internal battery level
NOTE:
If the ventilator has been stored for an extended period, allow it to acclimate to its environment before turning it on. This helps ensure that the ventilator powers up correctly.
Every 700 Series Ventilator System includes an internal battery as a backup power supply that comes standard with the ventilator. The internal battery can provide up to 2
½ hours of backup power,
depending on ventilator settings and battery charge level. The ventilator operates on backup power when ac power is lost or drops below minimum. The internal battery’s charge level is continuously indicated on the keyboard (see Figure 2-2). The ventilator alarms when it determines that only 5 minutes of power remain using the current settings.
Figure 2-2. Internal battery charge indicator
The optional external battery supplies up to 7 additional hours of backup power in case ac power is lost or falls below minimum. (See Appendix B for external battery ordering information.) If installed, the external battery is the first source of backup power. The ventilator uses the internal battery if the external battery is depleted or is not connected.
NOTE:
When the ventilator switches off because battery charge is inadequate, turn off the power switch to prevent the battery from being fully depleted. Connect the ventilator to ac power as soon as possible.
2-3
Page 52
Setting up the ventilator
Caution
The ventilator charges the internal and external batteries during ac power operation, or in standby mode when attached to ac power. (See Appendix C for battery specifications.) When the ventilator is operating on battery, you can use the MENU key to check the estimated operational time remaining until recharge is required for the internal and external batteries. (Chapter 6 tells you how to use the MENU key.) The ventilator must operate for 2 minutes before it can estimate battery time remaining.
You can disconnect and connect the external battery during normal operation. When operating the ventilator on battery power, you can install the external battery on the cart for easy transport. Follow these steps to connect the external battery:
1. Mount the external battery on the base of the cart (or othe table location).
sui
r
2. Plug the external battery’s cable into the connector (Figure 2-3).
3. Figure 2-4 shows you how to disconnect the external battery.
To ensure that the batteries retain their charge, store the ventilator in standby mode, with its power switch on, and connected to ac power. If the ventilator is not stored in this manner, check battery charge levels before using the ventilator.
If you plan to store the ventilator for more than 6 months, remove batteries before storage. Replace the internal battery before using the ventilator again. A qualified service technician must replace the battery according to the instructions in the 700 Series Ventilator System Service Manual.
If you turn on the ventilator after it has been unplugged for an extended period, the LOW EXT BATTERY and LOW INT BATTERY alarms may become active. If so, recharge the internal battery by leaving the ventilator plugged in and turned on in standby mode (with no patient connected) for up to 8 hours.
If the LOW EXT BATTERY, LOW INT BATTERY, or BAT NOT CHARGING alarm is still active after 8 hours, replace the batteries. If you turn off the ventilator when the battery is absent or inadequately charged, a LOSS OF POWER alarm sounds for at least 2 minutes.
2-4
Page 53
Setting up the ventilator
7-00055
External battery connector
External battery
NOTE:
The ON BATTERY indicator flashes when the ventilator is operating on battery power.
2-5
Figure 2-3. Plugging the external battery into the ventilator
Page 54
Setting up the ventilator
Warning
7-00051
Figure 2-4. Disconnecting the external battery

2.2 Connecting the electrical supply

• To avoid electrical shock hazard, connect the ventilator power cord into a grounded ac power outlet. If the integrit of
the ac ground is in question, operate the ventilator fr e internal or external battery.
th
• If used in the U.S., connect the ventilator to an ac mar
ked “Hospital Only” or “Hospital Grade” to ensure
grounding reliability.
y
om
receptacle
2-6
Every 700 Series Ventilator System is supplied with a power cord. A power cord retainer covers the connector and socket to protect against liquid spills or accidental disconnection.
When the ventilator is ready for operation on ac power, connect the power cord to ac power (see Figure 2-5).
Page 55
Setting up the ventilator
7-00056
Power cord retainer
Power cord
To ac power
7-00057
NOTE:
During operation on ac power, the power cord retainer must always be in place.
Figure 2-5. Connecting the ventilator power cord
When the power cord is not in use, you can wrap the power cord around the brackets on the back of the ventilator for convenient storage (see Figure 2-6).
2-7
Figure 2-6. Storing the power cord on the ventilator
Page 56
Setting up the ventilator
Warning
Warning
Caution

2.3 Connecting the oxygen supply

• To ensure proper oxygen concentration, do not obstruct the vent
ilator’s air intake.
• To ensure adequate oxygen delivery to the patient, us
vidien-supplied oxygen hoses only. Use of other oxygen
Co hoses could result in inadequate or inappropriate ox pr
essures or leaks at the oxygen inlet.
• When using a cylinder oxygen supply, point the cylind pr
essure relief device away from the vent
T
his helps avoid creating an oxygen-rich environmen
ithin the ventilator in the event that the cylind
w
gulator malfunctions.
re
The 700 Series Ventilator System can use oxygen from a cylinder or wall supply. Follow these steps to connect the oxygen supply:
e
ygen
er's
ilator air intake.
t
er oxygen
1. Ensure that the oxygen supply pressure is between 40 and 90 ps (
275 and 620 kPa) (50 psi (345 kPa) recommended).
Due to excessive restriction of certain hose assemblies (listed in Table B-1), reduced FiO pressures< 50 psi (345 kPa) are used. To maintain correct FiO
levels may result when oxygen inlet
2
2
levels, make sure that oxygen inlet pressure is 50 psi (345 kPa) when using these hose assemblies.
2. Connect the oxygen supply to the oxygen inlet connector on the side o
f the ventilator (see Figure 2-7). Make sure ventilator
con
figured with adapter(s) as required.
is
To prevent damage to the ventilator, ensure that the connection to the oxygen supply is clean and unlubricated.
i
2-8
Page 57
Setting up the ventilator
7-00058
Oxygen inlet connector
Oxygen hose (to oxygen supply)
Air intake
NOTE:
Whenever a pressurized oxygen source is connected to the ventilator, the oxygen regulator has a maximum bleed rate of 3 L/min, even when the ventilator is not in use. Always take this bleed rate into account when calculating oxygen cylinder volume and duration. This is especially important when using the ventilator and cart while transporting patients who are being ventilated at above-ambient FiO
levels.
2
Figure 2-7. Connecting the oxygen supply
2-9
Page 58
Setting up the ventilator
Warning
Caution

2.4 Connecting the ventilator breathing circuit

• To minimize the risk of bacterial contamination or comp
onent damage, inspiratory and expiratory filters must always be handled with care and connected to the ventilator during use.
• Use of an external pneumatically powered nebulizer with 700 Series V breathing circuit.
volumes, spirometry and spirometry alarms,
tidal Fi
O
, and patient triggering. (However, the Covidien EasyNeb
2
Nebu patient with data.)
To minimize the risk of patient injury, use only ventilato br
eathing circuits qualified for use in oxygen-enriched environments with 700 Series Ventilators. Do not us antistat tight connection, only use connectors and tubes with ISO­standard cone and socket fittings (or use adapters to connect barbed cuff fittings to ISO-standard fittings).
• Use ventilator breathing circuits identified by Covidien fo us
e with 700 Series Ventilators or circuits that ensu
th
e maximum resistance values specified in Table C-5 ar
exceeded.
event the ventilator from ventilating as long as th
pr vent
ilator passes SST.
entilator results in added flow to the ventil
This additional flow can af
lizer is designed to
deliver aerosol medications to the
out affecting ventilator performance or patien
ic or electrically conductive tubing. To ensure a leak-
Using a circuit with a higher resistance do
a
ator
fect delivered
delivered
t
r
e
r
re that
e not
es not
e
2-10
Adding accessories to the ventilator can increase system resistance. Ensure that any changes to the recommended ventilator circuit configurations do not exceed the values for resistance provided in
Appendix C.
Page 59
Setting up the ventilator
NOTE:
• Covidien recommends that you run SST every 15 days, between patients, and when you change the ventilator breathing circuit. Covidien recognizes that the protocol for running SST varies widely among healthcare institutions. It is not possible for Covidien to specify or require specific practices that will meet all needs, or to be responsible for the effectiveness of those practices.
• The ventilator uses an oxygen sensor to trigger an alarm if the delivered % O the % O
setting. You can view the ventilator’s % O
2
is 10 percentage points above or below
2
2
measurement by using the User settings function of the MENU key.
• Check the inspiratory and expiratory limbs of the ventilator breathing circuit and the collector vial and in-line water traps regularly for water buildup. Under certain conditions, they can fill quickly. In case of a partial occlusion alarm, check for water in the ventilator breathing circuit. In very severe cases, such a water occlusion can potentially cause the ventilator to enter occlusion cycling mode. Empty and clean the collector vial and in-line water traps as necessary.
• Covidien recommends using pediatric circuits when ventilating patients with 5-mm or smaller internal-diameter artificial airways.
Figure 2-8 shows how to connect the ventilator breathing circuit, including the inspiratory filter, humidifier (if used), inspiratory limb, patient wye, expiratory limb, collector vial, and expiratory filter.
2-11
Page 60
Setting up the ventilator
7-00059
Ventilator (to patient)
Inspiratory filter (arrow on filter indicates direction of flow)
Humidifier
Inspiratory limb of the ventilator breathing circuit (install EasyNeb between inspiratory filter and patient wye)
Patient wye
Expiratory
limb of the
ventilator breathing
circuit
Collector vial
Expiratory filter
(arrow on filter
indicates direction
of flow)
Ventilator
(from patient)
2-12
Figure 2-8 . Connecting the ventilator breathing circuit
Page 61

2.5 Installing the collector vial

7-00060
Bracket (on cart or shelf­mounting plate)
Collector
vial
Figure 2-9 shows you how to slide the collector vial into the bracket on the cart or shelf mounting plate.
Figure 2-9. Installing the collector vial
Setting up the ventilator

2.6 Installing the flex arm

Figure 2-10 shows you how to install the flex arm onto the threaded socket on the ventilator.
2-13
Page 62
Setting up the ventilator
7-00061
Flex arm
NOTE:
If the flex arm is too large to fit into a tight space, you can shorten the arm by removing the middle segment (Figure 2-11).
Figure 2-10. Installing the flex arm
2-14
Page 63
Setting up the ventilator
Remove
Remove
Remove
Figure 2-11. Shortening the flex arm

2.7 Installing the humidifier

Figure 2-12 shows you how to slide the humidifier into the bracket on the cart or shelf mounting plate.
NOTE:
The humidifier must be powered independently of the ventilator (regardless of whether the ventilator is running on ac or battery power).
2-15
Page 64
Setting up the ventilator
7-00063
Bracket (on cart or shelf­mounting plate)
Humidifier
Figure 2-12. Installing the humidifier
2-16
Page 65

2.8 Using the ventilator cart

Caution
Unlocked
Locked
7-00064
Unlocked
Ventilator cart base
The ventilator cart is equipped with a storage area for the external battery. You can lock and unlock the cart’s front wheels with your foot, as shown in Figure 2-13.
To prevent equipment damage, pull, rather than push, the ventilator over high thresholds when using the cart to transport the ventilator. Ensure that the flex arm is positioned at the front of the ventilator.
Setting up the ventilator
Figure 2-13. Locking and unlocking the cart’s front wheels
2-17
Page 66
2-18
This page is intentionally blank.
Page 67

Getting started

This section tells you how to:
• Power up the ventilator
• Select ventilator settings
• View and change alarm limits
• Enter and exit standby mode
NOTE:
Covidien recommends that before using the ventilator for the first time, you wipe the ventilator exterior clean and sterilize its components according to the instructions in Appendix A. Follow your institution’s protocol for cleaning and sterilizing the ventilator and its components.

3.1 Powering up the ventilator

NOTE:
To ensure battery charge(s), store the ventilator in standby mode with the power cord plugged into the mains and the power switch in the on (I) position whenever possible. The power switch controls ac mains and battery power to the ventilator.
Page 68
Getting started
Warning
7-00065
1. Turn the power switch on (see Figure 3-1). The ventilator disp
lays this mess
P
OST running...
age:
NPB 7x0 S/W Rev x PM due xxxxx hrs
“PM due” refers to the number of hours until a routine preventive maintenance procedure is due, based on the ventilator’s operational hours.
If a single beep is not audible during POST, the main audible alarm may be malfunctioning. Remove the ventilator from use and contact service.
NOTE:
• If the ventilator is running on ac but the battery is low, u wer up it may begin ventilating using th
po
ttings and an ABNORMAL RESTART alarm may occur.
se
• Do not switch off ventilator power during POST
e last valid
.
pon
3-
2
Figure 3-1. Turning the power switch on (the “I” position)
Page 69
Getting started
Warning
Caution
2. Once the power-on self-test (POST) is complete (several seconds), the ventilator displays the message:
Accept settings to start ventilation
The last valid settings are proposed, and the appropriate indicator lights to show the ventilator’s source of power.
Always let the ventilator run for 10 minutes before placing it on a patient to allow heaters to warm up. Failure to warm up the ventilator could result in failure of the SST flow sensor test, and higher measured exhaled tidal volumes.
Covidien recommends running short self-test (SST) and testing alarms to check proper operation of the ventilator before ventilation begins or according to your institution’s protocol. Chapter 4 tells you how to run SST. Appendix E tells you how to test alarms.
NOTE:
If the ventilator detects low oxygen pressure, a low battery level, or an occluded or missing air intake filter during POST, POST continues uninterrupted. These alarms become active when POST is complete, and you can override them by pressing the alarm reset key.
3-3
Page 70
Getting started
A/C
SIMV
SPONT
7-00066
Press each flashing key and turn the knob to adjust.

3.2 Selecting ventilator settings

NOTE:
Each keypress should result in a “blip” sound; otherwise, the main alarm speaker may be faulty. If a keypress does not result in this sound, turn the ventilator off then on to run POST and verify the proper functioning of the speaker.
1. Select the mode: To select a mode, press A/C, SIMV, or SPONT. (Appendix D gives a detailed explanation of these ventilatio mod
es.)
In assist/control (A/C) mode, you can only select volume control ventilation (VCV) or pressure control ventilation (PCV) settings (PCV is only available on the 760 Ventilator).
In synchronous intermittent mandatory ventilation (SIMV) mode, you can select both VCV or PCV (760 only) and pressure support ventilation (PSV) settings.
In spontaneous (SPONT) mode, you can only select PSV settings.
n
2. Confirm or change settings: For every flashing key light, yo
3-
4
mus
t touch the key (and adjust the setting if necess
can apply the settings. You can confirm or change setting
you in any orde
r.
Once you press a key, the proposed setting flashes to indicate that the knob is linked to the selected setting. The setting’s LCD window shows the setting and the message window displays the current setting, range, units, name of the setting, and associated parameter. Turn the knob to change the setting or press another key to leave the setting as is. (The setting then stops flashing.)
u
ary) before
s
Page 71
Getting started
PEEP/CPAP
TRIGGER
SENSITIVITY
L/min
% O2
ACCEPT
7-00067
3. Press CLEAR to cancel the most recently changed setting and return to the last valid setting. (This only works if you haven’t pressed ACCEPT yet.) Pressing CLEAR after you’ve pressed 100% O
stops the 100% O2 maneuver. The ventilator will beep
2
if you attempt to change a setting to a value above or below the setting range.
4. PEEP/CPAP, TRIGGER SENSITIVITY, and % O
: You
2
can change these settings in any mode. These keys don’t flash when you select a new mode, and you are not required to confirm these settings.
NOTE:
The O
% alarm is delayed for 30 seconds and 12 L of tidal
2
volume following a change in setting so that the oxygen concentration can stabilize.
5. Apply the settings: Press ACCEPT.
NOTE:
If apnea ventilation is possible, the ventilator displays this message when normal ventilation begins:
Review Ta = xx s
where Ta is the apnea interval and xx is the current setting in seconds (from 10 to 60 s). Chapter 5 tells you how to adjust apnea ventilation parameters. You are not required to review apnea parameters before normal ventilation begins.
3-5
Page 72
Getting started
Warning
7-00066
7-00068
Ventilation begins according to the displayed settings and the existing alarm limits. You should review all alarm limits and change them if necessary. (Section 3.3 tells you how.) You can change alarm limits or ventilator settings at any time during normal ventilation.
In PSV it is impossible to set support pressure + PEEP to deliberately cause a HIGH PRESSURE alarm. Although Covidien does not recommend doing so, it is possible in VCV to set a tidal volume that causes breath truncation and a HIGH PRESSURE alarm. To avoid activating a HIGH PRESSURE alarm, select the appropriate TIDAL VOLUME and HIGH PRESSURE alarm settings.

3.3 Viewing and changing alarm settings

You can view alarm settings at any time without affecting ventilation by pressing the appropriate key. The ALARM SETTINGS window displays the value of the alarm being viewed or changed. The alarm setting flashes for up to 30 seconds, and you can change an alarm setting any time its setting is flashing in the window.
All alarm settings are in effect during ventilation (except for LOW INSP PRESSURE, which is not active in SPONT mode). You can change alarm settings at any time during ventilation. You can only change one alarm setting at a time.
3-
6
To change a setting: touch, turn, ACCEPT.
1. Touch the key: The key light turns on, and the current alarm setting flashes in the window for 30 seconds. The message window shows the current setting, range, and units of the alarm.
2. Turn the knob to adjust the alarm setting (this only works when the alarm setting is flashing in the window). The ventilator will beep if you attempt to select a setting above or below the setting range.
Page 73
3. Press ACCEPT to apply the new
Warning
ACCEPT
7-00067
alarm setting se
tting stops flashing a
sage win
mes
sage:
mes
. The new alarm
nd the
dow displays this
Setting(s) accepted
• Inadvertent extubation in which the endotracheal tube remains pr
connected to the breathing circuit could result
essurization above the LOW INSP PRESSURE alarm setting.
The LOW INSP PRESSURE and low volume alarms sh
t appropriately.
se
• To ensure patient safety, always set the LOW VOLUME and
HIGH RATE alarms.
TIDAL
NOTE:
The O O and the low O the set % O
% alarm limits are based on the set % O2. The high
2
% alarm limit is 10 percentage points above the set % O2,
2
% alarm limit is 10 percentage points below
2
. The measured % O2 is shown in the message
2
window if its display is enabled (using the Oxygen sensor menu function, described in Chapter 6).
Getting started
in
ould be

3.4 Entering and exiting standby mode

Standby mode is a waiting state in which there is no ventilation, but the ventilator maintains its settings and battery charge(s).
3-7
Page 74
Getting started
Warning
• To avoid raising the oxygen concentration of room air, ensure that the ventilator is in an adequately ventilat ro
om if connected to an oxygen supply while in stan
e. Disconnect the oxygen supply if you do not intend to
mod use the ventilator immediately.
• To avoid patient injury due to lack of ventilatory support, do not enter standby mode with a patient atta
ilator. You must confirm that no patient is attached
vent
ched to the
before entering standby mode.
NOTE:
To maintain battery charge(s), the ventilator must be plugged into ac power with the power switch on while in standby mode.
Follow these steps to enter standby mode:
ed
dby
3-
1. Turn on the ventilator. If ventilation has already begun, mus
t turn off the ventilator, then turn it back on withou
st
arting ventilation. This allows you access to the Standby mode
you
t
menu function.
2. Press MENU.
3. Turn the knob to select standby mode.
4. Press ACCEPT. The ventilator displays this message:
Is pt disconnected? ACCEPT to proceed
5. Confirm that no patient is attached by pressing ACCEPT. (If you don’t press ACCEPT while this message is displayed, th vent
ilator will not enter standby mode.)
8
e
Page 75
Getting started
6. This message tells you that the ventilator is in standby mode:
In standby mode Clear to exit
The ON AC/BATTERY CHARGING, INTERNAL BATTERY LEVEL, and SAFETY VALVE OPEN indicators turn on. All other key lights and displays are off during standby mode. The only alarms that can become active during standby mode are FAN FAILED ALERT, BAT NOT CHARGING, LOSS AC POWER, and LOSS OF POWER.
Follow these steps to exit from standby mode and resume ventilation:
1. Press CLEAR.
2. The ventilator displays this message:
POST running...
3. After a few seconds, POST is complete and the ventilator begins ventilation using the last valid settings.
3-9
Page 76
3-10
This page is intentionally blank.
Page 77

Self tests (SST and EST)

Warning
All 700 Series Ventilators include these built-in self tests:
• Short self test (SST) is a short (about 2 to 3 minutes) and sequen the ventilator complia
• Extended self test (EST) is a thorough test of the operationa integ pne
• Before running SST or EST, you must disconnect the
• A fault identified in SST or EST indicates that the ventil
ce of tests that verifies proper ventilator op
breathing circuit for leaks, and measures the circui
nce. Power on self test (POST) is part of SST.
rity of the ventilator, both the electronics and
umatics. POST and SST are part of EST.
ventilator from the patient. Running SST or EST while the ventilator is connected to the patient can injure the patient.
an associated com or associated component should be repaired before th ventilator is returned to service, unless it can be determined with ce patient, hazards.
rtainty that the defect cannot create a hazard for
or add to the risks which ma
ponent is defective. A defective ventilator
y arise from other
eration, checks
simple
the
ator or
e
the
t
l
Page 78
Self tests (SST and EST)
Caution
To ensure accurate SST and EST operation, run SST and EST in room air (% O on for at least 10 minutes. (EST also requires an oxygen supply.)
Table 4-1 describes SST and EST.
Test name Purpose When test is run
is 21%) after the ventilator has been powered
2
Table 4-1: 700 Series Ventilator self tests
Short self test (SST)
Extended self test (EST)
An abbreviated version of EST to be run primarily by the operator. Makes detailed checks of the pneumatics and electronics. It also characterizes system leaks and system/tubing compliance to compensate during breath delivery. POST is run as part of SST.
Thoroughly tests the operational integrity of the ventilator, both the electronics and pneumatics. POST and SST are part of EST.
Before patient is connected to the ventilator or after ventilator breathing circuit or humidifier is changed.
• When the ventilator is serviced.
• As part of the ventilator performance verification.
• Following a ventilator inoperative.
NOTE:
The performance verification, described in the 700 Series Ventilator System Service Manual, is a more
thorough test of the ventilator to verify specifications are met. The technician runs a partial or full performance verification after servicing the ventilator and at regular intervals.
4-
2
Page 79

4.1 Short self test (SST)

Table 4-2 summarizes the functions of SST.
NOTE:
• Covidien recommends that you run SST every 15 da tween patients, and when you change the ventilat
be
eathing circuit. Covidien recognizes that the pr
br ru
nning SST varies widely among healthcare institutions
It is
not possible for Covidien to specify or require specif
pr
actices that will meet all needs, or to be responsible fo e effectiveness of those practices.
th
• When SST asks you to do something (for example, press
ACCEPT or block the patient wye), the ventilator will wa indefinit
Follow these steps to run SST:
ely for your response.
Self tests (SST and EST)
ys,
or
otocol for
. ic
r
it
1. Turn on the ventilator. If ventilation has already begun, must
turn off the ventilator, then turn it back on withou
st
arting ventilation. This allows you access to the Self test menu
function.
If you select the Self test function during ventilation th
ow displays this message:
wind
N/A in ventilation
2. Press MENU, turn the knob to select Self test, then press ACCEPT. The usual 30-second timeout for your input is disabled during the entry to the Self test function.
3. The window displays this message:
Short self test Turn knob to view
4. Turn the knob to select Short self test (you Extended self test). The
Short self test Vent warming 09:59 CLEAR to bypass
window displays this message:
can also select
you
t
e
4-3
Page 80
Self tests (SST and EST)
Warning
If the ventilator has already been running for more than 10 minutes before you select Short self test, this message is bypassed (see step 5). If the ventilator was recently running and is already warmed up, you can press CLEAR to override the Vent warming message and begin SST. Covidien cannot guarantee the accuracy of test results in this case, however.
5. The window displays this message:
6. Press ACCEPT to confirm that no patient is connected to the ventilator. (Or press CLEAR to back up through the menu.) The window displays this message:
7. Block the patient wye and then press ACCEPT.
8. The ventilator runs power-on self-test (POST) and displays this message for a few seconds:
Is pt disconnected? ACCEPT to proceed
Block wye
POST running...
4-
4
If a single beep is not audible during POST, the main audible alarm may be malfunctioning. Remove the ventilator from use and contact service.
NOTE:
• If the ventilator is running on ac but the battery is low, upon power up it may begin ventilating using the last valid settings and an ABNORMAL RESTART alarm may occur.
• During POST it is recommended that you not switch off ventilator power.
9. The message window then displays this message:
humidification device type
Choose humidifier
where the humidification device type is the most recently selected type, as follows:
Page 81
Self tests (SST and EST)
Warning
Warning
HME: Heat moisture exchanger (“artificial nose”)
Dual heated wire: Humidifier with heated wire on expiratory limb, or on both inspiratory and expiratory limbs
No heated wire: Conventional humidifier without heated wire circuit on the expiratory limb
Turn the knob to select the humidifier type attached, then press ACCEPT.
Incorrectly specifying the humidifier type during SST can affect the accuracy of spirometry calculations.
10. The message window then displays this message:
tubing type
Choose tubing type
where the tubing type is the most recently selected ventilator breathing circuit type.
Turn the knob to select the ventilator breathing circuit type (adult or pediatric tubing) attached to the ventilator, then press ACCEPT.
Incorrectly specifying the ventilator breathing circuit type during SST can cause an inappropriate sensitivity for the leak test and occlusion alarm. Covidien recommends using pediatric circuits when ventilating patients with 5-mm or smaller internal-diameter artificial airways.
NOTE:
To ensure that compliance compensation functions correctly, you must run SST with the circuit configured as intended for use on the patient.
4-5
Page 82
Self tests (SST and EST)
Warning
11. The message window then displays this message:
Turn the knob to select the appropriate ET (endotracheal tube) size, then press ACCEPT.
Specifying an ET tube size that is too large can cause premature termination of breaths on very small pediatric patients. During normal ventilation, you can use the MENU key’s User settings function to change the ET tube size (see Chapter 6).
12. The ventilator automatically starts the test sequence. A message indicat cannot be interrupted. In most cases you don’t need a required. When a test passes, SST automatically starts the next test.
NOTE:
The PRESSURE bar graph and the PRESSURE display window show the real-time expiratory pressure throughout SST.
ET size: x.x mm Choose ET size
es which test is being performed. Once started, a test
to do
nything unless a test result is fault or fail, or your action is
4-
6
Page 83
These keys are valid during SST:
ACCEPT
CLEAR
MANUAL
INSP
To confirm that a requested action is complete.
To repeat a test, return to the prompt at the start of a test, or indicate an inaudible alarm (during the Main alarm speaker test and Backup alarm test only).
To override an Incomplete or Fault test result.
To restart from the beginning of SST.
7-00424
To stop testing and skip to the end of SST.
Self tests (SST and EST)
7-00423 7-00072
NOTE:
If the tubing or humidifier type or ET tube size that you’ve selected has changed from the previous time SST was run, this key is temporarily disabled until the Circuit comp test is complete.
4-7
Page 84
Self tests (SST and EST)
Display:
Disconnect I tubing Reconnect I tubing
(Do not disconnect inspiratory filter.)
7-00069
Display:
Unblock wye Block wye
7-00070
Display:
Disconnect E tubing Reconnect E tubing
(Do not disconnect expiratory filter.)
7-00071
13. The I/E filter test prompts you to complete these actions (press ACCEPT when each action is complete):
4-
8
Page 85
Self tests (SST and EST)
14. When the final test in the sequence is complete, this message is displayed along with the overall SST result:
SST finished testing
Table 4-3 summarizes overall SST results and how to proceed in each case.
15. Unblock the patient wye, then press ACCEPT. The ventilator reruns POST.
Table 4-2: SST sequence of tests
Test Function Comments
Is pt disconnected?
Leak test Checks the ventilator
P
e Pcyl
compare
Auto-zero sol test
Circuit comp test
Asks for confirmation that a patient is not connected to the ventilator.
breathing circuit for leaks. Displays expiratory pressure on pressure bar graph.
Verifies proper function of pressure transducers in internal pneumatics and ventilator breathing circuit. Displays expiratory pressure on pressure bar graph.
Verifies the proper function (ability to open and close) of the P P
cyl
Determines the compliance of the ventilator breathing circuit. Displays expiratory pressure on pressure bar graph.
(expiratory) and
e
(cylinder) solenoids.
SST cannot proceed until you confirm that a patient is not connected.
Fault if not passed. Overriding a fault could
cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
Fault if not passed. Overriding a fault could
cause improper compliance compensation, or autocycling.
Failure if not passed.
Fault if not passed.
Overriding a fault could cause improper compliance compensation or inaccurate tidal volume delivery.
4-9
Page 86
Self tests (SST and EST)
Test Function Comments
Table 4-2: SST sequence of tests (continued)
Safety valve test
Verifies that safety valve relieves excess circuit pressure. Displays expiratory pressure on pressure bar graph.
PEEP system test
Verifies that the PEEP system can generate and maintain preset PEEP levels. Displays expiratory pressure on pressure bar graph.
Flow/O
2
sensor test
Checks the accuracy of the exhalation flow sensor and the oxygen sensor.
I/E filter test Checks the pressure drop
across the inspiratory and expiratory limbs of the entire patient system. Checks the pressure drop across the filters.
Failure if not passed.
Failure if not passed.
Failure if flow sensor test
does not pass. Fault if oxygen sensor test
does not pass. Use the MENU key’s Oxygen sensor function to recalibrate the oxygen sensor (see Chapter 6).
Overriding a fault (oxygen sensor test only) could result in inaccurate % O
2
measurement.
Fault if not passed. This test is optional: pressing CLEAR at the initial prompt skips this test.
Skipping this test or overriding a fault could result in inadequate bacteria protection or excessive resistance to inspiration or exhalation.
4-
10
Page 87
Self tests (SST and EST)
Table 4-2: SST sequence of tests (continued)
Test Function Comments
Heaters test Verifies correct operation of
the heaters in the exhalation subsystem.
Main alarm test
Backup alarm test
Sounds alarm at three volumes and verifies that user reports hearing it.
Checks the operation of the backup alarm circuit.
NOTE:
Faults detected during SST may require removing the ventilator from clinical use immediately and servicing it. A trained operator, however, must evaluate the situation and determine whether the ventilator can be used. Failures detected during SST require rerunning SST (using the alarm reset key). It is not possible to exit SST if SST fails. If SST continues to fail, the ventilator must be removed from service.
Fault if not passed. Overriding a fault could
cause inaccurate monitoring of exhaled volume or lead to damage to the ventilator’s flow sensor or exhalation valve.
Failure if not passed. If alarm is audible press
ACCEPT. If not, press CLEAR.
Fault if not passed. If alarm is audible press ACCEPT. If not, press CLEAR.
Overriding a fault could cause no backup audible alarm in case the main audible alarm fails.
4-11
Page 88
Self tests (SST and EST)
Table 4-3: Overall SST results
If the SST
result is:
It means: Do this:
SST passed All tests passed. Press ACCEPT to exit SST
or press the alarm reset key to restart SST from the beginning.
SST incomplete
SST fault One or more tests did not
All tests performed were passed, but some were skipped. The skipped tests were passed on a previous run.
pass. These tests might not compromise the ventilator’s ability to ventilate safely, based on the operator’s
Press the alarm reset key to restart SST from the beginning or press MANUAL INSP to override this result and exit SST.*
Press the alarm reset key to restart SST from the beginning or press MANUAL INSP to override this result and exit SST.*
evaluation of the situation.
SST failed One or more critical
problems were detected.
Press the alarm reset key to rerun SST. It is not possible to exit SST if SST fails. If SST continues to fail, remove the ventilator from use and contact service.
4-
* If you press MANUAL INSP to exit SST, you will be prompted to confirm that you
wish to override the fault or incomplete test. Confirm by pressing MANUAL INSP again.
12
Page 89

4.2 Extended self test (EST)

Warning
To run EST, you must have the hardware listed in Table 4-4.
Table 4-4: EST hardware requirements
Self tests (SST and EST)
Hardware description
Ventilator breathing circuit
Manufacturer/model or Covidien part number
To use as test circuit: G-061208-00 or equivalent (adult, reusable, without heated wire)
NOTE:
To ensure that compliance compensation functions correctly, the user must run EST or SST with the circuit configured as intended for use on the patient.
Stopper, wye (no. 2)
Stopper, inspiration port (no. 3)
Oxygen source, 40 to 90 psi (275 to 620 kPa)
Due to excessive restriction of Air Liquide™*, Australian, and Dräger™* hose assemblies, reduced FiO2 levels can result if you use oxygen inlet pressures < 50 psi (345 kPa). Make sure oxygen inlet pressure is 50 psi (345 kPa) when using these hose assemblies, to maintain correct FiO2 levels.
G-061574-00 or local supplier
G-061575-00 or local supplier
Local supplier
4-13
Page 90
Self tests (SST and EST)
NOTE:
Covidien recommends that you always run the full EST before placing the ventilator into operation following service and as part of the ventilator’s routine performance verification. For more information on performance verification, see the 700 Series Ventilator System Service Manual.
Follow these steps to run EST:
1. Set up the ventilator as for normal operation, co hu
midifier, if applicable, and leak-tight ventilator breathing
mplete with
circuit.
2. Turn on the ventilator. If ventilation has already begun, mus
t turn off the ventilator, then turn it back on withou
st
arting ventilation. This allows you access to the Self test menu
you
t
function.
3. Press MENU, turn the knob to select Self test, then press ACCEPT.
4. Turn the knob to select Extended self test. The window displays the messages shown in Table 4-5.
Table 4-5: EST setup messages
Message Description
Extended self test Vent warming xx:xx CLEAR to bypass
This message is displayed (followed by a countdown) if you just turned on the ventilator. The ventilator must be on for 10 minutes before running EST so that component temperatures can stabilize.
If the ventilator was recently running and is already warmed up, you can press CLEAR to override the Vent warming message and begin SST. Covidien cannot guarantee the accuracy of test results in this case, however.
4-
14
Is pt disconnected?
Block wye Install no. 2 stopper and then press ACCEPT.
Confirm that patient is disconnected by pressing ACCEPT; or press CLEAR to return to service menu.
Page 91
Self tests (SST and EST)
Warning
Table 4-5: EST setup messages (continued)
Message Description
POST running... Wait a few seconds until POST is completed.
NOTE:
• If a single beep is not audible during POST, the main audible alarm may be malfunctioning. Contact service.
• During POST do not switch off ventilator power.
humidification device type
Choose humidifier
Turn knob to select desired humidification device, then press ACCEPT. Device types include: HME (heat­moisture exchanger or “artificial nose”), Dual heated wire (humidifier with heated wires on both inspiratory and expiratory limbs), or No heated wire (humidifier without a heated wire on expiratory limb).
Incorrectly specifying the humidifier type during EST can affect the accuracy of spirometry calculations.
4-15
Page 92
Self tests (SST and EST)
Warning
Warning
Message Description
Table 4-5: EST setup messages (continued)
tubing type
Choose tubing type
ET size: x.x mm Choose ET size
Turn knob to select either Adult tubing or Pediatric tubing and then press ACCEPT.
Incorrectly specifying the ventilator breathing circuit type during EST can cause an inappropriate sensitivity for the leak test and occlusion alarm. Covidien recommends using pediatric circuits when ventilating patients with 5-mm or smaller internal-diameter artificial airways.
NOTE:
To ensure that compliance compensation functions correctly, run EST with the circuit configured as intended for use on the patient.
Turn knob to select appropriate ET (endotracheal tube) size, and then press ACCEPT.
Specifying an ET tube size that is too large can cause premature termination of breaths on very small pediatric patients. During normal ventilation, you can use the MENU key’s User settings function to change the ET tube size (see Chapter 6).
4-
16
Page 93
Self tests (SST and EST)
5. The ventilator automatically starts the test sequence (Table 4-6). Unless a test requires your intervention (and waits indefinitely for your response), you don’t need to do anything until a test result is fault or failed, or EST is complete. Use the keys listed in Table 4-7 as needed.
Table 4-6: EST test sequence
Test n u m b e r
and name
1. DAC-ADC loop test
2. Safety valve test
3. Motor sensor test
4. Zeroing of P
o
solenoids
5. O
2
test
Function Comments
A loopback test to check digital-to-analog (D/A) and analog-to-digital (A/D) converters’ operation.
Verifies that safety valve relieves excess circuit pressure.
Checks operation of rotary encoder.
Zeroes oxygen regulator pressure transducer.
Disconnect and reconnect oxygen supply, as directed.
Press ACCEPT to signify you have performed each requested action.
Checks oxygen solenoids and oxygen sensor.
Failure if not passed. Not in SST.
Failure if not passed. Also in SST.
Failure if not passed. Not in SST.
Fault if not passed. Not in SST.
NOTE:
Overriding this fault could result in inaccurate oxygen percentage delivery.
Fault if not passed. Not in SST.
NOTE:
Overriding this fault could result in inaccurate oxygen percentage delivery.
6. Flow sensor test
4-17
Checks accuracy of exhalation flow sensor.
Failure if not passed. Also in SST.
Page 94
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
7. Leak test Checks ventilator breathing circuit for leaks.
7. P
ePcyl
compare (also part of test 7)
Verifies proper functioning of pressure transducers and ventilator breathing circuit.
Function Comments
Fault if not passed. Also in SST.
NOTE:
Overriding this fault could cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
Fault if not passed. Also in SST.
NOTE:
Overriding this fault could cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
4-
18
7. Auto-zero
sol test (also part of test 7)
8. Circuit comp
test
Verifies the proper function of the Pe and P
Determines ventilator breathing circuit compliance.
solenoids.
cyl
Failure if not passed. Also in SST.
NOTE:
Overriding this fault could cause improper compliance compensation, inaccurate tidal volume delivery, or autocycling.
Fault if not passed. Also in SST.
Page 95
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
9. I/E filter test Checks pressure drop across inspiratory and expiratory limbs of entire patient system. Checks pressure drop across filters.
Disconnect and reconnect inspiration (I) tube from/to inspiratory filter outlet, and exhalation (E) tube from/to expiratory filter inlet, as directed. Block and unblock wye with a no. 2 stopper, as directed. Press ACCEPT to signify you have performed each requested action.
10. PEEP
system test
Verifies that PEEP system can generate and maintain preset PEEP levels within either of two sets of limits. If results are within the inner set of limits, test passes the first time. If the results are outside of the outer set of limits, test fails the first time. If the results are outside the inner limits but within the outer limits, the ventilator adjusts its calibration table in nonvolatile RAM (NVRAM) and repeats the test to verify the calibration. The test fails if the calibration cannot be verified after five attempts.
Function Comments
Fault if not passed. Also in SST. Mandatory in EST, but optional in SST.
NOTE:
Overriding this fault could result in inadequate bacteria protection or excessive resistance to inspiration or exhalation.
Failure if not passed. Also in SST, but not identical.
11. Check
valve test
4-19
Checks check valves on inlet and outlet ports of cylinder and on exhalation limb of ventilator breathing circuit.
Failure if not passed. Not in SST.
Page 96
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
12. Piston leak test
13. Lights­displays test
14. Keys test Verifies that every row and
Checks piston leak against a table of leak values stored in NVRAM.
First remove inspiratory filter; then block and unblock inspiratory filter port as directed, using no. 3 stopper. Press ACCEPT to signify you have performed each requested action. Replace inspiratory tubing at end of test.
Verifies that lights and displays function.
Observe each named display, and press ACCEPT to verify that it is lit. Press CLEAR if a display is not lit.
column of key matrix can be correctly read.
Press each named key (each key’s LED will also be lit). Press CLEAR if a key’s LED is not lit.
Function Comments
Failure if not passed. Not in SST.
Failure if not passed. Not in SST.
Failure if not passed. Not in SST.
4-
20
15. Main alarm test
Checks whether various alarm sounds are audible (at maximum volume).
Listen for each named sound, then press ACCEPT to verify that it is audible. Press CLEAR if a sound is not audible.
Failure if not passed. Also in SST.
Page 97
Self tests (SST and EST)
Table 4-6: EST test sequence (continued)
Test n u m b e r
and name
16. Backup alarm test
17. Fan test Verifies that main fan is
18. Heaters test
Checks operation of backup alarm circuit.
Press ACCEPT if alarm is audible or press CLEAR if alarm is not audible.
operating correctly by reading fan status bits.
Verifies correct operation of heaters in exhalation system.
Function Comments
Fault if not passed. Also in SST.
NOTE:
Overriding this fault could result in no audible alarm if the main speaker fails.
Failure if not passed. Not in SST.
Fault if not passed. Also in SST.
NOTE:
Overriding this fault could result in inaccurate monitoring of exhaled volume or damage to the ventilator’s flow sensor or exhalation valve.
At the end of each test, the test name and pass/fail/fault status are displayed. When a test passes, EST automatically starts the next test. You do not need to do anything unless a test result is Fault or Fail, or your action is required. If a test does not pass, you can still continue EST, although the overall EST result is not passed until the error condition is corrected. Since information on multiple errors can help troubleshooting, it can be useful to complete EST even with errors.
4-21
Page 98
Self tests (SST and EST)
ACCEPT
CLEAR
MANUAL
INSP
Table 4-7: Key functions during EST
Key Function
To confirm that a requested action is complete.
Repeat a test or return to prompt at start of a test.
To override an Incomplete or Fault test result.
To restart from the beginning of SST.
7-00424
To stop testing and skip to the end of SST.
4-
22
7-00423 7-00072
NOTE:
If the tubing or humidifier type or ET tube size that you’ve selected has changed from the previous time SST was run, this key is temporarily disabled until the Circuit comp test is complete.
6. Some tests display additional prompts, some of which require your
response. These are listed in Table 4-8.
Table 4-8: Prompts during EST
Prompt Your response
Disconnect O supply
Reconnect O supply
2
Disconnect ventilator from oxygen supply. Press ACCEPT.
2
Reconnect ventilator to oxygen supply. Press ACCEPT.
Page 99
Self tests (SST and EST)
Table 4-8: Prompts during EST (continued)
Prompt Your response
Disconnect I tubing Disconnect tubing from inspiratory filter outlet.
Leave inspiratory filter in place. Press ACCEPT.
I drop = x x.x at 100
or
I drop = x x.x at 40 Reconnect I tubing
This is the inspiratory filter pressure drop in cmH
O at
2
a flow of 100 L/min (adult tubing) or 40 L/min (pediatric tubing). Reconnect tubing to inspiratory filter outlet. Press ACCEPT.
Unblock wye Remove stopper from patient wye. Press ACCEPT.
Block wye Insert a stopper into patient wye. Press ACCEPT.
Disconnect E tubing Reconnect tubing to inspiratory filter outlet, and
disconnect tubing from expiratory filter inlet. Leave expiratory filter in place. Press ACCEPT.
E drop = x.x at 100
or
E drop = x x.x at 40 Reconnect E tubing
This is the expiratory filter pressure drop in cmH a flow of 100 L/min (adult tubing) or 40 L/min (pediatric tubing). Press ACCEPT. Reconnect tubing to expiratory filter outlet.
O at
2
Block insp filt port Remove inspiratory filter. Insert no. 3 stopper into
inspiratory filter outlet. Press ACCEPT.
Reconnect I tubing Remove stopper from inspiratory filter outlet.
Reinstall inspiratory filter. Press ACCEPT.
Name of a section of display
Upper MDW = blocks?
Press ACCEPT to confirm that all LEDs/LCDs in named section are on.
Press ACCEPT to confirm that all pixels are turned on in upper half.
Lower MDW = blocks?
Press ACCEPT to confirm that all pixels are turned on in lower half.
Name of a key Press named key.
High alarm sound Confirm that you hear alarm by pressing ACCEPT.
Medium alarm
Confirm that you hear alarm by pressing ACCEPT.
sound
4-23
Page 100
Self tests (SST and EST)
No sound Confirm that you do not hear alarm by pressing
Table 4-8: Prompts during EST (continued)
Prompt Your response
ACCEPT.
Is backup alarm ON?
Confirm that you hear alarm by pressing ACCEPT.
7. When the last test in the sequence is complete, the EST comp
letion status is displayed (Table 4-9). Respond as indicate
If cert
ain tests do not pass, safe ventilation
romised, and the ventilator is rendered inoperative
comp
paired. This is called a ventilator failure. If certain othe
is re tests do no (subje
t pass, but ventilation might not be compromise
ct to a trained operator’s evaluation), this is called a
may be
ventilator fault.
NOTE:
Faults detected during EST may not require the ventilator to be serviced or removed from use immediately. A trained operator, however, must evaluate the situation and determine whether the ventilator can be used. Failures detected during EST require immediate servicing and removal of the ventilator from clinical use.
d.
until it
r
d
4-
24
Loading...